Bedienungsanleitung CaliBox 200
Contents
1. 22 23 Wir danken Ihnen furs Vertrauen in unsere Produkte und w nschen Ihnen erfolgreiche Messungen und Kalibrierungen mit Ihrer CaliBox 200 und ihren vielseitigen Funktionen Fur allfallige Fragen steht Ihnen die Novasina oder eine ihrer Vertretungen gerne zur Verf gung Ihr Novasina Team Thank you for trusting Novasina and our products We wish you successful cali brations and measurements with your CaliBox 200 and its versatile functions For any question you can contact Novasina or your local Novasina distributor We remain at your complete disposal for any information Your Novasina team MORE THAN 50 YEARS OF Manufacturer Novasina AG Neuheimstrasse 12 C CH 8853 Lachen SZ Switzerland Tel 41 0 55 642 67 67 Internet www novasina com Fax 41 0 55 642 67 70 E Mail info novasina ch Technical data subject to modification Technische nderungen vorbehalten 260 0724 004059 00
2. Calibration with Indicator mode without pressure generator Remark The CaliBox 200 is used as reference measurement instrument with or without accredited certification Starting the calibration 1 Connect the pressure inlets of test sample and CaliBox 200 Static in parallel using a T fitting Pay attention that both inlet positive pressure sides are connected together 2 Connect the pressure outlets of test sample and CaliBox 200 Static in parallel using a T fitting Pay attention that both outlet negative pressure sides are connected together 3 Perform the calibration adjustment Remark During on site calibrations usually the outlet negative pressure side is connected to the reference duct 4 After completing the work switch off the calibra tor pull out the hoses from the nozzles close the nozzles and stow the hoses in the upper shell of the CaliBox 200 8 Configuration with software The CaliBox 200 can be configured and adjusted using the PascalTool WIN software supplied on CD This software runs on Windows 7 Vista XP 2000 and 98 operating systems Brief instruction to software installation 1 Put the CD in the PC disk drive and select the drive directory with the Windows Explorer 2 Proceed with the installation selecting the No vasina_PascalTool WIN exe data file gt finish the SETUP 3 The PascalTool WIN software is now installed Connection of PC Laptop to the CaliBox 200
3. The supplied data cable DB 9 to Stereo plug can be connected to the stereo socket Interface which can be found above the display as Stereo socket E for PascalTool WIN data cable Differential Pressure Operating the PascalTool WIN software Start the PascalTool WIN software as soon as the CaliBox 200 is connected to the PC laptop and all devices are switched on Brief instruction 1 Start the software gt main window appears 2 Activate the COM Ports is USB interface is used The comport can be selected in the menu Options gt Preferences gt COM Port gt OK 3 Read out the data gt Read Data afterwards the following device information appears data of integrated CaliBox 200 sensor Instrument no Pascal STVZ Serial No XXXXXXX dPrange 0 200Pa Pressure XXX XXPa currently measured value Remark In case of error messages please select another COM Port 4 Selection of one of the following registers for the activation of the required function Measurement Sensor calibration by gain adjustment Adjustment of low pass filter for smoothing of the actual value Analog Output Scaling of the analogue output Selection of the voltage output range 0 2 10V Analogue output calibration by gain and offset adjustment Selection of measurement value unit and display contrast LC Display 5 Press Write Data in order to save all device settings Remark The PascalTool
4. WIN software offers also the possibility to protect the settings by a personal password This can be activated by selecting the function Options gt Change protection code The settings can also be printed out This is done by the function File gt Print PascalTool WIN offers also a help program This function can be found in the main menu bar and con tains all relevant information and software functions 9 System messages AZ OFF auto zeroing off AZ ON auto zeroing on ZERO CAL Manual zeroing in progress dP UNDERFLOW Pressure below min level of measurement range lt 5 Pa dP OVERFLOW Pressure above max level of measurement range gt 200 Pa NOVAS SELFCAL Performance of self calibration self testing when device is switched on 10 Maintenance and support Maintenance cleaning The device can be wiped off with a damp cloth Don t use any aggressive cleaning agent Accreditation amp recalibration In case of a later certification of the device by an accredited calibration laboratory please contact you local Novasina distributor He will support you accordingly 2e2 Technical support Please get in touch with your local Novasina distri butor in case of any questions or problems For more information visit www novasina com 11 Accessories There are no accessories required 12 Technical data Type CaliBox 200 Part no 260 0538 Working measuring range O 200 Pa static a
