Home
Model 2800
Contents
1. Tube size mm diam x length 17x 121 17x 121 I7 X 121 17x 121 Swinging Maximum Speed rpm 4 000 4 000 4 000 4 000 bucket rotor Maximum RCF x g k1 2 590 2 580 q Bucket Code No 053 4910 053 5010 053 0040 RS 240 Buckets and adaptors are option Tube Rack Adaptor Code No 055 1280 S74 Autoclavable 121 C The number of times allowed for autoclaving are limited Please refer an instruction manual A Equipped with sealing caps patent pending Micro plate rotor Maximum Speed Maximum RCF rom xg Bucket Code No Test plates Nominal Capacity Plate size mm RMP 206 8 6 x Micro plates 2 x PCR plates 86 W x 128 D x 52 H 1 790 Equipped with rotor 2 x Deep well plates x The number of times allowed for autoclaving are limited Please refer an instruction manual KUBOTA 2420 2800 lt gt 122172 86 110 12 17 8 x 87 110 I F bE 85 110 12 17 2 x86 105 2 772 x 86 110 je 2508S e105 4 000 4 000 4 000 4 000 4 000 4 000 5 2 540 2 500 2 520 2 490 2 500 2 490 053 4910 053 4990 053 4900 053 4810 053 4980 053 7340 055 7400 6 024 1400 NU Rubber cushion NNA 27 37x95110 30 x 117 30 x 117 4 000 4 000 4 000 2 540 2 610 2 610 053 4990 053 0040 053 5010 A Equipped with sealing caps patent pending 1 To show the RCF exactly input the rotation radius At the f
2. e From 100 to 2 600 xg 100 xg increments RCF indication e Digital display from 0 to 3 000 xg 100 xg increments Timer setting and indication e Digital display with Hold 1 second to 99 seconds 1 sec increments 1 minute to 99 minutes 1 min increments Sounds for informing centrifuging stop amp 5 types and no sounding Power requirements Single phase 50 60Hz 110 115V 10 8A Single phase 50 60Hz 220 230V 10 5A Rated voltage and rated currente 110V 115V 2 5A 220V 230V 1 3A Power consumption and Heat output e 180W 0 65MJ h Dimensions weight e 35cm x 42cm x 32cm WxDxh 24kg Certificates of conformity e IEC 61010 2 020 CE Marking 220V 230V Model 2800 Maximum Speed e 4 000 rpm Maximum RCF 2 610 xg Maximum Capacity e 360 mL Control system e Microprocessor control brushless motor Speed RCF g Time Temperature Acceleration and Deceleration 3 memory banks Alarm display e Lid open Imbalance Over speed Abnormally high temperature Detected electrical abnormality in motor inverter lid interlock speed sensor and temperature sensor Acceleration Deceleration e Two selectable rates Rapid Slow 4 ways of setting Speed setting e From 300 to 4 000 rpm 100 rpm increments Speed indication e Digital display from 0 to 4 300 rpm 100 rpm increments RCF setting e From 100 to 2 600 xg 100 xg increments RCF indication e Digital display from 0 to 3 000 xg 100 xg increments Timer setting and indication e Digital display with HOLD func
3. EC SLOW SLOW deceleration Actual rotor radius can be set enabling accurate display of RCF values ACC SLOW SLOW acceleration A ACCEL V DECEL DEC RAPID RAPID deceleration A ACCEL V DECEL V or Z Pre cooling allows rotors to reach low temperature before experiments ACC SLOW SLOW acceleration DEC SLOW SLOW deceleration Memory flashing enables easy spin down Patented Deceleration characteristics Characteristics of Slow 1000 Natural deceleration Time Characteristics of rapid The switch over point of N to the natural deceleration can be selected from 0 to 4 000 rpm Operational Panel TEMP key SPEED display Speed RCF Knob Setting the temp speed RCF and time SPEED key A TIME key RE i TEMP display T A TIME display Ey STOP key Pre cooling key FLASHING key START key ACCEL DECEL key MEMORY selection key Rotor Bucket Tube Rack Table for Model 2420 2800 Flow Cytometer Tube size mm diam x lengih 9 5 11 x 36 60 12 17 x65 05 Swinging Maximum Speed rpm 4 000 4 000 bucket rotor Maximum RCF x g k1 1 950 2 410 Bucket Code No 053 5040 053 4930 RS 240 Buckets and adaptors are Tube Rack Adaptor Code No Autoclavable 121 C x The number of times allowed for autoclaving are limited Please refer an instruction manual
