Home
        MPA160 and MPA161 DigiMelt Student Melting Point System
         Contents
1.   requires little additional effort to measure the melting point of several compositions   Typically a 20 80  50 50 and 80 20   mixture of sample and reference is prepared and  the three tubes are run in the melting point apparatus  If the three melt at the same  temperature it is very likely the two compounds are one and the same     DigiMelt Student Melting Point System ASRS    Maintenance 29       Chapter 3    Maintenance    This chapter includes basic guidelines and recommendations for keeping your DigiMelt  in working condition  Detailed instructions are listed for      1  removal of broken capillaries   2  removal or replacement of viewing lens     3  fuse replacement    Broken Capillaries Inside Oven    With the oven cowl properly installed  capillaries rarely break off inside the oven   However  if this does happen  no tools are required to remove the broken capillaries   Furthermore  there is no need to remove the cowl     1  Wait for the oven to cool                                   Figure I  Remove oven cap  slide out lens holder to expose capillary bores  2  Loosen the thumbscrew on the top of the oven and remove the oven cap   3  Slide the lens holder assembly up and completely off the oven   4  The capillary slots are now exposed  use compressed air or other mechanical means  to remove the broken glass  Be careful not to damage the glass prism below the    capillary slots     5  Replace the lens holder making sure that the largest of the two parallel rectangula
2.  Cooling LED is lit      DigiMelt Student Melting Point System ASRS    Chapter 1 Getting Started 15       Detailed Setup and Operation    1  Connect the power    With the power switch in the off position  connect your DigiMelt to a grounded outlet  using the power cord provided     2  Turn on power    Turn the power switch on and wait for the brief Power on Self Test procedure to be  executed  The firmware revision is briefly displayed  followed by the current  temperature of the oven  The oven is off at this time     3  Prepare the sample s     Once powder is forced into the capillaries  they may be packed by using DigiMelt   s tube  tapper  Place the capillaries in the molded holes in the lower right side of the chassis and  press the Tube Tapper button  If your sample does not decompose or sublimate under  heating  you may choose to put the samples in the oven at this time  A detailed  discussion of the capillary melting point determination methodology  including sample  preparation steps  is included in Chapter 2 of this manual     4  Enter the start temperature    C     The start temperature is the temperature at which the heating ramp is begun     The start temperature is programmed by pressing the Start Temp button on the keypad   The amber START LED will flash and display the currently programmed start  temperature  Program the start temperature using the up and down arrow keys in the  Samples group  If you press and hold an arrow key  the temperature will move by sin
3.  at least 0 5   C  for  routine melting point ranges     e Report onset  meniscus and clear point to nearest 0 1   C for important melting point  ranges  such as those of new compounds     e Ifa single temperature is to be reported as the melting point  not recommended   specify whether it represents the meniscus or clear point     Melting Point Depression    Mixtures of substances  whose components are insoluble in each other in the liquid  phase  display a melting point depression and  instead of a sharp melting point  a melting  range  interval   The size of the melting point depression depends on the composition of  the mixture  The depression in melting point is used for determining the purity and  identity of compounds     Rule of thumb  1  of a foreign substance will result in a 0 5   C depression     This is the main reason why a melting point range is the preferred record of a melting  point determination  and is more useful than a single temperature report     A wide melting range usually indicates that a substance is impure  but it may also result  from the fact that the pure substance undergoes some decomposition prior to reaching its  melting point  Pure substances that decompose during heating form a mixture of the  parent substance and the byproducts and will also show a melting range  In some cases   the material undergoes a slight liquefaction and contraction at a temperature below the  true melting point  In others  the material may decompose and discolor so
4.  badly that a  definite melting point cannot be observed     Purity Tracking    The phenomenon of melting point depression can be applied to the evaluation of a  compound   s purity     ASRS DigiMelt Student Melting Point System    28 Melting Point Determination Chapter 2       In preparative organic chemistry the purity of a substance often has to be evaluated  without a pure reference sample being available  This is the case  for example  when a  new chemical compound is synthesized  The raw product is generally subjected to a few  purification steps  i e  recrystallization or resublimation  and the melting point is  determined at each stage  The onset point continues to increase  and the melting range  continues to decrease  until the substance is either pure or as pure as possible with a given  purification method     It is common practice to recrystallize synthetic products of reactions until no more  changes are detected in their melting point range     Careful reproduction of the sample preparation procedure is essential during purity  tracking determinations  Particular attention must be dedicated to grinding and drying all  samples reproducibly     Mixed Melting Point    If two compounds melt at the same temperature  a mixed melting point  determination can reveal if they are one and the same substance     The phenomenon of melting point depression can be applied to the identification of  unknown pure substances  For example  if you measure the melting point of a sam
