HILL-ROM avantguard Electric Bed Service Manual
Contents
1. Item Number Part Number Quantity Description 1 AG0028 1 Head and foot cover kit 2 AG0038 1 Base frame kit Casters 150 single wheel option 3 AG0057 3 Caster 150 single wheel kit AG0029 1 Caster 150 single wheel steer kit Casters 150 double wheel option AG0058 3 Caster 150 double wheel kit AG0030 1 Caster 150 double wheel steer kit Casters 150 single wheel anti static option 3 AG0057 1 Caster 150 single wheel kit 4 AG0029 1 Caster 150 single wheel steer kit 7 AG0031 2 Caster 150 single wheel anti static kit Casters 150 double wheel anti static option 5 AG0058 1 Caster 150 double wheel kit 6 AG0030 1 Caster 150 double wheel steer kit 8 AG0032 2 Caster 150 double wheel anti static kit 9 AG0036 2 Hilow column 40 80 kit 10 AG0037 2 Hilow column 45 85 kit 11 AG0035 2 Bilateral hilow pedal kit 12 AG0007 1 5th wheel kit 13 AG0033 1 Braking mechanism kit 14 AG0034 1 Brake on switch kit 15 AG0059 1 Braking rod kit NOTE All the kits contain their respective fixing systems as well any applicable adhesive labels Hardware and other small metal or plastic parts are available in the kits AvantGuard Electric Bed Service Manual QD1787 Page 5 9 Intermediate Frame Chapter 5 Parts List Intermediate Frame Figure 5 3 Intermediate Frame 1787_502 Table 5 3 Intermediate Frame Item Number Part Number Quantity Descript
2. Voltage 230V 10 15 AC voltage and earth Frequency 47 52 Hz Maximum power load 360 VA Refer to the schematic diagram at the beginning of this section to obtain specific information for the cabling The bed is operated by a PSU This unit controls the electric functions of the bed hilow H Col and F Col HS and TS with automatic contour see figure 3 2 on page 3 4 including the FS depending on the model see figure 3 3 on page 3 5 These functions are performed by actuators using very low voltages Power Supply Unit Characteristics Table 3 2 Power Supply Unit Characteristics Description Input characteristics Specification see table 3 1 on page 3 10 Fuse rating on primary 2 AT Maximum power load 360 VA Output voltage 18 to 40 V rectified filtered Output power 360 VA Intermittent operation 10 3 min every 30 min Page 3 10 AvantGuard Electric Bed Service Manual QD1787 Electrical System Chapter 3 Theory of Operation The power supply unit comprises of the following main sections e Power stage e Functions controlled by relays and transistors e Battery charging circuit e Control logic and specific functions The functions are activated after electronically unlocking each of them by using one of the following e Control pendant e Control panel on a flexible arm e Lateral caregiver unit e Hilow pedal HL only option or accessory
3. mark O of the chair function and check that the three sections return correctly to the low position Troubleshoot in the event of a malfunction see RAP 2 11 AvantGuard Electric Bed Service Manual QD1787 Page 6 11 Preventive Maintenance Chapter 6 General Procedures Function Mechanical foot section LI158Ax Procedure Activate the thigh section raise up to mid height Inspect the assembly of the foot actuator and its control lever Check that the pins and Truarc circlips are present and fully tightened Carry out a full raise lower cycle of the foot section by raising and lowering it manually when the lever has be used The foot section only must rise when the lever is operated A certain amount of force must be used when lowering the section manually Check that there is no friction or abnormal noises and that the movement correctly cushioned when lowering section Check that the section locks correctly when the lever is released Check that there is no oil on the actuator rod Replace the gas spring assembly if necessary refer to procedure 4 20 Trendelenburg reverse trendelenburg according to configuration Activate the function selection lockout trendelenburg reverse trendelenburg access sequence on the lateral caregiver unit Operate the sleep surface trendelenburg Check that there is no friction or abnormal noises and that no audible overload ind
4. NOTE Identification of components see figure 3 1 on page 3 3 NOTE Initial conditions the HL function works see Function Checks on page 2 6 1 Perform the following to check the foot siderail support a Remove the hard surfaces of the TS and the FS b Check that the screws pins and circlips are not damaged or missing from the foot siderail support mechanism All the screws pins and circlips are secure and undamaged Yes No Replace the missing screws pins or circlips refer to the relevant view of chapter 5 then go to Function Checks on page 2 6 2 Check that the foot siderail locks and unlocks correctly The locking unlocking is correct Yes No sl Replace the foot siderail support refer to procedure 4 23 on page 4 72 then go to Function Checks on page 2 6 3 The problem is solved Yes No y Contact your national Hill Rom representative see the last page of this manual 4 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 45 2 18 Head Siderail Locking Unlocking Malfunction Chapter 2 Troubleshooting Procedures 2 18 Head Siderail Locking Unlocking Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE Initial conditions the foot siderails are working see Function Checks on page 2 6 1 Check the locking and unlocking of the head siderail of the bed Set t
5. Replacement 1 Perform the removal procedure in reverse order Y SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the installation is not correct and or if the earth continuity is compromised 2 Check that the earth continuity of the accessible metal parts is correct see Earth resistance on page 6 7 in the Preventive Maintenance Schedule 3 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1 787 Page 4 77 4 27 Egress Bar Chapter 4 Removal Replacement and Adjustment Procedures 4 27 Egress Bar Identification see figure 3 1 on page 3 3 Time required 10 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench Removal 1 Apply the brakes raise the sleep surface to the high position the HS the TS and the FS to horizontal Store the foot siderails W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated
6. The PSU is defective Replace it refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 10 The TS actuator is defective Replace it refer to procedure 4 10 on page 4 33 then go to Final Actions on page 2 12 11 Test the function of the FS actuator intermediate switch Using an ohmmeter check the state of the FS actuator intermediate switch e When the rod is extended contacts 3 amp 4 see Figure 2 2 on page 2 32 of the connector socket 2 green must be short circuited e When the rod is retracted the circuit must be open The FS actuator intermediate switch is working Yes No The PSU is defective Replace it refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 12 The FS actuator is defective Replace it refer to procedure 4 11 on page 4 35 This solves the problem Page 2 32 AvantGuard Electric Bed Service Manual QD1787 2 9 Foot Section Malfunction Chapter 2 Troubleshooting Procedures Yes No y Contact your national Hill Rom representative see the last page of this manual 13 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2
7. 3 9 Foot Siderail Supports uxcrtimsciran tiara ad a andes 3 9 Head and Footboards 1 25 vaceepe aro ra a 3 9 Electrical System a oe dae AA Ce Pee cee a 3 10 Power Supply Unit Characteristics cutis sees ewow and vow es eee ees 3 10 Power ares 3 11 Function Control ca elt dd ee o Oe oe ae bl ar Ed 3 11 Battery Circuit ss a eed tan ee ee ieee ee ee edie Meee eee 3 12 Battery Characteristies oegojac n irrbraos Hebe do ta hie 3 12 Control Logic and Specific Functions 0 00 ee ee eee eee eee 3 13 Control Pendant seco mistica bed winter Bed wed 3 14 Control Panel on a Flexible Amor peje bes aw ds 3 14 Lateral Caresiver Unit 0200 23 oe ie be aes OL eR esl eee ae Be ee 3 14 Motor Characteristics ic iseracaetivadenee be ada Bes 3 14 Management of the Actuators 2 0 0 02 e eee eee eee eee 3 15 AvantGuard Electric Bed Service Manual QD1787 Page 3 1 Chapter 3 Theory of Operation Detection of Bed not in Low Position 0 0 0 eee eee eee 3 15 Automate CONTOUR tairis a bey aad A DERE y ORO 3 15 EasyChair Automatic Chair Function 0 0 00 eee ee eee 3 16 Emergency Trendelenburg Function ovina a es 3 16 Detection of Bed Connected to AC Power but Brakes not Applied Optional Depending on the Model 0 0 0 eee ee eee 3 16 Trendelenburg Reverse Trendelenburg 0 0002 e eee eee 3 16 Page 3 2 AvantGuard Electric Bed Service Manual QD1787 Componen
8. 33 2 10 Assisted Foot Section Malfunction Chapter 2 Troubleshooting Procedures 2 10 Assisted Foot Section Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE Initial conditions the TS function is working 1 Verify the attachment of the TS and the FS Remove the hard surfaces of the TS and FS Check that the screws bolts or pins are not damaged loose or missing from the control lever or the FS gas spring mechanism refer to assembly diagrams in chapter 5 All bolts screws and pins are secure and undamaged Yes No y Replace the missing or damaged parts then go to Function Checks on page 2 6 Perform the following to verify the adjustment of the FS gas spring a Press the TS raise E button until the TS reaches the high position see figure 1 11 on page 1 14 b Set the TS to high position c Check that the gas spring release adjustment is correct refer to procedure 4 20 on page 4 65 The adjustment is correct Yes No 4 Adjust the gas spring release refer to procedure 4 20 on page 4 65 then go to Final Actions on page 2 12 The FS gas spring operates correctly Yes No Replace the FS gas spring subassembly and cable and release handle refer to procedure 4 20 on page 4 65 then go to step 5 Go to Final Actions on page 2 12 This solves the problem Yes No y Contact your national Hill Rom representative see the las
9. Replacement tsue pe bs A wae ead 4 20 Battery Backup and Emergency Trendelenburg Function 4 22 Romo erase Bite OS Atte O Ud A Noe da 4 22 Replacement vc eee aad Ye chee ee Esko dd be bee IA a 4 23 Lateral Caregiver Unit sere seisus tapun tee thee id a 4 25 Removal ne Beet ath Ne Saath a eee ee edt eh a Ui a Dean ae 4 25 Replacement ec Rew hea ete ee aaa 4 26 Head Section Actuator asii este po cee oe eet ow A ao 4 27 RGMIOV al a Gu ox roe tae oe ch oe EY ec ewe tad Ca ee uN ee 4 27 Replacement sei bl li tine DA PA gee eee s 4 29 Head Section Release Cable 0er a a 4 30 Removal 22 nwo ook A E TA MeN oak aes eek 4 30 REPIMCEMENE uE 29 WOW Ba DOW es ENE PES x OO 4 31 PAOMMSUNCHU et adds id 4 31 Thigh Section ACtator lada ls a a ld oe ol e a i 4 33 Removal iio id A eden E A Laie ae poesia 4 4 33 Replacement os cris dorar ehh tach pira LS OREE taba beens 4 34 Foot Section Actuator Replacement LIIS9Ax only 0 000 4 35 Removal ci beds etapa detras da ita 4 35 Replacement Fao UE Sd A ese eee 4 36 Head Hilow Column o ce hee ea ee a ee ee eee A 4 37 Removal 2cica its bokybeade ough radio hteid eons tada Rhee 4 37 Replacement ves sota pian peed Adige a eek da 4 40 Poot HilowrG Ghani cuts eae ade as and coy a Sea ee wate 4 42 REMOVAL bu fie ea A SS hee EES thie eee E Ete ae 4 42 Replacement ce dada ads tados 4 46 o A A E AE 2 Get 4 48 Remoyal suo A IS AE eke i 4 48 Replacement ot
10. c Lock out the HL function W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur d Disconnect the bed from the electrical power supply If the bed has a battery ensure that no LED is lit up on the control pendant If this is not the case wait for around one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated Page 4 38 AvantGuard Electric Bed Service Manual QD1787 4 12 Head Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures ail CAUGHT HAZARD The elevation tool must be used to remove the column in safety Use the elevation tool otherwise personal injury could occur 9 Position the elevation tool at the level of the guide arms E and adjust the arms to the height corresponding to the bed frame Lift the arms situated at the head end of the bed see figure 4 20 on page 4 39 Figure 4 20 Upper Installation of Head HL Column Po 1680_445 10 Remove the two covers F situated at the base of the column see figure 4 22 on page 4 44 11 Using the 13 mm ratchet wrench loosen and remove the 4 screws securing the base of the column to the frame 12 Ensure that the bed frame is safely secured at its height then slide the column towards the head end of the bed
11. 33 Safety Tips Chapter 1 Introduction A A WARNING Only the facility authorised maintenance personnel should perform preventive maintenance on the on the AvantGuard Electric Bed Preventive maintenance performed by unauthorised personnel could result in personal injury or equipment damage WARNING Only the accessories specifically identified below are authorised to be used on the AvantGuard Electric Bed Any use of accessories not included in the list may compromise the safety of the AvantGuard Electric Bed WARNING Consult the AvantGuard Electric Bed User Manual before using the accessories to prevent any risk of injury to the patient or user Page 1 34 AvantGuard Electric Bed Service Manual QD1787 Identification and Caution Labels Identification and Caution Labels Figure 1 16 Identification and Caution Labels Hill Rom CLASS 1 BED IPX4 TYP B 230V 50 HZ 360VA MAX y INTERMITTENT WORKING 10 3min 30min Safe Working Load 185 KG Max Lift Up Load 250 KG F11282A 56330 PLUVIGNER FRANCE j LI158Ax 123456 00 00 0000 HRP000000001 56330 PLUVIGNER FRANCE LI159Ax 123456 00 00 0000 HRP000000001 56330 PLUVIGNER FRANCE PLUG INTO A RECOMMENDED PROTECTED INSTALLATION WITH DIFFERENTIAL CIRCUIT BREAKER MAX 30 mA 56330 PLUVIGNER FRANCE F11136 miil INDUSTR CONTINUITE A LA TERRE 0 177 EA
12. AvantGuard Electric Bed Service Manual QD1787 Page 4 41 4 13 Foot Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures 4 13 Foot Hilow Column Identification see figure 3 1 on page 3 3 Time required 40 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench with extension Elevation tool P N AG0052 Removal 1 Raise the sleep surface to the high position at least the head end if F Col cannot be used the casters in line with the bed and apply the brakes 2 Raise the TS and the FS to the high position and lower the HS to horizontal Raise and lock the siderails foot siderails in the folded position and remove the footboard Lower the foot siderail supports 3 Remove the hard surfaces of the TS and FS see Bed Frame Hard Surface Replacement on page 4 5 then remove the linen holder if fitted to the bed SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 4 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivat
13. Page 2 15 2 2 Control Unit Malfunction Chapter 2 Troubleshooting Procedures 2 2 Control Unit Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Comments e At the start of this troubleshooting procedure the CU is connected to the LCU e In this RAP the LCU is called X by convention whereas the LCU of the opposite side is called Y e The LCU is here understood to mean the Lateral Caregiver Unit and its cables 1 At least one of the other functions is working HL emergency trendelenburg etc Yes No Go to RAP 2 1 on page 2 13 2 Test for an open circuit at the LCU X Connect the CU to the other LCU Y The functions of the CU are working and the LEDs are lit Yes No uP Goto step 4 3 Replace the LCU X refer to procedure 4 7 on page 4 25 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 4 Test the control unit itself a Disconnect all the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AGO061 see Appendix on page 2 12 The functions of the CU are working Yes No Goto step 6 5 Replace the CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexi
14. Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Remove the hard surface of the FS see procedure 4 1 on page 4 5 4 Using the 13 mm ratchet wrench remove the two screws A supporting the egress bar and remove the assembly see figure 4 39 on page 4 79 Page 4 78 AvantGuard Electric Bed Service Manual QD1787 4 27 Egress Bar Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 39 Egress Bar Fixation ss 1680_449 Replacement 1 Perform the removal procedure in reverse order 2 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 79 4 27 Egress Bar Chapter 4 Removal Replacement and Adjustment Procedures NOTES Page 4 80 AvantGuard Electric Bed Service Manual QD1787 Chapter 5 Parts List Chapter Contents Warra t oo sich tices tete vies pto Rio da lee eke tii 5 3 Ordering Services Pais a AU la ge Leo da dd AE 5 4 Exchange Pol te s siase eye Jovan oboe ead ts oo ees 5 6 ine Watranty Exchanges eau n oes A ek Ge ees 5 6 Out of Warranty Exchanges ii eee Poe Pe eee eke 5 6 Recommended Spare Parts a eee toe e ados a ee de 5 7 Base Frame Elevation A id Aviles kee 5 8 Intermediate Frame 03 rra ie aes ant on ea eee ls 5 10 Sleep SUTIACE 2 cove he eee nie doen eae eee he See See eee eee 5 12 Sideral Assemblies e aaa ra
15. for reasons of force distribution Removal 1 Apply the brakes raise the sleep surface to the high position the TS and the FS to horizontal and the siderails lowered 2 Raise the HS to the high position W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 4 Remove the hard surface of the seat section refer to procedure 4 1 5 On each gas spring loosen the screw B securing the attachment plate C to the gas spring A by using the 13 mm combination spanner see figure 4 32 on page 4 64 AvantGuard Electric Bed Service Manual QD1787 Page 4 63 4 19 Head Section Gas Spring Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 32 Head Section Gas Spring 1680_410 6 Oneach gas spring remove the screw D securing the gas spring A by using the 13 mm ratchet wrench 7 Manually unscrew each gas spring from the corresponding attachment plate C then remove them from their housing NOTE The gas spring i
16. properly supported at the AC input terminal E 2 Adjust the sleep surface to reverse trendelenburg then to trendelenburg to ensure that the AC power cable can extend sufficiently without damage between the PSU and the strain relief bracket SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the installation is not correct and or if the earth continuity is compromised lt 3 Check that the earth continuity of the accessible metal parts is correct see Earth resistance on page 6 7 in the Preventive Maintenance Schedule Page 4 16 AvantGuard Electric Bed Service Manual QD1787 4 4 AC Power Cable Chapter 4 Removal Replacement and Adjustment Procedures W SHOCK HAZARD An abnormally high leakage current is symptomatic of a degradation within the AC power cable and the PSU Personal injury could result if the value of the leakage current is above 500 microamperes 4 Check that the leakage current of the bed is compliant see Leakage current on page 6 8 in the Preventive Maintenance Schedule 5 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 17 4 5 Power Supply Unit Chapter 4 Removal Replacement and Adjustment Procedures 4 5 Power Supply Unit Identification see figure 3 1 o
17. 10 on page 4 19 Initial conditions a The trendelenburg reverse trendelenburg function works see Function Checks on page 2 6 b The chair function works see Function Checks on page 2 6 c The emergency trendelenburg function works when the bed is connected to AC power see Function Checks on page 2 6 1 Check the configuration of the bed The bed is equipped with the battery backup function Yes No sl Go to step 8 2 Test the battery backup mode function Unplug the power connector from its power source and activate the various locked out functions The functions work with no AC power Yes No y Go to RAP 2 13 on page 2 37 3 Remove the cover of the PSU using the 13 mm combination spanner see figure 4 8 on page 4 16 Check the connections of the emergency trendelenburg battery pack G and the battery backup pack F see figure 4 10 on page 4 19 see table 4 1 on page 4 20 The connections are correct Yes No Connect the battery packs correctly and perform the Function Checks on page 2 6 4 The battery is not charging properly Plug the bed into an appropriate power source and ensure that it remains connected for at least 12 hours without using the batteries Unplug the bed afterwards Operate the emergency trendelenburg function The emergency trendelenburg function works after charging Yes No Go to step 6 AvantGuard Electric Bed Servi
18. Adjustment Procedures Adjustment Single and Double Band Casters Time required 10 min per caster Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench 10 mm Allen key Soft faced hammer Medium screwdriver Loctite adhesive 242 Elevation tool P N AG0052 1 Perform steps 1 to 5 of the caster removal procedure NOTE The braking adjustment is defined in the caster brake position by the number of turns made on this caster from the first point of contact between the band and the braking system The number of half turns depends on the type of caster A CAUTION When turning the double band caster it is very important not to turn the caster until it is no longer braked otherwise it may become impossible to adjust and damage may occur 2 Turn the caster body counterclockwise viewed from above by successive half turns until the caster is no longer braked 3 Turn the caster body clockwise viewed from above by successive half turns until the first point of contact between the band and the braking system 4 Ifthe double band caster body has been loosened too much it may be impossible to adjust it In this case perform the following a Unclip and remove the upper cover C with the small screwdriver place adhesive tape D around the shaft of the caster to block the two outer parts of the shaft see figure 4 30 on page 4 60 b Unclip the lower cov
19. Ae 7 9 Removal sch 94 baka II o e bad oer ee abe kage 7 10 Control Panel on a Flexible Arm Kit AD136A 20 0000 7 11 A A O E RO 7 11 Remo ostente O A AA AA 7 13 Hilow Pedal Kit ADI37A ooocoocoonorrormorrorrar ra 7 14 Sta e DS AAA A ed 7 14 REMOVAL AS ES SEA a 7 15 AvantGuard Electric Bed Service Manual QD1 787 Page 7 1 Chapter 7 Accessories NOTES Page 7 2 AvantGuard Electric Bed Service Manual QD1787 Accessories Chapter 7 Accessories Accessories A WARNING Only the accessories specifically identified below are authorised to be used on the AvantGuard Electric Bed Any use of accessories not included in the list may compromise the safety of the AvantGuard Electric Bed A WARNING Consult the AvantGuard Electric Bed User Manual before using the accessories to prevent any risk of injury to the patient or user Table 7 1 Accessories List Reference Description AC908A 2 receptacle bottle holder AC932A Stainless steel bottle holder 1 litre AC938A Stainless steel bottle holder 2 litres AC961A Stainless steel bottle holder 3 litres ADO16A Semi spherical urine holder France AC991A Triangular section urine holder AC959A Pivoting oxygen cylinder holder for B5 bottle diameter 150 mm AC962A Pivoting stainless steel bottle holder 3 litres ADI01A Pivoting oxygen cylinder
20. Perform the removal procedures in reverse order 2 Check the dates shown on the batteries and replace them if they are older than five years W SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the installation is not correct and or if the earth continuity is compromised 3 Check that the earth continuity of the accessible metal parts is correct see Earth resistance on page 6 7 in the Preventive Maintenance Schedule Page 4 20 AvantGuard Electric Bed Service Manual QD1787 4 5 Power Supply Unit Chapter 4 Removal Replacement and Adjustment Procedures W SHOCK HAZARD An abnormally high leakage current is symptomatic of a degradation within the AC power cable and the PSU Personal injury could result if the value of the leakage current is above 500 microamperes 4 Check that the leakage current of the bed is correct see Leakage current on page 6 8 in the Preventive Maintenance Schedule 5 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 21 4 6 Battery Backup and Emergency Trendelenburg Function Chapter 4 Removal Replacement and Adjustment Procedures 4 6 Battery Backup and Emergency Trendelenburg Function Identification The batteries are located in the PSU see figure 3 1 o
21. QD1787 2 1 No Functions Work Chapter 2 Troubleshooting Procedures 2 1 No Functions Work NOTE Identification of components see figure 3 1 on page 3 3 NOTE This RAP assumes that no functions work from the HL pedal the control pendant or control panel on a flexible arm NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 ed SHOCK HAZARD The use of electrical equipment carries a potential risk of electric shock Personnel must be trained and instructed in the inherent risks of electrical equipment 1 Verify that the AC power cable works properly The power connector is correctly plugged into the power outlet Yes No y Connect the plug check that it is correctly plugged in Then go to Function Checks on page 2 6 2 Unplug the power connector from the power outlet and use an appropriate test device to check for the presence of the voltage at the power outlet The voltage is correct Yes No y gt Inform the customer of the problem concerning the power outlet Connect the bed to a working power outlet then go to Function Checks on page 2 6 3 Inspect the AC power cable for any damage to the cable up to the PSU input no crushing apparent or cut visible The AC power cable is in satisfactory condition Yes No gt Replace the AC power cable refer to procedure 4 4 on page 4 15 4 Adjust the HS to the high position using the CPR cont
22. RAP 2 21 AvantGuard Electric Bed Service Manual QD1787 Page 2 5 Function Checks Chapter 2 Troubleshooting Procedures Function Checks 1 The Initial Actions have been performed Yes No Go to Initial Actions on page 2 4 Apply the brakes on the bed and raise and lock the siderails in the high position with the foot siderails folded Carry out the following instructions Unplug the power connector from the power outlet b Discover the causes of the problem for instance cables or connectors ripped out or damaged c Plug in the power connector to an appropriate power source d Switch on the power outlet switch if one is fitted No LED is lit on the control unit Yes No y Goto step 5 Operate the emergency trendelenburg function if the bed is equipped with it This function works Yes No y Go to RAP 2 1 on page 2 13 Unlock at least one bed function such as the HS The LED corresponding to the function is lit green Yes No y Go to RAP 2 2 on page 2 16 Lock out all the functions by applying the normal function management sequence All the control unit LEDs except reverse trendelenburg LED and non low position HL LED flash during the sequence and go off at the end of the sequence Yes No gt Goto RAP 2 3 on page 2 18 Unlock all the functions by applying the normal function unlocking sequence All the control unit LEDs except reverse trendelenburg
23. Section Three movements occur simultaneously when the HS elevation is activated elevation of the HS retraction of the HS towards the head end of the bed and the sliding extending movement of the HS hard surface on the HS frame These Page 1 16 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction movements enlarge the sitting area reduce pelvic pressure enable the patient to breathe more easily by reducing abdominal pressure limit the shearing forces and provide greater support for the head and neck CPR Emergency Release The AvantGuard Electric Bed has an emergency lower function for the HS The CPR release M is situated in the central part of the bed alongside the LCU K at either side under the sleep surface Handles integral with the HS hard surface support the section during the emergency release phase Activating the HS emergency release function Hold one side of the HS and pull one of the yellow CPR handles Keep the handle pulled to control the lowering of the HS Deactivating the HS emergency release function Release the CPR handle The HS raise A function will then be available when the relevant control is used see figure 1 11 on page 1 14 Thigh Section The TS of the AvantGuard Electric Bed can be adjusted independently to obtain the desired inclination up to its maximum value This function E is activated from the control pendant or contro
24. Support Bar Kit AD126A Chapter 7 Accessories 7 1 Accessory Support Bar Kit AD126A This kit contains the accessory support bars and the parts that enable them to be installed on the bed Note that only one accessory support bar can be installed on the bed if the use of a medical device support rail is required other kit The accessory support bar kit includes the following e Two accessory support bars one right hand and one left hand e Four guides e Four securing screws Figure 7 1 Accessory Support Bar Kit 1680_746 Identification see figure 3 1 on page 3 3 Time required 10 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required 6 mm Allen key 1 Allen is a trademark of Industrial Fasteners Inc AvantGuard Electric Bed Service Manual QD1787 Page7 5 7 1 Accessory Support Bar Kit AD126A Chapter 7 Accessories Installation 1 Check the composition of the kit 2 Apply the brakes adjust the bed to high position the sleep surface horizontal and the siderails to high position W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control unit If this is not the case wait for approximately one minute until the battery m
25. Test the function of the FS actuator intermediate switch The problem only occurs on the chair and TS functions Yes No Go to step 7 6 Go to step 11 7 Perform the following Remove the cover of the PSU b Cross connect the connectors of the TS amp FS actuators 5 brown 2 green on the power board see table 4 1 on page 4 20 c Press the emergency release of the sleep surface sections lower button O see figure 1 11 on page 1 14 The FS function works the TS actuator rod extends to mid travel Yes No J The FS actuator is defective Replace it refer to procedure 4 11 on page 4 35 AvantGuard Electric Bed Service Manual QD1787 Page 2 31 2 9 Foot Section Malfunction Chapter 2 Troubleshooting Procedures 8 Test the function of the TS actuator intermediate switch a Connect the FS actuator cable to its socket 2 green b Either connect new TS actuator in extended position or connect a TS actuator simulation connector RC P N AGO063 see Appendix on page 2 12 to the socket 5 brown on the power board or using an ohmmeter check the state of the intermediate switch of the TS actuator The actuator is extended the contacts 3 amp 4 of the 4 pin connector socket 5 brown must be short circuited see Figure 2 2 on page 2 32 and with the rod retracted the circuit must be open Figure 2 2 4 pin Connector 1680_202 9 The FS function works Yes No y
26. The overall length of the bed can be increased significantly by sliding the crossbar of the foot panel along its guide rails The mattress can be extended with the aid of a mattress extension cushion available as an accessory when the extension has been pulled out Two positions are available stored and extended To use the extension lift the handle located under the footboard then pull out the extension Ensure that the extension is secured in the notch on the guide rail To store the extension lift the handle under the footboard then push the extension Ensure that the extension is secured by the notch on the guide rail Linen Holder Optional The AvantGuard Electric Bed can be equipped with a linen holder that slides under the FS When in use it provides space for sheets and blankets when the bed is being made The maximum load supported is 15 kg To use the linen holder pull on the board located under the footboard then tip the metal stop towards you To store the linen holder tilt the metal stop away from you then push the board under the footboard Accessory Support Bar Optional or Accessory Two accessory support bars Hill Rom accessories urine bag drainage bag etc are available under the central section of the sleep surface The design enables the bags to be moved easily inwards or outwards by sliding them on their supports Medical Device Support Bar Optional or Accessory Two medical device suppor
27. and the FS to horizontal and remove the footboard 2 Store the extension and remove the linen holder table see procedure 4 25 on page 4 75 W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 4 Remove the hard surface of the FS see procedure 4 1 on page 4 5 5 Using the 13 mm ratchet wrench remove the 2 screws A supporting the extension while supporting this extension see figure 4 38 on page 4 77 Page 4 76 AvantGuard Electric Bed Service Manual QD1 787 4 26 Extension optional Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 38 Extension 1680_438 6 Lift the end of the extension without operating the control lever then remove the extension plus guide rail assembly from the support notches B on the bed NOTE The two sliderails are assembled with white petroleum jelly see Tools and Supply Requirements on page 6 18 Two of the extension slide pads are equipped with grounding spring blades
28. are acessable via AC power or a backup battery optional e Hilow sleep surface elevation raise lower e Head section raise lower Thigh section raise lower e Foot section raise lower LI159Ax only e Trendelenburg reverse trendelenburg e Emergency trendelenburg These controls are electrically operated except for the FS on LI158Ax and are accessible to the patient except for trendelenburg reverse trendelenburg and emergency trendelenburg or the caregiver from the control pendant control panel on a flexible arm lateral caregiver unit emergency trendelenburg only or the hilow pedal HL only AvantGuard Electric Bed Service Manual QD1787 Page 1 7 Introduction Chapter 1 Introduction Figure 1 4 AvantGuard Electric Bed A C x D 1680_117 1787_105 Page 1 8 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction Table 1 2 AvantGuard Electric Bed Feature Identification Letter Description TotalGuard Integral Siderail release TotalGuard Integral Siderail Control panel on a flexible arm Control pendant MobiBar Egress Bar Vertical removable footboard Integral bed extension Bumper at each corner PE is Ga ED mol A gt Foot siderail support with trendelenburg reverse trendelenburg angle and horizontal position indicator Emergency trendelenburg Electrical function mana
