Home
        Service Manual
         Contents
1.              0 50 100 150 200 250 300 fL  Figure 4 2  3  Ahistogram that does not show differential result because the WBC result is  less than 0 5        0 50 100 150 200 250 300 fL  WBC 0 4 x 1097  Figure 4 3  4  Ahistogram that does not show differential result because the peak is in the  middle of the histogram        0 50 100 150 200 250 300 fL  Figure 4 4  5  A histogram that does not show differential result because of increased  nucleated erythrocytes  or interference or inadequate hemolysis       BC 2800 Auto Hematology Analyzer  V1 1  4 1    Histograms and Pulse Graphs       0 50 100 150 200 250 300 fL  Figure 4 5  6  A histogram that does not show differential result because of severe  interference  check the pulse graph to determine whether there is  interference  in the WBC channel        0 50 100 150 200 250 300 fL  Figure 4 6  7  No lyse reagent or poor hemolysis          0 50 100 150 200 250 300 fL  Figure 4 7  8  Increased granulocytes    WBC               0 50 100 150 200 250 200 fL  Figure 4 8  9  Increased lymphocytes  WBC R2       O 50 100 150 200 250 300 fL  Figure 4 9  10  Tumor patient  WBC R4          _    O 50 100 150 200 250 300fL  Figure 4 10    11  Increased monocytes  WBC          7 Ss Seed Sn     0 50 100 150 200 250 300 fL  Figure 4 11       4 2 BC 2800 Auto Hematology Analyzer  V1 1     Histograms and Pulse Graphs                         4 2 Pulse Graphs   After each count  the system can save the original sampling pulses of this time  We can 
2.              Figure 6 15 A dialog box of Prepare to ship    Move the cursor to Yes and press  ENTER  if you want to proceed with this  procedure  or to No and press  ENTER  if you wan to abort this operation  If  there is any error present  the system will not perform this procedure  Follow the  instructions displayed on the screen to remove the error before trying to perform  the procedure again    After draining the tubing  follow the instructions displayed on the screen to put  the rinse  diluent and lyse tubing into distilled water and press  ENTER  to flush  this analyzer with the distilled water    When the washing is over  follow the instructions displayed on the screen to  remove the rinse  diluent and lyse tubing from the distilled water and press   ENTER  to drain the tubing again    Turn off the analyzer when the screen displays Turn off the analyzer    Wipe this analyzer dry and wrap it up for storage        6 16    BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 3 System Status    The items displayed in the Status screen reflect how the system is functioning  and contribute significantly to diagnosing system errors  You may follow the  instructions given below to check those items    Press  MENU  to enter the system menu and press the appropriate arrow keys  to move the cursor to Service     Status  as Figure 6 16 shows                     Current sample WBC 3 1 x 10 L  Lymph 3 7 x 10   L  Mi ct 0 7 x 10 7L Wh  ID 3118 Grane 47x tol BL  Time  01 24
3.             Figure 6 4 Rinse Priming screen       6 6 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 2 3 Lyse Prime    Biohazard  iS Wear standard laboratory attire  including rubber gloves  and  D4 follow safe laboratory procedures when handling any material in    the laboratory        You should perform the Lyse prime procedure to prime the lyse tubing with lyse  when      there are bubbles in the tubing  or    the lyse in the tubing is contaminated  or    the old lyse ran out and a new container of lyse is installed     Follow the steps given below to do so      At the Maintenance screen  press the appropriate arrow keys    t      J  lt         to move the cursor to Lyse prime      Press  ENTER  to prime the tubing with lyse and the priming progress will be  displayed at the bottom of the screen  as Figure 6 5 shows      When the priming is done  the screen will return to the initial state           Dilvent prime Lyse test   Rinse prime Clean bath  Drain bath   Zap aperture Drain tubing   Flush aperture Clean wipe black    Probe cleanser cleaning Prepare to ship    E Z cleanser cleaning       Priming lyse    Cyse prime                  Select service       Figure 6 5 Lyse priming screen    6 2 4 Zap Aperture    You can perform the Zap aperture procedure to unclog or prevent clogging   Follow the steps given below to do so       At the Maintenance screen  press the appropriate arrow keys    t         lt               BC 2800 Auto Hematology Analyzer  V1 1  6
4.    System Self Test eneee e eieiei 6 18   o oe E E EEPE PE EE PEPEE 6 23  6 6 System Configuration          2000000000 0000000000 00000000 01000000 0t000eeeeteeereeeteeeneneeee 6 25  6 7 Printing Management 2 24 2 02 1552 5  00ic0ck oon e eee ea 6 26  6 8 Adjusting Sample Probe and Replacing Probe Wipe                     08 6 28  Chapter7 TroubleSMOoting sic 2c  oct 2c nce acces at eds ad cec tad eSee sad ceded aed aedecels 7 1  GM Eror Godes ieee e eee e eE e ded eucde baa E E EERE AEE EEE 7 2  AZ     SOMOS  Alain natal au a E Me sede E T T 7 3  Chapters List of Spare parts sic  cael ccidvost ce didtesannne aectielstacatestacelsiees 8 1  Appendix BC 2800 Error MeSSaQ               ccccccccccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeseeaes 1    BC 2800 Auto Hematology Analyzer V 1 1       Hardware       Chapter1 Hardware    1 1 General    1 1 1       Schematic    puesane shesnneneesnsa               _    Analog board    Power drive       K      keyboard   hye Se    pannes       i  g   ig   i oe    Figure 1 1 Schematic of the CPU board    The CPU  FPGA and Super I O are the major components on the board  The  CPU carries out the instructions and functions as the core of the board  The  FPGA functions as the relay between the CPU and the Super IO  The Super  I O includes various interfaces that can be accessed by the CPU through the  FPGA  System memories are SDRAMs  The DOM is a Disk On Module that  stores the system software and test data  The RTC is a real time clock  System  confi
5.   Be sure not to  touch the power plug at the back of this analyzer with wet hands or wet rags    Be sure not use organic solvent or acid alkaline detergent to wash the surface of  this analyzer  Otherwise  the surface may fade or become corrupted    Be sure to avoid direct contact with the reagents that will hurt your eyes  skin and  diaphragm  In case you spill the reagents on you skin  be sure to wash them off  with much water  In case you spill the reagents into your eyes  be sure to  immediately wash your eyes with much water and go see a doctor for further  treatment        6 2    BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 1 2 Recommended Regular Maintenance    Table 6 1 Recommended Regular Maintenance       Maintenance  Period    Content of Maintenance       Everyday    If you are to use this analyzer 24 hours a day  be sure to perform the    E Z cleanser cleaning procedure everyday        Run the QC program everyday  S       Every three  days    If you are to use this analyzer 24 hours a day  be sure to perform the    Probe cleanser cleaning procedure every three days        Every Week    If you shut down your analyzer every day and follow the specified  shutdown procedure to do that  you need to perform the Probe    cleanser cleaning procedure every week        Every Month    You should use the supplied probe localizer to calibrate the  position of the probe to that of the wipe block  The analysis  result is sensitive to their alignment        As 
6.   amp  Time and set the time and date and  restart the analyzer to enable the new settings  If the error still remains  call       BC 2800 Auto Hematology Analyzer  V1 1  7 7    Troubleshooting       Mindray Customer Service Department or the distributor     7 2 24 Syringe Motor Error    Something is wrong with the motor controls the syringe that  aspirates dispenses samples and reagents    Solution    Access Service     Self test     Machine and test the motor The error will be  cleared if the result is normal  otherwise  call Mindray Customer Service  Department or the distributor     7 2 25 Rotation Motor Error    Something is wrong with the motor that rotates the sample probe    Access Service    Self test     Machine and test the motor  The error will be  cleared if the result is normal  Otherwise  call Mindray Customer Service  Department or the distributor     7 2 26 Elevator Motor Error    Something is wrong with the motor that controls elevation of the sample probe   Access Service     Self test     Machine and test the motor  The error will be  cleared if the result is normal  Otherwise  call Mindray Customer Service  Department or the distributor     7 2 27 A D Error    Something is wrong with the A D converter on the CPU board    Solution    Access Service     Self test     Circuit  Test the A D interrupt    The error will be cleared if the test result is normal  Otherwise  call Mindray  Customer Service Department or the distributor     7 2 28 Vacuum Error    The
7.   normal sample        BC 2800 Auto Hematology Analyzer  V1 1  4 5    Histograms and Pulse Graphs    priests   1270885 omet poma   MDM    a ap arabe  Ehi  aos page  O a a peer  ONTE oe  Figure 4 24 Interference occurs because  gain of PLT channel is too large and data  length increases  background     whe leagh   430038 Cure positon   7168       heute al      Press  1  to show rube   r  Jiro  to scrad   page   T1  00 sand 0 piges  ENTER  te ext    Figure 4 26 Slight interference in WBC  channel and data length does not increase  obviously  normal sample     ste lenge   41953 curent ponon   233064       Page  2  t show sier   JE  to porot 1 page   TI  to swot 30 pages   ENTER  i gt  enit          Figure 4 28 Slight interference in RBC  channel and data length does not increase  obviously  normal sample           p   lengh   moaca         current ponina   730618       JL Ju    Neve R  a show rin  GF  te emelt pags   IAE te etre 8D pages   ENTER  10 ait    Figure 4 25 Interference occurs because  gain of PLT channel is too large data  length increases  normal sample     pt lengh   aao curet portion   730518    j J r    were F  show rin    J  te emelt pags  N te rss pages   ENTERI at       na LaMar mee    Figure 4 27 Slight interference in PLT  channel and data length does not increase  obviously  normal sample     whe kagi   4008000 currest posica   3072    4       Press  8  to show nda i      to ron   page   T  L  te perot 0 pagea  ENTER  to ext    Figure 4 29 Inadequate or no 
8.   to check the events one by one or press  PgUp  or  PgDn  to check the events  on the previous or next screen  If you want to print out the displayed events   press  PRINT   If you want to acquire help information  press  HELP      For every recorded event  the NO  column displays the sequences of the  recorded events  the Time column displays the time when this event occurred   the Type column displays the event type  the error events are also marked the  corresponding error codes that are explained in Chapter 7 1   the Times  column displays how many times  1  255  this event occurred and if it occurred  more than 255 times  the excessive events will be recorded from 1 to another    log file  the Information column displays extra information regarding the event     This analyzer can save maximum 1000 log files and once the maximum number  has been reached  the newest log will automatically cover the oldest one        6 24    BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 6 System Configuration    To view the system configuration  press  MENU  to enter the system menu  and  then press the appropriate arrow keys to move the cursor to Service    Config     Press  ENTER  to enter the Config screen  as Figure 6 25 shows  where you  can only view the system configuration without changing it        1 0 0  Ac 000000  Storage capacity 10000    Discriminator Displayed    All info    Expiration date  Clean condition  PS 2 keyboard  Edit sample infa  Printer  Transmiss
9.   to control    the three switches that respectively control the aperture zapping  current  source and HGB LED     Others    The Super I O outputs 2 control signals to control the photo couplers of the  volumetric metering board and the buzzer of the keypad        BC 2800 Auto Hematology Analyzer  V1 1     Hardware       1 7 Driving Board    The driving board controls and drives the pumps  valves  and motors     1 7 1 Basic Functions    The driving board drives the valves  pumps and motors  It carries out the  following instructions sent by the CPU  to open close the pumps or solenoid  valves  to control the motors of the syringes  to control the movement of the  sample probe  to remain the torques of the motors when the analyzer has  entered the screen saver     1 7 2 Basic Blocks    The driving board mainly consists of a power block  switching block and motor  control block  See the figure below for the positions of each block of the PCBA     1 7 2 1 Power block  The power block includes a 5V  12V and 30V DC supply  The 12V and 30V  supply comes from the power interfaces  where two LEDs are installed to  respectively indicate the whether the 12V or 30V supply is connected  When  the LED is on  it indicates the corresponding power has been connected to the  driving board  The MC7805T converts the received 12V supply into the 5V  supply  as the following figure shows        SV  ts MC7805T             gt              Figure 1 7 How the 5V power is obtained    1 7 2 2 Switching
