        (IRIS) VITAL SIGNS PATIENT MONITOR Service Manual
         Contents
1.     Pulse rate  Measuring range    Alarm setup range    40 255  mmHg   10 195  mmHg   20 215  mmHg     40 200  mmHg   10 150  mmHg   20 165  mmHg     40 135 mmHg   10 110 mmHg   20 125 mmHg   1 mmHg  the average error  lt t5mmHg  the standard error  lt t8mmHg  software and hardware double protection  290 mmHg   220 mmHg   150 mmHg     double length infrared wave  0 100    70 100    1     2   70 100  adult child    3   70 100  neonate    4   40 69      25 250bpm  20 250bpm    PMS8210A Service Manual 1 3    Chapter 1 Complete Appliance Introduction    Resolution  Precision  Sweep speed  Alarm setup    Alarm method    1 4 4 Blood Oxygen  Nellcor     SpO2    Measuring method  Measuring Range  Alarm setup range  Resolution    Precision    Pulse rate    Measuring Range  Alarm setup range  Precision   Sweep speed  Alarm setup    Alarm method    1 4 5 Respiration    Measuring method  Measuring range  Resolution  Precision   Alarm setup  Alarm method    1 bpm    3 bpm Geostationary or 5 bpm Campaign   12 5mm s  25mm s   SpO2 overruns  pulse rate overruns    sound light alarm  and record the alarm status    for review    double length infrared wave   0 100    70 100    1    2   70 100  adult child    3   70 100  neonate   Unspecified  0 69      20 250bpm   20 250bpm    3 bpm Geostationary or 5 bpm Campaign   12 5mm s  25mm s   SpO2 overruns  pulse rate overruns    sound light alarm  and record the alarm status for    review    the thorax impedance method  used with ECG lead   0  10 120
2.    6  Make the network connection from a certain aspect disconnect the central station  can direct the machine to resume after the networking to connect and work normally   the time required  see the relevant technical indicators     3 1 6 Clock    Instructions correctly  walking evenly  not mutation  non stop     3 1 7 ECG waveform output    Connect IRIS to ECG lead and simulator  and connect D   A output  chassis behind  to  the oscilloscope  Universal oscilloscope   select the 5V or 1V respectively and compare  oscilloscope waveform with the output waveform of IRIS     3 1 8 Long time run    After long time run  observe the monitor state whether ok     3 1 9 Factory settings  The initial values set by software  except for the  factory settings  menu     3 1 10 Safety testing    1 The test of the protective earthing impedance  the impedance must not exceed 0 20  between the protective earth terminal in MAINS plug and any accessible MAINS part   Please refer to the standard of the clause 18 f  of IEC 60601 1 A1 A2    2 Earth leakage current  when the MAINS is 250V  the current must not exceed 0 5mA   under single fault conditions  the current must not exceed 1 0mA  Please refer to the  standard of the clause 19 of IEC 60601 1 A1 A2    3 The patient leakage current  when the MAINS is 250V  the current must not exceed  10uA Please refer to the standard of the clause 19 of IEC 60601 1 A1 A2     PMS8210A Service Manual 3 3    Chapter 3 Installation and inspection    3 2 The para
3.   Buffer amplifier circuit  complete the ECG signal s impedance transformation to  ensure that ECG has a high input impedance and a low output resistance    Right foot drive circuit  The mid point of the output buffer amplifier circuit to the RL of  the five lead ECG by reversed amplifying to ensure the human body in an  equipotential state  thereby reducing interference and improve the circuit s common  mode rejection ratio    Lead off Detection  According to lead off caused by a buffer amplifier output level  change  adjust the lead off via comparing device and converted into TTL level for  single chip detection    Lead connecting to the circuit  In the single chip control  enlarge the main release  according to the requirements of connecting different lead electrode access to the  main circuit    Main output circuit  By a standard consisting of three op amp instrumentation  amplifier     PMS8210A Service Manual    Chapter 2 Patient Monitor parameters and theory    g  Post processing circuit  The main function is to complete the coupling of ECG  signals  the size of programmable gain  filtering and level shifting  and the signals  amplification to a certain extent into the mode   the digital converters     2 2 1 3 Respiratory  RESP  parameters   IRIS vital sign monitor RESP measurement method is based on the principle of  impedance  The body chest is up and down when breathe  which is equivalent to  impedance changes between RL and LL  and through the ECG electrode RL and L
4.   appropriate government regulations   The instrument must be connected to the ground correctly     The monitor is used in accordance with the operation manual     3F is not responsible for the effects of safety  reliability  and performance  when     Spare Parts have been removed  debugged or readjusted       The monitor isn t used in accordance with the operation manual     Guarantee    The range of free service     The instrument  which is listed in the range of maintenance item can have the    free service     The range of charged service     The monitor which is not on the above free service list  and needs the  maintenance  3F will charge according to the service    If the maintained monitor is man made damaged  voltage out of the  required range  or force majuere by natural calamities  even it is within  warranty period  3F will charge for the service     3F is not responsible for the direct  indirect  completely damaging or delay    in such condition     Using instrument improperly    Tearing down  maintaining  debugging the instrument  replacing  components or changing electrical wires by the technician without 3F s  agreement or commission     Essential Guides for the Users    Warning  All Users must read following warnings and guide before  operating the Monitors  Any abnormalities or malfunction of the  monitor or body harms caused by the violations of the operational  guides will not be responsible by our company  nor any warranties  will be made by us   The inst