29. button at the desired angle or when maximum lowering is reached Automatic Contour Every AvantGuard Electric Bed has a special feature called automatic contour This feature enables the TS of the bed to be raised automatically to a preset angle when the HS is raised from its horizontal position provided that the TS is not locked out The object of this feature is to prevent the patient slipping towards the foot of the bed when the HS is raised It is activated from the control pendant C or control panel on a flexible arm B optional or accessory see figure 1 4 on page 1 8 Automatic contour raise this feature is systematically active for the raise function when the specific HS raise button is activated on the control pendant or the control panel on a flexible arm Automatic contour lower this feature is systematically active for the lowering function when the specific HS lower button is activated on the control pendant or control panel on a flexible arm The TS lower is only carried out when the HS reaches low position A specific button B enables the elevation of the HS only if the automatic contour feature is not required The automatic contour feature can also be disabled by locking out the TS from one of the lateral caregiver units If this feature is to be used later on the lockout must be deactivated see figure 1 11 on page 1 14 EasyChair Automatic Chair Position The combination of the electric controls of the s
30. caster one steer position then neutral then brake A CAUTION Ensure the caster is correctly positioned with respect to the movements of the hexagonal bar Disregarding this caution may cause the brake and steer system to malfunction 3 Position the socket in its location on the frame then using the 10 mm AllenTM key which you insert into the caster instead of the hexagonal bar set the caster to brake position then the steer position by rotating it in one direction or the other Note the direction of operation with respect to the hexagonal bar of the bed and turn the body of the caster in the socket so that the operating movements coincide Remove the 10 mm Allen key then fit the socket in its housing on the frame by inserting it into the upper opening of the caster in the hexagonal bar head end of bed or the brake bar foot end Set the single brake bar to brake position and check that the new caster is braked If the caster is not braked reset the bar to neutral remove the socket from the caster carry out a half turn with the rod of the caster then remount the socket and hexagonal bar Perform steps 1 to 5 of the removal procedure of the caster in the reverse order Perform the Function Checks on page 2 6 1 Allen is a trademark of Industrial Fasteners Inc Page 4 58 AvantGuard Electric Bed Service Manual QD1 787 4 17 Caster Brake and Steer Caster Chapter 4 Removal Replacement and
31. corners of the bed frame Restraining Strap Attachment Points The AvantGuard Electric Bed has six attachment points for restraining straps These points also support the hard surfaces of the SS TS and FS Protective Bumpers The protective bumpers are situated on the four corners of the bed on the bed frame AvantGuard Electric Bed Service Manual QD1787 Page 1 23 Introduction Chapter 1 Introduction I V Pole Sockets The AvantGuard Electric Bed has four IVP sockets placed at each corner of the bed An LV pole accessory can be inserted into each socket The two sockets at the patient s head also enable a patient helper to be attached Optional Features Hilow Pedal Optional or Accessory The AvantGuard Electric Bed has as an option or accessory a bilateral hilow pedal which enables nursing staff to raise or lower the sleep surface Battery Backup Optional The AvantGuard Electric Bed can be equipped with a battery backup that ensures the normal operation of the bed when it is moved or during AC power failure Detection of Bed Connected to AC Power but Brakes not Applied Optional The brake must always be applied when the bed is connected to a power outlet If the brake is not applied a beep will sound as a reminder This safety feature prevents the AC power cable from being pulled when the bed is moved The beep will stop when the brake is applied or when the power connector is u
32. correctly into the supports provided and unlock when the unlocking control is activated The locking unlocking operates correctly Yes No y Go to RAP 2 17 on page 2 45 Fold the foot siderails after having unlocked them Check that the head siderails lock correctly then operate the release controls to lower them The locking is correct and the siderails lower without any problem or abnormal noise Yes No y Goto RAP 2 18 on page 2 46 Try to move the bed with the brake still applied The four wheels are locked and prevent any movement Yes No 4 Goto RAP 2 19 on page 2 47 Page 2 10 AvantGuard Electric Bed Service Manual QD1787 Function Checks Chapter 2 Troubleshooting Procedures 33 For beds not equipped with 5th wheel system set the brake steer bar to steer position and trigger the lengthways locking of the steer caster The caster locks in the expected position Yes No sh Go to RAP 2 20 on page 2 48 34 For beds equipped with 5th wheel system push the bed in the lengthways direction to lock the 5th wheel in the direction of the bed movement and move the bed by veering to the left then to the right The caster unlocks in the expected position and remains locked when the bed turns to the right or left Yes No y Goto RAP 2 21 on page 2 49 35 For beds equipped with 5th wheel system push the bed sideways to move it diagonally The caster unlocks to allow sideways moveme
33. ee al ee 1 3 Organisation 2 iighohdeinlesbebtacew bebe aes ta adds 1 3 Chapter 1 Introduction i 36iehiee seh i rea e beac aged shee ds 1 3 Chapter 2 Troubleshooting Procedures 0 0 0 0 cece eee eee 1 3 Chapter 3 Theory of Operation 2 45 5 swt OS ay on gue ewe SOS Ga ees 1 3 Chapter 4 Removal Replacement and Adjustment Procedures 1 3 Chapters Parts Liste tea ey et dao c a be te e BR te E 1 3 Chapter 6 General Procedures 0 0 0 0 cece eee eee ee ene 1 3 Chapter 7 ACC SSOTIES osai ss tues taae pew PA deed gas Qed ee 1 4 Typographical Conventions 0 0c cece eee eee eens 1 5 Acronym Laa o a wee g be Lue ete A 1 6 AUC GUUCH OM ass 1 7 OCW IEA A DS Se ee Lee eee ae 1 7 Bed Positions sc6 i bi ms ta pen tine dagen Mode dhe hei ged fede 1 10 Operating Precautions 0 cee nesses eeeede veeaaee Lei eee eee eee a 1 14 Features 2 0 dw eed A ae aso Meh a 1 14 Hilow POCO siese oye A eons gare eee eee yee 1 16 Hedd Section set gu adele td pet ede na pida dass 1 16 Retract ble Head Sections Litas el Gan a eee le ae ees la 1 16 CPR Emergency Release vivir a 1 17 Thigh Section seriada o e heb takes bie tad 1 17 Poot Section auch evs nseri Ane A See ee eyes 1 17 Automatic Conlara dy eos 1 18 EasyChair Automatic Chair Position oooooooooooooooo o 1 18 Lateral Caresiver Unit A es A a da ai ees 1 19 TotalGuard Integral Siderails 0 0 00 eee eee ee 1 21 Brake ii DA E
34. four casters are locked by trying to move the bed The braking is correct Yes No Adjust the caster s refer to procedure 4 17 on page 4 56 then go to Final Actions on page 2 12 3 The problem is solved Yes No y gt Replace the defective caster s refer to procedure 4 17 on page 4 56 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 4 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 47 2 20 Steering Malfunction Chapter 2 Troubleshooting Procedures 2 20 Steering Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE Initial conditions none 1 Check the caster attachment Set the brake steer bar to brake position Check that the screws or supports are not damaged loose or missing from the caster attachments refer to the assembly diagrams of chapter 5 All the supports and screws are secure and undamaged Yes No y Replace the missing parts refer to the corresponding view of chapter 5 then go to Function Checks on page 2 6 Check the steering function Set the brake steer bar to steer position push the bed lengthways and check that the head right caster locks in the lengthways direction of the bed The caster locks in the lengthways direction Yes No Replace the cas
35. has a battery ensure that no LED is lit on the control pendant If this 1s not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 5 Press the central button A of the foot siderail while holding the other end then remove it see figure 4 36 on page 4 71 Page 4 70 AvantGuard Electric Bed Service Manual QD1 787 4 22 Siderail Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 36 Folded Foot Siderail 1680_435 Replacement 1 Perform the removal steps in reverse order A CAUTION Always set the foot siderail to the folded position Failure to do so could result in equipment damage 2 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 71 4 23 Foot Siderail Support Chapter 4 Removal Replacement and Adjustment Procedures 4 23 Foot Siderail Support Identification see figure 3 1 on page 3 3 Time required 10 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench Removal 1 Apply the brakes raise the sleep surface to the high position lower the HS to horizontal and raise the TS and the FS to the high position W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disc
36. holder for D bottle diameter 110 mm AD102A High pivoting oxygen cylinder holder for E bottle diameter 110 mm AC963A Syringe driver holder table AC968A Equipotential cable ADO80B Chrome plated patient helper AD081B Adjustable patient helper AD082A Stainless steel I V pole AC953A I V hooks on patient helper AD128A Monitor stand table AC923A Label holder AD126A Accessory support bar kit AD127A Medical device support rail kit AD066A Cliniplot HI Mattress ADO85A Primo Dynamic Low Pressure Mattress System AD117A01 Clinisert 2 Mattress AD118A01 Clinisert 2 Mattress AvantGuard Electric Bed Service Manual QD1787 Page 7 3 Accessories Chapter 7 Accessories Reference Description AD119A01 Clinisert 2 Mattress AD120A01 Clinisert 2 Mattress AD140A Duo 21M Multi Mode Mattress System AD052C Climat System mattress AD142A AvantGuard Electric Bed removable frame AD082A Stainless steel I V pole with elbow AC806A I V pole support 5 branch socket 9099007A Cane holder AD134A MobiBar Egress Bar kit AD135A Self steering 5th wheel kit AD136A Control panel on a flexible arm kit AD137A Hilow pedal kit AD138A Extension kit AD139A Linen holder kit Page 7 4 AvantGuard Electric Bed Service Manual QD1787 7 1 Accessory
37. installation is not correct and or if the earth continuity is compromised 10 Check that the earth continuity of the accessible metal parts is correct see Earth resistance on page 6 7 in the Preventive Maintenance Schedule 11 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 47 4 14 Hilow Pedal Chapter 4 Removal Replacement and Adjustment Procedures 4 14 Hilow Pedal Identification see figure 3 1 on page 3 3 Time required 15 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required Small screwdriver 13 mm ratchet wrench Removal NOTE The two pedals are independent The following procedure applies to both pedals 1 Apply the brakes Set the siderails to high deployed the TS and the FS to horizontal position and the sleep surface to the high position SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Free the foot frame
38. is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Free the head and foot frame cover A Note that it is secured by clips at each end see figure 4 25 on page 4 51 1 Allen is a trademark of Industrial Fasteners Inc Page 4 52 AvantGuard Electric Bed Service Manual QD1787 4 16 Single Brake Bar Chapter 4 Removal Replacement and Adjustment Procedures A CAUTION If the sleep surface cannot be raised to mid height for whatever reason the arms of the elevation tool must not be positioned at the level of the actuators Disregarding this caution may cause the head section and foot section actuators to malfunction A WARNING Never lift the entire bed with the elevation tool lts maximum load is 150kg Disregarding this warning may cause the elevation mechanism to malfunction and could cause injury 4 Position the arms of the elevation tool under the right hand side of the frame Figure 4 26 Removal of the Braking Bar 1680_444 5 For beds fitted with a 5th wheel system use the 6 mm Allen key to remove the screw A securing the braking bar stop B see figure 4 27 on page 4 54 1 Allen is a trademark of Industrial Fasteners Inc AvantGuard Electric Bed Service Manual QD1787 Pag
39. malfunction see RAP 2 14 In all cases check the date indicated on the battery it is recommended to replace the battery if the date is over five years Operation on battery backup according to configuration Ensure that the bed is unplugged from the power outlet and operate all the functions using the control unit Recharge the batteries or replace them if necessary Troubleshoot in the event of a malfunction see RAP 2 8 In all cases check the date indicated on the battery it is recommended to replace the battery if the date is over five years Siderail structure and attachment Check that the head and foot siderails are not bent or twisted Check that the rotation and release system of each foot siderail is in good condition Fold back the foot siderails and check that the release control of each head siderail operates correctly Check the condition of the elevation mechanism of each head siderail Test that they lock correctly in high position by pushing them sideways Note that the guide rods on the siderails are self lubricating Replace the foot siderails or head siderail subassembly if necessary refer to procedure 4 22 or procedure 4 23 AvantGuard Electric Bed Service Manual QD 1787 Page 6 13 Preventive Maintenance Chapter 6 General Procedures Function Foot siderail locking mechanism Procedure Fold back the foot siderails on the head siderails Inspect
40. nuts using the two 10 mm combination spanners 4 Gently pull on each handle to ensure that there is no play after the nuts have been tightened Page 4 32 AvantGuard Electric Bed Service Manual QD1787 4 10 Thigh Section Actuator Chapter 4 Removal Replacement and Adjustment Procedures 4 10 Thigh Section Actuator Identification see figure 3 1 on page 3 3 Time required 15 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench Removal 1 Apply the brakes raise the sleep surface to the high position the HS in the high position and the TS if possible and the FS to horizontal Fold the foot siderails then lower the siderails Remove the headboard 2 Remove the hard surfaces of the SS and the TS refer to procedure 4 1 W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 4 Using the 13 mm ratchet wrench loosen and remove the two screws A securing the cover B on th
41. of the first momentary push except for the trendelenburg reverse trendelenburg LED Trendelenburg Reverse Trendelenburg Trendelenburg reverse trendelenburg is obtained according to the following sequence Push momentarily the LCU key in the function management trendelenburg reverse trendelenburg access opening A of the LCU The trendelenburg reverse trendelenburg function LED is on continuously on the control unit Continuously press the button of the trendelenburg or reverse trendelenburg according to requirements of the control pendant or control panel on a flexible arm Release the button when the sleep surface reaches the desired angle or maximum angle Push momentarily the LCU key in the same opening to inhibit the function or function management will automatically de activate after approximately 1 min The LED of the trendelenburg reverse trendelenburg function goes off on the control unit Page 1 20 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction The sleep surface can be inclined in 2 ways in trendelenburg where the head of bed is lowered in reverse trendelenburg where the foot of bed is lowered The maximum scope of trendelenburg movements is obtained irrespective of the the initial height of the sleep surface This function is activated by the control pendant or control panel on a flexible arm after validation on the lateral caregiver
42. pedal optional optional A fe BM optional 1680_300 Page 3 4 AvantGuard Electric Bed Service Manual QD1787 Component Positioning and Electrical Diagrams LI158Ax LI159Ax Chapter 3 Theory of Operation Figure 3 3 Wiring Schematic for LI159Ax Management Battery indicator mode Emergency Emergency trendelenburg trendelenburg Se Battery indicator mode CU Control ae ASE pendant panel on a flexible arm optional HL pedal HL pedal po Ne FS 1680_302 AvantGuard Electric Bed Service Manual QD1787 Page 3 5 Theory of Operation Chapter 3 Theory of Operation Theory of Operation The operation of the AvantGuard Electric Bed is divided into 2 sub systems e Electrical e Mechanical The interaction between these sub systems allows all the operational functions to be used The electric functions are driven by low voltage actuators Each of the sub systems is described separately in the following sections The schematic diagrams are located at the beginning of this chapter Page 3 6 AvantGuard Electric Bed Service Manual QD1787 Mechanical System Chapter 3 Theory of Operation Mechanical System The mechanical sub system consists of two connected systems e Bed frame e Sleep surface Retractable Head Section The retractable head section feature is achieved by controlling a sliding mo
43. principle involves controlling the two hilow columns in opposing directions according to the position required irrespective of the initial height and position of the sleep surface As long as one of the two buttons remains pressed the two columns continue to move until their internal EOT is reached The maximum inclination can therefore be reached as long as AC power or battery backup is available Page 3 16 AvantGuard Electric Bed Service Manual QD1787 Electrical System Chapter 3 Theory of Operation NOTE The warranty will be rendered null and void if any electrical component is opened except where the PSU cover is removed by authorised personnel AvantGuard Electric Bed Service Manual QD1787 Page 3 17 Electrical System Chapter 3 Theory of Operation NOTES Page 3 18 AvantGuard Electric Bed Service Manual QD1787 Chapter 4 Removal Replacement and Adjustment Procedures Chapter Contents Bed Frame Hard Surface Replacement 0 0 0 eee eee eee 4 5 Removal cts Bs We ee Eee A a eh fam ok Ai 4 5 Replacements 72 iara ga iai ps a te ess Y Ee Bae a A E a ER 4 7 Control Pendant uri h2ta bad ie braces pita od beatle Koad 4 10 Removal corro rss 4 10 Replacement sta Cae nae ee Se ae Dee i eee eek 4 11 Control Panel on a Flexible Arm e ts 2S oe bes eee ees Soe yen 4 12 Removal sot hedaget tect egg tenet hoe A ae 4 12 Replacement nenien EY NO e e do de o e Oe Boe 4 13 A
44. relating to the use of the AvantGuard Electric Bed refer to the user manual WARNING Consult the user manual before using the AvantGuard Electric Bed Failure to do so could result in personal injury or equipment damage WARNING Only facility authorised personnel should troubleshoot the AvantGuard Electric Bed Troubleshooting by unauthorised personnel could result in personal injury or equipment damage A WARNING The column must be replaced if it becomes defective The column must never be dismantled Failure to observe this warning could result in personal injury A WARNING Two persons are required to test the articulation of the head section Disregarding this precaution may result in personal injury or equipment damage A WARNING Only the AC power cables listed in chapter 5 can be used on the AvantGuard Electric Bed Other cables may compromise the safety and conformity of the bed A WARNING Never lift the entire bed with the elevation tool lts maximum load is 150kg Disregarding this warning may cause the elevation mechanism to malfunction and could cause injury A WARNING Do not remove the identification label from the bed the references featuring on it are essential for identifying the bed and guaranteeing the safety of the maintenance operations Failure to observe this warning could result in personal injury or equipment damage AvantGuard Electric Bed Service Manual QD1787 Page 1
45. the bed into the power outlet and lock out all the functions except for the HL function 5 Operate the HL raise function with the control pendant so as to bring the upper section of the column to the level of the bed frame SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 6 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this 1s not the case wait for about one minute until the battery mode is deactivated 7 Perform the removal procedure in reverse order NOTE Leave as much play as possible at the level of the HL column cable between this cable and the PSU 8 Refit the two covers F situated at the base of the column see figure 4 22 on page 4 44 Page 4 46 AvantGuard Electric Bed Service Manual QD1 787 4 13 Foot Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures 9 Operate the trendelenburg function of the bed to ensure that there is no tension on the power supply cable of the column during operation If necessary adjust the play of the cable with respect to the PSU e SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the
46. the bed is in low position Replace the defective column s in the event of a malfunction Troubleshoot in the event of a doubt see RAP 2 4 AvantGuard Electric Bed Service Manual QD1787 Page 6 9 Preventive Maintenance Chapter 6 General Procedures Function Procedure Hilow pedals Activate the raise and lower controls of the bed by means of optional the pedals and check that the working of the corresponding functions Replace the pedals if necessary Retractable Head Inspect the actuator assembly check that the pins and Section RHS Truarc circlips are not missing Carry out a full raise lower cycle of the corresponding section Check that there is no friction or abnormal noises and that no audible overload indication can be heard during the movement Replace in the event of a malfunction Retractable Head Inspect the hard surface assembly check that the pins and Section RHS Truarc circlips are not missing and Carry out a full raise lower cycle of the head section corresponding Check that the articulation moves through approximately 10 hard surface cm along the plane of the bed frame during the elevation movement Check that the hard surface moves approximately 4 cm on the section during the movement Check that the hard surface and its drive system under the head section are in good condition Check that there is no friction or abnormal noises Replace any damaged parts ref
47. the elements of the battery backup Disregarding this caution will permanently damage the power supply P C board of the bed CAUTION Ensure that the elastic found on the new actuators is not wound around the actuator body before removing it cut If necessary turn the actuator end rod in the appropriate direction to unwind this elastic Indeed it is used to indicate the correct position of the switches and internal end of travels of the actuator The incorrect position of these elements may damage the actuator or the structure of the bed Page 1 30 AvantGuard Electric Bed Service Manual QD1787 Safety Tips Chapter 1 Introduction A CAUTION If the sleep surface cannot be raised to mid height for whatever reason the arms of the elevation tool must not be positioned at the level of the actuators Disregarding this caution may cause the head section and foot section actuators to malfunction A CAUTION Ensure the caster is correctly positioned with respect to the movements of the hexagonal bar Disregarding this caution may cause the brake and steer system to malfunction A CAUTION When turning the double band caster it is very important not to turn the caster until it is no longer braked otherwise it may become impossible to adjust and damage may occur A CAUTION Always set the foot siderail to the folded position Failure to do so could result in equipment damage 4 al CAUGHT HAZARD The elevat
48. the foot siderail suport assembly and check that the springs pins and Truarc circlips are fitted and that the securing screws are tightened Operate the unlocking latch of each siderail support so as to lower then raise it to high position to ensure that the support locks correctly Deploy the foot siderails on the supports and ensure that the siderail locking mechanism operates correctly on each support Check the operation of the trendelenburg reverse trendelenburg angle indicators Replace if necessary refer to procedure 4 23 Extension optional Operate the extension to check that its movement is correct and that it locks properly into the different positions when the lever is released Lubricate the guide rails with white petroleum jelly see Tools and Supply Requirements on page 6 18 Replace the defective or missing parts refer to procedure 4 26 Linen holder Check that the linen holder operates correctly optional Replace the damaged or missing parts if necessary refer to procedure 4 25 Articulated joints Check that the articulations of the bed operate without making any noise HL column bearings head and thigh sections Mobibar Deploy each egress bar and ensure that they lock in high Egress Bar position optional Operate each release lever and ensure that the corresponding lever folds correctly Replace if necessary Casters Check for cuts wear and quality of
49. the tread etc Replace if necessary refer to procedure 4 17 Braking and Check the brakes to see whether the bed moves when the steering brake bar is pressed Adjust if necessary refer to procedure 4 17 on page 4 56 Check the steering mechanism Replace or adjust the steering control elements of the steering caster if necessary Replace the caster if necessary refer to procedure 4 16 5th wheel steering Ensure that the 5th wheel locks correctly when the bed is system optional moved in a straight line and unlocks when a sideways force is applied Replace the system if necessary refer to procedure 4 18 Page 6 14 AvantGuard Electric Bed Service Manual QD1787 Preventive Maintenance Chapter 6 General Procedures Function Detection of Bed connected to AC Procedure Plug the bed into the power outlet and set the brake bar to neutral The alarm must be heard power but brakes Apply the brakes on the bed The alarm must stop sounding not applied Troubleshoot in the event of a malfunction refer to optional procedure 2 15 Accessory Check the position of each bar on the bed and that the support bar and securing screws are correctly tightened medical Replace any missing or damaged parts if necessary refer to equipment procedure 7 1 or procedure 7 2 support bar medical device support rail according to configuration Administrative procedures Carr
50. thigh 4 AG0002 1 Hard surface foot 5 AG0006 1 Linen holder kit 6 AGO0021 1 Actuator head section kit 7 AG0023 1 Actuator foot section kit 8 AG0022 1 Actuator thigh section kit 9 AG0050 2 Gas spring kit 10 AG0041 1 Control panel on a flexible arm LI158Ax kit 11 AG0045 1 Control panel on a flexible arm LI159A x kit 12 AG0026 1 Intermediate frame kit 13 AG0053 1 Head section kit 14 AG0027 1 Fracture kit 15 AG0039 1 Fixed entension kit 16 AG0040 1 Extension kit 17 AG0060 1 Surface support mattress 18 AG0056 1 CPR cable and handles kit NOTE AvantGuard Electric Bed Service Manual QD1 787 Page 5 13 Siderail Assemb lies Chapter 5 Parts List Siderail As semblies Figure 5 5 Siderail Assemblies 1787_504 Table 5 5 Siderail Assemblies Item Number Part Number Quantity Description 1 AG0008 1 Head right siderail kit 2 AG0009 1 Foot right siderail kit 3 AG0011 1 Head left siderail kit 4 AG0012 1 Foot left siderail kit 5 AG0010 1 Foot right siderail support kit 6 AG0013 1 Control pendant LIIS8Ax 7 AG0014 1 Control pendant LIIS9Ax 8 AG0018 1 Foot left siderail support kit NOTE All the kits contain their respective fixing systems as well any applicable adhesive labels Hardware and other small metal or plastic parts are available in the kits Page 5 14 AvantGuard Electric Bed Service Manual QD1787 Miscella
51. to remove it from its guide rails E on the bed frame then remove it from the bed Note that the power supply cable of the column comes out from the side of the PSU see figure 4 20 on page 4 39 13 If the bed is imobilised in the low position gradually bring the head end of the bed frame with the H Col to high position using the elevation tool so as to slide the H Col from its guide rails E AvantGuard Electric Bed Service Manual QD1 787 Page 4 39 4 12 Head Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures Replacement y SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the installation is not correct and or if the earth continuity is compromised 1 Remove the two covers F situated at the base of the column see figure 4 22 on page 4 44 2 Position the new HL column of the frame without mounting it and connecting its connector to 6 of the main board C of the PSU see figure 4 10 on page 4 19 see table 4 1 on page 4 20 3 Plug the bed into an appropriate power source and lockout all the functions except for the HL function 4 Operate the HL raise by using the control pendant in order to raise the upper section of the column to the level of the bed frame SHOCK HAZARD Electric current flows through the connections when
52. track situated under the central crossbar B by leaving as much play as possible on the cable at the level of the HS articulation 12 Close the cable track situated under the central crossbar B of the bed frame 13 Replace the cable cover A on the central rail of the bed frame see figure 7 8 on page 7 12 14 Perform the Function Checks on page 2 6 of this manual Removal Perform the installation procedure in reverse order AvantGuard Electric Bed Service Manual QD1 787 Page 7 13 7 5 Hilow Pedal kit AD137A Chapter 7 Accessories 7 5 Hilow Pedal Kit AD137A This kit contains the two hilow pedals and the parts that enable them to be installed on the bed The hilow pedal kit includes the following e Two hilow pedals e Two cable tracks e Two securing bolts Identification see figure 3 1 on page 3 3 Time required 15 min Number of people 1 Possibility of intervening with patient on the bed Possibility of intervening with patient on the bed yes Tools required Small screwdriver 13 mm ratchet wrench Installation 1 Apply the brakes adjust the bed to high position the head thigh and foot sections horizontal Lift the foot siderails 2 Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control unit If this is not the case wait for about one minute until the battery mode is deactivated SHOCK HAZARD Electric current f
53. vs a ITA 1787_120 AvantGuard Electric Bed Service Manual QD 1787 Page 1 13 Introduction Chapter 1 Introduction Operating Precautions A WARNING Driven bed mechanisms can cause serious injury Do not operate bed controls until all persons are clear of mechanisms A bed articulation function can be halted by releasing the control pressing and holding the switch for the opposite function or by unplugging the AC power connector Before you operate the bed make sure to read and understand the contents of the AvantGuard Electric Bed User Manual and this manual It is important that you read and strictly follow all safety guidelines in these manuals To help prevent falls by patients always leave the bed in the low position and set the brake when the patient is unattended Features e SHOCK HAZARD The use of electrical equipment carries a potential risk of electric shock Personnel must be trained and instructed in the inherent risks of electrical equipment Figure 1 11 Control Pendant LI159Ax 1787_100 Page 1 14 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction Figure 1 12 Control Pendant LI158Ax 1787_101 Table 1 3 Control Pendant Functions Letter Description Head section raise with automatic contour Head section raise without automatic contour Head section lower wit
54. working Yes No y Go to RAP 2 1 on page 2 13 Perform the following to verify the connection of the BM a Remove the foot cover b Check that the BM connector is properly connected to the F Col connector under the cable cover F see figure 4 22 on page 4 44 The function bed connected to AC power but brake not applied detection works Yes No yo Goto step 4 Perform the Function Checks on page 2 6 Perform the following to verify the function of the BM a Disconnect the 2 pin connector at the bottom of the F Col b Using a DMM set to ohms at the terminals of the BM check that the BM operates correctly Idle not braked circuit closed Active braked circuit open The switch operates correctly Yes No Replace the BM and brake detection cabling subassembly refer to procedure 4 15 on page 4 50 then go to to Final Actions on page 2 12 Perform the following to verify the cabling of the F Col a Reconnect the 2 pin connector to the bottom of the F Col b Disconnect the F Col extension B in the central rail of the bed frame see figure 4 21 on page 4 43 Page 2 42 AvantGuard Electric Bed Service Manual QD1787 2 15 Bed Connected to AC Power Brakes not Applied Detection Malfunction Chapter 2 Troubleshooting Procedures c The same verification applies to the terminals of the 8 pin connector in the central rail of the bed frame by checking the o
55. 2 48 Sth Wheel Steering Malfunction Doro a o 2 49 Chapter 3 Theory of Operation Component Positioning and Electrical Diagrams LIIS8Ax LIIS9Ax 3 3 Theory Of Operauon 6 0 ontwag ated das 3 6 Mechanical System sas ete dest td et Su ald e E ae hl a Buk Ph he 3 7 Retractable Head Section cunda ee eed ea veneer ee eek 3 7 CPR Emergency Release nnn ita te e ca a Ree RS 3 7 Assisted Poot Sections erora renen pe T nevete poet ace soe eyes eed 3 7 Hiedra be eed edad das 3 7 CASI eE ou ja IES Gl e e ES eS 3 8 BAT id eran AS RAS DA RRA DO A Pe Os E To 3 8 TECH voir ies e ta tio ae A 3 8 Sth Wheel Steering System ununun unurnar narar 3 8 TotalGuard Integral Siderail Construction 0 0 00 00 000 3 9 Boot Siderail Supports eien enine A aw Pek eRe GS WS a ed 3 9 Head and Footboards mis arias raean 3 9 Eleccion ed Oe aN fe 3 10 Power Supply Unit Characteristics 0 0 00 cee eee eee 3 10 AvantGuard Electric Bed Service Manual QD 1787 Page vii Table of Contents Power St A a o nem 3 11 Function Control 3 11 B ttery CA a de o a rl a a a ON e he 3 12 Battery Characteristics a bw ha ee oa So ed 3 12 Control Logic and Specific Functions 2 0 0 0 0 0 0 e eee eee eee 3 13 Control Pendant ti bit Rew eae a ae cae eo eee Meee uth See 3 14 Control Panel on a Flexible Arm 2 2 ce eee eee eee 3 14 Lateral Caregiver Units 2 iaa sreci be Se ase SA owe eee ewes 3 14 Mot
