        For US Physicians Only IVTM System Operation Manual
         Contents
1.      Figure 8 1  Location of Condensate Pan       3  The pan may be washed in hot soapy water  Never use solvents or  abrasive cleaners on the pan  Dry the pan when finished and reinstall it  in the System     IMPORTANT     Do not operate the System without a condensate collection pan properly  installed  Failure to do so may allow water to accumulate on the floor under  the System        Drain Coolant    It may become necessary to drain and clean the coolant well because the  coolant has become contaminated  or to prepare the System for shipment     Required Tools and Materials    You will need the drain line and connector assembly  supplied with the System    and a 2 liter waste container to perform this task     Procedure    To drain the coolant well  follow these steps in the indicated order   1  Unplug the System     7 600479 001 Rev2 137 147    Maintenance    Remove the cap on the coolant well and set it aside on a clean surface     Remove the front panel of the console by pulling the panel straight away  from the front of the machine  no tools are required to remove the panel    Near the bottom left corner of the console is a drain hose held by a    clamp and the drain coupler           Figure 8 2  Drain Hose and Drain Coupler     Place the open end of the drain hose into a suitable container or position  it above a floor sink or drain     Push the connector on the drain hose into the drain coupler on the       MFA  Tey      ATA ENEA     4 ease    i    n a    Perret ir
2.      STANDBY    Prepare data link     2006 03 05   18 26 Cancel       Figure 4 65  Download Error Message     8  When the data download is complete  the System screen briefly displays  the message    Download Complete        9  If you are downloading data after treatment has ended  the patient trend  data in the System will be deleted  After about two seconds  the  message is automatically cleared and the screen displays the message     Turn power off        Turn power off    2006 03 05   18 26          Figure 4 66  Turn Power Off Message     10  At the rear of the console  turn the power on off switch to OFF     53 600479 001 Rev2 84 147    Operation    11  Disconnect the serial interface cable from the System and the laptop  computer  Patient trend data downloading is now complete     New Patient     No Power Down    Change the Start up Kit    The system does not have to be powered down to start a new case  A new  catheter  patient temperature probe and Start up Kit are required for each patient      see Disposal of Used Components  To immediately start a new case without    power down    1  Place the System is Standby    2 Delete the previous Patient Data     see below    3  Verify the System Settings    4 You may select to Pre Cool or Pre Warm the coolant using the settings    menus   Connect the New Patient   6  Place the System in RUN when desired     Delete Previous Patient Data    To delete a previous patient s data and prepare the System for immediate use  with a new
3.      You will not see any screen messages unless there is an abnormal result in  which case the appropriate alarm state will be called     Sensor Checks    System Set Up    Check the following    Air Trap Roller Pump Lid    Check Prime Switch Coolant      2006 03 05   18 26    Figure 4 1  System Setup Screen       For the system to enter the next state  STANDBY  the following sensors must be  checked and found normal     6 600479 001 Rev2 37 147    Operation    Table 4 1  Setup Sensor Checks    Air Trap There is no Start up Kit installed or there is a large amount of air in the  Start up Kit chamber     Roller Pump Lid The clear plastic roller pump lid is not closed properly    Check Prime Switch   The Prime Switch is being depressed  as when you use it to prime the  system  or has been jammed in the    on    position     The coolant level is low  Top up the coldwell with coolant     If any of these sensor states are not correct you will see the    System Setup    and  the problem sensor state will be highlighted in red  see Figure 4 1 above   Once  you rectify the identified problems  the System will progress to STANDBY        NOTE  Change in Design     Patient Temperature Probes    In earlier versions  the System would not enter STANDBY without at least a  Primary Patient Temperature Probe  T1  connected to the System  For XP  Systems  the System will enter STANDBY without a T1 probe being  connected  however  the System will NOT enter RUN without at least the  Primary Pati
4.     If particulate matter is   Drain coolant  clean the coolant well  and   observed in the refill with 1 1 mixture of distilled water and   coolant propylene glycol  refer to the    Drain  Coolant    section of this chapter         Inspect Replace Condenser Filter    Required Tools and Materials    You will need a Phillips screwdriver  an inspection lamp  and a vacuum cleaner  with a crevice tool to perform these tasks     Procedure   To inspect and replace the condenser filter  perform these steps in the indicated   order    1  Assure that the System has been switched off and the power cord has  been disconnected    2  Use a screwdriver to release the screws securing the access door   Remove the access door from the panel    3  Remove and inspect the filter  Replace it if it is clogged with dust    4  Use the inspection lamp to inspect the area in and around the  condenser  Look for accumulations of dust and debris    5  If necessary  use the vacuum cleaner   s crevice tool to remove dust from  the condenser  Gently pass the crevice tool over all exposed surfaces of  the condenser only  Take care not to bend any of the cooling fins of the  condenser  Never use cleaning fluids or water on the condenser    6  When finished  replace the access door and secure the screws    7  Inspection and cleaning are complete  The System may be returned to    service     5 600479 001 Rev2 135 147    Maintenance    Temperature Accuracy Test    Required Tools and Materials    You will need a 
5.    ei  a        ect Piet   g pi om        Temi   2  econ   a                Figure 4 33 500ml Saline Bag on Hook     The Top Cover    2  Open the top cover of the System  Open the transparent top cover of the  roller pump     32 600479 001 Rev2 63 147    Operation       Figure 4 34  Covers Open     A tubing circuit diagram is printed on the inside of the System top cover   Refer to this diagram when installing the Start up Kit        Figure 4 35  Tubing Circuit Diagram     33 600479 001 Rev2 64 147    Operation    The Coldwell    4  Remove the cap from the coolant well and set it aside in a clean location        Figure 4 36  Coolant Well Cap     5  Check the level of the coolant  The liquid level should be between the  two indicator lines on the wall of the coolant well  If the level is below the  bottom indicator line  add distilled water until the liquid is at the top  indicator line  If the coolant level is unusually low because of spillage   add premixed propylene glycol and water until the liquid is at the top  indicator line                   Figure 4 37  Coolant Well Liquid Level Indicator Lines     34 600479 001 Rev 2 65 147    Operation    The Alsius Start up Kit    6  Open the System Start up Kit  For convenience and sterility  all items in  the kit are preconnected        Figure 4 38  System Start up Kit     ie Insert the heat exchanger coil into the coolant well        Figure 4 39  Installing the Heat Exchanger Coil     8  Temporarily slide the air trap into the h
6.   18 26             Figure 4 58  Patient Data Message     f  Choose    Download Now      8  The screen displays the message    Prepare data link        System Set Up    Prepare data link       2006 03 05   18 26          Figure 4 59  Prepare Data Link Message     9  Connect one end of a serial interface cable to the laptop computer s  serial interface connector     10  Attach the other end of the cable to the serial interface connector on the  System  The male 9 pin subminiature    D    connector is located on the  rear of the control head     48 600479 001 Rev 2 79 147    Operation       Figure 4 60  Location of the Serial Interface Connector   11  Start the TempTrend CSV program on the laptop computer  Refer to  Chapter 5   TempTrend CSV     12  If a problem occurs during downloading  the System screen will display  the message    Download error  Please check external computer     Check  the serial cable connections and the operation of the laptop computer  and choose    Try again     If repeated failures occur  choose    Cancel      Repeated failures indicate a problem with the TempTrend program or the  System  contact your ZOLL representative for assistance     System Set Up    Download error  Please check  external computer     Prepare data link     Cancel  2006 03 05   18 26    Figure 4 61  Download Error Message        13  When the data download is complete  the System screen briefly displays  the message    Download Complete     At this time  the patient trend data  s
7.   7 070 612 7 090 792 7 097 657 7 101 388 7 144 407    ZOLL Circulation  Inc    650 Almanor Avenue  Sunnyvale  California 94085  U S A     Telephone   1 408 541 2140  Facsimile   1 408 541 1030    600479 001 Rev2 2 147    Contents    1  Safety Information  2  Introduction  3  Receiving  Inspection  and Assembly    4  Operation    Overview   Alarms  amp  Alerts   Your First Case   Setup   Variations  Ending Treatment  Temperature Trend Data  Mechanical Components    Accessory  HMIA       600479 001 Rev2 3 147    Safety Information    Safety Information    1 600479 001 Rev2 4 147    Safety Information    Contents    safety Information 3    Overview 3  Warnings  Cautions  and Notes 3  Definitions of Symbols and Labels Used on the Product and in the Manual 4  General Safety Precautions 5  Shipping and Storage Conditions 6  Ignition of Flammable Anesthetic Mixtures 6  Electrical Hazards 6  Primary Patient Temperature Probe  T1  Failure 7  Configuration Changes 7  Priming the Saline Circuit 7  Air Entry Into the Tubing Circuit 8  Check the Integrity of the Catheter 8  Check the Integrity of the Start up Kit Tubing 8  Interference 9  Product Label 9    2 600479 001 Rev2 5 147    Safety Information    Safety Information    Overview    Safety is of primary concern to ZOLL Circulation  Inc   This chapter provides  information on safely using the System  You must read and understand the  information in this chapter before operating the System  Always follow the  warnings  cautions  and 
8.   Attached       Brake   Steering   Brake    Steering  Yes current  position  Free swivel  Will lock when  Left Front Yes next straight  ae    Rear  both  Not   Not Applicable o       1  Direction Locking Caster   Note dome above caster axle    Free Swivel Casters at rear of system    3  Kick Stop Caster  Free swivel caster with  brake        Figure 4 73  Casters on the System    63 600479 001 Rev2 94 147    ZOLL Circulation       Hospital Monitor Interface  Accessory  HMIA     1 600479 001 Rev2 95 147    Accesories   HMIA    Contents  Overview 3    Operating the HMIA 3    Connecting to the Hospital Monitor 3  To begin operation 4  Controls and indicators on the HMIA 5  Operating States 6   START UP 6   RUN 6  Calibration 6    Installation 6    Installation Equipment 7  Installation Procedure 7  Removal of the HMIA 10  Connection Cable Part Numbers 10    Data Download 11    Troubleshooting 11    Errors 12  Green LED OFF 12  Flashing Red LED 12  No Temperature Reported On Hospital Monitor 12  Incorrect Temperature Reported On Hospital Monitor 12  Temperature jumps between 26  C and 49  C On Hospital Monitor 13  Hirose Connector does not fit 13    YSI 400 Temperature vs  Resistance 14    2 600479 001 Rev2 96 147    Accesories   HMIA    Overview    This chapter explains the installation  operation  and maintenance of the Hospital  Monitor Interface Accessory  HMIA      The sole function of the HMIA module is to simulate the patient temperature  probe connected to the T1 port on 
9.   Figure 4 70  Select Interval     4  Turn the knob to highlight your desired selection  When it is highlighted   press the knob once     5  The screen displays the pop up menu again  Turn the knob to highlight  the    Cancel Exit    selection     6  Press the knob once  The menu will disappear and the temperature trend  graph will be displayed using the interval you selected     60 600479 001 Rev2 91 147    Operation    Mechanical Components    Top Cover    When access to the tubing or Coldwell of the System is required  lift the top cover  to a fully upright position        p  1 Kad    N  ad AT psi     j    2 y E    a   7   i           Figure 4 71  Top cover fully open     61 600479 001 Rev 2 92 147    Operation    Control Head Tilt    To modify the tilt of the Control Head or to lock the Control Head in place use the  lever located at the swivel point  Turn the lever clockwise to tighten or  counterclockwise to loosen  The position of the lever may be changed  without  adjusting the tightness  by pulling the lever towards you        Figure 4 72  Control Head Tilt Lever     62 600479 001 Rev2 93 147    Operation    Casters    There are three different types of casters  wheels  on the System     1  The right front has a free swivel and brake feature   2  The left front has a free swivel and a directional lock feature   3  The rear casters do not lock and are free to rotate at all times     The figure below shows the operation of the casters     Table 4 6  Caster Operation   
10.   Run Button    Rate Deg   HR Button   Press for Menu   Enter Knob       z       ALSILS    Figure 2 2  Controls and Display     Power Indicators    An indicator lamp on the control panel is illuminated when power is switched on   A second power on indicator is mounted directly above the power switch on the  rear of the console     Alarm Indicators    The System typically notifies users of alarm conditions in two ways  When an  alarm occurs  the screen displays an alarm message  and an alarm annunciator  produces an audible alarm tone  beep   The alarm tone can be temporarily  muted by the user  but it cannot be turned off     If the nature of the failure prevents the System from displaying an alarm  message  the alarm indicator on the control panel will be illuminated     Control Buttons    The display head features four pushbuttons that are used to control System  functions  To provide confirmation  each time a button is pressed  a    key click     sound is produced by the annunciator     Target Temp    Press the    Target Temp    button to display a screen that allows you to set the  patient   s target temperature  You may set a target temperature between 31   C  and 38   C  87 8   F and 100 4   F      Rate Deg Hr    Press the    Rate Deg Hr    button to display a screen that allows you to set the  cooling warming rate  expressed in degrees per hour   You may set a cooling   warming rate between 0 10   C hr and 0 65   C hr  0 18   F hr and 1 17   F hr      Standby Run    5 
11.   or control buttons fail to operate as  expected     e The System has been exposed to fire  flood  or hazardous  substances     e The System has suffered potential damage due to improper storage   rough handling  or being dropped     6 600479 001 Rev2 130 147    Maintenance    Maintenance    1 600479 001 Rev2 131 147    Maintenance    Contents    Maintenance 3    Overview 3  Safety Precautions 3  Required personal protective equipment  PPE  3  Required tools and materials 3  Scheduled Maintenance 4  Unscheduled Maintenance 5  Inspect Replace Condenser Filter 5  Required Tools and Materials 5  Procedure 5  Temperature Accuracy Test 6  Required Tools and Materials 6  Procedure 6  Clean Console and Condensate Pan 6  Required Tools and Materials 6  Procedure 6  Drain Coolant 7  Required Tools and Materials 7  Procedure 7  Spillage 9    2 600479 001 Rev2 132 147    Maintenance    Maintenance    Overview    To assure safe operation and a long service life  the System must be periodically  maintained  This chapter describes the scheduled and unscheduled maintenance  tasks that must be performed by the user     Scheduled maintenance tasks must be performed at least as frequently as  recommended in this chapter  If your System is subject to very long periods of  continuous use  you may be required to perform these tasks more frequently  than the schedule recommends     The frequency of performing unscheduled maintenance tasks depends upon the  manner in which you use the System  Most
12.  200818 002   MALINCKRODT MEDICAL FOLEY PROBE 12F  200818 003   MALINCKRODT MEDICAL FOLEY PROBE 14F  200818 004   MALINCKRODT MEDICAL FOLEY PROBE 16F  200818 005   MALINCKRODT MEDICAL FOLEY PROBE 18F  200818 006   MALINCKRODT MEDICAL GP RECTAL PROBE 9F  200820   001   SMITH MEDICAL FOLEY PROBE 10 FR  600224 001  SMITH MEDICAL FOLEY PROBE 12 FR  600224 002  SMITH MEDICAL FOLEY PROBE 14 FR  600224 003                  _ Or SMr    SMITH MEDICAL FOLEY PROBE 16 FR  600224 004  SMITH MEDICAL FOLEY PROBE 18 FR  600224 005   SMITH MEDICAL GP RECTAL PROBE 9 FR  600225 001   SMITH MEDICAL GP RECTAL PROBE 12 FR  600225 002     Rusch Sensor Temperature Foley Catheters  CH 8  10  12  14   16  18      Rusch P N 179360     Rusch Temperature Supracath  CH10  12  14    Rusch P N  179370     Model CG 500D    147 147    
13.  68 147    Operation       Figure 4 43  Pump Tubing Installation     13  Close the top cover on the pump  It will snap shut     Cover the Coldwell    14  Replace the cap on the coolant well  Position the cap so that the heat  exchanger coil and tubing fits into the notch  Press down on the cap to  create a tight seal        Figure 4 44  Replace the Cap on the Coolant Well     Spike the Saline Bag    15  Using aseptic technique  connect the priming line to the sterile saline  container  The priming line is equipped with a    spike    connector  Hang  the saline container on the hook provided     38 600479 001 Rev2 69 147    Operation       Figure 4 45  Making a Sterile Connection to the Saline Container     CAUTION  START UP KIT SPIKE    The spike on the Start up Kit is relatively long  Be careful not to puncture the  side wall of the saline bag when connecting to the Start up Kit        Prime  amp  Fill the Air Trap    16  Remove the air trap from its holder and hold it upside down  with the  tubing connections pointing downward         Figure 4 46  Hold the Air Trap Upside Down     17  Prime the air trap and the tubing circuit by pressing and holding the  PRIME switch  The roller pump will slowly start and take about 20    39 600479 001 Rev 2 70 147    Operation    seconds to come up to operating speed  Observe the movement of  saline until it fills the air trap and the entire length of the tubing  The    PRIME switch will only function after the system self tests are  compl
14.  80 60 HZ 6A TSJA  slow blow   amp  X 200M  230  220 240V  50 60 HZ 2 254 TEJA  slow blow  5 X 20MM 230  220 240V  S0 60 HZ 3A TJA  slow blow  5 A 20MM    wtp  Tic     K  ZOLL Roc   lb  ZOLL ote    ZOLL Circulation  inc  J ZOLL Circulation  Inc    3040754 fe O50 Amano Awanu 3040754 650 Almanor Avenue  CONFORMS TO ip lhe a CONFORMS TO llama  UL STD 60601 1 408 841 4040 ULSTD 60601 1 406 844 4080  CERTIFIED TO CERTIFIED TO    CANICSA STD EC REP aa d CANICSA STD    22 2 NO  601 1 2813BH Den Haag   he 22 2 NO  601 1                            iit AEW   MAT 15 001 AEW      Figure 1 1  Product Label      The label provides safety information and identifies the manufacturer  model  number  serial number  power requirements  fuse capacity  and manufacturing  date for the System     9 600479 001 Rev2 12 147    Introduction    Introduction    1 600479 001 Rev2   13 147    Introduction    Contents    Introduction 3    Use of the System 3  Operating Life 3  Functional Description 3  System Components 4  Controls and Display 4  Display 4  Power Indicators 5  Alarm Indicators 5  Control Buttons 5  Control Knob 6  Serial Interface Connector 6  Recirculating Chiller 6  Temperature Controller 7  Temperature Probe Connectors 7  Pump 7  Prime Switch 8  Start up Kit 9  Data Memory 10  Saline Circuit Diagram 10    2 600479 001 Rev2 14 147    Introduction    Introduction  Use of the System    WARNING  PATIENTS MUST BE CONTINUOUSLY MONITORED   Patients being treated with the System must be checked 
