Home
        BeneHeart R3/BeneHeart R3A Electrocardiograph Service Manual
         Contents
1.     Check that a network connector is available if the equipment needs to be connected to    the wired network     Environmental Requirements    To avoid explosion hazard  do not use the equipment in the presence of flammable    anaesthetics  vapours or liquids  The environment where the equipment will be used should    be reasonably free from vibration  dust and corrosive substances  If these conditions are not    met  the accuracy of the system may be affected and damage may occur     Enviornment    Check and make sure that there is no electromagnetic interference source around the  equipment  especially large medical electrical equipment such as radiological  equipment and magnetic resonance imaging equipment etc  Switch off these devices    when necessary     Keep the examination room warm  no less than 18  C  to avoid muscle action voltages    in ECG signal caused by cold     3 1    M Power supply    Check that power supply specification is met and the power cord is securely connected    if mains power is used  Use only power socket that is properly grounded     AC mains    Input voltage 100 240V   110      Input power    Battery    When powered by a new fully charged battery and at ambient temperature    2 GH   Gs  Battery Type      500 auto measurement reports  or 2 hours of continuous recording  or 6    hours of measurement without recording   With equipment power off and at ambient temperature 25 C 5 TC   Run time  lt 3 h to 90  capacity    lt 3 5 h to 100  capacity 
2.    Shutdown delay at least 5 minutes  after a low battery message first occurs        3 3 Turning On the Equipment    Once the equipment has been installed and checked  you can get ready for measurement and  recording     l  Connect the equipment with AC mains  If you run the equipment on battery power   ensure that the battery is sufficiently charged     2  Press the power on off switch        4 Hardware and Software Upgrade    4 1 Hardware Upgrade    Not applicable     4 2 Software Upgrade    You can upgrade the system by installing and running the Mindray Patient Monitor  Software Upgrade Tool  PN  110 000493 00  on a PC with Windows operating system     Connect the equipment to be upgraded and the PC running the upgrade tool to the same    network  or directly connect the equipment and the PC via a crossover network cable  Then    configure the IP address of the PC  You can upgrade the following software          System software package  Simplified  110 002535 00 l  System softwar Chinese   package System software package  Traditional  Chinese     110 002525 0 Power management program    M501 ECG module STM32 chip writing  Module software 110 002545 00  software    110 001201 00 TR8A recorder drive chip writing software  115 017992 00 Glasgow upgrade pack  FRU  Chinese   Software function l  115 018007 00 Glasgow upgrade pack  FRU  English   upgrade package  115 017996 00 PDF format upgrade pack  FRU     Note  No specific sequence is required for the upgrading of above so
3.   English     Operator   s manual  German     Operator   s manual  Czech     Operator   s manual  Italian     Operator   s manual  Portuguese     Operator   s manual  Portuguese for Brazil     Operator   s manual  Russian     L  2  L  e   2    2        L    53  54    57 040 001584 00 Patient cablem  IEC  12 lead  Clip  58 0010 10 12304 ECG electrode  10 pcs pack    9 2    59    61  62  63  64  65    67    70    71  72  73    9000 10 07469   ECG electrode  50 pes pack    ECG electrode  neonate  3M 2258 3  3  900E 10 04880 y y  pcs pack     PC management software package  for  115 017981 00 l    international     115 017992 00 Glasgow upgrade pack  FRU  Chinese   115 018007 00 Glasgow upgrade pack  FRU  English  y  115 017996 00 PDF format upgrade pack  FRU        9 3    9 2 Material Change    Not Applicable     9 4    A Electrical Safety Inspection    The following electrical safety tests are recommended as part of a comprehensive  preventive maintenance program  They are a proven means of detecting abnormalities that   if undetected  could prove dangerous to either the patient or the operator  Additional tests    may be required according to local regulations     All tests can be performed using commercially available safety analyzer test equipment   These procedures assume the use of a 601PROXL International Safety Analyzer or  equivalent safety analyzer  Other popular testers complying with IEC 60601 1 used in  Europe  such as Fluke  Metron  or Gerb  may require modificatio
4.   Then select  Start      Turn on the equipment to be upgraded  Press and hold  F1  key and  F4  key for 2 to 3  seconds  Then the equipment enters upgrade mode and starts software upgrade  automatically  Corresponding prompt messages are displayed to be upgraded on both    the equipment and PC     When software upgrade is finished  restart the equipment and check if the software is    correctly upgraded     For details of software upgrade  please refer to help and instructions for use of Mindray    Patient Monitor Software Upgrade Tool     ANCAUTION    Discomnect the equipment from the patient and make sure the important data are saved  before upgrade     Do not shut down or power off the equipment when upgrading the system software     Otherwise  it may cause the equipment to break down   Software upgrade should be performed by qualified service personnel only     Crossover network cable is recommended when a PC is connected for software upgrade     NOTE    Make sure the version of the upgrade package is what you desired  To obtain the latest    upgrade package  please contact Mindray Customer Service Department     4 4    5 Testing and Maintenance    5 1 Introduction    To ensure the equipment always functions normally  qualified service personnel should  perform regular inspection  maintenance and test  This chapter provides a checklist of the  testing procedures for the equipment with recommended test equipments and frequency   The service personnel should perform the te
5.  1  aime oro LU Coin CO  a essentielle 5 1  MESE tadas  5 1  5 1 2 AMES CIRC DOU en ee see 5 1  5 13 Preventative Maintenance 2    aaa A ine 5 1  5 1 4 Recommended Frequentes 5 2  3 2  Ylsyal Inspection aan ER AN 5 2  ISA OW EL ORTES eier 5 3    SAE CG Performance Test ann a at ed 5 3    3 3 Recorder check  ae aa Er a 5 3  OBE Check a nen area ABeeRn ae 5 4  SJ Ma  inlchance Mode asien eae eo ee ee 5 4  6 Ir0ubleshootins aus  ee 6 1  COVE IE ee ee en ze ee A 6 1  6 2  Paris Replac A a 6 1  6 3 Ounickhy Identity ime Failure ccs eset kee eit 6 1  64 General Problems AAA E OO nn OR 6 2  IMEI SS 5c cael is dct sto E se Gs eared Sec Ga een E E ciel of E ET ORR E E 6 3  O A A A E 6 4  60 2 Prompt Messase o bete ato de rebote ato de a boe a LO dE 6 6  6 6  Troubleshooting GUIAS aaa ma 6 8  6 6 1 Power On OL FAU ni ri 6 8  02 RECO rd Ca US een acid 6 9  6 6 5 WONO SYSE O a cada 6 10  7 Disassemply and RE Pah 3    022 5 ee a Hans slehsileteiihe 7 1  A AAA A EEE NEIERIEIRSUIE 7 1  T 2 PreparaviOns  Tor DIS e des e pios 7 1  7 3 Disassembling  the Main DU ea aa 7 2  7 3 1 Removing Bottom housing assembly                    cccccccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 7 2  7 3 2  REMOVE the Rower DO lat 7 3  71 3 3 Removing the Main Control Board iii 7 3  734 Removns the DA and 7 4  7 3 5 Removing the Parameter Board         cccccccnncnnccncnonnnnnnnnnnnnnononnnnnnnnnnnnnnnnnnnnnnnnnnnnnss 7 5  720 0  Removing the Key pad  Board se    7 5  1 3 1  Removing the Recorder Drive Bol    ER II
6.  7 6  7 3 8 Removing the Thermal Print Head                                               n 7 7  A NA 8 1  8 1  o rn ere sr en ren rr 8 1  8 2 Mam Ut A SS 8 2  A IOC A A A A A IAS 8 2  A A A ee a 8 3  8 3 Top Housing Assembly  115 017634 00 or 115 017 7983 00    0ooocccccccccocooooooooooonnnnnnnnnnnos 8 5  Sl Exploded VIEW anne 8 5  A tue ace E E E E E E E 8 5  8 4 Recorder door assembly  115 01 7633 00    0ooooccoconocononooooooononononononononnnnnnnnonnnnnnononnnnonnnnnos 8 6  Sl Exploded ies 8 6  8 42 Parts Lita areas 8 6  8 5 Bottom housing assembly  115 017792 00 or 115 01838 7 00   0ooooocoocconooooooooonnnnnononnnos 8 7  SEE 0110  A ee Ne Ne ON 8 7    A ve 1a IG secrete eames O 8 7    9 BeneHeart R3 R3A Replacement Partes             ccccccccccssssssssssssssssssssssscccccsccccccsssssssssssess 9 1  9 1 BeneHeart R3 R3A BOM  V1 0  for Repair  Upgrade and Accessories    9 1  92 Maitena CHAS cis 9 4  A Electrical Safety Inspection   a uuani uni A 1  AE POW T CSU PINTO aa aA A 1  A 2 Device Enclosure and Accessoris   AA AA AAA A AAA A 2  AA A O A te eats A 2  A4 Protective Barth Resistance  tetas A 2  IR Earth Lee ak as CGS ies ssssnorasoerversaceansatcacea E E S A 4  2 6 Patient beakace Current ie eet ea aha A 6  Ac  Mains on Applied Patt Le  kaee en AAA A 7  A 8  Patient Auxiliary Coffent rai RER IRRE A 9  A 9 Scheduled Electrical Safety Inspection esen E RAEE A 11  A 10 Electrical Safety Inspection after Repalr           oooccccccccnccnnnnonononononncnnnncnnnnnnnnnnnnnnn
