Home
        Innovet Select Service Manual pdf
         Contents
1.                                          23  8 6 Full kVp range inaccessible                   tu gm Et mM due dut m UM 23  8 7 Poor film                                                                                   23   9 0  WAHRANTY NOTHGE                                    qux SEU FERA UOCE RUE 25    InnoVet Select     Summit Industries  Inc        Installation and Service Manual Revisions History    RELEASED         ADDED    COLD    AND    READY    DISPLAY OCTOBER 2002  MESSAGES  EXPOSURE COUNT DISPLAY    ADDED 50 HZ  REFERENCE AND TIME JULY 2006  STATIONS       iii    InnoVet Select       Summit Industries  Inc        1 0    INTRODUCTION    Congratulations  You ve purchased                   Select Veterinary X Ray System   The InnoVet line represents a new concept in veterinary radiography by being  the first system to incorporate the x ray generator into the table  This major  innovation led to other significant design breakthroughs yielding the most  practical  efficient and compact unit ever created  The result of this innovation  and improvement is the InnoVet Select you have purchased     Because the InnoVet Select is so unique  its features and functions are a bit  different from those of traditional x ray systems  The purpose of this manual is to  familiarize you with the InnoVet Select and assist you with operating and  maintaining your new system  The manual is divided into indexed sections for  quick and easy access to information     Should a situation ar
2.                                 5 8  5 5 FOOT TREADLE INSTALLA                aep        etie 5 9  S OUHPIEMIBINTANSTABENTIOIC              oet ed        onte        5 9  5 7    GENERATOR INS TAEDA THON                   eoe EpL ga eet ee posi rite ee entres 5 10  ELECTRICAL                              6 1  6 1 LOCATION OF ELECTRICAL GONNEGTIGONYS          5 c troi rrr                 6 1  6 2 MAKING CONNECTIONS                   edo conde neca             A      6 3               VION              aM                                   A AN                7 1  71 PREGALIBRATION GHEGIS aped                     7 1  7 2 MA CALIBRATION                                              La pia          7 3  7 3                               a eaa a               dg e Ser              kash 7 6  7 4  CALIBRATION OF KVP WITH A MEANS TO MEASURE ACTUAL                            7 7  7 5 CALIBRATION OF        WITHOUT A MEANS TO MEASURE ACTUAL KVP              7 9  7 6 CONTROL BOARD DIP SWITCH SETTINGS                     eem 7 11  7 7 X RAY TERMINATION BEEPER VOLUME CONTROL                                            7 12  7 8 TUBESTAND DISPLAY MESSAGES                                             ete                           7 12  7 9 DETAILED TROUBLE SHOOTING GUIDE              uertit ntt rne 7 13    Innovet Select        Summit Industries  Inc  Installation and Service Manual  TABLE OF CONTENTS   8 0 FINAL ADJUSTMENTS AND QUALITY     0                                                   8 1
3.          100 92009    298   411               001 2013         son    667    jo        Moo          vQH    Z42Q                                                                   21  23    1 108254        2    955    008   610 01    00158215                     225 2200    2429 15011 1281  119 11 0S0 11 09                     82 5454 508    85   980 OL       66  64    26      37100 619                 V         Quvo8 1    15  YOLOW OL        2     2H              108   1 206654  gen         2  EN                 Y3MOd    20  21                TIWUA  2021    QNnON9 SOTWNY 20 6                         922         lt                                       213  60     9    oor   019  928 574 OL         ocu           092                   Ov     M3AOSSONO 0432             42974      wol Seu         92      l x66   9268                No i             001  OL            e           432              012 vor    3228 zey  59A 59A    135440    188 090    033    59A                               zea                          mol    162    2     iL     X00 TH3INI 7      012313                1113    OL                     PPA     PPA                                   74     5A  21                 476501     QN9           2    VCVI6ZWPN                    0 oo     t                      0121           7526    0350 10  2   6    m i  i 0350 10      522  ALOS           V           PPA  et           37650         8151          ONS      1191   VZLNGZWEN         D  VEGPNL    bawit           ton   
4.          V3SONSZNWPN LYA             20  02                        206                                           asn 754  d    36 6v                        2            001                     PPA                     10     1    001  ved    vid                   062     YILIN                 COMPONENT SILK SCREEN  REF  AWJ401G 01        1 1                       1 4 1 8        H10                3         3 BEEPER SUMMIT INDUSTRIES INC  878879     Mode in U S A       2                   Poon CONTROL BOARD iE  1    5   m       uoo       WEN   22  22                   Q0  SERIA      O       PER                             A C27 C28 C29 C30C31 C32 C33  42                    C 22 01 R3 C26 R77                             12 88                    C3 Ut2              22 2      011 9   8                                                ZENER2  BN         zE 2525                       Stone      1100005   18               mE  4  5                 6     2              2        680                                E                FIL2         R103            SES sss              LED5 L M R95 R96   e   ae xa     2 TP3 TP4     5     6     7 TP8                   39  40 016 H8  C9 nis C  AA M TP9     12    8 j 8 05420  ei      Oui         R106    897 Po  HH     6   1 eh uC  R98  d 86      03 04   920       0       C148          108         pagan P1 30 OFF P2 100 OFF P3 300 OFF PIO  400 OFFES       5 88 Ch i Sui R07 1 R99         TP  4      5v No C17 Gnd Gnd n TRIA    MOV   5      Out tors  6
5.          Wa  1      12 ae  dd  14  12  9           8     IO  dd       S  015  Ve    20 N                                  So     eau TA 39A    NI 39010                               6                          1502  3                  2001 2700  20001         84 288 tsa  32A     lt  2        e               ven mro    20701717  38    002 353        X 113        9992 8   8 59  2   86 50 t0  2S                             866 V    gt           avg   481      1993141     0  9 NI Viva    a    MOIVNOS34                                        340022       NI 2070       COMPONENT LAYOUT for   350 DISPLAY                b    B A  sd  at                     2 0            ES   u5 SUMMIT INDUSTRIES INC   Made   n 1 5 4     DISPLAY PCB   PCBS J351  REV              4350  REV         a                             R27  RB2 884 U4 828           00000      R29     81 883 821 822  R23  R25  R26  3    000                     Swi    R76 R76A  C6B C5A    nm  P    penes     ux    TP6     20vP 0        pde        00130 000  RC Timing Control    2   1   SUMMIT INDUSTRIES        RC TIMING CONTROL GND e    5VDC  ASSY  01030 000 REV MEE        4 SN          C  R4 R3 02 R2         Eme       TOLERANCE UNLESS O           015 LAN  OTIM            
6.        Under no circumstances  does this voltage need to be more than 56 VAC    Take exposures and measure the mA or mAs output  Readjust the  filament voltage as needed to bring mA into range     Innovet Select     Summit Industries  Inc  Installation and Service Manual       4     If the filament voltage is correct but the output is   gt  the expected mAs  the SCR or high voltage transformer may be half waving  See 11 0  below     B  Determine if the line voltage drop is excessive  this test is only  valid after the mA calibration has been verified      1     Place a voltmeter  set for min max reading or to the hundreds of volts  scale  between L1 and L2 of the control  Record the no load line  voltage  Measure and record the line voltage during an exposure of  90 kVp  300 mA  300 mAs  The drop should be less than 7   If the  lights dim  the supply wires rattle in the conduit or the wall breaker  buzzes  the wire gauge or distribution transformer may be undersized     C  Determine if high voltage breakdown is present  hypot test     T     Inspect the high voltage cables  cable ends  and high voltage  receptacles for carbon tracking and evidence of high voltage  breakdown       Verify that the anode rotates and has no visible melt marks     With the high voltage cables removed from the A700 transformer  fill    the receptacles with at least two inches of dielectric oil       Remove the jumper between TB3 19 and TB3 20 and insert a mAs    meter set for AC and at maximum range
7.        to the ground lug     Innovet Select     Summit Industries  Inc  Installation and Service Manual       7 0 CALIBRATION    7 1  lt  PRECALIBRATION CHECKS    1     It is recommended that the assembler read and understand the  information provided with the x ray tube prior to making any            exposures  Particular attention should be given to     A  Initial seasoning of the x ray tube  B  Single exposure tube ratings  C  Accumulated heat and anode ratings of tube    Line voltage adjustments     Terminals on TB2 are provided for coarse and fine adjustments of line  voltage  Measure the line voltage at the disconnect switch and relocate  wires marked  LVAC  and  LVAF  to two terminals where the sum of the  two terminals markings equal the measured line voltage   2         For  example  If line voltage is 240 VAC  connect              to  224            LVAF     to     16        Prior to turning   on  power  set each selector switch as follows     TABLE 6   Selector Switch Settings       Power    On Off                       kVp major and   Fully Counterclockwise  minor       mA selector 300L                Time selector 1 120 sec  for 60 Hz   or 1 100 sec  for 50 Hz         In order to prevent accidental production of x rays during initial check out   disconnect the leads marked P1 and P2 from TB 1 and install a jumper  from TP     13 to TP     4 of J400 board  this will hold the SCR open    Switch the power safety disconnect switch    On    and then switch the  contro
8.    1 180         MS           efi  3109381      650  15760  91   55901    SNOT 1  40 03004    8  12069             SI3COW 401  190713 AVA v 804 5      1                         WNOILdO    ANO 57134       90 4 1901713    GIDN3 10S 110  9108           1331               ll                             6677 an  1 30 0        59       3NON 3NON           31925    ANON                   1031405 002      OILVWAHOS    0289 889 611         lt 429 605  lt 4  131                       SAV               1  M 1062           2 1  5319      vO TF       600              0          060          oser    22    1   1             WHO 001    Say                8 8      WSXY WSSU         9 8H      WHO 001            TWAS           ANSY       F    8H 1779 8817    713                    9154    1 8   7 2 8      NOSN3S                                                     9   860        1  15        1    9         01              81       59015534        13    v 1x SX            HOLIMS AWIL  OLH     LITT   bist             00 00 000089  OSUAN    tdl  8 05010       HOLIMS 123135         dog     20 80 01 df      Gud          24 03007  67 2         66 12 2   dr  425 JO 6       3123    0234  928          86 51 90      35314          66vf   alva   A8  NOISIA3N  393   Y               5                   320 1H31NI                  390 1831NI            398v     qauvoa  7031409 NO    000 76100    532078 45           VAr                             1    1       00        143205         40  8                iW M
9.   8 1                                          25 242 cnn ues    8 1  8 2 40 INCH SID DETENTS                         Geo rete tend 8 1  8 3 VERIFICATION OF X RAY TO LIGHT FIELD                                                          8 1  8 4 VERIFICATION OF X RAY TO IMAGE RECEPTOR ALIGNMENT                             8 1  8 5 SECURING TABLE TO ElOOR      eot tee    utn tette 8 1   9 0 OPTIONS FOR FIELD INSTALLATION                     ecce 9 1          S                                    9 1  9 2 TABLE TOP EXTENSION        ttt esta teet oet a            accident 9 1  93 ANIMAURESTRAINTSSS E note    Sa wl EE      MEE 9 1  9 4 LATCHING PREP FOOT                        9 2   10 0 LOCATION OF FIGURES WITHIN THE              50 10 1                     Sa siot orate ober dte ete              SII AREA          ia ih het        eM LUE 10 2    Summit Industries  Inc     Innovet Select     Installation and Service Manual       1 0 MANUAL REVISION CHART    TABLE 1    InnoVet Select    Installation and Service Manual Text Revisions  K286 T           Revision    Date       Description of Revision    Ecr No        OCT 1998    REFORMATTED AND EXPANDED  TEXT FIGURES          JUNE 1999    REVISED TEXT IN SECTION 9 1          JUNE 2000    ADDDED 4   POTENTIOMETER FOR KVP  CALIBRATION          FEB 2002    CONVERTED TO MS WORD          APRIL 2002    REVISED TUBEARM  INSTALLATION LEVELING          JULY 2002    REVISED INSTALLATION INSRUCTIONS  FOR ADDING TRIM WEIGHT TO THE TUBE   STAND COUNT
10.   Connect a jumper between  TP12 and TP15  near kVp comp pots lower left of board        Remove U14 from the J400 board  Turn on the control and adjust P12    to achieve 1 8 VDC      0 1VDO  at TP15  referenced to TP1       Adjust tap switches for 170 VAC between kVp major and minor    common  This will produce a no load potential of 110 kVp on the  transformer       Take exposures  Does the breaker trip      is mA present  If yes  the    diode sticks      the transformer are shorting and must be replaced     no  there may be a problem with the tube or cables not found with  visual inspection     D  Identification of poor SCR gating or half waving transformer    1     2     Select 50 kVp  100 mA  and 100 mAs  During exposure there should  be about 108        measured between P1 and P2 of the transformer        the voltage at P1 and P2 is 108 VAC  replace the diode sticks  If the  voltage is near 54 VAC  verify the SCR gate leads are tight  replace  the SCR    A heavy groan from the transformer during exposure  that worsens as  time and power are increased  is likely due to half waving from the  SCR     7 15    Innovet Select   Summit Industries  Inc  Installation and Service Manual       3  Open diode sticks will appear as high current draw with one half the  expected mA  when the filament voltages are normal as described in        above  and the test equipment is used properly   Non invasive  kVp meters may indicate an 8 millisecond exposure regardless of  selected time  due