15.  HMIA    To remove the HMIA  simply apply the installation instructions in reverse  beginning with step 12     Connection Cable Part Numbers    The following connection cables are available to allow the connection of the  HMIA to your hospital monitor  If you have another type of connector or do not  know which connector to use  please contact ZOLL Customer Service     Table 2 Connection Cable Part Numbers    Part Connects to  Number    500628 001   Monitors using a 3 5mm phone jack  Bard       500629 001   Monitors using a 2 5mm phone jack  Terumo        500630 001   Monitors using a 2 pin Molex type connector with socket contacts  Mallinckrodt   or  Smiths Medical       500631 001   Monitors using a 2 pin Molex type connector with pin contacts  R  sch          10 600479 001 Rev2 104 147    Accesories   HMIA  500632 001   Monitors using an RCA type Phono plug  GE    WelchAllyn   or Mallinckrodt       Data Download    You may download data from the System at any time  Simply disconnect the  HMIA RS 232 connector from the RS 232 port on the back of the display head of  the System  Then download data as per the instructions in the TempTrend  section of the System manual     This procedure will not delete the data from the System  The data is simply  copied from the system to the laptop     Reconnect the HMIA RS 232 connector when you have finished   You cannot damage the HMIA by mistakenly    downloading data    into it   The HMIA does not store data     Note   When you disconn
16.  Icy     Quattro     or  Fortius    model catheter  is indicated for use     e in cardiac surgery patients to achieve and or maintain normothermia  during surgery and recovery intensive care  and    e to induce  maintain and reverse mild hypothermia in neurosurgery  patients in surgery and recovery intensive care     2 600479 001 Rev2 24 147    Receiving  Inspection  amp  Assembly    Receiving  Inspection     amp  Assembly    1 600479 001 Rev2 25 147    Receiving  Inspection  amp  Assembly    Contents    Receiving  Inspection  and Assembly 3    Overview 3  Inspection for Damage 3    Required Tools 3  Unpacking 3  Assembly 4  Hook 4  Handle 5    Control Head 5   Control Cable 6   Power Cord 7   Condensate Pan 7   Saline Container Insulating Jacket 7    2 600479 001 Rev2 26 147    Receiving  Inspection  amp  Assembly    Receiving  Inspection  and Assembly    Overview    This chapter provides information on how to receive  unpack  and assemble the  System  If your System was delivered and set up by an ZOLL representative  you  may skip this chapter and turn to Chapter 4     Inspection for Damage    Each System is carefully inspected before it is shipped  When the carrier  delivers your System  ensure that the shipping containers are not damaged   Visually inspect the outside of the shipping container for any damage  If  damage is detected  please notify ZOLL   s customer service department and  file a damage claim with the carrier     Required Tools  To safely unpack  inspect  a
17.  MAX WARMING    i    STANDBY    A    30 Minutes iin STANDBY    Click to clear  Settings to reset     MAX COOLING       Figure 4 18  STANDBY Timer Alert    21 600479 001 Rev 2 52 147    Operation    T1 T2 Behavior   New    XP Systems differ from earlier ZOLL Heat Exchange Systems in their T1 T2  behavior i e  in how the Primary and Secondary Temperature probes are  monitored and alarmed     WARNING  Dislodged Foley Catheter Temperature Probe    The System can detect when a patient temperature probe is dislodged  suddently from the patient     However it is possible for the Foley catheter patient temperature probe  T1  to    become dislodged from the bladder to rest on the perineum or within the fold of  the thighs  In such a position the System may not detect the dislodgement  from the patient and will underestimate the patient   s core temperature  As a  result the System may inappropriately heat the patient     Failure to use a second temperature probe may result in patient injury        First Use Warning     No T2 Probe    For each time the System is powered on    1  The System will check to see if the Primary  T1  and Secondary  T2   Temperature Probes are present  The System will enter STANDBY  without probes being present  The System will not enter RUN unless the  T1 probe is present    2  When the System is first put into RUN  it will check to see which probes  are present    3  If there is no T2 present  the System will ask the User to verify that this is  intentional 
18.  Prepare the hospital monitor such that it will monitor patient temperature     9 600479 001 Rev 2 103 147    Accesories   HMIA    14  Power ON the System   15  Verify that the Green LED of the HMIA is ON     16  Connect a TP 400 temperature simulator to the T1 connection on the  front of the System  Set to 35  C     17  Using an air trap test fixture to complete the set up procedure place the  System in STAND BY     18  Verify that the patient temperature displayed by the System is also  displayed on the hospital monitor     0 1 C     Note     There may be a delay  as long as 2 minutes  before the temperature displayed    by the hospital monitor agrees  within     0 1  C  with the temperature  displayed by the System        19  Set the TP 400 to 38  C     20  Verify that the patient temperature displayed on the System screen is  also displayed on the hospital monitor     0 1  C     21  Remove the TP 400 and verify that the hospital monitor recognizes the  loss of patient temperature  Typically the hospital monitor will indicate  that the probe is DISCONNECTED     22  Reconnect the TP 400    23  Verify that the patient temperature displayed on the System screen is  also displayed on the hospital monitor     0 1  C     24  Power OFF the System     25  Verify that the hospital monitor recognizes the lack of input  Typically the  hospital monitor will indicate that the probe is DISLODGED     26  The System with the HMIA module is now available for clinical use     Removal of the
19.  See image below     Secondary Temp  Probe  T2  Not Connected    If you do not use the T2 probe   you must use  another secondary means of monitoring the patient  temperature  Otherwise failure of the primary  temp probe  11  may result in patient injury     Do you wish to continue     2006 03 05   18 26       Figure 4 19  No T2 Probe Connected    The System remembers the User   s preference     22 600479 001 Rev2 53 147    Operation    T2 Probe Disconnection Reconnection    If the User connects a T2 probe at any time during the operation of the System   in STANDBY or RUN   the System will assume that the use of the T2 probe is  desired  If the T2 probe is then disconnected the System will alarm     see below     If the User disconnects a T2 Probe in STANDBY  the System will NOT assume  that this is intentional  When the System is placed into RUN WITHOUT a T2  Probe present  the User will again see the Warning as described above     Accidental Disconnection T1 T2 Probe    MAX WARMING    PATIENT TEM P  Disconnection of the T1 probe during RUN results in    an Alarm  Press the knob to silence the alarm  The    System moves to STANDBY  Treatment cannot 4  continue until the T1 probe has been replaced        Note that the patient temperature is displayed as             and the yellow warning banner covers the bath   temperature display  Primary temp probe  T1  disconnected  Press Enter to proceed    MAX WARMING  PATIENT TEMP    Absence or disconnection of the T1 probe during  STA
20.  The default    directory is  C  TempTrendCSV Data    SE    Save in   E TempTrendCS   Data   do ft ee EE     TestS  cs         File name  JF csv  Save as  ppe   esv    Cancel    fe    Figure 5 5  Select File Dialog Box  2  You can either select an existing file name or create a new one     e To select an existing file name  double click on it  All existing data within the  file will be over written with the new download     e To enter anew file name  replace the   csv with the file name of your  choosing  Type it in place of the   csv in the file name box     ry Temp TrendCS   P N103710 Rev  A    om Fort          CA Teamplrendcs Datasdemo  cew       Patient Log File Name        Select File Download    Figure 5 6  Fully Qualified File Name Displayed    The full path name of the patient data log file that you selected is displayed above the  Select File icon     7 600479 001 Rev2   115 147       TempTrend CSV    Download    Click on this icon to download data from the System  Both the Select File and  Download icons will grey out during the download  TempTrend    CSV will display a    complete message during download  After the download is complete  both the Select  File and Download icons will appear normal again     Ole    Click on this icon to exit from TempTrend    CSV     Using the Alsius Add in    The Alsius Add in contains Microsoft macro codes that allow the viewing of the patient  data file  These are accessed via the Alsius menu item in the EXCEL menu bar     EJ Micro
21.  The low patient  temperature alarm limit        4 600479 001 Rev2 7 147    Safety Information    General Safety Precautions    WARNING  SYSTEMIC HYPOTHERMIA RISKS     Systemic hypothermia may cause cardiac arrhythmia  patient shivering  or  other system or organ complications  Systemic hypothermia should only be  utilized under the supervision of a qualified physician        When treating a patient with the System  appropriately qualified medical staff  must routinely and closely monitor the patient and must comply with the following  procedures     e Audible and visual alarms generated by the System require the  authorized individual to remain in close proximity to the patient  throughout the procedure     e Always verify the function of the System prior to insertion of an Alsius  catheter  In the event of a malfunction  have other means of cooling  available     e When combining the use of the System and other adjunctive means  of cooling  ensure that close observation of the patient is maintained     e Do not use the ZOLL System in conjunction with other temperature  maintenance devices that have an automatic temperature controller   Temperature oscillations may occur that are dangerous to the  patient     e Performance of installation  operation  or maintenance procedures  other than those described in this manual may create hazards and  may cause the manufacturer s warranty to become void     e Sterile components are designed for a single use only  If  unauthorized disp
22.  but NOT inserted into the patient you will experience    Lo    patient  temperature alarms if the ambient temperature of the exposed probe is below  the alarm limit  To avoid the annoyance of the alarm under these conditions   simply unplug the T1 temperature probe from the System until you are ready to  connect to the patient  You may not enter RUN without the probe connected  to the System        Bath Pre set    The menu allows the user to select from either  Pre Cool  Pre Warm or None   This selection operates only in STANDBY and is cancelled when the system  enters RUN     SETTINGS    PATIENT TEMP   Sath Pre set     Pre Cool    Pre Warm    None    Exit    2006 03 05   18 26       Figure 4 13  Bath Pre Set Menu    17 600479 001 Rev 2 48 147    Operation    The System bath is cooled to its lowest permitted temperature and maintained  Pre Cool at that temperature    The System bath is heated to its highest possible temperature and maintained  Pre Warm at that temperature     The Bath Pre set is not activated or is cancelled  If cancelled  the bath is  maintained at the temperature that was measured at the time of cancellation     Exits the menu without a change in System status or programming        Time and Date  This menu displays the current time and date settings   PATIENT TEMP SETTINGS  Time and Date    Select Hour   Press enter to set     STANDBY    TARGET TEMP RATE 18 26    2006 03 05    2006 03 05   18 26       Figure 4 14  Time and Date Settings     The time is div
23.  cloth  moistened with water   Clean rollers and apply light lubri cating oil if  rollers have been in contact with saline  solution     Annually Drain and refill with new coolant mixture     Power cord Annually Inspect for wear or damage     Temperature accuracy Annually Perform a temperature accuracy test  refer to  the    Temperature Accuracy Tes     section of  this chapter      NOTE TO INFECTION CONTROL     50  propylene glycol coolant has been evaluated for antimicrobial  effectiveness and determined not to promote growth of gram negative rods   gram positive cocci  or yeast microbes  50  propylene glycol was found to  be comparable to 70  isopropyl alcohol        4 600479 001 Rev2 134 147    Maintenance    Unscheduled Maintenance    The following table lists unscheduled maintenance tasks for the System  These  tasks should be performed when indicated  Detailed procedures for some tasks  are listed later in this chapter     Table 8 2  Unscheduled Maintenance     When deterioration is Inspect mechanical and electrical   evident components for wear and loose or  deteriorated parts  Verify continuity of the  electrical ground connection     System console When soiled Clean the exterior of the console  refer to  the    Clean Console and Condensate Pan     section of this chapter      Coolant If contaminated Drain coolant  clean the coolant well  and  refill with 1 1 mixture of distilled water and  propylene glycol  refer to the    Drain  Coolant    section of this chapter  
24.  in the System coldwell  as  depicted above   or     e To present menu selections  The main menu of the System is  presented in the figure below     10 600479 001 Rev2 41 147    Operation    PATIENT TEM P    View Graphs    A    New Patient  End Procedure    Settings    Exit    2006 03 05   18 26       Figure 4 5  System Main Menu    Changing Target Temperature    1  Press the Target Temp button once  If you are in RUN  you will be taken  to STANDBY  The Target Temperature value may only be changed in  STANDBY     TARGET TEMP    PATIENT TEMP    To change Target  Temperature or Rate   system must be in  Standby  Would you  like to go to Standby  mode     2006 03 05   18 26       Figure 4 6  Change Target Temperature in RUN    2  To change the target temperature  turn the knob until the desired  temperature is displayed  You may choose a temperature from 31   C to  38   C  87 8   F to 100 4   F   When the correct selection is displayed   press the knob once to enter the selection     11 600479 001 Rev 2 42 147    Operation    3  The target temperature setting is updated on the Operating screen     Changing Treatment Mode    As explained above  there are three treatment modes     MAX        Controlled  Rate    and    FEVER     To select the Treatment Mode  including one particular  value of    Controlled Rate    you use the    Rate Deg Hr    button     1  Press the    Rate Deg Hr      r button once  If you are in RUN  you will be  taken to STANDBY  The Treatment Mode value may on
25.  iv       0021     001    CAN CSA STD    rPo97 9   Me ens  C22 2 NO 601 1    HMIA 601033    RS232 T1 out SYSTEM    Figure 3 Front Panel of the HMIA    The meaning of the symbols on the front panel are as described in the table  below    Table 1 HMIA Control Indicators    Symbol    RST Reset Button This is a reset button  Lightly push the tip of a  paperclip  or similar sized device  into the hole  until you feel a    click    and then withdraw the  paper clip    Caution LED This red light indicates that there is a  malfunction in the HMIA   Power LED This green light indicates that the HMIA is  C  receiving power from the System     RS 232 Serial port connector   This short extension cable is connected to the  cable RS 232 data port connector on the back of the  display head of the System     T1 out Patient temperature output This connector accepts the custom Alsius  connector interface cable     SYSTEM   System umbilical cord One connector is attached to the round  multi   connections pin connector on the back of the display head  of the System via a short cable  The other  connector mates with the umbilical cord of the  System        5 600479 001 Rev2 99 147    Accesories   HMIA    Operating States    The HMIA has the following operating states  START UP and RUN     START UP    When the System is powered on  illumination of the Green LED is indication that  power has been supplied to the HMIA  If there is no power to the HMIA the  Green LED is not lit and the hospital monit
26.  of these tasks are performed when  an inspection indicates that they are necessary  or after a problem is discovered     Failure to perform the maintenance tasks listed in this chapter may cause  degraded performance of the System or may reduce the operating life of its  parts     IMPORTANT     System troubleshooting must be performed only by qualified personnel     Observe all safety warnings and use the required safety equipment when  performing troubleshooting and repairs        Safety Precautions    Required personal protective equipment  PPE     e Safety glasses  e Latex or neoprene gloves    Required tools and materials    e Phillips screwdriver   e Screwdriver   e Inspection lamp or flashlight   e Vacuum cleaner with crevice tool    3 600479 001 Rev2 133 147    Maintenance    Scheduled Maintenance    The following table lists scheduled maintenance tasks for the System  Detailed  procedures for some of these tasks are listed later in this chapter     Table 8 1  Scheduled Maintenance     Coolant reservoir Before each use Inspect and refill with distilled water to  indicated line if necessary     Condensate pan After each use Inspect  Empty condensate if necessary  refer  to the    Clean Console and Condensate Pan     section of this chapter      Refrigeration condenser Every 6 months Inspect the filter and replace if necessary  filter  refer to the     nspect Replace Condenser  Filter    section of this chapter      Roller pump Monthly Clean rollers and tubing path with
27.  patient  without powering off the system  follow these steps in the  indicated order     Press the    Press for Enter Menu    knob once  The menu is displayed     2  Turn the knob until the    New Patient    selection is highlighted  Press the  knob once to enter the selection    3  The New Patient menu will appear  The screen displays the message     Select Yes to Delete Previous Patient Data     Press the knob once to  enter    Yes       4  The screen displays the message    Are you sure you want to delete the    previous patient data     Press the knob once to enter    Yes        5  The screen briefly displays the message    Previous patient data deleted     and then displays the standby screen     Disposal of Used Components    WARNING  SINGLE USE DEVICE   DO NOT REUSE     Alsius catheters and System Start up Kit components are single use devices    and may not be reprocessed or reused  The cyclical stresses of the peristaltic  pump on the catheter and Start up Kit cause fatigue failures     DO NOT Use Catheters or Start up Kits Beyond the Labeled Usage Time   Failure of the product will result        54 600479 001 Rev 2 85 147    Operation    CAUTION  AVOID CONTACT WITH USED COMPONENTS     Handle used components as medical waste        To remove used components from the System and dispose of them properly   follow these steps in the indicated order     1  Assure that the patient has been disconnected from the System and that  the power switch has been turned off     2  
28.  reasons  either     1  The System is not connected and powered on  In which case simply connect the  System  power it on and try again     or    2  You have a non standard Microsoft installation  In which case you must specify the  Com Port to use     To find out which port is being used  look inside    Device Manager     You will select the  Com Port that matches your USB Serial Adaptor     zox       Fie Action       a  lig Computer   Es Disk drives   H a Display adapters   Fl DWD CO ROM drives   H a Human Interface Devices   HE IDE ATA ATAPI controllers       Keyboards   Ty Mice and other pointing devices  fae Modems   i Monitors   Ee Network adapters   ly Ports  COM  amp  LPT     be Sitecom USE to Serial Adapter CN 1o4v2  Com2    COM   H Processors   pe Sound  video and game controllers  fl yf System devices   na Universal Serial Bus controllers    ee   ee      Figure 5 4  Device Manager Display    Select the communications port for your System cable connection  Click on the button     Com Port    and select the appropriate one of the ports listed     6 600479 001 Rev2 114 147    TempTrend CSV    HINT  It is safe to just try each port setting offered within TempTrend    CSV until  you find the one that works  TempTrend    CSV will not allow you to select a port  that is not connected to the System  You will simple receive a warning message as    per Figure 5 3           Select File  Click on this icon to   1  Select the directory into which you want to save the data logs 
29.  to support catheter insertion     CAUTION  REFER TO THE CATHETER IFU    Alsius heat exchange catheters are inserted via a Seldinger technique as a  central venous line  There are specific instructions for use included with each  catheter  Refer to these to understand the specific unique insertion  requirements of an Alsius heat exchange catheter        Preparing the System for Treatment    To prepare the System for treatment  follow these steps in the indicated order     1  Roll the System to a convenient position near the patients bedside  Plug  the power cord into a hospital grade receptacle    2  Lock the right front caster by stepping down on the tab above the wheel    3  At the rear of the console  near the upper left corner is the power on off    switch  Turn the switch ON        Figure 4 22  Power Switch and Power Indicator Lamp     4  The green power indicator lamp will be illuminated and the alarm will  beep one long beep followed by a shorter softer beep   9  The System performs a self test  During this interval  the self test screen    appears on the display  If the self test detects a problem  an error    26 600479 001 Rev 2 57 147    Operation    message will be displayed  If this occurs  refer to Chapter 6   Alarms and  Corrective Actions for assistance  At the end of the self test there will be  two short beeps     ALSIUS    Self Tests in Progress        Software Version 2 00   1B   36   1 57    2006 03 05   18 26          Figure 4 23  Self Test Screen     Whe