5. cken dieser Taste unterbrochen werden Die Membran des Messelements wird dadurch kurzgeschlossen Die Measure LED geht dabei aus Bemerkung Die Aktivierung dieser Funktion emp fiehlt sich vor der manuellen Nullung im Messmodus Manueller Nullpunktabgleich Bei l ngerem Dr cken dieser Taste wird die manuelle Nullung durchgef hrt werden Dabei spielt es keine Rolle in welchem Modus sich das Ger t befindet Auf dem Display wird kurzzeitig ZERO CAL angezeigt Bemerkung Beim Aus Wiedereinschalten wird der Betriebsmodus auf Werkseinstellung zur ckgesetzt Anschlussstutzen Druckerzeuger Auf der Frontplatte der CaliBox 200 befinden sich 4 Stutzen von denen jeweils 2 f r den Betrieb als Druckerzeuger Kalibrator Calibrator und 2 f r die Nutzung als Differenzdruck Referenz Messger t Indicator gebraucht werden Die Anschl sse f r den Betrieb als Drucker zeuger Kalibrator unterscheiden sich zwischen 5 dynamische Dynamic und statische Static Kalibrierung welche jeweils fur Massenfluss bzw fur Membran Differenzdruck Messgerate benotigt wird a Indicator Indicator Static A Static Calibrator Calibrator Dynamic Static Output Output Anschluss fur dynamische Kalibrierung Dieser Druckanschluss dient der Kalibrierung von Pr flin gen mit Massenfluss Messprinzip d h Ger te die den dynamischen Differenzdruck messen und einen kontinuierlichen Luftfluss ben tigen F r solche I
6. is neither an influence of the hose i d nor the hose length on the generated pressure Analoque signal output Analogue signal output of the actually measu red and displayed measured value Analogue OutputVDC Pressure Value Analogue Output WET Pressure Value 18 Analogue signal output This interface provides the actual measurement signal in 0 2 10VDC of the measurement device integrated in the CaliBox 200 In order to obtain a higher resolution the signal can be scaled and also adjusted These settings can be done using the delivered PascalTool WIN configuration software Remark The analogue signal can be used for the connection to an external device e g data logger etc Battery charger socket Battery charger socket Batt Charger for con nection of the XLR plug of the external battery charger Batt Charger e s Charger socket By this socket the integrated battery is fully re charged within 3 hours EMC notes 1 The device design has been adapted in order to keep all emissions within the EMC norms and regulations 2 The CaliBox 200 should not be placed near any wirings or devices emitting electromagnetic interferences 5 Initial operation Charging the battery The CaliBox 200 is always delivered with a comple tely charged battery and can be used immediately after reception The operational time of the battery if fully charged is min 24 up to max 48 hours de pending on
7. zur Bedienung und zu den Funktionen der PascalTool WIN Software 9 Systemmeldungen AZ OFF Automatische Nullung aus AZ ON Automatische Nullung ein ZERO CAL Manuelle Nullung wird durchgef hrt dP UNDERFLOW Druckerzeugung unter min Messbereich 5 Pa dP OVERFLOW Druckerzeugung ber max Messbereich 200 Pa NOVAS SELFCAL Durchf hrung der Selbst kalibrierung Selbsttest beim Einschalten des Ger ts 10 Unterhalt und Support Unterhalt Reinigung Das Ger t bei Bedarf nur mit einem feuchten Tuch abwischen Auf keinem Fall d rfen aggressive Rei nigungsmittel verwenden werden Akkreditierung amp Rekalibrierung F r eine nachtragliche Zertifizierung des Kalibrators durch ein akkreditiertes Labor setzten Sie sich bitte mit Ihrem lokalen Novasina Vertriebspartner in Ver bindung Dieser wird Sie entsprechen unterst tzen 9 Technischer Support Melden Sie sich bei Fragen oder allfalligen St run gen bei Ihrem lokalen Novasina Vertriebspartner Mehr Infos auf www novasina com 11 Zubehor Es wird kein zusatzliches Zubeh r ben tigt 12 Technische Daten Typ CaliBox 200 Artikel Nr 260 0538 Differenzdruck Arbeitsbereich Messbereich 0 200 Pa statisch und dynamisch Max Aufl sung Anzeige 0 01 Pa Messgenauigkeit bei 20 C typisch lt 0 15 Pa bzw 0 08 FS Full Scale gesamter Messbereich Temperatureffekt Einfluss Konstanttemperatur ty
8. 11 Kalibrierpunkten ber gesamten Arbeitsbereich von 0 bis 200 Pa CD mit PascalTool WIN Geratekonfigurations Software und Bedienungsanleitung CaliBox 200 Gedruckte Bedienungsanleitung CaliBox 200 berpr fen Sie die Vollst ndigkeit des Lieferumfangs Unvollst ndige Lieferungen k nnen von Ihrer Novasina Vertretung umgehend erg nzt werden 3 Sicherheit e ACHTUNG Das Ger t enth lt einen Akku vom Typ LiFe mit hoher Energiekapazitat Laden Sie das Ger t immer unter Aufsicht und nur mit dem mitgelieferten Ladegerat auf e Aufgrund des eingebauten Akkus gelten f r den Transport spezielle Richtlinien Dieses Ger t hat das Werk in sicherheitstech nisch einwandfreiem Zustand verlassen und darf unter keinen Umstanden manipuliert werden Beachten Sie die Hinweise und Warnschilder auf dem Instrument und in der Bedienungsanleitung ACHTUNG Aufgrund des internen Energiespei chers besteht bei Kurzschl ssen Brandgefahr Die CaliBox 200 wurde f r die berpr fung von statischen und dynamischen Differenzdruck Messinstrumenten in Reinluftumgebung ent wickelt und sollte nur f r diesen Zweck und innerhalb der angegebenen Spezifikationen und Betriebsbedingungen eingesetzt werden Bei unsachgem sser Benutzung wird vom Her steller Lieferant f r entstandene Schaden nicht gehaftet Das Risiko tr gt allein der Betreiber Die Handhabung des Kalibrators setzt entspre chend geschultes Fachpersonal voraus Lesen S
