Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

image

Contents

1. e The Simple Remote Front Panel see Simple Remote Front Panel on page 261 e The Browser Based Remote Front Panel see Browser Based Remote Front Panel on page 262 e The SCPI Command window applet for Remote Programming see Remote Programming via the Web Interface on page 263 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 259 20 Web Interface If Java is not installed on your PC you will be prompted to install the Sun Microsystems Java Plug in This plug in must be installed on the controlling PC for the web interface s Remote Front Panel or Remote Programming operations The SCPI Command window is useful for testing commands or entering a few commands interactively When creating automated programs for controlling the oscilloscope you will typically use the Agilent IO Libraries from within a programming environment like Microsoft Visual Studio see Remote Programming with Agilent IO Libraries on page 264 Real Scope Remote Front Panel To operate the oscilloscope using the web interface s Real Scope Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Real Scope Remote Front Panel After a few seconds the Remote Front Panel appears 3 Click keys or knobs that you would normally press on the oscilloscope s front panel Drag on edges of knobs to
2. 1 In the Save Menu press the File Name softkey You must have a USB storage device connected to the oscilloscope for this softkey to be active 2 In the File Name Menu use the Spell Enter and Delete Character softkeys to enter the file name Spell press this softkey and turn the Entry knob to select the character at the current position Enter press this softkey to enter characters and move the cursor to the next character position Pushing the Entry knob is the same as pressing the Enter softkey Delete Character press this softkey to delete the character at the current position 226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 17 You can use a connected USB keyboard instead of using the Spell and other character editing softkeys When available the Increment softkey can be used to enable or disable automatically incremented file names Auto increment adds a numeric suffix to your file name and increments the number with each successive save It will truncate characters as necessary when the file name length is at maximum and more digits are required for the numeric portion of the file name Recalling Setups Masks or Reference Waveforms 1 Press the Save Recall key 2 In the Save Recall Menu press Recall 3 In the Recall Menu press Recall then turn the Entry knob to select the type of file you want to recall Setup scp See To recall
3. takes to print oscilloscope images by changing the black background to white Invert Graticule Colors is the default mode Graticule Not Inverted Graticule Inverted e Form Feed Select this to send a form feed command to the printer after the waveform is printed and before the setup information is printed Switch Form Feed off if you want setup information printed on the same sheet of paper with the waveform This option only has an effect when the Setup Information option is selected Also if the amount of setup information will not fit on the same page with the waveform it will be printed on a new page regardless of the Form Feed setting e Landscape Select this to print horizontally on the page instead of vertically portrait mode 234 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Print Screens 18 To specify the palette option In the Print Configuration Menu press the Palette softkey to change the following options e Color Select this to print the screen in color The oscilloscope s print driver cannot print color images to color laser printers so the Color option is not available when connected to laser printers e Grayscale Select this to print the screen in shades of gray rather than in color Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 235 18 Print Screens 236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000
4. the trigger condition is not found after the pre trigger buffer has filled and no acquisitions are occurring Trig d not flashing trigger condition is found or pre trigger buffer is being filled When the oscilloscope is not running the trigger indicator area shows Stop Whento Use Auto The Auto trigger mode is appropriate when Trigger Mode Checking DC signals or signals with unknown levels or activity e When trigger conditions occur often enough that forced triggers are unnecessary When to Use The Normal trigger mode is appropriate when Normal Trigger Mod e You only want to acquire specific events specified by the trigger ode settings e Triggering on an infrequent signal from a serial bus for example I2C SPI CAN LIN etc or another signal that arrives in bursts The Normal trigger mode lets you stabilize the display by preventing the oscilloscope from auto triggering e Making single shot acquisitions with the Single key Often with single shot acquisitions you must initiate some action in the device under test and you don t want the oscilloscope to auto trigger before that happens Before initiating the action in the circuit wait for the trigger condition indicator Trig d to flash this tells you the pre trigger buffer is filled See Also Forcing a Trigger on page 115 e To set the trigger holdoff on page 140 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guid
5. Table9 Accessories Available for InfiniiVision 2000 X Series Oscilloscopes Model Part Description DSOXLAN LAN VGA connection module DSOXGPIB GPIB connection module N6456A Rack mount kit N6457A Soft carrying case and front panel cover N2786A 2 leg probe positioner N2787A 3D probe positioner 1180CZ Testmobile N6458A Hardcopy user s guide various Front panel overlays see Front Panel Overlays for Different Languages on page 35 N6459 60001 8 channel logic probe and accessory kit standard with MSO models and with DSOX2MSO upgrade Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Reference 21 You can find these items at www agilent com or at www parts agilent com Loading Licenses and Displaying License Information e Licensed Options Available on page 279 e Other Options Available on page 280 e Upgrading to an MSO on page 280 Licensed Options Available Many of the following licensed options can be easily installed without returning the oscilloscope to a Service Center Not all options can be installed on all models See data sheets for details Table 10 Licensed Options Available License Description After purchase model number notes DVM Digital Voltmeter Order DSOXDVM Provides provides 3 digit voltage and 5 digit frequency measurements using any analog channel EDK Educator s Kit Order DSOXEDK Provides training signals on the o
6. 0 1 1 to 1000 1 in a 1 2 5 sequence When measuring current values with a current probe the attenuation factor can be set from 10 V A to 0 001 V A When specifying the attenuation factor in decibels you can select values from 20 dB to 60 dB 60 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Vertical Controls 3 If Amps is chosen as the units and a manual attenuation factor is chosen then the units as well as the attenuation factor are displayed above the Probe softkey Channel 1 Probe Menu 0 005 Units To specify the probe skew When measuring time intervals in the nanoseconds ns range small differences in cable length can affect the measurement Use Skew to remove cable delay errors between any two channels 1 Probe the same point with both probes 2 Press one of the probes associated channel key 3 In the Channel Menu press Probe 4 In the Channel Probe Menu press Skew then select the desired skew value Each analog channel can be adjusted 100 ns in 10 ps increments for a total of 200 ns difference The skew setting is not affected by pressing Default Setup or Auto Scale Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 61 3 Vertical Controls 62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide o7 o ee 4 he Math Waveforms To display math waveforms 63 To perform a transform
7. 4 Press the Cursors key 5 Set the Y2 cursor to the top of the signal and set Y1 to the bottom of the signal Note the AY value at the bottom of the display In this example we are using the Y cursors but you could have used the X cursors instead 6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis Again note the AY value Agilent Acquisition 1 Channels Cursors 0 0V Cursors 2 Figure 4 Phase difference measurements automatic and using cursors 7 Calculate the phase difference using the formula below For example if the first AY value is 1 688 and the second AY value is 1 031 sind segond LY LOL 0 37 65 degrees of phase shift Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 49 2 Horizontal Controls Z Axis Input in XY Display Mode Blanking When you select the XY display mode the time base is turned off Channel 1 is the X axis input channel 2 is the Y axis input and the rear panel EXT TRIG IN is the Z axis input If you only want to see portions of the Y versus X display use the Z axis input Z axis turns the trace on and off analog oscilloscopes called this Z axis blanking because it turned the beam on and off When Z is low lt 1 4 V Y versus X is displayed when Z is high gt 1 4 V the trace is turned off To display the zoomed time base 50 Zoom formerly called Delayed sweep mode is a horizontally expanded version of the nor
8. 48 Z Z axis blanking 50 zoom and pan 44 zoom display measurement window with 188 Zoom key 31 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide
9. Cursors 12 Track Waveform As you move a marker horizontally the vertical amplitude of the waveform is tracked and measured The time and voltage positions are shown for the markers The vertical Y and horizontal X differences between the markers are shown as AX and AY values Binary Logic levels of displayed waveforms at the current X1 and X2 cursor positions are displayed above the softkeys in binary The display is color coded to match the color of the related channel s waveform Cursors Menu Mode Source Cursors Units Binary X1 D Hex Logic levels of displayed waveforms at the current X1 and X2 cursor positions are displayed above the softkeys in hexadecimal Cursors Menu Mode Source Cursors Hex X1 Manual and Track Waveform modes can be used on waveforms that are displayed on the analog input channels including math functions Binary and Hex modes apply to digital signals of MSO oscilloscope models In Hex and Binary modes a level can be displayed as 1 higher than trigger level 0 lower than trigger level indeterminate state ct or X don t care In Binary mode X is displayed if the channel is turned off In Hex mode the channel is interpreted as 0 if turned off 4 Press Source or X1 Source X2 Source in the Track Waveform mode then select the input source for cursor values 5 Select the cursor s to be adjusted Push the Cursors knob then turn the Cursors knob To finaliz
10. MSO models Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 An MSO lets you debug your mixed signal designs using analog signals and tightly correlated digital signals simultaneously The 8 digital channels have a 1 GSa s sample rate with a 50 MHz toggle rate e 8 5 inch WVGA display e Interleaved 2 GSa s or non interleaved 1 GSa s sample rate e Interleaved 100 Kpts or non interleaved 50 Kpts MegaZoom IV memory for the fastest waveform update rates uncompromised e All knobs are pushable for making quick selections e Trigger types edge pulse width pattern and video e Math waveforms add subtract multiply and FFT e Reference waveforms 2 for comparing with other channel or math waveforms e Many built in measurements e Built in license enabled waveform generator with sine square ramp pulse DC noise e USB ports make printing saving and sharing data easy e Optional LAN VGA module for connecting to a network and displaying the screen on a different monitor e Optional GPIB module e A Quick Help system is built into the oscilloscope Press and hold any key to display Quick Help Complete instructions for using the quick help system are given in Access the Built In Quick Help on page 40 For more information about InfiniiVision oscilloscopes see www agilent com find scope Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide In This Guide This guide shows how t
11. on page 25 Some user settings are not changed when recalling the default setup Factory Default recalls the oscilloscope s factory default settings You must confirm the recall because there are no user settings that are left unchanged Performing a Secure Erase 1 Press the Save Recall key 2 In the Save Recall Menu press Default Erase 3 In the Default menu press Secure Erase This performs a secure erase of all non volatile memory in compliance with National Industrial Security Program Operation Manual NISPOM Chapter 8 requirements You must confirm the secure erase and the oscilloscope will reboot when finished Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 229 17 Save Recall Setups Screens Data 230 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide eo o e ee 18 o9 Print Screens To print the oscilloscope s display 231 To set up network printer connections 232 To specify the print options 234 To specify the palette option 235 You can print the complete display including the status line and softkeys to a USB printer or a network printer when the DSOXLAN LAN VGA module is installed The Print Configuration Menu is displayed when you press the Print key The print option softkeys and the Press to Print softkey are ghosted not available until a printer is connected To print the oscill
12. turn the Entry knob to select Pattern 3 For each analog or digital channel you want to include in the desired 118 pattern press the Channel softkey to select the channel This is the channel source for the 0 1 X or edge condition As you press the Channel softkey or rotate the Entry knob the channel you select is highlighted in the Pattern line directly above the softkeys and in the upper right corner of the display next to Pat Adjust the trigger level for the selected analog channel by turning the Trigger Level knob Press the Digital key and select Thresholds to set the threshold level for digital channels The value of the trigger level or digital threshold is displayed in the upper right corner of the display For each channel you select press the Pattern softkey then turn the Entry knob to set the condition for that channel in the pattern Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 100 05 2 05 Agilent Acquisition 1 Channels Pattern al Pattern 1 Trigger Pattern 0 0 sets the pattern to zero low on the selected channel A low is a voltage level that is less than the channel s trigger level or threshold level 1 sets the pattern to 1 high on the selected channel A high is a voltage level that is greater than the channel s trigger level or threshold level X sets the pattern to don t care on the selected channel Any channel set to d
13. using waveform math 63 MATLAB binary data 282 Maximum measurement 174 maximum sample rate 149 Meas key 33 169 Measure controls 33 measure Quick Measure All 253 measurement category definitions 274 measurement definitions 171 measurement record 224 measurement thresholds 186 measurement window with zoom display 188 easurements 171 easurements automatic 169 easurements delay 172 easurements overshoot 172 easurements phase 172 easurements preshoot 172 easurements time 181 easurements voltage 173 MegaZoom IV 4 memory depth and sample rate 149 memory segmented 155 menu line 39 Minimum measurement 174 Mode Coupling key trigger 135 model number 251 257 Modify softkey 240 modulation waveform generator output 210 module installed 251 module slot 37 MSO 3 MSO feature upgrade 280 MSO license 279 Multicast DNS 239 Multiply math function 67 m m m m m m m m navigate files 241 Navigate key 31 navigating the time base 53 network configuration parameters 257 network printer connection 232 network connecting to 239 new label 109 294 noise rejection 139 noise waveform generator output 207 noise adding to waveform generator output 210 noise high frequency 139 noise low frequency 138 noisy signals 135 non volatile memory secure erase 229 normal acquire mode 150 normal mode 149 150 Normal trigger mod
14. you must place the oscilloscope on the network and set up its LAN connection Accessing the Web Interface To access the oscilloscope s web interface 1 Connect the oscilloscope to your LAN see To establish a LAN connection on page 239 or establish a point to point connection see Stand alone Point to Point Connection to a PC on page 240 It is possible to use a point to point connection but using a normal LAN connection is the preferred method 2 Type the oscilloscope s hostname or IP address in the web browser The oscilloscope s web interface Welcome Page is displayed 258 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Web Interface 20 Support Products Agilent Site Agilent Technologies LOSSES Welcome to your L 1 Web Enabled Oscilloscope MSO X 2024A Information about this Web Enabled Instrument Instrument MSO X 2024A Oscilloscope Serial Number US50210029 Description Agilent MSO X 2024A US50210029 Hostname a mx2024a 10029 cos agilent com IP Address 130 29 70 169 VISA TCP IP Connect String TCPIPO a mx2024a 10029 INSTR id Advanced information Identification off Oon Use the navigation bar on the left to access your oscilloscope and related information Agilent Technologies Inc 2006 2010 Browser Web Control The web interface s Browser Web Control page gives you access to e The Real Scope Remote Front Panel see Real Scope Remote Front Panel on page 260
15. 19 Utility Settings To check warranty and extended services status To learn the warranty status of your oscilloscope 1 Point your web browser to www agilent com find warrantystatus 2 Enter your product s model number and serial number The system will search for the warranty status of your product and display the results If the system cannot find your product s warranty status select Contact Us and speak with an Agilent Technologies representative To contact Agilent Information on contacting Agilent Technologies can be found at www agilent com find contactus To return the instrument Before shipping the oscilloscope to Agilent Technologies contact your nearest Agilent Technologies sales or service office for additional details Information on contacting Agilent Technologies can be found at www agilent com find contactus 1 Write the following information on a tag and attach it to the oscilloscope Name and address of owner Model number Serial number Description of service required or failure indication 2 Remove accessories from the oscilloscope Only return accessories to Agilent Technologies if they are associated with the failure symptoms 3 Package the oscilloscope You can use the original shipping container or provide your own materials sufficient to protect the instrument during shipping 4 Seal the shipping container securely and mark it FRAGILE 252 Agilent InfiniiVision 2000 X Series Oscilloscopes
16. 2000 X Series Oscilloscopes User s Guide X Cursors Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ee 12 6 Cursors To make cursor measurements 160 Cursor Examples 163 Cursors are horizontal and vertical markers that indicate X axis values and Y axis values on a selected waveform source You can use cursors to make custom voltage time phase or ratio measurements on oscilloscope signals Cursor information is displayed in the right side information area Cursors are not always limited to the visible display If you set a cursor then pan and zoom the waveform until the cursor is off screen its value will not be changed It will still be there when you return to its original location X cursors are vertical dashed lines that adjust horizontally and can be used to measure time s frequency 1 s phase and ratio The X1 cursor is the short dashed vertical line and the X2 cursor is the long dashed vertical line When used with the FFT math function as a source the X cursors indicate frequency In XY horizontal mode the X cursors display channel 1 values Volts or Amps The X1 and X2 cursor values for the selected waveform source are displayed in the softkey menu area The difference between X1 and X2 AX and 1 AX are displayed in the Cursors box in the right side information area a Agilent Technologies 159 12 Cursors Y Cursors Y cursors are horizontal dashed lines that ad
17. 213 16 Waveform Generator 50 008 Agilent Acquisition i Channels Waveform Generator Modulation Menu Modulation Type Waveform FM Freq Symmetry FM v Ramp 10kHz 100 To set up Frequency Shift Keying Modulation FSK In the Waveform Generator Modulation Menu under Wave Gen gt Settings gt Modulation 1 Press the Type softkey and turn the Entry knob to select Frequency Shift Keying Modulation FSK 2 Press the Hop Freq softkey and turn the Entry knob to specify the hop frequency The output frequency shifts between the original carrier frequency and this hop frequency 3 Press the FSK Rate softkey and turn the Entry knob to specify the rate at which the output frequency shifts The FSK rate specifies a digital square wave modulating signal The following screen shows an FSK modulation of a 100 kHz sine wave carrier signal 214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Waveform Generator 16 Agilent Acquisition h Channels Waveform Generator Modulation Menu Modulation Type Hop Freq FSK Rate FSK 20kHz 10kHz To restore waveform generator defaults 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu press the Default Wave Gen softkey The waveform generator factory default settings 1 kHz si
18. 3 V to 0 8 V IRE Institute of Radio Engineers shows vertical grids in IRE units labeled on the left from 40 to 100 IRE The 0 35 V and 0 7 V levels from the mV grid are also shown and labeled at the right When the IRE grid is selected cursor values are also shown in IRE units Cursor values via the remote interface are not in IRE units The mV and IRE grid values are accurate and match Y cursor values when the vertical scaling is 140 mV division and the vertical offset is 245 mV To select the grid type Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Display Settings 7 1 Press Display 2 Press the Grid softkey then turn the Entry knob a to select the grid type To adjust the grid intensity To adjust the display grid graticule intensity 1 Press Display 2 Press the Intensity softkey then turn the Entry knob a to change the intensity of the displayed grid The intensity level is shown in the Intensity softkey and is adjustable from 0 to 100 Each major vertical division in the grid corresponds to the vertical sensitivity shown in the status line at the top of the display Each major horizontal division in the grid corresponds to the time div shown in the status line at the top of the display To freeze the display To freeze the display without stopping running acquisitions you must configure the Quick Action key See Configuring the Quick Action Key on page 253 1
19. Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 209 16 Waveform Generator Softkey logic levels Low level High level 50 ohm High level high Z expected output load expected output load TTL OV 2 5 V TTL compatible 5V CMOS 5 0V OV Not Available 5V CMOS 3 3V OV 2 5 V CMOS 3 3 V compatible CMOS 2 5V OV 2 5V 2 5V ECL 1 7 V 0 8 V ECL compatible 0 9 V To add noise to the waveform generator output 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu press the Add Noise softkey and turn the Entry knob to select the amount of white noise to add to the waveform generator output Note that adding noise affects edge triggering on the waveform generator source see Edge Trigger on page 115 as well as the waveform generator sync pulse output signal which can be sent to TRIG OUT see Setting the Rear Panel TRIG OUT Source on page 246 This is because the trigger comparator is located after the noise source To add modulation to the waveform generator output Modulation is where an original carrier signal is modified according to the amplitude of a second modulating signal The modulation type AM FM or FSK specifies how the carrier signal is modified To enable and set up modulation for the
20. Bus values Analog displayed channel values at cursor Digital Channels 6 Agilent ition 1 4 8 1 4 B B ES Set all Digits B Digital channel values at cursor See Pattern Trigger on page 118 for more information on Pattern triggering Digital channel signal fidelity Probe impedance and grounding When using the mixed signal oscilloscope you may encounter problems that are related to probing These problems fall into two categories probe loading and probe grounding Probe loading problems generally affect the device under test while probe grounding problems affect the accuracy of the data to the measurement instrument The design of the probes minimizes the first problem while the second is easily addressed by good probing practices Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 93 94 Digital Channels Input Impedance The logic probes are passive probes which offer high input impedance and high bandwidths They usually provide some attenuation of the signal to the oscilloscope typically 20 dB Passive probe input impedance is generally specified in terms of a parallel capacitance and resistance The resistance is the sum of the tip resistor value and the input resistance of the test instrument see the following figure The capacitance is the series combination of the tip compensating capacitor and the cable plus instrument capacitance in parallel with the stray tip ca
21. Frequency Modulation FM In the Waveform Generator Modulation Menu under Wave Gen gt Settings gt Modulation 1 Press the Type softkey and turn the Entry knob to select Frequency Modulation FM 2 Press the Waveform softkey and turn the Entry knob to select the shape of the modulating signal Sine Square Ramp Sine Cardinal Exponential Rise Exponential Fall When the Ramp shape is selected a Symmetry softkey appears so that you can specify the amount of time per cycle that the ramp waveform is rising 3 Press the FM Freq softkey and turn the Entry knob to specify the frequency of the modulating signal 4 Press the FM Dev softkey and turn the Entry knob to specify the frequency deviation from the original carrier signal frequency When the modulating signal is at its maximum amplitude the output frequency is the carrier signal frequency plus the deviation amount and when the modulating signal is at its minimum amplitude the output frequency is the carrier signal frequency minus the deviation amount The frequency deviation cannot be greater than the original carrier signal frequency Also the sum of the original carrier signal frequency and the frequency deviation must be less than or equal to the maximum frequency for the selected waveform generator function plus 100 kHz The following screen shows an FM modulation of a 100 kHz sine wave carrier signal Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide
22. Guide 47 2 48 Horizontal Controls XY Time Mode Example The XY time mode converts the oscilloscope from a volts versus time display to a volts versus volts display using two input channels Channel 1 is the X axis input channel 2 is the Y axis input You can use various transducers so the display could show strain versus displacement flow versus pressure volts versus current or voltage versus frequency This exercise shows a common use of the XY display mode by measuring the phase difference between two signals of the same frequency with the Lissajous method 1 Connect a sine wave signal to channel 1 and a sine wave signal of the same frequency but out of phase to channel 2 2 Press the AutoScale key press the Horiz key then press Time Mode and select XY 3 Center the signal on the display with the channel 1 and 2 position knobs Use the channel 1 and 2 volts div knobs and the channel 1 and 2 Fine softkeys to expand the signal for convenient viewing The phase difference angle 0 can be calculated using the following formula assuming the amplitude is the same on both channels sind org Signal must jt becentered in gt Measuring Signals 90 degrees Signals phase difference out of phase in phase Figure 3 XY time mode signals centered on display Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Horizontal Controls 2
23. Math key are selected for the math waveform If the arrow to the left of the Math key is not illuminated press the key 2 Use the multiplexed scale and position knobs just to the right of the Math key to re size and re position the math waveform Math Scale and Offset are Set Automatically Any time the currently displayed math function definition is changed the function is automatically scaled for optimum vertical scale and offset If you manually set scale and offset for a function select a new function then select the original function the original function will be automatically rescaled See Also Units for Math Waveforms on page 65 Units for Math Waveforms Units for each input channel can be set to Volts or Amps using the Units softkey in the channel s Probe Menu Units for math function waveforms are Math function Units add or subtract VorA multiply V2 A or W Volt Amp FFT dB decibels See also FFT Units on page 73 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 65 4 Math Waveforms Math function Units When the FFT source is channel 1 2 3 or 4 FFT units will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 MQ FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 5002 FFT units will be displayed as dB for all other FFT sources or when a source channel s
24. Measurement category III is for measurements performed in the building installation Examples are measurements on distribution boards circuit breakers wiring including cables bus bars junction boxes switches socket outlets in the fixed installation and equipment for industrial use and some other equipment for example stationary motors with permanent connection to the fixed installation Measurement category IV is for measurements performed at the source of the low voltage installation Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units 274 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Reference 21 Transient Withstand Capability CAUTION A Maximum input voltage at analog inputs CAT 300 Vrms 400 Vpk transient overvoltage 1 6 kVpk With 10073C 10 1 probe CAT I 500 Vpk CAT II 400 Vpk With N2862A or N2863A 10 1 probe 300 Vrms CAUTION A Maximum input voltage at digital channels 40 V peak CAT l transient overvoltage 800 Vpk Environmental Conditions Environment Indoor use only Ambient Operating 0 C to 55 C non operating 40 C to 71 C temperature Humidity Operating Up to 80 RH at or below 40 C Up to 45 RH up to 50 C Non operating Up to 95 RH up to 40 C Up to 45 RH up to 50 C Altitude Operating and non operating to 4 000 m 13 123 ft Overvoltage This product is intended to be
25. N Cycles measurement 178 averaging acquire mode 149 153 Back Up key 29 bandwidth 251 bandwidth limit 58 bandwidth required oscilloscope 148 bandwidth oscilloscope 146 Base measurement 176 BIN file format 218 binary data bin 281 binary data file examples 285 binary data in MATLAB 282 binary data example program for reading 284 Blackman Harris FFT window 70 blanking 50 BMP file format 218 brick wall frequency response 146 brightness of waveforms 29 Browser Web Control 259 260 261 262 263 Browser Based Remote Front Panel 262 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide built in help 40 burst capture signal bursts 155 bus display mode 90 buttons keys front panel 28 BW Limit in DVM display 202 C calibration 248 calibration protect button 37 38 calibration status 269 capture signal bursts 155 Center FFT 69 channel labels 107 channel analog 55 channel bandwidth limit 58 channel coupling 57 channel invert 59 channel on off keys 34 channel position 57 channel probe units 60 channel skew 61 channel vernier 58 channel vertical sensitivity 57 characteristics 273 choosing values 30 cleaning 251 clear display 152 clear display Quick Clear Display 253 clear persistence 104 clock 246 CMOS threshold 89 compensate passive probes 27 34 Config softkey 239 240 connect probes digital 81 connection t
26. Noise rises above zero volts or the DC offset value DC N A Noise N A 208 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Waveform Generator 16 To specify the expected output load 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu press the Out Load softkey and turn the Entry knob to select 500 High Z The output impedance of the Gen Out BNC is fixed at 50 ohms However the output load selection lets the waveform generator display the correct amplitude and offset levels for the expected output load If the actual load impedance is different than the selected value the displayed amplitude and offset levels will be incorrect To use waveform generator logic presets With logic level presets you can easily set the output voltage to TTL CMOS 5 0V CMOS 3 3V CMOS 2 5V or ECL compatible Low and High levels 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey In the Waveform Generator Settings Menu press the Logic Presets softkey 4 In the Waveform Generator Logic Level Presets Menu press one of the softkeys to set the generated signal s Low and High voltages to logic compatible levels
27. Once the Quick Action key has been configured press it to freeze the display 2 To un freeze the display press Quick Action again Manual cursors can be used on the frozen display Many activities such as adjusting the trigger level adjusting vertical or horizontal settings or saving data will un freeze the display Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 105 7 Display Settings 106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Labels J e e i To turn the label display on or off 107 e To assign a predefined label toa channel 108 To define anew label 109 To load a list of labels from a text file you create 110 To reset the label library to the factory default 111 You can define labels and assign them to each analog input channel or you can turn labels off to increase the waveform display area Labels can also be applied to digital channels on MSO models To turn the label display on or off 1 Press the Label key on the front panel This turns on labels for the displayed analog and digital channels Labels are displayed at the left edge of the displayed traces The figure below shows an example of displayed labels iE Agilent Technologies 107 8 Labels Apply New Character Label 2 To turn the labels off press the Label key again To assign a predefined label to a channel 1 Pres
28. Only displayed data points are saved The maximum number of data points depends on these things Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 223 17 Save Recall Setups Screens Data e Whether acquisitions are running When stopped data comes from the raw acquisition record When running data comes from the smaller measurement record e Whether the oscilloscope was stopped using Stop or Single Running acquisitions split memory to provide fast waveform update rates Single acquisitions use full memory e Whether only one channel of a pair is turned on Channels 1 and 2 are one pair channels 3 and 4 are the other Acquisition memory is divided among the channels in a pair e Whether reference waveforms are on Displayed reference waveforms consume acquisition memory e Whether digital channels are on Displayed digital channels consume acquisition memory e Whether segmented memory is on Acquisition memory is divided by the number of segments e The horizontal time div sweep speed setting At faster settings fewer data points appear on the display e When saving to a CSV format file the maximum number of data points is 50 000 When necessary the Length control performs a 1 of n decimation of the data For example if the Length is set to 1000 and you are displaying a record that is 5000 data points in length four of each five data points will be decimated creating an output file 1000
29. USB compliant printer For more information about printing see Chapter 18 Print Screens starting on page 231 You can also use the USB port to update the oscilloscope s system software when updates are available You do not need to take special precautions before removing the USB mass storage device from the oscilloscope you do not need to eject it Simply unplug the USB mass storage device from the oscilloscope when the file operation is complete caution Do not connect a host computer to the oscilloscope s USB host port Use the device port A host computer sees the oscilloscope as a device so connect the host computer to the oscilloscope s device port on the rear panel See 1 0 Interface Settings on page 237 There is a second USB host port on the back panel Connect the digital probe cable to this connector MSO models only See Chapter 6 Digital Channels starting on page 81 21 Waveform generator output Outputs sine square ramp pulse DC or noise on the Gen Out BNC Press the Wave Gen key to set up the waveform generator See Chapter 16 Waveform Generator starting on page 205 Front Panel Overlays for Different Languages Front panel overlays which have translations for the English front panel keys and label text are available in 10 languages The appropriate overlay is included when the localization option is chosen at time of purchase To install a front pane
30. X Series Oscilloscopes User s Guide 121 122 Triggers For digital channels press the Digital key and select Thresholds to set the threshold level The value of the trigger level or digital threshold is displayed in the upper right corner of the display Press the pulse polarity softkey to select positive Ib or negative dD polarity for the pulse width you want to capture The selected pulse polarity is displayed in the upper right corner of the display A positive pulse is higher than the current trigger level or threshold and a negative pulse is lower than the current trigger level or threshold When triggering on a positive pulse the trigger will occur on the high to low transition of the pulse if the qualifying condition is true When triggering on a negative pulse the trigger will occur on the low to high transition of the pulse if the qualifying condition is true Press the qualifier softkey lt gt gt lt to select the time qualifier The Qualifier softkey can set the oscilloscope to trigger on a pulse width that is Less than a time value lt For example for a positive pulse if you set t lt 10 ns H 10 ns 10 ns gt Trigger Greater than a time value gt For example for a positive pulse if you set t gt 10 ns 10 ns 4 1 10 ns __ T igger Within a range of time values gt lt For example for a positive pulse if you set t gt 10 ns
31. X cursor For more information on the use of FFTs please refer to Agilent Application Note 243 The Fundamentals of Signal Analysis at http cp literature agilent com litweb pdf 5952 8898E pdf Additional information can be obtained from Chapter 4 of the book Spectrum and Network Measurements by Robert A Witte FFT Units 0 dBV is the amplitude of a 1 Vrms sinusoid When the FFT source is channel 1 or channel 2 or channel 3 or 4 on 4 channel models FFT units will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 MQ FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50Q FFT units will be displayed as dB for all other FFT sources or when a source channel s units has been set to Amps FFT DC Value The FFT computation produces a DC value that is incorrect It does not take the offset at center screen into account The DC value is not corrected in order to accurately represent frequency components near DC FFT Aliasing When using FFTs it is important to be aware of frequency aliasing This requires that the operator have some knowledge as to what the frequency domain should contain and also consider the sampling rate frequency Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 73 4 74 Math Waveforms span and oscilloscope vertical bandwidth when making FFT measurements The FFT resolution the quotient of the samp
