Home
        Scout Pro Service Manual
         Contents
1.                us                                               LCD DISPLAY e    C4 a us    CAL ES                                               Figure 3 4  Transducer Removal   3  Disconnect Connector J6 coming from Transducer  8      4  Install the replacement Transducer using the 4 screws previously removed and connect the  Transducer flexible cable to connector J6 on the Printed Circuit Board     5  Reassemble the balance   6  See Appendix A Service Tools Instructions and load temperature compensation data     7  Recalibrate the balance in accordance with instructions in Appendix B  Factory Calibration  B 2    8  Check the performance of the balance   3 4    CHAPTER 4 TESTING    4 1 TESTING   Before servicing the balance  an operational test and various performance tests should be made to  ascertain whether or not the balance meets specifications  Turn the balance on and allow it warm up  for at least five minutes before performing these tests  Make sure the test area is free from drafts and  thatthe balance rests on a level and vibration free surface  The masses used for final calibration must  be adjusted to ASTM Class 4 tolerance or better     4 1 1 Operational Test    1  Connect a functioning AC Adapter to the balance Power Jack located at  the rear of the balance     2  Plug the AC Adapter into a suitable power source  If the AC Adapter supplied  with the balance is rated for a different voltage  use an appropriate adapter to match the  supply voltage   4 1 2 Segment
2.             Figure    4   A 3    APPENDIX A SERVICE TOOL INSTRUCTIONS    A 5 USING CONFIGTOOL EXE   CONT      2  Ifthe communications parameters have to be set  click on the Com Porttab andthe screen  shown in Figure A 5 will appear  Adjust the parameters as required to match the balance     El ScoutPro Service Tool  x   Config  Test Command        port         m Settings       Port  com 1    Baudrate  2400 y  ByteSize        Parity  one y   DK         Figure    5     3  When the communications parameters are matched click on the Test Command tab and  test the communications     4  inthis example  see figure A 6  the  P  command resulted in the 0 029 being written in the  lower screen  The  V  command resulted in Sr  1 10 being written  The T command had  the effect of zeroing the balance  and the last P command resulted in the 0 00g being  written  If the balance does not reliably communicate the problem must be fixed before  continuing     E ScoutPro Service To OOO yyy    Config Test Command   com port            Figure A 6     A 4    APPENDIX A SERVICE TOOL INSTRUCTIONS    A 5 USING CONFIGTOOL EXE   CONT      5  When communications are reliable return to the Config tab  Press the read button  A  and  the data will be read from the balance  See Figure A 7        Config   Test Command   COM port                          User Enable Unit  Fet  Ge            JE Heatm MOr        I wt Polite     gt    Pe     Poe     Custom quotiety   Unit T select  ose      r User Menu Related  
3.        OHAUS  SERVICE MANUAL    SCOUT   PRO BALANCES    Ohaus Corporation 19A Chapin Road  P O  Box 2033  Pine Brook  NJ 07058 2033  973  377 9000           OHAUS  SERVICE MANUAL    SCOUT  PRO BALANCES       The information contained in this manual is believed to be accurate at the time of publication  but Ohaus  Corporation assumes no liability arising from the use or misuse of this material  Reproduction of this  material is strictly prohibited     Material in this manual is subject to change     O Copyright 2003 Ohaus Corporation  all rights reserved      Registered trademark of Ohaus Corporation     TABLE OF CONTENTS    CHAPTER 1 INTRODUCTION FARA   1 1                        ee              1 1   12  Service FACIES ui ida 1 1   1 3 Tools and Test Equipment Required    1 2   1 3 1 Special Tools and Test Equipment List                ii 1 2   132 Standard Tools and Test Equipment Lisi  e 1 2   1 4 Test Masses ReQquirSd     aueh u 1 2   13  SPECIICAlONS m AS 1 3  CHAPTER 2 DIAGNOSIS   di TroublBehoelndg susi 2 1   22  Diagnostic    ela 2 1   A           ae 2 1   23 Error ONS artes                 2 4    CHAPTER 3 REPAIR PROCEDURES    spl  Repair                                              een 3 1       EET MILI SEE 3 1  3 1 2 Printed Circuit Board Replacement        iioc renta rr nein ba cir rin 3 2  3 1 3 Transducer Replacement       ioci anne 3 4  CHAPTER 4 TESTING  41  ile 4 1    El Operation Test mrt 4 1  43 2  Segment Display Test  sont 4 1  L2  Pellormanes Tests sust
4.      2  Select a mass weighing near the maximum capacity of the balance and place the mass on  the center of the Pan  1   Observe and record the reading     3  Remove the mass  Reading should return to Og  1 count     4  Repeat this test for ten readings  If the standard deviation of the readings is less than  1  count  the balance passes the Repeatability Test     Adjustment   If the deviation for any set of readings  using the same mass placed on the center of the Pan  is greater  than  1 count  the balance does not meet the Repeatability Test specification  Inspect and correct the  following areas     1  Check for mechanical obstructions  Any foreign object touching any part of the moving Pan  can cause a balance to fail the Repeatability Test  Inspect and correct as necessary     2  An error inthe Off Center Load Test can affect the results of the Repeatability Test  Inspect  and correct if necessary  See Off Center Load Test     3  Foreign material or debris located in the balance between the Pan  1  and the Top Cover   6  can cause the balance to fail the test     4  Environmental influences such as vibrations  drafts or a non level surface can also cause  failures     5  Ifthe balance fails the test  refer to table 2 1 Diagnostic Guide for assistance     4 2    CHAPTER 4 TESTING    4 2 2 Off Center Load Test  The Off Center Load Test is used to determine whether displayed weight values are affected by  moving the sample to different areas of the Pan  6      Test  1  Pl