5.  holder into your hand  Clean the  lens or obtain a new lens     4  Replace the lens in the holder  The curved side should face out     5  Replace the retaining ring     DigiMelt Student Melting Point System ASRS    Chapter 3    Maintenance 31       Fuse Replacement    1   2   3   4     5     6     ASRS    Under normal conditions  DigiMelt should not    blow    a fuse  Fuse replacement  should be performed only by competent personnel  If the unit continues to blow  fuses immediately after fuse replacement  return the unit to Stanford Research  Systems for repair  NEVER BYPASS THE FUSE              Open fuse door  remove fuse holder  remove fuse    Determine and obtain the correct fuse  For units running on 115 VAC  use a  3AG  1 A 250V fuse  For units running on 230 VAC  use a 3AG  1 2 A 250V  fuse     Turn off the unit  Unplug the unit from line power     Remove the fuse holder from the power entry module  see photos above   A  thin blade screwdriver or similar prying tool may be required     There is a fuse on the    hot    and the    neutral    power lines  Remove both fuses  and replace with new ones     Replace the fuse holder     Attach a power line cord to the unit and turn the unit on     DigiMelt Student Melting Point System    32 Maintenance Chapter 3       DigiMelt Student Melting Point System ASRS    
6.  of possibly explosive composition or  contamination     ASRS DigiMelt Student Melting Point System    iv Safety and Preparation for Use       DigiMelt Student Melting Point System    Contents    Contents v    Safety and Preparation for Use i  Contents v   Instrument Overview vi   Rear Panel viii   Specifications ix    Chapter 1 Getting Started 11    Unpacking 11   Instrument Placement 11  Principle of Operation 12   Quick Start Instructions 13  Detailed Setup and Operation 15    Chapter 2 Melting Point Determination 19    Introduction 19   Capillary Method 20   Visual Observations 23  Melting Point Range 26  Melting Point Depression 27    Chapter 3 Maintenance 29    Broken Capillaries Inside Oven 29  Broken Capillaries Inside Tube Tapper 30  Lens Cleaning or Replacement 30   Fuse Replacment 31    DigiMelt Student Melting Point System    vi Overview       Instrument Overview        1  Cowl    The cowl protects the user from accidentally touching the hot oven  Never run  experiments with the cowl removed      2  Oven Cover    The oven cover retains the lens holder and is removed for cleaning  see Chapter 3 for  details on cleaning broken capillaries     DigiMelt Student Melting Point System ASRS    Overview vii       ASRS     3  Lens and Lens Holder    The lens is retained by a knurled retaining ring  It is easily cleaned  see Chapter 3 for  details  The lens magnifies the view of the samples      4  Keypad    Buttons allow the user to set parameters and store data  Light emit
7.  points     is a valuable indicator of purity of a solid compound     The Melting Point Range is the most popular melting point record listed in scientific  papers  standard procedures  reference tables and melting point standards  It is always  advantageous to record the entire melting range of a substance  especially with    1  unknown or new compounds   2  impure samples   3  mixtures with large melting  intervals  The observed range is an aid in identifying the substance and drawing  conclusions about purity and heat stability     Reporting the melting range   onset point  clear point  of a solid sample  along with the  ramp rate  is the preferred way to report the results of a melt  and is much more reliable  than a single number report     If a single temperature must be used  please specify whether it represents the clear or  meniscus point     The ramp rate affects the melting point range record  and must always be specified for  full compliance with GLP specifications                             Ramp Rate    C min    Onset Point   Clear Point    C    Temp  Range    C   0 1 133 7 134 2 0 5  0 2 133 8 134 4 0 6  0 5 134 0 134 9 0 9  1 134 1 135 4 1 3  2 134 3 136 2 1 9  5 134 9 137 9 3 0                Melt Point Range of Phenacetin at various ramp rates  Notice the larger effect of ramp rate on the clear point than on the onset point     The sample height inside the capillary also affects the melting point range  Since the  temperature displayed by the melting point a
8. Operation and Service Manual    MPA160 and MPA161    DigiMelt  Student Melting Point  System         RS Stanford Research Systems    Revision 1 9  September 2009     Certification    Stanford Research Systems certifies that this product met its published specifications at the time  of shipment     Warranty    This Stanford Research Systems product is warranted against defects in materials and  workmanship for a period of one  1  year from the date of shipment        Service    For warranty service or repair  this product must be returned to a Stanford Research Systems  authorized service facility  Contact Stanford Research Systems or an authorized representative  before returning this product for repair     Information in this document is subject to change without notice   Copyright O Stanford Research Systems  Inc   2009  All rights reserved   Stanford Research Systems  Inc    1290 C Reamwood Avenue   Sunnyvale  California 94089   408 744 9049   www thinksrs com    Printed in U S A     DigiMelt Student Melting Point System SRS    Safety and Preparation for Use i    Safety and Preparation for Use    CAREFULLY READ THE IMPORTANT SAFETY INSTRUCTIONS AND NOTES  INCLUDED IN THIS SECTION BEFORE USING THE DIGIMELT STUDENT    MELTING POINT SYSTEM        Within this section  the word  product  specifically refers to the DigiMelt Student  Melting Point System     This product is designed and built for use in teaching laboratories  It is intended to be  used to determine melting temper