10.  2004 13 55 Lymph  H 41 0     i i  Mid  8 0    WB   Al 1 f  Mode B   All Gran  510  0 oo cam  300 fL  HGB 124 g L  Name  RBC 4 08 x 1012 L  HCT L 36 0    V 88 3 fL       Count       Review Maintenance         Quality control                       Setup  gt   Self test x 10   L       Log fL T     Calibration Config   10 15 20 25 fL  Shutdown Print          Reagents for   1 1  7 Select sub menu   ENTER  Enter sub menu       Figure 6 16 Entering status screen    Press  ENTER  to enter the Status screen  as Figure 6 17 shows         l tem  Envir  Temp    C   HGB zerolv   HGB blarik v     Vacuum                   Reagents for          Figure6 17 Status screen       BC 2800 Auto Hematology Analyzer  V1 1  6 17    Maintenance       Note that you can only view the displayed status items without changing them  If  any of the displayed item exceeds the given range  see Chapter 7    Troubleshooting for solutions     6 4 System Self Test    The system self test is a major way to locate system errors  Follow the  instructions given below to view and check the available self test items     Press  MENU  to enter the system menu and press the appropriate arrow  keys   t         lt       to move the cursor to Service     Self test  as Figure 6 18  shows                tem vValue   Range    Envir  Temp   C  26 0 15 0   35 0  HGB zerolv  0 01 0 00   0 20  HGB blarik v  4 51 3 40   4 80  Vacuum 187 175   205           Review Maintenance  Quality contro  gt   Status       og          Calibration 
11.  37 1      Dg 50 250 fL    Quality contro Status            Setup  gt   Self test  Calibration  gt   Config 5 f  Shutdown Print                 gt 99 Counts    Figure 6 1 Entering maintenance screen    Press  ENTER  to enter the Maintenance screen  as Figure 6 2 shows             tem   Item       Rinse prime Clean bath   Lyse prime Drain bath   Zap aperture Drain tubing  Flush aperture Clean wipe black  Probe cleanser cleaning Prepare to ship  E Z cleanser cleaning       Prime the tubing with diluent                   Figure 6 2 Maintenance screen    If you want to exit this screen  press  MENU  to enter the system menu and access the  desired screen from there        6 4 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 2 1 Diluent Prime    Biohazard      Wear standard laboratory attire  including rubber gloves  and  D4 follow safe laboratory procedures when handling any material in    the laboratory        You should perform the Diluent prime procedure to prime the diluent tubing  with diluent when   there are bubbles in the tubing  or   the diluent in the tubing is contaminated  or   the old diluent ran out and a new container of diluent is installed     Follow the steps given below to do so   At the Maintenance screen  press the appropriate arrow keys    t         lt         to move the cursor to Diluent Prime    Press  ENTER  to prime the tubing with diluent and the priming progress will be  displayed at the bottom of the screen  as Figure 6 3 shows   When t
12.  50uL 10mL  Waste exit  L l l    Probe  Wipe  Syringe  assembly  Sample  probe          _  gt           Figure2 5 Washing module  washing of sample probe     2 4 3 2 Washing of the front and back baths  The part that washes the baths are shown in Figure2 6  The washing consists of  two procedures  washing the front bath and washing the back bath  The front  bath is washed by diluent and the back bath by rinse  The diluent and the rinse  are different and cannot replace each other        2 8 BC 2800 Auto Hematology Analyzer  V1 1     Fluidic System                   Yo  Aerial dust             50uL 10mL          Dispensing  tube             Syringe  a assembly                                       tube    o    l       l x     Metering  l   l   l   l    EES E               Negative  pressure chamber    Rinse    entrance  aa        lt  lt     ____   _ Oo j    a o                 Figure2 6 Washing module of fluidic system  washing of front and back baths      2 4 4 Hydraulic module    The hydraulic module is shown in Figure2 7  This module serves to establish the    vacuum and the positive pressure     To establish the positive pressure  When both valve 1 and the pump are open   the pump sucks atmospheric air into it  through the upper port and the NC end of  valve 1  The sucked pressure is stored in the pressure chamber to establish the  positive pressure  The pressure value is monitored by the pressure sensor    To establish the vacuum  When both valve 2 and the pump are open  
13.  7    Maintenance       to move the cursor to Zap aperture      Press  ENTER  to zap the aperture and the zapping progress will be displayed at  the bottom of the screen  as Figure 6 6 shows      When the zapping is done  the screen will return to the initial state           Diluent prime Lyse test  Rinse prime Clean bath    Lyse prime Drain bath    Zap aperture Drain tubing    Flush aperture Clean wipe block  Probe cleanser cleaning Prepare to ship    E Z cleanser cleaning       Zapping aperture  Zap aperture                  Select service       Figure 6 6 Zapping aperture    6 2 5 Flush Aperture    You can perform the Flush aperture procedure to assist zapping the aperture     Follow the steps given below to perform the procedure      At the Maintenance screen  press the appropriate arrow keys    t         lt            to move the cursor to Flush aperture      Press  ENTER  to flush the aperture and the flushing progress will be displayed  at the bottom of the screen  as Figure 6 7 shows      When the flushing is done  the screen will return to the initial state        6 8 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance             Diluent prime Lyse test  Rinse prime Clean bath    Lyse prime Drain bath    Zap aperture Drain tubing    Flush aperture Clean wipe black  Probe cleanser cleaning Prepare to ship    E Z cleanser cleaning       Flushing aperture                Figure 6 7 Flushing aperture    6 2 6 Probe Cleanser Cleaning    Biohazard  Wear standard laborat
14.  Fluidic System    2 1 Fluidic system    prepares diluent for the whole blood and prediluted modes    counts blood cells and measures HGB concentration    1  2  3  dispenses diluent  4 implements the flush  startup and cleaning cycles  5    controls vacuum pressure       BC 2800 Auto Hematology Analyzer  V1 1  2 1    Fluidic System    2 2 Construction of Fluidic System    The fluidic system consists of a sample probe assembly  syringe assembly  pump    assembly  volumetric metering unit  right valve panel and front panel     2 2 1 Sample Probe Assembly             a  OS No  Name  1 Rotation motor  D IN  a r  i 5 gt  Lae 2 Probe wipe       3 Probe holder    4 Sample probe       5 Elevator motor       6 Upper photo coupler       Sleeve       Left and right          xe    es  Le A photo couplers  9 Retaining rin  Figure 2 1 Sample probe assembly AAN anng  10 Mount                2 2 2 Syringe assembly  2 2 3 Pump Assembly  2 2 4 Volumetric metering unit    2 2 5 Right valve panel and front panel    See Chapter 3 Disassembling Instructions        2 2 BC 2800 Auto Hematology Analyzer  V1 1     Fluidic System       2 3    Composition of Fluidic System    The fluidic system consists of the following subsystems  sensor subsystem  bath  subsystem  lyse dispensing and mixing subsystem  diluting subsystem   volumetric metering subsystem  vacuum subsystem  pressure subsystem and  auxiliary subsystem    The key components of the fluidic system are the solenoid valve  syringe   aperture  s
15.  O RINGE       TR6A 20 09940    Recorder door       TR6D 30 16662    TR60 D drive board       3001 10 07048    Rotation motor       3001 10 07049    Elevator motor       3001 10 07059    Sample probe       3001 10 07252    Diaphragm pump       3001 20 07072    Transformer       3001 20 07245    Three way valve  ASCO        3001 20 07246    Two way valve  ASCO        3001 30 06880    Sample probe assembly       3001 30 07021    Vacuum chamber assembly       900E 10 04913    Inverter       3001 10 07068    Tubing ID1 16     OD1 8           M90 100035       Tubing IDO 02  ODO 06           M90 10007 1       Tubing ID3 32     OD5 32                 8 2    BC 2800 Auto Hematology Analyzer  V1 1        Appendix BC 2800 Error Message    CODE Error Message Information    0401 Envir  Temp  Abnormal  0402 Background abnormal    HGB error   HGB adjust   WBC clog   WBC bubbles   RBC clog   RBC bubbles   Communication error   Scanner error   Scanner communication error   Printer out of paper   Printer connection error  Recorder communication error   Recorder out of paper   Recorder too hot   Press bar up   Lyse out   Diluent expired   Rinse expired   Lyse expired   Filer error   Real time clock error   Syringe motor error  Rotation motor error   Elevator motor error   A D error   Vacuum error   Press error       400B Diluent out  400D Waste full    8002 Dynamic memory error    8001 File error    P N  2800 20 28832  1 1     
16.  See the block diagram  above for details        BC 2800 Auto Hematology Analyzer  V1 1     Hardware       4  Drive signal detection circuit of the position transducers    The control system judges the motor positions by the signals sent by the  position transducers  photocouplers   The photocouplers are driven by the  MCU through a 74LS07  The photocoupler sends the position signals to the  MCU through a 74LS14  inverter   See the figure below for the block diagram  of the position detecting circuit  The photocouplers are installed on the sample  probe assembly and the syringe assembly and feed the control and feedback  signals to the driving board through cables   1 7 2 4 Testable Signals   To test the signals  connect the grounding terminals of the oscillator and the  multi meter to the DGND or the PGND     The testable signals are  control signals of valves and pumps  sequence  signals of the motors  valves  position signals sent by the position transducers   serial communication signals  reset signals and voltage signals of the power  supplies     BC 2800 Auto Hematology Analyzer V1 1  1 13    Hardware    1 8 Display Unit    1 8 1 Function of the Adapter    The LCD adapter connects the LCD to the CPU board        9 interface board     _        gt  connecting cables    CPU board    Figure 1 9 Connection Schematic    1 8 2 Introduction of the adapter    The adapter incorporates two FPC FFC connectors  J2 and J3  The J3 is for the  BC 2800 display while the J2 is reserved f
17.  X4 and X2 are external crystal oscillators whose frequencies are  45MHz  45MHz and 24MHz respectively  The clock output of the CPU  BCLKO   is main clock signal of the CPU board        1 4 BC 2800 Auto Hematology Analyzer  V1 1     Hardware       1 4 CPU and Peripheral Devices                 DRAM   CRTL    Clock Driver    SRAM  BUS CONTROL    DRC                         gt  BUS Interface               Figure 1 4 CPU and peripheral devices    1 4 1 CPU     gt  The CPU is MOTOROLA MCF5307  external frequency 45MHz  operation  frequency 90MHz  processing speed as high as 75MIPS      gt  The MCF5307 features a 32 bit data bus and a 32 bit address bus  The board  uses a 24 bit addressing mode  reserving the most significant 8 bits as the  general purpose I Os for the FPGA     gt  The MCF5307 can be tuned through the BDM port  J18 of the CPU board      gt  The CPU board utilizes the built in   C and UART controllers of the MCF5307  to use the EEPROM and RTC as expanded serials ports     gt  The CPU boards utilizes the built in DRMA controller of the MCF5307 to use  the 2x8M SDRAM as the expanded memory     1 4 2 WDT    The Watch Dog Timer  WDT  is Tl TPS3828  It monitors the running of the  software  The CPU must send a feedback to the WDT every 1 6s  otherwise  the WDT will force the CPU to restart     BC 2800 Auto Hematology Analyzer V1 1  1 5    Hardware    MCF5307    PPY  yJ nRSTI       Figure 1 5 WDT    1 4 3 FLASH    The FLASH is TE28F160 2M bytes    The boot program is sto
18.  analyze the reason leading to the fault by viewing these original data   Common pulse graphs  Enter password    3210     after a count  you can view the WBC pulse graph of this  count by pressing    1    and RBC pulse graph by pressing    2    and PLT pulse graph  by pressing    3     Press    ENTER     to exit   When the instrument is working normally  the length of pulse data is related to  the concentration of the blood sample  The length of the pulse data should be  within a limit range  For general samples  the range should be  WBC   lt  1M  RBC   lt  600K PLT   lt  1M  Data length of abnormal sample will not lie in this range   Length of normal level controls data should be   WBC  400  700K RBC  250   450K PLT  300   600K  Normal pulse graphs         Ny TF ii     HAT ilh H     sah N MND N Ni i N li HL Mh          A   Nays    M Areeni f Kal  l tel AN   IAR MA   4  VINN   UAL   f Fress  a to sus    i  oe cnr                Figure 4 12 WBC pulses Figure 4 13 PLT pulses          rhe lengh   119549 asma penbon  Te vecin      ne Curemtpomma  3022    Te AN WH A                  i pag  TTA     vert a    Figure 4 14 RBC pulses Figure 4 15 Normal WBC background  pulse       BC 2800 Auto Hematology Analyzer  V1 1  4 3    Histograms and Pulse Graphs       Pee    Press B  mow ron  99  3  to show   P  Ju paoti pp 4  w ema o a T we  lew i o  0 scrol   page   TI  to sorot JOpazes  ENTEN torm EES             Figure 4 16 Figure 4 17  Pulse of normal RBC background check Pulse of normal 
19.  block  The switching block mainly consists the photocoupling circuit and driving  circuit of the pumps and valves  as the figure below shows          Photo coupling  Isolation Drive circuit  circuit    Control Drive    signal signal    Figure 1 8 Switching control circuit    1  Photocoupling circuit    The photocoupling circuit mainly consists of the photocoupler and resistors  It  provides 16 TTL outputs to the valves and pumps  The photocoupler     BC 2800 Auto Hematology Analyzer V 1 1  1 11    Hardware    2     TLP521 2  isolates the digital ground from the driving ground    Driving circuit of valves and pumps   The driving voltage of the valves and pumps are 12V  TTL   The circuit mainly  consists of ULN2068  In the BC 2800  the circuit can maximum drives 14    valves and 2 pumps  The fluidic system decides which pumps or valves are to  be actually used     1 7 2 3 Motor control block    The motor control block includes  serial communication circuit  control drive  circuit of the sample probe mechanism  control drive circuit of the syringe  motor  and drive signal detecting circuit of the position sensors       Serial communication circuit    Since the CPU board requires a 3 3V power supply while the driving board  requires a 5V power supply  a photocoupler  H11L1  is needed for the  purposes of conversion and isolation       Control Drive circuit of sample probe mechanism    The control Drive circuit of sample probe mechanism includes the control drive  circuit of the 