5.   higher   the voltage comparator outputs the high level     PMS8210A Service Manual 2 5       Chapter 3 Installation and inspection    Chapter 3 Installation and Inspection    3 1 Instrument installation    Please refer to the    PMS8210A operation manual    for the usual measurement of IRIS   So here we will not describe in detail  The following steps are to indicate the main point  of the measurement as well as some parts which is not carefully described     3 1 1 The checking for appearance and assembling    Take out the instrument from the package carefully  Check the aspect first  If there are  some occurrences due to transportation as that the aspect of monitor is damaged  the  LCD panel is broken  or there is abnormal sound when shaking the instrument mainly  caused by fall off of some components  please don   t plug into the power socket or try to  open the instrument to examine or repair  Instead  contact with the local dealer or the  customer service department of 3F as soon as possible     If the instrument looks well and buttons operate well  put in on the flat desk or fix it on  the bracket  Please put the electro cardio lead plug  blood pressure plug and blood  oxygen probe plug in the corresponding socket on the right panel  If one of the  parameters is not wanted  the probe or cable could be taken off from the side panel   Connect one side of the power wire to the power socket on the back side of the  instrument and the other side to the output of 220V AC pow
6.  Complete Appliance Introduction    1 5 Appearance  1 5 1 Front View of IRIS       Picture 1 1 IRIS Vital Sign Monitor       Switch  Power on off       Print Switch  Start Stop the printer     8  Measure Blood Pressure  Start Stop measuring the Blood Pressure     4  Real Time Waveform Recording  Freeze unfreeze the current ECG Waveform    5  Mute switch  Turn Off On the sound effects       Multi function Key  SD Card storage  and the input of patient data      Functional Indicator  Display the status of each function   Menu Key  Adjusting the system parameters  and the patient data retrospection     1 6 PMS8210A Service Manual    Chapter 1 Complete Appliance Introduction    1 5 2 Back View of IRIS    I    COS as       Indicator light  Alarm Indicator    Communication port  Nurse Call    Network interface  Standard RJ45 input net connection with 3F Center Station   Ground Cable interface  Interface for connecting the ground wire    Car DC input port  9V 15V    Power  100   240VAC 10   50 60Hz  3Hz     Lithium battery cabin  6 4V 8 4V    Printer  Out built printer           Q             ORS    I  Channels of temperature sensor   Il  Channels of NIBP Cuff    III  SPO2 sensor    IV  ECG Cable interface     PMS8210A Service Manual 1 7       Chapter 2 Patient Monitor parameters and theory    Chapter 2 Patient Monitor parameters and theory    2 1 Summary    IRIS adopts parameter signal on the base of parameter modules  and transmits the  result to the main control panel through the
7.  Service Manual 6 1       Annex A Electric connect diagram    Annex A Electric connect diagram  IRIS         SYSTEM BOAR              Ji   LCDBOARD   22    A    l   A     JP2   J1 ALARM   LED    aia PRINTER          J4       MODULE BOARD  TEMP        Sur prose    a PROBE  SPO2 J3    P02 PROBE    J6       5 DC INLET    JS               35 JU  ki  POWER BOARD      J6        CHANGE BOARD  Ji  J2    a JP1    3 NET BOARD        GROUND  ULE NINE o IS A NIBP  Illumination J7 PROS    LINE OF INNER LINE OF  MODULE INLET IN MODULE CONNECT BOARD INLET OF SHELL       PMS8210A Service Manual A 1       Annex B Electric diagram of modules    Annex B Electric diagram of modules    1  Host Assembly                                        Std Code Name  amp  Spec  Qty   1 M10 70 010 Front Housing Component 1  2 M10 07 000 Main Housing Component 1  3 M10 70 040 Rear Housing Component 1  4 GB T 818 2000   Pan Head M3X6 Screws 4  5 GB819 M4 8 Pan Head M4X8 Screws 2  6 GB T 823 1988   Pan Head M4X14 Screws 2          PMS8210A Service Manual    Annex B Electric diagram of modules                                                                            No Std Code Name  amp  Spec  Qty    1 M10 70 011 Front Housing Panel 1   2 M10 06 010 Network Interface Board 1   3 M10 55 000 Printer 1   4 M10 71 020 Rear Housing Cover 1   5 M10 92 020 Li ion Battery 1   6 M10 70 062 Li ion Battery Shell 1  M10 70 063   7 M71 01 020 Foot Pad 4   8 M10 05 010 Power Module 1   9 M10 03 010 Li ion Battery Interface B
8.  all set adult model   B In the non invasive blood pressure simulator select a group of blood pressure    measurement within the scope of value  such as     NS   90  NM   70  ND   60    C Inspection IRIS portable measuring whether the outcome of the actual simulator set  with consistent value     D change the simulator set pressure value  re measurement    3 2 3 SPO2 parameters test    3 2 3 1 Inspection Tools  SPOz2 Simulator    3 2 3 2 Inspection steps  A Connect SPO2 Simulator with IRIS   B Set SPO2 simulator parameter values are as follows   SPO2   98  PR 70  C Inspection IRIS portable SPO2 and PR display value is consistent with the simulator    Note  To observe PR value must be in the ECG heart rate menu will be selected as  the source of PETH    D Change in SPO2 Simulator SPO2 and RP settings   E Check IRIS display value consistent with the settings   F If SPO2 probe is off  IRIS shall immediately report     3 2 4 Temperature Parameters Test    3 2 4 1 Inspection Tools  Physiological signal simulator    3 2 4 2 Inspection steps  A Temperature sensor will take a simulator  then an IRIS portable TEMP mouth   Through the simulator settings  TEMP   34   C   B Inspection IRIS portable screen shows the value of the TEMP for 34   C   PMS8210A Service Manual 3 5    Chapter 3 Installation and inspection    C Change the simulator settings  TEMP   40      D Inspection IRIS portable screen shows the value of the TEMP for 40   C     3 6 PMS8210A Service Manual    Chapter 4 Maintena
9.  be possibilities as follows   the power line is not plug in well  the fuse is broken  the power module is broken or  the indicator light is bad             1 blue DC indicator light suitable for the instrument power supplied by the  battery   The instrument is not plugged in but if there is battery installed in it  press the     start stop    button for about 2 3 seconds when the instrument is off  the instruments  will be started  and the blue indicator light is on the blue DC indicator light is off   and  it shows that the instrument can work through battery supply  If the indicator light isn   t  on  there may be some possibilities as follows  the    start stop    button is loose contact   some malfunctions in the charging control board  DC indicator light is bad  the  capacity of the battery is not enough or the battery is damaged    gm       blue indicator light suitable for the instrument with battery installed supplied  by the AC power   When AC is supplied  the instrument will charge up automatically  If the blue light is  on  it shows that the battery is charged up  If the capacity of the battery is full  the  blue light will be off  Generally  the battery needs 4 8 hours to be charged up with full  capacity  If the time for charging up is above 24 hours and the blue light is on yet   there may be the possibilities as follows  the charging control board has malfunction  or the battery is damaged     4 2 2 Instrument malfunctions       Malfunction    phenomena    Pos
