Home
SERVICE MANUAL for Telescopic pole, Gripo Alu
Contents
1. Vw HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu Telescopic pole Gripo Alu Side 1 av 11 HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu HEPRO aid manufacturer HEPRO is a Norwegian industrial company that develops manufactures and sells technical aids for adults and children with mobility impairment The company was established in 1987 and has to day about 50 employees The company s headquarter and factory is located in Rognan Nordland Hepro AS vermoan 9 8250 Rognan Technical Service Services are performed in the factory at Rognan but also by partners in many parts of the country Our sales personnel are also trained in technical maintenance If errors on products or other queries please contact your supplier or importer Adaptations and Special Adaptations Hepro is a manufacturer and can therefore offer user customized solutions when required Customizations and custom solutions to meet individual needs is part of the daily tasks of the plant Warranty When used in accordance with the supplied assembly guide and the user reference manual Hepro provides a two year warranty against defects in material and workmanship Please address warranty complaints or other inquiries about the product to your supplier or importer Always state the product s serial number as a reference The product is designed tested and produced in conformity to following regulations and requir
2. 35130 20 Hand wheel 1 35123 21 Washer 10 2x40x2 1 70250 22 Bend 40mm 2 35122 23 Adapter bend 40mm 2 35128 24 Tooth washer 1 35121 25 Bend 22mm 2 35120 26 Adapter bend 22mm 2 35129 27 Tighten screw M10 1 35119 4 Change log Edt 01 31 01 2012 Adapted new provisions Edt 02 10 09 2012 Updated 1 3 1 Telescopic pole Gripo Alu Side 11 av 11
3. The pole is tested according to EN 12182 and approved for a horizontal load of 100 kg on the middle of the pole with maximum length 3 0 meter It is produced in the materials of aluminum and plastic and intended for clamping between the floor and ceiling of any room where the ceiling and floor have enough strength to withstand a point load of 100 kg In each end of the pole there is a plate shaped end plate with a friction ring installed The force between the plates is maintained by a gas spring with an average pressure of 100 kg This makes the pole possible to be installed without use of any form of screw or nail connection causing perforation of floor or ceiling and the dismantling of the bar leaves no marks With support of different accessories the pole can be adapted to many different pole use situations The pole can be used in rooms with headroom from 2 1 meter to 3 0 meter and it is delivered in only one model in which the aluminum tubes are in the colour nature aluminum and all plastic parts in black colour Main Product Type designation Product information covers the following model GRIPO ALU Structure Generally the product is structured by the following main components which will also appear in how the product is broken down in section 3 Part catalog Base plate Outer tube w lifting mechanism Inner tube w tighten bushing with conus and screw Between these a Quick Release clamp Top plate W gas spring Telesco
4. The product is built quite simple and by using the Service Manual any error could easily be detected e Hepro s Service Organization is of course also available for advice and guidance Telescopic pole Gripo Alu Side 8 av 11 HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu 3 Part Catalog In the following sections the article name and article number included in the standard product will be stated Prices are stated in an own spare parts price list 3 1 Main parts replace modules and components Pos Description No Hepro Item no Gripo Alu complete 1 35100 1 Top plate 1 35110 2 Washer 6 4x20x1 25 3 10277 3 Lock nut Nylock ELZ M6 2 70054 4 Gas spring Stabilus LiftOMat 900N 1 35137 5 Inner tube 1 35111 6 Brand foil red 1 35143 7 Quick release 1 35195 8 Upper bushing 2 35194 9 Cone aluminium 1 35126 10 Tighten bushing w thread adapter 1 35118 11 Tighten screw stainless steel M6x0 75 compl 1 35144 12 Setscrew stainless steel thread stift M5 1 70198 13 Outertube 1 35115 14 Foot pedal 1 35113 15 Lifting arm 1 35114 16 Base plate 1 35112 17 Furniture button K 15 9x15 M6 2 10276 18 Lock nut Nylock M6 1 70054 Telescopic pole Gripo Alu Side 9av 11 easy life SERVICE MANUAL for Telescopic pole Gripo Alu Telescopic pole Gripo Alu Side 10 av 11 HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu Hepro Pos Description No Item no Universal conection complete 1