5.      O               OUI  Je     PS    nm       11     Enter the Windows Control Panel and click on the System Icon   Click on the Hardware tab    Click on the Device Manager button    Expand Ports  COM  amp  LPT     Right Click on USB Serial Port  COM5  and choose Properties   Click on the Port Settings tab    Click on the Advanced button    Change the COM port number to an unused port    Click on OK twice    Close the System Properties box    Close the Control Panel Window     The virtual port will now be the com port you selected     A 5 USING CONFIGTOOL EXE     Ti    Open ConfigTool exe  you will see the screen shown in Figure A 4 after the communica   tions parameters have been checked  If the communications parameters are not matched   or  there is no balance connected  error messages will appear  If a balance is connected   it will be read automatically and the data will appear    El 2272  x    Config   Test Conmand   COM port       User Enable Unit      Tice                   Newton I Oz   0   Det   Lb        Pe I      Pe  enh      Custom divisor   quotiety   Unit T select   y       User Menu Related    Sample size         Auto Off Power On Unit      AZT       Mode   y         Stability   y      85232 Menu Related    Bandrate   Y  I Ser  Enable Parity   Y   Handshake   y   Auto Print   y     Print Stable      Serial Number    C Scale     Loadcell   OK                              0  7 59  20 00  60  00  83  35  100     50 00   Model  ET Reset Cal Lin Read Update Quit 
6.   Refer to Figure  5 1  Before disassembling the balance  it should be noted that components inside of the balance are  delicate and need to be handled with care  It is imperative that the Transducer  8  should never be  subjected to any excessive torque  stress  or abrasion as damage may result  Once the balance has  been disassembled  any small scratch or abrasion made to any of the Individual Strain Gauges will  render the entire Transducer  8  unusable  The Printed Circuit Board  9  contains integrated circuits  which employ CMOS technology  therefore  caution must be exercised so as not to subject any of these  components to static electricity discharge  When servicing  a wrist Ground Strap with a 10 Megohm  series resistor to earth ground is highly recommended  The Printed Circuit Board  9  should be  handled by grasping the edges only and never placing fingers on any of the runs or traces     3 1 1 Cover Removal  To disassemble the balance  proceed as follows     1  Turn the balance off and if using an AC Adapter  disconnect it from the balance   2  Remove the Pan  1  2 or 3  from the balance  Pan sizes and shapes vary   3  Place the Shipping Lock in the locked position     See Figure 3 1  This will help protect the  Transducer     Unlocked  Position       Figure 3 1  Shipping Lock     4  Ifabattery is being used  turn the balance over and remove the Battery Compartment Cover   12  as shown in Figure 3 2     Cover Screws       Figure 3 2  Bottom of Balance     3 1    C
7.  Display Test    1  Turn the balance on by pressing ON ZERO Off  all segments are enabled and displayed  briefly  then followed by a software revision number  See Figure 4 1 for full display     2  Tare the balance  The display should indicate a zero weight     51 III DEN    DODO    Figure 4 1  LCD Full Display        4 2 PERFORMANCE TESTS   Accurate performance of a Scout Probalance is determined by a series of three performance tests  The  displayed readings are compared with the tolerances listed in Table 4 1  Tolerance values are  expressed in counts  A one count change is equal to the last digit shown on the balance display     TABLE 4 1  TYPES OF PERFORMANCE TESTS  LPEBFORNMANGE es   TEST   REFRANES       4 1       CHAPTER 4 TESTING    4 2 PERFORMANCE TESTS  CONT     The following performance tests are used to evaluate the balance operation before and after repairs   Each balance tested must meet the requirements specified in each test as well as the specifications  listed in Table 1 2  Tolerance values are expressed in counts  A balance which passes each ofthese  three tests meets the manufacturing specifications     4 2 1 Repeatability Test   Repeatability is a word used in balance specifications meaning the Standard Deviation of a set of  similar weight readings  To determine whether a balance meets the calculated Standard Deviation  value in the Specification Table 1 2  perform the following test     Test  1  Tare the balance  The reading on the display should be 00
