1. 2 General information This hand held measuring instrument provides you with a thermo hygrometer a laser pyrometer or a combination of the two depending on the application mode The measuring instrument can be operated in three different modes In the TH mode the measuring instrument offers you all the functions of an electronic thermohygrometer In the IR mode you can use the measuring instrument as a laser pyrometer for surface temperature measurement with measurement location marking Operating manual English In the DP mode the measuring instrument unites the functions of a thermohygrometer with those of a pyrometer and at the same time indicates the dewpoint temperature and surface temperature As soon as the surface temperature is below the dewpoint tempera ture the measuring instrument alerts the user with an optical laser signal and an alarm tone Surfaces can be examined in no time and weakpoints detected quickly with this alarm function The alarm thresholds are individually configurable B 04 3 The display Upper menu Upper display Lower display Lower menu Operating modes Configuration types HOLD MAX MIN AVG A 4 Operation Unlike conventional hand held measuring instruments 1 this measuring instrument has no keypad but a so called thumb wheel on the left hand side The wheel allows a 15 rotation down and up and k can be pressed additionally
2. Insbesondere beim Standortwechsel von kalten zu warmen Umge bungsbedingungen zum Beispiel bei Verbringung in einen beheizten Raum nach der Lagerung ber Nacht im Auto kommt es je nach Raumluftfeuchte zu Kondensatbildung auf der Leiterplatine Dieser physikalische Effekt der sich konstruktionsseitig bei keinem Messger t verhindern l sst f hrt zu falschen Messwerten Deshalb zeigt das Display in dieser Situation keine Messwerte an Bitte warten Sie in solchen F llen ca eine Minute bis das Messger t aklimatisiert ist und setzen den Messvorgang dann fort 9 Hinweise zum Emissionsgrad Der Emissionsgrad ist ein Wert der benutzt wird um die Energieab strahlungs Charakteristik eines Materials zu beschreiben bungs ce bali Lager Betrieb Je h her dieser Wert desto h her ist die F higkeit des Materials lassi Strahlungen auszusenden Viele organische Materialien und Oberfla Temper 30 C 60 C 0 C 50 C chen haben einen Emissionsgrad von ca 0 95 zul ssige 95 rH Metallische Oberflachen oder gl nzende Materialien haben einen nied REM 95 o rH bzw 20 g m rigeren Emissionsgrad und liefern daher ungenaue Messwerte der kleinere Wert gilt zul ssige H he Bitte beachten Sie dies bei der Anwendung des Messger tes un 5 000 m 5 000 m Nes m Uber Zur Kompensation kann die Oberfl che gl nzender Teile mit Klebeband Bedienungsanleitung Deutsch oder mit matts
3. Alle Sensoren sind bereits werksseitig kalibriert und verf gen ber LI UI wm gt eine entsprechende Werkskalibrierkennlinie B BA Die Werkseinstellungen erh lt man durch Bei der Einpunktkalibrierung wird durch Angabe eines Abgleichwertes Offset eine globale Kalibrierkurvenverschiebung durchgef hrt die im gesamten Messbereich wirksam ist Der einzugebende Offset ist der Wert um den die Kalibrierkurve verschoben wird m m or ce Setzen des Offsets auf 0 0 Bedienungsanleitung Deutsch A 14 e eee CAL 2 Mit CAL 2 Einpunktkalibrierung stellt Gi i NO Offset Eingabe NEN man den Offset f r Sensor 2 relative Feuchte Kennlinie wid aa Lu f H ein Der Offset dreht die Kennlinie um den un verschoben E stax teren Abgleichpunkt 11 r H Der Abgleich oe _ punkt muss im Bereich von 30 bis 95 rH 2 H H liegen Der Offset wird in der oberen Displayan Gemessener Temperaturwert Li wn zeige dargestellt Maximal lassen sich 10 rH e einstellen CAL 2 ist nur anw hlbar in Ver b m m op oe bindung mit der Einheit r H und bei einer id i F Offset Eingabe vorherrschenden Luftfeuchte von mindestens 30 r H gedreht gt a E W hlen kann man mit 8 und d best tigt wird mit gt 2 m S Die Werkseinstellungen erh lt man Gemessener Feuchtewert durch Setzen des Offsets auf 0 0 A 15 Bedienungsanleitung Deutsch 7 Kalibrierung Die Kalibrierung der Klimasensoren is
4. B 18 2 General information B 04 9 Notes on emission facht B 19 3 Thedisplay sees B 05 10 Technical data B 21 4 Operation B 05 11 2 EE B 22 5 The upper menu sese eee eee B 07 This publication replaces all previous announcements No part of this publication may be reproduced processed using electronic systems replicated or distributed in any form without 6 Thelower menu sees B 08 our written authorisation Subject to technical changes All rights reserved Names of goods are used without guarantee of free usage keeping to the manufacturer s syntax The names 6 1 Operating moies nen B 09 of goods used are registered and should be considered as such We reserve the right to 6 2 Configuration n sse ervub Obes tele B 13 modify design in the interest of on going product improvement such as shape and colour A modifications The scope of delivery may vary from that in the product description All due 7 Calibration sssseseeeeeeeeeene B 16 care has been taken in compiling this document We accept no liability for errors or omissions B 01 Operating manual English 1 Read before starting This measuring instrument has been built according to the state of the art The instrument is compliant with the standards EN50082 2 EN61000 4 2 EN 61000 4 3 EN50081 2 EN 55011 and meets the requirements of the valid European and national regulations The con for
