ARGUS 707V Infusion Pump Service Manual

image

Contents

1. 11 REPAIR ORDER FORM ARGUS Medical AG Heimberg Switzerland REPAIR ORDER FORM Purchase order Proforma invoice number Customer name and address Name of contact person Tel number Device A414 ARGUS 100 P A400 ARGUS 100 M A404 ARGUS 600 S A200 ARGUS 707 V A300 ARGUS 708 V Serial Number Accessory Serial Number Production code Detailed failure or problem description Expected work repair to be done Repair Warranty repair Replacement Other Description Signature 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 48 48
2. 15 No Display VTBI volume to be infused No Yes 16 No Display Setz if only one setis enabled LI No Yes 117 Yes KVO KOR enabled 60 Ir NoYes 18 No Dropalarm only if bottle empty 10 No gt trA __ No Yes 19 4 No Buzzeratstart Nes 20 No MenuCir clearmlinf ch 1 No Yes Pok pg ENER aaa 22 No MenutrA transport 10 18 N0 No Yes 23 Yes MenuPrl pressure alarm lm No Yes 24 Yes MenuCAP battery capacity NoYes 25 No MenuSetFill primelVset ll 1 26 No MenulnF mlinf sincelastpoweron No Yes 27 No Menudlo data lock NoYes 1 28 No MenuStb standby NoYes 29 No Menu Med medication name No Yes 30 No MenutM timeralarm d NoYes st 32 No Menubol bolus release Noi Yes 33 No MenuboLr bolusrate sf 32 Yes 3 3 No Yes 34 No Menutot bolus totaly 32 Yes No Yes 28 pogo go NA gg po jp ol oo 1 lo 2 go go Tell CQ 38 No Automatic bolus application 32 34 Yes No Yes 39 No Neonatology mode accessible in PILmenu No Yes 40 occ o loo T1 1 1 41 No Clear ml h after infusion completed No Yes 42 No Clear ml total after infusion completed 41 Yes 3 3 No Yes 1 43 No Airvolumeaccumulated LL No Yes 44 5 Yes Automatic press
3. Service Manual for Volumetric Infusion Pump ARGUS 707 V Made in Switzerland i z poy Felt Dote san ann Th MIN C 0120 MINGUS MEDICAL ARGUS Medical AG CH 3627 Heimberg Switzerland a member of the CODAN group 14 203 B_A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 1 48 TABLE OF CONTENTS 1 INTRODUCT ION WEE 4 EE GSA aeea AA ANA AA 4 2 PUMP ei ei Lied RR LN 4 Ze la ICME RE 4 2 2 Interrogation mode without ARGUS service 5 2 3 Configuration mode without ARGUS service 6 2 4 First activation of a PIN code write protection aa T 29 Changing an Exising PIN COGE naaa ale a a A AS 8 2 6 Address list of the pump configuration without ARGUS service 9 2 7 Index list of the pump configuration with ARGUS service 10 2 8 Special configuration options with ARGUS service 14 2 9 Ee Uer L I EE 15 3 Serial communication of the pump ccceseesesseeceeseecenseeeeseeseneeseeseesenseesenseesones 18 Sie GENETTA RE 18 Id Sella COMMUMICATION ProloCO E 18 A ARGUS SEN ka Sana ka ka KA KO AA EG 19 Al GENE aa oR AIDE ANAL ALAT NANA ee ee 19 4 2 ARGUS service Confguraton a 20 4 3 ARGUS service Calibration volumetric pums only annannnannannnnnnennennnnnnnnnennnnnnnnnnnnne 21 4A ARGUS Service TODIDOX daan ANA AN NAA NAGA 22 5 SOFTWARE UPDATES mO KANAN KA ANAK EEN 26 e OSNA E 26 5 2 Requirements for a software update 26 99 O
4. Enter 30 in the target volume setting field enter the measured volume weight in the actual measurement reading field The deviation will be calculated by the PC software Apply this value to the IV Set you want to calibrate Perform a control measurement according to the steps 2 amp 3 above Repeat the calibration procedure if necessary 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 36 48 6 5 3 Volume calibration of set 1 with the pump integrated calibration program Needed equipment balance with a resolution of 0 1g at least 1 Insert a new IV set only recommended IV sets may be used see ap pendix in the user manual in the pump filled with water and connect the infusion line to a measuring cup on a balance Enter into the configuration mode as described in chapter 2 3 Enter at address 399 the value 123 Press key START to acknowl edge the entered value Now you can select your preferred set in our case set 1 4 Press the key START STOP The pump delivers a volume of 40 ml at an infusion rate of 999 9 ml h tArA is flashing When tArA stops flashing reset the balance to 0 Press the key START STOP The pump will display tM 432 infusion delivery time is 432 seconds It should now deliver a volume of 30 ml at an infusion rate of 250 ml h 7 After the delivery time has elapsed the pump stops and shows bAL 30 00 in the displays Enter now the value o
5. Check the green op ndear Door open the door indicator lights up d Door close the door indicator turns off ER Open the door and remove any IV set Indicator air bubble lights up i ER Install a water filled IV set close the door Indicator air bubble turns off E Set rate to 333 3 ml h press START The red indicator drop lights up delayed disconnect the drop detector The acoustical alarm turns on Check the external connector nurse call Relay contact switches see chapter 9 E Calibration of the pressure sensors See chapter 6 3 and 6 4 of the service 13 IV set type used manual Volume calibration See chapter 6 5 of the service manual HE Charge the battery while the pump is running 16 hours at a rate of 30 0 ml h Battery check at a rate of 30 ml h The green indicator battery lights up during 16 Run the battery test until the pump switches off this test automatically Printout the pump history refer to chapter 4 4 5 Check the pump internal time and data 17 Check the battery run time by checking the latest history entries gt 4h 30min 18 Electrical test according to EN60601 1 Visual check of mains connector all measurements made with a power cable 2 5m Measurements attached The pump has passed the SSC and Is safe for use Dato NAME er e AON O aaa ANAN ANAN ARGUS Medical AG 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 47 48
6. ROM test F 22 ROM check Runtime F 23 RAM test check F 24 XRAM test check F 25 CPU test F 26 Invalid function menu F 27 EEPROM aata invalid F 28 RTC data invalid no RTC etc F 29 Stepper motor power test delayed 5s F 32 5Volt supply out of range F 33 24Volt supply out of range delayed 5s F_37 Downstream pressure sensor test failed always gt 4 7V or lt 0 2V delayed 5s F 38 Upstream pressure sensor test failed always gt 4 7V or lt 0 2V delayed 5s F 39 Downstream pressure sensor test failed dynamical test failed F 40 Upstream pressure sensor test failed dynamical test failed F 44 Address invalid for config eeprom F 45 Address invalid for history eeprom F_46 Frequency from uC or RTC out of range F 47 Display print not present F 48 Key s too long active F 49 Sensor print not present F 50 AIL Air in line detector test failed F 51 Movement test failed Home Pulse lt Hall Home F 52x4 Movement test failed Home Pulse gt Hall Home F 53 More than one rotation at STOP without KVO F 54 Infused sum s7 Calculated sum Rotations F 55 Frequency calculation F 56 Invalid volume adjustment over time F 57 Rotation SW overflow F 58 Internal volume control 10 ml We recommend replacing the main board in case a fault code is not included in this list above Failure released only if configured with upstream sensor index 51 Yes kx Fault code F 52 may occur after a software update or