5. is wrong and the pole must not be used Installation and assembly Test before installation Beforehand the ceiling should be examined with regard to strength and opportunity for placement of the pole at the desired location At plate covered ceiling the pole should preferably be placed directly under the beams or nailing strip to prevent plate damage If the ceiling is made of concrete the pole can be placed anywhere If the wallboard is made of gypsum or other week material The Top plate Big has to be used or the ceiling has to be reinforced by for instance using a more solid wallboard between the pole s top plate and the ceiling Installation The pole is delivered with a loose base and top plate Place the pole on the base plate and set the top plate in place Position the pole in a suitable place with the base plate flat against the floor Pull out the foot pedal and open the Quick Release clamp Rotate the inner tube clockwise so it want slide down by it s own weight Extend the pole till the top plate reaches the ceiling by pulling the inner tube out of the outer tube Close the Quick Release clamp Press down the foot pedal Check that the red mark is no longer visible If not go back to point 4 If the pole is tilted pull out the foot pedal and adjust it vertical and press down the foot pedal again 9 Installation of accessories by using the Universal link can now be done 10 If the red mark still is visible this may be the reaso
6. the ceiling is solid and the Quick Release must be checked so that the red marking is not visible after installation Disassembly Pull out the foot pedal and open the quick release The inner tube can now be pushed down into the outer tube How to use Gripo Alu can be used in many situations where there is a need for support Use of equipment and methods will in cooperation with the user the therapist and the supplier s qualified person be best taken care of See also the illustrated brochure Error Indications and check out of these If the pole seems loose out of plumb or there are marks on the floor or the ceiling that indicates that the pole has displaced the assembly must be checked by a competent person If the red mark on the top of the inner tube appears the pole must not be used before new installation is done and checked This may indicate that the pole is either not installed correctly or the ceiling is not strong enough Inquiry by mistake If an error occurs using problems or suspected wrongly installation please contact your supplier By request always give the serial number affixed to the pole Telescopic pole Gripo Alu Side 7 av 11 HEPRO easy life PA HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu 2 Work Descriptions General It must never be changed to other parts than those stated in the split images in the 3 PART CATALOG with article numbers It ca
7. ements NS EN 12182 Technical aids for disabled persons 93 42 E F Equipment class 1 marked CE Telescopic pole Gripo Alu Side 2 av 11 HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu Table of contents 1 Short product description ssssseesessssesesssoeeesseseeseseseeseseseeseseosesesseeesees 4 TT MIn OCCU apre E E E O E EOE 4 ERE NE ESN NN 4 12 TUE hhv 4 1 2 Technical SPENN se 5 TD SEN NONNE iaia 6 151 VANG 6 1 3 2 Installation and assembly reroenonnonnoneenevnennennenneneenennennennenneneene 6 TS MSN NG 6 1 3 4 Function Verification and testing of product and accessories 7 1 3 5 DISSENS 7 1230 HOW OP 7 1 3 7 Error Indications and check out of these 7 2 Work DESCNPI0NS Lvov eee 8 2 1 Cleaning performed by user rrnronevnennennennenevnennennennennnnennennennennene 8 22 MANNEN ria 8 2 2 1 PGS 8 2 22 PICDaAralontorreuse zasori ioni 8 225 Troubleshooting civica rina 8 gt PALIN hhv 9 3 1 Main parts replace modules and components ccecesceccecceceees 9 Aria dI o 11 NENNE 11 Telescopic pole Gripo Alu Side 3av 11 HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu 1 1 1 1 1 1 1 2 Short product description Gripo Alu is classified in NS EN ISO 9999 as a medical device and within the product group 180806 and is defined as a TELESKOPIC POLE to carry support devices
