Model WT-5300A

image

Contents

1. Caution If a malfunction occurs in the warming system discontinue use Notify your sales service center of the malfunction Service is only to be done by qualified personnel Caution The Model WT 5300A should not be operated in the presence of electromagnetic fields that are greater than 3 Volts meter This could cause shutdown of the Model WT 5300A by the fail safe function within the equipment Caution Ensure that the patient is dry or the warming system may be ineffective Introduction Intended Use The WarmTouch Model WT 5300A Patient warming system warming unit and blanket is intended for prevention and treatment of hypothermia For example with the surgical patient the patient in the preoperative holding area the pregnant woman who shivers during epidural anesthesia due to hypothermia or any patient who is uncomfortable anywhere in the cold critical care environment Manual Availability The most recent revision of this manual is available on the Internet at http www mallinckrodt com respiratory resp Serv_Supp ProductManuals html Background Information WT 5300A There are numerous ways of warming your patient from cotton blankets to water mattresses Research has shown that low temperatures surrounding the patient are a major factor contributing to hypothermia 2 The warming system covers the patient with warm air and actively transfers heat across the skin The result is to achieve normal thermia
2. In creating this warm customized pocket of air around the hypothermic patient it is important to note that stagnant air even ifit is warm does not work as an effective heat transfer medium Stagnant air acts as an insulator preventing the boundary layer of molecules next to the skin surface from transferring heat Forced air warming causes warmed air molecules to flow over cooler skin surface It is this active flow of warmed molecules that acts as a heat transfer medium With the warming system air is warmed and delivered into a lightweight blanket CareOuilt or CareDrape Blanket that rests over or under the patient The CareQuilt and CareDrape blankets have many small perforations on the underside that allow air to exit the blanket and surround the patient 1 Morris RH Wilkey BR The Effects ofthe Ambient Temperature on Patient Temperature During Surgery Not Involving Body Cavities Anesthesiology 32 102 107 1970 2 Morris RH Influence of Ambient Temperature on Patient Temperature During Intra abdominal Surgery Annals of Surgery 173 230 233 1971 Introduction Safety Features The VVarming System is designed to give healthcare professionals more control over the patient s core body temperature There are several safety features of the warming system which make it safe and appropriate for such use Customized Warming Therapy Clinicians select temperature range setting at the onset of warming therapy to help ensu
3. Model WT 3300A Patient VVarming System Operator s Manual To obtain information about a warranty if any contact Nellcor s Technical Services Department 1 800 635 5267 or your local Nellcor representative Contents 53004 Contents COMEM 45004 GEN duc EEES N N VEE ER CO OL DA ON EE Ed i A seis id oi Gn AE mida ed eee de ONES ii ADOS c MO EN EO ENE PLUGGED ae ET N Cites Ee EP iii Safety MORIS sie sera ar presi ea AA DA 1 WAMINOS RE HH edie 1 or Te EG ee 2 ues ME PERE HE EE EE HEN ME rebase 3 Intended Use RR nennen ande 3 Manual Availability 0 0 0 ce ee Re AR ee AR Ee nnm eene nnns 3 Background Information ee AR ee AA AR ee ee ee AA ee ee nr 3 Safety Features se ee ee se ee ee Re ee ee Re ee ee ee ke ee ee ee Re ee ee ee ee ee ee 4 Customized Warming Therapy iese ee ee ee ee ee Ge ke ee ee 4 Automatic Temperature Stepdown ee ee ee ER ER AAR Re ee 4 Automatic Over Temperature Shutdown 4 Warning Light ee ee RR Ge AR Re AA ee Re ee mmn 4 HEPA Eier SE EE cla tt ea ec ai 4 Wheel Do Te CA 4 SYMBOIS ER EE 6 Description of the Warming System sss enn 7 INSTUSWIN errors nn oe OT ER EO ON 9 Cart Installation n ctr a 9 IV Pole Installation aan PES 10 Patient Bed Installation 4444424444444nnnnnnenen
4. Cleaning 17 Control panel 14 21 Customized warming therapy 4 D Disinfecting 17 E Electric shock 1 Electrical shock hazard 1 Explosion hazard 1 F Fire hazard 1 I Introduction 3 IV pole installation 10 WT 5300A M Main power 13 Main power switch 13 MRI warning 2 P Patient bed installation 11 Patient burns 1 Possible explosion hazard 1 Power supply cord 13 Principles of operation 21 R Routine maintenance 17 S Safety features 4 Safety information 1 Single patient use 2 Specifications 19 Surface cleaning 17 T Temperature control 14 U Using the vvarming system 13 W Warning light 4 Warnings 1 23 Blank Page Q NELLCOR Tyco Healthcare Group LP Nellcor Puritan Bennett Division 4280 Hacienda Drive Pleasanton CA 94588 USA Toll Free 1 800 635 5267 O 2006 Nellcor Puritan Bennett Inc All rights reserved Manufactured in the U S A P N 10017812A 1206 Tyco Healthcare