20.  gt   Config    Shutdown             Print    Figure 6 18 Entering self test screen    Press  ENTER  to enter the Self test screen  as Figure 6 19 shows  If you want  acquire help information  press  HELP         6 18 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance                Tubing   Result  O Machine i    O Valve          Circuit                                7       Figure 6 19 Self test screen    The available self test items are divided into four groups  the Tubing  Machine   Valve and Circuit  as shown on the left of the screen  You can switch among  the groups by pressing  F1  and once a group is selected  the included test  items  together with their test results  if available   will be displayed on the right  of the screen  You may press  PRINT  to print out the displayed results  Follow  the instructions given below to conduct every test and if you want to acquire  help information  press  HELP      6 4 1 Testing Tubing    At the self test screen  press  F1  to select the Tubing group  as Figure 6 19  shows  The following test items will be displayed on the right of the screen     Count Time    It measures the duration of a WBC and RBC count  namely how many seconds  it takes for the aspirated fluid flows from the first sensor to the second     Aperture v     It measures the voltage  v  over the aperture     Vacuum    It checks whether the vacuum system functions normally     Pressure    It checks whether the system flushes the aperture at a normal press
21.  may press  ENTER  to  stop the process any time  Note that a shortened soaking process may not be as  effective as a complete one      When the soaking is done  press  ENTER  to flush the bath and tubing  after  which the screen will return to the initial state    If you analyzer has been running continuously for 24 hours  a dialog box will pop  up to remind you to perform the Probe cleanser cleaning procedure  If you want  to do so  move the cursor to Yes and press  ENTER   Otherwise  move the cursor  to No and press  ENTER      6 2 8 Lyse Test    Biohazard  Wear standard laboratory attire  including rubber gloves  and    follow safe laboratory procedures when handling any material in  the laboratory        In case of any abnormal WBC counts or histograms  you can perform the Lyse  test procedure to check whether the lyse can be dispensed properly     Follow the steps given below to do so     Unscrew and remove the retaining screws with hands or screwdrivers  pointed  by the arrows shown in Figure 6 9  on the right plate     SMITH NMHYS     WA Cx             ZA    Figure6 9 Removing the two screws      Follow the arrow shown in Figure 6 10 to push and remove the right plate        BC 2800 Auto Hematology Analyzer  V1 1  6 11    Maintenance          Figure 6 10 Removing right plate      Remove the screws fixing the shielding box of the bath  as Figure 6 11 shows        Figure 6 11 Shielding box      Remove the shielding box to expose the bath  as Figure 6 12 shows        6 1
22.  so  move the cursor to Yes and press  ENTER   otherwise  move the cursor to  No and press  ENTER         Whole Blood AlIl    Current sample WBC L 0 3 x 10L  Lymph  wee 1 AL     Soha ras Midi reese x 10   W    21 Gran  see DAL    Time  01 28 2004 00 17 Lymph  t z       Mid       fe 0 100 200 300 fL  Gran  OR    Mode   WB   All    Name    Gender       Ape 150 250 fL  Next sample ROW CVY H 23 59    ROW SD 44 9 fL 5  a  n PLT H 443 x 10L    Wwe MPY 9 4 fL  N WB   All  inan B PDW H 17 4  PCT H 0 416   0 5 10 15 20 25 fL             Figure 6 8 Recommending probe cleanser cleaning    6 2 7 E Z Cleanser Cleaning    Biohazard    om  A Wear standard laboratory attire  including rubber gloves  and    follow safe laboratory procedures when handling any material in  the laboratory        You can use the E Z cleanser  an enzyme based  isotonic cleaning solution and  wetting agent  to clean the tubing and bath by performing the E Z cleanser  cleaning procedure     Follow the steps given below to perform the procedure     At the Maintenance screen  press the appropriate arrow keys    f      J  lt         to move the cursor to E Z cleanser cleaning     Present the cleanser to the probe and press  ENTER  to aspirate the cleanser   Remove the cleanser after the probe has risen up  This analyzer will  automatically soak the bath and tubing with the aspirated cleanser        6 10 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance         The soaking process will last about 10 minutes and you
23.  system does not reach the expected vacuum within the given time   Solution    Access Service     Self test     Tubing and test the vacuum  The error will  be cleared if the result is normal  otherwise  call Mindray Customer Service  Department or the distributor     7 2 29 Pressure Error    The vacuum chamber does not reach the expected pressure within the given  time    Solution    Access Service     Self test     Tubing and test the pressure  The error will  be cleared if the result is normal  otherwise  call Mindray Customer Service  Department or the distributor        7 8    BC 2800 Auto Hematology Analyzer  V1 1     Troubleshooting       7 2 30 Diluent Out    Possible causes  insufficient diluent or wrong diluent volume setting   Solution    Check if there is sufficient diluent left  If so  access Setup     Settings      Reagents and adjust the remaining diluent volume  if not  change a new  container of diluent     7 2 31 Rinse Out    Possible causes  insufficient rinse or wrong rinse volume setting    Solution    Check if there is sufficient rinse left  If so  access Setup  gt  Settings      Reagents and adjust the remaining rinse volume  if not  change a new container  of rinse     7 2 32 Waste Full    The waste container is full   Empty the container  or change a new container to receive the waste and re set  the waste container volume     7 2 33 File Error    Something is wrong with file saving   Turn off the analyzer and call Mindray Customer Service Department o
24.  up from the bottom of the bath and the mixing is done  thus    The on off interval of the valve is critical to effect of the bubble mixing  Either too  many or too few bubbles will affect the mixing  During the mixing process  the  airway should be well drained  or the trapped liquid will affect the quantity of the  bubbles as well as the dilution        2 10    BC 2800 Auto Hematology Analyzer  V1 1     Fluidic System       Positive pressure         Bath  assembly          Negative pressure  chamber    Figure2 8 Mixing module  2 4 6 Waste discharging module  The waste discharging includes  discharging waste of the probe wipe   discharging the waste of the bath  discharging the waste of pressure chamber   As Figure2 9 shows  once the negative pressure is established inside the  pressure chamber  valve 11 will open to discharge the waste of the bath to the  outside through the pressure chamber and the pump  Likewise  the waste of the  probe wipe is discharged to the outside when valves 9 and 10 are open        Bath  gt   assembly Negative pressure    chamber    Figure2 9 Waste discharging module       BC 2800 Auto Hematology Analyzer  V1 1  2 11    Fluidic System    2 5 Counting Procedure    2 5 1 Dilution    Usually in blood samples  the cells are too close to each other to be identified or  measured  For this reason  the diluent is used to separate the cells so that they  are drawn through the aperture one at a time as well as to create a conductive  environment for blood a
25. 0 fL   PLT  2 30 fL       BC 2800 Auto Hematology Analyzer  V1 1  2 13    Disassembling the Analyzer       Chapter3 Disassembling the Analyzer    3 1 Main Unit Structure    3 1 1 Main user interfaces                   Front view   1      LCD  2      Keypad    3      Recorder  4      Power Indicator    5       START  key  6      Sample Probe          Figure 3 1 Front view  N   m 5     p  lt 2 on an an a Rear view           1     Power switch  r   m 2     Equipotentiality  3     Waste Outlet  Red   4     Lyse Inlet  Orange      a 5     Rinse Inlet  Blue   6     Diluent Inlet  i i  Green   ob eri  A     d L    Figure3 2 Rear view             BC 2800 Auto Hematology Analyzer  V1 1  3 1       Disassembling the Analyzer          Left view                 PP    1     RS232 Port1   for the scanner   2     Parallel Port  3     RS232 Port 2   for the computer   4     Power Interface of Floppy  Disk Drive  5     Keyboard Interface                   Figure3 3 Left view       3 2    BC 2800 Auto Hematology Analyzer  V1 1           Disassembling the Analyzer  3 2 Disassembling the Main Unit    3 2 1 Removing the left  right and top covers    As the figure below shows  remove the retaining screws  two for each cover  with  a Philips screwdriver and pull the covers away        4 ep Pra j      P  i p   a l     l A  B  gt A      Gi              0    eae    J  2        q co  lt n  7 W J        iN     Z    Figure3 4 Removing the left  right and top covers  3 2 2 Removing the back plate and the f
26. 2 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 2 9       Figure 6 12 Bath    At the Maintenance screen  press the appropriate arrow keys    t         lt          to move the cursor to Lyse test   Press  ENTER  and this analyzer will automatically drain the bath and then  dispense 2ml lyse to the bath    Check the scale to see whether the lyse has reached the expected line  the  second from bottom   In case the dispensed lyse has failed to reach this line for  several consecutive times  check whether the lyse has run out or the lyse tubing  is not properly connected to this analyzer  If the lyse is still enough and the  tubing is well connected to the analyzer  contact the Mindray Customer Service  Department for repair    Press  ENTER  to flush the bath and finish the test  The screen will return to the  initial state     Clean Bath    When you keep finding abnormal results from the background check  follow the   steps given below to perform the Clean Bath procedure    At the Maintenance screen  press the appropriate arrow keys    t         lt         to move the cursor to Clean bath    Press  ENTER  to start the procedure    When the cleaning is done  the screen returns to the initial state     6 2 10 Drain Bath    The Drain Bath procedure checks whether the analyzer can drain the bath  within the given time  You can perform this procedure when you find three or  more of the HGB  WBC  RBC and PLT results are abnormal     Follow the first 4 steps of performi
27. 2051 whose resetting time is 470ms  It uses a 11 0592MHz  crystal oscillator    1 9 2 3 Keypad scanning  The keypad matrix is 5X4 one  incorporating 9 I O wires and 20 keys  Note that  the keys at line 5 and columns 3 and 4 are not used    1 9 2 4 Backlight control  The keypad adapter shuts off the backlight and blinks the power indicator when  instructed by the main board to do so  usually after the analyzer entered the  screen saver   The backlight board uses an independent 12V power supply  and receives the control signals through photocouplers  The transistor is used  to help control the LED so that the power indicator can be turned on even when  something is wrong with the MCU        1 16 BC 2800 Auto Hematology Analyzer  V1 1     Hardware          YOO       Figure 1 11 Control of LCD brightness    The LCD brightness is controlled by pot RV1  Adjusting the RV1 can force the  VBL to change within 0 5 3V  The voltage change is fed into the inverter and  causes the change of the drive current and hence the change of the brightness   Note that the smaller the voltage and the brighter the LCD     1 9 2 5 Buzzer control    The buzzer is controlled by a DC signal  5V DC  current lt 40mA   The 5V supply  of the buzzer is isolated from the VDD and the control signal is received through  a photocoupler  TLP521 2  that is controlled by a current around 10mA        AAT HALTI    Figure 1 12 Buzzer control    BC 2800 Auto Hematology Analyzer V1 1  1 17    Fluidic System       Chapter2
28. 8 indicate the system status as defined by the  software     BC 2800 Auto Hematology Analyzer V1 1  1 7    Hardware    1 5 Analog Inputs and Outputs    1 5 1 Signals of Blood Cell Counts    The CPU board has three A D converters  U10  AD7928   U11 AD7908  and  U14  AD7908   Both the AD7928 and AD7908 feature 8 channel and 1MSPS   only the former is 12 bit converter and the latter 8 bit  The U10 is actually  installed and powered by a 2 5V supply  while the U11 and U14 are reserved   The sampling speed is set to 500KSPS     1 5 2 Signals of System Monitoring    The Super I O monitors such system status as the  48V   12V and  12V  supplies of the analog board  the  3 3V and  12V supplies of the CPU board  itself and the temperature of the whole analyzer     1 5 3 Signals of LCD Contrast    The Super I O generates PWM signals that are then integrated to output a  0 2 5V analog signal to control the LCD contrast  The user can adjust the  contrast through the software interface        1 8 BC 2800 Auto Hematology Analyzer  V1 1     Hardware       1 6 Digital Inputs and Outputs    1 6 1 Serial Port    The analyzer has 6 serial ports  which are illustrated in Figure 1 6                 COMO  CELUI ais   COM    COM    FPGA eer 2   COMS             ae COMSB  Super I O  COMA    Figure 1 6 Serial ports       The CPU incorporates 2 UART controllers  3 3LVTTL   one to control the motor  of the driving board and the other communicates with the recorder  powered by  5VTTL   The FPGA implemen
