        95-113969 LA EN ES PT R1 STATIM 2000 5000 Operators Manual
         Contents
1.      3 2 Unit Placement    When placing the unit on a counter top  ensure the following     e The level indicator bubble  1  on the front panel should be  balanced in the front right quadrant of the target  Four o clock  position   This will ensure that the unit drains properly  Adjusting  the three leveler feet will help you move the bubble if necessary    e The unit should be stable and all four feet should be securely  in contact with the counter surface  This will prevent the unit  from moving freely     Installation 5                         Figure 3    STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     3 3 Connecting the Waste Bottle    The waste bottle  2  is used to collect the wastewater after it has been  converted to steam and then drained from the cassette  To connect the  waste bottle to the STATIM  follow these steps  see Figure 4      and connect tightly    2  Cut the tube to length and slide the waste bottle fitting  5  into place    3  Place the free end of the tube into the hole in the lid of the waste bottle  and hand tighten the fitting  Do not coil the exhaust tube    4  Unscrew the lid and copper condenser coil  6  assembly from the  waste bottle  The lid and coil should come out together    5  Fill the waste bottle with water to the MIN line and replace the lid and  copper condenser assembly  Empty the waste bottle often to avoid  unpleasant odors and discoloration of the contents   A low level  disinfectant  prepared according to the
2.     Message  MSD FLASH FULL  lt  gt   REPLACE MSD    The MSD is full  Export the data        Message  Missing lines of  Data on MSD FLASH    Refer to the STATIM screen to confirm successful sterilization     Reset the Data Logger by unplugging its power  disconnecting the MSD and waiting 10 seconds  Then re   connect the power adapter and insert the MSD into the Data Logger     If the problem persists  contact the SciCan Service Center        Message  The file or directory in the MSD is  corrupted or unreadable           Refer to the STATIM screen to confirm successful sterilization     The MSD may have been unplugged while data was being written to it  The MSD should not be unplugged  until after    SAFELY REMOVE MSD  lt  gt  MSD FLASH DETECTED     is displayed     The corrupted files or directories may be lost  Reformat the MSD on your computer        SciCan Ltd  Troubleshooting    28 STATIM 2000 5000 Cassette Autoclave Operator Manual       9  Spare Parts List                                                                                                                                                                                              01 100780S   Bumper 01 101766S Power Cord UK  2000  amp  5000    01 100782S   Push in Fitting  2000  01 101767S Power Cord Australia  5000    01 101658S   Cassette Handles  6000  01 101768S Power Cord Switzerland  2000  amp  5000   01 101755S   Push in Fitting  6000  01 101769S Power Cord Italy  2000  amp  5000    01 106080S   Cass
3.   C  Pressurizing F  Air Drying                COYCLE YOU SELECTEDI WARMING LIF    ECVELE VO SELEC TERS  sm E    CONDITIONING    Pee E Wau SeLERT etd    Leash E    DONELE VOU sia ses fala      CO a o ae Ji    VENTING CC  CLE YOU SELECTEDI  inet    iS ea SE    Vania       ng qn gos    pere E Gol SSeS  AIF Dt ias SEE BIE    EQS Aurai EEE GMP   Fines Silt To SSSI REMGUE CASSETTE       As the cycle progresses  the cycle stage and temperature of the chamber are displayed    During sterilization  the pressure of the chamber and the time remaining for that stage is also displayed   During the air drying phase  the time remaining is displayed    Message will vary depending on the cycle choosen     Instructions for Use 15 STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     As the cycle is running  an intermittent buzzing sound will be apparent as the pump injects water into the steam  generator  A random clicking sound will also be heard as the exhaust valve opens and closes  The buzzing noise  during the air drying stage is the compressor operating     If a sterilization cycle is successful  the reminder tone sounds and the amber light flashes until the STOP button is  pressed  or the cassette is removed from the unit and the display will appear similar to       gt   _       Be careful  The metal parts will be hot  and the cassette will contain hot steam  After the cassette is removed  from the unit  it should be opened to hasten cooling of instruments     If a mess
4.  103139S Conductivity Meter  2000  amp  5000    01 101647S   Power Cord North America   6000  01 103923 Condenser Additional Bottle    01 101649S   Cassette Seal  5000  01 103935 STAT DRI Plates  5000     01 101652S   Air Compressor Filter  5000  01 103945S Rack Tray Unwrapped Instrument Kit  2000     01 101657S   Thermal Paper  Box of 10 rolls  01 104104 Extended Length Cassette  5000     01 106071S   Extended Cassette Lid Handle  6000  01 104499 Extended Cassette Instrument Rack  5000     01 101614S   Tray Complete with Box  5000  01 104785S Instrument Holder   10 mm  5000     01 106438S   Cassette Lid  Extended 01 108263S Basket Medical  5000     01 106653   Mesh Tray  2000  01 210000 Kit Printer  5000     01 107767S   Extended Tray  5000  2OZPLUS STAT DRI Plus 2 oz  Bottle w  sprayer    01 101709S   Mesh Rack  5000  320ZPLUS STAT DRI Plus 32 oz  Bottle w  sprayer    01 101757S   Cassette Lid w  Box  5000  80ZPLUS STAT DRI Plus 8 oz  Bottle w  cap    01 1017638   Power Cord Japan  2000  amp  5000  99 108332 Chemical Emulator 134  C  3 5 min          SciCan Ltd     Spare Parts    29       STATIM 2000 5000 Cassette Autoclave Operator Manual    10  Warranty    SciCan Ltd     Limited Warranty    For a period of one year  SciCan guarantees that the STATIM 2000 5000  when manufactured by SciCan in new  and unused condition  will not fail during normal service due to defects in material and workmanship that are not  due to apparent abuse  misuse  or accident     The one year warr
5.  2000 5000 Cassette Autoclave Operator Manual    3  Installation    SciCan Ltd     3 1 Environmental Considerations    There are several factors that may affect the performance of your  STATIM  Please review these factors  and select a suitable location in  which to install the unit     Temperature and Humidity    Avoid installing your STAT M in direct sunlight or close to a heat source   e g  vents or radiators   The recommended operating temperatures are  15 25  C  69  F to 77  F  with humidity of 25 70      Spacing    The vents and openings on the STATIM should remain uncovered and  unobstructed  Leave a minimum of 50 mm 2    between the top  sides and  back of the unit and any wall or partition     Venting    The STAT M should be operated in a clean  dust free environment   Work Surface    The STATIM should be placed on a flat  level  water resistant surface  Never  install and operate the unit on a sloped surface     Electromagnetic Environment    The STAT M has been tested and meets applicable standards for  electromagnetic emissions  While the unit does not emit any radiation    it may itself be affected by other equipment which does  We recommend  that the unit be kept away from potential sources of interference     Electrical Requirements    Use properly grounded and fused power sources with the same voltage  rating as indicated on the label at the back of your STATIM  Avoid multiple  outlet receptacles  If using a surge suppressor power bar  plug in one  STATIM only
6.  STATIM 5000     Using the Cassette                           10  4 3 STATIM 5000     STAT DRI Plates                               11  4 4 Preparing and Loading Instruments                            11  4 5 Using Biological Chemical Indicators                             13  4 6 Instrument Weight Guide               ceseceeesseeeseceeeseeeesseeeees 13  47 CYCIO PrOtileS  ssscciccscucssstavavesncee se ptas sactiawrsecvancnesvarecerarsntease 13  4 8 Running A CY CIOS siaaa aointean araia 15  4 0 Stopping a Cy ClO ss sisices cesicesscenasasiceved cnaissseancesatsasessaiseacaig 16   D Maintena  ncCe     2  0 00525550 50855050005020555oasassiascenasssinsonasisis 17  5 1 Cleaning the Cassatt aissis 17  5 2 Cleaning the Water Reservoir Filter eeen 17  5 3 Cleaming the ResSarVOir   sssrinds igaren 17  5 4 Cleaning the Exterior Surfaces           c  cccccsscceeseseeeeerseees 17  5 5 Changing the STATIM 2000 Air Filter                              18  5 6 Changing the Bacteria Retentive Air Filters                      18  5 7 Replacing the Cassette Seal             cccsscessceeesereeesseeeees 19  5 8 Maintaining Fluid Levels                   se  19  5 9 Reading Water Quality  iacsccsss casi canesavevs ceteessiveestued sere n  s 19  5 10 Preventative Maintenance Schedule                              20   6  The Communication Port       ccccscccesseeeeeeceeseeeeeeeeeeeeees 21  6 1 The RS232 Communication Portistas 21  6 2 Installing the SciCan Data Logger    22  6 3 Cycle Printout Ov
