Home

Poet® Plus 8100 Patient Monitor Service Manual

image

Contents

1. Figure 2 5 Service Mode Window From this point you can access all service mode functions and calibrations set operation values and perform tests All the following functions are accessed from the Service Mode window shown above e NIBP Service Mode e CO O Service Mode e IBP Service Mode e Processor Software Download e ST Arrhythmia Keycode e Permanent Alarm Silence NOTE After you perform each service mode function you must turn off power to the monitor Press the ON OFF button on the front panel of the monitor NOTE See Addendum 1 Arrhythmia and ST Analysis in the Operator Manual for information Poet Plus 8100 Service Manual Page 2 3 Section 2 Service Windows NIBP Service Mode From the Service Mode window do the following NOTE To service the NIBP module you need specified equipment See Section 6 for details 1 To run a high time or pressure test for adults or infants or to run an inflate deflate speed test with the menu knob highlight Enter NIBP Service and press to select The value NO is highlighted 2 Rotate the menu knob to highlight YES and press to select The NIBP Service Mode Parameters window to service NIBP appears Run Leak Adult High Time Test Run Infant High Time Test Run Adult High Pressure Test Run Infant High Pressure Test Run Infl Defl Speed Test ELAPSED TIME MAX PRESSURE NO NO NO NO NO RESULT Figure 2 6 NIBP Service M
2. 4 Label and disconnect the two display cables from J3 and J4 of the Main Board See Figure 7 11 Remove the Front Housing 5 Note the orientation of the membrane cable in the ZIF connector Unlock the ZIF connector Label and remove the membrane cable from J7 of the Main Board See Figure 7 11 Remove the Front Housing Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 3 Section 7 Disassembly 6 Label and remove the menu knob cable from J8 of the Main Board See Figure 7 12 Main Board Component Side J1 J3 J4 Ground Cable J7 J8 7 Disconnect the CO tubing from the water trap manifold assembly NOTE Reconnect the Manifold Trap Side to the unit s left side port L Figure 7 13 Disconnect the CO Tubing Page 7 4 Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc Section 7 Disassembly 8 Lay the display front housing flat on the table lt ower gt eer Figure 7 14 Lay Display Flat 9 Remove the ground cable Phillips screw which is held in place with a Phillips screw 10 Remove the display ground cable by disconnecting it from the locking connector 11 Remove the front housing assembly Poet Plus 8100 Service Manual Page 7 5 Section 7 Disassembly Front Housing The following procedure uses examples from a monitor with an Disassembly 10 4 inch active TFT display Rotary Encoder Knob 1 With a flat screwdriver pry t
3. Section 7 Disassembly 26 f the assembly is a Temp NIBP Assembly Place the assembly in its slot on the left Attach the NIBP tubing to the quick connect fitting There should be a ground cable left over If the unit contains an IBP Board Mount the IBP connector panel using two screws pn 40995B001 and attach them to the rear housing If the unit is 2 Temp only Mount the board onto the standoff s that are provided and secure with two screws Item 28 pn 40995B001 27 There should be a ground cable left over Attach the GND cable from the multiground to the NIBP fitting pn 42014B001 and secure it with the nut pn 40284B005 Verify that the tubing from the NIBP runs along the Front chassis on the left side clearing all the ferrites under the transducer tubing 28 Connect the SpO cable from P1 of the SpO Board to P5 of the ECG Analog Board NOTE If the unit has CO then keep the CO cable behind and to the left of the SpO cable at this point 29 Wrap the CO cable over the SpO cable and fold the cable so that the head of the connector faces you Attach a tie wrap 40067B003 to the whole cable covering the both cables and pull tight Keep the head of the tie wrap near to the right side near the SpO board and tucking it in under the board 30 Attach the other end of the printer cable to the printer assembly Item 18 pn 95604A001 NOTE This can be only used if a A001 or A011 is present If you use Main Boa
4. es CO Module ReplaceMent cccccccccesescescsecssescnsenecsseseesecscscscsaessesseecacnees 8 6 Mai BOI iis cuscedicesbstcnseesunscajaceuerdesecessehiacetatreddbcenateniecesgtiednl claperececuistercetours unt fapenes Power Module es Fuse Replacement sssini eaa e a a arae araa A aa eai daS a Taan sarees Fan Replacement vccciccccavties cs civscosssreorneaeoedscentsavscevsercenantataceccavsrecanebovetaveedeveees Power Supply Replacement ce AC Power Component 0 0 cecccecceeeeeeeeeeeeeeeeeesecee seas sees seeesneeseeesseeseeesneeeee Poet Plus 8100 Service Manual Criticare Systems Inc Section 9 Drawings and Schematics LishotiDrawinGs ii sisi en he iia oles aad chee Ac aleua ex adv cue a E aa anii 9 1 Assembly BOMs and Drawing S 0 cccccceseee eee eeeeeeeeeeseeeseeseaeseeeseaeeseeeeeaes 9 1 Pneumatic Assemblies ue PCB Drawings and Schematics 0 0 0 0 eee eceee eens eens sense eeeeseeeseeeeeeeeaeeeee 9 2 Os GOs ASSOMDIY 2veccsisccsctseiutinescedwed tiie E pe a a 9 3 O Module Assembly 1 9 3 CO Module Assembly we 9 3 O gt 2 Manifold Resistor Assembly we 9 3 COs Pneumatics ASSEMDIY cccecec ce ee ecee eee ceeeeeseeseeseeseseeeseeseeeseseeesoasnss 9 3 O CO Pneumatics Assembly ccecccsccccsscsecectscsecsecseeseeseneetecssesseaseesaeeaeaae 9 3 Front Chassis Assemblies 0 Front Chassis Base Front Chassis COoz ae Front Ch
5. 3 Continue to increase the simulator voltage up to 14VDC Verify that as the voltage is gradually increased over 14VDC the current tapers down from 620ma 4 Increase the simulator voltage up to 15VDC Verify the trickle charge rate of 15ma 6ma The actual voltage across the simulator should be about 13 5VDC Poet Plus 8100 Service Manual Page 5 33 Section 5 Preventative Maintenance Page 5 34 Test Both Batteries 9 Set the simulator voltage back to less than 10VDC Switch Battery Select switch to BATT2 position Start with the simulated battery voltage at less than 10VDC The ammeter should be 15ma 6ma Slowly increase the voltage Verify the bulk rate is 620ma 100ma when the voltage is greater than 10VDC Continue to increase the simulator voltage up to 14VDC Verify that as the voltage is gradually increased over 14VDC the current tapers down from 620ma Increase the simulator voltage up to 15VDC 10 Verify the trickle charge rate of 15ma 6ma The actual voltage across the simulator is about 13 5VDC Set the simulator voltage back to less than 10VDC 11 Remove the battery load simulator 1 Attach the test battery simulator into each of the slots on the power supply module Connect both battery packs to the power supply Set the power supply to 12 5VDC Turn on the monitor and verify that the two green battery icons are displayed on screen Lower the voltage with the ad
6. DOX SpO Module CO Module CO Absorber Criticare Systems Inc Section 8 Troubleshooting The DOX SpO module consists of a separate board that is attached near the bottom edge of the ECG Analog Board There is a communication connection for a ribbon cable pn 90866A001 at the bottom edge of the ECG Analog Board The ribbon cable extends to the SpO Board There is also a short ground cable to the ECG Analog Board 90866A001 that also contains a shield pn 90866A001 The plethysmograph input signal enters the SpO Board through a shielded cable that runs between the boards and metal chassis to the opposite side of the monitor The DOX SpO module is composed of digital components that are not effected by noise There is also no signal degradation or loss of calibration due to analog components Interference or noisy signal can be due to a worn cable assembly or DB 9 connector Replace the connector and cable assembly if necessary Always check the sensor when intermittent problems are reported Stretched or otherwise damaged sensor cables must be replaced NOTE The line frequency setting of the monitor must be set correctly for optimal SpO monitoring The COs Module Assembly pn 91355A001 is not a field repairable component The CO module and board require factory calibration tables that can be corrupted by improper servicing The equipment and procedures for complete servicing are beyond the scope of this ma
7. Replace the sensor Replace receptacle Reseat the connection Replace Board s Trap error or CO mal function O not responding COs functioning ok Trap is plugged or leaks Tubes to the trap manifold are loose or pinched Bad valve or pump Op Cell is loose O Cell depleted Cable not connected Replace trap Check manifold connections Contact CSI Service Dept Tighten O Cell Replace O Cell Attach the cable to top of O Cell No sound from speaker Speaker scratchy or unclear Volume turned down Main Board connection is disconnected or loose Speaker wire broken at speaker terminals Metal debris on speaker Bad speaker Set volume higher in Configuration window Reconnect J20 on Main Board Check speaker wires for breaks or bad crimp pins Remove debris from speaker drum Replace speaker COM 1 serial interface doesn t work Set to internal printer Incorrect or bad cable Setting mismatch Connector is bad Reset monitor to external printer Replace with correct serial cable Reset baud rate and other settings Replace the Main Board No ECG trace displays ECG Cable may be bad The connection between Main Board J12 and ECG Analog Board P4 is loose or disconnected Main Board or ECG Analog Board may be bad Replace with known good cable Reseat the connection Replace Board s Page 8 2 Poet Plus 8100 Service Manual Criticare Systems I
8. c Check TP8 for a value of 0 122VDC 0 002VDC Use TP5 as ground d Adjust pot R92 with a small plastic tuning tool until TP8 is 0 122VDC aoa e Verify that the NIBP zero base point is 0 1 mmHg f Apply electrical varnish to seal the pot setting Page 6 6 Poet Plus 8100 Service Manual Criticare Systems Inc NIBP Performance Criticare Systems Inc Section 6 Service Testing and Calibration g Re attach the video screen cables and gas sample lines if you labeled and removed them h Close the monitor Since you opened the monitor case you must safety test the monitor before you place the monitor back in service 12 Select NEXT on the program This sets the 0 mmHg level Select NEXT again to actuate the pump and achieve the high pressure cal point The monitor automatically activates the pump after accepting the zero pressure 13 Calibrate the high pressure point Allow the pressure to stabilize should be between 240 to 270 mmHg before you make adjustments NOTE The reference manometer must be able to have 0 01 resolution 14 Adjust the Cuff Pressure measurement on the computer screen to match the measurement of the manometer Press the Up and Down arrow keys of the computer keyboard to calibrate the reading on the PC screen This should be within 0 5 mmHg of the manometer reading 15 Select FINISH to accept the high calibration point setting This calibration gain setting is set electroni
9. e Never enter the service menu while monitoring a patient Service Mode Window To access the Service Mode located in the CONFIG window 1 Press the ON OFF button on the front panel to turn on the power to the monitor 2 Rotate the menu knob to highlight CONFIG in the main menu and press the knob once to select it The CONFIG window appears 15 SEP 2000 17 47 Freeze timeout 2 minutes Standby timeout 30 minutes Standby tone ON Alarm tone warning ON Print Device Serial Language ENGLISH Line Frequency 60 Serial Format TEXT Baudrate 19200 Analog Out Select ECG I Restore Factory Defaults NO Enter Service Mode NO Enter Simulation Mode NO Figure 2 1 CONFIG Configuration Window Criticare Systems Inc Poet Plus 8100 Service Manual Page 2 1 Section 2 Service Windows 3 Rotate the menu knob to highlight Enter Service Mode and press to select The value NO is highlighted EXIT Date 15 SEP 2000 Time 17 47 Freeze timeout 2 minutes Standby timeout 30 minutes Standby tone ON Alarm tone warning ON Print Device Serial Language ENGLISH Line Frequency 60 Serial Format TEXT Baudrate 19200 Analog Out Select ECG I Resto Factory Defaul NO Enter Service Mode Mode Figure 2 2 CONFIG Window Enter Service Mode Highlighted 4 Rotate the menu knob to highlight YES and press to select The Enter Service Mode dialog box appears Enter Service Mode gt gt gt CAUTION lt lt lt All monitoring
10. Alarm Tone Warning on page 2 10 CAUTION A e Before you disable this safety feature contact your local distributor because the Permanent Alarm Silence tone is required in certain countries Page 2 8 Poet Plus 8100 Service Manual Criticare Systems Inc Section 2 Service Windows Service Settings Several settings on the CONFIG window are password protected You should not change these functions during use and only biomed technicians or supervisory personnel should adjust the settings These functions include Standby Timeout Alarm Tone Warning Line Frequency Enter Service Mode Enter Simulation Mode To set the functions listed above do the following 1 Rotate the menu knob to highlight CONFIG in the Main menu and then press the knob once to select it The CONFIG window appears 15 SEP 2000 17 47 Freeze timeout 2 minutes Standby timeout 30 minutes Standby tone ON Alarm tone warning ON Print Device Serial Language ENGLISH Line Frequency 60 Serial Format TEXT Baudrate 19200 Analog Out Select ECG I Restore Factory Defaults NO Enter Service Mode NO Enter Simulation Mode NO Figure 2 11 CONFIG Parameters window Rotate the menu knob to highlight the service function that you want and press to select Service Password To enter the service mode enter the password PIA418 by doing the following 1 Rotate the menu knob to select a character 2 Press the menu knob to enter each chara
11. CO Absorber Page 7 20 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly 5 Turn the monitor over and remove the two bottom screws NOTE Early model units have front rubber feet Later production units have a single front foot brace Both the feet and the brace use the same two screw mounts e Figure 7 38 Remove the Foot Foot Brace Screws 6 Gently pull the front chassis from the rear housing This exposes the gas module assembly if the monitor has the optional CO or CO2 O modules Figure 7 39 Remove the Front Chassis NOTE When you reinstall the metal chassis assembly reconnect all tubes Do not crush or pinch the tubes when you replace the chassis Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 21 Section 7 Disassembly 7 To expose the power cables carefully lay the front chassis assembly flat onto the bench top Page 7 22 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly ECG Analog Board The ECG Analog Board is located behind the Main Board The SpO Board is located near the lower right corner of the ECG Analog Board Removal Replacement Criticare Systems Inc 1 2 Se Detach the communication ribbon cable at P5 of the ECG Analog Board There is a hard wire ground between the ECG Analog Board and the SpO Board Desolder the ground wire at the ECG Analog Board Se j ee
12. Preventative Maintenance Page 5 32 BNC ECG Output Nurse Call You need a MedSim 300 patient simulator to perform this procedure 1 N ODO oO fF W PD 1 Connect a BNC cable with 0 1uf CAP signal to ground from the ECG BNC on the monitor to an input channel on an oscilloscope Select and press the PARAMS softkey Set Filter to Diagnostic and ECG Sensitivity to Medium Select and press the DISPLAY softkey Set Waveform 1 for ECG Lead II ECG IN on the screen Set the MedSim 300 patient simulator for Sq 2HZ Verify that the baseline is zero and signal is a peak to peak voltage of 1 2V 0 2 Connect the Smart Sat and monitor all the levels Insert the Nurse Call fixture into the back of the monitor while the Smart Sat simulator runs The lights on the fixture should illuminate when any limit alarm s triggered Press the Alarm SILENCE key Verify that the Nurse Call turns off Poet Plus 8100 Service Manual Criticare Systems Inc Battery and Power Testing Equipment Q Current Procedure Bulk Charge Test Criticare Systems Inc Section 5 Preventative Maintenance This procedure provides a method of testing the power regulation and battery charging circuits of the main board This test does not require you to open the monitor housing and break the warranty seal The following test recommends using a battery dummy block The dummy block is the same size as the battery
13. use Tubing Set pn 95632A001 e For monitors without O use Tubing Set pn 95632A002 You can replace failed or intermittent pump assemblies pn 94253A001 See drawing 92788A011 in Drawings and Schematics in Section 9 The CO Module Assembly CO only pn 91766A001 CO with Oz pn 91767A001 is available as a total replacement Replacement assemblies are pre tested as a finished assembly module Before you replace the CO module confirm that the problem is not due to a bad absorber poor cable connections leaky pneumatics or a failed pump assembly The new CO module must be field calibrated with the pump assembly that is installed with it You can recover the O module pn 91354A001 from the old CO module Recalibrate the O module after performing any service on the CO module The Main Board pn 91340A003 contains the primary processors storage memory NIBP circuitry communication interfaces audio driver vital signs module connections power regulation battery charge circuits and video generation and keypad inputs User I O The main ground cable pn 95637A001 and the display ground cable pn 95642A001 are part of the Main Board assembly The Main Board has an adjustable pot R29 that you set during NIBP calibration There are two CPUs on the Main Board that you must program from a CSI Service Download Station The programmable settings remain in non volatile memory Repair or improper handling of the Main Bo
14. 01 02 03 04 07 10 12 15 19 20 22 CSI Part 45113B001 95605A003 95638A001 45115B001 42141B001 45116B001 42140B001 46211B001 40284B005 40995B005 41371B001 Description HOUSING FRONT AY DISPLAY 10 4 NEC CBL FRNT HOUS TO MAIN BRD COVER WATER TRAP OPENING KNOB ROTARY ENCODER CUSTM MEMBRANE SWITCH 8100 ROTARY ENCODER LABEL NO CO2 NUT 10 32 KEPS PL P H M S 4 40X 25 SEMS TAPE INSUL KAPTON 2 WIDE Front Housing 93259A006 FRONT FRONT AY 8100 ACT Assembly with CO Criticare Systems Inc Item 01 02 03 05 07 10 12 14 19 20 21 22 CSI Part 45113B001 95605A003 95638A001 42177B001 42141B001 45116B001 42140B001 42270B002 40284B005 40995B005 40297B002 41371B001 Poet Plus 8100 Service Manual Description HOUSING FRONT AY DISPLAY 10 4 NEC CBL FRNT HOUS TO MAIN BRD COVER WATER TRAP MANIFOLD KNOB ROTARY ENCODER CUSTM MEMBRANE SWITCH 8100 ROTARY ENCODER WATER TRAP MANIFOLD NUT 10 32 KEPS PL P H M S 4 40X 25 SEMS TUBING SLCN 093IDX 3900D TAPE INSUL KAPTON 2 WIDE Page 9 11 Section 9 Drawings and Schematics Final Assemblies Final Assembly 93951A386 8100 FNL BASE 2 TEMP 6W 2Temp6Wave Item CSI Part 01 93258A001 03 93259A005 05 92788A001 07 95612A001 09 91340A003 10 91353A002 12 95618A006 15 90871A002 17 45114B001 19 41891B001 21 42225B001 22 42227B001 25 40284B001 26 40284B002 27 42065B004 28 40995B001 29 40995B006 33 40496B001 35 40876B001
15. 12 Page 5 10 Poet Plus 8100 Service Manual Criticare Systems Inc CO Absorber Replacement Criticare Systems Inc The The Section 5 Preventative Maintenance CO absorber removes COs from air samples during calibration se samples provide a base line for the measurement of ambient CO 1 2 3 4 Turn off the monitor Do not use the monitor for five minutes Remove the two Phillips screws from the rear access door Use a flathead screwdriver to pry the rear access door off the back of the monitor Look for the CO absorber It is a clear tube containing light colored granules a For early units the absorber is located to the left of the O board and tie wrapped to a valve assembly b For later units the absorber is attached directly to the access panel door CO Absorber we ee lt lt ui Figure 5 9 CO Absorber Newer Models The granules are normally white indicating a good absorber If the granules remain violet or blue the absorber is expired and you must replace it NOTE The absorber can change color during use but should always return to its original white color when not in use Remove the expired CO absorber where it attaches to the larger diameter tube Replace with pn 40570B003 For early units you must disassemble the monitor to remove and replace the absorber Refer to the disassembly instructions which are located in Disassembly in S
16. 2 Turn to Incr Decr Press to Save 02 C02 Wave Alignment Turn to Incr Decr Press to Save 02 Factory Defaults No Action Figure 2 7 CO O5 Service Mode Window 3 Press the ON OFF button on the front panel of the monitor to turn off power to the monitor Criticare Systems Inc Poet Plus 8100 Service Manual Page 2 5 Section 2 Service Windows IBP Service Mode From the Service Mode window do the following NOTE To service the IBP module you need specified equipment See Section 6 for details 1 To start or restart CAL store CAL results and set mmHg Point values with the menu knob highlight Enter IBP Service and press to select The value NO is highlighted 2 Rotate the menu knob to highlight YES and press to select The Service Mode window appears to provide functions and settings for IBP Select Channel Number 1 Start or Restart Cal NO Set mmHg Point NO 0 0 Set 256 mmHg Point NO 838 8 Store CAL Results Pressure Reported 0 Current CAL Values in EPROM OFFSET GAIN IBP 1 0 0 838 8 IBP 2 0 0 838 8 Figure 2 8 IBP Service Mode Window 3 Press the ON OFF button on the front panel of the monitor to turn off the power to the monitor Page 2 6 Poet Plus 8100 Service Manual Criticare Systems Inc Section 2 Service Windows Processor Software Use the Software Download parameter from the Service Mode Download window to download software for the following processors e Ma
17. 2Temp 6Wave eeeeeeeeeeeeeeneeeneeeaees 9 22 Final Assembly Printer CO Os IBP 2TOMD ccececesceecsseseeeeteeeceesetsetsenee 9 23 NIBP PUMpP ASSEMDIY lt ainaani ivsisnetecvicvevels iceusuuasDieea cise aaa e T rI 9 24 Printer MOdUIC vies ccscveccreencerscaccen inaen naaede riaka inidan 9 25 Display ASSOMDIY jie iyana veut an teverensetedecusdecibctucsunvantdussdsstadunauesanes 9 26 Power Module Assembly 0 eee cee cece eeceeeeeeeeeaeeceeeseaeeeaeseaeeseeesnaeeeaueseeeesaeeeaees 9 27 Connector Assemblies Connector Assembly Temp NIBP uu cee eeec cece eereeeeeeneeenaeeeaeeeeaeeeeeeeneeed 9 28 Connector Assembly 2IBP 2Tem NIBP ecceecceeee eee tesserae eee eeeeeeneeeeeeees 9 28 Connector Assembly 2T emp e cece cece eeneceeeeeeeeseeeeeeeeeaeeeseeseeeenaeseaeened 9 28 Tubing ASSEMDIICS eee cece eee eens eee eeeaeeceeeecaee cae seaeecseeseeeseseeeeesaeeseaeessaeseeeneees 9 29 Tubing ASSembly O pervs issues a e pandas ancl Ea 9 29 Tubing ASSeMbl COo z eher padaan e dernei ner Maheda r Apii dan 9 29 PAT Input Tubingsrtrnnanri toda edeti 9 30 Absorber TUDING Siani inia a A a NE 9 30 Pressure TUBIN tsi aaant irena i aaa a aeaa E 9 30 Bleed Check TUDING xcs ccesseatvecvnensoctcdincnesvadesinyaceecsercnstsvielvobacy dita a Aeeiiaea 9 30 O Valve Tubing Only with pn 95632A001 sssessseseesreeserrerrnrinerrnrresrnesenens 9 30 ValVeS PUMP TUBING aasar e aaa a aa a cua ii
18. 5A 10 5D 2 5VREF See Assembly Tests on page 7 40 4 If the values return to normal with the SpO Board removed replace the SpO Board For more information see procedure 91342T002 NOTE The ECG Analog Board provides power to the SpO Board The DOX SpOz Board can be replaced as a serviceable component The downloadable software settings associated with the SpO module are stored on the ECG Analog Board No calibration or SpO Board preparation is required Page 6 4 Poet Plus 8100 Service Manual Criticare Systems Inc NIBP Calibration Criticare Systems Inc Section 6 Service Testing and Calibration Equipment Required CAT 1320 kit which includes PC computer configured for use with CSI Service Software Install the service software according to the procedure that comes with it 8100 serial download cable 700cc dummy cuff ESD protected workbench Digital reference manometer Dynatech Nevada NIBP Analyzer Connect a serial download cable from the PC serial port to COM2 J10 on the Main Board Set the DIN download cable to NIBP USER COM2 i i III i 7 Open the service tool on the computer Select Start gt Programs gt CSI NIBP Service gt NIBPSvc Close the disclaimer dialogue box 3 Click on Connection and select Options 4 Select COM1 for the port and select Poet Plus 8100 for the model Click on OK 5 Click on Connection and select Open in Servi
19. 8100 46 40909B002 SPCR HEX 6 32X 625 NY M F 47 40496B006 P H M S 6 32X 43 SEMS 48 42132B001 ACCESS DOOR 02 SENSOR 50 42014B001 FITNG QUICK CON HH NIBP 51 40284B005 NUT 10 32 KEPS PL 52 83375B001 COVER CAN CONNECTOR 8100 53 40995B005 P H M S 4 40X 25 SEMS 60 40873B001 P H M S 8 32X 375 PL PH Page 9 20 Poet Plus 8100 Service Manual Criticare Systems Inc Final Assembly PrinteyCO O gt 2Temp6Wave Criticare Systems Inc Section 9 Drawings and Schematics 93951A686 8100 FNL CO2 02 2T 6BWAVE ltem CSI Part 93258A001 93259A006 92788A011 95612A001 91340A003 91353A002 95618A006 90871A002 95604A002 41891B001 42225B001 42227B001 40284B001 40284B002 42065B004 40995B001 40995B006 40496B001 40876B001 40873B004 42175B001 83278B001 40909B002 40496B006 42132B001 42134B003 42014B001 40284B005 83375B001 40995B005 40067B003 40873B001 Poet Plus 8100 Service Manual Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS FRNT 8100 CO2 02 AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CONN AY 2 TEMP AY CBL ECG ANALOG TO OUT PRINTER MODULE 8100 9100 DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 SPCR HEX 6 32X 625 NY M F P H M
20. Analog Board are not recommended for field repair and replacement You can replace the ECG Analog Board assembly as a serviceable component Replacement ECG Analog Board assemblies pn 91353A002 are pre tested software loaded and calibrated The monitor requires complete safety testing after reassembly NOTE Always check line frequency filter settings and the ECG lead cables as part of any ECG trouble shooting Always use the recommended simulators for performance testing There are three pots under the shielding at locations R112 R110 and R108 They are the baseline adjustments for ECG EGG Il and ECG V respectively These are factory set you should not adjust them Set the S1 switches off in the open position to allow communication with the DOX module See ECG Verification on page 5 21 The ECG Analog Board contains its own on board software that is factory loaded The program is stored in non volatile memory on the ECG Analog Board Repair or improper handling of the ECG Analog Board can corrupt this software The respiration circuit is located on the ECG Analog Board There is no separate calibration or testing for the impedance respiration See procedure 91342T002 for more information The temperature module is located on the ECG Analog Board There are no on board settings or service calibrations The temperature module is auto calibrating upon start up Poet Plus 8100 Service Manual Criticare Systems Inc
21. Before You Begin on page 7 1 Main Board Removal 1 Remove the front housing as described in Front Housing Replacement Removal on page 7 2 2 Remove the five Phillips screws shown in the picture below Figure 7 29 Main Board Screws 3 Disconnect the printer cable from J1 on the Main Board CAUTION A e The Main Board has a 10 pin connection J12 directly into the ECG Analog Board that is located behind it Do not bend the pins when you remove the Main Board NOTE Make sure that you remove the black coverplates from the COM1 and the Video connectors Page 7 14 Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc Section 7 Disassembly 4 Carefully pull the Main Board away from the front chassis The board only pulls forward slightly NOTE P2 PS1 PS2 J14 J16 J18 J19 and J20 are on the solder back side of the Main Board J20 Figure 7 30 Main Board Solder Side Connections 5 Label and disconnect the battery power cable at PS1 on the Main Board Label and disconnect the Power Supply Board cable at PS2 on the Main Board Label and disconnect the IBP ribbon cable at J14 on the back of the Main Board NOTE This cable is only present if the monitor has the IBP option Label and disconnect the CO ribbon cable at J16 on the back of the Main Board NOTE This cable is only present if the monitor has the CO O option Label and discon
22. CO using the principles of infrared absorption spectrometry An unknown concentration of gas CO3 is calculated by comparing its absorption of infrared light to that of a known standard The absorption of light is directly related to the concentration of the gas As infrared light passes through the sample gas chamber the light transmitted is converted to a voltage signal The patient monitor converts the voltage to CO concentration and expresses it as mmHg percent or kPa user selectable Infrared analysis of the gas samples is done using Beers Law The formula for Beer s Law Ihe A cd I Infrared value of measured sample Io Infrared value of light source e Exponential function A Extinction coefficient c Concentration of the gas sample d Distance measured through the sample The Beers Law calculation is performed by the patient monitors software Poet Plus 8100 Service Manual Criticare Systems Inc Conditions of Use Stability of Accuracy N2O Compensation Criticare Systems Inc Section 1 Introduction The Poet Plus 8100 patient monitor is calibrated with dry NIST traceable calibration gasses at room temperature and pressure 21C 740 mmHg Given the small effect of water vapor upon the CO measurement and the unit s built in temperature and pressure measurements and compensations this monitor s method of gas analysis per EN 864 is best described as ATPS Ambient Temperature an
23. N EA E A AA 5 33 Q Current ProCceduyre eeccecesceeseeseeeeceneeeeeeaeeeeeeeeaeeeeeaesaeesaeseseaeeaeeeaeeaeeneeeas 5 33 Bulk Charge Test wes iscceceid acess cevscecies a a ies teabiadlenesnde shop sukevadtdotnededtoreetevees 5 33 Test Both Batteries ccs culectsieccerecdectesvestchetecdocvedtertectevoeredecbacdeabeveetevsecestteys 5 34 Test Single Battery wcciccccrcsccccesccccnsciscaveoesetentececconcetensesutonsenccaseentatsasevers diaaa 5 35 Functional and Safety Testing Checklist cece ceeeceneeceeeeeeeseeeseeeseeeeeeeaees 5 36 Section 6 Service Testing and Calibration Field Service Testing Safety Field Service Test Matrix Field Calibration Procedures DOX SpO Performance Testing ec eeceeeceee eens eenee cee eesaeseaeeseeeseaeeeaeeeeeeenaeeeaeeege Programming the SmartSat Analyzer s Test Procedure ii laifi t inl AU Mie a ed eas eed aeia SpOvs Board Test Values cccccccccccscsscsecsscseesenseneeeeeessessnsescsaessessessnsenecsseaees NIBP Calibration ceeeeeeeees Manual Zero Point Calibration NIBP Performance 7 Temperature Input Check 0 0 se ecsccesesseessceecssneeseecseeseseesseeseceecsseeeseecseeseseeseaeseaeeones Gas Calibration 23 28 fone ich E ae ete eens re Calibration Kits Before you Begin CO Flow Check COs NV niliG atone ces tote ciel eens ete eaaa NSE Hates NEE aen EA COs User Calibration sci rer Ea aaa Aer i r a EnA LAERE E
24. N20 WC KH 1 11 0 39 ALARA PARAMS DI PLAYT ZERO IPI V000 NO ADMIT 21 10 5 64 ADM DIS PORRI PRINT ZERO IP2 Mm Mm Adult Mixed ENF E10 5 110 0 14 12 59 ART1 HIGH DIA Figure 2 12 Example of the Window in Simulation Mode Criticare Systems Inc Poet Plus 8100 Service Manual Page 2 11 Section 2 Service Windows Custom Profiles Setting Patient Size Page 2 12 The monitor has a set of default settings that you can configure to the needs of the facility These default settings are password protected The monitor is designed to look at the patient size information selected in the ADM DIS window and determine whether the monitor should use one of the following alarm settings while monitoring e Adult e Pediatric e Neonatal When you change the patient size in the ADM D S window or the ALARMS window the monitor determines which window appears when you select the ALARMS softkey When the user changes patient size the user effectively changes all the alarm limits for all of the following e SpO2 e ECG e NIBP e IBP e Temperature e Respiration e CO O e Maximum NIBP pressure limit To enter patient size do the following 1 Rotate the menu knob to highlight ADM DIS and press to select 2 Rotate the menu knob to highlight Update select YES and press to select You can now select the patient data field 3 Rotate the menu knob to highlight Patient Size and press to select 4 Rotate