8.  Extract the files in ScoutPro Tools zip into C  ScoutPro     5  Copythe file service txt into the C  ScoutPro  if required     6  You may want to put a shortcut on your desktop for easier access     A 1    APPENDIX A SERVICE TOOL INSTRUCTIONS    A 3 HARDWARE INSTALLATION  NOTE  The optional interface must be modified so that the EEPROM may be written to     1  Remove the screw and the housing as shown in  Figure A 1        Figure A 1  Screw and Housing Removal     2  Locate J3 on the interface board  See Figure A 2        3  Short the two pins of J3  see Figure         4  Replace the interface housing     5  Install the modified RS232   USB module in  the balance being repaired     NOTE  If using the customers interface  do not  forget to remove the modification after the  repair is complete        6  Plug the interface cable into the PC  pa    Figure A 3  Connector J3 Shorted     7  Make sure the communications parameters match  The parameters in the tools are   e Baud   2400  e Parity   7 None  e Handshake   None    8  Make sure the balance A off feature  under the  S E T U P  menu  is set to OFF  if this  setting is left ON  the balance will shut off during the re configuration     A 2    APPENDIX A SERVICE TOOL INSTRUCTIONS    A 4 SPECIAL INSTRUCTIONS FOR USB INTERFACE   After installing the interface using the instructions that come with the kit  the COM port will have to be   changed  The interface uses COM 5 as a default  Config Tool only recognizes COM 1 to COM 4   1
9.  Scout Pro  All   Adapter  230V AU  Scout Pro  All   Adapter  240V UK  Scout Pro  All   Adapter  230V EU  Scout Pro  All   Adapter  100 120V US  Scout Pro All    5 4    APPENDIX A SERVICE TOOL INSTRUCTIONS    A  SERVICE TOOLS INSTRUCTIONS  This tool is required when either a main PC board or a transducer is replaced in a Scout Pro balance   It is used to re configure the balance to it   s old parameters in the case of a board replacement or new  parameters in the case of a transducer replacement   A 1 REQUIREMENTS FOR SCOUT PRO TOOLS   e VGA 640x480 or higher resolution screen     Microsoft Windows NT 4 0 or later  or Microsoft Windows 98 or later   e Scout Pro RS232 Interface Kit or USB interface Kit       Scout Pro instruction manual       Scout Pro Tools zip  A 2 SOFTWARE INSTALLATION   1  If you need the software it can be downloaded at the following URL  HTTP      fileshare ohaus com docushare    a  Click on Login   b  Login   ServiceA  Password   aservice  Both are case sensitive  c  Click on the Xerox logo at the upper left   d  Click on Technical Support   e  Click on Service Notes   f  Click on Authorized Service Centers   g  Click on Scout Pro Tools   h  Download all 3 files    2  Create a folder in the C  drive and name it ScoutPro  If you choose another directory you  will have to open the file toledo ini and modify the path to the service txt file to agree with  the folder name and path you have chosen    3  Copy the file ScoutPro Tools zip into C  ScoutPro    4 
10.  This service manual contains the information needed to perform routine maintenance and service on  the Ohaus Scout    Pro Balances  The contents of this manual is contained in five chapters and are  listed as follows     Chapter 1 Introduction   Contains information regarding service facilities  tools  test equipment   calibration masses and specifications     Chapter 2 Diagnosis   Contains a diagnostic guide for troubleshooting problems and error code  tables     Chapter 3 Repair Procedures   Contains disassembly assembly and replacement procedures   Chapter 4 Testing   Contains an operational test  a segment display test and performance tests   Chapter 5 Drawings and Parts Lists   Contains an exploded view drawing and a parts list   Before servicing the balance  you should be familiar with the Instruction Manual which is packed with  every balance  The procedures in this manual assumes the technician performing them has a working    knowledge of the use of standard hand tools and the repair of precision instruments     1 2 SERVICE FACILITIES  To service the Scout  Pro balances  the service area should meet the following requirements     e Should be temperature controlled and meet the balance specifications for temperature  environmental requirements  See specifications for temperature ranges of the various  models     e Must be free of vibrations such as fork lift trucks close by  large motors  etc       Must be free of air currents or drafts from air conditioning heating 
11.  and use care when handling  components of the balance     2 2 DIAGNOSTIC GUIDE   Table 2 1 is a diagnostic guide designed to help locate the problem area quickly and easily  To use  the table  first locate the symptom that you are observing  Followthe symptom column and review the  probable cause column and remedy column  The probable causes are listed with the most common  cause first  If the first remedy does not fix the problem  proceed on to the next remedy  Before  attempting to repair the Balance  read all chapters of this manual to familiarize yourself with the  balance components and operation  Do not attempt repairs unless you fully understand the operation  of the balance     2 2 1 Diagnosis  1  Isolate and identify the symptom   2  Refer to Table 2 1 Diagnostic guide and locate the symptom   3  Follow the suggested remedies in the order that they appear   4  Perform the indicated checks  or see the appropriate section of the manual   5  Repair or replace the defective section of the balance   NOTE   If more than one symptom is observed  it is necessary to  approach one area at a time  and also remember  that the  symptoms may be interrelated   In the event that erratic or fluctuating weight readings are observed  it is necessary to isolate the  problem to either the mechanical area or the electronic area of the balance  The repeatability test will  quickly point out whether the Transducer  8  is operating properly or whether the problem is due to an    electronic 