30.  will be updated and the screen will display the    settings menu       C   F  Temperature Notation     This menu displays the options setting for temperature notation  The currently  selected setting is highlighted     PATIENT TEM P SETTINGS     CIF    Please select  Celsius or  Fahrenheit     STANDBY    Calsius    Fahrenheit    2006 03 05   18 26       Figure 4 15  Temperature Notation Settings   To keep the current selection  press the knob once  The current setting will not  be changed and the settings menu will be displayed     To change the current setting  turn the knob to highlight the desired setting   Press the knob once to enter the selection  The setting will be changed and the  settings menu will be displayed     19 600479 001 Rev2 50 147    Operation    NOTE  Change in Design     Pump Rate    In earlier versions of the System  there was the option to set the pump rate to  either 200 or 240 ml min  ALL Alsius catheters have been validated to  operate safely at a pump rate of 240 ml min  This control is not longer  presented to the operator     NOTE  Change in Design   Alarm Volume  amp  Key Press  Volume    In earlier versions of the System  there was the option to set the alarm volume  and the key press volume  In this version     1  The Alarm Volume is always set to High       2  The Key Press Volume is always set to Medium    Language    This menu displays the current setting for the language used for displayed text   The currently selected setting is highligh
31.  will reset  when the patient s  temperature falls below the  programmed Hi patient  temperature alarm value     The System alarm will reset  when the patient s  temperature rises above the  programmed Lo patient  temperature alarm value     Verify position of primary  temperature probe  Assure  that primary probe is plugged  into socket T1  or if an  interconnect cable is used  the  probe is plugged into the cable  and the cable is plugged into  Ti     Replace probe in patient        The pump cover is open  Close the cover     PUMP FAILURE     The pump has failed     Inspect the pump  Clear any  obvious faults and restart  If  the problem persists     discontinue use and contact  your ZOLL service  representative  see Chapter  9         600479 001 Rev 2    123 147    Alarms  amp  Corrective Actions    Alarm Message Cause Corrective Action    SECONDARY  TEMPERATURE  PROBE  DISCONNECTED     SECONDARY  TEMPERATURE  PROBE DISLODGED     TEMPERATURE  PROBES DO NOT  AGREE     SYSTEM  MALFUNCTION      The secondary temperature  probe is disconnected     Temperature output from the  secondary patient temperature  probe changed by    0 2   C or  more within 10 seconds     Temperature outputs from the  primary and secondary patient  temperature probes differ by  more than 2  C     May be caused by a variety of  failures     Verify position of secondary  temperature probe  Assure  that primary probe is plugged  into socket T2  or if an  interconnect cable is used  the  probe is plugged 
32. 600479 001 Rev 2 17 147    Introduction    Press the    Standby Run    button to toggle the operation of the System between  standby mode  the pump is stopped  or run mode     An alarm or fault can place the System into standby mode automatically  After  remedying the condition that caused the alarm  press this button to return to run  mode     Silence Alarm    Press the silence alarm button to silence the audible alarm tone for two minutes   120 seconds   If the alarm condition has not been cleared during this two minute  period  the audible alarm will sound again     Control Knob  The    Press for Menu Enter    control is a dual function control knob and  pushbutton     Press the knob to display a menu screen or to indicate the completion of a  selection     Turn the knob to scroll between selections or to scroll temperature trend graphs     Serial Interface Connector    A female 9 pin subminiature D connector is mounted on the lower left corner of   the rear of the display head  Use this connector to attach a serial interface cable  between the System and a laptop computer  Once connected  the computer can  download patient temperature trend data stored by the System        U a       Figure 2 3  Serial Interface Connector     Recirculating Chiller    The chiller consists of an air cooled refrigeration system  reservoir heater   circulation pump  stainless steel reservoir  reservoir cover  and a temperature  controller     6 600479 001 Rev2 18 147    Introduction    Tempera
33. A  Pushbuttons and knob   Thermistor  YSI 400 series   180   swivel  45   tilt   Serial RS 232C  9 pin sub D connector  Audible tones and displayed text messages  26  C     42  C    0 2  C    2 600479 001 Rev2 146 147    SALINE COOLANT CIRCUIT  Priming volume    Heat exchanger    Priming source    Patient connection     Pump tubing    Sterility    Saline alarm    Coolant circuit operating life     EQUIPMENT CLASSIFICATIONS  Type of protection against moisture     Type of protection against  electric shock     Protection class   Mode of operation     APPROVED PATIENT TEMPERATURE  PROBES    Temperature Probe Standard    YSI 400 Temperature probes verified by  ZOLL to work with the System    YSI 400 Temperature probes verified by  ZOLL to work with the System and Rusch  patient interconnection cable  Rusch P N   179516      ALSIUS START UP KIT    Tubing set including heat exchange coil   airtrap and roller pump tubing     3 600479 001 Rev 2    Specifications    200 ml  Disposable stainless steel coil  Sterile saline solution  hospital provided     Directional Luer connections on  72 in   183 cm  lines    Roller pump compatible with directional fittings  Gamma sterilized  Reservoir level detection  amp  alarm system    Replace disposable components after seven  7   days of continuous use    Ordinary    Type BF for temperature inputs  Type B for catheter connections    i  Continuous    YSI 400    MALINCKRODT MEDICAL FOLEY PROBE 8F  200818 001   MALINCKRODT MEDICAL FOLEY PROBE 10F 
34. A to ZOLL for service     Incorrect Temperature Reported On Hospital Monitor    1   2    3     Disconnect the hospital monitor from the HMIA interface cable   Using a DMM measure the resistance between the two pins     Using the chart YSI 400 Temperature vs  Resistance determine if the  resistance value measured corresponds to the temperature dispalyed on  the System display     If the resistance value is correct then the HMIA is functioning to  specifications     lf the resistance value is wrong  then test Alsius interface cable for shorts  or opens     12 600479 001 Rev2 106 147    Accesories   HMIA    If it passes the test  then press the reset button  RST     Wait for 20 seconds and then repeat step 3 above     If the resistance value is still wrong  then disconnect the RS232 cable  from the rear of the display housing  wait for 15 seconds and then  reconnect     9  Wait for an additional 20 seconds for the HMIA to re establish  communication with the System    10  If the resistance value is wrong then return the HMIA to ZOLL for service     Temperature jumps between 26  C and 49  C On Hospital    Monitor    There has been an internal component failue  Return the HMIA to ZOLL for  service     Hirose Connector does not fit    There are two versions of System Hirose connectors in clinical use  a 6 pin and  an 8 pin  If the Hirose connector does not fit  count the number of pins in the  HMIA cable  If this is not the same as the number of pins in the umbilical cable  of the 
35. Circulation  Inc   at  its sole discretion  shall repair or replace the defective parts without charge to the  purchaser     ZOLL Circulation  Inc  reserves the right to make any necessary repair at its  factory  at any authorized repair station  or at the facilities of the purchaser     Replacement parts may be new parts or remanufactured parts  Parts replaced  under warranty become the property of ZOLL Circulation  Inc      This warranty does not apply to repairs or replacements to correct damage  resulting from mishandling or abuse  failure to perform the required maintenance  described in the operator s manual  voltage surges  or casualty losses caused by  man or by nature     This warranty shall be void if repairs or attempted repairs have been performed  on the System by any person not authorized by ZOLL Circulation  Inc  to perform  repairs     The express warranty above is the sole warranty obligation of ZOLL  Circulation  Inc   and the remedies provided above are in lieu of any and all  other remedies  There are no other agreements  guarantees  or warranties   oral or written  expressed or implied  including without limitation any  warranty of fitness for a particular purpose or merchantability  ZOLL  Circulation  Inc  shall have no liability whatsoever for any improper use or  improper repair of the System     Technical Support and Resources    ZOLL Circulation  Inc  provides field  and factory based technical support for its  products  Our service hotline can answe
36. Cross connect the loose ends of the Start up Kit when you disconnect it  from the catheter  This will reduce the amount of fluid that has to be  cleaned up later     3  Position a large empty medical waste container or collection bag near  the System     CAUTION  DO NOT DISASSEMBLE STARTUP KIT     Do not disconnect the tubing connecting the components  They are removed  and disposed of as a complete unit  To avoid injury  do not disconnect the  saline solution container        4  Open the top cover of the System  Open the transparent top cover of the  roller pump     Remove the cap from the coolant well and set it aside in a clean location   Remove the insulating jacket from the saline container and set it aside   Lift the handle on the pump rollers     ONO    Grasp the pump tubing and gently pull it up and out of the channel while  rotating the pump head     9  Pull the tubing out of the pump head  Press the handle down onto the  rollers until it press into the detent and close the top cover of the roller    pump   10  Loosely coil the disconnected ends of the catheter tubing     11  Lift the heat exchanger coil out of the coolant well  Hold the coil above  the coolant well to allow coolant to drip back into the well     12  Pull the air trap up and out of the holder   13  Unhook the saline solution container from the hook     14  Bundle up the collection of components  still connected together  and  gently deposit them in the medical waste container  Take care to avoid  con
37. D  5  TEMPERATURE PROBES DO NOT AGREE  5  SYSTEM MALFUNCTION   5      Stops operation of the System     2 600479 001 Rev2   121 147    Alarms  amp  Corrective Actions    Alarms and Corrective Actions    Overview    This chapter lists alarm messages that may appear on the System display during  operation  If an alarm occurs  refer to the following table for information about the  Cause and corrective actions you can take to remedy the problem     Please note that in many cases the System will automatically restart when  the condition that generated the alarm has been cleared  In other cases   you must take corrective action and then restart the System     Alarms and Corrective Actions    Alarm messages are listed in alphabetical order     Stops operation of the System     Table 6 1  Alarm Messages  Causes  and Corrective Actions     Alarm Message Cause Corrective Action    R Air has been detected in the Verify that the air trap is  AIR TRAP FAULT air trap  completely filled with saline   Inspect tubing circuit for  leaking saline  Refer to    Air  entry into the tubing circuit    in  the Preface   If the problem persists   discontinue use and contact  your ZOLL service  representative  see Chapter  9    The air trap is full of saline Completely dry the outside of  but the saline level detector the air trap with a towel to  indicates a fault  remove any condensation   Verify that the air trap is  seated firmly at the bottom of  the air trap holder   If the problem persists   di
38. HMIA       Figure 5 The Hirose Connector     Rear of the Display Head    Verify that the connector type on the end of the Umbilical Cable is  identical to the connector on the HMIA  601033   001 for the 8 pin circular  connector and 601033   002 for the 6 pin circular connector  If the  connectors are incompatible then contact ZOLL to arrange for a  replacement     Remove the paper protecting the Velcro   strips found on the mounting  surface of the HMIA     Hang the HMIA on the stainless steel bracket located on the back of the  Display Housing  The connectors on the HMIA should be facing  downward     Press on the lower half the HMIA until the snaps engage   Connect the Umbilical Cable to the mating connector on the HMIA    labelled SYSTEM     Connect the cable with the circular connector to the mating connector on  the rear of the Display Housing     8 600479 001 Rev2 102 147    9        Accesories   HMIA       Figure 6 RS 232 Connector     Rear Display Head    Connect the HMIA RS232 connector  labelled RS232  to the mating  connector on the Rear of the Display Housing   See Figure 6         E       Figure7 Installed HMIA     Rear Display Head      Select the appropriate custom Alsius interface cable that mates with the    patient cable of the hospital temperature monitor       Insert the phone jack of the selected cable into the HMIA connector    labelled T1 out       Connect the other end of the Alsius interface cable to the patient cable of    the hospital monitor      
39. NDBY results in a warning display WITHOUT a  persistent audible alarm  The system CANNOT  enter RUN until a T1 probe has been connected   Note that the patient temperature is displayed as            and the yellow warning banner DOES NOT  cover the bath temperature display     MAX COOLING    f     MAX WARMING  Disconnection of the T2 probe during RUN results in PATIENT TEMP    an Alarm  Press the knob to silence the alarm  The N  System moves to STANDBY  If you attempt to    return to RUN without reconnecting the T2 probe     you will be asked to verify your intention     Note that the patient temperature will be displayed   correctly if the T1 probe remains connected  The   yellow warning banner covers the bath temperature  secondarytemp probe  12  Gaconnected  display  Press Enter to proceed         23 600479 001 Rev 2 54 147    Operation    Alarms  amp  Alerts    A detailed description of the alarms of the System is provided in later sections of  this manual  See     Section 6  Alarms and Corrective Actions  Section 7  Troubleshooting    Alerts    Alerts can be cleared by reverting the state that caused them to occur  For  example  a low coolant alert can be cleared by adding coolant to the coldwell of  the System  In most cases  the System will identify and notify you to rectify alert  states at power up     In STANDBY  Alerts are displayed across the lower half of the display screen   with the bath temperature display not obscured  against a yellow background   See Figur
40. System you have the wrong version of HMIA  If you have the wrong  version of the HMIA  then contact ZOLL to arrange for a replacement     Table 3 HMIA Part Numbers    601033    001   8 pin circular Hirose connector  601033    002   6 pin circular Hirose connector    13 600479 001 Rev2 107 147    Accesories   HMIA    YSI 400 Temperature vs  Resistance        F         ohms    omo   5   252  zs   26   as  sea   B   197  a2   3   ia    o   3   i5  o eas   a   733  s6   32   1667  o ota   3   159    932   a   383  so   a   o oun  o8   3   ma                             98 6 37 1355  1301   o3   tag    40 1200    1152  1107  1064  1023  983 8  946 2  910 2  8795 8  842 8  811 3    14 600479 001 Rev2 108 147    TempTrend CSV    TempTrend CSV       1 600479 001 Rev2 109 147    TempTrend CSV    Table of Contents  TempTrend CSV Overview 3    Installation 3    TempTrend    CSV Installation 3  Automated Installation   Autorun 3  Manual Installation 3  Alsius Add In Installation 3  Automated Installation     Standard    Windows    Setup 4  Manual Installation 4    Using TempTrend    CSV 4    What TempTrend    CSV Does 5  Starting TempTrend    CSV 5  Com Port 5    Select File 7       Download 8  Ol       Using the Alsius Add in 8    Open CSV 8  Graph CSV 8  Activate Scroll Control 9    The Scroll Control 10    Location 10   Zoom 10   Scroll 10   Limits of Scroll Control 11    Uninstalling TempTrend    CSV 11    2 600479 001 Rev2   110 147    TempTrend CSV    TempTrend CSV Overview    Temp
41. T be  monitored by a separate hospital patient temperature monitor     5  Place the Alsius catheter in the patient now  Refer to the Instructions for  Use for information about the catheter   6  The tubing to the catheter is supplied with the supply and return    connectors connected to each other  Using aseptic technique   disconnect the two connectors     42 600479 001 Rev 2 73 147    Operation       Figure 4 51  Disconnect the Connectors Using Aseptic Technique     7  Connect the female tubing connector to the male connector on the  catheter  The gender of the tubing connectors and the catheter  connectors assures that they cannot be connected backwards   ze h fst tm       Lia T        Figure 4 52  Connect the Tubing to the Catheter     8  Connect the male tubing connector to the female connector on the  catheter  Note that the return line is equipped with an inline flow   indicator    9  Safely route the tubing so that it is not kinked or obstructed and cannot    be easily dislodged by a patient   s movement   The System is now ready to begin treatment     Once treatment has begun  confirm that saline is flowing through the tubing and  catheter circuit by observing the rotation of the inline flow indicator  see figure   If    43 600479 001 Rev 2 74 147    Operation    the flow indicator does not rotate freely during patient treatment  inspect the  entire tubing circuit for kinks or other restrictions to flow  Tapping on the flow  indicator will liberate any trapped air bu
42. The next time the System is switched on  the self test program detects the saved  patient data and offers the user the option to download the data to a laptop  computer  If the user does not wish to download the saved data  the System will  delete the data before permitting set up for a new patient     To download patient trend data after an improper shutdown  follow these steps in  the indicated order     1  Plug in and switch on the System  The self test screen appears on the  display     ALSIUS    Self Tests in Progress        Software Version 2 00   1B   36   1 57    2006 03 05   18 26       Figure 4 55  Self Test Screen     2  When the self test is finished  the System Set Up screen displays the  message    Bath Pre Set        46 600479 001 Rev2 77 147    Operation    System Set Up    Bath Pre Set     Pre Warm  None      2006 03 05   18 26    Figure 4 56  Bath Pre Set           Choose    Pre Warm        Pre Cool    or    None    as desired     When the System detects the presence of saved patient trend data  the  System Set Up screen displays the message    New Patient        System Set Up    New Patient       2006 03 05   18 26    Figure 4 57  New Patient  Message              Choose    Yes        The screen displays the message    Previous patient data must be  downloaded or deleted to proceed        47 600479 001 Rev 2 78 147    Operation    System Set Up    Previous patient data must be  downloaded or deleted to proceed     Download Now  Delete  Exit      2006 03 05 
43. Trend CSV is used to download patient data logs from the System     The use of the CSV format is intended to allow the System user the greatest flexibility in  analyzing data downloaded from the System     The Graph CSV Add In converts the data log file from a     csv    format to an EXCEL     xls     format file  In this way  it facilitates graphing of the patient data that was downloaded  using TempTrend CSV     Installation    The TempTrend    CSV installation CD is designed to work with any version of Microsoft  Windows  Your system may have a configuration or security features that prevent the  installation from working  If you experience trouble with this installation  please send us  a description of the trouble you are having via email to     complaints zollcirculation com  We will respond quickly to your email     TempTrend    CSV Installation    Automated Installation   Autorun    TempTrend    CSV is supplied on a CD with an autorun feature  Simply put the CD into  a CD drive on your computer  A series of dialog boxes should guide you through  installation  If the install program has worked correctly  you should have TempTrend     CSV installed on your computer  See  Using TempTrend    CSV     Manual Installation    If your Microsoft Windows is a non standard setup  the Alsius Add In will have to be  loaded manually     1  Using Microsoft Windows Explorer  open the TempTrend    CSV installation CD  and locate the file  TempTrendCSV exe    Copy the file to a director