7.  plugged into the 601PRO  front outlet  If the DUT fails Earth Resistance  discontinue tests and label the    device defective     To Perform the Test  1  From the MAIN MENU  or with the outlet unpowered  plug the DUT into the 601PRO    front panel outlet     2  Attach the 601PRO RED input lead to the device   s Protective Earth terminal or an    exposed metal area   3  Press shortcut key 3  The Protective Earth Resistance test is displayed     4  Press SOFT KEY 3 to select a test current  25AMP   The selected test current is  displayed in the upper right corner of the display     Prot Earth Resistance  Test Current 1A  A    Ohm  Limit 0 000     START TEST CAL LEADS AMPERES    ay aw       e       5  Press START TEST to start the test  The test current is applied while resistance and    current readings are taken  This takes approximately 5 seconds     6  Press the print data key at any time to generate a printout of the latest measurement s      A 3    NOTE      When  Over  is displayed for Ohms  this signifies that a valid measurement was  not obtained because either an open connection was detected or that the  measurement was not within range  Readings greater than 9 999 Ohms will be    displayed as Over     In Case of Failure  Once it reaches the limitation  stop using and inform the Customer Service Engineer for    analysis and disposal     LIMITS  ALL COUNTRIES R   0 2 Q Maximum    A 5 Earth Leakage Test    Run an Earth Leakage test on the device being tested before perf
8.  to  IEC60601 1     SFC  1000uA    INSPECTION AND TESTING    Power Cord Plug    Device Enclosure and Accessories    Device Labeling    Protective Earth Resistance Q    Normal  condition NC     Earth    Leakage  Single Fault    condition SFC     A    IL       Normal  Patient condition NC   Leakage    Max   CF applied part   Current Single Fault NC 10uA  SFC  50uA    condition SFC     Max   CF applied part  SOUA    Mains on Applied Part Leakage    Normal condition NC     Patient Max     Auxiliary    CF applied part   NC 10uA  SFC  SOUA    Current Single Fault  condition SFC        Note  The equipment sold to the United States shall comply with the requirement of  UL60601 1  others shall comply with the requirement of IEC60601 1     Name  Signature     Date     FOR YOUR NOTES    P N  046 005000 00  1 0     
9.  two years     Refer to A 1  Following any repair or replacement    Electric safety  ve Electrical Safety   2  After the equipment drops   ests  Inspection 3  At least once every two years     Recorder check Following any repair or replacement of the recorder     1  When first installed   2  Whenever a battery is replaced     Battery check  Once per year  or if the battery run time reduced  Performance test oe  significantly     5 2 Visual Inspection    Functional test       Inspect the equipment for obvious signs of damage  The test is passed if the equipment has    no obvious signs of damage  Follow these guidelines when inspecting the equipment    NW Carefully inspect the case  display screen  keys  and knob for obvious signs of damage   Inspect the power cord  wall mount  and accessories for obvious signs of damage   Inspect all external connections for loose connectors  bent pins or frayed cables     Inspect all connectors on the equipment for loose connectors or bent pins     Make sure that safety labels and data plates on the equipment are clearly legible     5 3 Power on Test    This test is to verify that the equipment can power up correctly  The test is passed if the    equipment starts up by following this procedure     1  Install the battery  and connect the equipment to AC power  The AC mains indicator    and battery indicator are illuminated   2  Press the Power On Off switch to turn on the equipment     Then the equipment performs self test as soon as the equip
10. 15 017993 00 FRU  y y        orr o1eo 00       Power oaainsinor SY  fs  osioo7 00       TR recontra board ON  o  orons eana    Thermal print head  8 dot mm  72 mm  60  11   024 000134 00 y y  mm sec  FPC lead    E VN  rss   Bentca RA to housing emy N IN  ris  osmosis   Keypad Bent REAL Chin    N  o   1s 017693 00   record doraembly SY  20 aran   2 pogo pros N N  EMT re EE  a  115 017792 00   Bonom housings Sd  as   15 018987 00       Bonom housing asemby Berhe man      V  fas  uno er E    9 1    27 043 003451 00 Battery door  M501 for bidding   28 009 00323 1 00 Recorder cable    Cable between the main control board and  29 009 003325 00    parameter board    Connecting cable between main control board    30 009 003326 00    and keypad board    Cable between the main control board and    31 009 003327 00    power board  32  33    9211 20 87221  046 004650 00  34   046 004651 00  35 046 004737 00  36 046 004738 00 Operator   s manual  French   37   046 004739 00  38 046 004740 00  39 Operator   s manual  Spain   40   046 004742 00  41 046 004743 00  046 004744 00  046 004745 00 Operator   s manual  Turkish  y y     3   046 004817 00   Gago Piscis Guide  ng  O  N  so  osoz   Quick iden  O   so  02200012200   Lone 1 1V 2500 nab EMSOOLA      Y    Patient cablem  AHA  12 lead    4  banana  51 040 001582 00 y  connector  Patient cablem  IEC  12 lead    4  banana  52 040 001579 00  connector            Cable for AC input receptacle  Operator   s manual  Chinese   Operator   s manual
11. BeneHeart R3 BeneHeart R3A    Electrocardiograph    Service Manual    Intellectual Property Statement    SHENZHEN MINDRAY BIO MEDICAL ELECTRONICS CO   LTD   hereinafter called  Mindray  owns the intellectual property rights to this product and this manual  This manual  may refer to information protected by copyrights or patents and does not convey any license  under the patent rights of Mindray  nor the rights of others  Mindray does not assume any    liability arising out of any infringements of patents or other rights of third parties           IL       mora  and MINDRAY are the registered trademarks or trademarks owned             mindray    by Mindray in China and other countries     Revision History    This manual has a revision number  This revision number changes whenever the manual 1s  updated due to software or technical specification change  Contents of this manual are    subject to change without prior notice    NW Version number 1 0   NW Release time  April 2013      2013 Shenzhen Mindray Bio Medical Electronics Co   Ltd  All rights reserved     Preface    Manual Purpose    This manual provides detailed information about the assembling  dissembling  testing and  troubleshooting of the equipment to support effective troubleshooting and repair  It is not  intended to be a comprehensive  in depth explanation of the product architecture or  technical implementation  Observance of the manual is a prerequisite for proper equipment    maintenance and prevents equipment d
12. List 2    Data memory Data memory is unavailable or Contact your service personnel    unavailable    cannot detect the data memory    Data memory error    Unable to read or write the data Contact your service personnel   memory     RT clock need reset  The real time clock displays the Contact your service personnel     initial value because button cell  failed and reset  or button cell is not    available     RT clock error  Unable to read the real time clock Contact your service personnel   register    Battery error   Failure is detected when the battery   Contact your service personnel   1s being charged    Device abnormal The voltage of PCBA power supply   Contact your service personnel    voltage j 1s abnormal     Limb lead off       1  RL lead off or more than one 1  Check corresponding electrodes  limb lead off  and lead wires  Re apply the  2  Patient cable is detached from electrodes or reconnect the lead wires  the equipment  if necessary    2  Check that patient cable is    properly connected to the equipment     XX Lead off    The referred lead is off  Check corresponding electrodes and     XX refers to LA F  lead wires  Re apply the electrodes or    LL L  V1 C1 V6 C6  reconnect the lead wires if necessary        Noise or artifacts from lead I  II  Check the patient   V1  V2  V3  V4  V5  V6 is  detected     Noise       Recording    The recorder is working  To stop printing  press the  ECG  key   for auto measurement recording    or  Manual  key  for manual  measure