9. Ger t ON LED 1 aus gt Ger t OFF LED 2 leuchtet rot gt Laden ON LED 2 leuch tet grun gt Laden OFF bzw Akku geladen Vorgehensweise beim Laden 1 Das Ladegerat Uber IEC Dose am Netz mit 110 230VAC anschliessen gt LED1 leuchtet rot LED 2 leuchtet gr n 2 Ladegerat Uber XLR Stecker mit Calibox ver binden gt LED1 und LED 2 leuchten rot 3 Ladevorgang abwarten bis LED 2 gr n leuchtet Akku vollstandig geladen 4 XLR Stecker aus CaliBox 200 ziehen 5 Netzkabel aus IEC Dose ziehen Bemerkung Die vollst ndige Akkuladezeit bei lee rem Akku betr gt lt 3Std Tipps zur Akkupflege Wegen der Selbstentladung sollte der Akku nur voll geladen eingelagert werden Zum Schutz vor Tiefentladung sollte der Akku j hrlich min einmal voll geladen werden 6 Kalibrierung im Calibrator Modus mit Druckgenerator Anschliessen der Pr flinge Vor dem Anschliessen des Pr flings sollte dessen Messprinzip abgeklart werden Man unterscheidet dynamische mit Massenfluss und statische mit Membran Messverfahren Mit der CaliBox 200 k nnen alle Pr flinge kalibriert werden Je nach Messprinzip werden diese am Druckanschluss Dynamic oder Static angeschlossen Vorgehen bei Massenfluss Ger ten 1 Entfernen des Verschlusspfropfens beim Druck stutzens Dynamic gt L segabel ansetzen und mit einer Kippbewegung nach unten dr cken 2 Schlauchstutzen bis zum Einrasten in Druckstut zen ffnu
10. TASTE Calibrator Indicator m g we O Messung Park E LN e Zero Adjust Taste Calibrator Indicator mit Doppelfunktion f rs Umschalten zwischen Kalibrier Messmo dus kurz dr cken f rs Ein Ausschalten des automatischen Nullpunktabgleichs lang dr k ken Kalibriermodus Calibrator Der im Inneren der CaliBox 200 generierte Druck wird ber die 2 Stutzen Calibrator Dynamic Static nach aussen geleitet und auf dem Display angezeigt die Measure LED leuchtet rot die ndicator LED ist aus Messmodus Indicator Der Differenzdruck wird von aussen auf die 2 Stutzen Indicator Static Static angelegt und auf dem Display angezeigt die Measure sowie Indicator LED leuchten rot Ein Ausschalten der automatischen Nullung Werden beide Tasten gedr ckt gehalten schaltet die automatische Nullpunktjustierung aus Das Display zeigt kurzzeitig AZ OFF an und die Run Error LED blinkt orange Die CaliBox 200 arbeitet jetzt ohne automatische Nullung Bei erneutem l ngerem Dr cken beider Tasten wird die Funktion wieder eingeschaltet und dabei auch gleich eine Nullung durchgef hrt Das Display zeigt kurz AZ ON an und die Run Error LED blinkt wieder gr n Taste Park Zero Adjustment Measure mit Doppelfunktion f rs Umschalten zwischen Normalbetrieb Parken kurz dr cken f r den manuellem Nullpunktabgleich lang dr cken Parkmodus Der Normalbetrieb im Kalibrier oder Messmodus kann jederzeit durch kurzes Dr
11. battery type with energy capacity Do only charge the device under supervision and only with the delivered charger Due to the integrated battery special shipping rules must be considered for the CaliBox 200 This device has left the factory in an impeccable status from the point of view of safety and must not be manipulated under any circumstances Pay attention to the remarks and signs on the instrument and in the operating manual ATTENTION Because of the internal energy storage there is a danger of fire in case of a short circuit The CaliBox 200 has been developed for the inspection of static and dynamic differential pressure measurement instruments in clean air environment and should be only used for such purposes respecting the indicated specifications and operation conditions In case of inappropri ate use the manufacturer supplier declines any liability The risk is only on the operator s side The handling of this calibrator requires skilled personnel So please read this operating manual carefully Where occurrent failures or malfunctions might cause big damages on material or people ad ditional safety precautions have to be taken to assure a safe status in case of failure e g limit switch The CaliBox 200 is not suitable for potentially explosive processes and rooms Only a professional handling of the calibrator as well as a certain measurement technology know how allow a correct interpretatio
12. ckstutzen Static Output ange schlossen werden Dabei sollte wie in Punkt 1 und 2 verfahren werden Bemerkung Die CaliBox 200 kann auch als Referenz ger t zertifiziert werden In diesem Fall entf llt Pkt 5 aya 6 Sicherstellen dass sich die Calibox im Kalibrier modus befindet und dann den gew nschten Druck mit dem Drehregler einstellen 7 Kalibrierung Justierung durchf hren Fur weitere Druckpunkte Punkt 6 7 wiederholen 9 Die urspr ngliche Schlauchlange und ggf den neuen GAIN Wert mit Write data auf den Pr fling schreiben 10 Bei Bedarf weitere Pr flinge kalibrieren 11 Nach Abschluss der Arbeiten den Schlauch vom Pr fling und evtl vom Referenz Messger t abziehen und Kalibrator ausschalten 12 Die Schlauche mithilfe der L segabel von den Druckstutzen rausziehen Dazu Gabel zwischen Schlauchende und Stutzenring einstecken vertikal auf den Ring dr cken und gleichzeitig Schlauch rausziehen 00 13 Verschlusspfropfen bis zum Einrasten in die Stutzen einstecken und Schlauche im Deckel der CaliBox 200 verstauen Vorgehen bei Membran Ger ten 1 Entfernen des Verschlusspfropfens beim Druck stutzens Static Output gt L segabel ansetzen und Hebel mit einer Kippbewegung nach unten drucken 2 Schlauchstutzen bis zum Einrasten in Druckstut zenoffnung einstecken 3 Schlauch eingangsseitig positive Druckseite an Prufling anschliessen Nach Bedarf auch Aus gangsseite negative Druckseit
