Restek's Electronic Leak Detector
Contents
1. Fully charged off STEADY during the Fast Charge phase NOTE For best battery performance Restek recommends plugging the unit into the wall charger when not in use NOTE Replacement of the rechargeable cells in this unit is performed at the factory There are no serviceable parts in this unit Opening the case or tampering with the internal parts will void the factory warranty 3 www restek com RESTEK 3 0 Powering Up Depress and hold the power button Figure 1 until the unit responds with the wake up mode The leak detector will run through a self calibration sequence for approximately 15 seconds During this time DO NOT attempt to zero the unit 4 0 Zeroing the Unit After the LED lights stop flashing the unit is ready for use The instrument may need to be zeroed periodically between uses especially if it is moved from room to room or between areas of differing temperature or humidity Do not attempt to zero the unit while the probe is stored in the holder The probe MUST be removed from the storage container before zeroing the unit To re zero press the Zero button The unit will run a self calibration sequence for approximately 4 seconds When all LED lights stop flashing and the green LED light is lit the unit is ready for use Figure 1 Leak Detector schematic Sample Probe LED Leak Indicator Lights LED Battery Charge Indicator Light Power Button Zero Set Button Reference Gas Inlet Keep Clear
2. Sample Probe Stylus BACK VIEW FRONT VIEW NOTE To avoid false readings do not attempt to use or zero the unit while the self calibration sequence is in progress 5 0 Prior to Operation Verify the operation of the Leak Detector before each use by sampling gas from a GC split vent or other source of hydrogen or helium Also visually inspect the probe tip reference gas inlet and exhaust port for obstructions Figure 1 IMPORTANT Fittings being checked must be clean and dry liquid leak detecting agents dust and other debris may damage the Leak Detector if drawn into the probe The Leak Detector responds to almost any gas you can smell and many gases that you can t smell Solvent vapors split vent exhaust or even strong air currents around the probe or reference inlet can cause instability or false positive readings Be careful not to breathe into the reference inlet when checking for leaks or to cover block the inlet with your hand gt RESTEK www restek com 4 6 0 Detecting Leaks Slowly move the probe tip around fittings and other potential leak sources If the Leak Detector senses a gas other than air the LED bar graph will begin to light and an alarm will sound when the last LED light illuminates The red LED lights indicate helium and hydrogen leaks The yellow LED lights indicate a nitrogen argon or carbon dioxide leak Remove the probe from the vicinity of the leak and allow the unit to return to zero
3. 3 Outside the USA Contact your Restek representative Restek U S 110 Benner Circle Bellefonte PA 16823 phone 814 353 1300 800 356 1688 fax 814 353 1309 www restek com Restek France phone 33 0 1 60 78 32 10 fax 33 0 1 60 78 70 90 e mail restek restekfrance fr Restek GmbH phone 49 0 6172 2797 0 fax 49 0 6172 2797 77 e mail info restekgmbh de Restek Ireland phone 44 0 2890 814576 fax 44 0 2890 814576 e mail restekireland aol com Restek Japan phone 81 3 6459 0025 fax 81 3 6459 0025 e mail restekjapan restek com Thames Restek U K LTD phone 44 0 1494 563377 fax 44 0 1494 564990 e mail sales thamesrestek co uk