13. amage and personnel injury     Intended Audience    This manual is for biomedical engineers  authorized technicians or service representatives    responsible for troubleshooting  repairing and maintaining the equipments    Passwords  A password may be required to access different modes  The passwords are listed below   NW Maintain Mode  888888  E Demo Mode  2088    II    Contents    DGC A TN 1 1  A I  ORY A teste ti ieee tee 1 1  1  Sale Info is 1 1  ELA DANGER E E aeeeee 1 2  122 WaN iS een 1 2  A A A T 1 2  LS A at otras toe tat 1 3  ELSE Symbol O 1 3  2 Theory of Operation sii adsl 2 1  O E TON ee es 2 1  2 2 Connectors Tor External E en 2 2  23 Main Unit a hes Ba a cae ne ea Het la a Ina 2 4  2 4  lop housins as MD een ehe ueisieheie  2 5  2 42   Mam E00troL Board une ee ee i   2 5  242  Keypad DO nen ern 2 6  243 AQ DE Power Bostd  u  nun denies an ateros a 2 6  24 4 Recorder DIVE Boar rn en ee an en ee 2 6  PAo Parame er Dosen ran 2 6  3 Unpacking and Installation 3 1  a in 3 1  32 Prepatation Tor Installations  idad dabas 3 1  33 Turnins On tne Equipment ie cobos Nd  3 2  A Hardware and Software Upgrade s  cisesiscscscsicescseei essen 4 1  41  Hardware Ur bc a 4 1  RS A A A A IA 4 1  4 2 1 Installing Software Upgrade Tool               cccccccccnnncnnnnnnnnnnnnnnnnnnnnnnonononnnnnnnnnnnnnnos 4 2  4 2201 Ware   pgrade  Proceed  re mi 4 4   gt  Testing  and  MAINTENANCE ar cscs cs cscceendceciessscecdeeeiveudoscedecDoevivwedoceses bossuerenveceesaveetevstseoseeees 5 1  9
14. and operate it on battery power   if possible    eo Dispose of the package material  observing the applicable waste control regulations    and keeping it out of children   s reach     1 2 3 Cautions       CAUTION    Make sure that no electromagnetic radiation interferes with the performance of the  equipment when preparing to carry out performance tests  Mobile phone  X ray  equipment or MRI devices are a possible source of interference as they may emit    higher levels of electromagnetic radiation     Before connecting the equipment to the power line  check that the voltage and  frequency ratings of the power line are the same as those indicated on the    equipment   s label or in this manual     Protect the equipment from damage caused by drop  impact  strong vibration or    other mechanical force during servicing     1 2    1 2 4 Notes    NOTE    o Refer to Operation Manual for detailed operation and other information     1 3 A a Symbols    Caution  Attention  consult ON OFF part  accompanying documents  equipment  EN Battery indicator  gt  Alternating current  AC     DEFIBRILLATION PROOF  TYPE CF APPLIED PART    E remar eK mr  DATE OF  S N Serial number  MANUAFACTURE      Equipotentiality    Dispose of in accordance to   AY General symbol for    your country   s requirements E recovery recyclable    CCC Mark   China Chi et  ina etrolo   Compulsory Certification M GC l   2  Certification   mark   Environment friendly Used Authorized representative   Period per Chinese Stan
15. condition  Patient  NC  LICF uA Max   Leakage CF applied part   g   BF WA ppl  ed p  Current Single Fault     NC 10uA  SFC  50uA  condition  SFC  OCF uA  LIBF uA      Max   7 Mains on Applied Part Leakage    OCF uA CF applied part  SOUA  E BF uA  Normal condition  Patient  NC  LICF uA Max   Auxilia CF applied part   ry   u BF mn ppl  ed p  Current Single Fault condition T  NC 10uA  SFC  50uA   SFC  OCF uA       Note  The equipment sold to the United States shall comply with the requirement of  UL60601 1  others shall comply with the requirement of IEC60601 1     Name  Signature   Date     A 10 Electrical Safety Inspection after Repair    The following table specifies test items to be performed after the equipment is repaired     Repair with main unit not disassembled Test items  1  2  3    Repair with   When neither power supply Test items  1  2  3  4  main unit PCBA nor patient    disassembled   electrically connected PCBA is    repaired or replaced    When power supply PCBA is Test items  1  2  3  4  5  repaired or replaced    When patient Test items  1  2  3  4  6  7  8  electrically connected PCBA is    repaired or replaced   When both power supply Test items  1  2  3  4  5  6  7  8  PCBA and patient   electrically connected PCBA are    repaired or replaced       ELECTRICAL SAFETY INSPECTION FORM    Equipment  Control Number     Manufacturer  Model     Measurement equipment  SN  Date of Calibration   Max 0 2 Q    Max    NC  300uA  refer to  UL60601 1      NC  500uA  refer
16. could be used instead   Change another probe to confirm if the fail is caused by console     If the leakage current measurement tests fail on a new unit and if situation can not be  corrected  submit a Safety Failure Report to document the system problem  Remove    unit from operation     If all else fails  stop using and inform the Customer Service Engineer for analysis and    disposal     NW For CF applied parts  50 uA    A 8 Patient Auxiliary Current    Patient Auxiliary currents are measured between any selected ECG jack and the remaining    selected ECG jacks  All measurements may have a true RMS only response     Preparation    l     From the MAIN MENU  or with the outlet unpowered  plug the DUT into the 601 PRO    front panel outlet  and turn on the device   Attach the patient leads to the 601PRO ECG jacks     Define the Lead Types from the View Settings Option  refer to  Lead Type Definitions  in Section 5 of this chapter      Press shortcut key 8  The Patient Auxiliary Current test is displayed  and the test  begins immediately  Display values are continuously updated until another test is    selected     Patient Auxiliary Cur  RA AII  Outlet  Norm Pol  Earth  L2    OuA  Limit Inv        DUT OFF NO EARTH NOL2 APPLIED PART    aD ED AD AD AD ED    5  Press SOFT KEYS 1 4 to select leakage tests  6  Press APPLIED PART  SOFT KEY 4  at any time to select the desired applied part    leakage current     7  Modify the configuration of the front panel outlet by pressing the ap
17. dard in the European   SJ T 11363 2006 Community    The product bears CE mark indicating its conformity with the provisions of       en the Council Directive 93 42 EEC concerning medical devices and fulfils the    essential requirements of Annex I of this directive        NOTE      Some symbols may not appear on your equipment     1 3    FOR YOUR NOTES       2 Theory of Operation    2 1 Introduction    BeneHeart R3 BeneHeart R3A Electrocardiograph  hereafter referred to as    the equipment     or    the system     is intended to acquire  analyze  display  store  and record the patient   s    electrocardiographic information for clinical diagnosis and study     The equipment provides the following functions     M Display 12 lead ECG waveforms in real time  print manual  auto  and rhythm    measurements     NW Perform resting ECG interpretation  output measurements and diagnosis in auto    measurement mode   NW Save ECG data in internal memory and export the data through a USB memory     NW Enables program upgrade over the network        Mobile Viewer  r    2 1    The above figure shows the system consists of the electrocardiograph and its peripheral    devices  The electrocardiograph    NW Acquires  processes and reviews 12 lead resting ECG data   M Prints ECG reports    NW Runs on AC power source or a battery    ii    Exports data through a USB memory     2 2 Connectors for External Equipment    The connectors for external devices are located on the left and rear of the equip
18. e insulator     AC input receptacle    and cable       7 3 3 Removing the Main Control Board    1  Release the latch locking the flexible PCB connecting the display to the main control  board by pushing the latch from Position 1 to Position 2 as shown in the picture below   Disconnect the flexible PCB     1      l N    Te    cee eaten ages    a  se ral le men MARA AAA A            i a aN    1 1 i                  i                      NOTE    o Make sure to disconnect the flexible PCB between the main control board and the  display before removing the main control board     7 3    2  Unscrew the four M36 screws  Disconnect the recorder cable  the connecting cable  between the main control board and the parameter board  the connecting cable between    the main control board and the keypad board  and the connecting cable between the    main control board and the power board        7 3 4 Removing the Display    Before removing the display  remove the main control board as described in 7 3 3    Removing the Main Control Board     Then you can remove the display  follow this procedure   1  Unscrew the four PT3x6 tapping screws to remove the displayfixing kit   2  Remove the TFT display        7 4    7 3 5 Removing the Parameter Board    1  Unscrew the four M3x6 tapping screws     2  Disconnect the cable between the main control board and the parameter board        7 3 6 Removing the Keypad Board    Before removing the keypad board  remove the parameter board as described in Secti