7. ex works settings The current PIN a The new PIN to set Codes 889 can also be entered on the pump Please note The PIN code corresponds with the PIN Codes 10000 32767 only within AM Service code mentioned in chapter 2 3 2 If a PIN code greater SE than 9999 is entered the pump configuration can only Z be accessed using the ARGUS service PC tool Quit 4 43 ARGUS service Toolbox Service interval Use this feature to set a reminder alarm on the pump for the next service interval 4 A pending reminder alarm will be shown on the pump display after vs power up by a flashing CtrL text accompanied by an acoustic Service interval sound The point in time when an active reminder alarm occurs is given by the settings of the configuration 393 and 394 and the pump internal clock Any value higher than O on those indexes will release the reminder alarm after the service interval has elapsed Please check those settings first before you set the reminder alarm Please note By the ex works settings the reminder alarm is disabled 4 4 4 ARGUS service Toolbox Replicate s Use this feature to replicate fast and easily pump configuration from a saved configuration file or from a pump to another A con figuration can only be replicated if the saved configuration and RR pump type corresponds with the firmware of the connected pump E in the first 2 digits for e g 4 30 to 4 31 is possible Rep
8. if peristaltic is manu ally turned during service In this case turn the pump off then on again and restart the pump If a failure free start can be performed omit this fault code 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 38 48 7 REPLACEMENT OF PARTS 7 1 General CAUTION The ARGUS 707 V may only be used with accessories and spare parts which have been approved by ARGUS Medical AG for safe technical use CAUTION If a door a housing a pressure sensor or a main board is replaced a full calibration pressure sensors and volume calibration is required 7 2 Disassembling of the ARGUS 707 V CAUTION Disconnect the mains cable from the power outlet before opening the housing Observe the antistatic protection rules when disassembling the ARGUS 707 V the use of an antistatic table mat and a grounded clip are recommended 7 2 1 Disassembly of the case 2 Separate the front and the back side 3 Disconnect the battery the supply and the docking in terface connection 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 39 48 7 2 2 Removing the pump door 1 Gently press this hinge stopper backwards 2 Pull the shaft towards the centre Repeat point 1 and 2 for the other side of the door and then remove the door 7 2 3 Disassembling the pump unit 1 Unscrew these screws Remove the pump unit 7 2 4 Assembling the pump 1 Use the screw with two washers
9. it will never be visible again an OF 323 CAUTION After you switch the pump OFF and ON again you can enter into the configuration mode only if you enter the correct PIN code eg Cod k 3333 i Please note The interrogation mode can always be accessed without the PIN ra O O m TOI 2 5 Changing an existing PIN code Enter the configuration mode using present PIN select add O and set the new code Cod je 3333 r Enter actual PIN code and confirm with START key mw Press MODE key 0 The flashing decimal points Dis Ll E will change to the right hand display y wen Enter the new PIN code and press the START STOP 0 CECE le key to acknowledge the entered code 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 8 48 2 6 Address list of the pump configuration without ARGUS service The following list declares the possible configuration options which can be performed on the pump keypad without using the PC All these options can also be configured by the PC Software ARGUS service Index Default Function left display PC right display 1 2 No Key ON OFFonlyatstop valid O No 1 Yes 10 No Operation without drop detector 0 No 1 Yes 3 11 No Recallofthelastused ml h rate 0 No 1 Yes 4 19 No Buzzeratstart LL O No 1 Yes 5 44 Yes Automatic pressure release 0 No 1 Yes 6 49 N
10. not L aed Green pirmam ARGUS 707 2 Select address 399 3 Enter value 1234 4 Press START STOP and press MODE Display shows Set 1 6 3 3 Set selection before calibrating mi total E Before you select an ee me IV set check chap 2 8 100 ag a wf z et h Bem Gi 8 s RGUS707 2 Press START STOP Display shows CAL door 1 Press key 1 to select the set num ber you want to calibrate 1 to 4 6 3 4 Calibrating the offset of both sensors 1 Remove the IV set and open the door 2 Press START STOP to store the values for the up and downstream sensor 3 Press MODE Display shows CAL 1 0b 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 31 48 6 3 5 Calibrating the downstream sensor 1 Take a new IV set filled with water 2 Simulate a downstream occlusion by the 3 way stop cock before you insert the set 3 Open the roller clamp insert the new IV set and close the door 4 The pump generates a pressure build up automatically When the pressure 1 0 bar reaches press START STOP 5 Release the pressure and press MODE Display shows uPSt reAM 6 3 6 Calibrating the upstream sensor part 1 UPSt reAM 1 Install a clamping shears near the pump bottle side and wait 15 sec then press START STOP 2 Release the clamping shears and press MODE Display shows CAL 0
11. to chapter Pressure calibration has to be carried out again 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 35 48 6 5 Volume calibration 6 5 1 General On the A707 Volumetric pump it is possible to calibrate up to 4 different IV Sets For each IV set you can define a correction factor which calibrate the volume deliv ered There are two ways to calibrate the volume delivered by the ARGUS 707 Volumetric pump select one e By entering the correction factor e With the internal calibration program of the pump Needed equipment balance with a resolution of 0 19 at least 6 5 2 Volume calibration of the pump by defining the correction factor Please be sure which IV set you want to calibrate If it is not enabled in the configu ration please enable it first and select the according IV set in the normal mode of the pump in menu Set x 1 Insert a new IV set only recommended IV sets may be used see appendix in the user manual in the pump and perform a warm up infusion of 40 ml at an infusion rate of 999 9 m h 2 Infuse now a volume of 30 ml of water in a measuring cup on a zeroed balance at an infusion rate of 250 ml h 3 Determine the weight of the delivered water 4 Connect the pump to a PC and start the ARGUS service PC tool as described in chapter 4 go into the calibration part of this tool 5 Refer to chapter 4 3 select by VOLUME on the left side in the calibration win dow 6