4. Detector s LED light response range The greater Gas Leak Rate atm cc sec LED Light Color the number of red or yellow LED lights lit correlates in general to the size of Helium a Red the leak NOTE The Leak D F itative devi ther iti Hydrogen 10x10 Red e leak Ihe Le etector is not a quantitative device rather it is Nitrogen 14x10 Yellow designed to detect leaks in gas line connections commonly associated with Argon 1 0x10 Yellow laboratory equipment Carbon dioxide 10x10 Yellow Figure 5 LED light response Green LED and all 3 yellow LEDs and chart for the Leak Detector A abis COAC Green LED and 1 1 ratio Of prove reference indicates alarm significant 4 red LEDs no leak present Red LED lights leak detected small leak detected indicate the presence of one or more of the following gases helium or hydrogen Yellow LED lights indicate the presence of one or more of the following gases nitrogen argon or car Grean LED only Green LED and bon dioxide no leak detected all 7 red LEDs and audible continuous alarm significant leak detected CAUTION This unit is designed to detect TRACE AMOUNTS of hydrogen arising from a small leak in a nonflammable environment e g laboratory room air etc This unit is rated for use in a nonflammable atmosphere where the sample gas may become sufficiently high concentration to become explosive Tip drift Tip drift is the phenomenon when a false LED light response is registered as t
5. If a large amount of gas has entered the probe it may take a few seconds for the instrument to clear itself Please do not attempt to zero the unit while it is clearing out the gas from the probe This may cause the unit to malfunction Place the probe near the leak again to confirm its location The reference gas inlet Figure 1 must not be restricted or the unit will not operate correctly Similarly the exhaust port allows the gas being tested to exit the Leak Detector and must remain unobstructed The exhaust port is located in the probe docking station CAUTION This unit is designed to detect TRACE AMOUNTS of hydrogen arising from a small leak in a nonflammable environment e g laboratory room air etc This unit is rated for use in a nonflammable atmosphere where the sample gas may become sufficiently high concentration to become explosive 7 0 Specifications Power Rating 12 Volts DC battery charger supplied Warranty 1 year warranty Battery Rating 6 hours normal operation Certifications CE Ex and Japan Figure 2 Figure 3 Operating Temp Range 32 120 F 0 48 C Compliance WEEE RoHS Cap unscrewed Cap removed Humidity Range 0 97 and partially exposing probe tip removed brush for cleaning 8 0 Maintenance Avoid spilling liquids onto the unit or it may malfunction If a liquid is spilled onto the unit turn off the power immediately remove heavy liquids with a dry towel and let the unit sit until the liquid dries Dust and
6. Oberflachenbeschaffenheit Einspritzgeformtes Versaflex FX1040X 1 Pantone 426C Schwarz mit MT 11030 Oberflachenbehandlung F r Fragen Probleme Reperaturservice Innerhalb der USA Rufen Sie Restek Customer Service unter 001 814 353 1300 ext 3 an Au erhalb der USA Kontaktieren Sie Ihren Restek H ndler RESTEK www restek com 20 800 20 707 6002 20 70 ABA JO 10WDeI1q suey sewon 6 8CC AJ TOPPA JLT ET89OT Vd UWOPJPEI SPATZ Isuusg OTT uoneIodion Yo soy S007 ST 62009NA 9007 0 6Z009NH V SSEID EV M Z661 97EI9 94 6 76 PANIC XALV I3 801 7007 ANd IWY NA a nyeugig ed PL UONISOA uonereppadq sry Sunyew 107 aqisuodsai uosiag s Jaquinu Zoey quatudinbg Jo adA1 SSIIPPY SIE SUWeN SIA porepaq st AyTULIOFUOY YOIyM 0 sprepur g SJA punopo jo uoneorddy ALINYOANOD JO NOILVYY193QA 3 JUSH G Optional Accessories Soft Side Carry Storage Case Ideal for storing the leak detector in smaller spaces such as a tool box cat 22657 Small Probe Adaptor Verify hard to reach leaks using the small probe adaptor cat 22658 23 www restek com RESTEK Innovative Chromatography Products 2009 Restek Corporation All rights reserved Printed in the U S A 204 07 006 Rev date 05 12 9001 17025 Guide 34 For questions problems repair services Within the USA Call Restek Customer Service at 800 356 1688 or 814 353 1300 ext