25. NOTE More than one click clockwise or counter clockwise changes the flow setting Verify the flow with a flow meter Use a flowmeter to measure the flow The flow should be 50 3 ml min If the flow rate is not correct turn the rotary knob clockwise to increase the flow or counter clockwise to decrease the flow When the flow is within specification press the rotary knob to save the setting NOTE The low flow setting 50 ml min is not as stable as the high flow setting 150 ml min Rotate the rotary knob to highlight CO2 Zero Cal and press to select The value Request CO2 Zero Cal is highlighted Press the rotary knob to select The value CO2 Zero Cal is highlighted Wait for zero calibration to complete Poet Plus 8100 Service Manual Criticare Systems Inc CO Verification CO User Calibration Criticare Systems Inc 1 Section 6 Service Testing and Calibration Connect a sample line Cat No 625 and water trap Cat No 938 to the monitor Allow the monitor s power to remain on for at least one hour to warm up the monitor and stabilize the temperature From the PARAMS window select the CO unit of measure For greatest accuracy use percent CO for the CO unit of measure Confirm that N2O Compensation is turned off 4 Connect the bottle gas either 10 or 5 5 From the DISPLAY window set a waveform TYPE to ET CO2 EXIT from the DISPLAY window Verify that the CO funct
26. PRINTER MODULE 8100 9100 DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 43 SEMS ACCESS DOOR 02 SENSOR CONN MNT PLT 2TEMP FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS P H M S 8 32X 375 PL PH Criticare Systems Inc Final Assembly CO 2Temp 6Wave Item 01 03 05 07 09 10 12 15 18 19 21 22 25 26 27 Criticare Systems Inc CSI Part 93258A001 93259A006 92788A003 95612A001 91340A003 91353A002 95618A006 90871A002 95604A002 41891B001 42225B001 42227B001 40284B001 40284B002 42065B004 40995B001 40995B006 40496B001 40876B001 40873B004 42175B001 83278B001 46220B001 42220B001 40909B002 40496B006 42132B001 42134B003 42014B001 40284B005 83375B001 40995B005 40067B003 40873B001 Poet Plus 8100 Service Manual Section 9 Drawings and Schematics 93951A676 8100 FNL CO2 2T 6WAVE Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS FRNT 8100 CO2 AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CONN AY 2 TEMP AY CBL ECG ANALOG TO OUT PRINTER MODULE 8100 9100 DB9 COVER PLATE SUPPORT
27. Service Manual Criticare Systems Inc Section 1 Introduction Temperature The patient monitor measures body temperature using a thermistor Measurement temperature sensing elements in the temperature probe The thermistor can sense changes in body temperature by changing electrical resistance e Unusual fast artificial variations in temperature readings may occur with accompanying applications of an electrocautery system e Electrical leakage current of the cable when used with the monitor and sensor comply with IEC 601 1 EN 60601 1 The monitor is compatible with any YSI 400 or YSI 700 series temperature probe Criticare Systems Inc Poet Plus 8100 Service Manual Page 1 11 Section 1 Introduction Specifications Page 1 12 ECG Connectors Lead Selection ECG Sensitivity Frequency Response Electrosurgery Protection HF Equipment Protection Defibrillator Protection Pacer Detection Rejection Sp0 Range Resolution Accuracy Indications Method Modes Operation Sensor Wavelength Sensor Power Heart Rate Source Range Accuracy Pulse Tone Respiration Source Rate Range Resolution Accuracy Poet Plus 8100 Service Manual 3 or 5 Lead Standard AAMI 3 Lead Il Ill 5 Lead Il Ill aVR aVL aVF V Low Medium High Diagnostic 0 05 100 Hz 3db Monitor ESU 0 50 40 Hz 3db ST Filter 0 05 40 Hz 3db Yes Yes Yes Yes 1 99 1 7
28. Software Error Related Hazard Mediation 0 cccccccsssseceesssesseeeeeeesesseaeeees 1 18 Potential INterterene aisi nenna a iai aE EENE 1 18 Use of AnestheticSers cession rntarendtenaveden a a e 1 18 Biocompatibility EIEEE a AE E E EE E E A Criticare Systems Inc Poet Plus 8100 Service Manual Page iii Page iv Section 2 Service Windows SOrvice Mode WindOWsiesaieccsdceseetsocacaciceaveesectedaads ceaetteedoedasaadhacavevseavendscaadessssendeedeedeoesse 2 1 Service Mode WindOW cc cccccccecsesseceeeccecseneeeeeeescseeeeeeeeessesnsaeeeeesseesnaeeeeess 2 1 NIBP Service Mode a COg Op Service MOE n isseire nerian ei iienaa eee a ieni aiae 2 5 IBP S rvicS Modens a eerie aia ee en a a GS 2 6 Processor Software Download Permanent Alarm Silence Service Settings ee 2 9 Service PASSWOI eee eee cece ceeeceeeeeeeeeceeeseaeseaeeceaeseaeeseaeeseeeseaeeseeeeeeeeaeeeaees 2 9 Standby TIMCOUts 6 cccsececusstescencuenuecedatens i aeiia dae a 2 10 Alarm Tone Warning 2 10 Line Frequency 2 10 Simulation Mode wee Custom Profiles 0 ccc cece eene cece sees eeeeceaeeeeaeseaeeeeeeeeaeesaeseeeseaeseadeseeeesneaseneees 2 12 Setting Patient SiZe smira anren ia aaea N a A aa acieesvssavereciee 2 12 Creating Custom ProfileS ssessssssesresssssnereneruseeserureneresreneruerenernsensenntnasenernnne 2 13 Section 3 Theory of Operation POWerMOGUIC ricsticnetian dss ticcincs a a a a t
29. To replace a battery lift up the locking tab and sliding the battery into the bay The contact end of the battery goes in first aes Lift Locking Tab 4 Contacts 4 _ Go in First Figure 5 3 Replacing a Battery Page 5 6 Poet Plus 8100 Service Manual Criticare Systems Inc Fuse Removal Replacement Criticare Systems Inc Section 5 Preventative Maintenance There are two AC power fuses located at the rear of the monitor directly above the AC power entry socket e United States monitors use 1A 250V Slo Blo fuses e International monitors used in 220 240 volt environments use 0 5A 250V Slo Blo fuses 4 Press in the side clips at the same time with a tool and lift out the small black access cover The two fuse sockets are visible Clips Figure 5 4 Remove the Fuse Cover Figure 5 5 Fuses Exposed Poet Plus 8100 Service Manual Page 5 7 Section 5 Preventative Maintenance 5 Gently pull the fuses out of the fuse cover assembly eR il Figure 5 6 Fuses Pulled 6 Reassemble in reverse order For more information about troubleshooting power problems refer to Troubleshooting in Section 8 in this manual Page 5 8 Poet Plus 8100 Service Manual Criticare Systems Inc Oxygen Sensor Replacement A hisd Criticare Systems Inc Section 5 Preventative Maintenance CAUTION A Use a properly grounded ESD dissipating wrist strap durin
30. Verify that you get an audible response from the speaker A Trend window appears on the display Press the TREND key again to close the Trend window Or optionally press and hold the TREND key to open the TYPE INTERVAL SCREEN menu that you close by selecting EX T Press the STANDBY key Verify that you get an audible response from the speaker A standby confirmation menu should appear Press the STANDBY key again to close the menu Press the DEFAULT key Verify that you get an audible response from the speaker The Custom Default menu appears Select NO and then select EXIT to return to the Main window Poet Plus 8100 Service Manual Page 5 19 Section 5 Preventative Maintenance Alarms Verification To verify the alarm circuitry for SpO perform the following procedure 1 In the PARAMS window turn off all monitoring modules except for SpO Set the display to show a plethysmograph waveform Inthe PARAMS window set the SpO2 Low Limit Alarm setting to HIGH This sets the alarm priority for only this alarm parameter Use an SpOz simulator to set the monitor to display the plethysmograph waveform Confirm the heart rate and saturation reading In the ALARMS window adjust the SpO low alarm level above the saturation reading to cause an alarm condition Verify that the message LOW SPO2 appears at the top of the waveform channel in red letters Verify that you get an audible response The ala
31. alongside the LEDs measures the intensity of each wavelength transmitted through the monitoring site The light intensity is converted to an electrical signal which is input to the monitor The effects of skin pigmentation venous blood and other tissue constituents are eliminated by separating out the pulsating absorption data Poet Plus 8100 Service Manual Criticare Systems Inc SpO Clinical Testing and Heart Rate Criticare Systems Inc Accuracy Section 1 Introduction All Criticare oximeters DOX compatible have SpOz calibration tables which were originally generated by monitoring desaturated human patients or volunteers and matching their displayed SpO value to the value determined by sampling arterial blood and measuring functional SaO with a clinical laboratory grade multi wavelength optical oximeter i e CO oximeter The final SpO5 calibration curve was then generated based upon numerous patients data over the range of 40 to 99 SaO All accepted data were taken from patients with dyshemoglobin i e carboxyhemoglobin methemoglobin concentrations near zero This oximeter is a two wavelength device which is calibrated to measure functional SpO only when dyshemoglobin concentrations are near zero The accuracy specifications of this device will not be met with high concentrations of dyshemoglobins Significant concentrations of carboxyhemoglobin results in a higher displayed SpO value than is actu
32. and replace the display screen NOTE Be sure the display and the inside of the clear cover are clean before you re assemble the assembly 6 Reassemble in reverse order Page 7 12 Poet Plus 8100 Service Manual Criticare Systems Inc Printer Removal Replacement Criticare Systems Inc Section 7 Disassembly 1 Remove the paper roll from the unit 2 Remove the two Phillips screws that hold the printer assembly Figure 7 27 Remove the Printer Housing Screws 3 Disconnect the ribbon cable from CN1 on the the Printer Board 4 Remove the printer assembly 5 Replace the printer assembly and reassemble the monitor 6 Remove the four Phillips screws from the Printer Board STITT Pw ware ee ont ej J wa o M lt Figure 7 28 Printer Board Screws 7 Lift the Printer Board away from the unit 8 Label and disconnect the cables from CN3 and CN2 on the back of the Printer Board 9 Remove the metal plate and shielding 10 Remove the printer from the housing 11 Replace the printer and or Printer Board and reassemble in reverse order Poet Plus 8100 Service Manual Page 7 13 Section 7 Disassembly Rear Housing Follow the ESD cautions discussed in Electrostatic Discharge Disassembly Protection on page 7 2 Because the interior of the rear assembly contains exposed AC contacts you must disconnect the AC power before you begin the disassembly procedure See
33. breaks or shorting The AC to DC converter of the power supply should be able to generate 15 6 volts See Service Testing and Calibration in Section 6 for test procedures The power entry assembly pn 95641A001 is composed of two AC chokes a ferrite and power cable connections The source connections attach directly to the AC receptacle fuse assembly pn 80124B002 Replace broken cracked or loose AC receptacles Replace the receptacle if the external pins are bent or broken Do not attempt to remove or modify the ground pin Check the cables and chokes for heat damage before power module replacement Remove the assembly from the housing and inspect the back side of the board for scorching or damaged solder connections Replace the entire assembly if there are any signs of high current damage Poet Plus 8100 Service Manual Criticare Systems Inc Section 9 Drawings and Schematics List of Drawings Assembly BOMs and Drawings Criticare Systems Inc Title Op Module Assembly CO Module Assembly O gt Manifold Resistor Assembly CO Pneumatics Assembly O2 CO gt Pneumatics Assembly Front Chassis Base Front Chassis CO Front Chassis IBP Front Chassis CO gt 2 O gt 5 Front Chassis CO IBP Front Chassis CO O IBP Rear Housing Assembly Front Housing Assembly without CO Front Housing Assembly with CO Final Assembly 2Temp 6Wave Final Assembly CO 2Temp 6Wave Final Assembly I
34. cannot adhere properly 5 Remove the protective backing from the adhesive on the new membrane O Install the new membrane Line up the membrane along a straight edge and lay the membrane into place Press down firmly on the membrane when it is in position 7 Reattach the rotary encoder knob to the front housing Water Trap Manifold 8 Remove the CO cover from the manifold located on the front of Removal Replacement the unit CO Cover Figure 7 18 Remove the CO Cover 9 Remove the two Phillips screws from the back of the housing Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 7 Section 7 Disassembly 10 With a flat screwdriver release the two tabs to free the manifold and the water trap manifold cover Manifold Screws Figure 7 19 Remove the Water Trap Manifold 11 When you install the new manifold be sure to insert the lower tab into the recess first 12 Be sure the cover locks firmly in place 13 Replace the two Phillips screws Page 7 8 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly Display Assembly 1 Remove the four nuts from the interior frame of the front housing Removal Replacement Figure 7 20 Remove the Display Nuts 2 Lift the front up Allow the cables to slide through the access hole 3 Remove the four screws that hold the display housing together Figure 7 21 Remove the Display Housing Screws 4 Remove the b
35. computational functions and provides integrated I O functions including timers I O pins serial interfaces and analog I O The CONM module contains a microprocessor and associated circuitry for NIBP and the ability to control the optional Capnometer Module No electrical connection is made with the patient from the monitor for NIBP The vital signs parameters are passed to the main module using a serial channel The vital signs module is a single printed circuit board with connectors to the Main Board the ECG input leads the SPO connector and the temperature sensor The vital signs module contains a microprocessor and associated circuitry for pulse oximetry temperature ECG signals and TTI respiration ECG signals can be either 3 lead single channel or 5 lead triple channel Patient isolation for galvanic contact is required for the temperature and ECG functions Patient isolation for non galvanic contact is required for the pulse oximetry sensor A CAN bus passes the vital signs parameters to the main module The main module sends the vital signs module control information and the vital signs module sends the main module waveforms and hardware status The communication cycle is 20 Hz for the vital signs information from the VS module twice a second for the status of the VS module and as needed for the control information from the main module The speaker drive circuit is a DAC controlling speaker volume and an output pin to con
36. hemoglobin to the sum of oxygenated hemoglobin plus hemoglobin which is available for binding to oxygen as expressed in the following formula oxyhemoglobin oxyhemoglobin deoxyhemoglobin X100 percent oxygen saturation Dyshemoglobins such as carboxyhemoglobin and methemoglobin are not directly measured and therefore are not factored into the measurement The patient monitor does not use analog circuitry for signal processing Digital signal processing in the microprocessor results in lower noise from circuitry components resulting in a cleaner signal and better performance under low perfusion conditions There is also improved rejection of noise from the patient and environment due to the availability of the true unfiltered sensor signal for digital signal processing The digital pulse oximeter measures oxygen saturation and pulse rate using the principles of spectrophotometry and plethysmography The sensor is completely non invasive and there is no heat source that could burn the patient The pulse oximeter sensor contains two types of LEDs Each type emits a specific wavelength of light Since oxygenated hemoglobin and deoxygenated hemoglobin absorb light selectively and predictably we can determine the amounts of these two compounds by measuring the intensity of each wavelength that passes through the measuring site The light from the LEDs shines into a pulsating vascular bed A photodetector located opposite or
37. load that can be used for this test is a foam peanut used for packing Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance NIBP Verification The NIBP verification requires Dynatech Nevada NIBP Analyzer Connect the 8100 monitor to a Dynatech Nevada NIBP Analyzer set for the following operation NIBP Analyzer Settings Adult 120 80 90 Heart Rate 80 bpm Neonate 80 50 62 Heart Rate 80 bpm Pressure Adjustments Gain 100 Shift 4 1 Use the T connection with the dummy cuffs for these tests Connect them to the 0 300 mmHg port of the NIBP analyzer 2 Connect the monitor to an AC mains power source and turn on the monitor 3 Press and hold down the monitors NIBP CYCLE STAT key The monitor s cycle time window appears 4 Select the NIBP Cycle Time to two minutes 120 80 OPERATION SETTING 1 In the monitor s ADM DIS window set the Patient Size to Adult mode 2 Connect the NIBP hose to the fitting on the monitor 3 Press the NIBP key 4 Allow the monitor to take at least four 4 readings Calculate the average of the four 4 readings and verify that each reading does not vary by more than 4 or 4mmHg whichever is greater from the calculated average Each reading shall not vary more than 8 mmHg from the simulator setting 5 Recalibrate the monitor if necessary 80 50 OPERATION SETTING 1 Set the monitor to the Neonate mode 2 To verify STAT mode operation pr
38. monitor shuts down between 10 2VDC 5VDC Poet Plus 8100 Service Manual Page 5 35 Section 5 Preventative Maintenance Functional and Safety Testing Checklist Functional and Safety Testing Checklist Serial Number Software Revision Electrical Safety Testing Withstanding Voltage Hipot Hot Neutral to Chassis Ground 1mA 2600 VDC Pass ____ Hot Neutral to Secondary Ground 1mA 5656 VDC Pass ____ Hot Neutral to ECG Isolated Ground 1mA 6800 VDC Pass ____ Hot Neutral to IBP Ground 1mA 6800 VDC Pass __ Hot Neutral to Nurse Call Ground 1mA 6800 VDC Pass ___ ECG ISO Ground to Secondary Ground 1mA 4242 VDC Pass___ ECG ISO Ground to Chassis Ground 1mA 2600 VDC Pass __ Secondary Ground to Chassis Ground 1mA 707 VDC Pass ____ EPT to SpO Ground 1mA 2600 VDC Pass ____ EPT to Temp1 Temp2 Ground 1mA 2600 VDC Pass ____ Electrical Leakage Case Leakage Ground Conductor Normal polarity open ground lt 500 microamps Pass ____ Reverse polarity open ground lt 500 microamps Pass____ Normal polarity open neutral amp open ground lt 1 milliamp Pass ___ Reverse polarity open neutral amp open ground lt 1 milliamp Pass ___ ECG Mode to GND Normal 10 microamps Pass ____ ECG Mode to GND Reverse 10 microamps Pass ____ Leads all to Ground Normal polarity lt 10 microamps Pass __ Reverse polarity lt 10 microamps Pass __ Normal polarity open neutral lt 10 microamps Pass ____ Reverse polarity ope
39. operating and safety procedures refer to the Poet Plus 8100 Patient Monitor Operator s Manual Equipment Needed The following are needed to complete this procedure e Kikusui TOS 8750 Withstanding Voltage Tester or equivalent Hi Pot tester Withstanding Voltage The following test and set up is for the Kikusui TOS 8750 g g Hi Pot Test Equipment Withstanding Voltage Tester If you use an equivalent tester use the and Setup factory recommended test for that tester 1 Test the Withstanding Voltage Tester by connecting the red and black lead together Set the voltage to 1000 VAC AN 2 Set timer for 1 second and cutoff current for 1 milliamp 3 Press the Test button The NG light should come on A CAUTION e Ifthe NG light does not come on do not continue with this test You must repair or replace the withstanding voltage tester Criticare Systems Inc Poet Plus 8100 Service Manual Page 5 13 Section 5 Preventative Maintenance Withstanding Voltage 1 Hi Pot Test Procedure Connect the red lead of the Withstanding Voltage Tester to Hot Neutral and the black lead to Chassis Ground Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 2600VDC Connect the red lead of the Withstanding Voltage Tester to Hot Neutral and the black lead to Secondary Ground BNC Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 5656VDC for 1 minute Connect the red lead of the Withst
40. rights of others Criticare Systems Inc does not assume any liability arising out of any infringements of patents or other rights of third parties PROPERTY OF CRITICARE SYSTEMS INC ALL RIGHTS RESERVED DOX Digital Oximetry ComfortCuff NIBP WaterChek water trap SHELL and Multi Site SpO Sensors are trademarks of Criticare Systems Inc Page ii Poet Plus 8100 Service Manual Criticare Systems Inc Contents Copyrighta eaaa ee a dara AEE eSa daas IAE aata Siana EE i addni E ii Contents iii Warranty Service Return POlICY ssssessesssesseseesrerirerrrrnsrrerrnrresrnnrinsrentinrnnsrnstnnrnrnnrrennnnnn enea x EC Declaration of Conformity 0 00 eee cee eeee eee eeeeeeeeeeseeeceessaeeseeseseeseeatenees xi Section 1 Introduction DOSCMPUOM 2 e a i e a a AA eaa er da aaa a a a A 1 1 Intended USGsiresteiia rarua ra r a ta iA 1 1 Poet Plus 8100 Patient Monitor Options sssseseesseereesisresrneiresrnrinerrnrrnrrnsrrnnrnsrrerrnnrnne 1 1 Pulse Oximetry Measurement SpOb sssssssesssesisssrsrrssrsriesrnniiesrnrinnrnnrinrinsrrnnrnsrrennrrnnt 1 2 Definition z DOX Digital Oximetry sssesesssserreresrrerrusrureneruerenernereusensenstanrensrasrensrasrnntnnsene 1 2 Methoden ona aa a EE A a A S 1 2 SpO Clinical Testing and Accuracy 1 3 Heart Rate 1 3 ECG Measurement wee 1 4 Matho ecccccssccceverveetecatevenecsrentectecttectenecenesuinryeavecneadendehsotbendunededee
41. screws to mount the IBP to the standoffs If CO is present Add three standoffs Item 34 pn 42188B001 to the top of the IBP board and two screws pn 40995B005 to the IBP board Add three Standoffs Item 6 pn 42189C001 to the Front chassis NOTE See the assembly print for exact locations of these parts 12 IBP Attach the two Cables 2MM IDC strips Item 36 pn 83352B001 to the IBP board 13 CO CO O Install the O Module Item 13 pn 91354A001 onto the CO Module Item 26 pn 91355A001 Insert the O3 Module into the CO Board and secure the module with four 4 Hex Nuts Item 14 pn 40284B001 Refer to drawing number 91765A001 O 91766A001 CO3 or 91767A001 CO Oz or 91354A001 Oz and 91355A001 CO3 in Section 9 of this manual for the following step 14 CO Oz Add CO O Pneumatics pn 95632A001 to the bench assembly according to the assembly print 91767A001 COz Add CO Pneumatics pn 95632A002 to the bench assembly according to the assembly print 91355A001 Refer to drawing number 92788AXXxX in Drawings and Schematics in Section 9 of this manual for the following steps 15 COs CO2 Oz Mount the CO Item 29 pn 91766A001 or CO O Assembly Item 29 pn 91767A001 to the Front Chassis using six 4 Screws Item 5 pn 40995B005 Finish by installing the CO exhaust tubing assembly to the output of the pump 16 COz CO2 Oz Plug the CO Pump Assembly Item 1 into the J3 connector on
42. the battery door ee n ie Fan Figure 7 51 Remove the Power Supply Assembly Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 29 Section 7 Disassembly 3 Label and remove the ground cables to the rear assembly Label and remove the AC power cable at J2 of the Power Supply Board A DA J ps ir Figure 7 52 Remove the cable from J2 NOTE For more information about trouble shooting the power supply refer to Troubleshooting in Section 8 in this manual 4 Replace the power supply assembly 5 Reassemble the monitor Power Supply Board Do not calibrate or adjust the pot on the Power Supply Board Low voltage indicates a damage and unsafe power supply Do not reuse the old insulator when replacing power boards Order 80232C001 Power Supply Board w coating New Power Supply Boards are pre tested and ready for installation The calibration pot is factory set on new Power Supply Boards Check the connections to the power module and to the AC Power Entry Assembly pn 95641A001 for burn marks or AN discoloration Replace any assemblies that have been subject to heat damage 4 N Remove the fan from the Power Supply Board w Disconnect the cable from the AC Power Entry Assembly at J1 of the Power Supply Board 4 Remove the ground connections from the Power Supply Board Remove the DC power cable pn 90865A001 at J2 of the Power Supply Board ol 6 Remove the c
43. the tie wraps pn 40067B001 are your center point three ferrites on the backside and two on the front side but only in the space provided between the tie wrap holders NOTE Make sure all ferrites do not fall behind the front chassis assembly 18 Add a last tie wrap to other the tie wrap Pull the tie wraps tightly and secure all ferrites in place 19 Verify that the RTV is placed on the ECG Analog Board at T2 Solder Lugs to Ferrite Case 20 Attach the shielding pn 41875B004 to the ECG Analog Board pn 91342A002 by placing the four screws pn 40995B005 through four washers pn 40086B001 Attach insulator pn 83358B001 to the top of the shielding near the opening at T2 21 Plug the ECG cable assembly pn 90871A002 into the ECG Analog Board P1 and route the ferrite into the handle 22 Insert the ECG Analog Board with shielding Item 10 pn 91353A002 or 91353A001 onto the standoffs on the front chassis 23 Tighten the board down with the two standoffs Item 38 pn 42175B001 one to the bottom right and the other to the bottom left Tighten the standoffs down firmly Add a plastic standoff Item 46 pn 40909B002 to the top right hand corner 24 Add the last standoff to the upper left by placing it through the ring of the cap assembly on the ECG Analog Board 25 Solder the SpO GND cable from the SpO Board to the ECG Analog Board Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc
44. to connector PS2 on the Main Board 2 Insert an ammeter inline at PS1 on the Main Board 3 Verify the quiescent current the power is applied but the power control is in the OFF state is less than 1ma This quiescent current decreases to less than 1ma in a very short period of time This must be done at the Battery connector Remove the ammeter Connect a speaker to J20 on the Main Board Connect a membrane switch to J7 on the Main Board Attach a menu knob to the main board at J8 on the Main Board Verify that the monitor turns on after pressing the ON OFF key and that the following voltages are present Test Point Signal Value TP21 PWRUNSW 16 0V 0 5VDC TP4 5D 5 0V_ 0 1VDC TP18 5P 5 0V 0 1VDC TP1 3 3D 3 3V 0 05VDC TP17 8V 8 0V 0 25VDC TP7 5A 5 0V_ 0 25VDC T13 6 8 6 8V 0 32VDC TP9 5A 5 0V 0 3VDC TP14 5VREF 5 0V 0 025VDC TP3 5B 5 0V_ 0 25VDC Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly Before You Begin 1 Turn the monitor off 2 Disconnect the A C Mains power cord and remove the batteries WARNING e The procedures presented in this section are intended to be implemented by qualified biomedical engineering or field service personnel for PCB assembly replacement e The procedures presented in this section are not intended to be used for component level trouble shooting and repair of the PCB assemblies On
45. under the pump Mount the NIBP Module Assembly Insert the Standoffs for the CO Module Insert the Standoffs for the IBP Module Add the insulator to the Front Chassis Add the insulator to the back of the Front Chassis Add two tie wrap holders 9 Add the NIBP tubing conduit 10 Add tie wraps to hold the tubing 11 Install the IBP board with screws 12 Add additional screws if CO is not present 13 Add standoffs to the top of the IBP board 14 Attach the IBP cables to the board 15 Install the O gt Module 16 Add pneumatics for CO 05 17 Add pneumatics for CO3 18 Install bench with screws 19 Plug CO Pump into J3 on CO Board 20 Plug Main CO Cable into CO Board 21 Add cable support 22 Fold CO cable and pull it through front of chassis 23 Add tie wraps and holders 24 Insert plastic standoffs 25 Lie cable down with shielding underneath 26 Wrap cable in shielding 27 I nstall ferrite with tape 28 Plug the SpO to Analog cable into SpO board 29 Attach tubing from valve 2 to pump in 30 Attach exhaust tubing to pump out 31 Attach four screws to hold SpO board down 32 Insert SpO cable into J1 33 Add double sided tape to bushings 34 Install Insulator Bushings 35 Add shrink tubing to mounting flanges of Front Chassis 36 Add tie wrap to hold pneumatics for manifold intake ON OA AON Page 7 40 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly Final Assembly Check Off INITIALS 37 Verify th
46. 0 99 range 2 50 69 range 3 lt 50 unspecified Statistical represents one st dev 66 of clinical samples Plethysmograph Numerical Audible pulse tone pitch varies with SpO Dual wavelength LED Adult Pediatric Neonate Continuous Use 660nm 905nm lt 80mW Smart Switching ECG primary IBP Pleth NIBP 20 300 bpm ECG IBP Pleth 30 240 bpm NIBP 1 bpm or 1 Selectable On Off Selectable ECG CO or Smart Switching COs primary 6 to 150 breaths minute 1 breath minute gt of 1 or 1 breath minute Criticare Systems Inc Invasive Blood Pressure Channels Transducer Sites Pressure Range Transducer Sensitivity Resolution Accuracy Scale Range NIBP Technique Measurement Time Automatic Measurement Cycles Inflation Pressure Range Resolution Transducer Accuracy STAT mode Capnometry CO3 Units Display Method Calibration Waveform Scale Range Resolution Accuracy N20 Compensation Flow rate System Response Time Rise Time Delay Time Time from cold start Pneumatic Sound Pressure Criticare Systems Inc Poet Plus 8100 Service Manual Section 1 Introduction 2 Selectable ART PA LA RA CVP ICP 20mmHg to 300mmHg 5yuV V mmHg 1 mmHg 1 mmHg or 1 whichever is greater 10 to 10 0 to 20 0 to 30 0 to 40 0 to 60 0 to 100 0 to 150 0 to 200 0 to 300 and Auto Ranging Oscillometric measure
47. 0 NIBP PUMP 05 40995B005 P H M S 4 40X 25 SEMS 07 83311B001 INSULATOR FRONT CHASSIS 08 42211B001 INSULATOR CHASSIS BACK 09 40294B001 HOLDER TIE WRAP ADH BK SM 10 41602B001 TUBING POLYU 188 X 3125 11 40067B001 CABLE TIE 4 0 LG NYLON 16 40909B001 SPACER HEX 6 32X 5 NY M F 17 90866A001 SPO2 TO ANALOG BOARD CBL 18 82515B001 SPLIT FLAT CBL SUPP CORE 19 41955B001 FOAM TAPE 1 0 X 2 12 20 95624A001 AY CBL SPO2 TO DB9 24 42212B002 CABLE SUPPORT CO2 MAIN 28 91339A001 AY PCB 8100 SPO2 BRD 30 42228B002 SHIELD SPO2 ANALOG CBL 31 42230B002 INSULATING BUSHING 32 40041B009 TUBING SHRINK 375 CLR 33 42187C001 STANDOFF MODIFIED 35 91350A002 AY IBP BRD 36 83352B001 CBL 2MM IDC STRIPS 37 40683B001 TAPE TRANSFER ADH 5MIL 38 40067B002 CABLE TIE 5 5 LG NYLON Page 9 6 Poet Plus 8100 Service Manual Criticare Systems Inc Section 9 Drawings and Schematics Front Chassis CO O 92788A011 CHASIS FRNT 8100 CO2 02 Criticare Systems Inc Item 01 02 03 04 CSI Part Description 94253A001 BTC PUMP MOTOR AY 42131B001 CHASSIS FRONT 40496B002 P H M S 6 32X 250 SEMS 94252A001 AY 8100 NIBP PUMP 40995B005 P H M S 4 40X 25 SEMS 421890001 STNDOF M 4 40 X 1 312 HEX 83311B001 INSULATOR FRONT CHASSIS 42211B001 INSULATOR CHASSIS BACK 40294B001 HOLDER TIE WRAP ADH BK SM 41602B001 TUBING POLYU 188 X 3125 40067B001 CABLE TIE 4 0 LG NYLON 87225B001 CBL 2MM PITCH RIBBON 40909B001 SPACER HEX 6 32X 5 NY M F 90866A001 SPO2 TO ANALOG BOARD CBL 82515B001
48. 0601 2 27 ECG Safety Standard All specifications are subject to change without notice Criticare Systems Inc Poet Plus 8100 Service Manual Page 1 15 Section 1 Introduction Symbols Symbol Definition N Refer to Operator s Manual for Information C 0413 European Community Mark ly us Underwriters Laboratories Mark Not For Use with Flammable Anesthetic Gasses Do not dispose of in municipal waste Wheeled bin symbol indicates separate collection for electrical and electronic equipment WEEE Directive 2002 96 EEC Type CF Equipment defib proof AN Shock Hazard TEF ike PEN Alternating Current AC E Technical Support Phone Number Definitions for Warning and Caution symbols AN DANGER Designates an imminent danger Non observance is likely to lead to death or the most severe injuries Designates a possible dangerous situation WARNING Non observance may lead to death or the most severe injuries Designates a possible dangerous situation CAUTION A Non observance may lead to minor injuries or damage to the product Page 1 16 Poet Plus 8100 Service Manual Criticare Systems Inc Safety Leakage Current Voltage Fluctuations Criticare Systems Inc Section 1 Introduction WARNING Read this manual entirely before attempting clinical use of the monitor A possible explosion hazard exists Do not use the monitor in the presence of flammable anesthetics Cables cor