11.   Single Phase        Acceptable Line Voltage    Regulation at Maximum  Line Current    Not to exceed 5      Not to exceed 5         Maximum Line Current   240 VAC input     140A   300mA    187A   400mA       Technique Factors that  Constitute the Maximum    Line Current    300 mA   125 kVp     400mA   100kVp       Control Rating      Output Current    Control Rating      Output Voltage    100 mA and 300mA are  standard  Three mA  stations of 30  100  and  300 mA are available as an  option     40 to 125kVp    Three mA stations of 100   300  and 400 mA are  available as an option     40 to 125 kVp          Control Duty Cycle         100 mA  125 kVp 4   300 mA   125 kVp 1        100 mA  9125 kVp 4   300 mA   125 kVp 1   400 mA   100 kVp 1        4 1       Innovet Select   Summit Industries  Inc  Installation and Service Manual       4 2 SPACE REQUIRMENTS    1  For STANDARD Fixed Top tables            HEIGHT       82 1 2    5        ___      2  For tables with Float Top option    HEIGHT  Each Version  82 1 2        4 2    Summit Industries  Inc     Innovet Select     Installation and Service Manual       4 3    ELECTRICAL REQUIREMENTS    TABLE 3     ELECTRICAL POWER SUPPLY REQUIREMENTS       300mA Station    400 mA Station   Optional        Equipment Category    300mA   125 kVp   Single Phase    400mA   100 kVp   Single Phase       Nominal Line Voltage    240         50   60 Hz    Single Phase    240         50   60 Hz    Single Phase       Line Voltage Range  Allowed  Alternat
12.  3 2 2    3 2 3    3 2 4    3 2 5    3 2 6    TABLE   TUBESTAND    Enclosed  integrated table  Removable 24  x 53  top   Welded base construction    Full travel  lead lined grid cabinet    103 line  8 1 aluminum grid    Grid cabinet tubestand interlock    Tubestand mounted digital kVp  mAs and status displays   Attached  counterbalanced tubestand  travels full table length   Angulation dial and operator handles    Two position foot switch                             6    6           CASSETTE HOLDER     Easy fit  removable  ABS plastic holder       Accepts 8  x 10   10  x 12  and 14  x 17  cassettes in either  direction     TUBE      e 1 0 2 0 mm focal spots  140 000 H U     CABLES    e 10  Max Flex high voltage cables  Federal terminations     COLLIMATOR    Manual with light field    Swivel mount    90 second bulb  on  time    Gradual turn off warning  Soft Start Bulb Life feature       amp                        8    InnoVet Select       Summit Industries  Inc     3 3 Options    3 3 1    3 3 2    3 3 3    TABLE OPTIONS    Exposure foot switch    Extra lamp switches    Float top 33  x 57  or Float top 33  x 67     12  table top extension  Specify right or left  fixed top only    Upgrade grid to 10 1  103 line    Animal restraining device    Third mA Station 400  large filament  or 30  small filament    Cassette Storage Bin                                   TUBE                     OPTIONS     0 6 1 5 mm  200 000 H U      Programmable lamp times   30  60  90  and 120 seconds  
13.  MT2 uri    VR2 019    PETSDTSLP 65710075  86730075   51740075        m is          C20       5 8 1  08 80051        R53          0901             sie        gt _ lt  Riot          UG          m                                      E                        TN            2         12 PGND 1 1  gum TVS3 ee      Sut 022 1 25 is m  H2 MAJ   MIN mi      o9 RO SS HSS SESS H5 AN AN           nin Hvac  SF H3                 coi TENTER  PREP EXPO 1    SSE DISPLAY   H6 s    R32 TNI 1   6 2   1    R32J 853015 012 R34 R35A C254 F2  120                 ON9 110 Viva  100                SAL ASI  vc              0289 086   11                             TL  2 1  S319NV iTo   42 NS amego                                       52 50 rmi       060       20 6   DA SO 8  NPN                 cm 20  21    PPA  ZOOPNE  OGAS  10 201214    SH3ISYNX2 0900              WHO       180  VO     6             51    T 91  06        623 2         veou              NI  uj                                     OOOorzz voum oooozz vom  DN            ezoa                                    Th          9                 EN VM VM 9                     Sen      2 58     EE Eg SE             aur             823 99A  uo 1    1210104      joues                     08                                08        8510             30 0190   8  99A    213 154           SA     43    TIOHTL  azn    DA                       08                                  08             510                                                23 23    
14.  OTHER OPTIONS      Remote exposure hand switch       Standard manual tray with sliding jaws  accepts all cassette  sizes up to 14  x 17      Page 9    InnoVet Select       Summit Industries  Inc     4 0 FUNCTIONS    4 1    Getting Started    4 1 1    4 1 2    Opening Unit   Depress the Release Foot Pedal to open the front  access door     Power Switch   To turn the generator on  move the circuit breaker  switch to the up position  The switch is located at the left side of  the hinged front door     It is a good idea to turn the unit off at the end of each day  This  ultimately will extend your system   s life  If power saving mode is  selected  the unit turns off automatically     Generator Warm up   When the generator is first turned on after  being turned off for two hours or more  the tubestand display will  indicate  COLD   After ten minutes this display will change to   READY  which will scroll across the tubestand status window   The generator will take exposures when  COLD   but the outputs  will be most accurate when the display indicates  READY   For  consistent films wait for the  READY  indicator prior to taking  exposures     Tube Warm up   It is recommended that you warm up the unit  prior to the first higher powered exposure to prolong tube life  This  is done by taking a series of three  1 10 second exposures using  the 300 mA station  Start with a kVp of 70 and increase to 80  then  90  Exposures should be spaced at least 20 seconds apart     Storing Cassette
15.  The following schedule of maintenance for the InnoVet Select is required for  safety of operation  continued ease of use  and continued long life of the  product     The maintenance program should only be performed by service personnel  authorized by Summit Industries  Frequency of the service should be 30 days  after installation and annually thereafter     7 1    7 2    X Ray Generator Maintenance    7 1 1    Verify that all technique selector knobs are mounted solidly   Verify proper operation of kVp and mAs digital displays    Verify proper operation of audio and visual x ray termination   Verify time  mA  and kVp accuracy    Examine all electrical connections of the high voltage generator     Examine oil level of high voltage transformer and refill if fluid level  is greater than 3 4  from the top of the lid     Examine the high voltage cable ends for carbon tracking     Examine inside of power supply box for any possible fluid leakage     X Ray Tube Housing    7 2 1    7 2 2    7 2 3    7 2 4    Inspect for possible oil leakage     Assure that the housing is tightly fastened to the tube mount and  collimator     Inspect stator cable for fraying or damage     Inspect high voltage cable ends for carbon tracking  Clean and  Regrease HV Cable ends annually     Page 19    InnoVet Select       Summit Industries  Inc     7 3    7 4    7 5    Collimator   7 3 1 Verify accuracy of field size    7 3 2 Verify accuracy of light field to x ray field alignment   7 3 3 Check lamp on 
16.  and 10   This will allow measurement of the    no load    primary voltage     Turn on power and adjust the kVp selector switches to achieve 220  VAC on the voltmeter  Refer to Table 8 for correct kVp display  values    Reset the kVp tap switches to achieve 160 VAC on the meter   Refer to Table 9 for the correct kVp display values     TABLE 8  kVp mA Readings  2 220 VAC                   mA Station kVp display on tubestand  30  if included  116  100 114  300 87  400  if included  70             7 2    Innovet Select     Summit Industries  Inc  Installation and Service Manual       7 2    TABLE 9   kVp mA Readings   160                                   mA Station kVp display on tubestand  30  if included  82  100 80  300 53  400  if included       If the kVp meter varies by more than 5 kVp to the above values  proceed to the  kVp calibration procedure 7 4 before adjusting mA     4  Switch the disconnect switch to  Off and reconnect the leads   1  and P2 to      1 on the control  Be sure to remove the jumper  between TP   13 and TP     4 on      J400 PCB     mA CALIBRATION    Achieving maximum accuracy of tube current  mA  involves two types of  adjustments     e Overall mA level  bands of filament resistor RX   and  e mA balance throughout the useful kVp range  bands of space charge  compensating resistor RSCC      The leads connected to the resistor bands of resistor RX  the third resistor on the  right of the three at the right corner of the Control Board  are marked with l
17.  table  To attach the extension align notches in  extension with buttons mounted on end of table support and slide over buttons as  shown in Figure 19     TABLE TOP  EXTENSION    NOTCHES    BUTTONS    TABLE TOP SUPPORT       Figure 19  Table Top Extension   AWW430 08  ANIMAL RESTRAINTS    The animal restraints used are nautical style tie down cleats  They are screwed  to the underside of the table top frame at either end of the table  This prevents    lead aprons from snagging on the cleats     9 1    Summit Industries  Inc     Innovet Select   Installation and Service Manual       9 4    FOOT TREADLE    If the  nnoVet Select      has the latching prep Foot Treadle Option  the treadle  must be installed prior to mounting the control into the table  If the Foot Treadle  Option is being added to an existing system  the control will need to be removed  from the table base and laid flat on the floor to install the treadle     1     10     Remove the control from the table base by unclipping the truarc ring and  extracting the hinge pin which hold the control to the air cylinder  Lay the  control face down on cardboard and pull it slightly away from the table     Put the treadle in position across the front of the table base     Thread the spring guide pins located at either end of the treadle up  through the table base until they protrude out the top of the base  securing  each guide pin in place with cotter pin provided     Bolt the foot treadle brackets to the underside of the 
18.  the brackets until the  vertical slide has been freed to move  The cables will route directly into the  right rear of the table base  not beneath the lower tubestand rail     WARNING    Particular care should be taken when servicing the inside of the  tubestand  There is an extreme threat of mechanical pinching  between the vertical slide and counterweight due to their close  proximity and opposite directions of motion  All of the movable  assemblies and parts of x ray equipment should be operated with  care  Only properly trained and qualified personnel should be  permitted access to any internal parts     Remove the shipping bolts which lock the main counterweight and vertical  carriage together  taking care to insure that the counterbalance is adequate to  avoid uncontrolled motion of the vertical slide  The trim weights provided in  the hardware kit should be used to achieve proper counterbalancing of the  tubestand  Remove the back panel of the tubestand  and then insert trim  weights as needed into the channels at the top of the main counterweight     5 8    Innovet Select     Summit Industries  Inc  Installation and Service Manual       5 5    5 6    FOOT TREADLE INSTALLATION  skip ahead to 5 6 if using standard  foot switch     If the  nnoVet Select   has      latching prep Foot Treadle Option  the treadle  must be installed prior to mounting the control into the table   If the Foot  Treadle Option is being added to an installed system as a field upgrade  refer to  S
19.  to their resetting during the missing pulse     IIl  CIRCUIT BREAKER TRIPS WITH ROTATION OF KVP TAP SWITCH    A  Look for a broken connection or poor solder joint on the noise suppression  assembly between the common poles of the kVp major and minor tap  switches  Replace if suspect  as this problem can damage rectifier sticks in  the high voltage transformer     IV  FIL 1  ERROR  low filament current in standby idle mode     A  Open connection at H8 or H9 of J400 board  1  This could be a pin to plug or plug to board connection     B  Open loose terminal at either end of filament cable  XC  XL  XS   C  Open F1 or F2 fuse in control  D  Open between resistor and resistor band  RSCC or RX     E  Open or loose cathode cable  open filament of x ray tube   1  Remove the cathode cable from the tube end  Use a collimator bulb  as a dummy filament load by connecting it between XC XS or XC XL  pins and see if it will light     use standby  idle  mode only       the  collimator bulb lights the tube filament may be open      not  cathode  cable may be open    2  Switch cables anode for cathode  If collimator bulb now lights replace  the cable formerly in the cathode side     F  Check the J400 board  1  Once it is confirmed there are no open primary connections or open  secondary loads  check for about 1 7 VDC      0 2 V  between     12  and TP1  If the voltage is missing  confirm LED4 is illuminated  If the  light is OFF  jumper pin3 to pin4 of U18 optocoupler  Replace U18  optocouple
