1. 2 ENTER 3 4 2011 Kikkoman Corp 20150301 kikkomar ATP Hygiene Monitoring test kit Swab Test LuciPac I Instruction Manual CAUTION Lu
2. 2011 Kikkoman Corp 20150301
3. 20 lt 30 35C 2 8C
4. D 2 2 1 2 8
5. 1 4 2 1 2 1 1 A 2 C 1 2 B 1 1
6. pa RLU 1 I 1 aes RLU Hl ADE AIA 2 100 200 500 1 000 1 500 3 000
7. 20160709 I T 105 0003 2 1 1 Tel 03 5521 5490 Fax 03 5521 5498 E mail biochemifa mail kikkoman coJp http biochemifa kikkoman cojp 2011 Kikkoman Corp 20150301 I 20 35 20 F 7 0 ET
8. ATP ATP Os Mg AMP COs ATP ATP I 20 5 100 AY I
9. C 110 C 100N C 100 AF 70 1 20 35 EK I 2035 20 D O
10. Useitto swab the test subject with constant pressure as muchas possible Under suitable pressure the swab shaft will slightly bend as illustrated in the panel 3 of page 6 Return the swab stick to the the main body casing and push it all the way into the the main body casing Hold firmly onto the LuciPac II casing and shake it Allow the leftover luminescent reagent to thoroughly dissolve Put your thumb on one of the two welded zones and pull the tip of the test tube to break the welded bound Repeat this procedure for another welded zone of the opposite side Remove the test tube from the main body Be careful not to spill the releasing reagent Insert the test tube into the measuring chamber of Lumitester Press the ENTER key Results are obtained in 10 seconds 900 1 Regular tap water may contain slight amount of ATP We recommend that you wash or otherwise clean off the faucet in such cases and allow the water to run for a while before collecting any for use Refer to Direction for LuciPac II on page 8 for further details 2 Handling of data Benchmark levels 1 2 are set that are used to determine whether cleanliness levels are at acceptable levels or not see Fig 1 Cleanliness levels are designated as Pass A if the measured amount of luminescence falls below level 1 and as Fail C if the amount is higher than level 2 With Fail C cleaning will need to be redone and cleaning proced
11. 25C 14 30 5 2 20 2 8 CH 2 3 2011 Kikkoman Corp 20150301 20150410S
12. Symbols Used in the Packaging and Labeling of this product Symbol for temperature limitation The upper and lower temperature limits will be indicated on either side of the symbol Please store this product at the indicated temperature range Symbol for Caution or Attention for use Symbol for Lot Number This symbol shall be adjacent to the manufacturer s lot number e g 20150410S or description of its printed location Symbol for Use By This symbol shall be adjacent to the expiration date expressed as YY YYMMDD e g 20160709 or description of its printed loca tion Symbol for Consult Instructions Manual Symbol for Manufacturer This symbol shall be adjacent to the name and address of the manufac turer LuciPac and Lumitester are registered trademarks of Kikkoman Corporation in Japan and other countries Manufacturer Kikkoman Biochemifa Company 2 1 1 Nishi shinbashi Minato ku Tokyo 105 0003 Japan Phone 8 1 3 5521 5490 Fax 81 3 5521 5498 E mail biochemifa mail kikkoman co jp URL http biochemifa kikkoman co jp e 2011 Kikkoman Corp 20150301 LuciPac II Direction for LuciPac IJ lt Leave LuciPac II at room temperature 20 35 C for 20 minutes before use gt A Swab stick Pull the cotton swab stick When the test object is dry moisten Swabthetest object A outofthemainbody the swab A with tap water 3 or with the swab stick A B wet the object with
13. ciPac II Product code 61301 1 Do not attempt to drink the kit reagent or touch it with bare hands or allow it to splash into eyes 2 Please make sure to read the precautions and instructions in this Instruction Manual before attempting to use the kit and exercise extreme caution when using it The Kikkoman ATP hygiene monitoring test kit LuciPac II is a kit for testing cleanliness levels using bioluminescence techniques using firefly luciferase developed with Kikkoman s unique biotechnology Applications This kit can be used to test cleanliness levels using a swab test Do not use this device for any purpose other than what it was designed for as a tester of cleanliness levels Please be aware that this kit cannot be used to test or measure for counts of viable bacteria or more specific types of pathogenic bacteria This kit is used to indicate cleanliness levels in food production centers and medical treatment facilities This kit shall not provide any guarantee that a given test environment is free of bacterial contamination This kit cannot be used to test food or food products directly Measurement Principles This kit uses a luminescent reactions from firefly luciferase ATP Luciferin O2 Luciferase Mg AMP Pyrophosphoric acid Oxyluciferin CO light Firefly luciferase emits light in the presence of ATP and luciferin ATP is a source of energy necessary for various forms of life that is present in or