5. function flashes again To select the flashing function confirm with gt A confirmed function is shown statically in the display If you do not want to select a function but exit the upper menu enter dk If no inputs are made the menu is exited automatically after 20 seconds To deactivate a selected function enter gt Operating manual English 6 The lower menu The three operating modes TH thermohygrometer IR pyrometer and DP dewpoint alarm generator and the configuration areas CFG Unit 1 Unit 2 CAL 1 CAL 2 can be selected in the lower menu You go to the lower menu with d the first selectable menu item flashes You go to the next selectable menu item by entering Y again The menu items are only selectable one after another in one direction If you have skipped a menu item you wanted to select press d repeatedly until the menu item flashes again To select the desired menu item which is now flashing confirm with gt If you do not want to select a menu item but exit the lower menu enter If no inputs are made the menu is exited automatically after 20 seconds Operating manual English B 08 6 1 Operating modes TH The instrument can be used as a thermohygrometer when you select the TH mode R DP CFG In this mode the temperature is shown in the upper display and the humidity in the lower display The unit of temperature C F is set in configuration area Unit 1 in
6. 90 to 0 95 Eemer ars hottie ha eit ee dc ftd Ped Arn cd 0 90 to 0 95 Chalks i es are chou Rea Siete eet DNA ede tod dan 0 95 IB ir nas Sethe E pest EIE NES Bender 0 95 Operating manual English Enamel varnish back 0 95 Glass Seres e Ee 0 85 to 0 90 Marble nne aan de 0 90 to 0 95 Paint non metallic 0 0 eee nn 0 95 Plaster use ed i ton t eee ds 0 90 to 0 95 Plastics Set o oreet tul esee Eee dab me sorde dee e 0 90 Radiator paints ss ceti sepes REESE UD RR 0 95 Roofing felt eec etre ne ees tend 0 95 let ef nee 0 90 to 0 95 Textiles non metallic 0 95 Wallpaper non metallic 0 95 Wale SEN toe ee uate EC ale s 0 93 KU EE nett du ge On 0 90 to 0 95 10 Technical data Airtemperat re 2 sso ada rer pte anes CPF Measuring principle NTC Measuring range 20 to 50 C Resolution ar ta rien 0 1 C Accuracy 0 4 C at 0 to 40 C otherwise 0 7 C Air humidity are exa wees REA eR EX re Y nd r H g m Measuring principle capacitive Measuring range 5 to 95 rh EE KE EE 0 1 rh Le 3 rh Surface temperature 0 cece eee eee eee C F Measuring principle Thermopile Measuring range eee eee 20 to 60 C Measuring Jeng 8 1 Measuring spot size Min 20 mm Resolution 5 vind east 0 1 C EUREN sid asian ea I I AA PSOE dee d ond 42 C Emission factor 2 22 222er 0 95 permanently set Useful life approx
7. D Infrarot Thermohygrometer Bedienungsanleitung Deutsch A 1 Infrared thermohygrometer Operating manual English B 1 CE Version 1 0 Inhaltsverzeichnis 1 Vor der Inbetriebnahme lesen A 02 8 Hinweise zu Wartung und Betrieb A 18 2 Allgemeines een A 04 9 Hinweise zum Emissionsgrad A 19 3 Das Display A 05 10 Technische Daten A 21 4 Bedienung Eu A 05 11 belge DEE SE E EE A 22 5 Das obere Men A 07 Diese Ver ffentlichung ersetzt alle vorhergehenden Kein Teil dieser Ver ffentlichung darf in irgendeiner Form ohne unsere schriftliche Genehmigung reproduziert oder unter Verwendung 6 Das untere Ven ee 61 Betrlebsarten s A 09 Die venendeten Werenramen ed eingetragene und solien als coche beat werden A 13 Konstruktionsver nderungen im Interesse einer laufenden Produktverbesserung sowie Form 7 Farbveranderungen bleiben vorbehalten Lieferumfang kann von den Produktabbildungen ab Zum weichen Das vorliegende Dokument wurde mit der gebotenen Sorgfalt erarbeitet Wir ber 7 Kalibrierung esee nnn nn A 16 nehmen keinerlei Haftung f r Fehler oder Auslassungen A 01 Bedienungsanleitung Deutsch 6 2 Konfigurationsbereiche 1 Vor der Inbetriebnahme lesen Das vorliegende Messgerat wurde nach dem heutigen Stand der Technik gebaut Das Gerat entspricht den Standards EN5
8. DO NOT STARE INTO BEAM DIODE LASER lt 1mW Output at 675 nm CLASS Il LASER PRODUCT Bedienungsanleitung Deutsch Laserstrahlung kann irreparable Sch den am Auge hervorrufen Bei Messungen in der N he von Menschen muss der Laserstrahl deaktiviert werden 2 Allgemeines Mit diesem Handmessgerat verftigen Sie je nach Anwendungsmodus ber ein Thermohygrometer ein Laserpyrometer oder eine Kom bination aus beidem Das Messger t kann in drei unterschiedlichen Arbeitsmodi betrieben werden Im TH Modus bietet das Messger t Ihnen alle Funktionen eines elek tronischen Thermohygrometers Im IR Modus k nnen Sie das Messger t als Laserpyrometer zur Ober fl chentemperaturmessung mit Messort Markierung einsetzen Bedienungsanleitung Deutsch Im DP Modus vereint das Messger t die Funktionen eines Thermo hygrometers mit denen eines Pyrometers und zeigt gleichzeitig Tau punkttemperatur und Oberfl chentemperatur an Sobald die Oberfl chentemperatur unterhalb der Taupunkttempera tur liegt alarmiert das Messger t den Anwender durch ein optisches Lasersignal und einen Alarmton Durch diese Alarmfunktion lassen sich Oberfl chen in k rzes ter Zeit untersuchen und Schwachstellen schnell detektieren Die Alarmschwellen sind dabei individuell konfigurierbar 3 Das Display Q HOLD wax MN AVG 4 Bedienung A A A A Oberes Men Eee Im Gegensatz zu konventionellen Handmessger ten LI IL 41 besitzt d