20.  will produce about 2 68 VDC at TP12    2  Faulty or misused test equipment may give an improperly low reading  for mA  causing an over adjustment of the filament circuit and a    FIL4     error  Meters must be set to AC mA  and if equipped with an  adjustable range scale  20  200  2000  it must be set to maximum    3  Functional test equipment does not see the mA produced    a  Always remove the jumper between the two    AC mA    terminal  locations when measuring mA  and always replace the jumper  when finished    b  Ensure spark gap on top of the HV transformer at    M1    is not  shorted to ground    4  Failure of J400 board  This is by far the least likely possibility     7 17    Innovet Select   Summit Industries  Inc  Installation and Service Manual       VIII  DIGITAL DISPLAY PROBLEMS    A  Status display does not change or is at wrong state  1  On J400 board set DIP2 4 to OFF  down   disabling  ready cold   feature  If the display now updates  the    ready cold    feature is not  available with the display s software version  Update the software or  leave switch OFF   2  If locked in  wait  check for shorted footswitch     B  Scrambled characters missing segments    1  Verify that each of the conductors on the cable going into H6 of J400  board is secured to the plug by looking beneath the plug cover     2  Confirm the Mate  amp  Lock plug pins of the display cable at the rear of  the control are seated and making contact     3  Confirm the plug and wires are well se
21.  x ray control  Wire Size from Power for 50 feet use For 50 feet use  Transformer to  2 AWG   2 AWG   Disconnect Switch for 100 feet use for 100 feet use  800AWG  800AWG   for 200 feet use for 200 feet use  250 mcm 250 mcm       The information provided in Table 3 above and text below is taken from NEMA    standards  Minimum Power Supply Requirements for x ray Machines  and National  Electric Code     Connection to Supply Circuit  taken from N F P A  70 1984    A disconnecting means of adequate capacity for at least 5096 of the input required for  the momentary rating shall be provided in the supply circuit  The disconnecting means  shall be operable from a location readily accessible from the x ray control  Underwriters  Laboratories also requires that this disconnecting means to be mounted on either the  wall behind the x ray control or the wall directly adjacent to it     4 4       Summit Industries  Inc     Innovet Select   Installation and Service Manual       5 0  5 1    5 2    PRELIMINARY MECHANICAL ASSEMBLY    1     1     RADIOGRAPHIC TABLE    During shipment  the table top is banded to the table base  Cut loose the  banding and set the table top aside for assembly later  Loosen the hold down  screws underneath the front rail of the table to release the grid cabinet   Remove packing material from rear of grid cabinet     Install the five leveling feet by threading them into the holes at the two front  corners of the table and at the ends and middle of the back rail  T
22. 00   48 100  6 10  9 10  1 2 10  1 1 2    60 Hz  1 120  1 60  1 40  1 30  1 24  1 20  1 15  1 12     1 10  1 8  3 20  1 5  1 4  2 5  1 2  5 8  3 4  1             Making an X Ray Exposure     Note  If it is desired  to view the total number of X ray exposures  simply  turn the KVp major  KVp minor  amp  mA selector knobs fully counter   clockwise  the bottom display will indicate  EXPO   The KVp display and  the mAs display will indicate the total number of exposures taken on this  x ray control     4 5 1    4 5 2    Making Exposures with the Standard Foot Switch    There are two ways to make a x ray exposure with the 2 position  footswitch     Prep and Expose   When proper patient position is achieved  partially depress the  footswitch to activate the x ray tube prep cycle  When    PREP    is  displayed on the Tubestand  watch the patient for respiration  phase or proper restraint  At the desired moment  make the x ray  exposure by fully depressing the footswitch  An audible tone will  indicate an exposure has been made     Expose   When proper patient positioning is achieved  fully depress the  exposure footswitch  The x ray tube will begin its prep cycle and  upon completion of the prep cycle  the x ray tube will make an  exposure  An audible tone will indicate an exposure has been  made  Note that in this procedure an x ray exposure will not be  made immediately upon depressing the footswitch  A delay will be  experienced due to the x ray tube prep cycle     Making Exposur
23. 3Q0V3H    ONINLL      LH Ed VUL              yw              ON              HOLIMS         7102 QvOTH3AO  VW           val 70  Alddns         OL          vc  val    12          009    Z    H Us    000 06010 2 1      00         lt                    5  OVA vl     482                 5               9v  092             000 56220    YOUMS         p  s       109                   aS  VN               001  7102     5  35    S9ev  Joyeaig                330   NO                        IIO   lt  lt         o                        uu                      t                                5                                   54141510   LINNAS    MOSN3S                         6   005    30151405                                     431534 NOS    M3W303SNVUI 357                   AVIdSIG OL 9  90    110                                           1 50706711 91 vZUW1                          2070   509     1       0 11701 91            aaaav 410                             0780701           8 SVM 5    0  504 S 240 168          70 0  7        2                             H                  30151538 15008 30               AN3AVII3 01                 0054    4  90               PPA           620         vvin    904      1   to    siz         Lou                      620                106254                             ACIL 3VISP               6675 90           5  0n  PPA PPA    T           120           1  lt 06054  N                     0   234 014 01    110 22012    x00        893   810  
24. 4 2  Fig  2 Floor Space for Float Top                                                     4 2  Fig  3 Tubestand Stop at Lower Left           5 3  Fig  4 T  bestand                 eii re Eutr             5 4  Fig  5 Mounting Tubestand               5 4  Fig  6 Table Top Frame Screws  amp  Lock Release Tabs                   5 5  Fig  7 Elevated Frame  amp  Transverse Bearing Track                       5 6  Fig  9 Air Cylinder                            5 10  Fig  10 Mounting the Foot                          tiae enne 5 11  Fig  11 Removing the Front Cover                              eese 6 1  Fig  12 Interior of Control oin      ae aie          diens 6 2  Fig  13 Rear of Control Showing Cable Entry Points                         6 2  Fig  14 Table Interface Junction                     4  4  04  214  10 2        6 3  Fig  15        and     4             eise bote beri    d 2240 54214104 6 5  Fig  16 mA Tracking                                           7 4  Fig  17 Control Board Component Identification                                7 6  Fig  18       Switch Settings   oer aen                                   7 11  Fig  19 Table Top  Extensions  up opa                9 2  Fig  20 System Cable                                                                       10 2  Fig  21 High Voltage Transformer Diagram                                     10 3  K158 Schematic  InnoVetSelect Table   J499 Schematic  K200 Control   J402 Schematic  Control Board  J400    J352 Schematic  
25. AL CUTOUT YELLOW    SWITCH    ORANGE   BROWN         6    THERMAL CUTOUT  SWITCH   RUN WINDING OF  TUBE STATOR    BROWN   ORANGE    BLUE       START WINDING OF  TUBE STATOR          COM  LEAD OF TUBE  STATOR                If the system is using the standard two position Foot Switch  route the  cable  labeled S1  S2 and S3  through one of the openings in the control  and connect to TB   4 terminals 27  28 and 29 respectively  which are  labeled the same as the Foot Switch cable   Refer to Figure 15 below for  reference      6 4    Innovet Select                                                                          Summit Industries  Inc  Installation and Service Manual  AC Common   FF4B 3    23  CSW  lt  Collimator Light Switch        4   2 24       1    2   24            E 5  GND4  25     F 6 26   TEUNAN    F2B 7 51 27   Two Position   lt    2   8 52 28   Foot Switch         9 83 29  Tap Switches           10 T5 30    L MIN 1116 31  Interlocks         12 17 32  13  07  33       GND 14 08 34  gt  Rotor                Ground Reference 4 GND1 15   09 35                 18 XL 38  gt HVT  Filament Common                                Measurement      19 M 39          5                           20         5 40  HVT   Large Filament                      Figure 15   TB3 and     4 Terminal Designations    6  If the latching prep Foot Treadle option has been installed  push the air  hoses from the foot pads on the air cylinder nozzles protruding from the  rear of the control  Plu
26. Display Board  J350    01032 Schematic  RC Timer Board  01030 000     10 1            Innovet Select    Installation and Service Manual    Summit Industries  Inc     S318VO  H    OT  Z     NMOHS SHLONAT 3 S3ZIS 318V2             15  INVHOVIG 1                1  1               371919 WALSAS    HOLIMS  1004             8 X QNOD       V9 88           02      X ANOD S  V9 818  2801       X ANOD S  V9 8189       1ST X ANOD      V9 8          X ANOD        V9 818                                  37971    HOLIMS    193NNOOSIQ                   AV1dSI  d SNLVLS    AV1dsid         AW1dSId 4A     Figure 20  System Cable Routing   AWW430 09    10 2            Innovet Select    Installation and Service Manual    Summit Industries  Inc                        YAWYOASNVEYL JADVLIOA HOIH OOLV           YAWYOASNVYL                           H9IH           WIN 4      11VWS   IN3INNV I      WIN dX   39uv1   53314149334                    3    21115       S IVNIWWU3 L                                     14424     1    14423      1            AGONV   184          AGOHLVD    Figure 21         Voltage Transformer Diagram     AWW430 10    10 3           8 4 8813  gt  1841                   0299 989   1 XVJ    429 809   LL TAL  71                                            1062    CIWNOILd0 gt   HILIMS          dWVv            O  1431    LHOTY             57300    dol 190713    PAS 02071    AS AIO     10aLNO9          1331  Imm oO O          o   to        20        20  po  so  t  60 cg              
27. ERBALANCE          OCT 2002    ADDED RC TIMER BOARD  EXPOSURE  COUNTER INFORMATION          OCT 2004    EXPANDED TROUBLESHOOTING GUIDE          JUL 2006    ADDED 50 HZ  REFERENCE AND TIME  STATIONS                                                       Innovet Select   Summit Industries  Inc  Installation and Service Manual       2 0 UNPACKING    The InnoVet Select   Radiographic System is delivered in three packages   e High voltage transformer  control  x ray tube  palletized    e Table  collimator and accessories        Tubestand     It is the installer   s responsibility to inspect the shipment for damage and proper  count  Upon receipt of the merchandise  any visible damage to the cartons  should immediately be examined while the shipper is present  If the visible  damage to the cartons also includes damage to the merchandise  the installer is  responsible for making all claims with the shipping company     If there is hidden damage to the merchandise  it is the installer   s responsibility to    discover that damage within a reasonable amount of time and contact the  shipping company     2 1    Innovet Select     Summit Industries  Inc  Installation and Service Manual       3 0 RADIATION AND MECHANICAL ELECTRICAL WARNING    3 1    RADIATION  WARNING    X rays are dangerous to both operator and others in the vicinity unless  established safe exposure procedures are strictly observed     The useful and scattered beams can produce serious  genetic or potentially fatal  bodi
28. InnoVet    Select    Operator s Manual  K287 Rev E    SUMMIT INDUSTRIES INC   2901 W  Lawrence Avenue  Chicago  IL 60625  1 800 972 9776       InnoVet Select     Summit Industries  Inc     TABLE OF CONTENTS    1 0 INTRODUGTIQON                              1  2 0 PRECAUTIONS      2                           CLERI C RICE RM          2  272               OM ERR 2  3 0 GENERAL PRODUCT OVERVIEW               55655u ccs ako        Grana ES nere p nnmnnn nnmnnn 4  2 1 cMalor Components              a teria               coU UE 4  3 22                  8   eL EHE 7  9 9   ODUOINS             ooo an                                  Sin DEAS 9  4 0  FUNCTIONS    oneseceenence ceca sns exe eue  accen exo aq Ea Eia yn ex Ea  Yen aS Fu CU n exin ce exin                     10                                                          10  4 2 Positioning the                                                                                                       11  4 3  Collltnalor Open alone                aA fcm eccl 11  AA Seting                     uiae a                           12     5                             sri per fep repo                                    13  4 6 Prep Expose Tubestand Status                                                                           15  47                           ERR 15  5 0 TECHNICAL DATA                                                    16  5 1 Radiation Leakage  Meets Federal Government Standards                                16  5 2 Rate
29. V grooves of the pivot  rollers at the base of the table     3  Engage the notches on the bottom of the control with the pivot rollers and lift  the control so that it hinges up into the table base     4  Connect the air cylinder to the control using the pin and locking ring as shown  in Figure  9   Note that there is no other device holding the control    within the table  If the pin is ever to be removed from the air cylinder   exercise caution to maintain balance of the control within the table     RETAINING RING    CONTROL DOOR    FRONT LINKAGE    AIR CYLINDER CLEVIS PIN       Figure 9  Air Cylinder Attachment   AWW430 06    5 10    Innovet Select   Summit Industries  Inc  Installation and Service Manual       5  Bolt the foot loop to the bottom of the control as shown in Figure    10        FOOT TREADLE       Figure 10  Mounting the Foot Loop   AWW430 07    6  Center the control within the table opening by adjusting the two front feet  If  the left side of the control is too close to the left pedestal leg  raise the left  front corner of the table slightly  If the right side of the control is too close to  the right pedestal leg raise the right front corner slightly  Use the red or blue  shims as required to achieve a snug fit between the top of the leveling feet  and the bottom of the table for maximum stability     Innovet Select   Summit Industries  Inc  Installation and Service Manual       6 0 ELECTRICAL CONNECTIONS  6 1 LOCATION OF ELECTRICAL CONNECTIONS    1  Remo
