Home
Service Manual
Contents
1. MESA MULTI RAD ISSUEDBY DATE
2. Chassis Chassis GND GND BUCKY TABLE BUCKY CABLE A3091 XX See schematic 001 ATP CONSOLE BOARD ION CHAMBER AEC CABLE A3253 XX dE Preamplifier See Note 3 See Note 3 Note 3 AEC CABLE AEC INTERFACE CABLE AND AEC INTERFACE CABLE A3251 02 AEC ADAPTATION BOARD NOT USED FOR RAD ROOM STANDARD without ION See Note 3 See Note 3 CHAMBER AEC ADAPTATION BOARD GENERATOR CONSOLE VERTICAL BUCKY A3263 03 TOUCH SCREEN See schematic IM 015 ON UE MEE ri FD J1 J2 J3 INTERFACE BOX CABLE A7067 XX Preamplifier See Note 3 a See Note 3 J5 GENERATOR See Note 3 TPC SUPPLY CABLE A7059 XX TS1 AC SUPPLY WALL BUCKY CABLE A6703 XX POWER I F CABLE A3252 03 BUCKY TS1 2 GND CABLE A7023 XX See Note 3 Chassis Chassis See schematic I F 002 GND mH Lew GND CABLE A6383 XX Chassis Chassis GND GND THIS SHEET ONLY APPLIES with TPC SERIAL CONSOLE jp name DATE SHEET 08 pee TL REVISED ADIAZ 02 06 04 NET HIER E as RAD ROOM STANDARD A newinteface F Garc a 30 11 04 4 SEDECAL amp with AEC SYSTEM MESA TABLE TS2 i E ee Freno Bucky Bucky Lock Bucky Lock Cable A7096 01 Transv Solenoid Cable A7098 02 Solenoide Transversal GENERADOR COMPACT COMPACT GENERATOR Transversal Solenoid Long Solenoid Cable A7098 01 FRENOS 3 Solenoide Longitudinal LOCKS BOARD ie N Longitu
3. Technical Publication SM 1103R3 Service Manual General Purpose Rad Table 0120 This product bears CE marking in accordance with the provisions of the 93 42 EEC MDD dated June 14 1993 Este producto ostenta una marca CE de acuerdo con las disposiciones de la Directiva 93 42 CEE del 14 de Junio de 1993 sobre Productos M dicos Ce produit porte la marque CE de conformit aux r glements de la Directive 93 42 CEE du 14 juin 1993 relative aux Produits m dicaux The information comprised in this manual applies to the following equipments La informaci n contenida en este manual se aplica a los siguientes equipos L information contenue dans ce manuel est appliqu e aux quipements suivants MULT FWFTT Four Ways Floating Top Table MULTIRAD Manufactured by Fabricado por Fabriqu par SEDECAL Sociedad Espa ola de Electromedicina y Calidad S A Pelaya 9 13 Pol gono Industrial R o de Janeiro 28110 Algete Madrid Espa a Spain Phone 34 916 280 544 Fax 34 902 190385 www sedecal com General Purpose Rad Table Service Manual REVISION HISTORY REVISION DATE REASON FOR CHANGE 0 MAY 23 2011 First edition 1 AUG 09 2013 Design changes 2 SEPT 09 2013 General Update 3 NOV 05 2013 Ammended distance of anchoring holes This Document is the english original version edited and supplied by the manufacturer The Revision state of this Document is indicated in the code number shown at the bottom
4. 1 SAT A9657 02 Frame 1 5mm for Grid 1 4m SAT A10523 01 Frame 1 5mm for Grid 1 5m SAT A9656 02 Frame 1 5mm for Grid 1 8m SAT A10522 01 Frame 2mm for Grid 1m SAT A10667 01 Frame 2mm for Grid 1 3m SAT A10666 01 Frame 2mm for Grid 1 4m SAT A10665 01 Frame 2mm for Grid 1 5m SAT A10664 01 Frame 2mm for Grid 1 8m SAT A10663 01 Frame 3mm for Grid 1m SAT A10667 02 Frame 3mm for Grid 1 3m SAT A10666 02 Frame 3mm for Grid 1 4m SAT A10665 02 Frame 3mm for Grid 1 5m SAT A10664 02 Frame 3mm for Grid 1 8m SAT A10663 02 Vacutec lon Chamber 6695 21 AID lon Chamber 6695 11 Medys lon Chamber 6695 66 Claymount lon Chamber 6695 56 6695 57 Switch Kit BDC A520491 01 Bumper Rubber SAT 6533 01 This reference comprises 5 units Grid Blockage A520476 01 Fans Kit SAT A9721 01 30 SM 1103R3 General Purpose Rad Table Service Manual 31 SM 1103R3 General Purpose Rad Table Service Manual This page intentionally left blank 32 SM 1103R3 General Purpose Rad Table Service Manual SECTION 8 SYSTEM INTERCONNECTION MAPS Refer to the following maps for details of the wire connections SYSTEM INTERCONNECTION General Purpose Rad System Interconnection IM 341 Radiographic Table Interconnections 54303079 Not