9. already calibrated at the factory and have an appropriate CO CH RER factory calibration characteristic Li I E I confirmation with gt S In the one point calibration a global calibration curve shift is made by aL The factory settings are obtained by specifying an offset which is effective over the whole measuring range MN setting the offset to 0 0 The offset to be entered is the value by which the calibration curve is shifted Operating manual English B 14 CAL 1 characteristic is shifted CAL 2 characteristic is turned B 15 displayed temperature value displayed humidity value offset input gt 11 rH gt measured humidity value CAL 2 The offset for sensor 2 relative humi dity is set with CAL 2 one point calibration CH CH The offset turns the characteristic around the I i if lower calibration point 11 r H The calibra KA tion point must be in the range from 30 to CO 25 rH The offset is displayed in the upper CO Li I Li 1 display A maximum 10 r H can be set Ken A CAL 2can only be selected in connection with the unit r H and at a prevailing C humidity of at least 30 r H Selection can be made with 8 and d confirmation with gt The factory settings are obtained by setting the offset to 0 0 Operating manual English tL ati One point calibration r H with calibration block and 7 Calibration calibration ampoule Check
10. erm glicht die gleichzeitige Anzeige von Oberfl chen temperatur z B Wand sowie Taupunkttemperatur und dient zum Aufsp ren von kritischen Oberfl chen an denen Kondenswasserbil dung durch Taupunktunterschreitung stattfinden kann Die Taupunkttemperatur ist jene Temperatur bei der die Luft mit Wasserdampf ges ttigt ist Bei Erreichen bzw Unterschreiten dieser Temperatur zum Beispiel an kalten Wandoberfl chen kommt es zu Kondensatbildung A 11 In der oberen Displayanzeige wird in diesem Modus die Oberfl chen temperatur und in der unteren Displayanzeige die Taupunkttempera tur TdP des jeweiligen Umgebungsklima angezeigt Im DP Modus ist keine Funktion aus dem oberen Men anw hlbar Im Konfigurationsbereich Unit 1 wird die anzuzeigende Einheit C F f r beide Temperaturen identisch eingestellt Werksseitig ist C voreingestellt Sowohl Oberfl chentemperatur als auch die Taupunkttemperatur TdP werden in der ausgew hlten Einheit dar gestellt Sobald der DP Modus ausgew hlt wurde schaltet sich zur Messort Markierung das Laserlicht ein Der Laser leuchtet f r maximal 2 Minu ten und kann in dieser Betriebsart jederzeit ber ein gt f r jeweils weitere 2 Minuten aktiviert werden Bedienungsanleitung Deutsch Verwendung der Alarmfunktion Im DP Modus ist die Alarmfunktion automatisch aktiv Die Alarmaus l sung und Alarmintensit t wird bestimmt durch den oberen Alarm grenzwert und den unteren Ala
11. in the mid position All the settings for operation and instrument configuration can be made with these operating positions Operating manual English The three operating positions of the thumb wheel Middle position Rotation up Rotation down Symbol in further text Symbol in further text Symbol in further text gt T v Instructions for selecting the upper and lower menu and selecting the operating mode and instrument configuration can be found in the appropriate chapters Switching the instrument on and off Press the middle position gt of the thumb wheel to switch on To switch off press the mid position gt Automatic switch off function after 3 minutes Operating manual English of the thumb wheel for about 2 seconds B 06 5 The upper menu Depending on the selected mode the standard functions HOLD MAX MIN AVG can be selected in the upper menu Hold Hold freezes the measured value MAX MAX represents the maximum value in the active time period MIN MIN represents the minimum value in the active time period AVG AVG represents the arithmetic average in the active time period You go to the upper menu with f the first selectable function flashes You go to the next selectable function by entering f again The func tions are only selectable one after another in one direction If you have skipped a function you wanted to select press f repeatedly until the
12. surface The smaller the target the shorter the distance from the instrument must be The exact measuring spot size can be taken from the diagram For accurate measurements the target must be at least twice as big as the measuring spot E M 8 1 Operating manual English DP DP DewPoint The instru ment can be used in the dewpoint alarm generator mode by selecting TH IR CFG the DP mode The DP mode enables simultaneous display of surface temperature e g wall and dewpoint temperature and serves to trace critical surfaces on which condensation could form by dropping below the dewpoint The dewpoint temperature is the temperature at which air is saturated with water vapour On reaching or dropping below this temperature for example on cold wall surfaces condensation forms B 11 The surface temperature appears in the upper display and the dew point temperature TdP of the respective ambient climate in the lower display in this mode No function from the upper menu can be selected in the DP mode In configuration area Unit 1 the unit C F to be displayed is set identically for both temperatures C is set at the factory Both the sur face temperature and the dewpoint temperature TdP are displayed in the selected unit The laser light is switched on for measurement location marking as soon as the DP mode is selected The laser lights for a maximum of 2 minutes and can be activated for another 2 minute