9. System    Getting Started 11       Chapter 1  Getting Started       Unpacking    Read the entire Safety and Preparation for Use section of this manual before starting any  installation procedure     Read and follow all installation and operation instructions in this manual to ensure that  the performance of this instrument and the accuracy of your melting point determinations  is not compromised     Checklist    e Open the box es  and inspect all components of your DigiMelt System s    e Report any damage to Stanford Research Systems immediately     e Compare the contents of the shipping boxes against your original order and the list of  Standard Eguipment Supplies below  Report any discrepancies to Stanford Research  Systems immediately     Standard Eguipment Supplies    e Melting Point Apparatus  DigiMelt  SRS Part  MPA160 or MPA161   e Operation and Service Manual  e One  1  Power Cord    Instrument Placement    e Place your DigiMelt on a stable  clean  level surface and away from any solvents     e Place your DigiMelt away from water sources  i e  faucets  safety showers   eyewashes  rain  etc   Do not allow the unit to become wet     e No containers  chemicals or other appliances should be placed near the product     e Always operate the unit in its proper upright orientation  Do not operate the unit on  its side     ASRS DigiMelt Student Melting Point System    12 Getting Started Chapter 1       Principle of Operation    DigiMelt employs a microprocessor controlled heat
10. USP25 NF20   The clear point  described below  is  identified as the    melting point    of a substance instead  Notice that this is a significant  difference in interpretation between the British and US Pharmacopeias     DigiMelt Student Melting Point System ASRS    Chapter 2 Melting Point Determination 25       Clear or Liquefaction Point    The clear point corresponds to the stage of the melt at which the substance becomes  completely liquid     no more solid is left  i e  the last crystals are melted      KARGAS     P     p              The clear point    The clear point is more dependent on the ramp rate than the onset point  In general  the  clear point increases with increasing ramp rates  see below under Melting Point Range  for a table of ramp rates and corresponding clear points      The clear point corresponds to the high temperature record in the melting point range of a  substance  and is most often the single temperature listed in melting point tables     The clear point is the temperature most often listed in US based Melting Point tables and  the only one accepted by the US and International Pharmacopeias as the    single    melting  point of a substance     DigiMelt Student Melting Point System    26 Melting Point Determination Chapter 2       Melting Point Range    In a dynamic melting point determination  where true equilibrium between solid and  liquid phase is never achieved  the Melting Point Range     defined as the interval between  the onset and clear
11. age used in your location   The DigiMelt MPA160 is wired for 115V  The DigiMelt MPA161 is wired for 230 V  operation  The expected line voltage is clearly marked on the rear of the unit  If the  unit is not marked for the line voltage in your location  return it to SRS     Burn Risks    A common risk associated specifically with the operation of thermal    analysis instrumentation equipment is burns        e Acowl around the oven protects the user from accidental burns  DO NOT REMOVE  THE COWL  Do not use the cowl as a carrying handle  Chapter 3 shows  how to perform routine maintenance on your DigiMelt without removing the cowl     e Do not touch the oven while it is hot     e The unit will display the oven temperature on power up  Check this temperature  before handling the oven     DigiMelt Student Melting Point System ASRS    Safety and Preparation for Use iii       If the oven temperature exceeds 280 degrees Celsius  the temperature display will flash  and an alarm will sound  This indicates a serious control malfunction and the unit should  be returned to SRS for repair as soon as possible     Explosion Risks    Injury due to explosion is another important safety concern during the    operation of thermal analysis instrumentation        e This product is not compatible with application environments requiring  explosion   proof equipment  or compatibility with samples which may explode or ignite by heat   friction or spark     e Do not use this product to analyze samples
12. and Preparation for Use       Electrical Shock Risks    The most common risk associated with the operation of chemical    instrumentation equipment is electrical shock        e Itis your responsibility to install and operate this product in full  conformance with local electrical codes  Consult an experienced electrician  if necessary     e If the power cord becomes damaged  replace it immediately     e Dangerous voltages capable of causing injury are present during the operation of this  product  Do not remove the covers while the unit is plugged into a live  outlet     e Do not use this product if it has unauthorized modifications  Unauthorized  modifications may result in fire  electric shock and other hazards     e Do not install substitute parts or perform any unauthorized modifications to this  instrument     e Always use an outlet which has a properly connected protective  ground  Consult with an experienced electrician if necessary     e GFCI  Ground Fault Circuit Interrupter  protected outlets are often available in  laboratory environments  particularly in proximity to water sources  GFCTs are  generally regarded as an important defense against electrocution  However  the use  of a GFCI in conjunction with the DigiMelt must NOT be regarded as a  substitute for proper grounding and careful connections  GFCI s must  also be tested regularly to verify their functionality  Always consult an electrician  when in doubt     e Make sure that the unit is marked for line volt