29. 845    Locking nut of 7 5ml syringe       2800 20 28846    Body of 50ul syringe       2800 20 28847    Nut of 50ul syringe       2800 20 28848    Bolt of 50ul syringe       2800 20 28849    Connection wire of inverter       2800 21 28728    Shielding box cover assembly       2800 21 28788    Shielding box assembly       2800 30 28650    CPU board       2800 30 28652    Analog board       2800 30 28654  2800 30 28664    Driving board  Keypad                BC 2800 Auto Hematology Analyzer  V1 1     8 1    List of Spare parts          2800 30 28668    Volumemetric metering board       2800 30 28670    Power board       2800 30 28715  2800 30 28716    Screen assembly  Sampling assembly       2800 30 28746    LCD adaptor       2800 30 28767    Bath assembly       2800 30 28778    Syringe assembly       2800 30 28779    7 5ml syringe assembly       2800 30 28780    50ul syringe assembly       2800 30 28789    Bath mount assembly       2800 30 28815    HGB mount assembly       2800 30 28817    CAP component for LYSE       2800 30 28818    CAP component for DILUENT       2800 30 28819    CAP component for RINSE       2800 30 28839    Waste tubing assembly       3001 10 07252    Vacuum Pump             3001 20 07247  Localizer  3001 21 07182  Cap for LYSE  3001 30 06957  Probe wipe       59BR 10 08830    Thermal head                Mg90 000130     O RING  6x1    M90 100054     O RING 2 00X1 80   M90 100069     P N X420 1 10 32 Special Tapered Thread To 3 32   Mg90 100101     Y shaped
30. ARTICULAR PURPOSE     Exemptions    Mindray s obligation or liability under this warranty does not include any transportation or  other charges or liability for direct  indirect or consequential damages or delay resulting  from the improper use or application of the product or the substitution upon it of parts or  accessories not approved by Mindray or repaired by anyone other than a Mindray  authorized representative     This warranty shall not extend to any instrument which has been subjected to misuse   negligence or accident  any instrument from which Mindray s original serial number tag or  product identification markings have been altered or removed  or any product of any other    manufacturer     Safety  Reliability and Performance    Mindray is not responsible for the effects on safety  reliability and performance of the   BC 2800 Hematology Analyzer if    M assembly operations  extensions  re adjusts  modifications or repairs are carried out  by persons other than those authorized by Mindray     E Personnel unauthorized by Mindray repairs or modifies the instrument     BC 2800 Auto Hematology Analyzer V1 1  III    Copyright       Return Policy    Return Procedure    In the event that it becomes necessary to return a unit to Mindray  the following procedure  should be followed     1  Obtain return authorization  Contact the Mindray Service Department and obtain a  Customer Service Authorization  Mindray  number  The Mindray number must appear  on the outside of the sh
31. BC 2800  Auto  Hematology    Analyzer    Service Manual    Copyright       Copyright  Statement    SHENZHEN MINDRAY BIO MEDICAL ELECTRONICS CO   LTD   hereinafter called  Mindray  owns all rights to this unpublished work and intends to maintain this work as  confidential  Mindray may also seek to maintain this work as an unpublished copyright   This publication is to be used solely for the purposes of reference  operation  maintenance     or repair of Mindray equipment  No part of this can be disseminated for other purposes     In the event of inadvertent or deliberate publication  Mindray intends to enforce its rights to  this work under copyright laws as a published work  Those having access to this work may  not copy  use  or disclose the information in this work unless expressly authorized by  Mindray to do so     All information contained in this publication is believed to be correct  Mindray shall not be  liable for errors contained herein nor for incidental or consequential damages in  connection with the furnishing  performance  or use of this material  This publication may  refer to information and protected by copyrights or patents and does not convey any  license under the patent rights of Mindray  nor the rights of others  Mindray does not  assume any liability arising out of any infringements of patents or other rights of third  parties     Content of this manual is subject to changes without prior notice     PROPERTY OF  SHENZHEN MINDRAY BIO MEDICAL ELECTRONICS 
32. CO   LTD   ALL RIGHTS RESERVED    Responsibility on the manufacturer party    Mindray is responsible for safety  reliability and performance of this equipment only in the   condition that    e all installation  expansion  change  modification and repair of this equipment are  conducted by Mindray qualified personnel    e applied electrical appliance is in compliance with relevant National Standards     e the instrument is operated under strict observance of this manual     BC 2800 Auto Hematology Analyzer V1 1       Copyright       A Note AY    This equipment is not intended for family usage     This equipment must be operated by skilled trained clinical personnel     A Warning A    It is important for the hospital or organization that employs this equipment to carry  out a reasonable maintenance schedule  Neglect of this may result in machine    breakdown or injury of human health        Il BC 2800 Auto Hematology Analyzer  V1 1     Copyright       Warranty  Workmanship  amp  Materials    Mindray guarantees new equipment other than accessories to be free from defects in  workmanship and materials for a period of one year from date of shipment under normal  use and service  Mindray s obligation under this warranty is limited to repairing  at    Mindray   s option  any part which upon Mindray s examination proves defective     THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES   EXPRESSED OR IMPLIED  INCLUDING WARRANTIES OF MERCHANT ABILITY OR  FITNESS FOR ANY P
33. Drain tubing    Flush aperture       Probe cleanser cleaning Prepare to ship       E Z cleanser cleaning    Cleaning wipe block                Figure 6 14 Cleaning wipe block    When the soaking is done  wipe the bottom of the wipe block with a probe  cleanser dipped cloth that does not leave debris    Press  ENTER  to flush the block and the inner surface of the probe    After the flushing is done  the screen returns to the initial state    When the accumulated analyzed samples reach 4 000  a dialog box will pop up  to remind to clean the wipe block  You may move the cursor to Yes and press   ENTER  to start the cleaning  or to No and press  ENTER  to ignore the  message        BC 2800 Auto Hematology Analyzer  V1 1  6 15    Maintenance       6 2 13 Preparing to Ship    If this analyzer is not to be used for over two weeks  or is to be shipped  perform  the Prepare to ship procedure to flush and drain it     Follow the steps given below to do so    Remove the diluent  rinse and lyse tubing from their containers    At the Maintenance screen  press the appropriate arrow keys    t J       lt         to move the cursor to Prepare to ship    Press  ENTER  and a dialog box will pop up to ask you to confirm this operation   as Figure 6 15 shows                  Diluent prime Lyse test  Rinse prime Clean bath  Lyse prime Drain bath  Zap aperture   Flush aperture         Probe cleanser c   7  Prepare to Sip        E Z cleanser cle          Remove the tubin e and lyse containers      
34. PLT background check       4 4 BC 2800 Auto Hematology Analyzer  V1 1     Histograms and Pulse Graphs       Abnormal pulse graphs    Carer postion   15306              irene tinea timate    Pers F  t show nier  Pp to enog  page  TILI     eroe 20 pages   ENTIM  iora       Figure 4 18 Severe interference in WBC  channel and data length increases  obviously  background     ureri potice   13324    Press  7  te stow nas  Pp 80  scre8   page   T 1   eros 28 pages   ENTER  to ext       Figure 4 20 Severe interference in RBC  channel and data length increases  obviously  background     PEER E TA TTET    Paese F  to show rder  P   io smd  p    Figure 4 22 Severe interference in P  LT channel and data length increases  obviously  background        whe length   7205314    Cures poston   15640      T            LW  J o J  Press  7  le stow rer   lonni paga  MEL poras    papa  OTER po me    4          Figure 4 19 Severe interference in WBC  channel and data length increases  obviously  normal sample     st ength  1123968 current potna   T680    Vuh    i   da   j  Treva fi  abow ruler    CE  be enoa  paga  1I to rE 0 pages   ENTER 10 w          Figure 4 21 Severe interference in RBC  channel and data length increases  obviously  normal sample     pt leh   261000 eurez pores   128000                  4 j   Villesi   ENA bu  Pme F  t show rder   P  Tor  to panogi pagr  TI  to cre 0 papra   ENTEN IO exe       Figure 4 23 Severe interference in PLT  channel and data length increases  obviously
35. ample probe  pump  bath   metering tube  probe module   negative positive pressure chamber and hose    The solenoid valve is ASCO458  Totally 11 valves are used  7 three way valves  and 4 two way valves    The two syringes  10mL and 50uL  are made by Mindray    The aperture is O80    The sample probe has two layers and its surface is polished  making the probe    easy to clean        BC 2800 Auto Hematology Analyzer  V1 1  2 3    Fluidic System    2 4 Functional modules    The fluidic system can be divided into the following functional modules   aspiration dispensing module  counting module  washing module  hydraulic  module  mixing module and waste discharging module  See the figure below for  the interaction of these modules     Reagent                          Washing  module module       Counting                Quantification                      Mixing             module Sample                Hydraulic  module                      discharging          Figure2 2 Interaction of the functional modules    2 4 1 Aspiration dispensing module    The aspiration dispensing module includes a motor that drives the 50 uL and 10  mL syringes  the former for aspirating dispensing the whole blood sample and  the latter for aspirating dispensing the prediluted sample  diluent and lyse  See  the figure below for details        BC 2800 Auto Hematology Analyzer  V1 1     Fluidic System             Buffer  tubing         j          C    NC N  c NC Cc NO c  CRO  NO NO    Buffer          tubi
36. ample probe if  you have replaced the wipe block  or observed motor error  or wrong test result  Also   as required by regular maintenance  you should use the localizer to adjust the  position of the sample probe monthly     Figure3 17 Localizer    3 2 15 Adjusting Sample Probe Position    Biohazard  Wear standard laboratory attire  including rubber gloves  and    follow safe laboratory procedures when handling any material in  the laboratory           3 10 BC 2800 Auto Hematology Analyzer  V1 1     Disassembling the Analyzer       Warning  The probe is sharp and may contain biohazard materials  including    controls and calibrators  Avoid any unnecessary contact with the  probe and probe area        Follow the steps given below to adjust the sample probe position     1  Refer to the steps mentioned in section 3 2 1 to remove the right cover of the    analyzer     2  Access the Setup    Password screen and enter the administrator password to    obtain the administrator authority  Access the Service   Self test screen     Press  F1  to select the Machine group and press  f  or     to move the cursor to  Elevator motor  as the figure below shows              Tubing I Item    Result     Machine      Syringe motor       O Valve n  O Circuit Rotation motor  Elevator motor    Print                                Select test  ENTER  Execute       Figure3 18 Accessing elevator motor    Press  ENTER  and an elevator motor screen will pop up  as the figure below shows        BC 2800 Au
37. boa   81920             Fe Bjo owner   el to scro8 8 page   TNL  be serol 50 pates   ENTER  9 ext       Figure 4 38 Interference in WBC channel  from tubing  Feature  data length  increases  the base line of signal is not  stable     p   lengt   2043000 curent pcsion   151000       Veen J    M  Vna plat anny      Iarr aati OO       Pass  3  to show eater   J   tose o   I page   TE    to srol 30 pages   INTER  io ent       Figure 4 37 Interference in PLT channel  from tubing  Feature  data length  increases  the base line of signal is not  stable     the length   2043000 carent porhor   1152000          Prese     to tow nda   m     80 srol   pare   TLL  80 scroll  pages  ENTER  to ext       Figure 4 39 Interference in RBC channel  from tubing  Feature  data length  increases  the base line of signal is not  stable        4 8 BC 2800 Auto Hematology Analyzer  V1 1     Adjustment       Chapter5 Adjustment    This chapter deals with gain adjustment and hardware adjustment     5 1 Gain Adjustment    5 1 1 Adjust the channel gains if you replace the below parts     the bath  s    the aperture  s    the analog board   the Disk on Module   Method   1  Using normal patient fresh blood or calibrator to check the gains   2  Control the gains through the digital potential meter     Replacement of the analog board will affect WBC  whole blood   WBC  prediluted    RBC    HGB  PLT results    Replacement of the bath s  will affect WBC  whole blood   WBC  prediluted    RBC  HGB    PLT   Repla