5. Protective Earth Ground Earth Ground This symbol according to the European Directive indicates that the Waste of Electrical and Electronic Equipment WEEE must not be disposed of as unsorted municipal waste and must be collected separately Please contact an authorized representative of the manufacturer or an authorized waste management company for information concerning the decommissioning of your equipment Li Pb Cd Hg This separate collection symbol is affixed to a battery or its packing to advise that the battery must be recycled or disposed of in accordance with local or country laws The letters below the symbol indicate whether certain elements Li Lithium PB Lead CD Cadmium Hg Mercury are contained in the battery All batteries removed from the equipment must be properly recycled or disposed Please contact an authorized representative of the manufacturer or an authorized waste management company for information concerning the decommissioning of your equipment G Pollution Control Only applicable to People s Republic of China PRC This symbol indicates the product contains hazardous materials in excess of the limits established by the Chinese Standards It must not be disposed of as unsorted municipal waste and must be collected separately Please contact an authorized representative of the manufacturer or an authorized waste management company for information concerning the decommissioning
6. WALL BUCKY CABLE A6703 XX BUCKY See schematic I F 002 GND CABLE A6383 XX Chassi assis GND Chassis GND Note 3 AEC CABLE AEC INTERFACE CABLE AND AEC ADAPTATION BOARD NOT USED FOR RAD ROOM STANDARD without ION CHAMBER AEC INTERFACE CABLE A3251 XX GENERATOR CONSOLE ATP CONSOLE BOARD See Note 3 POWER CABLE A7021 XX COMMUNICATION CABLE A7022 XX INTERFACE CABLE A7069 XX GND CAB LE A6383 XX Chassis GND THIS SHEET ONLY APPLIES with CONVENTIONAL CONSOLE owe sa ET IM 341 DRAWING F GARCIA 05 05 04 1 2 REVISED A DIAZ 02 06 04 new interface A new interface interface Garc a 30 11 04 Garcia 30 11 04 2 SEDECAL ne DESCRIPTION ISSUED BY DATE ROT MOBILE COLUMN N u GENERATOR CABINET See schematic 54301085 for Interconnections A7014 XX for Low High Speed TUBE STATOR CABLE See Note 2 Note 1 Cable code TS2 HV A8410 01 for Compact Generator single phase CATHODE CABLE TANK A8411 01 for Compact Generator three phase 4 A8412 01 for Battery Generator A A VAC SUPPLY GND CABLE 6383 A Chassis Chassis GENERATOR SUPPLY CABLE See Note 1 GND GND AC LINE COLLIMATOR CABLE A3111 XX COLLIMATOR SUPPLY CABLE A6736 XX RAD TABLE See schematic 54302065 for interconnections es SUPPLY CABLE A7235 XX 3 GND CABLE A6383 XX
7. of your equipment SM 1103R3 General Purpose Rad Table Service Manual TABLE OF CONTENTS Section Page 1 INTRODUCTION Ja qasa a a ela eee ee 1 1 1 Objective and Scope of this Manual 1 1 2 Tools and Test Equipment 1 2 PRE INSTALLATION 2 22 20 aa aan 3 2 1 Pre Installation Check 3 2 2 Responsibilities of the Purchaser 3 2 3 Room Requirements 3 2 3 1 Environmental Requirements 3 2 3 2 Floor Requirements nai 4 2 3 3 Technical Specifications 5 3 UNPACKING o few 7 4 INSTALLATION ol eat ala ea ae hatches 9 4 1 Undertable Installation of the Generator option 12 5 ADJUSTMENTS eas eee ee eed a 15 5 1 Brake AdjgSImentu ua tg asar 15 5 2 Pedal Microswitch Replacing and Adjustment 16 6 PERIODIC MAINTENANCE 44444ssssssssessesesesses 19 6 1 Maintenance Activities Description 19 6 2 Checking of Radiographic Table 19 6 3 Check Off Maintenance Revision Table 20 7 RENEWAL PARTS odia arten beac eid eae eds 21