7.  STATIM 5000 is designed to hold a maximum of 12 autoclave bags  Care  must be taken to ensure that the combined weight of the loaded bags does not exceed the maximum load  capacity as stated in Appendix  Cycle Profile and Capacity     The Unwrapped instrument rack fitted with a maximum of 10 STAT DRI plates will hold 10 autoclave bags   Rubber and Plastic Instruments    The following materials can be sterilized in the STATIM     Nylon   Polycarbonate  Lexan      Polypropylene  PTFE  Teflon      Acetal  Delrin       Polysulfone  Udel      Polyetherimide  Ultem      Silicone rubber   Polyester       When loading rubber and plastic instruments in the tray  leave a space between the instruments and the cassette  walls  This ensures that steam reaches all surfaces  and will promote drying     The following materials cannot be sterilized in the STAT M     e Polyethylene   e ABS   e Styrene   e Cellulosics   e PVC   e Acrylic  Plexiglas      e PPO  Noryl       e Latex   e Neoprene    Use of these and similar materials may lead to instrument or equipment damage  If you are unsure of your  instrument s material or construction  do not load into your STATIM until you have checked with the instrument  manufacturer     All Instruments  The STATIM is not intended for sterilizing textiles  liquids or biomedical waste     Instruments will remain sterile after a successful cycle until the cassette is disengaged from the unit  Unwrapped  instruments  once exposed to ambient or external condi
8.  The front leveler feet  should be approximately 12 mm 0 5    from the edge    2  Lift the front left corner of the unit upward and remove the drain tube   3  from the clip located on the underside of the unit    3  Pull the drain tube outward so the free end can be positioned over a   water container    Fill the reservoir with steam process distilled water    5  Remove the plug  4  from the end of the drain tube  3  and allow water  to drain from the tube into a container for 30 seconds  When the water  flows in a steady stream  replace the stopper    6  Lift the front left corner of the unit upward and reinsert the tube into  the clip on the underside of the unit  Push the excess length of tubing  back into the space provided     A    Make sure the plug in the drain tube is secured     After installation  and before sterilizing any instruments  run two Wrapped cycles   For further instructions  see Section 3 10 Preparing Unit for Use     Installation 6                       N ne   STATIM 5000                                                 i    STAT M 2000    Figure 4       Figure 5                               Figure 6    STATIM 2000 5000 Cassette Autoclave Operator Manual    3 6 Setting the Time and Date    To set the time and date  follow these steps and watch the blinking cursor  on the LCD     1  Power the STATIM OFF   2  Press and hold the UNWRAPPED cycle button     3  While pressing the UNWRAPPED cycle button   power the STAT M ON  The LCD display reads     4  Use t
9.  and medical instruments designed to withstand steam sterilization   The STATIM is not designed to sterilize liquids  cloth loads  biomedical waste or materials not compatible with steam  sterilization  The processing of such loads may result in incomplete sterilization and or damage to the autoclave     For more information about instrument suitability for steam sterilization  consult the manufacturers    reprocessing  instructions     1 1 Disclaimers    Use only steam process distilled water in your STAT M  Deionized  demineralized  or specially filtered water should  not be used  Never use tap water     Do not permit any person other than certified personnel to supply parts for  service or maintain your STATIM    SciCan shall not be liable for incidental  special or consequential damages caused by any maintenance or services   performed on the STATIM by a third party  or for the use of equipment or parts manufactured by a third party    including lost profits  any commercial loss  economic loss  or loss arising from personal injury  Never remove the   cover of the unit and never insert objects through holes or openings in the cabinetry  Doing so may damage the unit  A and or pose a hazard to the operator     All elements of this book are common to both STATIM 2000 and STATIM 5000 except when noted   IMPORTANT  Follow local guidelines governing verification of the sterilization procedure   Drying Performance    The STATIM 2000 and 5000 have been designed to provide a comple
10.  for dirt and moisture  Replace if dirty  Call for service if wet        Water Filter    Check the water reservoir filter every week and clean  if necessary  Replace only if necessary        Cassette Seal    Replace every 500 cycles or six months  whichever is first   or whenever necessary        Every 6  Months    Biological and or Air Filter             Replace every 500 cycles or six months  whichever is first                                    Cassette e Check the tray  lid and seal for damage  Replace if necessary   Biological Filter e Inspect the biological filter for moisture   Solenoid Valve e Inspect the valve and clean if dirty  Replace the plunger if defective   E Pump e Clean the filters  replace if dirty   oO  z Check Valve e Remove the exhaust tube from the back of the unit during the air drying phase   o Check for air coming from the fitting   5 e Remove the air compressor tube from the check valve inlet while running a cycle   Make sure no steam is leaking from the valve  Replace if there are any leaks   Water Reservoir e Check the reservoir for dirt  Clean and rinse with  steam process distilled water if necessary   Calibration e Calibrate the unit   SciCan Ltd  Maintenance 20 STATIM 2000 5000 Cassette Autoclave Operator Manual    6  The Communication Port    SciCan Ltd      Not on all models        6 1 The RS232 Communication Port    Printer          Some models without an internal printer may be equipped with an 9 pin  RS232 communication port that will all
11.  manufacturer s instructions   may be added to the waste bottle to remedy this situation   As a  minimum  empty the waste bottle each time you refill the reservoir       1  Insert the exhaust  3  tube into the fitting  4  on the back of the unit    The tube can be routed through a hole   8 mm 0 3    in diameter  in the  counter top and secured with the provided nylon clamps     3 4 Filling the STATIM Reservoir     11  6  Place the waste bottle near the unit  Store the bottle below the unit     When filling the reservoir  ensure you only use steam processed  distilled water containing less than 5 ppm total dissolved solids  having  conductivity of less than 10 uS cm   The impurities and additives in other  water sources will cause an error reading on the LCD  If you have a water  conductivity meter  available from SciCan  order number 01 1031398   check each new water container before filling the reservoir  If your unit is  equipped with a Rev 7 PCB or higher  software SxxxR6xx   you can read  the water conductivity by refering to Section 5 9 Reading Water Quality     To fill the reservoir  follow these steps  see Figure 5      1  Remove the reservoir cap  2      2  Pour steam process distilled water into the reservoir until almost full   a maximum of 4L 1 U S  gal   Use a funnel to avoid spillage     3  Replace and secure the cap     3 5 Priming the STATIM Pump    To prime the STATIM pump  follow these steps  see Figure 6      1  Move the unit to the edge of the work surface 
12.  opened  the instruments must be handled in an aseptic manner for immediate use or resterilized     Dryness is important for unwrapped instruments for corrosion prevention   For wrapped instruments  a dry wrap is required to maintain sterility     The Air Dry Only cycle will run for up to 60 minutes    Heavy Duty Unwrapped Cycle  STATIM 5000 ONLY   DN 0 0 6   Sterilization Temperature  134  C  270  F    Sterilization Hold Time  6 minutes    The Heavy Duty Cycle  only available in STATIM 5000  is used to sterilize larger loads of unwrapped metal  instruments  Dental hand pieces can be sterilized in this cycle  The STATIM 5000 Extended Cassette  order no   01 104104  can be used to process loads up to 470 mm  18 5  long  such as rigid endoscopes  in this cycle     Instructions for Use 14 STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     4 8 Running a Cycle    To operate each cycle  follow these steps and watch the LCD     Turn the power switch at the back of the unit to ON  The LCD Display reads        Press the appropriate cycle button on the keypad        The display will read     BEMGUS YEO Se Jartam and LOSE VOL SELECT  ras pupilas then PRESS START       Once the start button has been pressed  the amber indicator light comes on to indicate  that the cycle is in progress  During the cycle  the phases of the cycle will be displayed similar to     di eso     A  Warming up D  Sterilizing       Hees SELE B  Conditioning   E  Venting  CYCLE HUMBER Hees        
