SERVICE MANUAL Rocket Digital Oocyte
Contents
1. Q Rocketmedical SERVICE MANUAL Rocket Digital Oocyte Aspiration Pump Rocket Medical plc Sedling Road Washington Tyne amp Wear NE38 9BZ England T 44 0 191 419 6988 F 44 0 191 4196989 E customerservices rocketmedical com www rocketmedical com Registered Office Rocket Medical plc Imperial Way Watford WD24 4XX England Company No 03276608 Rocket CRAFT Oocyte Aspiration Pumps Service Manual Il CONTENTS CONTENTO eee ete ne on ee oe a A eee 2 AC LssM E 3 2 1 Rocket Digital Oocyte Aspiration Pump ccccccccecssseeeceeeeeesaeeeseeeeeeeessaaeaeeeeeeeessaeeseseeeeeeessaeaaseeeeeeeeesaaaasess 9 2 2 Rocket Digital Oocyte Aspiration Pump Internal View ccccccccccceescseeesseeeeeeeeeeaeeeseeeeeeessueeaeeeeeeeeeesaeaeeees 4 SN GENERAL DESCRIPFON MENMEM uu 5 cM BIG TIIN 5 3 2 CONTRAINDICA TONG 5 es me OLAS IRCA TON eea E E E E E E E ee eee eee 5 A REFERENCES eran ER T m 5 4 GENERAL INFORMATION Eii 6 LX PECOILi cU M ccc E E E A A 6 LAB MN IS NRERNEREENR M P M 6 SEORSIM REVISION 6
2. WARNING Device can cause explosion in the presence of flammable gases Copyright 2013 15 Rocket Medical plc Rocket CRAFT Oocyte Aspiration Pumps Service Manual 4 6 SUPPLY VOLTAGE SELECTION The device is only for use with power supply Model Number MPUSOB 3 Attachment of any other power supply may severely damage the device CAUTION Disconnection from the mains supply can only be achieved with the removal of the mains power lead from the wall socket The device operates at a voltage 110V 60Hz or 220 240 VAC 9 50Hz 0 12A 4 7 ELECTROMAGNETIC COMPATIBILITY Rocket Craft Suction Pumps comply with the electromagnetic compatibility EMC limits for medical devices as specified by IEC 60601 1 2 2001 These limits are designed to provide a reasonable degree of protection against harmful interference found in typical medical installations Medical electrical equipment requires special precautions regarding EMC and the device must be installed positioned and operated according to the instructions contained in this manual to ensure continued electromagnetic compatibility The device must be operated according to the instructions contained in this manual to ensure continued electromagnetic compatibility 4 8 PACKAGING The packaging has been designed to allow secure transportation of the pump and its accessories After unpacking re assemble and retain the packaging for transport for servicing when required 4 9
3. The device must be transported in temperatures 5 C to 35 C The device is FRAGILE and must be transported in its original packaging to ensure protection If the original packaging is not available please contact your local Customer Services Agent who will provide replacement packaging Dimensions e W 248mm e H 86mm e D 194mm Weight e Unit 2 56Kg e Foot Switch 0 51Kg Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent 23 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 24 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 12 WARRANTY Rocket Oocyte Aspiration Pumps are sold by Rocket Medical plc under the warranties set forth in the following paragraphs Such warranties are extended only with respect to the purchase of the Products directly from Rocket Medical plc as new merchandise and are extended to the first Buyer thereof other than for resale For a period of twelve 12 months from the date of shipment the Products are warranted to be free from functional defects in materials and workmanship and to conform to the description of the Products contained in the operating manual and accompanying labels provided the same is properly operated under conditions of normal use that annual maintenance and service is performed at an authorised Rocket Medical plc servi
4. 44 MANUFAC TURER Mc 6 45 ERVICE AGENT ee E A E ae ee Mcd DE 6 4 6 SUPPEY VOLTAGE SELEG TION arero a E a E E EE E E Eos rceude 7 4 7 ELECTROMAGNETIC COMPATIBILITY ccccccccccceeceeeceeeeeesssseeececeeeeeeeeeeeeeeeaaaaaaussssseceeeeeeseeeesessesagaaaaaseeseees 7 LEN e dlclcm 7 4 9 POSITIONING and PLACEMENT of the DEVICE seesessesseeeeeseeeeenennnnnnnnnnnnnnnnn nnn nn nenne nnnnn nsns nn nnsa aa rrr n nnns nns 7 4 9 SYMBOLS USED ON OOCYTE ASPIRATION PUMP cccsseecceceeeeeeeeeeseeeeaeeeseessseecceceeeeeeessesesaaaeaaaseeeeees 8 4 10 SYMBOLS USED ON POWER SUPPLY UNIT seseeeeeeseeeeeeennnnnennnnnnnnnnnnnnnnnnn nnn nennen nennen nnns nnnnna sain nnns 9 4 11 SYMBOLS USED ON R57686 PATIENT CONNECTION SET ssesesessseseeeeeeeneeeeenennnnnnnnnn nnns 9 INIA ak Ne 10 5 1 Replacing the Inlet and exhaust fiters euis oos uo euis oa quud 9b De coo a a eoa Eo URE iiaa du ouo IET e PH IN D ednd 10 5 2 Replacing the pump valve amp diaphragm set cccccccccsseeccceescecceeseeeceeeseeceeaseeessaeeecseueeesseeeesssageeessageeesseass 12 5 9 Reassembling the chassis into the casing cccccseececceeeeeeceeeeeeecaeeeeecaeeeeeeseaueeesseueeessaeeeeseaeeeessanseessaeeeeenaegs 15 6 POST SERVICE INSPECTION cerr