44. ZOLL Circulation       Operation Manual       600479 001 Rev 2 1 147    Copyright    2009 ZOLL Circulation  Inc    All rights reserved  Printed in U S A     Trademarks   Alsius  CoolGard  CoolGard 3000  Thermogard XP  Cool Line  Fortius  and  Icy are trademarks of ZOLL Circulation  Inc    Mallinckrodt is a registered trademark of Mallinckrodt Inc    Windows is a registered trademark of Microsoft Corporation     Other products and names listed in this document may be trademarked by their  owners and no representation is made by ZOLL Circulation  Inc  as to rights  thereto     U S  Patents Pending    This product is covered by or for use under one or more of the following U S   patents  Other U S  and foreign patents pending     4 290 428 4 648 384 4 689 041 4 850 969 4 865 581  4 917 667 4 927 412 5 011 468 9 033 998 9 059 167  9 908 407 6 019 783 6 126 684 6 149 670 6 165 207  6 287 326 6 290 717 6 299 599 6 338 727 6 368 304  6 393 320 6 405 080 6 409 747 6 416 533 6 419 643  6 432 124 6 436 130 6 447 474 6 450 990 6 451 045  6 454 792 6 454 793 6 458 150 6 460 544 6 494 903  6 916 224 6 520 933 6 929 775 6 930 945 6 530 946  6 554 797 6 572 640 6 581 403 6 582 398 6 585 692  6 589 271 6 602 243 6 620 131 6 641 602 6 641 603  6 645 233 6 645 234 6 652 565 6 682 551 6 699 268  6 706 060 6 709 448 6 716 188 6 716 236 6 719 724  6 726 653 6 726 710 6 733 517 6 749 585 6 749 625  6 755 851 6  86 916 6 796 995 6 872 222 6 878 156  6 893 419 6 893 454 6 942 644 7 001 418 7 014 651
45. a log file are both  visible and    active    on your desk top  i e  both visible and both the focus for a mouse  click      Uninstalling TempTrend    CSV    The automated installation program installs an    Uninstall    option with TempTrend     CSV  To uninstall TempTrend    CSV     e Open the TempTrend CSV folder within your Program menu  e Select Uninstall    e You will be asked to confirm that you want to uninstall and given a progress  report at the end of the uninstall process       ZOLL Circulation  Inc  2006 TempTrend  GraphTrend  CoolGard 3000 are registered  trademarks of ZOLL Circulation  Inc  No part of this document or software may be  reproduced or reused without prior written authorization of ZOLL Circulation  Inc     Microsoft  Windows  EXCEL are registered trademarks of Microsoft Corporation     11 600479 001 Rev2 119 147    Alarms  amp  Corrective Actions    Alarms  amp  Corrective       Actions    1 600479 001 Rev2 120 147    Alarms  amp  Corrective Actions    Contents    Alarms and Corrective Actions 3    Overview 3   Alarms and Corrective Actions 3  AIR TRAP FAULT  3  COOLANT EMPTY  3  COOLANT LOW 3  DOWNLOAD ERROR  PLEASE CHECK EXTERNAL COMPUTER 4  PATIENT DATA LOG FULL  PLEASE DOWNLOAD NOW 4  Hi PATIENT TEMPERATURE ALARM 4  Lo PATIENT TEMPERATURE ALARM 4  PRIMARY TEMPERATURE PROBE DISCONNECTED  4  PRIMARY TEMPERATURE PROBE DISLODGED  4  PUMP DOOR OPEN  4  PUMP FAILURE  4  SECONDARY TEMPERATURE PROBE DISCONNECTED  5  SECONDARY TEMPERATURE PROBE DISLODGE
46. allation CD  and locate the file  Alsius xls    Double Click on  Alsius xls  Alsius xls should open in Microsoft EXCEL  Follow the instructions in this file     Using TempTrend    CSV    You must connect the System to your computer and power it on before you can use  TempTrend    CSV to download data logs  Make this connection as described in the  System Operations Manual  To do this you will need either     e An RS 232 serial cable if your computer has an RS 232 port  or    e AUSB Serial Adaptor cable if your computer does not have an RS 232 port  but does have a USB port     Your computer will assign a communications port to this connection  a    Com Port         TempTrend CSV will attempt to determine which Com Port is in use by System with each  download  However  if you have a non standard Microsoft installation  you may have to  specify the Com Port in use  See below  Com Port    There are MANY non standard USB Serial Adaptor cables in the personal computer  marketplace  Please ensure that     e You have loaded all the recommended drivers for your USB Serial Adaptor cable    and that it uses a communication port between Com1 and Com  s     e You do not have another device attempting to use the same communications  port as TempTrend    CSV        4 600479 001 Rev2   112 147    TempTrend CSV    What TempTrend    CSV Does    TempTrend    CSV downloads the data logs from the System  It creates three files in  comma separated values format  CSV   You are asked to supply a fi
47. ata and provides details about the format and structure of the  downloaded data     Displaying the Temperature Trend Graph    To display the temperature trend graph  follow these steps     1  Press the Press for Enter Menu knob once  The screen displays the  menu  The selection    View Graphs    will be highlighted     PATIENT TEM P    View Graphs    New Patient  End Procedure    Settings    Exit       Figure 4 67  Menu     2  Press the knob once  The screen will display the temperature trend graph     57 600479 001 Rev 2 88 147    Operation    Temperature Trend Graph    Temperature trend data can be displayed as an interactive graph on the screen   The display is a time series of patient temperature  from the primary temperature  probe  and System activity  plotted in two graphs  The patient temperature graph  plots temperature vertically and time horizontally  The System activity graph plots  cooling warming activity vertically and uses the same time scale horizontally  An  example of the patient temperature trend display is shown below     TempTrend v    Figure 4 68  Temperature Trend Display        Patient Temperature    The patient temperature graph  labeled    Patient Temp     is scaled for any  temperature between 31   C and 41   C  87 8      105 8   F   The time scale can be  set to any of four intervals  refer to Setting the Time Scale  later in this chapter   for details   Figure 4 68 shows the vertical temperature displayed in degrees  Celsius  and the horizontal t
48. aved from the improper shutdown is deleted     49 600479 001 Rev2 80 147    Operation    14  After about two seconds  System Set Up screen displays the message     Set up system now  Press Enter to proceed       15  Disconnect the serial interface cable from the System and the laptop  computer     Downloading of patient trend data is now complete  To set up the System for a  new patient  return to the section in this chapter titled Preparing the System for  Treatment     CAUTION  There is no    Undelete       If you choose    Delete     the patient s trend data will be permanently deleted  and cannot be recovered later        50 600479 001 Rev2 81 147    Operation    Ending Treatment    End Procedure    Use of the    End Procedure    menu option ensures that the patient data log is  closed and cleared so that the System is ready to receive the next patient  If the     End Procedure    menu option is not used to terminate a case  then the System is  programmed to ensure that you make active decisions as to what is to happen to  the patient data log when you next power on the system  See  Downloading  Data After Improper Shutdown Error  Reference source not found     We refer to the use of the END PROCEDURE menu option as    PROPER  SHUTDOWN     in this manual     Data Download    Data may be downloaded at the end of acase  To download patient trend data  from the System to a laptop computer  follow these steps in the indicated order     1        When you choose to end the proc
49. bar which displays the patient   s  temperature  the status of the System  and the target temperature for the data  point under the cursor     The patient s temperature and target temperature are displayed using the current  temperature notation setting  Celsius or Fahrenheit      The System status field uses colors to display one of nine status messages that  are described in the following table     Table 4 5  System Status Bar Messages     Status Message     ist Mt aie   Color Explanation      sTBY _ The System was in standby mode  pump off      wep  sd   The System wes warning    tow   Red        o   Black   The System was neither warming nor cooling     LOW Blue  MED The System was cooling     E nE E The System was turned off        Setting the Time Scale    The time scale displayed by the temperature trend graph can be set to any of  four intervals  4 hours  12 hours  24 hours  or 72 hours  To set the time scale   follow these steps     1  Display the temperature trend graph     2  Press the Press for Enter Menu knob once  The screen displays a pop   up menu with the choice    Set Time Scale    highlighted     59 600479 001 Rev 2 90 147    Operation    TempTrend v    Set Time Scale    Exit          Figure 4 69  Set Time Scale     3  Press the knob once  The screen displays the message    Select interval     followed by four interval choices  The most recent choice will be  highlighted     TempTrend v  Time Scale    Select interval     12 Hours  24 Hours  72 Hours        
50. bbles and return them to the saline bag        Figure 4 53  Inline Flow Indicator     WARNING  INVESTIGATE AIR TRAP ALARMS     If an    Air Trap Fault    alarm occurs  it is likely that there is a leak in the tubing  circuit or the catheter has failed  Refer to Chapter 6     Alarms and Corrective  Actions for assistance  Do not keep replenishing saline that is being rapidly  depleted   a problem exists that must be immediately remedied     WARNING  START UP KIT LIFETIME IS SEVEN  7  DAYS     The designed operating lifetime for Start up Kit components is seven  7  days  of continuous operation  If a patient must be treated for a longer period  a new  Start up Kit must be installed in the System  Failure to adhere to this time limit  may cause injury to the patient        44 600479 001 Rev2 75 147    Operation    Setup   Variations    System Setup Sequence    The System Setup sequence varies depending upon the following factors   1  The time from the last power down     2  Whether or not you used the    End Procedure    menu option when you  terminated the previous case     Time From Last Power Down    lf the System detects that it is less than 3 minutes since the last time it was  powered down it will dispense with some of its self tests depending upon the  state of the cooling engine at the time of power down  The idea being that  with  such a short time  it is more likely than not that the power down was either  unintentional  you tripped over the power cord  or an intention
51. calibrated temperature source  e g   Fogg System Co  model TP   400 or equivalent      Procedure    To perform a temperature accuracy test  follow these steps in the indicated order     1     Set the calibrated temperature source to exactly 37   C and plug it into  connecior 11     Start the System and proceed to the Standby screen     Observe the patient temperature displayed on the screen  It should  indicate 37   C   0 2   C     lf the displayed temperature is above or below the indicated range   contact your ZOLL service representative     lf the displayed temperature was within the indicated range  the test is  complete     Clean Console and Condensate Pan    Required Tools and Materials    You will need a soft  lint free cloth and a solution of mild detergent and water to  perform these tasks     Procedure    To clean the System console  follow these steps in the indicated order     1     Assure that the System has been switched off and the power cord has  been unplugged     Clean the exterior of the console using a soft cloth dampened with a mild  detergent and water mixture  Never use solvents or abrasive cleaners on  the console  Avoid vigorous scrubbing  especially on the front surface of  the display     Wipe down the outside surfaces with a water dampened cloth to remove  remaining spots or residue     Slide out and remove the condensate collection pan from under the front  of the console  see the following figure      6 600479 001 Rev2 136 147    Maintenance     
52. ciated with air embolism  should the catheter fail  Death or serious injury may result     Check the Integrity of the Catheter    To check the integrity of the catheter  perform these steps in the indicated order        if Stop the System    2  Using aseptic technique  disconnect the tubing from the catheter and  properly cap both the catheter and the tubing set    2  Fill a sterile 10 ml syringe with sterile saline    4  Connect the syringe to the INFLOW lumen of the catheter and  disconnect the outflow cap    9  Infuse the 10 ml of saline     it should flow out the outflow lumen    6  Cap the OUTFLOW lumen and pull and hold 5 ml of vacuum for at least    10 seconds  Approximately 4 ml of saline  but not blood  should enter the  syringe and you should be able to maintain the vacuum     re Ease the vacuum  disconnect the syringe  and recap the INFLOW lumen     Check the Integrity of the Start up Kit Tubing  To check the integrity of the start up kit tubing  perform these steps in the  indicated order   Stop the System   Look for obvious leakage     3  Remove the tubing from the pump and check it for damage  Replace it in  the pump if it is undamaged     4  Inspect the tubing from the pump to the patient for sources of fluid loss     8 600479 001 Rev2 11 147    Safety Information    e Look for damage to the tubing and or the presence of air inside the  tubing     e Inspect  and tighten if necessary  each Luer fitting  do not use  instruments to tighten the fittings      5  Ins
53. ck the plug in the socket        Figure 3 5  Plug the Control Cable Into the Socket     6 600479 001 Rev2 30 147    Receiving  Inspection  amp  Assembly    Power Cord   T  Plug the female end of the power cord into the recessed power inlet  connector  Wrap the power cord around the two cord hooks on the rear  of the console    Condensate Pan   8  Remove the condensate pan from its packaging and install it in the slot  under the front of the console  refer to figure 8 1     Saline Container Insulating Jacket    9  Remove the saline container insulating jacket from its packaging and  hang it on the saline container hook     Assembly is complete     7 600479 001 Rev2 31 147    Operation    Operation       1 600479 001 Rev2 32 147    Operation    Contents    Overview 5    Operating States in the System 5  SETUP 5  System Self Tests 5  Thermometer Functions Self Test 5  Cooling Engine Self Test 6  Sensor Checks 6  STANDBY 7  RUN   Treatment Modes 8  Max Power  MAX  8  Controlled Rate 8  FEVER  FVR  8  User Interface 9  System Controls 9  The System Display Screen 10  Changing Target Temperature 11  Changing Treatment Mode 12  System Menu System 14  Main Menu 14  Settings Menu 15  Hi Lo Patient Temperature Alarms 16  Setting the Alarms 16  Nature of the Alarm 16  Bath Pre set 17  Pre Cool 18  Pre Warm 18  None 18  Exit 18  Time and Date 18  Time Setting 18  Date Setting 19    C   F  Temperature Notation  19  Language 20  STANDBY Timer 20    T1 T2 Behavior   New 22    2 600479 001 R
54. ck to the patient  Clear all restrictions   indicator  Check the flow indicator to confirm flow     The saline flow rate is Set the pump flow rate as specified by the  incorrect  catheter   s  nstructions for Use     The catheter is impro  Place and position the catheter as directed  perly positioned or is not   by the  nstructions for Use   connected     Connect the saline circuit tubing to the  catheter and prime the saline circuit        4 600479 001 Rev2 128 147    Troubleshooting    Symptom Probable Cause Corrective Action    The patient does not  cool  The activity  monitor indicates     MAX COOLING    and  the patient s temper   ature is increasing     The patient does not  cool  The activity  monitor does not  indicate    MAX  COOLING     and the  patient   s temperature  is increasing after 45  minutes of sustained  activity     The patient s temper   ature is more than  0 5   C below the  setpoint     The patient does not  stop cooling  The ac   tivity monitor indi   cates    MAX COOLING     when the target tem   perature has been  reached     The patient is  warming  when set  for cooling      The heat exchanger coil is  not in the coolant  reservoir     The System is not  operating within specifi   cations     The System is not  operating within  specifications     The patient s resistance to  hypothermia has dimin   ished  resulting ina  temperature controller  undershoot     Machine malfunction     A cooling or temperature  controller fault has  occurred     A c
55. der     Hook   1  Attach the gray hook to the front of the mast using the short bolt  provided  Use a 7 64 inch Allen wrench to tighten the bolt securely  see  figure 3 2         Figure 3 2  System With Hook Attached     4 600479 001 Rev2 28 147    Receiving  Inspection  amp  Assembly    Handle    2  Attach the handle to the mast using the long bolt and the short screw  provided  Use a 3 16 inch Allen wrench to tighten the bolt securely  Use  a 5 32 inch Allen wrench to tighten the screw  see figure 3 3   Do not lift  the System by the handle        Figure 3 3  Use the Long Bolt and Center Screw to Attach the Handle     Control Head    3  Carefully remove the control head from its packaging     4  Attach the control head to the mast  Hold the pivot assembly perfectly  vertical and slide it into the mast opening by applying even  gradual  pressure  The pivot assembly fits into the mast only in one direction     5 600479 001 Rev 2 29 147    Receiving  Inspection  amp  Assembly       Figure 3 4  Attach the Control Head to the Mast     5  Secure the control head to the mast by installing four screws in the holes  provided at the top of the mast  Use a Phillips screwdriver to tighten the  screws securely     Control Cable    6  Connect the control cable to the socket on the lower right rear corner of  the control head  Align the plug with the socket and gently push the plug  into the socket until it is seated  Turn the retaining collar approximately  two full turns clockwise to lo
56. e 4 20     Right   Alerts in STANDBY are generally NOT accompanied  by an audible alarm tone     PATIENT TEMP MAX WARMING PATIENT TEM P MAX WARMING    STANDBY P    STANDBY    Primary probe  11  disconnected  MAX COOLING    Figure 4 20  T1 Disconnect alerts in RUN  left  and STANDBY  right     Primary temp probe  T1  disconnected  Press Enter to proceed       In RUN  Alerts are displayed across the full lower half of the display screen  against a yellow background with the bath temperature display is partially  obscured  See Figure 4 20     Left   If an alert occurs in RUN  the System will  generally revert to STANDBY and an audible alarm tone will sound  The  exception to this is that if you depress the prime switch during run     no audible  alarm will sound and the bath temperature display will not be obscured     In both STANDBY and RUN  a message on the screen will indicate the action  required to clear the alert     During an alert  the patient temperature display is still visible and the patient  temperature alarms are still active if the T1 Temperature probe is functional     NOTE  In the Event of an Alert  Investigate and rectify the cause   refer to    e Section 6  Alarms and Corrective Actions  e Section 7  Troubleshooting       If the alert persists  call ZOLL for service     24 600479 001 Rev2 55 147    Operation    Alarms    An alarm is more serious in nature than an alert and relates to an issues that will  typically require a service call  In most cases a text 
57. e weit  ES A T aS     n  Ee     A    ies       Figure 8 3  Push Hose Into Drain Coupler     As soon the connector is firmly seated in the coupler  coolant will  automatically begin to flow through the line     8 600479 001 Rev 2 138 147    Spillage    10     11   12     13   14     Maintenance    Wait until the coolant flow has stopped  Disconnect the drain line by  pressing down on the metal tab on the coupler  Gently pull out the drain  hose     Dispose of the used coolant     Clean the coolant well using distilled water  Do not use alcohol or any  harsh or abrasive cleansers     Reconnect the drain hose as described in steps 4 7 and drain the liquid  used for cleaning  Dispose of the liquid     Repeat this process while flushing the coolant well with distilled water     When cleaning is complete  you may leave the coolant well empty  to  prepare it for shipment   or refill it with a 1 1 mixture of distilled water and  propylene glycol  Refer to Preparation for First Use in Chapter 3 for  details     Replace the cap on the coolant well     Replace the drain hose in the clamp  Replace the front panel on the  console     Both saline and coolant are corrosive and electrically conductive  Although the  System is designed and tested to be in compliance with the spillage protection  requirements of ISO 60601  it is important to clean up spillage quickly     1     2     To ensure a safe work environment  Spillage  especially of coolant  can  result in a very slippery floor     To 