13.  sterilization and damage  the instruments  the cassette  and the STATIM  Lubricated instruments must  be wiped thoroughly and any excess lubricant should be removed before loading     Unwrapped Instruments    Arrange unwrapped instruments on the instrument rack in the tray so  that they do not touch one another  This ensures that steam reaches all  surfaces and will promote drying     Instruments must not be stacked or piled in the cassettes  as this will  impede the sterilization process     Wrapped Instruments  single wrap     Place the instruments into single layer autoclave bags according to the  manufacturer   s instructions  Orient the instrument rack in the cassette to  ensure that wrapped instruments rest approximately 6 mm 0 25    above  the cassette base  Place the wrapped instruments on the rack and arrange  them to avoid overlap  Ensure that all wrapped loads are dry before  handling and or storage to maintain sterility     The use of cloth wraps in the STATIM is not recommended     Instructions for Use 11                            Figure 10    STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     SciCan recommends the use of plastic paper autoclave bags such as SPS     Medi Plus    and Chex All II      Steri Stik    paper paper autoclave bags may also be used  The use of cloth wraps in the STAT M is not recommended   Loosely pack instruments in the bags to allow steam penetration to all instrument surfaces     The wrapped instrument rack for the
14.  the holes in the lid  13 c           Figure 13 b  NOTE  At every corner and at the holes in the lid  two square nibs should    be visible  The nibs should fit flush with the lid   s outer surface        5  Ensure the seal is completely inserted  Feel around the periphery to     ensure the seal is securely in place   NOTE  During a cycle  steam may appear between the lid and the tray   If this persists  remove the cassette and check that the seal is  correctly installed  13 d                 Be careful  The metal parts will be hot  and the cassette may contain hot steam  Figure 13 c       5 8 Maintaining Fluid Levels    Use only steam process distilled water containing less than 5 ppm   total dissolved solids  having conductivity of less than 10 uS cm  in the  STATIM  To fill the reservoir  remove the cap from the top of the unit and  fill the reservoir  We recommend using a funnel to minimize spills  Each  time you refill the reservoir  empty the waste bottle and refill with water  to the MIN line  Empty the waste bottle often to avoid unpleasant odors  and discoloration of the contents  A low level chlorine free disinfectant   prepared according to the manufacturer   s instructions  may be added to  the waste bottle                 Figure 13 d    5 9 Reading Water Quality    1  Power up unit while pressing the STOP button to access the User  Setup menu     2  Using the UNWRAPPED and WRAPPED buttons   scroll to Water Quality and select it by pressing the  RUBBER AND PLASTIC 
15. 1 102119S  check that the arrow   mark on the filter matches the direction of the arrow on the bracket   Push the left hand filter fitting into tube B  7     Gently press the replacement filter into the filter bracket  6            STATIM 2000             Figure 11       STATIM 5000                                                                      Figure 12    The arrow mark of the filter should be facing out and pointing to the left     Re connect tube A  4  to the right hand filter fitting     Maintenance 18    STATIM 2000 5000 Cassette Autoclave Operator Manual       5 7 Replacing the Cassette Seal    To ensure optimum performance of your STATIM cassette autoclave  change  the cassette seal every 500 cycles or every six months  whichever comes first     Replacement seals are available from SciCan  order number 01 100028S  for STATIM 2000 and 01 101649S for STATIM 5000                  Figure 13 a  To change the cassette seal  follow these steps        1  Place the cassette lid and the new seal on a clean work surface      Examine the position of the old seal in the cassette lid and arrange the  new seal in the same orientation  next to the lid    2  Remove the old seal and discard  Clean any residue out of the seal ea  channel and flush out the channel with distilled water  13 a     gt    3  Lubricate the new seal with the liquid seal lubricant provided  13 b     4  Insert the rounded edge of the seal under the round lip of the lid  Align  the holes in the new seal with
16. 3 5 minutes    The Unwrapped Cycle is a general purpose sterilization cycle used to sterilize loads of solid metal instruments  such as pliers  burrs  scalers and forceps  Dental handpieces may be sterilized in this cycle     Wrapped Cycle  D O  E     Sterilization Temperature  134  C  273  F   Sterilization Hold Time  10 minutes  STAT M 2000   6 minutes  STAT M 5000     The Wrapped Cycle is used to sterilize solid and hollow metal instruments which have been sealed in paper   paper  or paper plastic autoclave bags  Dental handpieces may be sterilized in this cycle     If the cassette contains wrapped instruments and the wraps are not dry when the cassette is opened  the  instruments must be handled in an aseptic manner for immediate use  or resterilized     Rubber and Plastics Cycle  0 00 0    Sterilization Temperature  121  C  250  F   Sterilization Hold Time  15 minutes    The Rubber and Plastics Cycle is used to sterilize solid unwrapped instruments constructed of metal or the  materials listed in Section 4 4 Preparing and Loading Instruments     Air Dry Only Cycle  STAT M 2000 ONLY     O 0090 0    This is not a sterilization cycle     At the end of a cycle  if the cassette has not been removed from the autoclave  the Air Dry Only cycle may be  used to promote further drying  If the cassette has been removed from the autoclave  it may NOT be reinserted  for the Air Dry Only cycle  If the cassette contains wrapped instruments and the wraps are not dry when the  cassette is
17. 3 Ibs     emp Pressure          Time    Rubber Plastic  Load Capacity  0 4 kg  0 9 Ibs     Temp Pressure          Time    Heavy Duty  Load Capacity  1 5 kg  4 4 Ibs     STATIM 2000 5000 Cassette Autoclave Operator Manual    For all service and repair inquiries   In Canada  1 800 870 7777   United States  1 800 572 1211  International   416  446 4500  techservice ca scican com    SciCan Inc    701 Technology Drive  Canonsburg  PA 15317   USA   Telephone   1 724 820 1600  Fax   1 724 820 1479   Toll free  1 888 632 6031    EU Representative     SciCan GmbH   Wangener Stra  e 78   88299 Leutkirch   Germany   Telephone   49  0 7561 98343 0  Fax   49  0 7561 98343 699    SciCan Medtech  Alpenstrasse 16   6300 Zug Switzerland  Telephone   41  0  41 727 7027  Fax   41  0  41 727 702    Manufactured by    SciCan Ltd  C      1440 Don Mills Road  0123  Toronto ON M3B 3P9    Canada   Telephone   416  445 1600  Fax   416  445 2727   Toll free  1 800 667 7733       STATIM Cassette Autoclave and STATIM are registered  trademarks and STAT DRI  Your Infection Control Specialist   DriTecR and DriTec   logos are trademarks of SciCan Ltd    All other trademarks referred to in this manual are the property  of their respective owners       2013 SciCan Ltd   All rights reserved    95 113969 LA EN R1 STATIM 2000 5000 Operator Manual    SciCan   Your Infection Control Specialist       
18. Considerations   Section 3 1     Try running another cycle  If the problem persists  record the cycle fault message number and contact your  dealer        Drying interrrupted   LOAD NOT STERILE    User may have pressed the STOP button during the drying phase  Repackage the instruments  if necessary  and re run another sterilization cycle        Excessive steam issuing from  the front of the machine     message          Remove and reinsert the cassette  Attempt another cycle  If the leak persists  remove and check the cassette  seal for misalignment or damage  Replace the seal if required  Be careful as the metal parts will be hot and  the cassette will contain hot steam        If the leak persists  turn the unit OFF  remove and unload the cassette and contact your SciCan dealer        SciCan Ltd  Troubleshooting    27 STATIM 2000 5000 Cassette Autoclave Operator Manual    Message  WATER QUALITY IS NOT  ACCEPTABLE  Machine will not start     You have used water which is not steam process distilled or is improperly distilled    Empty the reservoir and refill with steam process distilled water containing less than 5 ppm total disolved  solids  having conductivity of less than 10 uS   cm     If you have the water conductivity meter  check the quality of the water before refilling the reservoir    Refer to the steps described in Section 3 11 Shipping the Unit to empty the reservoir        Message  REFILL RESERVOIR   Machine will not start     The level of the water in the reservoi
19. STATIM 2000 5000    CASSETTE AUTOCLAVE       e Operator   s Manual  e Manual del Operador    e Manual do Opera    o       SciCan   Your Infection Control Specialist       Latin America    TABLE OF CONTENTS    1  INTPODU CH ON siiis 2  171 Disclaimers is asasisaiiasiaassasinsare gro cvadastegsauibassasengieaidesascersedensa 2  1 2 HOW to Use this BOOK sserrep destims 3  1 3 Whats in the BOX siso ssanirsseasocs pes cassasedirasasreantisoquangegalicass 3   2  STATIM Unit OVerVieWw        cescceceeeeeeeseeeeeseeeeeseeeeeseeeeeeeeees 4   Si INStal atom 4s snscnshnscheatenceanahesuannaansecesiea CORRA NA Rasa Sn RNA Rand E 5  3 1 Environmental Considerations                        ss sas 5  3 2 Unit Plemenit eisiea iasan aaaea aiaiai 5  3 3 Connecting the Waste Botile                           ss sas  6  3 4 Filling the STATIM Reservoir                      s sis 6  3 5 Primingthe STATIM Pump sascssssserassainesscusgado prcaigsa faia sera 6  3 6 Setting the Time and Date                           oo 7  3 7 Language SelECHONn saaquanasisinmata inissentesssuasi  aini  ispaneiaianat  s 7  3 8 Assigning Unit Identifier Number                        7  3 9 Adjusting the Cycle Drying Time                             ss  is  8  3 10  Preparing Unit for USS  iseseisana 8  Slironipping the UAir eee Pere errr ete eee ee ite peer emer ae 8   4  Instructions for Use    sssss   scsssssasseasessisseseeessasesssssssasnsess 9  4 1 STATIM 2000     Using the Cassette                           9  4 2