49. 1 42230B002 INSULATING BUSHING 32 40041B009 TUBING SHRINK 375 CLR 37 40683B001 TAPE TRANSFER ADH 5MIL 38 40067B002 CABLE TIE 5 5 LG NYLON Page 9 4 Poet Plus 8100 Service Manual Criticare Systems Inc Section 9 Drawings and Schematics Front Chassis CO 92788A003 CHASSIS FRNT 8100 CO2 Criticare Systems Inc Item 01 02 03 04 CSI Part Description 94253A001 BTC PUMP MOTOR AY 42131B001 CHASSIS FRONT 40496B002 P H M S 6 32X 250 SEMS 94252A001 AY 8100 NIBP PUMP 40995B005 P H M S 4 40X 25 SEMS 421890001 STNDOF M 4 40 X 1 312 HEX 83311B001 INSULATOR FRONT CHASSIS 42211B001 INSULATOR CHASSIS BACK 40294B001 HOLDER TIE WRAP ADH BK SM 41602B001 TUBING POLYU 188 X 3125 40067B001 CABLE TIE 4 0 LG NYLON 87225B001 CBL 2MM PITCH RIBBON 40909B001 SPACER HEX 6 32X 5 NY M F 90866A001 SPO2 TO ANALOG BOARD CBL 82515B001 SPLIT FLAT CBL SUPP CORE 41955B001 FOAM TAPE 1 0 X 2 12 95624A001 AY CBL SPO2 TO DB9 42212B002 CABLE SUPPORT CO2 MAIN 91339A001 AY PCB 8100 SPO2 BRD 91766A001 AY C02 8100 42228B002 SHIELD SPO2 ANALOG CBL 42230B002 INSULATING BUSHING 40041B009 TUBING SHRINK 375 CLR 40683B001 TAPE TRANSFER ADH 5MIL 40067B002 CABLE TIE 5 5 LG NYLON 40041B001 TUBING SHRINK 250 CLR Poet Plus 8100 Service Manual Page 9 5 Section 9 Drawings and Schematics Front Chassis IBP 92788A005 CHASSIS FRNT 8100 IBP Item CSI Part Description 02 42131B001 CHASSIS FRONT 03 40496B002 P H M S 6 32X 250 SEMS 04 94252A001 AY 810
50. 2 Add 1 inch of Shrink Tubing Item 32 pn 40041B009 to the mounting flanges of the Front Chassis per the assembly drawing 33 Add a Tie wrap Item 11 pn 40067B001 through the tie wrap holder on the bottom of the pump and secure the pneumatics for the manifold intake Final Assembly Rebuild Refer to drawing number 93951AXXxX in Drawings and Schematics in Section 9 of this manual for the following steps 1 Attach the power supply ground cable from the multi ground cable pn 90883A001 to the stud on the power module Item 7 pn 95612A001 using a keps nut Item 26 pn 40284B002 2 Attach the ground power supply cable from multi ground cable to the quick disconnect on the power module circuit board pn 80232C001 3 Attach wire harness from Isolation PCB to the input of power supply 4 Attach shielding with the last strip of tape All three cables should be routed around the shielding 5 Install the Power Module Assembly Item 7 pn 95612A001 into the rear housing Item 1 pn 93258A001 over the battery door hinge 6 Insert the four painted screws pn 42065B004 into the bottom of the rear chassis assembly pn 93258A001 to secure the power supply module 7 Before you tighten all the screws on the side by the battery door close battery door and then tighten the screws This allows the battery door to function properly 8 Route the white thermister black red fan blood pressure GND cable and speaker ca
51. 2225B001 SUPPORT FRONT PLASTIC 22 42227B001 RICHCO WIRE CLIP 25 40284B001 NUT 4 40 KEPS PL 26 40284B002 NUT 6 32 KEPS PL 27 42065B004 F H M S 4 40 X 375 PH PL 28 40995B001 P H M S 4 40X 312 SEMS 29 40995B006 P H M S 4 40X 38 SEMS 33 40496B001 P H M S 6 32X 375 SEMS 35 40876B001 P H M S 8 32X 25 PL SEMS 36 40873B004 P H M S 8 32X 750 PL PH 38 42175B001 STANDOFF 632X 625 25HEX 40 83278B001 BATTERY 8100 44 46220B001 LBL SERIAL 8100 45 42220B001 LBLS LASER 5167 SHEET 46 40909B002 SPCR HEX 6 32X 625 NY M F 47 40496B006 P H M S 6 32X 43 SEMS 48 42132B001 ACCESS DOOR 02 SENSOR 50 42014B001 FITNG QUICK CON HH NIBP 51 40284B005 NUT 10 32 KEPS PL 52 83375B001 COVER CAN CONNECTOR 8100 53 40995B005 P H M S 4 40X 25 SEMS 60 40873B001 P H M S 8 32X 375 PL PH Page 9 14 Poet Plus 8100 Service Manual Criticare Systems Inc Section 9 Drawings and Schematics Final Assembly 93951A398 8100 FNL CO2 02 2T 6WAVE CO Temp6Wave aCe Be Item CSI Part 01 93258A001 03 93259A006 07 95612A001 09 91340A003 10 91353A002 10 92788A011 12 95618A006 15 90871A002 17 45114B001 19 41891B001 21 42225B001 22 42227B001 25 40284B001 26 40284B002 27 42065B004 28 40995B001 29 40995B006 33 40496B001 35 40876B001 36 40873B004 38 42175B001 40 83278B001 44 46220B001 45 42220B001 46 40909B002 47 40496B006 48 42132B001 49 42134B003 50 42014B001 51 40284B005 52 83375B001 53 40995B005 57 40067B003 60 40873B001 Description AY HOUS
52. 3 SEMS ACCESS DOOR 02 SENSOR FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS CABLE TIE 8 LG NYLON P H M S 8 32X 375 PL PH Criticare Systems Inc Final Assembly PrinteyCO O IBP 2Temp Criticare Systems Inc Section 9 Drawings and Schematics 93951A708 8100FNL 8100 PRNTR CO2 O2 IBP 2T Item 01 03 05 07 09 10 12 15 18 19 21 22 25 26 27 CSI Part 93258A001 93259A006 92788A025 95612A001 91340A003 91353A002 95618A004 90871A002 95604A002 41891B001 42225B001 42227B001 40284B001 40284B002 42065B004 40995B001 40995B006 40496B001 40876B001 40873B004 42175B001 83278B001 46220B001 42220B001 40909B002 40496B006 42132B001 42014B001 40284B005 83375B001 40995B005 40067B003 40873B001 Poet Plus 8100 Service Manual Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS 8100 CO2 02 IBP AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CONN AY 2 IBP 2 TEMP NIBP AY CBL ECG ANALOG TO OUT PRINTER MODULE 8100 9100 DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 43 SEMS ACCESS DOOR O2 SENSOR FITNG QU
53. 36 40873B004 38 42175B001 40 83278B001 44 46220B001 45 42220B001 46 40909B002 47 40496B006 48 42132B001 49 42134B003 50 42014B001 51 40284B005 52 83375B001 53 40995B005 60 40873B001 Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS FRONT 8100 BASE AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CONN AY 2 TEMP AY CBL ECG ANALOG TO OUT COVER PRINTER OPENING DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 43 SEMS ACCESS DOOR O02 SENSOR CONN MNT PLT 2TEMP FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS P H M S 8 32X 375 PL PH Page 9 12 Poet Plus 8100 Service Manual Criticare Systems Inc Final Assembly CO 2Temp 6Wave Item 01 03 05 07 09 10 12 15 17 19 21 22 25 26 27 Criticare Systems Inc CSI Part 93258A001 93259A006 92788A003 95612A001 91340A003 91353A002 95618A006 90871A002 45114B001 41891B001 42225B001 42227B001 40284B001 40284B002 42065B004 40995B001 40995B006 40496B001 40876B001 40873B004 42175B001 83278B001 46220B001 42220B001 40909B002 40496B006 42132B001
54. 4 pn 94252A001 to the Front Chassis using two 4 Screws Item 5 pn 40995B005 CO COO Insert the six M F Standoffs Item 6 pn 42189C001 into the PEM nuts on the Front Chassis CO and IBP If IBP and CO2 are being built insert 5 Standoffs Item 33 pn 42187C001 into the PEM nuts on the Front Chassis Add Insulator Front Chassis Item 7 pn 83311B001 to the front surface of the Front Chassis Add Insulator Chassis Back Item 8 pn 42211B001 to the back surface of the Front Chassis Place one Tie Wrap Holder Item 9 pn 40294B001 below the manifold of NIBP Add a second Tie Wrap Holder approximately 4 inches to the right of the first one placed on an angle NOTE For BP units only the first tie wrap holder is added Take the Clear Tubing Item 10 pn 41602B001 and cut to approximately 4 inches long Slide over the thin piece of tubing coming from the NIBP manifold Add two Tie Wraps Item 11 pn 40067B001 to the Tie Wrap Holders on the back surface of the Front Chassis to retain NIBP Tubing Conduit NOTE For BP units just add the second tie wrap to hold the tubing Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly 10 Install the IBP board Item 35 pn 91350A002 to the top of the standoffs and attach with the two screws pn 40995B005 NOTE See the assembly print for details on the IBP 11 f no CO is present Add five
55. 42134B003 42014B001 40284B005 83375B001 40995B005 40067B003 40873B001 Poet Plus 8100 Service Manual Section 9 Drawings and Schematics 93951A388 8100 FNL CO2 2T 6WAVE Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS FRNT 8100 CO2 AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CONN AY 2 TEMP AY CBL ECG ANALOG TO OUT COVER PRINTER OPENING DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 43 SEMS ACCESS DOOR O2 SENSOR CONN MNT PLT 2TEMP FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS CABLE TIE 8 LG NYLON P H M S 8 32X 375 PL PH Page 9 13 Section 9 Drawings and Schematics Final Assembly 93951A392 8100 AY IBP 2T 6WAVE IBP2Temp6Wave Item CSI Part Description 01 93258A001 AY HOUSING REAR 8100 03 93259A005 HOUSING FRONT AY 8100 ACT 05 92788A005 CHASSIS FRNT 8100 IBP 07 95612A001 AY POWER MODULE 09 91340A003 AY PCB 8100 MAIN BRD 10 91353A002 AY SHIELDED ANALOG BRD 12 95618A004 CONN AY 2 IBP 2 TEMP NIBP 15 9087 1A002 AY CBL ECG ANALOG TO OUT 17 45114B001 COVER PRINTER OPENING 19 41891B001 DB9 COVER PLATE 21 4
56. 5632A001 ltem Tubing Assembly CO 95632A002 ltem Section 9 Drawings and Schematics AY 8100 TUBING SET W O2 CSI Part 46238B001 95625A001 95626A001 95627A001 95628A001 95629A001 95630A001 95631A001 95633A001 Description LABEL SET 8100 PNEU AY 8100 PAT INPUT TUBING AY 8100 ABSORBER TUBING AY 8100 PRESSURE TUBING AY 8100 BLEED CHECK TUBNG AY 8100 02 gt VALVE TUBING AY 8100 VLV2 gt PUMP TUBING AY 8100 VLV3 gt VLV2 TUBING AY 8100 PUMP EXHAUST AY 8100 TUBE SET NO 02 CSI Part 46238B001 95625A001 95626A001 95627A002 95628A001 95630A001 95631A001 95633A001 Description LABEL SET 8100 PNEU AY 8100 PAT INPUT TUBING AY 8100 ABSORBER TUBING AY 8100 PRESSURE TUBING AY 8100 BLEED CHECK TUBNG AY 8100 VLV2 gt PUMP TUBING AY 8100 VLV3 gt VLV2 TUBING AY 8100 PUMP EXHAUST The parts list for the assemblies listed above are found on the following pages Poet Plus 8100 Service Manual Page 9 29 Section 9 Drawings and Schematics PAT Input Tubing Absorber Tubing Pressure Tubing Bleed Check Tubing O Valve Tubing Only with pn 95632A001 Valve3 Pump Tubing Page 9 30 95625A001 AY 8100 PAT INPUT TUBING Item CSI Part Description 01 41700B001 TUBING POLYU 094IDX 1870 02 42063B003 TUBE PVC 1 8 X 1 16 95 DU 03 41585B002 TUBE CONECTOR 3 32 X 1 16 95626A001 AY 8100 ABSORBER TUBING Item CSI Part Description 01 41700B003 TUBE POLYU 125X 250 70DU 02 40570B003 A
57. 7B001 11 42170B001 12 42173B001 15 80124B002 16 80130B001 18 41158B001 20 95257A001 21 95645A001 24 83302B001 26 95641A001 27 42206B004 29 90883A001 31 42205B001 32 83340B001 34 83339B002 35 42210B004 36 42223B001 37 42223B002 40 40995B005 41 40123B003 42 40995B003 43 42065B002 44 42064B004 45 40878B001 48 40284B001 49 41164B001 52 40086B003 53 40007B002 54 42222B007 56 83111B002 57 41955B001 59 80131B002 Poet Plus 8100 Service Manual Description HOUSING REAR BATTERY DOOR HINGE BATTERY DOOR BATTERY DOOR PIN BED RAIL HOOK BRKT BED RAIL HOOK STANDOFF 188X 141 875L PWR RECEPT FUSE MODULE PWR RECPT FUSE DRAWER TERMINAL EQUIPOTENTIAL AY SPEAKER 8100 AY INSUL SPKER RETAINER 4 ROUND RUBBER FOOT AY POWER ENTRY INSULATOR PWR ENTRY BRD AY CBL MULTIPLE GROUND INSULATOR BATT TERMINAL DRIP SHIELD FRONT SNAPS DRIP GUARD PRINTER DECK INSULATOR BENCH AC INLET FOAM GASKT LOW O2 SEN DOR FOAM GASKT SIDEO2 SEN DOR P H M S 4 40X 25 SEMS P H M S 4 40X 375 NY WHT P H M S 4 40X 500 SEMS F H M S 4 40 X 250 PH PNT F H M S 4 40X 375 PH PL H H M S 6 32X1 25 PL NUT 4 40 KEPS PL NUT M6 X 1 0 KEPS PL WASHER FLAT PL 156X 375 WASHER LOCK 6 EXT TOOTH 1 4 INTRNL TOOTH LCK WSHR 8100 PRINTER CBL FOAM TAPE 1 0 X 2 12 FUSE 250V 1 AMP SB Criticare Systems Inc Front Housing Assemblies Section 9 Drawings and Schematics Front Housing 93259A005 HOUSING FRONT AY 8100 ACT Assembly without CO Item
58. ANSFER ADH 5MIL 38 40067B002 CABLE TIE 5 5 LG NYLON 39 40041B001 TUBING SHRINK 250 CLR Page 9 8 Poet Plus 8100 Service Manual Criticare Systems Inc Section 9 Drawings and Schematics Front Chassis CO O 92788A025 CHASSIS 8100 CO2 02 IBP Criticare Systems Inc IBP Item 01 02 03 04 CSI Part Description 94253A001 BTC PUMP MOTOR AY 42131B001 CHASSIS FRONT 40496B002 P H M S 6 32X 250 SEMS 94252A001 AY 8100 NIBP PUMP 40995B005 P H M S 4 40X 25 SEMS 42189C001 STNDOF M 4 40 X 1 312 HEX 83311B001 INSULATOR FRONT CHASSIS 42211B001 INSULATOR CHASSIS BACK 40294B001 HOLDER TIE WRAP ADH BK SM 41602B001 TUBING POLYU 188 X 3125 40067B001 CABLE TIE 4 0 LG NYLON 87225B001 CBL 2MM PITCH RIBBON 40909B001 SPACER HEX 6 32X 5 NY M F 90866A001 SPO2 TO ANALOG BOARD CBL 82515B001 SPLIT FLAT CBL SUPP CORE 41955B001 FOAM TAPE 1 0 X 2 12 95624A001 AY CBL SPO2 TO DB9 42212B002 CABLE SUPPORT CO2 MAIN 91339A001 AY PCB 8100 SPO2 BRD 91767A001 AY 02 C02 8100 42228B002 SHIELD SPO2 ANALOG CBL 42230B002 INSULATING BUSHING 40041B009 TUBING SHRINK 375 CLR 42187C001 STANDOFF MODIFIED 42188B001 STANDOFF 25 4 40 X 812LG 91350A002 AY IBP BRD 83352B001 CBL 2MM IDC STRIPS 40067B002 CABLE TIE 5 5 LG NYLON Poet Plus 8100 Service Manual Page 9 9 Section 9 Drawings and Schematics Rear Housing Assembly Page 9 10 93258A001 AY HOUSING REAR 8100 Item CSI Part 01 45112B001 03 42129B001 04 42130B001 05 42144B001 10 8326
59. B BATTERY CONTACT PWR SPLY CONFORMAL COATNG BULKHEAD BATT COMPARTMENT CHASSIS REAR BATTERY CAGE DBL BATTERY LATCH BATTERY EJECTOR EXTENSN SPRING BATT EJECT AY FAN 8100 W CONNECTORS FAN MOUNT POWER SUPPLY TO MAIN BRD THERMISTER TO MAIN BRD CLIP RICHCO P H M S 4 40X 25 SEMS P H M S 4 40X 38 SEMS P H M S 4 40 X 56 SEMS NUT 4 40 KEPS PL CABLE TIE 4 0 LG NYLON INSULATOR PWR SUPPLY BATTERY SLIDE 8100 POWER HOLDER TIE WRAP ADH BK SM LATCH SUPPORT Poet Plus 8100 Service Manual Page 9 27 Section 9 Drawings and Schematics Connector Assemblies Connector Assembly TempNIBP Connector Assembly 2IBP2Tem NIBP Connector Assembly 2Temp Page 9 28 95618A001 95618A004 95618A006 ltem 01 02 Poet Plus 8100 Service Manual CONN AY TEMP NIBP CSI Part Description 42133B001 CONN MNT PLATE TEMP NIBP 90872A001 AY CBL TEMP CONNECTOR 40119B003 WASHER LOCK 3 8 INT TOOTH 42237B001 NUT NYLON JAM 3 8 X 32 42014B001 FITNG QUICK CON HH NIBP 40284B005 NUT 10 32 KEPS PL 42236B001 WASHER MYLAR 3 8 015THK 40683B001 TAPE TRANSFER ADH 5MIL CONN AY 2 IBP 2 TEMP NIBP CSI Part Description 91351A003 IH AY CONNECT BRD 2IBP 2TEMP 90872A003 AY CBL 2TEMP CONNECTOR 42134B001 CONN MNT PLT 2 TEMP NIBP CONN AY 2TEMP CSI Part Description 91351A005 AY CONNECT BRD 2TEMP 90872A003 AY CBL 2TEMP CONNECTOR Criticare Systems Inc Criticare Systems Inc Tubing Assemblies Tubing Assembly O 9
60. BP 2Temp 6Wave Final Assembly CO O 2Temp 6Wave Final Assembly CO IBP 2Temp 6Wave Final Assembly CO O IBP 2Temp 6Wave Final Assembly Printer 2Temp 6Wave Final Assembly CO 2Temp 6Wave Final Assembly Printer IBP 2Temp 6Wave Final Assembly Printer CO O 2Temp 6Wave Final Assembly Printer CO2 IBP 2Temp 6Wave Final Assembly Printer CO O IBP 2Temp NIBP Pump Assembly Printer Module Display Assembly Power Module Assembly Connector Assembly Temp NIBP Connector Assembly 2IBP 2Temp Connector Assembly 2IBP 1Temp NIBP Connector Assembly 2Temp Poet Plus 8100 Service Manual Drawing Number 91354A001 91355A001 91765A001 91766A001 91767A001 92788A001 92788A003 92788A005 92788A011 92788A013 92788A025 93258A001 93259A005 93259A006 93951A386 93951A388 93951A392 93951A398 93951A402 93951A420 93951A674 93951A676 93951A680 93951A686 93951A690 93951A708 94252A001 95604A002 95605A003 95612A001 95618A001 95618A004 95618A005 95618A006 Page 9 1 Section 9 Drawings and Schematics Pneumatic Assemblies Title Page 9 2 PCB Drawings and Schematics Only on units with the CO or the CO O options Tubing Assembly O Tubing Assembly CO gt Drawing Number 95632A001 95632A002 The above tubing assemblies contain the following subassemblies PAT Input Tubing Absorber Tubing Pressure Tubing Bleed Check Tubing O Valve Tubing Only with pn 95632A001 Valve3 Pump
61. BSORBER IBENCH 03 41585B004 TUBE CONECTOR 1 8 X 1 16 04 42227B001 RICHCO WIRE CLIP 95627A001 AY 8100 PRESSURE TUBING Item CSI Part Description 02 40324B004 TEE TUBING 063X 063X 093 03 41570B001 ELBOW 063ID TUBING 04 42063B003 TUBE PVC 1 8 X 1 16 95 DU 05 40296B001 TUBING SILC BLU 040X 140 95628A001 AY 8100 BLEED CHECK TUBNG Item CSI Part Description 01 40214B002 TYGON TUBING 02IDX 060D 02 41700B003 TUBE POLYU 125X 250 70DU 03 40296B001 TUBING SILC BLU 040X 140 04 40324B008 T CONNECTOR WHT NYLON 06 41571B002 L CONNECTOR WHT NYLON 07 40318B002 VALVE CHECK FLUID VL1300 08 41840B002 ORIFICE RESTRICTOR 09 42081B004 ORIFICE RESTRICTOR 002 10 42084B001 FILTER 5 MICRON 1 16 BARB 11 41700B001 TUBING POLYU 094IDX 1870 95629A001 AY 8100 O2 gt VALVE TUBING Item CSI Part Description 01 41700B003 TUBE POLYU 125X 250 70DU 02 41569B001 ELBOW 094 063 TUBING 03 42063B003 TUBE PVC 1 8 X 1 16 95 DU 95630A001 AY 8100 VLV2 gt PUMP TUBING Item CSI Part Description 01 40296B002 TUBING SILC BLU 125X 250 Poet Plus 8100 Service Manual Criticare Systems Inc Section 9 Drawings and Schematics Valve3 Valve2 Tubing 95631A001 IHAY 8100 TUBING SET W 02 Item CSI Part Description 01 41700B003 TUBE POLYU 125X 250 70DU 02 41571B002 L CONNECTOR WHT NYLON 03 41700B001 TUBING POLYU 094IDX 1870 Pump Exhaust 95633A001 AY 8100 PUMP EXHAUST Item CSI Part Description 01 41585B003 TUBE CONECTOR 3 32 X 1 8 02 40296B002 T
62. C mains power sources User selectable filters are also available The accuracy of the ECG analog output bandwidth is equal to the frequency response specified in the ECG specifications The variable gain control is x200 x400 or x800 according to the ECG Sensitivity setting The propagation delay is 1000 milliseconds The accuracy of the synchronizing pulse amplitude is equal to 500 times that of leadview II The pulse shape and duration match those of leadview II The output impedance is 1000 ohms and propagation delay is less than 6 6 milliseconds With the pacemaker detector turned ON in the 8100 the system detects and rejects pacemaker pulses ranging from 2 to 700 mV amplitude and 0 1 to 2 0 ms duration Heart rates properly display over this range of pacemaker operation Pacer pulse markers are present if pace detect is on and appear in the ECG analog output as narrow positive spikes at the point of pace detection CAUTION e Ambient noise sources may induce artifactual triggers of the pacemaker pulse detector and display Poet Plus 8100 Service Manual Criticare Systems Inc Respiration Non Invasive Blood Pressure NIBP Comfort Cuff Technology Description of NIBP Measurement Criticare Systems Inc Section 1 Introduction Respiration is measured via the ECG electrodes or by using the respiration cycle data from the optional CO monitoring module To determine respiration from the ECG the pa
63. C cable with 0 1uf CAP signal to ground from the ECG BNC on the monitor to an input channel on an oscilloscope Select and press the PARAMS softkey Set Filter to Diagnostic and ECG Sensitivity to Medium Select and press the DISPLAY softkey Set Waveform 1 for ECG Lead II ECG IN on the screen Set the MedSim 300 patient simulator for Sq 2HZ N O oO fF WO PD Verify that the baseline is zero and signal is a peak to peak voltage of 1 2V 0 2 Page 5 24 Poet Plus 8100 Service Manual Criticare Systems Inc Diagnostic Monitor and ST Filter Modes Test Criticare Systems Inc 1 2 Section 5 Preventative Maintenance Test the DIAGNOSTIC MONITOR and ESU FILTER modes In the PARAMS window set Filter to Diagnostic On the simulator select PERF under ECG then WAVE and then s40 40 Hz sine wave Verify that the sine wave signal displayed has amplitude no less than 0 95 X the input amplitude in mV when the simulator is set to s10 10 Hz sine wave To accomplish this measure the peak to peak amplitude of the 10 Hz sine wave in waveform Channel 2 then measure the peak to peak amplitude of the 40 Hz sine wave There shall be no more than a 5 decrease 40 Hz amplitude or gt 0 95 X 10 Hz amplitude Select Monitor filtering The 40 Hz sine wave shall be 0 70 X the input amplitude in mV when the simulator is set to s10 10 Hz sine wave Measure as described in step 2 Select ST Filterin
64. C001X00ECX No CO Yes CAT81C005X00ECX No CO O Yes CAT81C024X00ECX No IBP Yes CAT81C025X00ECX No IBP CO Yes CAT81C028X00ECX No IBP CO O Yes CAT81COOOPOOECX Yes Standard Yes CAT81C001POOECX Yes CO Yes CAT81C005P00ECX Yes CO O Yes CAT81C024P00ECX Yes IBP Yes CAT81C025P00ECX Yes IBP CO Yes CAT81CO28PO0ECX Yes IBP CO O Yes The units shown above are listed as E in the Cat No and have an English language label set and instructions Contact customer service for additional language options The instructions in this manual cover the operation of each of the option packages listed above For those models that do not include a particular monitoring module i e CO or O3 the monitor functions as if that module had been turned off in the monitor Some Poet Plus 8100 patient monitors may not have an external module connection and may require updates to meet the specifications described in this manual Poet Plus 8100 Service Manual Page 1 1 Section 1 Introduction Pulse Oximetry Measurement SpO DOX Digital Oximetry Page 1 2 Definition Method The patient monitor uses Digital Oximetry DOX technology to measure blood oxygen saturation SpO Hemoglobin exists in the blood in several forms e Oxygenated Oxyhemoglobin e Reduced Deoxyhemoglobin e Dyshemoglobins carboxyhemoglobin and methemoglobin In the patient monitor SpO pulse arterial oxygen saturation is the ratio of oxygenated
65. CH 10 41188C201 CO2 DETECTOR 11 83304B001 GASKET DETECTOR IR SOURCE 91765A001 8100 O2 MANIFOLD AY Item CSI Part Description 01 10023B003 RESISTOR 30 OHM 1W 5 02 40298B001 ADHESIVE STYCAST CATALYST 03 40995B005 P H M S 4 40X 25 SEMS 04 42062B003 TUBING SSTL 0830D X 75L 05 42146B001 MANIFOLD 02 SENSOR 06 80059B001 THERMISTOR 10K 25C 07 91332A001 8100 02 BOARD 91766A001 AY C02 8100 Item CSI Part Description 01 91355A001 AY 8100 CO2 MODULE 02 95632A002 AY 8100 TUBE SET NO 02 91767A001 AY O2 CO2 8100 Item CSI Part Description 01 91355A001 AY 8100 CO2 MODULE 02 91354A001 AY O2 MODULE 8100 03 95632A001 AY 8100 TUBING SET W 02 Poet Plus 8100 Service Manual Page 9 3 Section 9 Drawings and Schematics Front Chassis Assemblies Front Chassis Base 92788A001 CHASIS FRNT 8100 BASE Item CSI Part Description 02 42131B001 CHASSIS FRONT 03 40496B002 P H M S 6 32X 250 SEMS 04 94252A001 AY 8100 NIBP PUMP 05 40995B005 P H M S 4 40X 25 SEMS 07 83311B001 INSULATOR FRONT CHASSIS 08 42211B001 INSULATOR CHASSIS BACK 09 40294B001 HOLDER TIE WRAP ADH BK SM 10 41602B001 TUBING POLYU 188 X 3125 11 40067B001 CABLE TIE 4 0 LG NYLON 16 40909B001 SPACER HEX 6 32X 5 NY M F 17 90866A001 SPO2 TO ANALOG BOARD CBL 18 82515B001 SPLIT FLAT CBL SUPP CORE 19 41955B001 FOAM TAPE 1 0 X 2 12 20 95624A001 AY CBL SPO2 TO DB9 24 42212B002 CABLE SUPPORT CO2 MAIN 28 91339A001 AY PCB 8100 SPO2 BRD 30 42228B002 SHIELD SPO2 ANALOG CBL 3
66. FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 43 SEMS ACCESS DOOR O2 SENSOR CONN MNT PLT 2TEMP FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS CABLE TIE 8 LG NYLON P H M S 8 32X 375 PL PH Page 9 19 Section 9 Drawings and Schematics Final Assembly 93951A680 8100 PRNTR IBP 2T 6WAVE PrinterABP2Temp 6Wave Item CSI Part Description 01 93258A001 AY HOUSING REAR 8100 03 93259A005 HOUSING FRONT AY 8100 ACT 05 92788A005 CHASSIS FRNT 8100 IBP 07 95612A001 AY POWER MODULE 09 91340A003 AY PCB 8100 MAIN BRD 10 91353A002 AY SHIELDED ANALOG BRD 12 95618A004 CONN AY 2 IBP 2 TEMP NIBP 15 90871A002 AY CBL ECG ANALOG TO OUT 18 95604A002 PRINTER MODULE 8100 9100 19 41891B001 DB9 COVER PLATE 21 42225B001 SUPPORT FRONT PLASTIC 22 42227B001 RICHCO WIRE CLIP 25 40284B001 NUT 4 40 KEPS PL 26 40284B002 NUT 6 32 KEPS PL 27 42065B004 F H M S 4 40 X 375 PH PL 28 40995B001 P H M S 4 40X 312 SEMS 29 40995B006 P H M S 4 40X 38 SEMS 33 40496B001 P H M S 6 32X 375 SEMS 35 40876B001 P H M S 8 32X 25 PL SEMS 36 40873B004 P H M S 8 32X 750 PL PH 38 42175B001 STANDOFF 632X 625 25HEX 40 83278B001 BATTERY
67. G BRD CONN AY 2 IBP 2 TEMP NIBP AY CBL ECG ANALOG TO OUT COVER PRINTER OPENING DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 43 SEMS ACCESS DOOR O2 SENSOR FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS CABLE TIE 8 LG NYLON P H M S 8 32X 375 PL PH Page 9 17 Section 9 Drawings and Schematics Final Assembly Printer2Temp 6Wave Page 9 18 93951A674 8100 FNL PRNTR 2T 6WAVE Item CSI Part 01 93258A001 03 93259A005 05 92788A001 07 95612A001 09 91340A003 10 91353A002 12 95618A006 15 90871A002 18 95604A002 19 41891B001 21 42225B001 22 42227B001 25 40284B001 26 40284B002 27 42065B004 28 40995B001 29 40995B006 33 40496B001 35 40876B001 36 40873B004 38 42175B001 40 83278B001 44 46220B001 45 42220B001 46 40909B002 47 40496B006 48 42132B001 49 42134B003 50 42014B001 51 40284B005 52 83375B001 53 40995B005 60 40873B001 Poet Plus 8100 Service Manual Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS FRONT 8100 BASE AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CONN AY 2 TEMP AY CBL ECG ANALOG TO OUT
68. H M S 4 40 X 375 PH PL 28 40995B001 P H M S 4 40X 312 SEMS 29 40995B006 P H M S 4 40X 38 SEMS 35 40876B001 P H M S 8 32X 25 PL SEMS 36 40873B004 P H M S 8 32X 750 PL PH 38 42175B001 STANDOFF 632X 625 25HEX 40 83278B001 BATTERY 8100 44 46220B001 LBL SERIAL 8100 45 42220B001 LBLS LASER 5167 SHEET 46 40909B002 SPCR HEX 6 32X 625 NY M F 47 40496B006 P H M S 6 32X 43 SEMS 48 42132B001 ACCESS DOOR 02 SENSOR 50 42014B001 FITNG QUICK CON HH NIBP 51 40284B005 NUT 10 32 KEPS PL 52 83375B001 COVER CAN CONNECTOR 8100 53 40995B005 P H M S 4 40X 25 SEMS 57 40067B003 CABLE TIE 8 LG NYLON 60 40873B001 P H M S 8 32X 375 PL PH Page 9 16 Poet Plus 8100 Service Manual Criticare Systems Inc Final Assembly CO vOBP 2Temp 6Wave Criticare Systems Inc Section 9 Drawings and Schematics 93951A420 8100 CO2 O02 IBP 2T 6 WAVE Item 01 03 05 07 09 10 12 15 17 19 21 22 25 26 27 CSI Part 93258A001 93259A006 92788A025 95612A001 91340A003 91353A002 95618A004 90871A002 45114B001 41891B001 42225B001 42227B001 40284B001 40284B002 42065B004 40995B001 40996B006 40496B001 40876B001 40873B004 42175B001 83278B001 46220B001 42220B001 40909B002 40496B006 42132B001 42014B001 40284B005 83375B001 40995B005 40067B003 40873B001 Poet Plus 8100 Service Manual Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS 8100 CO2 02 IBP AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALO
69. ICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS CABLE TIE 8 LG NYLON P H M S 8 32X 375 PL PH Page 9 23 Section 9 Drawings and Schematics NIBP Pump Assembly 94252A001 AY 8100 NIBP PUMP Item CSI Part Description 01 95576A003 AY PUMP 8100 W WIRES 03 95596A002 AY OKEN VALVE W CRIMPS 05 95595A001 AY VP NIBP MANIFOLD 08 87151B006 HOUS SOCKET 098 LOCKING 10 42104B001 BRACKET OEM NIBP OS VALVE 11 42111B001 SCREW M2 448 PH BINDER 13 40067B002 CABLE TIE 5 5 LG NYLON 15 40132B001 TAPE MICROFOAM ADH BACKED 18 40296B001 TUBING SILC BLU 040 140 19 40296B002 TUBING SILC BLU 125X 250 21 41579B003 CHECK VALVE 24 41585B003 TUBE CONECTOR 3 32 X 1 8 25 41700B001 TUBING POLYU 094IDX 1870 26 42069B001 TUBING 1 4 1 8 POLYU COIL 28 42081B002 ORIFICE RESTRICTOR 125 30 80111B001 FERRITE BEAD EMI SHIELD 32 40041B009 TUBING SHRINK 375 CLR 33 41700B003 TUBE POLYU 125 x250 70DU Page 9 24 Poet Plus 8100 Service Manual Criticare Systems Inc Printer Module Criticare Systems Inc Section 9 Drawings and Schematics 95604A002 PRINTER MODULE 8100 9100 CSI Part 83279B001 83280B001 42158B001 84005B001 42159B001 91341A002 40995B005 42171B001 Description PRINTER DOOR 8100 9100 PRINTER HOUSING 8100 9100 UPPER SHIELD PRINTER PRINTER SEIKO LOWER SHIELD PRINTER AY PCB 8100 PRINTER P H M S 4 40X 25 SEMS PRINTER MOUNT Poet Plus 8100 Service Manual Page 9 25 Section 9 D
70. II waveforms Check the following lead configurations for the following items e Heart rate 60 3 BPM e RESPIRATION RATE of 20 2 breaths per minute and e Correct waveform Poet Plus 8100 Service Manual Page 5 21 Section 5 Preventative Maintenance 7 Rotate the menu knob to select PARMETERS and press to enable the menu Select ECG for 3 LEADWIRE Set the monitor for LEAD II waveforms Check the following lead configurations for following items e heart rate 60 3 BPM and e correct waveform Ne ee ee I ee ie Hiss A III j j Page 5 22 8 Use the Electrode status error messages on the CRT to perform a lead off check in LEAD II a LL Red LL LEAD OFF b LA Black LEADS OFF c RA White RA LEAD OFF Connect all the leads back on the simulator and select ECG to 5 lead Set the monitor for LEAD II waveforms Disconnect the following leads one at a time and look for the proper status error message and for the waveform to continue or flatline Reconnect the lead and the message should disappear and the waveform should return if it flatlined a V1 Brown V LEADS OFF b LL Red LL LEADS OFF c LA Black LA LEADS OFF d RA White RA LEADS OFF e RL Green LEADS OFF Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance 10 Check the following lead configurations for following items e Heart rate 60 3 BPM e resp rate 20 2 BPM and e correct wa
71. ING REAR 8100 HOUSING FRONT AY 8100 ACT AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CHASSIS FRNT 8100 CO2 02 CONN AY 2 TEMP AY CBL ECG ANALOG TO OUT COVER PRINTER OPENING DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 43 SEMS ACCESS DOOR O2 SENSOR CONN MNT PLT 2TEMP FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS CABLE TIE 8 LG NYLON P H M S 8 32X 375 PL PH Criticare Systems Inc Poet Plus 8100 Service Manual Page 9 15 Section 9 Drawings and Schematics Final Assembly 93951A402 8100 CO2 IBP 2T 6WAVE COABP 2Temp 6Wave Item CSI Part Description 01 93258A001 AY HOUSING REAR 8100 03 93259A006 HOUSING FRONT AY 8100 ACT 05 92788A013 CHASSIS FRNT 8100 CO2 IBP 07 95612A001 AY POWER MODULE 09 91340A003 AY PCB 8100 MAIN BRD 10 91353A002 AY SHIELDED ANALOG BRD 12 95618A004 CONN AY 2 IBP 2 TEMP NIBP 15 90871A002 AY CBL ECG ANALOG TO OUT 17 45114B001 COVER PRINTER OPENING 19 41891B001 DB9 COVER PLATE 21 42225B001 SUPPORT FRONT PLASTIC 22 42227B001 RICHCO WIRE CLIP 25 40284B001 NUT 4 40 KEPS PL 26 40284B002 NUT 6 32 KEPS PL 27 42065B004 F
72. O gas absorbs certain wavelengths of infrared energy with the amount depending on the amount of CO gas A detector outfitted with an infrared bandpass filter only allows a certain wavelength of infrared energy that the CO gas absorbs As the CO concentrations in the sampled air passes through the sample cell the detector senses the infrared energy changes and converts them to an electrical signal This electrical signal is then amplified and conditioned and sent to a microprocessor for processing and then sent to a display The CO Board also uses a reference channel The purpose of this reference channel is to cancel the effects of IR source changes sample cell contamination temperature etc The only difference from the CO channel is that the detector is outfitted with an optical filter that is tuned to a band in the infrared spectrum that is transparent to CO and other gases typically found in the medical industry The signal from the reference channel is processed similarly to that of CO measurement channel Poet Plus 8100 Service Manual Page 3 3 Section 3 Theory of Operation Galvanic Oxygen Sub Module 02 Temperature Communications Printer Page 3 4 The CO and reference detectors work by an effect known as pyroelectric Pyroelectric effect works by changing energy levels either increasing or decreasing that are imposed on a sensing element The detector is outfitted with an optical bandpass f
73. Once the metal is used up the depleted cell can no longer sense oxygen The cell generates a voltage output proportional to the amount of oxygen in the sampled gas This oxygen cell has an internal thermistor and circuitry that adjusts the output voltage based on current temperature of the cell The voltage is read by the microprocessor and an O measurement is generated using predictive circuitry This predictive function enhances response time of the O monitoring module The relationship between gas concentration and pressure is calculated by the microprocessor The monitor displays the numerical value O2 Calculated generated using the following formula 20 9 uke O Calculated O Measured x O Ambient P The measured O predictive value is multiplied by a fixed value derived from room pressure divided by room ambient oxygen levels The measurement is further adjusted by multiplying the ratio of the pressure determined at calibration P with the current pressure Po There is a negligible effect on O measurements due to humidity The O function is appropriate for measuring respiratory Oo concentrations in all patient populations The O monitor is suitable for use in breathing systems and with the use of inhalation anesthetic agents The oxygen cell contains temperature correction circuitry The oxygen sensor temperature is maintained at a nominal 40 degrees Celsius to maintain a consistent performance Poet Plus 8100