7. debris can enter the probe tip of the Leak Detector and over time can clog the small bore tubing inside the unit To prevent this clean the probe tip periodically To clean the probe tip unscrew the cap to expose the brush Figures 2 and 3 Gently clean the probe using a small brush or your fingers to remove dust and debris then replace the cap Do not use liquids to clean the probe Liquids can damage the Leak Detector if drawn in through the probe Information on where to have the unit sent for maintenance or service is listed at the end of this d ocument 5 www restek com RESTEK 9 0 Troubleshooting Problem Possible Cause Suggested Solution Sensitivity decreased Probe clogged Clean the probe tip to remove any debris Probe line punctured Visually inspect probe line for holes Weak battery Recharge or return to Restek for battery replacement Response decreased Detector not zeroed Re zero detector LED bar graph stays lit during operation Detector re zeroed before unit was purged out Allow adequate time for detector to purge then re zero Reference gas inlet covered by hand or Remove obstruction other object Does not power up Batteries need to be charged Charge unit for 24 hours Contact Restek or your Restek representative for return instructions for servicing a damaged unit Additional charges may apply if the warranty has expired or the unit is damaged due to misuse RESTEK www restek co
8. restek com 14 0 Service The Restek Leak Detector carries a 1 year limited warranty from time of purchase Please have the Leak Detector serial number available when calling Restek with any concerns you may have Additional charges may apply if the warranty is expired or the damage is due to misuse Expected battery lifetime is 2 years from time of purchase Customers will need to return the unit to Restek for battery replacement At that time preventative maintenance services can also be performed on the unit A fee will be charged for servicing the unit For questions problems repair services Within the USA Call Restek Customer Service at 800 356 1688 or 814 353 1300 ext 3 Outside the USA Contact your Restek representative 10 Restek s Elektronischer m 1 Leak Detektor Kat 22839 Bedienungsanleitung F r beste Batterieleistung sollte das Ger t bei Nichtbenutzung im Ladeger t aufbewahrt werden x be Coa ae AN RESTEK www restek com V 1 0 20 3 0 40 5 0 CE AA 6 0 7 0 amp oo 9 0 10 0 11 0 12 0 13 0 14 0 2009 Restek Corporation Alle Rechte vorbehalten Kein Teil dieser Anleitung darf in jeglicher Form ohne vorherige Genehmigung von Restek Corporation vervielf ltigt werden Gedruckt in den USA Restek Corporation 110 Benner Circle Bellefonte PA 16823 N N Inhaltsverzeichnis Seite EINLEITUNG nn ee 13 AKKU AUNSdenNa 0 ee eier 13 Einscha
9. ED Das GR NE LED blinkt auch wenn das Ger t eingeschalten wird Definition der Auflademodi Langsamaufladung Dieser Modus ist aktiv wenn die Batterien unter einem vorgegebenen Schwellenwert entladen werden Die Langsamaufladung gibt dem Akku einen sehr langsamen Aufladepuls Dieser l dt den Akku sehr langsam auf bis er ber dem Schwellenwert ist Dann geht das Aufladeger t in den Schnellaufladungsmodus ber Langsamaufladung kann mehere Stunden dauern und l uft normaler weise ber Nacht Das Akku Aufladungslicht BLINKT einmal pro Sekunde w hrend der Langsamaufladung Auflademodi Auflade LED Schnellaufladung Dieser Modus ist nach der Langsamaufladung aktiv oder wenn das Langsamaufladung amp waltz Aufladeger t in das Ger t eingesteckt wird Akkuspannung noch ber dem vor nen De A ee be und uspannung noch ber dem vorgegebene Schnellaufladung COI st ndig an Schwellenwert ist In diesem Modus wird der Akku schnell aufgeladen Das Akku Aufladungslicht ist in diesem Modus STANDIG an Voll aufgeladen Aus Bemerkung Fiir beste Akku Leistung empfehlen wir den Leak Detektor bei Nichtgebrauch eingesteckt zu lassen Bemerkung Die Akkus im Ger t k nnen nur vom Hersteller ersetzt werden Es gibt keine Teile in diesem Ger t die Wartung ben tigen Wenn Sie das Geh use ffnen oder die internen Teile entfernen verf llt die Garantie des Herstellers AANS 13 www restek com RESTEK 3 0 Einschalten Die Power Taste Abbildu