5. Figure 1 Cart Wheel Lock WT 5300A 5 Introduction Symbols The symbols identified in Table 1 are the symbols used on the WT 5300A and all WT 5300A labeling Table 1 WT 5300A Symbols Attention symbol consult accompanying documentation Temperature Control Low 32 C Medium Temperature Control Medium 38 C 38 C High Temperature Control High 43 C Temperature Control Boost 45 C for 45 minutes I Power On O Power Off Protection Class I Protection Type BF Description of the Warming System Introduction The WarmTouch Model VV1 5300A Patient warming system Figure 2 and Figure 3 CareQuilt and CareDrape blankets are intended for prevention and treatment of hypothermia For example with the surgical patient the patient in the preoperative holding area the pregnant woman who shivers during epidural anesthesia due to hypothermia or any patient who is uncomfortable cold in the critical care environment Hose 2 Main Power Switch 4 Power Cord 5 Hour Meter 3 Nozzle 6 Control Panel Figure 2 Model WT 5300A Front View WT 5300A Introduction 1 Over Temperature Test Port 2 Bed Hook Bracket 4 Warning Label 5 Nozzle Strap with Clip 3 Filter Cover 6 Instruction Label Figure 3 Model WT 5300A Back View Installation Cart Installation The warming system
6. 4444444 nn 10 Warming System Patient Bed Installation ee ke ee 11 Main Power Switch netten erred tena 13 Low Temperature Light seeded decent 14 Control Panel ee ee ee ee ee nennen enne 14 Control Panel RE OR RE DR OE RE 21 Tables WT 5300A Table 1 WT 5300A Symbols Table 2 Model WT 5300A Blower Specifications Table 3 Model WT 5300A Cart Specifications Contents Blank Page Safety Information Warnings WT 5300A P P P P Pb pb pb WARNING Possible explosion hazard Do not use the Warm Touch Model WT 5300A patient warming system in the presence of flammable anesthetic agents WARNING Possible electrical shock hazard To reduce the risk of electrical shock do not remove the back case Servicing is only to be done by qualified personnel WARNING Possible electric shock hazard Grounding reliability can be achieved only when the warming system is connected to an equivalent socket marked HOSPITAL ONLY or HOSPITAL GRADE WARNING Possible fire hazard Prevent the blanket material from coming into contact with a laser or an electrosurgical active electrode rapid combustion could result WARNING Possible burn hazard Do not apply heat directly to open wounds All patient s wounds should be covered while using the warming system WARNING Possible patient burns Use caution and consider discontinuing use on patients du
7. e monitor The warming system may be disinfected using a soft cloth saturated with a 10 solution of chlorine bleach in tap water Other Routine Maintenance Refer to the VVT 5300A service manual for routine maintenance procedures that require a qualified technician to perform WT 5300A 17 Blank Page Specifications Warming System Specifications No special transport or storage methods are required for the warming system Table 2 Model WT 5300A Blower Specifications Dimensions 38 cm x 41 cm x 28 cm 15 inches x 16 inches x 11 inches Weight 6 8 kgs 15 Ibs Power Requirements 120 VAC 50 60 Hz 10A Automatic Temperature Stepdown Boost to High Temperature Power Supply Cord After 45 minutes of continuous use 4 26 m 14 feet Thermal Protection Thermostat internal 47 C 50 C 117 F 122 F Ambient Blower Operating Temperature Range 18 C 28 C 64 4 F 82 4 F Protection Against Ingress of Fluids IEC 60601 1 sub clause 44 6 for class IPXO Protection Class 1 Protection Type B Table 3 Model WT 5300A Cart Specifications Weight Height with Handle Extended 4 6 kg 10 1 pounds 87 6 cm 34 5 inches Height with Handle Retracted Width 70 5 cm 27 8 inches 31 8 cm 12 5 inches Depth 38 6 cm 15 2 inches 19 Specifications Transport and Shipping in Shipping Container 20 Te