10.  switch board  and display the data and  waveform on the screen  The commands of main control panel and module s  information are also transmitted through the switch board  The switch board can also  transfer the power     The framework of system is demonstrated as below     REIN EEE ged Medical IH eZ N  staff    Main  Control board Net port    Respiration    CO   pets                      PMS8210A Service Manual 2 1    Chapter 2 Patient Monitor parameters and theory    2 2 Parameters and theory    The five parameter modules do the real time monitoring on non invasive blood pressure     SpO2  ECG respiration   body  temperature through the cuff and cables  The results are    transmitted to the main control panel for processing and displaying  and input to print by    the thermal recorder     2 2 1 ECG    2 2 1 1 The IRIS main function of ECG     a   b     Qa o                gt  0    Lead  3 Lead  5 Lead   Lead measurement     Il  III  avR  avL  avF  V  CAL   Drive by right foot   Anti  defibrillation   Detection of lead off   Magnify dual channel ECG  while process the ECG signal by any two lead  measurement     2 2 1 2 The processing of ECG signals is achieved by circuit of ECG parameter  and the    circuit is constructed by several parts     a     b     d     Input electric circuit  ECG electrode is connected to the input circuit with the ECG  cable  The main function of the circuit is to protect the input port of ECG  and signal  filter out interference from the outside  
11. L s  high frequency signal into a high frequency signal  The signal is amplified to  electric signal output  with changes in respiration  and sent to mode digital converters   Respiratory parameters is combined by the respiration electric board and coupling  transformer  Circuit is included by  oscillometry  coupling  detection  early release   high gain amplification and so on     2 2 2 Noninvasive blood pressure  NIBP  parameters    NIBP adopts the theory of pulse oscillometry  Puff the cuff around the upper arm until the  pressure which generated by the cuff interdict the flow of the brachial artery  and then  deflated gradually according to a certain algorithm requirements  As the pressure  reduced  with the beat in the pulse  the arterial blood will generate a pulse signal  Filter  out the pulse signal and enlarge it with a high pass filter  about 1HZ   and convert it into  digital signal by the A   D  The systolic  diastolic and the mean blood pressure can  measure after the treatment by the single chip microcomputer     For neonate  pediatric and adult  we must choose the correct size of the cuff in order to  avoid the measurement error     NIBP prevents the excessive pressure with the protection electric circuit     Main work mode to NIBP    1  Adult Pediatric Neonate mode  choose according to the body figure  weight and age    2  Manual Auto Series  Manual measurement also means a single measurement  Only  measured once after choosing the Manual  Auto can measure onc
12. PMS8210A  IRIS   VITAL SIGNS PATIENT MONITOR    Service Manual    Shanghai 3F Electronics Co   Ltd     ADDRESS  Building 3  No  128  Jiujing Road  TEL   86 21 51695200  Shanghai 201615  China FAX   86 21 67696296  POSTALCODE  201615 DATE  2008 09    VERSION  V1 0       Intellectual Property Rights   Shanghai 3F Electronics Co   Ltd  has the copyright of this service manual and  also can require it as confidential data  This manual is only referred to the  maintenance and service of IRIS Vital Sign Monitor  This manual and  intellectual property rights  include copyright  belong to 3F Electronics Co    Ltd  Without the written approval of 3F  no one can use  publish or allowing  other people to get the information of this manual at any way  Without the  advanced permission of 3F  no one can take pictures  copy  print or translate  the content into other language     Announcement    gt  3F reserves the right of final explanation about this service manual     gt  3F reserves the right that update this manual without advanced notice     gt  3F reserves the right to change the monitor s specifications change  without advanced notice      gt  3F doesn t guarantee this manual in any form     3F is responsible for the effects of safety  reliability  and performance only  if     Assembly operations  extensions  readjustments  modifications  or  repairing is carried out 3F technical personnel or 3F s authorized agent   e The electrical installation complies with the requirements of the
13. ait about 10 minutes  the screen will show   Initializing  the system  please wait     And pls  Wait another 12 minutes  the monitoring picture  and waveform will appear  If the power if off during the working process  the power    indicator will extinguish  and the monitor stops working       Check out all the function  and confirm the monitor is working normally  then put the    cable probe to the related body place of the patient  The IRIS will start the real time    monitoring     PMS8210A Service Manual    Chapter 6 Cleaning    Chapter 6 Cleaning    A Warning   But can not dip the liquid into the monitor   Be forbidden the caustic chemical materials that will make erosion     6 1 Clean    Wipe the instrument and the cables by a clean wet cloth  if wanted  usual cleanser or  alcohol is available     6 2 Sterilize    After cleaning  the ECG leads  SPO probe  NIBP cuff and TEMP probe can be  sterilized sometimes     The material of sterilization can use alcohol  70    formaldehyde  35  37   or javelle   water  the concentration of chloros is 500ppm 5000ppm        gt  The ECG leads can be sterilized by formaldehyde  35  37       gt  The SPO  probe and NIBP cuff can be sterilized by alcohol  70   or formaldehyde   35  37       gt  The TEMP probe can be sterilized by javelle water  the concentration of chloros is  500ppm 5000ppm      6 3 Disinfection    After cleaning  the disinfection can be done sometimes  But the process must be  according to the hospital order     PMS8210A