4. KU_OTA Compact Table Top Centrifuge www centrifuge jp Model 2420 The small footprint Model 2420 can centrifuge 24 x 15mL tubes at 4 000rpm Compact size only 35cm W x 42cm D x 32cm H Rotor for PCR plates Micro plates Deep well plates can be used 2 x PCR plates 6 x Micro plates or 2 x Deep well plates can be centrifuged at a maximum of 1 790 x g with the RMP 206 rotor In the event of breakage all fragments will be safely retained inside the centrifuge unit Interlocking parts are protected against mechanical impact and employ a two stage electrical protection circuit No air intake vent on top of lid To prevent inspiration of foreign material the air intake vent is built into the lid Plastic tachometer port on the top of the lid enables easy checking of actual rotor speed and complies with GMP reguirements footprint and loaded with features es of Swinging Bucket 4 000rpm Rotor swinging bucket rotors for 24 x 15mL tubes otor for 2 x PCR plates 6 x Micro plates 2 x Deep well plates Image of bucket Comes with sealing cap included with bucket Adaptor for 15mL Conical tubes Code No 055 1280 Patent pending Autoclavable 121 C 4 x 50mL Conical tubes Code No 053 0040 Autoclave parameters should not be exceeded Refer to operation and service manual Useful functions for easy operation Three available memory settings with one key save and recall o
5. actory the rotation radius is pre set at 14 6cm The above Maximum RCF is the calculated value at the maximum speed at each rotation radius 2 Tube support can be produced for special tube sizes Contact our distributor 3 5 7mL Blood collecting tubes made by SEKISUI Insepack TI Termo Venoject II Nipro Neotube or Beckton Dickinson Vacutenar can be used lt 4 7 10mL Blood collecting tubes made by SEKISUI Insepack II Termo Venoject II Nipro Neotube or Beckton Dickinson Vacutenar can be used 5 Inthe case of 15mL taper tubes please operate at 3 000rpm or less Bucket 053 7340 is exclusively used with 2420 2800 6 In case of using glass tube rubber cushion 024 1400 is necessary 7 This is sealing bucket and it is equipped with sealing cap and O ring silicon Autoclavable at 121 C 8 Verify plate strength before centrifuging Plates in lower position may fracture Model 2420 Maximum Speed e 4 000 rpm Maximum RCF 2 610 xg Maximum Capacity e 360 mL Control system e Microprocessor control brushless motor Speed RCF g Time Acceleration and Deceleration 3 memory banks Alarm display e Lid open Imbalance Over speed Detected electrical abnormality in motor inverter lid interlock and speed sensor Acceleration Deceleration e Two selectable rates Rapid Slow 4ways of setting Speed setting e From 200 to 4 000 rpm 100 rpm increments Speed indication e Digital display from 0 to 4 300 rpm 100 rpm increments RCF setting
6. employ a two stage electrical protection circuit Plastic tachometer port on the top of the lid enables easy checking of actual rotor speed and complies with GMP reguirements Autoclavable 121 C Table for centrifuge with 4 x 50mL Conical tubes RMP 206 Plate Rotor casters and adjuster optional Code No 053 0040 6 x Micro plates Code No 047 3820 Image of bucket Comes with sealing Size 41cm w x 64cm d x 46cm h cap included with bucket Adaptor for 15mL Conical tubes Code No 055 1280 Patent pend l ng x Autoclave parameters should not be exceeded Refer to operation and service manual KUBOTA 2800 Useful functions for easy operation Acceleration Deceleration Function Rapid slow acceleration and rapid slow deceleration can be selected to suit sample type and experimental conditions The button operations are shown below RMP 206 rotor is available Three available memory settings with one key save and recall operation Numerous detectors ensure safe reliable operation If trouble is detected operation stops immediately Five different buzzer sounds allow easy discrimination when any centrifuge operation ends among more than ACC RAPID RAPID acceleration one centrifuge V DECEL DEC RAPID RAPID deceleration Complies with GMP requirements Actual rotor speed can be measured using the tachometer port on the top of the lid ACC RAPID RAPID acceleration D
7. he equipment or its parts to prevent occurrence of functional failure or accidents immediately turn off the power switch disconnect the power cable plug from the mains socket and contact your local distributor or Kubota Corporation if there is no local distributor in your neighbourhood The rotor or bucket appears to be damaged or corroded The equipment emits a burnt smell Me You receive a minor electric shock when you touch the DISCONTINUE Conditions listed at eguipment with your bare hand me RE ONE He TESTE OT Any other abnormality or indication of failure is noticed Always discontinue use if any of the For your safe use of the equipment be sure to Do not misplace the instruction manual Keep the instruction read the instruction manual carefully before you manual nearby so you can refer to it whenever necessary start operations Safety Precautions e The term of supplying spare parts for repair is 10 years after discontinuation of production except spare parts which we are unable to procure e This catalogue is not for distribution in the USA Canada and Mexico as products shown are not for sale in these countries KUBOTA CORPORATION http www centrifuge jp 29 9 Hongo 3 chome Bunkyo ku Tokyo 113 0033 Japan Tel 81 3 3815 1331 Fax 81 3 3814 2574 E mail f trade kubotacorp co jp Copyright 2012 KUBOTA CORPORATION N JUN Printed in Japan