38. cement of the aperture s  or gasket  s  will affect WBC  whole blood    WBC   prediluted    RBC PLT    Enter the Password screen and enter the password     3210              Count    Review       Quality contro     Service    Calibration       Shutdown       PW level service    Press  MENU  to enter the system menu  move the cursor to Setup     Gain and  press  ENTER  to enter the Gain screen        BC 2800 Auto Hematology Analyzer  V1 1  5 1    Adjustment       Setup    Settings         Caunt  Review    Quality contro         Service         Calibration    Shutdown           Reagent      Print  amp  comm     Date  amp  time   2 Wain    gt  Count      Ref  range       Other                CoOo0e000    Reagent    Digipot       Print  amp  comm  Rate    Date  amp  time  Gain   Count   Ref  range  Other       100 0   100 0        WBC WB   WBC PB   REC   HGB                109 7              P  a gt    A gt                    HGB blank V                          MENU              7 Move cursor  T        Displayed values  current gains  Range  adjustable range  Rate  changing factor compared to the old value     Reagents for    Select setting item  F1 Select setting group         gt 99 Counts    WBC whole blood   gain of the WBC channel in the whole blood mode  WBC  prediluted   gain of the WBC channel in the prediluted mode  RBC  gain of the RBC channel  HGB  gain of the HGB gain  PLT  gain of the PLT channel        BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       Cha
39. creen       6 26 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       You can perform the following operations at the Print screen   Press  f  or     to select the desired task    Press  DEL  to delete the selected task    Press  HELP  to display the help information    Press  MENU  to return to the system menu        BC 2800 Auto Hematology Analyzer  V1 1  6 27    Maintenance       6 8 Adjusting Sample Probe and Replacing Probe Wipe    See Chapter 3 3        6 28 BC 2800 Auto Hematology Analyzer  V1 1     Troubleshooting       Chapter7 Troubleshooting    The chapter deals with the codes  possible causes and solutions of the errors  If the error  remains after you have tried the recommended method  call Mindray Customer Service  Department or the distributor  This chapter consists of two parts  the first part dealing with  the errors and assigned error codes and the second possible causes and recommended  solutions        BC 2800 Auto Hematology Analyzer  V1 1  7 1    Troubleshooting       7 1 Error Codes    The errors recorded in the log are presented in the error codes  See Table 7 1  for the correspondence between the errors and error codes     Table 7 1 Errors and codes    Environmental  Background    Temperature 0403   HGB error  Abnormal    0404   HGB adjust WBC clog 0406   WBC bubbles  0407   RBC clog RBC bubbles    Communication Scanner  0801 Scanner error 0803 ar  error communication error      Printer  Printer out of l Recorder  1001 connection 1003 PEON  pa
40. e accordingly  After the  adjustment  test the actual count time again and make sure the difference is  within 2 seconds     If the error still remains  call Mindray Customer Service Department or the  distributor     RBC Bubbles    This error message occurs when the actual RBC count time is less than the  preset RBC count time by 2 seconds   Possible causes     1  Insufficient diluent or rinse    2  Loose tubing connection    3  Inappropriate RBC count time setting    Solution    Check if the diluent or rinse is sufficient  If not  change a new container of  diluent of rinse    Check the tubing connections  If necessary  reconnect the tubing    If the error still remains  access Setup     Password to gain the administrator    authority and then access Setup     Settings     Count and adjust the RBC  count time     If the error still remains  turn off this analyzer and call Mindray Customer  Service Department or the distributor        BC 2800 Auto Hematology Analyzer  V1 1  7 5    Troubleshooting       7 2 9    Communication Error    The received communication settings are different with the BC 2800    Solution    Access Setup     Settings     Print  amp  comm  and change the communication  settings accordingly     7 2 10 Scanner Error    Too long or invalid bar code    Solution    Check and make sure the bar code is correct  If the error still remains  call  Mindray Customer Service Department or the distributor     7 2 11 Scanner Communication Error    Something is wrong w
41. e unclogging is successful and you can return to the Count screen to continue  the analysis  Otherwise  Access Service     Maintenance to soak the bath  and tubing with probe cleanser    When the soaking is done  access Setup     Settings     Count to note down  the preset WBC count time  Access Service     Self test     Tubing and test  the actual WBC count time  If the tested time differs from the preset time by less  than 2 seconds  it means the unclogging is successful and you can return to the  Count screen to continue the analysis  If the difference is still greater than 2  seconds and stabilized around a certain value  access Setup     Settings  gt   Count to change the WBC count time accordingly  After the adjustment  test the  actual count time again and make sure the difference is within 2 seconds     If the error still remains  call Mindray Customer Service Department or the  distributor        7 2 6 WBC Bubbles  This error message occurs when the actual WBC count time is less than the  preset WBC count time by 2 seconds   Possible causes   Insufficient diluent or rinse   Loose tubing connection   Inappropriate WBC count time setting   Solution   Check if the diluent or rinse is sufficient  If not  change a new container of  diluent of rinse   Check the tubing connections  If necessary  reconnect the tubing   If the error still remains  access Setup     Password to gain the administrator  authority and then access Setup     Settings     Count and adjust the WBC  co
42. eeeeeeeeeeeeeeeeeeeeees 1 16  C  hapter2 Fluidie Sy Stem iccastecnesecc  duase vacetdanbadetdevtacesdautccetdevtacesdautacetdevtecunds 2 1  2 1  Eluidie systemi ion glad te  d oe  oi bed oi edo hed a ed id bed od bans 2 1  2 2 Construction of Fluidic System              2 2     c   c ceeeeeeeeeeeeneeceeeeseeeeeeeceee 2 2  2 3 Composition of Fluidic System             eee eeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeee 2 3  2 4 Functional modules 3  c xp  gscceaccesargeassect chen saevasevsneveneetustausseuavenerdasteneesuenaad 2 4  2 5 Counting Procedure ein cesta eee eee ee 2 12  Chapter3 Disassembling the Analyzer                  eeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeneeeeees 3 1  3 1 Main Unit Structure oad oder  cnredereat rath Seid ted ideadeacesecttededeutsaceiachentedeutdacatatts 3 1  3 2 Disassembling the Main Uniit c         cece cee ete cette tceeeceeeeeeeceeenes 3 3  Chapter4 Histograms and Pulse Graphs                 ccccccccceeeeeeeeeeeeeeeeeeeeeeeeeees 4 1  4 1   Histograms   crit tees Gaeta tive lite nie hee i e 4 1  4 2 Pulse Graphs ienei aa a aa aaa adnate anaes 4 3  Chapters Adj  StMENt onene ek reee EEEE EEEE EEEE EEEE E Eeh 5 1  5 1 Gain Adjustment    nenene 5 1  Chapter6 Maintenance ss  cas ccacd2pnes at snsgiasdonnecntosnieaudtosei esol nelaeeeieeletens 6 1  6 1 Regular Maintenance ccs c cote 2c odeoccedeoricedessesstuaeducateadu cates steaduende 6 2  6 2 System MAINIOMANCE noieeoe eoero eardere ieaie ardea 6 4  Bs  System StatUS re ie aaa a e ad i dd Ded 6 17  0 4
43. elevator motor and that of the rotation motor  The control system  of the sample probe motor consists of a AT89S51 MCU and ADM705 WDT   The AT89S51 also detects the signals coming from the position sensor when  controlling the motors       Control drive circuit of the elevator motor    The circuit includes the control part  a MCU system  and the drive part  as  shown in the figure below     The MCU system provides the sequence signals for the elevator and rotation  motors and controls the position sensors  as the figure below shows  The MCU  reset signal  RST_XY  is active high    The drive part mainly consists of a control device  L6506   drive device  L298N   and follow current device  UC3610   The drive voltage is 30V  The sequence  signal and the enable signal of the drive device come from the MCU       Control drive circuit of the rotation motor    The circuit mainly consists of a control part  MCU system  and a drive part   Refer to the previous introduction for the MCU system  The drive part is the  ULN2068B and the drive voltage is 12V  The sequence signal comes from the  MCU  as shown in the figure below     Control drive circuit of the syringe motor   The circuit mainly consists of a control part  MCU system  and a drive part   The MCU is the P87LPC762 with built in WDT  The MCU system executes the  aspirating and dispensing operation of the syringe and detects the signals sent  by the position transducers     The drive part is similar to that of the elevator motor 
44. ells passing the aperture  The detected pulses are then sent to the analog  board to be amplified  rectified  recognized  adjusted and counted    When the sample is mixed in the bath  valve 2 is open and the negative pressure  chamber introduces the atmospheric air to pass through the filter and valve 2 and  then to flush the metering tube  Valve 1 is open when the vacuum has been  established and the sample  blood cells  in the bath is drawn through the  aperture by the negative pressure to generate the counting pulses  The sample  keeps moving to push the rinse between the back bath and metering tube to  move through the tube  When the rinse passes the upper optical sensor mounted  on the metering tube  a start signal is generated and sent to the analog board   which starts the counting right away  and when the rinse passes the lower optical  sensor  a stop signal is generated and sent to the analog board  which stops the  counting right away    Volumetric metering  the volumetric metering ensures a relatively objective and  stable analysis cycle    Monitoring of the counting time  the volumetric metering enables the monitoring  of the counting time  By monitoring the counting time  the system can easily  know whether the aperture is clean or blocked and feed this information to the  service personnel in terms of the aperture voltage so that they can service the       2 6    BC 2800 Auto Hematology Analyzer  V1 1     Fluidic System       analyzer in time   2 4 3 Washing mod
45. gurations are stored in the EEPROM  The VRAM is the memory for video  display     1 1 2 Basic Functions of the CPU Board    To receive such analog signals as the WBC RBC PLT counts  HGB  measurement  aperture voltage vacuum pressure signals  etc     To monitor such system status as the  48V   12V and  12V supplies of the  analog board  the  3 3V and  12V supplies of the CPU board itself and the  temperature of the whole analyzer     To receive the keypad signal and control the keypad buzzer and LCD backlight     BC 2800 Auto Hematology Analyzer V1 1  1 1    Hardware    To generate control signals to control the valves  aperture zapping  HGB LED   current source and digital pot     To drive and turn on the LCD and adjust the contrast     To drive the keyboard  printer and floppy drive        1 2 BC 2800 Auto Hematology Analyzer  V1 1     Hardware       1 2 Power Supply    The CPU board is powered by two independent external power supplies  a  5V  supply and a 12V supply  Two 5A fuses are respectively installed on the two  power entries  The  5V supply is converted a  3 3V supply to power the digital  components and the  3 3V supply is also further converted into a  1 5V supply  to power the FPGA  The  12 8V supply serves the CPU board only     Power socket       Figure 1 2 Power distribution of the CPU board    BC 2800 Auto Hematology Analyzer V1 1  1 3    Hardware    1 3 RTC       VARM         SDRAM       Synchronous clock    Figure 1 3 Arrangement of the CPU Clock    The X1 
46. he priming is done  the screen will return to the initial state                      p    ent prime Lyse test    Rinse prime Clean bath    Lyse prime Drain bath    Zap aperture Drain tubing  Flush aperture Clean wipe block  Probe cleanser cleaning Prepare to ship    E Z cleanser cleaning               Priming diluent  Diluent prime               Select service          Figure 6 3 Diluent priming screen       BC 2800 Auto Hematology Analyzer  V1 1  6 5    Maintenance       6 2 2 Rinse Prime    Biohazard  Ca Wear standard laboratory attire  including rubber gloves  and  D4 follow safe laboratory procedures when handling any material in    the laboratory        You should perform the Rinse prime procedure to prime the rinse tubing with  rinse when    there are bubbles in the tubing  or  the rinse in the tubing is contaminated  or  the old rise ran out and a new container of rinse is installed     Follow the steps given below to do so    At the Maintenance screen  press the appropriate arrow keys    t JE  J  lt         to move the cursor to Rinse prime   Press  ENTER  to prime the tubing with rinse and the priming progress will be  displayed at the bottom of the screen  as Figure 6 4 shows   When the priming is done  the screen will return to the initial state                  tem    Diluent prime Lyse test       Lyse prime Drain bath   Zap aperture Drain tubing  Flush aperture Clean wipe block  Probe cleanser cleaning Prepare to ship    E Z cleanser cleaning       Priming rinse 