5. Pumps Service Manual 6 1 RESET Pump Timer to ZERO 1 Place a finger on the rotary control surface initial BEEP 2 HOLD for 7 secs until 2 BEEPS are heard 3 Hold for a further 2 secs until a continuous BEEP is heard 4 Display will show elapsed pump hours 5 CHECK using an anticlockwise circular motion to decrease the value to ZERO 6 HOLD rotary control until a single BEEP is heard 7 CONTINUE TO HOLD for further 7 seconds until a continuous BEEP is heard 8 CHECK the display returns to 100mmHg 9 CHECK the pump has returned correctly to the PRE SET mode by activating the black footswitch reading should show 500mmHg activate the white footswitch which should show 100mmHg PASS 10 If the display does NOT show static values of both 500 and 100mmHg hold down the rotary control until a single BEEP is heard 11 Continue to hold for a further 7 seconds until 3 BEEPS are heard Release the control 12 CHECK the pump has correctly returned to pre set mode 2 Copyright 2013 15 Rocket Medical plc Hold the rotary control down 7 second interval Two beeps Continuous Beep Display shows elapsed hours of pump operation Use rotary contro to decrement displayed value to zero Hold the rotary control down to confirm All rights reserved Beep 7 second interval Continuous Beep Display returns back to 100mmHc Check that the pump has retur
6. Supply Unit 12V DC connection polarity Dispose of this device in accordance with WEEE directive 2002 96 EC PSU conforms to EN60601 1 amp IEC60601 1 4 11 SYMBOLS USED ON R57686 PATIENT CONNECTION SET N Read the Instruction for Use before connection and use Device is for Single Use Only Batch number for sterile device ISTERILE EO Device is sterilised by Ethylene Oxide e The device is not manufactured with natural latex CE Mark Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 5 ANNUAL SERVICE 5 1 Replacing the inlet and exhaust filters 1 Lay the pump on a clean t A surface protecting the casing from scratches WARNING Observe precautions for 2 Remove 3 screws from handling electrostatic the rear casing and retain discharge sensitive devices 3 Turn the pump over and remove 3 screws from the front lower casing and retain 4 Remove white silicone retaining O ring from vacuum inlet port 5 CAUTION Avoid the use of sharp tools as these may damage the O ring O Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 6 Detach ribbon cable from the front panel vacuum controller PCB by lifting the tab 7 Carefully detach the ribbon from the front cover and store secur
7. POSITIONING and PLACEMENT of the DEVICE Rocket Digital Oocyte Aspiration Pumps must be placed on a secure level surface away from sources of heat water splashes mists or cooling vents Do not expose to direct sunlight Do not expose to flammable gases Operating temperature Range 5 C and 35 C All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 4 9 SYMBOLS USED ON OOCYTE ASPIRATION PUMP 12 V DC Power ON OFF This device operates on 12V DC power provided by a separate power supply module supplied with the device Disconnection from the mains supply can only be achieved with the removal of the mains power lead from the wall socket This device is Type B Connection point for filter set This device is powered by a Class Il power supply unit Indicates the Service Interval Indicator Suction Display Touch control symbol Indicates activation of the pre set Max 500ommHg mode Indicates activation of the user set 50 300mmHg mode Dispose of this device in accordance with WEEE directive 2002 96 EC CE Mark Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 4 10 SYMBOLS USED ON POWER SUPPLY UNIT Read the manual before connection and use WARNING Risk of electric shock The PSU should not be opened Device is only for use indoors Class Il Power