8. 03R3 General Purpose Rad Table Service Manual 7 Note 8 Note 9 10 Connect the Power Supply cable for the Table Locks from the Table TS1 to TB7 of the Lock Board in the Generator and connect the GND cable from the Table Base to the GND Stud in the Generator refer to schematics IM 341 and 54303079 This Table can be also power supplied by a 24 VDC External Power Supply instead of a Generator If applicable connect the Table Receptor Cable directly to the Terminal Block 3TS1 in the Generator and the Table lon Chamber Cable to the Generator as indicated in schematic IM 341 Refer to the Generator Service Manual for further information Route the cables through the rear cables outlet of the Table Base along the room for their later connection Re install the Front Cover of the Table Base Remove the two rubber stops of the same end of the Table Top 10 SM 1103R3 General Purpose Rad Table Service Manual 12 Install the Table Top from one of the Table sides right or left inserting the Bearings inside their Rails At least two people are required for this operation Be sure that the Table Top slides over the bearings and the CAUTION Longitudinal Lock Pin does not block its way 13 Remove the wire tie wrap of the Longitudinal Lock Pin 14 Re install the rubber stops of the Table Top end SM 1103R3 11 General Purpose Rad Table Service Manual 4 1 UNDERTABLE INSTAL
9. 1 5mm for Grid 1 8m SAT A10522 01 Frame 2mm for Grid 1m SAT A10667 01 Frame 2mm for Grid 1 3m SAT A10666 01 Frame 2mm for Grid 1 4m SAT A10665 01 Frame 2mm for Grid 1 5m SAT A10664 01 Frame 2mm for Grid 1 8m SAT A10663 01 Frame 3mm for Grid 1m SAT A10667 02 Frame 3mm for Grid 1 3m SAT A10666 02 Frame 3mm for Grid 1 4m SAT A10665 02 Frame 3mm for Grid 1 5m SAT A10664 02 Frame 3mm for Grid 1 8m SAT A10663 02 Vacutec lon Chamber 6695 21 AID lon Chamber 6695 11 Medys lon Chamber 6695 66 Claymount lon Chamber 6695 56 6695 57 Tray Blockage SAT 28213 01 Grid Blockage A525510 01 Switch Kit BDC A520491 01 Grid Guides BDC Kit A520492 01 Includes ITEM 2B4 Rubber Bumper SAT 6533 01 This reference comprises 5 units Handle Kit SAT A10634 02 Ball Bearing Slide SAT 54505012 Cable Support SAT 53530150 Includes 10 Units 26 SM 1103R3 General Purpose Rad Table Service Manual SM 1103R3 27 General Purpose Rad Table Service Manual DESCRIPTION REFERENCE REMARKS DETECTOR CABINET WITH ROTATING Frame 1 5mm for Grid 1m SAT A10521 01 Frame 1 5mm for Grid 1 3m SAT A9657 02 Frame 1 5mm for Grid 1 4m SAT A10523 01 Frame 1 5mm for Grid 1 5m SAT A9656 02 Frame 1 5mm for Grid 1 8m SAT A
10. 10522 01 Frame 2mm for Grid 1m SAT A10667 01 Frame 2mm for Grid 1 3m SAT A10666 01 Frame 2mm for Grid 1 4m SAT A10665 01 Frame 2mm for Grid 1 5m SAT A10664 01 Frame 2mm for Grid 1 8m SAT A10663 01 Frame 3mm for Grid 1m SAT A10667 02 Frame 3mm for Grid 1 3m SAT A10666 02 Frame 3mm for Grid 1 4m SAT A10665 02 Frame 3mm for Grid 1 5m SAT A10664 02 Frame 3mm for Grid 1 8m SAT A10663 02 Vacutec lon Chamber 6695 21 AID lon Chamber 6695 11 Medys lon Chamber 6695 66 Claymount lon Chamber 6695 56 6695 57 Rear bumper Kit Detector Detent Kit Switch Kit BCG SAT A10676 02 SAT A10677 01 A520490 01 Bumper Rubber SAT 6533 01 This reference comprises 5 units Grid Blockage A520476 01 Grid Guides BDG Kit A520493 01 Includes ITEM 3B7 Handle Kit SAT A10634 11 Holder for Detector SAT A10904 01 Only for detector without handle Ball Bearing Slide SAT 54505012 Cable Support SAT 53530150 Includes 10 Units Turning Kit SAT A10654 01 28 SM 1103R3 General Purpose Rad Table Service Manual 29 SM 1103R3 General Purpose Rad Table Service Manual DESCRIPTION QTY REFERENCE REMARKS DETECTOR CABINET FIXED DETECTO R AND REMOVABLE GRID Frame 1 5mm for Grid 1m 1 SAT A10521 01 Frame 1 5mm for Grid 1 3m