12. 100 247 1014 Correction at 75 ml h these values unless 900 1100 1 243 1008 Correction at 100 min gie trained 249 999 Correctionat150 m h 250 988 Correctionat200 mi 251 976 Correction at 250 mh 252 967 Correction at300 m h _ 253 952 Correction at400 m h 254 942 Correction at 500 m h 255 925 Correctionat750 ml h 256 910 Correction at999 m h 257 1000 Correction at 200 mp 258 1 TI 260 Co Ll 261 1065 Correctionat1m h 262 1054 Correctionat3m h 263 1044 Correctionat5mi h 264 1035 Correctionat10m h SETA 265 1027 Correctionat25m h 266 1020 Correction at 50 ml h Do not change ratio 900 1100 267 1014 Correction at 75 ml h ERE ee UNESE 268 1008 Correctionat100mih explicitly trained 269 999 Correctionat150m h 270 988 Correctionat200mlh 271 976 Correctionat250m h 272 967 Correctionat300m h 273 952 Correctionat400m h 274 942 Correctionat500m h 275 925 Correctionat750mVvh 276 910 Correctionat999mVv h 277 1000 Correctionat1200m h i cuh Ch CE EE C UA FT AU V PTI To NN CC TI To TA amo 9909 maxfowrae ooo am 999 Max prime rate Mam 1200 Max bolus rale Set 1 parameters a3 10 Max bolus to
13. 2b Remark By pressing key 1 the pressure signal in mV of the upstream sensor will be displayed after approximately 15 seconds the value does not decrease that fast any more at this point press START STOP to store the value refer to step 1 The software will not allow the value to drop lower than the previous saved door open offset value 50mV 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 32 48 6 3 7 Calibrating the downstream sensor CAL 0 2b 1 Simulate a downstream occlusion by the 3 way stop cock 2 The pump generates a pressure build up automatically When the pressure 0 2 bar reaches press START STOP 3 Press MODE Display shows CAL 1 2b CAL 1 2b 4 The pump continues on with the pressure build up When the pressure 1 2 bar reaches press STARI STOP Remark The pump must reach a pressure of at least 1 2 bar Otherwise there might be a mechanical problem Check the pump unit and the door 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 33 48 6 3 8 Calibrating the upstream sensor part 2 2 Open the roller clamp insert the IV set in the reverse direction and close the door 3 The pump generate a pressure build up automatically When the pressure 1 0 bar reaches press START STOP 6 The pump generates a pressure build up automatically Is the pres sure of 0 2 bar reached press START STOP 7 Press MODE Display show
14. O Wak leefeg TEE 26 Did OEC ASD CCS orina AA AA aa AA 28 6 MAINTENANCE AA AA AA AA AA 29 o SC AA AA AA 29 6 2 MRECAIBTAUON ama Aa ee ee a ea ALAGANG KR 29 0o Giele leie Lee 30 64 Pressure controliMmeasurEeMEN EE 35 6 5 VOolume calIbral0N aaa NAAN a dee 36 6 64 PUMP SDECINCaIONS Ama a AA AA 37 Gila SAU COC CS eos os ced teas cites ted went ete ee PADA TA 38 7 REPLACEMENT OF PARTS 1 0 020020002002a NANA ANN NNNNNNNNNN RAN RAN KE KEN RAN KEREN RER ERKENNEN KREE KE EN 39 e GEN dhan naa O A A AR OE ARA eae ee Reon OE O OPR EM 39 72 Disassembling ei RER 39 ki IPA DANI AN Na a NG ANA Aa bna 43 8 WIRING DIAGRAM Mi AA AA AA 46 9 BLOC SCHEMATIC u 1 namna m aman KA NAN KANKA KANAN RANE NAN SKO NEON NANESE ONONNNONENNNNN NON NONA RESEN Na EN 46 10 SAFETY STANDARD CHECK ana a naa ANN nauna anuna KEREN KEREN RER ENKEN R ENNER ENNEN nana E E 47 NB CA GT Re RI Ce d ee ANNA UA AA 48 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 1 48 ms si rr WW sli Drop detector Drop chamber E d i de l Air bubble detector Pressure sensor automatic Pressure sensor upstream free flow clamp downstream 14 203 B_A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 2 48 Universal clamp Staff alert and RS 232 connection Drop detector connection gt em Identification plate iss lt j with pump serial wil EB number AC power connection DC c
15. O71 O O Q1 CO o OO OO Oo OL OJOJO O 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 15 48 2 9 3 Defined medication index PC index PC Def Yes 660 No Fotemustine No Yes 601 602 606 No Ajmain No Yes 666 No Glucose n saline No Yes 609 No Alizapride No Yes 669 No Hartmanns No Yes 617 No Antibiotic No Yes 677 No KCL No Yes No Yes 690 No Methotrexate ss No Yes 696 No Mitoguazone No Yes 699 No Mivacurium chloride No Yes No Yes No Yes 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 16 48 727 No Propofol No Yes 768 No Vinorelbine No Yes Ritodrine hcl No No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes Streptokinase 3000 ie Streptokinase 6000 ie 731 No Remifentanyl No Yes No Yes No Yes No Yes No Yes No No Yes Tegafur No Y 789 No Y No Theopyllin Ba aaa REMARK Via barcode reader all medication names can be selected even if they are not released in the configuration 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 17 48 3 SERIAL COMMUNICATION OF THE PUMP 3 1 General The ARGUS 708 Volumetric pump has two serial interfaces on board One is wired to the docking interface connector and one is connected to the RS232 connector Im
16. PLACEMENT OF PARTS 10 127 Door complete A707 10 128 Casing back plane A707 30VAC complete y ee EB K a E m a 2c MU mu z 10 129 Casing forepart without door complete 11 221 Sealing A707 12 042 Cable staff alert 2m 12 011 Interface cable 8pol 2m 11 268 Alarm lights amp 12 012 Interface adapter 11 132 Label external 11 237 Identification plate A707 11 238 Label flow direction drop detector 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 44 48 10 135 Edge board complete A707 mi total mi int fh min gum a e 100 10 1 01 sma OO OO NG eo DA MODE STOP pri amaan HABI 12 035 Pressure gauge 11 219 11 239 11 247 with stopcock Short instructions Part no 11 222 Front panel A707 SW DE EN FR DK NL CZ PT SP IT 11 005 Bottle holder 45 cm 12 044 Plug 11 043 Bottle holder 60 cm 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 45 48 8 WIRING DIAGRAMM Display Board with Hall Switch Door X01 OOOOGOOOGO OOOOOOGO Drip SCI 16 Nurse Call Oooo 00000000 OOoOO0000 CH0 EI ee 5 D O Edge Board Cj KM xoz EI Fr Stepper Motor D 6 D X02 X01 ER i 2 Battery Main Board xo1 Sensor Board 10 with Hall Switch Motion Trafo Pressure Sensor Pressure Sensor Air Detector Stop Flow Protection i Input Output i AC Main LELTEETEEETTTTTTTETETETTTITITTETTTTITTTTTTI 9 BLOC SCHEMATIC nm HOH D
17. a 100 10 il Mode Start Stop Controller 3 na Alarm 2 Relay Nurse call Pressure HE pa KVO RS232 Il BEBA Display D o a RS 232 Driver Motor Driver Reset Watchdog a Power Switch Test Switcher 0 EEPROM AC Main CC D RTC L 512 Hz Motor Buzzer Ext DC Li Batt i 3 Volt DV Ni Cd Logic Air Bubble Detector Akku 12V Drop Detector Pressure Sensor in Door Control Pressure Sensor out Motion Control 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 46 48 10 SAFETY STANDARD CHECK Safety Standard Check SSC ARGUS 707 V en ae US Department ccc eeeceeeeeesaeeeeeeneeans CUSTOMER aNG NIANG NANDTO RAON AN AN AN AALANGAN The SSC has to be performed at least every 24 months or after 10 000 hours of operation The check has to be done in accordance to the user and service manuals BEE Check if a software upgrade is required 1 ee 2 Visual check for damage cleanness and com Housing labels accessories connectors pleteness power cable etc 3 ER Test the function of the stop flow clamp Proper movement of the clamp NE Keep MODE pressed while switching on the Display pump type and software release 4 pump Display of 2 4 7 F in numeric display Display of all operation and alarm indicators The indicator external supply turns on off KE 6 Teste drop deeco by simulating drons