19. ectly loaded          Set    Paper Type    correctly       6 6 3 Wrong system time       Displayed system time is  wrong                             Message     RT clock error     is presented     Replace main  control board       Y          Reset system time    yo OK  gt           No cell button or  cell button failed    Y      Replace button cell         N    y           Replace main control board  gt        6 10    F   Disassembly and Repair    7 1 Tools Required    To disassemble and replace the parts and components  the following tools may be required     Philips screwdrivers  Tweezers   Sharp nose pliers  Clamp    7 2 Preparations for Disassembly    Before disassembling the equipment  finish the following preparations     NW Stop measuring  turn off the equipment  and disconnect all the accessories and  peripheral devices    EW Disconnect the AC power source and remove the battery     N WARNING   o Before disassembling the equipment  be sure to eliminate the static charges first   When disassembling the parts labeled with static sensitive symbols  make sure you  are wearing electrostatic discharge protection such as antistatic wristband or gloves  to avoid damaging the equipment    e Properly connect and route the cables and wires when reassembling the equipment to  avoid short circuit    o Select appropriate screws to assemble the equipment  If unfit screws are tightened by  force  the equipment may be damaged and the screws or part may fall off during use   causing un
20. ent system status     Some messages  see Message List 1  are more important and urgent  and need you to  acknowledge or take actions in time  The system pops up a dialog box when these messages  happen  In this case  you cannot operate the equipment unless you press any key to clear the    messages or wait till the triggers disappear     Some messages  see Message List 2  are less important  These messages are shown in the    message area  They disappear automatically when the triggers disappear     The equipment can give a notification tone when a message is presented  The notification    tone is switched off by default  You can enable it by accessing the  System Setup  menu    However  the equipment always gives a notification tone when some messages occur    regardless of the setting of  Notification tone   Refer to the messages followed by an    asterisk in the Messages Lists below     6 3    6 5 1 Message List 1    Battery depleted        Recorder unavailable     1  When recording is needed or a  measurement is started  In this case  a dialog box pops up  You can press  any key to close the dialog box  It  does not affect the function   2  Recorder communication error or    does not work     ECG module error     failure causes ECG communication    error or communication stops     Paper type error i    different with the used paper type     The battery is too low     Damaged ECG board or software    The setting of  Paper Type  is       Connect the equipment to AC  mains t
21. er or owner to correct any deviations  As a work around  check the other    outlets to see if they could be used instead   NW Change another probe to confirm if the fail is caused by console     m Ifthe leakage current measurement tests fail on a new unit and if situation can not be  corrected  submit a Safety Failure Report to document the system problem  Remove    unit from operation     NW If all else fails  stop using and inform the Customer Service Engineer for analysis and    disposal     LIMITS    For UL60601 1     300 uA in Normal Condition    1000 uA in Single Fault Condition    For IEC60601 1     500 uA in Normal Condition  9 1000 uA in Single Fault Condition    A 6 Patient Leakage Current    Patient leakage currents are measured between a selected applied part and mains earth  All    measurements have a true RMS only response     Preparation    Perform a calibration from the Mains on Applied Part menu     The following outlet conditions apply when performing this test    NW Normal Polarity  Earth Open  Outlet ON Normal Polarity  Outlet ON   WE Normal Polarity  L2 Open  Outlet ON Reversed Polarity  Outlet ON   NW Reversed Polarity  Earth Open  Outlet ON Reversed Polarity  L2 Open  Outlet ON     NWARNING    o Ifall of the applied parts correspond to the instrument type  the applied parts  will be tied together and one reading will be taken  If any of the applied parts  differ from the instrument type  all applied parts will be tested individually   based on the type 
22. ftware  For detailed       information  please refer to 4 2 2 Software Upgrade Procedure     4 2 1 Installing Software Upgrade Tool    1  Find the installation program Ef SystemUpdateTool  exe  and double click it to start  installation     2  Select language     3  Click  Ok  and the following screen 1s displayed  Click  Next  to go to the next step     Welcome       Welcome to Mindray Patient Monitor System Update Tool install program  This program install  Mindray Patient Monitor System Update Tool into your computer     Click  Cancel  to Exit the install program     Click  Next  to continue the install program     The program can Upgrade the Patient Monitor of Mindray Co  Ltd     If you don t have Patient Monitor of Mindray the program can t work If you want purchase  patient monitor of mindray please contact with us     Authorization    The program need license  if you haven t now please contact with us to ask for one        cnoe         4  Enter User Name  Company name  and Serial Number  26582640   Then click  Next      Customer Information  Please enter your information        Please enter your name  the name of the company for whom you work and the product  serial number     User Name    ities  Company Name               o oo    Serial Number     cos       4 2    5  Specify the destination folder for installing this program  Then select  Next      Choose Destination Location  Select folder where Setup will install files        Setup will install Mindray Patient Monit
23. h voltage is present at applied part terminals while measurements are being    taken     To Perform the Test  l  From the MAIN MENU  or with the outlet unpowered  plug the DUT into the 601  2  Attach the applied parts to the 601PRO applied part terminals   3  Attach the red terminal lead to a conductive part on the DUT enclosure   4  Press shortcut key 7  The Mains on Applied Part test is displayed     Mains on Applied Part  All Earth  Outlet  Norm Pol  Earth  L2    Norm uA Rev UA  Limit Inv     START TEST CAL DUT OFF APPLIED PART    ED AD as   eas AD aD       5  Select the desired outlet configuration and applied part to test using the appropriate  SOFT KEYS     6  Press START TEST  SOFT KEY 1  to begin the test     7  Press the print data key to generate a printout of the latest measurement     NOTE    If all of the applied parts correspond to the instrument type  the applied parts  will be tied together and one reading will be taken  If any of the applied parts  differ from the instrument type  all applied parts will be tested individually   based on the type of applied part  This applies to Auto and Step modes only     In Case of Failure    LIMITS    Check any broken of the enclosure  Replace any defective part   Inspect wiring for bad crimps  poor connections  or damage     Test the wall outlet  verify it is grounded and 1s free of other wiring abnormalities   Notify the user or owner to correct any deviations  As a work around  check the other    outlets to see if they 
24. k whether the USB memory is  file s  are being exported  properly plugged  If yes  the file    system might be damaged  Format    the USB E ES and try again        Data export completed Patient data 1s successfully exported     Export configuration Configuration is successfully  successfully exported     Export failed Exporting patient data failed  1  Check that the settings are  correct and a correct option is    select     2  Check that USB memory is    properly inserted and file system is    not damaged     3  Check that the USB memory has    sufficient              File s  are being deleted     Deleted successfully A single file or all files are  successfully deleted     Deleting failed Deleting file s  failed  Check that deleting option is    selected  You can format the    internal memory 1f you want to  delete all the files         USB memory low The USB memory has insufficient Delete useless files stored in the  space when patient data or USB memory to release the  configuration is to be exported to the   memory space     USB memory     Formatting failed Formatting memory failed  Internal memory might be    damaged  Contact your service    mz    Formatting completed The memory is successfully         Formatting  Please wait    Please wait      The memory is   The memory is being formatted  formatted     Shutting down    The system is shutting down  BO    Note    means that the equipment always gives a notification tone when the message occurs     6 5 2 Prompt Message 