32. a background Disable the Transparent mode to show them with a gray background The Transparent setting affects mask test statistics measurement statistics and reference waveform information display Clear Display Clears acquisition data from the oscilloscope display To manually modify a mask file You can manually modify a mask file that you created using the Automask function 1 ao oO FF O N 8 9 Follow the steps 1 7 in To create a mask from a golden waveform Automask on page 189 Do not clear the mask after creating it Attach a USB mass storage device to the oscilloscope Press the Save Recall key Press the Save softkey Press the Format softkey and select Mask Press the second softkey and select a destination folder on your USB mass storage device Press the Press to Save softkey This creates an ASCII text file that describes the mask Remove the USB mass storage device and connect it to a PC Open the msk file your created using a text editor such as Wordpad 10 Edit save and close the file The mask file contains the following sections Mask File Identifier Mask Title Mask Violation Regions Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 195 14 Mask Testing Mask File Identifier Mask Title Mask Violation 196 Regions e Oscilloscope Setup Information The Mask File Identifier is MASK_FILE_548XxX The Mask Title is a string of AS
33. and ensure the desired analog channel is selected Automask Menu Source Units Create Divisions 0 10 0 10 Mask Adjust the mask s horizontal tolerance Y and vertical tolerance X These are adjustable in graticule divisions or in absolute units volts or seconds selectable using the Units softkey Press the Create Mask softkey The mask is created and testing begins Whenever the Create Mask softkey is pressed the old mask is erased and a new mask is created Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Mask Testing 14 7 750 Channels 9 To clear the mask and switch off mask testing press the Back Up key to return to the Mask Test Menu then press the Clear Mask softkey If infinite persistence display mode see To set or clear persistence on page 103 is on when mask test is enabled it stays on If infinite persistence is off when mask test is enabled it is switched on when mask test is switched on then infinite persistence is switched off when mask test is switched off Troubleshooting If you press Create Mask and the mask appears to cover the entire screen Mask Setup check the Y and X settings in the Automask Menu If these are set to zero the resulting mask will be extremely tight around the waveform If you press Create Mask and it appears that no mask was created check the Y and X settings They may be set so large that the mask is not visible Mask Test Setup O
34. and post trigger information available depends on the time div selected and memory depth The horizontal position knob works in the Normal time mode while acquisitions are running or when they are stopped When running adjusting the horizontal scale knob changes the sample rate When Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 45 2 Horizontal Controls stopped adjusting the horizontal scale knob lets you zoom into acquired data See Panning and Zooming Single or Stopped Acquisitions on page 46 Note that the horizontal position knob has a different purpose in the Zoom display See To display the zoomed time base on page 50 Panning and Zooming Single or Stopped Acquisitions 46 When the oscilloscope is stopped use the horizontal scale and position knobs to pan and zoom your waveform The stopped display may contain several acquisitions worth of information but only the last acquisition is available for pan and zoom The ability to pan move horizontally and scale expand or compress horizontally an acquired waveform is important because of the additional insight it can reveal about the captured waveform This additional insight is often gained from seeing the waveform at different levels of abstraction You may want to view both the big picture and the specific little picture details The ability to examine waveform detail after the waveform has been acquired is a benefit generally associated wit
35. and t lt 15 ns Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Pulse width trigger lt qualifier time set softkey Pulse width trigger gt qualifier time set softkey Video Trigger Triggers 9 10 ns H 15 ns y p 12 ns Trigger J Select the qualifier time set softkey lt or gt then rotate the Entry knob to set the pulse width qualifier time The qualifiers can be set as follows 2 ns to 10 s for gt or lt qualifier 5 ns to 10 s for 350 MHz bandwidth models 10 ns to 10 s for gt lt qualifier with minimum difference of 5 ns between upper and lower settings When the less than lt qualifier is selected the Entry knob sets the oscilloscope to trigger on a pulse width less than the time value displayed on the softkey When the time range gt lt qualifier is selected the Entry knob sets the upper time range value When the greater than gt qualifier is selected the Entry knob sets the oscilloscope to trigger on a pulse width greater than the time value displayed on the softkey When the time range gt lt qualifier is selected the Entry knob sets the lower time range value Video triggering can be used to capture the complicated waveforms of most standard analog video signals The trigger circuitry detects the vertical and horizontal interval of the waveform and produces triggers based on the video trigger settings you have selecte
36. are defined in the Oscilloscope Setup portion of the mask file The Y values normally voltage are normalized in the file using the following equation Ynorm Y YD AY where AY Y2 Y1 To convert the normalized Y values in the mask file to voltage Y Cm Du where AY Y2 Y1 The X values normally time are normalized in the file using the following equation Xnorm X X1 AX To convert the normalized X values to time X Xnorm AX X1 The keywords setup and end_setup appearing alone on a line define the beginning and end of the oscilloscope setup region of the mask file The oscilloscope setup information contains remote programming language commands that the oscilloscope executes when the mask file is loaded Any legal remote programming command can be entered in this section Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 197 14 Mask Testing The mask scaling controls how the normalized vectors are interpreted This in turn controls how the mask is drawn on the display The remote programming commands that control mask scaling are MTES MTES MTES MTES MTES SCAI SCAI SCAI SCAI SCAI Building a Mask File 198 Lis Li G Li Li BIND 0 X1 400 000E 06 XDEL 800 000E 06 Y 1 359 000E 03 Y2 2 35900E 00 The following mask uses all eight mask regions The most difficult part of creating a mask file is normalizing the X and Y values from the
37. display The time base is turned off Channel 1 amplitude is plotted on the X axis and Channel 2 amplitude is plotted on the Y axis You can use XY mode to compare frequency and phase relationships between two signals XY mode can also be used with transducers to display strain versus displacement flow versus pressure volts versus current or voltage versus frequency Use the cursors to make measurements on XY mode waveforms For more information about using XY mode for measurements refer to XY Time Mode on page 48 Roll causes the waveform to move slowly across the screen from right to left It only operates on time base settings of 50 ms div and slower If the current time base setting is faster than the 50 ms div limit it will be set to 50 ms div when Roll mode is entered In Roll mode there is no trigger The fixed reference point on the screen is the right edge of the screen and refers to the current moment in time Events that have occurred are scrolled to the left of the reference point Since there is no trigger no pre trigger information is available If you would like to pause the display in Roll mode press the Single key To clear the display and restart an acquisition in Roll mode press the Single key again Use Roll mode on low frequency waveforms to yield a display much like a strip chart recorder It allows the waveform to roll across the display Agilent InfiniiVision 2000 X Series Oscilloscopes User s
38. display to the right each time a particular trigger condition is met This provides stable display of periodic signals such as sine waves and square waves as well as nonperiodic signals such as serial data streams The figure below shows the conceptual representation of acquisition memory You can think of the trigger event as dividing acquisition memory into a pre trigger and post trigger buffer The position of the trigger event in acquisition memory is defined by the time reference point and the delay horizontal position setting see To adjust the horizontal delay position on page 45 Trigger Event Pre Trigger Buffer Post Trigger Buffer J DA Acquisition Memory Adjusting the Trigger Level 114 You can adjust the trigger level for a selected analog channel by turning the Trigger Level knob You can push the Trigger Level knob to set the level to the waveform s 50 value If AC coupling is used pushing the Trigger Level knob sets the trigger level to about 0 V The position of the trigger level for the analog channel is indicated by the trigger level icon Tp if the analog channel is on at the far left side of the display The value of the analog channel trigger level is displayed in the upper right corner of the display Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 The trigger level for a selected digital channel is set using the threshold menu in the Digital Channel Menu P
39. four would mean every waveform data point in the waveform record has been averaged at least four times The default value is 0 X Display Range A 32 bit float that is the X axis duration of the waveform that is displayed For time domain waveforms it is the duration of time across the display If the value is zero then no data has been acquired X Display Origin A 64 bit double that is the X axis value at the left edge of the display For time domain waveforms it is the time at the start of the display This value is treated as a double precision 64 bit floating point number If the value is zero then no data has been acquired X Increment A 64 bit double that is the duration between data points on the X axis For time domain waveforms this is the time between points If the value is zero then no data has been acquired X Origin A 64 bit double that is the X axis value of the first data point in the data record For time domain waveforms it is the time of the first point This value is treated as a double precision 64 bit floating point number If the value is zero then no data has been acquired X Units A 32 bit integer that identifies the unit of measure for X values in the acquired data 0 Unknown e 1 Volts 2 Seconds e 3 Constant 4 Amps e 5 dB 6 Hz Y Units A 32 bit integer that identifies the unit of measure for Y values in the acquired data The possible values are listed above unde
40. frequency Span to 500 kHz and window to Hanning Agilent on Channels Measurements Freq 1 74 63kHz Pk Pk 1 4 30V tit FFT Ch1 Scale 20dB Hfset 60 OO0dBY FFT Resolution 763Hz Window Vertical Units Hanning Decibels Auto Setup See Also To perform a transform function on an arithmetic operation on page 64 e FFT Measurement Hints on page 72 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 71 4 72 Math Waveforms e FFT Units on page 73 e FFT DC Value on page 73 e FFT Aliasing on page 73 e FFT Spectral Leakage on page 75 e Units for Math Waveforms on page 65 FFT Measurement Hints The number of points acquired for the FFT record can be up to 65 536 and when frequency span is at maximum all points are displayed Once the FFT spectrum is displayed the frequency span and center frequency controls are used much like the controls of a spectrum analyzer to examine the frequency of interest in greater detail Place the desired part of the waveform at the center of the screen and decrease frequency span to increase the display resolution As frequency span is decreased the number of points shown is reduced and the display is magnified While the FFT spectrum is displayed use the Math and Cursors keys to switch between measurement functions and frequency domain controls in FFT Menu FFT Resolution The FFT resolution is the quotient of the sampling rate and th
41. front panel of the mixed signal oscilloscope The digital probe cable is keyed so you can connect it only one way You do not need to power off the oscilloscope CAUTION A Probe cable for digital channels Use only the Agilent logic probe and accessory kit supplied with the mixed signal oscilloscope see Accessories Available on page 278 3 Connect the ground lead on each set of channels each pod using a probe grabber The ground lead improves signal fidelity to the oscilloscope ensuring accurate measurements 82 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 Channel Pod Ground Circuit Ground 4 Connect a grabber to one of the probe leads Other probe leads are omitted from the figure for clarity Grabber 5 Connect the grabber to a node in the circuit you want to test Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 83 6 Digital Channels 6 For high speed signals connect a ground lead to the probe lead connect a grabber to the ground lead and attach the grabber to ground in the device under test Ground Grabber 7 Repeat these steps until you have connected all points of interest Signals Ground 84 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 Acquiring waveforms using the digital channels When you press Run Stop or Single to run the oscilloscope the oscilloscope examines th
42. function on an arithmetic operation 64 To adjust the math waveform scale and offset 65 Units for Math Waveforms 65 Math Operators 66 Math Transforms 68 Math functions can be performed on analog channels The resulting math waveform is displayed in light purple You can use a math function on a channel even if you choose not to display the channel on screen You can e Perform an arithmetic operation like add subtract or multiply on analog input channels e Perform a transform function like FFT on an analog input channel e Perform a transform function on the result of an arithmetic operation To display math waveforms 1 Press the Math key on the front panel to display the Waveform Math Menu Waveform Math Menu tit Ch1 Ch2 Operator Source 1 Source 2 F a Agilent Technologies 25 4 Math Waveforms TIP 2 If f t is not already shown on the Function softkey press the Function sofkey and select f t Displayed 3 Use the Operator softkey to select an operator or transform For more information on the operators or transforms see Math Operators on page 66 Math Transforms on page 68 4 Use the Source 1 softkey to select the analog channel on which to perform math You can rotate the Entry knob or repetitively press the Source 1 softkey to make your selection If you choose a transform function FFT the result is displayed 5 If you select an arithmetic operator use the Source2 softkey to
43. gt AutoScale 2 In the AutoScale Preferences Menu you can Press the Fast Debug softkey enable disable this type of autoscale When fast debug is enabled autoscale lets you make quick visual comparisons to determine whether the signal being probed is a DC voltage ground or an active AC signal Channel coupling is maintained to allow easy viewing of oscillating signals Press the Channels softkey and turn the Entry knob to specify the channels to be autoscaled All Channels The next time you press AutoScale all channels that meet the requirements of AutoScale will be displayed Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 245 19 Utility Settings Only Displayed Channels The next time you press AutoScale only the channels that are turned on will be examined for signal activity This is useful if you only want to view specific active channels after pressing AutoScale Press the Acq Mode softkey and turn the Entry knob to select whether the acquisition mode should be preserved during autoscale Normal to make the oscilloscope switch to Normal acquisition mode whenever the AutoScale key is pressed This is the default mode Preserve to make the oscilloscope remain in the acquisition mode you have chosen when the AutoScale key is pressed Setting the Oscilloscope s Clock The Clock Menu lets you set the current date and time of day 24 hour format This time date stamp will appear
44. knob to select Edge 3 Select the trigger source Analog channel 1 to the number of channels Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 115 9 116 Triggers Digital channel on mixed signal oscilloscopes DO to the number of digital channels minus one External triggers on the rear panel EXT TRIG IN signal Line triggers at the 50 level of the rising or falling edge of the AC power source signal WaveGen triggers at the 50 level of the rising edge of the waveform generator output signal Not available when the DC or Noise waveforms are selected You can choose a channel that is turned off not displayed as the source for the edge trigger The selected trigger source is indicated in the upper right corner of the display next to the slope symbol 1 through 4 analog channels DO through Dn digital channels E External trigger input L Line trigger W Waveform generator Press the Slope softkey and select rising edge falling edge alternating edges or either edge depending on the selected source The selected slope is displayed in the upper right corner of the display Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 200 08 271V Agilent on Channels Alternating edge mode is useful when you want to trigger on both edges of a clock for example DDR signals Either edge mode is useful when you want to trigger on any activi
45. on hardcopy prints and directory information on the USB mass storage device To set the date and time or to view the current date and time 1 Press Utility gt Options gt Clock Clock Menu Year y Minute 2010 3 52 2 Press the Year Month Day Hour or Minute softkey then rotate the Entry knob to set to the desired number The hours are shown in the 24 hour format So 1 00 PM is hour 13 The real time clock only allows selection of valid dates If a day is selected and the month or year is changed so the day is invalid the day is automatically adjusted Setting the Rear Panel TRIG OUT Source 246 You can choose the source of the TRIG OUT connector on the rear panel of the oscilloscope Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 1 Press Utility gt Options gt Rear Panel 2 In the Rear Panel Menu press Trig Out then turn the Entry knob to select from Triggers Each time the oscilloscope triggers a rising edge occurs on TRIG OUT The rising edge is delayed 30 ns from the oscilloscope s trigger point The output level is 0 5 V into an open circuit and 0 2 5 V into 50 Q See Chapter 9 Triggers starting on page 113 Mask The pass fail status is evaluated periodically When the evaluation of the testing period results in a failure the trigger output pulses high 5 V Otherwise the trigger output remains at low 0 V See Chapter 14 Mask Testin
46. on the back until it is fully seated The module s spring tabs will latch into the slot keeping the module in place E LAN VGA Module ni i DEVICE oe E Module Slot GPIB Module The LAN VGA or GPIB module must be installed before powering on the oscilloscope Tilt the Oscilloscope for Easy Viewing There are tabs under the oscilloscope s front feet that can be flipped out to tilt the oscilloscope 22 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 Power On the Oscilloscope Power Requirements Ventilation Requirements To power on the oscilloscope Line voltage frequency and power e Line 100 120 Vac 50 60 400 Hz e 100 240 Vac 50 60 Hz e 100 W max The air intake and exhaust areas must be free from obstructions Unrestricted air flow is required for proper cooling Always ensure that the air intake and exhaust areas are free from obstructions The fan draws air in from the left side and bottom of the oscilloscope and pushes it out behind the oscilloscope When using the oscilloscope in a bench top setting provide at least 2 clearance at the sides and 4 100 mm clearance above and behind the oscilloscope for proper cooling 1 Connect the power cord to the rear of the oscilloscope then to a suitable AC voltage source Route the power cor
47. samples the oscilloscope will take 100 us x 4 GSa s 400K samples The oscilloscope is required to display those 400K samples using 640 pixel columns The oscilloscope will decimate the 400K samples to 640 pixel columns and this decimation keeps track of the min and max values of all the points that are represented by any given column Those min and max values will be displayed in that screen column A similar process is used to reduce the acquired data to produce a record usable for various analysis needs such as measurements and CSV data This analysis record or measurement record is much larger than 640 and may in fact contain up to 65536 points Still once the of acquired points gt 65536 some form of decimation is required The decimator used to produce a CSV record is configured to provide a best estimate of all the samples that each point in the record represents Therefore the min and max values may not appear in the CSV file 288 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Reference 21 Acknowledgements RealVNC HDF5 CUPS mDNSResponder RealVNC is licensed under the GNU General Public License Copyright C 2002 2005 RealVNC Ltd All Rights Reserved This is free software you can redistribute it and or modify it under the terms of the GNU General Public License as published by the Free Software Foundation either version 2 of the License or at your option any later version This soft
48. setup files on page 227 Mask msk See To recall mask files on page 228 Reference Waveform data h5 See To recall reference waveform files from a USB storage device on page 228 You can also recall setups and mask files by loading them using the File Explorer See File Explorer on page 241 You can also configure the Quick Action key to recall setups masks or reference waveforms See Configuring the Quick Action Key on page 253 To recall setup files Setup files can be recalled from one of 10 internal Agilent Flash locations or from an external USB storage device 1 Press Save Recall gt Recall gt Recall then turn the Entry knob to select Setup scp Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 227 17 Save Recall Setups Screens Data 228 2 Press the softkey in the second position and use the Entry knob to navigate to the file to recall See To navigate storage locations on page 226 3 Press the Press to Recall softkey A message indicating whether the recall was successful is displayed 4 If you would like to clear the display press Clear Display To recall mask files Mask files can be recalled from one of four internal Agilent Flash locations or from an external USB storage device 1 Press Save Recall gt Recall gt Recall then turn the Entry knob to select Mask msk Press the softkey in the second position and use the Entry kn
49. show what portion of the waveform is being measured The Y cursor shows the middle threshold point e To isolate an event for frequency measurement on page 182 To isolate an event for frequency measurement The following figure shows how to use Zoom mode to isolate an event for a frequency measurement You may need to change the measurement window setting so that the measurement is made in the lower Zoom window See Measurement Window with Zoom Display on page 188 If the waveform is clipped it may not be possible to make the measurement 50 005 1 00027 Auto 1 22V Agilent sition Channels x Measurements Freq 1 9 60kHz Measurement Menu Source ype Clear Meas Figure 35 Isolating an event for Frequency measurement Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 Width Width is the time from the middle threshold of the rising edge to the middle threshold of the next falling edge The X cursors show the pulse being measured The Y cursor shows the middle threshold point Width Width is the time from the middle threshold of the falling edge to the middle threshold of the next rising edge The X cursors show the pulse being measured The Y cursor shows the middle threshold point Duty Cycle The duty cycle of a repetitive pulse train is the ratio of the positive pulse width to the period expressed as a percentage The X cursors show the time period bein
50. signal that you would like to trigger on The Video trigger modes available are Field1 and Field2 Trigger on the rising edge of the first serration pulse of field 1 or field 2 interlaced standards only All Fields Trigger on the rising edge of the first pulse in the vertical sync interval All Lines Trigger on all horizontal sync pulses Line Field1 and Line Field2 Trigger on the selected line in field 1 or field 2 interlaced standards only Line Alternate Alternately trigger on the selected line in field 1 and field 2 NTSC PAL PAL M and SECAM only 9 If you select a line mode press the Line softkey then rotate the Entry knob to select the line number on which you want to trigger The following table lists the line or count numbers per field for each video standard Video standard Field 1 Field 2 Alt Field NTSC 1 to 263 1 to 262 1 to 262 PAL 1 to 313 314 to 625 1 to 312 PAL M 1 to 263 264 to 525 1 to 262 SECAM 1 to 313 314 to 625 1 to 312 The following are exercises to familiarize you with video triggering These exercises use the NTSC video standard e To trigger on a specific line of video on page 127 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 e To trigger on all sync pulses on page 128 e To trigger on a specific field of the video signal on page 129 e To trigger on all fields of the video signal on page 1
51. the expanded portion of the normal window around the glitch Averaging Acquisition Mode The Averaging mode lets you average multiple acquisitions together to reduce noise and increase vertical resolution at all time div settings Averaging requires a stable trigger The number of averages can be set from 2 to 65536 in power of 2 increments A higher number of averages reduces noise more and increases vertical resolution Avgs Bits of resolution 2 8 4 9 16 10 64 11 gt 256 12 The higher the number of averages the slower the displayed waveform responds to waveform changes You must compromise between how quickly the waveform responds to changes and how much you want to reduce the displayed noise on the signal To use the Averaging mode 1 Press the Acquire key then press the Acq Mode softkey until the Averaging mode is selected 2 Press the Avgs softkey and turn the Entry knob to set the number of averages that best eliminates the noise from the displayed waveform The number of acquisitions being averaged is displayed in the Avgs softkey Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 153 11 Acquisition Control 1469 Agilent Acquisition No 25 0h Channels DC 10 0 1 DC 10 0 1 cq Mode Segmented Normal g Figure 26 Random noise on the displayed waveform 14 Agilent Acquisition 1 Channels DC 10 0 1 DC 10 0 1 Acquir Acq Mode D Avgs Seg
52. the frequency span goes from 0 to the Nyquist frequency the best way to prevent aliasing is to make sure that the frequency span is greater than the frequencies of significant energy present in the input signal FFT Spectral Leakage The FFT operation assumes that the time record repeats Unless there is an integral number of cycles of the sampled waveform in the record a discontinuity is created at the end of the record This is referred to as leakage In order to minimize spectral leakage windows that approach zero smoothly at the beginning and end of the signal are employed as filters to the FFT The FFT Menu provides four windows Hanning Flat Top Rectangular and Blackman Harris For more information on leakage see Agilent Application Note 243 The Fundamentals of Signal Analysis at http cp literature agilent com litweb pdf 5952 8898E pdf Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 75 4 Math Waveforms 76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide eo o ee 5 a ni a Reference Waveforms e e j To save a waveform to a reference waveform location 77 e To display a reference waveform 78 To scale and position reference waveforms 79 To adjust reference waveform skew 79 To display reference waveform information 80 To save recall reference waveform files to from a USB storage device 80 Analo
53. the left side Demo 2 Probe Comp terminal How AutoScale Auto Scale analyzes any waveforms present at each channel and at the Works external trigger input This includes the digital channels if connected Auto Scale finds turns on and scales any channel with a repetitive waveform that has a frequency of at least 25 Hz a duty cycle greater than 0 5 and an amplitude of at least 10 mV peak to peak Any channels that do not meet these requirements are turned off The trigger source is selected by looking for the first valid waveform starting with external trigger then continuing with the lowest number analog channel up to the highest number analog channel and finally if digital probes are connected the highest number digital channel During Auto Scale the delay is set to 0 0 seconds the horizontal time div sweep speed setting is a function of the input signal about 2 periods of the triggered signal on the screen and the triggering mode is set to Edge Compensate Passive Probes Each oscilloscope passive probe must be compensated to match the input characteristics of the oscilloscope channel to which it is connected A poorly compensated probe can introduce significant measurement errors 1 Input the Probe Comp signal see Input a Waveform on page 25 2 Press Default Setup to recall the default oscilloscope setup see Recall the Default Oscilloscope Setup on page 25 3 Press Auto Scale to automatically configu
54. the softkey menu area To save a screen image showing the Save Recall Menu at the bottom press the Save Recall key twice before saving the image You can also save the oscilloscope s display image using a web browser See Get Image on page 267 for details See Also Adding an Annotation on page 254 To save CSV ASCII XY or BIN data files Data files can be saved to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select CSV data csv ASCII XY data csv or Binary data bin 220 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 17 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 226 Press the Settings softkey In the File Settings Menu you have these softkeys and options Setup Info when enabled setup information vertical horizontal trigger acquisition math and display settings is also saved in a separate file with a TXT extension Length sets the number of data points that will be output to the file For more information see Length Control on page 223 Save Seg when data is acquired to segmented memory you can specify whether the currently displayed segment is saved or all acquired segments are saved See also Saving Data from Segmented Memory on page 158 Finally press the Pr
55. time and voltage values This example shows an easy way to convert voltage and time to normalized X and Y values in the mask file Agilent Acquisition ormal Channels BW Limit Invert The following mask file produced the mask shown above MASK_FILE_548XX All Regions Region Number 1 Number of vertices 4 12 50 10 00 MAX 1 750 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 00 1 750 12 50 MAX Region Number 2 Number of vertices 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 3 Number of vertices 05 00 1 000 02 50 0 500 02 50 0 500 05 00 1 000 02 50 1 500 02 50 1 500 Region Number 4 Number of vertices 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 5 Number of vertices 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 6 Number of vertices 05 00 1 000 02 50 0 500 02 50 0 500 05 00 1 000 02 50 1 500 02 50 1 500 Region Number 7 Number of vertices 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 8 Number of vertices 12 50 MIN 10 00 1 750 10 00 1 750 12 50 MIN setup MTES ENAB 1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide xy x4 7 Mask Testing 14 199 14
56. to a reference waveform location 77 To display a reference waveform 78 To scale and position reference waveforms 79 To adjust reference waveform skew 79 To display reference waveform information 80 To save recall reference waveform files to from a USB storage device 80 6 Digital Channels To connect the digital probes to the device under test 81 Probe cable for digital channels 82 Acquiring waveforms using the digital channels 85 To display digital channels using AutoScale 85 Interpreting the digital waveform display 86 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 7 8 To change the displayed size of the digital channels 87 To switch a single channel on or off 88 To switch all digital channels on or off 88 To switch groups of channels on or off 88 To change the logic threshold for digital channels 88 To reposition a digital channel 89 To display digital channels as abus 90 Digital channel signal fidelity Probe impedance and grounding 93 Input Impedance 94 Probe Grounding 96 Best Probing Practices 98 To replace digital probe leads 98 Display Settings To adjust waveform intensity 101 To set or clear persistence 103 To clear the display 104 To select the grid type 104 To adjust the grid intensity 105 To freeze the display 105 Labels To turn the label display on or off 107 To assign a predefined label to a channel 108 To define a new label 109 To load a list of labels from a text fil
57. variations Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Cursors 12 1 004 Agilent Acquisition I Channels Cursors 167 12 Cursors 168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ee 13 Measurements To make automatic measurements 170 Measurements Summary 171 Voltage Measurements 173 Time Measurements 181 Measurement Thresholds 186 Measurement Window with Zoom Display 188 The Meas key lets you make automatic measurements on waveforms Some measurements can only be made on analog input channels The results of the last four selected measurements are displayed in the Measurements information area on the right hand side of the screen Cursors are turned on to show the portion of the waveform being measured for the most recently selected measurement bottom most on the right side measurement are Post Acquisition Processing In addition to changing display parameters after the acquisition you can perform all of the measurements and math functions after the acquisition Measurements and math functions will be recalculated as you pan and zoom and turn channels on and off As you zoom in and out on a signal using the horizontal scale knob and vertical volts division knob you affect the resolution of the display Because measurements and math functions are performed on displayed data you affect the res
58. waveform generator output 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 210 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Waveform Generator 16 3 In the Waveform Generator Settings Menu press the Modulation softkey 4 In the Waveform Generator Modulation Menu orm Generator Modulation Menu jodulation Type Waveform AM Depth Symmetry AM vy Ramp 50 100 Press the Modulation softkey to enable or disable modulated waveform generator output You can enable modulation for all waveform generator function types except pulse DC and noise Press the Type softkey and turn the Entry knob to select the modulation type Amplitude Modulation AM the amplitude of the original carrier signal is modified according to the amplitude of the modulating signal See To set up Amplitude Modulation AM on page 211 Frequency Modulation FM the frequency of the original carrier signal is modified according to the amplitude of the modulating signal See To set up Frequency Modulation FM on page 213 Frequency Shift Keying Modulation FSK the output frequency shifts between the original carrier frequency and a hop frequency at the specified FSK rate The FSK rate specifies a digital square wave modulating signal See To set up Frequency Shift Keying Modulation FS
59. www agilent com find 2000X Series Then select the Library tab followed by Specifications Or go to the Agilent home page at www agilent com and search for 2000 X Series oscilloscopes data sheet To order a data sheet by phone please contact your local Agilent office The complete list is available at www agilent com find contactus or on To contact Agilent on page 252 Measurement Category e Oscilloscope Measurement Category on page 274 e Measurement Category Definitions on page 274 l eg e Agilent Technologies aa 21 Reference e Transient Withstand Capability on page 275 Oscilloscope Measurement Category The InfiniiVision oscilloscopes are intended to be used for measurements in Measurement Category I WARNING Use this instrument only for measurements within its specified measurement category Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS Examples are measurements on circuits not derived from MAINS and specially protected internal MAINS derived circuits In the latter case transient stresses are variable for that reason the transient withstand capability of the equipment is made known to the user Measurement category II is for measurements performed on circuits directly connected to the low voltage installation Examples are measurements on household appliances portable tools and similar equipment
60. 1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index Simplified Chinese front panel overlays 36 Simplified Chinese user interface and Quick Help 41 sine waveform generator output 207 single acquisition 31 Single key 143 single shot acquisitions 137 single shot events 143 size 87 skew analog channel 61 ope trigger 115 Snapshot All measurements 173 snapshot all quick action 253 softkey labels 40 softkeys 6 29 software updates 280 software version 251 Span FFT 69 Spanish front panel overlays 36 Spanish user interface and Quick Help 41 specifications 273 spectral leakage FFT 75 square waveform generator output 207 square waves 146 stand alone connection 240 start acquisition 31 statistics mask test 194 status line 39 status User Cal 251 Std Deviation measurement 179 stop acquisition 31 storage locations navigate 226 subnet mask 239 subtract math function 66 sync pulse waveform generator 208 n T template front panel 35 test mask 189 theory sampling 145 threshold analog channel measurements 186 threshold digital channels 88 thumb drive 35 tilt for viewing 22 time measurements 181 time reference indicator 53 time re arm 158 295 Index timebase 47 times for saving data 224 Tools keys 30 Top measurement 175 tracking cursors 161 trademarks 2 Traditional Chinese front panel overla
61. 1 Video Trigger 123 A trigger setup tells the oscilloscope when to acquire and display data For example you can set up to trigger on the rising edge of the analog channel 1 input signal You can adjust the vertical level used for analog channel edge detection by turning the Trigger Level knob In addition to the edge trigger type you can also set up triggers on pulse widths patterns and video signals You can use any input channel or the External Trigger Input on page 140 BNC as the source for most trigger types Changes to the trigger setup are applied immediately If the oscilloscope is stopped when you change a trigger setup the oscilloscope uses the new specification when you press Run Stop or Single If the oscilloscope is running when you change a trigger setup it uses the new trigger definition when it starts the next acquisition You can use the Force Trigger key to acquire and display data when triggers are not occurring You can use the Mode Coupling key to set options that affect all trigger types see Chapter 10 Trigger Mode Coupling starting on page 135 a Agilent Technologies i 9 Triggers Triggers General Information You can save trigger setups along with the oscilloscope setup see Chapter 17 Save Recall Setups Screens Data starting on page 217 A triggered waveform is one in which the oscilloscope begins tracing displaying the waveform from the left side of the