58. ect the HMIA RS 232 connector from the display head of    the System  the HMIA will stop sending data to your hospital monitor  The  hospital monitor will detect this as a patient disconnection or a very low patient  temperature  Depending upon the set up of your hospital monitor  this may or  may not result in a patient temperature alarm on the hospital monitor        Troubleshooting    The HMIA has a very limited user interface  The only function of the HMIA is to  simulate the patient temperature probe connected to the T1 input port of the  System     This troubleshooting guide deals only with the HMIA  It is incumbant on the  technician doing troubleshooting to verify Items 1 through 4 below prior to  troubleshooting of the HMIA     1  The System is powered ON  operational  and reporting a valid  temperature on the Display Screen     2  The hospital monitor is ON and operational   3  The hospital monitor conforms to the YSI 400 specifications   All cables are connected   a  RS232 from the HMIA to the Display Head     b  The circular connector on Umbilical Cable of the System is  plugged into the HMIA     c  The HMIA is attached to the circular connector on the display  head     d  The hospital monitor is connected to one end of the Alsius  interface cable of the HMIA     e  The other end of the Alsius interface cable is connected to the  HMIA     11 600479 001 Rev2 105 147    Errors    Accesories   HMIA    Green LED OFF    The Green LED indicates that power is present on 
59. ed brief adjustment  that required the system to be powered off  you moved the System to a new  position      The System will give you the option to use the last programmed settings  If you  elect to do this you may still use the menu system to change system settings in  STANDBY     System Set Up    TARGET TEMP    37 0   MAX _    Operate at current settings     Yes No      2006 03 05   18 26    Figure 4 54  Operate at Current Settings          At power on of the system  if it has been less than 3 minutes since the last power  down  then you will be presented with the screen above     i  If you select    Yes    and press the knob  you will proceed directly to  STANDBY once the system self checks have completed     2  If you select    No    and press the knob  you will continue to be presented  with screens that allow you to select operating parameters for the  System prior to entering Standby starting at Figure 4 57     see below     45 600479 001 Rev2 76 147    Operation    If the System detects that it is more than 3 minutes since the last time it was  powered down it will conduct all its self tests  You will not be offered the option  of operating the system at the current settings     Downloading Data After Improper Shutdown    To help prevent the accidental loss of patient trend data  the System is designed  to store patient trend data if power to the system is interrupted or lost  If the  System has not been properly shut down  all patient trend data will be preserved   
60. edure  the screen displays the  message    Patient data must be downloaded or deleted to proceed        PATIENT TEMP   MENU    End Procedure    Patient data must  be downloaded or  deleted to    Aalia proceed     TARGET TEMP    Download Now    Download Later  Delete    Sa    Figure 4 62  Patient Data Message     Choose    Download Now      The screen displays the message    Prepare data link        51 600479 001 Rev 2 82 147    Operation    PATIENT TEMP MENU  End Procedure    Prepare data link     2006 03 05   18 26 Cancel       Figure 4 63  Prepare Data Link Message     Connect one end of a serial interface cable to the laptop computer   s  serial interface connector     Attach the other end of the cable to the serial interface connector on the  System  The female 9 pin subminiature    D    connector is located on the  rear of the display head        Figure 4 64  Location of the Serial Interface Connector     Start the TempTrend program on the laptop computer  Refer to Chapter  5     TempTrend CSV     If a problem occurs during downloading  the System screen will display  the message    Download error  Please check external computer      Check the serial cable connections and the operation of the laptop    52 600479 001 Rev 2 83 147    Operation    computer and choose    Try again     If repeated failures occur  choose     Cancel    and contact your ZOLL representative for assistance     PATIENT TEMP   MENU    End Procedure    Download error   please check  external computer
61. ending upon the state of the system when it was last turned off     You can pre set the bath temperature so that the system is either coolingor  warming the bath in SETUP     Patient temperature alarms are not active in SETUP     System Self Tests    The System performs self tests both at initial power on and then hourly during  operation  These tests allow the system to check its own performance     Thermometer Functions Self Test    The Primary Patient Temperature monitoring circuit  T1  is checked during  SETUP and then hourly against the System internal high resolution calibration  resistor  This is a quick test     5 600479 001 Rev2 36 147    Operation    Cooling Engine Self Test    During SETUP and then every four hours  during run  the System will  dynamically test its cooling engine  This test will take a few minutes     Cooling Engine Self Test   The Pump Will Stop    Each hour the System tests its cooling engine dynamically during RUN  These    tests are brief  The system will stop the pump during these tests to remove the  heat load from the patient during the test     THE PUMP WILL STOP DURING THIS TEST   THIS IS NORMAL        During these tests  the peristaltic pump is stopped  to remove the patient   s heat  load from the equation   The tests that are conducted depend upon the state of  the System at the time of the test  The test s  done is are  as follows     Coolant Temperature   Heating Test   Cooling Test    v   will be performed x   will not be performed  
62. ent Temperature Probe  T1  connected to the System  see below   T1 T2 Behavior   New         STANDBY    PATIENT TEM P MAX WARMING    STANDBY    TARGET TEMP    2006 03 05   18 26 MAX COOLING       Figure 4 2  System Operating Display   Standby  In STANDBY you can interact with the full user interface of the System  You can  select the target temperature  the rate and the treatment mode     Patient temperature alarms are active in STANDBY if the primary patient  temperature probe  T1  is connected     The coolant bath temperature graph is active in STANDBY  a new feature  compared to earlier versions of the System      7 600479 001 Rev2 38 147    Operation    From STANDBY you may toggle into and out of RUN  You may only return to  SETUP by powering down the system     RUN   Treatment Modes    Once you are ready to begin treatment  you may move from STANDBY to RUN   The System will become fully operational provided that all its sensors indicate  that it is ready  Alerts will trigger if it is not    There are three treatment modes in RUN     Max Power        Controlled Rate    and     FEVER        Max Power  MAX     In this treatment option  the System seeks to make the patient   s temperature the  same as the selected target temperature  It will keep the peristaltic pump  operating unless the patient s temperature    inverts     This occurs whenever     A  Bath Temperature  gt  Patient Temperature  gt  Target Temperature   OR  B  Bath Temperature  lt  Patient Temperature  lt  Ta
63. er  then connect the inflow and outflow  luer fittings of the saline circuit together        7 600479 001 Rev2 10 147    Safety Information    Air Entry Into the Tubing Circuit    Air entry may occur with the failure of any part of the start up kit  between the  saline bag and the outflow of the pump  In such cases  the integrity of the  catheter prevents air entry into the patient  In the rare event of a second   simultaneous failure of the catheter  air entry into the patient is possible     Air entry into the tubing circuit will usually  but not always  be associated with an  air trap alarm that will stop the System  Always investigate air trap alarms  The  cooling circuit is a closed loop   usually air trap alarms indicate a breach  somewhere in this closed loop  occasionally an air trap alarm can be caused by  condensation forming on the air trap exterior   With any air trap alarm  check both  the integrity of the catheter and the start up kit  see below      Periodically check the start up kit for significant air bubbles and replace the kit if  necessary     WARNING  NEVER CLINICALLY CIRCUMVENT THE AIR  TRAP ALARM    ZOLL supplies air trap    dummies    for testing  training and demonstration    purposes  These are fluid filled air trap assemblies that are separate from a  standard Start up Kit assembly  Never use this device  or other method  to  circumvent the air trap alarm when the system is connected to the patient   Doing so exposes the patient to the hazards asso
64. eted        Figure 4 47  Use the Prime Switch to Prime the Tubing Circuit     18  Continue to hold the PRIME switch  When the air trap is completely filled    with saline  tap it to dislodge any remaining air bubbles  Observe the 500  ml saline bag  When bubbles are no longer seen in the saline bag   priming is complete  Release the PRIME switch     NOTE  Air Trap    The air trap must be completely filled for proper system operation        19  Turn the air trap right side up and insert it in the holder     20  Place the tubing to the catheter in the two notches at the front of the  console  Place the priming line and the saline return line in the channels  leading to the rear of the console  Close the top cover of the System     CAUTION  Damage to Top Cover    Do not sit on the top cover or place heavy objects on the top cover        40 600479 001 Rev 2 71 147    Operation       Figure 4 48  Route the Tubing Out of the System     21  Lift the saline container off the hook and slip the insulating jacket around  the container  Carefully close the hook and loop fasteners at the top and  bottom of the jacket  Rehang the container on the hook     22  When the self tests are completed and the Start up Kit is loaded and  primed  the System will enter STANDBY  The System is ready to be  connected to the patient     PATIENT TEMP MAX WARMING    E    Figure 4 49  STANDBY Screen     T1 Probe Not Connected        41 600479 001 Rev 2 72 147    Operation    Connecting the Patient to the Sys
65. ev 2    33 147    Operation    First Use Warning     No T2 Probe 22  T2 Probe Disconnection Reconnection 23  Accidental Disconnection T1 T2 Probe 23    Alarms  amp  Alerts 24    Alerts 24  Alarms 25    Your First System Case 26    What you need 26  Preparing the System for Treatment 26  Installing the Start up Kit 32  Hang the Saline Bag 32  The Top Cover 32  The Coldwell 34  The Alsius Start up Kit 35  The Persitaltic Pump 36  Cover the Coldwell 38  Spike the Saline Bag 38  Prime  amp  Fill the Air Trap 39  Connecting the Patient to the System 42    Setup   Variations 45    System Setup Sequence 45  Time From Last Power Down 45  Downloading Data After Improper Shutdown 46    Ending Treatment 51  End Procedure 51  Data Download 51  New Patient     No Power Down 54  Change the Start up Kit 54  Delete Previous Patient Data 54  Disposal of Used Components 54    Temperature Trend Data 5 7    Overview 57  Displaying the Temperature Trend Graph 57    3 600479 001 Rev2 34 147    Temperature Trend Graph 58  Patient Temperature 58  System Activity 58  Cursor 59  Status Bar 59   Setting the Time Scale 59    Mechanical Components 61  Top Cover 61  Control Head Tilt 62  Casters 63    Operation    600479 001 Rev 2    35 147    Operation    Overview    This chapter explains how to start treatment and change target temperature and  rate settings during treatment  It provides instructions on the proper way to end  treatment  including how to download patient temperature trend data to a lapto
66. frequently  hourly     when the System is operating  It is possible for malfunctions or misuse of the  System to result in patient injury or death        The ZOLL Thermal Regulating System is comprised of an external heat  exchange system  CoolGard 3000 or Thermogard  and an Alsius endovascular  heat exchange catheter connected via a sterile heat exchanger and tubing circuit   the Alsius Start up Kit   These components together comprise a patient  temperature regulation apparatus employing feedback control  The catheter and  the Start up Kit  its heat exchange coil  air trap  and tubing  are single use  disposable devices     This manual provides operating instructions for the System and the start up kit   Catheter components are referenced where it is necessary to assure proper use  with the system components  Always refer to the catheter   s Instructions for Use  for additional specific information     Operating Life    The operating life of the catheters may vary according to design as indicated by  the model designation  Always refer to the catheter   s  nstructions for Use for  information about the catheter   s operating life     The disposable components of the Alsius start up kit are designed for continuous  use for a period not to exceed seven  7  days  After seven days of use  all start   up kit components must be removed and replaced with components from a new  start up kit     CAUTION  START UP KIT LIFETIME IS SEVEN  7  DAYS   The designed operating lifetime for s
67. he System keeps the bath cold and starts the peristaltic pump when the  patient temperature is greater than the target temperature       System Menu System    Main Menu    The main menu of the System is displayed in Figure 4 5 above  To call up the  main menu  press the knob from the operating screen  Some settings are  accessible only in Standby Mode     The options available to you are as follows     Table 4 3  System Main Menu    View Graphs Takes you to a display of the patient temperature data log  See   Temperature Trend Data    New Patient Indicates to the System that you are starting a new patient case   Downloading or deleteing the existing patient data log is required    End procedure Indicates to the System that you are finishing a patient case  You will be  prompted to save the existing patient data log  See  Ending Treatment  Takes you to the Settings menu  See below    Returns you to the operating display  The bath temperature meter display  will replace the menu        14 600479 001 Rev 2 45 147    Operation    Settings Menu    The settings menu allows you to modify the information displayed on the  operating display  It appears as in Figure 4 11     MENU  PATIENT TEM P    Settings    Hi Lo Alarms    Time and Date  aS i    Bath Pre Set  Language  Standby Timer  Exit    2006 03 05   18 26       Figure 4 11  System Settings Menu    The options offered by the settings menu are listed and described below     Table 4 4  System Settings Menu  Hi Lo Alarms Allows you 
68. ided into two fields  hours  designated    HH     and minutes     MM       The System uses only 24 hour time notation  e g   3 00 p m  is 15 00      The date is divided into three fields  year  designated    YYYY       month   designated    MM       day  designated    DD      For example  the 24  February  1959  would be shown as  1959 02 24     Time Setting    The screen first displays the message    Select Hour  Press Enter to set        2  The numbers displayed in the hour field will change as you turn the knob   When the correct hour is displayed  press the knob once to enter your  selection    3  The screen next displays the message    Select Minute  Press Enter to  set       4  Turn the knob until the correct minute is displayed  Press the knob once    to enter your selection     18 600479 001 Rev2 49 147    Operation    Date Setting    The screen displays the message    Select Day  Press Enter to set        2  The numbers displayed in the day field will change as you turn the knob   When the correct day is displayed in the field  press the knob once to  enter your selection     3  The screen next displays the message    Select Month  Press Enter to  set       4  Turn the knob until the correct month is displayed  Press the knob once  to enter your selection    5  The screen next displays the message    Select Year  Press Enter to set       6  Turn the knob until the correct year is displayed  Press the knob once to  enter your selection    7  The time and date settings
69. ime scale displayed for a 4 hour interval     System Activity    The System activity graph  labeled    System     is scaled for any System activity   from maximum cooling to maximum warming  The horizontal time scale is the   same as that set for the Patient Temperature graph  The vertical scale uses a   colored activity indicator     e The red zone indicates coolant temperatures between 36   C and 42   C     e The neutral point  between red and blue  indicates coolant temperature of  36   C and also indicates when the pump is not operating     e The blue zone indicates coolant temperatures between 0   C and 36   C     58 600479 001 Rev2 89 147    Operation    Cursor    The cursor is a fixed vertical line that runs through the center of both graphs   When the temperature trend graph is displayed  you may turn the Press for  Enter Menu knob to scroll the display to the left or right  As data scrolls under  the cursor  the top of the screen shows the time and date of the data under the  cursor     Turn the knob clockwise to scroll to the right  As you scroll to the right  the time  and date display will indicate later data  By scrolling right to the end  you can  display the most current data     Turn the knob counterclockwise to scroll to the left  As you scroll to the left  the  time and date display will indicate earlier data  By scrolling left to the end  you  can display the data collected when treatment began     Status Bar    Across the bottom of the display is a status 
70. into the cable  and the cable is plugged into  T2     Verify position of secondary  temperature probe     Determine the cause of the  discrepancy and correct it   One or both temperature  probes may need to be  replaced     Discontinue use and contact  your ZOLL service  representative  see Chapter  9         600479 001 Rev2 124 147    Troubleshooting    Troubleshooting    1 600479 001 Rev2 125 147    Contents    Troubleshooting 3  Overview 3  Symptoms and Remedies 4  Events Requiring Technical Assistance 6    Troubleshooting    600479 001 Rev 2    126 147    Troubleshooting    Troubleshooting    Overview    IMPORTANT     Qualified medical personnel must read and understand these instructions  before performing troubleshooting on the System        This chapter provides information about performing troubleshooting and service  for the System  A table of symptoms  probable causes  and recommended  corrective actions has been provided to assist troubleshooting by qualified users     Some test and repair tasks must be performed only by ZOLL trained service  personnel  If you encounter a problem that is not listed in this chapter  do not  attempt to make repairs or adjustments   contact your ZOLL service  representative for assistance     Chapter 9  Warranty and Service  provides information about how to obtain  assistance or service from an ZOLL representative     3 600479 001 Rev2 127 147    Troubleshooting    Symptoms and Remedies    Consult the following table for help with tr
71. istors and variable resistor test boxes  e g the  TP 400 FOGG Box  for testing  training and demonstration purposes  These    can be plugged into the Primary Patient Temperature Probe T1 connection on  the front of the System to represent a patient  Never use this device  or other  method  to circumvent the normal patient temperature feedback control when  the system is connected to the patient  Doing so exposes the patient to the  hazards associated with hypo  or hyper  thermia  Death or serious injury may  result        Configuration Changes    CAUTION  CONFIGURATION CHANGES MUST BE  CERTIFIED     Equipment connected to the analog and digital interfaces must be certified to  the respective IEC standards    e   IEC 950 for data processing equipment  and IEC 60601 1 for medical equipment   Furthermore  all configurations shall  comply with the system standard IEC 60601 1  Any person who connects  additional equipment to the signal input part or the signal output part con   figures a medical system and is therefore responsible that the system com   plies with the requirements of the system standard IEC 60601 1        Priming the Saline Circuit    WARNING  DO NOT PRIME THE SALINE CIRCUIT WHILE  CONNECTED TOA PATIENT    During the priming operation  the air trap alarm will be disabled  Air present in    the saline line may be circulated through the indwelling catheter     Before priming the circuit or during troubleshooting for possible leak   disconnect the heat exchange cathet