56. 5 5 Disconnect the 2 LCU la and 1b grey Either reconnect a new LCU and activate the emergency trendelenburg function Or connect the test control unit P N AG0061 see Appendix on page 2 12 and press the simulation button of the emergency trendelenburg function The emergency trendelenburg function works Yes No y The PSU is defective Replace it refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on AvantGuard Electric Bed Service Manual QD1 787 Page 2 37 2 13 Emergency Trendelenburg Function not Working with Bed Connected to AC Power Chapter 2 Troubleshooting Procedures page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 6 Replace the defective LCU refer to procedure 4 7 on page 4 25 This solves the problem Yes No Contact your national Hill Rom representative see the last page of this manual 7 Go to Final Actions on page 2 12 Page 2 38 AvantGuard Electric Bed Service Manual QD1787 2 14 Emergency Trendelenburg Function not Working with the Bed not Connected to AC Power Chapter 2 Troubleshooting Procedures 2 14 Emergency Trendelenburg Function not Working with the Bed not Connected to AC Power NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4
57. 9 Thigh section actuator Inspect the assembly of the actuator Check that the pins and Truarc circlips ar not missing Carry out a full raise lower cycle of the corresponding section Check that there is no friction or abnormal noises and that no audible overload indication can be heard during the movement Replace in the event of a malfunction refer to procedure 4 10 Automatic Contour Check that the TS Section rises up to mid height when the HS raise function is activated from the low position and that it lowers automatically when the HS lower function is activated and the HS section reaches low position Troubleshoot and replace the actuator or defective unit if necessary see RAP 2 7 Foot section Activate the thigh section raise up to mid height actuator Inspect the assembly of the foot actuator Check that the pins LI159Ax and Truarc circlips are not missing Carry out a full raise lower cycle of the foot section Check that there is no friction or abnormal noises and that no audible overload indication can be heard during the movement Replace in the event of a malfunction or abnormal noises refer to procedure 4 11 EasyChair Ensure that the Chair HS TS and FS functions are not Chair function locked out LI159Ax Activate the chair function and check that the three sections are correctly activated to reach the required position Activate the control exit from chair position
58. 9 Test control unit P N AGO0061 HL column simulation connector P N AG0062 TS and FS actuator simulation connector P N AG0063 HS actuator simulation connector P N AG0064 Loctite 242 adhesive Beige retouching paint bottle P N CO0400 Pot of white petroleum jelly approx 400 g P N CO1131 1 Torx is a registered trademark of Textron Inc 2 Allen is a trademark of Industrial Fasteners Inc 3 Loctite is a registered trademark of Loctite Corporation Inc Page 6 18 AvantGuard Electric Bed Service Manual QD1787 Tools and Supply Requirements Chapter 6 General Procedures Figure 6 4 Elevation Tool P N AG0052 y vw w 1680_600 AvantGuard Electric Bed Service Manual QD1787 Page 6 19 Tools and Supply Requirements Chapter 6 General Procedures NOTES Page 6 20 AvantGuard Electric Bed Service Manual QD1787 Chapter 7 Accessories Chapter Contents ACCESSONIECS sitiada ratico WORE badd ROGERS EE adhe Rhee 7 3 Accessory Support Bar Kit AD126A 0 0 eee eee 7 5 A hye CAPRA ae Be aye ee e Oe ee ae 7 5 Removal os eu es ba ee aes A ee es oe eas 7 6 Medical Device Support Rail Kit ADI27A 0 0 cee eee ee 7 7 HSA ETETE Hoes Grant ed o AN eee ee ee el oe Gt E 7 7 R moval ani es yd AA E SAG eee wha Ae Bw led eae ee 7 8 MobiBar Egress Bar Kit ADI34A 0 0 eee 7 9 Sa ALON aot Sieh eet Re atk acetate Uh een ih ae eta te La
59. AvantGuard Electric Bed and help to reduce downtime due to excessive wear failures An annual service of the bed is advised in order to maintain its characteristics and performance Every 3 years the electrical equipment should be inspected by an approved after sales service technician in order to obtain consistent equipment performance over time The life time of the hilow head section thigh section and foot section actuators is 5 years the power supply unit the control pendant the control panel on a flexible arm and the gas springs is 10 years for average use The actuators should be replaced after these periods AvantGuard Electric Bed Service Manual QD1787 Page 6 5 Preventive Maintenance Chapter 6 General Procedures Preventive Maintenance Schedule The following preventive maintenance schedule is a reminder designed to guide the technician through a normal preventive maintenance procedure on the AvantGuard Electric Bed Check each element of the schedule and carry out all the necessary adjustments during the preventive maintenance procedure The preventive maintenance schedule is intended to be used in conjunction with the preventive maintenance checklist following it This checklist is designed to keep a running history of maintenance and subsequent repair costs for the AvantGuard Electric Bed Keeping close records and maintaining the AvantGuard Electric Bed are two good ways to reduce downtime an
60. AvantGuard Electric Bed Service Manual QD1787 Page 2 29 2 8 Battery Backup Malfunction Chapter 2 Troubleshooting Procedures Yes No y Replace the PSU refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 7 Goto Final Actions on page 2 12 Page 2 30 AvantGuard Electric Bed Service Manual QD1787 2 9 Foot Section Malfunction Chapter 2 Troubleshooting Procedures 2 9 Foot Section Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions the TS actuator movements are working retraction and extension and the actuator rod is in extended position 1 At least one of the other functions is working Yes No y Goto RAP 2 1 on page 2 13 2 Perform the following to verify the proper operation of the remote control a Disconnect the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AGO0061 see Appendix on page 2 12 3 The functions are working Yes No Go to step 5 4 Replace the CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexible arm refer to procedure 4 3 on page 4 12 according to the option 5
61. C Power Cable soci s inca a PE a he bee ee ds kes 4 15 Removal siria edi ae pleas a o Re eae ages 4 15 Replacement of seven it tae cea ao Mena ER 4 16 Power SUPPLY Unit coords A 4 18 REMOVAL a SS aie EE shan 4 18 Replacement is A NS E ed ees DA 4 20 Battery Backup and Emergency Trendelenburg Function 4 22 Removals Aere ra fee hae eee see herein dee Legend eke 4 22 REPLICA nee A he Ba eae a Se 4 23 Lateral Caregiver Unit Vi A e went 4 25 Removal ocak ody mari de dad he wee ae cee bd eee 4 25 Replacements fen atte al el Sam ail Om ot Sle le la ot a 4 26 Head Section Actuator comia oie he hes eee igh Leia be ner ees 4 27 Removal seriden piro bata beige MATA e pa Rees 4 27 Replacement gt fcecuueyavse ee Sie A E 4 29 Head Section Release Cable oasis td nee add 4 30 AvantGuard Electric Bed Service Manual QD1787 Page 4 1 Chapter 4 Removal Replacement and Adjustment Procedures REMOVAL A e Cee RR ee saw ee eRe 4 30 REPIMCEMENE te xx outed e Ew SEES ae UW RIES woes CO EAE OOM Se 4 31 A 4 31 Thigh Section Act tor 2252 ane a o ad a A dl era Eso Ose 4 33 Removal tai A pe eed ee ee eee hs A DEEA A 4 33 Replacement eii ida a a Che TAGGe REA ERS 4 34 Foot Section Actuator Replacement LIIS9Ax only 0 0 4 35 Removal 22 bet pal od wets da 4 35 Replacement ss bas was ee Se at 4 36 Head Hilow Column 09 o TAS pl oe Be AA A ee ee 4 37 Removal ti caniteialebiveeabe
62. Chapter 1 Introduction Purpose This manual provides requirements for the AvantGuard Electric Bed normal operation and maintenance It also includes a parts list in chapter 5 for ordering replacement components Audience This service manual is to be used only by facility authorised personnel only Failure to observe this restriction can result in severe injury to people and serious damage to equipment Organisation This manual contains seven chapters Chapter 1 Introduction In addition to a brief description of this service manual chapter 1 also provides a product overview Chapter 2 Troubleshooting Procedures Repair analysis procedures are contained in this chapter Use these procedures to gather information identify the maintenance need and verify the effectiveness of the repair Chapter 3 Theory of Operation This chapter describes the application of the mechanical and electrical systems employed in this product Chapter 4 Removal Replacement and Adjustment Procedures Chapter 4 contains the detailed maintenance procedures determined necessary in chapter 2 Chapter 5 Parts List This chapter contains the warranty part ordering procedure and illustrated parts lists Chapter 6 General Procedures Cleaning preventive maintenance and other general procedures are described in this chapter AvantGuard Electric Bed Service Manual QD1787 Page 1 3 Organisation Chapter 1 Intro
63. Corporation MobiBar is a trademark of Hill Rom Services Inc Primo is a trademark of Hill Rom Services Inc Torx is a registered trademark of Textron Inc TotalGuard is a trademark of Hill Rom Services Inc Truarc is a registered trademark of Waldes Truarc Inc Page ii AvantGuard Electric Bed Service Manual QD1787 Revisions The information contained in this manual is subject to change without notice Hill Rom makes no commitment to update or keep current the information contained in this manual The only product warranty intended by Hill Rom is the express written warranty accompanying the bill of sale to the original purchaser Hill Rom makes no other watranty express or implied and in particular makes no warranty of merchantability or fitness for a particular purpose Additional copies of this manual can be obtained from Hill Rom To order additional copies of this manual call your national Hill Rom representative see back cover and place a parts order for part number QD1787 IMPORTANT Refer to the last page of this manual for the address of your National Hill Rom representative AvantGuard Electric Bed Service Manual QD1787 Page iii Revisions NOTES Page iv AvantGuard Electric Bed Service Manual QD1787 Table of Contents Chapter 1 Introduction PURPOSE 2 oan BSS ole iin co TS e Swe ae ee oe ke h t 1 3 A sadn 2s ca ce Cae eB eed pk ae BN se a cee ph
64. D IS be aa Pe ek a 1 22 AvantGuard Electric Bed Service Manual QD1787 Page v Table of Contents PICCLING phe wo Oh yes Gait iG gisele e tw Sane Chua wretch Sg Gate 1 22 Headboards and Footboards 0 0 0 0 cece ee eee 1 23 Hard Surfaces on the Head Seat Thigh and Foot Sections 1 23 CASI cia yy Peed A oh ieee ee A eked 1 23 Restraining Strap Attachment Points 0 0000 aee 1 23 Protective BUIMPCts se i44 54a ee kt ae een Ae ee ee ee 1 23 EN Pole SOBRA al 1 24 Optional ESAS ena ee asa A SRA owe et tows 1 24 Specifica ONS e n ni AA EPA EA DA E 1 26 Physical Description 2i c45i ee cla Rese PA gem ee es 1 26 Electrical Description 4 4 s a a 1 27 Power Supply Unitins nas See dae A eed oe 1 28 Control Unites ee vigis Sarees we adie a a SAA E weeks 1 28 Motor Driven Actuators rd 1 28 Regulations Standards and Codes 0 cece eee eee 1 29 Model Identifications ss 4 4 3 55 saaena yea ve Gat ee ieee bi we eres 1 30 Safety TIPS betis ler idee at e taba bees 1 30 Identification and Caution Labels estar 1 35 Chapter 2 Troubleshooting Procedures G tung Started ee E eee ste ee ee E E a Boe Baal 2 3 Tital ACUONS prais eria ds nye that oa ar 2 4 Rapid Problem Solution Identification Table 00055 2 4 Function Checks 1300 Saw E need DA E AS AD ees NR 2 6 FinalACuons 22iceuhtin dee rs Da id AA 2 12 Appendix qs ob O 2 12 Nov Function
65. E a 6 3 Lubrication Requirements vxi diran hig andi Saeed ee bee eked 6 4 Preventive Maintenance ropita rra a ta NS A atea 6 5 Preventive Maintenance Schedule ocio ee ee tees 6 6 Preventive Maintenance Check listy 21 0orei34 4 02058 e8s way peewee yes 6 16 Tools and Supply Requirements us ec ave seeks aver aaeg Sow da 6 18 AvantGuard Electric Bed Service Manual QD1787 Page 6 1 Chapter 6 General Procedures NOTES Page 6 2 AvantGuard Electric Bed Service Manual QD1787 Cleaning and Care Chapter 6 General Procedures Cleaning and Care e SHOCK HAZARD Unplug the bed before any cleaning or maintenance Failure to do so could result in personal injury or equipment damage Refer to the AvantGuard Electric Bed User Manual and to the specific chapters of this manual for any additional cautions and warnings General Cleaning We recommend that you clean the bed with detergent and warm water Do not use excessive liquid or harsh cleansers Steam Cleaning Do not use any steam cleaning device on the AvantGuard Electric Bed The excessive moisture can damage mechanisms and components in the bed Hard to Clean Spots To remove difficult spots or stains we recommend that you use standard household cleansers and a soft bristle brush To loosen heavy dried on soil you may first need to saturate the spot Disinfection Dilute disinfectants and or germicides as specified on the manufa
66. Egress Bar kit includes the following e Two MobiBar Egress Bars one right hand and one left hand e Four securing bolts e Four washers Figure 7 5 MobiBar Egress Bar Kit 1680_743 Identification see figure 3 1 on page 3 3 Time required 10 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench Installation 1 Apply the brakes adjust the bed to high position the head thigh and foot sections horizontal Lift the foot siderails AvantGuard Electric Bed Service Manual QD1 787 Page 7 9 7 3 MobiBar Egress Bar Kit AD134A Chapter 7 Accessories e SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control unit If this is not the case wait for about one minute until the battery mode is deactivated Figure 7 6 Installing the MobiBar Egress Bar 1680_747 3 Remove the hard surface of the foot section refer to procedure 4 1 4 Position one of the egress bars A at the location provided on the bed frame see figure 7 6 on page 7 10 5 Insert the two corresponding securing bolts B and washers C and lock these bolts by using the 13 mm ratchet wrench 6 Repeat steps 4 and 5 to install the o
67. Hill Rom representative to the attention of technical support refer to the last page of this manual AvantGuard Electric Bed Service Manual QD1787 Page 5 5 Exchange Policy Chapter 5 Parts List Exchange Policy The conditions specified below apply to any exchange carried out by Hill Rom during the in warranty period and out of warranty period In Warranty Exchanges Hill Rom may request that certain defective products or parts be returned for assessment You will then be held liable for responding with the 30 days following reception of the spare parts or products By default Hill Rom will invoice the full price of these parts or products NOTE This invoicing principle applies exclusively to the parts or products for which Hill Rom has requested the return In certain cases the invoice sent with the parts will show the full selling price of the parts This is for Hill Rom s internal use only and should not be confused with the price you will actually pay Out of Warranty Exchanges Not applicable Page 5 6 AvantGuard Electric Bed Service Manual QD1787 Recommended Spare Parts Chapter 5 Parts List Recommended Spare Parts For a recommended spare parts list to service 10 AvantGuard Beds for 3 years see table 5 1 on page 5 4 Table 5 1 Recommended Spare Parts Reference Quantity Description AG0013 1 Control
68. K ati de an CN A eR tada e a e AS 2 12 No Functions Works cuero rar a aa ta ai a eked 2 13 Control Unit Malfunction 2 02404 e e taa 2 16 Function Management Mode Malfunction o ooooooocoocoocoooro romo 2 18 Hilo Maluacuol dar ir a 2 20 Thigh Section Malic Onc rs poe e AA i 2 22 Head Section Malfunction excluding Automatic Contour 2 24 Automatic Contour Malfunction 0 0 0 eee eee eee 2 26 Battery Backup Malfunction 0 0 0 0 cece cece ee eens 2 29 Foot Section Malfunction 0 ekts9 dee eee dae pha Me Pee eee eee 2 31 Assisted Foot Section Malfunction 0 0 0 eee eee eee ee eee 2 34 EasyChair Automatic Chair Position Malfunction o oooo ooooooooo 2 35 Trendelenburg Reverse Trendelenburg Malfunction 04 2 36 Emergency Trendelenburg Function not Working with Bed Connected to AC Power vi A ones ae SE 2 37 Emergency Trendelenburg Function not Working with the Bed not Connected to AC Power won aed dager rad dev ad eee 2 39 Bed Connected to AC Power Brakes not Applied Detection Malfunction 2 42 CPR Malf nction AA ee Bekins a RM eta hee 2 44 Foot Siderail Locking Unlocking Malfunction 0 0 0 0 00000 2 45 Head Siderail Locking Unlocking Malfunction 000000 2 46 Braking Walhuncu On 0 E DS AE Lee DAA AA PONS 2 47 AvantGuard Electric Bed Service Manual QD1787 Page 2 1 Chapter 2 Troubleshooting Procedures Ste
69. LED and non low position HL LED flash during the sequence and remain lit constantly at the end of the sequence Yes No Go to RAP 2 3 on page 2 18 Press the raise control of the HL function then the left and right raise pedal controls Page 2 6 AvantGuard Electric Bed Service Manual QD1787 Function Checks Chapter 2 Troubleshooting Procedures The bed rises to the high position without stopping or the audible alarm sounding during the movement Yes No Go to RAP 2 4 on page 2 20 8 Press the lower control of the HL function then the left and right lower pedal controls The bed goes down to the low position without stopping or the audible alarm sounding during the movement In addition the bed in low position LED goes off when the bed reaches the maximum low position Yes No Go to RAP 2 4 on page 2 20 9 Press the raise button of the TS function The TS rises to the high position without stopping or the audible alarm sounding during the movement Yes No y Go to RAP 2 5 on page 2 22 10 Press the lower button of the TS function The TS goes down to the low position without stopping or the audible alarm sounding during the movement Yes No y Go to RAP 2 5 on page 2 22 11 Press the raise button of the HS function only note that there are two raise buttons for this function The HS rises to the high position without stopping or the audible alarm sounding during th
70. ND IMPLIED WARRANTIES INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS OF PURPOSE HILL ROM S OBLIGATION UNDER THESE WARRANTIES SHALL NOT INCLUDE ANY LIABILITY FOR LOSS OF PROFITS DIRECT INDIRECT OR CONSEQUENTIAL DAMAGES OR DELAYS Some states do not allow the exclusion or limitation of incidental or consequential damages so the above exclusion or limitation may not apply Any improper or negligent use any alterations or repairs not in accordance with Hill Rom s manuals or performed by others in such manner as in Hill Rom s judgement affects the product materially and adversely shall void these warranties These warranties do not cover material breakdown resulting from improper or abusive use negligence or an inadequate maintenance routine No employee or representative of Hill Rom is authorized to change these warranties in any way or grant any other warranty without the written and signed agreement of a member of the management These warranties provide specific legal rights however there may be other available rights which vary from state to state Revised 21 January 2001 Hill Rom B P 14 Z L du Talhou t 56330 Pluvigner France AvantGuard Electric Bed Service Manual QD1787 Page 5 3 Ordering Service Parts Chapter 5 Parts List Ordering Service Parts Use the parts list in this service manual to identify the part numbers you require Next note the product number and se
71. Power Stage The power stage is composed of a 230VAC transformer protected at the input by 2 antisurge fuses of 2A followed at the output by a bridge rectifier and filter capacitor The output voltage VDC1 is is approximately 24VDC and unregulated and depending on the power required The transformer is also protected by an internal thermal cutout that cannot be reset The available power enables several functions to be controlled simultaneously The VDC1 also passes through two regulators that supply voltages VDC2 15VDC 3V for the analogue components comparators battery charging control circuit etc and VDC3 3 3VDC 10 for the logic components control pendant etc The power stage is also protected against electrostatic discharges ESD of 20kV Function Control The actuators are controlled by relays In order to protect these relays from electrical sparks during switching the second FET transistor checks the delay power when a function is activated thus allowing the switches to close with no voltage In the same way when a function control is released the FET transistor cuts the power before the contact opens Overload limiting This is provided by permanently controlling the current at the output of each motor In the event of overload or mechanical blockage the supply of the relevant motor is cut off by a control transistor without affecting the operation of the other motors except for specific situations In thi
72. RTH CONTINUITY DAUERERDSCHLUSS Ohms COURANT DE FUITE A LA TERRE 0 121 EARTH LEAKAGE CURRENT ERDABLEITSTROM mA 56330 PLUVIGNER FRANCE Chapter 1 Introduction 1787_106 1680_107 1680_108 1787_114 1680_118 AvantGuard Electric Bed Service Manual QD1787 Page 1 35 Identification and Caution Labels Chapter 1 Introduction Safety batteries have to be mounted er QD1642 Batterie de s curit iy presence obligatoire Hilfsakkus Mussen vorhangen sein 001733 E gt lt A CAUTION TO PREVENT ELECTRIC SHOCK DO NOT REMOVE COVER ATTENTION AFIN D EVITER TOUT RISQUE D ELECTROCUTION NE PAS ENLEVER LE COUVERCLE ACHTUNG UM EINEN ELEKTRISCHEN SCHLAG ZU VERMEIDEN BITTE DEN DECKEL NICHT ENTFERNEN THIS BOX CONTAINS LEAD ACID BATTERIES CHECK BEFORE DISPOSAL FOR COMPLIANCE WITH THE LOCAL ENVIRONMENTAL POLICY CE BOITIER CONTIENT DES BATTERIES AU PLOMB NE PAS JETER DIESES GEHAUSER ENTH LT BATTERIEN MIT BLEI NICHT IN DEN HAUSM LL 1787_110 Page 1 36 AvantGuard Electric Bed Service Manual QD1787 Chapter 2 Troubleshooting Procedures Chapter Contents Getting Staricd visas pista 225805 bade bate led EGER pate Koes 2 3 Initial ACOSTA 2 4 Rapid Problem Solution Identification Table ooooo oooooo momo 2 4 F ncton OA a A AS Ea 2 6 A A A Oe 2 12 PAPE I
73. Service Manual AvantGuard Electric Bed Hill Rom A HILLENBRAND INDUSTRY QD1787 AvantGuard Electric Bed Service Manual Revisions Revision Number Original Issue Pages Affected Date August 2002 QD1787 AvantGuard Electric Bed Service Manual QD1 787 Page i Revisions Manufacturer HILL ROM Z I du Talhou t B P 14 56330 PLUVIGNER FRANCE 2002 by Hill Rom Services Inc ALL RIGHTS RESERVED No part of this text shall be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying recording or by any information or retrieval system without written permission from Hill Rom Services Inc Hill Rom The information in this manual is confidential and may not be disclosed to third parties without the prior written consent of Hill Rom Services Inc Hill Rom First Edition First Printed 2002 Printed in Europe Hill Rom is a registered trademark of Hill Rom Services Inc Allen is a trademark of Industrial Fasteners Inc AvantGuard is a trademark of Hill Rom Services Inc Climat System is a trademark of Hill Rom Services Inc Cliniplot is a registered trademark of Hill Rom Services Inc Clinisert 2 is a trademark of Hill Rom Services Inc Duo 2 is a trademark of Hill Rom Services Inc EasyChair is a trademark of Hill Rom Services Inc Loctite is a registered trademark of Loctite
74. a flexible arm b Connect the test CU P N AGO0061 see Appendix on page 2 12 c Press the function management control on the LCU All the LEDs flash Yes No gt Go to step 6 Page 2 18 AvantGuard Electric Bed Service Manual QD1787 2 3 Function Management Mode Malfunction Chapter 2 Troubleshooting Procedures 5 Replace the Control Unit CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexible arm refer to procedure 4 3 on page 4 12 according to the option fitted 6 Perform the following to test for a short circuit on the LCU Y a Remove the cover of the PSU using the 13 mm combination spanner see figure 4 9 on page 4 19 b Disconnect the LCU Y c Press the function management control on the LCU All the LEDs flash Yes No y Go to step 8 7 Replace LCU Y refer to procedure 4 7 on page 4 25 8 Perform the following to test for a short circuit test on the LCU X a Connect the LCU Y b Disconnect the LCU X c Connect CU to LCU Y d Press the function management control on the LCU All the LEDs flash Yes No a Replace the PSU refer to procedure 4 4 on page 4 15 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 9 Replace the LCU X This solves the problem Yes No y Contact your nati
75. abet idas 5 10 DICED lacas E a e sl Gee eee 5 12 Siderail Assemblies i252 h4 634005 S246 pe b8 i Shee ee be ae Pd eke ed 5 14 Miscellaneous Items not Illustrated 0 0 0 cece eee 5 15 Chapter 6 General Procedures Cleaning and Car csi at eta ti a add 6 3 General Clannad 6 3 Steam Cleaning lidad 6 3 Hard t Clean Spots 34s aww tees od big A ia 6 3 A 8 s ar o iae aa A 6 3 Storing the B d ecco crcoci sterge beta a tae add 6 3 Lubrication Requirements 0 0 0 ccc ee eee eee 6 4 Preventive Maintenance iii ba oie Sag Baan ye SoMa a ayes 6 5 Preventive Maintenance Schedule aii bes sai e were dayueess 6 6 Preventive Maintenance Checklist 02 0 eee eee eee ee eee 6 16 Tools and Supply Requirements dis 6 18 AvantGuard Electric Bed Service Manual QD1787 Page xi Table of Contents Chapter 7 Accessories PRC COSSOTIOS ex sois tipa 7 3 Accessory Support Bar Kit AD126A 0 eee eee eee 7 5 Installation A sks ke ene bE AAG aE ease EW na eee ew eal 7 5 Removal isser ion aria e a AER ges rata EA EE A 7 6 Medical Device Support Rail Kit ADI27A 0 2 eee eee ee 7 7 Insta AO 34 044 srt a tow hey ee hho tale eae 17 17 Removal da EE be ete 7 8 MobiBar Egress Bar Kit ADI34A 0 0 eee 7 9 Installation pads sagdu de a Bene ed pen be da AA eed oe 7 9 R mov l n hereke ade li A E A A eb eres 7 10 Control Panel on a Flexible Arm Kit AD136A ooooocoocoomoo
76. ables the catch to slide within its support and is designed to prevent a sudden unintended release e Each foot siderail is secured to the corresponding head siderail by a system of clips made of injection moulded plastic that enable rotation Attachment and removal of the foot siderail is only possible when the siderail is folded onto the head siderail The metal supports consist of three articulated arms one central arm and two lateral arms The siderails and supports thus form a single unit Foot Siderail Supports The foot siderail supports are retractable mechanisms integrating a catch for locking and unlocking the foot siderail Head and Footboards The head and footboards feature an extruded blow moulded plastic structure with chrome plated steel clasps for assembly and decorative laminated panels AvantGuard Electric Bed Service Manual QD1787 Page 3 9 Electrical System Chapter 3 Theory of Operation Electrical System The specifications of the bed are as follows Table 3 1 Specifications of the LI158Ax LI159Ax Bed Description Specification Type AvantGuard Electric Bed Reference LI158Ax LI159Ax Type of protection against electric shock Class I according to IEC 60601 1 Degree of protection against electric shock Type B according to IEC 60601 1 Protection against harmful ingress of water IPX4 according to IEC 60529 Intermittent operation 10 3 min 30 min
77. acement and Adjustment Procedures 4 For the TS hard surface grip the hard surface B on either side of the sleep surface and remove it from the bed 5 For the SS hard surface grip the hard surface C on either side of the sleep surface and remove it from the bed A CAUTION The head section must always be horizontal when removing or fitting the head section hard surface Failure to do so may give rise to an incorrect position and cause the elevation sliding system to malfunction 6 Perform the following to remove the HS hard surface NOTE If the bed fitted with a control panel on a flexible arm it must be removed first refer to procedure 4 3 a Ensure that the SS has been removed then remove the headboard and lower the HS to horizontal by using one of the CPR controls b Remove the 4 Truarc circlips on pins E and F securing the hard surface D to the HS and recover the spacers on the two lower Truarc circlips photo shown with HS section lifted for clarity see figure 4 2 on page 4 6 Figure 4 2 Installing the Head Section Hard Surface 1680_442 c Pass your hand through the openings G of the hard surface and remove the 2 securing pins F see figure 4 3 on page 4 7 NOTE The 2 pins are mounted from the inside to the outside but they can be reversed 1 Truarc is a registered trademark of Waldes Truarc Inc Page 4 6 AvantGuard Electric Bed Service Manual QD1787 4 1 Bed Fra
78. actuator switch is closed a Either replace the TS actuator with a new actuator in retracted position or simulate the TS actuator with the TS and FS actuator simulation connector P N AG0063 the switch simulating the actuator lt mid travel connected to the brown socket 5 b Press the HS raise with automatic contour button A see figure 1 11 on page 1 14 If new TS actuator the TS section rises If TS actuator simulation connector the raise LED lights up Yes No J Go to step 10 9 Replace the TS actuator refer to procedure 4 10 on page 4 33 10 Perform the following to verify that the PSU and TS actuator switch is open a Either connect a new actuator with rod in extended position or a switch simulating actuator to mid travel position b Press the HS raise with automatic contour button A If new TS actuator The TS section rises If TS and FS actuator simulation connector the raise LED lights up Yes No 4 Replace the TS actuator refer to procedure 4 10 on page 4 33 11 Replace the PSU refer to procedure 4 5 on page 4 18 12 Perform the following to troubleshoot without a new actuator or simulation tool possible only if power driven FS option Press the HS raise button B until this section reaches the high position b Press the TS and FS lower buttons F until these sections reach the low position c Press the HS raise with automatic contour button A AvantGuard Elect
79. age 3 3 Time required 20 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required 13 mm ratchet wrench Removal 1 Apply the brakes raise the sleep surface to the high position the sleep surface sections to horizontal 2 Raise and lock the head siderails with the foot siderails folded W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 4 While supporting the siderail press the release push button A and remove the foot siderail see figure 4 36 on page 4 71 5 Using the 13 mm ratchet wrench loosen and remove the four screws B securing the head siderail support and recover the spacers C see figure 4 35 on page 4 69 NOTE To make removal of the siderail support more simple it is assembled so as to rest on the HS structure even when the screws are removed Page 4 68 AvantGuard Electric Bed Service Manual QD1787 4 21 Head Siderail Including Elevation Arm Subassembly Chapter 4 Re
80. aise and lock the siderails Apply the brakes and raise the sleep surface to the high position Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this 1s not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated Identify connection A of the control pendant on the bed see figure 4 6 on page 4 11 NOTE This connection can be found both on the right or left of the sleep surface behind the LCU Page 4 10 AvantGuard Electric Bed Service Manual QD1 787 4 2 Control Pendant Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 6 Control Pendant Connection a 1680_425 6 Disconnect the connector A from the LCU by pulling no specific locking The cable does not pass through the conduit located under the crossbar B see figure 4 6 on page 4 11 Replacement 1 Identify the version of the control pendant to be replaced LI158Ax or LI159Ax to ensure that the replacement will allow the same functions 2 Perform the removal procedure in reverse order 3 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 11 4 3 Control Panel on a Flexible Arm Chapter 4 Removal R
81. al A see figure 6 1 e Point 2 Head side upper column screw B see figure 6 1 e Point 3 amp 4 Right and left hand siderail internal pins C see figure 6 2 e Point 5 Right or left hand foot accessory support screw D see figure 6 3 Use a Portable Appliance Tester PAT or a Digital MultiMeter DMM for this check This value must be less than 0 2 ohms Figure 6 1 Equipotential Terminal and Column Screw 1680_603 AvantGuard Electric Bed Service Manual QD1787 Page 6 7 Preventive Maintenan ce Chapter 6 General Procedures Function Leakage current Procedure Unplug the power connector from the power outlet Connect the power connector of the bed to the test device and connect this device to the power outlet Check that the bed is correctly supplied by the test device control unit LEDs lit unlock the functions if necessary Take leakage current measures but without operating any function Use a Portable Appliance Tester PAT or a Digital MultiMeter DMM for this check The value must be less than 500 uA Check the AC power cable and power supply unit if the value is outside of the specifications Replace the AC power cable or power supply unit if necessary Connect to an appropriate power source following of this test unit Lateral caregiver Check the condition of the labels Replace if necessary Check the condition of the female connector behind LCU Che