62. 1 Press the Ref key to turn on reference waveforms 2 In the Reference Waveform Menu press the Ref softkey and turn the Entry knob to select the desired reference waveform location 3 Press the Clear R1 R2 softkey to clear the reference waveform location Reference waveforms are also cleared by a Factory Default or Secure Erase see Chapter 17 Save Recall Setups Screens Data starting on page 217 To display a reference waveform 78 1 Press the Ref key to turn on reference waveforms 2 In the Reference Waveform Menu press the Ref softkey and turn the Entry knob to select the desired reference waveform location 3 Then press the Ref softkey again to enable disable the reference waveform display Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Reference Waveforms 5 1 7605 500 08 1 14V Agilent Acquisition Channels Refer eform Menu Display Ref Source Options R1 0 0s T One reference waveform can be displayed at a time See Also To display reference waveform information on page 80 To scale and position reference waveforms 1 Make sure the multiplexed scale and position knobs to the right of the Ref key are selected for the reference waveform If the arrow to the left of the Ref key is not illuminated press the key 2 Turn the upper multiplexed knob to adjust the reference waveform scale 3 Turn the lower multiplexed knob to adjust the reference waveform positi
63. 29 INSTR Advanced information Identification Ooff on Use the navigation bar on the left to access your oscilloscope and y lated information Agilent Technologies Inc 2006 2010 Identification option Instrument Utilities The Instrument Utilities page of the web interface lets you e View installed options 268 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide e View firmware versions e Install firmware upgrade files e View calibration status You can select these cababilities via a drop down menu Agilent Technologies KOSICE ZJ Instrument Utilities Installed Options Firmware Version Calibration Status _ talled Options Web Interface 20 trument Utilities FPGA Altera Power application Segmented Memory 200MHz Bandwidth 100MHz Bandwidth Education license Setting a Password Whenever you connect the oscilloscope to a LAN it is good practice to set a password The password prevents someone from remotely accessing the oscilloscope via a Web browser and changing parameters Remote users can still view the Welcome screen view network status etc but they can t operate the instrument or change its setup without the password Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 269 20 Web Interface To set a password 1 Access the oscilloscope s web interface see Accessing the Web Interface on
64. 30 e To trigger on odd or even fields on page 131 To trigger on a specific line of video Video triggering requires greater than 1 2 division of sync amplitude with any analog channel as the trigger source Turning the trigger Level knob in Video trigger does not change the trigger level because the trigger level is automatically set to the sync pulse tips One example of triggering on a specific line of video is looking at the vertical interval test signals VITS which are typically in line 18 Another example is closed captioning which is typically in line 21 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard NTSC 4 Press the Mode softkey and select the TV field of the line you want to trigger on You can choose Line Field1 Line Field2 or Line Alternate 5 Press the Line softkey and select the number of the line you want to examine Alternate Triggering If Line Alternate is selected the oscilloscope will alternately trigger on the selected line number in Field 1 and Field 2 This is a quick way to compare the Field 1 VITS and Field 2 VITS or to check for the correct insertion of the half line at the end of Field 1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 127 9 Triggers Line F1 Agilent Channels Mode Field Holdoff
65. 7 Learn the Oscilloscope Display 38 Access the Built In Quick Help 40 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 2 Horizontal Controls To adjust the horizontal time div scale 44 To adjust the horizontal delay position 45 Panning and Zooming Single or Stopped Acquisitions 46 To change the horizontal time mode Normal XY or Roll 47 XY Time Mode 48 To display the zoomed time base 50 To change the horizontal scale knob s coarse fine adjustment setting 52 To position the time reference left center right 52 Navigating the Time Base 53 To navigate time 53 To navigate segments 53 3 Vertical Controls To turn waveforms on or off channel or math 56 To adjust the vertical scale 5 7 To adjust the vertical position 57 To specify channel coupling 5 7 To specify bandwidth limiting 58 To change the vertical scale knob s coarse fine adjustment setting 58 To invert a waveform 59 Setting Analog Channel Probe Options 59 To specify the channel units 60 To specify the probe attenuation 60 To specify the probe skew 61 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms To display math waveforms 63 To perform a transform function on an arithmetic operation 64 To adjust the math waveform scale and offset 65 Units for Math Waveforms 65 Math Operators 66 Add or Subtract 66 Multiply or Divide 67 Math Transforms 68 FFT Measurement 68 5 Reference Waveforms To save a waveform
66. 74 Yes Period on page 181 Yes Phase on page 185 Measures between two sources Press Settings to specify the second source Preshoot on page 177 Rise Time on page 183 DC RMS on page 178 AC RMS on page 179 Top on page 175 Width on page 183 Yes Width on page 183 Yes Use the cursors to make other measurements on FFT Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 Snapshot All The Snapshot All measurement type displays a popup containing a snapshot of all the single waveform measurements 1 00V 1 00 500 05 Agilent Acquisition Channels x Measurements vasurement Menu Source Type Settings Clear Meas Snapshot Me ent oe You can also configure the Quick Action key to display the Snapshot All popup See Configuring the Quick Action Key on page 253 Voltage Measurements The following figure shows the voltage measurement points Maximum Measurement units for each input channel can be set to Volts or Amps using the channel Probe Units softkey See To specify the channel units on page 60 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 173 13 Measurements The units of math waveforms are described in Units for Math Waveforms on page 65 e Peak Peak on page 174 e Maximum on page 174 e Minimum on page 174 e Amplitude on p
67. A Active probe 800 MHz requires 1142A power supply 1145A Active probe 750 MHz 2 ch requires 1142A power supply N2772A Active differential probe 20 MHz 1 2 kVDC peak AC max requires N2773A power supply N2791A High voltage differential probe 25 MHz 700 V 1 MOhm termination 10 1 or 100 1 switchable N2792A Differential probe 200 MHz 10 1 50 Ohm termination N2793A Differential probe 800 MHz 10 1 15 V 50 Ohm termination Current Probes The following current probes can be used with the InfiniiVision 2000 X Series oscilloscopes Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 277 21 Reference 278 Table8 Current Probes Model Description 1146A Current probe 100 kHz 100 A AC DC N2774A Obsolete replaced by N2782A with N2775A power supply N2780A Current probe 2 MHz 500 A AC DC use with N2779A power supply N2781A Current probe 10 MHz 150 A AC DC use with N2779A power supply N2782A Current probe 50 MHz 30 A AC DC use with N2779A power supply N2783A Current probe 100 MHz 30 A AC DC use with N2779A power supply Accessories Available In addition to passive proves Passive Probes on page 276 differential probes Differential Probes on page 277 and current probes Current Probes on page 277 the following accessories are available for the InfiniiVision 2000 X Series oscilloscopes
68. Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide oa Agilent Technologies Notices Agilent Technologies Inc 2005 2012 No part of this manual may be reproduced in any form or by any means including elec tronic storage and retrieval or translation into a foreign language without prior agree ment and written consent from Agilent Technologies Inc as governed by United States and international copyright laws Manual Part Number 75015 97034 Edition Fifth edition July 2012 Printed in Malaysia Agilent Technologies Inc 1900 Garden of the Gods Road Colorado Springs CO 80907 USA Print History 75015 97000 January 2011 75015 97011 February 2011 75015 97012 June 2011 75015 97023 March 2012 75015 97034 July 2012 Trademarks Java is a U S trademark of Sun Microsys tems Inc Sun Sun Microsystems and the Sun Logo are trademarks or registered trademarks of Sun Microsystems Inc in the U S and other countries Warranty The material contained in this docu ment is provided as is and is sub ject to being changed without notice in future editions Further to the max imum extent permitted by applicable law Agilent disclaims all warranties either express or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchantability and fitness for a par ticular purpose Agilent shall not be liable for
69. CII characters Example autoMask CH1 OCT 03 09 40 26 2008 When a mask file contains the keyword autoMask in the title the edge of the mask is passing by definition Otherwise the edge of the mask is defined as a failure 50 008 Trig d 2 00V Agilent on Region 1 Channels Region 2 Up to 8 regions can be defined for a mask They can be numbered 1 8 They can appear in any order in the msk file The numbering of the regions must go from top to bottom left to right An Automask file contains two special regions the region glued to the top of the display and the region that is glued to the bottom The top region is indicated by y values of MAX for the first and last points The bottom region is indicated by y values of MIN for the first and last points The top region must be the lowest numbered region in the file The bottom region must be the highest numbered region in the file Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Oscilloscope Setup Information Mask Testing 14 Region number 1 is the top mask region The vertices in Region 1 describe points along a line that line is the bottom edge of the top portion of the mask Similarly the vertices in Region 2 describe the line that forms the top of the bottom part of the mask The vertices in a mask file are normalized There are four parameters that define how values are normalized e X1 e AX e YI e Y2 These four parameters
70. Default CAN I2C LIN UART RS232 SPI 2 Wire Serial SPI 8 Wire Serial SPI 4 Wire Serial When you choose any ALB format other than Default waveforms from the analog channels if displayed are also digitized and displayed as digital waveforms The analog channel s trigger level determines the point at which the voltage is considered a logic 1 or 0 If you connect your serial bus signals to the oscilloscope as shown in the following table the bus signal names in the Agilent Logic Analyzer application will be correct Otherwise you must re map the signals in the Agilent Logic Analyzer application 222 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Table5 Recommended Signal Mapping Save Recall Setups Screens Data 17 Label Signal Oscilloscope Maps to Logic Channel Analyzer Channel TxRS232 Tx Chi DO RxRS232 Rx Ch2 D1 12C Data Chi DO Clk Ch2 D1 SP12 2 wire Clk Chi DO Data Ch2 D1 SPI3 3 wire Chip Select Ch1 DO Clk Ch2 D1 Data Ch3 D2 SPI4 4 wire Chip Select Ch1 DO Clk Ch2 D1 Dataln Ch3 D2 DataOut Ch4 D3 CAN Data Chi DO LIN Data Ch1 DO 4 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed Length Control The Length control is available when saving data to CSV ASCII XY BIN or ALB format files It sets the number of data points that will be output to the file
71. K on page 214 To set up Amplitude Modulation AM In the Waveform Generator Modulation Menu under Wave Gen gt Settings gt Modulation 1 Press the Type softkey and turn the Entry knob to select Amplitude Modulation AM 2 Press the Waveform softkey and turn the Entry knob to select the shape of the modulating signal Sine Square Ramp Sine Cardinal Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 211 16 Waveform Generator 212 Exponential Rise Exponential Fall When the Ramp shape is selected a Symmetry softkey appears so that you can specify the amount of time per cycle that the ramp waveform is rising 3 Press the AM Freq softkey and turn the Entry knob to specify the frequency of the modulating signal 4 Press the AM Depth softkey and turn the Entry knob to specify the amount of amplitude modulation AM Depth refers to the portion of the amplitude range that will be used by the modulation For example a depth setting of 80 causes the output amplitude to vary from 10 to 90 90 10 80 of the original amplitude as the modulating signal goes from its minimum to maximum amplitude The following screen shows an AM modulation of a 100 kHz sine wave carrier signal 4 Channels Waveform Generator Modulation Menu Modulation Type Waveform AM Depth Symmetry AM v Ramp 50 100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Waveform Generator 16 To set up
72. LB data alb Binary data bin wy Instrument Mask msk Utilities OOOO You can click Preview to view the oscilloscope s current screen image When previewing you can use the New Acquisition check box to force a new acquisition before the preview With some formats you can click Save Setup Info to save setup information to an ASCII txt format file c Click Save The current acquisition is saved d In the File Download dialog click Save e In the Save As dialog navigate to the folder where you want to save the file then click Save Recalling Files via the Web Interface To recall setup files reference waveform data files or mask files from your PC via the oscilloscope s web interface 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface is displayed select the Save Recall tab from the left side of the Welcome screen 3 Click the Recall link 266 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Web Interface 20 4 On the Recall page a Click Browse b In the Choose file dialog select the file you want to recall then click Open c When recalling reference waveform data files select the To Reference Waveform location Agilent Technologies KOSICE l C templweb_interface scope h5 Browse To Reference Waveform iga v R1 R2 d Click Recall Get Image To save or print th
73. Line Fld1 O Oflds Figure 15 Example Triggering on Line 136 To trigger on all sync pulses To quickly find maximum video levels you could trigger on all sync pulses When All Lines is selected as the Video trigger mode the oscilloscope will trigger on all horizontal sync pulses 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select All Lines 128 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 Lines Agilent Channels Video Trigger Menu Standard Mode Field Holdoff NTSC All Lines O Oflds Figure 16 Triggering on All Lines To trigger on a specific field of the video signal To examine the components of a video signal trigger on either Field 1 or Field 2 available for interleaved standards When a specific field is selected the oscilloscope triggers on the rising edge of the first serration pulse in the vertical sync interval in the specified field 1 or 2 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select Field1 or Field2 Agilent InfiniiVision 2000 X Series Oscilloscope
74. Mask Testing CHAN1 RANG 4 00E 00 OFFS 0 0E 00 COUP DC IMP ONEM DISP 1 BWL 0 INV 0 CHAN1 LAB 1 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING CHAN2 RANG 16 0E 00 OFFS 1 62400E 00 COUP DC IMP FIFT DISP 0 BWL 0 INV 0 CHAN2 LAB 2 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING CHAN3 RANG 40 0E 00 OFFS 0 0E 00 COUP DC IMP ONEM DISP 0 BWL 0 INV 0 CHAN3 LAB 3 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING CHAN4 RANG 40 0E 00 OFFS 0 0E 00 COUP DC IMP ONEM DISP 0 BWL 0 INV 0 CHAN4 LAB 4 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING EXT BWL 0 IMP ONEM RANG 5E 00 UNIT VOLT PROB 1 0E 00 PROB STYP SING TIM MODE MAIN REF CENT MAIN RANG 50 00E 09 POS 0 0E 00 TRIG MODE EDGE SWE AUTO NREJ 0 HFR 0 HOLD 60E 09 TRIG EDGE SOUR CHAN1 LEV 75 00E 03 SLOP POS REJ OFF COUP DC ACQ MODE RTIM TYPE NORM COMP 100 COUNT 8 SEGM COUN 2 DISP LAB 0 CONN 1 PERS MIN SOUR PMEM1 HARD APR AREA SCR FACT 0 FFE 0 INKS 1 PAL NONE LAY PORT SAVE FIL mask_0 SAVE IMAG AREA GRAT FACT 0 FORM NONE INKS 0 PAL COL SAVE WAV FORM NONE MTES SOUR CHAN1 ENAB 1 LOCK 1 MTES AMAS SOUR CHAN1 UNIT DIV XDEL 3 00000000E 001 YDEL 2 00000000E 00 1 MTES SCAL BIND 0 X1 0 0E 00 XDEL 1 0000E 09 Y1 0 0E 00 Y2 1 00000E 0 0 MTES RMOD FOR RMOD TIME 1E 00 WAV 1000 SIGM 6 0E 00 MTES RMOD FACT STOP 0 PRIN 0 SAVE 0 end_setup How is mask testing done InfiniiVision oscilloscopes start mask testing by creati
75. Modulation FSK 214 To restore waveform generator defaults 215 17 Save Recall Setups Screens Data Saving Setups Screen Images or Data 217 To save setup files 219 To save BMP or PNG image files 219 To save CSV ASCII XY or BIN data files 220 To save ALB data files 221 Length Control 223 To save reference waveform files to a USB storage device 225 To save masks 225 To navigate storage locations 226 To enter filenames 226 Recalling Setups Masks or Reference Waveforms 227 To recall setup files 227 To recall mask files 228 To recall reference waveform files from a USB storage device 228 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Recalling Default Setups 229 Performing a Secure Erase 229 18 Print Screens To print the oscilloscope s display 231 To set up network printer connections 232 To specify the print options 234 To specify the palette option 235 19 Utility Settings 1 0 Interface Settings 237 Setting up the Oscilloscope s LAN Connection 238 To establish a LAN connection 239 Stand alone Point to Point ConnectiontoaPC 240 File Explorer 241 Setting Oscilloscope Preferences 243 To choose expand about center or ground 243 To disable enable transparent backgrounds 244 To load the default label library 244 To setupthe screensaver 244 To set AutoScale preferences 245 Setting the Oscilloscope s Clock 246 Setting the Rear Panel TRIG OUT Source 246 Performing Service Tasks 247 To per
76. Oscilloscope 251 AC channel coupling 58 AC RMS Full Screen measurement 179 AC RMS N Cycles measurement 179 accessories 19 276 278 acquire 143 153 Acquire key 33 acquisition memory 114 acquisition memory saving 223 acquisition mode 149 acquisition mode averaging 153 acquisition mode high resolution 155 acquisition mode normal 150 acquisition mode peak detect 150 acquisition mode preserve during AutoScale 246 acquisition modes 143 activity indicator 87 actual sample rate 149 adding digital channels license 280 addition math function 66 Addresses softkey 240 Agilent 10 Libraries Suite 264 aliasing 145 aliasing FFT 73 alternating edge trigger 117 AM amplitude modulation waveform generator output 211 Amplitude measurement 174 amplitude modulation AM waveform generator output 211 analog channel probe attenuation 60 analog channel setup 55 analog filters adjusting 69 Analyze key 30 Analyze Segments 156 157 annotation adding 254 ASCII file format 218 attenuation probe 60 attenuation probe external trigger 141 Auto Increment 227 Auto Scale key 32 Auto Setup FFT 70 Auto trigger mode 136 Auto trigger indicator 137 AutolP 239 240 automatic measurements 169 171 automatic setup 85 AutoScale preferences 245 AutoScale digital channels 85 AutoScale undo 26 Average Full Screen measurement 178 Average
77. To navigate segments on page 53 To navigate time When acquisitions are stopped you can use the navigation controls to play through the captured data 1 Press Navigate 2 In the Navigate Menu press Navigate then select Time 3 Press the Q O navigation keys to play backward stop or play forward in time You can press the or A keys multiple times to speed up the playback There are three speed levels To navigate segments When the segmented memory acquisition is enabled and acquisitions are stopped you can use the navigation controls to play through the acquired segments 1 Press Navigate 2 In the Navigate Menu press Navigate then select Segments Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 53 2 Horizontal Controls 3 Press Play Mode then select Manual to play through segments manually In the Manual play mode Press the QA back and forward keys to go to the previous or next segment Press the a softkey to go to the first segment Press the gt softkey to go to the last segment Auto to play through segments in an automated fashion In the Auto play mode Press the Q O navigation keys to play backward stop or play forward in time You can press the or keys multiple times to speed up the playback There are three speed levels 54 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User
78. User s Guide Utility Settings 19 Configuring the Quick Action Key The Quick Action key lets you perform common repetitive actions by pressing a single key To configure the Quick Action key 1 Press Utility gt Quick Action gt Action then select the action that should be performed Off disables the Quick Action key Quick Measure All displays a popup containing a snapshot of all the single waveform measurements The Source softkey lets you select the waveform source which also becomes the source selection in the Measurement Menu See Chapter 13 Measurements starting on page 169 Quick Print prints the current screen image Press Settings to set up the printing options See Chapter 18 Print Screens starting on page 231 Quick Save saves the current image waveform data or setup Press Settings to set up the save options See Chapter 17 Save Recall Setups Screens Data starting on page 217 Quick Recall recalls a setup mask or reference waveform Press Settings to set up the recall options See Chapter 17 Save Recall Setups Screens Data starting on page 217 Quick Freeze Display freezes the display without stopping running acquisitions or un freezes the display if currently frozen For more information see To freeze the display on page 105 Quick Trigger Mode toggles the trigger mode between Auto and Normal see To select the Auto or N
79. Vision 2000 X Series Oscilloscopes User s Guide Triggers 9 Table 4 Half field holdoff time Standard Time NTSC 8 35 ms PAL 10 ms PAL M 10 ms SECAM 10 ms Field1 Agilent Acquisition h BEEN Field Holdoff Field1 3 0000flds Figure 20 Using Field Holdoff to Synchronize to Color Field 1 or 3 Field 1 mode Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 133 9 Triggers 134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Noisy Signals Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ee 10 Trigger Mode Coupling To select the Auto or Normal trigger mode 136 To select the trigger coupling 138 To enable or disable trigger noise rejection 139 To enable or disable trigger HF Reject 139 To set the trigger holdoff 140 External Trigger Input 140 To access the Trigger Mode and Coupling Menu e In the Trigger section of the front panel press the Mode Coupling key Noise Rej HF Reject oldo External If the signal you are probing is noisy you can set up the oscilloscope to reduce the noise in the trigger path and on the displayed waveform First stabilize the displayed waveform by removing the noise from the trigger path Second reduce the noise on the displayed waveform 1 Connect a signal to the oscilloscope and obtain a stable display 2 Remove the noise from the trigger path by turning on high frequency rejection To
80. X Series Oscilloscopes User s Guide 19 Utility Settings 1 0 Interface Settings 237 Setting up the Oscilloscope s LAN Connection 238 File Explorer 241 Setting Oscilloscope Preferences 243 Setting the Oscilloscope s Clock 246 Setting the Rear Panel TRIG OUT Source 246 Performing Service Tasks 247 Configuring the Quick Action Key 253 Adding an Annotation 254 This chapter explains oscilloscope utility functions 1 0 Interface Settings The oscilloscope can be accessed and or controlled remotely via these I O interfaces e USB device port on the rear panel square shaped USB port e LAN interface when a LAN VGA module is installed in the rear panel module slot e GPIB interface when a GPIB module is installed in the rear panel module slot To configure the I O interfaces 1 On the oscilloscope s front panel press Utility 2 In the Utility Menu press 1 0 s Agilent Technologies 237 19 Utility Settings See Also 3 In the I O Menu press Configure LAN When a DSOXLAN LAN VGA module is installed you can use the LAN Settings and LAN Reset softkeys to configure the LAN interface See Setting up the Oscilloscope s LAN Connection on page 238 GPIB When a DSOXGPIB GPIB module is installed you can use the Address softkey to configure the GPIB address There are no configuration settings for the USB interface When an I O interface is installed remote control over that interface is always
81. You can also configure the Quick Action key to print the display See Configuring the Quick Action Key on page 253 See Also Adding an Annotation on page 254 To set up network printer connections When the DSOXLAN LAN VGA module is installed you can set up network printer connections A network printer is a printer attached to a computer or print server on the network 1 Press the Print key on the front panel 232 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Print Screens 18 2 In the Print Configuration Menu press the Printto softkey then turn the Entry knob to select the network printer you want to configure either 0 or 1 3 Press the Network Setup softkey 4 In the Network Printer Setup Menu press the Modify softkey then turn the Entry knob to select the network parameter you want to enter The setting that must be entered is Printer Address this is the printer or print server s address in one of the following formats IP address of a network enabled printer for example 192 168 1 100 or 192 168 1 100 650 Optionally a non standard port number can be specified following a colon IP address of a print server followed by the path to the printer for example 192 168 1 100 printers printer name or 192 168 1 100 650 printers printer name Path to a Windows network printer share for example server share When the Printer Address is a Windows network printer share
82. age 174 e Top on page 175 e Base on page 176 e Overshoot on page 176 e Preshoot on page 177 e Average on page 178 e DC RMS on page 178 e AC RMS on page 179 Peak Peak The peak to peak value is the difference between Maximum and Minimum values The Y cursors show the values being measured Maximum Maximum is the highest value in the waveform display The Y cursor shows the value being measured Minimum Minimum is the lowest value in the waveform display The Y cursor shows the value being measured Amplitude The Amplitude of a waveform is the difference between its Top and Base values The Y cursors show the values being measured 174 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 Top The Top of a waveform is the mode most common value of the upper part of the waveform or if the mode is not well defined the top is the same as Maximum The Y cursor shows the value being measured See Also To isolate a pulse for Top measurement on page 175 To isolate a pulse for Top measurement The following figure shows how to use Zoom mode to isolate a pulse for a Top measurement You may need to change the measurement window setting so that the measurement is made in the lower Zoom window See Measurement Window with Zoom Display on page 188 Agilent Channels x Measurements Top 1 Settings Clear Meas gt Figure 33 Isolating area for Top measuremen
83. al channels Bytes Per Point A 16 bit short that is the number of bytes per data point Buffer Size A 32 bit integer that is the size of the buffer required to hold the data points Example Program for Reading Binary Data To find an example program for reading binary data direct your web browser to www agilent com find 2000X Series examples and select Example Program for Reading Binary Data 284 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Examples of Binary Files Reference Single The following picture shows a binary file of a single acquisition with Acquisition multiple analog channels Multiple Analog File Header Number of Waveforms N 12 bytes Channels Waveform Header 1 Number of Waveform Buffers 1 140 bytes Waveform Data Header 1 12 bytes Voltage Data 1 buffer size Waveform Header 2 Number of Waveform Buffers 1 140 bytes Buffer Type 1 floating point Bytes per Point 4 Waveform Data Header 2 12 bytes Buffer Type 1 floating point Bytes per Point 4 Voltage Data 2 buffer size Waveform Header N Number of Waveform Buffers 1 140 bytes Waveform Data Header N 12 bytes Buffer Type 1 floating point Bytes per Point 4 Voltage Data N buffer size Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 285 21 Reference Single The following picture shows a binary file of a single acquisi
84. alog channel measurement thresholds You can also open the Measurement Threshold Menu by pressing Analyze gt Features and then selecting Measurement Thresholds 2 Press the Source softkey to select the analog channel or math waveform source for which you want to change measurement thresholds Each analog channel and the math waveform can be assigned unique threshold values Measurement Thre Source 3 Press the Type softkey to set the measurement threshold to percentage of Top and Base value or to Absolute absolute value Percentage thresholds can be set from 5 to 95 The units for absolute threshold for each channel is set in the channel probe menu When the Source is set to Math f t the threshold Type can only be set to Percent TIP Absolute threshold hints Absolute thresholds are dependent on channel scaling probe attenuation and probe units Always set these values first before setting absolute thresholds The minimum and maximum threshold values are limited to on screen values If any of the absolute threshold values are above or below the minimum or maximum waveform values the measurement may not be valid 4 Press the Lower softkey then turn the Entry knob to set the lower measurement threshold value Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 187 13 Measurements Increasing the lower value beyond the set middle value will automatically increase the middle value to
85. ame IP address Configuration subnet mask gateway IP DNS IP etc from your network administrator 2 Press Utility gt 1 0 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 239 19 Utility Settings 240 3 Press the LAN Settings softkey 4 Press the Config softkey then turn the Entry knob to select Automatic and press the softkey again to disable it If Automatic is not enabled the oscilloscope s LAN configuration must be set up manually using the Addresses and Host name softkeys Configure the oscilloscope s LAN interface a Press the Addresses softkey b Use the Modify softkey and the other softkeys and the Entry knob to enter the IP Address Subnet Mask Gateway IP and DNS IP values When you are done go back up in the menu hierarchy c Press the Hostname softkey Use the softkeys and the Entry knob to enter the Host name When you are done go back up in the menu hierarchy d Press the Apply softkey Connect the oscilloscope to the local area network LAN by inserting the LAN cable into the LAN port on the rear panel of the oscilloscope Stand alone Point to Point Connection to a PC The following procedure describes how to establish a point to point stand alone connection to the oscilloscope This is useful if you want to control the oscilloscope using a laptop computer or a stand alone computer 1 2 3 Press Utility gt 1 0 Press the LAN Settings softkey Press the Co
86. analog oscilloscopes When variable persistence is selected press the Time softkey and use the Entry knob to specify the amount of time that previous acquisitions are to be displayed The display will begin accumulating multiple acquisitions Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 103 7 Display Settings 3 To erase the results of previous acquisitions from the display press the Clear Persistence softkey The oscilloscope will start to accumulate acquisitions again To return the oscilloscope to the normal display mode turn off persistence then press the Clear Persistence softkey Turning off persistence does not clear the display The display is cleared if you press the Clear Display softkey or if you press the AutoScale key which also turns off persistence For another method of seeing worst case extremes of varying waveforms see Glitch or Narrow Pulse Capture on page 151 To clear the display 1 Press Display gt Clear Display You can also configure the Quick Action key to clear the display See Configuring the Quick Action Key on page 253 To select the grid type 104 When the Video trigger type is selected see Video Trigger on page 123 and the vertical scaling of at least one displayed channel is 140 mV div the Grid softkey lets you select from these grid types Full the normal oscilloscope grid mV shows vertical grids labeled on the left from 0
87. and measurement statistics if your oscilloscope supports them run only on waveforms that contain a mask violation Measurements are not affected by passing waveforms This mode is not available when the acquisition mode is set to Averaging Note that you can choose to Print or Save but you cannot select both at the same time All other actions may be selected at the same time For example you can select both Stop and Measure to cause the oscilloscope to measure and stop on the first error You can also output a signal on the rear panel TRIG OUT BNC connector when there is a mask test failure See Setting the Rear Panel TRIG OUT Source on page 246 Source Lock When you turn on Source Lock using the Source Lock softkey the mask is redrawn to match the source whenever you move the waveform For example if you change the horizontal timebase or the vertical gain the mask is redrawn with the new settings When you turn off Source Lock the mask is not redrawn when horizontal or vertical settings are changed Source If you change the Source channel the mask is not erased It is re scaled to the vertical gain and offset settings of the channel to which it is assigned To create a new mask for the selected source channel go back up in the menu hierarchy then press Automask and press Create Mask The Source softkey in the Mask Setup Menu is the same as the Source softkey in the Automask Menu Test All When enabled all displaye
88. annel 2 is being displayed to turn channel 1 off press 1 to display the channel 1 menu then press 1 again to turn channel 1 off Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Vertical Controls 3 To adjust the vertical scale 1 Turn the large knob above the channel key marked N N to set the vertical scale volts division for the channel The vertical scale knob changes the analog channel scale in a 1 2 5 step sequence with a 1 1 probe attached unless fine adjustment is enabled see To change the vertical scale knob s coarse fine adjustment setting on page 58 The analog channel Volts Div value is displayed in the status line The default mode for expanding the signal when you turn the volts division knob is vertical expansion about the ground level of the channel however you can change this to expand about the center of the display See To choose expand about center or ground on page 243 To adjust the vertical position 1 Turn the small vertical position knob to move the channel s waveform up or down on the display The voltage value momentarily displayed in the upper right portion of the display represents the voltage difference between the vertical center of the display and the ground level icon It also represents the voltage at the vertical center of the display if vertical expansion is set to expand about ground see To choose expand about center or ground on page 243 To s
89. ars in the graticule along with a scale and the frequency counter value The DVM scale is determined by the channel s vertical scale and reference level The scale s blue triangle pointer shows the most recent measurement The white bar above that shows the measurement extrema over the last 3 seconds a Agilent Technologies 201 15 Digital Voltmeter 202 6258 Agilent DVM Channels Mode Transparent Auto Range DVM AC RMS The DVM makes accurate RMS measurements when the signal frequency is between 20 Hz and 100 kHz When the signal frequency is outside this range lt BW Limit or gt BW Limit appears in the DVM display to caution you about inaccurate RMS measurement results To use the digital voltmeter 1 2 3 4 Press the Analyze key Press Features then select Digital Voltmeter Press Features again to enable the DVM measurements Press the Source softkey and turn the Entry knob to select the analog channel on which digital voltmeter DVM measurements are made The selected channel does not have to be on displaying a waveform in order for DVM measurements to be made Press the Mode softkey and turn the Entry knob to select the digital voltmeter DVM mode AC RMS displays the root mean square value of the acquired data with the DC component removed DC displays the DC value of the acquired data DC RMS displays the root mean square value of the acquired data Frequency displ
90. ave softkey A message indicating whether the save was successful is displayed Mask files have the extension MSK Masks are also saved as part of setup files See To save setup files on page 219 See Also Chapter 14 Mask Testing starting on page 189 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 225 17 Save Recall Setups Screens Data To navigate storage locations When saving or recalling files the softkey in the second position of the Save Menu or Recall Menu along with the Entry knob are used to navigate to storage locations The storage locations can be internal oscilloscope storage locations for setup files or mask files or they can be external storage locations on a connected USB storage device The softkey in the second position can have these labels e Press togo when you can push the Entry knob to navigate to a new folder or storage location e Location when you have navigated to the current folder location and are not saving files e Save to when you can save to the selected location e Load from when you can recall from the selected file When saving files e The proposed file name is shown in the Save to file line above the softkeys e To overwrite an existing file browse to that file and select it To create a new file name see To enter file names on page 226 To enter file names To create new file names when saving files to a USB storage device