47. hemolysis in  WBC channel and data length increases       4 6 BC 2800 Auto Hematology Analyzer  V1 1     Histograms and Pulse Graphs       pt length  2048000 arent porhon   4608    Ten       Figure 4 30 Too high sample  concentration in the RBC channel Figure 4 31 Too high concentration in the     it normally does not appear  PLT channel   it normally does not appear     wde length   2088000 Curent potion   564228 whe length   362076 cwrte potan   370668  wherval   Sou hgh   100    AAAI ANA A rae     Pues  i  D show ater DANI  te select rder   2  to Bade ndee   fji  te sero 1 page   T1  to srol 30 pages   ENTER  to est  hij tomore   opace   TL  to move 10 eposen  ENTER to est       Figure 4 32 Interference in WBC channel Figure 4 33 Measuring interference from  caused by inverter  Feature  sine wave inverter  with cycle of 20 26us    whe length   2379108 carent ponton   73767   rhc length   183886 Cusrest pontica   179712          AAAA aetna Aen nea tad hehe an ann ah tn no nen a a a    toler    f  te serol page   TK4  1o sered 30 pages   ENTER  1s est       Figure 4 34 Insufficient liquid in WBC bath Figure 4 35 Insufficient liquid in RBC bath  during count during count       BC 2800 Auto Hematology Analyzer  V1 1  4 7    Histograms and Pulse Graphs       phticngsh   739909 camem poston   204800     TT    N EEEE  fJ  te soroa   page   T11  to sarot 30 pages   DITER  io cut          Figure 4 36 Insufficient liquid in RBC bath  during PLT counting    p   kng   2233232 cune pon
48. ins  call Mindray Customer Service  Department or the distributor     7 2 18 Lyse Out    Possible causes  insufficient lyse or wrong lyse volume setting    Solution    Check if there is sufficient lyse left  If so  access Setup     Settings  gt   Reagents and adjust the remaining lyse volume  if not  change a new container  of lyse     7 2 19 Diluent Expired    The diluent has expired or its expiration date is not correctly set   Check whether the diluent has expired  If so  change a new container of diluent   if not  access Setup     gt  Settings     Reagents and adjust the expiration date     7 2 20 Rinse Expired    Possible causes  rinse has expired or its expiration date is not correctly set   Solution    Check the rinse  If it is expired  change a new container of rinse  if not  access  Setup     Settings     Reagents and adjust the expiration date     7 2 21 Lyse Expired    The lyse has expired or its expiration date is not correctly set   Check whether the lyse has expired  If so  change a new container of lyse  if not   access Setup     Settings     Reagents and adjust the expiration date     7 2 22 Filter Error    Something is wrong with the filter of the vacuum chamber    Solution    Access Service     Self test     Tubing and test the filter  The error is cleared  if the result is normal  If not  call Mindray Customer Service Department or the  distributor     7 2 23 Real Time Clock Error    Something is wrong with the clock   Access Setup     Settings     gt  Date
49. ion    Barcode                         Figure 6 25 Configuration    Every screen displays 13 items and you can press   t   or       to select the item  you want to see  or press PgUp  or  PgDn  to go to the previous or next screen   If you want to print out the configuration  press  PRINT   If you want to acquire  help  press  HELP         BC 2800 Auto Hematology Analyzer  V1 1  6 25    Maintenance       6 7 Printing Management    Press the  MENU  to enter the system menu and press the appropriate arrow  keys    t       i lt      to move the cursor to Service Print  as Figure 6 26  shows                 Software version 1 0 0   2 S N Ac 000000  Storage capacity 10000   gi Discriminator Displayed  Enter sample into All into   6 Search On          date       Review Maintenance        Status    Quality contro             Setup Self test    Og             Calibration Config          Shutdown       Figure 6 26 Entering print screen    Press  ENTER  to enter the Print screen  as Figure 6 27 shows  The printing  tasks are queued in this screen  where you can view the all tasks and delete  those waiting to be processed  Once something goes wrong with the printing  device  the task being processed will be deleted and the queued tasks will keep  waiting  Once the system finds the error has been removed  it will resume  printing and process the tasks from the first one  Note that you cannot change  the sequence of the queued tasks                                      Figure 6 27 Print s
50. ipping container  Return shipments will not be accepted if the  Mindray number is not clearly visible  Please provide the model number  serial    number  and a brief description of the reason for return     2  Freight policy  The customer is responsible for freight charges when equipment is    shipped to Mindray for service  this includes customs charges      Company Contact  Address  Mindray Building  Keji 12th Road South  Hi tech  Industrial Park  Nanshan  Shenzhen  P  R  China  Phone   86 755 26582888  Fax   86 755 26582680    Authorized Representative    Name  Shanghai International Holding Corp  GmbH  Europe   Address  Eiffestra B e 80 D 20537 Hamburg Germany  Phone   49 40 2513175   Fax   49 40 255726       IV BC 2800 Auto Hematology Analyzer  V1 1     Content       Content  Chapter 1 Hardware secs asssgcssaanckeattagatseaannbarn bank akgensbace dean eas eanands aaa 1 1  LCL  Generala r a E a E E E A A AE 1 1  k OW SI SUMO IY ih e eee deb ER e EE OK Eer KOEK E Eei 1 3  TI Oe ever e E E E E a 1 4  1 4 CPU and Peripheral DeVICES   cciccicscsaiscivccsiaccinstietivecciedcteshetiieciatesecdices 1 5  1 5 Analog Inputs and Outputs cc ccssers  catencceteee cance cate eater eee 1 8  1 6 Digital Inputs and Outputs sass vecks exci evsnznsiecaeenth essernevbeapentheosemmaveaomes ons 1 9  ET     Driving Boarne e Vleet e ven en vice Rene vents ye Rel v een vier 1 11  1 8     Display Unitins ene wn weGnees OAR  1 14  1 9 Functions of the Keypad Adapter                   cc cceeeeeeeee
51. ith the communication between the scanner and the  analyzer    Solution    Check the connection between the two devices     7 2 12 Printer Out of Paper    Possible cause  The printing paper has run out or is not correctly installed   Solution    Check whether the printer is out of paper  If so  load paper to the printer   otherwise  re install the existing paper    If the error still remains  please contact Mindray Customer Service Department  or the distributor     7 2 13 Printer Connection Error    Check whether the printer is well connected to the analyzer   If the error still remains  please contact Mindray Customer Service Department  or the distributor     7 2 14 Recorder Communication Error    Call Mindray Customer Service Department or the distributor     7 2 15 Recorder Out of Paper    Possible causes  recording paper has run out or is not correctly installed   Solution    Check whether the recording paper has run out  If so  load new paper  if not   re install the existing paper  If the error still remains  call Mindray Customer  Service Department or the distributor     7 2 16 Recorder Too Hot    Possible causes  the recording head overheats        7 6    BC 2800 Auto Hematology Analyzer  V1 1     Troubleshooting       Solution   Stop using the recorder  If the error repeats  call Mindray Customer Service  Department or the distributor     7 2 17 Press Bar Up    Possible causes  The press bar of the recorder is up    Solution    Push it back If the error still rema
52. ixing plate of the power  module     As the figure below shows  remove the retaining screws  totally 10  to remove  the plates        BC 2800 Auto Hematology Analyzer  V1 1  3 3    Disassembling the Analyzer                                                                                                                                                                   Figure 3 5    3 2 3 Removing the front cover    As the figure below shows  remove the retaining screws  totally 6  to remove the  cover                                                     Figure3 6       3 4 BC 2800 Auto Hematology Analyzer  V1 1     Disassembling the Analyzer       3 2 4 Removing the LCD assembly    As the figure below shows  remove the retaining screws  totally 4  to remove the  assembly        Figure3 7    3 2 5 Removing the keypad    As the figure below shows  remove the retaining screws  totally 7  to remove the  keypad        Figure3 8  3 2 6 Removing the LCD and adaptor    As the figure below shows  remove the retaining screws  4 for each part  to  remove the LCD and adaptor        BC 2800 Auto Hematology Analyzer  V1 1  3 5    Disassembling the Analyzer    NN S i    j AA bel         Figure3 9    3 2 7 Removing the power shielding box    As the figure below shows  remove the retaining screws  totally 3  to remove the  box        Figure3 10    3 2 8 Removing the driving board  analog board  CPU board and    power board    As the figure below shows  remove the retaining screws  totally 18  t
53. low the displayed instructions to remove all the tubes   except for the one for the waste  from this analyzer    Press  ENTER  to start the draining procedure    When the draining is done  the screen will display Turn off this analyzer and  you should turn off this analyzer as instructed        6 14 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 2 12 Cleaning Probe Wipe Block    After being used for a long time  the bottom of the probe wipe block may be  contaminated by blood and the inside of the block may also be contaminated by  the dirt sucked in  So you need to clean the wipe block regularly     Follow the steps given below to do so    At the Maintenance    screen  press the appropriate arrow keys   CJL  I  lt      to move the cursor to Clean wipe block    Present the probe cleanser to the sample probe and press  ENTER  to aspirate  the cleanser  Remove the cleanser after the probe has risen up    Unscrew and remove the retaining screws with hands or screwdrivers  pointed  by the arrows shown in Figure 6 9  on the right plate of this analyzer    Follow the arrow shown in Figure 6 10 to push and remove the right plate    Follow the instructions displayed on the screen to place an empty cup below the  sample probe    Press  ENTER  to soak the wipe block with the aspirated cleanser  The soaking  progress will be displayed on the screen  as Figure 6 14                        Diluent prime Lyse test  Rinse prime Clean bath  Lyse prime Drain bath  Zap aperture 
54. nalysis  This analyzer can process two types of blood  samples     whole blood samples and prediluted blood samples     2 5 1 1 Whole blood mode    When analyzing a whole blood sample  this analyzer aspirates 13  L of the  sample and follow the procedure presented in Figure 2 10 to dilute it before  proceeding to the actual analysis           Whole Blood Sample  13uL          Diluent 3mL    1  231 Dilution    Diluent 0 5mL         Diluent 3  5mL       For WBC Analysis   1  308 Dilution      For RBC Analysis   1  44862 Dilution           Figure 2 10 Diluting procedure of whole blood mode    2 5 1 2 Prediluted mode    When analyzing a prediluted sample  the operator should first collect 20 u L of  capillary sample and dispense 1 6mL of diluent from this analyzer to predilute the  sample  Then the operator should present the prediluted sample to the analyzer        2 12 BC 2800 Auto Hematology Analyzer  V1 1     Fluidic System       which will aspirate 0 7ml of the sample for further dilution  as Figure 2 11 shows                 Capillary Blood Sample  20uL    Diluent 1 6mL    1   1 Dilution    Diluent 2 1mL    1 324 Dilution        Diluent O0 44mL    Lyse 0 36mL  For WBC Analysis   1 417 Dilution     Diluent 3mL       For RBC Analysis   1 45004 Dilution     Figure 2 11 Diluting procedure of prediluted mode       2 5 1 3 Volume range of blood cells    After reacting with the diluent and lyse  the cell volumes mainly fall into the  following ranges    WBC  30 350 fL   RBC  25 25
55. nalyzer  It helps the service engineers diagnose system errors     Press  MENU  to enter the system menu and press the appropriate arrow keys     t JL   Ji lt        to move the cursor to Service Log  as Figure 6 23 shows        Count          Setup            Review    Quality contro  gt   Status    Calibration    Shutdown       Tubing         O Machine       O Valve         Circuit             Maintenance               gt   Self test    Config                        Print    Reagents for   gt 99 Counts    enu    A     ITER Enter sut          Figure 6 23 Entering log    Press  ENTER  to enter the Log screen  as Figure 6 24 shows     O Other             Settings    NO   Time   Type g g  Times   Informat ion        1 1 04 2005 15 34 Self test  2 09 04 2005 15 30 Others  3 09 04 2005 15 23 Print  amp  comm    4 03 04 2005 15 21 Self test 2                 MENU    F    1 Select log type   T  1 PgUp  PgOn  Scrol       Figure 6 24 Log screen       BC 2800 Auto Hematology Analyzer  V1 1     6 23    Maintenance       The recorded events are divided into three groups  All  Settings and Other   including setting discriminators  system self test and updating system  software   which are all listed on the left of the screen  All the recorded events  are listed on the right of the screen by default  You can press  F1  to select the  interested group and the right of the screen will display the events of the  selected group only  Every screen displays 10 events  You can press  ft  or  J  
56. needed       When you think the bath might be dirty  perform the Clean the bath    procedure        When the analyzed samples add up to 200  the system will remind    you to perform the Probe cleanser cleaning procedure        When the analyzed samples add up to 4 000  the system will remind    you to perform the Clean wipe block procedure        When this analyzer is not to be used for two weeks  be sure to  perform the Prepare to ship procedure to empty and wash the fluidic    system        To obtain reliable analysis results  this analyzer needs to work in a  normal status  Be sure to run the Self test items regularly to check    the status of this analyzer        When this analyzer gives alarms for clogging  you can perform the  Flush aperture or Zap aperture procedure  or press  F2  at the    Count screen to unclog the system           If you see other error messages  see Chapter 7 Troubleshooting  for       solutions           BC 2800 Auto Hematology Analyzer  V1 1  6 3    Maintenance       6 2 System Maintenance    Press  MENU  to enter the system menu and press the appropriate arrow keys     t JL 4 J  lt      to move the cursor to Service     Maintenance  as Figure 6 1  shows     Whole Blood AIl                Current sample WBC H 12 5 x 10L  Lymph H 48x10  X   Mid  H 10x10 W  0  120 Gran   6 7 x 10L     Time  01 27 2004 22 23 Lymph  39 0   x  WD Mid  8 1   A CaA 4 an    Mode   WB   All i  100 200 300 fL  i j Gran  53 9    HGR H 167gL  Name  gt  R  RAC awx wn R    HOT