20. age is displayed which gives a CYCLE FAULT code or a NOT STERILE message  the cassette contents  are not sterile  See Section 8 Troubleshooting for more information     If the cassette contains wrapped instruments and the wraps are not dry when the cassette is opened  the  instruments must be handled in an aseptic manner for immediate use  or resterilized     4 9 Stopping a Cycle    To stop a cycle press the STOP button     If the STOP button is pressed  or the cassette is removed  or the unit detects a problem while operating  the  cycle will stop and the amber active light will flash     Once a cycle has been stopped  the STOP button must be pressed before another cycle can be started     The display reads any of the following messages     or or       If the display shows the message  CYCLE FAULT or NOT STERILE  the cassette contents are not sterile   See Section 8 Troubleshooting for more information     If the air drying stage of the cycle is interrupted  the display reads the following messages        Instructions for Use 16 STATIM 2000 5000 Cassette Autoclave Operator Manual    5  Maintenance    SciCan Ltd     5 1 Cleaning the Cassette    Keeping the STAT M cassette clean is good clinical practice and assists in the function of the unit  SciCan  recommends that the interior surface be cleaned at least once a week     e Use dishwashing soap or a mild detergent that does not contain chlorine   e Scrub the inside of the cassette with a cleaning pad designed for use with Te
21. al if required     Be careful  The metal parts will be hot  and the cassette will contain hot steam     Attempt another cycle  If it still leaks attempt another cycle using a different cassette if possible     If the leak persists  turn the unit OFF  remove and unload the cassette  unplug the unit  and call your dealer        Instruments do not dry     Best drying occurs when the cycle continues to completion  Allow the cycle to finish  Make sure the  instruments are loaded correctly in the cassette  Refer to Section 4 4 Preparing and Loading Instruments     Check the unit leveling  Ensure the bubble level is set to Four o clock position   Check air filters and replace if dirty   For the STATIM 5000  ensure the Stat Dri plates are being used     Clean the inside of the cassette and treat with Stat Dri drying agent  Refer to Section 5 1 Cleaning the  Cassette     Examine the exhaust tube  tube to the waste bottle  for kinks     If kinked  straighten the tube  If the tube cannot be straightened  remove it from the push in fitting attached  to the STATIM  Depress the collar on the fitting and with the other hand pull firmly on the tube  Once the tube  is free of the fitting  cut the damaged section of tubing away using a sharp instrument  Be sure that you leave  enough tube to reach the unit when you re attach the tube to the exhaust fitting  If the tube is too short to  remove a section  contact your SciCan dealer for a replacement     For the STAT M 2000  make sure the compres
22. anty will cover the performance of all components of the unit except consumables such as the  cassette seal  the compressor filter and the microbiological filter  provided that the product is being used and  maintained according to the description in the user   s manual     In the event of failure due to such defects during this period of time  the exclusive remedies shall be repair or  replacement  at SciCan   s option and without charge  of any defected part s   except gasket   provided SciCan is  notified in writing within thirty  30  days of the date of such a failure and further provided that the defective part s   are returned to SciCan prepaid     This warranty shall be considered to be validated  if the product is accompanied by the original purchase invoice  from the authorized SciCan dealer  and such invoice identifies the item by serial number and clearly states the  date of purchase  No other validation is acceptable  After one year  all SciCan   s warranties and other duties with  respect to the quality of the product shall be conclusively presumed to have been satisfied  all liability therefore  shall terminate  and no action or breach of any such warranty or duty may thereafter be commenced against  SciCan     Any express warranty not provided hereon and any implied warranty or representation as to performance  and  any remedy for breach of contract which  but for this provision  might arise by implication  operation of law   custom of trade or course of dealin
23. button     CD   Conductivity yyy   engineering value  XX   micro S  value z z   parts per million value       SciCan Ltd  Maintenance 19 STATIM 2000 5000 Cassette Autoclave Operator Manual    5 10 Preventative Maintenance Schedule      To ensure trouble free performance  both the operator and the dealer must follow a preventative maintenance schedule     NOTE  Please refer to your National  Regional  State or Safety laws for any additional reoccurring user testing that    may be required     The schedules below describe the necessary actions     Water Reservoir    Daily    Replace water as needed   Check the reservoir for dirt or particles and clean if necessary     For opthalmic use  drain at the end of every workday  leave empty  and refill at the  start of the next workday        Waste Bottle    Cassette    Weekly    Empty the waste bottle every time you refill the water reservoir    Fill the bottle with water  up to the MIN line marking  You may also add some  chlorine free disinfectant    Wash the interior of the cassette with dishwashing soap or a mild detergent that  does not contain chlorine    Scrub the inside with a cleaning pad designed for   use with Teflon    coated surfaces    After removing all traces of the detergent  treat interior surfaces of the cassette  with the STAT DRI Plus drying agent to enhance the drying process    Order more STAT DRI Plus from your dealer quoting 2OZPLUS  80ZPLUST  or 320ZPLUS        Biological and or Air Filter    Check the filter
24. cals or disinfectants     Maintenance 17 STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     5 5 Changing the STATIM 2000 Air Filter    The filter should be replaced every six months in order to maintain an  adequate supply of clean air during the air drying cycle     To change the filter  follow these steps  see Figure 11      1   2     3     E    Turn the power switch at the back of the unit OFF    Remove and retain the screw  3  securing the filter plate  2    to the back of the compressor    Remove and discard the old foam air filter  1     Install the new filter  1   SciCan part no  01 1002078     Secure the filter plate  2  to the back of the compressor   using the screw  3  retained during the disassembly procedure     5 6 Changing the Bacteria Retentive Air Filters    The filters should be replaced every six months or after 500 cycles to  maintain an adequate supply of clean air during the air drying cycle     To change the bacteria retentive air filter on the STATIM 2000 and 5000   follow these steps  see Figures 11  amp  12      1   2     Power the STATIM OFF    Disconnect tube A  4  from the bacteria retentive filter  5  and remove  the filter from the filter bracket  6     As you remove the filter from the bracket  note the   orientation of the arrow mark on the filter    When the filter is free of the bracket    carefully disconnect tube B  7  from the filter    Before installing the replacement bacteria retentive filter  5    SciCan order no  0
25. close the lid  the hinge tab and slot will engage   Inserting the cassette into the STAT M 2000     1  Place the end of the cassette into the unit   2  Gently push inward until you hear a    click    sound     Never push the cassette into the STAT M with force as the interior  components could be damaged   Removing the cassette     1  Grasp the handle with two hands and pull away from the unit     2  Pull the cassette clear of the unit and set down on a firm surface     Disengaging the cassette    When not in use  the cassette should be disengaged        Cassette Latch                         Figure 8    To disengage the cassette  grasp the handle and pull the cassette out until  there is a 15 mm to 20 mm  4 to       gap between the front of the STATIM    2000 and the cassette handle     STAT DRI    Treatment of the interior surfaces of the cassette with the STAT DRI    drying agent  provided with your unit  will enhance the drying process      Replacement bottles are available from SciCan  order number 20ZPLUS     80ZPLUS  320ZPLUS      Instructions for Use 9    STATIM 2000 5000 Cassette Autoclave Operator Manual       4 2 STATIM 5000     Using the Cassette    When removing the cassette after a cycle  exercise caution as the metal  areas will be hot and the cassette may contain hot steam        To open the cassette     1  Push the carry handle  1  into the open position   2  Put your hands on either side of the cassette handle     3  Insert your forefingers in the slots and 