82. ase wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Remove the hard surfaces of the TS and the FS see procedure 4 1 on page 4 5 4 Position a support under the foot section of the bed frame so as to block it at the required height 5 Using the T25 Torx key loosen and remove the two screws A securing the control lever subassembly B see figure 4 33 on page 4 66 1 Torx is a registered trademark of Textron Inc AvantGuard Electric Bed Service Manual QD1787 Page 4 65 4 20 Gas Spring Subassembly with Cable and Foot Section Sleeve LI158Ax only Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 33 Assisted Foot Section Control 1680_419 6 Remove the Truarc circlips from the pin D at the tip of the gas spring then remove the pin while supporting the foot section of the bed frame and recovering the two spacers E see figure 4 34 on page 4 66 Figure 4 34 Assisted Foot Section Gas Spring 1680_428 7 Remove the Truarc circlips from the pin F at the base of the gas spring then remove the pin 1 Truarc is a registered trademark of Waldes Truarc Inc Page 4 66 AvantGuard Electric Bed Service Manual QD1787 4 20 Gas Spring Subassembly with Cable and Foot Section Sleeve LI158Ax only Chapter 4 Remova
83. ave to be lifted upwards to remove them from their sockets Replacement Install the boards by inserting the two chrome plated guides of each board into the sockets provided on the bed frame Page 4 74 AvantGuard Electric Bed Service Manual QD1787 4 25 Linen Holder Table optional Chapter 4 Removal Replacement and Adjustment Procedures 4 25 Linen Holder Table optional Identification see figure 3 1 on page 3 3 Time required 5 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required None Removal 1 Apply the brakes raise the sleep surface to the high position and the TS and the FS to horizontal 2 Remove the footboard 3 Pull out the linen holder then tilt it upwards to remove it from the two guides and pull it towards you to remove it from the bed Replacement 1 Perform the removal procedure in reverse order 2 Ensure that the board slides correctly from its stored to extended position AvantGuard Electric Bed Service Manual QD1787 Page 4 75 4 26 Extension optional Chapter 4 Removal Replacement and Adjustment Procedures 4 26 Extension optional Identification see figure 3 1 on page 3 3 Time required 10 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench White petroleum jelly Removal 1 Apply the brakes raise the sleep surface to the high position the TS
84. ays and check that the articulated arm supporting the 5th wheel unlocks The articulated arm unlocks correctly Yes No y Go to step 7 5 The problem is solved Yes No 4 Contact your national Hill Rom representative see the last page of this manual 6 Go to Final Actions on page 2 12 7 Replace the support and caster subassembly refer to procedure 4 17 on page 4 56 The problem is solved AvantGuard Electric Bed Service Manual QD1787 Page 2 49 2 21 5th Wheel Steering Malfunction Chapter 2 Troubleshooting Procedures Yes No y Contact your national Hill Rom representative see the last page of this manual 8 Goto Final Actions on page 2 12 Page 2 50 AvantGuard Electric Bed Service Manual QD1787 Chapter 3 Theory of Operation Chapter Contents Component Positioning and Electrical Diagrams LIIS58Ax LIIS9Ax 3 3 Theory of Operations at fh ave dees Sees ese is eet eee ee et wee eee 3 6 Mechanical System sra veal tas ee ae es 3 7 Retractable Hedd Secos is a eta Boeke hes 3 7 CPR Emergency Release fs fie ei Ort ba oe eae ee a eS 3 7 Assisted FOOL SECHOM eik a BON Stee o a a ol IO 3 7 HilOW sige cha a eed PAA ee ae A bed ek 3 7 CASLOIS 24 aa A Banke a ees 3 8 Brake ie E O A A ee 3 8 Steering a goda A II eet eerie the hoses 3 8 Sth Wheel SSA yew sree aoa vow es teeoss 3 8 TotalGuard Integral Siderail Constructi0N o oooooocoooooomoooo
85. ble arm refer to procedure 4 3 on page 4 12 according to the options fitted If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 6 Perform the following to test for a short circuit on the LCU Y circuit Page 2 16 AvantGuard Electric Bed Service Manual QD1787 2 2 Control Unit Malfunction Chapter 2 Troubleshooting Procedures a Remove the cover of the PSU using the 13 mm combination spanner see figure 4 8 on page 4 16 b Disconnect the LCU Y identify the cable by tracing it from the LCU to the connector Ensure the connector is unlocked before trying to free it The functions of the CU are working Yes No y Go to step 8 7 Replace the LCU Y refer to procedure 4 7 on page 4 25 8 Perform the following to test for a short circuit on the LCU X circuit a Reconnect the LCU Y b Disconnect the LCU X c Connect the CU to the LCU Y The functions of the CU are working Yes No Replace the PSU refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 9 Replace the LCU X This solves the problem Yes No Contact your national Hill Rom representative see the last page of this manual 10 Go to Final Actions
86. c is a registered trademark of Waldes Truarc Inc AvantGuard Electric Bed Service Manual QD1787 Page 4 29 4 9 Head Section Release Cable Chapter 4 Removal Replacement and Adjustment Procedures 4 9 Head Section Release Cable Identification see figure 3 1 on page 3 3 NOTE The HS release cable is a Y cable common to the two CPR handles The two uncovered branches are however of different lengths Time required Time required 15 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 10 mm combination spanners 2 required Small screwdriver Removal 1 Perform steps 8 to 11 of the HS actuator replacement refer to procedure 4 8 2 Itis important to note that each side of the CPR cable H runs behind the pivot point I of the bed frame crossbar see figure 4 15 on page 4 30 3 Comment on the cable position The end of the cable arriving at the pivot point I passes behind this point then continues to the left The end that arrives at the left of the pivot point I also passes behind this point then continues to the right Figure 4 15 CPR Handles and Cable JJ ll 1680_439 4 With the two 10 mm combination spanners loosen the adjuster E Page 4 30 AvantGuard Electric Bed Service Manual QD1 787 4 9 Head Section Release Cable Chapter 4 Removal Replacement and Adjustment Procedures 5 Pull on the CPR cable sleeving to
87. ce Manual QD1787 Page 2 39 2 14 Emergency Trendelenburg Function not Working with the Bed not Connected to AC Power Chapter 2 Troubleshooting Procedures 5 6 10 11 Go to Final Actions on page 2 12 Replace the battery pack of the emergency trendelenburg function with a new one refer to procedure 4 6 on page 4 22 The emergency trendelenburg function works Yes No The PSU is defective Replace it refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual Go to Final Actions on page 2 12 Perform the following to verify the operation of the emergency trendelenburg function WITHOUT battery backup a Remove the cover of the PSU using the 13mm combination spanner see figure 4 8 on page 4 16 b Check the connections of the battery packs of the emergency trendelenburg function G see figure 4 10 on page 4 19 The emergency trendelenburg batteries are properly connected see figure 4 10 on page 4 19 Yes No Connect the battery packs properly and perform the Function Checks on page 2 6 Perform the following to eliminate a battery charging problem a Plug the bed into an appropriate power source and ensure that it remains connected for at least 12 hours without using the batteries Unplug the bed afterwards b Operate the em
88. cedures 4 1 Bed Frame Hard Surface Replacement Time required 10 min Number of people 1 plus 1 occasionally Possibility of intervening with patient on the bed No Tools required 2 phillips head screwdriver control panel on a flexible arm only Removal 1 Apply the brakes raise the sleep surface and the HS to the high position lower the TS and the FS to horizontal Raise and lock the siderails with the foot siderails folded W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 For the FS hard surface remove the footboard from the bed then grip the foot hard surface on either side of the sleep surface A at the restraining strap attachment handles then remove the surface by tilting it towards the foot end of the bed see figure 4 1 on page 4 5 Figure 4 1 Bed Frame Hard Surface Overall View 1680_421 AvantGuard Electric Bed Service Manual QD1787 Page 4 5 4 1 Bed Frame Hard Surface Replacement Chapter 4 Removal Repl
89. ck that no object is in the openings for the two control units Activate the function lockout sequence and check that the LEDs flash on the control pendant Exit the lockout sequence and ensure that the LEDs do not flash any longer If emergency trendelenburg function is fitted optional operate the emergency trendelenburg function and ensure that the bed changes to trendelenburg correctly and that the head section lowers Make sure that the movement stops when the LCU key is released If the bed is fitted with Battery Backup option unplug the power connector from the power outlet and carry out cycles on different sections The Low battery LED must light up after a certain time Note that the number of cycles depends on the charged condition of the battery Carry out these operations on both LCU Replace if necessary or troubleshoot in the event of a doubt see RAP 2 2 RAP 2 13 or RAP 2 8 Page 6 8 AvantGuard Electric Bed Service Manual QD1787 Preventive Maintenance Chapter 6 General Procedures Function Control pendant Procedure Disconnect the control pendant of the lateral caregiver unit and check the condition of its connector Then reconnect or replace the control unit if necessary Activate the sequence of function selection lockout trendelenburg reverse trendelenburg access and lock out all the functions Ensure that the LEDs are then unlit except for the trendelenburg re
90. cover A Note that it is secured by clips at each end see figure 4 25 on page 4 51 4 Remove the cable cover F situated on the column next to the pedal to be removed see figure 4 22 on page 4 44 5 Using the small screwdriver open the two cable tracks situated under the frame between the pedal to be removed and the F Col 6 Remove the pedal cable from the cable tracks as well as the cable support situated at the level of the foot crossbars of the frame 7 Disconnect the cable from the pedal at the base of the F Col 8 Using the 13 mm ratchet wrench partly loosen the bolt securing the pedal to the frame then slide the pedal to remove it from the frame Page 4 48 AvantGuard Electric Bed Service Manual QD1787 4 14 Hilow Pedal Chapter 4 Removal Replacement and Adjustment Procedures Replacement 1 Perform the removal procedure in reverse order 2 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 49 4 15 Replacement of Microswitch Assembly for Bed Braking Detection Chapter 4 Removal Replacement and Adjustment Procedures 4 15 Replacement of Microswitch Assembly for Bed Braking Detection Identification see figure 3 1 on page 3 3 Time required 15 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required Small screwdriver 13 mm ratchet wrench Removal 1 Apply the brakes Raise and
91. cturer s label Storing the Bed For prolonged storage the bed must be in the low position covered damp proofed and at a temperature of between 0 and 40 C Do not use the bed as a support for working or storage during the storage period For a bed fitted with a battery during storage it is recommended to recharge the battery every three months After prolonged storage it is also advisable to plug the bed into the power outlet for 10 hours to fully recharge the battery before using the bed Always let the bed reach room temperature before using the bed after prolonged storage AvantGuard Electric Bed Service Manual QD1787 Page 6 3 Lubrication Requirements Chapter 6 General Procedures Lubrication Requirements Certain parts of the AvantGuard Electric Bed must be lubricated at least once a year for optimum performance Consult the preventive maintenance schedule Page 6 4 AvantGuard Electric Bed Service Manual QD1787 Preventive Maintenance Chapter 6 General Procedures Preventive Maintenance A WARNING Only the facility authorised maintenance personnel should perform preventive maintenance on the on the AvantGuard Electric Bed Preventive maintenance performed by unauthorised personnel could result in personal injury or equipment damage The AvantGuard Electric Bed should be subject to an efficient maintenance programme This will ensure the long life and productivity of the
92. d at the same time keep the nursing staff happy and efficient Table 6 1 Preventive Maintenance Schedule Function Procedure Overall condition Check the overall aspect of the bed frame and welded Check that the structure and welded assemblies are in good assemblies working condition and that there are no impacts or corrosion Carry out any necessary repairs or paint retouches see Tools and Supply Requirements on page 6 18 replace if necessary Observe the symmetry of the bed and check that the bed frame and base are not twisted Carry out any necessary repairs or paint retouches see Tools and Supply Requirements on page 6 18 Check that all labels are fitted replace them if necessary references see chapter 5 Head and Check the appearance attachment and safety of the head and footboards footboards Replace if necessary AC power cable Check that there are no cuts scrapes or breaks in the AC power cable or connector Replace the AC power cable if necessary Page 6 6 AvantGuard Electric Bed Service Manual QD1787 Preventive Maintenance Chapter 6 General Procedures Function Earth resistance Procedure Check the condition of the grounding straps of the bed Replace them if they are damaged or cut Measure the resistance to earth between the earth terminal of the power connector and the following points exposed metal e Point 1 Equipotential termin
93. da 5 14 Miscellaneous Items not Illustrated 0 0 0 0 0 0 cece eee 5 15 AvantGuard Electric Bed Service Manual QD1787 Page 5 1 Chapter 5 Parts List NOTES Page 5 2 AvantGuard Electric Bed Service Manual QD1787 Warranty Chapter 5 Parts List Warranty HERON LIMITED WARRANTY Hill Rom has a long tradition of providing superior quality products and service to our customer Our goal is Total Customer Satisfaction In that spirit Hill Rom is proud to offer the following warranty GENERAL WARRANTY APPLICABLE TO ALL CASES IN WHICH A SPECIFIC WARRANTY DOES NOT APPLY Hill Rom warrants to the original purchaser that its products and spare parts shall be free from defects in material and workmanship for a period of one 1 year after date of delivery Hill Rom s obligation under this warranty is expressly limited to supplying replacement parts and or service for or replacing at its option any product which is in the sole discretion of Hill Rom found to be defective In addition to the foregoing one year warranty Hill Rom warrants to the original purchaser that the frame and welds on its beds will be free from structural defects for the life of the bed SPECIFIC WARRANTIES CONSUMABLE WARRANTIES From the date of delivery a limited warranty of thirty 30 days applies to consumable items such as cushions blankets bed covers batteries program diskettes detector batteries incandescent
94. default a test column fully working column with its extension cable The control activates the test system or the test column Yes No y Replace the PSU refer to procedure 4 5 on page 4 18 15 Replace the 2 HL columns H Col refer to procedure 4 12 on page 4 37 and F Col refer to procedure 4 13 on page 4 42 This solves the problem Yes No Contact your national Hill Rom representative see the last page of this manual 16 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1787 Page 2 21 2 5 Thigh Section Malfunction Chapter 2 Troubleshooting Procedures 2 5 Thigh Section Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions the power driven FS function works if this option is fitted on the bed see Function Checks on page 2 6 1 At least one of the other functions is working Yes No y Go to RAP 2 1 on page 2 13 Perform the following to verify the proper operation of the CU a Disconnect the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AGO061 see Appendix on page 2 12 The functions are working Yes No y Go to step 4 Replace the CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexible a
95. derail structure Foot siderail supports Extension Sa 2 Linen holder 5 5 Articulation points a Egress bars Casters Braking and steering Sth wheel steering system KA Detection of Brake not applied E Accessory support and a medical device support rails Administrative procedures Labour Repair cost ased sy JO 350 8301 Inspected by ZOA APACE O O omo ul uu Il gg s Arat SPECE CEE oo oe eae X ca E gee o 3 2 m8 o AvantGuard Electric Bed Service Manual QD1 787 Page 6 17 Tools and Supply Requirements Chapter 6 General Procedures Tools and Supply Requirements To service the AvantGuard Electric Bed the following tools and supplies are required Ratchet wrench Facom P N S 151 Extension Facom P N S 210 13 mm rod 12 sided Facom P N S 13 6 mm combination spanner Facom P N 39 6 8 mm combination spanner Facom P N 39 8 10 mm combination spanner 2 required Facom P N 39 10 13 mm combination spanner Facom P N 39 13 17 mm combination spanner Facom P N 39 11 T25 Torx key Facom P N 89825 6 mm Allen key Facom P N 82H6 10 mm Allen key Facom P N 82H10 Small screwdriver Facom P N AR 3x75 2 phillips head screwdriver Facom P N AP 2x100 Soft faced hammer 8 mm pin punch Facom P N 249 8 Elevation tool P N AG0052 see figure 6 4 on page 6 1
96. djustment Procedures A CAUTION The head section must always be horizontal when removing or fitting the head section hard surface Failure to do so may give rise to an incorrect position and cause the elevation sliding system to malfunction 3 Grip the end handles of the HS hard surface operate one of the CPR handles and raise lower the HS to check that the hard surface moves without hindrance 4 Perform the Function Checks on page 2 6 Page 4 14 AvantGuard Electric Bed Service Manual QD1 787 4 4 AC Power Cable Chapter 4 Removal Replacement and Adjustment Procedures 4 4 AC Power Cable Identification see figure 3 1 on page 3 3 A WARNING Only the AC power cables listed in chapter 5 can be used on the AvantGuard Electric Bed Other cables may compromise the safety and conformity of the bed Time required 10 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 2 phillips head screwdriver 13 mm ratchet wrench Removal e SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 1 Unplug the power connector from the power outlet 2 Apply the brakes raise the sleep surface to the high position and lower the sleep surface sections to horizontal Use the battery supply 1f possible 3 Remove the headboard raise the HS and siderails to the hi
97. duction Chapter 7 Accessories A list of additional products that can be used in conjunction with the AvantGuard Electric Bed is available in chapter 7 Installation procedures for these accessories are also included Page 1 4 AvantGuard Electric Bed Service Manual QD1787 Typographical Conventions Chapter 1 Introduction Typographical Conventions This manual contains different typographical conventions designed to enhance the readability and understanding of its content These conventions are listed below e Standard text used for standard text throughout the manual e Boldface text emphasizes a word or phrase e NOTE sets apart special information clarification of an important instruction The symbol below highlights a WARNING or CAUTION Figure 1 1 Warning or Caution A A WARNING identifies situations or actions that may affect patient or user safety Disregarding a warning could result in patient or user injury A CAUTION points out special procedures or precautions that personnel must follow to avoid equipment damage The symbol below highlights a CAUGHT HAZARD WARNING Figure 1 2 Caught Hazard Warning Q D The symbol below highlights an ELECTRICAL SHOCK HAZARD WARNING Figure 1 3 Electrical Shock Hazard Warning AvantGuard Electric Bed Service Manual QD1787 Page1 5 Acronym List Chapter 1 Introduction Acronym List There are several acronyms a
98. e 4 53 4 16 Single Brake Bar Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 27 Braking Bar Stop 1680_051 D 6 Adjust the elevation tool so as to lift the frame by a few centimetres 7 Using the 13 mm ratchet wrench loosen the 4 screws C securing the head and foot end casters on the right hand side of the frame 8 Set the single brake bar E to neutral position horizontal see figure 4 28 on page 4 54 Figure 4 28 Braking System 1787_452 Page 4 54 AvantGuard Electric Bed Service Manual QD1 787 4 16 Single Brake Bar Chapter 4 Removal Replacement and Adjustment Procedures 9 Remove the caster plus support assembly D by using the soft faced hammer see figure 4 27 on page 4 54 NOTE Do not rotate the casters in their sockets This is to retain the adjustment 10 Using the soft faced hammer and the 8 mm pin punch head end punch out the brake and steer bar E and the head end hexagonal bar F from the right hand end of the bed viewed by a patient see figure 4 28 on page 4 54 NOTE The plastic articulation bushes H will be damaged during this step and must therefore be replaced 11 Remove the link bar G of the right hand longitudinal member of the chassis 12 Using the soft faced hammer and the screwdriver punch out the remaining pieces of the plastic articulation bushes H on the braking bar and the head hexagonal bar Replacement 1 In
99. e PSU recover the two screws and remove the cover B by unclipping it see figure 4 9 on page 4 19 5 On the main board D of the PSU disconnect the connector 5 of the TS actuator see table 4 1 on page 4 20 see figure 4 10 on page 4 19 6 Lift the cable cover C on the central rail of the sleep surface and remove the cable of the TS actuator see figure 4 9 on page 4 19 7 Remove the Truarc circlips from the pin at the end of the TS actuator rod E and remove the pin while supporting the actuator see figure 4 17 on page 4 34 NOTE Support the actuator then lay the TS flat if you could not lower this section to the horizontal position initially 1 Truarc is a registered trademark of Waldes Truarc Inc AvantGuard Electric Bed Service Manual QD1 787 Page 4 33 4 10 Thigh Section Actuator Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 17 Thigh Section actuator 1680_430 8 Remove the Truarc circlips from the pin supporting the TS actuator and remove the pin while supporting the actuator Replacement 1 Perform the removal procedure in reverse order A CAUTION Ensure that the elastic found on the new actuators is not wound around the actuator body before removing it cut If necessary turn the actuator end rod in the appropriate direction to unwind this elastic Indeed it is used to indicate the correct position of the switches and internal end of trave
100. e control is released EasyChair Automatic Chair Function This function combines the movements of the bed frame sections and uses the intermediate internal microswitches of the TS and FS actuators to control the interaction between these two sections Emergency Trendelenburg Function When the emergency trendelenburg control is activated by the LCU key on the LCU the PSU simultaneously moves the TS and FS to horizontal and adjusts the sleep surface to the trendelenburg position The two columns will be controlled until the maximum trendelenburg is reached or for as long as the control is activated The movement stops when the EOTs are reached The battery dedicated to this function guarantees movement when there is no AC power but is independent from the battery backup However the battery charging circuit is common Detection of Bed Connected to AC Power but Brakes not Applied Optional Depending on the Model The system operates by an NO switch mounted on the foot end crossbar and is activated by the braking hexagonal bar The PSU controls an audible beep if the following two elements occur together AC power present and switch not activated The beep only stops when the brake is applied or when the power connector is unplugged from the power outlet Trendelenburg Reverse Trendelenburg This fully electric function can only be accessed by the caregiver The function is enabled by the LCU key on the LCU and activated via the CU The
101. e movement In addition the hard surface moves upward on the section RHS Yes No y Goto RAP 2 6 on page 2 24 12 Press the lower button of the HS function The HS goes down to the low position without stopping or the audible alarm sounding during the movement do not consider the TS actuator In addition the hard surface moves downward on the RHS section Yes No y Go to RAP 2 6 on page 2 24 13 Press the raise button of the HS with automatic contour function button from the horizontal position of the base frame note that there are two raise buttons for this HS function The HS and TS rise simultaneously then the TS stops at an angle approaching 17 while the HS continues to rise to its highest position AvantGuard Electric Bed Service Manual QD1787 Page 2 7 Function Checks Chapter 2 Troubleshooting Procedures 14 15 16 17 18 19 20 Yes No y Go to RAP 2 7 on page 2 26 Press the lower button of the HS function from its highest position The HS goes down then the TS continues to drop when the HS reaches low position Yes No y Go to RAP 2 7 on page 2 26 If the bed is fitted with battery backup optional unplug the power connector from the power outlet and test the functions checked in step 12 to step 14 The bed works properly Yes No Go to RAP 2 8 on page 2 29 If the bed is fitted with a power driven FS operate the TS raise control until i
102. ection Microswitch 1787_413 Replacement 1 Perform the removal procedure in reverse order 2 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1 787 Page 4 51 4 16 Single Brake Bar Chapter 4 Removal Replacement and Adjustment Procedures 4 16 Single Brake Bar Identification see figure 3 1 on page 3 3 Time required 40 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench 10 mm Allen key 6 mm Allen key Sth wheel steering system only 8 mm pin punch Screwdriver Soft faced hammer Elevation tool P N AG0052 NOTE Removing the brake and steer system means breaking and therefore replaceing the two bushes that linking the braking bar link bar and hexagonal brake bar at the head end of the bed NOTE The single brake bar is part of a kit that also contains the head end hexagonal bar and the plastic articulation bushes Removal 1 Apply the brakes set the sleep surface to mid height and the TS and the FS to the horizontal position Raise and lock the siderails in high with the foot siderails folded SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED
103. ed Free the foot frame cover A Note that it is secured by clips at each end see figure 4 25 on page 4 51 Remove the cable cover on the central rail A of the sleep surface at the foot end of the bed and disconnect the cable from the F Col at the level of the connector B see figure 4 21 on page 4 43 Page 4 42 AvantGuard Electric Bed Service Manual QD1787 4 13 Foot Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 21 Foot Hilow Column Connection 1680_431 7 Perform the following if the bed is imobilised in the low position a Plug the bed into an appropriate power source and lockout all the functions except the HL function b Operate the HL raise using the control pendant to bring the H Col to high position NOTE When a faulty HL column is disconnected the other column will still function c Lock out the HL function SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur lt d Disconnect the power connector from the power outlet If the bed has a battery ensure that no LED is lit up on the control pendant If this is not the case wait for around one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 8 Remove the two cover
104. ed with a chair function press the non chair position button of the bed The HS TS and FS sections gradually return to the horizontal as long as the button is pressed Yes No Go to RAP 2 11 on page 2 35 22 Operate the HL raise until the intermediate position is reached then operate the sleep surface trendelenburg by using the lateral caregiver unit The sleep surface gradually tilts to maximum trendelenburg without any problem or abnormal noise Yes No y Go to RAP 2 12 on page 2 36 23 Operate the sleep surface reverse trendelenburg by using the lateral caregiver unit The sleep surface gradually tilts to maximum reverse trendelenburg without any problem or abnormal noise Yes No y Go to RAP 2 12 on page 2 36 24 If the bed is fitted with an emergency trendelenburg function optional Leave the bed in reverse trendelenburg position and operate the HS raise to high position Next enable the emergency trendelenburg function by using the two lateral caregiver units successively and carry out the movement in several stages The bed moves gradually to trendelenburg while the sections of the bed frame move to the horizontal position Yes No Go to RAP 2 13 on page 2 37 25 Adjust the bed to reverse trendelenburg with the HS section in high position then unplug the power connector from the power outlet Operate the emergency trendelenburg function again from one of the lateral caregiver units and ca
105. eeving stop F and free the CPR cable from its guide on the actuator see figure 4 14 on page 4 28 Figure 4 14 CPR Control on HS Actuator Ma 1680_429 Page 4 28 AvantGuard Electric Bed Service Manual QD1787 4 8 Head Section Actuator Chapter 4 Removal Replacement and Adjustment Procedures 11 Remove the cable and its pivoting attachment from the release cam G NOTE The release cam G has a slot enabling the cable to pass This slot must always point toward the articulation centre of the motor when installing so as to prevent it from retracting when in use 12 Remove the Truarc circlips from the pin at the end of the HS actuator rod and remove the pin while supporting the actuator 13 Remove the Truarc circlips from the pin supporting the actuator body and remove the pin while supporting the actuator and remove the actuator from the bed Replacement 1 Perform the removal procedure in reverse order A CAUTION Ensure that the elastic found on the new actuators is not wound around the actuator body before removing it cut If necessary turn the actuator end rod in the appropriate direction to unwind this elastic It is used to indicate the correct position of the switches and internal end of travels of the actuator The incorrect position of these elements may damage the actuator or the structure of the bed 2 Perform the Function Checks on page 2 6 1 Truar
106. egiver Unit Key 1680_104 AvantGuard Electric Bed Service Manual QD1787 Page 1 19 Introduction Chapter 1 Introduction The lateral caregiver unit is situated on each side of the bed under the seat section of the sleep surface The LCU key provides access A to the function lockout sequence validation of the trendelenburg reverse trendelenburg function and to the emergency trendelenburg function C optional on LI158Ax It also has a LED indicator B for low battery power availability of battery power or battery charging see figure 1 13 on page 1 19 The CPR control is separate from the LCU but is located next to it Electrical Function Management The functions are locked out unlocked according to the following sequence Push momentarily the LCU key in the electrical function management trendelenburg reverse trendelenburg access opening A of the LCU The LEDs flash on the control unit except for the trendelenburg reverse trendelenburg LED which is continuously illuminated Select the functions to be locked out the LEDs go off or unlocked the LEDs start flashing again by pressing the buttons of the relevant functions Push momentarily the LCU key in the same opening to validate or function management will automatically de activate after approximately 1 min Only the LEDs of the active functions remain lit NOTE If all the functions are initially locked out no LED is lit at the moment
107. en fitting the casters see figure 4 27 on page 4 54 Removal ly w 1 Apply the brakes raise the sleep surface to mid height the HS the TS and the FS to horizontal with the siderails raised and locked remove the headboard SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button on the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Free the head or foot frame cover A depending on the caster to be replaced Note that it is secured by clips at each end see figure 4 25 on page 4 51 1 Allen is a trademark of Industrial Fasteners Inc Page 4 56 AvantGuard Electric Bed Service Manual QD1787 4 17 Caster Brake and Steer Caster Chapter 4 Removal Replacement and Adjustment Procedures A WARNING Never lift the entire bed with the elevation tool lts maximum load is 150kg Disregarding this warning may cause the elevation mechanism to malfunction and could cause injury A CAUTION If the sleep surface cannot be raised to mid height for whatever reason the arms of the elevation to
108. eplacement and Adjustment Procedures 4 3 Control Panel on a Flexible Arm Identification see figure 3 1 on page 3 3 Time required 15 min Number of people 1 plus 1 occasionally Possibility of intervening with patient on the bed No Tools required 2 phillips head screwdriver Removal 1 Identify the version of the control panel on a flexible arm to be replaced LIL58Ax or LI159Ax to ensure that the replacement will allow the same functions Raise the HS to the high position then raise and lock the siderails and remove the headboard Apply the brakes raise the sleep surface to the high position and lower the sleep surface sections to horizontal Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated Identify the assembly of the control panel on a flexible arm on the HS It is fixed to the hard surface by the knob screw F its cable is maintained by the supports E and guided by the notch J see figure 4 7 on page 4 13 Page 4 12 AvantGuard Electric Bed Service Manual QD1787 4 3 Control Panel on a Flexible Arm Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 7 Ins