11. 8 SYSTEM INTERCONNECTION MAPS 33 SM 1103R3 i General Purpose Rad Table Service Manual SM 1103R3 General Purpose Rad Table Service Manual SECTION 1 INTRODUCTION 1 1 OBJECTIVE AND SCOPE OF THIS MANUAL This Service Manual is intended to describe the installation adjustments troubleshooting and periodic maintenance of the Radiographic Table OPERATOR AND SERVICE MANUALS SHOULD BE UT WARNING CAREFULLY READ AND UNDERSTOOD BY SERVICE PERSONNEL BEFORE USING AND SERVICING THE EQUIPMENT ESPECIALLY THE INSTRUCTIONS CONCERNING SAFETY REGULATORY DOSAGE AND RADIATION PROTECTION KEEP THE MANUALS WITH THE EQUIPMENT AT ALL TIMES AND PERIODICALLY REVIEW THE OPERATING AND SAFETY INSTRUCTIONS SERVICE PERSONNEL MUST HAVE SUFFICIENT Y WARNING KNOWLEDGE TO COMPETENTLY PERFORM THE SERVICE TASKS RELATED TO X RAY DEVICES AND PARTICULARLY TO THE EQUIPMENT DESCRIBED IN THIS MANUAL THIS KNOWLEDGE IS ACQUIRED THROUGH A VARIETY OF EDUCATIONAL METHODS FOR TECHNICIANS IN ACCORDANCE WITH LOCAL LAWS OR REGULATIONS INCLUDING SPECIFIC TRAINING ON THIS EQUIPMENT 1 2 TOOLS AND TEST EQUIPMENT The following tools and test equipment are required for the installation Standard service engineers tool kit Standard and extended levels 1 meter for Column Base Electric drill motor and assorted bits SM 1103R3 1 General Purpose Rad Table Service Manual This page intent
12. BRAKE ADJUSTMENT Note As brakes come factory adjusted field adjustment should only be necessary when replacing Brakes All brakes must have a space between the brake surface and the contact surface of less than or equal to one 1 millimeter lt 1mm If the space is too close the equipment may drag or movements may be difficult If the space is too large the brakes may not be strong enough to block movement To raise lower or level the Brake Assembly loosen the screws indicated in the following illustration and move the Brake Assembly to the desired position Illustration 5 1 Lateral and Receptor Brakes Screws Screws SM 1103R3 15 General Purpose Rad Table Service Manual 5 2 PEDAL MICROSWITCH REPLACING AND ADJUSTMENT 1 Remove the Table Front Cover 2 Disconnect Microswitch cables check previously the cable identification to connect the cables to the new switch in the same position 3 Remover nut 2 for that hold nut 1 with a 14 mm open wrench and loose nut 2 with another 14 mm Remove nut 2 4 Remove defective Microswitch 5 Install the new Microswitch 6 Adjust the Microswitch at 1 or 2mm from the cam to the wheel neck with the Table Pedal fully pressed Illustration 5 2 Microswitch Replacement Microswitch Wheel lt Adjust at 1 or 2 mm M Wheel neck 16 SM 1103R3 General Purpose Rad Table Service Manual 7 Check that the rolling wheel is pe
13. LATION OF THE GENERATOR OPTION Follow the steps below to install the Generator underneath the Table 1 Remove the Front Cover after unscrewing the four 4 screws FRONT COVER SCREWS 2 Disconnect the ground cable and dismount the front cover 3 Disconnect the three faston terminals of microswitch SW1 located inside the Table Base 4 Remove the four 4 allen screws of the pedal assembly at both sides of the Table Base a pair on each side and pull out the pedal assembly MICROSWITCH SW1 PEDAL ASSEMBLY SCREWS 12 SM 1103R3 General Purpose Rad Table GENERATOR CARRIAGE Note Service Manual Place the Generator Cabinet without covers on its carriage in front of the Table This component is freestanding below the Table Seismic areas and other conditions require the Generator Cabinet to be secured to the floor through the mounting holes on the bottom Do not slide the Generator under the Table until Configuration and Calibration procedures are completed Refer to the Generator Service Manual for further information Once configured and calibrated push the Generator inside the Table Reassemble all of the components parts and connections then replace the Front Cover of the Table SM 1103R3 13 General Purpose Rad Table Service Manual This page intentionally left blank 14 SM 1103R3 General Purpose Rad Table Service Manual SECTION 5 ADJUSTMENTS 5 1