57. ng                      Lyse entrance    Lyse Diluent  dispensing   Sample  dispensing  tube tube  probe  E       Diluent entrance    Ly             Syringe  assembly    Figure 2 3 Aspiration dispensing module    The lyse is aspirated and dispensed into the bath as described below    The valves 1  3 and 5 are open and the motor pulls the syringe plunger  downward to aspirate certain amount of lyse  The aspirated lyse is stored in the  buffer tubing of valves 1 and 3  Then valve 1 is closed and the motor runs  reversely to push the syringe plunger upward to dispense the stored lyse into the  bath    Since the capacity of the buffer tubing is far greater than the volume of the  aspirated lyse  the lyse will not overflow to the syringes through valves 3 and 4   Service tip  Note that the length and type of the buffer tubing shall not be  changed     2 4 2 Counting module    The counting module mainly consists of the bath  valve 1 and 2  vacuum filter   metering tube  negative pressure and other supporting components  as figure  below shows        BC 2800 Auto Hematology Analyzer  V1 1  2 5    Fluidic System    i r   ye r x ar dust  ra ll   2 ilter   eee     tis sssi N   Io o     l l   l   Metering   tube     l   l         Aperture  Negative Pressure  Front chamber  bath      Back  bath    Figure2 4 Counting module    The counting module is implements the most important function of the analyzer    counting  The electrodes installed on the bath detect the pulses caused by blood  c
58. ng the Lyse Test procedure to expose the  bath        BC 2800 Auto Hematology Analyzer  V1 1  6 13    Maintenance         At the Maintenance screen  press the appropriate arrow keys    t         lt          to move the cursor to Drain bath      Press  ENTER  to start the procedure and the screen will display the progress  bar       When the bath is drained  the screen displays Drain bath  as Figure 6 13  shows       Check the bath and its lower hose for residual liquid  If there is  turn off the  analyzer and call the Mindray Customer Service Department or the distributor for  repair  if not  press  ENTER  and the analyzer will prime the bath with diluent      When the priming is done  the screen will return to the initial state and you  should try to find other reasons the contribute to the abnormal results                            tem  Diluent prime Lyse test  Rinse prime Clean bath  Lyse prime Drain bath  Zap aperture Drain tubing  Flush aperture Clean wipe black  Probe cleanser cleaning Prepare to ship  E Z cleanser cleaning  Prime the bath with diluent  Press  ENTER  to continue              Select service          Figure6 13 Drain Bath    6 2 11 Draining Tubing    If this analyzer is not to be used for a long time or it is to be maintained  be sure  to perform the Drain tubing procedure     Follow the steps given below to do so      At the Maintenance screen  press the appropriate arrow keys   CIL  I  lt      to move the cursor to Drain tubing    Press  ENTER  and fol
59. o remove  the boards        3 6 BC 2800 Auto Hematology Analyzer  V1 1     Disassembling the Analyzer          Figure3 11    3 2 9 Removing the transformer    As the figure below shows  remove the retaining screws  totally 2  with a socket  wrench        Figure3 12    3 2 10 Removing valves    As the figure below shows  remove the retaining screws  two for each valve  to  remove the valves        BC 2800 Auto Hematology Analyzer  V1 1  3 7    Disassembling the Analyzer       Figure3 13    3 2 11 Removing the sample probe assembly     As the figure below shows  remove the retaining screws  totally 3  to remove the  assembly        SoZ    Figure3 14       3 8 BC 2800 Auto Hematology Analyzer  V1 1     Disassembling the Analyzer       3 2 12 Removing the syringe assembly    As the figure below shows  remove the retaining screws  totally 4  to remove the  assembly           Figure3 15    3 2 13 Removing the vacuum chamber and volumetric metering    tube assembly    As the figure below shows  remove the retaining screws  4 for each part  to  remove the assembly        BC 2800 Auto Hematology Analyzer  V1 1  3 9    Disassembling the Analyzer    i       _ re  F    z S  7  5    Figure3 16    3 2 14 Adjusting Sample Probe and Replacing Wipe block    The relative position between the sample probe and wipe block has influence on the  test results  In the accessory kit  there is a sample probe localizer  as Figure 8 28  shows  You need to use the localizer to adjust the position of the s
60. o tubing malfunctions  Therefore  testing the  valves is a major way to remove fluidic errors     At the Self test screen  press  F1  to select the Valve group  as Figure 6 21  shows  The following test items will be displayed on the right of the screen   Press the appropriate arrow keys    t         lt        to select the valve you want to  check and press  ENTER  to test it  If the valve goes through the Off On Off  sequence without making abnormal sound  it passes the test  Otherwise   something may be wrong with the valve              Tubing   item   Result I item   Result    O Machine vi   i Off V7 Off  Oo vi be     Valve 7 ka   fE     Circuit in   i pil  v3 Off yo Off  WA Off vid Off    v5 Oft vii ort                             Select test   ENTER  D          Figure 6 21 Testing valves    6 4 4 Testing Circuits    At the Self test screen  press  F1  to select the Circuit group  as Figure 6 22  shows  You can test the A D interrupt at this screen by pressing  ENTER  to  see whether the WBC  RBC and PLT signals can be properly converted into  digital signals        BC 2800 Auto Hematology Analyzer  V1 1  6 21    Maintenance                Tubing tem   Result    O Machine       O Valve         Circuit                         Reagents for     Select test  ENTER  Execute          Figure 6 22 Testing A D converter       6 22 BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       6 5 Log    The log records all the major events taking place during the running of this  a
61. omer Service Department or the distributor     HGB Error    HGB gain is not correct and HGB blank voltage is within OV 3 2V or 4 9V 5V   Solution    Access Setup     Password to gain the administrator authority  Access Setup      gt  Settings     Gain to adjust the HGB blank voltage to 3 4 4 8V  4 5V  recommended   If the error still remains  turn off this analyzer and call Mindray  Customer Service Department or the distributor     HGB Adjustment    HGB gain is not correct and HGB blank voltage is within 3 2   3 4V or 4 8  4 9V   Solution    Access Setup     Password to gain the administrator authority    Access Setup  gt  Settings     Gain to adjust the HGB blank voltage to 3 4   4 8V  4 5V recommended   If the error still remains  turn off this analyzer and  call Mindray Customer Service Department or the distributor     WBC Clog    This error message occurs when the actual WBC count time is greater than the  preset WBC count time by 2 seconds        BC 2800 Auto Hematology Analyzer  V1 1  7 3    Troubleshooting       Possible causes  clogged aperture  inappropriate WBC count time settings or  solenoid valve error   Solution     Access Service     Maintenance and do the Zap aperture and Flush  aperture procedures     After unclogging  access Setup     Settings      Count to note down the  preset WBC count time  Access Service     Self test     Tubing and test the  actual WBC count time     If the tested time differs from the preset time by less than 2 seconds  it means  th
62. or other Mindray analyzer  Only the  J2 is installed for the BC 2800  The J1 serves to connect the LCD signal cable        1 14 BC 2800 Auto Hematology Analyzer  V1 1     Hardware       For new BC 3000 For BC 2800               Figure 1 10 Schematic of the adapter    BC 2800 Auto Hematology Analyzer V1 1  1 15    Hardware    1 9 Functions of the Keypad Adapter    1  To scan the keypad  The keypad adapter scans the keypad and reports the scanned key code to the  main board    2  To control the LCD brightness  The keypad adapter receives the instructions from the main board to turn on off  the backlights and power indicator of the LCD and to control the brightness of  the backlights    3  To control the buzzer    The keypad adapter receives the instructions from the main board to turn on off  the buzzer     1 9  1 Architecture of the Adapter    The adapter mainly consists of a MCU  keypad matrix  backlight control  power  indicator control and buzzer     1 9 2 Detailed Description    1 9 2 1 Power supply  The main board provides a  12V and 3 3V supplies  which are isolated from  each other  The 3 3V supply is the main power of the adapter and the  12V is  passed to the backlight board  inverter  of the LCD and also converted to a 5V  supply to drive the buzzer and controls the on off of the backlight power the  adapter  Since the 3 3V and  12V are isolated  the MCU send the control  signals to the buzzer and backlight board through photocouplers    1 9 2 2 MCU  The MCU is AT89C
63. ory attire  including rubber gloves  and    follow safe laboratory procedures when handling any material in  the laboratory        You can soak the bath and tubing with the probe cleanser  an alkaline detergent   by performing the Probe cleanser cleaning procedure  If your analyzer is to  run 24 hours a day  you should perform this procedure every 3 days  If you  follow the shutdown procedure to turn off your analyzer everyday  you should    perform this procedure every week     Follow the steps given below to do so      At the Maintenance screen  press the appropriate arrow keys    t      J  lt         to move the cursor to Probe cleanser cleaning      Present the cleanser to the probe and press  ENTER  to aspirate the cleanser   Remove the cleanser after the probe has risen up  This analyzer will  automatically soak the bath and tubing with the aspirated cleanser      The soaking process will last about 5 minutes and you may press  ENTER  to  stop the process any time  Note that a shortened soaking process may not be as  effective as a complete one        BC 2800 Auto Hematology Analyzer  V1 1  6 9    Maintenance         When the soaking is done  press  ENTER  to flush the bath and tubing  after  which screen will return to the initial state     To make sure this analyzer functions normally  every time the accumulated  analyzed samples reach 200  a dialog box will pop up to remind you to perform  the Probe cleanser cleaning procedure  as Figure 6 8 shows  If you want to do 
64. per communication error  error    abnormal    R d t of R d t  1004 ecorder out o ecorder o0 1006   Press bar up  paper hot    2004   Lyse expired Filter error 2006   Real time clock error    Syringe motor Rotation motor  4004   Elevator motor error  error error    400B   Diluent out Rinse out 400D   Waste full    f Dynamic  8001   File error  memory error          7 2 BC 2800 Auto Hematology Analyzer  V1 1     Troubleshooting       7 2 Solutions    7 2 1    7 2 2    7 2 3    7 2 4    7 2 5    This chapter presents measures to be taken when the errors occur     Environmental Temperature Abnormal    Possible causes  abnormal environmental temperature or malfunctioning  temperature sensor    Solution    Access Service     Status and check the environmental temperature  If the  temperature exceeds the specified range by 15  C 35  C  you need to adjust the  work environment of this analyzer so that the analyzer works in the requested  environment  If the temperature is within the requested range and the error  remains  call Mindray Customer Service Department or the distributor     Background Abnormal    At least one parameter failed the background check    Solution    At the Count screen  press  F3  to do the startup procedure  If the error still  remains  access Service     Maintenance and do the Probe cleanser  cleaning procedure  When the cleaning is done  return to the Count screen  and check the background again to see whether the error is cleared  If not  call  Mindray Cust
65. pter6 Maintenance    Regular cleaning and maintenance are demanded to guarantee this analyzer operating  properly  This chapter introduces how to take care of this analyzer and check the system  status     Liquid overflow or leak during the operation of this analyzer will degrade the accuracy of  the analysis results  Once it occurs  immediately wipe off the spills  If it occurs inside this  analyzer  be sure to shut down the power immediately and call Mindray Customer  Services Department or the distributor  Otherwise  the service life of this analyzer may be  shortened        BC 2800 Auto Hematology Analyzer  V1 1  6 1    Maintenance       6 1    Regular Maintenance    To keep this analyzer in a good shape  regular maintenance is demanded     6 1 1 Special Notes    The parts in contact with blood are potentially infectious  Be sure to wear  standard laboratory attire  including rubber gloves  when maintaining or  operating this analyzer and wash your hands with detergent when you are  done     Be sure to keep you hair  clothes  cuff or hands away from the moving parts of  this analyzer    Be sure to use specified tools or parts to maintain this analyzer and be sure to  clean the used tools as instructed by their instruction manual when you are  done    Be sure to use soft and clean cloth  or neutral detergent soaked cloth  twisted  dry   or soft cloth washed by ethanol to clean the surface of this analyzer    Be sure to pay attention to the marks or symbols on this analyzer