13. de press Write data to write the data on the test sample more details see chapter 8 Important It is essential to compensate the hose length as otherwise due to the decrease of pres sure the value is falsified The compensated length value on the test sample must be reset to the original value after the calibration adjustment is completed So please do not forget to notify the originally compensated hose length before the calibration 5 Additionally a reference measurement device can be installed in parallel This is connected to the pressure nozzle Static Output Follow according the points 1 and 2 Remark The CaliBox 200 can also be certified by 19 an Official laboratory and used as reference With a certified CaliBox 200 the point 5 is not to be applied 6 Make sure the CaliBox 200 is in the calibration mode and set the required pressure using the rotary knob controller 7 Perform the calibration adjustment 8 For further reference pressure points repeat points 6 7 9 Write the original hose length and if applicable the new GAIN value with the Write data function on the test sample 10 If required calibrate other test samples 11 After conclusion remove the hose from the test sample and if applicable from the reference instrument and switch off the calibrator 12 Pull out the hoses from the pressure nozzles using the removing bracket Fix the bracket between hose end and nozzle ring push ver
14. den Akku Ladebuchse Akku Ladebuchse Batt Charger zum Einstecken des XLR Steckers des externen Ladeger ts Batt Charger e CA Ladebuchse Uber diese Buchse wird der einge baute Akku innerhalb einer Ladezeit von lt 3 Std wieder zu 100 aufgeladen EMV Hinweise 1 Das Geratedesign ist so konzipiert dass alle Emissionen Imissionen im Rahmen der ge setzlichen EMV Normen und Vorschriften erfullt werden 2 St rungsaussendende Leitungen im Bereich des tempor ren Standortes der CaliBox 200 sind zu meiden 5 Inbetriebnahme Laden des Akkus Die CaliBox 200 wird immer mit voll aufgeladenem Akku ausgeliefert und kann sofort in Betrieb ge nommen werden Die Akku Betriebszeit bei voller Ladung betr gt je nach Verwendung min 24 bis max 48 Stunden Der LiFe Akku kann jederzeit wieder voll aufgeladen werden d h er muss nicht zwingend zuerst voll entladen werden Die Ladekapazitat verringert sich dabei nicht Wichtig Beim Aufladen darf ausschliesslich das mitgelieferte Ladeger t verwendet werden Die CaliBox 200 darf w hrend des Ladevorgangs nicht zugedeckt werden und es muss immer f r eine ausreichende Bel ftung gesorgt werden Es gilt zu beachten dass das Ladeger t nach dem Aufladen zuerst vom Ger t und anschliessend vom Netz getrennt wird Lesen Sie vor der ersten Benutzung die beiliegende Ladeger t Bedienungsanleitung sorgf ltig durch Zustandsanzeige Ladegerat e LED 1 leuchtet rot gt
15. e anschliessen Static Pressure dabei ist evtl eine Verbindung mit einem T Stuck zum Referenzdruck Atmo sphare oder zur Referenzleitung n tig Bemerkung Beim Kalibrieren von Membransenso ren muss die Schlauchlange zwischen Kalibrator und Prufling NICHT kompensiert werden 4 Zusatzlich kann ein Referenz Messgerat am Druckstutzen Dynamic Output oder parallel uber ein T Stuck am gleichen Druckstutzen 8 wie der Pr fling Static Output angeschlossen werden Bemerkung Die CaliBox 200 kann auch als Referenz ger t zertifiziert werden In diesem Fall entf llt Pkt 4 5 siehe Massenfluss Messger te 6 siehe Massenfluss Messger te 7 siehe Punkt 8 Massenfluss Messger te 8 Siehe Punkt 10 Massenfluss Messger te 9 Siehe Punkt 11 Massenfluss Messger te 10 Siehe Punkt 12 Massenfluss Messger te 11 Siehe Punkt 13 Massenfluss Messger te 7 Kalibrierung im ndicator Modus ohne Druckgenerator Bemerkung Die Calibox 200 dient dabei als Re ferenz Messger t mit oder ohne akkreditierte Zertifizierung Kalibrierung beginnen 1 Mittels T St ck die Druckeing nge von Pr fling und CaliBox 200 Static parallel verschlauchen Dabei achten dass beide Eing nge positive Druckseite zusammen verbunden werden 2 Mittels T St ck die Druckausg nge von Pr fling und CaliBox 200 Static parallel verschlauchen Dabei achten dass beide Ausg nge negative Druckseite zusammen verbunden werden 3 Kalibri
16. erung Justierung durchf hren Bemerkung Bei der Kalibrierung im Feld wird in der Regel die Ausgangsseite negative Druckseite auf die Referenzleitung gef hrt 4 Nach Abschluss der Arbeiten CaliBox 200 ausschalten die Schl uche von den Stutzen abziehen und im Deckel verstauen 8 Konfiguration mittels Software Die CaliBox 200 kann mit der auf CD mitgelieferten PascalTool WIN Software konfiguriert und justiert werden Diese l uft auf Rechnern mit Betriebssystem Windows 7 Vista XP 2000 und 98 Kurzanleitung Softwareinstallation 1 CD in das Laufwerk des Rechners legen und im Window Explorer Laufwerk auswahlen 2 Durchf hrung der Installation auf Rechner mit der Novasina PascalTool WIN exe Datei gt SETUP abschliessen 3 Die PascalTool WIN wurde erfolgreich installiert Anschliessen des Rechners an die CaliBox 200 Das mitgelieferte Datenkabel DB 9 zu Klinkenste cker wird in die Klinkenbuchse Interface welche sich ber dem Display befindet eingesteckt ae Klinkenbuchse E f r Anschluss PascalTool WIN Differential Pressure Bedienung der PascalTool WIN Software Sobald die CaliBox 200 am Rechner angeschlossen und das Gerat eingeschaltet ist kann die PascalTool WIN Software gestartet werden Kurzanleitung 1 Software starten Hauptfester geht auf 2 Aktivierung des COM Ports falls USB Schnitt stelle benutzt wird Dieses kann im Men Opti ons gt Preferences gt COM Port gt OK akt