91. avigate segments on page 53 Infinite Persistence with Segmented Memory When data has been acquired to segmented memory you can also turn on infinite persistence in the Display Menu and press the Analyze Segments softkey to create an infinite persistence display The Analyze Segments softkey appears when the acquisition is stopped and the segmented memory feature is on Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 157 11 Acquisition Control 158 Segmented Memory Re Arm Time After each segment fills the oscilloscope re arms and is ready to trigger in about 8 us Remember though for example if the horizontal time per division control is set to 5 us div and the Time Reference is set to Center it will take at least 50 us to fill all ten divisions and re arm That is 25 us to capture pre trigger data and 25 us to capture post trigger data 2 0005 500 08 Stop 1 13 Agilent Saving Data from Segmented Memory You can save either the currently displayed segment Save Segment Current or all segments Save Segment All in the following data formats CSV ASCII XY and BIN Be sure to set the Length control to capture enough points to accurately represent the captured data When the oscilloscope is busy saving multiple segments progress is displayed in the upper right area of the display For more information see To save CSV ASCII XY or BIN data files on page 220 Agilent InfiniiVision
92. ays the frequency counter measurement Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Voltmeter 15 6 Press Transparent to toggle between a transparent and shaded background for the DVM display 7 If the selected source channel is not used in oscilloscope triggering press Auto Range to disable or enable automatic adjustment of the DVM channel s vertical scale vertical ground level position and trigger threshold voltage level used for the counter frequency measurement When enabled Auto Range overrides attempted adjustments of the channel s vertical scale and position knobs When disabled you can use the channel s vertical scale and position knobs normally Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 203 15 Digital Voltmeter 204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide e e e ee 16 a ni a Waveform Generator e e e i To select generated waveform types and settings 205 e To output the waveform generator sync pulse 208 To specify the expected output load 209 To use waveform generator logic presets 209 To add noise to the waveform generator output 210 To add modulation to the waveform generator output 210 To restore waveform generator defaults 215 A waveform generator is built into the oscilloscope It is enabled by Option WGN or the DSOX2WAVEGEN upgrade The wavef
93. be more than the lower value The default lower threshold is 10 or 800 mV If threshold Type is set to the lower threshold value can be set from 5 to 93 Press the Middle softkey then turn the Entry knob to set the middle measurement threshold value The middle value is bounded by the values set for lower and upper thresholds The default middle threshold is 50 or 1 20 V If threshold Type is set to the middle threshold value can be set from 6 to 94 Press the Upper softkey then turn the Entry knob to set the upper measurement threshold value Decreasing the upper value below the set middle value will automatically decrease the middle value to be less than the upper value The default upper threshold is 90 or 1 50 V If threshold Type is set to the upper threshold value can be set from 7 to 95 Measurement Window with Zoom Display 188 When the zoomed time base is displayed you can choose whether measurements are made in the Main window portion of the display or the Zoom window portion of the display Press the Meas key In the Measurement Menu press the Settings softkey In the Measurement Settings Menu press the Meas Window softkey then turn the Entry knob to select from Auto Select The measurement is attempted in the lower Zoom window if it cannot be made there the upper Main window is used Main The measurement window is the upper Main window Zoom The measurement wind
94. byte characters AG that indicate the file is in the Agilent Binary Data file format Version Two bytes that represent the file version File Size A 32 bit integer that is the number of bytes that are in the file Number of A 32 bit integer that is the number of waveforms that are stored in the file Waveforms It is possible to store more than one waveform in the file and each waveform stored will have a waveform header When using segmented memory each segment is treated as a separate waveform The waveform header contains information about the type of waveform data that is stored following the waveform data header Header Size A 32 bit integer that is the number of bytes in the header Waveform Type A 32 bit integer that is the type of waveform stored in the file 0 Unknown e 1 Normal 2 Peak Detect e 3 Average 4 Not used in InfiniiVision oscilloscopes e 5 Notused in InfiniiVision oscilloscopes e 6 Logic Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Reference 21 Number of A 32 bit integer that is the number of waveform buffers required to read the Waveform data Buffers Points A 32 bit integer that is the number of waveform points in the data Count A 32 bit integer that is the number of hits at each time bucket in the waveform record when the waveform was created using an acquisition mode like averaging For example when averaging a count of
95. ck Help 41 K keyboard USB 110 227 234 245 255 keys front panel 28 knobs front panel 28 Korean front panel overlays 36 Korean user interface and Quick Help 41 L Label key 34 label list 111 label list loading from text file 110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index labels 107 labels auto increment 110 labels default library 111 LAN connection 239 LAN interface remote control 237 LAN port 37 LAN Settings softkey 239 240 LAN VGA module 22 37 LAN VGA module installation 22 andscape mode 234 anguage user interface and Quick Help 41 ength control 223 ength softkey 221 222 evel trigger 114 LF Reject 138 ibrary labels 109 icenses 279 280 ine voltage 23 oad file 241 Load from 226 ocalized front panel overlay 35 Location 226 242 ogic presets waveform generator 209 ogic threshold 88 ow frequency noise rejection 138 M mask files recall 228 MASK license 279 mask test trigger output 193 247 mask testing 189 mask TRIG OUT signal 247 Math key 32 math operators 66 ath transforms 68 ath 1 2 67 ath 1 2 67 ath addition 66 ath divide 67 ath FFT 68 ath functions 63 ath multiply 67 ath offset 65 ath scale 65 ath subtract 66 ath transform functions on arithmetic operations 64 m m m m m m m m m m m m 293 Index math units 65 math
96. consult the InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 99 6 Digital Channels 100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide eo o e e 7 209 Display Settings To adjust waveform intensity 101 To set or clear persistence 103 To clear the display 104 To select the grid type 104 To adjust the grid intensity 105 To freeze the display 105 To adjust waveform intensity You can adjust the intensity of displayed waveforms to account for various signal characteristics such as fast time div settings and low trigger rates Increasing the intensity lets you see the maximum amount of noise and infrequently occurring events Reducing the intensity can expose more detail in complex signals as shown in the following figures 1 Press the Intensity key to illuminate it This key is located just below the Entry knob 2 Turn the Entry knob to adjust the waveform instensity Waveform intensity adjustment affects analog channel waveforms only not math waveforms referencewaveforms digital waveforms etc a Agilent Technologies me 7 Display Settings Agilent Acquisiti Channels 10 0 1 3 Figure 13 Amplitude Modulation Shown at 100 Intensity Agilent Acquisition Channels 10 0 1 10 0 1 Edge Figure 14 Amplitude Modulation S
97. cope probes or BNC cables to these BNC connectors In the InfiniiVision 2000 X Series oscilloscopes the analog channel inputs have 1 MQ impedance Also there is no automatic probe detection so you must properly set the probe attenuation for accurate measurement results See To specify the probe attenuation on page 60 18 Demo 2 Ground and Demo 1 terminals Demo 2 terminal This terminal outputs the Probe Comp signal which helps you match a probe s input capacitance to the oscilloscope channel to which it is connected See Compensate Passive Probes on page 27 With certain licensed features the oscilloscope can also output demo or training signals on this terminal Ground terminal Use the ground terminal for oscilloscope probes connected to the Demo 1 or Demo 2 terminals Demo 1 terminal With certain licensed features the oscilloscope can output demo or training signals on this terminal 34 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 20 USB Host port Digital channel inputs This port is for connecting USB mass storage devices or printers to the oscilloscope Connect a USB compliant mass storage device flash drive disk drive etc to save or recall oscilloscope setup files and reference waveforms or to save data and screen images See Chapter 17 Save Recall Setups Screens Data starting on page 217 To print connect a
98. creen 231 print landscape 234 print Quick Print 253 printer USB 35 231 printing the display 231 probe attenuation 60 probe attenuation external trigger 141 probe compensation 34 probe units 60 probes 276 278 probes connecting to oscilloscope 24 probes current 277 probes differential 277 probes digital 81 probes passive 276 probes passive compensating 27 programmer s guide 265 pulse polarity 122 pulse waveform generator output 207 pulse width trigger 121 Q qualifier pulse width 122 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide uick Action key 30 253 uick Clear Display 253 uick Freeze Display 253 uick Help 40 uick Help language 41 uick Measure All 253 uick Print 253 uick Recall 253 uick Save 253 uick Trigger Mode 253 Cooeovooo00Oo ramp waveform generator output 207 random noise 135 ratio X cursor units 162 ratio Y cursor units 162 raw acquisition record 224 Real Scope Remote Front Panel 260 rear panel connectors 37 re arm time 158 recall 253 recall files via web interface 266 recall mask files 228 recall setups 227 recall Quick Recall 253 Rectangular FFT window 70 Ref key 32 77 reference point waveform 243 reference waveforms 77 remote control 237 Remote Front Panel 263 remote programming Agilent 10 Libraries 264 remote programming web interface 263 replacement parts 98 req
99. d The oscilloscope s MegaZoom IV technology gives you bright easily viewed displays of any part of the video waveform Analysis of video waveforms is simplified by the oscilloscope s ability to trigger on any selected line of the video signal Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 123 9 Triggers It is important when using a 10 1 passive probe that the probe is correctly compensated The oscilloscope is sensitive to this and will not trigger if the probe is not properly compensated especially for progressive formats 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 200 08 Vid Field Agilent Acquisition val Channels Source Settings Video ae 3 Press the Source softkey and select any analog channel as the video trigger source The selected trigger source is displayed in the upper right corner of the display Turning the Trigger Level knob does not change the trigger level because the trigger level is automatically set to the sync pulse Trigger coupling is automatically set to TV in the Trigger Mode and Coupling Menu 124 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 Provide Correct Matching Many video signals are produced from 75 Q sources To provide correct matching to these sources a 75 Q terminator such as an Agilent 11094B should be connected to the oscilloscope input 4 Pres
100. d DEVICE is provided for controlling the oscilloscope over USB See the Programmer s Guide for more information The oscilloscope s internal file system under Agilent Flash consists of 10 locations for oscilloscope setup files and four locations for mask files To use the File Explorer 1 Press Utility gt File Explorer 2 In the File Explorer Menu press the softkey in the first position and use the Entry knob to navigate Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 241 19 Utility Settings 500 08 2 Agilent ition al MEMAN i E o B a a Ba E a B Oct 01 2010 Filip Locatio SM502000 The softkey in the first position can have these labels Press to go when you can push the Entry knob to navigate to a new folder or storage location Location when pointing to a directory that is currently selected Selected when pointing to a file that can be loaded or deleted When this label appears you can press the Load File or Delete File softkeys to take the action Pushing the Entry knob is the same as pressing the Load File softkey A file that has been deleted from a USB storage device cannot be recovered by the oscilloscope Use your PC to create directories on a USB storage device USB Storage Most USB mass storage devices are compatible with the oscilloscope Devices However certain devices may be incompatible and may n
101. d analog channels are included in the mask test When disabled just the selected source channel is included in the test Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 193 14 Mask Testing Mask Statistics From the Mask Test Menu press the Statistics softkey to enter the Mask Statistics Menu Agilent Acquisition i Channels Show Stats When you enable Show Statistics the following information is displayed Current mask name of mask Channel number date and time of Tests total number of mask tests executed Status Passing Failing or Untested e Accumulated test time in hours minutes seconds and tenths of seconds And for each analog channel Number of failures acquisitions in which the signal excursion went beyond the mask e Failure rate percentage of failures Sigma the ratio of process sigma to maximum achievable sigma based on number of waveforms tested 194 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Mask Testing 14 Reset Statistics Note that statistics are also reset when Mask Test is switched on after being switched off Clear Mask softkey is pressed e An Automask is created Additionally the accumulated time counter is reset whenever the oscilloscope is run after the acquisition was stopped Transparent Enable Transparent mode to write measurement values and statistics on the screen without
102. d so the oscilloscope s feet and legs do not pinch the cord Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 23 1 Getting Started 2 The oscilloscope automatically adjusts for input line voltages in the range 100 to 240 VAC The line cord provided is matched to the country of origin Always use a grounded power cord Do not defeat the power cord ground 3 Press the power switch The power switch is located on the lower left corner of the front panel The oscilloscope will perform a self test and will be operational in a few seconds Connect Probes to the Oscilloscope CAUTION 1 Connect the oscilloscope probe to an oscilloscope channel BNC connector 2 Connect the probe s retractable hook tip to the point of interest on the circuit or device under test Be sure to connect the probe ground lead to a ground point on the circuit A Maximum input voltage at analog inputs CAT 300 Vrms 400 Vpk transient overvoltage 1 6 kVpk With 10073C 10 1 probe CAT 500 Vpk CAT II 400 Vpk With N2862A or N2863A 10 1 probe 300 Vrms CAUTION A Do not float the oscilloscope chassis Defeating the ground connection and floating the oscilloscope chassis will probably result in inaccurate measurements and may also cause equipment damage The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord If you need to measure between two live points use a differential probe with suffic
103. d to select items from menus and to change values The function of the Entry knob changes based upon the current menu and softkey selections Note that the curved arrow symbol 4 above the entry knob illuminates whenever the entry knob can be used to select a value Also note that when the Entry knob a symbol appears on a softkey you can use the Entry knob to select values Often rotating the Entry knob is enough to make a selection Sometimes you can push the Entry knob to enable or disable a selection Pushing the Entry knob also makes popup menus disappear Tools keys Trigger controls The Tools keys consist of Utility key Press this key to access the Utility Menu which lets you configure the oscilloscope s I O settings use the file explorer set preferences access the service menu and choose other options See Chapter 19 Utility Settings starting on page 237 Quick Action key Press this key to perform the selected quick action measure all snapshot print save recall freeze display and more See Configuring the Quick Action Key on page 253 Analyze key Press this key to access analysis features like mask testing see Chapter 14 Mask Testing starting on page 189 trigger level setting measurement threshold setting or Video trigger automatic set up and display Wave Gen key Press this key to access waveform generator functions See Chapter 16 Wavefo
104. data points in length When saving waveform data the save times depend on the chosen format Data File Format Save Times BIN ALB fastest ASCII XY medium CSV slowest See Also Binary Data bin Format on page 281 e CSV and ASCII XY files on page 287 e Minimum and Maximum Values in CSV Files on page 288 224 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 17 To save reference waveform files to a USB storage device 1 Press the Save Recall key 2 In the Save Recall Menu press the Save softkey 3 In the Save Menu press the Format softkey and turn the Entry knob to select Reference Waveform data h5 4 Press the Source softkey and turn the Entry knob to select the source waveform 5 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 226 6 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed To save masks Mask files can be saved to one of four internal Agilent Flash locations or to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select Mask msk 2 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 226 3 Finally press the Press to S
105. digital waveform display The following figure shows a typical display with digital channels 86 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 Delay Time Trigger Trigger time div mode or type and run status source 0s Threshold Agilent level Acquisition Channels i we Digital lt _ Activity channel indicators identifiers KELELE Turn off Thresholds Bus D7 DO v w Waveform Turn Turn Threshold size individual groups of menu key channels channels on off on off Activity indicator When any digital channels are turned on an activity indicator is displayed in the status line at the bottom of the display A digital channel can be always high Cy always low or actively toggling logic states 1 Any channel that is turned off will be grayed out in the activity indicator To change the displayed size of the digital channels 1 Press the Digital key 2 Press the size n JL Ib softkey to select how the digital channels are displayed Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 87 6 Digital Channels The sizing control lets you spread out or compress the digital traces vertically on the display for more convenient viewing To switch a single channel on or off 1 With the Digital Channel Menu displayed rotate the Entry knob to select the desired channel from the popup menu 2 Push the Entry knob or press the softkey that is directly below the pop
106. e 136 notices 2 Nyquist frequency 74 Nyquist sampling theory 145 0 operators math 66 options print 234 oscilloscope bandwidth 146 oscilloscope bandwidth required 148 oscilloscope rise time 147 oscilloscope sample rate 147 output load expected waveform generator 209 output trigger 246 overlay localized 35 Overshoot measurement 172 176 overvoltage category 275 P palette 220 pan and zoom 44 parts replacement 98 passive probes 276 passive probes compensating 27 password network reset 271 password network setting 269 pattern trigger 118 PC connection 240 peak detect mode 149 150 151 Peak peak measurement 174 Period measurement 181 persistence 103 persistence clearing 104 persistence infinite 144 Phase measurement 172 185 phase X cursor units 162 PNG file format 218 point to point connection 240 pollution degree 275 pollution degree definitions 275 Portuguese front panel overlays 36 Portuguese user interface and Quick Help 41 position digital channels 89 position knob 89 position analog 57 post processing 169 post trigger information 45 power consumption 23 power cord connector 37 power requirements 23 power supply 37 power switch 24 29 power on 23 predefined labels 108 Preshoot measurement 172 177 Press to go 226 242 pre trigger information 45 print 253 Print key 33 print options 234 print s
107. e This lets you type on a keyboard rather than edit the label list using the oscilloscope s controls You can create a list of up to 75 labels and load it into the oscilloscope Labels are added to the beginning of the list If more than 75 labels are loaded only the first 75 are stored To load labels from a text file into the oscilloscope Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Labels 8 1 Use a text editor to create each label Each label can be up to ten characters in length Separate each label with a line feed 2 Name the file labellist txt and save it on a USB mass storage device such as a thumb drive 3 Load the list into the oscilloscope using the File Explorer press Utility gt File Explorer Label List Management When you press the Library softkey you will see a list of the last 75 labels used The list does not save duplicate labels Labels can end in any number of trailing digits As long as the base string is the same as an existing label in the library the new label will not be put in the library For example if label A0 is in the library and you make a new label called A12345 the new label is not added to the library When you save a new user defined label the new label will replace the oldest label in the list Oldest is defined as the longest time since the label was last assigned to a channel Any time you assign any label to a channel that label will move to the newest in the lis
108. e waveform The bandwidth limit also limits the trigger signal path of any channel that has BW Limit turned on To change the vertical scale knob s coarse fine adjustment setting 1 Push the channel s vertical scale knob or press the channel key and then the Fine softkey in the Channel Menu to toggle between fine and coarse adjustment of the vertical scale 58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Vertical Controls 3 When Fine adjustment is selected you can change the channel s vertical sensitivity in smaller increments The channel sensitivity remains fully calibrated when Fine is on The vertical scale value is displayed in the status line at the top of the display When Fine is turned off turning the volts division knob changes the channel sensitivity in a 1 2 5 step sequence To invert a waveform 1 Press the desired channel key 2 In the Channel Menu press the Invert softkey to invert the selected channel When Invert is selected the voltage values of the displayed waveform are inverted Invert affects how a channel is displayed However when using basic triggers the oscilloscope attempts to maintain the same trigger point by changing trigger settings Inverting a channel also changes the result of any math function selected in the Waveform Math Menu or any measurement Setting Analog Channel Probe Options 1 Press the probe s associated channel key 2 In the Channel Menu press the Pr
109. e 137 10 Trigger Mode Coupling e To position the time reference left center right on page 52 To select the trigger coupling 138 1 Press the Mode Coupling key 2 In the Trigger Mode and Coupling Menu press the Coupling softkey then turn the Entry knob to select DC coupling allows DC and AC signals into the trigger path AC coupling places a 10 Hz high pass filter in the trigger path removing any DC offset voltage from the trigger waveform The high pass filter in the External Trigger input path is 50 Hz for all models Use AC coupling to get a stable edge trigger when your waveform has a large DC offset LF low frequency Reject coupling adds a high pass filter with the 3 dB point at 50 kHz in series with the trigger waveform 0 dB 3 dB down point Pass Band DC 50 kHz Low frequency reject removes any unwanted low frequency components from a trigger waveform such as power line frequencies etc that can interfere with proper triggering Use LF Reject coupling to get a stable edge trigger when your waveform has low frequency noise TV coupling is normally grayed out but is automatically selected when TV trigger is enabled in the Trigger Menu Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Trigger Mode Coupling 10 Note that Trigger Coupling is independent of Channel Coupling see To specify channel coupling on page 57 To enable or disable trigger
110. e information on settings that affect record length see Length Control on page 223 144 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control 11 Overview of Sampling To understand the oscilloscope s sampling and acquisition modes it is helpful to understand sampling theory aliasing oscilloscope bandwidth and sample rate oscilloscope rise time oscilloscope bandwidth required and how memory depth affects sample rate Sampling Theory The Nyquist sampling theorem states that for a limited bandwidth band limited signal with maximum frequency fmax the equally spaced sampling frequency fg must be greater than twice the maximum frequency fmax in order to have the signal be uniquely reconstructed without aliasing fmax fs 2 Nyquist frequency fy folding frequency Aliasing Aliasing occurs when signals are under sampled fg lt 2fyax Aliasing is the signal distortion caused by low frequencies falsely reconstructed from an insufficient number of sample points Figure 21 Aliasing Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 145 11 Acquisition Control Oscilloscope Bandwidth and Sample Rate An oscilloscope s bandwidth is typically described as the lowest frequency at which input signal sine waves are attenuated by 3 dB 30 amplitude error At the oscilloscope bandwidth sampling theory says the required sample rate is fg 2fpw However the theory assu
111. e input voltage at each input probe When the trigger conditions are met the oscilloscope triggers and displays the acquisition For digital channels each time the oscilloscope takes a sample it compares the input voltage to the logic threshold If the voltage is above the threshold the oscilloscope stores a 1 in sample memory otherwise it stores a 0 To display digital channels using AutoScale When signals are connected to the digital channels be sure to connect the ground leads AutoScale quickly configures and displays the digital channels e To configure the instrument quickly press the AutoScale key 100 08 Agilent Acquisition h A enu Undo Fast Debug Channels Acq Mode Autoscale All Normal Figure 8 Example AutoScale of digital channels MSO models only Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 85 6 Digital Channels Any digital channel with an active signal will be displayed Any digital channels without active signals will be turned off e To undo the effects of AutoScale press the Undo AutoScale softkey before pressing any other key This is useful if you have unintentionally pressed the AutoScale key or do not like the settings AutoScale has selected This will return the oscilloscope to its previous settings See also How AutoScale Works on page 27 To set the instrument to the factory default configuration press the Default Setup key Interpreting the
112. e number of FFT points fs N With a fixed number of FFT points up to 65 536 the lower the sampling rate the better the resolution Decreasing the effective sampling rate by selecting a greater time div setting will increase the low frequency resolution of the FFT display and also increase the chance that an alias will be displayed The resolution of the FFT is the effective sample rate divided by the number of points in the FFT The actual resolution of the display will not be this fine as the shape of the window will be the actual limiting factor in the FFTs ability to resolve two closely space frequencies A good way to test the ability of the FFT to resolve two closely spaced frequencies is to examine the sidebands of an amplitude modulated sine wave For the best vertical accuracy on peak measurements e Make sure the probe attenuation is set correctly The probe attenuation is set from the Channel Menu if the operand is a channel Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Math Waveforms 4 e Set the source sensitivity so that the input signal is near full screen but not clipped e Use the Flat Top window e Set the FFT sensitivity to a sensitive range such as 2 dB division For best frequency accuracy on peaks e Use the Hanning window e Use Cursors to place an X cursor on the frequency of interest e Adjust frequency span for better cursor placement e Return to the Cursors Menu to fine tune the
113. e oscilloscope s display from the web interface 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface is displayed select the Get Image tab from the left side of the Welcome screen After a delay of several seconds the oscilloscope s screen image will be displayed 3 Right click on the image and select Save Picture As or Print Picture 4 Select a storage location for the image file and click Save Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 267 20 Web Interface Identification Function The Identification web interface feature is useful when trying to locate a specific instrument in a rack of equipment 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface Welcome Page is displayed select the Identification on radio button An Identify message is displayed on the oscilloscope you can either select Identification off or press the OK softkey on the oscilloscope to continue Support Products Agilent Site Another web enabled instrument from Agilent Technologies Welcome to your L 1 Web Enabled Oscilloscope MSO X 2024A Information about this Web Enabled Instrument Agilent Technologies KOSICE S3 Instrument MSO X 2024A Oscilloscope Agilent MSO X 2024A US50210029 atten alton 130 29 70 169 TOPIPO a mx2024a 100
114. e you create 110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide To reset the label library to the factory default 111 9 Triggers Adjusting the Trigger Level 114 Forcing a Trigger 115 Edge Trigger 115 Pattern Trigger 118 Hex Bus Pattern Trigger 120 Pulse Width Trigger 121 Video Trigger 123 To trigger on a specific line of video 127 To trigger on all sync pulses 128 To trigger on a specific field of the video signal 129 To trigger on all fields of the video signal 130 To trigger on odd or even fields 131 10 Trigger Mode Coupling To select the Auto or Normal trigger mode 136 To select the trigger coupling 138 To enable or disable trigger noise rejection 139 To enable or disable trigger HF Reject 139 To set the trigger holdoff 140 External Trigger Input 140 Maximum voltage at oscilloscope external trigger input 141 11 Acquisition Control Running Stopping and Making Single Acquisitions Run Control 143 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Overview of Sampling 145 Sampling Theory 145 Aliasing 145 Oscilloscope Bandwidth and Sample Rate 146 Oscilloscope Rise Time 147 Oscilloscope Bandwidth Required 148 Memory Depth and Sample Rate 149 Selecting the Acquisition Mode 149 Normal Acquisition Mode 150 Peak Detect Acquisition Mode 150 Averaging Acquisition Mode 153 High Resolution Acquisition Mode 155 Acquiring to Segmented Memory 155 Navigating Segments 157 Infinite P
115. e your selection either push the Cursors knob again or wait about five seconds for the popup menu to disappear Or Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 161 12 Cursors 162 Press the Cursors softkey then turn the Entry knob The X1 X2 linked and Y1 Y2 linked selections let you adjust both cursors at the same time while the delta value remains the same This can be useful for example for checking pulse width variations in a pulse train The currently selected cursor s display brighter than the other cursors To change the cursor units press the Units softkey In the Cursor Units Menu Use X Cursors Y Units As 360 Base You can press the X Units softkey to select Seconds s Hz 1 s Phase when selected use the Use X Cursors softkey to set the current X1 location as 0 degrees and the current X2 location as 360 degrees Ratio when selected use the Use X Cursors softkey to set the current X1 location as 0 and the current X2 location as 100 You can press the Y Units softkey to select Base the same units used for the source waveform Ratio when selected use the Use Y Cursors softkey to set the current Y1 location as 0 and the current Y2 location as 100 For phase or ratio units once the 0 and 360 degree or 0 and 100 locations are set adjusting cursors will display measurements relative to the set locations 7 Adjust the selected cursor s by
116. efault the oscilloscope is not password protected See Setting a Password on page 269 to set a password Any time you modify the oscilloscope s hostname it breaks the connection between the oscilloscope and the LAN You need to re establish communication to the oscilloscope using the new hostname To establish a LAN connection Automatic 1 Press Utility gt 1 0 Configuration Press the LAN Settings softkey N Press the Config softkey then turn the Entry knob to select Automatic and press the softkey again to enable it If your network supports DHCP or AutoIP enabling Automatic lets the oscilloscope use those services to get its LAN configuration settings 4 If your network provides Dynamic DNS you can enable the Dynamic DNS option to let the oscilloscope register its hostname and use the DNS server for name resolution 5 You can enable the Multicast DNS option to let the oscilloscope use Multicast DNS for name resolution on small networks without a conventional DNS server 6 Connect the oscilloscope to the local area network LAN by inserting the LAN cable into the LAN port on the rear panel of the oscilloscope In a few moments the oscilloscope will connect to the network automatically If the oscilloscope does not automatically connect to the network press Utility gt 1 0 gt LAN Reset In a few moments the oscilloscope will connect to the network Manual 1 Get the oscilloscope s network parameters host n
117. el key or softkey or right click softkey when using web browser remote front panel Quick Help remains on the screen until another key is pressed or a knob is turned 40 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide To select the user interface and Quick Help language Getting Started To select the user interface and Quick Help language 1 Press Help then press the Language softkey 2 Repeatedly press and release the Language softkey or rotate the Entry knob until the desired language is selected The following languages are available English French German Italian Japanese Korean Portuguese Russian Simplified Chinese Spanish and Traditional Chinese Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 41 1 42 Getting Started Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls To adjust the horizontal time div scale 44 To adjust the horizontal delay position 45 Panning and Zooming Single or Stopped Acquisitions 46 To change the horizontal time mode Normal XY or Roll 47 To display the zoomed time base 50 To change the horizontal scale knob s coarse fine adjustment setting 52 To position the time reference left center right 52 Navigating the Time Base 53 The horizontal controls include e The horizontal scale and position knobs e The Horiz key
118. elect a bus Rotate the Entry knob then press the Entry knob or the Bus1 Bus2 softkey to switch it on Use the Channel softkey and the Entry knob to select individual channels to be included in the bus You can rotate the Entry knob and push it or push the softkey to select channels You can also press the Select Deselect D7 D0 softkey to include or exclude groups of eight channels in each bus Digital Bus Menu D ttt ttt D Bus Channel Deselect Base Bust D7 DO Hex If the bus display is blank completely white or if the display includes you need to expand the horizontal scale to allow space for the data to be shown or use the cursors to display the values see Using cursors to read bus values on page 91 The Base softkey lets you choose to display the bus values in hex or binary The buses are shown at the bottom of the display 90 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 Agilent tion Channels 1101 111 1110 0100 1110 0110 X 1110 0100 X11 D ttt t D Deselect Base D7 DO Binary Bus values can be shown in hex or binary Using cursorsto To read the digital bus value at any point using the cursors read bus values 4 Turn on Cursors by pressing the Cursors key on the front panel 2 Press the cursor Mode softkey and change the mode to Hex or Binary 3 Press the Source softkey and select Bus1 or Bus2 4 Use the Entry knob and the X1 and X2 softkeys to p
119. elete the character You can use a connected USB keyboard instead of using the Spell and other character editing softkeys Use the Clear softkey to delete all annotation characters Press OK to save the annotation edits 6 Press the Text Color softkey and turn the Entry knob to select the annotation color You can choose white red or colors that match analog channels digital channels math waveforms reference waveforms or markers 7 Press the Background softkey and turn the Entry knob to select the annotation background Opaque the annotation has a solid background Inverted the annotation s foreground and background colors are switched Transparent the annotation has a transparent background Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 255 19 Utility Settings 1 5005 500 08 1 70V Channels Annotation Menu Annotation Text Color Background Inverted See Also To save BMP or PNG image files on page 219 e To print the oscilloscope s display on page 231 256 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ee 20 6 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Web Interface Accessing the Web Interface 258 Browser Web Control 259 Save Recall 265 Get Image 267 Identification Function 268 Instrument Utilities 268 Setting a Password 269 When the Agilent InfiniiVision X Series oscilloscopes have the DSOXLAN LAN VGA option module i
120. ement Menu 2 Press the Source softkey then turn the Entry knob to select the first analog channel source 3 Press the Type softkey then turn the Entry knob to select Delay 4 Press the Settings softkey to select the second analog channel source and slope for the delay measurement The default Delay settings measure from the rising edge of channel 1 to the rising edge of channel 2 5 Press the Back Up key to return to the Measurement Menu 6 Press the Add Measurement softkey to make the measurement The example below shows a delay measurement between the rising edge of channel 1 and the rising edge of channel 2 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 1 00V 1 00V 500 08 2 08V Agilent on Channels x Measurements Freq 1 499 65Hz Pk Pk 1 Delay 14 24 191 0us Measurement Menu Source Add Settings Clear Meas Measurement lt gt Phase Phase is the calculated phase shift from source 1 to source 2 expressed in degrees Negative phase shift values indicate that the rising edge of source 1 occurred after the rising edge of source 2 Delay Phase Source 1 Period i4 Period gt Source 1 _ fy ff S lt Delay Source 2 NO ee eee 1 Press the Meas key to display the Measurement Menu 2 Press the Source softkey then turn the Entry knob to select the first analog channel source 3 Press the Type softkey then turn the Entry knob t