74. PARAMS Use the rotary encoder knob to scroll to select the IBP1 ZERO selection on the display Observe that a 0 zero appears on the screen next to the line that contains ART MMHG Return to the Medsim and select WAVE and execute Select ART for ARTERIAL and press EXE to execute the function Observe the waveform on the screen Verify that the pressure reads 120 80 2MMHG Attach the IBP cable from BP2 on the simulator to IBP 2 on the connector panel Select BP on the simulator Select Channel 1 Select STAT Select ZERO by using the up and down arrows on the MED SIM 300 scroll through the menu until STAT appears and press to select Rotate the rotary encoder knob to highlight PARAMS and move down to IBP2 Press the menu knob once Set the SITE for CVP Press the rotary encoder knob again Adjust the RANGE This should be set for 0 20 Press the rotary knob until you reach IBP2 again Exit out of PARAMS 10 Exit out of PARAMS Use the rotary encoder knob to scroll to select the IBP1 ZERO selection on the display Observe that a 0 zero appears on the screen next to the line that contains CVP MMHG 11 Return to the Medsim and select WAVE and execute Select Page 5 30 CVP and press EXE to execute the function Observe the waveform on the screen Verify that the pressure reads 12 1MMHG Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance Com
75. Poet Plus 8100 Patient Monitor Service Manual C E oss Cat No 1071 Part No 39125B101 Date 08 05 Revision 4 Copyright COPYRIGHT CRITICARE SYSTEMS INC 2003 2005 CRITICARE SYSTEMS INC Criticare owns all rights to this unpublished work and intends to maintain this work as confidential Criticare may also seek to maintain this work as an unpublished copyright This publication is to be used solely for the purposes of reference operation maintenance or repair of Criticare equipment No part of this publication may be reproduced in any manner or disseminated for other purposes In the event of inadvertent or deliberate publication Criticare intends to enforce its rights to this work under copyright laws as a published work Those having access to this work may not copy use or disclose the information in this work unless expressly authorized by Criticare to do so All product specifications as well as information contained in this publication are subject to change without notice All information contained in this publication is believed to be correct Criticare Systems Inc shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing performance or use of this material This publication may refer to information and products protected by copyrights or patents and does not convey any license under the patent rights of Criticare Systems Inc nor the
76. Press the ISO button and monitor the meter for less than 20 microamps For inline voltages of 264V meter for less than 50uA Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance 13 Keep the power cord plugged into the unit under test and the Dynatech Unplug the 8100 ECG cable from the front of the monitor Unplug the ECG test leads from the Dynatech Connect the IBP cable end to any of the BROWN ECG test terminals on the Dynatech Connect the other end to the IBP connector on the EUT MODE switch should be on ECG Change the LEADS switch to ISO Set the POLARITY switch to NORMAL Press the ISO button and monitor the meter for less than 20 microamps For inline voltages of 264V meter for less than 50uA 14 Verify that the green AC LED lights up on the front membrane Criticare Systems Inc Poet Plus 8100 Service Manual Page 5 17 Section 5 Preventative Maintenance Ground Continuity Test This portion of the test requires a ROD L Model 25 Ground Continuity Tester or equivalent Do the following before turning on the ground tester 1 Set the ground continuity tester to the resistance trip point setting of 0 1 ohms 2 Set test time of 6 seconds Set the tester to ground test only Set the test current to 25 AMP s 3 Turn on the ground tester 4 Plug in the AC cord from the ground continuity tester to the monitor being tested 5 Connect the ground cable to the cha
77. S 6 32X 43 SEMS ACCESS DOOR O2 SENSOR CONN MNT PLT 2TEMP FITNG QUICK CON HH NIBP NUT 10 32 KEPS PL COVER CAN CONNECTOR 8100 P H M S 4 40X 25 SEMS CABLE TIE 8 LG NYLON P H M S 8 32X 375 PL PH Page 9 21 Section 9 Drawings and Schematics Final Assembly PrinterCO IBP 2Temp 6Wave Page 9 22 93951A690 8100 CO2 PRNTR IBP 2T 6W Item CSI Part 01 93258A001 03 93259A006 05 92788A013 07 95612A001 09 91340A003 10 91353A002 12 95618A004 15 90871A002 18 95604A002 19 41891B001 21 42225B001 22 42227B001 25 40284B001 26 40284B002 27 42065B004 28 40995B001 29 40995B006 33 40496B001 35 40876B001 36 40873B004 38 42175B001 40 83278B001 44 46220B001 45 42220B001 46 40909B002 47 40496B006 48 42132B001 50 42014B001 51 40284B005 52 83375B001 53 40995B005 57 40067B003 60 40873B001 Poet Plus 8100 Service Manual Description AY HOUSING REAR 8100 HOUSING FRONT AY 8100 ACT CHASSIS FRNT 8100 CO2 IBP AY POWER MODULE AY PCB 8100 MAIN BRD AY SHIELDED ANALOG BRD CONN AY 2 IBP 2 TEMP NIBP AY CBL ECG ANALOG TO OUT PRINTER MODULE 8100 9100 DB9 COVER PLATE SUPPORT FRONT PLASTIC RICHCO WIRE CLIP NUT 4 40 KEPS PL NUT 6 32 KEPS PL F H M S 4 40 X 375 PH PL P H M S 4 40X 312 SEMS P H M S 4 40X 38 SEMS P H M S 6 32X 375 SEMS P H M S 8 32X 25 PL SEMS P H M S 8 32X 750 PL PH STANDOFF 632X 625 25HEX BATTERY 8100 LBL SERIAL 8100 LBLS LASER 5167 SHEET SPCR HEX 6 32X 625 NY M F P H M S 6 32X 4
78. SPLIT FLAT CBL SUPP CORE 41955B001 FOAM TAPE 1 0 X 2 12 95624A001 AY CBL SPO2 TO DB9 42212B002 CABLE SUPPORT CO2 MAIN 91339A001 AY PCB 8100 SPO2 BRD 91767A001 AY 02 C02 8100 42228B002 SHIELD SPO2 ANALOG CBL 42230B002 INSULATING BUSHING 40041B009 TUBING SHRINK 375 CLR 40683B001 TAPE TRANSFER ADH 5MIL 40067B002 CABLE TIE 5 5 LG NYLON Poet Plus 8100 Service Manual Page 9 7 Section 9 Drawings and Schematics Front Chassis CO IBP 92788A013 CHASSIS FRNT 8100 CO2 IBP Item CSI Part Description 01 94253A001 BTC PUMP MOTOR AY 02 42131B001 CHASSIS FRONT 03 40496B002 P H M S 6 32X 250 SEMS 04 94252A001 AY 8100 NIBP PUMP 05 40995B005 P H M S 4 40X 25 SEMS 06 42189C001 STNDOF M 4 40 X 1 312 HEX 07 83311B001 INSULATOR FRONT CHASSIS 08 42211B001 INSULATOR CHASSIS BACK 09 40294B001 HOLDER TIE WRAP ADH BK SM 10 41602B001 TUBING POLYU 188 X 3125 11 40067B001 CABLE TIE 4 0 LG NYLON 15 87225B001 CBL 2MM PITCH RIBBON 16 40909B001 SPACER HEX 6 32X 5 NY M F 17 90866A001 SPO2 TO ANALOG BOARD CBL 18 82515B001 SPLIT FLAT CBL SUPP CORE 19 41955B001 FOAM TAPE 1 0 X 2 12 20 95624A001 AY CBL SPO2 TO DB9 24 42212B002 CABLE SUPPORT CO2 MAIN 28 91339A001 AY PCB 8100 SPO2 BRD 29 91766A001 AY C02 8100 30 42228B002 SHIELD SPO2 ANALOG CBL 31 42230B002 INSULATING BUSHING 32 40041B009 TUBING SHRINK 375 CLR 33 42187C001 STANDOFF MODIFIED 34 42188B001 STANDOFF 25 4 40 X 812LG 35 91350A002 AY IBP BRD 36 83352B001 CBL 2MM IDC STRIPS 37 40683B001 TAPE TR
79. Service Testing and Calibration Page 6 16 8 Set the simulator to 256 mmHg and the toggle switch to The Pressure Reported value changes to a positive integer that is close to the value of 256 9 Select Set 256 mmHg Point and then select Yes A corrective offset value appears at the end of the line An example is 852 8 The Pressure reported value changes to 256 NOTE If the value at the end of the line is unchanged then the calibration CAL failed Restart calibration and begin at step 4 10 Once the pressure has stabilized 256 to store your calibration select Store CAL Results and then select Yes Verify that the IBP 1 GAIN and offset values change to match the values from the set 0 and set 256 lines 11 Insert the BP 600 into the bottom IBP connector Set the BP 600 to a pressure of 256MMHG and the switch set to ZERO Select Channel 2 12 For calibration select Start or Restart Cal and then select Yes 13 Allow the pressure to stabilize before you select YES to the step below Select Set 0 mmHg Point and then select Yes NOTE A set of corrective offset values appears at the end of the line Example 11 2 14 Set the simulator to 256 mmHg and the toggle switch to The Pressure Reported value changes to a positive integer that is close to the value of 256 15 Select Set 256 mmHg Point and then select Yes A corrective offset value appears at the end of the line Example 852 8 The Pressure reported val
80. Tubing Valve3 Valve2 Tubing Pump Exhaust Title CO Module PCB Os Module PCB SpO PCB SpO PCB Schematic Main PCB Main PCB Schematic Printer Interface PCB Printer Interface PCB Schematic ECG Analog PCB ECG Analog PCB Schematic IBP PCB IBP PCB Schematic Poet Plus 8100 Service Manual 95625A001 95626A001 95627A001 95628A001 95629A001 95630A001 95631A001 95633A001 Drawing Number 91330A001 91332A001 91339A001 91339S001 91340A003 91340S003 91341A002 913415002 91342A002 913428002 91350A002 913508002 Criticare Systems Inc O2 CO Assembly Section 9 Drawings and Schematics O Module Assembly 91354A001 AY O2 MODULE 8100 ltem 01 CSI Part 90399A002 40995B005 91765A001 83147B001 80078B001 42147B001 40067B001 40284B001 Description AY CABLE 02 SENSOR P H M S 4 40X 25 SEMS AY 02 W MANIFOLD amp RESIS FERRITE BD 2750HMS 100 02 SENSOR INSULATOR 02 SENSOR CABLE TIE 4 0 LG NYLON NUT 4 40 KEPS PL CO Module Assembly 91355A001 AY 8100 CO2 MODULE O Manifold Resistor Assembly CO Pneumatics Assembly O CO Pneumatics Assembly Criticare Systems Inc Item CSI Part Description 01 40995B005 P H M S 4 40X 25 SEMS 02 41258B002 P H M S 2 56X 375SEMS PH 03 41457B001 P H M S 2 56X 25 PL PH 04 42006B001 SPACER 093ID X 1880D X 05 42186B001 STANDOFF 4 40 X 218 HEX 06 42193B001 CO2 SHIELD 8100 MONITOR 07 91330A001 AY CO2 PCB 8100 08 95611A001 AY 8100 CO2 BEN
81. UBING SILC BLU 125X 250 03 40296B001 TUBING SILC BLU 040X 140 04 95599A002 AY HYDROPHOBIC DISK CO2 Criticare Systems Inc Poet Plus 8100 Service Manual Page 9 31
82. Will be stopped Enter password Figure 2 3 Password Window for Service Mode 5 To enter the service mode enter the password PIA418 by doing the following e Rotate the menu knob to select a character e Press the menu knob to enter each character e Ifa mistake is made advance to the end of the password and start over When you enter the correct password the Service Mode window appears See the next step Page 2 2 Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc Section 2 Service Windows If you entered an invalid password the window below appears Check the validity or spelling of the password you entered and repeat the password entry Enter Service Mode gt gt gt CAUTION lt lt lt All monitoring will be stopped Enter password nvalid password Figure 2 4 Invalid Password Window 6 When you enter a valid password the Service Mode window appears EXIT Enter NIBP Service NO Enter C02 02 Service NO Enter IBP Service NO Software Download No action taken ST Arrhythmia Keycode 111P03 Valid Keycode Permanent Alarm Silence Enable Current Software Revisions Main Display Processor Rev 4 2D NIBP CO2 02 Processor Rev 4 0B Vital Signs Processor Rev 5 1B IBP 1 2 Processor Rev 5 0A IBP 3 4 Processor Not Present Agent Bench Processor Not Present Mezzanine Processor Not Present Unit Serial Number 411111111 Unit Part Number 91340A001
83. able at J1 of the CO Board if replacing the CO Board or J2 of the O Board if replacing the O Board Remove the O cell Figure 7 44 O Board 6 Reassemble either or both boards in reverse order Replacement of either the O Board or the CO Board requires user calibration for both boards See Section 6 for calibration procedures NOTE If you need to replace the IBP Board or the NIBP Pump or Valve perform these procedures before you reassemble the unit See IBP Board Removal Replacement on page 7 26 and or NIBP Pump Valve Removal Replacement on page 7 27 for instructions Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 25 Section 7 Disassembly IBP Board Removal Page 7 26 Replacement 1 With the CO O2 removal complete remove the three standoffs that hold the IBP in place 2 Label and remove the cables on the IBP Board NOTE The IBP Board cables are not unique Label the cables so that you can reconnect them properly Figure 7 45 IBP Board 3 Remove the two Phillips screws that hold the IBP on the chassis 4 Replace the IBP Board 5 Perform an IBP calibration if you replaced the IBP Board 6 Reassemble in reverse order Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly NIBP Pump Valve 1 Follow CO2 02 Removal Replacement on page 7 24 The Removal Replacement NIBP Pump Assembly is located underneath NOTE If your u
84. ack half of the display housing Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 9 Section 7 Disassembly Inverter Board Removal 1 Label and disconnect the cables Item 1 on the photo on the Replacement Inverter Board 2 Remove the two small Phillips screws and plastic washers Inverter Figure 7 22 Remove the Inverter Board 3 Remove and replace the Inverter Board 4 Fasten the Inverter Board with the two screws and nylon washers you removed earlier See 2 in Figure 7 22 Remove the Inverter Board NOTE When you install the new Inverter Board be sure the nylon standoffs are in place 5 Reattach the cables See 1 in Figure 7 22 Remove the Inverter Board Page 7 10 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly Display Screen Removal 1 Remove the four Phillips screws Replacement Figure 7 23 Remove the Display Housing Screws 2 Label and disconnect the cable at Label 2 from the Inverter Board Inverter Figure 7 24 Remove the Inverter Board Connection Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 11 Section 7 Disassembly 3 Disconnect the ribbon cable from the display Figure 7 25 Remove the Ribbon Cable Display Screws 4 Turn the display over and remove the four Phillips screws Le a gt la fe e 6 Figure 7 26 Remove the Display Screws 5 Remove
85. age causes the monitor to revert to battery power Very high line voltage may cause damage to the charger circuits The monitor is designed with circuitry that turns the unit off before spurious readings can be caused by a low battery condition Poet Plus 8100 Service Manual Page 1 17 Section 1 Introduction Software Error Related Hazard Mediation Potential Interference Page 1 18 Use of Anesthetics Biocompatibility Latex Content Criticare Systems Inc has quality control practices and procedures in place to review potential hazards as they relate to software The monitor is Year 2000 Compliant and utilizes a 4 digit year for all date time and leap year calculations This device has been successfully tested to IEC 601 1 2 specified levels for emissions of and resistance to electromagnetic energy fields External disturbances which exceed these levels may cause operational issues with this device Other devices which are sensitive to a lower level of emissions than those allowed by IEC 601 1 2 may experience operational issues when used in proximity to this device MAGNETIC FIELDS Use of the monitor in an MRI environment may interfere with MRI image quality Use of MRI may interfere with the monitor RADIO FREQUENCY INTERFERENCE The monitor conforms with IEC 1000 4 3 for radio frequency interference and will operate with negligible adverse effects CONDUCTED TRANSIENTS The monitor conforms with IEC 1000 4 4 a
86. alibration Calibrate if necessary Change the O cell Check the CO absorber Change if necessary Check the gas flow rate Adjust if necessary Criticare Systems Inc Poet Plus 8100 Service Manual Page 5 1 Section 5 Preventative Maintenance Monitor Safety Testing ta Page 5 2 You can perform the following tests as part of a periodic safety check The following safety tests are designed so that the monitor s warranty seal does not have to be broken If the monitor fails any portion of these tests contact the CSI service department The contents of this section Preventative Maintenance include the following verifications and safety tests e Withstanding Voltage Hi Pot Electrical Leakage Ground Continuity Interface Inspection Manual Controls Check Alarm Verification Vital Sign Modules Verification e ECG SpOo e NIBP e Temperature e Gas Flow CO O e IBP e Internal Printer e Communications e Battery and Power Criticare recommends that you run a serviced monitor for the following time periods before you perform the tests listed above and place the monitor back into service e Cat 8100 Family with CO O requires 48 hours e Cat 8100 Family without CO O requires 24 hours Repaired modules could require more extensive testing than what is shown in this list Always follow ESD precautions when performing a procedure discussed in this section WARNING e Because te
87. all electrostatic discharge ESD protection as described Perform the disassembly procedure on an antistatic mat that is grounded Check the ground cable to insure that it is connected to a good earth ground Use a grounded soldering iron Wear a wrist grounding strap Connect the wrist strap and mat through a resistor 1 megohm typical to the same ground source Test the wrist ground straps on a daily basis Temporarily store components in metal or antistatic containers Do not store components in plastic dishes Store circuit boards in sealed antistatic bags or cover them in antistatic boxes Do not store electronic boards directly in cardboard boxes Front Housing Removal Follow the ESD cautions discussed in Electrostatic Discharge Protection on page 7 2 1 Page 7 2 Remove the eight Phillips screws from the back of the unit as it sits on the table Figure 7 10 Screws on the Back of the Monitor Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly 2 Carefully pull off the front housing in order to avoid pulling any wires NOTE The unit may be hard to open due to the curing of the adhesive on the water ingress shield Carefully separate the housing taking care that you don t crack the housing Front Housing Figure 7 11 Remove the Front Housing 3 Label and disconnect the printer ribbon cable from J1 of the Main Board See Figure 7 11 Remove the Front Housing
88. ally present in the patient The heart rate is determined primarily from the ECG waveform data A beat detection algorithm is used to identify QRS beats The patient monitor has a user selectable smart heart rate function It automatically uses alternate sources to determine heart rate if the primary source becomes unmeasurable e The IBP waveform is used if the ECG waveform data is unavailable e The plethysmograph SpO waveform is used if the IBP waveform data or ECG heart rate is unavailable e Inthe absence of SpOs IBP and ECG data the NIBP oscillometric data is the final default source for a heart rate measurement Response times for ECG heart rate meter change from 80 BPM to 40 BPM and from 80 BPM to 120 BPM is less than or equal to 10 seconds The alarm for tachycardia is less than or equal to 10 seconds per EC 13 NOTE The accuracy of the heart rate depends upon the source The range of the measurable NIBP based heart rate does not extend as far as the range available in other modules used by the smart heart rate feature NOTE The NIBP based heart rate is not a continuous measurement and is only current during an NIBP measurement Poet Plus 8100 Service Manual Page 1 3 Section 1 Introduction ECG Measurement Method Stability of Accuracy Pacemaker Pulse Rejection Page 1 4 The electrocardiogram ECG or EKG records the changing potential generated by electrical activity of the heart To obta
89. ance Testing Programming the SmartSat Analyzer Test Procedure Criticare Systems Inc Section 6 Service Testing and Calibration This test requires a SmartSat Pulse Oximetry Analyzer Model SS 100A The SmartSat is a programmable simulator and probe analyzer The SmartSat is the recommended device for testing the DOX SpO module The SmartSat model SS 100A is available from Clinical Dynamics Corp of Wallingford CT The SmartSat comes standard with a Lemo style connections The Cat No 913A adapter that converts Lemo to DB 9 style SpOz connections is needed for the 8100 series The analyzer also has a custom port designed for testing DOX SpOz sensors The SmartSat can be used for spot checking SpO values using the manual settings The manufacturer recommends using a timed and programmed sequence to ensure that there is optimal performance For older revisions of SmartSat Analyzer that does not have a program for the Model 8100 monitor program the SmartSat as specified below Verify that the monitor s display value matches the corresponding value in the test procedure not the program settings Auto SEQ MobEL 8100 Oximeter DOX Step Time Mod Setting Rate 1 18s 5 00 99 40 2 20s 5 00 96 60 3 20s 5 00 89 80 4 20s 5 00 73 100 5 20s 5 00 52 120 6 20s 5 00 42 180 7 20s 5 00 26 300 8 END Recent versions of the SmartSat Analyzer are pre loaded with a different program but utilize the same test 1 Verify the SpO m
90. and or J5 Make sure all Inverter connections are secure Replace the inverter NIBP air leak NIBP not functioning Defective cuff Defective hose Pneumatic tube leaks Defective valve Pump disconnected Pump motor failure Main Board pump driver bad NIBP calibration error Replace Cuff Replace Hose Re seat tube or replace pneumatics Contact CSI Service Department Reconnect pump to P2 on Main Board Contact CSI Service Dept Contact CSI Service Dept Contact CSI Service Dept SpO not functioning SpO gt Board disconnected ECG S1 switch set wrong Software problem Reconnect to P5 on ECG Analog Board and check the connection at P1 on SpO Board Adjust switches to correct module 1 4 OFF for DOX SpO Contact CSI Service Dept Criticare Systems Inc Poet Plus 8100 Service Manual Page 8 1 Section 8 Troubleshooting Symptom Problem Solution ECG noise or intermittent function Poor electrode placement Bad wrong ECG Cable Line Frequency Filter wrong ECG connection worn Cable to ECG Analog Board bad Reposition replace electrodes Replace ECG cable Check ECG settings Replace ECG port Replace cable assembly at P1 on ECG Analog Board SpOz is noisy Sensor may be bad Receptacle may be damaged The connection between Main Board J12 and ECG Analog Board P4 is loose or disconnected Main Board or ECG Analog Board may be bad
91. anding Voltage Tester to Hot Neutral and the black lead to ECG Ground Isolated Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 6800VDC Connect the red lead of the Withstanding Voltage Tester to Hot Neutral and the black lead to IBP Ground special test connector is required Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 6800VDC Connect the red lead of the Withstanding Voltage Tester to Hot Neutral and the black lead to the Nurse Call Ground special test connector is required Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 6800VDC Connect the red lead of the Withstanding Voltage Tester to the ECG ISO Gnd and the black lead to the Secondary GND BNC Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 4242VDC for 1 minute Connect the red lead of the Withstanding Voltage Tester to the ECG ISO Ground special test connector is required and the black lead to the Chassis AC GND Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 2600VDC Connect the red lead of the Withstanding Voltage Tester to the Secondary GND BNC and the black lead to the Chassis AC GND Set the Withstanding Voltage Tester to a Trip Current of 1mA and test 707VDC for 1 minute Connect the red lead of the Withstanding Voltage Tester to the EPT and the black lead to the SPO2 Ground special test connector is required S
92. ard can corrupt this software See procedure 91340T003 for more information There are two replacable 3 amp fuses pn 82004B001 for the battery input at PS1 There is one replaceable fuse pn 82004B001 for the power supply module at PS2 F4 protects the supply output If you replace the Main Board you are required to cal check the NIBP module NIBP pneumatics assemblies and Main Boards may not be changed unless re calibrated The Main Board is not a field repairable component Replacement Main Boards are pre tested programmed and calibrated Poet Plus 8100 Service Manual Criticare Systems Inc Power Module AN Fuse Replacement Fan Replacement A Criticare Systems Inc Section 8 Troubleshooting The power module pn 95612A001 consists of the battery compartment bulkhead and chassis the fan cable assemblies and the Power Supply Board The Power Supply Board AC to DC converter is not a field repairable component If the board is damaged or fails test replace the entire board pn 80232C001 DANGER A e High Voltage Repairing and testing the power supply requires working with dangerous voltages The power supply module should only be serviced by experienced electrical repair technicians The main fuses pn 80131B002 for the power module is located in an external compartment above the AC power cord connection Replace only with e 0 5A 230 volt Slo Blow fuses 240 volt units e 1A 110 volt Slo Bl
93. assiS IBP cee cece cess eeeeeeeeeeeeeeeeeeesesaeseaeeseeeseaeeeaeeeeeeenaeeeaeenee Front Chassis CO O5 Front Chassis CO IBP Front Chassis CO O2 IBP Rear Housing Assembly 0 Front Housing Assemblies ee ee ee cece ecee eens eee cee eeeeseaeeseeessaeesaaeseaeessaeseaeeneges Front Housing Assembly without COo eee cece ceeeeeneeeeeeseaeessaeeneeneees 9 11 Front Housing Assembly with CO Final ASSOMDIIGS isnende es eieiei aiie Final Assembly 2Temp 6Wave sesseseesseriesisererirrinsrnrinsrrerinrinsirerinrinerrnreent Final Assembly CO2 2TeMmp 6Wave sssesssssseseesrreinesinrinsrrerrnrinnrrerinsrrerrnrenes 9 13 Final Assembly IBP 2TeMmp 6Wave eeceeccec cee cess ceeeeceeeeeeeeseeesseeesaeenaees 9 14 Final Assembly C0O3 09 2Temp 6Wave eccececceeeeee cess senses seats seeeeesieees 9 15 Final Assembly COo IBP 2T emp 6Wave ou cece eee ceee cee eeeeeeneeseneeeneeeenees 9 16 Final Assembly COo Oo IBP 2TEMP 6WaAVE cccccccccecseeseeeetectscteesetsetsenee 9 17 Final Assembly Printer 2Temp 6Wave cc cece eeseeeseeeeeeeeneeeeeeeneeeseneeeaes 9 18 Final Assembly COo 2TeMmp 6Wave ceccceceeeceeeeeenseeeeeeseetseesseessaeeeeeneaes 9 19 Final Assembly Printer IBP 2T emp 6Wave 0 00 0 eee cee eeeeeeeeeeseeseneeeneeeeaees 9 20 Final Assembly Printer COo Oo 2TEMP 6WaVE ccceccesceseeseteeeeeeeeeeneees 9 21 Final Assembly Printer COo IBP
94. at the ground cableis attached with the nut to the power module 38 Verify that the quick connect is connected to power module 39 Verify that wire harness is attached to the input of the power supply 40 Attach shielding with tape 41 Install power module assembly 42 Install four screws to hold Power Supply Assembly in place 43 Verify that battery door has been adjusted 44 Verify cables are routed properly and pulled through window 45 Position Front Chassis into rear housing 46 Route printer cable above front chassis 47 Verify CO tubing is routed to the left side 48 Verify Battery cable GND and NIBP hoses are routed to left side of chassis 49 Verify that AC cable is routed through the top of the chassis 50 Place Front chassis into rear and secure with top screws 51 Connect front support with screws into chassis 52 Insert SpO cable into SpO board 53 Attach ground wire to front chassis 54 Slide up all ferrites so that they lie under the tie wrap 55 Secure all ferrites under tie wrap and pull tight 56 Verify that RTV is applied to T2 solder lugs 57 Verify shielding is attached 91342A001 only 58 Verify bare wire is attach from TP1 to shield 91342A001 only 59 Attach ECG cable to J1 of the Analog board 91342A001 only 60 Assemble Shielding to ECG board using screws and washers 91342A002 _ 61 Plug ECG cable 90871A002 into analog board 62 Install ECG board with standoffs 63 Verify 2 standoffs and 1 plastic standoff 64 Add last standoff th
95. at the monitor be returned to a CSI authorized service facility if pump replacement is required The transducer SENS1 and the zero point calibration pot R92 are not field serviceable components Improper replacement can result in permanent damage to the Main Board Criticare Systems Inc Poet Plus 8100 Service Manual Page 8 3 Section 8 Troubleshooting ECG Impedance Respiration Module Board Settings and Module Page 8 4 Software Respiration Temperature The ECG module is located on the ECG Analog Board assembly pn 91353A002 that is attached to the metal chassis behind the main board The ECG Analog Board fits directly on to the Main Board through a socketed 10 pin connection on the left side of the board assembly There is also a shielded cable assembly that extends from the top of the board to the external ECG port Check each of these connection points if the monitor fails to produce an accurate ECG waveform The ECG Analog Board has a foil shield to reduce noise from other portions of the monitor Noise can be introduced by worn cable assemblies and degraded power supplies Test the analog ECG waveform signal and the input power where they enter the ECG Analog Board for interference noise There is a replacable 2 amp fuse pn 80004B003 at location F1 on the ECG Analog Board The fuse provides protection in the power circuit leading from the P4 connection to the Main Board Surface mount components on the ECG
96. b to select a character e Press the menu knob to enter each character e If you make a mistake advance to the end of the password and start over e After you press the menu knob for the last character of the password the monitor indicates acceptance of the password by displaying the Set Unit Defaults window e When the Set Unit Defaults window displays both SELECT and CUSTOM DEFAULTS are highlighted Set Unit Defaults Set current configuration and alarm settings as new unit default settings SELECT EDIT CUSTOM DEFAULTS ACCEPT ALTERNATE CARE STD DEFAULTS STD DEFAULTS STD DEFAULTS CANCEL Figure 2 13 Set Profile Before Modifications Poet Plus 8100 Service Manual Page 2 13 Section 2 Service Windows ENTERING THE NEW PROFILE TITLE 3 With both SELECT and CUSTOM DEFAULTS highlighted in the Set Unit Defaults window rotate the menu knob to highlight EDIT and press to select e As you rotate the menu knob CUSTOM DEFAULTS remains highlighted e The first letter of CUSTOM DEFAULTS is highlighted 4 Change the label CUSTOM DEFAULTS for example to the title FACILITY DEFAULTS by doing the following e With the letter C highlighted the first letter of CUSTOM DEFAULTS rotate the menu knob to highlight the letter F the first letter of the example FACILITY DEFAULTS and press to select The next letter automatically highlights e Continue for the remaining letters of the new title e After completin
97. battery completely after extended battery use to ensure a fully charged battery is available for the next use Explosion hazard Keep lighted cigarettes sparks and flames away from the battery The battery contains sulfuric acid electrolyte that can cause severe burns and eye damage as well as illness from sulfur oxide fumes Do not crack cut burn or dissolve with solvents the battery case Damaging the battery case can cause the release of sulfuric acid If sulfuric acid is released from the battery wear eye protection and rubber gloves to handle the battery and use a solution of baking soda in water to neutralize the sulfuric acid The used battery is a potential environmental hazard and must be disposed of properly Dispose the old battery in accordance with local and federal laws Do not incinerate Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance Battery Replacement The battery access door is on the lower left side of the monitor 1 Open the battery access door 2 Lift up a battery s locking tab and then slide it out of its individual Figure 5 1 Raise the Lock Tab to Remove a Battery As you remove each battery note that the battery contacts are located at the end of the battery a Battery Contacts e y Figure 5 2 Battery Contact Location Criticare Systems Inc Poet Plus 8100 Service Manual Page 5 5 Section 5 Preventative Maintenance 3