10. Restek s pS Electronic Xe Leak Detector cat 22839 iy 4 Instruction Manual keep unit on the charger when not in use ATNA RESTEK www restek com r aW Table of Contents Page IN 1 0 HITOdUCHONn ee er r 3 A 20 Battery Charta ee etesen aia 3 3 0 Powering Bauern ee 4 l 4 0 Zeroing CIS MMIC ana ee 4 5 0 Prior to Operation 4 C A 6 0 Detecting Leaks era ee 5 7 0 SPEcifiCatiON nee 5 Ex 8 0 MaintenaNCE nee 5 9 0 Troubleshooting 6 10 0 Technology u 7 11 0 Interpretation of Results ueesssesssssesssnensnnennnnenennnnennnennen 8 12 0 Back Label Explanation eeesseressseessnsneensnnnnennnnnnennnnnnnnenn 9 13 0 Case Specifications anne 10 14 0 Servieren 10 Declaration of Conformity sssssssesssssesersssrssereerersresrerse 21 2009 Restek Corporation All rights reserved No part of this manual may be reproduced in any form or by any means without the prior agreement or permission by Restek Corporation Printed in the USA Restek Corporation 110 Benner Circle Bellefonte PA 16823 Operating Instructions 1 0 Introduction Restek s portable Leak Detector is specifically designed for use with gas chromatography GC systems It detects minute leaks of any gas with a thermal conductivity different from air The reference gas inlet Figure 1 draws in ambient air for comparison to air drawn into the sample probe A leak is indicated by both an LED light display and an audible a
11. he unit is quickly turned or swept in dramatic arc movements Tip drift is inherent to all dual thermistor leak detector technology and is based in large part on the asymmetry of the flow cells shaking or tipping the unit influences the air flow profiles which impacts the rates of heat exchange If the device is functioning normally the LED light signal will return to zero in 3 5 seconds after the unit is held still In extreme cases the unit may require another zero cycle before using To avoid tip drift be sure to hold the unit steady while making measurements RESTEK www restek com 8 12 0 Back Label Explanation Pe Ne oe 10 11 12 Product name Product catalog number Product serial number Warning note This plastic case does not exhibit adequate surface resistance properties suitable for high electric fields DO NOT CHARGE THIS DEVICE IN A HAZARDOUS AREA Definition of symbols see table at right This unit conforms to EU EMC Directive 2004 108 EC Standards to which Conformity is declared include 61326 1997 w A3 Class A Unit is WEEE compliant Unit is RoHS compliant This complies with part 15 ofthe FCC Rules ExnC Operation is subject to the following 2 conditions 1 This device may not cause harmful interference and 2 this device must accept any interference NG received including interference that may cause 36 undesired operation This Class A digital apparatus complies with Canad
12. ian ICES 003 Units must be sent back to Restek Corporation for service Manufacturer company name address and contact information 2009 Definition of back label symbols EN60079 0 2006 Electrical apparatus for explosive gas atmospheres Part 0 General Requirements EN60079 15 2005 Electrical apparatus for explosive gas atmospheres Part 15 Construction test and marking of type of protection n electrical apparatus Group II applies to areas above ground environments Gas Group IIC relates to hydrogen and related gas types Category 3 relating to gas analysis normal safety measure Sufficient safety during normal operation Normal operation described as monitoring for trace amounts of flammable or explosive gases in a nonflammable environment While testing this unit neither internal nor external elements exceed 85 C Additional information Operating range 32 F lt Tam lt 120 F 0 C lt Tam lt 48 C This unit is designed to detect TRACE AMOUNTS of hydrogen arising from a small leak in a nonflammable environment e g laboratory room air etc This unit is rated for use in a nonflammable atmosphere where the sample gas may become sufficiently high concentration to become explosive Year of construction www restek com RESTEK 13 0 Case Specifications D With MT 11090 SURFACE TEXTURE INJECTION OVERMOLD VERSAFLEX 1063 1 PANTONE 4260 MACK DED SURFACE TREATMENT A RESTEK www