121. enable or disable trigger HF Reject on page 139 low frequency rejection To select the trigger coupling on page 138 or To enable or disable trigger noise rejection on page 139 3 Use Averaging Acquisition Mode on page 153 to reduce noise on the displayed waveform a Agilent Technologies 10 Trigger Mode Coupling To select the Auto or Normal trigger mode 136 Triggering and the Pre and Post Trigger Buffers When the oscilloscope is running the trigger mode tells the oscilloscope what to do when triggers are not occurring In the Auto trigger mode the default setting if the specified trigger conditions are not found triggers are forced and acquisitions are made so that signal activity is displayed on the oscilloscope In the Normal trigger mode triggers and acquisitions only occur when the specified trigger conditions are found To select the trigger mode 1 Press the Mode Coupling key 2 In the Trigger Mode and Coupling Menu press the Mode softkey then select either Auto or Normal See the following When to Use Auto Trigger Mode on page 137 and When to Use Normal Trigger Mode on page 137 descriptions You can also configure the Quick Action key to toggle between the Auto and Normal trigger modes See Configuring the Quick Action Key on page 253 After the oscilloscope starts running after pressing Run or Single or changing the trigger condition the oscilloscope first fills the pre
122. enabled Also the oscilloscope can be controlled via multiple I O interfaces for example USB and LAN at the same time e Chapter 20 Web Interface starting on page 257 when the oscilloscope is connected to a LAN e Remote Programming via the Web Interface on page 263 e The oscilloscope s Programmer s Guide e Remote Programming with Agilent IO Libraries on page 264 Setting up the Oscilloscope s LAN Connection 238 When the DSOXLAN LAN VGA module is installed you can place the oscilloscope on the network and set up its LAN connection Once that is done you can use the oscilloscope s web interface or remotely control the oscilloscope via the LAN interface The oscilloscope supports methods for automated LAN configuration or manual LAN configuration see To establish a LAN connection on page 239 It is also possible to set up a point to point LAN connection between a PC and the oscilloscope see Stand alone Point to Point Connection to a PC on page 240 Once the oscilloscope is set up on the network you can use the oscilloscope s web page to view or change its network configuration and access additional settings like the network password See Chapter 20 Web Interface starting on page 257 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 When you connect the oscilloscope to a LAN it is a good practice to limit access to the oscilloscope by setting a password By d
123. ented memory acquisitions 5 Press the of Segs softkey and turn the Entry knob to select the number of segments into which you would like to divide the oscilloscope s memory Memory can be divided into as few as two segments and as many as 25 segments 6 Press the Run or Single key The oscilloscope runs and fills a memory segment for each trigger event When the oscilloscope is busy acquiring multiple segments the progress is displayed in the upper right area of the display The oscilloscope continues to trigger until memory is filled then the oscilloscope stops If the signal you are measuring has more than about 1 s of inactivity consider selecting the Normal trigger mode to prevent AutoTriggering See To select the Auto or Normal trigger mode on page 136 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control 11 1469 Segment 3 Agilent Acquisi i Progress indicator Sample rate 3 0 0s Current Seg of Segs 1 25 See Also Navigating Segments on page 157 Infinite Persistence with Segmented Memory on page 157 e Segmented Memory Re Arm Time on page 158 e Saving Data from Segmented Memory on page 158 Navigating Segments 1 Press the Current Seg softkey and turn the Entry knob to display the desired segment along with a time tag indicating the time from the first trigger event You can also navigate segments using the Navigate key and controls See To n
124. erfaces Connectivity Guide For a printable electronic copy of the Connectivity Guide direct your Web browser to www agilent com and search for Connectivity Guide Save Recall You can save setup files screen images waveform data files or mask files to your PC via the oscilloscope s web interface see Saving Files via the Web Interface on page 265 You can recall setup files reference waveform data files or mask files from your PC via the oscilloscope s web interface see Recalling Files via the Web Interface on page 266 Saving Files via the Web Interface To save setup files screen images waveform data files or mask files to your PC via the oscilloscope s web interface 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface is displayed select the Save Recall tab from the left side of the Welcome screen 3 Click the Save link 4 On the Save page a Enter the name of the file you are saving to b Select the format Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 265 20 Web Interface Support H Agilent Technologies KOSICE ISE fle J Page i en E irr E J ie ee cokol C New Acquisition nT am n bit Bitmap image bmp Save Recall 24 bit Bitmap image bmp PNG 24 bit image png e a CSV data csv 5 mage ASCII XY data csv ull Reference Waveform data n5 A
125. eriod are averaged and the average value is stored Use this mode for reducing random noise See High Resolution Acquisition Mode on page 155 Normal Acquisition Mode In Normal mode at slower time div settings extra samples are decimated in other words some are thrown away This mode yields the best display for most waveforms Peak Detect Acquisition Mode In Peak Detect mode at slower time div settings when decimation would normally occur minimum and maximum value samples are kept in order to capture infrequent and narrow events at the expense of exaggerating any noise This mode displays all pulses that are at least as wide as the sample period For InfiniiVision 2000 X Series oscilloscopes which have a maximum sample rate of 2 GSa s a sample is taken every 500 ps sample period See Also Glitch or Narrow Pulse Capture on page 151 e Using Peak Detect Mode to Find a Glitch on page 152 150 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control 11 Glitch or Narrow Pulse Capture A glitch is a rapid change in the waveform that is usually narrow as compared to the waveform Peak detect mode can be used to more easily view glitches or narrow pulses In peak detect mode narrow glitches and sharp edges are displayed more brightly than when in Normal acquire mode making them easier to see To characterize the glitch use the cursors or the automatic measurement capabilities of the osc
126. errors or for incidental or consequential damages in connec tion with the furnishing use or per formance of this document or of any information contained herein Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the sep arate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accor dance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcon tract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or asa commercial item as defined in FAR 2 101 a or as Restricted computer soft ware as defined in FAR 52 227 19 June 1987 or any equivalent agency regulation or contract clause Use duplication or disclo sure of Software is subject to Agilent Tech nologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will receive no greater than Restricted Rights as defined in FAR 52 227 19 c 1 2 June 1987 U S Government users will receive no greater than Limited Rights as defined in FAR 52 227 14 June 1987 or DFAR 252 227 7015 b 2 November 1995 as applicable in any techn
127. ersistence with Segmented Memory 157 Segmented Memory Re Arm Time 158 Saving Data from Segmented Memory 158 12 Cursors To make cursor measurements 160 Cursor Examples 163 13 Measurements To make automatic measurements 1 0 Measurements Summary 171 Snapshot All 173 Voltage Measurements 173 Peak Peak 174 Maximum 174 Minimum 174 Amplitude 174 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Top 175 Base 1 6 Overshoot 176 Preshoot 177 Average 178 DCRMS 178 ACRMS 179 Time Measurements 181 Period 181 Frequency 182 Width 183 Width 183 Duty Cycle 183 Rise Time 183 Fall Time 183 Delay 184 Phase 185 Measurement Thresholds 186 Measurement Window with Zoom Display 188 14 Mask Testing To create a mask from a golden waveform Automask 189 Mask Test Setup Options 191 Mask Statistics 194 To manually modify a mask file 195 Building a Mask File 198 How is mask testing done 200 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 14 15 Digital Voltmeter 16 Waveform Generator To select generated waveform types and settings 205 To output the waveform generator sync pulse 208 To specify the expected output load 209 To use waveform generator logic presets 209 To add noise to the waveform generator output 210 To add modulation to the waveform generator output 210 To set up Amplitude Modulation AM 211 To set up Frequency Modulation FM 213 To set up Frequency Shift Keying
128. erved not erased but all channel labels will be set to their original names Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 1 Getting Started In the Save Recall Menu there are also options for restoring the complete factory settings see Recalling Default Setups on page 229 or performing a secure erase see Performing a Secure Erase on page 229 Use Auto Scale Use Auto Scale to automatically configure the oscilloscope to best display the input signals 1 Press Auto Scale You should see a waveform on the oscilloscope s display similar to this 200 08 Agilent Acquisition Channels Fast Debug Channels Acq Mode Autoscale All Normal 2 If you want to return to the oscilloscope settings that existed before press Undo AutoScale 3 If you want to enable fast debug autoscaling change the channels autoscaled or preserve the acquisition mode during autoscale press Fast Debug Channels or Acq Mode These are the same softkeys that appear in the AutoScale Preferences Menu See To set AutoScale preferences on page 245 26 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 If you see the waveform but the square wave is not shaped correctly as shown above perform the procedure Compensate Passive Probes on page 27 If you do not see the waveform make sure the probe is connected securely to the front panel channel input BNC and to
129. es User s Guide Getting Started 1 5 Tools keys _ 6 Trigger controls 7 Horizontal controls 8 Run Control keys oe 10 Auto Scale key 11 Additional waveform controls 4 Entry knob 3 Intensity key 12 Measure controls 2 Softkeys 13 Waveform keys 14 File keys 15 Help key 21 Waveform 20 Digital 19 USB 18 Demo 2 Ground 17 Analog 16 Vertical controls 1 Power switch generator channel Host and Demo 1 channel output inputs port terminals inputs 1 Power switch Press once to switch power on press again to switch power off See Power On the Oscilloscope on page 23 2 Softkeys The functions of these keys change based upon the menus shown on the display directly above the keys The O Back Up key moves up in the softkey menu hierarchy At the top of the hierarchy the Back Up key turns the menus off and oscilloscope information is shown instead 3 Intensity key Press the key to illuminate it When illuminated turn the Entry knob to adjust waveform intensity You can vary the intensity control to bring out signal detail much like an analog oscilloscope Digital channel waveform intensity is not adjustable More details about using the Intensity control to view signal detail are on To adjust waveform intensity on page 101 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 29 1 30 Getting Started Entry knob The Entry knob is use
130. ess to Save softkey A message indicating whether the save was successful is displayed See Also Binary Data bin Format on page 281 e CSV and ASCII XY files on page 287 e Minimum and Maximum Values in CSV Files on page 288 To save ALB data files Agilent logic analyzer binary ALB format data files can be viewed and analyzed offline on a PC using the Agilent Logic Analyzer application and the Agilent B4610A Data Import Tool See the Agilent Technologies B4610A Data Import Tool for Offline Viewing and Analysis Data Sheet publication number 5989 7834EN ALB data files can be saved to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select ALB data alb Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 226 Press the Settings softkey In the File Settings Menu you have these softkeys and options Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 221 17 Save Recall Setups Screens Data Setup Info when enabled setup information vertical horizontal trigger acquisition math and display settings is also saved in a separate file with a TXT extension Length sets the number of data points that will be output to the file For more information see Length Control on page 223 Alb Format Available ALB serial formats include
131. f the box The Horizontal position delay time knob sets the left to right position of the zoom window The delay value which is the time displayed relative to the trigger point is momentarily displayed in the upper right portion of the display when the delay time lt knob is turned Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 51 2 Horizontal Controls Negative delay values indicate you re looking at a portion of the waveform before the trigger event and positive values indicate you re looking at the waveform after the trigger event To change the time div of the normal window turn off Zoom then turn the horizontal scale sweep speed knob For information about using zoom mode for measurements refer to To isolate a pulse for Top measurement on page 175 and To isolate an event for frequency measurement on page 182 To change the horizontal scale knob s coarse fine adjustment setting 1 Push the horizontal scale knob or press Horiz gt Fine to toggle between fine and coarse adjustment of the horizontal scale When Fine is enabled turning the horizontal scale knob changes the time div displayed in the status line at the top of the display in smaller increments The time div remains fully calibrated when Fine is on When Fine is turned off the Horizontal scale knob changes the time div setting in a 1 2 5 step sequence To position the time reference left center right 52 Time refe
132. for accessing the Horizontal Menu e The zoom key for quickly enabling disabling the split screen zoom display The Search key for finding events on analog channels e The Navigate keys for navigating time search events or segmented memory acquisitions The following figure shows the Horizontal Menu which appears after pressing the Horiz key ie Agilent Technologies 43 2 Horizontal Controls Trigger Time Delay Time Trigger Trigger level point reference time div source or threshold N 100 05 Agilent Acquisition Sample rate Channels XY or Roll LULUN UU mode x Il IUU Time Mode Time Ref Normal Center Normal Zoomed Fine Time time mode time base control reference Figure 2 Horizontal Menu The Horizontal Menu lets you select the time mode Normal XY or Roll enable Zoom set the time base fine control vernier and specify the time reference The current sample rate is displayed above the Fine and Time Ref softkeys To adjust the horizontal time div scale 1 Turn the large horizontal scale sweep speed knob marked N Ag to change the horizontal time div setting 44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Horizontal Controls 2 Notice how the time div information in the status line changes The V symbol at the top of the display indicates the time reference point The horizontal scale knob works in the Normal time mode while acquisi
133. form test 189 GPIB address 238 GPIB interface remote control 237 GPIB module 22 37 GPIB module installation 22 grabber 83 84 graphical user interface language 41 graticule intensity 105 graticule type 104 grid intensity 105 grid type 104 ground level 56 Ground terminal 34 Hanning FFT window 70 hardware self test 250 Help key 33 help built in 40 hex bus trigger 120 Hex softkey 120 HF Reject 139 high frequency noise rejection 139 high resolution mode 149 155 holdoff 140 hop frequency FSK modulation 214 Horiz key 31 43 48 50 152 Horizontal controls 31 47 horizontal Navigate key 31 horizontal position control 31 horizontal position knob 45 horizontal scale fine adjustment 52 horizontal Search key 31 horizontal sweep speed control 31 horizontal time div control 31 horizontal Zoom key 31 host name 239 257 Host name softkey 240 1 0 interface settings 237 identification function web interface 268 impedance digital probes 93 indeterminate state 161 infinite persistence 103 144 151 information area 39 installed licenses 251 installed options 268 Instrument Utilities web page 268 intensity control 101 Intensity key 29 invert graticule colors 220 invert waveform 59 IP address 239 257 Italian front panel overlays 36 Italian user interface and Quick Help 41 J Japanese front panel overlays 36 Japanese user interface and Qui
134. form user calibration 248 To perform hardware self test 250 To perform front panel self test 251 To display oscilloscope information 251 To display the user calibration status 251 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 16 20 21 Web Interface Reference To clean the oscilloscope 251 To check warranty and extended services status 252 To contact Agilent 252 To return the instrument 252 Configuring the Quick Action Key 253 Adding an Annotation 254 Accessing the Web Interface 258 Browser Web Control 259 Real Scope Remote Front Panel 260 Simple Remote Front Panel 261 Browser Based Remote Front Panel 262 Remote Programming via the Web Interface 263 Remote Programming with Agilent I0 Libraries 264 Save Recall 265 Saving Files via the Web Interface 265 Recalling Files via the Web Interface 266 Get Image 267 Identification Function 268 Instrument Utilities 268 Setting a Password 269 Specifications and Characteristics 273 Measurement Category 273 Oscilloscope Measurement Category 274 Measurement Category Definitions 274 Transient Withstand Capability 275 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide A Maximum input voltage at analog inputs 275 A Maximum input voltage at digital channels 275 Environmental Conditions 275 Probes and Accessories 276 Passive Probes 276 Differential Probes 277 Current Probes 277 Accessories Available 278 Loading Licenses and Disp
135. g starting on page 189 Waveform Generator Sync Pulse All of the waveform generator output functions except DC and Noise have an associated Sync signal The Sync signal is a TTL positive pulse that occurs when the waveform rises above zero volts or the DC offset value See Chapter 16 Waveform Generator starting on page 205 The TRIG OUT connector also provides the User Cal signal See To perform user calibration on page 248 Performing Service Tasks The Service Menu under Utility gt Service lets you perform service related tasks Service Menu Hardware Front Panel About User Cal rt User Cal Self Test Self Test Oscilloscope Status e To perform user calibration on page 248 e To perform hardware self test on page 250 e To perform front panel self test on page 251 e To display oscilloscope information on page 251 e To display the user calibration status on page 251 For other information related to oscilloscope maintenance and service see Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 247 19 Utility Settings e To clean the oscilloscope on page 251 e To check warranty and extended services status on page 252 e To contact Agilent on page 252 e To return the instrument on page 252 To perform user calibration 248 Perform user calibration e Every two years or after 4000 hours of operation e If the ambient temperature is gt 10 C from the calibra
136. g channel or math waveforms can be saved to one of two reference waveform locations in the oscilloscope Then a reference waveform can be displayed and compared against other waveforms One reference waveform can be displayed at a time When the multiplexed knobs are assigned to reference waveforms this happens when you press the Ref key and the LED to the left of it is lit you can use the knobs to scale and position reference waveforms There is also a skew adjustment for reference waveforms Reference waveform scale offset and skew information can optionally be included on the oscilloscope display Analog channel math or reference waveforms can be saved to a reference waveform file on a USB storage device You can recall a reference waveform file from a USB storage device into one of the reference waveform locations To save a waveform to a reference waveform location 1 Press the Ref key to turn on reference waveforms ie Agilent Technologies 11 5 Reference Waveforms To cleara reference waveform location 2 In the Reference Waveform Menu press the Ref softkey and turn the Entry knob to select the desired reference waveform location 3 Press the Source softkey and turn the Entry knob to select the source waveform 4 Press the Save to R1 R2 softkey to save the waveform to the reference waveform location Reference waveforms are non volatile they remain after power cycling or performing a default setup
137. g measured The Y cursor shows the middle threshold point Duty cycle awh x 100 Rise Time The rise time of a signal is the time difference between the crossing of the lower threshold and the crossing of the upper threshold for a positive going edge The X cursor shows the edge being measured For maximum measurement accuracy set the horizontal time div as fast as possible while leaving the complete rising edge of the waveform on the display The Y cursors show the lower and upper threshold points Fall Time The fall time of a signal is the time difference between the crossing of the upper threshold and the crossing of the lower threshold for a negative going edge The X cursor shows the edge being measured For maximum measurement accuracy set the horizontal time div as fast as possible while leaving the complete falling edge of the waveform on the display The Y cursors show the lower and upper threshold points Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 183 13 Measurements Delay 184 Delay measures the time difference from the selected edge on source 1 and the selected edge on source 2 closest to the trigger reference point at the middle threshold points on the waveforms Negative delay values indicate that the selected edge of source 1 occurred after the selected edge of source 2 Source 1 COCO AS NO ee aoe Delay Source 2 ALSS A wN y 1 Press the Meas key to display the Measur
138. gnal If less than three edges are present the measurement shows No edges The X cursors show the interval of the waveform being measured The Full Screen Std Deviation measurement interval variation is an RMS measurement across the full screen with the DC component removed It shows the standard deviation of the displayed voltage values The standard deviation of a measurement is the amount that a measurement varies from the mean value The Mean value of a measurement is the statistical average of the measurement The following figure graphically shows the mean and standard deviation Standard deviation is represented by the Greek letter sigma o For a Gaussian distribution two sigma 1o from the mean is where 68 3 percent of the measurement results reside Six sigma 30 from is where 99 7 percent of the measurement results reside Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 179 13 Measurements 180 mean 30 20 10 0 10 20 30 68 3 95 4 99 7 The mean is calculated as follows N z diet N where e x the mean e N the number of measurements taken e x the ith measurement result The standard deviation is calculated as follows A Z where o the standard deviation e N the number of measurements taken e x the ith measurement result e x the mean Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measu
139. h digital oscilloscopes Often this is simply the ability to freeze the display for the purpose of measuring with cursors or printing the screen Some digital oscilloscopes go one step further by including the ability to further examine the signal details after acquiring them by panning through the waveform and changing the horizontal scale There is no limit imposed on the scaling ratio between the time div used to acquire the data and the time div used to view the data There is however a useful limit This useful limit is somewhat a function of the signal you are analyzing Zooming into stopped acquisitions The screen will still contain a relatively good display if you zoom in horizontally by a factor of 1000 and zoom in vertically by a factor of 10 to display the information from where it was acquired Remember that you can only make automatic measurements on displayed data Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Horizontal Controls 2 To change the horizontal time mode Normal XY or Roll 1 Press Horiz 2 In the Horizontal Menu press Time Mode then select Normal the normal viewing mode for the oscilloscope In the Normal time mode signal events occurring before the trigger are plotted to the left of the trigger point W and signal events after the trigger plotted to the right of the trigger point XY XY mode changes the display from a volts versus time display to a volts versus volts
140. he set threshold are high 1 and values less than the set threshold are low 0 If the Thresholds softkey is set to User press the User softkey for the channel group then turn the Entry knob to set the logic threshold There is one User softkey for each group of channels To reposition a digital channel 1 Make sure the multiplexed scale and position knobs to the right of the key are selected for digital channels If the arrow to the left of the Digital key is not illuminated press the key 2 Use the multiplexed Select knob to select the channel The selected waveform is highlighted in red 3 Use the multiplexed Position knob to move the selected channel waveform If a channel waveform is repositioned over another channel waveform the indicator at the left edge of the trace will change from Dnn designation where nn is a one or two digit channel number to D The WAN Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide indicates that two channels are overlaid 89 6 Digital Channels To display digital channels as a bus Digital channels may be grouped and displayed as a bus with each bus value displayed at the bottom of the display in hex or binary You can create up to two buses To configure and display each bus press the Digital key on the front panel Then press the Bus softkey Digital Channel Menu D ttt t2t2D om m lt Channel Turn off Thresholds Bus v D7 DO w w Next s
141. hecked the oscilloscope mechanically and electrically e Verify that you received the following items and any optional accessories you may have ordered a Agilent Technologies Li 1 20 Getting Started InfiniiVision 2000 X Series oscilloscope Power cord country of origin determines specific type Oscilloscope probes gt Two probes for 2 channel models Four probes for 4 channel models Documentation CD ROM Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 senate iviion SORDA mee saa Z 0m a OKO N2862B probes Qty 2 or 4 Digital Probe Kit MSO models only Power cord Based on country of origin N6459 60001 Digital Probe Kit contains N6459 61601 8 channel cable qyt 1 01650 82103 2 inch probe ground leads qyt 3 5090 4832 Grabber qty 10 US t Digital probe replacement parts are listed in the Digital Channels chapter See Also Accessories Available on page 278 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 1 Getting Started Install the Optional LAN VGA or GPIB Module If you need to install a DSOXLAN LAN VGA module or a DSOXGPIB GPIB module perform this installation before you power on the oscilloscope 1 If you need to remove a module before installing a different module pinch the module s spring tabs and gently remove the module from the slot 2 To install a module slide the module into the slot
142. hown at 40 Intensity 102 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Display Settings 7 To set or clear persistence With persistence the oscilloscope updates the display with new acquisitions but does not immediately erase the results of previous acquisitions All previous acquisitions are displayed with reduced intensity New acquisitions are shown in their normal color with normal intensity Waveform persistence is kept only for the current display area you cannot pan and zoom the persistence display To use persistence 1 Press the Display key Display Menu Persistence ture Intensity 20 2 Press Persistence then turn the Entry knob to select between Off turns off persistence When persistence is off you can press the Capture Waveforms softkey to perform a single shot infinite persistence A single acquisition s data is displayed with reduced intensity and it remains on the display until you clear persistence or clear the display co Persistence infinite presistence Results of previous acquisitions are never erased Use infinite persistence to measure noise and jitter to see the worst case extremes of varying waveforms to look for timing violations or to capture events that occur infrequently Variable Persistence Results of previous acquisitions are erased after a certain amount of time Variable persistence gives you a view of acquired data that is similar to
143. ical data Safety Notices CAUTION A CAUTION notice denotes a haz ard It calls attention to an operat ing procedure practice or the like that if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly per formed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indicated condi tions are fully understood and met Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide InfiniiVision 2000 X Series Oscilloscopes At a Glance SE Alant Technologies InfinilVision MSOX 20248 Agilent igre Table 1 2000 X Series Model Numbers Bandwidths Bandwidth 70 MHz 100 MHz 200 MHz 2 Channel 8 Logic MSO X 2002A MSO X 2012A MSO X 2022A Channels MSO 4 Channel 8 Logic MSO X 2004A MSO X 2014A MSO X 2024A Channels MSO 2 Channel DSO DSO X 2002A DSO X 2012A DSO X 2022A 4 Channel DSO DSO X 2004A DSO X 2014A DSO X 2024A The Agilent InfiniiVision 2000 X Series oscilloscopes deliver these features e 70 MHz 100 MHz and 200 MHz bandwidth models e 2 and 4 channel digital storage oscilloscope DSO models e 2 8 channel and 4 8 channel mixed signal oscilloscope
144. ient dynamic range Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 WARNING Do not negate the protective action of the ground connection to the oscilloscope The oscilloscope must remain grounded through its power cord Defeating the ground creates an electric shock hazard Input a Waveform The first signal to input to the oscilloscope is the Demo 2 Probe Comp signal This signal is used for compensating probes 1 Connect an oscilloscope probe from channel 1 to the Demo 2 Probe Comp terminal on the front panel 2 Connect the probe s ground lead to the ground terminal next to the Demo 2 terminal Recall the Default Oscilloscope Setup To recall the default oscilloscope setup 1 Press Default Setup The default setup restores the oscilloscope s default settings This places the oscilloscope in a known operating condition The major default settings are Table 2 Default Configuration Settings Horizontal Normal mode 100 ps div scale 0 s delay center time reference Vertical Analog Channel 1 on 5 V div scale DC coupling 0 V position Trigger Edge trigger Auto trigger mode 0 V level channel 1 source DC coupling rising edge slope 40 ns holdoff time Display Persistence off 20 grid intensity Other Acquire mode normal Run Stop to Run cursors and measurements off Labels All custom labels that you have created in the Label Library are pres
145. igger mode trigger level measurements cursors and math function settings that tell the oscilloscope how to make a particular measurement See To save setup files on page 219 8 bit Bitmap image bmp The complete screen image in a reduced color 8 bit bitmap format See To save BMP or PNG image files on page 219 24 bit Bitmap image bmp The complete screen image in a 24 bit color bitmap format See To save BMP or PNG image files on page 219 24 bit image png The complete screen image in a 24 bit color PNG format that uses lossless compression Files are much smaller than the BMP format See To save BMP or PNG image files on page 219 CSV data csv This creates a file of comma separated values of all displayed channels and math waveforms This format is suitable for spreadsheet analysis See To save CSV ASCII XY or BIN data files on page 220 ASCII XY data csv This creates separate files of comma separated values for each displayed channel This format is also suitable for spreadsheets See To save CSV ASCII XY or BIN data files on page 220 Reference Waveform data h5 Saves waveform data in a format that can be recalled to one of the oscilloscope s reference waveform locations See To save reference waveform files to a USB storage device on page 225 ALB data alb This creates an Agilent proprietary format file that can be imported by the Agilent Logic Analyzer app
146. illoscope Agilent l isition Channels Acquire Menu Acq Mode Segmented Normal gt Figure 24 Sine With Glitch Normal Mode Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 151 11 Acquisition Control Agilent Acquisition Peak D 62 Channels Segmented q Peak Detect v Figure 25 Sine With Glitch Peak Detect Mode Using Peak Detect Mode to Find a Glitch 1 Connect a signal to the oscilloscope and obtain a stable display 2 To find the glitch press the Acquire key then press the Acq Mode softkey until Peak Detect is selected 3 Press the Display key then press the oo Persistence infinite persistence softkey Infinite persistence updates the display with new acquisitions but does not erase previous acquisitions New sample points are shown at normal intensity while previous acquisitions are displayed at reduced intensity Waveform persistence is not kept beyond the display area boundary Press the Clear Display softkey to erase previously acquired points The display will accumulate points until Persistence is turned off 4 Characterize the glitch with Zoom mode a Press the zoom key or press the Horiz key and then the Zoom softkey b To obtain a better resolution of the glitch expand the time base 152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control 11 Use the horizontal position knob 4 to pan through the waveform to set
147. illoscope s digitizer is sampling faster than is required to fill memory For example suppose an oscilloscope s digitizer has a sample period of 1 ns maximum sample rate of 1 GSa s and a 1 M memory depth At that rate memory is filled in 1 ms If the acquisition time is 100 ms 10 ms div only 1 of every 100 samples is needed to fill memory To select the acquisition mode 1 Press the Acquire key on the front panel 2 In the Acquire Menu press the Acq Mode softkey then turn the Entry knob to select the acquisition mode The InfiniiVision oscilloscopes have the following acquisition modes Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 149 11 Acquisition Control Normal at slower time div settings normal decimation occurs and there is no averaging Use this mode for most waveforms See Normal Acquisition Mode on page 150 Peak Detect at slower time div settings the maximum and minimum samples in the effective sample period are stored Use this mode for displaying narrow pulses that occur infrequently See Peak Detect Acquisition Mode on page 150 Averaging at all time div settings the specified number of triggers are averaged together Use this mode for reducing noise and increasing resolution of periodic signals without bandwidth or rise time degradation See Averaging Acquisition Mode on page 153 High Resolution at slower time div settings all samples in the effective sample p
148. ist of Math key provides access to math add subtract etc waveform functions See Chapter 4 Math Waveforms starting on page 63 Ref key provides access to reference waveform functions Reference waveforms are saved waveforms that can be displayed and compared against other analog channel or math waveforms See Chapter 5 Reference Waveforms starting on page 77 Digital key Press this key to turn the digital channels on or off the arrow to the left will illuminate When the arrow to the left of the Digital key is illuminated the upper multiplexed knob selects and highlights in red individual digital channels and the lower multiplexed knob positions the selected digital channel If a trace is repositioned over an existing trace the indicator at the left edge of the trace will change from Dn designation where n is a one digit channel number from 0 to 7 to D The indicates that two channels are overlaid You can rotate the upper knob to select an overlaid channel then rotate the lower knob to position it just as you would any other channel For more information on digital channels see Chapter 6 Digital Channels starting on page 81 Serial key This key is not currently used with the 2000 X Series oscilloscopes Multiplexed scale knob This scale knob is used with Math Ref or Digital waveforms whichever has the illuminated arrow to the left For math and reference
149. ithout scroll bars use a monitor resolution greater than 800 x 600 on your computer s display Browser Based Remote Front Panel To operate the oscilloscope using the web interface s Browser Based Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Browser Based Remote Front Panel After a few seconds the Remote Front Panel appears 3 Click keys or knobs that you would normally press on the oscilloscope s front panel Buttons have been added for turning knobs Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Web Interface 20 EEEE Run Control SCE Agilent Tochaologies InfiniiVision MSO X 2024A med 007 Remote Programming via the Web Interface To send remote programming commands to the oscilloscope via the SCPI Commands applet window 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Remote Programming The SCPI Commands applet appears within the browser web page Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 263 20 Web Interface 264 Agilent Technologies KOZIE Remote Programming If you press the SCPI Commands button below you can send remote programming comma
150. just vertically and can be used to measure Volts or Amps dependent on the channel Probe Units setting or they can measure ratios When math functions are used as a source the measurement units correspond to that math function The Y1 cursor is the short dashed horizontal line and the Y2 cursor is the long dashed horizontal line The Y cursors adjust vertically and typically indicate values relative to the waveform s ground point except math FFT where the values are relative to O dB In XY horizontal mode the Y cursors display channel 2 values Volts or Amps When active the Y1 and Y2 cursor values for the selected waveform source are displayed in the softkey menu area The difference between Y1 and Y2 AY is displayed in the Cursors box in the right side information area To make cursor measurements 160 1 Connect a signal to the oscilloscope and obtain a stable display 2 Press the Cursors key The Cursors box in the right side information area appears indicating that cursors are on Press the Cursors key again when you want to turn cursors off 3 In the Cursors Menu press Mode then select the desired mode Manual AX 1 AX and AY values are displayed AX is the difference between the X1 and X2 cursors and AY is the difference between the Yl and Y2 cursors Cursors Menu ode Source Cursors Units Y1 E S X2 Y2 1 4 Manual Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide
151. l line above the softkeys and in the Spell softkey Labels can be up to ten characters in length 4 Press the Enter softkey to enter the selected character and to go to the next character position 5 You may position the highlight on any character in the label name by successively pressing the Enter softkey 6 To delete a character from the label press the Enter softkey until the letter you want to delete is highlighted then press the Delete Character softkey You can use a connected USB keyboard instead of using the Spell and other character editing softkeys 7 When you are done entering characters for the label press the Apply New Label softkey to assign the label to the selected channel When you define a new label it is added to the nonvolatile label list When you assign a label ending in a digit such as ADDRO or DATAO the oscilloscope automatically increments the digit and displays the modified label in the New label field after you press the Apply New Label softkey Therefore you only need to select a new channel and press the Apply New Label softkey again to assign the label to the channel Only the original label is saved in the label list This feature makes it easier to assign successive labels to numbered control lines and data bus lines To load a list of labels from a text file you create 110 It may be convenient to create a list of labels using a text editor then load the label list into the oscilloscop
152. l overlay 1 Gently pull on the front panel knobs to remove them 2 Insert the overlay s side tabs into the slots on the front panel Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 35 1 Getting Started 3 Reinstall the front panel knobs Front panel overlays may be ordered from the following part numbers www parts agilent com using Language 2 Channel Overlay 4 Channel Overlay French 75019 94324 75019 94316 German 75019 94326 75019 94318 Italian 75019 94323 75019 94331 Japanese 75019 94311 75019 94312 Korean 75019 94329 75019 94321 Portuguese 75019 94327 75019 94319 Russian 75019 94322 75019 94315 Simplified Chinese 75019 94328 75019 94320 Spanish 75019 94325 75019 94317 Traditional Chinese 75019 94330 75019 94310 36 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 Learn the Rear Panel Connectors For the following figure refer to the numbered descriptions in the table that follows 8 USB Device port l wi 3 LAN VGA HH K use AD 7 USB Host port option module ia our ATEN 6 EXT TRIG IN Ss connector tai a 5 Calibration a protect K button 3 GPIB 4 TRIG OUT E connector MAN ANTA 3 Module slot Tian aaou en 2 Kensingt
153. laying License Information 279 Licensed Options Available 279 Other Options Available 280 Upgrading toan MSO 280 Software and Firmware Updates 280 Binary Data bin Format 281 Binary Data in MATLAB 282 Binary Header Format 282 Example Program for Reading Binary Data 284 Examples of Binary Files 285 CSV and ASCII XY files 287 CSV and ASCII XY file structure 288 Minimum and Maximum Values in CSV Files 288 Acknowledgements 289 Index Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide eo o e e 1 o9 Getting Started Inspect the Package Contents 19 Tilt the Oscilloscope for Easy Viewing 22 Power On the Oscilloscope 23 Connect Probes to the Oscilloscope 24 Input a Waveform 25 Recall the Default Oscilloscope Setup 25 Use Auto Scale 26 Compensate Passive Probes 27 Learn the Front Panel Controls and Connectors 28 Learn the Rear Panel Connectors 37 Learn the Oscilloscope Display 38 Access the Built In Quick Help 40 This chapter describes the steps you take when using the oscilloscope for the first time Inspect the Package Contents e Inspect the shipping container for damage If your shipping container appears to be damaged keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have c
154. lication software using the Agilent B4610A Data Import tool for offline viewing and analysis See To save ALB data files on page 221 Binary data bin This creates a binary file with a header and data in the form of time and voltage pairs This file is much smaller than the ASCII XY data file See To save CSV ASCII XY or BIN data files on page 220 218 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 17 Mask msk This creates a mask file in an Agilent proprietary format that can be read by Agilent InfiniiVision oscilloscopes A mask data file includes some oscilloscope setup information but not all setup information To save all setup information including the mask data file choose Setup scp format instead See To save masks on page 225 You can also configure the Quick Action key to save setups screen images or data See Configuring the Quick Action Key on page 253 To save setup files Setup files can be saved to one of 10 internal Agilent Flash locations or to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select Setup scp 2 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 226 3 Finally press the Press to Save softkey A message indicating whether the save was successful is displa
155. ling rate and the number of FFT points is displayed directly above the softkeys when the FFT Menu is displayed Nyquist Frequency and Aliasing in the Frequency Domain The Nyquist frequency is the highest frequency that any real time digitizing oscilloscope can acquire without aliasing This frequency is half of the sample rate Frequencies above the Nyquist frequency will be under sampled which causes aliasing The Nyquist frequency is also called the folding frequency because aliased frequency components fold back from that frequency when viewing the frequency domain Aliasing happens when there are frequency components in the signal higher than half the sample rate Because the FFT spectrum is limited by this frequency any higher components are displayed at a lower aliased frequency The following figure illustrates aliasing This is the spectrum of a 990 Hz square wave which has many harmonics The sample rate is set to 100 kSa s and the oscilloscope displays the spectrum The displayed waveform shows the components of the input signal above the Nyquist frequency to be mirrored aliased on the display and reflected off the right edge Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Math Waveforms 4 O 0V Agilent Acquisition iT Channels tit FFT Ch3 S Function Operator Source 1 Center More FFT tit FFT 25 0kHz Sg Scale 20dB ffset 33 50dBV FFT Resolution 1Hz Figure 7 Aliasing Because
156. lloscope will perform the action specified by the On Error softkey Note that this is a test sigma the max achievable process sigma assuming no defects for a certain number of tested waveforms as opposed to a process sigma which is tied to the amount of failures per test The sigma value may exceed the selected value when a small sigma value is chosen The actual sigma is displayed Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Mask Testing 14 On Error The On Error setting specifies the action s to take when the input waveform does not conform to the mask This setting supersedes the Run Until setting Stop The oscilloscope will stop when the first error is detected on the first waveform that does not conform to the mask This setting supersedes the Minimum of Tests and Minimum Time settings Save The oscilloscope saves the screen image when an error is detected In the Save Menu press Save Recall gt Save select an image format bmp or png destination on a USB storage device and file name which can be auto incrementing If errors occur too frequently and the oscilloscope spends all its time saving images press the Stop key to stop acquisitions Print The oscilloscope prints the screen image when an error is detected This option is only available when a printer is connected as described in To print the oscilloscope s display on page 231 Measure Measurements
157. lloscopes with a Gaussian type frequency response have an approximate rise time of 0 35 fpw based on a 10 to 90 criterion An oscilloscope s rise time is not the fastest edge speed that the oscilloscope can accurately measure It is the fastest edge speed the oscilloscope can possibly produce Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 147 11 Acquisition Control Oscilloscope Bandwidth Required See Also 148 The oscilloscope bandwidth required to accurately measure a signal is primarily determined by the signal s rise time not the signal s frequency You can use these steps to calculate the oscilloscope bandwidth required 1 Determine the fastest edge speeds You can usually obtain rise time information from published specifications for devices used in your designs 2 Compute the maximum practical frequency component From Dr Howard W Johnson s book High Speed Digital Design A Handbook of Black Magic all fast edges have an infinite spectrum of frequency components However there is an inflection or knee in the frequency spectrum of fast edges where frequency components higher than fknee are insignificant in determining the shape of the signal fknee 0 5 signal rise time based on 10 90 thresholds fknee 0 4 signal rise time based on 20 80 thresholds 3 Use a multiplication factor for the required accuracy to determine the oscilloscope bandwidth required Re
158. lly change the FFT scale or offset settings when you turn the horizontal scale knob the span and center frequency settings will automatically change to allow optimum viewing of the full spectrum If you do manually set scale or offset turning the horizontal scale knob will not change the span or center frequency settings allowing you see better detail around a specific frequency Pressing the FFT Auto Setup softkey will automatically rescale the waveform and span and center will again automatically track the horizontal scale setting Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Math Waveforms 4 3 To make cursor measurements press the Cursors key and set the Source softkey to Math f t Use the X1 and X2 cursors to measure frequency values and difference between two frequency values AX Use the Y1 and Y2 cursors to measure amplitude in dB and difference in amplitude AY 4 To make other measurements press the Meas key and set the Source softkey to Math f t You can make peak to peak maximum minimum and average dB measurements on the FFT waveform You can also find the frequency value at the first occurrence of the waveform maximum by using the X at Max Y measurement The following FFT spectrum was obtained by connecting a 4 V 75 kHz square wave to channel 1 Set the horizontal scale to 50 us div vertical sensitivity to 1 V div Units div to 20 dBV Offset to 60 0 dBV Center frequency to 250 kHz
159. lticast DNS ON OFF ON Agilent MSO X 2024A T a InfiniiVision US50210029 Agilent MSO X 2024A InfiniiVision Multicast DNS Description eee Enter GPIB Address 7 Set to blank for factory default value When accessing the password protected oscilloscope the user name is the IP address of the oscilloscope Toresetthe Do one of these things to reset the password password Using the keys on the front panel of the oscilloscope press Utility gt 1 0 gt LAN Reset e Using the web browser select the Configure Network tab select Modify Configuration erase the Password and select Apply Changes Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 271 20 Web Interface 272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide eo o amp ee 21 a9 Reference Specifications and Characteristics 273 Measurement Category 273 Environmental Conditions 275 Probes and Accessories 276 Loading Licenses and Displaying License Information 279 Software and Firmware Updates 280 Binary Data bin Format 281 CSV and ASCII XY files 287 Acknowledgements 289 Specifications and Characteristics Please see the InfiniiVision oscilloscope data sheets for complete up to date specifications and characteristics To download a data sheet please visit
160. mal display When Zoom is selected the display divides in half The top half of the display shows the normal time div window and the bottom half displays a faster Zoom time div window The Zoom window is a magnified portion of the normal time div window You can use Zoom to locate and horizontally expand part of the normal window for a more detailed higher resolution analysis of signals To turn on or off Zoom 1 Press the zoom key or press the Horiz key and then the Zoom softkey Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Horizontal Controls 2 These markers show the Time div Time div Delay time beginning and end of the for zoomed for normal momentarily displays Zoom window window window when the Horizontal N i position knob is turned 2 21V Agilent Acquisition Normal lt window Channels Signal anomaly expanded in zoom window Zoom A window Horizontal Menu Time Mode Time Ref Normal Center Select Zoom The area of the normal display that is expanded is outlined with a box and the rest of the normal display is ghosted The box shows the portion of the normal sweep that is expanded in the lower half To change the time div for the Zoom window turn the horizontal scale sweep speed knob As you turn the knob the zoomed window time div is highlighted in the status line above the waveform display area The Horizontal scale sweep speed knob controls the size o
161. ment For example you can choose to enter amplitude and offset values or you can choose to enter high level and low level values Or you can choose to enter frequency values or period values Keep pressing the softkey to select the type of adjustment Turn the Entry knob to adjust the value Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 207 16 Waveform Generator Notice that you can select between coarse and fine adjustments for frequency period and width Also pushing the Entry knob is a quick way to toggle between coarse and fine adjustments The Settings softkey opens the Waveform Generator Settings Menu which lets you make other settings related to the waveform generator See To output the waveform generator sync pulse on page 208 To specify the expected output load on page 209 e To use waveform generator logic presets on page 209 To restore waveform generator defaults on page 215 To output the waveform generator sync pulse 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key In the Waveform Generator Menu press the Settings softkey In the Waveform Generator Settings Menu press the Trig Out softkey and turn the Entry knob to select Waveform Generator Sync Pulse Waveform Type Sync Signal Characteristics All waveforms except The Sync signal is a TTL positive pulse that occurs when the waveform DC and
162. mented Averaging 128 gt _ gt Figure 27 128 Averages used to reduce random noise 154 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control 11 See Also Chapter 10 Trigger Mode Coupling starting on page 135 High Resolution Acquisition Mode In High Resolution mode at slower time div settings when decimation would normally occur extra samples are averaged in order to reduce random noise produce a smoother trace on the screen and effectively increase vertical resolution High Resolution mode averages sequential sample points within the same acquisition An extra bit of vertical resolution is produced for every factor of 4 averages The number of extra bits of vertical resolution is dependent on the oscilloscope s time per division setting sweep speed and the displayed sample rate The slower the time div setting the greater the number of samples that are averaged together for each display point High Resolution mode can be used on both single shot and repetitive signals and it does not slow waveform update because the computation is done in the MegaZoom custom ASIC High Resolution mode limits the oscilloscope s real time bandwidth because it effectively acts like a low pass filter Displayed Sample Rate sr per Displayed Sample Rate sr Bits of channel 1 Gsa s max interleaved 2 Gsa s max resolution 250 MSa s lt sr lt 1 Gsa s 500 MSa s lt sr lt 2 G
163. mes there are no frequency components above fmax pw in this case and it requires a system with an ideal brick wall frequency response lt __ Attenuation fN fs Freq u en C Yom Figure 22 Theoretical Brick Wall Frequency Response However digital signals have frequency components above the fundamental frequency square waves are made up of sine waves at the fundamental frequency and an infinite number of odd harmonics and typically for 500 MHz bandwidths and below oscilloscopes have a Gaussian frequency response 146 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control 11 Aliased frequency i components gt lt Attenuation fs 4 fN fs Frequency Limiting oscilloscope bandwidth faw to 1 4 the sample rate fs 4 reduces frequency components above the Nyquist frequency fn Figure 23 Sample Rate and Oscilloscope Bandwidth So in practice an oscilloscope s sample rate should be four or more times its bandwidth fy 4fpw This way there is less aliasing and aliased frequency components have a greater amount of attenuation SeeAlso Evaluating Oscilloscope Sample Rates vs Sampling Fidelity How to Make the Most Accurate Digital Measurements Agilent Application Note 1587 http cp literature agilent com litweb pdf 5989 5732EN pdf Oscilloscope Rise Time Closely related to an oscilloscope s bandwidth specification is its rise time specification Osci
164. mon mode voltage longer ground returns also degrade the pulse fidelity of the probe system Rise time is increased and ringing due to the undamped LC circuit at the input of the probe is also increased Because the digital channels display reconstructed waveforms they do not show ringing and perturbations You will not find ground problems through examination of the waveform display In fact it is likely Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 97 Digital Channels you will discover the problem through random glitches or inconsistent data measurements Use the analog channels to view ringing and perturbations Best Probing Practices Because of the variables L di and dt you may be unsure how much margin is available in your measurement setup The following are guidelines for good probing practices The ground lead from each digital channel group D15 D8 and D7 D0 should be attached to the ground of the device under test if any channel within the group is being used for data capture When capturing data in a noisy environment every third digital channel probe s ground should be used in addition to the channel group s ground High speed timing measurements rise time lt 3 ns should make use of each digital channel probe s own ground When designing a high speed digital system you should consider designing dedicated test ports that interface directly to the instrument s probe system This will ease meas
165. nabling 280 gital channels logic threshold 88 gital channels probing 93 gital channels size 87 gital display interpreting 86 igital probes 81 93 gital probes impedance 93 igital voltmeter DVM 201 isplay key 33 splay multiple acquisitions 144 splay area 39 splay interpreting 38 splay persistence 103 splay signal detail 101 splay softkey labels 40 splay status line 39 isplayed channels AutoScale 245 stortion problems 69 Divide math function 67 DNS IP 239 Duty cycle measurement 183 DVM digital voltmeter 201 DVM license 279 Dynamic DNS 239 E ECL threshold 89 edge speeds 148 edge triggering 115 EDK license 279 either edge trigger 117 English user interface and Quick Help 41 Entry knob 30 Entry knob push to select 30 erase secure 229 expand about 57 243 expand about center 244 expand about ground 243 exporting waveform 217 EXT TRIG IN as Z axis input 50 EXT TRIG IN connector 38 external memory device 35 external trigger 140 external trigger input impedance 141 external trigger probe attenuation 141 external trigger probe units 141 F f t 64 factory default settings 229 Fall time measurement 183 fast debug AutoScale 245 FFT aliasing 73 FFT DC value 73 FFT measurement hints 72 FFT measurements 68 FFT resolution 72 FFT spectral leakage 75 FFT units 73 FFT vertical units 70 FFT window 70 file e
166. nal Trigger Input on page 140 for an explanation of this feature This port is for connecting the oscilloscope to a host PC You can issue remote commands from a host PC to the oscilloscope via the USB device port See Remote Programming with Agilent 10 Libraries on page 264 7 USB Host port This port functions identically to the USB host port on the front panel USB Host Port is used for saving data from the oscilloscope and loading software updates See also USB Host port see page 35 Learn the Oscilloscope Display 38 The oscilloscope display contains acquired waveforms setup information measurement results and the softkey definitions Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Analog channel sensitivity Trigger point Status line Trigger level Analog channels and ground _ levels i Digital channels SS Menu line Softkeys Figure 1 time reference Getting Started 1 Time div l Delay time Run Stop Trigger status type Trigger _ source __ Trigger level or Agilent digital threshold Acquisition Normal Information area Channels Cursors defining measurement 2 Measurements Freq 1 Pk P 1 aconv BANN DIZ Measurements UU OL UIN WL UL T w Interpreting the oscilloscope display Status line Display area Information area The top line of the display contains
167. nds directly to your For more information about the commands see the Programming Commands Quick Reference Command IDN x r Successfully connected to Agilent MS0X20244 InfiniiVision US50210029 gt IDN lt AGILENT TECHNOLOGIES MS0 X 20244 U550210029 01 20 2011050300 Remote Programming with Agilent 10 Libraries While the SCPI Commands applet window lets you enter and remote programming commands remote programming for automated test and data acquisition is typically done using the Agilent IO Libraries which are separate from the instrument s web interface The Agilent IO Libraries let a controller PC communicate with Agilent InfiniiVision oscilloscopes via their USB LAN when the LAN VGA option module is installed or GPIB when the GPIB option module is installed interfaces The Agilent IO Libraries Suite connectivity software to enables communication over these interfaces You can download the Agilent IO Libraries Suite from www agilent com find iolib Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Web Interface 20 Information about controlling the oscilloscope through remote commands is contained in the Programmer s Guide which is included on the documentation CD supplied with this oscilloscope You can also access this document on the Agilent web site For more information about connecting to the oscilloscope refer to the Agilent Technologies USB LAN GPIB Int
168. ne wave 500 mVpp 0 V offset High Z output load are restored Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 215 16 Waveform Generator 216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide on e 17 7 e Save Recall Setups Screens Data Saving Setups Screen Images or Data 217 Recalling Setups Masks or Reference Waveforms 227 Recalling Default Setups 229 Performing a Secure Erase 229 Oscilloscope setups reference waveforms and mask files can be saved to internal oscilloscope memory or to a USB storage device and recalled later You can also recall default or factory default setups Oscilloscope screen images can be saved to a USB storage device in BMP or PNG formats Acquired waveform data can be saved to a USB storage device in comma separated value CSV ASCII XY binary BIN and Agilent logic analyzer binary ALB formats There is also a command to securely erase all the oscilloscope s non volatile internal memory Saving Setups Screen Images or Data 1 Press the Save Recall key 2 In the Save Recall Menu press Save 3 In the Save Trace and Setup Menu press Format then turn the Entry knob to select the type of file you want to save a Agilent Technologies a 17 Save Recall Setups Screens Data Setup scp The oscilloscope s horizontal timebase vertical sensitivity tr
169. nfig softkey then turn the Entry knob to select Automatic and press the softkey again to enable it If your network supports DHCP or AutoIP enabling Automatic lets the oscilloscope use those services to get its LAN configuration settings Connect your PC to the oscilloscope using a cross over LAN cable such as Agilent part number 5061 0701 available on the web at www parts agilent com Cycle power on the oscilloscope Wait until the LAN connection is configured Press Utility gt 1 0 and wait until the LAN status shows configured Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 This may take a few minutes Now the instrument is connected and the instrument s web interface or remote control via LAN may be used File Explorer The File Explorer lets you navigate the oscilloscope s internal file system and the file systems of connected USB storage devices From the internal file system you can load oscilloscope setup files or mask files From a connected USB storage device you can load setup files mask files license files firmware update cab files label files etc Also you can delete files on a connected USB storage device The USB port on the front panel and the USB port on the rear panel labeled HOST are USB Series A receptacles These are the receptacles to which you can connect USB mass storage devices and printers The square receptacle on the rear panel labele
170. ng a database that is 200 x 640 for the waveform viewing area Each location in the array is designated as either a violation or a pass area Each time a data point from a waveform occurs in a violation area a failure is logged If Test All was selected every active analog channel is tested against the mask database for each acquisition Over 2 billion failures can be logged per channel The number of acquisitions tested is also logged and displayed as of Tests The mask file allows greater resolution than the 200 X 640 database Some quantization of data occurs to reduce the mask file data for display on screen 200 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ee 15 Digital Voltmeter The Digital Voltmeter DVM analysis feature provides 3 digit voltage and 5 digit frequency measurements using any analog channel DVM measurements are asynchronous from the oscilloscope s acquisition system and are always acquiring To enable the digital voltmeter analysis feature order Option DVM at time of oscilloscope purchase or order DSOXDVM as a stand alone item after oscilloscope purchase The DVM display is a seven segment readout like you would see on a digital voltmeter It shows the selected mode as well as the units Units are selected using the Units softkey in the channel s Probe Menu After pressing the Analyze key the DVM display also appe
171. nnels is proportional to the current Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 67 4 Math Waveforms 500 05 2 Agilent Acquisition I Channels fft Chi Cho j Function Operator Source 1 Source 2 tit n Figure 6 Example of Multiply Channel 1 by Channel 2 See Also Units for Math Waveforms on page 65 Math Transforms Math transforms perform a transform function FFT on an analog input channel or on the result of an arithmetic operation e FFT Measurement on page 68 FFT Measurement FFT is used to compute the fast Fourier transform using analog input channels or an arithmetic operation g t FFT takes the digitized time record of the specified source and transforms it to the frequency domain When the FFT function is selected the FFT spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency The readout for the horizontal axis changes from time to frequency Hertz and the vertical readout changes from volts to dB 68 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Math Waveforms 4 Use the FFT function to find crosstalk problems to find distortion problems in analog waveforms caused by amplifier non linearity or for adjusting analog filters To display a FFT waveform 1 Press the Math key press the Function softkey and select f t press the Operator softkey and select FFT f t FFT Ch1 cale 20dB fset 60 00 FFT Resol
172. noise rejection Noise Rej adds additional hysteresis to the trigger circuitry By increasing the trigger hysteresis band you reduce the possibility of triggering on noise However this also decreases the trigger sensitivity so that a slightly larger signal is required to trigger the oscilloscope 1 Press the Mode Coupling key 2 In the Trigger Mode and Coupling Menu press the Noise Rej softkey to enable or disable To enable or disable trigger HF Reject HF Reject adds a 50 kHz low pass filter in the trigger path to remove high frequency components from the trigger waveform 3 dB DC 50 kHz You can use HF Reject to remove high frequency noise such as AM or FM broadcast stations or noise from fast system clocks from the trigger path 1 Press the Mode Coupling key 2 In the Trigger Mode and Coupling Menu press the HF Reject softkey to enable or disable Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 139 10 Trigger Mode Coupling To set the trigger holdoff Trigger Holdoff Operating Hints Trigger holdoff sets the amount of time the oscilloscope waits after a trigger before re arming the trigger circuitry Use the holdoff to trigger on repetitive waveforms that have multiple edges or other events between waveform repetitions You can also use holdoff to trigger on the first edge of a burst when you know the minimum time between bursts For example to get a stable trigger on the re
173. ns Run Control 143 Overview of Sampling 145 Selecting the Acquisition Mode 149 Acquiring to Segmented Memory 155 This chapter shows how to use the oscilloscope s acquisition and run controls Running Stopping and Making Single Acquisitions Run Control There are two front panel keys for starting and stopping the oscilloscope s acquisition system Run Stop and Single e When the Run Stop key is green the oscilloscope is running that is acquiring data when trigger conditions are met To stop acquiring data press Run Stop When stopped the last acquired waveform is displayed When the Run Stop key is red data acquisition is stopped Stop is displayed next to the trigger type in the status line at the top of the display To start acquiring data press Run Stop To capture and display a single acquisition whether the oscilloscope is running or stopped press Single The Single run control lets you view single shot events without subsequent waveform data overwriting the display Use Single when you want maximum memory depth for pan and zoom 7 ot a Agilent Technologies m3 11 Acquisition Control When you press Single the display is cleared the trigger mode is temporarily set to Normal to keep the oscilloscope from auto triggering immediately the trigger circuitry is armed the Single key is illuminated and the oscilloscope waits until a trigger condition occurs before it displays a
174. nstalled you can access the oscilloscope s built in web server using a Java enabled web browser The oscilloscope s web interface lets you View information about the oscilloscope like its model number serial number host name IP address and VISA address connect string Control the oscilloscope using the Remote Front Panel Send SCPI Standard Commands for Programmable Instrumentation remote programming commands via the SCPI Commands applet window Save setups screen images waveform data and mask files Recall setup files reference waveform data files or mask files Get screen images and save or print them from the browser Activate the Identification function to identify a particular instrument by causing a message to be displayed or a front panel light to blink View installed options view firmware versions and install firmware upgrade files and view calibration status via the Instrument Utilities page View and modify the oscilloscope s network configuration 7 eg is Agilent Technologies ant 20 Web Interface The web interface for InfiniiVision X Series oscilloscopes also provides help for each of its pages Microsoft Internet Explorer is the recommended Web browser for communication and control of the oscilloscope Other Web browsers may work but are not guaranteed to work with the oscilloscope The Web browser must be Java enabled with Sun Microsystems Java Plug in Before you can use the web interface
175. o a PC 240 connectors rear panel 37 control remote 237 controls front panel 28 copyright 2 coupling channel 57 coupling trigger 138 291 Index crosstalk problems 69 CSV file format 218 CSV files minimum and maximum values 288 urrent probes 277 ursor measurements 159 ursor units 162 ursors key 33 ursors knob 33 ursors binary 161 ursors hex 161 ursors manual 160 ursors track waveform 161 eagaanananeaa D D 32 89 damage shipping 19 data sheet 273 DC channel coupling 58 DC RMS Full Screen measurement 178 DC RMS N Cycles measurement 179 DC signals checking 137 DC waveform generator output 207 dead time re arm 158 decibels FFT vertical units 70 decimating samples 149 decimation for measurement record 288 decimation for screen 288 default configuration 25 default label library 111 default setup 25 229 Default Setup key 31 defaults waveform generator 215 delay knob 45 Delay measurement 172 184 delay time indicator 53 delayed sweep 50 delete character 226 delete file 241 Demo 1 terminal 34 Demo 2 terminal 34 depth AM modulation 212 deviation FM modulation 213 DHCP 239 240 differential probes 277 Digit softkey 120 digital bus mode 90 digital channel controls 32 292 gital channel inputs 35 gital channel menu 88 gital channels 88 gital channels AutoScale 85 gital channels e
176. o select Delay Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 185 13 Measurements 4 Press the Settings softkey to select the second analog channel source for the phase measurement The default Phase settings measure from channel 1 to channel 2 5 Press the Back Up key to return to the Measurement Menu 6 Press the Add Measurement softkey to make the measurement The example below shows a phase measurement between the channel 1 and the math d dt function on channel 1 1 00 1 00 Agilent Acquisition Normal Channels x Measurements Freq 1 Pk Pk 1 Delay Phase 1 2 35 99 Measurement Menu Source Type Add Settings Clear Meas Phase Measurement o a gt Measurement Thresholds Setting measurement thresholds defines the vertical levels where measurements will be taken on an analog channel or math waveform 186 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 Changing default thresholds may change measurement results The default lower middle and upper threshold values are 10 50 and 90 of the value between Top and Base Changing these threshold definitions from the default values may change the returned measurement results for Average Delay Duty Cycle Fall Time Frequency Overshoot Period Phase Preshoot Rise Time Width and Width 1 From the Measurement Menu press the Settings softkey then press the Thresholds softkey to set an
177. o use the InfiniiVision 2000 X Series oscilloscopes When unpacking and using the Chapter 1 Getting Started starting on page 19 oscilloscope for the first time see When displaying waveforms and e Chapter 2 Horizontal Controls starting on page 43 acquired data see e Chapter 3 Vertical Controls starting on page 55 e Chapter 4 Math Waveforms starting on page 63 Chapter 5 Reference Waveforms starting on page 77 e Chapter 6 Digital Channels starting on page 81 e Chapter 7 Display Settings starting on page 101 e Chapter 8 Labels starting on page 107 When setting up triggers or changing Chapter 9 Triggers starting on page 113 how data is acquired see e Chapter 10 Trigger Mode Coupling starting on page 135 e Chapter 11 Acquisition Control starting on page 143 Making measurements and analyzing Chapter 12 Cursors starting on page 159 data e Chapter 13 Measurements starting on page 169 Chapter 14 Mask Testing starting on page 189 e Chapter 15 Digital Voltmeter starting on page 201 When using the built in waveform e Chapter 16 Waveform Generator starting on page generator see 205 When saving recalling or printing Chapter 17 Save Recall Setups Screens Data see starting on page 217 e Chapter 18 Print Screens starting on page 231 When using the oscill
178. ob to navigate to the file to recall See To navigate storage locations on page 226 Press the Press to Recall softkey A message indicating whether the recall was successful is displayed If you would like to clear the display or clear the recalled mask press Clear Display or Clear Mask To recall reference waveform files from a USB storage device Press the Save Recall key 2 In the Save Recall Menu press the Recall softkey 3 In the Recall Menu press the Recall softkey and turn the Entry knob to select Reference Waveform data h5 Press the To Ref softkey and turn the Entry knob to select the desired reference waveform location Press the softkey in the second position and use the Entry knob to navigate to the file to recall See To navigate storage locations on page 226 Press the Press to Recall softkey A message indicating whether the recall was successful is displayed Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 17 7 If you would like to clear the display of everything except the reference waveform press Clear Display Recalling Default Setups 1 Press the Save Recall key 2 In the Save Recall Menu press Default Erase 3 In the Default Menu press one of these softkeys Default Setup recalls the oscilloscope s default setup This is the same as pressing the front panel Default Setup key See Recall the Default Oscilloscope Setup
179. obe softkey to display the Channel Probe Menu This menu lets you select additional probe parameters such as attenuation factor and units of measurement for the connected probe robe Menu 1 00 1 Probe Probe Check Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 59 3 Vertical Controls The Probe Check softkey guides you through the process of compensating passive probes such as the N2862A B N2863A B N2889A N2890A 10073C 10074C or 1165A probes See Also To specify the channel units on page 60 e To specify the probe attenuation on page 60 e To specify the probe skew on page 61 To specify the channel units 1 Press the probe s associated channel key 2 In the Channel Menu press Probe 3 In the Channel Probe Menu press Units then select Volts for a voltage probe Amps for a current probe Channel sensitivity trigger level measurement results and math functions will reflect the measurement units you have selected To specify the probe attenuation The probe attenuation factor must be set properly for accurate measurement results To set the probe attenuation factor 1 Press the channel key 2 Press the Probe softkey until you have selected how you want to specify the attenuation factor choosing either Ratio or Decibels 3 Turn the Entry knob Ito set the attenuation factor for the connected probe When measuring voltage values the attenuation factor can be set from
180. od Fine Amplitude High Level and Offset Low Level softkeys to set the sine signal parameters The frequency can be adjusted from 100 mHz to 20 MHz Square Use the Frequency Frequency Fine Period Period Fine Amplitude High Level Offset Low Level and Duty Cycle softkeys to set the square wave signal parameters The frequency can be adjusted from 100 mHz to 10 MHz The duty cycle can be adjusted from 20 to 80 Ramp Use the Frequency Frequency Fine Period Period Fine Amplitude High Level Offset Low Level and Symmetry softkeys to set the ramp signal parameters The frequency can be adjusted from 100 mHz to 100 kHz Symmetry represents the amount of time per cycle that the ramp waveform is rising and can be adjusted from 0 to 100 Pulse Use the Frequency Frequency Fine Period Period Fine Amplitude High Level Offset Low Level and Width Width Fine softkeys to set the pulse signal parameters The frequency can be adjusted from 100 mHz to 10 MHz The pulse width can be adjusted from 20 ns to the period minus 20 ns DC Use the Offset softkey to set the DC level Noise Use the Amplitude High Level and Offset Low Level to set the noise signal parameters For all waveform types the output amplitude into 50 Q can be adjusted from 10 mVpp to 2 5 Vpp or from 20 mVpp to 5 Vpp into an open circuit load Pressing a signal parameter softkey can open a menu for selecting the type of adjust