98. bles through the opening in the front chassis NOTE Pull all ferrites to the ends of the cables and through the opening of the window 9 Position the Front chassis Item 5 pn 92788AXXX into the rear housing 10 Route the printer cable above the front chassis 11 Route the tubing from the CO bench to the left side Also route the Battery cable NIBP hose GND wire NIBP and NIBP transducer hose through the left side of the chassis NOTE All ferrites are to be as close to the connectors as possible Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 35 Section 7 Disassembly Page 7 36 12 Route the AC cable through the top half of the window of the front chassis This cable does not have to be secured down with the others 13 Place the front chassis into the rear housing and secure with two screws Item 35 pn 40876B001 through the back of the top of the chassis 14 Place Front Support Item 21 pn 42225B001 onto bottom of chassis insert two screws Item 36 pn 40873B004 into holes and attach to chassis Attach a Clip Item 22 pn 42227B001 to bottom of Front Support 15 Be sure that the SpO cable assembly fits into slot on side of rear housing and attach it to the SpO board 16 Attach the front chassis ground wire from multiple ground cable to the stud located on the bottom surface of the front chassis using one nut Item 25 pn 40284B001 17 Slide all ferrites up to the point at which
99. bration only if necessary You should have the monitor on and open the Service window 1 O Ko Rotate the rotary knob to highlight the CONFIG softkey Press the rotary knob to open the CONFIG menu Rotate and press the rotary knob to select Enter Service Mode NO is highlighted Turn and press the rotary knob to select YES The Enter Service Mode window appears Enter the service mode password PIA418 to enter the service mode Rotate and press the rotary knob to select Enter CO2 02 Service NO is highlighted Turn and press the rotary knob to select YES Rotate and press the rotary knob to select 150 ml min Flow Rate The value Turn to Incr Decr Press to Save is highlighted Rotate the rotary knob one click clockwise and one click counter clockwise then press the rotary knob NOTE More than one click clockwise or counter clockwise changes the flow setting Verify the flow with a flow meter Use a flowmeter to measure the flow The flow should be 150 2 ml min If the flow rate is not correct turn the rotary knob clockwise to increase the flow or counter clockwise to decrease the flow When the flow is within specification press the rotary knob to save the setting Rotate and press the rotary knob to select 50 ml min Flow Rate The value Turn to Incr Decr Press to Save is highlighted Rotate the rotary knob one click clockwise and one click counter clockwise then press the rotary knob
100. cally on the Main Board 16 Turn the monitor s power off This stores the calibration settings into non volatile memory written to flash 17 Exit the service program on the PC 1 Restart the service program on the PC Select Start gt Programs gt CSI NIBP Service gt NIBPSvc Close the disclaimer dialogue box 2 Turn the monitor s power on again and establish a new communications link between the monitor and the PC service program 3 To perform the service test for speed click on the Speed button then click on Start When the unit passes the speed test press Done 4 To perform the service test for leakage click on the Leak button then click on Start When the unit passes the leakage test press Done 5 To perform the service test for inflation safety click on the Safety button then click on Start When the unit passes the safety test press Done 6 Turn the monitor off Click on Abort if an error message appears on the computer once the monitor is off Poet Plus 8100 Service Manual Page 6 7 Section 6 Service Testing and Calibration Temperature 1 Input Check Exit the service program on the computer Remove the download cable Click No to exit the program click Yes to save the data to a user specified location Connect the temperature simulator TP400 to the one of the connectors attach the cable to P3 of the ECG Analog Board being tested Verify the Temperature on the display is w
101. ce Mode 6 Press the Poet Plus 8100 monitor ON OFF key to start the monitor Connect a 700cc dummy cuff and manometer to the NIBP fitting on the unit On the LCD window of the monitor verify that the message NBP SERVICE TOOL and the pressure reading appears Verify that the monitor s serial number appears on the ID box Click OK Poet Plus 8100 Service Manual Page 6 5 Section 6 Service Testing and Calibration 10 Click on the Calibrate button to calibrate the NIBP module 11 Check the cuff pressure as indicated on the service screen of the computer The cuff pressure should be zero 1 mmHg If the zero base point is out of range the NIBP module must be calibrated If the zero base point is out of range manually set the NIBP module If the zero base point is within 1 mmHg of zero proceed to the next step In this case you skip the instructions which are found in the topic Manual Zero Point Calibration Manual Zero Point Calibration CAUTION A A e Utilize ESD protections at all times a Open the case of the monitor to expose the Main Board A Follow the disassembly procedures in Disassembly in Section 7 of this manual The Main Board can remain attached to the rear assembly b Allow the monitor to remain on You can label and remove the screen power and data cables to gain access to the Main Board You can label and disconnect interior gas sample lines as the case is pulled open
102. ces Hardware under software control performs filtering for monitor vs diagnostic modes lead selection lead off detection and pacer pulse detection Software performs gain control notch filtering to remove line frequency interference pacer artifact handling noise detection beat detection heartrate calculation and serial CAN bus communications The TTI respiration circuit uses a transformer to generate a low current high frequency stimulus through the Lead electrodes The resulting voltage variations reflect impedance variations across the thorax which in turn reflect respiration activity as well as other phenomena including cardiogenic artifact and motion artifact The TTI respiration software performs low pass filtering high pass filtering breath detection and calculates breath rate Poet Plus 8100 Service Manual Criticare Systems Inc Pulse Oximeter Module SpO Capnometer Module CO3 Criticare Systems Inc Section 3 Theory of Operation The pulse oximeter is based upon the principle of dual wavelength photoplethysmography Red and infrared LEDs are alternately driven to illuminate some relatively well perfused tissue site and a photodetector receives either the transmitted or reflected light resulting in a current flow This input signal is demultiplexed with sampling in the monitor to separate the red and infrared components The wavelengths are chosen to permit discrimination of oxygenated and d
103. cter 3 If a mistake is made advance to the end of the password and Criticare Systems Inc start over Poet Plus 8100 Service Manual Page 2 9 Section 2 Service Windows Standby Timeout Alarm Tone Warning Line Frequency Page 2 10 You can set the Standby Timeout value to disable the monitor from 5 minutes to 2 hours Factory default value is 30 minutes If you select OFF you cause the STANDBY key to disable the monitor permanently until you press the STANDBY key again Set the Alarm Tone Warning value to ON to allow the monitor to produce a low pitched double beep every 2 minutes after a new alarm activates Set the Alarm Tone Warning value to OFF to allow the monitor to suspend any audible alarm remain silent after a new alarm activates Thus you can silence alarms with a long push held in for 2 seconds on the monitor s SILENCE key NOTE As a safeguard measure the Permanent Alarm Silence function Service Mode window can override the Alarm Tone Warning function The Disable value of the Permanent Alarm Silence function can prevent the ability of the Alarm Tone Warning function s OFF value to suspend audible alarms In this case pushing the Silence button has no effect the monitor s full audio alarm would continue Set this safeguard according to your facility protocols and according to local safety regulations for medical devices Override from the Configuration Window Serv
104. d Pressure Saturated 21C 750 mmHg 100 Humidity Saturated The monitor is suitable for sustained pressure breathing circuit monitoring environments and has been tested per Clause 102 Sustained Pressure of EN 864 The monitor has an internal barometer and thermistor that allow compensation for changes over a range of temperature and atmospheric pressures The monitor complies with EN 864 standards for cyclical pressure The module as installed in the Poet Plus 8100 patient monitor has been clinically tested for performance with a variety of patients The patient monitor has a manual N2O compensation feature for a fixed N20 value of 60 The user may select NoO compensation when 40 80 N30 is in use Poet Plus 8100 Service Manual Page 1 9 Section 1 Introduction Measuring Oxygen 03 Method Conditions of Use Stability of Accuracy Page 1 10 The Poet Plus 8100 patient monitor uses the sidestream method of measuring O The gas is sampled from the same gas intake system used in CO monitoring The water trap removes moisture and particulate matter from the gas samples The gas sample is measured using a reactive oxygen cell The oxygen sensor is a galvanic electrochemical cell that works by a process known as oxidation reduction Oxygen from the air comes in contact with a highly reactive metal reacts with the metal and produces a current As the oxygen reacts this reactive metal is gradually used up
105. ds and leadwires may present a risk of entanglement or strangulation Verify safe and proper positioning of these items after patient application Unapproved modifications to the monitor may cause unexpected results and present a hazard to the patient Risk of electrical shock Do not remove cover Refer servicing to qualified personnel U S Federal law restricts this device to sale by or on the order of a physician CAUTION A Use the monitor only with recommended accessories Use of unapproved accessories may cause inaccurate readings Equipment accuracy may be affected at extreme temperatures Do not store equipment at extreme temperature Temperatures exceeding specified storage temperatures could damage the system Do not press on the buttons with surgical instruments or other tools Sharp or hard objects could damage the buttons Use only your fingertips to press on the buttons Changes or modifications not expressly approved by Criticare Systems Inc may void the user s authority to operate the equipment and may also void the warranty The monitor complies with leakage current limits required by medical safety standards for patient connected devices Conforms to EN 60601 1 A hazard caused by the summation of leakage currents is possible when several pieces of equipment are interconnected When operated in the line voltage range specified in this manual any fluctuation has a negligible effect Very low line volt
106. e Poet Plus 8100 Patient Monitor Options Criticare Systems Inc The Poet Plus 8100 patient monitor interprets and displays real time physiological data including waveforms and numerical data The monitor is designed for multi parameter measurements including ECG NIBP SpOz temperature and respiration with optional IBP CO and Os For all these vital parameters the Poet Plus 8100 patient monitor has limit alarms and alerts The monitor also prints strip chart recordings and stores tabular trends for review The Poet Plus 8100 patient monitor monitors physiological parameters of patients within clinical care settings It is intended that the user is a professional health care provider Physiological data system alarms and patient data analysis are available to the care provider from the monitor The user is responsible for the interpretation of the monitored data that is made available Physiological data should be reviewed by a qualified clinical personnel prior to any medical intervention The Poet Plus 8100 patient monitor comes standard with 5 Lead ECG ComfortCuff NIBP DOX SpOs and two temperature channels for monitoring Optional features include two IBP channels an internal printer and COs and Op monitoring A color TFT screen with a six waveform display is standard on all Poet Plus 8100 patient monitors Additional External Catalog Number Printer Features Module Port CAT81COOOXO00ECX No Standard Yes CAT81
107. e The Main Board processes error messages from the Printer Board and displays them on the screen The internal printer module is a Sieko LTP1245 type mechanism Poet Plus 8100 Service Manual Criticare Systems Inc Section 3 Theory of Operation Block Diagram The following block diagram aids in tracing communication and data routes within the system Connection numbers are given to show the relationship between the boards The diagram is keyed to the theory of operations found on the previous pages 02 i F co2 x aien i l Temp i Probe fi Power tremor gr Module J Assembly EN l Battery a Main ia COMI Ej GND Con AC Line Criticare Systems Inc Poet Plus 8100 Service Manual Page 3 5 Section 4 Cleaning and Disinfecting Cleaning and Disinfecting Pulse Oximeter Sensors Blood Pressure Cuffs Criticare Systems Inc WARNING e Shock Hazard Before cleaning a monitor and sensor turn the power off and disconnect the AC power cable e Shock Hazard Never immerse the monitor because the monitor has an internal power source that is active even when the unit is unplugged Do not use abrasive cleaners on the monitor or on any sensors or probes Abrasive cleaners damage the monitor sensors and probes With the exception of the display screen wipe clean the exterior surface of the monitor with alcohol Dry it with a soft dry cloth Because paper t
108. e a a INSP440 SRA t 4 sas Pere OT AR o 8 I o a C Figure 7 41 ECG Analog Board r a 9 p s E s 4 Label and disconnect the ECG cable assembly from P1 on the ECG Analog Board Remove the four standoffs holding the ECG Analog Board in place NOTE The standoff in the upper right corner is plastic and this standoff MUST return to this position in reassembly Lift the ECG Analog Board from the chassis NOTE The upper left corner has a capacitor that must pass through the eye of the standoff Remove the four screws and washers and remove the shielding from the ECG Analog Board Replace the ECG Analog Board and reassemble in reverse order Poet Plus 8100 Service Manual Page 7 23 Section 7 Disassembly CO O Removal NOTE Make sure that when you order a CO Board that the CO Replacement Board is factory calibrated 1 Remove the six Phillips screws from the standoffs _ Screws i v EN V Screws Figure 7 42 Pneumatic ee ly 2 Disconnect the pump wires from J3 on the CO Board Figure 7 43 CO Board 3 Disconnect the ribbon cable from J2 on the CO Board 4 Disconnect the PUMP IN tube from the pump Page 7 24 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly 5 If an O Board with manifold is attached remove the four nuts that hold the O Board to the CO Board if necessary Remove the c
109. e POLARITY switch in the NORMAL position press OPEN GROUND and monitor the readout for less than 500 microamps 2 Change the polarity switch to the REVERSE position press OPEN GROUND and monitor the meter for less than 500 microamps 3 With the POLARITY switch in the NORMAL position press OPEN NEUTRAL and OPEN GROUND switches and monitor the meter for less than 1 milliamp 4 Change the POLARITY switch to the REVERSE position and press the OPEN NEUTRAL and OPEN GROUND switches and monitor the meter for less than 1 milliamp 5 Connect the Dynatech 232 to the 8100 using the 5 lead ECG cable marked Cat 1075 S Change the MODE switch to ECG For the rest of the Leakage test insure that the leads are dressed away from the ground planes and the AC power cords and the chassis Set the LEADS selector switch to ALL TO GND Set the POLARITY switch to NORMAL Monitor the meter for less than 5 microamps 6 Set the POLARITY switch to REVERSE and monitor the meter for less than 5 microamps Criticare Systems Inc Poet Plus 8100 Service Manual Page 5 15 Section 5 Preventative Maintenance Page 5 16 7 Set the POLARITY switch to NORMAL and press the OPEN NEUTRAL switch and monitor the meter for less than 10 microamps 8 Set the POLARITY switch to REVERSE and press the OPEN NEUTRAL switch and monitor the meter for less than 10 microamps 9 Set the POLARITY switch to NORMAL and press the OPEN GROUND switch and monitor the me
110. e and click the encoder again You are now in the 150ml range 10 The 10 CO and 95 Os readings display Turn the rotary knob to select CO2 Gain Adjustment and press the rotary knob Adjust the rotary knob clockwise to increase the CO reading or counter clockwise to decrease the CO reading Adjust for a CO reading of 9 7 to 10 1 Press the knob to save the gain NOTE If the waveform on O looks abnormal check the pressure from the O tank There should be only 100ml of flow coming out of the tanks 11 Turn the rotary knob to select O2 Gain Adjustment Press the rotary knob Adjust the rotary knob for an O reading of 95 0 2 0 Press the rotary knob to save the setting Criticare Systems Inc Poet Plus 8100 Service Manual Page 6 13 Section 6 Service Testing and Calibration 12 Once you adjust the CO and Os press the rotary knob then turn the rotary knob until you highlight CO2 Zero Cal Press the rotary knob and select Abort CO2 Zero Cal Press the rotary knob turn the rotary knob one click counter clockwise to select Request CO2 Zero Cal and press the rotary knob again NOTE After the first zero cal request you must first abort before you request a zero cal Every time you make an adjustment you must request a zero cal 13 Once you are satisfied with both numbers after the zero cal turn the rotary knob to highlight 50m min Flow Rate If you are not repeat until satisfied Press the rotary knob click it on
111. ection 7 of this manual NOTE The end of the absorber has a small opening that must remain unblocked as a sampling port during calibration Poet Plus 8100 Service Manual Page 5 11 Section 5 Preventative Maintenance Long Term Monitor and Battery Storage Monitor and Accessories Disposal Page 5 12 No special preparation is necessary for long term storage of the monitor Although you do not have to remove the battery from the monitor for long term storage the battery drains to an unrecoverable state after 3 months without periodic charging At the end of its useful life you can dispose of the monitor and its accessories according to your institution s policies and procedures for disposal of patient contact medical waste Alternately you can return the monitor and its accessories to Criticare Systems Inc for safe disposal The shipping address is Criticare Systems Inc 20925 Crossroads Circle Suite 100 Waukesha WI 53186 Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance Electrical Safety Perform this test each time you open the monitor housing and before Testing you use the monitor on patients WARNING e Shock Hazard Because dangerous and lethal voltages are present during the withstanding voltage leakage and ground continuity test all leakage and voltage testing must be done with the monitor housing in place For complete information about the proper
112. egulator valve on and off Page 6 12 simultaneously while you perform this procedure _ Connect the Poet Plus 8100 monitor to the Respiration circuit box if it is not already connected N Connect the 10 CO to a sample line on the unit w Turn the rotary knob to select the CONFIG menu then turn the rotary knob until Enter Service Mode is highlighted Press the rotary knob and select YES 4 Enter the password PIA418 then press the rotary knob Turn the rotary knob to select Enter CO2 O2 Service Press the rotary knob and select YES The CO and O waveforms should display The CO2 Operating Mode option should be set at User Mode 5 Turn the rotary knob to select O2 Wave Response POT 1 Press the rotary knob Rotate the knob slowly clockwise while you observe the O waveform Watch the rising edge of the waveform Adjust the knob until the rising edge just starts to overshoot From that point adjust the knob counter clockwise until the rising edge no longer overshoots The positive plateau of the waveform should be flat Press the rotary knob to save the setting Poet Plus 8100 Service Manual Criticare Systems Inc Section 6 Service Testing and Calibration 6 Turn the rotary knob to select O2 Wave Response POT 2 Press the rotary knob Rotate the knob slowly clockwise while you observe the O waveform Watch the falling edge of the waveform Adjust the knob until the zero plateau is as flat as po
113. emoval Replacement ssessesseesserierinerrerinrrnsrrerrnrrrerenrrnes SpO Cable SpO Board Removal Replacement 7 18 Front Chassis Removal seseeeeeseseerrerreerrernree wi 7 20 ECG Analog Board Removal Replacement COo Op Removal Replacement cccccccccccsceesecssescescnecsscecssesensctseseeseseaees IBP Board Removal Replacement 0 cccceecccsecseeceneecseseneeeeeeeeeeeneeeeaeeeaas NIBP Pump Valve Removal Replacement P Power Module Assembly Removal Replacement c cessesseeseeseeeneeees 7 29 Power Supply Board sssssessssssseeerruernssresensrneeenniusrasrenennsreneruseenenunenerasrensnnsrnae 7 30 Reassembling the Monitor 000 7 32 Reassembling the Front Chassis 7 32 Final Assembly Rebuild we 7 35 ASSOMDIY ToStS eruan a AE tates 7 40 Section 8 Troubleshooting Troubleshooting and Repair sessesssessesiesrssresrresrnsrrerinrinsrnnrtninnrrnsinntrnstneinnrnnrinnrnanene ne 8 3 ComfortCuff NIBP Module ssessessessrssnessnsenereerenreerrueruseneennreasranrensrasreatenstnnennseneenneanne 8 3 ECG Impedance Respiration Module s sssssssnsissiesnrisnrrisrnsinninnnniesnninninnnrinninrnnnen ne 8 4 Board Settings and Module Software 8 4 Respiration 8 4 Temperature 18 4 DOX ES POs MOU let cisiccr cists a a a aAa aaa aaa a aaa AE eiaa 8 5 OFRO I e E E E ES T 8 5 CO Absorber 8 5 CO Pneumatics CO Pump Replacement
114. encuacerinnsdhenta lia upcasncvicesigieveensatncdmteess Main CPU and System Vital Signs Module Speaker Driver ee EGG MOGUile ss a a A EAR EAE maces AA EEn Impedance Respiration cccccsscsessccsessesssessccseesseessseecseesssessseeseseveneseaeees Pulse Oximeter Module SpO gt oe Capnometer Module CO5 c cccccecececssecesseesesesssesseeseesetesasnsnsutusesuseveeseevsenstavsevseteeeenenes Galvanic Oxygen Sub Module Oo e cee eseesee cess eens eeeeeeeeeseeseaeessaeseaeeeeeeseneaseneeeaes 3 4 TOMPeCrature eee cece eee eerste Communications Printer cm Block Diagrami asn e aanne ienen aaneen AE ae aE Aa E AEA OANE ANEN aa EN AE Section 4 Cleaning and Disinfecting Cleaning and Disinfecting ssessesesesserrerresirsirerinsireirtinsrnstinrinettnrinerretnntinernnrrnernere nt 4 1 P lse Oximeter SONSOMS i ipsia aoia e a aa a aa 4 1 Blood Pressure C ffSieni is n a a a a Ee ATE 4 1 ECG Cable 0 42 Temperature Cable 4 2 Accidental Wetting cs ics ie nectcecgarednorseinensdechedesrint decncsineparenenbctsnegudenbatenurtticeteteenanecranebece 4 4 Section 5 Preventative Maintenance Incoming Inspection Maintenance Schedule ae Monitor Satety Testing seics cc iccsseicaciiescesadercanseycneteh csaudensepeesent astansbtenstbasteedecensundeannrecteds EqQuipment and TOMS c siseviccsnrcesectesctsneetescttscendascapnrssatsdietesdeseinensdesasendoades
115. eoxygenated hemoglobin based on the difference in their absorbance spectra In a pulse oximeter a high passed AC component is extracted from each of the red and infrared signals and digitally processed to detect the pulsations corresponding to blood flow in the illuminated tissue The AC and DC components in red and infrared measured near the time of pulsation are combined in a relatively simple ratiometric formula obtained with many assumptions by solving Beer s law equations with empirically derived coefficients to yield a figure proportional to SpO The pulse rate is also calculated from the pulse intervals detected The capnometer contains the circuitry for monitoring the sidestream CO level by actively sampling from the patient s airway It contains provisions for interfacing with galvanic oxygen monitoring option sub modules No electrical connection is made with the patient from the monitor for capnometry The CONM module controls the capnometer The CO Board contains an assembly known as a bench This bench contains the optics IR source detector sample cell and optical filters The purpose of this bench is to convert CO levels in a gas sample to electrical signals The bench assembly utilizes the principle of infrared absorption spectrometry to measure the amount of CO in patients breath Infrared absorption spectrometry is widely used in gas analysis Using a phenomenon explained in Beers Law C
116. es hisd A Power Supply Board Test 1 Criticare Systems Inc DANGER High Voltage Board testing requires working with dangerous voltages The rear assembly that contains the power supply module should only be serviced by experience electrical technicians Remove the batteries from the monitor Open the case Follow the disassembly procedures refer to Disassembly in Section 7 in this manual to expose the power module Do not remove the power module Disconnect the DC power cable at PS2 of the Main Board Leave the other end attached to the Power Supply Board Using a controlled AC source apply 115 VAC at 60 Hz to the AC receptacle The source should be able to supply a minimum of 3 amps Use a power cable that has suitable connections to the AC source Measure the DC power cable end The power module should supply 15 6 VDC 0 1V Proceed to the next step if the voltage is within specification If it is a new board under test it might not be calibrated The trim potentiometer on calibrated boards should be varnished or otherwise sealed Remove the power module Verify that the trim potentiometer measures 15 6 VDC Reassemble the monitor Poet Plus 8100 Service Manual Page 6 17 Section 6 Service Testing and Calibration Main Board Testing Page 6 18 Test the following conditions without batteries in the monitor 1 Connect a 16VDC 3A power supply
117. ess and hold the NIBP CYCLE STAT key 3 Allow the monitor to take at least four 4 readings Calculate the average of the 4 readings and verify that each reading does not vary by more than 4 or 4mmHg whichever is greater from the calculated average Each reading shall not vary more than 8 mmHg from the simulator setting Criticare Systems Inc Poet Plus 8100 Service Manual Page 5 27 Section 5 Preventative Maintenance Temperature Verification 1 Select Temperature ON from the PARAMS window and TEMP F C Centigrade Connect the temperature to T1 on the unit Insert the simulator TP 400 2 Verify that simulator settings correctly display for all temperatures that range from 20 C to 45 C with accuracy of 0 1 C 3 Select Temperature ON from the menu setup and TEMP F C Centigrade Connect the temperature to T2 on the unit Insert the simulator TP 700 4 Verify that simulator settings correctly display for all temperatures that range from 20 C to 45 C with accuracy of 0 1 C Verify that TEMP INVALID appears when the temperature is at 0 5 Verify that the message TEMP INVALID appears when the temperature is at 0 for both TEMP 1 and TEMP 2 Gas Flow CO O5 Turn on the monitor and then rotate the menu knob to select the Verification PARAMS window Select the unit of measure for PERCENT Select EXIT to return to the Main menu where you select CONFIG and press the menu knob Rotate the menu knob to hi
118. et the potentiometer Hipoter to a trip point of 1mA and test for 2600VDC 10 Connect the red lead of the Withstanding Voltage Tester to the Page 5 14 EPT and the black lead to the Temp1 or Temp 2 Ground special test connector is required Set the potentiometer Hipoter to a trip point of 1mA and test for 2600VDC Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance Leakage Test Equipment Perform all testing with the cover on The following test requires a and Setup Dynatech 232D leakage tester or equivalent 1 Perform a self test if applicable on the DYNATECH 232D by plugging it into 110 VAC outlet Turn on the leakage tester Set the mode switch to self test Display should read 1000 20 and the CURRENT SOURCE ACTIVE lamp should be on CAUTION A e lf these conditions are not met do not continue with this test 2 Set the mode switch on the Dynatech 232 to L1 L2 line voltage The display should read the line voltage 10 Set the mode switch to L1 GND The display should read no more than 5 of L1 L2 on a grounded power system Set the mode switch to L2 GND NOTE This should read about the same as the L1 L2 reading for a grounded system Leakage Test Procedure The monitor must be powered up during all of the isolation tests 1 Turn the mode switch to CASE LEAKAGE GROUND CONDUCTOR Connect the AC cord from the Dynatech 232D to the monitor and turn on the unit With th
119. etime counter clockwise and click the rotary knob again You are now in the 50ml range Go back to Steps 12 to 14 and follow the same procedure for the 50ml flow 14 Request a zero cal if you are satisfied with both numbers after zero cal then go back to the 150ml flow Now flip back and forth between the two flows If no adjustment is needed you are ready to leave the service screen The flow should be back on the 150ml when you are ready to leave Verify that nothing is highlighted then power cycle the monitor off and on 15 Keep respirating the gas into the unit Observe the inspired O3 reading The reading should be 21 0 2 0 16 Verify that in normal operation the 5 and 95 read correctly on the 150ml flow 17 Verify that in normal operation the 5 and 95 read correctly on the 50ml flow 18 If these numbers are satisfactory for you then proceed to the cal check Return to 150ML of flow Flow Rate Verification Follow the procedure found in Gas Flow CO2 O2 Verification on and Calibration page 5 28 Page 6 14 Poet Plus 8100 Service Manual Criticare Systems Inc Section 6 Service Testing and Calibration IBP Calibration You must perform the IBP factory calibration procedure in the Service Mode The Service Mode password is PIA418 You need a Fogg BP 600 Pressure Transducer Simulator to perform this calibration The main focus of the following steps is the monitor with the power turned ON 1 In the monito
120. euentesecesieeted Valve3 Valve2 Tubing Pump Exhaust Criticare Systems Inc Poet Plus 8100 Service Manual Page vii Warranty Workmanship amp Materials Exemptions Safety Reliability amp Performance In Case of Emergency Contact ry Criticare Systems Inc Criticare Systems Inc CSI warranties new equipment to be free from defects in workmanship and materials for a period of one 1 year from date of shipment under normal use and service The 940 Series Multi Site Sensor and the 644 Op Sensor carry a six month warranty CSI s obligation under this warranty is limited to repairing or replacing at CSI s option any part which upon CSI s examination proves defective EXCEPT AS DESCRIBED IN THE PARAGRAPH ABOVE CSI MAKES NO WARRANTIES EXPRESS OR IMPLIED INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE CSl s obligation or liability under this warranty does not include any transportation or other charges or liability for direct indirect or consequential damages or delay resulting from the improper use or application of the product or the substitution upon it of parts or accessories not approved by CSI or repair by anyone other than a CSI authorized representative This warranty shall not extend to any instrument which has been subjected to misuse negligence or accident any instrument from which CSl s original serial number tag or product identificat
121. ews 83 Attach one screw through Cap assembly 84 Connect GND wire from front housing to main board 85 Install last screw to hold gnd wire lug 86 Attach female tab to male tab on display 87 Plug printer cable into J1 88 Attach encoder to J8 89 Attach membrane switch to J7 90 Connect display cable to J2 if it is a Passive 91 Connect display cable to J3 if it is a Active 92 Connect inverter cable to J4 93 Connect trap side to manifold 94 Connect manifold side to manifold 95 Insert ECG cable into rear housing 96 Secure front housing to back housing with 8 screws 97 Connect exhaust tubing to O door fitting 98 Attach CO absorber to cord clip on door 99 Attach the door with 2 screws 100 Insert battery into battery compartment 101 Add connector to back of housing 102 Add serial label to bottom of the unit 103 Add screws to bottom of unit o sTa o 29D Page 7 42 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly Assembly Electronic Test Plug an AC cord into the unit The green AC light should be on Hit the Power key The display should show CSI with current revision of software Check for proper operation of pump by hitting the NIBP key To insure the membrane buttons are operable hit the following keys in sequence and listen for the beep FREEZE NIBP CYCLE STAT NIBP PRINT SILENCE DEFAULT TREND STANDBY Rotate the Encoder knob and verify the cursor