14. bolts The Drill Templates of the anchoring holes are shown in the next illustrations Potential for Injury and or Equipment Damage Floor anchors m CAUTION must be a minimum of 150 mm 6 from any concrete edge including ducts and cracks In addition the general condition of the concrete in the immediate mounting area should be inspected to ensure that anchors will be set in good quality concrete The floor bearing must be concrete and the thickness to be determined by a Structural Engineer to properly support the equipment loads The anchors require a minimum embedment of 57 2 mm 2 1 4 into the concrete If the floor thickness is less than 101 6 mm 4 itis recommended that the unit be secured using a through bolt method with a reinforcement plate on the back side Note gt The maximum tolerance for the Floor levelling should be 1 5 mm per meter 0 02 per feet Illustration 2 1 Drill Template 1240 48 87 DETAIL A 4 SM 1103R3 General Purpose Rad Table Service Manual 2 3 3 TECHNICAL SPECIFICATIONS POWER LINE REQUIREMENTS Input Voltage 24 VDC Input Power 50 VA DIMENSIONS AND WEIGHT Height RE 770 mm 30 3 Length eng 2200 mm 86 6 WIG an 825 mm 32 4 Weight sec o dto hg e 150 kg 330 71b Dimensions of Floating Table Top 2200 x 825 mm 86 6 x 32 4 Table Top Receptor
15. dinal Solenoid Cable Alimentaci n Power Supply Cable A7235 XX 24 VDC LOCKS EA 0 VDC LOCKS J1 Cable Freno Multirad A15118 01 Freno Transversal 2 Transversal Lock Foot Switch Cable A7095 01 o swt PEDAL FOOT SWITCH BUCKY DETECTOR WIRING Lock Button Cable See pag 2 Tracking Bucky Tracking Cable A7099 01 SW3 n Nota La funci n de tracking con el SW3 y el Cable Bucky Tracking es opcional Si no se usa esta opci n hacer puente entre TS1 5 y TS1 6 Note The tracking function with the SW3 and the Bucky Tracking Cable is optional Make jumper between TS1 5 and TS1 6 when that option is not used LI T r IT T HERE 54303079 lt SEDECAL MESA MULTI RAD A B C D E IN CASE OF BUCKY ASSEMBLY A7094 01 Lock Button Cable PULSADOR Lock Button sw2 IN CASE OF DETECTOR ASSEMBLY TS1 A6387 07 Bucky Ground Cable Chassis GND 5 CABINET DETECTOR A15066 01 In case of BDG docking A15077 01 In case of BDC docking A15080 01 Cable SW gt A6387 57 Bucky Ground Cable PULSADOR Lock Button Chassis GND ee eee MW DATE TE SHEET OF OF DRAWING Hermosilla 11 06 13 2 2 54303079 REVISED F DIAZ 11 06 13 A lt RE j _ o A NC 13 204 SEDECAL Hermosilla 11 06 13 REV DESCRIPTION
16. distance 102 mm 4 4 0 15 Table Top Attenuation lt 1 0 mm Al eq Longitudinal travel of Table Top 900mm 35 Transverse travel of Table Top 230 mm 9 Longitudinal travel of Receptor 518 mm 20 3 Grid standard 40 lines cm 12 1 Cassettes Detector sizes from 18 to 43 cm from 7 to 17 SM 1103R3 5 General Purpose Rad Table Service Manual Illustration 2 2 Dimensions 769 ol 6 SM 1103R3 General Purpose Rad Table SECTION 3 Service Manual UNPACKING The system is shipped in different crates to facilitate transport and installation Upon receipt of components and parts of the equipment inspect all shipping crates for signs of damage If damage is found immediately notify the agent or carrier 1 Place crates close to the location in the room where the system is intended to be placed and remove each wood packed part Do not discard any packing material envelopes boxes bags until all parts are identified as listed in the packing list AT LEAST TWO PEOPLE ARE REQUIRED TO REMOVE ALL HEAVY COMPONENTS FROM THE SHIPPING PALLET 2 When the equipment is unpacked check the part and serial number of each component s identification label Inspect all pieces for visible damages Ifany damaged part is found repair it or order a replacement to prev