8. asing 33 Re install the white silicone retaining O CAU TONE AVSA Hie ring to the vacuum inlet use of sharp tools as ort these may damage the Pore O ring l6 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 6 POST SERVICE INSPECTION 1 Attach 240V Main cable to 12V Power Supply Unit PSU Model SUAM number MPU30B 3 and sone aa connect to an electrical supply Mee LL 240VAC 50Hz The power ANC E EF supply module is self regulating Xn to generate 12VDC NACE NW 2 Attach 12V lead to rear input port on the Rocket Digital Oocyte Aspiration Pump RDOAP 3 Attach the footswitch air tubing to the male female ports on the underside of the pump casing 4 CHECK for security of attachment 5 Turn the power switch 0 I rear panel to on 6 CHECK The green power LED indicated on the front panel is illuminated 7 The main display will display an initial value of 100mmHg Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 8 CHECK the main display holds a steady value of 100mmHg On Start Up the pump is pre set to 100mmHg suction 9 Place a finger on the control surface and using a rotating motion The central blue dot is NOT active 10 CHECK to see the display changes in 5 m
9. ce facility Removal of any QC seal voids the warranty The foregoing warranties shall not apply if the Products have been repaired other than by Rocket Medical plc or other than in accordance with written instructions provided by Rocket Medical plc or altered by anyone other than Rocket Medical plc or if the Products have been subject to misuse negligence or accident Rocket Medical plc s sole and exclusive obligation and Buyer s sole and exclusive remedy under the above warranties is limited to repairing or replacing free of charge at Rocket Medical plc s option Products which are reported to Rocket Medical plc by mail telephone or email and which if so advised by Rocket Medical plc is thereafter returned with a statement of the observed deficiency not later than seven 7 days after the expiration date of the warranty to Rocket Medical plc during normal business address transport charges prepaid and which upon Rocket Medical plc s examination is not found to conform with the above warranties Rocket Medical plc shall not be otherwise liable for any damages including but not limited to incidental damages consequential damages or special damages There are no express or implied warranties which extend beyond the warranties herein above set forth Rocket Medical plc makes no warranty of merchantability or fitness for a particular purpose with respect to the Products or parts thereof 13 DISPOSAL At the end of th
10. e service life of the equipment this device should be handled and disposed of in accordance with local hospital policy and with regard to all applicable regulations including but without limitation to those pertaining to human health amp safety and care of the environment Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 14 TECHNICAL SPECIFICATIONS 14 1 CLASSIFICATION IEC 60601 1 Type of protection against electric shock Class II Degree of protection against electric shock Type B Vacuum type high vacuum low volume Suitable for continuous operation Not suitable for use in the presence of flammable gases Not suitable for use in conditions which expose the device to the ingress of water Not suitable for sterilisation 14 2 SPECIFICATIONS Power Input to Pump 12VDC Universal Power Supply 100 240VAC Frequency 50 60 Hz Maximum current 2 5A 12V Environmental conditions 5 C to 35 C Service interval 1000hrs of pump operation Dimensions e W 248mm e H 86mm e D 194mm Weight e Unit 2 56Kg e Foot Switch 0 51Kg Suction Ranges e User Set 50mmHg to 300mmHg in 5bmmHg increments e Pre set 500mmHg nominal Suction Range Accuracy 10 at operating range 100 500mmHg 25 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspirat
11. ely 8 Detach patient outlet filter from tube 9 Unclip filter from spring clip discard and replace filter Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 5 2 Replacing the pump valve amp diaphragm set 9 Unscrew two securing nuts and lift out chassis to gain access to pump head 10 Angle chassis to support on casing 11 Unscrew 4 screws form the pump head ina crosswise manner to prevent distortion of the pump head 12 Discard and replace the pump head valve module 13 CAREFULLY remove the diaphragm by unscrewing the centre IMPORTANT screw arrowed 2 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 14 Remove the diaphragm to reveal the piston head 15 Ensure the piston is free to move and that there is no contamination or corrosion evident 16 Discard old diaphragm and replace 17 Centralise piston into housing and re attach diaphragm plate and screw DO NOT OVERTIGHTEN SCREW IMPORTANT IMPORTANT Note valve head orientation 18 Discard old head gasket and replace A with new component IMPORTANT 3 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pum