122. g The 40 Hz sine wave shall be 0 70 X the input amplitude in mV when the simulator is set to s10 10 Hz sine wave Measure as described in step 2 Set Filter back to Diagnostic On the simulator use the gt arrow to select 0 05 for a 0 05 Hz sine wave In the monitor s DISPLAY window select a sweep speed of 12 5 mm sec The 0 05 Hz sine wave shall be 0 70 X the input amplitude in mV when the simulator is set to s10 10 Hz sine wave Measure as described in step 2 Select Monitor filtering The 0 05 Hz sine wave is no longer present flat baseline On the simulator select s0 5 0 5 Hz sine wave and verify that a 0 5 Hz sine wave with amplitude equal to 0 70 X the 10 Hz sine wave amplitude is present Set the simulator back to s0 05 Select ST Filtering On the simulator select s0 05 0 5 Hz sine wave and verify that a 0 05 Hz sine wave with amplitude equal to 0 70 X the 10 Hz sine wave amplitude is present Set Filter back to Diagnostic Poet Plus 8100 Service Manual Page 5 25 Section 5 Preventative Maintenance SpO Verification The following test requires a SmartSat SpO analyzer If you service the monitor i e you open the housing use the longer SpO Performance test For more information refer to DOX SpO2 Performance Testing on page 6 3 NOTE For older revision SmartSat analyzers that do not have a Model 8100 selection use the general Criticare selection in the SmartSat simulator Chec
123. g the replacement process to prevent static damage to the sensitive electronic circuits Turn the monitor s power off Remove the two Phillips screws from the rear access door 3 Use a flathead screwdriver to pry the rear access door off the back of the monitor _ 02 ca G A Figure 5 7 O Cell Replacement NOTE Later model units have the CO absorber mounted directly on the inside of the rear access door It is not necessary to disconnect the absorber or the exhaust port tube With your fingernail carefully pull the brown connector from the O cell body Unscrew the expired Op cell and dispose of it NOTE Use only replacement O Cell CSI Catalog 644 CLASS R 22CC OXYGEN SENS Figure 5 8 Oxygen Sensor O Cell Carefully screw in the replacement O cell Do not over tighten this screw 14 turn is recommended Observe the polarity and carefully plug the cable back into the top of the O cell 6 Check the exhaust line and absorber tube for proper connection 7 Replace the rear access door Tighten the screws Poet Plus 8100 Service Manual Page 5 9 Section 5 Preventative Maintenance 8 Turn the monitor s power on and let it warm up for at least one hour before you make response or gain adjustments Failure to allow this warm up can result in inaccurate adjustments 9 Perform an O User Calibration for both the 50 ml min and 150 ml min See O2 Calibration on page 6
124. g the last letter in the title in this example it would be the letter S press the menu knob until the entire label name highlights This indicates that the entry is complete but not accepted ACCEPTING THE NEW PROFILE TITLE 5 To accept the new label name rotate the menu knob to highlight ACCEPT and press to select The message Action Completed appears Set Unit Defaults Set current configuration and alarm settings as new unit default settings SELECT EDIT FACILITY DEFAULTS ACCEPT ALTERNATE CARE STD DEFAULTS STD DEFAULTS STD DEFAULTS CANCEL Figure 2 14 Set Profile After Modifications 6 Press the menu knob Your set of defaults are saved in memory 7 Press the ON OFF button on the front panel of the monitor to turn off the power to the monitor Page 2 14 Poet Plus 8100 Service Manual Criticare Systems Inc Section 3 Theory of Operation Power Module Main CPU and System Vital Signs Module Speaker Driver Criticare Systems Inc The patient monitor employs a universal switching power supply for international use In order to monitor the status of the line power and condition of the battery pack the internal power supply circuitry drives several status lines available to the main module The power management watchdog circuit provides a low battery level detection mechanism and a hardware based check on software operation watchdog timer The microprocessor circuitry handles the Main Board
125. ghlight ENTER SERVICE MODE and press the menu knob to select it Rotate the menu knob to highlight YES The password dialog box appears where you enter PIA418 After you enter the password the Service window appears where you select CO2 O2 mode FLOW CALIBRATION 1 Connect a flow meter to the end of the sample line and measure the flow 150 10 milliliters If not in that range adjust by using the menu knob to select 150 ml min flow and adjust for 150ml Once stabilized press the menu knob to accept any changes 2 Select 50m and press the encoder knob and turn the encoder knob clockwise one click to activate the 50ml flow Once stabilized turn the encoder knob to adjust and then press the encoder knob to accept 50ml Then change back to 150ml 3 Remove the sample line from the flow meter 4 Power cycle the monitor OCCLUSION TEST 1 Connect a vacuum gauge 0 25 inches of mercury to the sample line There should be more than 12 5 inches of mercury of vacuum 2 The message CO2 O0CCULSION should appear The pump should turn on Page 5 28 Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc Section 5 Preventative Maintenance EXHAUST TEST 1 Place the flow meter on the exhaust port and measure the flow The flow should be more than 140 milliliters CO TEST 1 Check the operation of the COs Use 10 CO gas Simulate a breath rate between 10 20 BPM for 150ml min flow and 5 10BPM f
126. he rotary encoder knob from the Removal Replacement unit p N Rotary d Encoder 4 f Knob o lt e Figure 7 15 Remove the Rotary Encoder Knob 2 Remove the retaining nut and lock washer from the front of the housing 3 When you replace the rotary encoder knob be sure to insulate with Kapton tape 4 Reinstall and tighten the nut and lock washer 5 Install the new rotary encoder knob Membrane Keypad Use this procedure only if you need to replace the membrane keypad Removal Replacement on the monitor NOTE Once you remove a membrane from the front panel you cannot reattach it to the front panel 1 With a flat screwdriver pry the rotary encoder knob from the unit ig N Rotary Q Encoder 4 y Knob w ee Figure 7 16 Remove the Rotary Encoder Knob Page 7 6 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly 2 Push a flat screwdriver through the ribbon cable slot to begin peeling the membrane off the front of the unit Do not attempt to use tools to peel the membrane from the front as you may mar the case and be unable to properly attach the new membrane Use a screwdriver to push membrane away from housing through the ribbon cable slot Figure 7 17 Removing Membrane a Peel the membrane away from the front housing iN Remove any excess glue from the front housing Use a cotton cloth dampened with alcohol If any glue remains on the housing the new membrane
127. hical Span Data Types Printer Recorder Type Data Formats Paper Speed Page 1 14 Poet Plus 8100 Service Manual Percent Inspired Oo expired O Numerical values only Oxidation reduction galvanic cell 0 100 1 3 vol 0 100 1 7 seconds 150 ml min 10 90 600 milliseconds 150 ml min 50 ml min or 150ml min User Selectable 2 68 113 F 20 45 C 68 94 9 F 20 34 9 C 0 5 95 107 5 F 35 41 9 C 0 2 107 6 113 F 42 45 C 0 5 YSI 400 or YSI 700 EN 475 Adjustable Audible Visual High Medium Low Informational User Defaults Hospital Defaults Factory Defaults Adult Pediatric Neonate High and low limit settings for each mode User Adjustable 1 10 Yes 2 minutes or permanent Tabular and Graphical 24 hours 30 sec 1 2 5 10 30 min 1 2 4 hrs User Selectable Serial output includes 1 2 5 10 15 and 20 seconds 2 4 8 12 or 24 hours BPM HR SpO Temp Resp ETCO INCO Expired Os Inspired O NIBP Systolic Diastolic Mean IBP Systolic Diastolic Mean Internal thermal line printer Single or dual waveform Tabular 12 5 or 25mm sec continuous Snapshot at 50mm sec Criticare Systems Inc Controls Screen Resolution Waveforms Waveform Display Gain Waveform Sweep Speed Keys Rotary knob Languages System Outputs Com Ports Analog Output Nurse Call Defibrillation Sync Vide
128. hield Fold one side of the shielding on top of the bare wire Pull the other half and fold it on top of that Bend the cable and shield to form an L shape NOTE Keep the connector flush with the edge of the shielding 24 Install the Ferrite Package Item 18 pn 82515B001 onto the SpO to Analog cable Place a piece of Foam Tape Item 19 pn 41955B001 to the back of the Ferrite and adhere this to the back surface of the SpO Board 25 Plug the SpO to Analog Cable Item 17 pn 90866A001 into the SpO Board Item 28 pn 91339A001 26 CO2 CO Oz Attach the tubing from V2 C center valve to the input port on the CO Pump Item 1 see arrows on Pump body 27 CO COVO Attach the CO Exhaust Tubing Assembly pn 95633A001 to the output port on the CO Pump 28 Mount the SpO Module Item 28 pn 91339A001 to the Front Chassis Assembly with the four 6 Screws Item 3 pn 40496B002 onto the plastic Standoffs Item 16 29 Plug the SpO to DB9 Cable with the Plastic Insert Item 20 pn 95624A001 into J1 of the SpO Board Route the cable between the mounting standoffs for the Analog Board 30 Cut four 1 x 1 Double Sided Tape Item 37 pn 40683B001 and place them onto the Insulating Bushings pn 42230B002 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly 31 Install the Insulating Bushings Item 31 pn 42230B002 over the four stand offs that mount the Analog Board 3
129. i Op Calibration ccccccccscceteeecteeteeees Flow Rate Verification and Calibration sessessesineseesrreinerierrnrrnsrrerinsrrerrnrenes 6 14 IBP Calibration pin v toinen ea e aeaa raes aatia raataa oe devour Narea eat ainat 6 15 Monitor Fault Testing ccd deiccccivcatcsuseezonnctlncrssadansinyaneusaperaneyseruetntandenentandndptaeevaatcesbes 6 17 Power Supply Board Test cece eescceeeceeeeceee sees seeeseaeseseesseeseeeseeeeeeesaees 6 17 Main Board TOSI eean a a a ea a a aE a datbesbed 6 18 Poet Plus 8100 Service Manual Page v Page vi Section 7 Disassembly Before VOU Begin sce cisco Ra ETE DEMVICS SAlety E A T E E T cadaheataeedsttisdss Electrostatic Discharge Protection se Front HOUSING REMOVAL isi ccivagescev segs e aaa aa aa sti ae aeran Front Housing Disassembly cccccvccrecascescetcesenccuessccstecsscavrececesevetestncevecssoneoseevacenes Rotary Encoder Knob Removal Replacement Membrane Keypad Removal Replacement Water Trap Manifold Removal Replacement ni Display Assembly Removal Replacement esssessesseeresreerrerrnerrerisrrerrrerrerrnrrnee Inverter Board Removal Replacement sessessesreesisriesrrsrresrnrinrrnerinrresrrnrinerne Display Screen Removal Replacement Printer Removal Replacement Rear Housing Disassembly 000 Main Board Removal Replacement ssesssesserierieeirerrerinsrrerrnrinrrnerinrrerrrerrsren Connector Board R
130. ice Mode Window Alarm Tone Warning Permanent Alarm Silence ON OFF ENABLE double beep always silent DISABLE full audio alarms full audio alarms In the CONFIG menu the frequency which is rejected is either 50 Hz or 60 Hz Set the frequency to the local AC mains power which is 60 Hz in the U S and 50 Hz in most of Europe Contact your local distributor for more information The password to change the frequency is PIA418 Poet Plus 8100 Service Manual Criticare Systems Inc Section 2 Service Windows Simulation Mode Simulation software is only used for demonstrations and not intended for servicing the monitor The service password is required to activate this demonstration function The words SIMULATION MODE appear in large letters on the monitor screen when in use In the sample window shown below all function parameters are active and reflected in the waveforms and numerical values that could be useful for a demonstration or training session NOTES The data for halothane and nitrous oxide is shown in the lower right hand corner of the sample window Though it is equiped to display anesthetic gas values the monitor must receive that data from an external module or device that is not included with the monitor For more information contact your local CSI dealer or the Criticare customer service department The screen only displays functions that are available on the unit BPM E pO R P Br IBP1 ART 145 105 mK
131. ilter that passes only a narrow band of a determined wavelength of infrared energy The infrared source is modulated at 15 hertz Hz to provide varying energy for the detector and amplifier circuits Correction for temperature and pressure effects on the CO sample is implemented to maintain accuracy over a wide temperature and atmospheric pressure range The galvanic oxygen monitoring sub module contains the control and signal acquisition circuitry for monitoring the O level in the airflow drawn by the capnometer module It is an optional daughter board on the capnometer module No electrical connection is made with the patient from the monitor for oxygen monitoring The temperature module measures probe resistance and uses an internal data table for looking up temperature conversions Depending on the user s selected settings the monitor calculates the temperature in Fahrenheit or Celsius The conversion value displays on the video screen The monitor provides two patient isolated external RS 232 serial data ports TX RX and ground e The DB 9 COM1 port is configured for compatibility with ASCII printer applications e The DIN 8 COM2 port is custom configured with one pin reserved for analog output The main module drives a patient isolated external analog output It provides a 0 to 1 v output capable of driving a high level input such as a defibrillator or chart recorder The Main Board powers and controls the printer modul
132. in Display Processor e IBP 1 2 Processor e NIBP CO2 O Processor e Vital Signs Processor From the Service Mode window 1 Select the processor you wish to download 2 Rotate the menu knob to highlight Software Download and press to select The value NO is highlighted Are you sure NO Figure 2 9 Initial Software Download Dialog Box 3 Rotate the menu knob to highlight YES and press to select The following window appears Are you sure YES Run PC Download Application Figure 2 10 Software Download Response Message If you choose not to select YES allow NO to remain selected and then press the menu knob to return to the Service Mode window The following message then appears next to the Software Download parameter No action taken 4 Press the ON OFF button on the front panel of the monitor to turn off the power to the monitor Criticare Systems Inc Poet Plus 8100 Service Manual Page 2 7 Section 2 Service Windows Permanent Alarm Silence From the Service Mode window 1 To disable the permanent alarm silence feature with the menu knob highlight Permanent Alarm Silence and press to select Enable is highlighted 2 Rotate the menu knob to highlight Disable and press to select 3 Press the ON OFF button on the front panel of the monitor to turn off the power to the monitor NOTE Read about the override capability of the Permanent Alarm Silence function over the Alarm Tone Warning function in
133. in an overall view of the heart s electrical activity three or five electrodes attached to lead wires detect electrical impulses from the patient s heart to the skin The patient monitor calculates the difference in electrical force between two electrode sites Electrode polarity positive negative or ground depends on the cable receptacle the lead wire is attached to and the lead selected on the patient monitor screen The ECG design uses the standard conventional bi polar limb leads leads I Il Ill and uses a 3 lead or 5 lead cable accessory With the 5 lead cable accessory you may also view leads aVR aVL aVF and V lead The monitor has user selectable automatic lead switching capability when using the 5 lead settings If a lead becomes detached or is unmeasurable the monitor can automatically display an alternate lead view using the remaining leads The patient monitor is equipped with pacemaker detection and user selectable pacer rejection There are no known safety hazards due to the operation of a cardiac pacemaker or other electrical stimulators when used with this patient monitor The accuracy of the patient monitor is not affected by arrhythmia or other physiological conditions where the electrocardiogram amplitude and heart rate are within the detectable limits specified for the monitor The patient monitor has user selectable signal filtering in the 60 Hz and 50 Hz bands that reduce electrical interference from the A
134. ing function when the condition occurs IBP Clinical Testing and The invasive blood pressure module is in compliance with applied Accuracy standard IEC 601 2 34 and has been tested to meet the performance specifications listed in this manual The module as installed in the Poet Plus 8100 patient monitor has been clinically tested for performance with a variety of patients Criticare Systems Inc Poet Plus 8100 Service Manual Page 1 7 Section 1 Introduction Capnometry Measurement of CO Method of Measurement Page 1 8 The Poet Plus 8100 patient monitor uses the sidestream method of measuring COs Gas is aspirated through a nasal cannula or a ventilation circuit adapter The gas sample enters from a sampling tube into a water trap which removes water vapor and particulate matter from the gas sample The gas then enters the CO detector where it is analyzed The patient monitor measures CO concentrations and displays them in a continuous waveform The patient monitor also detects end tidal and fraction Inspired CO levels displaying them numerically End tidal CO ETCO 3 is defined as the maximum CO concentration at the end of expiration The patient monitor measures and displays this numerical value of CO concentration The ETCO gt value is updated continuously with each breath cycle The amount of CO in the gas mixture inhaled in by the patient is the fractional Inspired CO FICO3 The patient monitor measures
135. ion displays Squirt gas from the bottle into the system Verify that the parameter display box shows 10 0 4 or 5 0 2 NOTE If the numbers do not meet the specifications go to CO2 User Calibration on page 6 11 to calibrate the monitor for the correct values Perform a user calibration at 150 ml min and at 50ml min Perform the user calibration only if necessary You should have the monitor on and open the Service window 1 Rotate the rotary knob to highlight the CONFIG softkey Press the rotary knob to open the CONFIG menu Rotate and press the rotary knob to select Enter Service Mode NOis highlighted Turn and press the rotary knob to select YES The Enter Service Mode window appears Enter the service mode password PIA418 to enter the service mode Rotate and press the rotary knob to select Enter CO2 02 Service NO is highlighted Turn and press the rotary knob to select YES Verify that the CO2 Operating Mode is in User Mode The display shows a 10 CO reading Connect 10 CO calibration gas Cat No 622 to the monitor sample line Calculate the ideal reading for 10 CO Rotate the rotary knob to highlight CO2 Operating Mode Verify that User Mode highlights If not press and rotate the rotary knob to User Mode and the press the rotary knob The 10 CO2 reading displays Rotate the rotary knob to highlight CO2 Gain Adjustment and press to select Turn the rotary kn
136. ion markings have been altered or removed or any product of any other manufacturer Criticare Systems Inc is not responsible for the effects on safety reliability and performance of the 8100 Patient Monitor if assembly operations extensions readjustments modifications or repairs are carried out by persons other than those authorized by Criticare Systems Inc or the 8100 Patient Monitor is not used in accordance with the instructions for use or the electrical installation of the relevant room does not comply with NFPA 70 National Electric Code or NFPA 99 Standard for Health Care Facilities Outside the United States the relevant room must comply with all electrical installation regulations mandated by the local and regional bodies of government CRITICARE SYSTEMS INC Telephone 262 798 8282 20925 Crossroads Circle Tech Support 800 458 2697 Waukesha WI 53186 Orders 800 458 4615 USA Fax 262 798 8290 Internet www csiusa com Poet Plus 8100 Service Manual Page ix Service Return Policy Return Procedure Page x In the event that it becomes necessary to return a unit to Criticare Systems Inc the following procedure should be followed Obtain return authorization Contact the CSI Service Department at 800 458 2697 to obtain a Customer Service Authorization CSA number Outside the US call 001 262 798 8282 The CSA number must appear on the outside of the shipping container Return shipments
137. ithin 0 1 C for temperature simulator settings of 20 25 30 35 36 37 38 39 40 and 45C Disconnect the Temperature simulator from P3 and verify that the Temperature on the display is Verify TEMP INVALID appears on the display Repeat steps 1 3 for the other connector on the temperature assembly Switch off the power to the Main Board Disconnect all cables Gas Calibration This manual provides a field calibration procedure for the gas module of the monitor The equipment for calibration is available from CSI in the form of two kits that can be supplemented with a few common measuring devices Equipment required for calibrating the COs module Cal Kit Cat No 612 Sample Line 8 feet Cat No 625 Replacement Water Trap WaterChek Plus Cat No 938 Barometer with unit of measure in mmHg and Flowmeter with 50 ml min and 150 ml min capability Additional equipment required for oxygen calibration 95 Oo balance 5 COs Cylinder Cat No 716 Replacement Oxygen Sensor Op Cell Cat No 644 ESD grounding strap required for O cell replacement and Phillips Screwdriver to remove access panel of O cell NOTE Follow the O replacement procedure in Oxygen Sensor Replacement on page 5 9 Page 6 8 Poet Plus 8100 Service Manual Criticare Systems Inc Section 6 Service Testing and Calibration Calibration Kits CAUTION A e Calibration kits contain compressed gas cylinder
138. justment of the power supply to a voltage of 11 8VDC 0 3VDC Two yellow battery icons appear Lower the voltage to 11 5VDC 0 3VDC and verify that two black battery icons appear Continue to decrease the voltage and verify that the monitor shuts down 10 0VDC 0 5VDC Poet Plus 8100 Service Manual Criticare Systems Inc Test Single Battery Criticare Systems Inc 1 Section 5 Preventative Maintenance Attach the single test battery block into slot 1 of the power supply module Attach the other end to a power supply with 12 5VDC Turn on the monitor and verify that the green battery icon appears Lower the voltage with the adjustment of the power supply to a voltage of 12 2VDC 0 3VDC A yellow battery icon appears Lower the voltage to 11 9VDC 0 3VDC and verify that a black battery icon appears Continue to decrease the voltage and verify that the monitor shuts down 10 2VDC 5VDC Attach the single test battery block into slot 2 of the power supply module Remove the first test battery block Attach the other end to a power supply with 12 5VDC Turn on the monitor and verify that the green battery icon appears Lower the voltage with the adjustment of the power supply to a voltage of 12 2VDC 0 3VDC A yellow battery icon appears Lower the voltage to 11 9VDC 0 3VDC and verify that a black battery icon appears Continue to decrease the voltage and verify that the
139. k the SpO value of 52 3 with the analyzer set to 42 1 Page 5 26 Verify that the message SPO2 NO SENSOR appears when there is nothing connected to the SpO connector Verify that the message SPO2 HIGH AMBIENT appears by introducing a higher than normal amount of ambient light on the SPO2 sensor detector Verify the message disappears within 10 seconds after you remove the ambient light Verify that the message SPO2 SENSOR appears when you plug the finger sensor in but without a finger actually inserted into the sensor Use an SpOsz finger sensor and verify that the heart rate and plethysmograph operation display on the LEDs within 15 seconds Verify that SpO error messages do not appear alarm violations can occur depending on individual readings and monitor set up Verify the LCD displays a plethysmograph waveform Connect cable CAT 931A to COM2 Plug BNC A of the cable into a oscilloscope channel In the CONFIG window ANALOG OUT SELECT select PLETH Use the SpOz finger sensor to verify that the SpO waveforms appear on the scope The waveform should be centered on 0 volts and register 1 volt peak to peak Remove the CAT 931A cable Verify that the pulse tone increases as you change the tone from 7 to 10 in the Alarm Setting window Use an optical load to simulate a small signal and verify that the message SPO2 SEARCH or LOST appears after search delay time 5 seconds An optical
140. ld trap side to the water check outside fitting 48 CO Oz Connect the manifold sample side to the water check inside fitting 49 Verify that the ECG cable is inserted into the groove provided at the top of the case with the connector key facing up 50 Secure the front and rear housings together with eight screws Seven Item 47 pn 40496B006 are placed three in the handle two in middle and two in the bottom outside One Item 33 pn 40496B001 at the bottom center 51 CO O Attach the exhaust tubing from the CO bench to the O sensor door pn 42132B001 52 CO Oz Attach the CO absorber to the cord clip PN 42227B001 on the O sensor door 53 Attach the door with two screws Item 29 pn 40995B006 54 Insert a battery Item 40 pn 83278B001 into the unit 55 Add the Hypertronics connector Item 52 pn 83375B001 to the back of the 8100 56 Make sure the label is still attached to the bottom of the unit The label should contain the Cat number for the first line the serial number for the second line and the drawing number for the unit It should look like this 81XYYYZAABCD B E FOR ENGLISH XXX XXXXXX 93951 AXXX 57 Insert screws Item 60 pn 40873B001 into the bottom of the unit These are used for mounting the roll stand Poet Plus 8100 Service Manual Page 7 39 Section 7 Disassembly Assembly Tests Front Chassis Check Off Sheet INITIALS Mount the CO Pump Add the tie wrap holder
141. libration Field Service Testing Safety A A Field Service Test Matrix Field Calibration Procedures Criticare Systems Inc AN WARNING e Service testing procedures require working with exposed electrical circuits and should only be attempted by experienced electrical or biomedical technicians e When a monitor is altered through repair or hardware adjustment it should be fully tested before use A CAUTION A e Always follow ESD precautions when performing any procedure discussed in this section e The manufacturer recommends that a serviced monitor be allowed to run for 24 hours before the monitor is placed back into operation e Modules and PCBs that have been repaired may require more extensive testing than what is described in this manual The pre assembly testing of printed circuit boards PCBs is not covered in this manual Disassembly of surface mounted components on PCBs is not recommended Tests provided here are only for the identification of damaged or degraded PCBs Any time you open a monitor s case you must perform electrical safety tests before you return the monitor to operation If you service the monitor you should also perform the functional tests You should perform additional tests that are specific to modules and assemblies when you service adjust calibrate or disassemble assemblies See the following table Some of the following procedures require specialized equipment that co
142. lip holding the white thermocouple wire Be careful not to damage the tip of the wire 7 Remove the old Power Supply Board Leave the old protective insulator on the board assembly Page 7 30 Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc Section 7 Disassembly 8 Re assemble the new board onto the power module chassis The calibration pot is factory set 9 Re attach all wires and cables to the Power Supply Board Reassemble the power module 10 Replace the board by following all of the steps sections before this to reassemble the unit 11 Replacement or repair of the power module requires Hi Pot and Leakage testing See Service Testing and Calibration in Section 6 for full test procedures Poet Plus 8100 Service Manual Page 7 31 Section 7 Disassembly Reassembling the Monitor Page 7 32 Reassembling the Front Chassis NOTE Every step applies for a base unit All other steps are identified by COs COo Op or IBP Refer to drawing number 92788AXXxX in Drawings and Schematics in Section 9 of this manual for the following steps 1 O Ko CO CO2 Oz Mount the CO Pump Item 1 pn 94253A001 to the Front Chassis Item 2 pn 42131B001 using two 6 Screws Item 3 pn 40496B002 NOTE Do not over tighten CO2 CO2 Oz Add Tie Wrap Holder Item 9 pn 40294B001 to the bottom of the CO2 pump Mount the NIBP Module assembly Item
143. ly CSI service personnel are authorized to perform component level repair e Possible safety hazard Read the battery safety information provided in Safety on page 1 17 CAUTION A e Opening a monitor and breaking the quality seal can void your manufacturer warranty Before breaking the seal on a CSI monitor contact the CSI service department The repair procedures for the monitor are included here for the determination of damaged or unusable assemblies CSI does not recommend attempting the field repair of the circuit boards This monitor requires the installation of a factory calibration lookup table in the CO gas module This lookup table is specific to the bench and gas module assembly The factory calibration of the gas module cannot be performed by the user Service Safety DANGER A e The monitor contains an internal AC power supply that contains potentially lethal voltages e High Voltage The internal Power Supply Board has a protective shield that should not be removed during service If the power supply is defective replace the entire Power Supply assembly Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 1 Section 7 Disassembly Electrostatic Discharge The procedures in this section require handling of electrostatic Protection sensitive components Microprocessors and other electronic components can be permanently damaged by attempting repairs at an unprotected workbench A A Use
144. mits maximum possible cuff pressure to 332 mmHg in adult pediatric mode and 166 mmHg in neonatal mode Cuff pressure is allowed to remain above 30 mmHg for a maximum of 2 minutes The patient monitor automatically deflates the cuff if the time limit is violated The patient monitor contains hardware protection for overpressure conditions pressure transducer failures or microprocessor and pump control circuit failures 140 B P Cuff Inflation Pressure Shown during ia Systolic Pressure 110 N pie Actual Blood 00 U ii Pressure Waveform 90 Diastolic Pressure lt Cuff deflates rapidly after J the monitor determines A the systolic pressure Pressure in mmHg Pulse Waveform Measured from B P nt RNA nea Cuff Pressure Fluctuation Diastolic Mean Systolic Pressure Pressure Pressure Figure 1 1 NIBP Cuff Pressure and Pulse over Time Page 1 6 Poet Plus 8100 Service Manual Criticare Systems Inc Section 1 Introduction Invasive Blood The IBP monitoring module is intended for direct pressure monitoring Pressure IBP using an intra arterial catheter a continuous flushing system and an IBP transducer The IBP module provides analog to digital conversion of the transducer signal The IBP module processes continuous digital patient data that can be displayed as waveform and numerical values The module supports up to two separate IBP channels simultaneously For arterial and pulmonar