14. ase be familiar with their definition and meaning    Warning  Instructions to avoid potential danger and incorrect operation  Obey the  instructions  otherwise death and serious injury may be caused    Caution  Instructions to avoid potential danger and incorrect operation  Obey the  instructions  otherwise injury and product loss may be caused    Note  Operational instructions or other useful information to use the instrument fully     m Symbols  The list of the symbols used in this instrument is as follows   N Symbol for    caution  consult accompanying documents       C     CE mark  when a device has been past CE test  we must put the CE mark on  the device  And the CE number can be after the svmbol     a  This svmbol shall be accompanied bv the manufacturer s serial number  The  ISN  serial number shall be after or below the symbol  adjacent to it    b  The relative size of the symbol and the serial number are not specified    c  So far our product serial number is made of seven digits  two digits for the   year and two digits for the month and three digits for sequence number  So the    serial number could be show the date of made     a  The symbol shall be accompanied by the name and the address of the   sl manufacturer  adjacent to the symbol  The address is not required with the  symbol on an immediate container as specified in EN375 AND EN376  except  when the immediate container is also the outer container     b  The relative size of the svmbol and the name and addr
15. e in the selected  time intervals  period can be 1  2  3  4  5  10  15  30  60  90  120  180  200 or 280  minutes  Series measurement starts the fast consecutive measurements in five  minutes  and can be effectively measure the blood pressure changes     PMS8210A Service Manual 2 3    Chapter 2 Patient Monitor parameters and theory    2 2 3 SPO2 parameter    SPO2 is obtained through describing the pulse wave of fingertip  which passing through  the specific algorithm and looking up the clinical data table     The SO2 probe is the surveying sensor  built of two light emitter diodes and an  electro optical battery part  Two light emitter diodes conclude a red diode and an  infrared diode  They light alternately according to certain succession  When fingertip s  blood capillary hyperemia repeatedly along with heart s pump blood  the blood vessel  and the organization absorb the light of the emitter diode  and project on to the solar  battery  the solar battery may sense luminous intensity along with the pulse blood  change  and be formed of changed electrical signal  The ratio of current and alternating  current component is according to the content of oxygen in blood  The correct value of  oxygen saturation degree is obtained though the specific algorithm  simultaneously the  arteries rate may be calculated according to the blood oxygen pulse profile     The circuit module of SPO2 mainly includes four parts    1  Probe parts  Luminescence tubes launch infrared and red radiant t
16. e key board or speaker        The printer does  not print     1  Printer electric line off or bad  connection   2  Printer failure   3  Printer not set    4  Print paper not set     1  Reconnect or replace the printer  electric line   2  Replace the printer   3  Add the paper        The print record  is tilted     The installation of printer is not well     Adjust the printer           The buttons or the  knob is invalid        1  the keyboard signal lines is loose  contact or damaged  2  malfunction of keyboard PCB       1  re plug the signal line  2  replace the keyboard       4 4    PMS8210A Service Manual          Chapter 4 Maintenance    1     offline    is set in the menu   2  the net wires are loose contact or  damaged   3  the connect board is damaged    Not connect to  the center station    WN           set    online    in the menu    rep plug or replace the net wires    replace the connect board       4 2 5 Power board malfunction       Malfunction phenomena    Possible cause    Means to dispose       The fuse burn when power  is on     The power or other parts is short     Open the instrument to check        Cut the load but The fuse  burn still    The power is damaged     Replace the power board        The fuse burn when the  part connects     The part is short     Replace the part        The LED of power and  main board is light  but the  fan is not work or the LED  of trans board is not light     The  12VDC power is damaged     Replace the power           The fa
17. ecially for neonate  1  ECG Measurement    Warning  Away from the throats of the neonate to avoid asphyxiation    The electro cardio cables and the electrode must be used correctly  because neonate   s body is shorter and tender  Avoid the injury of the  electrode  And check on time and change with the finger     2  Blood Pressure Measurement    Warning  The pattern of blood pressure measurement for neonate must be  selected before the neonate measures  otherwise accidents will be caused   The cuff that fits neonate must be used to take the measurement of  neonate  Set correctly the parameters of the air pressure of the cuff and  alarm when neonate takes the blood pressure  Please select the default   Especially  note these parameters that neonate fit if you have to adjust  parameters     3  Blood Oxygen Measurement   Caution  Blood oxygen value  SpO2  may be not obtained precisely  because the neonate moves  To measure precisely  please keep neonate at  rest    Use correctly the blood oxygen probe to neonate  Do not place the blood  oxygen probe on the fingers with skin injury  edema or fragile  Do not place  the probe on the same finger over 2 hours to prevent discomfort of the  finger  Check on time and change with the finger     4  Temperature Measurement    Use the inside temperature probe when taking neonate   s temperature     Terms for safety and symbols    m Terms for safety   In this manual  warning  caution  and notice are used to describe the level of danger   Ple