109. er E of the caster with a screwdriver to be inserted at each side of the caster between the bands Lift and remove the caster cover d Position the 10 mm combination spanner on the central shaft of the caster which must always be braked then turn the caster body clockwise viewed from above until the bands begin to be braked 1 Allen is a trademark of Industrial Fasteners Inc 2 Loctite is a registered trademark of Loctite Corporation AvantGuard Electric Bed Service Manual QD1787 Page 4 59 4 17 Caster Brake and Steer Caster Chapter 4 Removal Replacement and Adjustment Procedures e Refit the caster covers Figure 4 30 Caster 1680_416 5 Depending on the caster type adjust the caster body as follows a Non antistatic single band caster 4 half turns clockwise viewed from above b Antistatic single band caster 3 half turns clockwise viewed from above c Antistatic and non antistatic double band casters 3 half turns clockwise viewed from above 6 Fit the two securing screws of the caster and tighten them with the 13 mm ratchet wrench NOTE When adustments are complete apply Loctite 242 adhesive to the two securing screws of the caster to ensure that they are properly locked 7 Lower the elevation tool to place the bed on its casters 8 Apply the brakes on the bed and check the braking efficiency by pushing the bed along a level floor with a force of 200 newton
110. er connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 4 Using the 13 mm ratchet wrench loosen and remove the two screws A securing the cover B to the PSU recover the two screws and remove the cover B by unclipping it see figure 4 9 on page 4 19 5 Remove the cable cover C 6 On the main board D of the PSU disconnect the connector 1a or 1b from the LCU cable to be replaced see table 4 1 on page 4 20 and remove the cable from the central rail of the bed frame see figure 4 10 on page 4 19 7 Open the cable track located under the crossbar E of the bed frame and free the cable from the LCU see figure 4 12 on page 4 26 AvantGuard Electric Bed Service Manual QD1787 Page 4 25 4 7 Lateral Caregiver Unit Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 12 Installing the Lateral Caregiver Unit 1680_424 8 Disconnect the connector F of the control pendant or control panel on a flexible arm if this is connected to the LCU that is to be replaced see figure 4 12 on page 4 26 9 Using the 13 mm ratchet wrench remove the screw G securing the LCU H to the bed frame and remove it from the bed Replace
111. er to procedure 4 1 Hard surfaces of Remove each hard surface and check its condition refer to the bed frame procedure 4 1 Replace if necessary CPR release Inspect the assembly of the handles cable and CPR mechanism at the RHS actuator Check that the screws are present and fully tightened Check the condition of the two CPR release handles cable and CPR mechanism at the RHS actuator Operate the HS raise up to the high position by using the control pendant then activate one of the CPR releases Ensure that the head section lowers Carry out the same tests on the other side of the bed Release the CPR handles and check that the mechanism relocks correctly by operating the RHS raise for a few seconds The elevation movement must be observed Replace the RHS actuator assembly in the event of a malfunction or broken cable Page 6 10 AvantGuard Electric Bed Service Manual QD1787 Preventive Maintenance Chapter 6 General Procedures Function Head section gas springs Procedure Inspect the assembly of the gas springs Check that the screws are present and fully tightened Check that there is no oil on its shaft Operate the HS raise up to the high position by using the control pendant then activate one of the CPR releases Ensure that the HS lowers rapidly to the intermediate position then gradually to the low position Replace the gas springs if necessary refer to procedure 4 1
112. ergency trendelenburg function The emergency trendelenburg function works after charging Yes No y gt Goto step 11 Go to Final Actions on page 2 12 Perform the following a Replace the emergency trendelenburg batteries with new ones refer to procedure 4 6 on page 4 22 b Operate the emergency trendelenburg function The emergency trendelenburg function works Yes No The PSU is defective Replace it refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on Page 2 40 AvantGuard Electric Bed Service Manual QD1787 2 14 Emergency Trendelenburg Function not Working with the Bed not Connected to AC Power Chapter 2 Troubleshooting Procedures page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 12 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1787 Page 2 41 2 15 Bed Connected to AC Power Brakes not Applied Detection Malfunction Chapter 2 Troubleshooting Procedures 2 15 Bed Connected to AC Power Brakes not Applied Detection Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions the HL function works see Function Checks on page 2 6 1 At least one of the other functions is
113. ering MALICIA Sa ee ORES lt 2 48 a Wheel Steering Malfunction 22 2 eyes sree A 2 49 Page 2 2 AvantGuard Electric Bed Service Manual QD1787 Getting Started Chapter 2 Troubleshooting Procedures Getting Started Begin each procedure in this chapter with step 1 Follow the sequence outlined each step assumes that the previous steps are correct Each step is the normal operational event of the product and can be confirmed by answering Y yes or N no to the question Your response leads either to another step in the procedure a Repair and Analysis Procedure RAP or the replacement of a component If a number of components are mentioned replace them in the order given Start with initial actions to collect the initial information concerning the problem Perform the function checks to isolate identify a problem and verify the repair after completing each corrective action replacing or adjusting a part installation of a cable etc Perform the final actions after the function checks in order to verify the repair If troubleshooting procedures do not isolate the problem ask for technical support from your national Hill Rom representative refer to the last page of this manual A WARNING Only facility authorised personnel should troubleshoot the AvantGuard Electric Bed Troubleshooting by unauthorised personnel could result in personal injury or equipment damage NOTE The purpose of this troubleshooting me
114. esigned period of use see Preventive Maintenance on page 6 5 Consequently the columns can be replaced in the event of a failure but never dismantled or repaired except by Hill Rom personnel AvantGuard Electric Bed Service Manual QD1787 Page 3 7 Mechanical System Chapter 3 Theory of Operation Casters The bed has four casters three brake and one brake steer each with a 30 inclination activation mechanism and three positions e Brake caster Brake Neutral Neutral positions e Steer caster Brake Neutral Steer positions Antistatic brake casters are available as an option They are always fitted in pairs when this option is installed and are identified by a yellow dot on the tyre single band or in the centre of each caster double band They are located to the left if the steering caster is on the right hand of the head end or to the right of the sleep surface if the steering caster is on the left hand of the foot end Braking The braking system comprises four brake casters controlled by two hexagonal bars with a connecting bar situated in the righthand box section of the lower frame The brake control bar is directly mounted on the foot end hexagonal bar of the bed Steering The head end caster on the right viewed by the patient or foot end left optional is a steering caster It is also controlled by the brake steer bar 5th Wheel Steering System This automatic steering system comprises a supp
115. f the bed AvantGuard Electric Bed Service Manual QD1 787 Page 1 27 Specifications Chapter 1 Introduction Power Supply Unit The AvantGuard Electric Bed is equipped with electric actuators for the HS TS and FS except LI158Ax and HL functions which are controlled by via the power supply unit The power supply unit operates with 230VAC 50Hz for tolerances see table 3 2 on page 3 10 and can be equipped with a 24VDC battery backup optional This battery controls the functions of the bed when it is not supplied with AC power A specific audible alarm can be heard when the battery power is low It is set off when a function of the bed is activated after having enabled the battery mode When the alarms sounds the bed plugged into an appropriate power source so that the battery can be recharged The AvantGuard Electric Bed complies with the IEC standard 60601 2 38 and ammendment 1 for details see Regulations Standards and Codes on page 1 29 Every bed undergoes operating tests before delivery and tests are carried out for earth conninuity potential equalisation and leakage current Control Unit Control Pendant The control pendant provides access to bed electric functions and also provides indicators of function status It operates at a very low safe voltage and comprises the buttons and LEDs that enable the bed to be controlled except for low battery It is connected to the rear of the LCU o
116. g Elevation Arm Subassembly 4 68 Removals arias rae toe ates bo Sl ete a 4 68 Replacement dadas acne tt he dot at de dar td Bed agra 4 69 Siderail sitan ne SS e ada ease See ow tng eG ea le Se Acca elena es sade 4 70 REMOVAL da oa ee ee AL EA BR ees ee ee Be ee ia 4 70 Replacement vrs niera bits a ta a e ada E 4 7 Foot Siderail Support sucio beige ebb i siiegsee debe runer 4 72 Removal doe fag nade A Dek E deen eed 4 72 Replacements ence e a ES og Dee o h elie Soe tite 4 73 Head and Footboard estais wee betwee eel sett Gea dd ge atedasas 4 74 Removal rt SAU ths ein EY dd ei LO othe o ON he ae Ea 4 74 Replacement id s sng ee bean Pd eRe ee ee a kw ee ees de 4 74 Linen Holder Table optional 0 0 0 eee eee eee 4 75 Removal area ioe eee ad cha ea tere a ee es 4 75 Replica Sede eee We ka eee eds 4 75 EXtension Optional esser gigi ss oyna wy bw eee towards eae Se eels 4 76 Removal a AAA ADS See Sees wha tae ee A 4 76 Replacement lesa ee eee Gam al a del lar al de 4 77 Betess Bar cain cebu Wane ie ea ele owed ante Liang iia dacs 4 78 Removal 200 ceded A a ene BOO oe ee eS 4 78 Replacement castas e A te ee 4 79 AvantGuard Electric Bed Service Manual QD1787 Page 4 3 Chapter 4 Removal Replacement and Adjustment Procedures NOTES Page 4 4 AvantGuard Electric Bed Service Manual QD1787 4 1 Bed Frame Hard Surface Replacement Chapter 4 Removal Replacement and Adjustment Pro
117. g the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control unit If this is not the case wait for about one minute until the battery mode is deactivated 4 Remove the hard surfaces of the seat and thigh sections refer to procedure 4 1 5 Remove the cable cover A of the central rail of the bed frame see figure 7 8 on page 7 12 Figure 7 8 Installing the Control Unit on Flexible Arm B C D A 1680_403 6 Position the arm of the control unit with respect to the hard surface and fit the supporting knob screw B with its yoke Tighten the assembly 7 Press slightly on the centre of the headsection hard surface so as to move it away from the HS frame and pass the connector through the specific notch C 8 Install the two supports D and tighten the corresponding screws with the medium Phillips head screwdriver 9 On the right hand or left hand lateral caregiver unit note the connection socket available for the control panel on a flexible arm socket A according to the control pendant position see figure 4 6 on page 4 11 10 Open the cable track located under the righthand or lefthand central crossbar B of the bed frame Page 7 12 AvantGuard Electric Bed Service Manual QD1787 7 4 Control Panel on a Flexible Arm Kit AD136A Chapter 7 Accessories 11 Position the cable of the control panel on a flexible arm in the central rail of the bed frame then in the cable
118. gement Lateral caregiver unit on each side Single band casters diameter 150 mm CPR release Removable headboard Head section angle indicator Patient position indicator 6 attachment points for restraining straps Linen holder Brake and steer control 5th wheel Hilow pedal Double band casters diameter 150 mm Accessory support bar I V pole sockets at each corner N lt x lt lt c 4 7 o lo z z r x gt I V pole and patient helper sockets at each head end corner a Optional b Not shown AvantGuard Electric Bed Service Manual QD1787 Page 1 9 Introduction Chapter 1 Introduction Bed Positions Figure 1 5 EasyChair Automatic Chair Position 1787_115 Figure 1 6 Reverse Trendelenburg 1787_116 Page 1 10 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction Figure 1 7 Emergency Trendelenburg 1787_117 Figure 1 8 EasyChair Automatic Chair Position and Reverse Trendelenburg 1787_118 AvantGuard Electric Bed Service Manual QD1787 Page 1 11 Introduction Chapter 1 Introduction Figure 1 9 Trendelenburg 1787_119 Page 1 12 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction Figure 1 10 Hilow Positions High Position Low Position a a CP ea eo UU Se AR RE A ces
119. gh position Use the battery supply if possible or the CPR function 4 Lock out all the functions If the bed has a battery wait for about one minute until the battery mode is deactivated Next ensure that no LED is lit on the control pendant Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 5 Using the 13mm ratchet wrench loosen and remove the two screws A securing the cover B on the PSU recover the two screws and remove the cover B by unclipping it see figure 4 9 on page 4 19 6 Using the medium phillips head screwdriver remove the two screws securing the strain relief bracket C of the AC power cable on the right hand side view of patient lying down of the frame The AC power cable has a special straight section F for its correct positioning with the strain relief bracket C see figure 4 8 on page 4 16 AvantGuard Electric Bed Service Manual QD1787 Page 4 15 4 4 AC Power Cable Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 8 Installing the AC Power Cable 1680_423 7 Note the passage of the AC power cable and the connection of the wires on the AC input terminal E inside the PSU see figure 4 10 on page 4 19 8 Unclip the AC power cable from the AC input terminal E 9 Remove the AC power cable Replacement 1 Perform the removal procedure in reverse order checking that the wires are
120. gure 4 10 on page 4 19 Initial conditions the bed works normally when connected to an appropriate power source see Function Checks on page 2 6 1 Perform the following to verify the proper operation of the control unit a Disconnect the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AGO0061 see Appendix on page 2 12 The functions of the CU are working Yes No y Go to step 3 2 Replace the CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexible arm refer to procedure 4 3 on page 4 12 according to the option 3 Verify the battery pack connections F see figure 4 10 on page 4 19 Remove the cover of the PSU using the 13 mm combination spanner see figure 4 8 on page 4 16 The battery pack is properly connected see table 4 1 on page 4 20 Yes No y Connect the battery pack properly and go to Function Checks on page 2 6 4 There is a battery charging problem Plug the bed into an appropriate power source and ensure that it remains connected for at least 12 hours without the batteries being used Unplug the bed afterwards The bed operates correctly on battery backup after charging Yes No y Go to step 6 5 Go to Final Actions on page 2 12 6 Verify the battery backup pack by replacing the battery pack with a new battery pack The bed operates correctly with the new battery backup kit
121. h automatic contour DQ a gt LED indicator for locking unlocking of the function available for each function Thigh section raise Thigh section lower Foot section raise Foot section lower Bed raise sleep surface Bed lower sleep surface Bed non low position indicator Reverse trendelenburg Trendelenburg EasyChair Automatic Chair position Emergency release of the sleep surface sections VIO ZZ O AISIT A a a Activation of battery backup mode AvantGuard Electric Bed Service Manual QD1787 Page 1 15 Introduction Chapter 1 Introduction Hilow Function The sleep surface of the AvantGuard Electric Bed can be raised or lowered to move the patient or to facilitate getting on off the bed The column based elevation system frees space under the sleep surface and enables trendelenburg reverse trendelenburg to be obtained automatically by the control unit In order to raise the sleep surface press and hold the hilow raise button on the control pendant control panel on a flexible arm I or the hilow pedal Release the button or pedal at the desired height or when the maximum height is reached see figure 1 11 on page 1 14 In order to lower the sleep surface press and hold the hilow lower button on the control pendant control panel on a flexible arm J or the hilow pedal Release the button or pedal at the desired height or when the mi
122. has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 4 Using the 13 mm ratchet wrench loosen and remove the two screws A securing the cover B on the PSU recover the two screws and remove the cover B by unclipping it see figure 4 9 on page 4 19 5 Free the head end frame cover Note that it is secured by clips at each end 6 Note the passage of the H Col cable then on the main board D of the PSU disconnect the connector 6 of the H Col see table 4 1 on page 4 20 and free the cable from the H Col see figure 4 10 on page 4 19 AvantGuard Electric Bed Service Manual QD1787 Page 4 37 4 12 Head Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 19 Head HL Column Supply 1680_400 7 Disconnect the earthing conductor D of the H Col by unclipping it from the AC input board see figure 4 19 on page 4 38 8 Perform the following if the bed is imobilised in the low position a Plug the bed into an appropriate power source and lockout all the functions except the HL function b Operate the HL raise using the control pendant to bring the F Col to high position NOTE When a faulty HL column is disconnected the other column will still function
123. he foot siderail to folded position Check that the screws pins and circlips are not damaged or missing from the locking mechanism refer to the corresponding view of chapter 5 All the screws pins and circlips are secure and undamaged Yes No y Replace the missing parts refer to the corresponding view of chapter 5 then go to Function Checks on page 2 6 Check that the siderail locks and unlocks correctly The locking unlocking is correct Yes No y Replace the head siderail and the elevation arm subassembly refer to procedure 4 21 on page 4 68 The problem is solved Yes No y Contact your national Hill Rom representative see the last page of this manual 4 Goto Final Actions on page 2 12 Page 2 46 AvantGuard Electric Bed Service Manual QD1787 2 19 Braking Malfunction Chapter 2 Troubleshooting Procedures 2 19 Braking Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE Initial conditions none 1 Check the caster attachment Check that the screws or supports are not damaged loose or missing from the caster attachments refer to the assembly diagrams of chapter 5 All the supports and screws are secure and undamaged Yes No Replace the missing parts refer to the corresponding view of chapter 5 then go to Function Checks on page 2 6 2 Check the brake bar Set the brake steer bar to brake position then check that the
124. he power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 4 Using the 13 mm ratchet wrench loosen and remove the two screws A securing the cover B on the PSU recover the two screws and remove the cover B by unclipping it see figure 4 9 on page 4 19 5 On the main board D of the PSU disconnect the connector 3 of the HS actuator see table 4 1 on page 4 20 6 Lift the cable cover C of the central rail and remove the cable from the HS actuator 7 Operate the CPR control see CPR Emergency Release on page 1 17 and lower the HS to horizontal AvantGuard Electric Bed Service Manual QD1787 Page 4 27 4 8 Head Section Actuator Chapter 4 Removal Replacement and Adjustment Procedures 8 With the two 10 mm combination spanners loosen the adjuster E see figure 4 13 on page 4 28 NOTE Adjusting the CPR cable involves compensating the play at the level of the CPR handle and the HS actuator by the adjuster without creating any previous tension of the cable Figure 4 13 CPR Cable Stop 1680_439 9 Pull on the CPR cable sleeving to separate the adjuster E from its support 10 Using the two 10 mm combination spanners loosen the nut of the sl
125. he siderail including the folded foot siderail Gently push the siderail inwards with one hand and activate the unlocking button A on the siderail then rotate the siderail downward into the low position see figure 1 4 on page 1 8 e To deploy a foot siderail only if the head siderail is in high position Grip the mobile section of the foot siderail and pivot it towards the FS of the bed then position it in the support H provided for this purpose The position is secure when the siderail clicks into place AvantGuard Electric Bed Service Manual QD1787 Page 1 21 Introduction Chapter 1 Introduction To store a foot siderail With one hand grip the upper mobile section of the siderail at the foot end of the bed and with the other hand press the release button of the siderail support H then raise the siderail and pivot it towards the head end of the bed Then position it in the support provided for this purpose on the head siderail Patient Positioning Locator A locator triangle has been moulded on the inside of each head siderail C at the level of the foot siderail articulations to indicate the correct position of the patient s hips while on the bed Foot Siderail Supports e To raise a foot siderail support H Grip the support and move it upward until it locks The position is secure when the foot siderail support clicks into place e To lower a foot siderail support Ens
126. ic Bed Rapid Problem Solution Identification Table If a problem with the bed is identified rapidly use the following table to identify the appropriate troubleshooting procedure Table 2 1 Rapid Problem Solution Identification Table Problem Solution No Functions Work RAP 2 1 Control Unit Malfunction RAP 2 2 Function Management Mode Malfunction RAP 2 3 Hilow Malfunction RAP 2 4 Thigh Section Malfunction RAP 2 5 Head Section Malfunction excluding Automatic Contour RAP 2 6 Automatic Contour Malfunction RAP 2 7 Battery Backup Malfunction RAP 2 8 Foot Section Malfunction RAP 2 9 Assisted Foot Section Malfunction RAP 2 10 EasyChair M Automatic Chair Position Malfunction RAP 2 11 Page 2 4 AvantGuard Electric Bed Service Manual QD1787 Initial Actions Chapter 2 Troubleshooting Procedures Problem Solution Trendelenburg Reverse Trendelenburg Malfunction RAP 2 12 Emergency Trendelenburg Function not Working with Bed Connected to RAP 2 13 AC Power Emergency Trendelenburg Function not Working with the Bed not RAP 2 14 Connected to AC Power Bed Connected to AC Power Brakes not Applied Detection Malfunction RAP 2 15 CPR Malfunction RAP 2 16 Foot Siderail Locking Unlocking Malfunction RAP 2 17 Head Siderail Locking Unlocking Malfunction RAP 2 18 Braking Malfunction RAP 2 19 Steering Malfunction RAP 2 20 5th Wheel Steering Malfunction
127. ication can be heard during the movement Operate the trendelenburg Check that there is no friction or abnormal noises and that no audible overload indication can be heard during the movement Exit the function selection lockout sequence Replace the defective part in the event of a malfunction or abnormal noises Troubleshoot in the event of a malfunction see RAP 2 12 Page 6 12 AvantGuard Electric Bed Service Manual QD1787 Preventive Maintenance Chapter 6 General Procedures Function Emergency trendelenburg function according to configuration Procedure Adjust the bed to Chair position and the bed frame horizontal Operate the emergency trendelenburg function successively from the two lateral caregiver units Ensure that the sleep surface moves into trendelenburg that the three head thigh and foot sections move to horizontal without friction or abnormal noises and that no audible overload indication can be heard during the movement Operate the adjustment of the sleep surface to horizontal position then unplug the power connector from the power outlet Operate the Emergency trendelenburg function again and ensure that the function works correctly without AC power Reconnect the battery backup if the bed is equipped with one Recharge the emergency trendelenburg function battery or replace it if the function only works with the AC power applied Troubleshoot in the event of a
128. ics Table 3 3 Battery Backup Characteristics Description Specification Components 2 components of 12V Capacity of each component 1 2Ah Voltage 24VDC Maximum charging time 12 hours Discharging time storage battery Minimum 10 days connected to bed Discharging time storage battery not 3 months connected to bed Table 3 4 Emergency Trendelenburg Function Battery Characteristics Description Specification Components 2 components of 12V Capacity of each component 0 7Ah Voltage 24VDC Maximum charging time 10 hours Page 3 12 AvantGuard Electric Bed Service Manual QD1787 Electrical System Chapter 3 Theory of Operation Description Specification Discharging time storage battery Minimum 10 days connected to bed Discharging time storage battery not 3 months connected to bed These are sealed lead acid batteries connected in series on the main board and protected by fuses on the connection wires This battery type has no memory effect and can thus be recharged regularly without any particular constraints Reminder concerning the memory effect Some batteries types e g nickel cadmium must be fully discharged before being recharged Failure to observe this precaution will significantly reduce their lifetime The capacity of the battery backup after full charging is a minimum of 3 raise and lower cycles for HL HS TS and FS opti
129. ify the proper operation of the remote control a Disconnect all the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AG0061 see Appendix on page 2 12 The functions are working Yes No y Go to step 4 3 Replace the CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexible arm refer to procedure 4 3 on page 4 12 according to the option 4 Operate the actuator in raise mode The TS section rises Yes No y Go to step 7 5 Perform the following to verify the proper operation of the PSU and its correct usage a Operate the HS raise until the section reaches high position b Operate the HS lower The HS section lowers Page 2 26 AvantGuard Electric Bed Service Manual QD1787 2 7 Automatic Contour Malfunction Chapter 2 Troubleshooting Procedures Yes No y Replace the PSU refer to procedure 4 5 on page 4 18 6 The problem is due to incorrect usage 7 Troubleshoot using the following two possibilities a A new actuator or a TS and FS actuator simulation connector P N AG0063 see Appendix on page 2 12 is available Go to step 8 b A new actuator or a TS and ES actuator simulation connector possible only for power driven FS option is NOT available Go to step 12 8 Troubleshoot with a new actuator or TS and FS actuator simulation connector Perform the following to verify that PSU and TS
130. ion 1 AG0001 2 Head or foot board LI158Ax 2 AG0065 2 Head or foot board LI159Ax 3 AG0046 2 Endboard label LI158Ax 4 AG0066 2 Endboard label LI159Ax 5 AG0048 1 Power supply unit kit 6 AG0049 1 LCU left without emergency trendelenburg option 7 AG0051 1 LCU left with emergency trendelenburg option 8 AG0054 1 LCU right without emergency trendelenburg option 9 AG0055 1 LCU right with emergency trendelenburg option 10 AG0043 1 Bumper kit NOTE All the kits contain their respective fixing systems as well any applicable adhesive labels Hardware and other small metal or plastic parts are available in the kits Page 5 10 AvantGuard Electric Bed Service Manual QD1787 Intermediate Frame Chapter 5 Parts List NOTES AvantGuard Electric Bed Service Manual QD1787 Page 5 11 Sleep Surface Chapter 5 Parts List Sleep Surface Figure 5 4 Sleep Surface 1787 503 AvantGuard Electric Bed Service Manual QD1787 Page 5 12 Sleep Surface Chapter 5 Parts List Table 5 4 Sleep Surface All the kits contain their respective fixing systems as well any applicable adhesive labels Hardware and other small metal or plastic parts are available in the kits Item Number Part Number Quantity Description 1 AGO0005 1 Hard surface head section kit 2 AG0004 1 Hard surface seat 3 AG0003 1 Hard surface
131. ion tool must be used to remove the column in safety Use the elevation tool otherwise personal injury could occur 4 al CAUGHT HAZARD The bed frame must be secured lengthways to prevent it coming away from the rail at the level of the head HL column Disregarding this hazard may cause injuries e SHOCK HAZARD The use of electrical equipment carries a potential risk of electric shock Personnel must be trained and instructed in the inherent risks of electrical equipment e SHOCK HAZARD An abnormally high leakage current is symptomatic of a degradation within the AC power cable and the PSU Personal injury could result if the value of the leakage current is above 500 microamperes AvantGuard Electric Bed Service Manual QD1 787 Page 1 31 Safety Tips Chapter 1 Introduction Y SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the installation is not correct and or if the earth continuity is compromised SHOCK HAZARD Ensure that the elastic found on the new actuators is not wound around the actuator body before removing it cut If necessary turn the actuator end rod in the appropriate direction to un
132. l Replacement and Adjustment Procedures Replacement 1 Perform the removal procedure in reverse order 2 Perform the Function Checks on page 2 6 Adjustment Tools required 6 mm combination spanner 8 mm combination spanner 17 mm combination spanner NOTE The system is delivered fully assembled and adjusted However adjustment may be required during the life of the product if play is observed in the release handle C This adjustment must be made at the the control lever B see figure 4 33 on page 4 66 1 Check that the cable is positioned correctly at the level of the release head G of the gas spring see figure 4 34 on page 4 66 2 Using the 17 mm combination spanner ensure that the locknut is correctly tightened on the release head G of the gas spring 3 Using the 6 mm and 8 mm combination spanners loosen the adjuster located at the level of the control lever subassembly B see figure 4 33 on page 4 66 4 Tighten the adjuster again to compensate for the play in the cable NOTE The cable must not have any prior tension 5 Lock the adjuster of the control lever assembly B using the 6 mm and 8 mm combination spanners AvantGuard Electric Bed Service Manual QD1 787 Page 4 67 4 21 Head Siderail Including Elevation Arm Subassembly Chapter 4 Removal Replacement and Adjustment Procedures 4 21 Head Siderail Including Elevation Arm Subassembly Identification see figure 3 1 on p
133. l panel on a flexible arm optional or accessory To raise the TS press and hold the TS raise button E on the control pendant or the control panel on a flexible arm Release the button at the desired angle or when maximum elevation is reached To lower the TS press and hold the TS lower button F on the control pendant or the control panel on a flexible arm Release the button at the desired angle or when the low position is reached Foot Section The FS of the AvantGuard Electric Bed can be electrically LI159Ax or mechanically LI158Ax adjusted independently to obtain the desired inclination up to its maximum value when the TS is at a minimum of mid height This function G is activated from the control pendant optional or accessory LI159Ax or from the control handle situated on either side of the sleep surface LIL58Ax To raise the FS of the LI159Ax model press and hold the TS raise button G of the control pendant or control panel on a flexible arm until the TS reaches a minimum of mid height Next press the FS raise button on the same control unit Release the button at the desired angle or when maximum elevation is reached AvantGuard Electric Bed Service Manual QD1787 Page 1 17 Introduction Chapter 1 Introduction To lower the FS of the LIIS9Ax model ensure that the TS is at least at mid height and press the FS lower button H of the control pendant or control panel on a flexible arm Release the
134. lamps for overhead lighting fluorescent tubes heating elements temperature probes filtering material and microspheres This warranty is limited to the replacement of the items covered SUPPLY OF PARTS AND SERVICES Call your approved Hill Rom Distributor the address of your national Hill Rom representative are located on the last page of this manual In order to expedite service we request you furnish the following information customer account number product model number serial number and description of problem A qualified Specialist will provide via telephone troubleshooting assistance for hospital personnel and provide necessary parts to make repairs If telephone troubleshooting determines the need for on site technical service a qualified Territory Service Representative will be dispatched Replacement of non technical items will be the responsibility of the customer At the initiative of Hill Rom any products or parts subject to a claim within the scope of this warranty can be returned to the Hill Rom factories with carriage paid AVAILABILITY OF PARTS For new and remanufactured parts Hill Rom will make the parts available for a period of ten 10 years from the date of manufacture of the last model Note Some manufacturer components and assemblies may no longer be available In this case elements of equivalent functionality can be substituted THE FOREGOING WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES A