145. moves Press the encoder and verify a menu is selected Press the encoder again to exit Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 43 Section 8 Troubleshooting This section lists the possible causes of monitor problems Use this table and the schematics in Drawings and Schematics in Section 9 to identify and locate components that malfunction Symptom Problem Solution Unit won t power up On Off LED is dark Battery is discharged AC power cord is not securely connected to monitor AC outlet off or unpowered Fuses are blown Located above AC connection Main Board is disconnected Bad membrane Main Board fuse blown Battery fuses are blown Connect AC power cord Connect AC power cord to monitor Connect power cord to a live outlet Replace AC fuses pn 80131B002 Reconnect DC power cable at PS2 on Main Board Replace the keypad membrane Replace F3 on Main Board Replace F1 and F2 on Main Board Cannot activate hardkey function or turn on moni tor with AC power On Off LED is green Membrane panel is disconnected or loose Keypad membrane is bad Reconnect ribbon cable to J7 on Main Board Replace keypad membrane panel Display is blank or not readable Video screen cables to Main Board are disconnected or loose Inverter not plugged in Bad inverter Reconnect data and inverter cables to connectors on Main Board J2 J3 J4
146. munication Testing Perform the following tests of the output ports of the monitor Video Port Use the external video cable and connect it to the DB 15 port COM2 Port Analog Port 1 Criticare Systems Inc r i 2 e e x Em DB 15 a Port i CRT monitor specification IBM VGA compatible 640 x 480 RGB color Analog monitor 75 ohms impedance Vertical Sync Freq must support 60 Hz 5 and 85 Hz 5 Horizontal Sync must support 31 5KHz to 45KHz 15 pin DB connector input Turn the monitor s power on Select the DISPLAY softkey When the Display menu appears select External Display and select ON Verify that the port communicates to the external video display The on screen image of the vital signs monitor should match the external video display NOTE External video displays do not produce AMMI compliant waveforms graphics due to the variety of video aspect ratios To check the COM2 port configure the monitor for SoO monitoring Select and press the CONFIG softkey scroll down to Analog Out Select on the Config menu and select ECG II Connect a DIN 8 to BNC cable pn 90840A002 to the COM2 port of the monitor Attach the other end of the cable to an oscilloscope In the CONFIG window select Analog Out Select on the Config menu and select PLETH Using an SpOz finger sensor verify that the SpO waveform appears on the scope Poet Plus 8100 Service Manual Page 5 31 Section 5
147. n neutral lt 10 microamps Pass ____ Leads to ISO position Normal polarity ISO Test lt 50 microamps Pass ____ Normal SpO z ISO Test lt 50 microamps Pass ____ Temp Cable Normal polarity ISO Test lt 50 microamps Pass ____ IBP Cable Normal polarity ISO Test lt 50 microamps Pass Ground Continuity Pass____ Functional Testing Interface Inspection Pass ____ Manual Controls Check Pass Alarm Verification Low Level Alarm Pass ____ Medium Level Alarm Pass ____ Alarm Off Indicator LED Flashes Pass ____ Alarm Volume Control Pass ____ Printer Check Pass____ Page 5 36 Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance Monitoring Module Verification ECG Verification ECG Analog Output Diagnostic Monitor and ST Filter Modes SpO Verification NIBP Verification Check Adult Cuff at 120 80 Check Neonatal Cuff at 80 50 Temperature Verification Gas Flow CO Oz Verification Flow Calibration Occlusion Test Exhaust Test CO Test O Test IBP Calibration Communication Testing Video Port COM2 Port BNC ECG Output Nurse Call Battery and Power Test Charger no load output 13 6 VAC 5 AC power LED Battery power operation CERTIFICATION OF FUNCTIONAL AND SAFTEY TESTING TECHNICIAN This checklist may be copied as needed to record field testing Page 5 37 Poet Plus 8100 Service Manual Criticare Systems Inc Section 6 Service Testing and Ca
148. nc Section 8 Troubleshooting Troubleshooting and Before attempting to repair a damaged monitor the technician must Repair fully understand the functions of the monitor The service technician should have read and understood all the previous sections of this manual You must fully repair calibrate and test all damaged monitors before you use them on patients The monitor must meet the original operating specifications before you use it on a patient rae Always follow ESD precautions when performing any of the procedures discussed in this section ComfortCuff NIBP The NIBP module consists of a pneumatic pump assembly Module pn 95576A003 and control circuitry located on the Main Board pn 91340003 The pneumatics assembly is mounted on the back side of the metal chassis plate A large diameter tube pn 41602B001 connects to the Quick Connect hardware fixture A small diameter tube connects to the transducer on the Main Board The transducer port faces out the back side of the Main Board through a hole The remaining pneumatics links connect the two control valves pn 95596A002 the pump pn 94252A001 and the check valve pn 41579B003 As the tubing ages it may lose its seal Old tubing can be re seated or replaced as necessary When any major pneumatic component such as a pump or valve is replaced the NIBP module must be re calibrated as in NIBP Calibration on page 6 5 of this manual The manufacturer recommends th
149. nd IEC 1000 4 5 for conducted transients and operates with negligible adverse effects X RAY The monitor operates with negligible adverse effects in an x ray environment However the monitor should not be placed directly in the x ray beam which could damage the internal electronics of the monitor OTHER INTERFERENCE There is a negligible adverse effect to the monitor from electrocautery and electrosurgery infrared energy and defibrillation Do not use this device in conjunction with flammable anesthetics such as cyclopropane and ether The monitor can sample from pure oxygen environments but the monitor itself should never be placed inside an oxygen tent or gas containment apparatus Proper anesthetic gas waste recovery should be used All patient contact or user contact materials in this monitor and it s accessories have passed ISO 10993 5 10 amp 11 biocompatibility tests or have been in use in clinical environments in large numbers over an extended period of time predating these standards All Criticare Systems Inc products including patient monitors and accessories are free from latex in any location that may result in patient contact Poet Plus 8100 Service Manual Criticare Systems Inc Section 2 Service Windows Service Mode Window The monitor has a Service Mode window that is accessible through the CONFIG window of the monitor WARNING A e Never service a monitor while it is attached to a patient
150. nect the NIBP Pump Valve cable assembly at P2 on the Main Board Poet Plus 8100 Service Manual Page 7 15 Section 7 Disassembly 10 Label and disconnect the thermister cable J18 on the Main Board NOTE J18 and J20 use similar connectors Plugging these in the opposite connections results in a Main Board fault 11 Label and disconnect the fan cable at J19 on the Main Board 12 Label and disconnect the speaker cable at J20 on the Main Board NOTE J18 and J20 use similar connectors Plugging these in the opposite connections results in a Main Board fault 13 Remove the blue tygon tubing from SENS1 on the solder side of the Main Board SENS1 itself appears on the component side The connection to SENS1 is on the solder side SENS1 Tubing Connector on solder side Figure 7 31 SENS1 Tubing 14 Replace in reverse order Page 7 16 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly Connector Board Removal To remove the Connector Board Replacement 1 Remove the two Phillips screws holding the Connector Board to the unit FY E a o za keta o A t AE a aa S Soe EE E A a Figure 7 32 Remove the Connector Board Screws 2 On monitors with IBP label and disconnect the ribbon cable at P1 on the Connector Board 3 Label and disconnect the temperature cable from P3 on the ECG Analog Board 4 Disconnect the NIBP tubing from the NIBP fitting 5 Remove the nut from
151. nit has IBP you need to remove the IBP Board See IBP Board Removal Replacement on page 7 26 Figure 7 46 Remove CO O Module 2 Remove the pump valve assembly v PA Figure 7 47 NIBP Pump 3 Disconnect the clear tubing from the NIBP connector To remove the entire NIBP assembly expose the screws holding it to the chassis 4 Remove the four bushings that are on the chassis Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 27 Section 7 Disassembly 5 Pull back the insulator from the upper left corner to expose and remove the two Phillips screws Ss nd Figure 7 48 NIBP Assembly Screws 6 Remove the NIBP pump assembly from the chassis Pump with tie Valves Figure 7 49 NIBP Pump See Troubleshooting in Section 8 for more information about troubleshooting the NIBP assembly 7 Reassemble with a replacement NIBP pump assembly in reverse order Page 7 28 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly Power Module Assembly 1 Turn the unit over and remove the six Phillips screws four small Removal Replacement tan colored screws and two larger unpainted screws from the bottom of the unit b D i Figure 7 50 Remove the Bottom Screws 2 Turn the monitor upright and tilt the power supply module out of the rear assembly You need to carefully lift the power supply assembly from the standoffs by
152. nual A limited number of procedures are approved for field service Follow the disassembly procedures as described in Disassembly in Section 7 of this manual After you reassemble and calibrate the monitor you must perform safety testing as described in Service Testing and Calibration in Section 6 Replacement of any of the following CO components must be followed by a gas field calibration as described in Gas Calibration on page 6 8 of this manual You may replace the CO absorber pn 40570B003 without affecting the factory settings of the module If the CO absorber is degraded the monitor may not calibrate properly The COz absorber is a 2 inch long cylinder located along the left edge of the CO Board or mounted on the rear access panel Remove and replace the CO absorber cylinder as necessary One end of the absorber must be open as a sampling port during calibration A field calibration of the gas module must be performed to ensure that the module was reassembled correctly Poet Plus 8100 Service Manual Page 8 5 Section 8 Troubleshooting CO Pneumatics CO Pump Replacement CO Module Replacement Main Board Page 8 6 Obviously damaged leaking or degraded tubing can be replaced without affecting the factory settings of the CO module See drawing 95628A001 in Drawings and Schematics in Section 9 of this manual for details on the pneumatics assembly e For monitors with COs and O
153. o Port External Module Port Mechanical Electrical Weight Size Mechanical Shock Vibration Power Requirements Voltage Number of Batteries Battery Life Recharge time Environmental Operating Temperature Storage Temperature Operating and Storage Humidity Type of Protection Degree of Protection Protection against ingress Section 1 Introduction 10 4 active color TFT 640 x 480 pixels 6 maximum 0 5x 1x 2x 4x user selectable 6 25 12 5 25 or 50 mm sec selectable 9 membrane activated Push and rotate 24 steps turn Eng Fr Ger It Port Span RS 232 compatible digital DB9 COM 1 mini DIN8 COM 2 mini DIN8 Selectable waveform output Range 2 to 3 volts ECG Waveform 1 volt signal Plethysmograph or Capnogram Contact switch audio jack 1 8 inch 24V 100 ma maximum switching BNC connector Serial VGA Compatible Hypertronics connection 12 Ib 5 45kg 12 H x 10 5 W x 7 5 D 30 5cm H x 26 7 cm W x 19 1cm D Negligible effect up to 40G Negligible effect up to 0 5G at 200Hz 35W typical 100 240 VAC 50 60 Hz Up to 2 sealed lead acid batteries will operate with one battery 1 hr typical one battery without using gas feature 4 5 hours 59 95 F 15 35 C 23 122 F 5 50 C 15 to 90 non condensing Class Equipment Type CF Defibrillator Proof Ordinary Complies with the ingress protection test specified in EN 6
154. ob clockwise to increase the CO2 reading or counter clockwise to decrease the reading Adjust for a CO reading of 9 7 to 10 1 Press the rotary knob to save the gain Poet Plus 8100 Service Manual Page 6 11 Section 6 Service Testing and Calibration 8 Once you have adjusted the COs press the rotary knob then turn the rotary knob until until CO2 Zero Cal is highlighted Press the rotary knob and select Abort CO2 Zero Cal 9 Press the knob then turn the rotary knob one click counter clockwise to select Request CO2 Zero Cal and press the rotary knob again NOTE After the first zero cal request you must first abort before you request a zero cal Every time you make an adjustment you must request a zero cal 10 Once you finish the zero cal turn the rotary knob until 50m min Flow Rate is highlighted Press the rotary encoder turn the rotary knob one click counter clockwise and press the rotary knob again The monitor is now in the 50ml range 11 Repeat steps 7 through 10 for the 50ml flow rate 12 Adjust for 10 Request a zero cal if you are satisfied with the number after the zero cal Go back to the 150ml flow and flip back and forth between the two flows If no adjustment is needed you are ready to exit the service screen Return the flow to 150ml when you are ready to exit the service screen Verify that nothing is highlighted then power cycle the monitor off and on O Calibration NOTE To respirate turn the r
155. ode Parameters Window 3 Rotate the menu knob to highlight the test you want to run and press to enter The test value NO is highlighted 4 Rotate the menu knob to highlight YES and press to begin the test When the test is executing the following message appears on the bottom of the NIBP service window Test in progress select YES to cancel 5 Press the ON OFF button on the front panel of the monitor to turn off power to the monitor Page 2 4 Poet Plus 8100 Service Manual Criticare Systems Inc Section 2 Service Windows CO O Service Mode From the Service Mode window NOTE To service the CO2 O2 module you need specified equipment See Section 6 for details 1 To calibrate and adjust the gas module of the monitor with the menu knob highlight Enter CO2 O Service and press to select The value NO is highlighted 2 Rotate the menu knob to highlight YES and press to select The Service Mode window appears to provide functions and settings to calibrate and adjust the CO and Op modules of the monitor The window is shown below c02 Operating Mode User Mode C02 Zero Cal Abort C02 Zero Cal C02 Gain Adjustment Turn to Incr Decr Press to Save 50 ml min Flow Rate Turn to Incr Decr Press to Save 150 ml min Flow Rate Turn to Incr Decr Press to Save c02 Factory Defaults No Action 2 Gain Adjustment Turn to Incr Decr Press to Save 02 Wave Response POT 1 Turn to Incr Decr Press to Save 02 Wave Response POT
156. odule For verification refer to SpO2 Verification on page 5 26 2 Attach the monitor to the SmartSat 3 Start the SmartSat programmed sequence SmartSat Auto Sequence Verify that the monitor s reported SpO gt values are within the limits specified Level Saturation Limits Heart Rate Limits 1 99 97 99 40 bpm 39 41 bpm 2 96 94 98 60 bpm 59 61 bpm 3 90 88 92 80 bpm 79 81 bpm 4 78 76 80 100 bpm 99 101 bpm 5 61 58 64 120 bpm 119 121 bpm 6 52 49 55 180 bpm 179 181 bpm 7 40 37 43 190 bpm 297 191 bpm Poet Plus 8100 Service Manual Page 6 3 Section 6 Service Testing and Calibration 4 Even if the monitor fails only one level rerun the sequence after 30 seconds Only if the monitor is successful the second time do you pass the monitor 5 If the monitor fails again contact the CSI service department SpO gt Board Test Values While the DOX SpO Board is in operation it should conform with the following test values The operational current draw of the board should be less than 50 ma Use TP4 as ground for each measurement Test Point Signal Value TP3 5A 5V 0 05V C9 10 10V 0 20V TP2 5D 5V 0 05V VR1 4 2 5VREF 2 5V 0 03V 1 If the any of these values fail to test correctly isolate the problem to either the SpO Board or the power supply circuits 2 Remove the SpO Board from the monitor 3 Test the equivalent signals on the ECG Analog Board circuits
157. or 50ml min flow The following parameters can be used with the Poet Plus 8100 e READING in MMHG CO2 gas X barometric pressure in mmhg e READING in CO2 gas 0 4 for 10 0 2 for 5 O TEST 1 Attach 100 nitrogen to the sample line The O value should go to 1 or 0 This indicates there are no leaks in the sample cell side of the pneumatics NOTE You must be in the Service Mode window to see this happen The Service Mode password is PIA418 Verify proper operation of the Os Use 10 COs balance No gas Simulate a breath rate between 10 20BPM for 150ml min flow and 5 10BPM for 50ml min of flow The unit should give a reading in 2 3 breaths Accept if INS Oz is less than 3 and the EXP is between 19 and 23 Poet Plus 8100 Service Manual Page 5 29 Section 5 Preventative Maintenance IBP Verification The main focus of the following steps is the MED SIM 300 or its equivalent 1 Attach the IBP cable from BP1 on the simulator to IBP 1 on the connector panel Select BP on the simulator Select Channel 1 Select STAT Select ZERO by using the up and down arrows on the MED SIM 300 scroll through the menu until STAT appears and press to select Rotate the rotary encoder knob to highlight PARAMS and move down to IBP1 Press the menu knob once Set the SITE for ART Press the rotary encoder knob again Adjust the RANGE This should be set for 0 200 Exit out of
158. ow fuses 120 volt units Check the monitor ID label to determine which voltage the unit uses You do not have to open the monitor case to change the main power fuses See Fuse Removal Replacement on page 5 7 WARNING e Shock hazard Substitution of fuses or jumping the fuse contacts can endanger the patient and damage the monitor The fan pn 95619A001 is designed to turn on only when the monitor is warm The monitor may operate for more than an hour before it becomes warm enough to activate the fan Much of the heat generated will be due to charging the batteries If the bottom rear housing becomes hot to the touch the fan may not be operating properly Turn off the CO module to shut off the pump and listen for the fan motor The Main Board and a thermal sensor controls the fan The thermistor wire pn 90875A001 extends from the Main Board to the power module The wire is clipped to the Power Supply Board Check the wire for breaks damage or poor connection to the Main Board The end of the wire should be coated If the tip is broken replace the wire Poet Plus 8100 Service Manual Page 8 7 Section 8 Troubleshooting Power Supply Replacement AC Power Components Page 8 8 AN You can test the fan directly with a 5 volt DC source Replace the fan if necessary Fan replacement requires you remove the power module The DC output from the power supply enters the Main Board at PS2 Check the cable for
159. owels or tissues can scratch the surface of the display use a cotton cloth to clean the monitor Do not use full strength alcohol on the display screen Repeated use of strong cleaners damages the screen Clean the display window by wiping it clean with a soft lint free cloth sprayed with common glass cleaner hwoever do not spray glass cleaner directly on the display CAUTION A e Do not immerse any Criticare pulse oximeter sensor connector in any liquid because it damages the connector The SpOz sensor can be wiped clean with alcohol Disinfect the SpO sensor by placing the paddles and cable in a 2 glutaraldehyde solution Only the sensor paddles and cable are placed in the solution You can clean the reusable blood pressure cuff by wiping it with a damp cloth or sponge If necessary disinfect the cuff by wiping it with 70 alcohol mild bleach solution or other disinfectant Disposable blood pressure cuffs are for single patient use and are not intended to be disinfected You can sterilize the cloth cuff and neoprene bag with commercially available disinfectants such as ethylene oxide ETO Rinse thoroughly to remove residual disinfectants Do not allow liquids to enter the neoprene bag The cloth cuff can also be sterilized in an autoclave Poet Plus 8100 Service Manual Page 4 1 Section 4 Cleaning and Disinfecting Page 4 2 ECG Cable Temperature Cable If the cuffs become grossly soiled with blood or othe
160. r the more damage it can do It is important to service the monitor immediately after any liquid is spilled into it Page 4 4 Poet Plus 8100 Service Manual Criticare Systems Inc Section 5 Preventative Maintenance Incoming Inspection You must inspect monitors coming back from service for shipping damage before you place them into operation just like newly purchased monitors The monitor should be free from dents cracks or other physical damage The quality inspection seal of the monitor should be unbroken indicating that the monitor has been tested according to manufacturer s specifications If further incoming inspection or testing is required the manufacturer recommends that you use the Alarms Verification located in this section as an incoming inspection test You can perform additional electrical safety testing in this section as part of an incoming inspection in accordance with the policies of the health care provider Maintenance Schedule Every Patient e Clean and disinfect the sensor cables and sensor e Inspect the accessories and cables for damage e Change the gas sampling device and sampling line Every Day e Charge the monitors batteries Every Week e Change the water trap Or change as needed Every 3 Months e Clean the exterior of the unit or clean as needed e Check the O cell Change if necessary Every Year Perform the annual safety tests provided in this section Verify the CO auto c
161. r body fluids you should launder the cloth cuffs by hand or machine You can launder or sterilize the dacron cloth cuff by first removing the neoprene inflation bag Feed the inflation tube back through the hole and then pull out the cloth flap Flap Figure 4 1 Remove Inflation Bag from Cuff Roll up the inflation bag and slide it out the open slot in the cloth cuff The following laundering precautions apply only to non disposable cuffs You should not launder disposable cuffs and neoprene inserts e Remove the inflatable bag from the cuff before laundering or sterilizing the cuff Close the Velcro fastener before you launder the cuff e Strong bleach solutions damage the cuff e Temperatures over 275 F 135 C damage the cuff e Soaking the cuff in dark colored solutions may stain or discolor the cuff Hand laundering as opposed to machine laundering prolongs the life of the cuff Wash the cuff in warm soapy water Rinse the cuff thoroughly After you clean the cuff allow the cuff to air dry then insert the inflation bag in the cuff You can clean the ECG cable and leads with alcohol Do not immerse the connections Clean the cable according to the hospital protocol for cleaning of reusable equipment cables Typically this protocol consists of the following steps 1 Disconnect the cable from the monitor and temperature sensor 2 Wipe the cable with a nonabrasive cloth moistened with a mild detergent and wa
162. rawings and Schematics Display Assembly 95605A003 AY DISPLAY 10 4 NEC Item CSI Part Description 01 45119B002 DISPLAY BEZEL FRNT POET 02 45120B001 DISPLAY BEZEL REAR 03 83342B002 DISPLAY GASKET 05 42166B001 DISPLAY MNT UPPER 06 42167B001 DISPLAY MNT LOWER 07 83313B002 PLATE FERRITE MOUNT 10 4 09 81511B001 LCD ACTIVE NEC 10 83265B002 BACKLIGHT INVERTER 12 90859A002 AY CBL MAIN BRD INVRTR 2 14 90879A001 AY CBL ACTIVE LCD TO MAIN 15 95643A002 AY CBL DISPLAY GROUND 16 42127B001 WINDOW FILTER 10 4 18 42006B001 SPACER 093ID X 1880D X 20 40008B006 NUT HEX 10 32 PL 22 42183B001 H H M S 10 32 X 625 24 40496B001 P H M S 6 32X 375 SEMS 25 40995B003 P H M S 4 40X 500 SEMS 26 42338B001 PHMS 2 56 X 312 NY PATCH 27 40995B001 PH M S 4 40X 312 SEMS 28 40294B002 HOLDER TIE WRAP ADH BK LG 29 42182B001 WASHER NYLON 2 33 83148B001 FERRITE BD 2000HMS 100 36 40067B001 CABLE TIE 4 0 LG NYLON 37 40067B003 CABLE TIE 8 LG NYLON 40 83338B002 DISPLAY FILM 41 40132B001 TAPE MICROFOAM ADH BACKED Page 9 26 Poet Plus 8100 Service Manual Criticare Systems Inc Power Module Assembly Criticare Systems Inc Section 9 Drawings and Schematics 95612A001 AY POWER MODULE CSI Part 91352A001 80232C001 83276B001 42148B001 83275B001 83274B002 83273B001 83272B001 95619A001 42145B001 90865A001 90875A001 42191B001 40995B005 40995B006 40995B010 40284B001 40067B001 42204B002 42208B001 40294B001 83322B001 Description AY PC
163. rd A002 or better use printer assembly Item 18 pn 95604A002 31 Use the two screws Item 29 pn 40995B006 to secure the printer assembly to the rear housing 32 f the unit is a non printer place two screws Item 29 pn 40995B006 through the cover printer opening pn 45114B001 33 Turn the unit back around with the circuit boards facing you 34 Verify that all the ferrites are between the window of the chassis 35 Make sure all pneumatics are left hanging out There should be only two tubes on monitors without CO on monitors with CO2 there should be 4 tubes Poet Plus 8100 Service Manual Page 7 37 Section 7 Disassembly 36 Place the Main Board down with J11 BNC upwards and to the right Most of the connectors face upward Connect the following to the Main Board a Connect the speaker cable to J20 b Connect the fan cable to J19 c Connect the thermister cable to J18 d Connect the CO cable to J16 If monitor has COs e Connect the NIBP cable to P2 f Connect the IBP cable from the P1 connector by the fuse on the IBP Board to J14 Connect the other IBP cable from the P2 connector on the IBP Board to the Connector Board g Connect the Power Supply cable to PS2 NOTE Do not route this cable through the window h Connect the Battery cable to PS1 i Connect the blue tygon tubing to the pressure transducer at SENS1 NOTE Make sure that when you place the Main Board into the unit that thi
164. ribbon cable with the ferrite case is held on with double sided foam tape Carefully pull the ferrite casing away from the board 8 Disconnect the ribbon cable at P1 on the SpO Board THOS OD gt 6 e 6 6 amp amp KIA He aon a OF ro CRITICARE SYSTEMS INC 91339A001 REV p MADE IN USA Figure 7 35 P1 on the SpO Board 9 Clean away the old foam tape Use new foam tape to reattach the ribbon cable assembly 10 Reverse these steps to place a new SpO Board in the monitor and reassemble the monitor Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 19 Section 7 Disassembly Front Chassis Removal The Main Board is attached to a metal chassis along with the vital signs monitoring modules You can remove these components as an assembly NOTE Criticare advises that you remove the Main Board prior to removing the Front Chassis See Main Board Removal Replacement on page 7 14 1 Remove the printer from the monitor See Printer Removal Replacement on page 7 13 2 Remove the two Phillips screws hidden beneath the printer module A Screws Figure 7 36 Remove the Chassis Screws 3 Remove the two Phillips screws that hold the O sensor door 4 Remove the exhaust tubing and the CO absorber from the O sensor door See Oxygen Sensor Replacement in Section 5 Exhaust Tubing J el E Figure 7 37 Remove the Exhaust Tubing
165. rm should be a high priority alarm tone consisting of 3 beeps followed shortly by two beeps Inthe PARAMS softkey window change the SpO2 Low Limit Alarm setting to MEDIUM Verify that the message LOW SPO2 appears at the top of the waveform channel in yellow letters Verify that you get an audible response The alarm should be a medium priority alarm tone consisting of 3 beeps ALARM VOLUME TEST 1 From the ALARMS window vary the ALARM VOLUME through the ranges from OFF to 10 Verify that the speaker volume changes according to the setting of the ALARM VOLUME Press the alarm SILENCE key once Verify that the low heart rate alarm stops for 2 minutes and that the 2 minute Alarm Silence icon appears in the system status area of the screen Press and hold the alarm SILENCE key for at least 2 seconds Verify that the low heart rate alarm stops and the Alarm Suspend icon appears in the system status area of the screen Printer Check The printer only prints if the green battery light is ON or the AC mains cord is plugged in 1 Lift the paper feed lever up and insert paper Close the lever and the paper should automatically feed 2 Set the printer for INTERNAL PRINTER in the CONFIG window Set print type for TABULAR in the PRINT window Press and hold the PRINT key on the front of the monitor 3 Verify that the correct date and time print Page 5 20 Poet Plus 8100 Service Manual Criticare Sy
166. rm water or a disinfectant 3 Dry thoroughly Poet Plus 8100 Service Manual Criticare Systems Inc Section 4 Cleaning and Disinfecting 4 Do not use alcohol or solvents to clean the cable 5 Do not allow the cable connectors and contact points to come in contact with liquids 6 Do not fully immerse the cable in liquids 7 Do not autoclave or ETO sterilize the cable Criticare Systems Inc Poet Plus 8100 Service Manual Page 4 3 Section 4 Cleaning and Disinfecting Accidental Wetting WARNING A e Shock Hazard Because a monitor is an AC powered device an immersed monitor presents a danger to anyone who attempts to handle the device Immediately perform the following steps following an accidental wetting of equipment 1 Turn the power off Disconnect the AC power cord from the monitor 2 If the monitor is monitoring a patient transfer the patient to another monitor as quickly as possible 3 Use a clean dry towel or cloth to remove the liquid from the monitor housing 4 A service technician should inspect the monitor as soon as possible 5 If the internal mechanism is saturated allow the liquid to drain out for 24 hours before shipping 6 If liquid has entered the monitor it must be dried and cleaned internally Full testing is required before you can use the monitor Contact the CSI service department as soon as possible Time is critical The longer any liquid remains in the monito
167. rnrrinsrnerinernetrnrrneee 5 19 Alarms Verification coi ciestev dnates detec devaveccestvesutvescevacetuahaveeoses cduncncrepesteevdnecdis coat 5 20 Printel CHECK A cass canst faxectsepscuesntue codheintantetoiaecusens2intasdesuteenhdese satectesstaansed 5 20 Module Verification sssrinin a daiar 5 21 ECG Verifications a a a T N A TAS 5 21 ECG Analog Output Test vc se ccccccccsescteossntesecuceerseveuaverscicestcensttecntvosooresreentsvsees 5 24 Diagnostic Monitor and ST Filter Modes Test cececeeeeeeeeeeseeeeeeeeeeeeeee 5 25 SPOog Verification sececceeescesqeseesersseeenseeneseaeeneescasatsecesaesesseraseesaetasueaeanenten 5 26 NIBP Verification Temperature Verification cece cee cceeeenee seas ceeeeseaeseeeesseeesaeeseaeessaeeeaeenaes 5 28 Gas Flow COo Opo Verification 0 0 cece esses cee ceeeeseseeeesceeseeseeaseeseeaseesseseaes 5 28 IBP2V GrifiCertion ss iceiseiseccde eae tisk oteaeevacsnsts cvenisubdesescvardssavcdsussesitusacbessiestecneetoduneye 5 30 Communication Testing 0 ccc cece ene cee cess cesses eseeeeeeseeesseessaeseeeseneeseeeseaeees 5 31 VIGGO RO sg a e a a AANA SEARED 5 31 COM2 Port Analog Port sssssseessssssiesiseinesirrinrrinrininsrrerinsrretnrinernntrnrnnrinrrnerrent 5 31 BNC ECG Output ianaeaieo i pia 5 32 NU PSO Cale iinta a A rr A AEAEE 5 32 Battery and Power TeSting cscccsscssscesscscsscesssesssescseesesessssesescesseessseseseessneenseaesees 5 33 Eq ipment egaa aaaea ea A
168. rough Cap assembly 65 Solder ground from SpOz board to Analog board 66 If TEMP NIBP place it on the left 67 Install the correct connector assembly 68 Mount the plate and the connector board and attach cable to ECG Bd 69 Mount The IBP connector assembly to the rear housing 70 Mount board for 2 temp units 71 Connect GND cable to NIBP fitting and nut 72 Connect interface cable from the SpO to the Analog board 73 Attach Tie wrap to CO cable 74 Attach the printer cable to printer assembly 95604A001 only 75 Attach the printer cable to Printer assembly 95604A002 only 76 Secure printer assembly with two screws Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 41 Section 7 Disassembly Final Assembly Check Off Con t INITIALS 77 1f unit is non printer attach non printer door 78 Turn unit around 79 Verify that all the ferrite are between the window of the chassis 80 If CO is applicable make sure both tubing s are to the left 81 Place Main board face down with J11 facing upwards Verify that the speaker is attached J20 Verify that the fan cable is attached J19 Verify that the thermister is attached J18 Verify that the CO cable is attached J16 Verify that the NIBP cable is attached P2 Connect IBP to J14 Connect IBP to connector board Verify that the AC cable is attached PS2 Verify that the Battery cable is attached PS1 Verify that the blue tygon is attached to pressure xdr 82 Secure main board with six scr