18. er supply     Complete marks  with correct content  And standard configuration and fixed socket is ok     Now the instrument can be used     If there is battery supply  the instrument can work relying on the internal power when it  isn   t connected to the 220V AC  When the AC220V is connected  the instrument uses  the AC power supply and the internal battery is charged by the AC power     A Warning     1  Can   t be put under direct sunlight  to avoid damages by the high  temperature in the machine box     2  Don   t use the instrument in the environment with poisonous  flammable  or caustic gas     3  Pay attention to the unsteady of the local voltage  if the voltage is over the  permitted range  we advise to add voltage stabilizing device     PMS8210A Service Manual 3 1    Chapter 3 Installation and inspection    4  Don t use keyboard in order to avoid damage to internal information and  procedures     3 1 2 Power on    Before power on  you should check    1  The power voltage meets the requirement of the instrument    2  The three wire power cable is necessary  and the power inlet must match three wire   Do not use the two wire AC power supply    3  When IRIS is used together with other medical devices  the protective ground should  be reliably connected to other device    4  Do not put monitor below the perfusion bag or some other places with the liquids  leakage to avoid the liquids enter into the monitor     When the power is on  check whether the patient monitor can f
19. ess is not specified     a  The svmbol shall be accompanied bv the name and the address of the    m  o    authorized representative in the European Community  adjacent to the symbol   The address is not required with the symbol on an immediate container as    VI    specified in EN375 AND EN376  except when the immediate container is also  the outer container     b  The relative size of the symbol and the name and address is not specified     Q  Smbol for    do not reuse        single use        use only once       X Symbol for    temperature limitation       D    BA Symbol for    consult instructions for use        consult operating instructions     ad Symbol for    biological risks    0  Defibrillation proof type CF applied part   on medical equipment to identify a    defibrillation proof type CF applied part     Equipotentiality  to identify the terminals which  when connected together  bring  the various parts of an equipment or of a system to the same potential  Not  necessarily being the earth  ground  potential  e g  for local bonding     The symbol indicating separate collection for electrical and electronic  equipment consists of the crossed out wheeled bin  The symbol must be  printed visibly  legibly and indelibly     m The interface of networking    CONTENT    Chapter 1 Complete Appliance Introduction    1 1  el ON CL VICW TEE i a a T E 1 1  O CS HEUER EEE 1 1  1 3  Tag SUCCES ee Be ee i a 1 1  1 4 Parameters        cesenseesssensssenennsnnnennnnennnenennonnnenenne
20. he value of    Cuff is tied outside the clothes    Tie the cuff after taking off the  clothe       blood pressure  cannot be read  or the value is  incorrect     Air leakage for the cuff    Change the new cuff       The connection between the cuff tube  and the plug is not tight    Connect again and ensure its  tightness       Rubber tube of the cuff twisting    Release the tube to keep the  airway smooth          Patients belongs to the one who is not  allowed to measure blood pressure       Take the value repeatedly       Blood pressure measurement error code contrast table                                        Error Code Description  00 No error  06 Wrong cuff  Cuff not connected or not tied   07 Air leakage  08 Air pressure wrong unstable pressure   09 Signal weak  10 Upper measurement  11 Too much motion  12 Over pressure tested  13 Signal saturated  19 Over time testing          4 2 4 Operation  record  displayed and network malfunction                Malfunction   B  phenomena Possible cause Means to dispose  1  Malfunction of parameter modules t  Replace tne module   Initialization 2  The communicational signal line or zus PIO or I  a i    9 communicational signal lines or  connect board is loose or damaged   module  The parameters probe or electrode is  Lead off damaged  or the electrode is BOP La en Pepe  BIELLSSE OT    A electrode   invalidation   The speaker is  1  The key board is damaged     hoarse or invalid     2  Speaker or the line is damaged     Replace th
21. ion monitor lead   standard lead  Gain 5mm mv  10mm mv  Sweep speed 12 5mm s  25mm s  50mm s  Wave gain 5   Range of heart rate monitoring 0  20 350 bpm  Resolution 1 bpm  Precision 5   Alarm setting the limit of alarm  setup range  20 350 bpm   and  leads off alarm display   Aalarm mode audible and visual alarm  and record the data during  alarm for the retrospection  Input resistance  5 MO  CMRR   89 dB  S T detecting range  1 00 1 00mv  Heart disorder analysis NO  Anti polarized voltage  500 mV  Baseline renewing time  lt 5 s after the defibrillation  ECG mode mode 1  mode 2  mode 3  mode 4  Frequency characteristic 0 5Hz 40Hz  Safeguard 4000V high voltage isolation  anti defibrillation  anti     high frequency electrical bistoury    1 4 2 Blood Pressure  NIBP     Method Oscillometric   Mode Manual  Auto  Continuous   Measuring Interval in auto mode 0 480 min     Data unit mmHg   kPa optional   Data storage review 4000 blood pressure value at most   Alarm setup SYS  DIA  MAP the range is the same as parameter  measurement range    Alarm method sound light alarm  and record the alarm status for   review    1 2 PMS8210A Service Manual    Chapter 1 Complete Appliance Introduction    Measuring range  Adult Mode  SYS  DIA  MAP  Child Mode  SYS  DIA  MAP  Neonate Mode  SYS  DIA  MAP  Resolution  Precision    Overpressure Protection  Adult Mode  Child Mode  Neonate Mode    1 4 3 Blood Oxygen  3F   SpO2  Measuring method  Measuring range  Alarm setup range  Resolution    Precision
22. m left to the right  when viewing from IRIS front panel      1       START STOP  Power on off button     2   RECORDER  Start stop printing     3  Y  NIBP  Start stop the NIBP measurement     4   O   MENU  enter the system menu  the main set up button    for the system function   5  bid  FREEZE  Freeze unfreeze the current waveform      6  X  MUTE  Mute turn on suspended the alarm sound    5 1 2 Screen Displaying    There are three parts on the screen  indication  waveform  and parameters field      Indication field   It is on the top side of the screen  can display the bed number  system time date   indication information  and alarm information      Waveform field   This field can show 2 waveforms at most  On each waveform s top left  there is a name   When select these names by the knob  users can control the name  gain  and  monitoring type   All these control is limited to ECG waveform       Parameter field   It is on the upper half of the front panel  Users can select the parameters by the knob   and enter into related menu  and operate each item on the menu to reach the demand     PMS8210A Service Manual 5 1    Chapter 5 Basic instruction    5 2 Monitor Usage  Users should take the steps as below when take measurement     1   2     5 2    Read carefully the user s Operation Manual   Check carefully the possible damages caused by transportation  and bad  connection of power cable  ECG cable  NIBP cuff  SpO2 probe  slots  sockets  and  ports    Press the power on button  w