30. aight  Replace the back cover     If the light field is out of alignment with the grid cabinet  call an  authorized service representative for re alignment     Page 22    InnoVet Select       Summit Industries  Inc        8 4    8 5    8 6    8 7    No digital display    If  after turning on the circuit breaker  digital displays do not light up  this  indicates lost power  The source may be a blown fuse within the unit or  a problem with incoming line power  Call an authorized service  representative     Circuit tripping    If the circuit breaker shuts off on its own  this indicates mA overload   This could be due to miscalibration of the mA or another malfunction     In the event that the unit is miscalibrated  you may still be able to  operate the unit by moving to 100 mA  This will allow you to take X   Rays until an authorized service representative is called     Full kVp range inaccessible     If this occurs  it means that incoming line voltage is too low  Call an  authorized service representative to correct     Poor film quality    There are many interrelated factors in the radiographic process which  impact the degree of darkening visible on a film  The following is a list of  items to check whenever you notice a significant change in film quality     8 7 1 Film Processing   Has chemistry been changed recently  Is a new  brand type being used  Have developing temperatures varied  All  of these affect film quality     8 7 2 Technique   Were proper technique factors use
31. ated at J350 board in tubestand     C  Blank display  1  Verify the Display cable connections at the H6 header of the J400  board  the pins in the Mate  amp  Lock plug at the back of the control  and  at the J350 display board   2  Confirm  20 VDC at          12VDC at TP5 of J400     D  kVp changes without tap switch adjustment  1  Select 100 mA and 143 VAC between the leads of MOV1 on the J400  board  located near H2  70 kVp should be displayed  If the voltage  varies check H2 and kVp tap switch connections   2  Confirm a stable 1 VAC between the junction of R35A R36 and TP1  If  this voltage is unstable replace the J400 board  If the voltage is stable  and the kVp is still variable replace the J350 board     E   HEAT  Error  1  Verify TB4  T5 T6  tube thermal switch connections are closed   2  Verify zero ohms on J400 board between U22 pin 1 and pin 2   3  Replace U22 or J400 board            TUBE    Error  1  Select 70 kVp and 10 mAs  The  Tube  Error should go away   2  If 70 kVp and 10 mAs cannot be selected  the J400 board may not be  getting a valid kVp  time or mA input  or the board may be bad   a  Adjust kV major minor tap switches to achieve 180 VAC  between the two leads of MOV1 on the J400 board  If no    7 18    Innovet Select   Summit Industries  Inc  Installation and Service Manual       voltage is present inspect H2 and kVp tap switch common poles  for open connection    b  Remove the H10 plug from the J400 board  Place a jumper  between pins 1 and 3 on the b
32. ct     Summit Industries  Inc  Installation and Service Manual       7 2    mA Calibration  continued     In the following procedures  use the SM  MD or LG bands as appropriate for the  mA stations available per this unit s configuration according to Table 10 above   Calibration is typically started with the 100mA station     1  Connect    Dynalyzer or      mAs meter to perform the following  procedure  An AC mAs meter can be inserted into the circuit by removing  the jumper between TB3 19             and TB3 20  MA   and connecting the  meter to those terminals  AC readings only   Be certain to replace the  jumper after removing meter     2  Select 100 mA and 80 kVp    3  Adjust the appropriate band of filament resistor RX for the mA station  chosen to produce the desired mA  Moving the band upward increases  mA    4  Select 50 kVp and note the mA produced    5  Select 120 kVp and note the mA produced    6  If the mA at 120 kVp is higher than the mA at 50 kVp  move the band of  RSCC appropriate for the selected mA station upward  If lower  move the  band downward  Achieve relative balance per Figure 16 above  The unit  is capable  by fine tuning  to track within  5  of selected value    7  Repeat steps  3  through  6  until no further adjustments are necessary     8  Repeats steps 2 through 7 above for the one or two remaining mA options  available on your station     9  Replace jumper after removing the AC mAs meter     7 5    Innovet Select   Summit Industries  Inc  Instal
33. d  Consult with a    Summit dealer or call Summit for information on setting up a  technique chart     Page 23    InnoVet Select       Summit Industries  Inc        8 7 3    8 7 4    8 7 5    Screens Cassettes   Screens within the cassettes degrade with  time  They typically lose 10  of their light emitting capability per  year  It is recommended that cassettes be re screened every five  to seven years  All screens should be replaced at one time to  maintain uniform exposure     Film   Has a new brand type speed of film been used  Is the film  beyond its expiration date  If so  new film may be purchased or  technique factors may need adjustment     Foot switch   If thorax chest films are blurring more than usual  it    could be that the exposure switch is being used incorrectly  See  section 4 5 to review how to make exposures     Page 24    InnoVet Select       Summit Industries  Inc        9 0    WARRANTY NOTICE    The InnoVet Select is warranted as follows  any Summit manufactured part  proving defective will be repaired or replaced free of charge  F O B  factory  if  the defective part is returned for inspection  freight charges prepaid  This  warranty covers parts only and does not include any on sight labor costs     This warranty is five years from the date of installation     This warranty does not apply to collimator glassware or x ray tubes or damage  caused by accident  misuse  neglect  or shipment  and is void if service is  performed by persons other than authori
34. d line voltage  240         50   60 Hz   single phase                                       16  5 3 Range of line voltage regulation at max line current  1  to 596                           16  5 4 Maximum line current  based on input voltage of 240 VAC   140 AMP                16  5 5 Technique factors that constitute the max  line current  300 mA  9 125 kVp       16  5 6 Maximum Deviation of technique                                       16  6 0 CLEANING THE  nnoVet 5                                                                                                             18  7 0 SCHEDULED DEALER                                                                     19  7 1 X Ray Generator                            ac                   ca dap                               19  7 2 X Ray Tube Housing                                                19         leerlo                                    nenene 20   E dX                  m eem 20  355  OUDBE cei tot rep tetanic sain          Evae edu                                                  20    InnoVet Select     Summit Industries  Inc     8 0                                                                     aya e aoo aero cc aka                                     21  9 1              88406  5 Dm 21  Ba NO EXDOSUEO    C E                  UL eu 21  8 3 Collimator light bulb burned                           21  55220110 GIGI Al display e                                   23  8 55                                        
35. e  the SID markers indicate  40  positions from the table top or the grid cabinet  During vertical  motion  you will note a pause at one height  which is the pre set  height for optimal exposure  set by adjusting the detents during  installation     Float Top  option    Push in float top unlock button and move the  table  Release the button to lock the top position     Angulation   Release the lock  which is located on the right side of  the tube arm  Grasp the handle bars and turn the tube to proper  angle  Turn the lock handle clockwise to hold tube position  The  degree of rotation is indicated by the dial located between the  handle bars     Collimator Operation    4 3 1    Light Field   Once the tube is positioned relative to the grid  cabinet  press the light button on the front of the collimator or one  of the optional         hands  collimator switches on the front or ends  of the table  A light field will appear  WARNING  DO NOT LOOK  DIRECTLY INTO LIGHT SOURCE AS RETINA DAMAGE COULD  RESULT     Page 11    InnoVet Select       Summit Industries  Inc     4 4    4 3 2    4 3 3    4 3 4    Programmable Lamp Times   Lamp gradually turns off in one of  the following selectable times  30  60  90  120 seconds  The  factory setting is 90 seconds  If you push the button during the  lamp   s power down ramp  a new    on    cycle starts     Collimator Adjustment   Use the sliding knobs to move the  shutters  Scales indicate knob positions that correspond to  common cassette siz
36. e  tubestand  There is an extreme threat of mechanical pinching  between the vertical slide and counterweight due to their close  proximity and opposite directions of motion     Page 3    InnoVet Select     Summit Industries  Inc     3 0 GENERAL PRODUCT OVERVIEW  3 1 Major Components    2 S    2 X           2    gt         OOO      gt   e            q r s   on tilt out     p D ee                       4    InnoVet Select     Summit Industries  Inc     a  Tubestand   Supports collimator and x ray tube  Moves both horizontally and  vertically     b  X Ray Tube   c  Angulation Dial   Displays degree of x ray tube angulation    d  Angulation Lock   When loosened  allows angulation of x ray tube    e  Handle Bars   Used to move tube vertically or to angle the tube    f  Light Switch Button   Turns collimator bulb on and off    g  Source Image Distance  SID  Indicator   When marker aligns with label on  tubestand  SID  the distance between the tube focal spot and the film  is 40      There is a position for tabletop work and one for grid cabinet applications     h  Grid Cabinet   Lead lined cabinet holds cassette tray and grid     i  Grid  mounted to upper carriage of grid cabinet  absorbs scatter radiation for  improved film clarity     j  Cassette Tray   Molded indentions hold cassettes in correct position    k  Cassette Box  optional    Opens to store cassettes  Lead lined to protect film      Rustproof Support Feet   Adjust to level the table    m  Exposure Foot Switch   Used t
37. e installed before the control  is put in place  lt is also recommended that the transformer be    installed before the tubestand is mounted to the table for maximum  stability during installation     5 3 TUBESTAND INSTALLATION    The sequence of assembly is slightly different between Fixed Top and Float Top  tables     For Fixed Top  standard  tables proceed directly to section 5 3 1 below   For Float Top  optional  tables the table top frame must first be removed before the    tubestand can be mounted  Skip ahead to section 5 3 2 for instructions for this type  of table     5 2    Innovet Select   Summit Industries  Inc  Installation and Service Manual       5 3 1 For Standard Fixed Top Table    1  Remove tubestand stop from the left end of the table   s lower rear rail  See figure 3        Figure 3  Tubestand stop at lower left rail    5 3    Summit Industries  Inc        Figure 4     Tubestand Positioning    Innovet Select   Installation and Service Manual    Position the tubestand so that it is a few  inches from the end of the back rail of  the table with the lower tubestand  bearings in line with the back rail  See  Figure 4      The tubestand shown in Figures 4 and 5 is for a Float Top table  but mounting the  tubestand to the rail is done the same for either configuration         Figure 5     Mounting Tubestand on Rail    Carefully tilt the tubestand away from  the table until the elevated tubestand  bearing can be pushed into the lower  bearing rail  Now straighten 
38. e line    voltages which will provide  normal operation      194 284 VAC     194 284 VAC        Maximum Momentary    140 amperes    187 amperes       Line Current  at 240 VAC   at 240 VAC   Note  Maximum l2   140 x 240 lo   187 x 240  momentary line current at V2 V2    alternate line voltages can  be determined using the  formula     V2   alternate line voltage  maximum line current  at the alternate line  voltage     V2        l2      V2   alternate line voltage    l    maximum line current  at the alternate line voltage                        Line Voltage Regulation  under load       The line voltage drop under  load is not to exceed 5  at  maximum line current        The line voltage drop under  load is not to exceed 5  at  maximum line current        4 3       Innovet Select              Summit Industries  Inc  Installation and Service Manual  Calculating line voltage  VNL           100                  100  regulation VL VL          line voltage under         line voltage under  no   no load  conditions load  conditions       line voltage under       line voltage under    full     full load  conditions load  conditions  Minimum Over Current 5096 of maximum line 5096 of maximum line current  Protection Rating current rating or greater  rating or greater   100 Amps   Service disconnect   100 Amps is is recommended      recommended         Distribution Transformer   Minimum 30 kVA dedicated   Minimum 37 5 kVA dedicated                Requirements to the x ray control to the
39. e red and black wires from  air switches at TB5 1 and TB5 4 and recheck measurement to  determine if the short is in the air or strip switches    2  If a two position foot switch is installed  confirm open contacts between  51  S2 and 53 on TB3    3  Remove H5 at J400 board  If problem continues replace J400     7 19    Innovet Select     Summit Industries  Inc  Installation and Service Manual                   DISPLAY IS INCORRECT FOR SETTINGS    A  Verify mA switch inputs  1  Confirm proper number of switch positions  two and three mA station  units are available   If three positions are available on a two mA  station unit  rotate switch fully CCW and remove the switch  With shaft  pointing upwards  move the stop tab to the fully CCW position   2  Remove H11 plug on J400 to force in a good low mA station selection   Jumper pin 1 to pin 2      J400 board H11 for medium mA station   Jumper pin 1 to pin 3 for high mA station  if third station exists   If mAs  display responds to change of mA station replace mA switch  if not   replace J400 board     B  Verify time selector switch inputs  1  Select 100 mA  Remove H10 plug from J400 board  Without H10  connected mAs display should be 0 83  Jumper pin 1 to pin 3 on the  board  and display should change to 8 3  If it does  change time  selector switch  If not  replace J400 board     XII  NO mA MEASURED WITH RADIATION PRESENT    A  Spark gap is shorted to top of transformer     1  The metal tab should not make contact with transfor