169. rs CONFIG window select Enter Service Mode and then select Yes Enter the correct password PIA418 in the Enter Service Mode dialog box 2 Verify that the software revision of the IBP Board appears on the screen 3 Select Enter IBP Service and answer Yes The IBP factory calibration window appears 4 Verify current contents of the factory calibration CAL window as follows Select Channel Number 1 Start or Restart Cal NO Set 0 mmHg Point NO 0 0 Set 256 mmHg Point NO 838 8 Store CAL Results NO Pressure Reported X where X is currently interpreted input level can be Current CAL Values in EEPROM OFFSET GAIN IBP 1 0 0 838 8 IBP 2 0 0 838 8 NOTE If a Main Board from a different unit has been used for the IBP calibration the offset and gain numbers are different and the cal is stored on the Main Board chosen The main focus of the following steps is the BP 600 or its equivalent 5 Insert the BP 600 into the top IBP connector Set the BP 600 to a pressure of 256MMHG and the switch set at the ZERO position Select Channel 1 6 For calibration select Start or Restart Cal and then select Yes 7 Allow the pressure to stabilize before you select YES to the step below Select Set 0 mmHg Point and then select Yes NOTE A set of corrective offset values appears at the end of the line Examples are 11 and 2 Criticare Systems Inc Poet Plus 8100 Service Manual Page 6 15 Section 6
170. s Read and obey all precautions listed on the gas cylinder labels The Cat No 612 kit is for monitors with CO only The Cat No 715 kit provides an additional gas canister for calibrating the Op module Items from the Cal Kits can be ordered separately CAT NO 612 Supplied with this Cal Kit e 5 COs balance No Cylinder Cat No 621 e 10 COs balance No Cylinder Cat No 622 e Regulator Cat No 623 e Regulator Tubing Cat No 613 CAT NO 715 for monitors with oxygen monitoring Supplied with this Cal Kit e 5 COs balance No Cylinder Cat No 621 e 10 COs balance N2 Cylinder Cat No 622 e 95 O2 balance 5 COs Cylinder Cat No 716 e Regulator Cat No 623 e Regulator Tubing Cat No 613 Before you Begin Make sure the CO and O calibration gas cylinders are not depleted or near depleted the pressure meter should be above zero and you can hear a hissing sound when in use WARNING A e Never attach intravenous tubes to gas sampling connections because gas sampling lines could inadvertently be connected to intravascular fluid systems thus allowing air into a blood vessel CAUTION A e Do not use if the CO and or O calibration gas cylinder is beyond its expiration date Criticare Systems Inc Poet Plus 8100 Service Manual Page 6 9 Section 6 Service Testing and Calibration Page 6 10 CO Flow Check Perform a flow check at 150 ml min and at 50ml min Perform the user cali
171. s tubing is clear from being kinked Route the tubing above the battery cables 37 Secure the Main Board to the chassis with the five screws Item 33 pn 40496B001 NOTE Make sure the connector for ECG is secure The left side middle is not installed until later 38 Attach the same screw through the Cap assembly on the Main Board pn 95653A001 This is found on the right hand side of the board 39 Position the front housing assembly Item 3 pn 92759AXXX in front of the rear housing and connect the ground wire to the Main Board mounting point 40 Insert the last screw Item 33 pn 40496B001 through the ground wire lug from the front housing and secure it to the middle left hand side of the Main Board NOTE Make sure the lug faces outward 41 Attach the quick connect tab from PS1 to the female tab of the display 42 lf the unit has a printer plug the printer cable into the Main Board at J1 Page 7 38 Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc Section 7 Disassembly 43 Attach the encoder cable to J8 Make sure that the cable is folded flat back over the connector and not pinched in between the enclosures 44 Attach the membrane switch to J7 Make sure that the lock is seated all the way down NOTE The dark blue side faces to the right 45 Connect the display cable to the Main Board at J3 46 Connect the inverter cable to the Main Board at J4 47 CO Oz Connect the manifo
172. ssaeas 5 3 Accessory Testing 5 3 Battery Safety D4 Battery Replacement 5 5 Fuse Removal Replacement cccccccescesceeeceseeseesensecseessesseessaeseneeseeeesseeseeeeaees 5 7 Oxygen Sensor Replacement cccccceeeeeceeceeeeeeeeeeeaeeeeseaeeeseaeeeseeeeseeeeeeeenaeeetenaeeess 5 9 CO Absorber Replacement ccccsssssesnsecseseseseesseessseeesseseseeeseeseseesseessnerseessetens 5 11 Poet Plus 8100 Service Manual Criticare Systems Inc Criticare Systems Inc Long Term Monitor and Battery Storage oo eee ccc eee eeeee eras eee eeeeeenseeseetenaes 5 12 Monitor and Accessories Disposal Electrical Safety TeSting cccctcccscccecavescavessecvadecceocercostevectecetensseucabeseuceodedvesceaveteosedonenes Equipment Needed irrena eriin innana aaea iaa iaaa Withstanding Voltage Hi Pot Test Equipment and Setup ee 5 13 Withstanding Voltage Hi Pot Test Procedure 0 0 ieee eee ceee eee eeneeeee 5 14 Leakage Test Equipment and Setup eee eee eeeecee ese seeeeeeeeneeeseneeeaes 5 15 Leakage Test Procedure c esecsecscssessesstesssnessaeeeseeceeeesseeseeesesnessneeesesesees 5 15 Ground Continuity TSt inseneri enia aaie aa aa aia iaeia 5 18 F nctional Testings aa 5 19 Interface INSpection ss sssesisseserrsserernreusennriusruerentrusenstnusenecauneneraerenerasreneensrnant 5 19 Manual Controls CheCKk sesesssssesreesrerirerrerinsrnerrnriesirsrinsrre
173. ssible Press the rotary knob to save the setting 7 Turn the rotary knob to select O2 CO2 Wave Alignment Press the rotary knob Rotate the knob one click counter clockwise and then one click clockwise One of the waveforms should invert and both waveforms should appear to be in phase Observe the two numbers in the O display column The top number is the delay between the O and CO waveforms in milliseconds The bottom number is the confidence level that quantifies how much the waveforms are in sync Adjust the rotary knob one click at a time while observing the delay and confidence level numbers The delay changes by 0 5 ms and the confidence reading shifts The confidence reading changes slightly from cycle to cycle so observe the reading for several cycles Adjust the rotary knob for the greatest confidence reading possible Press the rotary knob to save the setting The maximum confidence level should be greater than 9 0 If it is not possible to adjust to a reading greater than 9 0 the you should repair the unit 8 Once you are satisfied with the results then proceed to highlight the 50ML flow rate If you are not repeat until satisfied Press the rotary knob turn it one click counter clockwise and click the rotary knob again You are now in the 50ml range 9 Repeat steps 1 through 8 Once you are satisfied with the results then proceed to highlight the 150ML flow rate Press the encoder knob click it onetime clockwis
174. ssis ground on the rear of the monitor 6 Press the Reset button on the tester Press the Start button 7 The monitor passes if the Fail Light does not come on or if the resistance does not exceed 0 1 ohms during the test Page 5 18 Poet Plus 8100 Service Manual Criticare Systems Inc Functional Testing Section 5 Preventative Maintenance Interface Inspection This procedure is for the monitor s TFT active display 1 Manual Controls Check 1 Criticare Systems Inc Verify that a readable screen displays For a guide to proper screen layout refer to the Poet Plus 8100 Patient Monitor Operator s Manual for sample screens Verify that the screen display is clear and bright when you directly face the monitor Verify the image quality There should be no intermittence or jittering of pixels The maximum amount of pixels missing from any display are as follows e Maximum number of bad dots allowed per screen is 7 e Maximum number of bad dots within any 20mm circle is 2 Verify that rotating the menu knob moves the cursor through the menu items Verify that when you press the menu knob on a highlighted menu item the monitor displays a sub menu Use the menu knob to enter the CONFIG window and set the date and time for the real time clock Press the FREEZE key Verify that you get an audible response from the speaker Press the FREEZE key again to unlock the waveforms Press the TREND key
175. st procedures require working with exposed electrical circuits only experienced electrical or biomedical technicians should perform the procedures e After a monitor is altered through repair or hardware adjustment you must fully test it before use Poet Plus 8100 Service Manual Criticare Systems Inc Equipment and Tools Accessory Testing Criticare Systems Inc Section 5 Preventative Maintenance The following procedures assume that the technician has an ESD safe workbench a set of electronic hand tools and a digital multimeter with a 10 amp setting At the beginning of each test special equipment can be listed A variety of customized cables clips and test fixtures are also needed to complete the tests For more information contact the CSI Service Department Check patient cables e g temperature cables printer cables monthly for damage loose wires connections loose connectors cracked housing etc Check the electrical safety of the power cord as part of the monitor safety testing Check the cuffs for leakage as part of the NIBP verification Poet Plus 8100 Service Manual Page 5 3 Section 5 Preventative Maintenance Battery Safety Page 5 4 The batteries require no maintenance For more information refer to the Poe Plus 8100 Patient Monitor Operator s Manual CAUTION A Do not short circuit the battery terminals The resulting high current discharge can cause burns Charge the
176. stems Inc Module Verification Section 5 Preventative Maintenance ECG Verification This test requires a Dynatech Nevada Model 300 simulator 1 Criticare Systems Inc Select the menu to set up for ECG PARAMS to 5 Lead ECG ON Filters MONITOR SENSITIVITY to HIGH Insert a shorted ECG cable into the ECG connector Use the menu knob to enter the DISPLAY window Set the following and use the current defaults for the size They are 25MM except waveform 3 e Waveform 1 ECG x 4 0 25 0 e Waveform 2 ECG II x 4 0 25 0 e Waveform 3 ECG V x 4 0 25 0 Exit the DISPLAY window Verify that the baselines are centered Verify that the base line signal is no greater than 3 pixels in height Select the most solid line as the reference line and verify that this reference is less than 3 pixels from the ECG y axis center line If they are not refer to ECG Analog Output Test on page 5 24 to adjust them Remove the shorted cable Verify that the message LEADS OFF appears at the top of the window when a cable is not connected Connect the 3 lead ECG patient cable CAT 1075 S to the ECG cable assembly and use a Dynatech Nevada Model 300 simulator to turn on ECG normal sinus 60BPM amp 1 0mV axis int Respiration rate 20 and the impedance set for 0 5 ohms and base impedance for 2K Rotate the menu knob to select PARMETERS and press to enable the menu Select ECG for 3 LEADWIRE Set the monitor for LEAD
177. tdenustictarereees 1 4 Stability Of ACCULACY irisse aau aeea a a capes ENEE AE ETa EKERN 1 4 Pacemaker Pulse Rejection wi 1 4 Respiration c cece cece eee eee eee wae 1 5 Non Invasive Blood Pressure NIBP wi 1 5 Comfort Cuff Technology cccccccccecsccascsisesccvcccesasceseesvserescevssassarcoveseconesvectscaresbes 1 5 Description of NIBP Measurement 0 cece cee cess cee eeeeeteeeeeaeeeeeeenaeeeaeeee 1 5 NIBP Clinical Testing and Accuracy wi 1 6 Cuff Inflation and Pressure Protection 0 0 ec ee ceeeeeeeceeecseeeeaeseseeseeeneeeseaes 1 6 Invasive Blood Pressure IBP c cece eee ceeeeeeeeenee sees eeaeeseaeseneetneeseeeeseeeeeeeaeeeae 1 7 Method of Measurement cesseccscsseesseesseessscesseeeeseecseeecneesseessseessneesseaeseeees 1 7 IBP Clinical Testing and Accuracy o eee cece cee cere cee sens eeeeeteaeeseeeeeeeeaeeeee 1 7 Capnometry Measurement of CO gt wt Method of Measurement secssccscssesseesssessssessneseseecseeeseessaasseeesneeeseaeseeees Conditions Of WSC cccscesicsncpchevcnscaduisuercebetenseeticeteterdaisesenensacensherdelvenmarndadeunarecess Stability of Accuracy N20 Compensation Measuring Oxygen Ob Methods nanen naa eaa e N aA aR SSe ela ECAI T a EEEE EA T ETEA T ET ATTA Stability of Accuracy Temperature Measurement Specifications Leakage Current Voltage FIUCtUAtIONS 00 eee cece cece e ree trae eeeee sees seeeeseeesaeseaeeneaeeeeeeeeeeed
178. ter for less than 10 microamps For inline voltages of 264V meter for less than 20uA 10 Set the POLARITY switch to REVERSE and press the OPEN GROUND switch and monitor the meter for less than 10 microamps For inline voltages of 264V meter for less than 20uA 11 Turn the LEADS selector switch to the ISO Position Turn the POLARITY switch to NORMAL press the ISO Test button and monitor the meter for less than 20 microamps For inline voltages of 264V meter for less than 50uA Keep the power cord plugged into the unit under test and the Dynatech Unplug the 8100 ECG cable from the front of the 8100 Unplug the ECG test leads from the Dynatech Connect the Sp0 cable end to any of the BROWN ECG test terminals on the Dynatech Connect the other end to the SPO2 connector on the EUT MODE switch should be on ECG Change the LEADS switch to ISO Set the POLARITY switch to NORMAL Press the ISO button and monitor the meter for less than 20 microamps For inline voltages of 264V meter for less than 50uA 12 Keep the power cord plugged into the unit under test and the Dynatech Unplug the 8100 ECG cable from the front of the monitor Unplug the ECG test leads from the Dynatech Connect the Temp cable end to any of the BROWN ECG test terminals on the Dynatech Connect the other end to the TEMP connector into T1 or T2 connector of the EUT Set the MODE switch to ECG Change the LEADS switch to ISO Set the POLARITY switch to NORMAL
179. the CO board Remove the black covers from the Pump 17 COz CO2 Oz Insert the CO to the Main Cable Item 15 pn 87225B001 into the header of the CO Board 18 Flip the Front Chassis over and slide the Cable Support Item 24 pn 42212B002 over the four standoffs on the Front Chassis Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 33 Section 7 Disassembly Page 7 34 19 COz CO Oz Flip the front chassis back over and fold the CO cable to the left fold the cable at the point where the cable lines up with the Cable support Fold the creases so they are flat Insert the cable through that opening The cable should now be in the front of the chassis 20 Attach a Tie Wrap Holder Item 9 pn 40294B001 to the Front Chassis on the right and left hand side of opening These are across from each other Insert a tie wrap Item 31 pn 40067B002 from the right hand side leaving the head on the right hand side This should pass through both tie wrap holders 21 Install the four plastic Standoffs Item 16 pn 40909B001 onto the front surface of Front Chassis These parts support the SpO Board 22 Lay shielding Item 30 pn 42228B002 down with the V shape cut to your left Obtain the SpO to Analog Cable Item 17 pn 90866A001 and place it on top of the shielding with the connectors up 23 Place the cable along the line that runs horizontally V shape Place the bare wire into the middle of the s
180. the connector plate 6 Replace the board assembly and reassemble in reverse order Criticare Systems Inc Poet Plus 8100 Service Manual Page 7 17 Section 7 Disassembly SpO gt Cable SpO Board 1 Desolder the ground connection from the ECG Analog Board if Removal Replacement not done already Desolder Ground Wire ECG Analog rae Board co ia o 6 o SpO Board wa 8 at ied CRITICARE Systems inc CNO ch 91339A001 REV p MADE IN USA Figure 7 33 ECG Analog and SpO Boards 2 Detach the communication ribbon cable at P5 of the ECG Analog Board 3 Disconnect the SpO cable from J1 of the SpO Board NOTE If you must replace the SpO cable you must remove the ECG Analog Board see ECG Analog Board Removal Replacement on page 7 23 Then replace and reconnect to J1 of the SpO Board 4 If necessary on monitors with COs cut the cable tie on the P5 cable assembly Page 7 18 Poet Plus 8100 Service Manual Criticare Systems Inc Section 7 Disassembly 5 Remove the four Phillips screws holding the SpO Board to the chassis no ee cS ee 66 6e 6 amp Penns aon 03 cat ah CRITICARE Systems ino NO Q4 R 913394001 Rev N MADE IN USA Figure 7 34 Screws Holding the SpO Board 6 Remove the SpO gt Board 7 The communication cable to the ECG Analog Board is attached to the back side of the SpO Board The
181. the menu knob to highlight Adult Pediatric or Neonate as required Optional Enter a value for unit label only if you use central station communication 5 Select EXIT to return to the Main window 6 Exit the ADM DIS window Poet Plus 8100 Service Manual Criticare Systems Inc Creating Custom Profiles Criticare Systems Inc Section 2 Service Windows The default mode you last completed is the default patient size for custom defaults If you want a different default size mode change the order in which the monitor saves alarm limits In the example below the first patient size setting created is NEONATE Update the custom default alarm limits for the pediatric and adult modes by repeating the steps in this procedure two more times once for PEDIATRIC and once for ADULT as the patient size setting ACCESSING THE DEFAULTS DIALOG BOX After setting the patient size e g NEONATE start from the main window to access and set the defaults for your facility s custom profile 1 Press and hold the DEFAULT button on the front panel for approximately two seconds The Set Unit Defaults dialog box appears NOTE If the Set Unit Defaults dialog box that prompts you to enter a password did not appear you did not hold the button in properly In this case with the menu knob highlight and select CANCEL or you could lose the settings you just created 2 Enter the password LIA608 by doing the following e Rotate the menu kno
182. tient monitor measures patient respiration by impedance pneumography As the patient s chest changes size and shape during inspiration and expiration the resistance between two chest or abdomen ECG electrodes changes Respiration rate is calculated from this change in resistance The user may select the ECG which uses the impedance measurement based off of lead or the CO respiration that is based of the capnogram There is also a selectable smart respiration function that can automatically switch sources if there is an interruption of waveform data The CO data is the primary source for the smart respiration function and it defaults to the CO source if it is available The patient monitor uses ComfortCuff technology to determine non invasive blood pressure by means of oscillometry The oscillometric method detects volume displacements within the artery and senses pressure variations within the blood pressure cuff during inflation The patient monitor uses cuffs ranging in size from neonate cuffs to thigh cuffs ComfortCuff technology measures NIBP while the cuff inflates Consequently a measurement is obtained more quickly and with less discomfort than with monitors which measure NIBP during cuff deflation The NIBP cuff begins to inflate at the beginning of the NIBP measurement cycle As the cuff pressure approaches the diastolic pressure of the patient the cuff pressure waveform begins to indicate the pulse waveform The c
183. trol frequency A software controlled timer generates a square wave signal that drives the speaker Hence any drive frequency has harmonic components and generates no pure tones Applicable standards recommend against the use of pure tones in alarm situations to reduce the risk of an individual with a notch hearing deficit missing an alarm Poet Plus 8100 Service Manual Page 3 1 Section 3 Theory of Operation ECG Module Impedance Respiration Page 3 2 The ECG module monitors from either three lead or five lead surface electrode configuration providing one or three concurrent ECG leadviews respectively The ECG module processing includes heart rate ECG lost pacer pulse and leadoff detection The main software is able to control lead gain scale and filter selection as well as pacer mode operation Shielding both on the PCB and external to it reduces the impact of ESU noise on signal quality and heart rate calculation The following issues were important in the design of the ECG module e Diagnostic quality signal bandwidth e reliable beat detection and heart rate calculation e ESU noise rejection e pacemaker pulse detect and e artifact rejection The ECG module uses a differential amplifier to extract surface ECG activity from a pair of electrodes Input to the amplifier is protected against damage from excessive voltages Passive filtering reduces high power RF interference from electrosurgical devi
184. ue changes to 256 NOTE If the value at the end of the line is unchanged then the calibration CAL failed Restart calibration and begin at step 10 16 Once the pressure has stabilized 256 to store the calibration select Store CAL Results and then select Yes Verify that the IBP 1 GAIN and offset values change to match the values from the set 0 and set 256 lines 17 Press the monitors ON OFF key to exit the Service window and turn off power to the monitor 18 Press the ON OFF key to turn on power to the monitor again Attach the IBP cable from BP1 of the simulator to IBP 1 on the connector panel Use the menu knob to select PARMS and then select BP1 Press the menu knob once Set the SITE to ART Press the menu knob again Adjust the RANGE to 0 200 Press the menu knob again until you reach ZERO Adjust the ZERO to ALL Rotate the menu knob to highlight ALL and press to select Observe that a 0 zero appears on the window next to the line that contains ART MMHG Poet Plus 8100 Service Manual Criticare Systems Inc Section 6 Service Testing and Calibration Monitor Fault Testing The following procedures are used to determine whether boards and assemblies are damaged New boards and mechanical assemblies should be tested as part of any replacement repair Board level tests require opening the monitor housing which can void your product warranty Always follow ESD precautions when performing these procedur
185. uff pressure at this point is equal to the patient s diastolic pressure which is stored by the patient monitor As cuff pressure continues to increase the pulse waveform as measured from BP cuff pressure fluctuation becomes stronger reaching its maximum at the patient s mean arterial pressure i e when cuff pressure mean BP The monitor stores this value as mean pressure As cuff pressure increases further it approaches the patient s systolic pressure and the cuffs pulse waveform decreases in amplitude The cuff pulse waveform disappears at the point where cuff pressure is equal to the patient s systolic pressure Poet Plus 8100 Service Manual Page 1 5 Section 1 Introduction When the patient monitor determines that the cuff waveform has decreased to zero amplitude it stores the cuff pressure value as the systolic pressure and releases the pressure from the cuff This typically occurs at about 10 mmHg over the patient s systolic pressure The cuff then rapidly deflates NIBP Clinical Testing and This device was clinically tested per the requirements of EN 1060 and Accuracy AAMI SP 10 The NIBP module as installed in the Poet Plus 8100 patient monitor is tested to meet the performance specifications listed in this manual Cuff Inflation and Pressure The maximum cuff inflation rate is 15 mmHg sec The software limits Protection inflation to 300 mmHg adult 300 mmHg pediatric or 150 neonate A secondary circuit li
186. uld be available only to larger service facilities Calibration kits are provided for field calibration of the CO2 O module The field gas calibration is not intended to replace factory calibration settings Damaged or otherwise faulty CO O modules could require factory calibration that can only be performed at Criticare Systems Criticare recommends that gas calibration NIBP calibration and calibration of the ECG Analog Board be performed at CSI authorized service facilities Poet Plus 8100 Service Manual Page 6 1 Section 6 Service Testing and Calibration Buse peog ulen uoneJqleo dgl UOHEIGIED seg yooeyyD amnp adw a UOHEIGIeD NPON dgIN is eoURWOed ZOdS Bus uoleorunwwoyg UOHBONUSA dal UONBOYUEA O OO MO 4 SED UOHBOYUSA Sinyesodwe uoe dgIN UuOHeodUEA OdS uoe HOA Buysa peuonouny js AinuyUoD punoly ebeyee7 e9993 jodiH eBbeyYo BulpueysuypA VVVvd VviviVv VvVVV VV VV VV VV evvvvvevy VeVvVVVVV Vv Y ve eveeeevv 680 ev Recommended Test Required Test Pneumatics Pump 02 Cell Field Service Testing No Fault case opened NIBP Pump SPO2 Board Display Inverter Board Printer and or Board Power Module Instrument Dropped ECG Analog Board CO2 Absorber Main Board Key Pad Software Download Annual Safety Test Gas Criticare Systems Inc Poet Plus 8100 Service Manual Page 6 2 Perform
187. upon inflation lt 40 seconds average standard adult cuff 2 3 5 10 15 30 60 min 2 4 hrs Adult 0 to 300 mmHg Pediatric 0 to 300 mmHg Neonatal 0 to 150 mmHg 1 mmHg 2 mmHg or 2 of reading 5 min of consecutive readings mmHg Percent kPa Torr Inspired CO Expired CO End Tidal Numerical values capnogram and breath by breath ETCOs bar graph Sidestream Non dispersive Infrared Auto calibrating Manual Calibration Selectable percent only 0 to 3 13 6 25 12 5 or 25 0 to 99 mmHg 0 to 12 5 0 to 12 5 kPa 0 to 99 Torr 1 mmHg 0 1 0 1 kPa 1 0Torr 2 mmHg 0 3 vol 0 3 kPa 2 Torr 150 ml min amp RR lt 35 Br min User Selectable 50 ml min or 150ml min User Selectable 1 7 seconds 150 ml min 4 6 seconds 50 ml min 10 90 350 milliseconds 150 ml min 500 milliseconds 50ml min 1 3 seconds 150 ml min 4 0 sec 50ml min 15 sec including auto calibration to first reading 30 sec to 1 0 vol 2 5 min to 0 5 vol 5 min to full accuracy 54 dBa maximum 1 meter Page 1 13 Section 1 Introduction Oxygen Monitoring 03 Units Display Method Range Resolution Accuracy System Response Time Rise Time Flow rate Temperature Channels Range Accuracy Probe Type Alarms Characteristics Indication Levels Settings Alarm Modes Volume Silence Trend Reports Types Trend memory Tabular Intervals Grap
188. used in the monitor and provides an extension to the battery contacts inside the battery cage WARNING e Shock Hazard Do not attempt to power the monitor with an external DC source while sensors are attached to a person Always use vital signs simulators when using alternate power sources during testing The dummy battery can be used with an external DC power source Use BC precision Model 1646 or equivalent supply rated to 16 volts at 10 amps Contact the CSI Service Department for more information The test also requires using a battery load simulator The simulator is used with an ammeter to test charge current For more details about battery load simulators for the monitor contact the CSI Service Department 1 Insert the battery dummy block and verify the quiescent current is less than 1mA in BATT1 slot 1 2 Verify that the on current is less than 1100mA without the printer or NIBP pump running with an input voltage of 15VDC 3 If CO O option is included the current should be less than 1660ma with the CO pump running and the heater turned on 4 With AC mains plugged in connect the battery load simulator to the main board at PS1 or a fake battery 1 Switch the Battery Select switch to the BATT1 position Start with the simulated battery voltage at less than 10VDC The ammeter should be 15ma 6ma 2 Slowly increase the voltage Verify the bulk rate is 620ma 100ma when the voltage is greater than 10VDC
189. veforms aVR 11 Verify that the respiration rate changes from 20 to 30bpm when changing on the MED SIM 300 Criticare Systems Inc Poet Plus 8100 Service Manual Page 5 23 Section 5 Preventative Maintenance 12 Rotate the menu knob to highlight Display and press to select Under waveforms 3 6 Select e RESP as TYPE e GAIN as 1 0 e SWEEP as 25MMW SEC and size The size might not be changeable on some waveforms Select Exit and verify that anon symmetrical sine wave displays on the window 13 Set the Dynatech MED SIM300 to a heart rate of 300BPM Verify that the ECG HR reads 300BPM 1BPM 14 Turn off the simulator and verify that the message ECG LOST appears at the top of the window 15 Turn the simulator back on and set the simulator for 6 OBPM LEAD II and verify that the gain settings are effective at x 5 x2 and x4 for 1mv ECG input 16 Select on the simulator for ECG performance Cursor through the menu until you obtain and execute twice the PACED RYTHMS Verify that PACE is turned on in the menu screen of the 8100 You should see white spikes on the waveform and 75BPM 17 Select ECG on the simulator select base then BPM and 60 This gives a normal sinus rhythm at this time Select ECG then ARRH then VNT2 and then VFIB The unit should detect an erratic baseline and display the message ECG LOST ECG Analog Output Test You need a MedSim 300 patient simulator to perform this procedure 1 Connect a BN
190. will not be accepted if the CSA number is not clearly visible Please provide the model number serial number and a brief description of the reason for return Freight policy The customer is responsible for freight charges when equipment is shipped to CSI for service this includes customs charges Loaner service In the U S If it is necessary to provide a loaner system CSI will ship a loaner by overnight courier The loaner system must be returned to CSI at the customer s expense within one week after receipt of the repaired goods If the unit is not returned to CSI within that time the customer will be invoiced for the full purchase price of the equipment Outside the U S No loaners are available from CSI internationally Contact your local CSI representative Poet Plus 8100 Service Manual Criticare Systems Inc EC Declaration of Conformity Poet Plus 8100 Patient Monitor Representative in the European Union Criticare Systems Inc To view the Declaration of Conformity visit the Criticare website at www csiusa com A copy of the Declaration can also be faxed Contact Criticare s customer service department at 262 798 8282 to obtain a faxed copy of the Declaration Frank Henaghan Frank Henaghan Criticare Systems Inc Fax 44 151 342 7208 P O Box 48 Frank Henaghan csiusa com Heswall CH616TZ United Kingdom Poet Plus 8100 Service Manual Page xi Section 1 Introduction Description Intended Us
191. y artery IBP measurements the numerical values for systolic diastolic and mean pressure are computed and displayed Method of Measurement The IBP module uses standard invasive pressure monitoring techniques A bridge transducer is measured by a highly accurate analog to digital converter The measuring circuit uses a 5 volt excitation voltage that provides a transfer function of 25 micro volts per mmHg of pressure The signal is buffered and brought to the A D converter which is configured for a full scale signal of 25 millivolts The circuit has a potential measuring range of 1000 mmHg of pressure This range is restricted to 20 to 300 mmHg for actual monitoring purposes There is no analog amplification of the transducer signal The module uses a rate of 50 samples per second for each of the two independent IBP channels The resulting measurement resolution is slightly less than 0 5 mmHg The numerical IBP value is reported by the Poet Plus 8100 monitor at a resolution of 1 mmHg Non pulsatile IBP sites LA RA CVP or ICP are reported as mean numerical values The non pulsatile mean value is an average of the last 200 samples taken over a period of approximately 4 seconds The values for pulsatile sites ART and PA are reported as an average of the last 8 calculated values The IBP heart rate is reported as an average of the last 16 heart rate values The software also scans for sudden changes in heart rate and overrides the averag

Download Pdf Manuals

image

Related Search

Related Contents

  FHYC-01取扱説明書  fiche technique Masque complet  Sandisk Digital Photo Viewer User's Manual  series  PDF File  manual de instruções do termômetro digital modelo td  Transmisión - Centro Renault  American Standard Afwall Wall 1.6 GPF Flush Valve Wall-Mounted Toilets 2258 User's Manual  ISTRUZIONI PER L`USO - Pdfstream.manualsonline.com  

Copyright © All rights reserved.
Failed to retrieve file