23. meters of testing and calibration    At least once a year  the following parameters are tested and calibrated in order to  guarantee the accuracy of IRIS  After repair  the various parameters shall re calibration     3 2 1 ECG and respiratory parameters testing    3 2 1 1 Tools  Simulator of human body physiology parameter    3 2 1 2 Test steps  A Connect IRIS to ECG simulator   B Confirm the number of ECG waveforms on the screen  and accord with the selection  of EC menu   C Configure the default values of ECG1   and ECG2  when ECG2 exist    Ensure the waveform of ECG and RESP to be ok   E Set parameter values of simulator are as follows   HR   30  gain   4   RR   15  F Check to the waveform or values of ECG  RESP  HR and RR  are correct   G To change the simulator s configuration   HR   240  RR   120  H Check the ECG  RESP  HR  RR values set with the simulator parameter values  consistent     If ECG leads is off  IRIS should be reported immediately     U    3 2 2 Non invasive blood pressure parameter test    3 2 2 1 Inspection Tools    Non invasive blood pressure simulator    3 2 2 2 Test steps   Calibration function with the use of non invasive blood pressure simulator  in  accordance with the  user manual  give the calibration method and calibrated to  determine the accuracy of measurement of blood pressure pump  lf the calibration    3 4 PMS8210A Service Manual    Chapter 3 Installation and inspection    passes  the following tests are performed   A Simulator and monitor
24. n is work and the  LED of trans board is light  but the LED of power and  main board is not light        The  5VDC power is damaged     Replace the power           4 2 6 Parameters malfunction                      Malfunction Possible cause Means to dispose  phenomena  1  Electrode patch is bad contact   2  CAL self check does not detect  1  Stock the good electrode patch   No ECG waveform square wave  2  Replace the ECG RESP module   3  RL electrode hangs in the air  3  Re plug RL electrode   4  ECG RESP module is damaged   1  The electrode connects error  1  Connect the electrode right   5 2  The AC does not connect the  2  Adopt the three wire power  ECG waveform is    abnormality or ground  IRB   interter  d  3  The electrode mav hang  3  Movethe free electrode patch   4  Filter mode of ECG is error  4  Select the right filter mode   5  ECG RESP module is damaged    5  Replace ECG RESP module   No RESP waveform or 1  The electrode connects error  1  Connect the electrode right   abnormity 2  The patient move frequent  2  Keep patient still     3  ECG RESP module is damaged    3  Replace ECG RESP module   ER  er oh se not The measurement bad connect  Adjust the ECG electrode   correct   The value of TEMP is   1  The probe connects badly  1  Re fix the probe   error 2  The performance of probe is bad   2  Replace the probe        NIBP is not puff     Air windpipe may be twisted     Adjust the windpipe        Sometimes  NIBP can  be not measured     The cuff is loose or the 
25. nce    Chapter 4 Maintenance    4 1 The assembly parts drawing    4 1 1 The front assembly view    4 1 2 The front frame view    Inlet module NIBP Pump    N                 f  4  i  i    modules HA T Main  Control board    l Battery  back board          PMS8210A Service Manual    Chapter 4 Maintenance    4 1 3 The back assembly view    4 1 4 The back frame view    4 2 Malfunction guide    On the process of transport  storage and usage  IRIS may produce some malfunctions   If the instrument has malfunctions in use  the problems can be disposed as shown in the  table below  If the malfunction is not debugged  please contact with our dealer or service  department     4 2 1 Simple and apparent malfunction checking  When you check the malfunctions of the monitor  two problems must be checked first   1  Whether the battery is full  if the monitor own the re changeable battery   2  If it is supplied by the AC  check if the monitor is plugged in and if the power line is  connected to the monitor   If there is no problem with power supply  observe the three power indictor lights on    4 2 PMS8210A Service Manual    1     2     Chapter 4 Maintenance    the front panel to judge where the malfunction exits        blue indicator light for AC  If AC power is supplied  press the    start stop    button for about 2 3 seconds when the  instrument is off  the instruments will be started  and the blue indicator light is on  the  blue DC indicator light is off   If the light is not on  there may
26. nnennnennnenennennennensnenennensnenensenennensnensnsnensnenenseensnsnensnn 1 2  IS Appeatatiet iii e e a ees l 6   Chapter 2 Patient Monitor parameters and theory           sssssesenzenzenzinzinnzanzinzinzznnznnzznzznzzzzzanrznzzzzzanznnzrnzzzzza 2 1  2 1 Summary O AE 2 1  2 2 Parameters and theory iii e Meere 2 2   Chapter 3 Installation and Inspection    3 1  3 1 Instrument installation                                                cc Mb  ms 3 1  3 2 The parameters of testing and calibration    3 4   Chapter 4 Maintenance iii 4 1  CES NI RA  ss 4 1  4 2 Malfunction guide     cooooconococononcnoonnonnnnnononononnnonnnononrn ono nn non cnn nro cnn nn nn nro nn nn nan rr nn nr nan rr none rca nn rnann rr nanrninnss 4 2  4 3 P  nodic Checa O Nic 4 6   Chapter 5 Basic instruction    ico cette ells MM lg Vescssssssensesscsscensenscsaces 5 1  MEE TEE en one re 5 1  5 2 Monitor Usage    47 2    AS AA 5 2   Chapter 6 Cleaning EE encres snreneenesnsenennennsensenrenneensee 6 1  AME EP    SAS 7 RA A A CE   RRR RE T RER 6 1  AS ii sissa 6 1  ASI  A  2    AA RT 6 1   Annex A Electric connect diagram ss A 1   Annex B Electric diagram of modules       ooocoocccononicocococanononncononnonnnnnonannnnonnonnnnnonnnon non nonannn non nonnna non nonncnnnnos B 1    VIII    Chapter 1 Complete Appliance Introduction    Chapter 1 Complete Appliance Introduction    1 1 Overview    IRIS Vital Sign Patient monitor is one of the PMS8 monitor series products of the 3F  Medical Company  and is born on 