26. d between the cover and the door  The power  button and the paper advance button must protrude   through the openings in the cover and operate freely    Using a  1 Phillips screwdriver  secure the printer cover to the printer  door with the three screws retained during the disassembly procedure   Do not over tighten these screws    Plug your STATIM 5000 into the wall receptacle  Turn the power switch  to the ON position    Press the printer power button to the ON position  Load paper into the  printer following the procedures in Section 7 1 Installing Paper on the  Internal Printer                    ado PA D 10             12       Figure 16    STATIM 2000 5000 Cassette Autoclave Operator Manual    8  Troubleshooting    Unit does not power ON     Check that the unit is plugged into a properly grounded outlet and that the power cord is firmly seated at the  rear of the machine     Try another circuit  Power unit OFF for 10 seconds and then power ON again     Check the condition of the line circuit breaker or fuse        There is water under the machine     A  A    Check that water has not spilled when refilling the reservoir  Make sure the plug in the exhaust tube is  secured  Remove and reinsert the cassette   Attempt another cycle     Be careful  The metal parts will be hot  and the cassette will contain hot steam     The cassette is leaking if water drips from the underside of the unit during operation  Check the cassette seal  for misalignment or damage and replace the se
27. erview          cccceseceseseeeessceeeseeeeseseeeees 24   T  Internal Printer iissa 25  7 1 Installing Paper on the Internal Printer                             25  7 2 Removing Internal Printer Paper Jams                         26   8  Troubleshooting     s    ccessasssscesssasssaceasassssecesmasssacessmasieo 27   9  Spare  Parts iSense 29   10  Warranty sisiseccccccsssecctvecssesiececsssececieersianiieeessiieereesesieens 30   Ws SpecificationsS  arccos nnna 31   12  Appendix  Cycle Profile and Capacity                       32    SciCan Ltd  Table of Contents 1 STATIM 2000 5000 Cassette Autoclave Operator Manual    1  Introduction    Congratulations on your selection of the STAT M Cassette Autoclave      We are confident that you have purchased the finest equipment of its type  The STATIM   is a compact  counter top  unit that features a number of sterilizing cycles designed to meet your needs and suitability for steam sterilization     The details of installing  operating and servicing your STATIM are all contained within this operator s manual   To ensure years of safe  trouble free service please read these instructions before operating this unit and keep them  for future reference  Operational  maintenance and replacement instructions should be followed for the product  to perform as designed  Contents of this manual are subject to change without notice to reflect changes and  improvements to the STATIM product     The STAT M is suitable for the sterilization of dental
28. ette Lid Handle  2000  01 101779S Power Cord Europe  2000  amp  5000    01 104472 Extended Cassette Handles 01 101783S Reservoir Cap  2000  amp  5000    01 100008A   Kit Cassette Final  2000  01 101787S Water Reservoir Cap and Filter  2000  amp  5000   01 100028S   Cassette Seal  2000  01 109300S Water Reservoir Filter   01 100204S   Exhaust Tube  2000  amp  5000  01 102119S Filter Biological  2000  amp  5000    01 100207S   Compressor Filter 01 103475S Tray  2000    01 100271A   Kit Tray Cassette w  Mesh  Box Stat 01 103555S8 Power Cord Replacement Denmark   01 100572S   Power Cord North America  2000  01 103557S Power Cord Replacement Denmark  detached   01 100584S   Power Cord German  2000  01 103865S Seal Lubricant  2000  amp  5000    01 100585S   Power Cord Italian  2000  01 104093S Exhaust Tube 3m Long   01 100586S   Power Cord U K   2000  01 104284S Biological Filter Bracket  2000  amp  5000    01 100612S   English Manual Tubing Indicators  2000  01 1043438 Plug Drain Tubing  5000    01 100724S   Condenser Bottle w o Condenser  2000  amp  5000    01 100730S   Power Cord Japan  2000    01 100735S   Waste Water Bottle Fitting  2000  amp  5000  01 101613S Cassette Complete with Box  5000    01 100799S   Power Cord Switzerland  2000  01 106325 Container Endoscope Complete   01 100812S   Condenser Bottle  2000  amp  5000  01 108262S Basket Medical  2000    01 100834S   Cassette Lid  2000  01 102054S Rack Wrapped Instruments  5000    01 101553S   Leveler  2000  amp  5000  01
29. faces of the cassette with the STAT DRI  drying agent  provided with your unit  will enhance the drying process    Replacement bottles are available from SciCan  order number 2OZPLUS   80ZPLUST  320ZPLUS                                                SciCan Ltd  Instructions for Use 10 STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     4 3 STATIM 5000     STAT DRI Plates    The STATIM 5000 cassette uses STAT DRI plates to enhance the drying  process for wrapped instruments  The adjustable plates are designed for  the unwrapped instruments rack and eliminate the need for a separate  wrapped instrument rack  Up to ten plates can be arranged along the  length of the rack  Additional plates may be ordered from SciCan  SciCan  part number 01 103935      To install and adjust a STAT DRI plate  follow these steps   1  Hold a plate over the desired position  tilted slightly forward with the  tabs down   2  Insert the tabs between the mesh of the unwrapped instrument rack  1      3  Each tab has an elongated slot  Slide the plate until the wire is at the  end of each slot     4  Release the plate to the resting position   5  Prepare and load instruments for sterilization  2      4 4 Preparing and Loading Instruments    Before loading any instruments into the STATIM   consult the manufacturer   s reprocessing instructions     Clean Instruments    Clean and rinse all instruments before loading them into the cassette   Disinfectant residues and solid debris may inhibit
30. flon    coated surfaces   e After scouring  rinse thoroughly with water to remove all traces of the detergent     Cleaning the inside of your cassette is very important if you regularly sterilize lubricated instruments     Coating the entire inside surface with STAT DRI STAT DRI PLUS drying agent induces water to form an even coat  on the inside surface  without beading  The water in contact with the hot cassette surfaces also evaporates much  more efficiently  Spotting is minimized and instruments dry much better     STAT DRI STAT DRI PLUS should be applied every 10 cycles  or when you notice water beading on the lid  and  after every cassette cleaning     5 2 Cleaning the Water Reservoir Filter    The water reservoir filter should be cleaned at least once a week or when required  The filter can easily be removed  and cleaned by placing the filter upside down under running water to wash away the particles until clean  and then  placed back into the reservoir opening  If a replacement water reservoir filter is required  order part number 01 109300S     5 3 Cleaning the Reservoir    Check the reservoir for dirt or particles  The reservoir may be cleaned by draining followed by cleaning and rinsing  with steam process distilled water ONLY  Use of chemicals or cleaning agents is not reccommended and could  cause the unit damage     5 4 Cleaning the Exterior Surfaces    Use a soft cloth moistened with soap and water to clean all exterior surfaces  Do not use harsh cleaning  chemi
31. g   including any implied warranty of merchantability or of fitness for particular  purpose with respect to all and any products manufactured by SciCan is excluded and disclaimed by SciCan    If you would like to learn more about SciCan products and features  visit our website at www scican com     Warranty 30 STATIM 2000 5000 Cassette Autoclave Operator Manual    11  Specifications    Machine Dimensions  L x W x H      485 mm x 415 mm x 150 mm    550 mm x 410 mm x 190 mm          19    x 16 3    x 5 9    22    x 16    x 7 5     Cassette Size  External  410 mm x 195 mm x 40 mm 495 mm x 195 mm x 80 mm   L x W x H   includes Handles   lo renda eta IS SPEEDS 2A  Cassette Size  Internal   L x W x H     280 mm x 180 mm x 35 mm 380 mm x 180 mm x 75 mm  11    x 7 1    x 1 4  15    x 71  x3   Extended Section  L x W x H   N A 110 mm x 180 mm x 16 mm    4 3    x 5 1  x 0 63        Sterilization Chamber Volume     1 8L  61 fl  oz   U S     51 L  170 fl  oz   U S        Reservoir Volume     4 0 L  140 fl  oz   U S     4 0 L  140 fl  oz  U S        Weight  Without water      Clearance required   Top Sides Back Front      21 kg  46 Ibs     50 mm  1 9     50 mm  1 9       50 mm  1 9     480 mm  18 9           33 kg  73 lbs     50 mm  1 9     50 mm  1 9       50 mm  1 9    570 mm  22 4           Optional Internal Printer     Type  N A   Print  N A   Print Speed  N A   Paper Roll Capacity  N A    Type  Thermal Printer   Print  20 characters per line   Print Speed  1 line per second   Paper R