14. ganic residues such as microorganisms food residue and biological substances that originate from other living organisms This method allows you to measure and detect organic residues at high speed and high sensitivity by detecting ATP using luciferase which is why itis widely used in determining cleanliness levels in food production centers and medical treatment facilities This kit is a simple integrated testing instrument that contains both the test reagent and the swab device required for testing cleanliness levels Contents The LuciPac I kit comes with five aluminum bags each containing 20 swab devices for a total of 100 swab devices Swab stick Red NN Main body Cotton swab Welded zone Test tube Container for releasing reagent Reagent name Main component Luminescent reagent Luciferin Luciferase Magnesium acetate Releasing reagent Surfactant Benzalkonium chloride Precautions for Use Please make sure to follow the items outlined below in order to obtain optimal performance from this kit Do not use products that are past the expiry date Expired products may not yield accurate results the expiry date is printed on the bottom of the aluminum bag holding the swab devices Be sure to use only designated products when taking measurements for amounts of luminescence This kit cannot be used with non designated products 8 The swab devices should be allowed about 20min to reach ro
15. kikkoma I I 60375 2 1 II ATP A
16. om temperature 20 C to 35 C 68 F to 95 F prior to testing if they are from refrigerated stock Measurement values may read lower than actual if the swab devices are used while still cold Use the swab devices assoonas possible once they havereturnedtoroom temperature Do not let the kit sit out in temperatures exceeding 35 C 95 F High temperatures may cause product performance to drop 4 Wherever possible be sure to use up all swab devices from a single bag that has been opened at one time If you absolutely must store leftover swab devices once finished with a test session firmly close the aluminum bag and store it in a refrigerated environment 2 C to 8 C 35 6 F to 46 4 F High temperatures may cause product performance to drop Do not subject the kit or any part of it to direct sunlight for long periods of time Strong light may cause product performance to drop Do not touch any of the parts inside the kit particularly not any part of the cotton swab itself with a finger or other object before use Touching the parts may affect cleanliness levels making them hard to determine M Do not drop the kit or any of its parts or allow any parts to be struck or jolted The inner aluminum sheets and other parts in the kit may become damaged causing product performance to drop Do not use the kit if any parts become damaged such as the inner aluminum sheet Such damage may affect product performance causing cleanline
17. parts used in this kit are listed below No PVC materials were used in the making of the plastics in this kit Swab stick orange part Polypropylene Main body casing Polypropylene Swab shaft Polypropylene Polypropylene Container for releasing reagent Aluminum bag with Aluminum Polyethylene dehumidifying function Polyethylene terephthalate Precautions for Handling Please observe the following items to ensure safe use of this product Be careful not to let the reagents or other substances in the kit get into your mouth or eyes or onto bare hands before or after use Rinse your mouth out thoroughly with water if any of the substances get into your mouth rinse off your skin with copious amounts of water if any get onto your skin and rinse out affected eyes thoroughly with copious amounts of water should it get into your eyes and immediately contact a physician for advice and follow the instructions given 2 Exercise enough caution when storing and disposing of the kit and its reagents to ensure that none of the substances become mixed in with food and other products Be careful not to get fingers caught when inserting the swab stick into the main body casing Please make sure to store this kit and its parts out of the reach of young children Note that the releasing reagent used in this kit contains cationic surfactants benzalkonium chloride 2 Take precautions when disposing of this kit after use