27. nt  all other  parts can   t be connected with patients  except the defibrillation ECG cable  supplied by pacemaker of 3F    The high frequency electrical bistoury can   t be touched the electrode when  use it together with the monitor  in order to avoid of burning the patients   Forbidden to place the electrode onto the injured or edematous site to  prevent infection    Blood pressure measurement mode are forbidden to use in the blood  pressure monitoring or test for neonates to prevent injuries    Do not measure the blood pressure on the limbs with catheter or infusion   Do not put on the cuff at or near the wounded position    Local bleeding may be caused when using the blood pressure monitoring  in the patients with severe bleeding tendency  Be careful to use on the  patients with sickle cell disease    Forbidden to place the blood oxygen sensor onto the injured skin   edematous or fragile tissues    Discomfort or pain may be caused by the continuous use of the clip type of  blood oxygen sensor especially in patients with micro circulation disorder   Better not to place the sensor over 2 hours at the same place   Non disposable accessories should be sterilized before it is used on next  patient to prevent cross infection    Opening the apparatus by the unauthorized personnel of our company is  forbidden    Don   t need to note the user when the types of the accessories are  changed    To the disposal of package waste  please deal with by the local waste law       Note esp
28. o measured  position alternately  phototubes will receive the light into electrical signals    2  Signal processing parts  The electric signal is amplified by the measuring amplifier   and after high pass filter and program controlled amplification by A D conversion to  digit  It is through D A switch to control signal baseline  and amplify to the  communication signal  to get an appropriate pulse wave  A D conversion for the  digital quantity is processed by SCM    3  Light emitting diode drivers control parts  led transferring and light degree is  controlled by the sequential circuits and DAC  SCM control the size of the electric  current according to algorithm requirements    4  SCM parts  conclude by CPU  RAM  ROM and interface circuit     2 2 4 Temperature  TEMP  parameters    Temperature measurement is to convert the temperature to the electrical signal by a  sensor  enlarge it by the amplifier  and process into data  The circuit is made of  proportional amplifier which is constituted by operational amplifier  Turned into voltage  signal by the thermistor probe  the temperature is magnified and sent to the A D    converter     1 4 PMS8210A Service Manual    Chapter 2 Patient Monitor parameters and theory    Probe examination electric circuit is constituted to voltage comparator by operational  amplifier  When probe off  the input voltage is lower than the comparison voltage  and  the voltage comparator outputs the low level  when insert the probe  the input voltage is  
29. oard 1   10 M10 55 001 Print Paper 1   11 M10 92 050 Printer Slot 1   12 M10 71 032 AC  DC Input 1 1  M10 45 020   13 GB T 845 1985   Self Tapping M3X8 Screws 1   14 M10 92 040 Name Plate 1   15 UL F002 Functional Ground Port 1   B 2 PMS8210A Service Manual          Annex B Electric diagram of modules    Sos                                                                           No Std Code Name  amp  Spec  Qty   1 M10 70 011 Front Housing Facing 1  2 M10 70 010 Front Housing Cover 1  3 M10 04 010 1 Alarm and Optic sense Eye Board 1  4 M10 04 010 2 Speaker Board 1  5 M10 70 050 Alarm Window 1  6 M10 01 010 Display and Key Board 1  7 M10 02 010 LCD Driver Board 1  8 G240320LTSW   LCD 1  9 M10 70 030 Knob 1  10 M10 70 020 Key 1  11 M10 45 010 Knob Head 1  12 GB41 M3 Screws  amp  Nuts 9  13 7SEG3 0 8  3 8   LED Digital Tube 7  14 SMDLED LED Indicator Light 8  PMS8210A Service Manual    B 3    Annex B Electric diagram of modules                                                       No Std Code Name  amp  Spec  Qty   1 M10 71 010 SD Card Interface Panel 1  2 SD 1GDB SD Card 1  3 M10 55 001 Printer Interface 1  4 M10 07 010 Signals Process Board 1  5 M10 71 020 Signals Interface Panel 1  6 M10 92 013 ECG SpO2 NIBP TEMP interface   1  7 M00 06 020 Pump 1  8 M00 06 050 SD Card Slots 1  9 GB T9074 5 2004 Screws 2  10 M10 92 010 Right Name Plate 1  B 4 PMS8210A Service Manual       
30. patient  moves     Keep patient still and adjust the cuff       The measurement   1     The size of cuff is not fit     error is large     2  NIBP module is bad     1  Select the correct size   2  Replace the NIBP module           No SPO  waveform       The probe or module of SPO  is  damaged        Replace the probe or module        PMS8210A Service Manual    4 5          Chapter 4 Maintenance          The SPO  wave is ia  Tie Dane ee 1  Keep patient still    A 2  The ray of ambience is too A  interfered  strong 2  Weaken the rav of ambience   Measurement of SPO    The patient may be injected staining   Before measuring  exclude the  is not correct  material  factor of the staining material             4 3 Periodic Check    1  The using life of the instrument is designed at 5 years    2  Clean the instrument and accessories often    3  Check the instrument once a year in its using life    4  Calibrate all kind of parameters once a year in its using life    5  Check the accessories once half a year in its using life    6  If the instrument with battery hasn t been used for a long time  then it needs to be    charged up once half a year at least  Otherwise  the battery performance will be  affected or invalid  The means to charge up is that plug into the power socket for  about 4 8 hours at least    4 6 PMS8210A Service Manual       Chapter 5 Basic instruction    Chapter 5 Basic instruction    5 1 Basic instruction    5 1 1 Functional Button    There are such below buttons fro
31. rpm  1 rpm  the bigger one between  2 rpm or  2    Resp rate overruns  asphyxiation  sound light alarm  and record the alarm status for review    PMS8210A Service Manual    Chapter 1 Complete Appliance Introduction    1 4 6 Temp  Channel  Measuring mode  Measuring and Alarm Range  Resolution    Precision    Actualization interval  Average Time Constant    Data unit    1 4 7 Others    Power supply  Ambulance power supply  Internal battery input    Power consumption    1   Thermal   0 50  C  32 122  F   0 1   C     0 1   C  25 45  C     0 2   C  Others    1  Sec      lt  10  Sec     Cie    100 240VAC   10    50 60Hz  3Hz  40VA  9V 15V   6 4V 8 4V    lt AOVA    Display LED 3 2  colorful TFT  LCD  320x240   Input NIBP  ECG  Temperature  SpO2  Output Printer and internet   CLASS     According to the type of protection against electric shock  Class   internal power  supply    According to the degree of protection against electric shock  Anti defibrillation CF  type   According to the degree of protection against ingress of water as detailed in the  current edition of IEC 529  Ordinary equipment sealed equipment without liquid  proof    According to the degree of safety of application in the presence of a FLAMMABLE  ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE  not  suitable for use in the presence of a FLAMMABLE ANAESTHETIC MIXTUREWITH  AIR or WITH OXYGEN OR NITROUS OXIDE     According to the mode of operation  Continuous     PMS8210A Service Manual 1 5    Chapter 1