13. atures and melting ranges between ambient and 260  C  temperatures     Though long experience has proven melting point instrumentation to be remarkably safe   hazards are always associated with analytical eguipment  The most effective way to  minimize risk to yourself and others is to read and follow all safety instructions     Safety Instructions and Warnings    e SAFETY PAYS  Safety instructions must be strictly followed during all stages of  installation  operation and service of this product  Failure to comply with these  precautions and warnings violates the safety standards expected of users of this  product     e This manual is a component of the product and must remain readily available to all  laboratory personnel with access to the product     e If you have any doubts about how to use this product safely  contact Stanford  Research Systems using the contact information provided in this manual  see inside  cover      e Do not use this product for any purpose other than its intended usage   e Retain these safety and operating instructions for future reference   e Identify and adhere to all warnings posted on the product and throughout this manual     e Failure to comply with these instructions may result in serious personal injury   including death  as well as significant property damage     e Wear protective garments such as lab coat and goggles at all times     e Refer servicing to qualified personnel only     ASRS DigiMelt Student Melting Point System    ii Safety 
14. e is the leading cause of inaccurate and  irreproducible results in melting point determinations     Any substance being loaded into a melting point capillary must be  1  fully dry    2  homogeneous and  3  in powdered form     Moist samples must be dried first     48 hours over POs  in a dessicator usually works  quite well     The primary requirement for a good melting point determination is that the sample be in a  fine powder form  This assures efficient and reproducible heat transfer into the sample  and enhances the overall appearance of the sample for easier detection of the melt   Coarse crystalline and non homogeneous samples must be crushed into a fine powder in  a mortar  An agate  glass or alumina mortar and pestle are recommended     To fill a capillary tube with a sample  the open end of the capillary is pressed gently into  the substance several times  The powder is then pushed to the bottom of the tube by    DigiMelt Student Melting Point System ASRS    Chapter 2    Melting Point Determination 21          repeatedly tapping the bottom of the capillary against a hard surface  The integral tube  tapper in DigiMelt does this very well  Place each capillary in the molded form on the  chassis  see photo below   Press the tube tapper button and your samples will be packed   If desired the tube tapping sequence may be repeated  Alternatively  the capillary tube  can be dropped onto a table through a glass tube of  1m in length              Loading a sample into a cap
15. eterminations are more than just a classroom exercise in  the organic chemistry laboratory  The test is still an important technigue for gauging  purity of organic and pharmaceutical compounds     The determination of melting points is one of the oldest identification and test methods  for organic substances  The melting point is easy to measure  tabulate and classify   Extensive collections of tables give the exact values of many pure compounds  The  melting point determination is a fast and cost effective technigue and remains a strong  link to the vast pre instrumental chemistry literature     ASRS DigiMelt Student Melting Point System    20 Melting Point Determination Chapter 2       Capillary Method    In this methodology  a thin glass capillary tube containing a compact column of the  substance to be determined is introduced into a heated stand in close proximity to a high  accuracy thermometer  The temperature in the heating stand is ramped  at a user   programmed fixed rate  until the sample in the tube transitions into the liquid state   While determining a melting point  several observations and the temperatures in each  case  are recorded              Capillary tubes with solid sample     Precision melting point capillaries specifically designed to fit the sample slots and  provide the most uniform and repeatable results are available from SRS  part number  O100MPC  and other laboratory supply distributors     Sample Preparation    Careless preparation of the sampl
16. g Point System ASRS    Chapter 2 Melting Point Determination 23       Visual Observations    Several noticeable changes take place in the capillaries during a melting point  determination  Subjectivity in the interpretation of the physical and chemical changes  observed during the heating ramp can be an important factor affecting the reproducibility  of melting point results     The following events should be noted  and their temperatures recorded  to provide a  complete record of the changes observed in the samples during the melt     First signs of change  Record the first signs of change in the samples  Early changes may be due to   1  loss of solvent  dehydration    2  change in crystallization state  shriveling    3  slow onset of decomposition  darkening or change of color   4  condensation of solvent in the coolest points of the tube    5  individual isolated crystals starting to melt without the liquid showing up as a  cohesive phase     i e  Sintering Point     Onset Point    The onset point is generally considered the    official    start of the melt  liguid clearly  appears for the first time as a separate phase in coexistence with the crystals  It must not  be confused with the    sintering point    which corresponds to melting a few surface    crystals            AA     a                The onset point  also called collapse point    ASRS DigiMelt Student Melting Point System    24 Melting Point Determination Chapter 2       The onset point corresponds to the lowe