32. he cycle buttons to select and change the selected field   s value   To increase a field   s value  press the UNWRAPPED cycle button    Hold the button down to increase the value    To decrease the value  press the WRAPPED cycle button    To select the next field  press the RUBBER AND PLASTICS cycle button     7  To save changes and return to the regular operating mode   press the STOP button     8  To quit without making changes  power the STATIM OFF        Display when setting the Date   Time    D       3 7 Language Selection    The messages displayed on the LCD can be changed to a number of HM  Ele  different languages  To change the current language selections  follow  these steps        1  Turn the power switch at the back of the unit OFF  Display when scrolling through languages    2  Press and hold the WRAPPED cycle button   3  While depressing the WRAPPED cycle button   turn the power switch at the back of the unit to ON   4  Press the UNWRAPPED cycle button  to scroll to the next language selection   Press the WRAPPED cycle buttonto scroll to the previous language   6  When the desired language is displayed  press the STOP button  to save the selection and return to the regular operating mode     a    3 8 Assigning Unit Identifier Number    1  Power the STAT M OFF    Press and hold the RUBBER PLASTIC cycle button    3  While pressing the RUBBER PLASTIC cycle button   power the STAT M ON    4  Using the cycle buttons  select a maximum of 3 digits to be used as  the un
33. hemical indicators  please  consult the product literature accompanying the indicators or contact the manufacturer directly     To use the indicators with the STATIM  follow these steps     1  Place the appropriate biological indicator in the STATIM chamber    2  Process the load in the sterilizer according to your usual practice    3  Ensure that the message    Cycle Complete    is displayed on the LCD after the cycle is finished   4    Recover the biological and or chemical indicator and process further according to the literature that accompanied  the indicator     At the first indication of a potential sterilization failure     1  Do not process any more instruments until favourable test results have been returned    2  Ensure the correct indicator type was chosen    3  Ensure the cassette was not overloaded  Consult the earlier portion of this section for proper loading instructions   4    If the results do not change  do not process any more instruments within the STAT M and contact your SciCan  dealer for further assistance     4 6 Instrument Weight Guide                               Scissors 30g 0 96 oz   Dental scalers 20 g 0 64 oz  Forceps 15 g 0 48 oz  Dental handpiece 40 60 g 1 29 to1 92 oz  Wrapped instrument rack 260 9 8 35 oz  Unwrapped instrument rack 225 9 7 23 oz  Suction cannula 10 9 0 32 oz  Plastic mouth mirror 8 9 0 25 oz  Impression tray 15 to 45 g 0 48 to 1 45 oz  Plastic x ray positioning ring 20 g 0 64 oz                NOTE  The above weights are t
34. it   s identifier  The UNWRAPPED button will increase the selected  value and the WRAPPED cycle button will decrease it    Use the RUBBER PLASTIC to move to the next digit    5  To save changes and return to the regular operating mode    press the STOP button        N    Display when assigning unit number    SciCan Ltd  Installation 7 STATIM 2000 5000 Cassette Autoclave Operator Manual    3 9 Adjusting the Cycle Drying Time    The STAT M has preset dryng time of 10 minutes for Unwrapped cycles  and 30 minutes for Wrapped cycles  You can adjust the time between 5  to 60 minutes for Unwrapped cycles and 5 to 60 minutes for Wrapped    cycles     1   2     3     Power the STATIM OFF    While pressing and holding the STOP cycle   button  power the STATIM ON    Using the Wrapped and Unwrapped cycle buttons scroll to DRYING  UNWRAPPED or DRYING WRAPPED and   press the RUBBER PLASTIC button           or    Using the Wrapped and Unwrapped cycle buttons  scroll to the desired    drying time and select it by pressing the RUBBER PLASTIC button    Unwrapped Drying Time applies to Unwrapped   Heavy Duty  and Rubber Plastic cycles      To save changes and return to the regular operating mode  press the  STOP button     3 10 Preparing Unit for Use       Once the unit is installed and before any instruments are sterilized  run two  Wrapped cycles  see Section 4 8 Running a Cycle   Remove the cassette    once it has cooled  Clean the top  lid  and bottom  tray  sections using a    soft cloth t
35. k of the unit  Follow the steps below before connecting  the Data Logger     User Setup Menu    Please ensure Time Date and Unit ID are set as per Section 3 6 and  Section 3 8  To install the SciCan Data logger  access the user setup  menu by powering up the unit while pressing the STOP button   Follow the instructions below to complete installation steps     STEP 1  Selecting the USB Flash MSD Option         1  Power up unit while pressing the STOP button to access the User  Setup menu           USER SETUP MENU    2  Using the UNWRAPPED and WRAPPED buttons  scroll to RS232 and    select it by pressing the RUBBER AND PLASTIC button     3  From the RS232 menu  use the UNWRAPPED and  WRAPPED buttons to scroll down to the USB Flash MSD option  and press the RUBBER AND PLASTIC button to  select and return to the User Setup menu     STEP 2  Setting the Serial Port Bitrate to 9600       1  From the User Setup menu  use the UNWRAPPED and WRAPPED  buttons to scroll to the Serial Port Bitrate menu and select it using the  RUBBER AND PLASTIC button     2  From the Serial Port Bitrate menu  use the UNWRAPPED and  WRAPPED buttons to scroll down to 9600 and press the  RUBBER AND PLASTIC button to select and  return to the User Setup menu     STEP 3  Setting the Printer user Character  i e  134  C          1  From the User Setup menu  use the UNWRAPPED and WRAPPED       buttons to scroll to the Printer user    char menu and select it using the    RUBBER AND PLASTIC button     2  From the P
36. ll the paper into the printer  follow these steps     FON a    q    Power the STATIM 5000 ON    Open the printer door  1  by pushing on the top half of the door   Power the printer ON  2     Unroll some paper from the thermal paper roll  3  and trim the corners  using the paper cutting template included with each box    Move the paper roll arm  4  into the loading position  Place the paper  roll  3  on the arm so the paper strip feeds from the top of the roll and  then carefully insert it into the paper feed slot  5  until it stops     If the paper does not feed from the top  the heat sensitive side of the  paper will not be in contact with the print head and the printer will not    print     6     With one hand  continue to gently feed the paper strip into the paper  feed slot  With the other hand  press the paper advance button  6   until the paper feeds by itself     Keep the paper straight when feeding it into the printer or it may jam  Do  not force the paper into the slot  If the paper will not feed into the slot   pre cut the end of the roll again and reload the paper     fa    Continue to press the paper advance button  6  until the paper feeds  through the paper exit slot on the front of the printer  Then  move the  paper roll  3  and arm into the operating position and close   the printer door  1   The printer is now ready to operate     When you see a red line on one side of the paper  it is time to replace the    roll     If a paper jam occurs  and the paper canno
37. nd leave the arm  in the loading position    Using a  1 Phillips screwdriver  remove the three screws  7  from the  printer cover  8  and remove the cover    Note the orientation of the exposed printed wiring board  15  and the  paper roll arm  4  assembled on the printer door  1     Gently lift the printed wiring board upwards and away from the printer  door  Exercise care while handling the board  The printer is integral to  the wiring board  Do not place strain on the connections of the ribbon  cable soldered to the board  Do not remove the connector of the  flexible cable from the connector header on the board  The paper drive  mechanism on the underside of the wiring board is now exposed   Using a pair of tweezers or fine needle nosed pliers  carefully remove  the paper from the mechanism     When the paper is removed  reassemble the printer     1     2     Internal Printer 26    Carefully snap the paper roll arm  in the loading position  back into the  clips on the printer door  1     Place the printed wiring board back into position on the printer door   Note the alignment of the mounting holes  9  in the wiring board and the  mounting bosses  13  on the printer door  The black plastic printer body  rests between the locating ribs  10  on the inside of the printer door   Ensure that the flexible cables  11    14  are not pinched between the  printer door and the wiring board    Place the printer cover on the printer door  Ensure that the flexible  cables are not pinche
38. o be used as reference only   For exact weights of your instruments  consult the manufacturer   s specifications   4 7 Cycle Profiles    0 0     The STATIM s have multiple sterilization cycles  each designed to sterilize a specific type of instrument   The instruments will remain sterile after a successful cycle until the cassette is removed from the autoclave   At the end of each sterilization cycle  microbiologically filtered air drying will commence           Unwrapped instruments  once exposed to ambient or external conditions  cannot be maintained in a sterile state   If sterile storage is desired  wrap the instruments to be sterilized in autoclave bags according to the instrument  manufacturer   s instructions  and allow the wrapped cycle to run until the air dry phase is complete     It is important not to overload the chamber as this can inhibit steam access to all instrument surfaces  The types  of instruments and sterilization requirements for each cycle are described over the next few pages  Consult   the Instrument Weight Guide in the previous section for information on how to make up an appropriate load for  the masses specified for individual cycles  The maximum load capacity and a graph depicting each cycle are  provided in the Appendix  Cycle Profile and Capacity     Instructions for Use 13 STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     Unwrapped Cycle  AM tO mt      Sterilization Temperature  134  C  273  F     Sterilization Hold Time  