18. a cmH20 Psi Enum 1 5 a a 369 5 Automatic menu fall back delay time sec 5 30 oU o po Po po o o Lo o o TT eee eee 390 0 Lastservice dateinyear yar 0 99 391 O Lastservicedateinmonth mon 0 12 392 0 Last service dateinday day 0 31 393 0 Service interval in months disabled month 0 24 394 0 Service interval in hours of operation 0 disabled hour 0 10000 35 CoC ol CL 396 0 Pumpserialnumber L XXXB yyy xx8yy NEE EE NEE NEE ee Se 398 po a o ro 399 707 Enter the calibration menu clears protectionkey 1234 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 13 48 Permissions for serial communication 0 8 ne T Enum 0 3 none 1 query only 2 parametrising 3 remote control 522 No Allow change of ml total while infusing 65 No Ir No Yes _ 523 15 Alarmlamp brightness ooo SO level 2 15 524 Display a department info text after power up har 1 16 ASCII Using the ARGUS service tool the complete and detailed pump configuration can be done 2 8 Special configuration options with ARGUS service Enabling a new IV set Follow the steps below to enable and configure a new IV set a write Yes in the corresponding index 101 102 103 104 b make sure the set definitions index 201 to 277 correspond to the IV sets used In case of do
19. ast technical failure numbers are listed in this part ARGUS part 4 This part contains ex works settings e g pump serial number AMService Configurat al 1 El Configuration t Ab Calibration al Statistic ff ARGUS 14 203 B A707 en SM V4 3X AMService Configuration AYO 4 00 ol Configuration Menues Display Handling Functionality Techrical IV Det 5 Lutves Communication Medication vk 6660P 20 22 30 32 34 369 1 7 8 15 16 45 46 48 56 366 2 5 14 39 49 361 365 11 13 17 38 40 42 60 3 06 PJ 20 48 6 10 18 19 43 44 47 50 51 351 362 363 368 390 394 101 310 318 102 320 328 103 330 338 104 340 348 201 217 221 237 241 257 261 277 55 514 517 600 800 561 592 4 2 2 How to edit a configuration The following procedure describes how to edit a pump configuration 1 Press the green Edit button 2 The software will ask for the pump PIN code as next The button Edit changes its colour and will be renamed into Download 3 If you want to import a configuration from a file press the Import button other wise skip this point 4 Select Configuration in the structure tree in the left upper frame 5 Select the index group you want to modify by selecting the according structure tree and the according index 6 Modify the according index within the given restrictions shown Each value number must
20. at these places torque 1 0 Nm 2 Use the screw without a washer at these places torque 1 0 Nm 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 40 48 3 Reinstall the door Make Make sure that this part sure that a washer has touches nowhere otherwise been installed on each side noises may be generated 5 Reconnect the back side with the front side battery power amp docking interface 6 Make sure that both ca bles are correctly placed in the notch before the main board is inserted 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 41 48 7 Use the short screw at these places torque 1 1 Nm Et a pat will Bi 7 Use the long screw at these places torgue 1 1 Nm Mind the torque for the screws listed below What Case Pump unit Combination clamp Pressure sensor Stop flow Air detector Sensor board Torque in Nm Intofront 0 60 Intofront 0 60 0600 Into front Into pump unit ARGUS Medical AG 28 03 06 PJ 42 48 14 203 B A707 en SM V4 3X 7 3 Spare parts 10 087 Combination clamp 10 089 External drop detector 10 093 Interface cable docking pumps m m IW 88868 ECK 10 119 Battery NIMH 12V 1500mAh 10 120 Display board A707 10 121 Pressure sensor 10 122 Air detector 10 123 Stop flow A707 10 125 Sensor board A707 10 126 Pump unit A707 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 43 48 RE
21. at200m h 211 976 Correction at 250 ml h 212 967 Correctionat300m h 213 952 Correctionat400 min 214 942 Correctionat500m h 215 925 Correctionat750m h 216 910 Correction at 999 m h 217 1000 Correctionat 1200m h LL LI 221 1065 Correctionat1m h 222 1054 Correctionat3m h 223 1044 Correctionat5m h 224 1035 Correctionat10m h sid 225 1027 Correctionat25m h 226 1020 Correctionat50m h 227 1014 Correctionat75m h SET 2 Do not change these values unless 228 1008 Correction at 100 ml h explicitly trained 229 999 Correction at 150 ml h 230 988 Correction at 200 m h 231 976 Correction at 250 m h Correction at 750 ml h Correction at 999 ml h 1000 Correction at 1200 ml h 14 203 B A707 en SM V4 3X Correction at 500 ml h ARGUS Medical AG 28 03 06 PJ Z o Yes o Yes Z o Yes o Yes o Yes Z o Yes o Yes No Yes No Yes No Yes No Yes 900 1100 ratio 900 1100 o S 11 48 zgg p o 239 Coco lo 240 OT 241 1065 Correctionat1 mih 242 1054 Correctionat3 mih 243 1044 Correctionat5 mih 244 1035 Correctionat10m h 245 1027 Correction at25 mh 900 1100 900 1100 246 1020 Correction at 50 ml h Do not change 900 1
22. be acknowledged by the green Enter button Go through point 5 amp 6 for all further indexes you want to modify 7 Press the Download button if you want to save the modified configuration on the pump Otherwise you can save the modified configuration into a file by pressing the Save button 8 Make a functional check on each pump you have configured Important remark If a config has been edited performed point 1 and 2 once do not switch off the pump Otherwise the pump will change always into the PC configuration mode automatically 4 3 ARGUS service Calibration volumetric pums only Use this feature to easily calibrate the volume delivered by the volumetric pump Calibration ARGUS service Volume Calibration To do this it IS possible to enter Calculator the measured rate of an infu Higa Po bk ka sion device tester e g or from Actual measurement reading ml h C by VOLUME a measured volume of a bal Deviation 215 0 x ance to define a correction fac el earn ae tor according chapter 6 5 2 j Set 1 disabled Set 3 disabled Current Current Current Current la Correction factors 850 1150 OD v os 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 21 48 The calculated correction factor can be applied to IV sets which are released in the config index 101 104 only Please be sure which IV set you want to calibrate before you press the apply but
23. change back to the left hand display 10 To quit the configuration mode press the 5 0 log ON OFF key at least 2 seconds Changes in configuration become active after the pump is switched on normally again Important remark Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address 2 4 First activation of a PIN code write protection The activation of a PIN code allows you to protect the configuration from unauthor ized access To activate the PIN code enter into the configuration mode 1 To enter into the configuration mode switch the pump on while keeping the keys MODE and START STOP pressed 2 Press key START mi total mi int A min aie ai sme 4 Press key START At again The display will peg g show Cod PRGUS 707 teta Niet Cod je z 5 Press MODE key Code 0 will be set then START key to acknowledge the entered PIN code 0 3 Press key MODE ohe 0 ofer Ok Disse 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 7 48 6 Press the MODE key The flashing decimal me points will change to the mi inf right hand display A min CEC See VU GE m eag di OD 8 Press key START to naaa Ha acknowledge to new PIN 7 Enter now the new PIN Code max 4 digit number Please remember this code