17. gle  degrees  then by five degrees  When satisfied with the setting  you can press the Start  Temp key again to return to the current temperature display or go on to the next step     The Start temperature should be a few   C below the nominal melting point of the  sample s   If no keys are pressed for 30 seconds  the display reverts to showing the  current oven temperature     5  Enter the ramp rate    C per minute   The ramp rate dictates how much time is required to raise the oven temperature from the    start temperature to the stop temperature     The ramp rate menu is selected by touching the Ramp Rate button on the keypad  Once  the key is pressed  the amber RAMP LED flashes  Use the up and down arrow keys in  the Samples group to select a rate between 0 5 and 20   C per minute     ASRS DigiMelt Student Melting Point System    16 Getting Started Chapter 1       The ramp rate is the most important instrument parameter affecting the accuracy and  reproducibility of measurement of melting points  Rates up to 2   C per minute are  reasonable for routine determinations  Higher rates are only recommended for quick  determinations on substances with unknown melting points  Purity determination and  precision measurements are performed at a maximum heating rate of 0 5   C per minute   If no keys are pressed for 30 seconds  the display reverts to showing the current oven  temperature     6  Enter the stop temperature    The stop temperature is the temperature at which the heati
18. hat the internal wiring and fuses are different for the  nominal  115 VAC version  and the  nominal  230 VAC version  Your DigiMelt will be clearly marked indicating    which line voltage is expected  If your unit indicates it expects a different line voltage  for your region  DO NOT use the unit  contact Stanford Research Systems     DigiMelt Student Melting Point System ASRS    Overview ix       Specifications    Display    Type   Readout     Temperature    Resolution   Accuracy     Reproducibility    Oven    Range    Ramp rates   Heat up time   Thermometer     Temperature control     Safety shut off     Capillaries    Dimensions   Capacity    Fill height   Sample packing     General    Power   Environment   Humidity   Weight   Dimensions     ASRS    4 digit LED  Start  stop or ramp rate  oven temperature  and stored measurement data    0 1   C      0 45   0 0045 T    C  where Tis in degrees Centigrade  Typical   0 6   C   lt 200   C    1 0   C  2200   C    0 2   C    50   C to 260   C   0 5  1 0  2 0  5 0  10  20   C per minute    2 5 minutes to 100   C   6 5 minutes to 250   C  Built in platinum RTD   Closed loop digital PID   Automatic after 30 minutes of idle time    1 4 mm to 1 8 mm outside diameter  100 mm length  3 tubes   2 mm to 3mm   Automatic  tube tapper     115 or 230 VAC  60   50 Hz  75W  0   C to 30   C  non condensing    lt  90    2 Ibs    6 5  x 9 5  x 5 25   WxHxD     DigiMelt Student Melting Point System    x Overview       DigiMelt Student Melting Point 
19. illary tube  loading tube tapper    In addition to tight packing  maintaining a fixed level in the fill is also a very important  reguirement  Taller samples reguire extra heat to completely melt and usually display  larger melting ranges than their shorter counterparts     A sample height between 2 and 3 mm is recommended in the DigiMelt for optimum  results and reproducibility     It is good practice to wipe the outside surface of capillary tubes with a clean cloth before  inserting them into the heating stand     If your sample is hygroscopic  or sublimates at high temperatures  the open end of the  capillary tube must be sealed by heating  Hygroscopic samples must be stored in a  dessicator between tests  This is particularly critical in humid lab environments     The sample tubes are loaded into the DigiMelt by inserting them into one of the sample  position slots located on top of the instrument  Up to three samples can be  accommodated by the heating block  Loading three capillaries with the same substance  and melting them at the same time and averaging their melting points provides the fastest  and simplest way to improve the repeatability and accuracy of melting point  determinations     DigiMelt Student Melting Point System    22 Melting Point Determination Chapter 2                            Loading capillaries into DigiMelt     Never force a capillary into the heating block  See Chapter 3 if there is an obstruction to  the capillaries     DigiMelt Student Meltin
20. ing ramp  A built in resistance  temperature detector  RTD  provides feedback to the microprocessor so the control loop  continuously adjusts power as needed to maintain the desired temperature or ramp rate   The microprocessor also continuously displays the temperature to the student on the  numeric display  The keypad provides three keys to record temperatures  up to four for  each of the three capillaries   the student need never take their eyes off their samples   Front panel LEDs indicate important events such as temperature stabilization or end of  melt     The inclusion of three independent sample slots in the heating block allows simultaneous  analysis of up to three samples  thus ensuring high sample throughput     A typical analysis only takes a couple minutes to set up   e Select a start temperature    C   e Select a stop temperature    C     e Select a ramp rate    C minute     In a typical analysis  the Start temperature is programmed a few degrees below the  expected onset of the melting point  and the Stop temperature is set a few degrees above  the melting point ranges of the compound s  being tested  Programmable ramp rates  from 0 5   C per minute to 20   C per minute provide measurement flexibility     The unit first preheats the oven to the Start temperature  The oven remains at that  temperature until the user is ready to carry out the test  During analysis  the temperature  is ramped at the specified rate between the start and stop temperatures  Visual a