12.  replacement  4000g  Scout Pro  SP4001  SPX4001  Kit  Transducer replacement  12000  Scout Pro   Kit  Transducer replacement  600g  Scout Pro  SPx01  SPXx01   Kit  Transducer replacement  400g  Scout Pro  SP402  SPX402  Kit  Transducer replacement  2000  Scout Pro  SP202  SPX202  Kit  Transducer replacement  120g  Scout Pro  SP123  SPX123    Kit  PCB w LCD replacement  Full Bridge Hi  Scout Pro  For 120g   400g  4000g  SP123  SPX123  SP402  SPX402  SP401  SPX401    Kit  PCB w LCD replacement  Half Bridge  Scout Pro  For 600g   6000g  SPx01  SPXx01  SP6000  SPX6000    Kit  PCB w LCD replacement  Full Bridge Lo  Scout Pro  For 200g   1200g  2000g  SP202  SPX202  JS1200  SP2001  SPX2001    5 3    CHAPTER 5 DRAWINGS AND PARTS LISTS    TABLE 5 1  REPLACEMENT PARTS LIST  Cont      71164070  71160301  71160412  71160305  71160303  71160304  71147994  71160313  71160323  12102033  71160309  71160413  71160311  71160312  12102323  12102322  12102321  12102320    Base w threaded inserts  Scout Pro  SPGxxx  SPSxxx  SPxxxN  Base  Scout Pro  All with fixed feet   Adjustable feet  with rubber pad  Scout Pro  SPGxxx  SPSxxx  SPxxxN  Foot  Non adjustable  Scout Pro  All   Battery cover  Scout Pro  All   Interface cover  Scout Pro  All   LCD  Scout Pro  All   Cal Lock  Scout Pro  All   Connector  Battery  Scout Pro  All   Weigh Below cover  Scout Pro  All   Lock  sub platform  Scout Pro  All   Level  Scout Pro  SPGxxx  SPSxxx  SPxxxN   LCD support  Left   Scout Pro  All   LCD support  Right  
13. E  Cont      SYMPTOM PROBABLE CAUSE   REMEDY       4  Defective Transducer   4  Replace Transducer   5  PC board defective  5  Replace PCB     1  Defective Transducer   1  Replace Transducer     Displays  Err2   Cont       Exceeds off center load limit     Inaccurate readings     Fluctuating readings     Drifting display     1  Improper calibration   Span or Linearity      2  Not zeroed before  weighing     Defective Transducer     PC Board defective       Unstable environment     Debris in balance       PC Board defective     Defective transducer       Transducer Lock en   gaged or misaligned       Balance not warmed up       Defective Transducer     PC Board defective     1  Perform linearity calibra     tion     2  Press Zero On with no    weight on the Pan  then  weigh       Replace Transducer     Replace PC Board       Check area for vibrations     leveling and drafts       Disassemble and clean if    necessary       Replace PC Board     Replace Transducer       Unlock or realign       Allow balance to stabi     lize       Replace Transducer     Replace PC Board        Error code  Err   on display       Balance has detected an  error condition     2  Defective Transducer   3  PC Board defective        2 3      See Table 2 2  Error    Codes       Replace Transducer     Replace PC Board     CHAPTER 2 DIAGNOSIS    TABLE 2 1  DIAGNOSTIC GUIDE  Cont      SYMPTOM PROBABLE CAUSE  RS232 Not functioning     wa models equipped with   232      tings     2  RS232 not turned on   3  
14. HAPTER 3 REPAIR PROCEDURES    3 1 1 Cover Removal  Cont    5  Remove the 4 AA batteries from inside the battery compartment and replace the Battery  Compartment Cover  12      6  Slide the Lockswitch to to the unlocked position to gain access to one of the cover screws   Remove the 3 Cover Screws as shown in Figure 3 2     7  Turn the balance over in an upright position and carefully lift the Top Cover  6  from the  balance Base  10      8  After repairs and or adjustments have been made  reassemble the balance in the reverse  order     3 1 2 Printed Circuit Board  9  Replacement  In an effort to keep service costs down  it is suggested that if the Printed Circuit Board  9  is suspected  of being faulty  it should be replaced rather than repaired     To replace the Printed Circuit Board  9    proceed as follows   1  Remove the balance cover  refer to paragraph 3 1 1     CAUTION  WHEN HANDLING THE PRINTED CIRCUIT BOARD   HANDLE BY EDGES ONLY  DO NOT TOUCH FOIL SIDE  OF BOARD  STATIC DISCHARGE MAY DAMAGE SOME  COMPONENTS     2  The Printed Circuit Board  9  is fastened in place by 2 clips which are part of the Base  10   and 2 screws located on top of the Printed Circuit Board  In addition  a Calibration Lock   15  small plastic piece  is positioned at the front of the Printed Circuit Board  See Figure  3 3                 10 BASE                      J6 CONNECTOR                                              CLIP   T T xi EUR CLIP    J2 BATTERY je rom S J11 RS232 CONNECTOR  CON