25. ll else fails  stop using and inform the Customer Service Engineer for analysis and    disposal     LIMITS    For CF applied parts  9  10uA in Normal Condition    50uA in Single Fault Condition    A 7 Mains on Applied Part Leakage    The Mains on Applied Part test applies a test voltage  which is 110  of the mains voltage   through a limiting resistance  to selected applied part terminals  Current measurements are  then taken between the selected applied part and earth  Measurements are taken with the  test voltage  110  of mains  to applied parts in the normal and reverse polarity conditions    as indicated on the display     The following outlet conditions apply when performing the Mains on Applied Part test   M Normal Polarity     NW Reversed Polarity    Preparation    To perform a calibration from the Mains on Applied Part test  press CAL  SOFT KEY 2    1  Disconnect ALL patient leads  test leads  and DUT outlet connections   2  Press CAL to begin calibration  as shown     Disconnect All Patient Leads  and OUTLET connections    CAL    PREVIOUS  dG Tr wo   DD       If the calibration fails  the previously stored readings will be used until a passing    calibration has occurred  Also  the esc stop key has no effect during calibration     3  When the calibration is finished  the Mains on Applied Part test will reappear      NWARNING      A 2 beep per second signal indicates high voltage present at the applied part    terminals while a calibration is being performed     o Hig
26. ment     Side View       l  Patient cable connector  connects the patient cable for ECG acquisition    2  Recorder door latch  opens the recorder door       1  Network connector  connects the equipment to the network for software upgrade    2  USB connector  connects USB disc for data transfer    Back View       1  AC power input  2   Equipotential Grounding Terminal  When the equipment and other devices are to be    used together  their equipotential grounding terminals should be connected together to    eliminate the potential difference between them     Bottom View       1  Battery compartment    2 3 Main Unit    The equipment consists of the following parts     M Top housing assembly  consists of main control board  keyboard assembly  display     power switch  power board  AC DC   recorder  and parameter board     M Lower housing  includes a battery compartment     The following figure shows the main unit architecture                                                                                                     Thermal head  module  W7  Damen hoard Recorder drive  Keypad board board  Wil   W3  W2  USB   T  RJ45 Main control board W4 Parameter Board      9    WS W6  B Wi Fi  En LCD  reserved                                2 4 Top housing assembly    Bosses for screwing  top and bottom    housings together    Recorder drive board Power board    Bosses for screwing  top and bottom    housings together       Keypad board Parameter board Main control board    2 4 1 Main C
27. ment is powered on     5 4 ECG Performance Test    Tool required   Patient simulator  eg  Medsim300B or equivalent equipment  Test Method     1  Connect the patient cable to the equipment under test and the lead wires to the patient    simulator     2  Set the patient simulator as follows  ECG sinus rhythm  HR   80 bpm  with the  amplitude as 1 mV  default setting      3  Check that the displayed ECG waveforms are clear and complete without obvious    noise     4  Press the  ECG  key to start automatic ECG measurement  Print out the waveforms    and diagnosis     5  Check that the displayed ECG waveforms are clear and complete with the amplitude of  lead II waveform about 10 mm  assumed the Gain is set to 10mm mV   and diagnosis    indicating sinus rhythm and HR 80 bpm     6  Disconnect each of the leads in turn and observe the corresponding lead off message    displayed on the screen     5 5 Recorder check  Tool required   E None    1  Print ECG waveforms  The recorder should print correctly and the printout should be    clear     2  Simulate some recorder problems  such as out of paper  paper jam  and etc   and then  the equipment should present the message  Check Recorder   After the problem is    removed  the recorder should be able to work correctly     5 3    5 6 Battery Check    Tool required     E None    Functional Test    1  Ifthe equipment is installed with a battery  remove the battery first     i    Verify that the equipment works correctly when running powered f
28. ment recording  or the   Rhythm  key  for rhythm    measurement recording      Analyzing    The algorithm is analyzing acquired      ECG data     Analyzing Failed The algorithm fails to analyze Refer to   12 Lead ECG Interpretive       acquired ECG data and is unable to   Program Physician   s Guide     PN   give disagnoses  046 004817 00         ECG data insufficient In the situation that pre acquisition   Wait till sufficient data is acquired   is enabled  the equipment has not  acquired 10 seconds of ECG data    when auto measurement is started     Acquiring    The equipment is acquiring Wait till 60 seconds of countdown 1s  60 second ECG data when rhythm   reached  To stop acquisition  press  measurement is started  the  Rhythm  key     Please check recorder 1  Recording paper is not loaded or   1  Verify that recording paper is  runs out  loaded and sufficient     2  Recorder door is not properly 2  Verify recorder door is properly    closed  closed     Recorder head hot   Print head has heated up do to Stop printing and wait till the  aie tale OO use  aa disappears     IP IP address conflict   conflict IP IP address conflict    conflict  Contact your service Contact your service personnel     The memory space Left memory space is less than 10 Delete useless historic files   insufficient files     Low battery Battery charge is low  Connect the equipment to AC mains    to run the equipment and charge the       battery     Note    means that the equipment always gives a no
29. nnanos A 12    FOR YOUR NOTES    1 General    1 1 Revision History    Manual Version Description of Changes Change Time   SN Since  Premios         1 2 Safety Information     NDANGER    o Indicates an imminent hazard that  if not avoided  will result in death or serious    injury      N WARNING    eo Indicates a potential hazard or unsafe practice that  if not avoided  could result in    death or serious injury      NCAUTION    eo Indicates a potential hazard or unsafe practice that  if not avoided  could result in    minor personal injury or product property damage     NOTE      Provides application tips or other useful information to ensure that you get the    most from your product     1 1    1 2 1 DANGER    There are no dangers that refer to the product in general  Specific    Danger    statements may    be given in the respective sections of this manual     1 2 2 Warnings    NWARNING   eo All installation operations  expansions  changes  modifications and repairs of this  product are conducted by authorized personnel    eo Disconnect the equipment from external power source and remove the battery  before disassembling the equipment    eo When you disassemble reassemble the equipment  a patient leakage current test  must be performed before it is put in use    o The equipment must be connected to a properly installed power outlet with  protective earth contacts only  If the installation does not provide for a protective  earth conductor  disconnect it from the power line 
30. ns to the procedure  Please    follow the instructions of the analyzer manufacturer     The consistent use of a safety analyzer as a routine step in closing a repair or upgrade is  emphasized as a mandatory step if an approved agency status is to be maintained  The  safety analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of    line voltage and grounding  as well as total current loads     A 1 Power Cord Plug    Test Item Acceptance Criteria  The power plug pins No broken or bent pin  No discolored pins   The plug body No physical damage to the plug body     The power plug l No physical damage to the strain relief  No plug  The strain relief 8  warmth for device in use     No physical damage to the cord  No deterioration to   the cord    For devices with detachable power cords  inspect  The power cord   the connection at the device    For devices with non detachable power cords    inspect the strain relief at the device        A 2 Device Enclosure and Accessories    A 2 1 Visual Inspection    Test Item Acceptance Criteria    No physical damage to the enclosure and accessories     No physical damage to meters  switches  connectors  etc     The enclosure and accessories No residue of fluid spillage  e g   water  coffee  chemicals     etc          No loose or missing parts  e g   knobs  dials  terminals  etc       A 2 2 Contextual Inspection    Test Item Acceptance Criteria    No unusual noises  e g   a rattle inside the case        No unusual 