32. rument is not therapeutic instrument   The instrument must be operated under the direction of the professional  medical staff   All of the monitoring parameters are used as a reference and should not be  used as the clinical diagnosis  For abnormalities occurred  clinical  methods should be used to check out the reasons   The instrument should not be operated in the circumstance with flammable  gas or corrosive gas   Prevent the liquid or electrical conductive substance entering into the  instrument   The instrument must be grounded correctly  and the power supply must be  in accordance with the specified requirement   Delete all the previous data when monitoring a new patient  Only one  patient once   If monitor connects to the other instrument  the leakage circuit must be  tested by qualified biological technician before use  and must accord to  IEC 60601 1   It can be used on many parts of this monitor together  and the security  meets the demand of IEC 60601 1   Make sure the connection is not dangerous to the patient or circumstance  before the monitor connects to the other instrument   Must use the defibrillation ECG cable supplied by 3F  or the monitor can   t  be use together with defibrillator   Checkout the alarm system periodically   Do not touch the patient in defibrillation  Otherwise  it may lead to serious  injury or death     III    All cables must be away from patient   s throat to avoid asphyxia    When using together with pacemaker or other electric equipme
33. sible causes    Means to dispose       The screen is black  and power is not on    the fuse is damaged  the power is damaged  Other components are short     1  Replace the fuse   2  Replace power   3  Confirm the component and replace        The screen is black  when power on    1   2   3   1   2     LCD or mainboard is damaged   The lines of LCD is loose  contact or damaged    1  Replace the LCD or mainboard   2  Re plug or replace the lines       The word is ok but  the waveform is not  continuous    Mainboard or parameters module  is loose contact or damaged     Base on the error display  replace the  module        The operation or  measurement is fault    Mainboard or parameters module  is damaged     Base on the error display  replace the  module           1     The MAINS fluctuates     1  Check the MAINS and GROUND       ine performance pl power    gt  Replace the power board or main  Halt on sometime board or main board is not well   board   3  The connector of power board       or main board is not well        3  Replace or repair the connector        PMS8210A Service Manual    4 3    Chapter 4 Maintenance    4 2 3 Error code displayed on the Screen  The error code is only used in blood pressure measurement in this monitor        Malfunction  phenomena    Possible cause    Means to dispose          Cuff tied incorrectly    Tie the cuff again according to 2 3       The patient   s arm with cuff is moving    Keep the patient quiet when  taking the blood pressure       T
34. the base of the 3F PMS series multi parameter  monitor  Fully considered the increasing need of clinical applications  3F develops IRIS  with the latest technology  making it as a new generated medical electronic product     IRIS monitor can monitor the patient s vital physiology signals  including  ECG  RESP   NIBP  SPO2 and TEMP  It can be used for adults  pediatric and neonatal     IRIS uses 50HZ  100V or 220V main power supply  displaying real time data and  waveform with 3 2 inches high definition color LCD and LED  It can display one  waveform or two and all monitor parameters  can be connected to the central network  monitoring system  It equips with a 50mm thermal sensitive recorder  and optional with  a 8V built in replaceable  amp  rechargeable lithium battery  It has many advantages  full  parameters  compact size  clear observation  convenient usage  and can be used for    emergency monitoring     1 2 Model  MODEL TRADE NAME SPECIFICATION CONFIGURATION  l B NIBP  TEMP  Nellcor SpO    l A NIBP  TEMP  3F SpO    PMS8210A IRIS g  a    II B NIBP  TEMP  Nellcor SpO2  ECG RESP  II A NIBP  TEMP  3F SpO gt   ECG  RESP    1 3 Tag Structure    PMS8210A II II  Product configuration code     Without ECG  Il  With ECG RESP rate    Product type code  Vital Signs Patient Monitor  LCD LED    IRIS PMS8210A   s trade name    PMS8210A Service Manual 1 1    Chapter 1 Complete Appliance Introduction    1 4 Parameters    1 4 1 ECG  Lead 3 lead RA LA LL    5lead RA RL LA LL V    Lead opt
35. unction properly  the  screen display is OK  and no error message  the two waveforms can show correctly the  parameters are refreshed every second  the tracking time is precise too     3 1 3 Configuration    Users can setup related function in    MENU     including the switch of the recorder  ECG   type of the lead  network bed number setting and the switch of arrhythmia     3 1 4 Running Normally    1  When operate any key  the patient monitor can run according to the operation  manual s guidelines    2  When rotate or press the knob  the patient monitor can run according to the  operation manual s guidelines    3  Operate the patient monitor as the user   s manual  and check sound and audio alarm  normal or not  alarm mute and turn off alarm is working or not    4  Check the machine to make sure that the memory function is OK     3 1 5 Network function    1  Start Central Station and the bedside monitor  make sure the net is connected  and  the bed number shows correctly    2  The transfer of data between IRIS and central station normally    3  Central Station receives waveform parameters  alarm level  alarm limits  and the    3 2 PMS8210A Service Manual    Chapter 3 Installation and inspection    patient information correctly    4  Waveform parameters  alarm level  alarm limits  and the patient information correctly  turned to the central station    5  IRIS portable shutdown correctly after the Central Station response time required   see the relevant technical indicators 
    