8. good judgment when selecting a temperature If unsure of proper setting consult with the attending physician The warming system contains a HEPA filter rated 99 97 efficient at 0 3 micron particle size See Routine Maintenance on page 17 for filter replacement procedure Self Supporting Air Hose Air is delivered into the CareQuilt or CareDrape blankets through a wire reinforced hose attached to the warming system Using WarmTouch CareQuilt and CareDrape Blankets WT 5300A WarmTouch CareQuilt and CareDrape blankets come in a range of sizes and configurations to fit patient s individual sizes surgical procedures and comfort needs Follow the instructions provided with all WarmTouch blankets for specific information concerning their recommended use 15 Blank Page Routine Maintenance instructions regarding disposal or recycling of filter and device components or end lt gt Caution The institution should follow local governing ordinances and recycling of life of the product Cleaning the Warming System Caution Do not spray pour or spill any liquid on the warming system its accessories connectors switches or openings in the case For surface cleaning and disinfecting follow your institution s procedures or The warming system may be surface cleaning by using a soft cloth dampened with either a commercial nonabrasive cleaner or a solution of 7096 alcohol in water and lightly wiping the surfaces of th
9. is shipped installed on the warming system cart See Figure 4 The warming system should not be placed on the cart with the handle any higher than 61 6 cm 24 25 inches At this height the warming system and cart pass a 10 degree tilt test Ensure that the three blower cart clamps are tight tr dum i a Hmmm mann as Je en dj 24 25 inches 61 6 centimeters Maximum 2 Blower cart clamps Figure 4 Warming System Cart Installation WT 5300A Installation IV Pole Installation Figure 5 illustrates the installation of the warming system installed on an IV pole The warming system should not be installed on the IV pole with the handle higher than 76 centimeters 30 inches 30 inches 76 centimeters Maximum Figure 5 Warming System IV Pole Installation 10 Installation Patient Bed Installation Figure 6 illustrates the installation of the warming system on a patient bed The patient rail connectors will fit on bed rails up to 1 4 inches 3 6 centimeters wide Bed Rails 1 4 inches 3 6 centimeters Figure 6 Warming System Patient Bed Installation WT 5300A 11 Blank Page Using the Warming System Power Supply Cord Plug the warming system power cord into a hospital grade or suitable mains outlet Main Power Power is controlled through one main rocker switch located on the front of the warming system To begin operation p
10. mperature 40 C to 70 C 40 F to 158 F Altitude 390 m to 6 096 m 1 280 ft to 20 000 ft Barometric Pressure 1 060 hPa to 500 hPa 375 mmHg to 795 mmHg Relative Humidity 15 to 95 non condensing Principles of Operation Principles of Operation Control Panel The control panel features four manually switched temperature settings The selected temperature is indicated by an LED indicator Warmlouch III neLrLcor Patient Warming System Figure 10 Control Panel Each temperature setting represents the average temperature of air entering the blanket The four temperature settings are Low 32 C 89 6 F Medium 38 C 100 4 F e High 43 C 109 4 F e Boost 45 C 113 F for 45 minutes There is a warning light on the control panel See Figure 10 The warning light indicates that the warming system s safety thermostats have disabled the heater The warming system will continue to blow air until the output temperature has decreased below the thermostat threshold temperature of 50 C Ifthere is no malfunction in the warming system it should function normally Ifthe warning light illuminates again the blower should be replaced and the faulty blower should be checked by qualified hospital personnel WT 5300A 21 Blank Page Index A Air filter 15 Air hose 15 Automatic shutdown 4 B Bed installation 11 Burn hazard 1 C CareDrape 15 CareQuilt 15
11. nennnnnnnnnnnnennnnnnnnannnnnnnen 11 Using the Warming System AE MR TEE HA IT 13 Power Supply COND decision diia Ke ek ci RR ERE B dX AE RN 13 MEd ER MH R 13 Temperature Control s sisien EE Ee ee AG ee ee ee ee Ee ede a ee ee ee ea ee 14 da AR 15 Self Supporting Air Hose ii esse se nennen 15 Using WarmTouch CareQuilt and CareDrape Blankets iss ese ee ee ee ee 15 Routine Maintenance EE sauna ER DA DRR RR RR DR ER Ed OR HR AR OR 17 Cleaning the Warming System ie ee ee Ee AR AA AA Ge ee ee ee ee Ge AR ee ee ee 17 Other Routine Maintenance see ee ee ee Ee AR AA Re ee ee ee ee ee ee Ee Re Re ee ee 17 ese ei RES EE ER eR UP PEC EG 19 Warming System Specifications assin AA ee ee AR Re eme 19 Transport and Shipping in Shipping Container nn 20 Principles Of Operation y eque ace x race aha dran doa os qd d Eloi a da ders no EN 21 Control Panel ge 21 MOOK aa ae EE TTE Da a DES DES DO me eee Go Sr DE DE DES de a N 23 WT 5300A i Contents Figures Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Cart Wheel LOCK SR trece ied cr te eter nidi A 5 Model WT 5300A Front View ees ee ee ee mmm 7 Model WT 5300A Back View seem emm 8 Warming System Cart Installation eeeeeee 9 Warming System IV Pole Installation 22