40. ection 9  Options for Field Installation       1  Put the treadle in position across the front of the table base     2  Thread the spring guide pins located at either end of the treadle up through  the table base and lock each guide pin in place with the cotter pin provided     3  Bolt the foot treadle bracket to the table base using the hex head bolts  provided   4  Plug the foot treadle cable into the S7 location of the junction box at the right    side of the table frame  Refer to Figure 14 as needed     FILM BIN INSTALLATION     skip ahead to 5 7 if no film bin is to be installed     The film bin must be installed prior to mounting the control into the table base  The  film bin comes in either a right hand or left hand configuration  Slide the PEM studs  of the film bin through the holes in the front edge of the appropriate side of the table  and secure it using the hardware provided  Bolt the back side of the film bin to the  rear of the table using the bolts provided  Install the plastic organizer in the top  drawer of the film bin  Note that for left hand film bins the control serial tag may be  partially obscured  Be certain to record the control s serial number in the user and  installation manual if this is the case     5 9    Innovet Select   Summit Industries  Inc  Installation and Service Manual       5 7 GENERATOR INSTALLATION    1  Lay the control face down on cardboard  centered in front of the table     2  Adjust the table legs to measure 34 5 8  between the 
41. es  Use the pointer to quickly move shutters  into relative position  then fine tune to collimate the x ray beam to  the area of radiographic interest     Swivel Movement   When doing table top work  you may want to  swivel the collimator to avoid awkward patient positioning  Simply  swivel the collimator to the preferred position  WARNING  Do not  touch the black lamp cover at the rear of the collimator as it gets  quite hot  The collimator hesitates at the 0   and 90   positions for  correct re alignment  This is in order to help find these position  accurately     Setting Techniques    4 4 1    4 4 2    mA   Adjust mA by moving the pointer to one of two  or three   stations  1005 or 300L  305  400L      When set at 100 mA  the small tube focal spot  1 0 mm in standard  tube  is used  yielding a more finely detailed image  At 300 mA   the large focal spot  2 0 mm in standard tube  is used     kVp   Two knobs control kVp   one for coarse adjustments of 15  kVp increments and one for fine tuning with 2 kVp adjustments     Turn the knobs to select the desired kVp     The kVp value is displayed on the tubestand mounted display     Page 12    InnoVet Select       Summit Industries  Inc     4 5    4 4 3    Time   The time dial can be set at any of 18 stations  these are  marked both in fractions of seconds and in natural numbers        Line Frequency   Time Stations Available       50 Hz  1 100  2 100  3 100  4 100  5 100  6 100  8 100   1 10  12 100  18 100  24 100  30 100  36 1
42. es with the Optional Foot Treadle    First depress the Exposure Foot Treadle until the display reads     WAIT     This starts the tube prep cycle  After 1 to 2 seconds the  tubestand display will read    PREP     The tube is now ready for an  exposure  The exposure will start when you depress the Exposure  Foot Treadle  The generator will instantaneously activate and  sound an audible    beep    to indicate that an exposure was made     Page 13       InnoVet Select       Summit Industries  Inc        If the Exposure Foot Treadle is initially depressed and then held   the unit will automatically go through its two second prep cycle and  then expose  Hence  you will note a slight pause between  pressing the foot switch and hearing the exposure signal     Page 14    InnoVet Select       Summit Industries  Inc     4 6    4 7    Prep Expose Tubestand Status Messages    WAIT Tube is not yet warmed up  Please wait about 2 seconds   Exposures are not allowed yet     PREP Tube is ready  Exposure starts immediately after initiating  with foot switch     X RAY X Ray exposure is in progress  Radiation is present   DONE Exposure is completed   COLD Unit needs to warm up for about ten minutes to achieve the    most accurate exposure outputs   READY Unit is warmed up and ready for taking exposures   EXPOSURE COUNT  With the kVp major  kVp minor and mA selector knobs  turned fully counter clockwise  the bottom display will  indicate  EXPO   The KVp display and the mAs display will  indicate 
43. etters  which correspond to the mA stations  small SM  medium MD  large LG   Moving  one of these bands upwards increases the mA for the corresponding mA station     The leads connected to the bands of RSCC are also marked with SM  MD  LG  which correspond to the mA stations  Moving one of these bands upwards  increases the space charge compensation  which increases mA at low kVp  and  reduces mA at high kVp  Moving one of these bands down reduces the space  charge compensation  decreasing mA at low kVp and increasing mA at high kVp     Normal mA tracking will result in values slightly high at 80 kVp and mA values  roughly equal to one another but slightly low at the high and low ends of the kVp  range  125 and 50   See Figure 16 below for a visual representation of this  effect      2  terminals  of TB3  are marked  mA   term  19  amp  20   Remove the factory    installed jumper from these 2 terminals  amp  connect an          meter to these  terminals to measure tube current     7 3    Innovet Select     Summit Industries  Inc  Installation and Service Manual            596    5        SELECTED mA          Figure 16   mA Tracking Example    For the following steps  use mA values and band designations in the table below     Table 10     mA Stations Band Labels                            mA Station 100 300 30 100 300 100 300 400  Configuration Configuration Configuration  30mA    SM      100     5        5    300mA MD LG MD  400mA         LG             7 4       Innovet Sele
44. ff Off   Off  4     gt   DIP1 DIP2  Power  Tube Select Beep Saver  Tube  1 2 Duration   6 7 Delay 213  E239         0 2Sec    E 1Hor           n n  E742       0 5Sec       2          M           T    108         4Hous     m               20   8    Shu  8                                                       Figure 18   Dip Switch Settings for Configuring Options and Features    Innovet Select     Summit Industries  Inc  Installation and Service Manual       7 7    X RAY TERMINATION              VOLUME CONTROL   See Figure 17     Electrical Control Board Component Identification     The volume of the x ray termination beeper can be adjusted by P7  located on  the control board  Turning P7 counter clockwise increases the volume from quiet  to very loud  Note  The front cover of the control will quiet the sound  considerably so test the sound with the cover in place     7 8 TUBESTAND DISPLAY MESSAGES  The following Messages are displayed by the bottom display on the InnoVet  Select    tubestand  The COOL and READY features can be disabled via DIP2 4   see FIG 18     WAIT PREP has been requested but the tube is not warmed up yet  The  message clears as soon as the tube is ready for an exposure  This  typically takes between one and two seconds    PREP The tube is warm and now ready for an exposure    XRAY Exposure is occurring  lonizing radiation is present    GOOD JOB Exposure completed successfully  The message is cleared when  the exposure switch is released    TUBE Indicates 
45. for the mA station selected    For 100 mA offset adjust P2   until the        meter reads 50 kVp    Readjust the kVp tap switches to achieve the voltage which should result  in 120 kVp per Table 12   For example  for the 100 mA station adjust the  tap switches to achieve 230 VAC     Adjust the appropriate slope pot until the        meter reads 120 kVp   Repeat steps 2 through 5 until no further adjustments are required    Repeat these steps  adjusting tap switch voltages  then offset and slope    pots  until the kVp display indicates the appropriate value across the  operating range for each mA station     7 10    Innovet Select   Summit Industries  Inc  Installation and Service Manual       7 6 CONTROL BOARD DIP SWITCH SETTINGS    The InnoVet Select    has a variety of dip switch settings to enable or disable  various functions and to configure the system for the type of x ray tube used in  the system  These switches are typically set at the factory  but can be changed  to accommodate the desires of the operator or the needs of the application  The  dip switch combinations are shown in Figure 18 below                       DIP SWITCH SETTINGS   nuammuuE  unum  1 2 3 4 5 6 7 8 1 2 3 4 5  50Hz 23 Latch 30 mA On On    Extra Power             Cold    Simu                           Ben      Power   Saver   Saver   Ready lator                                                                mA  Time    Saver   Delay  Delay           Stations Station            60Hz 18 2Pos  400 mA O
46. g the two conductor cable from the treadle into the  Table Interface Junction Box at S4  refer to Figure 14 above      7  At the Table Interface Junction Box plug the interface cable into 57   Route this cable through an opening in the rear of the control and connect  it to TB   5 as show in Table 5 below   Note that there are no connections  made to terminal 1 or 3 on TB   5      6 5    Innovet Select   Summit Industries  Inc  Installation and Service Manual       Table 5 Interface Cable From Table to Control                                     S7 Interface Cable Wire Goes To Control   Color   Brown TB5 2   Blue TB5 4   Red TB5 5   Green TB5 6   Orange TB5 7   Violet TB5 8   Black TB5 9   White TB 5  10                            8  All options  such as table locks  collimator hip switches  latching prep foot  treadle  plug into the Table Junction Box  The Junction Box is designed  so that each cable can only plug into the proper connector  If a plug will fit  into more than one connector then it is a parallel circuit and can go in  either connector  Refer to the Electrical Schematic Section 10 for detailed  information about how the Junction Box is wired and how it fits into the  overall electrical system as needed     9  Connect the high voltage cables  taking note of the following important  points   A  Be careful of the three pins at each end of the cables as they can  break if the cable is dropped     B  Route the cables into the rear of the table so that they do not li
47. gain access to the four small Phillips head  screws which secure the table top frame s bearing track at the exterior  head end of the table  Remove these four screws  sliding the bearing  track forward  See Figure 7        Figure 7   Table Top Raised  Bearing Track Moved Forward For Mounting Tubestand    C   Manually release the transverse lock solenoid pins and slide the table  frame forward to expose the upper tubestand bearing rail     5 6    Innovet Select   Summit Industries  Inc  Installation and Service Manual       3  Remove the tubestand stop on the left end of the table s lower rear rail   As  viewed from the front   See Figure 3     Tubestand Stop on page 5 3 for  reference  Position the tubestand so that it is a few inches from the end of  the back rail of the table with the lower tubestand bearings in line with the  back rail     See Figure 4     Tubestand Positioning  Page 5 4 for reference     4  Carefully tilt the tubestand away from the table until the lower tubestand  bearing can be pushed into the bearing rail  Now straighten the tubestand  and push it completely into the upper and lower bearing rails  See Figure 5      Mounting Tubestand on Bearing Rail  Page 5 4 for reference  Reinstall the  tubestand stop on the lower rail  A well leveled table will allow smooth motion  of the tubestand with no drift in either direction     5  Reassemble the table frame by repeating step 2 above in reverse  Replace  the tubestand stop on the lower rail  If not done previ
48. h the control turned OFF  select DEMO mode by setting DIP2 5 of  J400 board to the ON  up  position     If the control turns on  look for voltage between P1 and P2  If more  than 4 VAC is present  replace the SCR  If control does not turn on  see C below     C  Confirm valid kVp input and J400 board    1     If the breaker still trips when in DEMO mode  disconnect the wire from  terminal W2 on the J400 board and turn the control ON  If the display  is stuck in  EXPO  mode  the kVp tap switch input voltage is missing at  the J400 board header H2  Check the wire to plug and plug to board  connections at H2  and the kVp tap switches for open resistive  contacts or connections    If the kVp display is not stuck in  EXPO  mode  and shows a valid  selection between 40 kVp and 125 kVp  replace the J400 board     Il  CIRCUIT BREAKER TRIPS AT EXPOSURE    The circuit breaker will trip due to either excessive mA or excessive line current   It may appear to be related to a specific mA station  or occur above a specific  kVp range  or occur at longer time stations     A  Determine if the mA calibration is valid    1     3     Remove the jumper between TB3 19 and TB3 20 and connect a mAs  meter between these points  ensuring that the meter is set to read AC  mA  The range of the meter scaling must be set to maximum    Typical filament voltage between XC XS for 100 mA is 39 VAC during  PREP      2          Typical filament voltage between XC XL for 300  mA is 51        during PREP      2   