17. e gt For System Interconnection Maps I F 001 I F 003 1 F 045 IM 015 refer to Generator Service Manual Installation Chapter SM 1103R3 33 General Purpose Rad Table Service Manual This page intentionally left blank 34 SM 1103R3 ROT MOBILE COLUMN Note 2 Cable code See schematic 54301085 A3052 XX for only Low Speed GENERATOR CABINET for interconnections A7014 XX for Low High Speed TUBE STATOR CABLE See Note 2 CATHODE CABLE Note 1 Cable code A84 A84 A84 A84 10 01 for Compact Generator single phase 11 01 for Compact Generator three phase 12 01 for Battery Generator 13 01 for Capacitor Generator VAC SUPPLY GENERATOR SUPPLY CABLE See Note 1 C C A A GND CABLE A6383 XX AC LINE Chassis Chassis GND GND COLLIMATOR COLLIMATOR CABLE A3111 XX SUPPLY CABLE A6736 XX RAD TABLE See schematic 54302065 for interconnections Les SUPPLY CABLE A7235 XX Chassis GND GND CABLE A6383 XX Chassi assis GND BUCKY TABLE BUCKY CABLE A3091 XX AEC ADAPTATION BOARD See schematic I F 001 A3263 03 See schematic IM 015 ION CHAMBER AEC CABLE A3253 XX Preamplifier See Note 3 J1 J2 J3 See Note 3 TB1 See Note 3 VERTICAL BUCKY TS1 TS3 See Note 3 ION CHAMBER AEC CABLE A3253 XX Preamplifier See Note 3 Chassis See Note 3 GND POWER I F CABLE A3252 03
18. ent unnecessary delay in installation 3 Verify that all items on the customer order are present 4 Leave a free working area around the equipment until installation is complete SM 1103R3 General Purpose Rad Table Service Manual This page intentionally left blank 8 SM 1103R3 General Purpose Rad Table SECTION 4 Note gt Note Service Manual INSTALLATION Follow Installation sequence process as described At least two or three people are required for the following tasks Remove the transport screws that secure the Radiographic Table to the pallet and move away the Table Remove the front cover screws of the Table in order to access inside the Table Power ON the Collimator Lamp and align perfectly the Receptor axes with the Collimator light beam During alignment place the Tube Collimator Assembly and the Receptor Carriage at different positions to check that it is correctly aligned Mark on the floor the four anchoring holes of the Table Base Move away the Table Base and prepare its anchors Position the Table Base at its final place and anchor the Table Base to the floor Check that it is properly leveled placing levels in different points of the Upper Frame Use leveling plates to level the Base before attaching fixing installing definitively the Table Use an extended level 1 meter or several standard levels placed in different points for leveling the Table Base SM 11
19. ionally left blank 2 SM 1103R3 General Purpose Rad Table Service Manual SECTION 2 PRE INSTALLATION 2 1 2 2 2 3 2 3 1 PRE INSTALLATION CHECK Prior to beginning installation it is recommended to inspect the site and verify that the X ray room complies with Pre installation requirements such as incoming line main switch es safety devices conduits and space requirements Note gt For more Pre installation requirements also refer to Pre Installation Manual of the Generator RESPONSIBILITIES OF THE PURCHASER The purchaser is responsible for the completion of Pre Installation This includes the procurement and installation of all required materials and services