21. light at this time  To retrieve the data  push the sample buttons   see photo  below   The DATA LED and the LED near the sample button will flash to  show which sample data are displayed  Pressing the sample button again will retrieve the  next flagged datum  After four data are displayed  the display    wraps    back to displaying  the first stored data point  If less than four data are recorded            is displayed for each  null datum  If no keys are pressed for 30 seconds  the display reverts to showing the  current oven temperature        DigiMelt     gt  START    lt   RAMP           gt  STOP    lt     DATA    C OR   C MINUTE    TUBE  TAPPER    FORD RESEARCH SYSTEMS       Retrieving Flagged Data    12  Turn Oven Off    DigiMelt automatically turns off the oven when the stop temperature is reached  or if the  user stops the melt with the Start   Stop button  In either case  the COOLING LED will  illuminate  The oven is not actively cooled  the COOLING LED simply indicates that  the oven is off and cooling to ambient temperature     Also note that the DigiMelt will automatically turn off the oven after 30 minutes of idle   i e  no buttons are pushed  time  This is a safety feature  it prevents accidentally leaving  the oven at elevated temperatures indefinitely  If the automatic shut down occurs   DigiMelt also turns off the white LEDs illuminating the sample to conserve their useful  life  Simply press a key to return the DigiMelt to normal operation     DigiMelt Stude
22. lt   gt  DATA  C or  CIMHUTE  a N          TUBE  TAPPER    7    1  Push Start Temp and use the 4Y2 and   3 buttons to set the starting temperature  generally 5  degrees below the expected melting point               SJ ID                      4       Schematic of DigiMelt Keypad    2  Push Ramp Rate and use the 4Y2 and W 3 buttons to set the ramp rate  2 deg min is suggested      3  Push Stop Temp and use the 4Y2 and W 3 buttons to set the stop temperature  at least 5 degrees  above the expected melting point      4  Push Stop Temp again to return to the current temperature display     5  Load capillaries with sample  Insert capillaries into the chassis holes near the Tube Tapper button   Press the Tube Tapper button to pack your samples     6  Push Start Stop to preheat the block to the starting temperature  The Preheat LED will light     7  When the Ready LED becomes lit  the oven is holding at the start temperature  Insert your samples  into the DigiMelt oven     8  Push Start Stop to begin ramping the temperature at the ramp rate  The Melt LED will light   9  Observe your samples during the ramp     10  Push the 1  42 and W 3 buttons to record data  up to 4 temperatures per sample  during the melt    To end the experiment before the stop temperature is reached  push the Start   Stop button      11  When the Cooling LED is lit  the experiment is over  If you recorded data  the Data LED is also lit     12  To read back the data  push the 1  42 and   3 buttons  make sure the
23. nalysis  takes place during this time        Preheat  Cool Down    Maximum  Heating Rate    Stop Temp    StartTemp    Ambient  Temp       Time    DigiMelt Student Melting Point System ASRS    Chapter 1    Getting Started 13       When the stop temperature is reached  or when the user aborts the ramp   the oven turns  off and begins to cool down to the ambient temperature     The ramp rate during a preheat cycle is roughly 50   C per minute  If the melting point is  completely unknown  the preheat cycle  set starting temperature to 250   C to give the  maximum heating rate  can be used to obtain an estimate of the melting point  followed  by a more accurate determination with a controlled ramp rate and a reduced temperature  range     Auick Start Instructions    ASRS    The instructions on the next page are designed to give the user the fastest route to running  an experiment  Emphasis was placed on creating a single page so that instructors could  photocopy the page and hand it out to students as part of their pre lab documentation   The quick start makes the assumption that the instructor has already placed the  instrument in a suitable location and powered it up  A detailed setup section follows the  quick start     DigiMelt Student Melting Point System    14 Getting Started Chapter 1       DigiMelt Quick Start Guide                            PREHEAT  lt   gt           lt  gt  START  READY  lt   gt    Lf    lt   gt  RAMP  MELT   gt          lt   gt  STOP  COOLING   gt     