8. ion Characteristics of rapid The switch over point of N to the natural deceleration can be selected from 0 to 4 000 rpm key RMP 206 Plate Rotor 6 x Micro plates Oro naroa oo se rs TONA RILI A Tore 2 ee ste re s og O A LIL wette eg Autoclave parameters should not be exceeded Refer to operation and service manual Many researchers want to use microplates so that large numbers of samples can be easily centrifuged using less energy and less time The 2420 can use the RMP 206 rotor for centrifuging 6 x Micro plates at 1 790 x g Model 2800 can also use the RMP 206 rotor Model 2800 Table top refrigerated centrifuge with a small footprint Two Types of Swinging Bucket 4 000rpm Rotor e Multiple swinging bucket rotors for 24 x 15mL tubes e Plate rotor for 2 x PCR plates 6 x Micro plates 2 x Deep well plates The small footprint Model 2800 can centrifuge 24 x 15mL tubes at 4 000rpm Compact size only 37cm W x 64cm D x 37cm H and refrigerated Rotor for PCR plates Micro plates Deep well plates can be used 2 x PCR plates 6 x Micro plates or 2 x Deep well plates can be centrifuged at a maximum of 1 790 x g with the RMP 206 rotor Sample temperature is maintained at 4C at maximum speed 25 C ambient temperature In the event of breakage all fragments will be safely retained inside the centrifuge unit Interlocking parts are protected against mechanical impact and
9. peration The memory function makes repetitive operations using the same speed and time values easy Numerous detectors ensure safe reliable operation If trouble is detected operation stops immediately Buzzer sounds at the end of operation Five different buzzer sounds allow easy discrimination when any centrifuge operation ends among more than one centrifuge Complies with GMP requirements The tachometer port on the top of the lid enables easy checking of actual rotor speed and complies with GMP requirements Actual rotor radius can be set allowing accurate display of RCF values Operational Panel SPEED display Knob for setting Speed Speed RCF TIME display RCF and Time a a X 100 rpm o 8 x100g FUNCTION STOP key ACCEL DECEL key FUNCTION key MEMORY selection START key KUBOTA 2420 Acceleration Deceleration Function Rapid slow acceleration and rapid slow deceleration can be selected to suit sample type and experimental conditions The button operations are shown below RAPID RAPID acceleration DEC RAPID RAPID deceleration ACC DEC RAPID RAPID acceleration SLOW SLOW deceleration ACC DEC SLOW SLOW acceleration RAPID RAPID deceleration ACC DEC SLOW SLOW acceleration SLOW SLOW deceleration Deceleration characteristics Characteristics of Slow 1000 Natural decelerat
10. tion continuous operation and flashing 1 second to 99 seconds 1 sec increments 1 minute to 99 minutes 1 min increments Sounds for informing centrifuging stop 5 types and no sounding Temperature setting indication e Digital display 9 C to 40 C 1 C increments Pre cooling Refrigerante R134a CFC free Power requirements e Single phase 50 60Hz 110V 10 115V 10 15A Single phase 50 60Hz 220V 10 230V 10 10A Rated voltage and Rated currente 110V 4 5A 115V 4 4A 220V 2 4A 230V 2 3A Power consumption and Heat output e 380W 1 4MJ h Dimensions weight 37cm x 64cm x 37cm WxDxH 110V 115V 48kg 220V 230V 51kg Certificates of conformity e IEC 61010 2 020 CE Marking 220V 230V K1 Not including projections All specifications and external appearance are subject to change without notice Use only rotors adaptors and tube racks made by KUBOTA Do not attempt to use rotors adaptors or tube racks made by other manufacturers Follow all instructions in the relevant Instruction amp Service Manual when operating centrifuges Kubota has acquired ISO 9001 and ISO 13485 certification Products in this catalogue are designed for use by people having the requisite technical knowledge and must always be used with care People who do not have adequate technical knowledge or training should use our products only under appropriate supervision or else hazards are likely to occur In the event of any observed abnormality in t