72. lant delivery  lines  saline container connectors  catheter connectors  and the roller pump  tubing  These components are described in detail in later chapters of this  document  The components in the start up kit are designed to operate  continuously for seven days  after which they must be replaced        Figure 2 7  Start up Kit     9 600479 001 Rev2 21 147    Introduction    CAUTION  START UP KIT LIFETIME IS SEVEN  7  DAYS     The designed operating lifetime for start up kit components is seven  7  days    of continuous operation  If a patient must be treated for a longer period  a new  Start up kit must be installed in the System  Failure to adhere to this time limit    may cause injury to the patient        Data Memory  The System is capable of continuously recording patient temperature and system  activity for up to 21 days  This stored data can be downloaded to an attached   computer over a serial interface using optional software furnished by ZOLL     Saline Circuit Diagram    A flow diagram is printed on the inside of the top cover  Use this diagram to  assure that the start up kit has been installed correctly        Figure 2 8  Flow Diagram     10 600479 001 Rev2 22 147    Introduction    Indications for Use   USA    The Indications for Use listed below have clearance within the USA for the  following models of ZOLL Thermal Regulation System     e The CoolGard 3000  e The Thermogard XP    These systems can be used with any of the Alsius Catheters  The indications f
73. layer of the Scroll Control  On the left is a text box   Enter the offset in minutes of the section of the patient data log that is of interest to you   Then click the right hand box with the X in the center of the scroll symbols   gt  gt X lt  lt    The  Scroll Control will center the graph  at your existing zoom  with the entered value in the  center of the graph     If you are already at MAX zoom  there will be no change in the display     Zoom    The X axis of the graph will be scaled depending upon which of the buttons in the middle  layer of the Scroll Control that you click     If you are at MAX zoom  clicking on one of the three other buttons will result in the graph  remaining centered where it is and the span of the graph being modified to either 4 hours   12 hours or 24 hours  The graph will not move beyond the ends of the graph by more  than half the width of the displayed window     Scroll    There are two buttons in the scroll layer of the Scroll Control  Clicking on the left hand  facing triangle will scroll the graph half a zoom distance to the left  i e  to earlier in the  record   Clicking on the right hand facing triangle will scroll the graph half a zoom  distance to the right  i e  to later in the record   The graph will not move beyond the ends  of the graph by more than half the width of the displayed window     10 600479 001 Rev2 118 147    TempTrend CSV    Limits of Scroll Control    Scroll Control was designed to work only when it and the graphed dat
74. le name for the  patient data log  This is called the root file name      lt rootfilename gt       TempTrend    CSV  creates three files based upon this root file name   lt rootfilename gt    It varies the suffix on  the  lt rootfilename gt  depending upon the purpose of the file     Table 5 1  Installation File Paths    The main patient data log   lt roottiliename gt _Event csv   The events data log        lt roottilename gt _Info csv The system information    If you are asked to download the datalogs as part of a service call  please send all three  files     Starting TempTrend    CSV    To start TempTrend    CSV click one of the Alsius    A    shortcuts that have been  created on your computer        The TempTrend    CSV control box will open on your screen  TempTrend    CSV has  only four controls and is very simple to use        Ea Temp TrendCs   P N103710 Rey  A zj    Lom Port      Patient Log File Mame        Select File Download E mit    Figure 5 2  TempTrend Program Screen    Com Port    TempTrend CSV will attempt to determine which Com Port is in use by your computer to  connect to the System with each download  If it cannot find the connection you will  receive an error message     5 600479 001 Rev2 113 147    TempTrend CSV       fa Temp Trend CSY P N103710 Res        i  Y        Com Pork      TempTrendCs   x     Patient Log File Name  C  Temp    Unable to open CammPart1       Select File       Figure 5 3  Com Port Error    This error is typically due to one of two
75. ly be changed in  STANDBY     PATIENT TEM P    To change Target  Temperature or Rate   system must be in  Standby  Would you  like to go to Standby    2006 03 05   18 26       Figure 4 7  Change Rate Mode Temperature in RUN    2  Turn the knob until the desired Treatment Mode selection is displayed     When the correct selection is displayed  press the knob once to enter the  selection     12 600479 001 Rev2 43 147    Operation    PATIENT TEM P    Select Mode     STAN D BY Controlled    2006 03 05   18 26       Figure 4 8  Select Mode    If you select Controlled Rate  a second menu screen will be presented   To change the controlled cooling rewarming rate  turn the knob until the  desired selection is displayed  You may choose a rate from 0 10   C hr to  0 65   C hr  When the correct selection is displayed  press the knob once  to enter the selection     PATIENT TEM P    Select Rate     eee    TARGET TEMP    2006 03 05   18 26       Figure 4 9  Select Controlled Rate    If you have selected FEVER Mode  you will be asked to confirm your  selection with the reminder that FEVER Mode only cools and does not  warm as per the figure below     13 600479 001 Rev2 44 147    Operation    Fever Mode    Fever Mode provides cooling ONLY  There is no  warming with Fever Mode     Do you wish to continue     2006 03 05   18 26       Figure 4 10  FEVER Mode Confirmation message    5  The Operating screen will be updated to reflect your selection     NOTE  Target Temperature in FEVER Mode    T
76. message  specific to the  alarm  identifies the system code for the alarm  For example  the screen might  announce    TCM ID 01    or    MID 23    in addition to the text in the figure below     System error     Please power off and restart     Contact your Alsius service representative  if problem persists     TCM Id 01       Figure 4 21  Alarm Screen    In some cases  however  the alarm may be cleared by power cycling the system   For example  such an alarm would occur if the pump tubing were to become  jammed in the peristaltic pump causing the pump to slow down     If the reason for the alarm is not cleared by power cycling  the alarm will repeat  each time the System is turned back on     During an alarm  the patient temperature display and the patient temperature  alarms are NOT active     NOTE  In the Event of an Alarm    Investigate and troubleshoot the cause   refer to      Section 6  Alarms and Corrective Actions    e Section 7  Troubleshooting    Power cycle the System   If the alarm persists  call ZOLL for service        25 600479 001 Rev2 56 147    Operation    Your First System Case  What you need    You will need the following for each System case   e Anew ALSIUS Startup Kit  e Anew 500ml bag of normal saline     e A YSI 400 compatable temperature probe e g  a Foley catheter   rectal or esophageal temperature probe    e The blue patient connection cable to connect the temperature probe  to the System    e An Alsius heat exchange catheter  e An aseptic work area
77. minimize the risk of corrosion or damage to the System     Spillage is most likely under the following conditions     ih    During set up of the Start up Kit  Spillage of saline in this case should be  cleaned up as with the handling of any infusion fluid     During filling or emptying of the coldwell  The coolant may be safely  wiped with paper towel and the towel disposed of in the standard trash   Propylene glycol Water mix is slippery on sealed floors     In the event of an air trap alarm  With any air trap alarm  investigate to  see if there is spillage into the raceway  Remove the rotor and blot dry  the raceway and motor shaft     9 600479 001 Rev2 139 147    Warranty  amp  Service    Warranty  amp  Service    1 600479 001 Rev2 140 147    Warranty  amp  Service    Contents    Warranty and Service 3    Warranty 3   Technical Support and Resources 3  Obtaining Service from ZOLL 3  Packing and Shipping Instructions 4    Disposal of the System 5    2 600479 001 Rev2 141 147    Warranty  amp  Service    Warranty and Service    Warranty    ZOLL Circulation  Inc  warrants that the System shall be free from defects in  materials and workmanship for a period of one year after the purchase date or  the date first installed  whichever date occurs later  when the System has been  properly installed  maintained  and used for its intended purpose     This warranty applies only to the original purchaser of the System     lf  during the warranty period  a System requires repair  ZOLL 
78. n the self test is finished  the System Set Up screen displays the  message    Bath Pre Set        System Set Up    Bath Pre Set     Pre Warm      2006 03 05   18 26       Figure 4 24  System Pre Cool     To start cooling or heating the coolant reservoir immediately  choose the  desired option and press the    Press for Menu Enter    knob  the    knob      once to enter the selection  If you do not wish to begin cooling or  warming the coolant reservoir now  choose    None    and press the knob  once to enter the selection     2  600479 001 Rev2 58 147    Operation    8  You may be asked questions relating to the downloading of patient data   This is a new patient  Delete any old data left from the inservice you  received     System Set Up    New Patient       2006 03 05   18 26    Figure 4 25  New Patient  Message     9  Choose    Yes        10  The screen displays the message    Previous patient data must be  downloaded or deleted to proceed              System Set Up    Previous patient data must be  downloaded or deleted to proceed     Download Now    Exit      2006 03 05   18 26    Figure 4 26  Patient Data Message           11  Choose    Delete     You will be asked to confirm your choice to delete the  data file  Choose    Yes     A brief confirmation message will appear and  then automatically close  See below     28 600479 001 Rev 2 59 147    Operation    System Set Up System Set Up    Are you sure you want to delete previous    i    patient data  Previous patient da
79. nd assemble the System  you will need the following  tools   e Phillips screwdriver  included in shipping container    e 3 16 inch Allen wrench  included in shipping container    e  5 82 inch Allen wrench  included in shipping container    e 7 64 inch Allen wrench  included in shipping container      e Scissors or box knife   Unpacking    CAUTION  AVOID LIFTING INJURY     The System weighs 115 Ib  52 kg   Never attempt to lift the equipment without  assistance  Use safe lifting practice when handling the equipment        To unpack the System  follow these steps in the indicated order   1  Remove the straps from the carton and pallet     2  Open the top flaps of the carton and remove the inner carton containing  the control head  This carton also contains the handle  attachment  hardware  spare fuses  the condensate pan  the saline container  insulating jacket  and other miscellaneous parts     3  Remove the protective inserts and lift the outer carton up and off    4  Use scissors or a box knife to carefully cut away the moisture barrier bag  Surrounding the console  Use care to avoid scratching the console    5  With the help of an assistant  grasp the base of the console just above  the casters  carefully lift the console off the platform  and set it on the  floor     3 600479 001 Rev2 27 147    Receiving  Inspection  amp  Assembly       Figure 3 1  System Unpacked and Ready for Assembly     Assembly    To assemble the unpacked System  follow these steps in the indicated or
80. notes throughout this document     If you have questions about the safe or effective use of the System  please  contact the manufacturer     Warnings  Cautions  and Notes    This document uses the following conventions to indicate important information     WARNING     Warnings are accompanied by symbols surrounded by a triangle and are  printed in the text in bold italics  Warnings indicate events or conditions  that can result in serious injury or death or severe damage to the  equipment     CAUTION     Cautions are accompanied by symbols surrounded by a triangle and are  printed in the text in bold italics  Cautions indicate information for safe  operation  proper performance  or avoiding actions that may result in  damage to the equipment     NOTE     Notes are accompanied by a symbol of the letter    i    surrounded by a circle  and are printed in the text in italics  Notes clarify understanding  aid in the  proper operation of the product  and prevent problems or errors from  occurring        3 600479 001 Rev2 6 147    Safety Information    Definitions of Symbols and Labels Used on  the Product and in the Manual    Attention  Consult This symbol designates  accompanying the connector for the data  documents  acquisition cable     Electrical warning or Indicates notes intended  electrical caution  to clarify understanding     Temperature and  pressure range     Do not allow liquids to spill  on the product or  package     Type BF applied part   Defibrillator protected    
81. older     35 600479 001 Rev2 66 147    Operation       Figure 4 40  Air Trap Placed In Holder     The Persitaltic Pump    WARNING  FINGER INJURIES    Be careful when inserting the pump tubing that you do not catch your fingers  with the roller     When the System is operating DO NOT attempt to circumvent the safety  interlocks on the peristaltic pump lid  DO NOT place fingers or foreign objects  into the pump raceway when the pump is turning  The peristaltic pump has  sufficient torque to severely damage a finger        9  Locate the pump tubing and route it to the right side of the roller pump   Do not stretch or pull on the tubing  The tubing lengths and flanged  connector allow the tubing to fit into the pump in only one direction     10  Lift the handle on the pump rollers     36 600479 001 Rev 2 67 147    Operation       Figure 4 41  Lift the Pump Rollers Handle     11  Place the flanged connector of the pump tubing into the slot on the right  side of the pump head     1  Place flanged connector  into socket on right hand  side of pump raceway        Figure 4 42  Flanged Connector Fits Into Recess     12  Load the pump tubing around the rollers and into the channel of the  pump head  You must turn the handle counterclockwise as you feed the  tubing into the channel  Press down firmly on the tubing until it settles  into the bottom of the channel  Once the tubing is installed  press the  handle down onto the rollers until it presses into its detent     37 600479 001 Rev 2
82. ooling or temperature  controller fault has  occurred     Place the heat exchange coil in the  reservoir     Verify operation by performing the cooling  engine power test described in the System  Service Manual     Download the TempTrend data file  If the  coolant temperature is above the limits  below  perform the cooling engine power  test described in the System Service    Manual   Catheter    Coolant  Temperature    Used    ICY 11  C  26C  Monitor the patient s temperature  If the  patient s temperature drops more than one  degree below the target temperature   ensure that the System is operating normal   ly  If the coolant temperature is at or below  the patient s temperature  the activity moni   tor should indicate warming and the pump  should be off  Use warming blankets to  stabilize the patient s temperature     See above for correct machine function  Use  conventional methods to rewarm the patient  and discontinue use of the System     If the activity monitor indicates    MAX COOL   ING    and the patient   s temperature is one  degree below the target temperature  dis   continue treatment and contact ZOLL     If the target temperature is below the  patient s temperature  and the activity  monitor does not indicate cooling activity   discontinue treatment and contact ZOLL        600479 001 Rev2 129 147    Troubleshooting    Symptom Probable Cause Corrective Action    The patient does not The warming capacity of Use warming blankets to supplement  warm  The activity 
83. or  use are specific to the catheter  Please refer to the Indications for Use statement  in the catheter specific Instructions for Use     Indications for Use     Cool Line Catheters    The ZOLL Thermal Regulation System using the Alsius Cool Line    Catheter is  indicated for use in fever reduction  as an adjunct to other antipyretic therapy  in  patients with cerebral infarction and intracerebral hemorrhage who require  access to the central venous circulation and who are intubated and sedated     Warning     Fever Reduction    The safety of this device has not been demonstrated for fever reduction in  patients presenting with subarachnoid hemorrhage or primary traumatic brain  injury  The safety and effectiveness of this device was examined in a random   ized controlled trial of 296 patients  The mortality results reported in this trial   for the four patient cohorts enrolled  are presented in the table below  Cl      cerebral infarction  ICH     intracerebral hemorrhage  PTBI     primary traumatic  brain injury  SAH     subarachnoid hemorrhage      Table 1 1  Mortality by Diagnosis  ITT    Cool Line Control     Fischer   s exact test    n  e fw  ae  7    ase   10    For more details on the results of this study please refer to Physician s Manual         Normothermia for the Neuro critically Ill Stroke Patient     101416 001        1 600479 001 Rev2 23 147    Introduction    Indications for Use     All Other Catheters    The ZOLL Thermal Regulation System  using one of the
84. oring system  if attached to the HMIA   should detect this as a probe disconnection or temperature probe error     Once powered  the HMIA immediately establishes its link with the System and  performs a series of self tests including self calibration     see below     RUN    The HMIA is designed to provide a variable resistance value that is interpreted by  YSI 400 compliant hospital monitors as patient temperature  The patient  temperature is obtained via the T1 connection of the System     Calibration    HMIA performs a self calibration at START UP  It will also perform a self   calibration upon reconnection of the RS 232 port  for example  you might  temporarily disconnect the HMIA to allow access for data download   Self   calibration takes less than half a second  During self calibration  it may be  detected by the hospital monitor as a disconnected probe  At the end of self   calibration the HMIA will automatically return to RUN     Installation    The HMIA is quickly and easily installed     Prior to installation of the HMIA  verify that the System has the correct software  to support the HMIA  The software version is displayed during initial start up of  the System as per the figure below  The first number in the Software Version  must be equal to or greater than 1 04  See Figure 4      6 600479 001 Rev2 100 147       Accesories   HMIA    ALSIUS    Self Tests in Progress        Software Version 1 04   XX   XX   X XX    2006 03 05   18 26    Figure 4 System Software Ve
85. osable components are used  proper operation  cannot be guaranteed and harm to the patient may result     e Proper aseptic technique must be used while making all sterile  connections to the System     e Never operate damaged or leaking equipment   e Never operate the equipment without coolant fluid in the coolant well     e Never use pure water  pure propylene glycol  or alcohol as a coolant  fluid     e Never operate the equipment while smoking or in the presence of  open flame     5 600479 001 Rev2 8 147    Safety Information    Shipping and Storage Conditions    When shipping or storing the System  follow these recommendations   e Temperature range   20  C   4  F  to 60  C  140  F    e Atmospheric pressure range  50 kPa to 106 kPa   e Do not allow liquids to spill on the System or its packaging   e Humidity range  10  to 90  noncondensing   e Fragile contents  handle with care   e Always handle and store with the top facing up     Ignition of Flammable Anesthetic Mixtures    The System is not category AP or APG equipment and must not be used in  environments where flammable anesthetic gas mixtures are present     Electrical Hazards    This equipment has been tested and found to comply with the EMC limits of the  international standard EN 60601 1 2  These limits are designed to provide  reasonable protection against interference in a typical medical installation  The  equipment can radiate radio frequency energy if not installed in accordance with  the instructions  and ma