13. ie sich mit Restek oder Ihrem Restek H ndler in Verbindung wenn Sie Anleitungen brauchen um ein besch digtes Ger t zur Reperatur einzuschicken Es k nnen zus tzliche Geb hren auftretten wenn die Garantie abgelaufen ist oder das Ger t durch Mi handlung besch digt wurde RESTEK www restek com 16 10 0 Technik Die Leak Detektor Messtechnik beruht auf dem Vergleich der W rmeleitf higkeit zwischen dem gemessenen Gas und Luft Das Ger t benutzt eine Dual Thermistor Technik die das Verh ltnis der W rmewechselwerte Probengas Luft mi t und das Ergebnis auf einer LED Anzeige bertr gt Abbildung 4 Unter ideallen Bedingungen ziegt ein Verh ltnis von 1 1 an dass sich das gleiche Gas in der Pr fspitze wie auch in der Referenzzelle befindet und es daher kein Leck hat Abbildung 4 Diagramm der Leak Detektor Technik LINKS Dualanalyse wird dadurch erreicht dass sich die W rmeelmente in getrennten Flu r umen angebracht sind RECHTS Pr fspitze und Referenzgasflu werden simultan auf W rmeleitf higkeit berwacht Unterschiede in der Zusammnesetzung der Gase wird Puto Reference durch Unterschiede im Stromflu der ar uow ar Heater Heater z clamara dament W rmeelmente angezeigt Probe Reference Probe Reference curent curent prode Reterence Da es kleine Unterschiede in der Lufttemperature und oder Luftfeuchtigkeit zwischen dem Referenzinlet Abbildung 1 und der Pr fspitze geben kann gilt eine kleine Resp
14. larm Restek s Leak Detector is manufactured by Restek so you are assured of the same Restek quality and service you have come to recognize with the Restek name Should you require assistance at anytime regarding our Leak Detector please contact Restek Customer Service at 1 800 356 1688 or 814 353 1300 ext 3 2 0 Battery Charging The Leak Detector should be fully charged prior to use Only use the wall charge unit provided When the unit s charge is low the GREEN LED located between the red and amber LEDs will begin to flash If unit is off the GREEN LED will flash when the power button is depressed Define Charging Modes Trickle Charge This mode is active when the batteries have discharged below a predetermined threshold The Trickle Charge gives the batteries a very slow charging pulse This restores the batteries at a very slow rate until the batteries are above the threshold Then the unit goes into Fast Charge Mode Trickle Charge Mode can take several hours to complete usually over night The Battery Charging light FLASHES once a second during the Trickle Charge phase Charging Modes Charge LED Fast Charge Mode This mode is active after the Trickle Charge mode or if the instrument is Tape Stenger ee ape h plugged into the charger and the battery voltage is above the predetermined threshold During Fast charge mode CO Steady on this mode the batteries are being charged at a high current rate The Battery Charging light is on
15. lgenden Gase vorhan Nur gr ne LED kein Gr ne LED und alle 7 roten den sind Stickstoff Argon oder Leck detektiert LEDs und ein kontinuierlich h rbarer Alarm Signifikantes Kohlendioxyd Leck detektiert VORSICHT Diese Ger t wurde entwickelt um Spuren von Wasserstoff von einem kleinen Leck in einer nicht brennbaren Umgebung nachzuweisen z B in Laborluft usw Diese Ger t ist geignet f r eine nicht brennbare Atmosph re in der das Probengas zu einer Konzentartion ansteigen kann wo es explosiv werden kann Pr fspitzenabweichung Pr fspitzenabweichung ist das Ph nomen wenn eine falsche LED Licht Response registriert wird wenn das Ger t schnell bewegt wird Pr fspitzenabweichung kommt bei allen Dual Thermistor Leak Detektoren vor Sie basiert auf der Asymmetry der Flu zellen sch tteln oder neigen des Ger ts beinflusst die Luftstr mungsprofile was sich wiederum auf die W rmewechselraten auswirkt Wenn das Ger t normal funk tioniert kehrt das LED Lichtsignal in 3 5 Sekunden wieder auf Null zur ck sobald das Ger t wieder still steht In extremem F llen kann es sein dass man das Ger t vor dem n chsten Gebrauch wieder auf Null einstellen muss Um Pr fspitzenabweichung zu vermeiden sollte man das Ger t w hrend der Messungen still halten RESTEK www restek com 18 12 0 Erkl rung des Aufklebers auf der iO D f u 10 11 12 R ckseite Name des Produktes Artikel Nummer des Produktes Serien N