13. the calibration block for foreign bodies or residue from previous cali d brations and clean the block if necess Calibration of the climate sensors is only necessary in rare cases If high demands on the accuracy are made we recommend an annual one point calibration of the sensor You can perform one point calibration yourself but we advise against pers I ary doing so because professional reference values are not usually N mg Ditferent calibration fluids are available available SE nr l for the moisture values 35 50 and Instead make use of the possibility of certified calibration according to 80 Only the 50 calibration fluid DKD and ISO Further information about this can be obtained from x should be used for standard calibrations your dealer Operating manual English B 16 Please note the data and test values specified in the leaflet accompanying the calibration ampoules Break off the head of the ampoule Take the calibration block in your hand so that you can push in the calibration ampoule from the bottom Then place the calibration block on a level surface and make sure that the calibration fluid runs into the calibration block Now insert the sensor head of the measuring instrument into the calibration block carefully as far as it will go Wait two hours acclimatisation time before performing the cali bration according to the one point calibration procedure described under CAL Remove the sens
14. 0082 2 EN61000 4 2 EN 61000 4 3 EN50081 2 EN 55011 und erf llt die Anforderungen der geltenden europ ischen und nationalen Richt linien Die Konformit t wurde nachgewiesen die entsprechenden Erkl rungen und Unterlagen sind beim Hersteller hinterlegt Um diesen Zustand zu erhalten und einen gefahrlosen Betrieb sicher zustellen m ssen Sie als Anwender diese Bedienungsanleitung beachten Bedienungsanleitung Deutsch Vor der Verwendung des Ger tes ist diese Bedienungsanleitung aufmerksam zu lesen und in allen Punkten zu befolgen Niemals an spannungsf hrenden Teilen messen Messbereiche der Messwertaufnehmer beachten berhitzen kann zur Zerst rung f hren Temperatur und Feuchteabgleich nur mit geeigneter Referenz durchf hren Bei Standortwechsel mit unterschiedlichem Klima ben tigt das Ger t eine Angleichphase von mehreren Minuten A BestimmungsgemaBe Verwendung Das Messger t darf nur innerhalb der spezifizierten technischen Daten betrieben werden Das Messger t darf nur unter den Bedingungen und f r die Zwecke eingesetzt werden f r die es konstruiert wurde Die Betriebssicherheit ist bei Modifizierung oder Umbauten nicht mehr gew hrleistet Richten Sie den Laserstrah A Laser Warnhinweis Dieses Ger t ist mit einem Laser der Klasse 2 ausgestattet nie direkt oder indirekt durch reflektie rende Oberfl chen auf das Auge CAUTION LASER RADIATION
15. 150 h in the IR DP mode approx 10 h Dimensions 175 x 48 x 39 mm Net weight without batteries approx 100 g Operating manual English 11 Accessories optionally available Calibration block moisture ZB 911 9004 Calibration ampoules for calibration block ZB 911 9005 available for 35 50 and 80 moisture Stainless steel sinter Hier ZB 911 9003 Note on stainless steel sinter filter Since there is frequently a heavy accumulation of dust and dirt under practical conditions which can lead to falsification of the measuring results and shortening of the sensor life the measuring instrument is equipped with a metal grid filter as a standard Use the stainless steel sinter filter available as an optional exchangeable safety cap under conditions of heavy soiling Operating manual English LINI L0 HLI V8
16. In the CFG mode the values for the up LI L per threshold Hi and the lower threshold Lo I H ab are entered This menu item can only be selected if the DP mode is active The value range for Hi and Lo is from 0 0 to 9 9 The units C or F v are taken from the temperature setting T D CFG currently selected for the DP mode lo eg IE o L1 The minimum difference between Hi and Lo must be 1 0 The values for Hi upper display and Lo lower display are entered and confirmed one after another Unit 1 You select the unit of temperature with Unit 1 C and F are available for selection Selection can be made with and 4b confirmation with gt Unit 2 This menu item can only be selected if the TH mode is active With Unit 2 you select the unit of absolute humidity g m relative humidity r H or the dewpoint temperature dp C dp F r H is set at the factory Selection can be made with 4 and d confirmation with gt B 13 Operating manual English One point calibration of temperature CAL 1 The offset for sensor 1 temperature and relative humidity is set with CAL 1 one point calibration ICH CT The offset is shown in the lower display cu With CAL a one point calibration can be carried out for the sensor A maximum 10 C or 10 F can be set displays of temperature CAL 1 and relative humidity CAL 2 C Selection can be made with and d All sensors are