12. ion Pumps Service Manual 15 DRAWINGS 15 1 Set up Flow Chart Attach power cable Activate BLACK foot switch Attach Filter Set Attach both foot switch leads Display defaults to Check display to 100mmHg 500mmHg Check BLUE LED Pump sound READY FOR USE Track dial L amp R to set required value HOLD wait for tone to SET Activate BLACK foot switch Activate WHITE foot switch Check display returns to set value Check BLUE LED Pump sound 26 15 07 15 Rev 3 ZDOCK258 All rights reserved Copyright 2013 15 Rocket Medical plc
13. isplay Prevent any fluid from entering the device 8 YEAR OF MANUFACTURE The year in which the device was manufactured is indicated by the first 2 numbers of the serial number For example a serial number starting 11180776 indicates the device was manufactured in 2011 Units manufactured after 2014 have the year of manufacture stated on the rear label 9 RETURNING THE PUMP FOR FACTORY REPAIR OR SERVICE All devices to be returned must be prepared as described below for the protection of the servicing team and for safety during transport 1 Surface clean the pump as described in the Section above 2 Sealin a plastic bag and seal within a second plastic bag 3 Place in the original packaging 4 Enclose the following information e Contact name e Centre address e Decontamination Certificate e Description of the fault or service required e Accompanying Order to authorise servicing contact your local Customer Services Team for details All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 10 STORAGE The device must be stored and operated in temperatures 5 C to 35 C The device must be stored in a clean dry condition ideally in its original packaging which should be retained to return the unit for servicing Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent
14. l plc Rocket CRAFT Oocyte Aspiration Pumps Service Manual 4 GENERAL INFORMATION 4 1 COPYRIGHT This manual contains information that is subject to copyright All rights reserved This manual should not be photocopied duplicated or distributed completely or in part without the written approval of Rocket Medical plc 4 2 MODEL NUMBER Rocket Digital Suction Pump Complete 110v R29700 Power supply MPU30B 3 4 3 MANUAL REVISION Revision 1 First Publication 05 11 13 Revision 2 Update of LRQA CE mark 27 08 14 Revision 3 Pump timer zero procedure 15 07 15 4 4 MANUFACTURER Rocket Medical plc Sedling Road WASHINGTON Tyne amp Wear NE38 9BZ UK 4 5 SERVICE AGENTS Rocket Digital Pumps typically require only annual maintenance however it is recommended they are serviced and calibrated annually at a Rocket Medical plc approved service facility Failure to service the pump at the indicated intervals may invalidate the Warranty Customer Services Rocket Medical plc Sedling Road WASHINGTON NE38 9BZ ENGLAND Tel 44 0 191 419 6988 Fax 44 0 191 419 6989 Email homesales rocketmedical com All rights reserved ZDOCK258 Rev 3 15 07 15 WARNING ELECTRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IEC CEC and NEC requirements WARNING ELECTRIC SHOCK HAZARD Do not immerse the device b iy S oN 3 S E
15. mHg increments down to 50mmHg and up to a maximum of 300 mmHg 11 CHECK the active dial zone will sense the clockwise motion to increase the display in 5mmHg increments 12 CHECK using an anticlockwise circular motion to decrease the value 13 CHECK that a value of 150 mmHg can be set by selecting the value and holding the finger still for 2 seconds when an audible beep will confirm the new setting 14 CHECK If the new setting is not correctly saved the value will return to its previous set value in 5 10 seconds y 8 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual The footswitch control operates on a toggle switch press once for ON and depress again for OFF 15 Depress the WHITE a pedal once 16 CHECK The warning tone sounds to indicate activation of the footswitch 17 CHECK that the blue LED has illuminated and is flashing rapidly to confirm suction is active 18 Stop the suction by depressing WHITE pedal once again 19 CHECK the pump stops immediately and the blue LED has gone out Repeat the process for HIGH Vacuum 20 Depress the BLACK pedal once 21 CHECK The warning tone sounds to indicate activation of the footswitch 22 CHECK that the blue LED has illuminated and is flashing rapidly to confirm suction is active 23 CHECK that the ORANGE LED arrowed right has also illumi