to get the room ready for the installation of the product ROOM REQUIREMENTS ENVIRONMENTAL REQUIREMENTS Note F STORAGE values only refer to equipment that is still in shipping containers If the equipment is partially or completely installed refer to IN USE values RELATIVE HUMIDITY Non Condensing TEMPERATURE ATMOSPHERIC PRESSURE STORAGE STORAGE STORAGE MAX MIN MAX MIN MAX 10096 700 hPa 1060 hPa 500hPa 1060 hPa Note gt These environmental conditions do not include the Digital Detector Refer to the Digital Detector Documentation SM 1103R3 General Purpose Rad Table Service Manual 2 3 2 FLOOR REQUIREMENTS The method of installing the Table is anchor to floor with 4 x M10
20. it SAT A9871 01 Receptor Bumper SAT 51865P04 Includes 2 units Receptor Carriage Bearing Left Kit SAT A9869 XX XX 01 Bucky standard XX 02 BDC and BDG Receptor Carriage Bearing Right Kit SAT A9870 XX XX 01 Bucky standard XX 02 BDC and BDG Tabletop Brake SAT S0025036 Lock Pin SAT A11007 01 Microswitch SAT 50605015 Spring SAT 11330 01 Overlay Pedal SAT 55801403 22 General Purpose Rad Table Service Manual SM 1103R3 23 General Purpose Rad Table Service Manual ra A12 A14 A15 24 SM 1103R3 General Purpose Rad Table DESCRIPTION REFERENCE Service Manual REMARKS BUCKY ASSEMBLY Vacutec lon Chamber 6695 21 AID lon Chamber 6695 11 Medys lon Chamber 6695 66 Claymount lon Chamber 6695 56 Bucky USX Ray SAT 6685 11 Sensing Tray SAT 6694 05 Manual Tray SAT A8894 01 Push button SAT 50613014 Handle SAT 54402014 SM 1103R3 25 General Purpose Rad Table Service Manual DESCRIPTION QTY REFERENCE REMARKS DETECTOR CABINET WITH NON ROTATING TRAY Frame 1 5mm for Grid 1m 1 SAT A10521 01 Frame 1 5mm for Grid 1 3m SAT A9657 02 Frame 1 5mm for Grid 1 4m SAT A10523 01 Frame 1 5mm for Grid 1 5m SAT A9656 02 Frame
21. n Non ionizing electromagnetic radiation Radiation of Laser apparatus Do not stare into beam Only applicable to equipment with Laser Pointer Dangerous voltage SM 1103R3 General Purpose Rad Table Service Manual General warning caution risk of danger Warning lonizing radiation Warning Non ionizing radiation Warning Laser beam Warning Dangerous voltage Warning Do not place fingers between mobile and fixed parts of the equipment it may cause serious injuries to patient or operator As well make sure the patient extremities are correctly positioned into limit areas during operation movement of parts may cause serious damages to patient Electrostatic sensitive devices No pushing No sitting No stepping on surface A AN AN A A dus SM 1103R3 General Purpose Rad Table Service Manual Stop of action Emergency stop ON power ON OFF push push Each position ON or OFF is a stable position Alternating current Y Y O Three phase alternating current 3 N NI Three phase alternating current with neutral conductor N Connection point for the neutral conductor on Permanently Installed equipment Direct current SM 1103R3 General Purpose Rad Table Service Manual Both direct and alternating current
22. of this page SM 1103R3 General Purpose Rad Table Service Manual ADVISORY SYMBOLS The following advisory symbols will be used throughout this manual Their application and meaning are described below DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT 7 ANGER IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS PERSONAL INJURY OR DEATH DATA Advise of conditions or situations that if not heeded or avoided could cause personal injury or damage to equipment or data ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT UT WARNING HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL INJURY OR CATASTROPHIC DAMAGE OF EQUIPMENT OR CAUTION Note Alert readers to pertinent facts and conditions Notes