17. Messbereich 5 bis 95 rH AUTIOSUNG DEE Ee 0 1 rH Genauigkeit use 3 rH Oberfl chentemperatur C F Messprinzip se ee RR eraser os Thermopile Messbereich 2 cece cece eee eee nee 20 bis 60 C Messoptik cess ntact ck bial teen wate cel UE 8 1 Messfleckgr e mim 20 mm AUNOSUNG eg oy ds ae eee ge 0 1 C Genauigkeit se nenn 2 C Emissionsgrad cee eee el 0 95 fest eingestellt Standzeit 4 ca 150 h im IR DP Modus ca 10 h Abmessungen 0 00 cece eee 175 x 48 x 39 mm Leergewicht ohne Batterie 2 222 ca 100 g Bedienungsanleitung Deutsch 11 Zubeh r optional erh ltlich Kalibrierblock Feuchte ZB 911 9004 Kalibrierampullen f r Kalibrierblock ZB 911 9005 lieferbar f r 35 50 und 80 Feuchte Edelstahlsinterfiler 0 eee ZB 911 9003 Hinweis zum Edelstahlsinterfilter Da es im Praxiseinsatz h ufig zu Staub und Schmutzbelastung kommt was zu einer Verf lschung der Messergebnisse und Verk rzung der Sensorlebensdauer f hren kann ist das Messger t serienm ig mit einem Metallgitterfilter aus gestattet Verwenden Sie bei hoher Schmutzbelastung den als optio nale Wechsel Schutzkappe erh ltlichen Edelstahlsinterfilter Bedienungsanleitung Deutsch Table of Contents 1 Read before starting cece ee eee B 02 8 Notes on maintenance and operation
18. as Messger t kein Tastenfeld sondern ein Obere Displayanzeige K OH CO C CO sogenanntes Daumen Rad auf der linken Seite Untere Displayanzeige C F des Ger tes I Das Rad lasst eine 15 Drehbewegung nach unten und Unteres Men oben zu und kann in der Mittelstellung zus tzlich gedr ckt werden Betriebsarten Modi Mit diesen drei Bedienungspositionen lassen sich alle Einstellungen Konfigurationsarten zum Betrieb und zur Gr tekonfiguration vornehmen A 05 Bedienungsanleitung Deutsch Die drei Bedienungs Positionen des Daumen Rad Position mitte Drehbewegung Drehbewegung Symbol im nach oben nach unten weiteren Text gt Symbol im Symbol im weiteren Text M weiteren Text d Anweisungen zur Anwahl des oberen und unteren Men s sowie zur Auswahl der Betriebsart und Geratekonfiguration finden Sie in den entsprechenden Kapiteln Einschalten und Ausschalten des Ger tes Zum Einschalten dr cken Sie kurz die Mittelposition gt des Daumenrads Zum Ausschalten dr cken Sie ca 2 Sekunden die Mittelposition gt des Daumenrads Automatische Abschaltfunktion nach 3 Minuten Bedienungsanleitung Deutsch 5 Das obere Men Im oberen Men k nnen je nach gew hlter Betriebsart die Standard Funktionen HOLD MAX MIN AVG gew hlt werden Hold Hold friert den Messwert ein MAX MAX stellt den maximalen Wert im aktiven Zeitraum dar MIN MIN stellt den
19. chwarzer Farbe bedeckt werden Das Gerat kann nicht durch transparente Oberflachen wie z B Glas messen Stattdessen misst es die Oberfl chentemperatur des Glases Emissionsgrad Liste verschiedener Materialien f r den Bereich 0 bis 200 C ASD OSE rese esr est Qo te EA Bi alte 0 95 Asphalt m ee een 0 90 bis 0 95 Beinamen asian aaa nn 0 95 EE eaaa 2 ee ee 0 98 bis 1 00 Dachpappe pi ee ere einigen 0 95 Emaillelack schwarz 0 95 Erden ae fer 0 95 Farbe nichtmetalisch 0 00 cece cece eee eee 0 95 Bedienungsanleitung Deutsch Gipsy are vos Ep E CU EM EEE e 0 90 bis 0 95 Glas Wenige woe ee nen eat 0 85 bis 0 90 Heizk rperlack eu eg su 0 95 LEET 0 90 bis 0 95 Kalkstein ns esse nenne 0 95 tt ee A eet te 0 90 bis 0 95 Kunststoffe en ators Rhe Rees 0 90 Marmor oie reete ter emet ene een t 0 90 bis 0 95 P lz nie en EE 0 90 bis 0 95 Tapeten nichtmetallisch 0 95 Textilien nichtmetallisch 0 95 NOW aos AAT Gs nen ran er 0 95 Wassers ee sr ea 0 93 ZEMENT sn dc A decet ate el daturus tid 0 90 bis 0 95 Ziegel rau un epe dees EN gel 0 90 bis 0 95 10 technische Daten Lufttemperatur mire ee C F Me ssplirizip Ad de ante nmn Dn RES NIC Messbereich 20 6 06 c eee eee eee 20 bis 50 C Aufl SUhng uv dE extet ep rive e des 0 1 C Genauigkeit 0 4 C bei 0 bis 40 C sonst 0 7 C Luftfeuchte 22 0 o e v el r H 96 g m Messprinzip kapazitiv
20. e jetzt den Sensorenkopf des Mess ger tes bis zum Anschlag vorsichtig in den Kalibrierblock Warten Sie zwei Stunden Angleichzeit bis Sie den Abgleich laut der unter CAL beschriebenen Vorgehensweise zur Einpunktkalibrierung durchf hren A 17 Entfernen Sie den Sensorkopf aus dem Kalibrierblock Entsorgen Sie die Ampulle und reinigen den Kalibrierblock mit destilliertem Wasser Wichtig Kalibrierampullen nur einmal verwenden Wahrend der An gleichzeit darf sich die Temperatur nicht ndern Nur bei Zimmertem peraturen von 20 bis 21 C kalibrieren Kalibrierungen diirfen nur mitge eigneten Referenzwerten und von geschultem Personal ausgef hrt werden Bedienungsanleitung Deutsch 8 Hinweise zu Wartung und Betrieb Batteriewechsel Erscheint im Display die Anzeige BAT verbleibt je nach Betriebs modus eine Standzeit von einigen Stunden ffnen Sie den Batteriedeckel auf der Vorderseite des Ger tes Entnehmen Sie die leere Batterie und ersetzen Sie diese durch eine neue Verwenden Sie bitte ausschlie lich Batterien von Typ 9V E Block PP3 Keine Akkus verwenden Achten Sie bitte beim Einlegen der Batterie auf die korrekte Polung und verwenden Sie ausschlie lich hochwertige Batterien Bedienungsanleitung Deutsch Pflege Reinigen Sie das Ger t bei Bedarf mit einem feuchten Tuch Verwen den Sie keine Reinigungsmittel sondern nur klares Wasser zum An feuchten des Tuches Standortwechsel