17. generator calibrator Calibrator und the other 2 are for the use as differential pressure reference measurement instrument Indicator The nozzles for pressure generator calibrator f Indicator Static Calibrator Dynamic Output operation are distinguished between dyna mic Dynamic and static Static calibration which are used for mass flow respectively for membrane differential pressure instruments a E Indicator Indicator Static i Calibrator Static Output Connection for dynamic calibration This pressure nozzle can be used for the calibration of instruments working with the mass flow mea surement principle which means devices detec ting the dynamic differential pressure requiring a continuous air flow For such instruments it is recommended to use the included hose with i d 8mm as otherwise the loss of pressure will be too big Remark Due to the mass flow principle and the loss of pressure inside the hose the max hose length is limited to 30 m based on a hose with i d 8 mm On smaller diameters the max length is reduced dramatically Connection for static calibration This pressure nozzle is used for the calibration of instruments working with the membrane measurement princip le which means devices detecting the static dif ferential pressure In this case itis recommended to use the included hoses with i d 5 mm or 4 mm Remark In the case of static pressure there
18. he outside to the 2 nozzles Indicator Static Static and is shown on the dis play the Measure as well the Indicator LED flash red Switching on off the auto zeroing If both but tons are kept pushed the auto zeroing is switched off The display shows briefly AZ OFF and the Run Error LED blinks orange Now the CaliBox 200 works without auto zeroing function If both buttons are kept pushed again the function is re activated and a zeroing is performed immediately The display shows briefly AZ ON and the Run Error LED blinks green Button Park Zero Adjustment Measure with double function for switching between standard operation park mode push briefly for manual zeroing push long Park mode The standard operation in the calibrati on or measurement mode can be interrupted at any time by pushing briefly this button In this mode the membrane of the sensor is bypassed The Measure LED is switched off Remark The activation of this function is recommen ded prior to the manual zeroing in the measurement mode Manual zeroing If this button is pushed for a longer time the manual zeroing is performed independent from the actually selected mode The display shows briefly ZERO CAL Remark When the device is switched off and on again the operation mode is reset to factory settings Connection nozzles pressure generator There are 4 nozzles on the CaliBox 200 front panel of which 2 are for the operation as pressure
19. ie diese Bedienungsanleitung deshalb sorgfal tig durch Dort wo auftretende Fehler oder Fehlfunktionen grosse Material oder Personenschaden verur sachen k nnen sind zus tzliche Schutz und Sicherheitsvorkehrungen zu treffen so dass im Fehlerfall ein sicherer Betriebszustand gew hr leistet wird z B Grenzwertschalter Die CaliBox 200 ist nicht f r den Einsatz in ex plosionsgef hrdeten Raumen geeignet Nurderfachgerechte Umgang mit dem Kalibrator sowie entsprechende Messtechnikkenntnisse erlauben die richtigen R ckschl sse auf die effektive Messgenauigkeiten und Toleranzen e Der Kalibrator sollte periodisch berpr ft werden Dies kann intern oder durch unabhangige akkre ditierte Messstellen SCS DKD usw erfolgen Verwenden Sie ausschliesslich Originalzubeh r und Originalersatzteile von Novasina Das Offnen oder Umbauen des Kalibrators ist zu unterlassen und zieht die sofortige Aufhebung der Garantieleistung mit sich Dies gilt auch f r den Austausch des Akkus 4 Geratebeschrieb Hauptschalter Druckregler Hauptschalter Power On Off zum Ein und Ausschalten des Ger ts Drehregler mit Arretierung Pressure Adjust ment zur Einstellung des generierten Differenz druckes Power On OFF Druckeinstellung Pressure Adjustment Drucktasten f r Betriebmodus Auswahl Die CaliBox 200 verf gt ber 2 Tasten mit de nen s mtliche Betriebsmodi eingestellt werden k nnen
20. ion with Calibrator mode with pressure generator Connection of test sample Before connecting the test sample its measurement principle should be verified There are 2 principles the dynamic mass flow and static membrane measurement method With the CaliBox 200 all test samples can be calibrated Depending on the principle they are connected to the Dynamic or Static nozzle Procedure with mass flow devices 1 Remove the plug of the Dynamic Output pressure nozzle gt fix the plug removing bracket and press downwards by a tilting movement 2 Insert the hose in the pressure nozzle until it clicks into place 3 Connect the hose to the test sample inlet side positive pressure and use the reduction nozzle i d 8 5mm if required Leave the test sample outlet side negative pressure open 4 Electronic compensation of the complete hose length This consists of the hose length between calibrator and test sample 200 cm hose length inside the CaliBox 200 24 cm an eventual length of hoses connected to the outlet side This total length is entered for compensation into the PascalTool WIN software and stored in the test sample For this purpose the test sample must be connected to a PC laptop via the delivered data cable and the software must be started Afterwards press Read data the connection to the test sample is established Subsequently select the register Environment and enter the hose length To conclu