12. re the appropriate setting 1s selected for every patient Automatic Temperature Stepdown The warming system provides a 45 minute temperature stepdown feature When in the Boost mode blower temperature will automatically drop to the high mode after each 45 minutes of use The temperature may be reset to the Boost mode by selecting the Boost temperature on the unit front panel to start another 45 minute cycle Automatic Over Temperature Shutdown The automatic temperature controller and two back up systems help to ensure temperature will not reach excessive levels If necessary the back up systems automatically turn off the heater and illuminates the warning light The warming system heater will start producing heat when the warming system temperature drops to approximately 34 C to 37 C Warning Light When the warning light on the control panel is illuminated it indicates that one of the two thermostats has turned off the heater The reason for the heater turn off should be investigated as soon as possible after use HEPA Filter The system s High Efficiency Particulate Air Filter is 99 97 efficient at 0 3 micron particle size Wheel Locks The cart is equipped with two wheel locks The wheel locks prevent the cart from moving while in use The wheel locks must be released when moving the cart Introduction Press the wheel lock arm down to lock the wheel Lift the wheel lock arm to release the wheel lock See Figure 1
13. ress the Main Power switch to the ON position Figure 7 The Low temperature light on the control panel will illuminate and the unit will begin to blow air in the low temperature mode See Figure 8 Applying heat to a patient directly from the blower nozzle will cause patient 1 WARNING No free hosing Do not use the blower without a blanket burns 1 Main Power Switch Figure 7 Main Power Switch WT 5300A 13 Using the Warming System Warmlouch enELLCOR Patient Warming System 1 Low Temperature Light Figure 8 Low Temperature Light Temperature Control The heating temperature 1s controlled through the panel of buttons This panel has four temperature settings Low 32 C 89 6 F Medium 38 C 100 4 F e High 43 C 109 4 F e Boost 45 C 113 F for 45 minutes The temperature settings represent the range of air temperature entering the blanket For patients with severe hypothermia requiring rapid warming select the Boost temperature setting To help prevent hypothermia or for treatment of mild hypothermia select High Medium or Low The selected temperature light will illuminate See Figure 9 Warmlouch III NELLCOR Patient Warming System Medium High 38 C 43 C Figure 9 Control Panel 14 A Air Filter Using the Warming System WARNING The patient must be closely monitored for rewarming Vasodilation and possible hypotension can occur Use
14. ring vascular surgery when an artery to an extremity is clamped Do not apply the warming system to ischemic limbs This could possibly cause the patient to be burned WARNING Possible patient burns Use caution and monitor closely if used on patients with severe peripheral vascular disease This could possibly cause the patient to be burned Safety Information P PP gt Cautions 9 9 994 WARNING Possible patient burns Do not warm patients with the warming system hose only Thermal injury may result Always attach the hose to a WarmTouch blanket before providing warming therapy WARNING WarmTouch CareDrape and CareQuilt blankets are for single patient use only WARNING MRI scanning The warming system is not suitable for use during magnetic resonance imaging MRI scanning The warming system may affect the MRI image WARNING Continuously monitor the patient s temperature Reduce the air temperature or discontinue therapy when normothermia is reached Caution Federal U S A law restricts the use of the warming system to sale by or on the order of a physician Caution The warming system is fitted with an air filter however airborne contamination should be considered when using the warming system Caution If the warming system is mounted on the intravenous IV pole it should be installed with the top of the unit s handle less than 76 cm 30 inches above the floor to prevent the IV pole from tipping over