15. NECTOR Wee a  16                                                                                  MOUNTING MOUNTING  SCREW 9 15 SCREW  PCB CAL LOCK    Figure 3 3  Printed Circuit Board Removal        3 2    CHAPTER 3 REPAIR PROCEDURES    3 1 2 Printed Circuit Board  9  Replacement  Cont      3     When an RS232 option is installed  remove the RS232 Connector from J110n the Printed  Circuit Board  9      Disconnect the Transducer cable from Connector J6 on the Printed Circuit Board  9    Disconnect the battery cable from Connector J2 on the Printed Circuit Board  9     Remove the 2 mounting screws on top of the Printed Circuit Board  9     Push the Printed Circuit Board towards the back as far as it will go and apply a small forward  pressure to the Cal Lock  15   Lift up on the front of the Printed Circuit Board  9  to remove  it     Install the new Printed Circuit Board  9   Make sure that the Printed Circuit Board is  positioned under the 2 clips on the Base  10      Connect the flexible cable from the Transducer to Connector J6 on the Printed Circuit  Board  9      10  Connect the RS232 to Connector J11 on the Printed Circuit Board  9  if the RS232 is    installed     11  Connect the battery cable to connector J2 on the Printed Circuit Board  9      12  Replace the 2 mounting screws on top of the Printed Circuit Board  9   make sure the board    is positioned as far forward as possible to ensure the calibration lock switch will operate  properly     13  Reassemble the Bala
16. RS232 does not stay on   4  Loose or disconnected    RS232 option   5  RS232 option defective        2 3 ERROR CODES  Scout Pro balances are equipped with software which will display an error condition when it occurs   Table 2 2 Error Codes  describes the various error codes which can appear on the display and  specifies the probable reason and remedy     1  Improper Print Menu set     1  Check all settings   Menu  item Power must be set  to ON      2  Turn the option on   See  instruction manual      3  If batteries are being  used  the RS232 turns  off when the unit is turned  off  Either turn the option  on or use the ac adapter     4  Check connection   5  Replace RS232 option                                TABLE 2 2  ERROR CODES  DISPLAY REASON ACTION  Err 1 Invalid checksum data 1  Perform factory calibration   2  Restore data file   3  Check transducer   4  Check PCB   Err2 Overload or Underload 1  Make sure the right pan is on the balance   2  Perform service LIN calibration per B 1   3  Check transducer   4  Check PCB   Err 4 Incorrect calibration mass 1  Calibrate using correct mass   Err5 Parts counting error 1  Sample is less than 1d  Change sample quantity  or increase sample   Err8 RS232 buffer is full 1  Check printing device   Err9 Internal data error 1  Restore data file   2  Check PCB   3  Check transducer        2 4       CHAPTER 3 REPAIR PROCEDURES    3 1 REPAIR PROCEDURES   This section of the manual contains detailed disassembly procedures of the balance
17. Repeatedly press and release the PRINT Unit button until the display shows END     Press and release the On Zero Off button  The balance will return to the weighing mode     B 2 FACTORY CALIBRATION  This calibration must be used when either a PC board or a transducer is replaced     1     2     Remove power from the balance     Install an RS232 option with the pins shorted  See Appendix        If the RS232 option is  not available  pins 2 and 7 of IC9 may be shorted using a micro alligator clip  See Figure  B 1     CAUTIONS    1  Be very careful that no other pins are shorted  this could destroy the IC   2  Do not attempt to solder the jumper     B 1    APPENDIX B FACTORY CALIBRATIONS    B 2 FACTORY CALIBRATION  CONT                                   E             S    PDS          SF    J6    O              PIN 24  PPIN 7    IC U9    LCD DISPLAY          CAL           Figure B 1  IC U9 Connections     You may need to use some kind of draft shield since the top cover has to be left off   Apply power to the balance   Start with the unit off     Press and hold both the ON ZERO Off and the PRINT Unit buttons until the display reads  rAMP     Press and release the PRINT Unit button  The display will show Lin     Press and release the ON ZERO Off button  The display will show  C  blinking and then  the first mass value     Place masses totaling this value on the balance and press the ON ZERO Off button     Press and release the ON ZERO Off button  The display will show  C  blinkin
18. Sample size  10 pes y     Auto ose Power On Unit  c         0 5 d Y  Mode  ose z ri  Stability 1d zl    RS232 Menu Related  Bandrate  2400       Ser  Enable Parity T None v Handshake  None     Auto Print fore       Print Stable       Serial Number     T Scale        Loadcell   0K                 49 85                             Figure A 7   6  Ifthe PC Board has been replaced  refer to           7  Ifthe load cell was replaced  follow the same procedure as above except enter the load  cell serial number and click on Load cell  C  in the serial number box     8  The bar  G  is a graphical indication of the portion of the transducer output that is being  used     A 6 USING CONFIG TOOL AFTER REPLACING THE MAIN PC BOARD    1  Using ConfigTool after a new Main PC board has been installed  make sure the balance  lockout switch is in the unlocked position     2  Install an RS232 option with the factory jumper  J2 on the RS232 option  in the balance   3  The AC adapter must be used    4  Turn the balance on     A 5    APPENDIX A SERVICE TOOL INSTRUCTIONS    A 6 USING CONFIG TOOL AFTER REPLACING THE MAIN PC BOARD  CONT      The balance displays Err 9   Start the ConfigTool software  see A 5    An error message will be displayed    User Enable Units  amp  User Variable Check Sum  Verify Error     click on OK     Set RS232 to the correct com port  Baudrate   2400  ByteSize   7  Parity   None   The balance will respond to the    V    command but not the others    The bar graph will resp
