Home
Kendall SCD™
Contents
1. Error Service Required The system could not sense a pressure rise of more than 5 mmHg during an inflation cycle or during start up Service technician only Check the transducer tube inside the controller and ensure it is neither kinked nor disconnected E13 Low Battery Error Recharge Battery There is less than 15 minutes of battery charge remaining The pump and valves will continue to operate for as long as there is enough power Plug the controller into an AC power outlet Consult the Operation and Service Manual as your warranty may require certain repairs to be performed by an authorized Covidien service technician
2. Kendall SCD 700 Series Controller error Key When the microprocessor detects a fault condition it interrupts the normal operation of the controller deactivates all valves to vent the air from the garment s displays a fault code and sounds an audible error Errors can be reset once the user confirms the cause of the error and remedies the problem If a Garment Mismatch error is triggered the user may remedy the problem by pressing the corresponding A and B Button s Some errors will remain active until the controller is turned off or the battery runs out of charge if operating on battery power The Kendall SCD 700 series controller has three error types User Resettable Manual Reset Required and Service Required User Resettable This type of error allows the user to remedy the issue and resume operation by pressing the A and B button s corresponding with the port affected without powering the unit down For this type of error a check mark will be shown indicating which port is the area of concern A yellow triangle indicates a low concern error If the triangle is red it is indicative of an error related to abnormally high pressure If the error persists then the controller requires service Manual Reset Required An error that can be corrected by the user but requires the device to be powered off and on If the error persists then the controller requires service Service Required Error code is present because of a fa
3. iled internal component It cannot be addressed by the user Error Code Error Type Description Troubleshooting Garment Mismatch Error User Resettable The garment detection procedure has detected a garment configuration leg or foot flashing green that does not match the user selected configuration leg or foot red Press the port configuration button s to turn the foot selection on off depending on what type of garment is connected to the controller If the proper garment is selected and the problem persists have the controller serviced by a professional A B System High Pressure Error Manual Reset required System pressure has exceeded 90 mmHg leg sleeve or 180 mmHg foot cuff Check for kinked tubes or patient interference with the garments like pressing foot against foot board A B E1 High Pressure Leg Sleeves User Resettable Leg sleeve pressure is greater than 47 mmHg for 10 consecutive cycles or pressure is above 65 mmHg for 5 consecutive cycles Check for a tight leg sleeve and adjust fit appropriately Also check for a partially occluded tube A B E2 High Pressure Foot Cuffs User Resettable Foot cuff pressure is greater than 135 mmHg for 10 consecutive cycles or pressure is above 160 mmHg for 5 consecutive cycles Check for a tight foot cuff and adjust fit appropriately Also check for a partially occluded tube A B E2 Low Pressure Leg Sleeves User Rese
4. le The pressure in a garment is greater than 20 mmHg at the end of any vent period Check tubing for kink or occlusion Check garment application too loose or tight Service technician only Check for kinked internal tubing A B E8 A B E3 A B E2 A B E2 COVIDIEN COVIDIEN with logo and Covidien logo and are U S and internationally registered trademarks of Covidien AG Other brands are trademarks of a Covidien company 2013 Covidien 13MAN0079 VTH0304 12 13 Error Code Error Type Description Troubleshooting Temperature Error Manual reset required If the internal case temperature of the controller drops below 5 C 41 F or exceeds 55 C 131 F High temperature Make sure the controller is not covered by bedding and that the fan port located near the power cord is not obstructed Low temperature Allow the system to warm to room temperature E9 E9 Battery Error Service Required Safe battery operation of the controller cannot be ensured Service technician only Ensure that an unauthorized battery pack has not been used with the controller Replace pack or return to Covidien for service E10 Tubing Disconnect Error User Resettable Pressure measured in the inflatable garment is below 10 mmHg for 10 consecutive cycles or no garments are detected during startup Check for disconnected tube sets or garments and reconnect A B E12 Pressure Transducer
5. ttable Leg Sleeve pressure is less than 43 mmHg for 10 consecutive cycles Check for leaks in the sleeve or the tube connections Low Pressure Foot Cuffs User Resettable Foot Cuff pressure is less than 125 mmHg after 10 consecutive cycles Check for leaks in the cuff or the tube connections A B E2 A B E2 Error Code Error Type Description Troubleshooting Low Pressure Leg Sleeves User Resettable Leg Sleeve pressure is not between 35 and 55 mmHg for 12 consecutive cycles Check for leaks in the sleeve or the tube connections Low Pressure Foot Cuffs User Resettable Foot Cuff pressure is not between 110 and 150 mmHg for 12 consecutive cycles Check for leaks in the cuff or the tube connections Valve Feedback Error Service Required If a valve electrically malfunctions this error will be displayed Service technician only Verify that the valve assembly wires are properly connected and confirm solenoid actuation E5 Software Error Service Required Upon startup and periodically during operation the microprocessor performs diagnostic tests If a software error is detected this error will be triggered Return to Covidien for service E6 Compressor Error Service Required If the compressor electrically malfunctions this error will be displayed Service technician only Verify that the compressor wires are properly connected E7 Vent Error User Resettab
Download Pdf Manuals
Related Search
Related Contents
Samsung GALAXY Tab 8.9 WLAN版<br>(P7310/CM16) 用户手册(Icecream_China) Troy-Bilt Super Bronco XP__Lawn Tractor User's Manual 18.2040.00.00 Health Fryer PMC261 User's Guide - CBU Documentation Portail Intel StorageWorks EK-KZPCM-UG User's Manual Beko DPU 8360 W Clothes Dryer User Manual Bedienungsanleitung Philips 69191/14/PH Builders Edge 130120005089 Installation Guide Copyright © All rights reserved.