49. he two feet  with a 4  long body should be installed at the front corners of the table  The  three feet with a 1  long body should be installed at the back rail   In the  accessory bag there are six blue and six red  U  shaped plastic spacers   When leveling the table fully tighten the foot until the shoulder of the foot is  firmly against either these spacers or the bottom of the table back rail   Use  of the spacers makes the table base leveling feet union a solid           piece   structure  minimizing any wobble or table base motion     Position the table close to its final location  using the leveling feet and spacers  to bring the table close to level at this time  Final leveling can be done at end  of assembly     HIGH VOLTAGE TRANSFORMER    Verify that the oil level in the HV transformer is approximately 12    to 3     below  the underside of the cover  Refill with  Shell Diala oil AX  or remove oil as  necessary     5 1    Innovet Select   Summit Industries  Inc  Installation and Service Manual       2  Using a piece of cardboard to protect the floor  push the transformer onto the  transformer shelf in the table by tilting the transformer and sliding it up onto  the shelf  Once the transformer is in place  loosen the screw on the vent plug  to allow for oil expansion  Cover the vent plug and flange with one of the  loose fitting plastic covers provided with the HV transformer to prevent  contaminants from entering the oil     Note  The high voltage transformer must b
50. ibed in detail in section 6 2     Figure 14     Table Interface Junction Box       6 2 MAKING CONNECTIONS    1     Connect the Primary cable  labeled P1  P2 and GND  and the Filament  cable  labeled XS  XL  XC and GND  to the High Voltage transformer  terminals which carry the same markings     Route the collimator cable  labeled CSW1  CSW2  C1  C2 and GND   through one of the openings in the rear of the control and connect it to TB      4 terminals 21  22  23  24 and 25 respectively  which are labeled the  same as the cable   Refer to Figure 15 for reference      6 3    Summit Industries  Inc     Innovet Select   Installation and Service Manual       DESCRIPTION  CONNECTIONS    TERMINAL  TB4  MARKINGS    Route the tubestand display cable out of the tubestand and into the black  connector on the back of the generator  The connectors are keyed for  proper orientation  This cable  as the stator and collimator cables  must  not limit nor be strained by tubestand motion     Route the x ray tube stator cable  labeled T5  T6  07  08 and 09  through  one of the openings in the control and connect to TB  4 terminals 31  32   33  34 and 35 respectively  which are labeled the same as the cable   Refer to Table 4 and Figure 15 for reference  Remove jumper between  terminals T5  amp  T6 of TB4 when connecting tubes thermal cut out SW     Table 4   Common Stator Cable Designations    COLOR CODE FOR STATOR CABLE    EUREKA MACHLETT SUMMIT  amp   TUBES TUBES TOSHIBA  TUBES         5    THERM
51. ion   In the following procedures  an mA will be chosen  a kVp tap switch voltage will  be set  and the kVp display will be adjusted to indicate the value shown in the  column on the left side of Table 12 below     Typically if the 50 kVp and highest  120 or 100  kVp are set in this manner  all  values in between will also be correct  The tap switch voltages for kVp outputs  between 50 and 100 kVp shown below are not needed for this procedure  but are  provided for reference     TABLE 12   TYPICAL KVP TAP SWITCH VOLTAGES                                  30 mA Station 100 mA 300 mA 400 mA  Unloaded Tap Station Station Station  kVp Switch Unloaded Tap   Unloaded Tap   Unloaded Tap  Voltage Switch Switch Switch  Voltage Voltage Voltage  50 104 108 155 183  60 121 125 173 201  70 139 143 190 218  80 156 160 208 236  90 174 178 225 253  100 191 195 243 271  110 209 213 260    120 226 230 278                     7 9       Innovet Select     Summit Industries  Inc  Installation and Service Manual       7 5 CALIBRATION OF kVp WITHOUT MEANS TO MEASURE ACTUAL  kVp  continued     1     Connect an AC voltmeter capable of measuring 0 300 VAC to the  common terminals of the minor and major kVp tap switches  or to TB3  terminals 10 and 11     Select an mA station  Adjust the kVp tap switches to achieve the voltage  which should result in 50 kVp per Table 12 above   For example  at the  100 mA station adjust the tap switches to achieve 108 VAC      Adjust the kV offset pot shown in Table 11 
52. ise which cannot be resolved through the troubleshooting  procedures described in this manual  please contact the Summit dealer where  you purchased your machine for assistance  or call Summit directly at 1 800 972   9776     Page 1    InnoVet Select       Summit Industries  Inc        2 0 PRECAUTIONS    2 1 Installation    The InnoVet Select must be installed and serviced by an authorized  Summit dealer  All warranties will be void if such installation or service  is performed by persons not authorized by Summit     2 2 Operation    2 2 1    2 2 2    Radiation    X rays are dangerous to both operator and others in the vicinity  unless established safe exposure procedures are strictly observed     The useful and scattered beams can produce serious  genetic or  potentially fatal bodily injuries to any persons in the surrounding  area if used by an unskilled operator  Adequate precautions must  always be taken to avoid exposure to the useful beam  as well as to  leakage radiation from within the source housing or to scattered  radiation resulting from the passage of radiation through matter     Those authorized to operate  test participate in or supervise the  operation of the equipment must be thoroughly familiar and comply  completely with the currently established safe exposure factors and  procedures described in publications such as Sub Chapter J of Title  21 of the Code of Federal Regulations  Diagnostic X Ray Systems  and their Major Components  and the National Council 
53. l   s    on off    switch to    On     Observe the following     TABLE 7     Display Readings  mAs display indicates   2 5    kVp display indicates    lt 40                7 1    Innovet Select     Summit Industries  Inc  Installation and Service Manual       5  Rotate kVp knobs so that display indicates   40 kVp  While observing the  tube filaments through the port of the x ray tube  verify that the correct  filament is lit for each mA station  The small focal spot should illuminate  for 30 or 100 mA  and the large focal spot should illuminate for 300 or 400    mA     6  Verification of Rotor circuit     A     B     Traditional Foot Switch   Depress the foot switch to the 1  stage  PREP   or the    PREP     button on the optional remote switch  and verify anode rotation and  filament boost  The bottom status display should show  WAIT   momentarily  then  PREP      Foot Treadle with Latching Prep  Depress and release the foot switch  The bottom status display on    the tube stand should show  WAIT   then  PREP   Verify anode  rotation and filament boost  The unit will automatically remain in  PREP mode for about 20 seconds  then return to idle condition     7  The kVp meter circuit is factory adjusted  but display accuracy should be  checked prior to other calibration steps     1     With power            connect an AC voltmeter capable of reading 0 300  VAC across the wipers  center terminals of tap switches  of the kVp  selector switches  major and minor  of TB3 terminals 9
54. lation and Service Manual       73 kVp CALIBRATION    The  nnoVet Select    kVp compensation circuit has four offset potentiometers   one for each possible mA station   and four slope potentiometers  These  adjustment pots are located on the J400 control board  See Figure 17      Electrical Control Board Component Identification for component orientation  below        Figure 17     Control Board Component Identification    kV POTS   8 TOTAL     Innovet Select   Summit Industries  Inc  Installation and Service Manual       7 4 kVp CALIBRATION WITH A MEANS TO MEASURE ACTUAL kVp    Note       order to make a valid comparison between the actual kVp output and  the kVp meter s preread indication  the tube current must be calibrated  accurately per section 7 2  Failure to do so will result in miscalibration  of kVp     For the following steps  use values in the table 11 below   TABLE 11   Adjustment Pot Identification                               Adjustment Pot mA Configuration  P1 30 mA offset  P2 100 mA offset  P3 300 mA offset  P10 400 mA offset  P4 30 mA slope  P5 100 mA slope  P6 300 mA slope  P11 400 mA slope          In the following procedure  fine tuning of the kVp display calibration depends on the  assumption that mA has been calibrated     1  Offset Adjustment     a  Select 50 kVp and 100 mA  While monitoring mA and kVp output   compare the measured kVp to the kVp meter s readings     b  Adjust the kVp offset pot  from Table 11 above  that corresponds to  the mA of you
55. ly injuries to any persons in the surrounding area if used by an unskilled  operator  Adequate precautions must always be taken to avoid exposure to the  useful beam  as well as to leakage radiation from within the source housing or to  scattered radiation resulting from the passage of radiation through matter     Those authorized to operate  test  participate in or supervise the operation of the  equipment must be thoroughly familiar and comply completely with the current  established safe exposure factors and procedures described in publications such  as Sub Chapter J of Title 21 of the Code of Federal Regulations   Diagnostic x   ray Systems and their Major Components   and the National Council on  Radiation Protection  NCRP        33   Medical x ray and Gamma Ray protection  for energies up to 10 MeV Equipment Design and Use   as revised or replaced in  the future     Failure to observe these warnings may cause serious  genetic or potentially fatal  bodily injuries to the operator or those in the area     3 1    Innovet Select     Summit Industries  Inc  Installation and Service Manual       3 2    3 3    ELECTRICAL    WARNING    Failure to comply with the following may result in serious or potentially fatal bodily  injuries to the operator or those in the area     Only properly trained and qualified personnel should be permitted access to any  internal parts  Live electrical terminals may be deadly  be sure line disconnect  switches are opened and other appropriate preca
56. mer lid  Normal  spacing is equal to the thickness of radiographic film     B  Jumper is still present between TB3 19 and TB3 20     C  Faulty mA   mAs meter or leads     XIII  UNIT WILL NOT GO INTO SLEEP MODE    A  Ensure DIP2 1 on J400 is set to UP  ON  to enable feature     B  Poor connection between  W2  on 4400 and mA overload coil of  ON OFF breaker     C  Shorted spark gap on top of transformer between M1 and GND   1  Proper spacing is the thickness of a piece of radiographic film     D  High line voltage fluctuation  new software may be less sensitive     7 20    Innovet Select     Summit Industries  Inc  Installation and Service Manual       8 0  8 1    8 2    8 3    8 4    8 5    FINAL ADJUSTMENTS AND QUALITY ASSURANCE  LEVELING THE TUBE ARM    An adjustment screw for leveling the tube arm is provided on the front of  tubestand  just below the tube arm base  Turn the Allen screw clockwise to  raise the tube arm or counterclockwise to lower the tube arm     40 INCH SID DETENTS    To position the tube 40  from either the table top or the film cassette  detents are  provided  These detents are factory adjusted  however some adjustment is  provided if needed  Access to the adjustments are from the rear of the  tubestand     VERIFICATION OF X RAY TO LIGHT FIELD COINCIDENCE    The coincidence of x ray to light field must be verified  Specific instructions can  be found in the manual for the G800 collimator     VERIFICATION OF X RAY TO IMAGE RECEPTOR ALIGNMENT    Any left 
57. mit  or restrict tubestand motion and they are not stressed by the  tubestand motion     C  The pins at the end of the HV cables should not be pinched tightly  together but be slightly spread apart  about the thickness of a utility  knife blade   For old style split pins only      D  The HV cable terminals must be thoroughly cleaned and then  coated with Vapor Proofing Compound  normally provided with x     6 6    Innovet Select     Summit Industries  Inc  Installation and Service Manual       10     ray tube  prior to insertion into the x ray tube or HV transformer  receptacles  It is common to use dielectric      in the HV  transformer  and vapor proofing compound in the x ray tube     E  Be sure that  Anode  of HV transformer connects to  Anode  of x   ray tube  and    Cathode    to    Cathode     Insert the HV cable  terminals into the appropriate receptacle and screw the cable nut  as tightly as possible by hand     Do not use tools for tightening     NOTE  Be certain to verify that the control s on off switch is in the   OFF  position  and that the main incoming power at the  service disconnect is also in the  OFF  position  It is strongly  recommended that a meter be used to confirm no voltage is  present before connecting the Line cable to the service  disconnect switch     Route the line cable under the lower tubestand rail and wire it into the  safety disconnect  Connect  L1  to one leg of the incoming line power    L2  to the other incoming leg of line power and   
58. o make exposures     n  Release Foot Pedal   When depressed  opens front of table to access  control panel         Table Base   All steel  welded construction     p  Circuit Breaker   Turns generator on off  Located on the side of the tilt out  panel      q  kVp Dials   One for major adjustment  15 kVp increments   One for minor  2  kVp increments      r  mA Dial   Adjusts mA  Two stations  300 and 100 when using small tube    focal spot and 300 when using large  Optional third mA station  either 400   large focal spot  or 30  small focal spot      Page 5    InnoVet Select       Summit Industries  Inc     5     2     Time Dial   Adjusts time  18 stations available    Line Frequency   Time Stations Available   50 Hz  1 100  2 100  3 100  4 100  5 100  6 100  8 100   1 10  12 100  18 100  24 100  30 100  36 100   48 100  6 10  9 10  1 2 10  1 1 2             60 Hz  1 120  1 60  1 40  1 30  1 24  1 20  1 15  1 12   1 10  1 8  3 20  1 5  1 4  2 5  1 2  5 8  3 4  1             Drain Trough   Minimizes spillage behind table   Table Top  float table top    24  x 53  of Formica  33  x 57  of Formica      Tubestand Grid Cabinet Interlock   When engaged  allows tubestand  grid  cabinet and cassette tray to move together       Digital Displays     Upper one for kVp  Center one for mAs  and Lower one for    status  amp  error messages     Collimator   Controls size and rectangular shape of primary beam emitted  from x ray tube and provides coincident light field     Shutter Adjustment Kn