135. leep surface automatically adjusts the bed to the chair position This function N is activated from the control pendant or control panel on a flexible arm optional or accessory on LIIS9Ax model only To adjust the bed to the EasyChair Automatic Chair position press and hold the button N on the control pendant or control panel on a flexible arm Release the button at the desired angle or when the maximum chair position is reached To exit the chair position press the EasyChair Automatic Chair position exit button O on the control pendant or control panel on a flexible arm Release the button at the desired angle or when the the sleep surface sections reach low position Page 1 18 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction Lateral Caregiver Unit Figure 1 13 Lateral Caregiver Unit with Emergency Trendelenburg Function A 1680_102 Figure 1 14 Lateral Caregiver Unit without Emergency Trendelenburg Function A ge 1 Sl 1680_103 Table 1 4 Lateral Caregiver Unit Identification Letter Description A Function management opening trendelenburg reverse trendelenburg access opening Low battery battery charge LED indicator Emergency trendelenburg function access opening Lateral caregiver unit key storage mIa w Emergency trendelenburg function not available on LI158Ax model Figure 1 15 Lateral Car
136. ll reverse trendelenburg poistion and the HS in the high position NOTE The technician may lie on the bed during these tests AvantGuard Electric Bed Service Manual QD1787 Page 4 23 4 6 Battery Backup and Emergency Trendelenburg Function Chapter 4 Removal Replacement and Adjustment Procedures 3 The batteries should be fully recharged at the end of the test or if the low batery signal is heard during the tests 4 Perform the Function Checks on page 2 6 Page 4 24 AvantGuard Electric Bed Service Manual QD1787 4 7 Lateral Caregiver Unit Chapter 4 Removal Replacement and Adjustment Procedures 4 7 Lateral Caregiver Unit Identification see figure 3 1 on page 3 3 Time required 15 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench Removal 1 Identify the functions of the LCU to be replaced emergency trendelenburg present or not to ensure that the replacement will allow the same functions 2 Apply the brakes raise the sleep surface to the high position the sleep surface to horizontal then raise the siderails and the HS to the high position Remove the headboard then remove the hard surfaces of the SS and the TS W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the pow
137. lock the siderails and deploy the foot siderails lower the TS and the FS to horizontal and the sleep surface to the high position W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Free the foot frame cover A Note that it is secured by clips at each end see figure 4 25 on page 4 51 4 Remove the cable cover F situated on the F Col next to the BM to be removed and disconnect the cable from the BM see figure 4 22 on page 4 44 5 Open the cable track located under the foot crossbar of the frame and remove the cable from the BM 6 Using the 13 mm ratchet wrench loosen and remove the screw G securing the BM support H to the frame NOTE The two cables are connected to the external terminals of the BM terminals 1 and 2 NC Page 4 50 AvantGuard Electric Bed Service Manual QD1787 4 15 Replacement of Microswitch Assembly for Bed Braking Detection Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 25 Braking Det
138. lows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Free the foot frame cover A depending on the caster to be replaced Note that it is secured by clips at each end see figure 4 25 on page 4 51 4 Position each pedal in its support A and fit the corresponding bolts by using the 13 mm ratchet wrench see figure 7 9 on page 7 15 Page 7 14 AvantGuard Electric Bed Service Manual QD1787 7 5 Hilow Pedal Kit AD137A Chapter 7 Accessories Figure 7 9 Installing the Hilow Pedal 1680_742 5 Position the two cable tracks supplied on the two tubes B of the base frame between the pedals and the foot HL column 6 Remove the two covers C situated at the base of the column 7 Open the cable tracks situated on the base frame tubes and on the foot crossbar using the small screwdriver and insert the pedal cables Close the cable tracks 8 Connect the pedal cables to the connector provided at the base of the foot HL column 9 Refit the two covers C to the base of the foot HL column 10 Perform the Function Checks on page 2 6 of this manual Removal Perform the installation procedure in reverse order AvantGuard Electric Bed Service Manual QD1787 Page 7 15 7 5 Hilow Pedal Kit AD137A Chapter 7 Accessories NOTES Page 7 16 AvantGuard Electric Bed Service Manual QD1787 O France Hill Ro
139. ls of the actuator The incorrect position of these elements may damage the actuator or the structure of the bed 2 Perform the Function Checks on page 2 6 1 Truarc is a registered trademark of Waldes Truarc Inc Page 4 34 AvantGuard Electric Bed Service Manual QD1787 4 11 Foot Section Actuator Replacement LI159Ax only Chapter 4 Removal Replacement and Adjustment Procedures 4 11 Foot Section Actuator Replacement LI159Ax only Identification see figure 3 1 on page 3 3 Time required 20 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench Removal 1 Apply the brakes raise the sleep surface to the high position the HS in the high position and the TS in the intermediate position Fold the foot siderails then lower the siderails 2 Remove the head and footboards then the hard surfaces of the SS the TS and the FS refer to procedure 4 1 Remove any of accessories that may be fitted to the FS W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the con
140. m Z I du Talhou t BP 14 56330 Pluvigner E 33 0 2 97 50 92 12 Service Clients 33 0 802 01 23 45 Fax 33 0 2 97 50 92 00 Nederland Hill Rom Medical Services BV Hagenweg 1c Postbus 173 4130 ED Vianen E 31 0 347 32 35 32 Fax 31 0 347 32 35 00 ED United Kingdom Hill Rom Clinitron House Ashby Park Ashby de la Zouch Leicestershire LE65 1JG T 44 0 1530 411333 Fax 44 0 1530 411555 Italia Hill Rom S p A Via Ambrosoli 6 20090 Rodano 39 0 02 950541 Fax 39 0 02 95328578 Sverige Hill Rom Medibed AB PO BOX 3035 Djursholmsv gen 29 18303 T by 46 0 8 756 07 07 Fax 46 0 8 732 69 38 e mail se marketing hill rom com GD Ireland Hill Rom Unit 1 Kilcarberry Business Park Nangor Road Dublin 22 T 353 0 1 413 6005 Fax 353 0 1 413 6030 email dublin sales hill rom com Suisse Schweiz Hill Rom SA Chemin du Vallon 26 1030 Bussigny T 41 0 21 706 21 30 Fax 41 0 21 706 21 33 e mail hrch info hill rom com Portugal Hill Rom Iberia S L Edificio Oc ano 1 1 4B Parque de Negocios Mas Blau 08820 El Prat de Llobregat Barcelona T 34 0 93 478 77 49 Fax 34 0 93 379 48 91 O Deutschland Hill Rom Kurhessenstrasse 11 64546 M rfelden T 49 0 6105 9320 Fax 49 0 6105 932 182 gt a Osterreich Hill Rom Austria GmbH Buropark Donau Haus 8 Inkustrasse 1 7 3400 Kloster
141. main principles e Filtering and amplification of inputs for electrostatic protection and signal formatting e Multiplexing for the management of the buttons e Demultiplexing for the control of the LEDs Data from the buttons and LEDs are transmitted in series link type SPI so as to reduce the number of wires The clock signal that enables these exchanges is controlled by the power card All the data is sent in tens of milliseconds When the system is in standby mode the power is cut off and the clock signal is no longer controlled The battery mode activation button is therefore directly wired so as to be continually operational Control Panel on a Flexible Arm The control panel on a flexible arm is available in two configurations just like the control pendant and is identical in design Lateral Caregiver Unit The LCU is available in two configurations depending on whether or not the emergency trendelenburg function is available This unit is situated on either side under the sleep surface and has one opening two openings when the emergency trendelenburg function is fitted for the LCU key a charging indicator for the battery function and a connector for the control pendant or control panel on a flexible arm Each opening has a NO switch that activates function management and trendelenburg reverse trendelenburg functions and the emergency trendelenburg function for the other opening A single 12 wire bundle connected to the PSU g
142. me Hard Surface Replacement Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 3 Installing the Head Section Hard Surface 1680_409 d Manually move the brackets H on each side of the HS hard surface and slightly raise the hard surface photo shown with HS section lifted for clarity see figure 4 4 on page 4 7 Figure 4 4 Head Section Hard Surface Support 1680_450 e Remove the hard surface from its supports I by lifting it slightly at the base and pulling it towards the head end of the bed see figure 4 2 on page 4 6 Replacement 1 Perform the removal procedure in reverse order AvantGuard Electric Bed Service Manual QD1787 Page 4 7 4 1 Bed Frame Hard Surface Replacement Chapter 4 Removal Replacement and Adjustment Procedures A CAUTION The head section must always be horizontal when removing or fitting the head section hard surface Failure to do so may give rise to an incorrect position and cause the elevation sliding system to malfunction NOTE The SS TS and FS hard surfaces are supported by the restraining strap attachment points on either side of the sleep surface NOTE The hard surface of the SS can only be installed when the HS is in high position Figure 4 5 Head Section Hard Surface Positioning i y ci F 1680_422 2 To replace the HS hard surface perform the following Fit the HS hard surface with the HS in the horizontal position b In
143. ment 1 Identify the functions of the LCU to be replaced emergency trendelenburg present or not to ensure that the replacement will allow the same functions 2 Perform the removal procedure in reverse order 3 Perform the Function Checks on page 2 6 Page 4 26 AvantGuard Electric Bed Service Manual QD1787 4 8 Head Section Actuator Chapter 4 Removal Replacement and Adjustment Procedures 4 8 Head Section Actuator Identification see figure 3 1 on page 3 3 For clarity two separate procedures are used for the removal of the HS actuator and its CPR release cable However they are part of the same kit one reference available Time required Time required 30 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 10 mm combination spanners 2 required 13 mm ratchet wrench Removal 1 Apply the brakes raise the sleep surface to the high position the HS in high position the HS the TS and the FS to horizontal Fold the foot siderails then lower the siderails Remove the headboard NOTE If the HS actuator is defective use the CPR control to raise the HS 2 Remove the hard surfaces of the SS and the TS refer to procedure 4 1 W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the power connector from t
144. mooo 7 11 Installation 7 11 Removal add eds de ddr dida barda 7 13 Hilow Pedal Kite AD137A 2 O a ote 7 14 Installation A A Ware ae be AA 7 14 R mol st tated bet ob ii ta e a bt hee 7 15 Page xii AvantGuard Electric Bed Service Manual QD1787 Chapter 1 Introduction Chapter Contents PUTOS ide niche Meta od racks kb at da al Ad eke hose 1 3 AU aa Jue a A SO de a dl e rata OO 1 3 e A owe ead be eR aad ban fray ae oe a we 1 3 Chapter 1 hitrodueuon 5 05 a AS ii 1 3 Chapter 2 Troubleshooting Procedures 0 0 cece eee eee 1 3 Chapter 3 Theory of Operation a e AS AA 1 3 Chapter 4 Removal Replacement and Adjustment Procedures 1 3 Chapter 5 Parts Liso va scr seu tad AS RRA 1 3 Chapter 6 General Procedures sake es A et 1 3 Chapter ACCESO ance oven daa rd aaa 1 4 Typosraphical C onventions adas oops LI za 1 5 ACTOR LS A sl ADS AE AS A A A AA 1 6 INTOGdUCHON psp la a o e E Ad lo A Rees 1 7 OVEIVIEW A A A AA AA 1 7 Bed POSMIO S itv ral it oh Re Be eee ee 1 10 Operating Precautions ss S awe nag dew eee od 1 14 A ae dy eee oan eh oe tot oh eit haan tech ed ural 1 14 ilo w Funcion bs od a o ald oe do da dl hr ol 1 16 Head Section a a ei 1 16 Retractable Head Section s c40iinbcscae beeen da hues 1 16 CPR Emergency Release cueste dar ae 1 17 Thigh Section segonna Stet A ed aot eh he doers tate acetate Beet DEO 1 17 POOL Sectors 4 3 aicctu aie E Se ae ta E og fee et oie Ge t
145. moval Replacement and Adjustment Procedures Figure 4 35 Installing the Head Siderail 1680_436 NOTE The photo is shown with the HS hard surface removed for clarity 6 Slightly tilt the head siderail assembly and its support inwards then remove it from the bed Replacement 1 Set the head siderail to high position with respect to its support and position the head siderail plus support assembly on the HS structure 2 Perform the removal steps in reverse order 3 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 69 4 22 Siderail Chapter 4 Removal Replacement and Adjustment Procedures 4 22 Siderail Identification see figure 3 1 on page 3 3 Time required 5 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required None Removal 1 Apply the brakes raise the sleep surface to the high position then lower the sleep surface sections to horizontal 2 Raise and lock the relevant siderail A CAUTION Always set the foot siderail to the folded position Failure to do so could result in equipment damage 3 Set the foot siderail to the folded position e SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 4 Lock out all the functions Unplug the power connector from the power outlet If the bed
146. n 1 17 AO ACOMODO ot en Be eee AL AS ee ls 1 18 EasyChair Automatic Chair Position 0 0000 e eae 1 18 Lateral Caregiver Ult vo daa 1 19 AvantGuard Electric Bed Service Manual QD1787 Page 1 1 Chapter 1 Introduction TotalGuard Integral Siderails yas 24 2 fs5 lt axe aS a 1 21 Brakes tg eou tie owe eee eee Aine owes oe aw yet 1 22 SECO dd adds ds daria ada 1 22 Headboards and Footboards 0 0 cece eee ene 1 23 Hard Surfaces on the Head Seat Thigh and Foot Sections 1 23 Casters ca ie o ta iia 1 23 Restraining Strap Attachment Points oooooocooocoooooooooo 1 23 Protective BUMPErS tira eee ee eed net 1 23 EV Pole Sockets cl ye tee oy Saas ete ds S 1 24 Optional Feattfes Y Ti de i ERA A 4 Gra A E 1 24 Specifications vacios ls dos ew ta ti a aia a 1 26 Physical Description 2 250 b0 e5 050d chic ebbd da a a da 1 26 Electrical Description oy DA Binder ye bes Ag arenas 1 27 Power supply Unitat A eee ea ieee 1 28 Control Wiles 4 ted weve hoe tw eee hie ai whee Seles 1 28 Motor Driven Actuators oye ane oh hats oe SA BEG ESA dd 1 28 Regulations Standards and Codes 2 2 0c eee eee eee eee ee 1 29 Model Identification Woods bees thee Ree Gad heeded Bedwas 1 30 Satety UES O le ie et A OM ha ea 1 30 Identification and Caution Labels 2 0 0 0 0 eee eee eee 1 35 Page 1 2 AvantGuard Electric Bed Service Manual QD1787 Purpose
147. n page 3 3 Time required 10 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench Removal 1 Apply the brakes raise the sleep surface to the high position and the HS to the high position Raise the siderails to the high position and remove the headboard SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet Ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Using the 13 mm ratchet wrench loosen and remove the two screws A securing the cover B to the PSU recover the two screws and remove the cover B by unclipping it see figure 4 9 on page 4 19 4 Remove the battery backup F or emergency trendelenburg function G battery from their housing in the PSU see figure 4 10 on page 4 19 5 If the battery backup is being replaced note the connection of the wires on the batteries red wire H on the outer and red terminals black wire I on outer terminal and the short wire connecting the two central terminals then disconnect them and lift away the batteries
148. n page 3 3 Time required 20 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench NOTE The batteries are not supplied with the PSU Check the dates shown on the batteries and replace them if they are older than five years Removal 1 Apply the brakes raise the sleep surface to the high position and the HS to the high position Raise the siderails to the high position NOTE If necessary use the CPR control to raise the HS SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Using the 13 mm ratchet wrench loosen and remove the two screws A securing the cover B of the PSU recover the two screws and remove the cover B by unclipping it Note the position of the cable cover C see figure 4 9 on page 4 19 Page 4 18 AvantGuard Electric Bed Service Manual QD1 787 4 5 Power Supply Unit Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 9 Power Sup
149. n page 4 12 according to the option fitted This solves the problem Yes No Contact your national Hill Rom representative see the last page of this manual Go to Final Actions on page 2 12 Page 2 36 AvantGuard Electric Bed Service Manual QD1787 2 13 Emergency Trendelenburg Function not Working with Bed Connected to AC Power Chapter 2 Troubleshooting Procedures 2 13 Emergency Trendelenburg Function not Working with Bed Connected to AC Power NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions a The trendelenburg reverse trendelenburg function works see Function Checks on page 2 6 b The chair function works see Function Checks on page 2 6 1 At least one of the other functions is working Yes No Go to RAP 2 1 on page 2 13 2 Try to activate the emergency trendelenburg function on the other LCU The emergency trendelenburg function works Yes No Go to step 5 3 Remove the cover of the PSU using the 13mm combination spanner see figure 4 8 on page 4 16 a Cross connect the 2 LCU b Activate the emergency trendelenburg function The emergency trendelenburg function works Yes No Ar Replace the PSU refer to procedure 4 5 on page 4 18 4 Replace the LCU refer to procedure 4 7 on page 4 2
150. n the right or left of the sleep surface and is inserted easily into the siderails for storage Two versions are available according to the bed model Depending on whether or not the controls are locked out they are accessible by the caregiver or patient except for trendelenburg reverse trendelenburg and several functions can be enabled simultaneously Control Panel on a Flexible Arm Optional or Accessory The AvantGuard Electric Bed can be equipped with a control panel on a flexible arm This unit has the same functions as the control pendant and also comes in two versions It is connected to the rear of the LCU on the right or left of the sleep surface simultaneously with the control pendant but necessarily to the opposite LCU to the control pendant s connection Motor Driven Actuators CAUTION The usage cycle intermittent operation of the actuators must not exceed 3min 30min otherwise equipment damage could result The operation of the HS TS and FS LI159Ax only is provided by three motor driven actuators and two columns for hilow They are all designed in such a way as to be protected against the ingress of liquids under normal conditions of use Page 1 28 AvantGuard Electric Bed Service Manual QD1787 Specifications Chapter 1 Introduction The motor driven actuators operate under a nominal 24VDC and have a connecting plug to facilitate removal replacement should problems arise Each movement is onl
151. neous Items not Illustrated Chapter 5 Parts List Miscellaneous Items not Illustrated Table 5 6 Miscellaneous Items not Illustrated Item Number Part Number Quantity Description Not shown AG0015 1 Power cord Europe grey Not shown AG0016 1 Power cord UK Not shown AGO0017 1 Power cord Switzerland Not shown AG0019 1 Emergency trendelenburg battery and bat tery backup kit Not shown AG0020 1 Emergency trendelenburg battery kit Not shown AG0024 1 Battery backup kit Not shown AG0025 1 Mechanical foot section kit Not shown AG0044 1 Hardware kit screws nuts etc Not shown AG0047 1 Plastic parts kit Not shown AG0042 1 Labels set NOTE All the kits contain their respective fixing systems as well any applicable adhesive labels Hardware and other small metal or plastic parts are available in the kits AvantGuard Electric Bed Service Manual QD1 787 Page 5 15 Miscellaneous Items not Illustrated Chapter 5 Parts List NOTES Page 5 16 AvantGuard Electric Bed Service Manual QD1787 Chapter 6 General Procedures Chapter Contents Cleaning and Care uu pista bereti radar ub bade ROS ba a Dies EEE 6 3 General Cleans eee a de a hd e da Sel os sta 6 3 Steam Cleaning ir aye re AR tas ln ta ee 6 3 Hard te Clean Spots es tek AA eA Soe Doa 6 3 Disinfection saae eeaeee E ea tg Od tg ow apes UPE ER 6 3 Stonne the Bed aati ie oe ee eM oe e a
152. neuburg T 43 0 2243 28550 Fax 43 0 2243 28550 19 e mail austria hill rom com Ell s Patient Care Solutions Chatziioannidis Nikolaos G lordanidi 2 Str 54635 Thessaloniki T 30 0 31 960 303 Fax 30 0 31 960 087 Hill Rom A HILLENBRAND INDUSTRY O Belgique Belgi Distribal Admiral De Boisot Straat 19 21 2000 Antwerpen T 32 0 32 38 43 15 Fax 32 0 32 16 06 86 Danmark Grendorf Medical a s Holmevej 30 DK 2970 H rsholm T 45 0 45 57 08 81 Fax 45 0 45 57 06 50 e mail Info gm medical dk Espa a Hill Rom Iberia S L Edificio Oc ano 1 1 4B Parque de Negocios Mas Blau 08820 El Prat de Llobregat Barcelona T 34 0 93 478 77 49 Fax 34 0 93 379 48 91 Norge Avalon Medical Biscop Gens Nilssongate 5 0659 Oslo E 47 0 23 03 63 70 Fax 47 0 23 03 63 71 International HILL ROM CO INC INTERNATIONAL DEPARTMENT 1069 STATE ROUTE 46 EAST J55 BATESVILLE INDIANA 47006 USA T 1 0 812 934 8173 Fax 1 0 812 934 7191 WEBSITE www hill rom com e mail international hill rom com
153. nimum height is reached Bed Non Low Position Indicator An LED lights up on the control unit when the bed is no longer in the low position in order to heighten caregivers awareness of the risks inherent to the sleep surface height refer to the user manual Head Section The HS of the AvantGuard Electric Bed can be adjusted independently in order to obtain the desired inclination up to its maximum level This function is activated from the control pendant or control panel on a flexible arm optional or accessory There are two specific buttons for the elevation one for the HS only B the other for the operating with the automatic contour A see figure 1 11 on page 1 14 To raise the HS press and hold one of the HS raise buttons on the control pendant or the control panel on a flexible arm A Release the button at the desired angle or when maximum elevation is reached To lower the HS press and hold one of the HS lower buttons on the control pendant or the control panel on a flexible arm C Release the button at the desired angle or when the low position is reached Head Section Angle Indicators A HS angle indicator with ball bearing O is available on each head siderail The labels are graduated for approximately every 10 of HS inclination and from 0 to 90 in one direction for normal use and from 0 to 20 in the other when the bed is in trendelenburg position see figure 1 4 on page 1 8 Retractable Head
154. not wound around the actuator body before removing it cut If necessary turn the actuator end rod in the appropriate direction to unwind this elastic Indeed it is used to indicate the correct position of the switches and internal end of travels of the actuator The incorrect position of these elements may damage the actuator or the structure of the bed 2 Perform the Function Checks on page 2 6 1 Truarc is a registered trademark of Waldes Truarc Inc Page 4 36 AvantGuard Electric Bed Service Manual QD1787 4 12 Head Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures 4 12 Head Hilow Column Identification see figure 3 1 on page 3 3 Time required 40 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench with extension Elevation tool P N AG0052 Removal 1 Raise the sleep surface to the high position at least the FS if H Col cannot be used the casters in line with the bed and apply the brakes 2 Lower the TS and the FS to the low position and raise the HS to the high position Raise and lock the siderails in high position and remove the head and footboards W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Lock out all the functions Unplug the power connector from the power outlet If the bed
155. nplugged from the power outlet MobiBar Egress Bar Optional The AvantGuard Electric Bed can be equipped with egress bars that assist the patient in getting out of bed They can be folded away so as not to hinder the caregiver or patient when not in use To use the MobiBar Egress Bar pull outwards and rotate towards the head of the bed To ensure the locking mechanism is secure try to rotate it towards the foot of the bed To store the MobiBar Egress Bar pull outwards and rotate towards the foot of the bed and under the sleep surface then release 5th Wheel Steering System Optional on LI158Ax The 5th wheel system enables the AvantGuard Electric Bed to be moved more easily by providing a central steering point under the frame It is activated or deactivated automatically according to the type of bed movement If a 5th wheel steering system is fitted there will not be a standard steer caster To set the 5th wheel in the bed moving position position the bed in the direction of the movement required and move it in a straight line The 5th wheel automatically sets to the required position Page 1 24 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction To disengage the moving position and enable sideways movements push the bed sideways and the 5th wheel will automatically unlock Extension Optional The AvantGuard Electric Bed is designed for tall patients
156. nt Yes No y Goto RAP 2 21 on page 2 49 36 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 11 Final Actions Chapter 2 Troubleshooting Procedures Final Actions 1 Perform the required preventative maintenance procedures See Preventive Maintenance Checklist on page 6 16 2 Carry out all necessary administrative tasks Appendix e A test control unit P N AGO061 is recommended for most of the following investigation procedures e This unit contains all the functions of the control pendant CU all options and Lateral Caregiver Unit LCU of the bed and can be connected inside the power supply unit PSU connector la or 1b grey mark D see figure 4 10 on page 4 19 e However for all of these tests a control pendant CU QD1604 can be used The control pendant can be connected at the rear of the LCU see figure 4 6 on page 4 11 e A HL column simulation connector P N AG0062 allows the working of the PSU column output to be checked 2 Red and Green LEDs enable you to determine whether the actuator is being controlled for raise or lower e An actuator simulation connector P N AG0063 allows the working of the outputs on the PSU to be checked for the TS or FS e Another actuator simulation connector P N AG0064 allows the working of the HS actuator output on the PSU to be checked Page 2 12 AvantGuard Electric Bed Service Manual
157. ode is deactivated 4 Position the two guides A in the spaces provided on the accessory support bars B and position the assembly below the holes to the right or left of the bed frame Ensure that the socket C of each support bar is located at the head end of the bed see figure 7 2 on page 7 6 Figure 7 2 Installing the Accessory Support Bars 1680_741 5 Fit the two securing screws D of the accessory support bars and lock them using the 6 mm Allen key 6 Repeat steps 4 and 5 to install the other accessory support bar on the other side of the bed frame Removal Perform the installation procedure in reverse order 1 Allen is a trademark of Industrial Fasteners Inc Page 7 6 AvantGuard Electric Bed Service Manual QD1787 7 2 Medical Device Support Rail Kit AD127A Chapter 7 Accessories 7 2 Medical Device Support Rail Kit AD127A This kit contains the medical device support rails and the parts that enable them to be installed on the bed Note Only one medical device support rail can be installed on the bed if the use of an accessory support bar is required other kit The medical device support rail kit includes the following e Two medical device support rails one right hand and one left hand e Four guides e Four securing screws Figure 7 3 Medical Device Support Rail Kit 1680_745 Identification see figure 3 1 on page 3 3 Time required 10 min Number of people 1 P
158. ol must not be positioned at the level of the actuators Disregarding this caution may cause the head section and foot section actuators to malfunction 4 When replacing a caster at the head end of the bed position the elevation tool under the head end crossbar of the bed see figure 4 29 on page 4 57 Or when replacing a caster at the foot end of the bed position the elevation tool under one of the sides of the frame see figure 4 26 on page 4 53 Then adjust the elevation tool so as to raise the casters by a few centimetres Figure 4 29 Caster Installation 1680_434 5 Using the 13 mm ratchet wrench loosen the 2 screws A securing the caster to be removed see figure 4 29 on page 4 57 6 Set the brake steer bar to neutral position 7 Remove the caster and its socket B possibly by using the soft faced hammer 8 Remove the caster from its socket AvantGuard Electric Bed Service Manual QD1787 Page 4 57 4 17 Caster Brake and Steer Caster Chapter 4 Removal Replacement and Adjustment Procedures Replacement 1 Before installing the new caster set the single brake bar to brake position then steer position Note the direction of rotation of the hexagonal bar when moving from one position to the other then reset the bar to neutral 2 Insert the new caster into the socket NOTE Each caster type has three positions as follows e Brake caster two neutral positions and one brake position e Steer
159. on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 17 2 3 Function Management Mode Malfunction Chapter 2 Troubleshooting Procedures 2 3 Function Management Mode Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Comments The symptom of the fault searched for is after pressing the function management button of the LCU none of the LEDs flash e At the start of this troubleshooting procedure the CU is connected to the LCU the LCU is called X by convention whereas the LCU of the opposite side is called Y e LCU is here understood to mean the Lateral Caregiver Unit and its cables which cannot be separated see parts list in Chapter 5 NOTE The trendelenburg reverse trendelenburg LEDs and low position HL LED do not flash in the steps that include the asterisk symbol 1 Atleast one of the other functions is working Yes No y Go to RAP 2 1 on page 2 13 2 Verify the proper connection in LCU X a Connect the CU to the other LCU Y b Press the function management control on the LCU All the LEDs flash Yes No y Go to step 4 3 Replace the LCU X refer to procedure 4 7 on page 4 25 4 Perform the following to verify the operation of the control unit a Disconnect the CU s control pendant or control panel on
160. onal with the safe working load on the bed 185kg The capacity of the emergency trendelenburg function battery after full charging is a minimum of 3 emergency trendelenburg position cycles from the full reverse trendelenburg position and HS in high position with the safe working load on the bed 185kg Control Logic and Specific Functions e Combined activations of the actuators provide the following functions Emergency trendelenburg see Emergency Trendelenburg Function on page 3 16 EasyChair Automatic contour Emergency trendelenburg function e Lockout management e Control unit interfacing e Indication of Bed Not in Low Position see Detection of Bed not in Low Position on page 3 15 e Indication of Bed Connected to AC Power but Brakes not Applied see Detection of Bed Connected to AC Power but Brakes not Applied Optional Depending on the Model on page 3 16 1 The emergency trendelenburg function battery is only fitted if the emergency trendelenburg option AND the optional battery backup are fitted AvantGuard Electric Bed Service Manual QD1787 Page 3 13 Electrical System Chapter 3 Theory of Operation All of these functions are integrated into a CPLD Complex Programmable Logic Device Control Pendant The control pendant is available in two configurations depending on the bed model Each unit operates at 3 3VDC supplied by the power card and integrates three
161. onal Hill Rom representative see the last page of this manual 10 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 19 2 4 Hilow Malfunction Chapter 2 Troubleshooting Procedures 2 4 Hilow Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 1 Atleast one of the other functions is working Yes No Go to RAP 2 1 on page 2 13 2 Identify the relevant column s Only one column is not working Yes No y Goto step 5 SHOCK HAZARD Electric current flows through the connections when the bed is switched on Trials conducted under these conditions carry a risk of electric shock Do not touch any wire or terminal which conducts an electric current otherwise personal injury could result 3 Swap the connectors 6 and 4 yellow on the power supply P C board see table 4 1 on page 4 20 The faulty column can now be activated Yes No y Replace the defective column H Col refer to procedure 4 12 on page 4 37 or F Col refer to procedure 4 13 on page 4 42 4 Replace the PSU refer to procedure 4 5 on page 4 18 5 Perform the following to verify the proper operation of the remote control a Disconnect the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AG0061 see Appendi
162. onnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure that the battery mode is properly deactivated 3 Set the foot siderail to the folded position 4 Remove the hard surfaces of the TS and the FS see procedure 4 1 on page 4 5 5 Using the 13 mm ratchet wrench remove the 2 screws A securing the siderail support and remove the assembly see figure 4 37 on page 4 73 Page 4 72 AvantGuard Electric Bed Service Manual QD1787 4 23 Foot Siderail Support Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 37 Foot Siderail Support 1680_417 Replacement 1 Perform the removal procedure in reverse order 2 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 73 4 24 Head and Footboard Chapter 4 Removal Replacement and Adjustment Procedures 4 24 Head and Footboard Identification see figure 3 1 on page 3 3 Time required 5 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required None Removal As the head and footboards are not locked they merely h