8. n e tis too big distance between the ceiling and the top plate when the rotation of the inner tube started e The quick release could have been too loose Tighten the nut a few millimeters oe ooo Telescopic pole Gripo Alu Side 6 av 11 SERVICE MANUAL for Telescopic pole Gripo Alu e The ceiling is too weak and is lifted by the force in the pole s gas spring e The pole was ice cold during assembly See also the Assembly guide illustrated with pictures 1 3 4 Function Verification and testing of product and accessories 1 3 5 1 3 6 1 3 7 1 3 8 e Afterthe installation is completed it must be checked that the red mark in the upper part of the inner tube just below the top plate is not visible This should also be checked regularly by the use of the pole If the red marking is apparent the cause may be o Too week ceiling structure that causes the power of the pole pushing the ceiling up Then use the Top plate Big or the ceiling must be reinforced or the pole must be moved to an area where construction is strong enough o The tightening screw on the quick release may have loosened Tighten the nut on the quick release till it sits tight e When recycling the pole all surfaces must be inspected for damage Take special care of the list in the top and base plate which make the friction ring If this shows sign of darkning it should be replaced e After this control the pole must be installed in a room where
9. n in worst case cause the product not to satisfy its acceptance criteria and be changed from being an aid to become a dangerous tool for the user or the environment If for some reason it is desirable to deviate from this rule Hepro has to be contacted for review and possible approval 2 1 Cleaning performed by user METHODE e Normal dusting and cleaning with wet cloth and soap is sufficient 2 2 Maintenance 2 2 1 Periodic inspection GENERAL It is basically sufficient with inspection when recycling For poles older than 5 years the producer recommend inspection as for recycling once a year PRODUCTS WITH SPECIAL LOADS The product has no particular need for periodic inspection if it is a standard product used by a person who in any way exposes the product for special or extreme loads If the product is exposed to special loads it would be an advantage for all that each case will be considered and necessary maintenance Is established Hepro can of course be of help with advice and guidance if contacted 2 2 2 Preparation for reuse To ensure that the user gets a secure aid when recycling it is important that the aid undergo a thorough technical control and is properly cleaned before it comes to a new user As a manufacturer and supplier we believe that the control described in section 2 3 2 is sufficient Worn parts must be checked and replaced if they are considered to lead to reduced safety for the user 2 2 3 Troubleshooting
10. pic pole Gripo Alu Side 4 av 11 HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu 1 2 Technical specifications Gripo Alu DIMENSIONS Unit Average gas spring pressure Nm Top plate Base plate Max horizontal loa don the middle of pole Product weight without accessories Pkg AT Max user weight ke 150 MATERIALS AND COMPONENTS Inner and outer tube Aluminium 6063 T6 Surface protection tube Elox Gas spring Stabilus LiftOMat Top plate Plastic PA6 GF15 Base plate Plastic PA6 GF15 Tubes for accessories Aluminium 6063 T6 Elox Telescopic pole Gripo Alu Side 5 av 11 HEPRO SERVICE MANUAL for easy life Telescopic pole Gripo Alu 1 3 1 3 1 1 3 2 1 3 3 Safety provisions The installation of the pole has to be done by a trained person or by one who has set himself thoroughly in the assembly instruction Warning e The product should not be adopted without prior training and before the user reference manual has been carefully read and understood e All kinds of support devices may involve danger if incorrect assembled or wrongly used e Ifthe pole indicates any sign for loosening it shall not be used and help be called e The pole must not be installed in rooms with headroom exceeding the approved or be placed where there can be any risk for pinching e When used in showers friction mat has to be used under the base plate e Ifthe red warning sign is visible the assembly
Download Pdf Manuals
Related Search
Related Contents
Typhoon Office Laser Mouse Infinity CPG-INFVER01 User's Manual Repair Express Cement Coleman 2000006664 flashlight StarTech.com 10 ft Ultra Thin USB VGA 2-in-1 KVM Cable Samsung HBGR-401N User Manual raspador piezo ultrasónico uds k led manual de instrucciones PO6400: Base Interface Rede PROFIBUS Copyright © All rights reserved.