39. o wipe the inside surfaces and then rinse thoroughly with tap  water  Once the cassette is clean and dry  coat the inside surfaces with  STAT DRI     3 11 Shipping the Unit    Before you move the unit  you will need to drain the reservoir  To do so   follow these steps     1   2     3     SciCan Ltd  Installation    Place a water container below the unit     Use the drain tube  5  to empty the contents of the reservoir into the  water container     Remove any remaining water from the reservoir with a non linting   absorbant towel     Screw in the three leveler feet  6  found underneath the unit     Repack the unit in the original packing materials and include all  accessories originally included with the unit     Specify heated and insured shipping                                                     Figure 7    STATIM 2000 5000 Cassette Autoclave Operator Manual    4  Instructions for Use    A    SciCan Ltd     4 1 STATIM 2000     Using the Cassette    When removing the cassette after a cycle  exercise caution as the metal    areas will be hot and the cassette may contain hot steam     To open the cassette     1  Hold the cassette handle with your thumbs facing inward on the  cassette latch     2  Push downward on the cassette latch   3  Raise the cassette lid upwards and disengage the hinge   4  Rest the lid on its outer surface     To close the cassette     1  Align the hinge tab on the cassette lid with the hinge slot on the  rear of the bottom tray     2  As you begin to 
40. oll Capacity  approx  80 sterilization cycles  per roll       Maximum Steam Temperature     138  C  280  F     138  C  280  F        Maximum Operating Pressure     341 kPa abs  49 5 psia     341 kPa abs  49 5 psia        Electrical Rating      10      see serial number label for requirements specific  to your unit     100 V  50 60 Hz  11 A  110 V  50 60 Hz  11 A  220 240 V  50   60 Hz  6 A    100 V  50 60 Hz  11A  110 V  50 60 Hz  11 A  220 240 V  50 60 Hz  6 A       Protection Class              Protection  Covered  indoor use only  Covered  indoor use only    Ambient Operating Temperatures   15  C to 25  C  59  F to 77  F  and 25  to 70  15  C to 25  C  59  F to 77  F  and 25  to 70   and Humidity    Altitude  Up to 2000 meters  6600 ft  Up to 2000 meters  6600 ft           Installation Category        1    1          SciCan Ltd  Specifications    31    STATIM 2000 5000 Cassette Autoclave Operator Manual    12  Appendix  Cycle Profile and Capacity    Temp Pressure       Temp Pressure    Temp Pressure    STATIM 2000       Unwrapped  Load Capacity  1 0 kg  2 2 lbs        Wrapped  Load Capacity  1 0 kg  2 2 lbs     Rubber Plastic  Load Capacity  0 4 kg  0 9 Ibs        Time          Time                A  Warming up D  Sterilizing  B  Conditioning E  Venting  C  Pressurizing F  Air Drying          SciCan Ltd  Appendix    32    STATIM 5000    Temp Pressure       Time  Unwrapped  Load Capacity  0 5 kg  1 1 Ibs      Pressure    emp          Time  Wrapped  Load Capacity  1 5 kg  3 
41. ow you to connect to an external  printer or to a SciCan Data Logger              For printing  you will need to purchase a recommended printer  see list  below  from your local computer or electronics store  For data storage  you  can purchase SciCan   s USB Data Logger to record and store cycle  information onto a mass storage device  MSD  such as a USB Flash Drive  or SD memory card      Printer modei Endofline RAR SerialPort Bate Forener   Figure 4                              eee  C31C515603  SM   disie  E 10 40 FF 120B R ga Ea   E ada 20 eR iii rs  ao a 200 e e ces  Se ONDA End e                             For Mass Storage Device  CR LF 9600 82  Oxd2          Due to a variety of software programs viewers that can be used to view  the data files stored on a mass storage device such as a USB memory  stick  we recommend the above settings     For your STATIM to communicate to a specific device  you must enable  this function through the User Setup menu  Follow the instructions   in Section 6 2 Installing the SciCan Data Logger below to enable  communication to either an external printer or the SciCan Data Logger     The Communication Port 21 STATIM 2000 5000 Cassette Autoclave Operator Manual    6 2 Installing the SciCan Data Logger    SciCan   s USB Data Logger can record and store cycle information onto  amass storage device  MSD  such as a USB Flash Drive or SD memory  card     These instructions are for units with a 9 pin communication  RS232  port  located at the bac
42. place your thumbs  on the thumb pads     4  Press down with your thumbs and pull up with your forefingers  until the lid opens     5  Raise the cassette lid and disengage from the tray   Rest the lid on its outer surface     To close the cassette     1  Align the hinge tab on the lid with the hinge slot on the tray   2  As you begin to close the lid  the hinge tab and slot will engage   3  Place the carry handle into the closed position     Inserting the cassette into the STATIM 5000   1  Hold the cassette handle in one hand and the carry handle  in the other as shown in Figure 9     2  Place the end of the cassette into the unit  and drop the carry handle into its closed position     3  Gently push the cassette inward until you hear a click sound     Never push the cassette into the STAT M with force as the interior  components could be damaged              Removing the cassette     1  Grasp the cassette handle with one hand and pull out from the unit     2  As the cassette emerges from the unit  grasp the carry handle with  your free hand and lift it upwards     3  Pull the cassette clear from the unit and set down on a firm surface              Disengaging the cassette             To disengage the cassette  grasp the handle and pull the cassette out  until there is a 15 mm to 20 mm  42 to 34     gap between the front of the  STAT M 5000 and the cassette handle     STAT DRI         When not in use  the cassette should be disengaged  Figure o    Treatment of the interior sur
43. r is low  Refill the reservoir  Refer to the steps described in Section 3 4  Filling the Reservoir        Message  PRINTER FAULT displayed on  LCD Printer is not printing     Check for a paper jam  If the paper is jammed  follow the paper removal procedures outlined in Section  7 2  Power unit OFF for 10 seconds and then power ON again  If the message still appears  follow the  disassembly procedures outlined in Section 7 2 Removing Paper Jams        The printer does not work     Make sure that the printer cable is connected securely with the connector on the back of the STAT M and  the Statprinter  Make sure that the printer is powered ON  Power unit OFF for 10 seconds and then power  ON again        The printer appears to work  but nothing is  printed on the paper     Ensure that the paper is loaded properly  refer to Section 7 1 Installing Paper on the Internal Printer   Check  to see that the paper leaves the paper roll from the top of the roll  This means that the treated surface of the  coated thermal paper will be in contact with the thermal print head        Time and date are incorrect     The time and date have not been set  See Section 3 6 Setting the Time and Date        Message  MSD NOT CONNECTED  lt  gt   INSERT MSD FLASH    Check the serial cable connection    Check the power connection    Ensure the lower red LED is lit    Check that the Mass Storage Device is properly inserted     Repeat the instructions for Installing the SciCan Data Logger on your STATIM    
44. rinter user    char menu  using the UNWRAPPED button to    increase the displayed value by one  and the WRAPPED button to  increase that value by ten  enter the value 32  0x20    and press the RUBBER AND PLASTIC button to   accept and return to the User Setup menu     SciCan Ltd  The Communication Port 22    STATIM 2000 5000 Cassette Autoclave Operator Manual    STEP 4  Save and Exit          A Save and Exit must be performed after the above settings are  completed  If this is not done  the information will revert to its prior    settings   1     2   3     From the User Setup menu  use the UNWRAPPED and WRAPPED  buttons to scroll to Save and Exit     Select it by pressing the RUBBER AND PLASTIC button     The LCD will display time and date information and a sequence of  messages  See images after STEP 5 below      STEP 5  Connecting the SciCan Data Logger    1     Ensure that both the STAT M unit and the SciCan Data Logger are off           2  Connect the SciCan Data Logger to the STATIM  unit by using the serial cable   3  Power up the SciCan Data Logger   4  Power up the STATIM unit   5  The LCD will display the following sequence of messages   6  Insert USB Flash Drive or SD memory card   7  After a few seconds the LCD will display the  following sequence of messages   SciCan Ltd  The Communication Port 23    STATIM 2000 5000 Cassette Autoclave Operator Manual    6 3 Cycle Printout Overview       Model  STATIM 2000              Unit Identifier  autoclave has  been set up a