163. oot section from the 223 2 horizontal If thigh section in mid position 12 5 2 Maximum assisted foot section from the 14 42 horizontal If thigh section in mid position 5 5 42 Page 1 26 AvantGuard Electric Bed Service Manual QD1787 Specifications Chapter 1 Introduction Specification Sleep surface inclination from the horizontal trendelenburg reverse trendelenburg Dimension 16 16 Maximum height of siderail above sleep surface 395 mm 15 6 Maximum weight of bed without mattress or accessories 150 kg 331 Ib Safe working load 185 kg 408 Ib Maximum lifted load 250 kg 551 Ib add 50 mm for a bed featuring the hilow 45 to 85 cm option Table 1 6 Transport and Storage Environmental Conditions Condition Range Temperature Between 10 and 50 C Humidity 20 80 at 30 C Pressure 500 hPa to 1060 hPa 0 to 2000m altitude Table 1 7 Environmental Conditions for Use Condition Range Temperature Between 0 and 40 C Humidity 20 80 at 30 C Pressure 700 hPa to 1060 hPa 0 to 2000m altitude Electrical Description y SHOCK HAZARD The use of electrical equipment carries a potential risk of electric shock Personnel must be trained and instructed in the inherent risks of electrical equipment The electrical circuit is physically isolated from the metal parts o
164. or Characters OS is ieee oxy eo E ae A A 3 14 Management of the Actuators 0 0 0 cece eee eee eee 3 15 Detection of Bed not in Low Position 0 0 e cece ene 3 15 Automatic Contout 4 2 04 seas Gu A ae Re So ee ee 3 15 EasyChair Automatic Chair Function 0 0 0 0 ee eee eee ee 3 16 Emergency Trendelenburg Function 0 0 0 ee eee eee eee 3 16 Detection of Bed Connected to AC Power but Brakes not Applied Optional Depending on the Model 0 cee eee no 3 16 Trendelenburg Reverse Trendelenburg 0 0 0 0 cee ee eee 3 16 Chapter 4 Removal Replacement and Adjustment Procedures Bed Frame Hard Surface Replacement 23 06 6s 4 atada at 4 5 Removal opa e SES EEES SE 4 5 Replacement iers ars ras ata la 4 7 Control Pendant aiaa te Se Bae Al Re util e a o da rad Derbi 4 10 Removal 00 da a ee ee ise eee eee a Meee eee es 4 10 Replace Meira Baud cave sea dae eek Reh na ge eee 4 11 Control Panel on a Flexible Arm a aGeek 6 ee Ss OS Bee 4 12 Removal dados ad bed wot editar Bed a 4 12 Replacements tim tc ete de e O a E te ed dl Can E 4 13 AC Power Cable riire is cae ATA DE a RA A AI Re a 4 15 Removaliisc biaces alo dais tt heed Sees aa ROY 4 15 Replacements id NE A oe tet es 4 16 Power Supply Unit ra Loan ete oe eas baa oe ae eee 4 18 Removal e eon seu pa ra Eke Ge ES e tS Soo Lies 4 18 Page viii AvantGuard Electric Bed Service Manual QD1787 Table of Contents
165. ort crossbar fitted across the lower frame and an articulated mechanism that supports a 100mm diameter caster This mechanism contains two articulations The upper articulation enables the longitudinal indexing of the mobile subassembly This indexing is obtained by the pressure of two opposing cams around the vertical rotation axis The pressure results from the presence of a spring in the mobile part The mobile subassembly can be rotated as soon as a significant lateral force is exerted creating a torque at the level of the vertical rotation axis and compensating for the resulting force of the cams e The lower articulation of the mobile part combined with the gas spring enables doorway thresholds to be passed without incident NOTE If this 5th wheel steering system is installed on the bed the four casters are identical There is no steer caster fitted and the braking bar only has two postions neutral or braked Page 3 8 AvantGuard Electric Bed Service Manual QD1787 Mechanical System Chapter 3 Theory of Operation TotalGuard Integral Siderail Construction The siderail subassemblies comprise two head and foot sections in fire resistant extruded blow moulded polypropylene and a metal support of the malleable parallelogram type e Each head siderail integrates a double security release system e The siderail is lowered by sliding the unlocking button and slightly pushing the siderail inwards This last action en
166. ossibility of intervening with patient on the bed Yes Tools required 6 mm Allen key 1 Allen is a trademark of Industrial Fasteners Inc AvantGuard Electric Bed Service Manual QD1787 Page 7 7 7 2 Medical Device Support Rail Kit AD127A Chapter 7 Accessories Installation 1 The medical device support rails are located in the same sockets as the accessory support bars and are installed in the same manner 2 Position the two guides A in the spaces provided on the medical device support rail B and position the assembly below the holes to the right or left of the bed frame Ensure that the rod C of each medical device support rail is be located at the head end of the bed see figure 7 4 on page 7 8 3 Fit the two securing screws D of the medical device support rail and lock them using the 6 mm Allen key 4 Repeat steps 2 and 3 to install the other medical device support rail on the other side of the bed frame Figure 7 4 Installing the Medical Device Support Rails 1680_740 Removal Perform the installation procedure in reverse order 1 Allen is a trademark of Industrial Fasteners Inc Page 7 8 AvantGuard Electric Bed Service Manual QD1787 7 3 MobiBar Egress Bar Kit AD134A Chapter 7 Accessories 7 3 MobiBar Egress Bar Kit AD134A This kit contains the two egress bars and the parts that enable them to be installed on the bed The MobiBar
167. page 4 45 NOTE For example the photo shows an HL column with the 45 to 85 cm option Figure 4 24 Foot Hilow Column M 1680_408 13 Recover the BM assembly J optional AvantGuard Electric Bed Service Manual QD1787 Page 4 45 4 13 Foot Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures 14 Using the 13 mm ratchet wrench loosen and remove the four screws L on each side of the column and recover the grounding spring M 15 Slightly raise the bed frame foot section using the elevation tool tilt the column outwards then slide its upper section so as to remove it from the guide rails Remove the column from the bed Replacement y SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the installation is not correct and or if the earth continuity is compromised 1 Remove the two covers F situated at the base of the new column see figure 4 22 on page 4 44 2 Free the intermediate power supply cable from the column of the central rail A through the lower openings of this element so as to bring the connector B to one of the sides of the bed see figure 4 21 on page 4 43 3 Position the new HL column of the frame on the same side without mounting it and connect its connector to the intermediate connector B 4 Plug
168. pe eee 4 69 Siderail iie i045 taken Roti Moody bese bid ee heed Rees 4 70 Removal uan a peered ie ae ees Piece he bee E eked 4 70 Replacement sc i294 satin co Oe A At ee eon 4 71 Foot Sideral SUpp rt eS dente Sa 4 RSG AS ae sea 4 72 R moval a cease th a ye the kage 4 72 Replacement arog eh la cee A dee e el e ul Dek Goan 4 73 Head and Footboard mexica pa che ened vee a eee beaded ska 4 74 Removal cvc2csouttinigrdiectseen De eeeneew dead AEA eae 4 74 Replacements ren e A AR 4 74 Linen Holder Table optional sist a Qe abo haces 4 75 RemOyal ivomeatuvcweeiee tee eyes aye mess dE pd otk 4 75 Page x AvantGuard Electric Bed Service Manual QD1787 Table of Contents Replacement SS ed Soe ecg ews awn oat wea ean eA 4 75 Extension optional ovarios tias Rew dnt G aboe Peale fe 4 76 Romoyal ada eds FADE eR E dea o a AN Bd sa 4 76 Replacement ysis AAA AAA oe ee AAA 4 77 Egress Bat potasa dive ed pew teh tale id Bede 4 78 Removal eei Seen A eeu a eee a oe ee 4 78 Replacement aria hes Sheds died tae 4 79 Chapter 5 Parts List A se n A ad ene Oe pea He ed es 5 3 Ordering Service Parts lt iaun A o eee aes 5 4 Exchange PONCY e ee cesses A ape ced E S eee ee 5 6 In Warranty Exchanges ovigiciais ida ple ral a at e a 5 6 Out of Warranty Exchanges 0 0 eee eee eee ene 5 6 Recommended Spare Parts 4 62242454 te bee nace A eed adhe ones 5 7 Base Frame Rlevaion ss acan a A islas 5 8 Intermediate Plame 2 6 sew ceed
169. pendant LI158Ax AG0014 1 Control pendant LI159Ax AG0041 1 Control panel on a flexible arm LI158Ax AG0045 1 Control panel on a flexible arm LI159Ax AG0049 1 Lateral caregiver unit left without emergency trendelenberg option AG0051 1 Lateral caregiver unit left with emergency trendelenberg option AG0054 1 Lateral caregiver unit right without emergency trendelenberg option AG0055 1 Lateral caregiver unit right with emergency trendelenberg option AG0015 1 AC Power cable Europe grey AG0016 1 AC Power cable U K AG0017 1 AC Power cable Switzerland AG0048 2 Power supply unit AG0036 1 Hilow column 40 80 cm AG0037 1 Hilow column 45 85 cm AG0021 1 Head section actuator AG0022 1 Thigh section actuator AG0023 1 Foot section actuator AG0025 1 Mechanical foot section kit NOTE The spare parts listed above include the different versions of LI158Ax or LIIS9Ax beds and options In order to confirm the version of bed that you have check the identification label See Product Identification Label Location on page 5 4 AvantGuard Electric Bed Service Manual QD1 787 Page 5 7 Base Frame Elevation Chapter 5 Parts List Base Frame Elevation Figure 5 2 Base Frame Elevation 1680_501 Page 5 8 AvantGuard Electric Bed Service Manual QD1787 Base Frame Elevation Chapter 5 Parts List Table 5 2 Base Frame Elevation
170. peration of the BM from terminals 1 and 6 of the F Col see figure 2 3 on page 2 43 Figure 2 3 8 pin connector 1680_201 The column cabling and BM continuity is correct Yes No 4 Replace the F Col refer to procedure 4 13 on page 4 42 then go to Final Actions on page 2 12 6 Perform the following to verify the F Col extension a Remove the cover of the PSU using the 13mm combination spanner see figure 4 8 on page 4 16 b Disconnect connector 3 8 pin blue on the power board D see figure 4 10 on page 4 19 c The same verification applies to the terminals of connector 3 8 pin blue at the level of the PSU by checking the operation of the BM from terminals 1 and 6 see figure 2 3 on page 2 43 of the F Col extension connector The extension column cabling and switch continuity are correct Yes No y Replace the F Col and extension subassembly refer to procedure 4 13 on page 4 42 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 7 Replace the PSU refer to procedure 4 5 on page 4 18 This solves the problem Yes No y Contact your national Hill Rom representative see the last page of this manual 8 Goto Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 43 2 16 CPR Malfunction Chapter 2 Troubleshooting Procedu
171. ply Unit Cover 1680_426 4 Note the passage of the cables in the PSU and their connection on the main board D and the AC input terminal E see figure 4 10 on page 4 19 Figure 4 10 Power Supply Unit 1680_400 5 Free the AC power cable by unclipping it from the AC input terminal E 6 Disconnect the various connectors of the main board D of the PSU Mark the different connectors using the colour codes and markings on each connector see table 4 1 on page 4 20 AvantGuard Electric Bed Service Manual QD1787 Page 4 19 4 5 Power Supply Unit Chapter 4 Removal Replacement and Adjustment Procedures Table 4 1 Power Supply Unit Connections oe Marking Colour Identification 0 0 Red Not used la la Grey Right hand LCU lb 1b Grey Left hand LCU 2 2 Green Foot section actuator 3 3 Blue Foot section actuator 4 4 Yellow Foot hilow column 5 5 Brown Thigh section actuator 6 6 Yellow Head hilow column 7 Manually crimp the two earth wire terminal connectors H of the intermediate frame and the hilow column to the AC input terminal E and free the wires see figure 4 10 on page 4 19 NOTE The figure shows a bed with the 45 to 85 cm HL column for illustration only 8 If they are fitted disconnect the cables of the autonomy F and emergency trendelenburg function G batteries and remove them from the PSU 9 Remove the PSU from the bed Replacement 1
172. r NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions none 1 At least one of the other functions is working Yes No Go to RAP 2 1 on page 2 13 Perform the following to verify the proper operation of the remote control a Disconnect all the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AG0061 see Appendix on page 2 12 The functions are working Yes No y Go to step 4 Replace the CU control pendant refer to procedure 4 2 on page 4 10 or control Panel on a flexible arm refer to procedure 4 3 on page 4 12 according to the option Perform the following to verify the proper operation of the HS actuator a Press the HS raise without automatic contour button and raise the HS to the high position If this cannot be accomplished electrically disengage the HS actuator using the CPR handle and raise the HS manually 2 people required b Remove the cover of the PSU using the 13 mm combination spanner see figure 4 8 on page 4 16 c Either exchange the HS actuator with a new actuator or connect a HS actuator simulation connector P N AG0064 see Appendix on page 2 12 d Press the HS raise then lower The HS function works Yes No y The HS actuator is defective Replace it
173. r head end of the bed depending on the configuration In order to engage the steering position raise the brake and steer bar with your foot until it is tilting upwards To disengage the steering position step on the brake and steer bar until it is parallel with the floor The bar remains slightly tilted downwards when the bar is released Headboards and Footboards The AvantGuard Electric Bed is equipped with shaped endboards at the head and foot of the bed They can be removed vertically and aid the mobility and steering of the bed by the use of integrated handles They have slots which allow them to be quickly inserted into plastic brackets placed on bars at the head and foot ends The head and foot boards are removed by pulling them upwards to lift them from the sockets at the head and foot end of the bed Hard Surfaces on the Head Seat Thigh and Foot Sections To remove the hard surfaces from the FS and TS hold each section individually on each side of the sleep surface at the restraining strap attachment points then remove them by pulling upward To install the hard surfaces of the FS and TS line them up on each side of the sleep surface with respect to the restraining strap attachment points then fit them in the relevant positions To remove and install the hard sections of the SS and HS refer to procedure 4 1 Casters The AvantGuard Electric Bed is equipped with four swivelling casters situated at the four
174. re 4 3 on page 4 12 according to the option fitted This solves the problem Yes No Contact your national Hill Rom representative see the last page of this manual 11 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1787 Page 2 35 2 12 Trendelenburg Reverse Trendelenburg Malfunction Chapter 2 Troubleshooting Procedures 2 12 Trendelenburg Reverse Trendelenburg Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions the HL function works see Function Checks on page 2 6 1 At least one of the other functions is working Yes No Go to RAP 2 1 on page 2 13 Perform the following to verify the proper operation of the control unit a Disconnect the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AG0061 see Appendix on page 2 12 The trendelenburg reverse trendelenburg function works Yes No a The PSU is defective Replace it refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual Replace the CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexible arm refer to procedure 4 3 o
175. refer to procedure 4 8 on page 4 27 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual Page 2 24 AvantGuard Electric Bed Service Manual QD1787 2 6 Head Section Malfunction excluding Automatic Contour Chapter 2 Troubleshooting Procedures 5 The PSU is defective Replace it refer to procedure 4 5 on page 4 18 This solves the problem Yes No y Contact your national Hill Rom representative see the last page of this manual 6 Goto Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 25 2 7 Automatic Contour Malfunction Chapter 2 Troubleshooting Procedures 2 7 Automatic Contour Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions a The HS function without automatic contour works see Function Checks on page 2 6 b The TS function works see Function Checks on page 2 6 NOTE The TS only lowers after pressing the HS lower for the Ist time To make it lower even further you must first reset the function by raising the HS 1 At least one of the other functions is working Yes No Go to RAP 2 1 on page 2 13 2 Perform the following to ver
176. res 2 16 CPR Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE Initial conditions the HL function works see Function Checks on page 2 6 1 Check the mechanics of the CPR system Remove the hard surfaces of the SS the TS and the FS refer to procedure 4 1 on page 4 5 Check that the screws or pins are not damaged or missing from the release handles or release mechanism of the HS actuator refer to the relevant exploded view of chapter 5 All the pins and screws are secure and undamaged Yes No y Replace the missing or damaged parts then go to Function Checks on page 2 6 Check that the release cable is not broken The cable is intact Yes No y Go to step 4 Check that the adjustment of the release cable is correct refer to procedure 4 9 on page 4 30 The adjustment is correct Yes No i Adjust the release cable refer to procedure 4 9 on page 4 30 Replace the HS actuator refer to procedure 4 8 on page 4 27 and release cable subassembly refer to procedure 4 9 on page 4 30 This solves the problem Yes No Contact your national Hill Rom representative see the last page of this manual Go to Final Actions on page 2 12 Page 2 44 AvantGuard Electric Bed Service Manual QD1787 2 17 Foot Siderail Locking Unlocking Malfunction Chapter 2 Troubleshooting Procedures 2 17 Foot Siderail Locking Unlocking Malfunction
177. rial number shown on the product identification label A see figure 5 1 on page 5 4 Figure 5 1 Product Identification Label Location 1787_500 A WARNING Do not remove the identification label from the bed the references featuring on it are essential for identifying the bed and guaranteeing the safety of the maintenance operations Failure to observe this warning could result in personal injury Call the after sales service of your national Hill Rom representative refer to the last page of this manual To help expedite the processing of your parts order please have the following elements ready when you call e Customer account number e Purchase order number if available e Product code e Serial number of the bed e Part reference number s Page 5 4 AvantGuard Electric Bed Service Manual QD1787 Ordering Service Parts Chapter 5 Parts List For your convenience a fax number is available for ordering parts requesting prices or chasing up an order refer to the last page of this manual Postage will be invoiced for any orders less than 30 Euros Delivery within France e Standard delivery time 48 hours e We can also provide an EXPRESS delivery time of 24 hours e Postage paid dispatch e Dispatch charges are included in the price of the parts for a standard delivery time within 48 hours in France e Parts to be manufactured have a delay of 3 weeks Please address all orders to your national
178. ric Bed Service Manual QD1787 Page 2 27 2 7 Automatic Contour Malfunction Chapter 2 Troubleshooting Procedures The TS section raises or lowers until mid travel then stops Yes No y gt Goto step 14 13 The problem is due to incorrect usage 14 Switching the connectors of the TS FS actuators Switch the connectors of the TS and FS actuators 2 green and 5 brown a Press the FS raise button G the rod of the TS actuator extends fully b Press the TS lower button F the rod of the FS actuator retracts fully c Press the HS raise button A The FS actuator starts up and its rod stops after approximately 15 mm of travel Yes No Replace the PSU refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 15 Replace the TS actuator refer to procedure 4 10 on page 4 33 This solves the problem Yes No Contact your national Hill Rom representative see the last page of this manual 16 Go to Final Actions on page 2 12 Page 2 28 AvantGuard Electric Bed Service Manual QD1787 2 8 Battery Backup Malfunction Chapter 2 Troubleshooting Procedures 2 8 Battery Backup Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see fi
179. rm refer to procedure 4 3 on page 4 12 according to the option Perform the following to verify the proper operation of the TS actuator a Remove the cover of the PSU using the 13 mm combination spanner see figure 4 9 on page 4 19 b Cross connect the TS and FS actuator connectors 5 brown amp 2 green or connect new TS actuator or connect a TS actuator simulation connector P N AG0063 see Appendix on page 2 12 c Press the FS keys raise G then lower H see figure 1 11 on page 1 14 The TS function works The actuator rod retracts and opens correctly Yes No The TS actuator is defective Replace it refer to procedure 4 10 on page 4 33 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual The PSU is defective Replace it refer to procedure 4 5 on page 4 18 This solves the problem Page 2 22 AvantGuard Electric Bed Service Manual QD1787 2 5 Thigh Section Malfunction Chapter 2 Troubleshooting Procedures Yes No y Contact your national Hill Rom representative see the last page of this manual 7 Go to Final Actions on page 2 12 AvantGuard Electric Bed Service Manual QD1 787 Page 2 23 2 6 Head Section Malfunction excluding Automatic Contour Chapter 2 Troubleshooting Procedures 2 6 Head Section Malfunction excluding Automatic Contou
180. rol 2 people required AvantGuard Electric Bed Service Manual QD1 787 Page 2 13 2 1 No Functions Work Chapter 2 Troubleshooting Procedures W SHOCK HAZARD Electric current flows through the connections when the bed is switched on Trials conducted under these conditions carry a risk of electric shock Do not touch any wire or terminal which conducts an electric current a Remove the PSU cover using the 13 mm combination spanner see figure 4 9 on page 4 19 b Plug the unit into a power outlet c On the AC power input connector E see figure 4 10 on page 4 19 check for the presence of the AC voltage using a DMM set to AC volts see figure 2 1 on page 2 15 The voltage is correct Yes No Replace the AC power cable refer to procedure 4 4 on page 4 15 5 Verifiy that the remote control functions properly Connect the test CU P N AG0061 see Appendix on page 2 12 The functions are operational Yes No y Replace the PSU refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 6 Goto RAP 2 2 on page 2 16 Page 2 14 AvantGuard Electric Bed Service Manual QD1787 2 1 No Functions Work Chapter 2 Troubleshooting Procedures Figure 2 1 Measuring the AC Input Voltage 1680_200 AvantGuard Electric Bed Service Manual QD1787
181. roups the various signals NOTE The LCU is only available as an assembly which includes the wiring harness switches and connections Page 3 14 AvantGuard Electric Bed Service Manual QD1787 Electrical System Motor Characteristics The motors have the following characteristics Chapter 3 Theory of Operation Table 3 5 Motor Characteristics and Dimensions Description oot HS motor TS motor FS motor motor Power 96 W x2 120 W 84 W 84 W Voltage 18 to 40VDC 18 to 40VDC 18 to 40VDC 18 to 40VDC Intermittent 10 10 10 10 service 3min 30min 2min 18min 6min 60min 6min 60min Actuator 400 mm 252 mm 60 mm 45 mm column travel 5mm 15 75 2mm 9 91 2mm 2 36 2mm 1 77 Retracted rod 458 mm 310 mm 325 mm centre distance 2mm 18 03 2mm 12 2 2mm 12 8 Management of the Actuators The actuators are supplied by the power supply unit Their extreme positions are detected by upper and lower internal EOT devices in series These devices directly cut off the power supply for the head section thigh section and foot section actuators In addition the head section and foot section actuators have intermediate internal microswitches in order to control the automatic contour and chair position The two hilow columns also have upper and lower internal EOT devices For safety reasons they are both backed up at each end If one of these safety EOTs is activated
182. rry out the movement in one stage The bed moves gradually to trendelenburg while the sections of the bed frame move to the horizontal position Yes No a Go to RAP 2 14 on page 2 39 26 Plug the bed into an appropriate power source If the bed is fitted with a detection of bed connected to AC power but brakes not applied optional for LI158Ax set the brake bar to neutral AvantGuard Electric Bed Service Manual QD1787 Page 2 9 Function Checks Chapter 2 Troubleshooting Procedures 27 28 An audible alarm can be heard Yes No 4 Goto RAP 2 15 on page 2 42 Set the bar to brake position The audible alarm stops sounding Yes No 4 Goto RAP 2 15 on page 2 42 Press the HS raise button until the section reaches high position NOTE Someone should be lying on the bed so that this test can be carried out conclusively 29 30 31 32 Pull each of the CPR head section emergency release handles in turn The HS descends quickly up to the mid travel then the movement is cushioned until the low position is reached The movement occurs without any problem or abnormal noise Yes No y gt Go to RAP 2 16 on page 2 44 Release the CPR handles and operate the HS raise on the control unit to check that the HS actuator drive mechanism is working correctly The HS rises without any problem or abnormal noise Yes No Y Goto RAP 2 16 on page 2 44 Unfold the siderails Check that they lock
183. s F of the cable s exiting from the base of the Head column 1 cable for the BM plus 2 HL pedal cables optional and disconnect the relevant cable s see figure 4 22 on page 4 44 AvantGuard Electric Bed Service Manual QD1787 Page 4 43 4 13 Foot Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 22 Foot Hilow Column Base 1680_432 9 Open the cable tracks situated under the foot crossbar of the frame and free the cables sl CAUGHT HAZARD The elevation tool must be used to remove the column in safety Use the elevation tool otherwise personal injury could occur 10 Position the elevation tool and adjust its height so that the arm G supports the foot crossbar H of the bed frame see figure 4 23 on page 4 45 sl CAUGHT HAZARD The bed frame must be secured lengthways to prevent it coming away from the rail at the level of the H Col Disregarding this precaution may cause injuries 11 Ensure that the frame is correctly locked lengthways by the arm G of the elevation tool which secures it at the level of the crossbar through its U shape Page 4 44 AvantGuard Electric Bed Service Manual QD1787 4 13 Foot Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 23 Foot Section Support 1680_443 12 Using the 13 mm ratchet wrench loosen and remove the 4 screws I securing the base of the HL column see figure 4 24 on
184. s The bed must not move Steer Caster There is no specific adjustment as regards the steering system of the casters only the position of these casters on installation and braking are to be checked Please refer to the replacement and adjustment sections of this procedure 1 Loctite is a registered trademark of Loctite Corporation Inc Page 4 60 AvantGuard Electric Bed Service Manual QD1787 4 18 5th Wheel Steering System Chapter 4 Removal Replacement and Adjustment Procedures 4 18 5th Wheel Steering System Identification see figure 3 1 on page 3 3 Time required 20 min Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench 13 mm combination spanner Elevation tool P N AG0052 Removal 1 Apply the brakes raise the sleep surface to mid height the HS to the high position then the TS and the FS to horizontal Raise and lock the siderails and remove the headboard e SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this 1s not the case wait for about one minute until the battery mode is deactivated Press a function button of the control pendant not the battery button to ensure
185. s Work se eien na at Shae De Bee caw ae 2 13 Control Unit Malunchon s ario oye pew atte beens Sees tees 2 16 Function Management Mode Malfunction 0 0 0 0 e ee eee eee 2 18 Hilow Malfunction G27 eaa aa tg SANT ee alld a e e 2 20 Thigh Section Malfunction siii sees esate ees eet ea vee eee i 2 22 Head Section Malfunction excluding Automatic Contour 2 24 Automatic Contour Malfunction 2 0 0 cece cc eee nee 2 26 Page vi AvantGuard Electric Bed Service Manual QD1787 Table of Contents Battery Backup Malfunction ri oa RSS Eee ee 2 29 Foot Section Malfunction eri rias toi rei 2 31 Assisted Foot Section Malfunction 0 0 0 eee eee ee eee 2 34 EasyChair Automatic Chair Position Malfunction 04 2 35 Trendelenburg Reverse Trendelenburg Malfunction 04 2 36 Emergency Trendelenburg Function not Working with Bed Connected to AC POW Sr ii A ADA AA SA da 2 37 Emergency Trendelenburg Function not Working with the Bed not Connected to ACG PowWeb iriri gerent A ewe aes ees 2 39 Bed Connected to AC Power Brakes not Applied Detection Malfunction 2 42 CPR Malfunction ta se walter ae x wee da oy pat eee 2 44 Foot Siderail Locking Unlocking Malfunction oooooocoocooocooooooo 2 45 Head Siderail Locking Unlocking Malfunction 00000 2 46 Braking Malfunction 00 popa dd es 2 47 Steering Malf nction errn er ride Dt te it
186. s case the overload is indicated by an audible alarm This overload indication also enables failure of a actuator end of travel to be detected except for the HL columns which have a backup end of travel AvantGuard Electric Bed Service Manual QD1787 Page 3 11 Electrical System Chapter 3 Theory of Operation Battery Circuit The bed can be fitted with a battery backup optional and or an emergency trendelenburg function battery backup optional These batteries are internal to the power supply unit The emergency trendelenburg function battery is only available for this specific function The common charging circuit situated in the power supply unit has current limiting according to the level of the battery charge Charging begins as soon as the unit is switched on At the end of charging the load current is very low When the PSU is disconnected from the AC power battery operation is activated by the Battery Activation on the control pendant The battery charge level is controlled In this way a flashing LED and an audible low battery alarm by buzzer are activated during movement if the battery voltage drops this is a continuous beep of several seconds that is heard each time a button is pressed After reconnecting to the power outlet this LED continues to flash until the battery has recharged Battery operation is automatically inhibited 30 seconds after the end of the last movement Battery Characterist
187. s screwed flat against the attachment plate 8 Remove the attachment plate C NOTE The attachment plate is delivered with the gas spring Replacement 1 Perform the removal procedure in reverse order 2 Perform the Function Checks on page 2 6 Page 4 64 AvantGuard Electric Bed Service Manual QD1787 4 20 Gas Spring Subassembly with Cable and Foot Section Sleeve LI158Ax only Chapter 4 Removal Replacement and Adjustment Procedures 4 20 Gas Spring Subassembly with Cable and Foot Section Sleeve LI158Ax only Identification see figure 3 1 on page 3 3 NOTE Only the full subassembly is available as a spare part It is supplied ready assembled and adjusted Time required 20 min Number of people 1 Possibility of intervening with patient on the bed No Tools required T25 Torx key NOTE The bed has only one release lever located on the right hand side from the view point of a patient lying on the bed Removal 1 Apply the brakes raise the sleep surface to the high position and lower the sleep surface sections to horizontal Fold the foot siderails W SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 2 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this is not the c
188. see figure 4 11 on page 4 23 6 Ifthe emergency trendelenburg battery is being replaced each battery has a specific cable with special connector disconnect the connectors of both batteries and lift them away Page 4 22 AvantGuard Electric Bed Service Manual QD1 787 4 6 Battery Backup and Emergency Trendelenburg Function Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 11 Battery Backup Connections ee 1680_427 Replacement 1 Perform the removal in reverse order while ensuring that the correct polarities are used when connecting new batteries NOTE To prevent connection errors the connections of the battery backup and the emergency trendelenburg function battery are different NOTE When connecting the batteries the red wire must be connected last A spark is produced when it is connected but is not dangerous for the power supply P C board or the technical personnel A CAUTION The polarities must be respected when installing the elements of the battery backup Disregarding this caution will permanently damage the power supply P C board of the bed 2 Disconnect the bed from the power outlet then perform the following tests according to the batteries fitted on the bed Battery backup after full charge minimum of 3 HL HS TS and FS LI159Ax full raise and lower cycles Emergency trendelenburg battery minimum of 3 full trendelenburg cycles from the fu
189. separate the adjuster E from its support 6 Using the two 10 mm combination spanners loosen the nut of the sleeving stop F and free the CPR cable from its guide on the actuator see figure 4 14 on page 4 28 7 With the adjuster E removed from its support pull on one of the CPR handles and remove the end of the cable from the cable stop K on the CPR handle with the small screwdriver then repeat on the other CPR handle see figure 4 16 on page 4 31 Figure 4 16 Fitting the CPR Cable 1680_440 8 Remove the two ends of the CPR cable from the bed frame J while noting their position with respect to the pivot point I see figure 4 15 on page 4 30 Replacement 1 Perform the removal procedure in reverse order 2 Return to the following adjustment procedure in the event of incorrect operation 3 Perform the Function Checks on page 2 6 Adjustment 1 Using the two 10 mm combination spanners loosen the adjuster nuts E 2 Apply a gentle pull on the cable at the adjuster E to compensate for the play at the level of the CPR handles and the HS actuator Manually tighten the adjuster nuts E AvantGuard Electric Bed Service Manual QD1787 Page 4 31 4 9 Head Section Release Cable Chapter 4 Removal Replacement and Adjustment Procedures NOTE There should not be any tension on the cable merely compensation for the play at the level of the CPR handles and the HS actuator 3 Lock the