16. nated to confirm HIGH Vacuum is now active 24 CHECK that the display indicates 500 mmHg 25 Stop the suction by depressing BLACK pedal once again 26 CHECK the pump stops immediately and the blue LED has gone out 9 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 27 Connect a calibrated vacuum test gauge to the patient port on the pump 28 RECORD the Test Gauge serial number and calibration date in the table below WARNING Ensure the vacuum levels are within the specified accuracy tolerance before returning the pump to clinical service 29 Set the pump display to each of the indicated values on the table in turn 30 Activate the WHITE low vacuum Test Gauge pedal for the lower values and the BLACK pedal for the Max vacuum Serial Number value 31 RECORD the values on the gauge in Calibration Date the table below 32 Acceptance criteria 10 of displayed value DIES o FNAME TTE INITIAL TO CONFIRM mmHg COMPLETED ee Confirm bleed to OmmHg in 2 0 seconds TEST iiir i ana Confirm bleed to OmmHg in 2 0 seconds Recorded Value VAGY ZAT BLEED Confirm bleed to 0mmHg in 2 0 seconds TEST MAX 500mmHg Midi Confirm bleed to 0mmHg in 2 0 seconds MEM 20 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Hev 3 15 07 15 Rocket CRAFT Oocyte Aspiration
17. ned to the pre set mode of 500mmHg amp 100mmHg by activating the dual footswitch Hold the rotary Passed Inspectio control down Three beeps 15 07 15 ZDOCK258 Rev 3 WARNING ELECTRIC SHOCK HAZARD Do not immerse the device or power supply WARNING Do NOT attempt to sterilise the device WARNING Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent HAZARD Do NOT include used consumables as these pose a significant contamination risk e IMPORTANT A decontamination certificate MUST be included with every returned pump Repair or servicing cannot be commenced unless the service agent is in possession of this certificate 22 Copyright 2013 15 Rocket Medical plc Rocket CRAFT Oocyte Aspiration Pumps Service Manual 7 CLEANING THE PUMP CASING At the end of each clinical session turn off the device at the rear panel and disconnect the PSU from mains power supply Using an aqueous 70 alcohol solution eg IMS or isopropyl BP solution moisten a cloth and wipe all external surfaces of the device If the surface has become contaminated with proteinaceous material remove with a light detergent solution before surfacing cleaning with an alcohol solution Do NOT use a 100 alcohol or any other solvent to clean the device as this may cause damage to the casing surface and d
18. ockwise to increase anticlockwise to decrease the set value Suction Display in mmHg Patient connection port only for use with R57686 Rocket Oocyte Aspiration Pump Patient Connection Sets Power On Indicator LED Green 12VDC Footswitch connection ports User Set Suction Indicator LED Blue 50 300mmHg Pre set Max Suction Indicator LED Orange 500mmHg service Indicator LED Yellow 9 O I 12V Power On Off switch 10 Dual footswitch air controlled SO iW eS IE Sm Se Not shown 11 Power Supply Unit PSU Model MPU30B 3 12 Power cords IEC UK amp EC types LA Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 2 2 Rocket Digital Oocyte Aspiration Pump Internal View O I Mains Power On Off Vacuum display 0 500mmHg normal range 50 300mmHg Motor 12V Pump head including valve housing Vacuum outlet port Vacuum inlet filter Vacuum control solenoid Internal silencer with filter chamber A Internal silencer with filter chamber B e 49 24 eat eS Se PS 0 Internal exhaust vent and filter 1 Control PCB with ribbon cable to vacuum control touch wheel Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 3 GENERAL DESCRIPTION The Rocket Oocyte Aspiration Pump has been developed to pr
19. or ee eene editus Rec tu icteodaihsincaedenmindedenineatetontawduhostendadenseedstenieatel 17 6 1 RESET FU Timer TO ZERO PER 21 nmeuz wlsemdldsensuccm E 22 hz m8 dub iB eslcc 22 9 RETURNING THE PUMP FOR FACTORY REPAIR OR SERVICE esses enne nennen nennen 22 iX VIO C 23 CE TRANSPORTATION c ee 23 INT LIDIBnM c 24 jo eB ee Qo MU ee 24 T4 es ey ns al onl U7 24 Serene ne ee ee ee ee E ee ee eee eee 25 TET CLASIFICA TON p s 25 IDEM Idee es c 25 To DRAWING E 26 pow eue R6 ee E S 26 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Darkat CCDAETTM lynnvta Acniratinn E WNC C annoa Maniio Rocket CRAFT Oocyte Aspiration Pumps Service Manual 2 GENERAL ASSEMBLY 2 1 Rocket Digital Oocyte Aspiration Pump Rocket Digital Oocyte Aspiration Pump Touch sensitive Suction Control Dial cl