24. ng ramp is terminated     The stop temperature is programmed by touching the Stop Temp button on the keypad   The amber STOP LED will flash and display the currently programmed stop  temperature  Program the stop temperature using the up and down arrow keys in the  Samples group  If you hold an arrow key  the temperature will move by single degrees   then by five degrees  When satisfied with the setting  you can press the Stop Temp key  again to return to the current temperature display  If no keys are pressed for 30 seconds   the display reverts to showing the current oven temperature     7  Preheat the oven to the start temperature    Touch the Start   Stop button on the keypad to begin preheating the oven to the start  temperature  The PREHEAT LED will turn on     The READY LED is turned on as soon as the start temperature is reached and becomes  stable enough to begin a ramp at the specified rate     8  Insert the sample capillaries    Insert the capillary sample tube s  into the oven  Never force a capillary into the heating  block  If the capillary doesn   t slide into the oven  there is probably an obstruction  See  Chapter 3 on broken capillaries if you suspect there is an obstruction in the oven     Note that if your sample decomposes or sublimates upon melting  you can minimize the  exposure to high temperature by inserting the capillaries after the oven preheats     9  Initiate Heating Ramp    Once the oven is preheated and the READY LED is on  touch the Start   St
25. nt Melting Point System ASRS    Melting Point Determination 19       Chapter 2    Melting Point Determination    This chapter includes basic guidelines and recommendations designed to maximize the accuracy of  melting point determinations with DigiMelt  It also includes in depth details on sample preparation   visual observations        Introduction    A few basic guidelines must be carefully followed to avoid errors during melting point  determinations with DigiMelt  How the sample is prepared and the instrument is  configured have the greatest influence on the accuracy and reproducibility of a melting  point measurement  Visual melting point determinations are inherently subjective   Result reproducibility between different users may be compromised unless all users agree  on the same visual cues     Background    The melting point of a substance is the temperature at which the material changes from a  solid to a liguid state  Pure crystalline substances have a clear  sharply defined melting  point  During the melting process  all of the energy added to a substance is consumed as  heat of fusion and as a result  the temperature remains constant throughout the phase  transition     Determining the melting point is a simple and fast method used in many diverse areas of  chemistry to obtain a first impression of the purity of a substance  This is because even  small guantities of impurities change the melting point or at least clearly enlarge its  melting range  Melting point d
26. op button on  the keypad again to initiate the temperature ramp  The MELT LED will illuminate   After a brief delay the oven temperature will commence to rise at the specified ramp rate   The capillary tubes must not be disturbed while the ramp takes place  The heating ramp  is terminated when  1  the stop temperature is reached or  2  when the Start   Stop  button is pressed     Touch the Start   Stop button at any time to terminate the heating process before the stop  temperature is reached     DigiMelt Student Melting Point System ASRS    Chapter 1 Getting Started 17       10  View the Melt    Visualization of the melt during the heating ramp is via the lens located on the front of  DigiMelt  All three capillaries can be observed simultaneously     A  4  ot  Si  T  am   gt    vi  S                     Onset  Meniscus  and Clear points of Vanillin     Important changes that take place in the capillary tubes can be manually flagged by  touching the individual blue    sample buttons    on the keypad  Sample channel    1    is the  left most sample in the block  channel    2    is the center sample  and channel    3    is the  right most sample        Flagging a point during the melt    ASRS DigiMelt Student Melting Point System    18 Getting Started Chapter 1       11 Recall the data    Once the melt has completed  either by reaching the stop temperature or by the user  pressing the Start   Stop button  the COOLING LED will light  If data were stored  the  DATA LED will also 
27. ple at  160   C  you will find from the melting point tables that this is the melting point for  several different reference compounds  The substance can be identified by determining  its mixed melting point     the sample is mixed one by one with small amounts of the  references and the mixed melting point is determined in each case  Whenever the  melting point of the sample is depressed by mixing a small amount of a reference with it   the two substances cannot be identical  However  if the melting point of the mixture does  not drop  the reference substance that was added was identical to the sample  i e  the  sample has been identified      The mixed melting point technique is the main reason why most high quality melting  point measurement systems can accommodate a minimum of three capillaries in their  heating blocks     In its most common implementation three melting points are determined   1  sample     2  reference and  3  reference sample    1 1  If the melting point of the mixture remains  the same  then the two substances are identical  If the melting point is lowered then they  are two different substances     The requirements for precision and reproducibility are not as high here as when doing a  high precision single melting point determination  Heating rates as large as 5   C min are  acceptable     A few pairs of substances show no melting point depression when mixed  but more  frequently the failure to depress may be observed only at certain compositions  It