16. lten ee een 14 Nullabstininungeeene een 14 Vor Inbetriebnahme uurssessseesensessnnennnennnnennnennnnnnen 14 LORS linden 15 SpezifikatioNe ee 15 ATARE IANS seen 15 Troublesho6ting un uu u u0s000ane a 16 Technik a a 17 Interpretation der Ergebnisse uuensssenssserssensnnnnenenennnn 18 Erkl rung des Labels auf der R ckseite sesen 19 Spezifikationen des Geh uses eueesessneeesnneeneennenennnenenn 20 SElVI le elek 20 Konformitatserkl tund nsaseiseissanene 21 Bedienungsanleitung 1 0 Einleitung Restek s tragbarer Leak Detektor wurde speziell f r die Benutzung in Gaschromatographie GC Systemen entwickelt Er detektiert sehr kleine Leaks von jedem Gas dass eine andere w rmeleitf higkeit als Luft hat Der Referenzgas Inlet Abbildung 1 saugt Luft aus der Umgebung an die zum Vergleich mit dem Gas in der Pr fspitze benutzt wird Ein Leck wird sowohl durch eine LED Lichtanzeige wie auch durch ein h rbares Signal angedeutet Restek s Leak Detektor wird von Restek hergestellt Dadurch wird die Qualit t gew hrleistet die Sie mit dem Namen Restek identifizieren Sollten Sie jemals Hilfe mit Ihrem Leak Detektor ben tigen wenden Sie sich bitte an den Restek Customer Service 2 0 Akku Aufladen Der Leak Detektor sollte vor Gebrauch vollst ndig aufgeladen werden Dazu nur das mitgelieferte Aufladeger t benutzen Wenn der Akku fast leer ist blinkt ein GR NES LED es befindet sich zwischen den roten und braunen L
17. m 6 10 0 Technology The Leak Detector measurement is based on thermal conductivity comparisons between the probe air and a reference air The device employs a dual thermistor technology which measures the ratio of probe reference heat exchange values and displays the results on an LED scale Figure 4 Under ideal operating conditions a ratio of 1 1 indicates identical air samples for both probe and reference and therefore no leak is present Figure 4 Schematic layout of the Leak Detector technology LEFT Dual analysis is achieved with heater elements positioned in separate flow chambers RIGHT Probe and reference air streams are simultaneously monitored for thermal conductivity Differences in air composition are indicated by differences in the heater Probe Reteronce sac ace ar tow tw element currents ii P 3 R moen Probe Reference curent curent lProde Reference Because of slight differences in air temperature and or humidity between the reference inlet Figure 1 and the probe tip a small response indicated by a single red or yellow LED light is generally insufficient to positively identify a gas leak Small to moderate leaks are reliably indicated with 4 red LED lights larger leaks are indicated with all red LED lights or yellow LED lights lit and the continu ous alarm is audible Paa 7 www restek com RESTEK 11 0 Interpretation of Results Minimum Detectable Indicating Figure 5 illustrates the Leak
18. ng 1 dr cken bis das Ger t mit dem Aufwachmodus anspricht Der Leak Detektor f hrt dann eine Abbildung 1 Leak Detektor Diagramm Selbstkalibriersequenz durch die 15 Sekunden dauert W hrend dieses Ablaufs auf keinen versuchen das Ger t auf Null einzustellen Priifspitze 4 0 Nulleinstellung 2 m u Referenzgas Wenn die LED Lichter mit Blinken aufh ren ist das Ger t zur Benuztung ra sia Inlet indikator Lichter A bereit Das Ger t muss von Zeit zu Zeit wieder frisch auf Null eingestellt weten warden Besonders dann wenn es in einem andren Raum benutzt wird reat eg oder wenn sich die Temperatur oder Luftfeuchtichkeit ndert Niemals eine Nulleinstellung versuchen wenn sich die Pr fspitze in der Halterung befindet Die Pr fspitze MUSS vor der Nulleinstellung aus der Halterung genommen Nalleinsellandiaate werden Zur Nulleinstellung die Zero Taste dr cken Das Ger t f hrt eine Ansicht Selbstkalibriersequenz durch die etwa 4 Sekunden dauert Wenn alle LED vorne Lichter mit Blinken aufh ren und das gr ne LED Licht aufleuchtet ist das Ger t zur Benutzung bereit Pr fspitze Ansicht hinten Bemerkung Um falsche Messwerte zu vermeiden sollten Sie nicht versuchen das Ger t w hrend der Selbstkalibriersequenz zu benutzen 5 0 Vor der Inbetriebnahme berpr fen Sie den Betrieb des Leak Detektors vor jedem Gebrauch indem Sie ihn mit Gas vom GC Split Vent oder anderen Quellen von Wasserstoff oder Heli