21. minimalen Wert im aktiven Zeitraum dar AVG AVG stellt den arithmetischen Mittelwert im aktiven Zeitraum dar In das obere Men gelangen Sie mit f die erste ausw hlbare Funk tion blinkt Zur n chsten ausw hlbaren Funktion gelangen Sie durch erneute Ein gabe von f Die Funktionen sind nacheinander nur in einer Rich tung auswahlbar Wenn Sie eine Funktion die Sie auswahlen wollten bersprungen haben geben Sie solange ein bis die Funktion Ih rer Wahl wieder blinkt Zur Auswahl der gew nschten nun blinkenden Funktion best tigen Sie mit gt Eine best tigte Funktion wird statisch im Display ange zeigt M chten Sie keine Funktion ausw hlen und das obere Men verlas sen dann geben Sie ein Werden keine Eingaben get tigt wird das Men nach 20 Sekunden automatisch verlassen Um eine ausgew hlte Funktion wieder zu deaktivieren geben Sie gt ein A 07 Bedienungsanleitung Deutsch 6 Das untere Men Im unteren Men k nnen die drei Betriebsmodi TH Thermohygro meter IR Pyrometer und DP Taupunktalarmgeber sowie die Konfigurationsbereiche CFG Unit 1 Unit 2 CAL 1 CAL 2 gew hlt werden In das untere Men gelangen Sie mit d der erste ausw hlbare Men punkt blinkt Zum n chsten auswahlbaren Men punkt gelangen Sie durch erneute Eingabe von d Die Men punkte sind nacheinander nur in einer Rich tung ausw hlbar Wenn Sie einen Men punkt den Sie ausw hlen wollten bersp
22. mity has been proven the appropriate declarations and documents are kept by the manufacturer To maintain this status and ensure safe operation you the user must observe the instructions in this user guide Operating manual English Please read this user guide carefully before using the apparatus and follow all instructions Never make measurements on live parts Observe the measuring ranges of the measured value sensors overheating can destroy the instrument Only adjust temperature and humidity with a suitable reference When moving to a location with a different climate the instrument needs to acclimatise for several minutes A Use for the A Laser warning intended purpose This instrument is fitted with a class 2 laser Never direct the The measuring instrument must only be operated within the laser beam directly or indirectly from reflecting surfaces at the eyes bounds of the specified technical data 1 CAUTION Laser radiation can cause irreparable The measuring instrument must only be used under correct damage to the eyes The laser beam operating conditions and for the purpose for which it was vu det eh eam Must be deactivated when making constructed DIODE LASER measurements close to persons lt 1mW Output at 675 nm CLASS Il LASER PRODUCT e The operational safety is no longer guaranteed in the event of modifications or conversions ERG Operating manual English
23. ne Anwahl dieses Men punktes ist nur LO LI m glich wenn der DP Modus aktiv ist Der Wertebereich f r Hi und Lo geht von 0 0 bis 9 9 Die Einheiten C oder F werden von der aktuellen fiir den DP Modus ausge w hlten Temperatureinstellung bernommen Die minimale Differenz zwischen Hi und Lo muss 1 0 betragen Die Werte f r Hi obere Displayanzeige und Lo untere Displayan zeige werden nacheinander eingegeben und best tigt Unit 1 Mit Unit 1 w hlt man die Einheit der Temperatur Zur Auswahl stehen C und F W hlen kann man mit 8 und d best tigt wird mit gt Unit 2 Eine Anwahl dieses Men punktes ist nur m glich wenn der TH Modus aktiv ist Mit Unit 2 w hlt man die Einheit f r die absolute Luftfeuchte g m relative Luftfeuchte r H oder die Taupunkttem peratur dp C dp F Werksseitig ist r H voreingestellt W hlen kann man mit und d best tigt wird mit gt A 13 Bedienungsanleitung Deutsch Einpunktkalibrierungen von Temperatur CAL 1 Mit CAL 1 Einpunktkalibrierung und relativer Feuchte Stellt man den Offset f r Sensor 1 Tempe Mit CAL kann eine Einpunktkalibrierung f r die Sensoranzeigen m H r ratur ein Der Offset wird in der unteren von Temperatur CAL 1 und relativer Feuchte CAL 2 durchgef hrt L ULI Displayanzeige dargestellt Maximal lassen werden sich 10 C bzw 10 F einstellen ee M W hlen kann man mit 4 und Ay best tigt