181. of the levels of the waveform samples divided by the number of samples Xi Average j Where x value at ith point being measured n number of points in measurement interval The Full Screen measurement interval variation measures the value on all displayed data points The N Cycles measurement interval variation measures the value on an integral number of periods of the displayed signal If less than three edges are present the measurement shows No edges The X cursors show what interval of the waveform is being measured DC RMS is the root mean square value of the waveform over one or more full periods n x2 i l RMS dc V 5 Where x value at ith point being measured n number of points in measurement interval The Full Screen measurement interval variation measures the value on all displayed data points Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 The N Cycles measurement interval variation measures the value on an integral number of periods of the displayed signal If less than three edges are present the measurement shows No edges The X cursors show the interval of the waveform being measured AC RMS AC RMS is the root mean square value of the waveform with the DC component removed It is useful for example for measuring power supply noise The N Cycles measurement interval measures the value on an integral number of periods of the displayed si
182. olution of functions and measurements a Agilent Technologies ich 13 Measurements To make automatic measurements 170 1 Press the Meas key to display the Measurement Menu Settings Clear Meas w 2 Press the Source softkey to select the channel running math function or reference waveform to be measured Only channels math functions or reference waveforms that are displayed are available for measurements If a portion of the waveform required for a measurement is not displayed or does not display enough resolution to make the measurement the result will display No Edges Clipped Low Signal lt value or gt value or a similar message to indicate that the measurement may not be reliable Press the Type softkey then rotate the Entry knob to select a measurement to be made Agilent Acquisition nal Channels Measurements UL Period gt LL Frequ rd w JTL Width Measurement Menu Source Type Settings Clear Meas Freq ement gt w Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 For more information on the types of measurements see Measurements Summary on page 171 4 The Settings softkey will be available to make additional measurement settings on some measurements 5 Press the Add Measurement softkey or push the Entry knob to display the measurement 6 To turn off measurements press the Meas key again Measurements are erased from
183. om time to time Agilent Technologies releases software and firmware updates for its products To search for firmware updates for your oscilloscope direct your web browser to www agilent com find 2000X Series sw To view the currently installed software and firmware press Help gt About Oscilloscope Once you have downloaded a firmware update file you can place it on a USB storage device and load the file using File Explorer see File Explorer on page 241 or you can use the Instrument Utilities page of the oscilloscope s web interface see Instrument Utilities on page 268 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Reference 21 Binary Data bin Format The binary data format stores waveform data in binary format and provides data headers that describe that data Because the data is in binary format the size of the file is approximately 5 times smaller than the ASCII XY format If more than one source is on all displayed sources will be saved except math functions When using segmented memory each segment is treated as a separate waveform All segments for a channel are saved then all segments of the next higher numbered channel are saved This continues until all displayed channels are saved When the oscilloscope is in the Peak Detect acquisition mode the minimum and maximum value waveform data points are saved to the file in separate waveform buffers The minimum value data points are
184. on To adjust reference waveform skew Once reference waveforms are displayed you can adjust their skew Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 79 5 Reference Waveforms 1 Display the desired reference waveform see To display a reference waveform on page 78 2 Press the Skew softkey and turn the Entry knob to adjust the reference waveform skew To display reference waveform information 1 Press the Ref key to turn on reference waveforms 2 In the Reference Waveform Menu press the Options softkey 3 In the Reference Waveform Options Menu press the Display Info softkey to enable or disable reference waveform information on the oscilloscope display 4 Press the Transparent softkey to enable or disable transparent information backgrounds This setting is also used for other oscilloscope information on the display like mask test statistics etc To save recall reference waveform files to from a USB storage device Analog channel math or reference waveforms can be saved to a reference waveform file on a USB storage device See To save reference waveform files to a USB storage device on page 225 You can recall a reference waveform file from a USB storage device into one of the reference waveform locations See To recall reference waveform files from a USB storage device on page 228 80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscill
185. on lock hole oH X qe 1 Power cord connector 1 Power cord Attach the power cord here connector 2 Kensington This is where you can attach a Kensington lock for securing the g y g g lock hole instrument 3 Module slot No modules are included with the oscilloscope A DSOXLAN LAN VGA module may be ordered and installed separately LAN port lets you communicate with the oscilloscope and use the Remote Front Panel feature using the LAN port See Chapter 20 Web Interface starting on page 257 and Accessing the Web Interface on page 258 VGA video output lets you connect an external monitor or projector to provide a larger display or to provide a display at a viewing position away from the oscilloscope The oscilloscope s built in display remains on even when an external display is connected The video output connector is always active For optimal video quality and performance we recommend you use a shielded video cable with ferrite cores Also a DSOXGPIB GPIB module may be ordered and installed separately Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 37 1 Getting Started 4 TRIG OUT connector Trigger output BNC connector See Setting the Rear Panel TRIG OUT Source on page 246 5 Calibration protect button See To perform user calibration on page 248 6 EXT TRIG IN connector 8 USB Device port External trigger input BNC connector See Exter
186. on probes and accessories see www agilent com for e Probes and Accessories Selection Guide 5989 6162EN e 5000 6000 and 7000 Series InfiniiVision Oscilloscope Probes and Accessories Data Sheet 5968 8153EN Passive Probes Passive probes are included with the 2000 X Series oscilloscopes for each analog channel The following passive probes can be used with the InfiniiVision 2000 X Series oscilloscopes Any combination of passive probes can be used Table6 Passive Probes Model Description 1165A Passive probe 10 1 600 MHz 1 5 m 10070C D Passive probe 1 1 20 MHz 1 5 m 10073C Passive probe 10 1 500 MHz 1 5 m 10074C Passive probe 10 1 150 MHz 1 5 m 10076A B Passive probe 100 1 4 kV 250 MHz 276 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Table 6 Reference 21 Passive Probes continued Model Description N2771A B Passive probe 1000 1 30 kV 50 MHz N2862A B Passive probe 10 1 150 MHz 1 2 m N2863A B Passive probe 10 1 300 MHz 1 2 m N2889A Passive probe 10 1 1 1 350 MHz 1 2 m N2890A Passive probe 10 1 500 MHz 1 2 m Differential Probes The following differential probes can be used with the InfiniiVision 2000 X Series oscilloscopes Table 7 Differential Probes Model Description 1141A Active differential probe 200 MHz 200 VDC peak AC max requires 1142A power supply 1144
187. on t care is ignored and is not used as part of the pattern However if all channels in the pattern are set to don t care the oscilloscope will not trigger The rising edge 4 or falling edge t softkey sets the pattern to an edge on the selected channel Only one rising or falling edge can be specified in the pattern When an edge is specified the oscilloscope will trigger at the edge specified if the pattern set for the other channels is true If no edge is specified the oscilloscope will trigger on the last edge that makes the pattern true Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 119 9 120 Triggers Specifying an Edge in a Pattern You are allowed to specify only one rising or falling edge term in the pattern If you define an edge term then select a different channel in the pattern and define another edge term the previous edge definition is changed to a don t care Hex Bus Pattern Trigger You can specify a bus value on which to trigger To do this first define the bus See To display digital channels as a bus on page 90 for details You can trigger on a bus value whether you are displaying the bus or not To trigger on a bus value 1 Press the Pattern key on the front panel 2 Press the Channel softkey and rotate the Entry knob to select Bus1 or Bus2 3 Press the Digit softkey and rotate the Entry knob to select a digit of the selected bus 4 Press the Hex softkey and rotate the Entr
188. onnect a BNC tee to the equal length cables Then connect a BNC f to BNC f also called a barrel connector to the tee as shown below Long cable to TRIG OUT AF A To Channel 1 To Channel 2 Figure 36 User Calibration cable for 2 channel oscilloscope For a 4 channel oscilloscope connect BNC tees to the equal length cables as shown below Then connect a BNC f to BNC f barrel connector to the tee as shown below Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 249 19 Utility Settings To Channel 1 To Channel 2 Be a han To Channel 3 To Channel 4 Longer cable to TRIG OUT Figure 37 User Calibration cable for 4 channel oscilloscope 4 Connect a BNC cable 40 inches maximum from the TRIG OUT connector on the rear panel to the BNC barrel connector 5 Press the Utility key then press the Service softkey 6 Begin the Self Cal by pressing the Start User Cal softkey To perform hardware self test 250 Pressing Utility gt Service gt Hardware Self Test performs a series of internal procedures to verify that the oscilloscope is operating properly It is recommended you run Hardware Self Test e After experiencing abnormal operation e For additional information to better describe an oscilloscope failure e To verify proper operation after the oscilloscope has been repaired Successfully passing Hardware Self Test does not guarantee 100 of the oscilloscope s functionali
189. orm generator gives you an easy way to provide input signals when testing circuitry with the oscilloscope Waveform generator settings can be saved and recalled with oscilloscope setups See Chapter 17 Save Recall Setups Screens Data starting on page 217 To select generated waveform types and settings 1 To access the Waveform Generator Menu and enable or disable the waveform generator output on the front panel Gen Out BNC press the Wave Gen key When waveform generator output is enabled the Wave Gen key is illuminated When waveform generator output is disabled the Wave Gen key is off The waveform generator output is always disabled when the instrument is first turned on 7 eg a Agilent Technologies 205 16 Waveform Generator The waveform generator output is automatically disabled if excessive voltage is applied to the Gen Out BNC 2 In the Waveform Generator Menu press the Waveform softkey and turn the Entry knob to select the waveform type lenu Waveform Frequency AV Sine 1 000kH2 500 05 Agilent Acquisition rmal 11 Channels EE AV Sine Amplitude Settings 500mV pp 3 Depending on the selected waveform type use the remaining softkeys and the Entry knob to set the waveform s characteristics 206 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Waveform Generator 16 Waveform Type Characteristics Sine Use the Frequency Frequency Fine Period Peri
190. ormal trigger mode on page 136 Quick Clear Display clears the display see To clear the display on page 104 Once the Quick Action key is configured you simply press it to perform the selected action Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 253 19 Utility Settings Adding an Annotation You can add an annotation to the upper left corner of the oscilloscope s display The annotation is useful for documentation purposes to add notes before capturing screens To add an annotation 1 On the oscilloscope s front panel press Utility 2 In the Utility Menu press Annotation 3 In the Annotation Menu press Annotation to enable the annotation 4 Press Edit 5 In the Edit Annotation Menu 1 500 500 027 Trig d 1 70 Agilent Acquisition Channels Edit Annotation Menu Spell Delete Character Use the Spell and Delete Character softkeys to enter the annotation text Spell press this softkey and turn the Entry knob to select the character at the current position press this softkey to enter characters and move the cursor to the next character position Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 gt _ press this softkey to enter characters and move the cursor to the previous character position Delete Character press the or softkeys until the desired character is highlighted then press this softkey to d
191. oscope s display 1 Connect a printer You can Connect a USB printer to either the USB port on the front panel or the rectangular USB host port on the rear panel For the most up to date listing of printers that are compatible with the InfiniiVision oscilloscopes please visit www agilent com find InfiniiVision printers Set up a network printer connection See To set up network printer connections on page 232 2 Press the Print key on the front panel 3 In the Print Configuration Menu press the Printto softkey then turn the Entry knob to select the desired printer ie Agilent Technologies ai 18 Print Screens 4 Press the Options softkey to select the print options Print Configuration Metra Print to Options Palette Network Setup Press to NETPRTO Invert Grat Color ae Print See To specify the print options on page 234 5 Press the Palette softkey to select the print palette See To specify the palette option on page 235 6 Press the Press to Print softkey You can stop printing by pressing the Cancel Print softkey The oscilloscope will print the last menu visited before you pressed the Print key Therefore if you have measurements Amplitude Frequency etc showing on the display before you press Print the measurements will be shown on the printout To print the display showing the Print Configuration Menu at the bottom press the Print key twice then press the Press to Print softkey
192. oscope s utility Chapter 19 Utility Settings starting on page 237 functions or web interface see e Chapter 20 Web Interface starting on page 257 For reference information see e Chapter 21 Reference starting on page 273 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 5 TIP Abbreviated instructions for pressing a series of keys and softkeys Instructions for pressing a series of keys are written in an abbreviated manner Instructions for pressing Key1 then pressing Softkey2 then pressing Softkey3 are abbreviated as follows Press Key1 gt Softkey2 gt Softkey3 The keys may be a front panel Key or a Softkey Softkeys are the six keys located directly below the oscilloscope display Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Contents 1 InfiniiVision 2000 X Series Oscilloscopes At a Glance In This Guide 5 Getting Started Inspect the Package Contents 19 Install the Optional LAN VGA or GPIB Module 22 Tilt the Oscilloscope for Easy Viewing 22 Power On the Oscilloscope 23 Connect Probes to the Oscilloscope 24 A Maximum input voltage at analog inputs 24 Do not float the oscilloscope chassis 24 Inputa Waveform 25 Recall the Default Oscilloscope Setup 25 Use Auto Scale 26 Compensate Passive Probes 27 Learn the Front Panel Controls and Connectors 28 Front Panel Overlays for Different Languages 35 Learn the Rear Panel Connectors 3
193. oscopes User s Guide ee e e 6 o9 Digital Channels To connect the digital probes to the device under test 81 Acquiring waveforms using the digital channels 85 To display digital channels using AutoScale 85 Interpreting the digital waveform display 86 To switch all digital channels on or off 88 To switch groups of channels on or off 88 To switch a single channel on or off 88 To change the displayed size of the digital channels 87 To reposition a digital channel 89 To change the logic threshold for digital channels 88 To display digital channels as a bus 90 Digital channel signal fidelity Probe impedance and grounding 93 To replace digital probe leads 98 This chapter describes how to use the digital channels of a Mixed Signal Oscilloscope MSO The digital channels are enabled on MSOX2000 X Series models and DSOX2000 X Series models that have the DSOX2MSO upgrade license installed To connect the digital probes to the device under test 1 If necessary turn off the power supply to the device under test s Agilent Technologies 81 6 Digital Channels Turning off power to the device under test would only prevent damage that might occur if you accidentally short two lines together while connecting probes You can leave the oscilloscope powered on because no voltage appears at the probes ov 2 Connect the digital probe cable to the DIGITAL Dn DO connector on the
194. osition the cursors where you want to read the bus values Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 91 6 Digital Channels X1 cursor X2 cursor Bus values J EGQEDED 02 Ce es Os 42 Go es Oss Source Cursors X 5 000000us Bus1 X2 Set cursors Select Bus1 Bus values mode to or Bus2 source at cursors Binary or Hex shown here When you press the Digital key to display the Digital Channel Menu the digital activity indicator is shown where the cursor values were and the bus values at the cursors are displayed in the graticule Bus values are The bus values are also displayed when using the Pattern trigger function displayed when Press the Pattern key on the front panel to display the Pattern Trigger using Pattern Menu and the bus values will be displayed on the right above the trigger softkeys The dollar sign will be displayed in the bus value when the bus value cannot be displayed as a hex value This occurs when one or more don t cares X are combined with low 0 and high 1 logic levels in the pattern specification or when a transition indicator rising edge 4 or falling edge t are included in the pattern specification A byte that consists of all don t cares X will be displayed in the bus as a don t care eey 92 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Trigger Pattern Trigger pattern definition Channel Bus1 gee
195. ot be able to be read or written to When the USB mass storage device is connected to the oscilloscope s front or rear USB host port a small four color circle icon may be displayed briefly as the USB device is read 242 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 You do not need to eject the USB mass storage device before removing it Simply ensure that any file operation you ve initiated is done and remove the USB drive from the oscilloscope s host port Do not connect USB devices that identify themselves as hardware type CD because these devices are not compatible with the InfiniiVision X Series oscilloscopes If two USB mass storage devices are connected to the oscilloscope the first one is designated usb and the second one is designated usb2 See Also Chapter 17 Save Recall Setups Screens Data starting on page 217 Setting Oscilloscope Preferences The User Preferences Menu under Utility gt Options gt Preferences lets you specify oscilloscope preferences e To choose expand about center or ground on page 243 e To disable enable transparent backgrounds on page 244 e To load the default label library on page 244 e To set up the screen saver on page 244 e To set AutoScale preferences on page 245 To choose expand about center or ground When you change a channel s volts division setting the waveform display can be set to expand or compre
196. ow is the lower Zoom window Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide eo o e e 14 a amp Mask Testing J e e i To create a mask from a golden waveform Automask 189 e Mask Test Setup Options 191 Mask Statistics 194 To manually modify a mask file 195 Building a Mask File 198 One way to verify a waveform s compliance to a particular set of parameters is to use mask testing A mask defines a region of the oscilloscope s display in which the waveform must remain in order to comply with chosen parameters Compliance to the mask is verified point by point across the display Mask test operates on displayed analog channels it does not operate on channels that are not displayed To enable mask test order Option LMT at time of oscilloscope purchase or order DSOX2MASK as a stand alone item after oscilloscope purchase To create a mask from a golden waveform Automask A golden waveform meets all chosen parameters and it is the waveform to which all others will be compared 1 Configure the oscilloscope to display the golden waveform 2 Press the Analyze key 3 Press Features then select Mask Test 4 Press Features again to enable mask testing ie Agilent Technologies io 14 Mask Testing 190 Failures Channel 1 O Automask gt Press Automask In the Automask Menu press the Source softkey
197. pacitance to ground While this results in an input impedance specification that is an accurate model for DC and low frequencies the high frequency model of the probe input is more useful see the following figure This high frequency model takes into account pure tip capacitance to ground as well as series tip resistance and the cable s characteristic impedance Z 8 5 pF 100 kQ Figure 9 DC and Low Frequency Probe Equivalent Circuit Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 250 Q 7 5 pF 1 pF 100 kQ 150 Q Figure 10 High Frequency Probe Equivalent Circuit The impedance plots for the two models are shown in these figures By comparing the two plots you can see that both the series tip resistor and the cable s characteristic impedance extend the input impedance significantly The stray tip capacitance which is generally small 1 pF sets the final break point on the impedance chart Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 95 96 Digital Channels 100 k 10k High Frequency Model oO 1k Typical 100 Model 10 1 10kHz 100kHz 1MHz 10MHz 100 MHz 1 GHz Frequency Figure 11 Impedance versus Frequency for Both Probe Circuit Models The logic probes are represented by the high frequency circuit model shown above They are designed to provide as much series tip resistance as possible Stray tip capacitance to ground is minimized b
198. page 258 2 When the oscilloscope s web interface is displayed select the Configure Network tab from the instrument s Welcome page 3 Click the Modify Configuration button Agilent Technologies KORICE S Current Network Configuration Modify Configuration Configuration mode Dynamic DNS ON ON Configure Multicast DNS ON Network 5 isa tab ieee keene Agilent MSO X 2024A InfiniiVision US50210029 al 130 29 70 169 Subnet Mask 255 255 248 0 Default Gateway 130 29 64 1 130 29 64 128 130 29 64 128 a mx2024a 10029 GPIB Control GPIB Address USB Control 2 a D oO pd gt 21 2 o e FE E 2 8 3 z 2 3 z 5 ajo A E A E ole s amp lzZi lie Z re SIs izisizie 3 8 g FE ER FE E G 5 g E ES E Ea G E A a 4 Enter your desired password and click Apply Changes 270 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Web Interface 20 Agilent Technologies KOSICE IS Modify Network Configuration Undo Changes Factory Defaults Configured Value IP Settings may be configured using the following Automatic ON IP Settings to use in non automatic mode IP Address 169 254 254 254 Subnet Mask 255 255 248 0 Default Gateway 169 254 254 254 Domain name and name service settings DNS Server 0 0 0 0 0 0 0 0 Hostname a mx2024a 10029 a mx2024a 10029 Dynamic DNS ON OFF ON Mu
199. peak AC With 10073C or 10074C 10 1 probe CAT I 500 Vpk CAT II 400 Vpk The external trigger input impedance is 1M Ohm This lets you use passive probes for general purpose measurements The higher impedance minimizes the loading effect of the oscilloscope on the device under test To set the EXT TRIG IN units and probe attenuation 1 Press the Mode Coupling key in the Trigger section of the front panel Trigger Mode and Coupling Menu Mode Coupling Noise Rej HF Reject Holdoff External Auto DE 40 000ns ee 2 In the Trigger Mode and Coupling Menu press the External softkey External Trigger Menu 1 00 1 3 In the External Trigger Menu press the Units softkey to select between Volts for a voltage probe Amps for a current probe Measurement results channel sensitivity and the trigger level will reflect the measurement units you have selected 4 Press the Probe softkey then turn the entry knob to specify the probe attenuation The attenuation factor can be set from 0 1 1 to 1000 1 in a 1 2 5 sequence The probe attenuation factor must be set properly for measurements to be made correctly Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 141 10 Trigger Mode Coupling 142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ee 11 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control Running Stopping and Making Single Acquisitio
200. pecify channel coupling Coupling changes the channel s input coupling to either AC alternating current or DC direct current Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 57 3 Vertical Controls If the channel is DC coupled you can quickly measure the DC component of the signal by simply noting its distance from the ground symbol TIP If the channel is AC coupled the DC component of the signal is removed allowing you to use greater sensitivity to display the AC component of the signal 1 Press the desired channel key 2 In the Channel Menu press the Coupling softkey to select the input channel coupling DC DC coupling is useful for viewing waveforms as low as 0 Hz that do not have large DC offsets AC AC coupling is useful for viewing waveforms with large DC offsets AC coupling places a 10 Hz high pass filter in series with the input waveform that removes any DC offset voltage from the waveform Note that Channel Coupling is independent of Trigger Coupling To change trigger coupling see To select the trigger coupling on page 138 To specify bandwidth limiting 1 Press the desired channel key 2 In the Channel Menu press the BW Limit softkey to enable or disable bandwidth limiting When bandwidth limit is on the maximum bandwidth for the channel is approximately 20 MHz For waveforms with frequencies below this turning bandwidth limit on removes unwanted high frequency noise from th
201. per Point 4 Header N 12 bytes Voltage Data N buffer size CSV and ASCII XY files e CSV and ASCII XY file structure on page 288 e Minimum and Maximum Values in CSV Files on page 288 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 287 21 Reference CSV and ASCII XY file structure In CSV or ASCII XY format the Length control selects the number of points per segment All segments are contained in the CSV file or in each ASCII XY data file For example If the Length control is set to 1000 points there will be 1000 points rows in the spreadsheet per segment When saving all segments there are three header rows so the data for the first segment starts at row 4 The second segment s data starts at row 1004 The time column shows the time since the trigger on the first segment The top row shows the selected number of points per segment BIN files are a more efficient data transfer format than CSV or ASCII XY Use this file format for fastest data transfer Minimum and Maximum Values in CSV Files If you are running a Minimum or Maximum measurement the minimum and maximum values shown in the measurement display may not appear in the CSV file Explanation When the oscilloscope s sample rate is 4 GSa s a sample will be taken every 250 ps If the horizontal scale is set to 10 us div there will be 100 us of data displayed because there are ten divisions across the screen To find the total number of
202. petitive pulse burst shown below set the holdoff time to be gt 200 ns but lt 600 ns Holdoff Oscilloscope triggers here I gt gt gt gt gt ra p gt 200 ns 600 ns To set the trigger holdoff 1 Press the Mode Coupling key 2 In the Trigger Mode and Coupling Menu press the Holdoff softkey then turn the Entry knob to increase or decrease the trigger holdoff time The correct holdoff setting is typically slightly less than one repetition of the waveform Set the holdoff to this time to generate a unique trigger point for a repetitive waveform Changing the time base settings does not affect the trigger holdoff time With Agilent s MegaZoom technology you can press Stop then pan and zoom through the data to find where the waveform repeats Measure this time using cursors then set the holdoff External Trigger Input 140 The external trigger input can be used as a source in several of the trigger types The external trigger BNC input is on the rear panel and is labeled EXT TRIG IN Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Trigger Mode Coupling 10 CAUTION A Maximum voltage at oscilloscope external trigger input CAT 300 Vrms 400 Vpk transient overvoltage 1 6 kVpk 1 M ohm input For steady state sinusoidal waveforms derate at 20 dB decade above 57 kHz to a minimum of 5 Vpk With N2863A 10 1 probe CAT I 600 V CAT II 300 V DC
203. powered by MAINS that comply to Overvoltage Category Category Il which is typical of cord and plug connected equipment Pollution Degree The InfiniiVision 2000 3000 X Series oscilloscopes may be operated in environments of Pollution Degree 2 or Pollution Degree 1 Pollution Degree Pollution Degree 1 No pollution or only dry non conductive pollution occurs Definitions The pollution has no influence Example A clean room or climate controlled office environment Pollution Degree 2 Normally only dry non conductive pollution occurs Occasionally a temporary conductivity caused by condensation may occur Example General indoor environment Pollution Degree 3 Conductive pollution occurs or dry non conductive pollution occurs which becomes conductive due to condensation which is expected Example Sheltered outdoor environment Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 275 21 Reference Probes and Accessories This setion lists the probes and accessories that are compatible with the 2000 X Series oscilloscopes e Passive Probes on page 276 e Differential Probes on page 277 e Current Probes on page 277 e Accessories Available on page 278 Because the 2000 X Series oscilloscopes do not have a ring around the BNC connector for identifying probes you must set the probe attenuation factor manually See To specify the probe attenuation on page 60 See Also For more information
204. ptions From the Mask Test Menu press the Setup softkey to enter the Mask Setup Menu Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 191 14 Mask Testing 192 Run Until The Run Until softkey lets you specify a condition on which to terminate testing Forever The oscilloscope runs continuously However if an error occurs the action specified using the On Error softkey will occur Minimum of Tests Choose this option and then use the of Tests softkey to select the number of times the oscilloscope will trigger display the waveform s and compare them to the mask The oscilloscope will stop after the specified number of tests have been completed The specified minimum number of tests may be exceeded If an error occurs the action specified using the On Error softkey will occur The actual number of tests completed is displayed above the softkeys Minimum Time Choose this option and then use the Test Time softkey to select how long the oscilloscope will run When the selected time has passed the oscilloscope will stop The specified time may be exceeded If an error occurs the action specified using the On Error softkey will occur The actual test time is displayed above the softkeys Minimum Sigma Choose this option and then use the Sigma softkey to select a minimum sigma The mask test runs until enough waveforms are tested to achieve a minimum test sigma If an error occurs the osci
205. quired accuracy Oscilloscope bandwidth required 20 few 1 0 x fknee 10 few 1 3 X fknee 3 few 1 9 x fknee Choosing an Oscilloscope with the Right Bandwidth for your Application Agilent Application Note 1588 http cp literature agilent com litweb pdf 5989 5733EN pdf Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Acquisition Control 11 Memory Depth and Sample Rate The number of points of oscilloscope memory is fixed and there is a maximum sample rate associated with oscilloscope s analog to digital converter however the actual sample rate is determined by the time of the acquisition which is set according to the oscilloscope s horizontal time div scale sample rate number of samples time of acquisition For example when storing 50 us of data in 50 000 points of memory the actual sample rate is 1 GSa s Likewise when storing 50 ms of data in 50 000 points of memory the actual sample rate is 1 MSa s The actual sample rate is displayed in the right side information area The oscilloscope achieves the actual sample rate by throwing away decimating unneeded samples Selecting the Acquisition Mode When selecting the oscilloscope acquisition mode keep in mind that samples are normally decimated at slower time div settings At slower time div settings the effective sample rate drops and the effective sample period increases because the acquisition time increases and the osc
206. r X Units Date A 16 byte character array left blank in InfiniiVision oscilloscopes Time A 16 byte character array left blank in the InfiniiVision oscilloscopes Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 283 21 Reference Frame A 24 byte character array that is the model number and serial number of the oscilloscope in the format of MODEL SERIAL Waveform Label A 16 byte character array that contains the label assigned to the waveform Time Tags Segment Index A 64 bit double only used when saving multiple segments requires segmented memory option This is the time in seconds since the first trigger A 32 bit unsigned integer This is the segment number Only used when saving multiple segments Waveform Data A waveform may have more than one data set Each waveform data set Header will have a waveform data header The waveform data header consists of information about the waveform data set This header is stored immediately before the data set Waveform Data Header Size A 32 bit integer that is the size of the waveform data header Buffer Type A 16 bit short that is the type of waveform data stored in the file 0 Unknown data e 1 Normal 32 bit float data 2 Maximum float data e 3 Minimum float data 4 Not used in InfiniiVision oscilloscopes 5 Notused in InfiniiVision oscilloscopes e 6 Digital unsigned 8 bit char data for digit
207. re the oscilloscope for the Probe Comp signal see Use Auto Scale on page 26 4 Press the channel key to which the probe is connected 1 2 etc 5 In the Channel Menu press Probe 6 In the Channel Probe Menu press Probe Check then follow the instructions on screen Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 27 1 Getting Started If necessary use a nonmetallic tool supplied with the probe to adjust the trimmer capacitor on the probe for the flattest pulse possible On the N2862 63 90 probes the trimmer capacitor is the yellow adjustment on the probe tip On other probes the trimmer capacitor is located on the probe BNC connector Perfectly compensated Over compensated Under compensated S a 7 Connect probes to all other oscilloscope channels channel 2 of a 2 channel oscilloscope or channels 2 3 and 4 of a 4 channel oscilloscope 8 Repeat the procedure for each channel Learn the Front Panel Controls and Connectors On the front panel key refers to any key button you can press Softkey specifically refers to the six keys that are directly below the display The legend for these keys is directly above them on the display Their functions change as you navigate through the oscilloscope s menus For the following figure refer to the numbered descriptions in the table that follows 28 Agilent InfiniiVision 2000 X Series Oscilloscop
208. rements 13 Time Measurements The following figure shows time measurement points Rise Time Fall Time i lt i k Thresholds I Lower Width Width ha Period gt The default lower middle and upper measurement thresholds are 10 50 and 90 between Top and Base values See Measurement Thresholds on page 186 for other percentage threshold and absolute value threshold settings e Period on page 181 e Frequency on page 182 e Width on page 183 e Width on page 183 e Duty Cycle on page 183 e Rise Time on page 183 e Fall Time on page 183 e Delay on page 184 e Phase on page 185 Period Period is the time period of the complete waveform cycle The time is measured between the middle threshold points of two consecutive like polarity edges A middle threshold crossing must also travel through the lower and upper threshold levels which eliminates runt pulses The X cursors show what portion of the waveform is being measured The Y cursor shows the middle threshold point Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 181 13 Measurements 182 Frequency See Also Frequency is defined as 1 Period Period is defined as the time between the middle threshold crossings of two consecutive like polarity edges A middle threshold crossing must also travel through the lower and upper threshold levels which eliminates runt pulses The X cursors
209. rence is the reference point on the display for delay time horizontal position 1 Press Horiz 2 In the Horizontal Menu press Time Ref then select Left the time reference is set to one major division from the left edge of the display Center the time reference is set to the center of the display Right the time reference is set to one major division from the right edge of the display A small hollow triangle V at the top of the display grid marks the position of the time reference When delay time is set to zero the trigger point indicator V overlays the time reference indicator Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Horizontal Controls 2 The time reference position sets the initial position of the trigger event within acquisition memory and on the display with delay set to 0 Turning the Horizontal scale sweep speed knob expands or contracts the waveform about the time reference point V See To adjust the horizontal time div scale on page 44 Turning the Horizontal position 4 knob in Normal mode not Zoom moves the trigger point indicator W to the left or right of the time reference point V See To adjust the horizontal delay position on page 45 Navigating the Time Base You can use the Navigate key and controls to navigate through e Captured data see To navigate time on page 53 e Segments when segmented memory acquisitions are turned on see
210. ress the Digital key on the front panel then press the Thresholds softkey to set the threshold level TTL CMOS ECL or user defined for the selected digital channel group The threshold value is displayed in the upper right corner of the display The line trigger level is not adjustable This trigger is synchronized with the power line supplied to the oscilloscope You can also change the trigger level of all channels by pressing Analyze gt Features and then selecting Trigger Levels Forcing a Trigger The Force Trigger key causes a trigger on anything and displays the acquisition This key is useful in the Normal trigger mode where acquisitions are made only when the trigger condition is met In this mode if no triggers are occurring that is the Trig d indicator is displayed you can press Force Trigger to force a trigger and see what the input signals look like In the Auto trigger mode when the trigger condition is not met triggers are forced and the Auto indicator is displayed Edge Trigger The Edge trigger type identifies a trigger by looking for a specified edge slope and voltage level on a waveform You can define the trigger source and slope in this menu The trigger type source and level if applicable are displayed in the upper right corner of the display 1 On the front panel in the Trigger section press the Trigger key 2 In the Trigger Menu press the Trigger softkey and use the Entry