31. nts     Cr O  Er  SS  Dr  Dr    OO  OO    Equipment under test   EUT      Hardware version     Software version J  PO O  PO E  IO E  IO AO HE  PE E  PS E  A    5 5       Test Result  Test Contents Test Record     Pass Fail        Visual Inspection    The housing  display screen  keys  power cord  and accessories have       no obvious signs of damage     The external connecting cables are not frayed and the connector pins  aoa AS   Tnsemlomnesmwenstkosseriir mama  i  The safety labels and data plate are clearly legible  S       Power on Test    The power on self test is passed  The power indicator works  correctly and the equipment starts up properly     Performance test          ECG performance test and calibration    The ECG waveforms displayed and printed out should be clear and  complete without obvious noise  The amplitude of lead II waveform  is about 10 mm and diagnosis indicate sinus rhythm and HR 80    ECG Lead Off prompt behaves correctly  a    Recorder check    The recorder can print ECG waves correctly and the printout is  clear    Simulate recorder problems  such as out of pater and paper jam  the  message  Check Recorder  shall be presented           Battery Check    The equipment can operates correctly from battery power when an  AC power failure accidentally occurs     The equipment can operate independently on a single battery     Test conclusion     Tested by  Test date     5 6    6 Troubleshooting    6 1 Overview    In this chapter  equipment problems are li
32. o run the equipment and    charge the battery     1  Verify that recording paper is   properly loaded    2  Verify the platen of the print head  is in position    3  Verify that the recorder does not   stop due to hot print head    4  If the message persists after   above actions have taken  contact    your service personnel     Contact your service personnel     Verify the setting of  Paper Type   is correct by selecting  Menu      System Setup      Configuration loaded The configuration is successfully     successfully loaded    Configuration restored Default configuration 1s successfully       successfully i restored    Loading configuration Main control software or hardware Contact your service personnel   failed    fails     USB memory not found   The system fails to find the USB    memory     Configuration file not Configuration file is not found in the  found    USB memory when  Load    Configuration  is selected     Exporting data  Please The system 1s exporting patient data     wait     X Y  X refers to the number of files having  been exported  while Y refers the    total number of files to be exported        1  Verify the USB memory is  properly plugged   2  If the message persists  format    the USB memory and try again     1  Verify that correct configuration  file is stored in the USB memory   2  Check whether the file system is  damaged  If yes  contact your    service personnel     Failed to create file s           The system failed to create files when   Chec
33. of applied part  This applies to Auto and Step modes only     To Perform the Test    l  From the MAIN MENU  or with the outlet unpowered  plug the DUT into the 601PRO  front panel outlet  and turn on the device     2  Attach the applied parts to the 601PRO s applied part terminals     3  Press shortcut key 6  The Patient Leakage test is displayed  and the test begins    immediately     Patient Leakage  All Earth  Outlet  Rev Pol  Earth  No L2    OuA  Limit Inv        DUT OFF NO EARTH L2 APPLIED PART    m aw ep ee awe Gp    4  Press APPLIED PART  SOFT KEY 4  at any time to select the desired applied part    leakage current     5  Modify the configuration of the front panel outlet by pressing the appropriate SOFT  KEY on the 601PRO     6  Press the print data key at any time to generate a printout of the latest measurement     A 6    In Case of Failure    M Check any broken of the enclosure  Replace any defective part   NW Inspect wiring for bad crimps  poor connections  or damage     M Test the wall outlet  verify it is grounded and is free of other wiring abnormalities   Notify the user or owner to correct any deviations  As a work around  check the other    outlets to see if they could be used instead   NW Change another probe to confirm if the fail is caused by console     m Ifthe leakage current measurement tests fail on a new unit and if situation can not be  corrected  submit a Safety Failure Report to document the system problem  Remove    unit from operation     NW  fa
34. om housing assembly   assembly  Masa Bottom housing assembly a   BeneHeart R3A     15   022 000122 00 000122 00 Li ion battery  11 1V  2500 mAh        8 3    SN   PN Description FRU part   Remarks  number    ar GE HE BE  16  o43 003114 00   Battery door      043 003451 00 _  Battery door M501 forbidding       17   043 002861 00   Z fold paper jam protector     18   115 017633 00   Recorder door assembly     19   043 00311200   Recorder tink    o AAA    AAA    BEE EEE  EEE  EEE  BEE EEE    mm  60 mm sec  FPC lead       2  2  2  2  2    8 3 Top Housing Assembly  115 017634 00 or  115 017793 00   8 3 1 Exploded View       8 3 2 Parts List    OT 047 010209 00   Display lens a  E 043 003107 00 Upper housing       Display dust proof strip  short      Display dust proof strip  long  9211 20 87221 AC input receptacle and cable       8 5    8 4 Recorder door assembly  115 017633 00     8 4 1 Exploded View          8 4 2 Parts List  S  1 j      Z fold paper jam  protector    ing  rn       8 6    8 5 Bottom housing assembly  115 017792 00 or  115 018387 00   8 5 1 Exploded View       O    8 5 2 Parts List    FRU part  SN   PN Description Remarks  number    a Recorder instruction labelling rm  apoo O VE VE  a    Device Iabelling  Chinese  CA DEE       8 7    FOR YOUR NOTES    8 8       9 BeneHeart R3 R3A Replacement Partes    9 1 BeneHeart R3 R3A BOM  V1 0  for Repair  Upgrade    and Accessories    051 001353 01   Main control board  512MB  y    Main control board  512M Glasgow PDF   2 1
35. on    7 3 5 Removing the Parameter Board     Then follow this procedure to remove the keypad board   1  Unscrew the four PT3x6 tapping screws     2  Disconnect the cable between the main control board and the keypad board  Then  remove the keypad board     j     i    8  F d r y   k  y e   a  TTT TE i a    AAA a   A A O  I a       7 3   Removing the Recorder Drive Board    1  Unscrew the two M3x6 screws  Disconnect the recorder cable       NOTE    o Make sure to the flexible PCB which is located under the TR8A recorder and  connects the thermal print head before disassembling the recorder drive board        2  Release the latch locking the flexible PCB connecting thermal print head to the  recorder drive board by pushing the latch from Position 1 to Position 2 as shown in the    picture below  Disconnect the flexible PCB and remove the recorder drive board        7 3 8 Removing the Thermal Print Head    Before removing the thermal print head  remove the recorder drive board as described in    Section 7 3 7 Removing the Recorder Drive Board    Then follow this procedure to remove the thermal print head     1  Unscrew the two M3x6 screws  and the take out the thermal print head        NOTE    e Open the recorder door before removing and installing the thermal print head     FOR YOUR NOTES    7 8       O Parts    8 1 Introduction    This chapter contains the exploded views and parts lists of the main unit  It helps the  engineer to identify the parts during disassembling the equi
36. ontrol Board    The main control board provides the following functions     Drive display and control back light   Store data   Process printing   Implement serial port communication   Drive buzzer   Provide USB and wired network connection  Process data   Implement power supply conversion  Charge the battery    Implement power management    2 4 2 Keypad Board    The keypad board drives the keys  The power on off switch  AC indicator and battery    indicator are also located on the keypad board     2 4 3 AC DC Power Board    The power board transforms AC power to DC power     2 4 4 Recorder Drive Board    The recorder drive board drives the recorder module     2 4 5 Parameter Board    Parameter board is responsible for ECG data acquisition     2 6    3 Unpacking and Installation    3 1 Unpacking the Equipment    Open the package and take out the packing list  Check that all the articles included in the    packing list are available and the quantity and specification are correct     All the optional parts purchased by the customer shall also be checked     Notify the supplier 1f provided components are not correct as compared to the packing  list     In case of damage during transportation  keep the packing material and notify the    supplier immediately     Keep the packing material till the equipment is accepted     3 2 Preparation for Installation    Ensure that the site meets all safety  environmental and power requirements  Check that required power sockets are available 