24. or head from the calibration block Dispose of the ampoule and clean the calibration block with distilled water Important Only use the calibration ampoules once The tempera ture may not change during the acclimatisation time Only calibrate at room temperatures of 20 to 21 C Calibrations may only be made with suitable reference values and by trained personnel B 17 Operating manual English 8 Notes on maintenance and operation Changing batteries If BAT appears in the display a service life of a few hours remains depending on the operating mode Open the battery cover on the front of the instrument Take out the empty battery and replace it with a new one Please only use batteries of the type 9V E block PP3 Do not use rechargeable batteries Please ensure that the battery is inserted the right way round correct polarity and always use high quality batteries Care Clean the instrument when necessary with a damp cloth Do not use cleaning agents just clean water to moisten the cloth Changing locations Especially when moving from a cold to a warm environment for example taking inside a heated room after storing overnight in a car condensation may form on the circuit board depending on the humidity of the room This physical effect which cannot be prevented in any measuring instrument by constructional means leads to false measured values Therefore the display shows no measured values in this situa
25. rmgrenzwert Diese beiden Alarmgrenzwerte errechnen sich jeweils aus der gemes senen Taupunkttemperatur TdP und den individuell im Konfigurations bereich CFG definierten Schwellwerten dem oberen Schwellwert Hi und dem unteren Schwellwert Lo Die Additionssumme aus Taupunkttemperatur TdP und oberem Schwellwert Hi bildet den oberen Alarmgrenzwert TdP Hi Die Subtraktionssumme aus Taupunkttemperatur TdP und unterem Schwellwert Lo bildet den unteren Alarmgrenzwert TdP Lo Bedienungsanleitung Deutsch Sinkt die Oberfl chentemperatur unter den oberen Alarmgrenzwert TdP Hi wird ein akustischer Alarm Summer und optischer Alarm Laser blinkt ausgel st und proportional zur Ann herung an den un teren Alarmgrenzwert in seiner Intensivit t gesteigert Je weiter die Oberfl chentemperatur sinkt desto schneller steigt die Wiederholfrequenz des Summers und des Lasers Die maximale Wiederholfrequenz wird bei Erreichen des unteren Alarmgrenzwertes TdP Lo ausgel st Beispiel Die aktuelle Taupunkttemperatur TdP betr gt 2 C Sie de finieren den oberen Schwellwert Hi mit 5 C und den unteren Schwell wert Lo mit 5 C Der Alarm beginnt dann bei 7 C TdP Hi und erreicht seine st rkste Intensit t bei 3 C TdP Lo 6 2 Konfigurationsbereiche CFG Im CFG Modus k nnen die Werte f r H l I den oberen Schwellwert Hi und den unte ren Schwellwert Lo eingegeben werden I C T Ei
26. run gen haben geben Sie solange ein bis der Men punkt Ihrer Wahl wieder blinkt Zur Auswahl des gew nschten nun blinkenden Men punktes best tigen Sie mit gt M chten Sie keinen Men punkt auswahlen und das untere Men verlassen dann geben Sie ein Werden keine Eingaben get tigt wird das Men nach 20 Sekunden automatisch verlassen Bedienungsanleitung Deutsch A 08 6 1 Betriebsarten TH Durch Auswahl des TH Mo dus l sst sich das Ger t in der Be triebsart Thermohygrometer ein R DP CFG setzen In der oberen Displayanzeige wird in diesem Modus die Temperatur und in der unteren Displayanzeige die Luftfeuchte angezeigt Die Einheit f r die Temperatur C F wird in dieser Betriebsart im Konfigurationsbereich Unit 1 eingestellt Werksseitig ist C voreinge stellt Die Rechengr en f r die Luftfeuchte r H g m dp C dp F werden im Konfigurationsbereich Unit 2 eingestellt Werksseitig ist r H voreingestellt Eine Offsetkorrektur von Temperatur und Feuchtewerten kann in den Konfigurationsbereichen CAL 1 und CAL 2 vorgenommen werden IR Durch Auswahl des IR Modus l sst sich das Ger t in der Betriebs art Pyrometer einsetzen In die sem Modus kann die ber hrungs lose Messung der Temperatur von Oberfl chen durchgef hrt werden Das Ger t wurde ausschlie lich f r pyrometrische Oberfl chentemperaturmessungen in Innenr umen konzipiert TH DP CFG In der oberen Displa
27. s respectively with a gt at any time in this mode Operating manual English Using the alarm function The alarm function is automatically active in the DP mode The alarm triggering and intensity is determined by the upper alarm limit and the lower alarm limit These two alarm limits are calculated respectively from the measured dewpoint temperature TdP and the thresholds the upper threshold Hi and the lower threshold Lo defined individually in the configu ration area CFG The addition sum of the dewpoint temperature TdP and the upper thres hold Hi forms the upper alarm limit TdP Hi The subtraction sum of dewpoint temperature TdP and lower threshold Lo forms the lower alarm limit TdP Lo Operating manual English If the surface temperature drops below the upper alarm limit TdP Hi an acoustic alarm buzzer and optical alarm laser flashes is trig gered and increased in intensity proportional to approaching the lower alarm limit The further the surface temperature drops the faster the repetition frequency of the buzzer and laser The maximum repetition frequency is triggered on reaching the lower alarm limit TdP Lo Example The current dewpoint temperature TdP is 2 C You define the upper threshold Hi with 5 C and the lower threshold Lo with 5 C The alarm then begins at 7 C TdP Hi and reaches maximum intensity at 3 C TdP Lo 6 2 Configuration areas CFG