21. iviert werden 3 Daten auslesen Read Data danach werden folgende Ger tedaten des in der CaliBox 200 eingebauten Differenzdruck Messger ts ange zeigt Instrument no Pascal STVZ Serial No XXXXXXX dPrange 0 200Pa Pressure XXX XXPa akt Messwert Bemerkung Falls eine Fehlermeldung erscheint sollte ein anderer COM Port ausgew hlt werden 4 Auswahl eines der folgenden Register f r die Ausf hrung der gew nschten Funktion Measurement Sensorjustierung mittels Einstellung des Gain Einstellung des Tiefpass Filters f r Gl ttung des aktuellen Messwertes Analog Output Skalierung des analogen Ausgangs Auswahl des Spannungs bereichs 0 2 10V Justierung des analogen Ausgangs mittels Einstel lung von Gain und Offset Auswahl der Messeinheit und Kontrast auf Anzeige LC Display 5 Nach erfolgter Konfiguration bzw Einstellung jeweils auf Write Data dr cken um die Daten auf das Ger t zu speichern Bemerkung Die PascalTool WIN Software bietet auch die M glichkeit die eingestellten Parameter zu mit einem Passwort zu sch tzen Dieses kann in der Men leiste unter Options gt Change protection code aktiviert werden Die Ger teeinstellungen k nnen auch ausgedruckt werden Dies wird in der Men leiste unter File mit der Funktion Print durchgef hrt PascalTool WIN bietet zudem auch ein Hilfe pro gramm Die Help Funktion in der Men leiste enth lt alle wichtigen Informationen
22. measurement inst rument which can be even used as an accredited reference device Specific characteristics High versatility for static and dynamic differenti al pressure transducers High accuracy and stability thanks to the auto zeroing function Usable as calibrator as well as measurement instrument and reference with certificate Battery powered with extended running time Easy calibration and adjustment with Pascal Tool WIN software Integrated scalable analogue outputs 2 Scope of delivery 16 The scope of delivery of the CaliBox 200 contains Mobile differential pressure calibrator with ro bust plastic case and integrated LiFe battery External LiFe charger with 230 V plug RS 232 configuration cable e 2m hose inner diam 8mm with connection nozzle 2m hose inner diam 5mm with connection nozzle e 2m hose inner diam 4mm with connection nozzle e 2x 2m hose with inner diameter 4mm 2x plug removing bracket e 4x nozzles for hose reduction i d 8 5mm Factory calibration certificate with 11 calibrati on points over the working span of 0 to 200 Pa CD containing the PascalTool WIN configurati on software and operating manual for CaliBox 200 Printed operating manual for CaliBox 200 Please check the completeness of the delivery Incom plete deliveries can be completed by your local Novasina distributor 3 Safety e ATTENTION The device contains a LiFe
23. n of the real accuracy and tolerances The calibrator should be inspected and calibrated periodically This can be performed internally or by an accredited laboratory e g SCS DKD NIST COFRAC etc Use only original Novasina accessories and spare parts The opening and altering of the calibrator is strictly forbidden and revokes immediately the warranty This means also the battery replace ment 4 Device description Main switch pressure controller Main switch Power On Off to switch on and off the device Rotary knob controller with lock Pressure Adju stment for pressure adjustment Power On Off Power On OFF _ 28 Pressure adjust e ment D Pressure Adjustment Push button for operation mode selection The CaliBox 200 has 2 buttons with which all available operation modes can be selected BUTTON Measure baca Calibrator Indicator BUTTON Gi a Measure Park AA Cantor Zero Adjust Button Calibrator Indicator with double function for switching between calibration measurement mode push briefly for switching on off the auto zeroing push long Calibration mode Calibrator The pressure ge nerated inside the CaliBox 200 is carried outwards by the 2 nozzles Calibrator Dynamic Static and is shown on the display the Measure LED flashes red the ndicator LED is switched off Measurement mode Indicator The differential pressure is carried from t
24. nd dynamic 200 Pa Max display resolution 0 01 Pa Measurement accuracy at 20 C typical lt 0 15 Pa respectively 0 08 FS Full Scale Temperature effect influence temperature constant typical lt 0 03 Pa C resp 0 015 FS C influence temperature change 10 C h typical lt 0 6 Pa resp 0 3 FS Offset drift typical lt 0 15 Pa year automatic zero point adjustment Hysteresis typical lt 0 15 Pa Reference device sensor element piezo resistive membrane differential pressure sensor Max permissible overpressure 100 500 kPa Operating temperature 5 50 C Environmental media non condensing non corrosive fresh air Power supply Chargeable battery 25 6V 10 Ah LiFe PO4 with BMS Charging time lt 3 hours Operation time at full battery load gt 24 48 hours Analogue output 0 2 10V load gt 10 kOhm Accuracy of analogue output typical 0 05 of full scale FS Digital interface RS 232 Carrying case dimensions 360 x 304 x 194 mm Protection class open IP 30 Protection class closed IP 54 Total weight 7 3 kg Storage temperature range 10 60 C non condensing 13 Spare parts Description Article no Reduction nozzle for hose Inner diameter 8 5mm 111 6332 14 Factory settings Display 0 200 Pa AOUT 0 10V Password deactivated Factory calibration calibrated standard with certificate