19. ace 1 2 of the balance capacity in the center of the Pan  1      2  Note the reading     3  Move the mass halfway  between the center and the edge  to the front ofthe Pan  1   Note  any differences in the displayed weight reading     4  Repeat this test for the back  left  and right positions of the Pan     5  Maximum allowable change in displayed weight readings is   2 count for each of the four  positions  If this reading is exceeded  it usually indicates a defective Transducer  8      4 2 3 Linearity Test  This test is used to determine the linearity of the unit throughout its operating range  The masses used  to perform this test must be adjusted to ASTM Class 4 Tolerance  or better     NOTE   The balance must pass the Off Center Load Test and Re   peatability Test before the Linearity Test may be performed     Test  Loads do not have to be test weights  They can be anything that totals the load value  The test mass  can be anything that weighs near the test mass value   TABLE 4 2  TEST MASSES                      Capacity Test mass Load 1 Load 2 Load 3 Load 4  200g 1g 50g 100g 150g 190g  400g x 0 01 1g 100g 200g 300g 390g  400g x 0 1 5g 100g 200g 300g 400g  600g 5g 150g 300g 450g 600g  2000g 5g 500g 1000g 1500g 2000g  4000g 5g 1000g 2000g 3000g 4000g  6000g 10g 1500g 3000g 4500g 6000g                1  Place the test mass on the balance  record the weight and then remove   2  Place Load 1 on the balance and press the ON ZERO Off button   3  Place the test mass on the bal
20. ance  record the weight and then remove     4  Place Load 2 on the balance and press the ON ZERO Off button     CHAPTER 4 TESTING    4 2 3 Linearity Test  Cont     Test  Cont    5  Place the test mass on the balance  record the weight and then remove   6  Place Load 3 on the balance and press the ON ZERO Off button   7  Place the test mass on the balance  record the weight and then remove   8  Place Load 4 on the balance and press the ON ZERO Off button     9  Place the test mass on the balance  record the weight and then remove     10  The difference in the weights of the test mass should be within the tolerance in table 4 1   If not  perform a linearity calibration  see Appendix B and do the test again     4 3 Down Stop Adjustment   When the balance fails the Off Center Load test and or the Linearity test  the Transducer may require  checking and adjusting of the down stop  To make adjustments  the Transducer must be removed from  the balance  Perform procedures 3 1 1 and 3 1 3 and remove the Transducer from the balance   NOTE  This procedure will require feeler gauges ranging from 0 40mm 0 16in  to 0 65mm 0 026in   Checking and adjusting the down stop     1  Place masses equaling mass 1 as indicated in Table 4 3 on the platform  You should still  be able to feel some movement with a slight downward push     2  Place masses equaling mass 2 as indicated in Table 4 3 on the platform  You should feel  no movement with a slight downward push     3  Ifthe above conditions are no
21. antivo 4 1  4 2 1  Bepestabllily TOL au duo tina ia 4 2  42 2 Off Center Load Testori iii bei 4 3  4 23 Lineanty A A 4 3  43  CDOWmotop AdUSINCIE  aaa 4 4  CHAPTER 5 DRAWINGS AND PARTS LISTS  5A IDEA WIN           di                           5 1  5 2 Pans Liste                                        5 3  APPENDIX A SERVICE TOOL INSTRUCTIONS  A   Sevice Tools Ihsirlielians        re A 1  A 1 Requirements for Scout Pro Tools u    nu A 1  A2 Solware Insrallallgb         die A 1  AS Hardware Installation    ass a li A 1  AA Special Instructions for USB Interface    nennen A 2  AS Using  dre quietos Nus  cR Tm A 3  AB Final Service Calibration eiii A 5    APPENDIX B FACTORY CALIBRATIONS    B     TABLE NO   1 1 Calibration Points  1 2 Specifications  2 1 Diagnostic Guide  2 2 Error Codes  4 1  4 2 Test Masses  4 3 Test Mass Values  4 4  5 1 Replacement Parts List  FIGURE  3 1 Shipping Lock  3 2 Bottom of Balance  3 3  3 4 Transducer Removal  4 1 LCD Full Display  4 2  5 1  A 1  A 2 Connector J3 Location  A 3 Connector J3 Shorted  A 4 Screen 1  A 5 Screen 2  A 6 Screen 3  A 7 Screen 4  B 1 IC U9 Connections    Factory Calibrations  B 1 Service Calibration  B 2 Factory Calibration    Types of Performance Tests    Transducer Front View   Down Stop Adjustment  Scout Pro Exploded View  Screw and Housing Removal    LIST OF TABLES    TITLE    LIST OF ILLUSTRATIONS    TITLE    Transducer Adjustment Tolerances    Printed Circuit Board removal       CHAPTER 1 INTRODUCTION  1 1 INTRODUCTION 
22. chnician will be available  to provide assistance  Outside the U S A   please contact     Ohaus Corporation   19 Chapin Road   Pine Brook  NJ 07058  USA  www ohaus com   Tel   973  377 9000    Fax   973  593 0359    5 1    CHAPTER 5 DRAWINGS AND PARTS LISTS                                                                                                                                        Figure 5 1  Scout Pro Exploded View     5 2    5 2 PARTS LISTS  This section of the manual contains the replaceable parts for the Scout Pro Balances     KEY NO            OO                               N QO Ci dI A OQO    CHAPTER 5 DRAWINGS AND PARTS LISTS    TABLE 5 1  REPLACEMENT PARTS LIST    PART NO   71160314    71160308    71154611  71147378  71160395  71160346  71160302  71160347  71160477  71160408  71160407  71160406  71160405  71160404  71160403  71160402  71160411    71160410    71160409    DESCRIPTION    Square Pan  6 5  x 5 5  SS  Scout Pro  SPx01  SPxx01  SPxx00   SPXx01  SPXxx01  SPXxx00    Pan Support  6 5  x 5 5  Square  Scout Pro  SPx01  SPxx01  SPxx00   SPXx01  SPXxx01  SPXxx00    Round pan Assy  4 7   120mm  dia  SS  Scout Pro  SPx02  SPXx02  Specific Gravity Kit  Scout Pro  SPExxx   Function Label  Scout Pro  SPExxx   Function Label  Scout Pro  All but SPExxx   Top Cover  Scout Pro  All   Wire Assembly  Power  Scout Pro  All   Kit  Transducer replacement  2000g  Scout Pro  SP2001  SPX2001  Kit  Transducer replacement  6000g  Scout Pro  SP6001  SPX6001  Kit  Transducer