24. ctions displayed on the PC The firmware will be installed and the pump will be switched off automatically 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 27 48 8 Switch the pump ON while keeping key 10 pressed Start the ARGUS service tool and select the according COM port e 9 Import configuration from backup The H Seen calibration values and configuration of last Import nab connected pump will be imported 10 Download it to the pump by pressing the Report Ej Download button EMail to ARGUS Petpet Perform a standard safety check see chap 10 the calibration values maybe lost during the upgrade procedure 5 4 Safety aspects Be aware of the following points For medical device traceability your local distributor or ARGUS Medical AG needs to be informed about every device updates serial number you performed Do not make any software updates when the device is used and or connected to a patient CAUTION A standard safety check see chap 10 has to be performed after every software update 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 28 48 6 MAINTENANCE 6 1 6 2 General CAUTION Only authorized persons who have been trained by ARGUS Medical AG or by the local distributor are allowed to service the ARGUS 707 V infusion pump In case of repair request send the unit with the filled out repair order form see chapter 11 to the local distrib
25. f the balance e g 29 80g This value must be within the range of 25 50 34 50 Otherwise switch the pump off and re start the volume calibration Press the key START STOP to acknowledge the entered value The pump displays the new correction factor e g Cor 993 10 Press the key START STOP to store the new correction factor ac knowledged by a buzzer sound The pump display changes back to SE 1 again 11 Switch the pump off Do not remove the infusion set 12 Perform a control measurement with an infusion rate of 250 ml h and an infusion total of 30 ml Repeat the calibration procedure if necessary For an optimized long time accuracy over 24 hours we recommend to do a control measurement over 24 hours on a rate of 25 mI h 6 6 Pump specifications Please refer to the user manual for the specifications chapter 9 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 37 1 48 6 7 Fault codes A technical failure will be signalled by the pump with a continuous alarm display and a continuous sound During this state the fault code which causes the pump to fail can be displayed by pressing the key MODE If the pump was switched OFF after a detected failure the fault code will be stored in the configuration of the pump see index 380 389 via ARGUS service PC tool The possible fault codes registered in the configuration are listed in the table below Fault Code Failure F 21
26. he Serial Infrared Link Access Protocol Ir LAP Version 1 1 Please contact your local distributor or ARGUS Medical AG for the complete serial communication protocol description 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 18 48 4 ARGUS SERVICE 4 1 General The new ARGUS service utility is a high and user friendly PC software which can configure and upgrade pumps over PC serial COM port With this Windows based software you can also set pump clock change PIN code read and print out history and easily replicate pump configurations and so on The modern and clearly struc tured design of this self describing PC tool allows a very easy and rapid modifica tion of the ARGUS 600 Syringe pump the ARGUS 707 amp 708 Volumetric pump and the ARGUS docking station This software may be available from your local distribu tor or directly from ARGUS REMARK ARGUS service may only be used with software versions greater or equal to 4 00 CAUTION The infusion pump has to be disconnected from the patient before and while the se rial interface cable is connected to the pump The connection of the A707 over the serial interface RS 232 can be done by con necting the interface cable REF 10 093 to the serial interface outlet of the serial PC COM port ARGUS service ARGUS Medical AG v4 3 1 10 O NRGUS service Start the ARGUS service Please select device WT LZ War te Press the button which confirm to your D
27. licate Configuration Replicate Configuration Load Source Configuration From f Currently connected pump Connect target pump now File PIN code of target pump Replicate O Please note The pump internal clock and remainder alarm settings must be done individually on each pump 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 23 148 4 4 5 ARGUS service Toolbox Pump history Each registered event has his own date amp time stamp An event is registered on each pump status change Please refer to the com plete list mentioned in chapter 4 4 6 below View history 4 4 6 History messages Possible messages appearing in the description of each history event No information available Pump start Pump start in remote mode Battery defective Pump stop KVO Rate change during remote mode Battery low prealarm Pump off Pump off in remote mode Battery low pump stop Pump on Enter setup mode Airbubble pump stop Transport off Enter PC configuration mode Door open pump stop Transport on Exit setup or PC configuration mode No drops pump stop Data lock off Downstream occlusion pump stop Not enough drops pump stop Data lock on Upstream occlusion pump stop Too many drops pump stop Rate change Timer alarm pump stop KVO Neonatology mode set in PrL Inf Set change PC communication timeout reached PrLimit change Infusm cleared Logon in PC configuration mode Bolus total reached Bolus start Logoff in PC configura
28. mware gt V4 xx Important remark The actual calibration and configuration will be stored in a file on the PC please be sure you will restore the correct file into the pump after the firm ware update Otherwise invalid calibration values will be stored on the pump 1 Connect the pump to the serial interface of your PC Duve Please remember the COM port number where you have pi lo connected the pump Mio 2 Switch the pump ON while keeping key 10 pressed Set 3 Start the PC configuration tool ARGUS service and se lect the according COM port Go into the configuration part and save the present pump configuration incl calibration to a file Close the ARGUS service and switch the pump OFF 6 Perform point 3 again go into the Update center 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 26 48 7 Select the requested pump firmware file by pressing the button This will update the pump firmware user program part Press Update the pump firmware Follow the instructions displayed on the PC The firm ware will be installed and the pump will be switched off automatically 9 Go into the configuration part again refer to point 2 3 Press the Edit button and enter the pump PIN code de fault PIN after firmware update is O Restore the old configuration incl calibration from the previous created file Restore the configuration by pressing the Downl