190. sert the new articulation bushes H into the openings of the link bar G 2 Perform steps 9 to 11 of the removal procedure in reverse order 3 Set the single brake bar to the brake position and check that the four casters brake 4 If the casters are not braked reset the single brake bar to neutral remove the caster socket rotate the caster rod by half a turn then refit the socket and the hexagonal bar 5 Perform steps to 7 of the removal procedure in reverse order 6 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 55 4 17 Caster Brake and Steer Caster Chapter 4 Removal Replacement and Adjustment Procedures 4 17 Caster Brake and Steer Caster Identification see figure 3 1 on page 3 3 Time required 15 min without adjustment Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench 10 mm Allen key Soft faced hammer Medium screwdriver Elevation tool P N AG0052 NOTE When ordering a caster from a Hill Rom spare parts service remember that the bed may be fitted with the antistatic caster and or foot steering caster options see Braking on page 3 8 When the bed is fitted with the 5th wheel steering system see Casters on page 3 8 the braking bar is limited by a stop B and will have only two positions neutral or brake activated Take this information into account wh
191. sert the two support pins F at the head end of the bed with respect to the guide rails J see figure 4 5 on page 4 8 c Pass your hand through the openings G of the hard surface and fit the 2 securing pins F see figure 4 3 on page 4 7 d Fit the 4 Truarc circlips fit the two spacers to the lower pins on pins E and F securing the hard surface D to the HS see figure 4 2 on page 4 6 A WARNING Two persons are required to test the articulation of the head section Disregarding this precaution may result in injuries Page 4 8 AvantGuard Electric Bed Service Manual QD1787 4 1 Bed Frame Hard Surface Replacement Chapter 4 Removal Replacement and Adjustment Procedures 3 Grip the end handles of the HS hard surface operate one of the CPR handles and raise lower the HS to check that the hard surface moves without interference 4 Perform the Function Checks on page 2 6 AvantGuard Electric Bed Service Manual QD1787 Page 4 9 4 2 Control Pendant Chapter 4 Removal Replacement and Adjustment Procedures 4 2 Control Pendant Identification see figure 3 1 on page 3 3 Time required 5 min Number of people 1 Possibility of intervening with patient on the bed Yes Tools required None Removal 1 Identify the version of the control pendant to be replaced LI158Ax or LI159Ax to ensure that the replacement will allow the same functions If possible raise the HS then r
192. ssociated with the AvantGuard Electric Bed For a list of acronyms see table 1 1 on page 1 6 Table 1 1 Acronym List Acronym Definition CPR Emergency release of head section for Cardio Pulmonary Resuscitation CU Control Unit control pendant or control panel on a flexible arm DMM Digital Multimeter EOT End of Travel F Col Foot Hilow Column BM Brake Microswitch gly Green Yellow H Col Head Hilow Column HL Hilow HS Head Section HSA Head Section Adjustment IVP I V Pole LCU Lateral Caregiver Unit LED Light Emitting Diode FS Foot Section NC Normally Closed NO Normally Open PAT Portable Appliance Tester P N Part Number PSU Power Supply Unit RAP Repair Analysis Procedure RHS Retractable Head Section S A Sub Assembly SS Seat Section TP Test Point Tr Rev Tr Trendelenburg Reverse Trendelenburg TS Thigh Section VAC Volts Alternating Current VDC Volts Direct Current Page 1 6 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction Introduction Overview The AvantGuard Electric Bed comprises e Three adjustable sections head thigh and foot e Height adjustment of the sleep surface level e Inclination of the bed frame in trendelenburg head down or reverse trendelenburg feet down The AvantGuard Electric Bed has a dual control system Functions
193. t reaches the high position Next press the FS function raise button The FS rises to the high position without stopping or the audible alarm sounding during the movement Yes No Ji Go to RAP 2 9 on page 2 31 If the bed is fitted with a power driven FS press the lower button of the FS function The FS drops to the low position without stopping or the audible alarm sounding during the movement Yes No Go to RAP 2 9 on page 2 31 If the bed is fitted with an assisted FS operate the control lever of the gas spring and raise the section to high position Release the lever The FS rises without hindrance and remains in the high position Yes No y gt Goto RAP 2 10 on page 2 34 If the bed is fitted with an assisted FS operate the control lever of the gas spring and lower the section to low position Release the lever The FS drops without hindrance and remains locked when the lever is released Yes No Go to RAP 2 10 on page 2 34 Adjust the TS to horizontal If the bed is fitted with a chair function press the chair position function of the bed The HS and TS sections rise as long as the button is pressed until the required position is reached and the HS lowers when the TS reaches an angle approaching 17 Page 2 8 AvantGuard Electric Bed Service Manual QD1787 Function Checks Chapter 2 Troubleshooting Procedures Yes No sl Go to RAP 2 11 on page 2 35 21 If the bed is fitt
194. t Positioning and Electrical Diagrams LI158Ax LI159Ax Chapter 3 Theory of Operation Component Positioning and Electrical Diagrams LI158Ax LI159Ax Figure 3 1 Component Positioning LI158Ax LI159Ax Control panel F Control pendant on a flexible i arm Footboard Linen holder not visible Foot siderail support ig Headboard Hilow columns Lateral caregiver unit Foot end frame cover Head end frame Braking bar cover 1680_303 Double band caster diameter 150 mm Hilow pedal 5th wheel steering system not shown Medical equipment support rail Head siderail Foot siderail Head section CPR cable damping device Extension i Head section actuator Head hilow column Foot hilow column AC Power Assisted foot cable f section control not shown Braking detection 1680_304 switch Power supply Thigh section Accessory Foot section Egress bar unit actuator support bar actuator not shown AvantGuard Electric Bed Service Manual QD1787 Page 3 3 Component Positioning and Electrical Diagrams LI158Ax LI159Ax Chapter 3 Theory of Operation Figure 3 2 Wiring Schematic for LI158Ax Management mode Emergency trendelenburg optional Battery indicator Emergency trendelenburg optional Management Battery indicator mode CU M e Control ontro dant panel on a ee flexible arm optional HL
195. t bars are available under the central section of the sleep surface Their position and movements are identical to the accessory suport bars AvantGuard Electric Bed Service Manual QD1787 Page 1 25 Specifications Chapter 1 Introduction Specifications Physical Description For the AvantGuard Electric Bed specifications see table 1 5 on page 1 26 Table 1 5 AvantGuard Electric Bed Specifications Specification Overall length in high position basic version Overall length in high position with extension optional in use Dimension 2185 mm 5 mm 86 2365 mm 5 mm 93 1 Overall width siderails up 1000 mm 39 4 Overall width siderails down 940 mm 39 0 Height of sleep surface low position depending on caster model 400 mm 5 mm 15 8 double band 415 mm 5 mm 16 3 single band Height of sleep surface high position depending on caster model 800 mm 0 10 mm 31 5 double band 815 mm 0 10 mm 32 1 single band Maximum space under bed bed with 150 mm 5 9 hilow pedal double band casters 163 mm 6 4 single band casters Maximum space under bed bed without 187 mm 7 4 hilow pedal double band casters 200 mm 7 9 single band casters Maximum head section inclination 68 Thigh section inclination automatic 20 42 contour Maximum thigh section inclination 36 12 Maximum electric f
196. t page of this manual Go to Final Actions on page 2 12 Page 2 34 AvantGuard Electric Bed Service Manual QD1787 2 11 EasyChair Automatic Chair Position Malfunction Chapter 2 Troubleshooting Procedures 2 11 EasyChair Automatic Chair Position Malfunction NOTE Identification of components see figure 3 1 on page 3 3 NOTE To access the PSU see figure 4 9 on page 4 19 and connections see figure 4 10 on page 4 19 NOTE Initial conditions The TS function works see Function Checks on page 2 6 b The FS function works see Function Checks on page 2 6 c The HS with automatic contour function works see Function Checks on page 2 6 7 At least one of the other functions is working Yes No Go to RAP 2 1 on page 2 13 8 Perform the following to verify the proper operation of the control unit a Disconnect the CU s control pendant or control panel on a flexible arm b Connect the test CU P N AG0061 see Appendix on page 2 12 9 The functions are working Yes No The PSU is defective Replace it refer to procedure 4 5 on page 4 18 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual 10 Replace the control unit CU control pendant refer to procedure 4 2 on page 4 10 or control panel on a flexible arm refer to procedu
197. talling the Control Panel on a Flexible Arm 1680_403 6 Remove the hard surfaces of the SS and the TS refer to procedure 4 1 7 Remove the cable cover C of the central rail of the bed frame and free the cable from the control panel on a flexible arm Note that the cable must always be fitted with as much play as possible at the level of the HS articulation see figure 4 7 on page 4 13 8 Trace the cable to the correct LCU and disconnect the connector A from the rear of the LCU see figure 4 6 on page 4 11 9 Open the cable track located under the crossbar B of the bed frame and free the cable of the control panel on a flexible arm 10 Using the 2 phillips head screwdriver loosen and remove the screws from the two supports E Remove the supports see figure 4 7 on page 4 13 11 Remove the connector of the control panel on a flexible arm from the notch J by pushing gently underneath the HS hard surface so as to move the HS frame 12 Unscrew the knob screw F securing the control panel on a flexible arm and remove it from the HS hard surface Replacement 1 Identify the version of the control panel on a flexible arm to be replaced LI58Ax or LI159Ax to ensure that the replacement will allow the same functions 2 Perform the removal procedure in reverse order AvantGuard Electric Bed Service Manual QD1787 Page 4 13 4 3 Control Panel on a Flexible Arm Chapter 4 Removal Replacement and A
198. ter refer to procedure 4 17 on page 4 56 If this solves the problem perform the Final Actions on page 2 12 otherwise contact your national Hill Rom representative see the last page of this manual The problem is solved Yes No y Contact your national Hill Rom representative see the last page of this manual Go to Final Actions on page 2 12 Page 2 48 AvantGuard Electric Bed Service Manual QD1787 2 21 5th Wheel Steering Malfunction Chapter 2 Troubleshooting Procedures 2 21 5th Wheel Steering Malfunction NOTE Identification of components see figure 3 1 on page 3 3 Initial conditions none 1 Check the attachment of the 5th wheel Check that the screws or supports are not damaged loose or missing from the support Sth wheel and caster attachments refer to the assembly diagrams of chapter 5 All supports and screws are secure and undamaged Yes No y Replace the missing or damaged parts then go to Function Checks on page 2 6 2 Check the 5th wheel Push the bed lengthways and check that the articulated arm supporting the 5th wheel locks in the lengthways direction of the bed The articulated arm locks in the lengthways direction Yes No Go to step 7 3 Move the bed several metres and check that the 5th wheel remains in contact with the floor The 5th wheel remains in contact with the floor Yes No sh Go to step 7 4 Push the bed sidew
199. that the battery mode is properly deactivated A WARNING Never lift the entire bed with the elevation tool lts maximum load is 150kg Disregarding this warning may cause the elevation mechanism to malfunction and could cause injury 3 Position the elevation tool under the head crossbar of the bed and adjust it so as to raise the casters by a few centimetres see figure 4 29 on page 4 57 4 Using the 13 mm combination spanner and 13 mm ratchet wrench loosen and remove the two bolts A securing the 5th wheel steering system and remove the assembly Recover the two bolts A see figure 4 31 on page 4 62 AvantGuard Electric Bed Service Manual QD1787 Page 4 61 4 18 5th Wheel Steering System Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 31 Installing the 5th Wheel Steering System 1787_433 Replacement 1 Perform the removal procedure in reverse order 2 Perform the Function Checks on page 2 6 Page 4 62 AvantGuard Electric Bed Service Manual QD1 787 4 19 Head Section Gas Spring Chapter 4 Removal Replacement and Adjustment Procedures 4 19 Head Section Gas Spring Identification see figure 3 1 on page 3 3 Time required 10 min per gas spring Number of people 1 Possibility of intervening with patient on the bed No Tools required 13 mm ratchet wrench 13 mm combination spanner NOTE The bed has two HS gas springs It is advisable to replace both of them
200. the bed is powered Disconnect from the power outlet otherwise personal injury could occur 5 Lock out all the functions Unplug the power connector from the power outlet If the bed has a battery ensure that no LED is lit on the control pendant If this 1s not the case wait for approximately one minute until the battery mode is deactivated 6 Perform the removal procedure in reverse order NOTE Leave as much play as possible at the level of the H Col between this column and the PSU 7 Refit the two covers F located at the base of the column see figure 4 22 on page 4 44 8 Operate the trendelenburg function of the bed to ensure that there is no tension on the power supply cable of the column during operation If necessary adjust the play of the cable with respect to the PSU Page 4 40 AvantGuard Electric Bed Service Manual QD1 787 4 12 Head Hilow Column Chapter 4 Removal Replacement and Adjustment Procedures W SHOCK HAZARD The equipotential link and the grounding of the metal parts of the bed depend on the correct installation of the earth conductor and grounding spring blades Personal injury could result if the installation is not correct and or if the earth continuity is compromised 9 Check that the earth continuity of the accessible metal parts is correct see Earth resistance on page 6 7 in the Preventive Maintenance Schedule 10 Perform the Function Checks on page 2 6
201. the column cannot be used The other functions remain operational WARNING The column must be replaced if it becomes defective The column must never be dismantled Failure to observe this warning could result in personal injury Complementing these EOTs the overload limiting function of the PSU provides protection in the event of mechanical failures obstacles or overload Detection of Bed not in Low Position This type of detection is obtained by taking into account the state of the low EOT of the HL columns The power supply unit controls the corresponding LED of the control unit as soon as one of the two EOTs is released AvantGuard Electric Bed Service Manual QD 1787 Page 3 15 Electrical System Chapter 3 Theory of Operation Automatic Contour The PSU when it received a command from the CU controls the automatic contour HS and TS operation of the bed This function is carried out through an intermediate switch of the TS actuator When the HS raise function is activated from the horizontal position the HS and TS rise immediately until the intermediate internal microswitch of the TS actuator cuts its power at approximately 20 The HS actuator continues to work until its upper EOT is reached or when the control is released When the head section lower function is activated and the corresponding section reaches the low position the TS actuator is controlled for lowering until its lower EOT is reached or th
202. ther egress bar on the other side of the bed frame Removal Perform the installation procedure in reverse order Page 7 10 AvantGuard Electric Bed Service Manual QD1787 7 4 Control Panel on a Flexible Arm Kit AD136A Chapter 7 Accessories 7 4 Control Panel on a Flexible Arm Kit AD136A This kit contains the control panel on a flexible arm and the parts that enable it to be installed on the bed The control panel on a flexible arm kit includes the following e One contol panel on a flexible arm e One knob securing screw e Two cable plastic sleeve supports e Four cable sleeve securing screws Figure 7 7 Control Panel on a Flexible Arm 1680_744 Identification see figure 3 1 on page 3 3 Time required 15 min Number of people 1 plus 1 occasionally Possibility of intervening with patient on the bed No Tools required Medium phillips head screwdriver Installation 1 Apply the brakes adjust the bed to high position the sleep surface horizontal 2 Adjust the head section to high position the siderails to high position and remove the headboard AvantGuard Electric Bed Service Manual QD1787 Page 7 11 7 4 Control Panel on a Flexible Arm Kit AD136A Chapter 7 Accessories al SHOCK HAZARD Electric current flows through the connections when the bed is powered Disconnect from the power outlet otherwise personal injury could occur 3 Unplu
203. thod is to limit as much as possible the number of spare parts necessary and the cost incurred by a breakdown analysis NOTE It is recommended that you lock out the power driven functions not concerned by the repair analysis before beginning or unplug the bed if the repair is not connected to a power driven function In this case also lock out all functions if the bed has a battery NOTE Refer to table 1 1 on page 1 6 for all acronyms used in the following chapters AvantGuard Electric Bed Service Manual QD1787 Page 2 3 Initial Actions Chapter 2 Troubleshooting Procedures Initial Actions Use Initial Actions to gather information from operators concerning problems with the AvantGuard Electric Bed Note the symptoms or other information concerning the problem that the operator describes This information helps identify the probable cause 1 Someone who can explain the problem is available Yes No Go to Function Checks on page 2 6 2 Ask this person to demonstrate or explain the problem The problem can be duplicated Yes No a Go to Function Checks on page 2 6 3 The problem is a result of improper operator action Yes No y Refer to table 2 1 on page 2 4 or go to Function Checks on page 2 6 4 Instruct the operator to refer to the procedures in the AvantGuard Electric Bed Perform the Function Checks to ensure proper operation of the AvantGuard Electr
204. trol pendant not the battery button to ensure that the battery mode is properly deactivated 4 Using the 13 mm ratchet wrench loosen and remove the two screws A securing the cover B on the PSU recover the two screws and remove the cover B by unclipping it see figure 4 9 on page 4 19 5 On the main board D of the PSU disconnect the connector con2 of the FS actuator see table 4 1 on page 4 20 see figure 4 10 on page 4 19 6 Lift the cable cover C on the central rail of the sleep surface and remove the cable of the FS actuator see figure 4 9 on page 4 19 7 Support the FS and remove the Truarc circlips from the pin at the end of the FS actuator E see figure 4 18 on page 4 36 1 Truarc is a registered trademark of Waldes Truarc Inc AvantGuard Electric Bed Service Manual QD1787 Page 4 35 4 11 Foot Section Actuator Replacement LI159Ax only Chapter 4 Removal Replacement and Adjustment Procedures Figure 4 18 Foot Section Actuator 0 B 5 1680_407 8 Remove the pin while supporting the FS and the FS actuator 9 Remove the Truarc circlips from the pin supporting the FS actuator body and remove the pin from the actuator 10 Lay the FS on the intermediate frame or on a support that will maintain it in a horizontal position Replacement 1 Perform the removal procedure in the reverse order A CAUTION Ensure that the elastic found on the new actuators is
205. unit and the control key Emergency Trendelenburg Function Optional on LI158Ax The emergency trendelenburg function is obtained in the following manner Fully insert the LCU key into the opening of the emergency trendelenburg function C and keep pressing it The bed adjusts to the trendelenburg position and the sections of the bed frame are flat as long as the key is pressed Remove the LCU key and place it in its storage opening D to return to normal operation This emergency function simultaneously controls the lowering of the sleep surface and adjustment to the trendelenburg position It is activated from the LCU and is available in the event of AC power failure via a battery TotalGuard Integral Siderails The AvantGuard Electric Bed has siderails situated on either side of the bed at the HS and FS The FS has a specific foldaway support located under the sleep surface The foot siderail can also be removed if necessary When the foot siderail is folded away each head siderail has two positions high and low and can be raised or lowered separately These siderails have special openings that enable the control pendant to be stored or moved and integrate a HS angle indicator e To raise the head siderail including the folded foot siderail Pull the siderail out from the bed by rotating it until it reaches the high position The position is secure when the siderail clicks into place e To lower t
206. ura tr AEA ti ti eae eae S 4 49 Replacement of Microswitch Assembly for Bed Braking Detection 4 50 Removal bo hav ee larisa Raw crt he hea alte 4 50 REDACTORES ae WANE SE ee meters 4 51 AvantGuard Electric Bed Service Manual QD1787 Page ix Table of Contents Single Brake arcs a 4 52 Removals restato rires Toae sr ehi e a htcheeadieeseeee E E A 4 52 Replace aT i AE ad da A E hd os on 4 55 Caster Brake and Steer Caster 1 0 0 ccc cee A ee ee eens 4 56 Removal seis agil Rene EA sew PA A ages 4 56 Replacement lt 4 2a ta rd elo e y tla ee heh rita eta OS 4 58 Adjustment Single and Double Band CasterS oo oo oooooomomo o 4 59 Steer Caster seein wow na eee ey yey aia aetna eal pe eee wa 4 60 Sth Wheel Steering SIM is cas SONS EA ee Oe Rees PaaS 4 61 Removal cctitvenaeivicedl ae bless Rene ed eee Ree ate eed gee ages 4 61 Replace opecs erasa a a a ad aS 4 62 Head Section Gas PTE AA AAA A A A EA 4 63 Removal use PD a eat bee eee whet 4 63 Replacement tad spa Oaphnia aad ween eee eel mae a beeen 4 64 Gas Spring Subassembly with Cable and Foot Section Sleeve LI158Ax only 4 65 R moval s serene e redeem A a ee eee ees 4 65 Replacement vaci pira ta Me eagew RRO CAMO e at Rees 4 67 PROMUSUMICHL GS dene eth eh ete e hat ote ae e gs bets ete 4 67 Head Siderail Including Elevation Arm Subassembly 4 68 Removal ui a oe ee o eed ee Ge its 4 68 Replacements os cs A tee a any
207. ure that the corresponding foot siderail is folded away grip the foot siderail support with one hand and with the other operate the release button then rotate the support downwards until it reaches the low position Trendelenburg Reverse Angle Indicators A sleep surface inclination indicator is available on each side of the bed on the foot siderail support H The label is graduated for every 5 steps of inclination from 0 to 15 in both inclination directions Brakes The AvantGuard Electric Bed has a central braking system which is foot controlled by means of a simple bar S Pressing it simultaneously brakes the four casters In order to apply the brakes on the bed step on the brake and steer bar until it is tilted towards the floor To disengage the brakes raise the brake and steer bar with your foot until it is parallel with the floor The bar remains slightly tilted downwards when the bar is released Steering The AvantGuard Electric Bed has a steering system which improves the handling of the bed while moving or transferring a patient The right hand caster at the head of the bed patient viewpoint or left hand caster at the foot of the bed optional is fixed in a longitudinal position with respect to the bed when the bar is Page 1 22 AvantGuard Electric Bed Service Manual QD1787 Introduction Chapter 1 Introduction raised In this way the bed can be steered easily by pushing the foot o
208. utiuetadee bl ad a 4 37 Replacement vull piera GE i cede EAAS A A 4 40 Foot Hilo COMINO AA a Sa ee a 4 42 Removal sE e Sy let E R e s a sa 4 42 Replacement rada bad 4 46 Hilo Pedal sed ea de Ge A da dd a ON de ns 4 48 REMOVAL vi E a AA ee ees A i 4 48 Replacement lt 4 ii A ad Ad 4 49 Replacement of Microswitch Assembly for Bed Braking Detection 4 50 REMOVAL a eels cated cael AA bed eae 4 50 REPIMCEMENE 74 ER SAA a oe 4 51 Single Brake ar A ones ow ous E e A a Ee Ew ei 4 52 Removal La ed a iN ad fam ote ome ta Oe td ed 4 52 Replacement os ncaa Vag awe cage saa AAG eee enews 4 55 Caster Brake and Steer Caster ooooooooooo eee eee ees 4 56 Removal cts oo ery eo Sue eee Re ee eat a 4 56 Replacement iria dad id 4 58 Adjustment Single and Double Band Casters ooooooooooomooo 4 59 IA oe Aa are ade O 4 60 Sth Wheel Steering System evVrpiadis ibid heed one iden deta 4 61 A O 4 61 Replacement de ea rd St RS 4 62 Head Section Gas Spring ve o ida 4 63 NI E a pe ee eee 4 63 Page 4 2 AvantGuard Electric Bed Service Manual QD1787 Chapter 4 Removal Replacement and Adjustment Procedures Replacement rars iena IE Ea il hs aR oe ee ees 4 64 Gas Spring Subassembly with Cable and Foot Section Sleeve LI158Ax only 4 65 Removal eea tad sort dada eed wees 4 65 Replacements tt o ele dd ee Said E ae lll ar Eds 4 67 Adj stment sehe e a eee dda Meee eee es 4 67 Head Siderail Includin
209. vement of the HS frame with two sliding blocks in channels at each side of the intermediate frame which causes the HS frame to extend with respect to the SS The hard surface also extends with respect to the HS frame towards the head end this is achieved by the two elevation arms of the HS frame that terminate in gear teeth interfaced with the notches in the rear of the HS hard surface and move the hard surface towards the head end as the HS is raised CPR Emergency Release The CPR release handle located on either side under the sleep surface adjacent to the LCU enables the driving gear of the head section motor inside the actuator to be disengaged This allows the head section to drop under the pull of gravity assisted cushioned by a two gas springs Assisted Foot Section The LI158Ax bed features a gas spring assisted foot section The movement is controlled by the lever situated to the right of the sleep surface that is connected via a bouden cable to release the FS the movement is assisted cushioned by a gas spring Hilow The hilow system features two columns each composed of three sections that slide in relation with each other by a pantograph system This system is driven by a motor fixed at the base of the mechanism This system enables a travel of 400mm to be obtained Upper and lower End Of Travel EOT switches that are backed up ensure electrical security The system does not require any special maintenance during its d
210. verse trendelenburg which remains continuously lit Without exiting the function selection lockout sequence unlock all the functions and ensure that all the LEDs flash except for the trendelenburg reverse trendelenburg LED which remains continuously lit Test the two buttons of the trendelenburg reverse trendelenburg function to check that they activate the correct function Exit the sequence of function selection lockout trendelenburg reverse trendelenburg access Ensure that the LED of each function remains continuously lit and that the trendelenburg reverse trendelenburg LED goes out Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds Each movement must be continuous Replace the control unit if necessary Troubleshoot in the event of a doubt see RAP 2 2 Control unit on a flexible arm optional Perform the same tests as for the control pendant Replace if necessary Troubleshoot in the event of a doubt see RAP 2 2 Hilow columns Inspect the column assembly for the presence and tightness of the attachment screws Carry out a full raise lower cycle of the bed frame Check that there is no friction or abnormal noises and that no audible overload indication can be heard during the movement Check that the non low bed position LED lights up on the control pendant and extinguishes when
211. wind this elastic It is used to indicate the correct position of the switches and internal end of travels of the actuator The incorrect position of these elements may damage the actuator or the structure of the bed SHOCK HAZARD Unplug the bed before any cleaning or maintenance Failure to do so could result in personal injury or equipment damage Refer to the AvantGuard Electric Bed User Manual and to the specific chapters of this manual for any additional cautions and warnings SHOCK HAZARD Electric current flows through the connections when the bed is switched on Trials conducted under these conditions carry a risk of electric shock Do not touch any wire or terminal which conducts an electric current Failure to observe this warning could result in personal injury or equipment damage WARNING Driven bed mechanisms can cause serious injury Do not operate bed controls until all persons are clear of mechanisms A bed articulation function can be halted by releasing the control pressing and holding the switch for the opposite function or by unplugging the AC power connector Page 1 32 AvantGuard Electric Bed Service Manual QD1787 Safety Tips Chapter 1 Introduction A WARNING This manual describes the overall functions of the AvantGuard Electric Bed and their use but it is in no way intended to replace the user manual the purpose of which is to describe these functions in detail For precise information
212. x on page 2 12 The functions are working Yes No y Go to step 7 6 Replace the CU the control pendant refer to procedure 4 2 on page 4 10 or the control panel on a flexible arm refer to procedure 4 3 on page 4 12 according to the option 7 Is the bed is equipped with the optional HL pedal Page 2 20 AvantGuard Electric Bed Service Manual QD1787 2 4 Hilow Malfunction Chapter 2 Troubleshooting Procedures Yes No y Go to step 14 8 The HL fault is only noted on ONE HL pedal Yes No Go to step 10 9 Replace the defective pedal refer to procedure 4 14 on page 4 48 10 Verify the operation of the right hand pedal Disconnect the right hand pedal The HL function works from the CU or the other pedal Yes No Go to step 12 11 Replace the right hand pedal refer to procedure 4 14 on page 4 48 12 Verify the operation of the left hand pedal Disconnect the left hand pedal The HL function works from the CU or the other pedal Yes No y gt Go to step 14 13 Replace the left hand pedal refer to procedure 4 14 on page 4 48 14 Perform the following to troubleshoot the PSU or HL columns a Remove the cover of the PSU using the 13 mm combination spanner see figure 4 9 on page 4 19 b Connect a HL column simulation connector P N AG0062 see Appendix on page 2 12 to the socket 6 then 4 of the power board D see figure 4 10 on page 4 19 or by
213. y out all the necessary administrative procedures a Truarc is a registered trademark of Waldes Truarc Inc AvantGuard Electric Bed Service Manual QD 1787 Page 6 15 Preventive Maintenance Chapter 6 General Procedures Preventive Maintenance Checklist Table 6 2 Preventive Maintenance Checklist Date Functions Overall condition Head and footboards AC power cable Earth resistance wod fH 191398 NUBIA Leakage current Power supply unit Lateral caregiver unit Control pendant Q 2 Control panel on a 5 5 flexible arm a Hilow columns Hilow pedal Head section actuator Y Head section Bed frame hard E surfaces a CPR release l Automatic contour Foot section actuator Chair function Labour Repair cost a3ed sy 10 S09 BJO Inspected by ZOPaA gt ZO E IT uu ll gg Js e gPageces a Be lt 8 3 E amp 4 E R 2 RE Q 5 Sa o Page 6 16 AvantGuard Electric Bed Service Manual QD1787 Preventive Maintenance Chapter 6 General Procedures Date Functions Mechanical foot section a 518 Trendelenburg reverse 5 trendelenburg 3 Emergency trendelenburg function Battery backup Si
214. y possible by keeping the button of the corresponding function pressed no automatic movement Depending on the function a button corresponds to the movement of one or more actuators Regulations Standards and Codes The AvantGuard Electric Bed complies with CE Directive 93 42 CEE applicable to class I medical equipment CE The AvantGuard Electric Bed complies with the following standards e NF S 90 312 e NF EN 60601 1 e NF EN 60601 1 2 1993 e NF EN 60601 2 38 Amendment 1 MEDICAL AvantGuard Electric Bed Service Manual QD1 787 Page 1 29 Model Identification Chapter 1 Introduction Model Identification For AvantGuard Electric Bed model identification see table 1 8 on page 1 30 Table 1 8 Model Identification Model Number LI158Ax AvantGuard Electric Bed without EasyChair Automatic Chair function mechanical FS LI159Ax AvantGuard Electric Bed with EasyChair Automatic Chair function electrical FS Description Safety Tips A A CAUTION The usage cycle intermittent operation of the actuators must not exceed 3min 30min otherwise equipment damage could result CAUTION The head section must always be horizontal when removing or fitting the head section hard surface Failure to do so may give rise to an incorrect position and cause the elevation sliding system to malfunction CAUTION The polarities must be respected when installing