18. ss levels to fail to be measured correctly You can tell if the aluminum sheet is damaged or not by checking to see if the releasing reagent is leaking Set the benchmark level at which the required cleanliness levels can be obtained If the benchmark level is not set correctly cleanliness levels may not be evaluated correctly Secure the test tube of the LuciPac II ensure that it does not become shifted or displaced before measurement It may be difficult to accurately determine cleanliness levels if the test reagent begins leaking because the tube has been shifted 2011 Kikkoman Corp 20150301 Designated Product for Measurement Lumitester C 110 and Lumitester C 100 Manufacturer Kikkoman Biochemifa Company Be sure to use only designated products when taking measurements Measurement Methods 1 Measurement procedures Complete the procedures listed below within a temperature range of 20 C to 35 C 68 F to 95 F Make sure to always run measurement tests at the same temperature to maintain repeatability from comparison Remove the LuciPac II from refrigerator and wait until they have reached room temperature about 20 min 20 C to 35 C 68 F to 95 F Use the swab devices as soon as possible once they have returned to room temperature Remove the swab stick from the main body casing If running the measurement test on a dry subject first moisten the cotton swab or the test subject with tap water
19. to ensure that such substances do not become mixed in with food products at food production centers and similar facilities 2 Benzalkonium chloride is a disinfectant and antiseptic commonly used in hand and finger sterilizer solutions Storage 1 Kit storage Kits are to be stored at a low temperature 2 C to 8 C 35 6 F to 46 4 F for long term storage The kit can be stored below 25 C 77 F for up to 14 days or below 30 C 86 F for up to Sdays before opening an aluminum bag without any adverse effect on the long term stability Do not freeze the kit 2 We recommend that you use all 20 swab devices in a single bag at one time after opening an aluminum bag Ifyou have leftover swab devices that you must store after opening a bag be sure to store them at the recommended low temperature 2 C to 8 C 35 6 to 46 4 F and use them within two weeks from when the bag is opened 3 Expiry date Printed on the aluminum bag 2011 Kikkoman Corp 20150301 Warranty Kikkoman Biochemifa Company warrants the products in this kit to have a certain level of quality This warranty guarantees that Kikkoman Biochemifa Company shall replace defective products should any be found This warranty does not provide any other guarantees Kikkoman Biochemifa Company shall not be liable for any damages including special or consequential damages or expenses arising directly or indirectly from the use of this product
20. ures may need to be revised Cleanliness levels where the measured amount of luminescence is above level 1 but below level 2 are designated as Caution B and it is recommended that cleaning be redone and cleaning procedures revised in such cases Examples of benchmark setting values for food production centers are listed in Table 1 The benchmark setting values for cleanliness level control shown in Table 1 are provided merely for reference and may not necessarily apply to all situations and cases The benchmark levels are important guidelines for determining cleanliness levels and should be set as appropriate according to required cleanliness levels Level 1 Level 2 fall Low Relative Light Unit RLU High Clean Cleanliness level Contaminated Fig 1 Setting Benchmark Level for Cleanliness Level Control Table 1 Example of Setting Benchmark Level for Cleanliness Level Control in Food Production Centers Relative Light Unit RLU Substance under Test Level 1 Level 2 Stainless steel surface 100 200 Resin surface 500 1 000 Hands and fingers 1 500 3 000 Pa LuciPac II Disposal Methods This kit contains no hazardous materials This kit can be disposed of as regular garbage but when disposing of it 1t would be better to separate the parts and dispose of each one properly in accordance with the local regulations outlined by prefecture and city town or village for proper disposal of waste materials The main materials and
21. water 3 Put the swab stick A back to the main body Shake the whole body afew times so Gently shake the whole body B and pushitthroughallthe way by that any remaining liquid in the capsule so that the luminescent reagent putting the tip of the test tube C ona palm falls into the test tube C Otherwise is entirely dissolved Otherwise of hand or table to avoid having the tube the the remaining liquid may spill out when the resulting value may become tip part of the body come off the main body you remove the test tube in step 8 lower than the accurate value Put your thumb on one of the two Remove the test tube from the Insert the test tube into the measuring welded zones and pull the tip of the test main body Be careful not to spill chamberofLumitester Press the tube to break the welded bound Repeat the releasing reagent ENTER key Results are obtained this procedure for another welded zone in 10 seconds of the opposite side 3 If the luminescence level of tap water shows high we recommend you to let the water flow a few minutes and or clean up the outlet of the _Q 2011 Kikkoman Corp 20150301