37. or System Update Tool in the following folder     To install to this folder  click Next  To install to a different folder  click Browse and select  another folder        CA    AMindray Patient Monitor System Update Tool Browse           Destination Folder          InstallShield     lt  Back   Next  gt    Cancel         6  Select Program Folder  Then select  Next      Select Program Folder    Please select a program folder        Setup will add program icons to the Program Folder listed below  You may type a new folder  name  or select one from the existing folders list  Click Next to continue     Program Folders      s ystem Update Tool    Existing Folders     Accessories  Administrative Tools  Cisco    InstallShield        lt  Back Cancel         7  Click  Finish  to complete installation     Mindray Patient Monitor System Update Tool install sucessfully   Thank you for selecting Mindray product we will provide more  service and surport for you        Bact Cancel         4 2 2 Software Upgrade Procedure    l     Connect the equipment to be upgraded and the PC running the upgrade tool to the  same network  or directly connect the equipment and the PC via a crossover network    cable   Set the IP address of the PC to  77 77 1 XX  and subnet mask to  255 255 255 0      Run Mindray Patient Monitor Software Upgrade Tool on the PC and set Machine to       On the Mindray Patient Monitor Software Upgrade Tool screen  select  Select  Package  and then the packages you want to upgrade
38. orm an AC source     W    Verify that battery   s installed  If not  install the battery     4  Remove the AC power cord and verify that the equipment still works correctly     Performance Test    Perform the test by referring to the chapter 9 Battery in the Operator   s Manual and verify    the operating time of the battery meets the product specification     5 7 Maintenance Mode    The equipment includes maintenance mode for you to check the equipment conveniently     To enter the maintenance mode  select  Menu     System Setup      gt  Maintain    Mode    enter required password     The password for the maintenance mode is 332888 when the equipment left the factory  You  cannot change this password  If you have problem to enter the maintenance mode  please  contact your service personnel  You can perform the following tests in the maintenance    mode     Display test  Sound test  Keypad test    Lead wire check    Internal communication test    E   E   al   a   M Recorder test  A   NW USB test  Mm    Internal storage Test    Select a test item you want to perform  and follow the on screen prompt     5 4    In the maintenance mode  you can also  M View device information   Check work log   Set the frequency of AC filter  Format the internal memory    Input the equipment s serial number    Import hospital and department name    To exit the maintenance mode  turn off the equipment     Maintenance and Test Report     See the above sections for detailed test procedures and conte
39. orming any other leakage    tests     Leakage current is measured the following ways   NW Earth Leakage Current  leakage current measured through DUT outlet Earth    NW Earth Leakage Current AP EARTH  ALL Applied Parts connected to Earth   leakage  current measured through DUT outlet Earth    There is no need to attach a test lead  the 601PRO automatically connects the measuring    device internally     To Perform the Test    l  From the MAIN MENU  or with the outlet unpowered  plug the DUT into the 601PRO    front panel outlet  and turn on the device   2  Attach the device s applied parts to the 601PRO applied part terminals if applicable     3  Press shortcut key 4 The Earth Leakage test appears on the display  and the test begins    immediately     Earth Leakage   Outlet  Rev Pol  No Earth  L2  No AP Earth OuA  Limit Inv     DUT OFF EARTH NO L2 AP EARTH    A O O OD O C       SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse   SOFT KEY 2 toggles the DUT outlet from Earth to No Earth    SOFT KEY 3 toggles the DUT outlet from L2 to No L2    SOFT KEY 4 toggles the AP to Earth to No AP to Earth      gt  E E E E    Press the print data key at any time to generate a printout of the latest measurement     In Case of Failure    M Check any broken of the enclosure  Replace any defective part   NW Inspect wiring for bad crimps  poor connections  or damage     M Test the wall outlet  verify it is grounded and is free of other wiring abnormalities   Notify the us
40. pment and replacing the parts   This manual is based on the maximum configuration  Your equipment may not have same  parts and the quantity of the screws or stacking sleeves etc  may be different with those    included in the parts lists     NOTE    o The part number listed in the Parts List is only for checking the FRU part number  which is also included in the Parts List  Please provide the FRU parts number if    you want to purchase the spare parts     8 1    8 2 Main Unit    8 2 1 Exploded View       8 2    8 2 2 Parts List    SN   PN Description FRU part   Remarks  number    en 021 000108 00   Display EEE EEE    051 001353 01 1 001353 01 Main control board  512MB  N    115 017993 00 Main control board   512M Glasgow PDF  FRU    115 017994 00 Main control board  512M PDF   FRU    115 017995 00 Main control board   512M Glasgow  FRU     049 000542 00 Keypad  BeneHeart R3 Chinese  a    049 000543 00 Keypad  BeneHeart  a A Chinese     0051500   Keypa Benito      3  ormon  Damien    s ponos   ac impreso   SSS    115 017634 00 BeneHeat R3 top housing  assembly   115 017793 00 BeneHeart R3A top housing  aca Y Erre    17   051 001354 01 1 001354 01 Keypad board board    009 003326 00 Cable between main board IN BEN  keypad board    EXISTEN rro O A  047 010840 00 Power board insulator  M501  OA    11   009 003327 00 Cable between the main control  board and power board  12   115 018611 00   Parameter Board  FRU  aa    ee e a  board and parameter board  115 017792 00 Bottom   Bott
41. predictable equipment damage or human injury    e Follow correct sequence to disassembly the equipment  Otherwise  the equipment  may be damaged permanently    o Be sure to disconnect all the cables before disassembling any parts  Be sure not to  damage any cables or connectors    o Be sure to place removed screws and disassembled parts properly  preventing them  from being lost or contaminated      Place the screws and parts from the same module together to facilitate reassembling    o To reassemble the equipment  first assemble the assemblies  and then the main unit     Carefully route the cables     7 3 Disassembling the Main Unit    NOTE    eo To disassemble the equipment  place the equipment on a work bench free from  foreign material  avoiding damaging the screen     e All the operations should be performed by qualified service personnel only   When disassembling the PCBAs and parts labeled with static sensitive symbols   make sure you are wearing electrostatic discharge protection such as antistatic  wristband or gloves to avoid damaging the equipment     7 3 1 Removing Bottom housing assembly    1  Lay the equipment on the work bench  Unscrew the M3x6 screw to open the battery    compartment door  Then remove the battery        Ze       7 3 2 Remove the Power Board    Unscrew the four M36 screws  Disconnect the AC input receptacle and the cable   Disconnect the connection cable between the main control board and the power board  Then    take out the power board and th
42. propriate SOFT  KEY on the 601PRO     8  Press the print data key at any time to generate a printout of the latest measurement     In Case of Failure    M Check any broken of the enclosure  Replace any defective part   NW Inspect wiring for bad crimps  poor connections  or damage     m  Test the wall outlet  verify it is grounded and is free of other wiring abnormalities   Notify the user or owner to correct any deviations  As a work around  check the other    outlets to see if they could be used instead   NW Change another probe to confirm if the fail is caused by console     m Ifthe leakage current measurement tests fail on a new unit and if situation can not be  corrected  submit a Safety Failure Report to document the system problem  Remove    unit from operation     NW  fall else fails  stop using and inform the Customer Service Engineer for analysis and    disposal     LIMITS    For CF  amp  applied parts   9  10uA in Normal Condition  9 50uA in Single Fault Condition    A 10    A 9 Scheduled Electrical Safety Inspection    For scheduled electrical safety inspection  test items 1  2  3  4  5  6  7  and 8 included in the  ELECTRICAL SAFETY INSPECTION FORM shall be performed   ELECTRICAL SAFETY INSPECTION FORM       Device Enclosure and Accessories A  Protective Earth Resistance EE Max 0 2 Q  Normal condition uA Max   NC      NC  300uA  refer to    Earth UL60601 1     Leakage NC  500uA refer to  Single Fault LA 1EC60601 1     condition  SFC  SFC  10001 A  LIBF uA  Normal 