25. ng einstecken 3 Schlauch eingangsseitig positive Druckseite an Pr fling anschliessen Gegebenenfalls mit gelieferten Gummi Reduktionsstutzen ID 8 5mm am Schlauchende einschieben Ausgangsseitig negative Druckseite idealerweise nichts an den Pr fling anschliessen 4 Elektronische Kompensation der gesamten Schlauchlange Diese besteht aus der Lange zwischen Kalibrator und Pr fling 200 cm L n ge des im Innern der CaliBox 200 eingebauten Schlauchs 24 cm evtl Lange des ausgangsei tig angeschlossenen Schlauchs Diese Lange wird mittels der PascalTool WIN Software auf den Pr fling geschrieben und somit bei der Mes sung kompensiert Dazu muss der Pr fling ber das mitgelieferte Datenkabel an einen Rechner angeschlossen und Software gestartet werden Danach Read data dr cken die Verbindung zum Pr fling wird erstellt Anschliessend die Registerkarte Environment auswahlen und die Schlauchlange eingeben Abschliessend Write data dr cken um die Daten auf den Pr fling zu schreiben mehr Details siehe Kap 8 Wichtig Die Schlauchl nge muss zwingend kompensiert werden da sonst der Druckabfall im Schlauch die Messung verfalscht Die kompensierte L nge muss nach erfolgter Kalibrierung Justierung unbedingt wieder auf den urspr nglichen Wert zur ckgesetzt werden Deshalb unbedingt die urspr nglich kompensierte Schlauchl nge for der Kalibrierung notieren 5 Zusatzlich kann ein Referenz Messgerat parallel an den Dru
26. novasina The Art of Precision Measurement Bedienungsanleitung CaliBox 200 Diffenzdruck Kalibrator Messgerat zur Kalibrierung und Pr fung von statischen und dynamischen Luftdifferenzdruck Messinstrumenten Messbereich 0 bis 200 Pa www novasina com by Novasina AG 2011 Technical data subject to modification Printed in Switzerland Technische Anderungen vorbehalten Inhaltsverzeichnis eem ge MT PS 4 IECIT ee 4 SEENEN 4 SETI DCS CIC BMC 5 A A 6 K librierung im Calibrator MOQUS seite ti eene een 7 Kalibrierung im A GICAIOF MODUS RNC Be 8 Konfiguration mittels Software eese isis esee nnne cr 8 SYSIENINEIOUNGEN MMC 9 Unterhalt Und Support MP OE E E O E R ETR 9 A OE 10 egg ceo PR ERE rcp 10 Sera AMA QA el A ee Eee 10 WES EMISIONES 10 1 Produkt Beschreibung Vielen Dank dass Sie sich fur dieses Novasina Qualitatsprodukt entschieden haben Wir danken fur Ihr Vertrauen Der mobile Differenzdruck Kalibrator CaliBox 200 von Novasina ist ein universelles Kalibrier und Messger t f r die berpr fung von statischen Membran sowie dynamischen Massenfluss Luft Differenzdruck Messgeraten mit Messbereich von 0 bis 200 Pa Das Ger t kann in Reinr umen eingesetzt werden und wird mittels Akku betrieben Eine externe Spei
27. nstrumente empfiehlt Novasina den mitgelieferten Schlauch mit ID 8 mm zu verwenden da sonst der Druckabfall im Schlauch zu gross wird Bemerkung Bei der Kalibrierung von Massen fluss Messger ten ist infolge des Druckabfalls im Schlauch dessen max L nge eingeschr nkt Basierend auf einen ID 8mm ergibt sich eine ma ximale L nge von 30 m Bei kleineren Schlauch ID verringert sich diese Maximall nge erheblich Anschluss f r statische Kalibrierung Dieser Druckanschluss dient der Kalibrierung von Pr flin gen mit Membran Messprinzip d h Ger te die den statischen Differenzdruck messen Hier verwendet man mit Vorteil einer der mitgelieferten Schlauche mit ID 5 mm oder 4 mm Bemerkung Beim statischen Druck hat weder der Schlauch Innendurchmesser noch die L nge einen Einfluss auf den generierten Druck Analoger Signalausgang Analoger Signalausgang des aktuellen gemes senen und angezeigten Messwerts Analogue OutputVDC Pressure Value Analogue Output WK Pressure Value Analoger Signalaussgang Uber diese Schnitt stelle wird das aktuelle Messsignal des in der Ca liBox 200 integrierten Messger ts in 0 2 10VDC ausgegeben Zur Erh hung der Aufl sung kann es entsprechend skaliert und zudem justiert werden Diese Einstellungen k nnen ber die mitgelieferte PascalTool WIN Konfigurationssoftware durchge f hrt werden Bemerkung Das analoge Signal kann z B f r den Anschluss eines Datenschreibers genutzt wer
28. pisch lt 0 03 Pa C bzw 0 015 FS C Einfluss Temperaturanderung 10 C h typisch lt 0 6 Pa bzw 0 3 FS Offset drift typisch lt 0 15 Pa Jahr automatischer Nullpunkt Abgleich Hysterese typisch lt 0 15 Pa Messelement im Messger t Piezoresistiver Membran Differenzdruck Sensor Max zul ssiger berdruck 100 500 kPa Betriebstemperatur 5 50 C Umgebungsmedium nicht kondensierende nicht korrosive Luft Betriebsspannung Akku 25 6V 10 Ah LiFe PO4 mit BMS Akkuladezeit lt 3 Stunden Akkubetriebszeit bei voller Aufladung gt 24 48 Stunden Analoge Ausgange 0 2 10V Last gt 10 kOhm Genauigkeit der analogen Ausgange typisch 0 05 vom gesamten Messbereich FS Digitale Schnittstelle RS 232 Abmessungen Gehause 360 x 304 x 194 mm Schutzklasse ge ffnet IP 30 Schutzklasse geschlossen IP 54 Gesamtgewicht 7 3 kg Lager Temperaturbereich 10 60 C nicht kondensierend 13 Ersatzteile Bezeichnung Artikel Nr Stutzen f r Schlauch ID Reduktion 8 5mm 111 6332 14 Werkseinstellungen Anzeige 0 200 Pa AOUT 0 10V Passwort Deaktiviert Werksjustierung Vorkalibriert mit Zertifikat 10 11 12 novasina The Art of Precision Measurement Operating manual CaliBox 200 Differential pressure calibrator measurement device for calibration and inspection of static and dynamic Air differential pressure measurement instruments Measuremen