66. r the  distributor     7 2 34 Dynamic Memory Error    Something is wrong with the system memory   Turn off the analyzer and call Mindray Customer Service Department or the  distributor        BC 2800 Auto Hematology Analyzer  V1 1  7 9    List of Spare parts       Chapters List of Spare parts       P N    Description       0000 10 10891    Keyboard       0000 10 10907    Disk on module  32M        0010 10 12316    Hitachi STN screen  640X480        0200 10 05528    Bearing  for syringes        0200 20 05530    Plunger       1800 30 19455    Indicator board       2000 20 03124    Ruby Red Cell Counter 80um RB 22084       2800 20 32054    START key       2800 30 32057    Recorder assembly       2800 20 28674    Keypad overlay       2800 20 28696    Syringe press plate       2800 20 28699    Plunger pin          2800 20 28700    7 5ml bolt       2800 20 28701    Upper shielding plate for the analog board       2800 20 28702    Lower shielding plate for the analog board       2800 20 28719    Guide bushing       2800 20 28724    Sealing bushing       2800 20 28725    Sealing ring       2800 20 28739    Connection wires for the indicator board       2800 20 28740    LCD adaptor wire flexible FFC        2800 20 28754  2800 20 28757    Connection wire between the keypad and the recorder  Connection wire of the position sensor       2800 20 28770    Connection wire for the LCD       2800 20 28843    Body of 7 5ml syringe       2800 20 28844    Flange of 7 5ml syringe       2800 20 28
67. red in the FLASH   so the FLASH is also called the BootROM  Every time the system is powered  on  the CPU first executes the boot program that initializes the system and  loads the control software from the DOM  The FLASH also contains such  information as the FPGA configuration  FPGA version and LCD contrast     1 4 4 SDRAM  The system memory consists of two 8M byte memories     1 4 5 DOM    The CPU board uses a 32M DOM that is powered by a 3 3V supply  the DOM  can also be supplied by 5V supply   The DOM is only operational after the  FPGA is configured     1 4 6 RTC    The CPU board uses a real time clock  RTC  to record the time  The RTC is  connected to the I C bus of the CPU board and synchronized by a 32 768KHz  crystal oscillator  When the analyzer is powered on  the RTC is powered by the  CPU board  when the analyzer is powered off  it is powered by the built in  battery     1 4 7 EEPROM    The CPU board uses an 8K byte EEPROM to store such information as system  configurations and settings  It is connected to the I2C bus of the CPU and can  be written by CPU on line     1 4 8 LEDs    When D1 is on  it means  3 3V is functioning properly  When D9 is on  it  means  12 8V is functioning properly  When D5 is on  it means the system is       1 6 BC 2800 Auto Hematology Analyzer  V1 1     Hardware       reading or writing the DOM  When D7 is on  it means the FPGA has been  configured and is functioning properly  When D20 is on  it means the FPGA is  restarting  The D11 D1
68. sary contact with the  probe and probe area        Follow the steps given below to replace the wipe block    1  Follow steps 1 6 according to Section 3 3 1    2  Pull the loosen sample probe upward until it out of the wipe block  then remove  the wipe block and disconnect the tubings from the it  pay attention to the  correspondence between the tubings and the connectors   as the figure below  shows  Install a new wipe block and connect the tubings  the tubing marked with  a black line should connect to the bottom connector   When the connection is  done  place the wipe block back to its original position     3  Follow steps 7 9 according to Section 3 3 1 to fix the sample probe        3 14 BC 2800 Auto Hematology Analyzer  V1 1     Disassembling the Analyzer                   Figure3 24 Installing probe wipe       BC 2800 Auto Hematology Analyzer  V1 1  3 15    Histograms and Pulse Graphs       Chapter4 Histograms and Pulse Graphs    4 1 Histograms    This chapter presents several WBC histograms you may encounter frequently     1  A normal histogram                                  a G d    0 50 100 150 200 250 300 fL  Figure 4 1  NOTE  Blood cells lain between the first and the second discriminators are  lymphocyte  those between the second and the third discriminators are  mid sized cells  those between the third and the fourth discriminators are  granulocyte  The fourth discriminator is the fixed line   2  A histogram that is too narrow to indicate the differential result
69. the liquid  and air in the pressure chamber is expelled to the outside through the NO end of  valve 1 and the pump and the NC end of valve 2 to establish the vacuum  The    pressure value is monitored by the pressure sensor     The pump is an imported American product  whose P N is 3001 10 07252        BC 2800 Auto Hematology Analyzer  V1 1     2 9    Fluidic System    Pressure Pressure       sensor    sensor     lt                                            Probe Wipe       Le    Negative pressure    Negative pressure  chamber    chamber       NO       Waste exit                       gt  Waste exit    yc   gt     Pump    AY                                                                      NC Probe Wipe NC          Figure2 7 Pressure vacuum module   the left establishes the pressure  the right establishes the vacuum     2 4 5 Mixing module    The types of mixing is available  mixing by the horizontal movement of the  sample probe and bubble mixing  As Figure2 8 shows  the aspirated sample  needs to be diluted before the counting  Once the sample is dispensed into the  bath containing certain amount of dilute  the system will inform the mixing  module to work  Then the sample probe initiates a mild horizontal movement  inside the bath and the positive pressure is established inside the pressure  chamber  The valve is then open briefly and separate the air in the pressure  chamber into several air segments and expel them into the bath to introduces  bubbles  The bubbles pop
70. to Hematology Analyzer  V1 1  3 11    Disassembling the Analyzer             Tubing Item Result       Machine    Syringe motor       O Valve  O Circuit Rotation motor  Elevator motor    Print                         Reagents for           Select test  ENTER  Execute       Figure3 19 Elevator motor screen    Press  f  to move the sample probe upward and press       to move the probe to the  top of the bath  as the figure below shows                 Figure 3 20 Sample probe on the top of the bath    3  Loose the retaining screw by a Philips screwdriver  as the figure below shows        3 12 BC 2800 Auto Hematology Analyzer  V1 1     Disassembling the Analyzer             Figure 3 21 Removing screws    4  Remove the probe from the wipe block and insert the localizer into the wipe block  from the bottom  as the figure below shows           N    Figure3 22 Insert localizer    5  Insert the probe into the wipe block until it reaches the localizer  as the figure  below shows        BC 2800 Auto Hematology Analyzer  V1 1  3 13    Disassembling the Analyzer       Figure3 23 Inserting sample probe into wipe block    6  Re tighten the screw to fix the probe and remove the localizer   3 2 16 Replacing Wipe block    Biohazard  Wear standard laboratory attire  including rubber gloves  and  follow safe laboratory procedures when handling any material in  the laboratory     Warning  The probe is sharp and may contain biohazard materials  including  controls and calibrators  Avoid any unneces
71. ts 2 UART  3 3VTTL   one to connect the  keypad and the other reserved to control the pump  Another 2 UARTs  RS232   are implemented inside the Super I O to connect the scanner and to  communicate with the PC     1 6 2 Parallel Port and PS 2 Port    The Super I O provides a DB25 parallel connector to connect to connect a  printer or a floppy drive  the power supply of the floppy drive is supplied by the  PS 2   The software will automatically adapt to the connected printer or the  floppy drive     The Super I O provides a keyboard interface and a mouse interface  COM3  and COM4   Note that the BC 2800 does not support the mouse yet     1 6 3 GPIOs    1 Signals of the Start key    The FPGA detects the input signal  which will turn low when the Start key is  pressed    2 Volumetric metering Signals  The FPGA detects the signals sent by the start transducer and the end    transducer     BC 2800 Auto Hematology Analyzer V1 1  1 9    Hardware    Signals of level detection   The BC 2800 has not level sensors   Digital pot   The SPI bus interface implemented by the FPGA controls the 4 digital  potential meters on the analog board to control the HGB gain    Signals controlling valves and pumps   The Super I O outputs 20 control signals to control the valves and pumps    through the driving board  Since the BC 2800 only has 1 pump and 11 valves   the redundant lines and ports are reserved     Signals controlling bath    The Super I O outputs 4 control signals  through the analog board
72. ule    The flushing module includes  washing the exterior and interior of the sample  probe  washing the front and back bath  flushing the aperture and flushing the  fluidic lines    2 4 3 1 Washing exterior and interior of sample probe  The part that washes the sample probe is shown in Figure2 5  The washing  consists of two procedures     washing the exterior and washing the interior    1  To wash the exterior  The plunger of the 10mL syringe moves downward to  aspirate diluent from the NO end of valve 6  Then the plunger moves upward and  valves 6 and 5 are open so that the aspirated diluent is dispensed  through the  NC end of valve 6  NC end of valve 5 and valves 4 and 3 to the lower part of the  probe wipe  Then the pump functions to open valves 9 and 10 to introduce a  negative pressure on the upper part of the probe wipe  Because of this negative  pressure  the diluent dispensed to the lower part of the probe wipe is transferred  to the upper part and washes the exterior through the up and down movement of  the sample probe    2  To wash the interior  The plunger of the 10mL syringe moves downward to  aspirate diluent from the NO end of valve 6 Then the plunger moves upward to  dispense the aspirated diluent  through valves 5 and 6 and the buffer tubing  to  the sample probe to wash the interior        BC 2800 Auto Hematology Analyzer  V1 1  2 7    Fluidic System                                                                                                           
73. unt time   If the error still remains  turn off this analyzer and call Mindray Customer  Service Department or the distributor   7 2 7 RBC Clog  This error message occurs when the actual RBC count time is greater than the  7 4 BC 2800 Auto Hematology Analyzer  V1 1     Troubleshooting       7 2 8    preset RBC count time by 2 seconds    Possible causes  clogged aperture  inappropriate RBC count time settings or  solenoid valve error    Solution     Access Service     Maintenance and do the Zap aperture and Flush aperture  procedures     After unclogging  access Setup     Settings     Count to note down the  preset RBC count time  Access Service     Self test     Tubing and test the  actual RBC count time     If the tested time differs from the preset time by less than 2 seconds  it means  the unclogging is successful and you can return to the Count screen to continue  the analysis  Other wise  Access Service     Maintenance to soak the bath  and tubing with probe cleanser    When the soaking is done  access Setup     Settings     Count to note  down the preset RBC count time  Access Service     Self test     Tubing and  test the actual RBC count time  If the tested time differs from the preset time by  less than 2 seconds  it means the unclogging is successful and you can return  to the Count screen to continue the analysis  If the difference is still greater  than 2 seconds and stabilized around a certain value  access Setup  gt   Settings   Count to change the RBC count tim
74. ure        BC 2800 Auto Hematology Analyzer  V1 1  6 19    Maintenance       Filter    It checks whether the filter functions normally     6 4 2 Testing Motors and Recorder Printer    At the Self test screen  press  F1  to select the Machine group  as Figure 6 20  shows  The following test items will be displayed on the right of the screen     Press   t   or       to select the desired item and press  ENTER  to conduct the  test              Tubing   tem   Result     Machine E    Syringe motor       O Valve       Circuit Rotation motor  Elevator motor    Print                                Select test  ENTER  Execute       Figure 6 20 Testing mechanic parts    Syringe motor    The syringe motor controls the aspiration volume  This test checks whether the  motor functions normally     Rotation motor    The rotation motor rotates the sample probe inside the analyzer  This test  checks whether the motor functions normally     Elevator motor    The elevator motor controls elevation of the sample probe  This test checks  whether the motor functions normally     Print    This test checks whether the recorder or printer functions normally  If normal   when you press  ENTER   the recorder or printer will print out test page  if  abnormal  the screen will display the corresponding error message and you can       6 20    BC 2800 Auto Hematology Analyzer  V1 1     Maintenance       see Chapter 7 Troubleshooting for solutions     6 4 3 Testing Valves    Malfunctioning valves will lead t
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
FY-08PDX8W-W の取付工事説明書  Wireless Network Video Server Model: NC6200 User`s Manual  Phonix S3586TFG mobile phone case  Samsung HT-C550 Uporabniški priročnik  AspadTools User`s Manual    車載アタッチメント取扱説明書 - Support  Blizzard® Blizzard®  CENTERLINE 2100 Motor Control Centers  User Manual - Quest Engineering    Copyright © All rights reserved. 
   Failed to retrieve file