29. o Alarmacknowledge with key MODE 0 No 1 Yes KG o PT WC TO Lcd A0 362 8 v Display brightness ll 2 15 Buzzer volume os 316 Default pressure iit erwart 1 10 104 318 20 Droprate window centre IV set1 Drops ml_ 10 65 ee AA NG AN Cl o o Pp pog o vv v vv z v z 399 707 Enterthe calibration menu volume 123 go 707 Enter the calibration menu pressure 4 1234 Note The address does not correspond with the index used by the ARGUS service tool 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 9 48 2 7 Index list of the pump configuration with ARGUS service Index PC Default __ Yes Run indication by running decimal point LI No Yes 2 1 No KeyON OFFonlyatstopvalid No Yes 1 3 No RatechangeallowedonlyatSTOP Nol Yes 1 4 No Key STOP delayed SGT NolYes 5 No Second entry of rate 3 Yes No Yes 6 No Static alarm staff alerting system No Yes 7 L No Display elapsed time in run mode 8 No NolYes 8 No Displayremainingtime L tze No Yes pogo fo jj NA 10 2 No Operation withoutdropdetector LL No Yes 11 3 Yes Recall of ml h rate at next power on No Yes 12 No Recallofmitotalatnextpoweron o No Yes 113 No Recallofmlinf atnextpoweron No Yes 114 No SBS step by step function No Yes
30. oad but ton 12 Perform a standard safety check see chapter 10 nor mally the calibration will not be destroyed if the procedure is carefully performed step by step 5 3 3 Upgrade of a pump with firmware lt V4 xx With the Update center it is also possible to upgrade pump firmware older than V4 xx Important remark The actual calibration and configuration will be stored temporary on the PC please perform the upgrade procedure pump by pump Otherwise invalid cali bration values will be stored on the pump It is urgent necessary to perform a standard safety check see chapter 10 ARGUS Update Center COMB o 1 Go into the Update center see point 1 4 Before ZP Baan patay 4x of ch a pte r 5 S 3 z 2 n8 oY v1 3x i E SAT Wa 2 Switch the pump ON by keeping the key Backup legacy configuration ARGUS 600 v3 01 1 0 pressed 3 Backup the legacy configuration present EE configuration before the firmware update This may take several seconds 4 Switch the pump OFF EH 5 Press Upgrade to bootloader 4 Follow sma ii the instructions displayed on the PC The bootloader will be upgraded then Wett Upgrade to bootloader 4 Update the pump firmware 6 Select the requested pump firmware file This will revert to the legacy bootloader z 46 Revert to the legacy bootloader for using with AMFlasher by pressing the button 7 Press Update the pump firmware Follow the instru
31. ocking Station previously connected device ARGUS pump or docking station Hold the 10 key while switching ON the eg na pump connect te SA Select Start Configuration see next E SN chapter or Start Update Center see chapter 5 3 2 2 Select your PC serial COM port SIN Hold the 10 key while switching ON the pump to enter PC Configuration mode 8 04 00 Start Configurator Switch OFF the pump for updates Start Update Center 14 203 B A pat NG 28 03 06 PJ 19 48 ARGUS service ARGUS Medical AG ER ARGUS708 4 30 Configuration 052 91512 Toolbox 4 2 ARGUS service Configuration Important remark After configuration change a function check Select the next step by pressing one of the buttons configuration calibration or toolbox Configuration and control measurement has to be done 4 24 Configuration tree structure The configuration is split into 4 parts Configuration part 1 All configuration possibilities indexes mentioned in chapter 2 7 can be modified herein in its own tree structure as shown below All indexes which are different from the pump firmware default are high lighted Calibration part 2 Details of the pump calibration can be read out of the pump The calibration cannot be modified herein Statistic part 3 Details about last used infusion parameters total of infused volume and infusion time and so on are shown Also the l
32. on mode without ARGUS service With the interrogation mode you can read the present keypad configuration of the pump without the possibility to modify any configurations For a complete overview please take the ARGUS service PC tool 1 To enter into the interrogation mode switch the pump on while keeping the keys MODE and START STOP pressed mi total mil int bh min 740 r4 DA g e CZ L GG Da Pe LS zeg ore gine ARGUS 707 mi total 2 mil ini hi min t z S e IO ada Bi B ant STARTS Green stream gt ur ARGUS 707 si 3 Enter the reguested address see chapter 2 6 e g 5 The pro 4 To quit the interro grammed value of the address 5 gation mode press appears in the right hand display the ON OFF key bei at least 2 seconds Gi 0 en Flashing decimal points indicate which display is ready to accept an input by the keys 100 1081 In the interrogation mode the left hand display shows the address and the right hand display shows the according value configured at this address Please refer to chapter 2 6 where the meanings of the addresses are explained To modify any configuration data you have to go into the configuration mode 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 5 48 2 3 Configuration mode without ARGUS service The configuration mode permit you to modify the pump keypad configuration manually using the keypad Please refer
33. onnection amp QUICK 100 Docking Station interface 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 3 48 1 INTRODUCTION 1 1 General IMPORTANT This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 707 V infusion pump The service manual is meant to be used together with the user manual IMPORTANT ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 V infusion pump by a non authorized person CAUTION The ARGUS 707 V infusion pump may only be used with spare parts accessories consumables and IV sets with Luer Lock connections recommended by ARGUS Medical AG The functional safety of the pump is not guaranteed if non approved materials are used The safety of the patient may be endangered This manual contains the latest data available It is subject to further modifications in accordance with technical improvements 2 PUMP CONFIGURATIONS 2 1 General CAUTION The configuration possibilities with the ARGUS service PC utility tool and without PC assistance constitute a modification of the pump and may only be carried out by authorized persons CAUTION After changing the configuration a function check and a control measurement has to be performed 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 4 48 2 2 Interrogati