86. oubleshooting the System  If the  problem you are experiencing is not listed in the table  contact your ZOLL service  representative for assistance     Table 7 1  Symptoms  Probable Causes  and Corrective Actions     Symptom Probable Cause Corrective Action    Red LED on control System or control failure    Discontinue use and contact ZOLL   panel is illuminated   and audible alarm   sounds     Red LED on control Electrical power was Turn off power and wait 30 seconds  Turn  panel is blinking and momentarily interrupted    on the System and restart  If problem  audible alarm beeps persists  discontinue use and contact ZOLL   repeatedly     System will not start  No electrical power  Verify that the System is plugged into a  working circuit of correct capacity     Assure that the power cord is securely  plugged into the power inlet connection of  the System     The patient does not The patient may be The fever has overcome the cooling cap   cool  The activity febrile  ability of the catheter  Use additional patient  monitor indicates    MAX cooling methods as needed until the  COOLING    and the patient s temperature is stabilized     patient s temperature  The temperature control    Verify that the target temperature and that  ler setting is incorrect  the cooling rate option is not used     is increasing   The saline flow is Inspect the entire length of tubing  from the  obstructed   indicated by   patient   s catheter connection to the System  no rotation of the flow and ba
87. p  computer and how to remove used components and dispose of them safely     Subsections at the end of this chapter provide detailed procedures for recovering  from improper shutdowns  including special procedures for downloading patient  temperature trend data     In this Overview section  we will present the main features of system as they are  presented to the user on the display of the system  This section will help you be  familiar with the features of the System before starting to use it     After the Overview  you will find sections that describe in detail the operation of  the System   see Operating the System  The sequence of events that you must  pass through in starting the System varies with the way it was last turned off   Within this section we will review the different patterns of interaction you will  experience  Specifically we will review     e Your first case with the System   e Variations in the SETUP sequence of the System  e Ending procedures     Operating States in the System    SETUP    The System has three operating states  SETUP  STANDBY and RUN     When the System is first powered up  it goes through a sequence of self tests  It  tests its own electronics and internal sensors  These tests are called the POST   Power On Self Test   It is normal to hear two beeps during POST  Then the  System tests its thermometer functions and its cooling engine  The extent of the  testing during SETUP and the interactions required of you by the system both  vary dep
88. pect the tubing that returns to the pump from the patient    6  Examine the saline bag to ensure that it has not been accidentally  compromised  for example  the spike may have damaged the bag wall     7  Inspect the tubing from the saline bag to the saline reservoir and the  pump     Interference    If this equipment does cause interference with other devices  which can be  determined by turning the equipment off and on  the user is encouraged to try to  correct the interference by one or more of the following measures     e Reorient or relocate the receiving device   e Increase the separation between the equipment     e Connect the equipment into an outlet on a circuit different from that  to which the other device s  is connected     Product Label    An identifying label is attached to the outside of the System console near the  power cord inlet  The label is illustrated in Figure 1 1     To avoid electric shock  disconnect all JN To avoid electric shock  disconnect all  A WARNING cords before removing access panels  WARNING cords before removing access panels     AN WARNING Nouser serviceable pare inside  Refer AN WARNING Nouser serviceable pare inside  Refer    all servicing to qualified personnel  all servicing to qualified personnel     mod  COOlGard 3000    sa  Thermogard XP       3000 XXXXX XAXXXKK m   TGXP KXXXX MRMRRE    Power Regts  Fuse Regts             Power Regts  Fuse Regts              175  100 1207   S0 60 HZ 5A TAJA  slow blow   amp  X 200M 115  100 1207  
89. peration    ALSIUS    Self Tests in Progress        Software Version 2 00   1B   36   1 57    2006 03 05   18 26             Figure 4 31  Self Test Screen     18  Install the Start up Kit now  If the system finishes its self test before you  complete installing the Start up Kit  you will see the following Check  Screen  An item in RED requires your attention     System Set Up    Check the following    Air Trap Roller Pump Lid    Check Prime Switch Coolant    l 2006 03 05   18 26    Figure 4 32  Check the Following Screen           31 600479 001 Rev 2 62 147    Operation    Setup Sensor Checks    Air Trap There is no Start up Kit installed or there is a large amount of air in the  Start up Kit chamber     Roller Pump Lid The clear plastic roller pump lid is not closed properly     Check Prime Switch   The Prime Switch is being depressed  as when you use it to prime the    system  or has been jammed in the    on    position   The coolant level is low  Top up the coldwell with coolant        Installing the Start up Kit    Hang the Saline Bag    To install the System Start up Kit  follow these steps in the indicated order     1  Obtain an IV bag or bottle of sterile normal saline solution  The bag or    bottle must contain no more than 0 5 liter  500 ml  of solution  Hang the  saline container on the hook mounted on the rear of the display post     The container hangs inside the circumference of the handle   p        WE cauite      Au    i ws on      p gi  A 2   E A bes  ake see 
90. r questions  provide guidance  and  schedule service for your System     Upon request  ZOLL Circulation  Inc  can provide circuit diagrams  parts lists  and  service documentation to authorized users     Obtaining Service from ZOLL    To obtain service from ZOLL Circulation  Inc   contact their service hotline at   877  225 7487  Ext  3  You may send inquiries to us by FAX at  408  541 1030     3 600479 001 Rev2 142 147    Warranty  amp  Service    Packing and Shipping Instructions    Please call the service hotline to obtain a Return Goods Authorization  RGA   number before returning items to ZOLL  The RGA number must be marked on  the outside of shipping containers     ltems to be returned to ZOLL Circulation  Inc  must be packed carefully to avoid  damage during shipment     ltems may be shipped to us at     ZOLL Circulation  Inc    650 Almanor Avenue  Sunnyvale  California 94085  U S A     4 600479 001 Rev2 143 147    Warranty  amp  Service    Disposal of the System    Do not dispose of the  system as unsorted  municipal waste     When you wish to dispose of the System call ZOLL Circulation  Inc  At no  expense to you  we will arrange for the crating  removal and disposal of the  System  You will have to arrange a date for the pick up and allow us access to  crate and pick up the system        The functional life of the System will be best extended if you properly maintain  the system  Please refer to your service manual  Do not attempt to service the  cooling engine your
91. rget Temperature     Controlled Rate    In this treatment option  the System will attempt to move the patient   s  temperature to the target temperature at the programmed rate of heat exchange     C  hr   When the patient reaches the target temperature  the System will revert  to the MAX treatment option i e  it will attempt to make the patient s temperature  the same as the selected target temperature     NOTE  Controlled Rate    Controlled rate operates in both warming and cooling modes        FEVER  FVR     In this treatment option  the System will start cooling the patient once the patient  temperature is above the target temperature  It does this by keeping the bath at  its coldest permissible temperature and then operating the peristaltic pump  whenever the patient s temperature moves above the target temperature     WARNING     Lo    patient temperature alarm limit with    FEVER     The System will NOT heat the patient when the    FEVER    treatment option has  been selected  The    Lo    patient temperature alarm limit ensures that an alarm    occurs should the patient stop regulating his her own body temperature  Such  patients will cool to room temperature  This can occur when the patient dies  or becomes comatose     INVESTIGATE ALL PATIENT TEMPERATURE ALARMS        8 600479 001 Rev2 39 147    Operation    User Interface  The primary controls for the System are mounted in the display head of the  system  The next sections describe   e The controls available to the 
92. rsion    If your System does not have the compatible software version  then the correct  software will need to be installed for the HMIA to function  Failure to install the  correct software prevents the HMIA from functioning but it has no adverse affects  on the System     Installation Equipment    The following items are necessary to install a HMIA     1   2     System with correct software     There are two different configurations of umbilical cables in the field  One  has a 6 pin connector and the other has an 8 pin connector  The  configuration of the HMIA corresponds to the type of connector on the  Umbilical Cable of the System being modified  HMIA part number  601033    001 is for the 8 pin circular connector and HMIA part number  601033    002 is for the 6 pin circular connector    TP 400 test fixture with adapter cable  Alsius Part number 500341 001   that mates to the T1 connector of the System    Air trap assembly filled with water  Alsius Part number 500395 001      The standard Hospital Monitor patient temperature cable that is used to  connect the Hospital Monitor to a Foley or rectal patient temperature  sensor     Hospital Monitor that conforms to the YSI 400 requirements     Optionally a Digital Multimeter may be used if a hospital monitor is not  available     Installation Procedure    1   2     Disconnect power to the System    Disconnect the Umbilical Cable from the rear of the Display Housing   See Figure 5     7 600479 001 Rev2 101 147    Accesories   
93. scontinue use and contact  your ZOLL service  representative  see Chapter    9      The coolant well is empty  Fill with mixture of propylene  COOLANT EMPTY    The coolant welis empty  glycol and distilled water     The coolant level is low  Add distilled water until liquid    COOLANT LOW level reaches top fill line in    coolant well        3 600479 001 Rev2 122 147    Alarms  amp  Corrective Actions    Alarm Message Cause Corrective Action    DOWNLOAD ERROR   PLEASE CHECK  EXTERNAL  COMPUTER    PATIENT DATA LOG  FULL  PLEASE  DOWNLOAD NOW    Hi PATIENT  TEMPERATURE  ALARM    Lo PATIENT  TEMPERATURE  ALARM    PRIMARY  TEMPERATURE  PROBE  DISCONNECTED     PRIMARY  TEMPERATURE  PROBE DISLODGED     Patient trend data download  has failed or there is no  computer connected to the  System     Patient trend data log is full   Only two hours of space  remain     The patient s temperature is  above the programmed Hi  patient temperature alarm  value     The patient s temperature is  below the programmed Lo  patient temperature alarm  value     The primary patient  temperature probe is  disconnected     Temperature output from the  primary patient temperature  probe changed by    0 2  C or  more within 10 seconds     Check the connection  between the computer and the  System and select    Try again      If problem persists  select     Cancel    and contact ZOLL  technical support     Download patient trend data   If data is not downloaded  it  will be overwritten     The System alarm
94. selected by the user     7 600479 001 Rev2 19 147    Introduction    WARNING  FINGER INJURIES    Be careful when inserting the pump tubing that you do not catch your fingers  with the roller     When the System is operating DO NOT attempt to circumvent the safety  interlocks on the peristaltic pump lid  DO NOT place fingers or foreign objects  into the pump raceway when the pump is turning  The peristaltic pump has  sufficient torque to severely damage a finger        If a tubing leak or failure occurs in the pump raceway  the saline solution will  cause corrosion in the moving parts of the rotor     Remove the pump rotor  refer to Pump Rotor removal instructions in  System Service Manual      Rinse the rotor in clean water and thoroughly dry the rotor   3  Apply a few drops of light machine oil to the moving parts of the rotor     4  Reinstall the rotor  refer to pump rotor installation instructions in System  Service Manual         Figure 2 5  Pump     Prime Switch    The prime switch is located next to the pump under the top cover  The switch is  used to operate the pump to prime the tubing with sterile saline solution from the  saline source  When the switch is held down  the pump runs  when the switch is  released  the pump stops     8 600479 001 Rev2 20 147    Introduction       Figure 2 6  Prime Switch        Start up Kit    The Alsius start up kit contains the sterile disposable components for this  system  Each kit contains a heat exchanger coil  air trap  saline coo
95. self as this may release R134a refrigerant into the  atmosphere     The System contains   e Solder that contains Lead  e Adischarge vapor bulb  e Refrigerant R134a  e Mercury in some of its switches     Mercury and lead are long lasting organic metal pollutants that can cause direct  toxic effects in humans  R134a is a green house gas     The symbol above is fixed to the System to remind you     DO NOT dispose of the  System as unsorted municipal waste     5 600479 001 Rev2 144 147    Specifications    Specifications    1 600479 001 Rev2 145 147    PHYSICAL  Dimensions     Weight   ELECTRICAL    Configuration    100 120 VAC  50 60 Hz  5 A  220 240 VAC  50 60 Hz  2 25 A    ENVIRONMENTAL  Operating temperatures   Operating humidity   Atmospheric pressure   CHILLER and HEATER  Reservoir volume    Pump capacity   Temperature range   Coolant    Refrigerant   CONTROLLER AND DISPLAY  Screen display   Controls    Temperature input   Articulation    Data interface     Alarms   Displayed temperature range   Displayed temperature accuracy     Specifications    Height  45 in   114 cm   Width  17 in   43 cm   Depth  30 in   76 cm     115 Ib   52 kg     Voltage Fuse protection  115 V T6 3A  slow blow  5 x 20 mm  230 V T6 3A  slow blow  5 x 20 mm    10  C     27  C  50   F     81  F   30  to 75  noncondensing  70 kPa to 106 kPa    2 0 liters  0 5 gal     7 lpm at O m head  0 ft     0  C  42  C   1 1 mixture of propylene glycol and distilled water  HFC 134a    6 4 in   16 25 cm  LCD color VG
96. soft Excel   Book1            4lsius   Adobe PDF  Open CS    File Edit View Insert Format Tools Data Window Help    DSH SIiSGRAlIFHa   sae FI9     Arial  10   B Z7 U 8 amp        ZE gE I        Graph cs             Activate Scroll Control       Figure 5 7  The Alsius Menu with CSV Functions    Open CSV  This menu item allows you to open a CSV file  It will open ANY CSV file     Graph CSV    The Graph CSV Add In converts the data log file from a     csv    format to an EXCEL     xls     format file  It also saves the new format with the words    Copy of    appended as a prefix to  the original file name  In this way the original downloaded data logs are unchanged  You  are free to modify and experiment with your data analyses    This menu item will create a graph from a patient data log file  The graph will display the    following data series as an XY Cartesian graph with the X axis set as minutes of elapsed  time from the start of the data log and the Y axis as temperature in   C     1     2   d   4    Patient Temperature   Target Temperature   Coolant Temperature   System RUN STANDBY   e A graph value equivalent to 8   C indicates that the system was in RUN     e A graph value equivalent to 0   C indicates that the system was either in  STANDBY or powered off     8 600479 001 Rev 2 116 147    TempTrend CSV    CoolGard 3000 Patient Data          e T i  ares    o     a  on   OT  9260 9400 9520 9640 9760 9880 10000 10120 10240 10360 1048  Minutes       Figure 5 8  Graph CSV Di
97. splay with Scroll Control    Activate Scroll Control    The Scroll Control is automatically activated after you have graphed a patient data log   The Scroll Control allows you to     1  Select a location in the record that is of interest and display the data on either  side of that point in time    2  Change the degree of expansion on the X axis  the time base     3  Scroll forward or backward in time     The Scroll Control will be deactivated by EXCEL depending upon what you do within  EXCEL and you may close it yourself by clicking the Close Button  If it is deactivated   select this menu item to reactivate     Within the Chart Menu Bar  the Alsius menu selection has only one option  Activate  Scroll Control     EJ Microsoft Excel   Copy of demo xls    EH  File Edit wiew Insert Format Tools Chart Window Help   Alsius   Adobe POF     LJ eS m  fa     fe    ed ol   tt EF A       WJ     5 Activate Scroll Control Ep  g5  s   E  0   B  z u SE  SE mm  Ss    gt  co  fe    Value CY  Axi              CoolGard 3000 Patient Data    Figure 5 9  Alsius Menu in the Chart window          MT  ola     a      9 600479 001 Rev 2 117 147    TempTrend CSV    The Scroll Control     lt  lt  lt  SCROLL gt  gt  gt        Figure 5 10    The Scroll Control Dialog Box  The Scroll Control is divided into four layers     1  The location layer  2  The zoom layer  3  The scroll layer  4  The Close Button    To close the Scroll Control click the Close Button     Location    There are two boxes in the upper 
98. t the    Hi Lo Alarms    option    4  The    Lo    Alarm value will be presented  Turn the knob until the desired  value is displayed and press the knob    oF The    Hi    Alarm value will be presented  Turn the knob until the desired    value is displayed and press the knob     Nature of the Alarm    MENU    PATIENT TEM P z  Settings    Hi Lo Alarms    Time and Date  CIF  Bath Pre Set    Language  Standby Timer  Exit    2006 03 05   18 26       Figure 4 12     Hi    Patient Temperature Alarm    The alarms are both visual and audible  The alarms will not clear until the  patient s temperature no longer triggers the alarm state     The audible alarm may be temporarily muted for 2 minutes by pressing the Mute  button  The alarm will continue after that time unless it has cleared     The visual alarm is effected by writing the patient temperature area on the screen  in RED text and flashing the text  The visual alarm will not stop until the alarm  has cleared     16 600479 001 Rev2 47 147    Operation    WARNING  Change to Initial Specification of System     In the initial release of the System there were no settable patient temperature  alarms  There were fixed high and low patient temperature alarms  These  fixed alarms have been made programmable     The patient temperature alarms MAY NOT be deactived  They may be muted  for a 2 minute period     NOTE  Patient Temperature Alarms in STANDBY    If the System is in STANDBY and the T1 temperature probe is connected to  the System
99. ta deleted       2006 03 05   18 26 2006 03 05   18 26             Figure 4 27  Delete Previous Patient Data    12  The System Set Up screen then displays the message    Select Target  Temp        System Set Up    Select Target Temp  Press Enter to set     TARGET TEMP    2006 03 05   18 26          Figure 4 28  Select Target Temp Message     13  Turn the knob until the target patient temperature is displayed  When the  desired value is displayed  press the knob once to enter the selection     14  The System Set Up screen displays the message    Select Treatment  Mode    Note that you have three choices     Max Power        Controlled  Rate    or    FEVER        29 600479 001 Rev2 60 147    Operation    System Set Up    Select Treatment Mode    Max Power    Controlled Rate  Fever    2006 03 05   18 26    Figure 4 29  Select Treatment Mode Message           15  Turn the knob to highlight the desired mode  Press the knob once to  enter the selection  Controlled rate should not be selected when using  the Cool Line catheter     16  If you select    Controlled Rate    for the Treatment Mode  you will be  prompted to    Select Rate  Press Enter to Set     Use the knob to scroll  through to the desired rate and then press the knob to select it     System Set Up    Select Rate  Press Enter to set       2006 03 05   18 26    Figure 4 30  Select Rate Screen       17  If the System has not yet finished its self tests  the Self Test screen is  displayed     30 600479 001 Rev2 61 147    O