23. ducts  open windows   people walking by  fans  etc       Area must be clean and air must not contain excessive dust particles   e Work surface must be stable and level     e Work surface must not be exposed to direct sunlight or radiating heat sources     i3    CHAPTER 1 INTRODUCTION  1 3 TOOLS AND TEST EQUIPMENT REQUIRED    In order to properly service the Scout Pro balances  certain Ohaus special tools and test items are  required in addition to standard electronic tool kits  These items are listed as follows     1 3 1 Special Tools and Test Equipment List    1  Alternate voltage Power Adapter if local power requirements do not match Balance  Adapter voltage ratings     2  Scout Pro RS232 option PN 71147376    3  APC running Microsoft Windows NT 4 0 or later  or Microsoft Windows 98 or later   3  Scout Pro instruction manual    4  Scout Pro Tools zip    5  Set of Feeler gauges ranging from 0 40mm 0 16in  to 0 65mm 0 026in     6  Micro sized alligator clip     1 3 2 Standard Tools and Test Equipment List  1  Standard Electronics Tool Kit   2  Digital Voltmeter  DVM  Input impedance of at least 10 megohms in the 1 Volt dc position   3  Masses totaling up to 6000 grams are required  Ohaus makes various calibration sets    available  Please contact your nearest Ohaus dealer for further details     1 4 TEST MASSES REQUIRED  The masses required to test the Scout Pro balances must meet or exceed the requirements of ASTM  Class 4 Tolerance  The calibration points are listed in Tabl
24. e 1 1     TABLE 1 1  CALIBRATION POINTS                      CAPACITY SPAN CAL POINT   LINEARITY CAL POINTS  200g 200g 100g  200g  400g 200g 200g  400g  600g 300g 300g  600g  2000g 2000g 1000g  2000g  4000g 4000g 2000g  4000g  6000g 6000g 3000g  6000g       1 5 SPECIFICATIONS    CHAPTER 1 INTRODUCTION    Specifications for the Scout Pro balances are listed in Table 1 2  When a balance has been serviced   it must meet the specifications listed in the table  Before servicing the balance  determine what    specifications are not met     TABLE 1 2  SPECIFICATIONS                                           Capacity  g  200 400 400 600 2000   4000 6000  Readability  g  0 01 0 1 1 0   Repeatability  Std  Dev    g  0 01 0 1 1 0   Off Center Loading  g   0 01  0 1  1 0   Linearity  g   0 01  0 1  1 0   Tare range To capacity by subtraction   Over Range Capacity Capacity   90d   Stabilization Time 3 seconds       Operating Temp  Range    50   to 104   F 10   to 40   C       Power Requirements    AC Adapter  included  4 AA Batteries  not included        Item Numbers       SPx202    SPx402   SPx401   SPx601       SPx2001  SPx4001  SPx6000          CHAPTER 1 INTRODUCTION    CHAPTER 2 DIAGNOSIS    2 1 TROUBLESHOOTING  This section of the manual specifies problem areas of the balance which can occur  Information is  contained to isolate specific problems using Table 2 1  Diagnostic Guide  and Table 2 2  Error Codes   Follow all directions step by step  Make certain thatthe work area is clean
25. g and then  the second mass value     Place masses totaling this value on the balance and press the ON ZERO Off button     Press and release the ON ZERO Off button  The display will show  C  blinking and then  donE followed by Lin        APPENDIX B FACTORY CALIBRATIONS    B 2 FACTORY CALIBRATION  CONT      13     Repeatedly press and release the PRINT Unit button until the display shows END   Press andrelease the ON ZERO Off button  The balance will return to the weighing mode   Press and hold the ON ZERO Off button until the display shows OFF    Remove power from the balance   Remove either the short on U9 or the short on the RS232 option     Replace the cover     B 3    APPENDIX B FACTORY CALIBRATIONS    B 4    OHAUS      8 0 2 3 09 7 3       P N 80250973 SERVICE MANUAL SCOUT PRO ELECTRONIC BALANCES    
26. malfunction     Ifa problem arises that is not covered in this manual  contact Ohaus Corporation for further information     2 1    CHAPTER 2 DIAGNOSIS  TABLE 2 1  DIAGNOSTIC GUIDE    SYMPTOM REMEDY    No display    No power    Check AC Adapter       Power Jack defective    Replace Power Jack     2  3  Weak battery    Replace     4  Battery connector   poor  4  Clean or replace   contact       PC Board defective    Replace PC Board   Low BAT displayed    Weak battery    Replace battery       AC Adapter defective or  2  Replace AC Adapter   wrong type being used       PC Board defective    Replace PC Board     Stays in segment check    Weak battery    Replace battery       PC Board defective    Replace PC Board     Always displays zero    Possible power surge    Perform linearity calibra   tion       Down stop improperly set   2  Set per specifications       Transducer not con    Plug in Transducer   nected       Defective Transducer    Replace Transducer     PC Board defective    Replace PC Board     Cannot calibrate    Incorrect calibration   Use correct masses   masses       Transducer not con    Plug in Transducer   nected       Defective Transducer    ReplaceTransducer     PC Board defective    Replace PC Board     Displays  Err2    Missing Pan    Put pan on balance      Down Stop improperly   Check stops and reset if  set  necessary     3  Transducer zero has shifted   3  Perform factory calibra   tion B 1 or B 2       TABLE 2 1     CHAPTER 2 DIAGNOSIS    DIAGNOSTIC GUID