34. portant remark Only the RS232 connector is galvanic separated The docking interface on the pump is a non galvanic isolated interface Do not use the docking interface on the pump together with the interface cable part 10 093 on a patient If the pump is intended to be monitored as a standalone pump on a patient use the RS232 connector RJ45 beside the drop detector connector To use it configure this connector as the default serial interface refer index 55 in the configuration If the pump is docked into a docking station ARGUS 60 M or ARGUS 100 M the software switches automatically to the docking interface and the docking station builds the separation device galvanic isolation then 3 2 Serial communication protocol The following characteristics are basics for all the ARGUS devices volumetric pumps syringe pumps docking stations with V4 xx and PCs which are intended to communicate with the device mentioned in this service manual e Full duplex RS232 currently 4800Baud for single pumps 9600 Baud for docking stations also on master slave link e Simple master host PC slave device communication host does polling e The host has to repeat the request if there is no valid response e Uses a checksum CRC 8 e Binary data transmission thus no ASCll text parsing e Fast amp direct communication with pumps on ARGUS docking station e Specified timeouts during remote mode e Basic framing technique used as in t
35. red Needed equipment a manometer with a resolution of 0 1 bar a 3 way stop cock and a clamping shears the chosen IV set refer to the list including the recommended IV set in the user manual CAUTION The pressure calibration procedure always includes all of the follow ing steps including a control measurement Key functions during pressure calibration By the MODE key you can switch to the next calibration step START By the START STOP key you can store the value of the involved pressure STOP sensor this will be acknowledged by a sound beep By pressing the 1 key you can switch off and on the actual value of the pressure sensor in the display The value is displayed in mV G OG Upstream sensor Downstream sensor mi total 2 mi inf z h min up Set C sexo ee ee N a cd green girpam ARGUS 707 By pressing the 10 key you can switch off and on the motor Don t use it by the standard calibration By pressing the 100 key the pump rate can be changed from 100 to 200 ml h and back toggle action It is recommended to reduce the rate when the pressure is near 1 2 1 0 bar this allows a more precise calibration 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 30 48 6 3 2 Enter into the pressure calibration mode 1 Enter into the configuration mode mi total mil ini Hi miri see chapter 2 3 e l m Aoo Z 10 E r04 STARI d St 00 00 ada
36. s CAL 1 2b 8 The pump continues on with the pres sure build up When the pressure 1 2 bar reaches press START STOP 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 34 48 6 4 Pressure control measurement 6 4 1 Downstream control measurement 1 Insert the IV set and close the door Open the roller clamp 2 Switch on the pump and set the infusion rate at 200 ml h Start the infusion After 10 seconds you have to simu late a downstream occlusion 3 At 700 mbar the pump must stop and release an alarm The toler ance is 150 50 mbar Remark A pressure limit of 700 mbar must not be the default value in menu PrL configur able please select 700 mbar as default for this control measurement If the result of this control measurement does not fulfil the stated requirement a pressure calibration according to chapter Pressure calibration has to be carried out again 6 4 2 Upstream control measurement Start an infusion at a rate of 200 ml h according the user manual Simulate an upstream occlusion by installing a clamping shears see chapter 6 3 6 or by kinking the tube To avoid the drop alarm release simulate falling drops on the drop detector The pump must stop after several seconds and an upstream alarm right hand LED in the pressure bar graph must be activated If the result of this control measurement does not fulfil the stated requirement a pressure calibration according
37. start 2005 06 30714 41 mm d 900 0 700 1 1010100 15 e Ka Rate SEEE 2005 06 20T14 40 18 950 0 40 900 0 700 1 15 e bh Pump start 2005 06 3071 4 40 14 250 0 39 900 0 700 1010100 13 380 All pre alarms alarms and technical failures are high lighted in a different colour 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 25 48 5 SOFTWARE UPDATES 5 1 Did 5 3 General This chapter describes the procedure to perform a software update on the ARGUS 707 infusion pump To check the installed software release in your ARGUS 707 V press the MODE key while switching on the pump Please refer to your local distributor or ARGUS Medical AG to determine the latest software release able to run on your device hardware Requirements for a software update To update an ARGUS Medical device the following items are needed e PC with Microsoft Windows 2000 or newer NET Framework must be installed e RS 232 serial interface cable part no 10 093 e PC configuration tool ARGUS service e Latest firmware included in a text file named A707_xxx txt Cox is the placeholder for the firmware version Those items are available from your local distributor or from ARGUS Medical AG Software update procedure 5 3 1 General Please carefully check the software present installed on the pump If you have a firmware lt version 4 xx please follow chapter 5 3 3 to upgrade the firmware 5 3 2 Update of a pump with fir
38. ta Mata 10 Upstream occlusion sensitviy O eaor 0 60 as 20 Ar summation volume 3199 opo go o Lo 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 12 48 320 9999 Max flow rate 1 999 9 321 999 Max primerate 322 1200 Max bolusrae Set 2 parameters 323 324 10 Upstream occlusion sensitivity factor 0 60 325 20 Airsummationvolume ee 327 250 Airbubblesize 328 20 Dropratewindowcenter BEE ess 330 999 9 Max flow rate S 331 999 Max primerate 332 1200 Max bolusrate Set 3 parameters 333 10 Max bolus total 334 10 Upstream occlusion sensitivity Z actor 0 60 335 20 Airsummationvoume 336 7 Downstream default pressure limit Pr 337 250 Airbubblesize oOo NE ME ee eee MNE baa EE lo gao esu rale a 999 Maxprimerte 342 1200 Maxbolusrae Set 4 parameters 343 10 Max bolustotal 344 10 Upstream occlusion sensitivity tao 0 60 345 20 Airsummationvolume 346 7 Downstream default pressure limit Dr ee 20 Drop rate window center drops 10 65 se Ll od TF a 351 4 Airvolumesummationtime 8min 1 8 z 362 101 8 Displaybrighmess lwe 2 15 ggg Tol o po o o TO o 366 1 Pressure display unit mbar mmHg kP
39. tion mode Pump has detected failure Bolus stop Pump start ext changed parameters Total volume reached pump stop External power on Any defaults written in EEPROM area Total VTBI change External power off CRC error in PC configuration module 4 4 7 History printout example 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 24 48 CB AMService Histo ry REA ion 4 00 Gs i Lal 044 6 529 EMail ARGUS Description lu li ee tgal les Pil vsah Fee coef 3 EL Enter PC configuration mode 2005 DE 30T1 4 45 74 UU UU 0 0 0 0 10000 O Pump off 2005 06 30T 14 45 20 160 0 24 5 840 0 900 1 10101 E0 Pump stop EMU 2005 06 30T 14 45 20 160 0 24 5 840 0 900 10101 Pbb Pump start ib a sali 4 sa a 160 0 24 3 840 0 900 10100 bib Rate change 2005 DE s011 4 44 58 160 0 23 8 640 0 900 bbt Pump start 2005 06 30T 14 44 44 250 0 23 0 840 0 900 RO Pump stop KY0 2005 06 30T 14 44 34 250 0 22 9 900 0 900 PPP Pump start 2005 06 30T 14 44 22 250 0 23 900 0 900 a External power on 2005 06 30T 14 44 04 250 0 21 2 i Battery low prealarm 2005 06 30T14 43 56 250 0 20 8 a External power off 2005 06 30T 14 43 55 250 0 20 5 PrL Pl wot change 2005 06 3071 4 43 46 250 0 20 0 PrL Pl wot change 2005 06 30T 14 43 44 250 0 20 0 Pbb eg start 2005 06 30T 14 43 38 NU TRE 2005 06 30T14 43 02 173 900 0 760 388 pa IE start 2005 06 3071 4 42 32 9 9 900 0 700 1010100 387 be Pan a start 2005 e ar Gig 900 0 700 1010100 13 35 pa Print