45. s number 000    Time Date   10 47 am 25th October 2013    Cycle Name   and Parameters   UNWRAPPED  for 135  C 3 5 min    Warm up complete   start of the conditioning  phase is 1 19    see cycle graph       A    phase complete   start of    B    phase     Temp  Press   amp  Time  printed at 30 second  intervals during  sterilization     D    phase               Cycle completion time     SciCan Ltd  The Communication Port    16 42     24                       Software  S204R616    Cycle counter  the number  __    off cycles having been run  on the unit   2    Cycle clock  starting at 0 00    Start Time of Pressurization     1 27  start of    C    phase        Start Time of Sterilization   2 27  start of    D    phase       Time Venting started   TTT  5 57  start of    E    phase      T  Time Air Drying started   6 42  start of    F    phase        STATIM 2000 5000 Cassette Autoclave Operator Manual    7  Internal Printer     Not on all models     7 1 Installing Paper on the Internal Printer    Use only paper approved for use with the optional internal printer on the  STATIM 5000  The use of any other paper will damage the printer and will    void the warranty  Thermal paper is available from your dealer  SciCan  order no  01 101657S      SS    Do not operate the printer without paper  If you run out of thermal paper   or if you do not wish to use the printer  turn it OFF     Never pull the paper backwards through the printer  This will damage the  printer mechanism     To insta
46. sor is working  To check  remove the exhaust tube from the  waste bottle  Start the Air Drying Only Cycle  and place the free end into a glass of water  If there is not a  strong  steady flow of bubbles  the compressor is not functioning properly  Contact your SciCan dealer        Cycle interrupted     NOT STERILE   Cycle aborted     NOT STERILE and    CYCLE FAULT messages     Record the CYCLE FAULT number  Wait a few minutes and attempt another cycle before proceeding to   the next solution  If the message persists  remove the cassette  Be careful  The metal parts will be hot and  the cassette will contain hot steam  Inspect the cassette to ensure that the holes in the back of the seal are  perfectly aligned  and that the flexible lip of the seal is completely free  Check the exhaust tube for kinks or  obstructions  If kinked  straighten the tube    If the tube can not be straightened  remove it from the push in fitting attached to the STATI  Depress the  collar on the fitting and  with the other hand pull firmly on the tube  Once the tube is free of the fitting  cut the  damaged section of tubing away using a sharp instrument  Be sure that you leave enough tube to reach the  unit when you re attach the tube to the exhaust fitting  If the tube is too short to remove a section  contact  your SciCan dealer for a replacement     Check that the STATIM has not inadvertently been exposed to any electrical interference   Refer to the Installation section dealing with Environmental 
47. t be removed by pressing the  paper advance button  6   do not pull the paper backwards through the    printer     Never put a utensil or tool into the paper exit slot   For full instructions on how to remove paper jams   see Section 7 2 Removing Internal Printer Paper Jams     To replace the paper roll  3   follow these steps     J     With scissors  cut the paper between the roll and the paper feed slot  5      2  Remove the roll from the arm and discard the unused portion   3  Press the paper advance button  6  to feed the paper that remains in  the printer out of the slot at the front of the printer   4  Install the new thermal paper roll by following the instructions described  in this section   SciCan Ltd  Internal Printer 25                   Figure 15a                Figure 15 b                   STATIM 2000 5000 Cassette Autoclave Operator Manua    Figure 15 c       SciCan Ltd     7 2 Removing Internal Printer Paper Jams    If paper jams in the printer and cannot be removed by pressing the paper  advance button  6   the printer must be disassembled  Do not pull the  paper backwards through the printer and never put utensils or tools into  the paper exit slot     When paper is jammed in the printer  follow these steps to remove the jam   refer to figures 15c  amp  16      1   2    3     Power the STATIM 5000 OFF and unplug the unit    Using scissors  cut the paper between the   roll and the paper feed slot  5     Remove the paper roll  3  from the paper roll arm  4  a
48. te sterilization solution for your unwrapped  and wrapped instrument needs  rapid sterilization balanced with rapid drying  by using the SciCan Dri Tec Drying  Technology        p To dry instruments  the STAT M 2000 uses a combination of forced filtered air and convection heat  The convection  Dri Tect heat is derived from utilizing the remaining heat in the system after the sterilization phase has been completed   The heat is then captured and released into the cassette to rapidly dry a properly loaded cassette        The STAT M 5000 also uses forced filtered air and heat generated by the sterilization phase to dry the instruments   The heat generated by the sterilization phase is absorbed by the drying plates and then transferred directly to the  load  This then results in an accelerated  rapid drying of a properly loaded STATIM cassette         Dri Tec      Please refer to this operator   s manual for instructions on the proper arrangement of instruments in the cassette and  the use of STAT DRI    plates  STATIM 5000   By carefully following these directions on how to properly load the    cassette chamber  rapid drying of the load will be achieved     SciCan Ltd  Introduction 2 STATIM 2000 5000 Cassette Autoclave Operator Manual    1 2 How to Use this Book    The following symbols appear in the margins of this book     Warning icons        gt      _                         A potential A situation Important Caution  Caution  Caution  Alternating  hazard to which may informa
49. tion Hot Surface Risk of Refer to manual between  the operator  lead toa and or electrical shock  for details messages   mechanical Hot Steam Disconnect  failure  supply before  servicing  Cycle icons  Unwrapped Wrapped Rubber Plastic Air Cycle START STOP  Cycle Cycle Cycle  STATIM 2000  Button Button  or  Heavy Duty  Unwrapped   STATIM 5000     1 3 What s in the Box    When you receive your STATIM packing carton  the items listed below will be included   If any of the items are missing  contact your dealer immediately so that the situation can be corrected        Waste Bottle and Bottle Lid STAT DRI PLUS Spray  Fitting  el  f  A S5      SO  Rea       Tube Mounting Hardware       Wrapped Instrument Rack Operator s Manual Exhaust Tube   STATIM 5000                       SciCan Ltd  Introduction 3 STATIM 2000 5000 Cassette Autoclave Operator Manual    2  STATIM Unit Overview    STATIM 2000                               Figure 1a jj Figure 1 b       STAT M 5000    3                                                                                                             Figure 2 a Figure 2 b  1  Level indicator 9  Leveler feet  2  Keypad 10  Exhaust tube port  3  Reservoir cap water filter 11  Compressor  4  LCD 12  Cassette  5  Power indicator 13  Biological air filter  6  Active indicator 14  Optional internal printer  7  Power switch  STATIM 5000   8  Power cable port 15  RS232 port  not on all models    16  Serial Model Number Label       SciCan Ltd  Unit Overview 4 STATIM
50. tions  cannot be maintained in a sterile state  If sterile storage  is desired  wrap the instruments to be sterilized in autoclave bags  according to the instrument manufacturer   s  instructions  and then allow the wrapped cycle to run until the air dry phase is complete     Best Practice  Allow instruments  wrapped or unwrapped  to dry completely prior to handling  Instruments must  not touch each other to promote drying and enable effective sterilization     SciCan recommends the final user carefully choose the most appropriate sterilization cycle according to the  recommendations of their leading infection control authorities and local regulatory guidelines  recommendations     Routine Monitoring    Chemical process indicators suitable for steam sterilizers should be included in or on each package or load being  sterilized  In addition  the weekly use of biological indicators  which allow you to ascertain whether the instruments  have been exposed to sterilization conditions  is recommended  SciCan recommends using the 3M Attest     biological monitoring system for routine monitoring  This system consists of self contained biological indicators and  incubators  It is important to select the correct biological indicator for the cycle being tested     Instructions for Use 12 STATIM 2000 5000 Cassette Autoclave Operator Manual    SciCan Ltd     4 5 Using Biological Chemical Indicators    For detailed instructions on how to handle  use and dispose of both the biological and c
    