28. t nur in seltenen Fallen not wendig Falls hohe Anforderungen an die Genauigkeit erforderlich sind empfehlen wir einmal j hrlich eine Einpunktkalibrierung des Sensors durchzuf hren Grunds tzlich k nnen Einpunktkalibrierungen selbst durchgef hrt werden wovon wir aber abraten da professionelle Referenzwerte meistens nicht verf gbar sind Nutzen Sie stattdessen die M glichkeit einer zertifizierten Kalibrie rung nach DKD und ISO N here Informationen hierzu erhalten Sie durch Ihren H ndler Einpunktkalibrierung r H mit Kalibrierblock und Kalibrierampulle berpr fen Sie den Kalibrierblock auf Fremdk rper bzw R ckst nde von vorherigen Kalibrierungen und reinigen Sie den Block gegebenenfalls kopf a ec im Zur Kalibrierung stehen drei verschiedene ampulle r Kalibrierfl ssigkeiten f r die Feuchtewerte i 35 50 und 80 zur Verf gung F r Standardkalibrierungen sollte nur die 50 Kalibrierfl ssigkeit verwendet werden Sensor Bedienungsanleitung Deutsch A 16 Bitte beachten Sie unbedingt die auf der Beilage der jeweiligen Kalibrierampulle angegebenen Daten und Pr fwerte Brechen Sie den Kopf der Kalibrierampulle ab Nehmen Sie den Ka librierblock so in die Hand dass Sie die Kalibrierampulle von unten einschieben k nnen Stellen Sie dann den Kalibrierblock auf einer ebe nen Fl che ab und vergewissern sich dass die Kalibrierfl ssigkeit in den Kalibrierblock l uft Stecken Si
29. this mode C is set at the factory The units for humidity r H g m dp C dp F are set in configuration area Unit 2 r h is set at the factory An offset correction of temperature and humidity values can be made in the configuration areas CAL 1 and CAL 2 IR The instrument can be used as a pyrometer when you select the IR mode In this mode you can mea TH DP CFG sure the temperature of surfaces without contact The instrument has been designed exclusively for interior pyrometric surface temperature measurements The surface temperature is indicated in the upper display in this mode The corresponding HOLD MAX MIN or AVG value is indicated in the lower display if one of these functions was selected in the upper menu B 09 Operating manual English The unit of surface temperature C F is set in configuration area Unit 1 in this mode C is set at the factory An offset correction of the temperature values can be made in confi guration area CAL 1 The laser light is switched on for measurement location marking as soon as the IR mode is selected The laser lights for a maximum of 2 minutes and can be activated for another 2 minutes respectively with a gt at any time in this mode Distance and measuring spot size E M To achieve accurate measuring results the target must be larger than the measuring spot of the instrument The determined temperature is the average temperature of the measured
30. tion Operating manual English B 18 Please wait about a minute until the instrument has acclimatised before continuing the measuring process Ambient conditions Storage Operation Permissible R temperature 30 C 460 C 0 C 50 C issi lt 95 rH resp Permissible 95 rH lt 20 g m the smaller humidity value applies Permissible height above 5 000 m 5 000 m Sea level 9 Notes on emission factor The emission factor is a value used to describe the energy radiation characteristic of a material The higher the value the higher the capacity of the material to emit radiation Many organic materials and surfaces have an emission factor of about 0 95 Metal surfaces or glossy materials have a low emission factor and therefore give inaccurate measured values Please take this into account when using the measuring instrument Operating manual English For compensation the surface of glossy parts can be covered by adhesive tape or matt black paint The instrument cannot measure through transparent surfaces such as glass Instead it measures the surface temperature of the glass Emission factor list for different materials for the range 0 to 200 C ASDOSI0S scies ie owed E d d rd 0 95 Asplialtz ce tee denotes E 0 90 to 0 95 Bitumen ceres eov s oe Rene eR d 0 98 to 1 00 Brick rough reed an denen 0 90 to 0 95 Cemerss e e ere eres te ea E 0
31. yanzeige wird in diesem Modus die Oberfl chen temperatur angezeigt In der unteren Displayanzeige wird der ent Bedienungsanleitung Deutsch sprechende HOLD MAX MIN oder AVG Wert angezeigt falls eine dieser Funktionen im oberen Men ausgew hlt wurde Die Einheit f r die Oberfl chentemperatur C F wird in dieser Betriebsart im Konfigurationsbereich Unit 1 eingestellt Werksseitig ist C voreingestellt Eine Offsetkorrektur der Temperaturwerte kann im Konfigurations bereich CAL 1 vorgenommen werden Sobald der IR Modus ausgew hlt wurde schaltet sich zur Mess ort Markierung das Laserlicht ein Der Laser leuchtet f r maximal 2 Minuten und kann in dieser Betriebsart jederzeit ber ein gt f r jeweils weitere 2 Minuten aktiviert werden Bedienungsanleitung Deutsch Entfernung und Messfleckgr e E M Um genaue Messergebnisse zu erzielen muss das Messobjekt gr er als der Messfleck des Ger tes sein Die ermittelte Temperatur ist die Durchschnittstemperatur der gemessenen Fl che Je kleiner das Messobjekt ist desto k rzer muss die Entfernung zum Ger t sein Die genaue Messfleckgr e k nnen Sie dem Schaubild entnehmen F r genaue Messungen sollte das Messobjekt wenigstens doppelt so gro wie der Messfleck sein strahl E M 8 1 DP DP DewPoint Taupunkt Durch Auswahl des DP Modus l sst sich das Ger t in der Betriebsart TH P CFG Taupunktalarmgeber einsetzen Der DP Modus