59. oard  If the  Tube  error is  cleared inspect the connections at the cable plug  replace time  selector    c  Remove the H11 plug from the board  and jumper pin 1 to pin 2   If the error clears  inspect the switch and connections  replace  the mA selector    d  If the error still exists replace the J400 board     G   ZCO  ERROR    1  Verify 14 VAC between TP1 and TP2      J400 board  If voltage is  present switch the wires in      pins 2 and 3    error is still present   switch the wires back to their original locations and replace the J400  board     IX LONG  WAIT  DISPLAY     ROTR    ERROR AT EXPOSURE ATTEMPT    A  Verify rotor circuit   1  The A455 motor start board controls rotor delay time  240 VAC is  applied between stator main  07  and common  09  for 1 5 seconds   and then the voltage drops to 50 VAC  The red LED on the motor start  board turns on at that time  If the voltage stays at 240 VAC for longer  than 1 5 seconds  adjust R17  replace C4  or replace the board    2  If the red LED on the A455 board does not light  confirm 240 50 VAC  at 07 09 stator connections of the control and at tube end  Verify  stator resistance  30 ohms 07 black to 09 white  60 ohms 08 red to 09   white  90 ohms 07 black to 08 red   replace A455 board if voltages and  connections are correct     X  EXPOSURE BEEP or PREP AT TURN ON  A  Check for shorted exposure switch     1  If Latching Prep option is installed  verify open contacts between TB5 2  and TB5 4  If a short is measured  remov
60. obs   One each to determine width and depth of  primary beam     Swivel Mounting Plate   Allows collimator to be rotated     aa Lamp Switches  optional    Turns Collimator bulb on and off  Located on the    front and both ends of the table     bb Float Top Unlock Button  optional   Allows transverse travel of the table top     Located on the front of the table at both corners     Page 6    InnoVet Select       Summit Industries  Inc     3 2 Specifications    3 2 1    INTEGRATED GENERATOR     amp   amp      amp              amp                                     6                       Tilt out control    300 mA at 125 kVp    Dual digital timer with 18 time stations  1 100 to 1 1 2 second  for 50 Hz  or 1 120 to 1 second for 60 Hz      Adjustable solid state audio exposure signal    SCR contactor    Programmable tube protect system    Tube overload sensor    Independent kVp and mA adjustment circuits for each mA  station    Two or three mA stations  1005 and 300L  305  400L Optional     mA overload protection    64 kVp stations  40 to 125     kVp selection  major 15 kVp increments  minor 2 kVp  increments    mA  time and kVp knobs    Built in kVp compensation    Easy access exposure count display   Filament stabilization    16 foot line cable  4 foot control to transformer cable    Integral 24 Volt A C  power supply    Programmable line frequency 50 or 60 Hz    Programmable power saving mode    Latching prep option     Page 7    InnoVet Select       Summit Industries  Inc    
61. off switch for proper operation   7 3 4 Inspect collimator cable for fraying or damage   Tubestand   7 4 1 Verify set screws are securely holding tubearm   7 4 2 Inspect counterweight cables for fraying or damage     7 4 3 Inspect all tubestand movements for binding or interferences     Other    Inspect high voltage cable and all other cabling for damage     Page 20    InnoVet Select       Summit Industries  Inc     8 0 TROUBLESHOOTING    8 1    8 2    8 3    Error Message    Upon error detection  the InnoVet Select will display one of the following  error codes     FOOT Early release of the exposure foot switch or foot treadle  during exposure  Films are likely to turn out light  Foot  must be depressed for the entire selected exposure time     TUBE Single shot tube power limits were exceeded by the  selection technique  Reduce kVp  mA and   or time   PREP and EXPOSE locked out   Message clears when  the techniques change     HEAT Tube heat exceeded  PREP and EXPOSE locked out  until technique reduced or tube cools down    ROTR Tube rotor error  Call for service    ZCO Equipment component  hardware  error  Call for service    2222  Power saver mode only   Unit will go to sleep within 5  minutes    FIL1 FILA Hardware error  Call for service    COLD The unit needs to warm up for best output accuracy  but    exposures can be taken if needed  This display should  clear about ten minutes after first turn on of the day     No Exposure   If  after fully depressing the foot switch  
62. on  Radiation Protection  NCRP  No 33   Medical X Ray and Gamma   Ray protection for energies up to 10 MeV Equipment Design and  Use   as revised or replaced in the future     Failure to observe these warnings may cause serious  genetic or  potentially fatal bodily injuries to the operator or those in the area     Electrical    Only properly trained and qualified personnel should be permitted  access to any internal parts  Live electrical terminals may be  deadly  Be sure line disconnect switches are opened and other  appropriate precautions are taken before opening access doors   removing enclosure panels  or attaching accessories     Page 2    InnoVet Select       Summit Industries  Inc        2 2 3    Do not remove the flexible high tension cables from the x ray tube  housing or high tension generator or the access covers from the  generator until the main and auxiliary power supplies have been  disconnected     When disconnecting high voltage cables  they must be grounded  immediately in order to dissipate any electrical charge that may  remain on the cables or the tube     Failure to comply with the foregoing may result in serious or  potentially fatal bodily injuries to the operator or those in the area     Mechanical    All of the movable assemblies and parts of x ray equipment should  be operated with care  Only properly trained and qualified  personnel should be permitted access to any internal parts     Particular care should be taken when servicing the inside of th
63. ously  remove the  phenolic table top panel at this time via the nuts located at the underside  corners of the table top frame     5 7    Innovet Select     Summit Industries  Inc  Installation and Service Manual       5 4 TUBE ARM  X RAY TUBE AND COLLIMATOR INSTALLATION    1     Note     Install the tube mount by inserting the 1 1 2  diameter tube arm of the tube  mount into the collar on the face of the tubestand  Be sure to fully insert the  tube arm  Secure the tube arm by tightening the two set screws on the  outside diameter of the collar     Do not remove the two hex head shipping bolts which lock the main  counterweight and vertical carriage together until the tube  collimator and  high voltage cables have been mounted     Mount the x ray tube and collimator  When using the standard tube and  collimator  two 1 8  spacers will achieve the proper field size  coincidence  The standard tube has a part number of C458  and the  collimator G800  refer to their respective manuals for detailed instructions as  needed   Refer to Final adjustments Section 8 for central ray to grid cabinet  adjustments after x ray to light field verification and generator calibration is  completed later in this manual     Install the high voltage cables  Cable drape brackets are included with the  hardware accessories bag  The cable drape brackets are installed above and  below the handle on the right side of the tubestand  Loosely drape the HV  cables with the collimator and stator cables within
64. r     Page 16    InnoVet Select     Summit Industries  Inc     5 6 1 Peak Tube Potential  kVp     The Maximum Deviation of the Peak Tube Potential during an  exposure from its pre indicated value is   5 kVp     5 6 2 Tube Current  The Maximum Deviation of the tube current during an exposure  from its pre indicated value is   10  of selected value within a  kVp range of 50 125 kVp  inclusive   and   20  of selected  value within a kVp range of 40 50 kVp  exclusive     5 6 3 Exposure Time    The Maximum Deviation of the measured exposure time from its  pre indicated value           Max Deviation Time Range     1 pulse 1 100     12 100 sec  for 50Hz   1 120     1 10 sec  for 60Hz     15  18 100     1 1   2 sec  for 50 Hz   1 8     1 sec  for 60 Hz                 Page 17    InnoVet Select       Summit Industries  Inc        6 0    CLEANING THE INNOVET SELECT      The InnoVet Select was designed to withstand the sometimes extreme  conditions of a busy veterinary practice  In addition to the drip trough and rust   proof table legs  you ll note that there are no grooves or holes to collect debris     To periodically clean the unit  use a sponge and a non abrasive cleanser   Knobs on the control panel are removable for this purpose     The system must be located in a different room than the processor  Failure to  observe this could lead to premature corrosion not covered by warranty     Page 18    InnoVet Select       Summit Industries  Inc     7 0    SCHEDULED DEALER MAINTENANCE   
65. r Station  For a 100 mA station the pot is P2  until the  kVp meter s preread indication matches the actual measured kVp  output     2  Slope Adjustment   C  Select 120 kVp  Make an x ray exposure and adjust the slope pot   See Table 11 above  until the kVp meter s preread indication    matches the measured kVp output  Repeat steps 1 and 2 until no  further adjustments are necessary     7 7       Innovet Select   Summit Industries  Inc  Installation and Service Manual       3  Verify that the kVp display tracks well throughout the kVp range     4  Repeat steps 1  2 and 3 above for each mA Station  using Table 11 as a  guide for identifying the appropriate adjustment potentiometer     The offset and slope adjustments are interdependent  as an adjustment in the offset is  made  it will affect the slope  and as an adjustment is made in the slope it will affect the  offset  By anticipating the interactive response it will reduce the number of adjustments  required to achieve the desired accuracy of kVp display at each end of the kVp range     Under normal conditions the kVp meter s indication will not deviate from the measured  kVp output by more than 5 kVp     7 8    Summit Industries  Inc     Innovet Select            Installation and Service Manual       7 5 CALIBRATION OF kVp WITHOUT A MEANS TO MEASURE  ACTUAL kVp    The preceding Tables 10      mA Stations Band Labels  and 11      Adjustment Pot  Identification  contain settings and readings referred to in the following sect
66. r if function returns  or the J400 board     7 16    Innovet Select   Summit Industries  Inc  Installation and Service Manual       V   FIL 2  ERROR  high filament current during standby idle     A  Confirm J400 operation   1  Filament voltage during idle is typically about 24 VAC for small  28  VAC for large spot as measured XC to XS or XL  This idle voltage  typically produces 1 6 to 1 8 VDC at TP12  Turn pot P9 clockwise 1 2  turn  If error goes away  reduce the idling voltage by moving the  bottom band on the boost resistor as needed    2  If FIL2 is still present  remove H8 header from the J400 board and  cycle power  The error should change to FIL1  and voltage at TP12  should be near zero  If FIL2 error is still present replace the board            3  ERROR  low filament current during prep   A  Confirm contact of boost resistor band to the resistor    B  Confirm connection at H8 pins 7 and 8    1  These contacts close to bypass most of the boost resistor  If contacts  are good at H8 but closure does not occur at prep  replace A455 motor  start board     VII     FIL 4    ERROR  high filament current during prep     A  Gross miscalibration    1  Filament voltage is set too high    a  54 VAC at prep  XC to XS  is more than enough to achieve  300mA  A resistive cathode cable  filament cable or a miswire  of XC XL XS  causing both filaments to be illuminated at one  time  would require setting the filament voltage above this level  and create the error    b  54 VAC at prep
67. s   Cassettes can be stored in the optional  storage box or boxes mounted to either side of the table     Loading Cassettes   Three sizes of cassettes can be used  8  x  10   10  x 12  and 14  x 17   Simply place the cassette into the  specially formed grooves of the tray  It will accept the cassette in  either lengthwise or crosswise position     IMPORTANT     Although the storage box is  lead lined  it will not protect  fiim        fogging after  excessive exposure     Page 10    InnoVet Select       Summit Industries  Inc     4 2    4 3    Positioning the Tube    4 2 1    4 2 2    4 2 3    4 2 4    4 2 5    Moving the Grid Cabinet   The grid cabinet can be moved in  tandem with the tubestand  assuring alignment of the primary  beam to the center of the film  The grid cabinet can be positioned  anywhere along the entire length of the table  Grasp the  tubestand handle bars and move to the desired location     Disengaging the Grid Cabinet   Alternatively  the grid cabinet may  be moved independent of the tubestand  Release the interlock by  sliding the spring loaded lever extending from the grid cabinet to  the left  The grid cabinet may then move while the tubestand  remains stationary  To re engage the interlock  position grid  cabinet in line with tubestand  The interlock is engaged when you  hear a metallic  click  sound and feel the interlock mechanism  move into place     Vertical Movement   The tube can be moved from 10 1 2  to 40   above the table top  For convenienc
68. table base using the  hex head bolts provided     Plug the foot treadle cable into the S4 location of the junction box at the  right side of the table frame     Slide the control so that the notches engage the pivot rollers  lift the  control up into the table opening and resecure the control to the air  cylinder using the pin and locking ring     Remove the front panel of the control  Install the air switch packets using  the screws and nuts provided so that the nozzle of each exits the rear of  the control     Connect the wires from the air switches to TB5 2 and TB5 3 in the bottom  of the control  Connect the air hoses to the air switches by pushing them  over the nozzles which exit the rear of the control     Place a jumper between TB5     1 and TB 5     2  and another jumper  between TB5 3 and TB5 4  Verify on the J400 control board that DIP 1 5  is in the UP position   To enable the Latching Prep mode as shown in  Figure 18  page 7 12      If any existing exposure switches are to be left in the system  alert the  user that they will now operate like the Treadle  that is step on and release  to start PREP  then step on and release to start EXPOSURE  Generally   removing these old two position switches is recommended     9 2    Innovet Select     Summit Industries  Inc  Installation and Service Manual       10 0 ELECTRICAL SCHEMATICS AND DIAGRAMS    Figure Drawing No  Description   Page   Fig  1 Floor Space for Fixed Top                                                    