28. pparatus does not correspond to the exact  temperature in the melting substance but to that of the oven  higher clear point values are  obtained for taller samples  The sample height recommended for the DigiMelt systems is  2 3 mm     The capillary geometry  diameter and wall thickness  affects the melting point range   Thinner capillaries load smaller amounts of sample  but also provide decreased thermal    coupling with the block     A large majority of pure organic compounds melt within a range of 1 25  0 5   C or melt  with decomposition over a narrow range of temperature   2   C  at heating rates about    DigiMelt Student Melting Point System ASRS    Chapter 2 Melting Point Determination 27       1   C per minute  Many organic compounds melt with decomposition over a considerable  range of temp  amino acids  salts of acids  salts of amines  carbohydrates  etc     Impure substances  1 e  mixtures  also melt over a larger temperature range     Melting Point Report    A complete melting point report should include enough information to make it possible  for somebody else to reproduce the determination and compare results  Very useful  reporting guidelines  compatible with modern GLP reguirements  were set forth by Carter  and Carter  J  Chem  Ed   72  1995  647  and are listed here     e Report all instrument settings  especially heating rate  so they can be duplicated or  reasonable adjustments made     e Report onset and clear point temperatures to the nearest 0 1   C  or
29. r  faces of the prism is vertical and up against the front face of the oven  The large    DigiMelt Student Melting Point System    30 Maintenance Chapter 3       angled face of the prism should be facing up and slant down away from the oven   Replace the oven cap and gently tighten the oven cap thumbscrew with your fingers     Broken Capillaries Inside Tube Tapper    If capillaries are left in the tube tapper holes instead of in the oven  a careless gesture can  snap off the capillaries  If the capillaries break above the molded holes in the chassis   simply grasp them carefully with a tool and remove them  It is unusual  but the capillary  can break below the molded holes  If this occurs     1  Make sure the thumbscrew securing the oven cap is finger tight  This prevents the  lens holder from falling out in the next step     2  Turn the instrument upside down  The capillaries will fall out of the tube tapper  holes     Lens Cleaning or Replacement    Under normal conditions  the lens should rarely require replacement  However  the  DigiMelt is designed to make removal of the lens for cleaning very simple  You can  remove the lens without using any tools  Furthermore  you can remove the lens without  removing the lens holder from the instrument  and without removing the cowl     1  Wait for the oven to cool                    Remove retaining ring to access lens  2  Unscrew and remove the knurled lens retaining ring     3  Tilt the instrument so that the lens falls out of the
30. r temperature recorded in the Melting Point  Range of a substance     The US and International Pharmacopeias describe the Onset Point as  the temperature at  which the column of the substance under test is observed to collapse definitely against the  side of the tube     This is also defined as the    collapse point        Meniscus Point    The meniscus point corresponds to the stage of the melt when the meniscus of the liquid  becomes visible  there is a solid phase at the bottom and a clear liquid phase on top with a  well defined visible meniscus  This point is readily detectable except occasionally when  air bubbles from the bottom push unmelted solid to the surface              The meniscus point    Since the meniscus point represents the time during which liquid and solid coexist it is  often considered a good representation of the    thermodynamic    melting point of a  substance  However  this correlation is only accurate at very low ramp rates     The meniscus point is often the temperature listed in European Melting Point tables and  the preferred value of the British Pharmacopeia methodology     In an attempt to remove subjectivity from its detection  the Laboratory of the Government  Chemist  LGC  defines the meniscus point as    the point where a definite meniscus is  visible and there are equal volumes of solid and liquid in the capillary        The meniscus point is not specifically mentioned by the US Pharmacopeia Melting Point  methods  Method  lt 741 gt  of 
31. ting diodes  LEDs   indicate the state of the instrument  The numeric display indicates the current  temperature or experimental parameters      5  Tube Tapper    Samples within a capillary may be packed using an integral vibrating board inside the  chassis  There are molded holes within the chassis that guide the capillaries to the board   A button on the keypad activates the vibration     DigiMelt Student Melting Point System    viii Overview       Rear Panel    100 to 132 VAC 1 AMP E    MADE IN U S A TA 5010 60 Hz    DigiMelt MPA160  git STen    ASRS 5  Made in the USA  uer  S       WARNING    NO USER SERVICEABLE PARTS INSIDE a   REFER SERVICE AND FUSE REPLACEMENT complies with   TO QUALIFIED TECHNICIAN  UL 61010 1   INPUT VOLTAGE IS NOT USER SELECTABLE  UL61010A 2 010   SEE OPERATION MANUAL FOR SAFETY NOTICE CSA C22 2 NO 61010 1    Certification   E112896       Power Entry Module  The Power Entry Module is used to connect the DigiMelt to a power source through the  power cord provided with the instrument  The center grounding pin ensures the entire  chassis is grounded  Connect DigiMelt to a properly grounded outlet  Never bypass the  safety ground  Consult an electrician if necessary   DigiMelt is turned on by flipping the power switch on the Power Entry Module   The power requirements are either   100     132 VAC  1 AMP  common in USA and elsewhere   OR  200     250 VAC  32 AMP  common in Europe and elsewhere   The power line frequency can vary from 47 to 63 Hz   Note t
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Library Binding Preparation Manual  Tripp Lite USB 3.0 Laptop Dual Head Docking Station - HDMI and DVI Video, Audio, USB Hub Ports and Ethernet    Bedienungsanleitung  CATALOGO DE PRODUCTOS CATALOGO DE PRODUCTOS  Guide to "Finding Your Feet" (or what you should know, when and  Samsung Galaxy S5 mini Bruksanvisning  Data Exchange Benutzerhandbuch    Copyright © All rights reserved. 
   Failed to retrieve file