100. tact with the IV container    spike        15  Use a disposable tissue to wipe up any spilled coolant from the top of the  coolant well  Place the tissue in the medical waste container     16  Replace the insulated cap on the coolant well  Replace the insulating  jacket on the hook     Removal and disposal of the used components is complete  The System may  now be stored or moved to its next treatment location     55 600479 001 Rev2 86 147    Operation    NOTE  Disposal of System    When the System itself has reached the end of its useful life  it must be  disposed of in accordance with local governing ordinances and recycling plans  for refrigerated appliances        56 600479 001 Rev2 87 147    Operation    Temperature Trend Data    Overview    During operation  the System continuously collects and stores temperature trend  data  storing a record each minute  This data is stored in memory and can be  downloaded to an attached laptop computer for later analysis or plotting     The memory will hold up to 21 days of data  Data can be collected for 21 days  before the memory is filled and a download of data is necessary to preserve all of  the stored data  If data is not downloaded when memory is full  the newest data  will overwrite the oldest data and only data from the past 21 days will be stored     At any time during operation  the complete record of temperature trend data can  be displayed as a graph on the screen  This chapter explains how to display  temperature trend d
101. tart up kit components is seven  7  days    of continuous operation  If a patient must be treated for a longer period  a new  start up kit must be installed in the System  Failure to adhere to this time limit  may cause injury to the patient        Functional Description    The System can be described in terms of three major components  a  recirculating chiller  a sterile fluid roller pump  and a temperature control system   The System is connected to the temperature controlled catheter by two small   bore plastic tubes  One tube supplies temperature controlled sterile saline  solution to the catheter  and the other tube returns the saline solution to the  System  The sterile saline is pumped through a continuous recirculating loop by a  peristaltic pump inside the console  The saline solution acts as an intermediate  heat transfer medium between the patient and the System  Sterile saline is used  because it is biologically compatible with the patient and in the unlikely event of a  leak in the catheter  the possibility of harming the patient is reduced to a practical  minimum     3 600479 001 Rev2 15 147    Introduction    Patient temperature feedback is used to control the system  The patient   s  temperature is measured by an indwelling YSI 400 thermistor temperature  sensor  In response to the patient   s measured temperature  the System employs  both cooling and heating  Cooling occurs when the patient s temperature is  above the set point target temperature  Heating occ
102. ted     PATIENT TEM P SETTINGS  Language    Set system  language     eens Cli    TARGET TEMP    Francais  Espafiol  Italiano  2006 03 05   18 26 Portugu  s       Figure 4 16  Language Settings     To keep the current selection  press the knob once  The current setting will not  be changed and the settings menu will be displayed     To change the current setting  turn the knob to highlight the desired setting   Press the knob once to enter the selection  The setting will be changed and the  settings menu will be displayed     STANDBY Timer    The STANDBY Timer provides the User with an alarm as a reminder when the  System has been left in STANDBY for 15 or 60 minutes  The STANDBY Timer    20 600479 001 Rev2 51 147    Operation    presents the User with the opportunity to select from the following values  No  Alarm  15 minutes  60 minutes     SETTINGS    PATIENT TEM P Standby Timer    A    Select Delay   Press enter to set     15    Minutes in  standby  until warning    2006 03 05   18 26       Figure 4 17  STANDBY Timer Menu    If the system has been left in STANDBY for more than the specified time  an  alarm will sound to remind the User that the System remains in STANDBY   Pressing the knob will reset the timer  The Standby Timer function will continue  until either   e The System is placed into RUN   e The STANDBY Timer Menu is used to deactivate the STANDBY  Timer     The STANDBY Timer alert may be silenced for two  2  minutes by pressing the  mute button     PATIENT TEM P
103. tem    CAUTION  Verify System Function First    Ensure proper functioning of the System and initiate pre cool or pre warm  if  applicable  prior to placing the catheter in the patient        When the System has been prepared as directed in the preceding sections  it  may be moved to the patient s bedside and connected to the patient  Follow  these steps in the indicated order     1  Position the System near the patient   s bed  It must be close enough so  that the temperature probe cables and the tubing can conveniently reach  the patient  Route the cables and tubing safely     2  If the primary and secondary patient temperature probes have not been  placed in the patient  this should be done now  Refer to the  nstructions  for Use that accompany the temperature probes for information about the  probes     3  Connect the blue patient temperature cable to the YSI 400 primary  temperature probe  e g  Foley catheter  rectal or esophageal    Connect  the plug at the end of the blue patient temperature cable into the  connector labeled    T1    on the front of the System console        Figure 4 50  Temperature Probe Connections     4  If you are using a secondary patient temperature probe  connect the blue  patient temperature cable to the YSI 400 secondary temperature probe   Connect the plug at the end of the blue patient temperature cable into  the connector labeled    T2    on the front of the System console  If you are  not using a secondary temperature probe  the patient MUS
104. the HMIA circuit board     If the Green LED is not on then the LED has either burned out or the internal  power connector is dislodged  Return the HMIA to ZOLL for service     Flashing Red LED    The Red LED flashes only when there is a problem with the CPU or the internal  self calibration fails  Using a small solid wire  paperclip   press the reset button     RST   If the Red LED continues to flash return the HMIA to ZOLL for service     No Temperature Reported On Hospital Monitor    1     11     12     Observe Red LED   If it is flashing then follow instructions for Flashing Red LED above     If the Red LED is not flashing then disconnect the hospital monitor from  the HMIA interface cable     Using a DMM measure the resistance between the two pins     Using the chart YSI 400 Temperature vs  Resistance determine if the  resistance value measured corresponds to the temperature displayed on  the System display     If the resistance value is correct then the HMIA is functioning to  specifications     lf the resistance value is wrong  then test Alsius interface cable for shorts  or opens     If it passes the test  then press the reset button  RST     Repeat step 5 above       If the resistance value is still wrong  then disconnect the RS232 cable    from the rear of the display housing  wait for 15 seconds and then  reconnect     Wait for an additional 20 seconds for the HMIA to re establish  communication with the System    lf the resistance value is wrong then return the HMI
105. the catheter has reached   catheter performance until the patients  monitor indicates its upper limit  temperature has been stabilized       MAX WARMING    and   the patient s temper     ature continues to  decline     The patient is The warming rate is Reset the target temperature to hold or cool  warming too fast incorrect  the patient s temperature  Verify the warm    when set for ing rate setting    warming         Events Requiring Technical Assistance    CAUTION   The System has multiple internal alarm states   see Alarms  amp  Corrective  Actions     As described in this manual  some are remedied by the user action e g  Filling  the coldwell with coolant will rectify the coolant level alarm     There are others that may be alleviated  in some cases  by power cycling the  System  i e  turning it off and then back on      If neither user action nor power cycling the system clears the alarm     DO NOT  USE THE SYSTEM  Call ZOLL Service Representative for service of any  alarm that does not clear        Immediately discontinue using the System and seek advice from your ZOLL  service representative if any of the following events occur     e The System repeatedly trips an outlet equipped with a ground fault  circuit interrupter  GFCI      e The protective fuse repeatedly blows    e Fluid is observed leaking from the console    e The System emits an unusual odor when operating    e The System produces loud or unusual noises when operating     e The display screen  control knob
106. the front of the System  Connection of the  HMIA to a hospital temperature monitor using one of the Alsius custom cables   will permit the display of an identical patient temperature on both the System and  the hospital temperature monitor  while using only a single patient temperature  probe     WARNING     The HMIA is not a replacement for the T2 Temperature probe on the System     It simply simulates the T1 Temperature probe  Use of the HMIA does not  obviate the need for a second patient temperature monitoring method  Failure  to use a second patient temperature monitoring method can result in injury to  the patient in the event of a T1 Temperature Probe failure     Operating the HMIA    Connecting to the Hospital Monitor    The HMIA simulates a standard YSI 400 temperature sensor  such as in a patient  temperature sensing Foley catheter or rectal probe        The HMIA is connected to a hospital monitor via an Alsius custom interface  cable included with the HMIA  These cables have connectors that are identical to  the connectors of Foley and rectal patient temperature probes     One end of the Alsius interface cable plugs directly into the standard patient  cable of the hospital monitor  In this way  the hospital monitor connects to the  HMIA interface cable in the same fashion as it would connect to a Foley or rectal  temperature probe   See Figure 1      3 600479 001 Rev2 97 147    Accesories   HMIA             Figure 1 Cables and Connections  The other end of the Alsi
107. to modify the low  Lo  and high  Hi  patient temperature alarms     Time and Date Allows you to modify the System time and date       C   F Allows you to toggle between displaying temperatures in degrees Celsius  or degrees Farhenheit    Bath Pre Set Allows you to set the system to either HEAT or COOL during STANDBY   Language Allows you to select the display language     Standby Timer Provides you with a warning alarm after the system has been in STANDBY  for either 15 or 60 minutes  This alarm can be deactivated    Returns to the operating display     NOTE   There is no Pump Rate Selection in ZOLL XP systems     ZOLL has validated that the Cool Line catheter can be safely used at a  pump rate of 240 ml min        15 600479 001 Rev 2 46 147    Operation    Hi Lo Patient Temperature Alarms    The System features two patient temperature alarms     Hi    and    Lo     If the alarms  are set  the System will alarm whenever the patient   s temperature is higher than  the    Hi    Patient Temperature Alarm value and whenever the patient s  temperature is lower than the    Lo    Patient Temperature Alarm value  The range  of values for both alarms is 28   C     45   C  82 4   F     113 0   F      Setting the Alarms    The alarm values are set as follows in either STANDBY or RUN mode  Set the     Lo    and    Hi    Patient Temperature Alarms to appropriate values     1  Press the knob and the Main Menu will appear    2  Select the    Settings    option and press the knob    3  Selec
108. ture Controller    Pump    The temperature controller uses input from the patient   s temperature probe and  the operator selected patient temperature setpoint to regulate the coolant  temperature of the recirculating chiller  The temperature controller constantly  adjusts the coolant temperature by means of a closed loop control system  The  operator enters a setpoint that represents the patient   s target temperature  The  controller cools or heats the coolant  in a range between 0   and 42   C  32   and  107 6   F  to optimally achieve and maintain the target temperature  The controller  constantly displays the measured patient temperature and the target  temperature     An optional mode can command the controller to approach the target  temperature at a user selected rate     Temperature Probe Connectors    The front of the System console features two connectors  labeled    T1    and    T2     which are used for connection to patient temperature probes  The primary patient  temperature probe is plugged into connector T1  The secondary probe is plugged  into connector T2   a                  e9   Al es           Figure 2 4  Temperature Probe Connectors     Sterile saline solution is circulated through the heat exchanger coil and the  catheter by a high performance  compact roller pump  It pumps by peristaltic  action on the tubing installed in the pump head  The pump rotation speed is  accurately controlled by an electronic speed control system  The pump flow rate  can be 
109. urs when the patients  temperature is below the set point target temperature  The amount of heating or  cooling power is proportional to the difference in temperature between the set  point target temperature and the patient s measured temperature     A basic diagram of the System is illustrated in the following figure   Saline bag   Peristaltic pump  Air trap F  Heat exchange coil le  Coldwell            Coolant pump   ai LAN  Chiller  amp  Heater ilo o  Catheter LJe UZ    Patient 1J    G      ii    ee ee E R P A O    0 Pin wheel flow indicator D    Figure 2 1  Simplified Flow Diagram     System Components    Controls and Display    The operator s display panel contains four buttons and one knob used to access   functions and adjust settings with the aid of the menus and messages displayed   on the screen  The controls and display are illustrated in figure 2 2 and explained  in the following text     Display    The display is a backlit color LCD panel that can be easily read in all ambient  lighting conditions  It is used to display status  menus  messages  alarms  and  patient temperature trend graphs     The display head is attached to the mast by an adjustable swivel tilt mounting  clamp  The user can adjust the tilt and rotation of the display head and lock it into  position by using this clamp     4 600479 001 Rev2 16 147    ot ee Se ee I    Introduction          Display Screen   Alarm Indicator LED    Mute Button   Power On Indicator LED   Target Temp Button   Standby 
110. us interface cable plugs into the HMIA connector    labelled T1 out   See Figure 2    a    SUITE 150  CUDAN By RIVAL  CA T7511  ee TEL  49  453 060        CONFORMS TO     lt  DA r        UA STD 60801 1 p51 6H Den Haag FAX  340  483 0220  34        CERTHED TO  CANTSA STD    7 2 NO AY        lt c       1  l  E  RS232 T1 out CG3000    Figure 2 HMIA with T1 out connection    To begin operation    The operation of the HMIA is very simple  There is no independent on off switch   As soon as the System is switched on  the HMIA automatically begins  functioning     Connect the custom Alsius interface cable to the patient cable of the hospital  monitor   See Figure 1   Plug the other end of the Alsius interface cable into the  HMIA connector labelled    T1 out        4 600479 001 Rev2 98 147    Accesories   HMIA    The set up is now complete     Controls and indicators on the HMIA    The HMIA has the following control indicators  a reset switch  a green power  indicator and a red warning light     The front panel of the HMIA is illustrated in the figure below     RST HOSPITAL MONITOR   l INTERFACE F  a  C ZOLL Circulation ee l  c us 7  ly    ee  dy hee    U344 ZOLL Circulation  line  Lat fa   650 Almanor Avenue   404 OSA E MERI af  EUR IPE bt view i  i a kin ae  I aT J es   r E L TE Sunnyvale CA 940845  PaT N   Tm  ea oh   Mslanetraal 15 j  CONFI IRMS TO eeii aa IS t 408 541 2140  UL STD 60601 1 25 1J68H Den Haag    ane had aan  CERTIFIED TO The Netherlands 408 541 1030  a m i f     ir
111. user  e The System display screen  e The System menu system available to the user     System Controls                 Display Screen   Alarm Indicator LED   Mute Button   Power On Indicator LED   Target Temp Button   Standby   Run Button   Rate Deg   HR Button     Press for Menu   Enter  Target Temp  Knob               oO Oe e SY    Rate Deg  Hr Press for Menu   Enter      Standby Run    ALSIUS         00       Figure 4 3  Controls and Display     The use of each of the controls is described in the table below     Table 4 2  System Controls    Flashes red during alarm states    Rate Deg Hr button Press this to change the rate and or treatment mode        Press  amp  Rotate    knob This knob operates the menus on the System display   Rotate the knob to scroll between selections  Press and  release the knob to select        9 600479 001 Rev 2 40 147    Operation    The System Display Screen    The System features a color display screen  The figure below shows the screen  as It appears during RUN     MAX WARMING  PATIENT TEM P    2006 03 05   18 26 MAX COOLING       Figure 4 4  Operating Screen     The upper left hand of the screen displays    e The patient temperature   e The    Lo    and    Hi    patient temperature alarm values   The lower left hand of the screen displays    e The programmed Target Temperature for the System   e The programmed treatment mode controlled rate value  The right hand side of the screen is used either     e To display the temperature of the coolant
112. y cause harmful interference to other devices in the  vicinity  There is no guarantee that interference will not occur in a particular  installation  Always comply with the following     e To avoid the risk of electrical shock  do not remove any panels of the  product     e Refer servicing to qualified personnel   e Never operate equipment with damaged power line cords   e Refer service and repairs to a qualified technician     CAUTION  ELECTRIC SHOCK HAZARD     Electric shock hazard  Always turn off the System and disconnect the  power line cord from the source before performing any service or  maintenance procedures  or before moving the System        6 600479 001 Rev2 9 147    Safety Information    Primary Patient Temperature Probe  T1   Failure    The System relies upon the patient temperature reading from a YSI 400 type  thermistor connected to the Primary Patient Temperature Probe  T1   There are  rare failures of this type of thermistor that cannot be detected by the System with  100  reliability  Failure of the T1 can result in either patient hypo  or hyper   thermia  Death or serious injury to the patient may result  A secondary patient  temperature probe  T2  connection is therefore built into the system  For patient  safety  either use both the T1 and T2 connections or employ the T1 probe with  an independent frequent check of patient core temperature     WARNING  NEVER CLINICALLY USE A RESISTOR IN PLACE  OF THE T1 TEMPERATURE PROBE    ZOLL supplies fixed value res
113. y on your computer    Right Click the copied file and select  Create Shortcut  Move the shortcut to your desktop or other desired location   Double click on the shortcut to start TempTrend    CSV     Alsius Add In Installation    TempTrend    CSV is designed to use the features provided within Microsoft EXCEL  To  view the data in the patient data log you will use an Add In loaded into EXCEL  This  Add In is called  Alsius xla  Once it is installed  it will appear as a new item  called  Alsius  in the menu bar of EXCEL     M S  amp     3 600479 001 Rev 2 111 147    TempTrend CSV    E Microsoft Excel   Booki                Es  File Edit View Insert Format Tools Data Window Help  Je WdsighisaivGls  ae J7 i9     Arial  10   B Z7 UJ EBZSSZEBAL I    Adobe PDF  Open CS    Alsiiis          Graph cs    Activate Scroll Control    Figure 5 1  The Alsius Menu with CSV Functions    Automated Installation     Standard    Windows    Setup    If your Microsoft Windows is a standard setup  the Alsius Add In has already been placed  in the appropriate directory  To complete installation you should     Open Microsoft EXCEL    From the main menu bar  select  Tools     Add Ins   Click the box next to the word  Alsius   Click  OK   In the main menu bar  you should see the new item  Alsius    aia NY a    Manual Installation    If your Microsoft Windows is a non standard setup  the Alsius Add In will have to be  loaded manually     1  Using Microsoft Windows Explorer  open the TempTrend    CSV inst
    