27. nce     14  Replace the batteries if previously removed     15  See Appendix A Service Tools Instructions and load temperature compensation data     16  Check the performance of the Balance and perform Factory Calibration B 2     3 3    CHAPTER 3 REPAIR PROCEDURES    3 1 3 Transducer  8  Replacement   There are 8 Transducer Kits that are available for replacement purposes  Each Kitis fully assembled  and consists of metal base plate  a transducer mounted to the base plate and a subplatform mounted  to the transducer  All adjustments such up stop and down stop  limits travel of the transducer to safe  limits  have been factory set  Transducer kits are available in the following ranges  120g  200g  400g   600g  1200g  20009  4000g and 6000g  Checkthe bottom of the balance and verify the capacity before    ordering a new transducer kit   CAUTION    Extreme care must be exercised so as not to twist or  deform the Transducer in any way  Do not drop or hit  the Transducer  Any damage to the Transducer can  render it inoperative     1  Remove the cover in accordance with the procedures in paragraph 3 1 1     2  Remove the 4 screws that secure the Transducer  8  to the Base  10   The screws are  accessible at the bottom of the balance  See Figure 3 4        TRANSDUCER  MOUNTING SCREWS                                              _10 BASE   8  TRANSDUCER  TRANSDUCER     CONNECTOR J6  MOUNTING      E                                                             SCREWS LIA                
28. ond to pressure on the platform    Enter the balance serial number    Click on OK    Click on READ    The balance should read the correct model and serial number    Perform the factory calibration  See B 2    The unit is finished  Remove the RS232 option or the jumper on the customers RS232  option     A 6    APPENDIX B FACTORY CALIBRATIONS    B  FACTORY LINEARITY CALIBRATIONS  The Scout Pro service and factory linearity calibrations use three calibration points  zero  mid scale  and full scale     B 1 SERVICE CALIBRATION   This calibration is used when an error 2 has been displayed by the balance  The calibration is  performed with no error checking and is saved as a user calibration  It is not a factory calibration  because no data is written to the EEPROM     1     2     9     10     Start with the unit off     Press and hold both the ON ZERO Offand the PRINT Unit buttons until the display reads  rAMP     Press and release the PRINT Unit button  The display will show Lin     Press and release the On Zero Off button  The display will show  C  blinking and then the  first mass value     Place masses totaling this value on the balance and press the On Zero Off button     Press and release the On Zero Off button  The display will show  C  blinking and then the  second mass value     Place masses totaling this value on the balance and press the On Zero Off button     Press and release the On Zero Off button  The display will show  C  blinking and then  donE followed by Lin     
29. t met   adjust the screws on the base plate using the data in  Table 4 4  See Figure 4 2 for location of adjustment screws        DOWN STOP DOWN STOP  ADJUSTING SCREW ADJUSTING SCREW       Figure 4 2  Transducer Front View   Down Stop Adjustment Screws    4 4    CHAPTER 4 TESTING    4 3 Down Stop Adjustment  Cont      TABLE 4 3  TEST MASS VALUES    SPx202  SPx402  SPx401    SPx601   SPx2001  SPx4001  SPx6000       TABLE 4 4  TRANSDUCER ADJUSTMENT TOLERANCES    SPx123  SPx202  SPx402  SPx401    SPx601   SPx2001  SPx4001  SPx6000    Min mm in   0 40 0 016  0 60 0 024  0 50 0 020  0 75 0 030  0 65 0 026  0 65 0 026  0 60 0 024  0 50 0 020    Max mm in   0 45 0 018  0 70 0 028  0 60 0 024  0 90 0 035  0 75 0 030  0 90 0 035  0 70 0 028  0 60 0 024       After the transducer has been adjusted  assemble the balance retest and recalibrate     CHAPTER 4 TESTING    4 6    CHAPTER 5 DRAWINGS AND PARTS LISTS    5 1 DRAWINGS  This section of the manual contains an exploded view and parts lists  The exploded view drawing is  designed to identify the parts which can be serviced on the balance in the field     NOTE   In all cases where a part is replaced  the balance must be  thoroughly checked after the replacement is made  The  balance MUST meet the parameters of all applicable  specifications in this manual     If further technical information is needed  in the United States call Ohaus Aftermarket toll free 1 800   526 0659 between 8 00 a m  and 4 00 p m  EST  An Ohaus factory service te
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
ISL8117EVAL2Z User Guide  Ver PDF - American UPS  IFT3912 - Remise finale  A physics lecture for the 21st century  "取扱説明書"  Samsung Galaxy Gio Black Kullanıcı Klavuzu  PDF file - Deep Blue - University of Michigan    Baumatic BOD890BL Technical Drawing  SW-CAI - Elettronica Veneta    Copyright © All rights reserved. 
   Failed to retrieve file