represent information that is important to know but which do not necessarily relate to possible injury or damage to equipment SM 1103R3 General Purpose Rad Table Service Manual SAFETY SYMBOLS The following safety symbols may appear in the equipment Their meaning are described below Caution Consult accompanying documents Safety Symbol Follow instructions for use especially those instructions identified with Advisory Symbols to avoid any risk for the Patient or Operator Only applies to IEC 60601 1 Standard Third edition General Mandatory action Type B applied part Protection against harmful ingress of water or particulate matter IP Classification Ordinary lonizing radiatio
23. rfectly aligned with the cam movement to ensure proper activation of the wheel 8 Avoid wrong assembling of Microswitch as shown in illustrations below Illustration 5 3 Wrong Installation of Microswitch Rolling wheel not aligned Rolling wheel not aligned Rolling wheel too close to cam microswitch breaks 9 Re install the Table Front Cover SM 1103R3 17 General Purpose Rad Table Service Manual This page intentionally left blank 18 SM 1103R3 General Purpose Rad Table Service Manual SECTION 6 PERIODIC MAINTENANCE The maintenance service has to be performed every twelve 12 months after the installation 6 1 MAINTENANCE ACTIVITIES DESCRIPTION Electrical cables and connections Visually check for electrical cables proper isolation and proper connection inside and outside the unit Functional Check Perform a functional check of the equipment motion of Table top locks brakes and detents Adjust or replace as needed Adjustments Perform adjustments as explained in the Adjustments Section of this manual Lubrication Lubricate mobile components such as guides bearings etc Cleaning Clean from dust or moist inner components as grids bearings and rails 6 2 CHECKING OF RADIOGRAPHIC TABLE 1 Turn OFF the system and mains Check that there is not voltage at Table 2 Disassemble Front Table cover 8 Check anchoring screws of the Table they should be strongly tighten 4 Turn On
24. the system and press the mushroom shaped switch Verify that table is disconnected and the Table travels are blocked 5 Release the mushroom shaped switch 6 Connect Table power and check Table Top brakes and correct motion by pressing the Table Pedal 7 Assemble the Front Table cover 8 Release the horizontal Receptor brakes and check correct traveling 9 Take out the Cassette film tray and check correct functioning SM 1103R3 19 General Purpose Rad Table Service Manual 6 3 CHECK OFF MAINTENANCE REVISION TABLE Hospital name System Name System Number ID Customer Field Engineer Name Radiographic Table Maintenance activity Comments Electrical cables and connections Brakes locks and detents Functional Check Adjustment Replacement Lubrication Cleaning 20 SM 1103R3 General Purpose Rad Table Service Manual SECTION 7 RENEWAL PARTS A RADIOGRAPHIC TABLE B RECEPTOR ASSEMBLY SM 1103R3 21 General Purpose Rad Table Service Manual DESCRIPTION REFERENCE REMARKS RADIOGRAPHIC TABLE Tabletop Kit SAT A11008 01 Tabletop Bumper SAT 31617 01 Includes 4 units Eccentric Alignment Bearing Kit SAT A9874 01 Concentric Alignment Bearing Kit SAT A9873 01 Axial Bearing Kit SAT A9872 01 Tabletop Lineal Bearing Kit SAT A9842 01 Receptor Brake K
Download Pdf Manuals
Related Search
Related Contents
Kenmore 580.53650 Dehumidifier User Manual Samsung Airtrack Som 3D Bluetooth HW-E551 manual de utilizador Intel Core i7-2600 1 - サンケン電気 HP SC101 Cleaning Solution Genius SP-S105 印刷はこちら - Boysen Speco Technologies CVC71HRB surveillance camera Español - Graphic Communications PERFECTION - C2wireless.net Copyright © All rights reserved.