19. onse nur ein einziges rotes oder gelbes LED Licht im Allgemeinen nicht als positives Zeichen f r ein Leck Kleine bis mittelm ssige Lecks warden zuverl ssig durch 4 rote LED Licheter angezeigt und bei grossen Lecks leuchten alle roten LED Lichter oder gelbe LED Lichter auf und ein kontinuierlicher Alarm ist h rbar Paa 17 www restek com RESTEK 11 0 Interpretation der Resultate Minimale nachweisbare Angezeigte Abbildung 5 veranschaulicht den Bereich der LED Lichter des Leak Gas Leckrate atm cc sec LED Licht Farbe Detektors Je mehr rote oder gelbe LED Lichter aufleuchten desto gr sser Helium z 10x iv Rot ist das Leck Bemerkung Der Leak Detektor ist kei titati Wasserstofi 1 0x10 Rot ist das Leck ung Der Leak Detektor ist keine quantitative Stickstoff 14x10 Gelb Vorrichtung Er wurde entwickelt um Lecks in solchen Gasleitungen zu Argon 1 0x10 Gelb finden wie sie gew hnlich in anaytischen Labors vorkommen Kohlendioxyd 10x10 Gelb Abbildung 5 LED Licht Response Gr ne LED und alle 3 gelben Tabelle f r den Leak Detektor LEDs und ein kontinuierlich Gr ne LED und 4 s is A r h rbarer Alarm Signifikantes rote LEDs kleines Ein Verh ltnis von 1 1 zwischen L ck detektiert oo IPr fspitze Referenz zeigt an dass es kein Leck hat Rote LED Lichter Pr fspitze zeigen an dass ein oder mehere der l Referenz folgenden Gase vorhanden sind Helium oder Wasserstoff Gelbe LED Lichter zeigen an dass ein oder mehere der fo
20. pft werden Um das zu vermeiden sollte die Pr fspitze von Zeit zu Zeit periodisch gereinigt werden Um die Pr fspitze zu reinigen schrauben Sie den Deckel ab bis Sie die B rste sehen Abbildungen 2 und 3 Die Pr fspitze vorsichtig mit der B rste oder den Fingern reinigen Staub und R ckst nde entfernen und dann den Deckel wieder aufschrauben Keine Fl ssigkeiten zum Reinigen benutzen Fl ssigkeiten k nnen den Leak Detektor besch digen wenn Sie in durch die Pr fspitze in das Ger t gesaugt werden Information ber Wartungs oder Reperaturstellen finden Sie am Ende dieser Anleitung AANS 15 www restek com RESTEK 9 0 Troubleshooting Problem M gliche Ursache Vorgeschlagene L sung Verminderte Empfindlichkeit Pr fspitze verstopft Die Pr fspitze reinigen um R ckst nde zu entfernen Pr fspitzenleitung hat ein Loch Die Leitung auf L cher inspizieren Schwacher Akku Akk aufladen oder bei Restek einen neuen Akku bestellen Verminderte Response Detektor nicht auf Null eingestellt Detektor wieder auf Null einstellen LED Lichter bleiben w hrend der Benutzung Detektor auf Null eingestellt bevor das Ger t voll Genug Zeit zum Aussp len lassen dann wieder auf st ndig an st ndig ausgesp llt war Null einstellen Referenzgasinlet mit der Hand oder anderem Behinderung entfernen Gegenstand blockiert Ger t sch llt nich ein Akku muss aufgeladen werden Das Ger t 24 Stunden aufladen Setzen S
21. rheit w hred des Normalbetriebs Normalbetrieb heisst die berwachung von Spuren von brennbaren oder explosiven Gasen in einer nicht brennbaren Umgebung W hrend diese Ger t gepr ft wurde hat die Temperatur der externen und internen Teile nie 85 C berschritten Zus tzliche Information Betriebsbereich 0 C lt Tam lt 48 C Diese Gerat wurde entwickelt um Spuren von Wasserstoff von einem kleinen Leck in einer nicht brennbaren Umgebung nachzuweisen z B in Laborluft usw Diese Ger t ist geignet f r eine nicht brennbare Atmosph re in der das Probengas zu einer Konzentartion ansteigen kann wo es explosiv werden kann Herstellungsjahr www restek com RESTEK 13 0 Spezifikationen des Geh uses in 14 0 Service i Der Restek Leak Detektor hat eine beschr nkte Garantie von 1 Jahr vom Verkaufsdatum Haben Sie bitte die Leak Detektor Seriennummer zur Hand wenn sie Restek mit Fragen anrufen Es kann eine zus tzliche Geb hr erhoben werden wenn die Garantie abgelaufen ist oder das Ger t misshandelt wurde Einspritzgeformtes fi ABS Plastik FRI 7991 ff grau mit MT 11030 Oberfl chenbeschaffenheit Die Lebenserwartung des Akku ist 2 Jahre vom Verkaufsdatum Kunden m ssen zum Akku Austausch das Ger t an Restek zur ckschicken Gleichzeitig kann dann ein Wartungsdienst geleistet werden F r diesen Wartungsservice wird eine Geb hr erhoben Einspritzgeformtes Versaflex 1040X 1 Pantone 426C Schwarz mit MT 1055