43. rsists after    corrective actions have been taken  please contact your service personnel     The equipment does   1  The equipment is not connected   1  Check that the power cord is securely  not power up  to AC mains  connected   2  Battery is not installed or has 2  Check that the battery is installed and has  no charge  sufficient charge  Otherwise  connect the  equipment to AC mains to run the    equipment and charge the battery     ECG data displays 1  Patient movement  1  Tell the patient not to move during ECG  unacceptable noise    2  Improper filter setting  acquisition   3  Poor skin preparation  2  Check the settings of the filters are    4  Electrode problem  appropriate   5  Patient cable problem  3  Prepare the patient before ECG    isition   6  Wrong accessories are used or ae    mix electrode types and brands  4  Verify the electrodes are applied  correctly  Check for defective or expired    electrodes   5  Check for defective  broken or  disconnected patient cable   6  Use specified accessories  Do not mix  electrode types or brands   The recorder does 1  Paperless recording is enabled    1  Select  Menu     Record Setup  and set  not work  2  Recording paper is not loaded     Paperless Recording  to  Off    3  Recorder door is not properly 2  Verify recording paper is properly loaded   closed  3  Verify recorder door is properly closed   4  Print head is too hot  4  Wait till the print head cools down   5  Recorder is disabled due to 5  Connect the equipmen
44. smells  e g   burning or smoky smells   The enclosure and accessories    particularly from ventilation holes      No taped notes that may suggest device deficiencies or    operator concerns        A 3 Device Labeling    Check the labels provided by the manufacturer or the healthcare facility are present and    legible   NW Main unit label    E Integrated warning labels    A 4 Protective Earth Resistance    Protective Earth Resistance is measured using the RED test lead attached to the DUT  Protective Earth terminal or enclosure  Select the test current by pressing SOFT KEY 3 to  toggle 25AMP  The front panel outlet power is turned off for this test     The following conditions apply  L1 and L2 Open     Preparation    1  First select the test current that will be used for performing the Protective Earth  Resistance test by pressing AMPERES  SOFT KEY 3      2  Connect the test lead s  between the RED input jack and the GREEN input jack   3  Press CAL LEADS  The 601PRO will measure the lead resistance  and if less than    0 150 Ohms  it will store the reading and subtract it from all earth resistance readings    taken at the calibrated current     Calibration in Progress    Iscale 1 004  Lead Resistance Too High    Connect test lead from RED input jack  To GREEN input jack    SEED E  ep p    GD    If the calibration fails  the previously stored readings will be used until a passing calibration       has occurred      NWARNING    o During Earth Resistance testing  the DUT must be
45. sted along with possible causes and recommended  corrective actions  Refer to the tables to check the equipment  identify and eliminate the    troubles     The troubles we list here are frequently arisen difficulties and the actions we recommend  can correct most problems  but not all of them  For more information on troubleshooting     contact our Customer Service Department     6 2 Parts Replacement    The equipment s PCBs  major parts and components are replaceable  Once you isolate a  PCB you suspect defective  follow the instructions in Chapter 7 Disassembly and Repairto  replace the PCB with a known good one and check that the trouble disappears or the  equipment passes all performance tests  Defective PCB can be sent to us for repair  If the  trouble remains  exchange the replacement PCB with the original suspicious PCB and    continue troubleshooting as directed in this chapter     To obtain information on replacement parts or order them  refer to 9 BeneHeart R3 R3A    Replacement Partes     6 3 Quickly Identifying Failure    Some troubleshooting tasks may require you to identify the hardware version and status of  your equipment  To check equipment information    Select  Menu      System Setup      Maintain Mode     Enter the required password to  enter the maintenance mode  In the maintenance mode  you can perform related tests to    primarily identify the failure     6 4 General Problems    This chapter lists the problems that are likely to occur  If the problem pe
46. sting and maintenance procedures as required    and use appropriate test equipments     The testing procedures provided in this chapter are intended to verify that the equipment  meets the performance specifications  If the equipment or a module fails to perform as  specified in any test  repairs or replacements must be done to correct the problem  If you    have any question  contact our Customer Service Department        CAUTION    eo All tests should be performed by qualified service personnel only     Oo Service personnel should acquaint themselves with the test tools and make sure that    test tools and cables are applicable     5 1 1 Test Equipment    See the following sections     5 1 2 Test Report    Upon the completion of the tests  the table of preventative maintenance test reports and the    table of maintenance test reports in this chapter should be kept properly     5 1 3 Preventative Maintenance    Below are preventative maintenance tests which need to be performed on the equipment     See the following sections for detailed maintenance procedures     5 1    5 1 4 Recommended Frequency    Visual inspection 1  When first installed or reinstalled     1  When first installed or reinstalled     Power on test 2  Following any repairs or replacement of any main    unit parts     Performance Test   1  If the user suspects that the measurement is incorrect   2  Following any repairs or replacement of relevant  ECG test  Module module   Calibration 3  At least once every
47. t to AC mains to  depleted battery  run the equipment and charge the battery   Paper jammed or 1  Unapproved paper is used  1  Use approved recording paper   misaligned  2  Recording paper is not properly   2  Take out the paper and tear off the  loaded  jammed part  Reload the paper as described  in 3 2 3 Loading Paper of the Operator   s    Manual        Partially missing 1  Dirty print head  1  Clean the print head   printout or printout   2  Some thermal points on print 2  If the problem persists  contact your    not clear  head are damaged  service personnel     The equipment 1  Auto shutdown is enabled  1  Check the setting of  Auto Shut Down   automatically shuts   2  The battery is depleted when by selecting  Menu      System Setup   In    down  the equipment runs on battery the case that any of limb leads is off  the    power  equipment automatically shut down if there  is no operation when the defined time is  reached   2  Connect the equipment to AC mains to    run the equipment and charge the battery     The display is Auto Standby is enabled  In the case that any of limb leads 1s off  the  completely blank  equipment automatically turns off the  display to save power if there is no    operation when the defined time is reached     To exit the standby mode  press any key        The screen display   The setting of brightness 1s low  Adjust screen brightness     is too dark to be    seen clearly     6 5 Messages    The equipment prompts messages to indicate the curr
48. tification tone when the message occurs     6 6 Troubleshooting Guide    6 6 1 Power On Off Failure          System not power on by  pressing power switch    AC indicator on  N                       Check if AC  power is OK     as power     N  problem                   disconnecting recorder  drive board or  arameter board    Replace power  Y board  ar recorder eplace main  drive board or   board  parameter board    Power switch  indicator on                        A beep is heard within 10  after system powered on              Display back lighted     Black screen  replace main N  board  or replace display       Y    y       White screen  replace the display       6 8    6 6 2 Recorder Failures      Recorder not work D    Message                                 Cable between  recorder and main  board loose or  failed                 Reconnect or  replace cable       Recorder  unavailable     presented              Ok after recorder drive  board replaced           1  Check platen in position  Ssage    Ple  2  Check paper correctly loaded  AY    3  Close battery door     check recorder     resented                    N    y              Message    Please        check recorder     presented     N   Replace main board                        erform keypad test in  maintenance mode   Replace keypad board if  failed     Recorder runs by pressing     Manual    key     End    Replace  print head    N               Message     Paper type error     presented     Recording paper  incorr
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Rollei PDF-S 200 SE  Quick-startGuide forYRDKRX63N v3.3    33U/33U CO/33U F/33U CO F  Master Professionnel « Anthropologie & Métiers du Développement  Chèque Territoires d`Innovation, mode d`emploi    Copyright © All rights reserved. 
   Failed to retrieve file