69. that the singe shot tube power limits are exceeded by the  selected techniques  It is accompanied by an error beep and prep is  not allowed  The message clears when allowed techniques are  selected    EXPOSURE With the kVp major  kVp minor and mA selector knobs turned fully   COUNT counter clockwise  the kVp display the bottom display will indicate   EXPO   The mAs and the kVp display will indicate the total number  of exposures taken on this x ray control    HEAT Indicates tube heat interlock  The message is accompanied by an    error beep and cleared as soon as the x ray tube cools  PREP is  locked out     Innovet Select     Summit Industries  Inc  Installation and Service Manual                             ROTR    FIL1    FIL2    FIL3    FIL4    COLD    READY    2222    Line voltage zero crossing signal not detected  Accompanied by  error beep  PREP locked out  Message clears as soon as zero  crossing signal detected     Foot switch released before exposure is finished  Accompanied by  error beep  The message is cleared when any of the techniques  change or PREP is requested     Tube rotor error  Accompanied by error beep  PREP is terminated   The message is cleared when any of the techniques change or  PREP is requested     Standby filament current too low  or not present  Accompanied by  error beep  PREP locked out  Message clears when standby  filament current is acceptable     Standby filament current too high  Accompanied by error beep   PREP locked out  Filament circ
70. the  tubestand and push it completely into  the upper and lower bearing rails  See  Figure 5     4  Reinstall the tubestand stop on the lower rail  If leveled properly in step 5 1  the  tubestand should glide easily from end to end and stay in position without drifting to    either side     5 4    Innovet Select     Summit Industries  Inc  Installation and Service Manual       For Optional Float Top Table  Refer to photos on preceding pages  for reference      Mount the tubestand bearing carriage to the tubestand using the four  4  14   20  x 3 8  hex head screws provided  Figures 4 and 5 on page 5 3 show a  tubestand with a mounted bearing carriage  Verify that the safety rollers  are mounted on the tubestand bottom to prevent the table from tipping  backwards from the weight of the tubestand   This may occur if the high  voltage transformer is not yet installed      To mount the tubestand  the table top frame will need to be moved forward for  access to the upper tubestand bearing rail at the left end of the table  Refer  to Figures 6 and 7 below for reference     A  Remove the single  large Phillips head screw found at the front and  rear interior corners of the table top frame  Their location is shown in  Figure 6  front only shown      Figure 6     Table Top Frame Mounting Screws       5 5    Innovet Select   Summit Industries  Inc  Installation and Service Manual       B  Now that the table top frame is free from the bearing tracks  you can  raise the frame slightly to 
71. the total number of exposures taken on this x ray  control   Exposure Foot Treadle Latching PREP  option   4 7 1 Press exposure foot switch      treadle momentarily and release   4 7 2 WAIT is displayed for 2 seconds while tube is warmed up     4 7 3 PREP is displayed for 20 seconds  Exposure can be made any  time during this interval     4 7 4 For exposure push foot switch or treadle down and hold for  duration of exposure     4 7 5  f after 20 seconds no exposure is made  PREP is released     Page 15    InnoVet Select       Summit Industries  Inc     5 0 TECHNICAL DATA    5 1    5 2    5 3    5 4    5 5    5 6    Radiation Leakage  Meets Federal Government Standards     5 1 1 From the tube housing  less than 50 mR hr at 1m from focal spot    4mA  125kVp     5 1 2 From collimator  less than 50 mR hr at 1m   4mA  125 kVp   Rated line voltage  240 VAC  50   60 Hz   single phase    Range of line voltage regulation at max line current  196 to 596   Maximum line current  based on input voltage of 240 VAC   140 AMP    Technique factors that constitute the max  line current  300 mA   125  kVp     5 51 Generator rating   5 5 1 1  Output current  100 300 mA   5 5 1 2  Output voltage  40 125 kVp   5 5 2 Generator duty cycle   5 5 2 1  100 mA   125 kVp  4   5 5 2 2  300 mA   125 kVp  1   Maximum Deviation of technique factors     The maximum deviation statements provided below are broad and  considered  worst case  conditions     Typical performance will normally be significantly bette
72. there is no audible tone  your unit  may not be turned on  Use the circuit breaker to turn on power  If there is  still no exposure  call a service representative    Collimator light bulb burned out    When the collimator light bulb burns out  you may either call a service  person for replacement realignment or may choose to replace the bulb    yourself     The replacement bulb must be FCS type  150 Watt at 24 V with a pin  base     Page 21    InnoVet Select     Summit Industries  Inc     WARNING  1     DO NOT LOOK DIRECTLY INTO THE  COLLIMATOR LIGHT SOURCE AS RETINA  DAMAGE COULD RESULT     WARNING  2     TURN THE CIRCUIT BREAKER OFF  BEFORE ATTEMPTING BULB  REPLACEMENT TO AVOID POSSIBLE  ELECTRIC SHOCK     First  remove the back cover of the collimator  The bulb is now visible   WARNING  3     THE BULB  WHEN LIT  GETS VERY HOT   ALLOW IT TO COOL SEVERAL MINUTES  TO AVOID BURNING YOUR FINGERS     Pull the bulb straight out from the ceramic base   WARNING  4     DO NOT MAKE DIRECT CONTACT WITH  THE REPLACEMENT BULB AS OIL FROM  YOUR FINGERS WILL CREATE HOT  SPOTS ON THE BULB WHICH  SUBSEQUENTLY MAY LEAD TO  CRACKING     To avoid direct contact between bulb and fingers  the replacement bulb  is normally packaged in a cellophane wrapper  Tear open one end so  that prongs are exposed  hold the glass portion  still in its wrapper  and  insert the prongs into the ceramic base  Make sure that each prong is  securely fastened into its socket so that the resulting light field is  str
73. to right adjustment required to achieve coincidence between the central  ray and the center of the image receptor can be achieved by adjusting the grid  cabinet to tubestand interlock bracket  located on the lower front surface of the  tubestand     1  Any front to back adjustment of the central ray necessary  after the arm  has been leveled from 8 1 above  can be achieved by the slotted holes in  the tube mounting platform  Slightly loosen the collimator mounting bolts   shift the tube and collimator as required  and re tighten the mounting bolts     SECURING TABLE TO FLOOR    Once the table is in final position and all checks have been made  it can be  secured t the floor using the four  4  clamps provided with the feet  Merely slide  the clamp into the notch at the base of each foot  turn at 45  to the table and  secure with a lag screw     8 1    Innovet Select     Summit Industries  Inc  Installation and Service Manual       9 0    9 1    9 2    9 3    OPTIONS FOR FIELD INSTALLATION    Many options for the  nnoVet Select     will arrive at the jobsite pre installed   However  all of the following features can be added after the installation is  complete if the user so desires     FILM BIN    The film bin comes in either a right hand or left hand configuration  See  instructions that are provided with film bin kit     TABLE TOP EXTENSION  For Fixed Top tables only     The table top extension adds 12  of additional table top surface and can be  mounted on either end of the
74. uit disabled  Message will not clear  without rebooting system and resolving cause of high idle current     Boost filament current too low  Accompanied by error beep  PREP  locked out  The message is cleared when any of the techniques  change  PREP is requested  or cause of low boost current is found  and resolved     Boost filament current too high  Accompanied by error beep  PREP  locked out  The message is cleared when any of the techniques  change  PREP is requested  or cause of high boost current is found  and resolved     The control is cold  Exposures are possible  but the outputs will not  be at their maximum accuracy   COLD  will change to  READY   once the unit is on for about ten minutes     READY scrolls across the tubestand display once the unit has  warmed up to a level that delivers the most accurate outputs  about  ten minutes after initial turn on     Sleep timer is 5 minutes away from turning off system   Accompanied by error beep  Active in power saving mode only     7 13    Innovet Select     Summit Industries  Inc  Installation and Service Manual       7 9 Detailed Troubleshooting Guide for K200 Generator       CIRCUIT BREAKER TRIPS AT TURN ON  A  Confirm mechanical latching of breaker    1            OFF main input power at the wall breaker    2  Flip the control circuit breaker ON and OFF several times to confirm    the breaker will mechanically latch in the ON position  If not  replace  breaker     B  Check for shorted or leaking SCR    1     2     Wit
75. utions are taken before opening  access doors  removing enclosure panels  or attaching accessories     Do not remove the flexible high voltage cables from the x ray tube housing or  high voltage transformer or the access covers from the generator until the main  and auxiliary power supplies have been disconnected     When disconnecting high voltage cables  they must be grounded immediately in    order to dissipate any electrical charge that may remain on the cables or the  tube     MECHANICAL  WARNING  Particular care should be taken when servicing the inside of the tubestand     There is an extreme threat of mechanical pinching between the vertical slide and  counterweight due to their close proximity and opposite directions of motion        The control is secured to the table with an air cylinder  hinge pin and Truarc  retaining ring only  Once the ring and hinge pin are removed  care must be taken  to insure that the control does not swing forward in an uncontrolled manner     All of the movable assemblies and parts of x ray equipment should be operated    with care  Only properly trained and qualified personnel should be permitted  access to any internal parts     3 2    Innovet Select     Summit Industries  Inc  Installation and Service Manual       4 0 SPECIFICATIONS    41 TECHNICAL RATINGS    TABLE 2     TECHNICAL RATINGS       300mA Station    400 mA Station   Optional        Rated Line Voltage        240         50   60 Hz    Single Phase     240         50   60 Hz  
76. ve the technique select knobs  each of which are secured to their  respective switch shafts by two Allen set screws  Remove the two Phillips  head screws with rubber gaskets at the top of the control cover  Remove  the two shoulder bolts from the bottom corners of the control cover and lift  cover away as shown in Figure 11 below        Figure 11  Removing the Front Cover    Once the front cover has been removed  the interfacing terminal strips  within the control will be accessible as shown in Figure 12 below     6 1    Innovet Select     Summit Industries  Inc  Installation and Service Manual       TB  2    TB  3    TB  4    TB  1       Figure 12     Interior of Control    If not already done  remove the phenolic table top panel for access to the  rear of the control  The table top panel can be removed by a hex nut at  each underside corner of the table top  The Line  Primary  Filament   Collimator  Stator  Foot Switch and Table Interface cables will all enter the  control as shown in Figure 13 below        Figure 13     Rear of Control Showing Cable Entry Points    6 2    Innovet Select     Summit Industries  Inc  Installation and Service Manual       There is a  Table Interface Junction Box  where the cables for all Options  will plug in   Refer to Figure 14 below   There will be one cable from this  Junction Box which will enter the control and connect to TB     5  acting as  the interface between the control and all possible table options  The  connections will be descr
77. zed Summit dealers or if equipment is  interconnected with components not approved by Summit for compatibility     The x ray tube manufacturer s warranty is one year  pro rated  Summit will  extend this warranty to five years  pro rated on manufacturer s list price at time  of claim  provided the tube is purchased and used exclusively with an InnoVet  Select system     Please call your Summit dealer to coordinate the return of a  warranty part or call Summit direct at 1 800 972 9776     Page 25    INNOVET SELECT      Veterinary Radiographic  System    Installation  amp  Service Manual    K286  Rev J    SUMMIT INDUSTRIES INC   2901 W  Lawrence Avenue  Chicago  IL 60625  773 588 2444  www summitindustries net       Innovet Select     Summit Industries  Inc  Installation and Service Manual       1 0  2 0  3 0  4 0  5 0    6 0    7 0    Table of Contents    MANUAL REVISION CHART             eee ee en ee te ea 1 1  UNPACKING Geta tact T  tame sa a                         2 1  RADIATION AND MECHANICAL ELECTRICAL WARNING                                 3 1  EQUIPMENT SPECIFICATIONS               E Seu coU M bx dE  4 1  PRELIMINARY MECHANICAL ASSEMBLY                                                          5 1  5 1 RADIOGRAPHIC TABLE               tee eme tie ee ipao  5 1  5 2 HIGH VOLTAGE THANSFORBM ER                  paa area             5 1  53 TUBESTAND INSTAELATION       tte                                  5 2  5 4 TUBE ARM  X RAY TUBE  CABLES AND                             
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  Guía de instalación Router Fibra Óptica Comtrend  Remington S6500 hair stylers  ロ ロ 之 R ロ  GeMS - Susan G Komen® Northeast Ohio  Smashweld 318 Smashweld 318 TopFlex Smashweld 318 Plus      Copyright © All rights reserved. 
   Failed to retrieve file