40. to chapter 2 6 where the meaning of the adresses are explained To have access to all configuration options please use the ARGUS service PC tool 2 3 1 Step 1 1 To enter into the configuration mode switch the pump on while keeping the keys MODE and START STOP pressed O er eae eee ee 2 Press key START D LS mi total mil inf h min ai start MODE STOP 4 Press key START again The display will show Cod 3 Press key MODE 2 3 2 Step 2 5 Enter the PIN Code If no PIN code was configured no entry is reguired sle 6 Press key START 2 mi inf to acknowledge the lien entered PIN code e Dk Ol VV MVE Ane ai accepted the display ARGUS 707 f will change back to l point 5 If the PIN was not You have now access to all addresses in the list of chapter 2 6 Select therefore any address in the left display see next page 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 6 48 2 3 3 Step 3 8 Enter now the value on the right hand display The After entering an address e g 5 press the range of the value is given by the table in chap 2 6 key MODE The flashing decimal points C 5 6 E C 5 TI Es will change to the right hand display mi total mi int Hi min 100 ao NA 1 04 04 Ce Camp 9 Press key START to ac knowledge the value en woolen ai tered The flashing decimal ARGUS 707 points
41. ton Per ex works settings only IV Set number 1 is calibrated Important remark It is mandatory to perform also a pressure calibration according chapter 6 3 if a new IV set will be used We recommend to do a control measurement after changing the correction factor 4 4 ARGUS service Toolbox With the ARGUS service PC tool you can set the pump clock change PIN code read and print out history etc Toolbox Select the next step by pressing one of the but GE me N tons set clock change PIN code view history O e service interval or replicate Nga wf ZK vr Set clock Change PIN code View history Replicate 4 4 1 ARGUS service Toolbox Pump clock Use this feature to synchronize to pump internal clock with your PC time Please note The pump internal clock will set to the central European time zone Bern CET GMT 1 00h as ex works settings the pump Purp ne PEN internal clock will not switch automatically between diference 0000505156250 summer and winter time All history logs ENE gee a refer to chapter 4 4 5 will base on this time Quit PC time RUSE EN 4 4 2 ARGUS service Toolbox PIN code Change PIN code Use this feature to set the pump PIN code 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 22 48 AMService Set PIN Code The setting of a PIN code prevents access to the pump configuration of third persons The default PIN code is 0 by
42. ubt contact ARGUS Medical AG or your local dealer c configure the set parameters index 310 to 349 and 351 d perform a complete set calibration as per chapters 6 3 to 6 5 14 203 B_A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 14 48 2 9 Medication list 2 9 1 General To display medication names index 29 menu MEd must be set to Yes The se lected medication name can be displayed also in alternate mode rate ml inf med name during infusion for that set index 56 to Yes After enabled special function Med the following medication names can be se lected via pump keypad see user manual 2 9 2 User medication 32 user medication names can be custom defined Choose between capital and small letters for a better displayed medication name Because of the 7 segment pump display some characters maybe difficult to read Index PC Default Function Unit Range 61 User med name 1 char a 63 char O1 d O NO O 32 Di Q l D Q mg Q S 5 5 5 Q CIC ha h char h 70 char ha char C O1 O1 O O1 O O1 O 0 O G O D CO CO N oy amp C Q s CI d N O 3 CI N N 3 N LA ha 74 char a a char O CI 3 OJO Dy Sis ha char ha char ha 83 char h 85 char ha ha char ha char char 592 char O O CI CO d Co NO O gt CI O1 Co d Q 5 O1 OJO sis Q1 O1
43. ure release after occlusion No Yes 45 Yes Pressure display ON LED bar graph 20 40 60 80 100 No Yes 46 No Pressure display withindicator 45 Yes No Yes 47 No Standby and battery pre alarm low volume LI No Yes 1 48 No Flashingnumericdisplayatalam LL No Yes 49 6 No Alarm acknowledge only with key MODE LI No Yes 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 10 48 50 No Startwith21bar pressure allowed 51 Yes Enable upstream occlusion detection SI II s pv poo go 54T 55 No Select RJ 45 connector for serial interface 56 Yes Medication display alternate 29 Yes TH WE 60 No KvO only afterinfusion completed ll 65 No Clearandcontinue 15 No Ul ee 101 Yes Set1 enee 102 No Set2enabled 108 No Set3enabled 104 No Set4enabed S S O 105 No Automatic return on default set 1 after power up 101zYes TH 7 po oos 201 1065 Correction att mh 202 1054 Correctionat3 mih 203 1044 Correctionat5 mh 204 1035 Correctionat10m h SETI 205 1027 Correction at25 ml h 206 1020 Correction at 50 mn Do not change 207 1014 Correction at mm these values unless 1 208 om Correction at 100 min explioiily trained 209 999 Correctionat150 ml h 210 988 Correction
44. utor Further information is available from ARGUS Medical AG CH 3627 Heimberg Switzerland E mail info argusmedical com CAUTION The safety standard check SSC has to be performed at least every 24 month or after 10 000 hrs of operation The check has to be done in accordance to the chap ter 10 No special maintenance of the ARGUS 707 V infusion pump is necessary There are no wear and tear parts Recalibration 6 2 1 General Unless otherwise specified by the customer the ARGUS 707 V has been calibrated by the manufacturer with the CODAN L86 PVC REF 43 4304 infusion set If a dif ferent infusion set is used see recommended list in the appendix of the user man ual a recalibration is required CAUTION A new set calibration always requires pressure sensors up downstream and a volume calibration see chapter 6 3 amp 6 5 It is mandatory to execute first the calibration procedure of the pressure sensors and afterwards the volume calibration 14 203 B A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 29 48 6 3 Pressure calibration 6 3 1 General The volumetric infusion pump ARGUS 707 V contains two pressure sensors One upstream sensor bottle side left input One downstream sensor patient side right output A pressure calibration becomes necessary if the pressure control measurement is not accurate enough a pressure sensor is replaced the main board is replaced or a new set is configu