20. ovide smooth low volume high suction vacuum at a pre determined negative pressure Suction is activated by a foot operated toggle air switch controlled by the surgeon performing the oocyte collection The range of suction is variable from 50 300mmHg and at a pre set 500mmHg in Max suction mode The Rocket Oocyte Aspiration Pump requires a disposable filter set for attachment of the pump to the oocyte collection needle The filter set is supplied separately sterile and for single patient use You will also require e A suitable oocyte aspiration needle such as Rocket SX Single Lumen Oocyte Aspiration Set R57602 00 90 e Suitable collection tubes for use with oocyte needle sets such as B D Falcon test tube No 2001F 17 x 100mm e Flushing media 3 1 INDICATIONS For the generation of high vacuum low volume suction between 50 500 mmHg to permit the aspiration of follicular fluid oocytes and ovarian fluid as part of the treatment of infertility relating to IVF and other gynaecological procedures 3 2 CONTRAINDICATIONS Not intended for use where ovarian aspiration or the aspiration of ovarian fluid is contraindicated For short term operation only The device is NOT intended for continuous drainage 3 3 CLASSIFICATION e IEC 60601 1 2005 e Type of protection against electric shock Class II e Degree of protection against electric shock Type B e Vacuum type high vacuum low volume e Suitable for continuous opera
21. ps Service Manual 19 Re attach inlet vacuum tubing to TOP port 20 Re attach exhaust filter to BOTTOM port 21 Relocate exhaust filter in bottom of case 22 Tighten screws in a crosswise fashion to avoid distorting the pump head IMPORTANT DO NOT OVER TIGHTEN 23 Return chassis into casing 24 Re tighten two securing nuts 4 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 5 3 Reassembling the chassis into the casing 25 Pass patient outlet filter underneath tubing and re attach filter to VERY patient connection port IMPORTANT tube 26 Re clip filter into spring clip 2 7 Ensure tubing is not trapped by spring clip and correctly routed to avoid occlusion 28 Re attach ribbon cable from PCB to the front panel vacuum controller PCB by lifting the tab 29 Carefully inserting the ribbon and closing the tab securely 5 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 30 Lay the pump ona clean surface protecting the casing from scratches Ce 31 After re assembling the upper casing to the IMPORTANT lower frame re install 3 DO NOT OVER screws to the rear TIGHTEN THE casing SCREWS 32 Turn the pump over and re install 3 screws to the front lower c
22. tion e Not suitable for use in the presence of flammable gases e Not suitable for use in conditions which expose the device to the ingress of fluids e Not suitable for sterilisation 3 4 REFERENCES Craft McLeod F Edmonds K Human embryo transfer technique Lancet 1961 ii 1104 5 Craft Diahanbakch O McLeod F et al Human pregnancy following oocyte and sperm transfer to the uterus Lancet 1992 i 1031 3 Craft I 1984 Clinical Methodology British Journal of Hospital Medicine 90 102 Reeves G Scott H T et al 1989 Journal of Assisted Heproduction and Genetics Volume 6 Number 6 December 1989 Copyright 2013 15 Rocket Medical plc All rights reserved ZDOCK258 Rev 3 15 07 15 N WARNING READ THIS MANUAL CAREFULLY Please familiarise yourself with the contents of this manual before attempting to use the device Failure to observe these instructions may result in damage to the pump or cause injury to the patient or user This device should only be used by suitably qualified personnel WARNING ELECTRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IEC CEC and NEC requirements N CAUTION Any adjustment modification or repairs to the equipment should be carried out by authorised service agents Disposal of this device must be undertaken with regard to the WEEE directive 2002 96 EC Copyright 2013 15 Rocket Medica
Download Pdf Manuals
Related Search
Related Contents
Jura Impressa X90/X95 - Gastronomie Service Dahmen Multi-SeaLite® P/N 710-040-601 User Manual, Rev. 10/23/02 Salin Plus - Suki 16 Litre Page 13: Femme ネットワーク接続ガイド Jwin JX-CD3400 Shelf System Válvulas Fisherr ET e ED de Grandes Dimensões NPS 12 a 16 e Inhalt - bei Dual Nikon COOLSCAN IV ED LS40 Guía de usuario! Copyright © All rights reserved.