22. ringen Sie die Pr fspitze nun noch mals an das Leck um die genaue Position zu best tigen Der Referenzgasinlet Abbildung 1 darf nicht eingeschr nkt sein sonst funktioniert das Ger t nicht einwandfrei Genauso muss der Gasoutlet offen und nicht behindert sein Der Gasoutlet befindet sich in der Halterung der Pr fspitze VORSICHT Diese Ger t wurde entwickelt um Spuren von Wasserstoff von einem kleinen Leck in einer nicht brennbaren Umgebung nachzuweisen z B in Laborluft usw Diese Ger t ist geignet f r eine nicht brennbare Atmosph re in der das Probengas zu einer Konzentartion ansteigen kann wo es explosiv werden kann 7 0 Spezifikationen Power Werte 12 Volt DC Aufladeger t wird mitgeliefert Garantie 1 Jahr Garantie Akku Werte 6 Stunden Normalbetrieb Zertifikate CE Ex und Japan Betriebstemperaturbereich 0 48 C Befolgungen WEEE RoHS Luftfeuchtigkeitsbereich 0 97 Abbildung 2 Abbildung 3 Deckel Deckel entfernt 8 0 Wartung aufgeschraubt man sieht die kleine Keine Fl ssigkeiten auf das Ger t versch tten es k nnte dann zu St rungen kommen Falls eine und B rste zum Reinigen Fl ssigkeit auf das Ger t versch ttet wird sollten Sie dieses sofort ausschalten Die Fl ssigkeit mit teilweise einem trockenen Handtuch entfernen und das Ger t dann trocknen lassen Staub und andere entfernt R ckst nde k nnen in die Pr fspitze eindringen und sich dort ansammeln Dadurch kann die kleine Leitung in dem Ger t vertso
23. um testen Sie sollten auch die Pr fspitze Referenzgasinlet und Gasoutlet auf Behinderungen untersuchen Abbildung 1 WICHTIG Zu berpr fende Dichtungen m ssen sauber und trocken sein fl ssige Reagenzien zum detektieren von Lecks Staub und andere R ckst nde k nnen den Leak Detektor besch digen wenn sie in die Pr fspitze gezogen werden Der Leak Detektor spircht auf fast alle Gase riechbar und nicht riechbar an L sungsmitteld mpfe Split Vent Abgase oder sogar starke Luftstr mungen um die Pr fspitze oder den Referenzinlet k nnen zu Instabilit t oder falschen Messwerten f hren Beachten Sie dass Sie w hrend der Messung nicht in den Referenzinlet atmen oder mit der Hand den Inlet blockieren As RESTEK www restek com 14 6 0 Lecks Finden Die Pr fspitze langsam and Dichtungen oder anderen potentiellen Leckquellen vorbeif hren Wenn der Leak Detektor auf ein Gas dass sich von Luft unterscheidet anspricht f ngen der LED Lichter zu leuchten an und wenn das letzte Licht erleuchtet h rt man einen Alarm Die roten LED Lichter bedeuten entweder Helium oder Wasserstoff Leaks Die gelben LED Lichter zeigen Stickstoff Argon oder Kohlendioxyd Leaks an Entfernen Sie die Pr fspitze von dem Leak und lassen Sie das Ger t auf Null zur ckkommen Wenn viel Gas in der Pr fspitze ist kann das ein paar Sekunden dauern Bitte versuchen Sie nicht das Ger t w hrend dieser Phase auf Null einzustellen Das kann zu St rungen f hren B
24. ummer des Produktes Warnung Dieses Plastikgeh use hat keinen ausreichenden Oberfl chenwiderstand f r starke elektrische Felder DIESES GER T NICHT IN GEF HRLICHEM UMFELD AUFLADEN Definition der Symbole siehe Tabelle rechts Dieses Ger t entspricht EU EMC Direktive 2004 108 EC andere Standards denen es entspricht sind u A 61326 1997 w A3 Class A Ger t erf llt WEEE Ger t erf llt RoHS Dadurch wird Teil 15 der FCC Regeln erf llt Betrieb ist abh ngig von den folgenden 2 ic Bedingungen 1 Diese Ger t darf zu keinen sch dlichen St rungen f hren und 2 diese Ger t 3G muss empfangene St rungen akzeptieren was auch St rungen beinhaltet die zu unerw nschtem Betrieb f hren k nnen T6 Dieser Klasse A digitale Apparat erf llt Kanadische x ICES 003 Ger te m ssen zum Service an Restek Corporation zur ckgeschickt werden Hersteller Name Addresse und Kontaktinformation 2009 19 Definition der Symbole auf der R ckseite EN60079 0 2006 Elektrischer Apparat f r Atmosph ren mit explosivem Gas Teil 0 Allgemeine Anforderungen EN60079 15 2005 Elektrischer Apparat f r Atmosph ren mit explosivem Gas Teil 15 Konstruktion Test und Demarkierung des Schutztyps n Elektrischer Apparat Gruppe Il ist f r Umgebungen ber Grund Gas Gruppe IIC ist f r Wasserstoff und verwandte Gasarten Kategorie 3 im Verh ltnis zur Gasanalyse normale Sicherheitsvorkehrungen Ausreichende Siche