29. sung entfallt somit und eine gr sstm gliche Mobilitat ist gewahrleistet Die CaliBox 200 ist ein hochpraziser Druckgenerator sowie Messger t und erzeugt einen genauen und kontinuierlichen Luftstrom welcher sich ber einen Drehregler einstellen l sst Dank automatischem Nullpunktabgleich erreicht das Gerat eine sehr hohe Reproduzierbarkeit und Langzeitstabilitat Nebst Differenzdruck Erzeuger ist dieser Kalibrator auch ein vollwertiges Messinstrument das sogar als akkreditiertes Referenzmessger t eingesetzt werden kann Besondere Eigenschaften Universell einsetzbar f r statische und dynami sche Differenzdruck Messumformer Hohe Genauigkeit und Stabilitat dank automati schem Nullpunktabgleich Als Kalibrator sowie als Messgerat einsetzbar auch zertifizierbar Akkubetrieben mit langer Betriebszeit Einfache Justage ber PascalTool Win Softwa re Integrierte skalierbare analoge Ausgange 2 Lieferumfang Der Lieferumfang der CaliBox 200 beinhaltet Mobiler Differenzdruck Kalibrator im robusten Kunststoffkoffer mit integriertem LiFe Akku e Externes LiFe Ladeger t mit 230 V Stecker RS 232 Konfigurationskabel e 2m Schlauch ID 8mm mit Anschlussstutzen 2m Schlauch ID 5mm mit Anschlussstutzen e 2m Schlauch ID 4mm mit Anschlussstutzen e 2x 2m Schlauch ID 4mm e 2x L segabeln f r Druckstutzenpfropfen 4x Stutzen f r Schlauch ID Reduktion 8 5mm Werkskalibrierzertifikat mit
30. t range 0 to 200 Pa 13 14 Table of contents nn O O E EU OA 16 nao do do E 16 rj A 16 EE o NH 17 ERT EE 18 CAMISA with calibrator MODE Drm 19 Calibration With Indicator Mode nn ee 20 Brei eme 20 E A ee eee 27 Maintenance and SUDDOM EE 21 PICS MEET 22 Technical data ausser 22 Spare DIAS serie kenis einir EnTra E AEE E AE E E laa 22 AAPP 0 22 15 1 Product description Thank you for choosing this Novasina quality pro duct We thank you for your trust The mobile differential pressure calibrator CaliBox 200 from Novasina is an universal calibration and measurement device for the inspection of static membrane as well as dynamic mass flow air dif ferential pressure measurement instruments within the range of 0 to 200 Pa This device can be used in clean rooms and labora tories on site and is powered by a battery An external power Supply is not necessary and thus a maximum mobility and versatility is granted The CaliBox 200 is a highly accurate pressure generator and also a measurement device which generates an accurate and stable continuous air flow which is controlled by a rotary knob Thanks to its auto zeroing function it achieves a very high reproducibility and long term stability In addition to a differential pressure generator this calibrator is also a professional
31. the use The LiFe battery can be re charged at any time and has not to be completely empty before the re charge The battery capacity is not reduced by any re charging process Important Use only the supplied external charger for the the battery re charge The CaliBox 200 must not be covered or closed during the charging process and a sufficient aeration must be assured After the charging process it should be paid attention to first remove the charger from the device and then form the mains Read the charger s manual before the first use Status indicator charger LED 1 flashes red gt device ON LED 1 off gt device OFF e LED 2 flashes red gt charging ON LED 2 flas hes green gt charging OFF battery charged Charging procedure 1 Connect the charger with the IEC socket to the mains 110 230VAC gt LED1 flashes red LED 2 flashes green 2 Connect the charger with the CaliBox 200 via the XLR plug gt LED1 and LED 2 flash red 3 Wait for the charging process until the LED 2 flashes green battery completely charged 4 Remove XLR plug from the CaliBox 200 5 Remove power cable from the IEC socket Remark The charging time of an empty battery is lt 3 hours Battery maintenance Due to the self discharge the battery should be stored only completely charged In order to protect the battery from a total dischar ge it should be charged completely min once a year 6 Calibrat
32. tically on the ring and at the same time pull out the hose 13 Insert the pressure nozzle plug until it clicks into place stow the hoses inside the upper shell of the CaliBox 200 Procedure with membrane devices 1 Remove the plug of the Static Output pressure nozzle gt fix the plug removing bracket and press downwards by a tilting movement 2 Insert the hose in the pressure nozzle until it clicks into place 3 Connect the hose to the test sample inlet side positive pressure If required connect also the outlet negative pressure side Static Pressure thereby it is maybe necessary to connect via a T fitting to the reference pressure atmosphere or reference duct Remark For the calibration of membrane devices it is NOT necessary to compensate the hose length 4 Additionally a reference measurement device can be installed in parallel and connected to the pressure nozzle Dynamic Output or in parallel gt to the same nozzle as the test sample Static Output using a T fitting Remark The CaliBox 200 can also be certified by an official laboratory and used as reference With a certified CaliBox 200 the point 4 is not to be applied 5 see mass flow instruments 6 see mass flow instruments 7 see point 8 of mass flow instruments 8 see point 10 of mass flow instruments 9 see point 11 of mass flow instruments 10 see point 12 of mass flow instruments 11 see point 13 of mass flow instruments 7