211. rm Generator starting on page 205 These controls determine how the oscilloscope triggers to capture data See Chapter 9 Triggers starting on page 113 and Chapter 10 Trigger Mode Coupling starting on page 135 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 Horizontal controls The Horizontal controls consist of Horizontal scale knob Turn the knob in the Horizontal section that is marked N Ay to adjust the time div sweep speed setting The symbols under the knob indicate that this control has the effect of spreading out or zooming in on the waveform using the horizontal scale Horizontal position knob Turn the knob marked lt p to pan through the waveform data horizontally You can see the captured waveform before the trigger turn the knob clockwise or after the trigger turn the knob counterclockwise If you pan through the waveform when the oscilloscope is stopped not in Run mode then you are looking at the waveform data from the last acquisition taken Horiz key Press this key to open the Horizontal Menu where you can select XY and Roll modes enable or disable Zoom enable or disable horizontal time division fine adjustment and select the trigger time reference point Zoom key Press the zoom key to split the oscilloscope display into Normal and Zoom sections without opening the Horizontal Menu Search key Lets you
212. s Guide ee 3 Vertical Controls To turn waveforms on or off channel or math 56 To adjust the vertical scale 57 To adjust the vertical position 57 To specify channel coupling 57 To specify bandwidth limiting 58 To change the vertical scale knob s coarse fine adjustment setting 58 To invert a waveform 59 Setting Analog Channel Probe Options 59 The vertical controls include e The vertical scale and position knobs for each analog channel e The channel keys for turning a channel on or off and accessing the channel s softkey menu The following figure shows the Channel 1 Menu that appears after pressing the 1 channel key e Agilent Technologies a 3 Vertical Controls Channel Trigger Trigger level Volts div source __ or threshold x cs 100 05 Agilent Acquisition Channels Channel 1 ground level Channel 2 ground ___ level Channel 1 Menu Coupling BW Limit oc The ground level of the signal for each displayed analog channel is identified by the position of the P icon at the far left side of the display To turn waveforms on or off channel or math 56 1 Press an analog channel key turn the channel on or off and to display the channel s menu When a channel is on its key is illuminated Turning channels off You must be viewing the menu for a channel before you can turn it off For example if channel 1 and channel 2 are turned on and the menu for ch
213. s User s Guide 129 9 Triggers 200 05 o Vid Field1 Agilent Channels Mode Field Holdoff Field1 0 0flds Figure 17 Triggering on Field 1 To trigger on all fields of the video signal To quickly and easily view transitions between fields or to find the amplitude differences between the fields use the All Fields trigger mode 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select All Fields 130 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 200 05 o Vid All Fld Agilent Channels Video Trigger Menu Standard Mode Field Holdoff NTSC All Fields O Oflds Figure 18 Triggering on All Fields To trigger on odd or even fields To check the envelope of your video signals or to measure worst case distortion trigger on the odd or even fields When Field 1 is selected the oscilloscope triggers on color fields 1 or 3 When Field 2 is selected the oscilloscope triggers on color fields 2 or 4 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select Field1 or Field2 The trigger circui
214. s the Label key 2 05V Agilent on Channels Library ACK 2 Press the Channel softkey then turn the Entry knob or successively press the Channel softkey to select a channel for label assignment 108 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Labels 8 2 05 Agilent on Channels Library Character ACK The figure above shows the list of channels and their default labels The channel does not have to be turned on to have a label assigned to it 3 Press the Library softkey then turn the Entry knob or successively press the Library softkey to select a predefined label from the library 4 Press the Apply New Label softkey to assign the label to your selected channel 5 Repeat the above procedure for each predefined label you want to assign to a channel To define a new label 1 Press the Label key 2 Press the Channel softkey then turn the Entry knob or successively press the softkey to select a channel for label assignment The channel does not have to be turned on to have a label assigned to it If the channel is turned on its current label will be highlighted 3 Press the Spell softkey then turn the Entry knob to select the first character in the new label Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 109 8 Labels Label Assignment Auto Increment Turning the Entry knob selects a character to enter into the highlighted position shown in the New labe
215. s the More FFT softkey to display additional FFT settings Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 69 4 70 Math Waveforms f t FFT Ch1 Scale 20dB ffset 60 00 j FFT Resolution 763Hz Window Hanning Auto Setup Window selects a window to apply to your FFT input signal Hanning window for making accurate frequency measurements or for resolving two frequencies that are close together Flat Top window for making accurate amplitude measurements of frequency peaks Rectangular good frequency resolution and amplitude accuracy but use only where there will be no leakage effects Use on self windowing waveforms such as pseudo random noise impulses sine bursts and decaying sinusoids Blackman Harris window reduces time resolution compared to a rectangular window but improves the capacity to detect smaller impulses due to lower secondary lobes Vertical Units lets you select Decibels or V RMS as the units for the FFT vertical scale Auto Setup sets the frequency Span and Center to values that will cause the entire available spectrum to be displayed The maximum available frequency is half the FFT sample rate which is a function of the time per division setting The FFT resolution is the quotient of the sampling rate and the number of FFT points fs N The current FFT Resolution is displayed above the softkeys Scale and offset considerations If you do not manua
216. s the sync polarity softkey to set the Video trigger to either positive C TD or negative CL sync polarity 5 Press the Settings softkey Video Trigger Menu Standard NTSC Mode Field Holdoff Field1 O Oflds Auto Setup 6 In the Video Trigger Menu press the Standard softkey to set the video standard The oscilloscope supports triggering on the following television TV and video standards Standard Type Sync Pulse NTSC Interlaced Bi level PAL Interlaced Bi level PAL M Interlaced Bi level SECAM Interlaced Bi level 7 Press the Auto Setup softkey to automatically set up the oscilloscope for the selected Source and Standard Source channel vertical scaling is set to 140 mV div Source channel offset is set to 245 mV Source channel is turned on Trigger type is set to Video Video trigger mode is set to All Lines Display Grid type is set to IRE when Standard is NTSC or mV see To select the grid type on page 104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 125 9 Triggers Video Triggering Examples 126 Horizontal time division is set to 10 us div for NTSC PAL SECAM standards Horizontal delay is set so that trigger is at first horizontal division from the left You can also press Analyze gt Features and then select Video to quickly access the video triggering automatic set up and display options 8 Press the Mode softkey to select the portion of the video
217. sa s 8 62 5 MSa s lt sr lt 250 MSa s 125 MSa s lt sr lt 500 MSa s 9 12 5 MSa s lt sr lt 62 5 MSa s 25 MSa s lt sr lt 125 MSa s 10 2 5 MSa s lt sr lt 12 5 MSa s 5 MSa s lt sr lt 25 MSa s 11 sr lt 2 5 MSa s sr lt 5 MSa s 12 Acquiring to Segmented Memory You can purchase the oscilloscope with the segmented memory option factory installed Option SGM or you can install a license to enable it order model number DSOX2SGM Segmented Memory Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 155 11 Acquisition Control To acquire to segmented memory 156 When capturing multiple infrequent trigger events it is advantageous to divide the oscilloscope s memory into segments This lets you capture signal activity without capturing long periods of signal inactivity Each segment is complete with all analog channel and digital channel on MSO models data When using segmented memory use the Analyze Segments feature see Infinite Persistence with Segmented Memory on page 157 to show infinite persistence across all acquired segments See also To set or clear persistence on page 103 for details 1 Set up a trigger condition See Chapter 9 Triggers 113 for details Press the Acquire key in the Waveform section of the front panel starting on page N 3 Press the Segmented softkey 4 In the Segmented Memory Menu press the Segmented softkey to enable segm
218. saved first then the maximum value data points are saved BIN data using When saving all segments each segment has its own waveform header see segmented Binary Header Format on page 282 memor 7 In BIN file format data are presented as follows e Channel 1 data all segments e Channel 2 data all segments e Channel 3 data all segments e Channel 4 data all segments e Digital channel data all segments e Math waveform data all segments When not saving all segments the number of waveforms is equivalent to the number of active channels including math and digital channels with up to seven waveforms for each digital pod When saving all segments the number of waveforms is equal to the number of active channels multiplied by the number of segments acquired Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 281 21 Reference Binary Data in MATLAB Binary data from an InfiniiVision oscilloscope can be imported to The MathWorks MATLAB You can download the appropriate MATLAB functions from the Agilent Technologies web site at www agilent com find 2000X Series examples Agilent provides the m files which need to be copied into the work directory for MATLAB The default work directory is C MATLAB7 work Binary Header Format File Header Waveform Header 282 There is only one file header in a binary file The file header consists of the following information Cookie Two
219. scilloscope s Demo terminals and a lab guide tutorial for education environments MASK Mask Limit Test Order DSOX2MASK Lets you create a mask and test waveforms to determine whether they comply to the mask MSO Mixed Signal Oscilloscope MSO Upgrade a DSO to an Order DSOX2MSO MSO The digital probe cable kit is supplied with Adds 8 digital channels You do not have to install any the MSO license hardware SGM Segmented Memory Order DSOX2SGM Lets you capture infrequent or burst signals with great resolution by eliminating the capture of your signal s inactivity Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 279 21 Reference Table 10 Licensed Options Available continued License Description After purchase model number notes WAVEGEN Waveform Generator Order DSOX2WAVEGEN Other Options Available Table 11 Calibration Option Option Order A6J ANSI Z540 Compliant Calibration Upgrading to an MSO A license can be installed to activate the digital channels of an oscilloscope that was not originally ordered as a mixed signal oscilloscope MSO A mixed signal oscilloscope has analog channels plus 8 time correlated digital timing channels For information about upgrading your oscilloscope through licensing contact your local Agilent Technologies representative or see www agilent com find 2000X Series Software and Firmware Updates 280 Fr
220. search for events in the acquired data Navigate keys Press this key to navigate through captured data Time search events or segmented memory acquisitions See Navigating the Time Base on page 53 For more information see Chapter 2 Horizontal Controls starting on page 43 Run Control keys When the Run Stop key is green the oscilloscope is running that is acquiring data when trigger conditions are met To stop acquiring data press Run Stop When the Run Stop key is red data acquisition is stopped To start acquiring data press Run Stop To capture and display a single acquisition whether the oscilloscope is running or stopped press Single The Single key is yellow until the oscilloscope triggers For more information see Running Stopping and Making Single Acquisitions Run Control on page 143 Default Setup key Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Press this key to restore the oscilloscope s default settings details on Recall the Default Oscilloscope Setup on page 25 31 1 Getting Started 10 Auto Scale key 11 Additional waveform controls When you press the AutoScale key the oscilloscope will quickly determine which channels have activity and it will turn these channels on and scale them to display the input signals See Use Auto Scale on page 26 The additional waveform controls cons
221. select the second source for the arithmetic operation The result is displayed 6 To re size and re position the math waveform see To adjust the math waveform scale and offset on page 65 Math Operating Hints If the analog channel or math function is clipped not fully displayed on screen the resulting displayed math function will also be clipped Once the function is displayed the analog channel s may be turned off for better viewing of the math waveform The vertical scaling and offset of each math function can be adjusted for ease of viewing and measurement considerations The math function waveform can be measured using Cursors and or Meas To perform a transform function on an arithmetic operation 64 To perform a transform function FFT on an arithmetic operation add subtract or multiply 1 Press the Function softkey and select g t Internal 2 Use the Operator Source 1 and Source 2 softkeys to set up an arithmetic operation 3 Press the Function softkey and select f t Displayed 4 Use the Operator softkey to select a transform function FFT Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Math Waveforms 4 5 Press the Source 1 softkey and select g t as the source Note that g t is only available when you select a transform function in the previous step To adjust the math waveform scale and offset 1 Make sure the multiplexed scale and position knobs to the right of the
222. ss about the signal ground level or the center of the display To set the waveform expansion reference point 1 Press Utility gt Options gt Preferences gt Expand and select Ground The displayed waveform will expand about the position of the channel s ground This is the default setting The ground level of the signal is identified by the position of the ground level gt icon at the far left side of the display The ground level will not move when you adjust the vertical sensitivity volts division control Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 243 19 Utility Settings If the ground level is off screen the waveform will expand about the top or bottom edge of the screen based on where the ground is off screen Center The displayed waveform will expand about the center of the display To disable enable transparent backgrounds There is a preference setting for whether measurements statistics reference waveform information and other text displays have transparent or solid backgrounds 1 Press Utility gt Options gt Preferences 2 Press Transparent to toggle between transparent and solid text display backgrounds To load the default label library See To reset the label library to the factory default on page 111 To set up the screen saver The oscilloscope can be configured to turn on a display screen saver when the oscilloscope has been idle for a specified length of
223. ss the Save Recall key to save or recall a waveform or setup See Chapter 17 Save Recall Setups Screens Data starting on page 217 The Print key opens the Print Configuration Menu so you can print the displayed waveforms See Chapter 18 Print Screens starting on page 231 15 Help key Opens the Help Menu where you can display overview help topics and select the Language See also Access the Built In Quick Help on page 40 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 33 1 Getting Started 16 Vertical controls The Vertical controls consist of e Analog channel on off keys Use these keys to switch a channel on or off or to access a channel s menu in the softkeys There is one channel on off key for each analog channel e Vertical scale knob There are knobs marked Ny N for each channel Use these knobs to change the vertical sensitivity gain of each analog channel Vertical position knobs Use these knobs to change a channel s vertical position on the display There is one Vertical Position control for each analog channel Label key Press this key to access the Label Menu which lets you enter labels to identify each trace on the oscilloscope display See Chapter 8 Labels starting on page 107 For more information see Chapter 3 Vertical Controls starting on page 55 17 Analog channel inputs Attach oscillos
224. t Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 175 13 Measurements 176 The Base of a waveform is the mode most common value of the lower part of the waveform or if the mode is not well defined the base is the same as Minimum The Y cursor shows the value being measured Overshoot Overshoot is distortion that follows a major edge transition expressed as a percentage of Amplitude The X cursors show which edge is being measured edge closest to the trigger reference point local Maximum D Top Rising edge overshoot Amplitude x 100 Base D local Minimum Falling edge overshoot Amplitude x 100 Overshoot aoe a Ba local Minimum Overshoot Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Measurements 13 Agilent Acquisition I Channels Over 1 Settings Clear Meas w Figure 34 Automatic Overshoot measurement Preshoot Preshoot is distortion that precedes a major edge transition expressed as a percentage of Amplitude The X cursors show which edge is being measured edge closest to the trigger reference point local Maximum D Top 7 Rising edge preshoot Amplitude 100 _ Base D local Minimum Falling edge preshoot Amplitude x 100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 177 13 Measurements 178 Average DC RMS s Preshoot local Maximum N oN e Preshoot local Minimum Average is the sum
225. t Thus after you use the label list for a while your labels will predominate making it easier to customize the instrument display for your needs When you reset the label library list see next topic all of your custom labels will be deleted and the label list will be returned to its factory configuration To reset the label library to the factory default Pressing the Default Library softkey will remove all user defined labels from the library and set the labels back to the factory default Once deleted these user defined labels cannot be recovered 1 Press Utility gt Options gt Preferences 2 Press the Default Library softkey This will delete all user defined labels from the library and set the labels in the library back to the factory default However this does not default the labels currently assigned to the channels those labels that appear in the waveform area Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 111 8 Labels NOTE Defaulting labels without erasing the default library Pressing Default Setup sets all channel labels back to the default labels but does not erase the list of user defined labels in the library 112 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ee 9 Triggers Adjusting the Trigger Level 114 Forcing a Trigger 115 Edge Trigger 115 Pattern Trigger 118 Pulse Width Trigger 12
226. the Modify softkey also lets you enter these settings Network Domain this is the Windows network domain name Username this is your login name for the Windows network domain Password this is your login password for the Windows network domain To clear an entered password press the Clear Password softkey 5 Use the Spell Enter and Delete Character softkeys to enter the network printer settings Spell press this softkey and turn the entry knob to select the character at the current position Enter press this softkey to enter characters and move the cursor to the next character position Delete Character press the Enter softkey until the desired character is highlighted then press this softkey to delete the character Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 233 18 Print Screens You can use a connected USB keyboard instead of using the Spell and other character editing softkeys 6 Press the Apply softkey to make the printer connection A message appears to tell you whether the connection was successful To specify the print options In the Print Configuration Menu press the Options softkey to change the following options e Setup Information Select this to print oscilloscope setup information on your printout including vertical horizontal trigger acquisition math and display settings e Invert Graticule Colors Select this to reduce the amount of black ink it
227. the display 7 To stop making one or more measurements press the Clear Meas softkey and choose the measurement to clear or press Clear All Clear Measurements Menu Clear Meas 1 Clear Meas 2 Freq 1 Pk Pk 1 After all measurements have been cleared when Meas is pressed again the default measurements will be Frequency and Peak Peak Measurements Summary The automatic measurements provided by the oscilloscope are listed in the following table All measurements are available for analog channel waveforms All measurements except Counter are available for math waveforms other than FFT A limited set of measurements is available for math FFT waveforms and for digital channel waveforms as described in the following table Measurement Valid Valid for Notes for Digital Math Channels FFT Snapshot All on page 173 Amplitude on page 174 Average on page 178 Yes Full Screen Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 171 13 Measurements 172 Measurement Base on page 176 Delay on page 184 Valid for Math FFT Valid for Digital Channels Notes Measures between two sources Press Settings to specify the second source Duty Cycle on page 183 Yes Fall Time on page 183 Frequency on page 182 Yes Maximum on page 174 Yes Minimum on page 174 Yes Overshoot on page 176 Peak Peak on page 1
228. time 1 Press Utility gt Options gt Preferences gt Screen Saver to display the Screen Saver Menu Screen Saver Menu Saver Wait Logo 180min Preview 2 Press the Saver softkey to select the screen saver type The screen saver can be set to Off to display any of the images shown in the list or can display a user defined text string If User is selected press the Spell softkey to select the first character of the text string Use the Entry knob to choose a character Then press the Enter softkey to advance to the next character and repeat the process 244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 You can use a connected USB keyboard instead of using the Spell and other character editing softkeys The resultant string is displayed in the Text line above the softkeys Text AGILENT TECHNOLOGIES Saver Wait User 180min Preview 3 Press the Wait softkey then turn the Entry knob to select the number of minutes to wait before the selected screen saver activates When you turn the Entry knob the number of minutes is displayed on the Wait softkey The default time is 180 minutes 3 hours 4 Press the Preview softkey to preview the screen saver you have selected with the Saver softkey 5 To view the normal display after the screen saver has started press any key or turn any knob To set AutoScale preferences 1 Press Utility gt Options gt Preferences
229. tion temperature e If you want to maximize the measurement accuracy The amount of use environmental conditions and experience with other instruments help determine if you need shorter User Cal intervals User Cal performs an internal self alignment routine to optimize the signal path in the oscilloscope The routine uses internally generated signals to optimize circuits that affect channel sensitivity offset and trigger parameters Performing User Cal will invalidate your Certificate of Calibration If NIST National Institute of Standards and Technology traceability is required perform the Performance Verification procedure in the Agilent Infinii Vision 2000 3000 X Series Oscilloscopes Service Guide using traceable sources To perform user calibration 1 Disconnect all inputs from the front and rear panels including the digital channels cable on an MSO and allow the oscilloscope to warm up before performing this procedure 2 Press the rear panel CAL button to disable calibration protection Connect short 12 inch maximum equal length cables to each analog channel s BNC connector on the front of the oscilloscope You will need two equal length cables for a 2 channel oscilloscope or four equal length cables for a 4 channel oscilloscope Use 50W RG58AU or equivalent BNC cables when performing User Cal Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 For a 2 channel oscilloscope c
230. tion with all Acquisition All pods for the logic channels saved Pods Logic Pile Header Number of Waveforms 2 12 bytes Channels Waveform Header 1 Number of Waveform Buffers 1 140 bytes Waveform Data Header 1 12 bytes Pod 1 Timing Data buffer size Waveform Header 2 Number of Waveform Buffers 1 140 bytes Buffer Type 6 unsigned char Bytes per Point 1 Waveform Data Header 2 12 bytes Buffer Type 6 unsigned char Bytes per Point 1 Pod 2 Timing Data buffer size Segmented The following picture shows a binary file of a segmented memory Memory acquisition on one analog channel Acquisition on One Analog Channel 286 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Reference 21 Fil Header Number of Waveforms N Number of Segments 12 bytes Number of Waveform Buffers 1 Index 1 Time Tag 0 0 Waveform Header 1 140 bytes Waveform Data Header 1 12 bytes Voltage Data 1 buffer size Waveform Header 2 140 bytes Buffer Type 1 floating point Bytes per Point 4 Number of Waveform Buffers 1 Index 2 Time Tag time between segment 1 and 2 Waveform Date Buffer Type 1 floating point Bytes per Point 4 Header 2 12 bytes Voltage Data 2 buffer size Number of Waveform Buffers 1 Index N Time Tag time between segment 1 and N Waveform Header N 140 bytes Waveform Data Buffer Type 1 floating point Bytes
231. tions are running or when they are stopped When running adjusting the horizontal scale knob changes the sample rate When stopped adjusting the horizontal scale knob lets you zoom into acquired data See Panning and Zooming Single or Stopped Acquisitions on page 46 Note that the horizontal scale knob has a different purpose in the Zoom display See To display the zoomed time base on page 50 To adjust the horizontal delay position 1 Turn the horizontal delay position knob lt p The trigger point moves horizontally pausing at 0 00 s mimicking a mechanical detent and the delay value is displayed in the status line Changing the delay time moves the trigger point solid inverted triangle horizontally and indicates how far it is from the time reference point hollow inverted triangle V These reference points are indicated along the top of the display grid Figure 2 shows the trigger point with the delay time set to 200 us The delay time number tells you how far the time reference point is located from the trigger point When delay time is set to zero the delay time indicator overlays the time reference indicator All events displayed left of the trigger point happened before the trigger occurred These events are called pre trigger information and they show events that led up to the trigger point Everything to the right of the trigger point is called post trigger information The amount of delay range pre trigger
232. trigger buffer Then after the pre trigger buffer is filled the oscilloscope starts searching for a trigger and sampled data continues to flow data through the pre trigger buffer in a first in first out FIFO manner When a trigger is found the pre trigger buffer contains the events that occurred just before the trigger Then the oscilloscope fills the post trigger buffer and displays the acquisition memory If the acquisition was initiated by Run Stop the process repeats If the acquisition was initiated by pressing Single the acquisition stops and you can Pan and Zoom the waveform In either Auto or Normal trigger mode a trigger may be missed if the event occurs while the pre trigger buffer is being filled This may be more likely for example when the horizontal scale knob is set to a slow time div setting such as 500 ms div Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Trigger Mode Coupling 10 Trigger Indicator The trigger indicator at the top right of the display shows whether triggers are occurring In the Auto trigger mode the trigger indicator can show Auto flashing the trigger condition is not found after the pre trigger buffer has filled and forced triggers and acquisitions are occurring e Auto not flashing the trigger condition is found or the pre trigger buffer is being filled In the Normal trigger mode the trigger indicator can show Trig d flashing
233. ts look for the position of the start of Vertical Sync to determine the field But this definition of field does not take into consideration the phase of the reference subcarrier When Field 1 is selected the trigger system will find any field where the vertical sync Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 131 9 132 starts on Line 4 In the case of NTSC video the oscilloscope will trigger on color field 1 alternating with color field 3 see the following figure This setup can be used to measure the envelope of the reference burst 424 0 500 0 Vid Field Agilent Acquisition Channels Video Trigger Menu Standard Mode Field Holdoff NTSC Field1 O 0flds Figure 19 Triggering on Color Field 1 Alternating with Color Field 3 If a more detailed analysis is required then only one color field should be selected to be the trigger You can do this by using the Field Holdoff softkey in the Video Trigger Menu Press the Field Holdoff softkey and use the Entry knob to adjust the holdoff in half field increments until the oscilloscope triggers on only one phase of the color burst A quick way to synchronize to the other phase is to briefly disconnect the signal and then reconnect it Repeat until the correct phase is displayed When holdoff is adjusted using the Field Holdoff softkey and the Entry knob the corresponding holdoff time will be displayed in the Trigger Mode and Coupling Menu Agilent Infinii
234. turn them 260 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Web Interface 20 XIE Agilent Technologies InfiniiVision MSO X 20248 Med Segre Ow atoe cea Simple Remote Front Panel To operate the oscilloscope using the web interface s Simple Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 258 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Simple Remote Front Panel After a few seconds the Remote Front Panel appears 3 Use the Main Menu and the Function Keys to control the oscilloscope To view Quick Help right click on a softkey Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 261 20 262 Web Interface Main Menu Function Keys VNC Agilent MSOX2024A InfiniiVision US50210029 Remote Front Panel File Run Control Setup Horizontal Trigger Measure Waveform Tools Run Stop Fault Force Single Trig Horizontal TAE Hint appears lt when you roll Trig teva over with vA mouse Cursors e C8 Intensity va Digital Entry Knob GOR Analog chi G Softkeys left click to select right click for Quick Help Channels All Fast Debug Scrolling and Monitor Resolution When using a monitor resolution of 800 x 600 or less on the remote computer you need to scroll to access the full remote front panel To display the remote front panel w
235. turning the Cursors knob Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Cursor Examples Source Cursors Manual sil Cursors 12 1 00 Agilent Acquisition ormal Channels Cursors Figure 28 Cursors used to measure pulse widths other than middle threshold points Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 163 12 Cursors Agilent Acquisition Channels Cursors 171 000ns 4 18100V nu Mode Source Cursors X1 216 000ns Manual X2 Figure 29 Cursors measure frequency of pulse ringing Expand the display with Zoom mode then characterize the event of interest with the cursors 164 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Cursors 12 1 00 Agilent Acquisition Cursors 64 400ns Source Cursors Units Manual Y2 5 Figure 30 Cursors track Zoom window Put the X1 cursor on one side of a pulse and the X2 cursor on the other side of the pulse Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 165 12 Cursors 500 08 1 00 Agilent tion Channels Cursors Source Cursors Manual X2 Figure 31 Measuring pulse width with cursors Press the X1 X2 linked softkey and move the cursors together to check for pulse width variations in a pulse train 166 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Source Cursors Manual X1X2 Figure 32 Moving the cursors together to check pulse width
236. ty Hardware Self Test is designed to provide an 80 confidence level that the oscilloscope is operating properly Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Utility Settings 19 To perform front panel self test Pressing Utility gt Service gt Front Panel Self Test lets you test the front panel keys and knobs as well as the oscilloscope display Follow the on screen instructions To display oscilloscope information Press Help gt About Oscilloscope to display information about your oscilloscope e Model number e Serial number e Bandwidth e Module installed e Software version e Installed licenses See also Loading Licenses and Displaying License Information on page 279 To display the user calibration status Pressing Utility gt Service gt User Cal Status displays the summary results of the previous User Cal and the status of probe calibrations for probes that can be calibrated Note that passive probes do not need to be calibrated Results User Cal date Change in temperature since last User Cal Failure Comments Probe Cal Status To clean the oscilloscope Remove power from the instrument 2 Clean the external surfaces of the oscilloscope with a soft cloth dampened with a mixture of mild detergent and water 3 Make sure that the instrument is completely dry before reconnecting it to a power source Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 251
237. ty of a selected source All modes operate up to the bandwidth of the oscilloscope except Either edge mode which has a limitation Either edge mode will trigger on Constant Wave signals up to 100 MHz but can trigger on isolated pulses down to 1 2 oscilloscope s bandwidth Using AutoScale The easiest way to set up an Edge trigger on a waveform is to use to Set Up Edge AutoScale Simply press the AutoScale key and the oscilloscope will Triggers attempt to trigger on the waveform using a simple Edge trigger type See Use Auto Scale on page 26 MegaZoom Technology Simplifies Triggering With the built in MegaZoom technology you can simply AutoScale the waveforms then stop the oscilloscope to capture a waveform You can then pan and zoom through the data using the Horizontal and Vertical knobs to find a stable trigger point AutoScale often produces a triggered display Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 117 9 Triggers Pattern Trigger The Pattern trigger identifies a trigger condition by looking for a specified pattern This pattern is a logical AND combination of the channels Each channel can have a value of 0 low 1 high or don t care X A rising or falling edge can be specified for one channel included in the pattern You can also trigger on a hex bus value as described on Hex Bus Pattern Trigger on page 120 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then
238. uired oscilloscope bandwidth 148 reset network password 271 return instrument for service 252 Rise time measurement 183 rise time oscilloscope 147 rise time signal 148 roll mode 47 Run Control keys 31 runt pulses 181 Russian front panel overlays 36 Russian user interface and Quick Help 41 S safety warning 25 sample rate 4 sample rate and memory depth 149 sample rate current rate displayed 44 sample rate oscilloscope 146 147 sampling theory 145 sampling overview 145 save 253 save file 241 save files via web interface 265 save segment 221 save setup files 219 save times data 224 Save to 226 save Quick Save 253 save recall from web interface 265 Save Recall key 33 saver screen 244 saving data 217 SCPI Commands window 263 screen image via web interface 267 screen saver 244 Search key 31 secure erase 229 segmented memory 155 segmented memory re arm time 158 segmented memory saving segments 221 select digital channels 89 select knob 89 Selected 242 selecting values 30 self test front panel 251 self test hardware 250 serial decode controls 32 serial number 251 257 service functions 247 Set all Digits softkey 120 setup files saving 219 setup automatic 85 setup default 25 setups recall 227 SGM 155 SGM license 279 shipping damage 19 shipping precautions 252 Sigma minimum 192 Simple Remote Front Panel 26
239. units has been set to Amps A scale unit of U undefined will be displayed for math functions when two source channels are used and they are set to dissimilar units and the combination of units cannot be resolved Math Operators 66 Math operators perform arithmetic operations add subtract or multiply on analog input channels e Add or Subtract on page 66 e Multiply or Divide on page 67 Add or Subtract When you select add or subtract the Source 1 and Source 2 values are added or subtracted point by point and the result is displayed You can use subtract to make a differential measurement or to compare two waveforms If your waveforms DC offsets are larger than the dynamic range of the oscilloscope s input channels you will need to use a differential probe instead Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Math Waveforms 4 500 05 2 Agilent Acquisition I Channels ft Chi Ch2 j Function Operator Source 1 Source 2 tit Figure 5 Example of Subtract Channel 2 from Channel 1 See Also Units for Math Waveforms on page 65 Multiply or Divide When you select the multiply or divide math function the Source 1 and Source 2 values are multiplied or divided point by point and the result is displayed The divide by zero case places holes that is zero values in the output waveform Multiply is useful for seeing power relationships when one of the cha
240. up menu to switch the selected channel on or off To switch all digital channels on or off 1 Press the Digital key to toggle the display of digital channels The Digital Channel Menu is displayed above the softkeys If you want to switch the digital channels off and the Digital Channel Menu is not already displayed you must push the Digital key twice to switch the digital channels off The first push displays the Digital Channel Menu and the second push switches the channels off To switch groups of channels on or off 1 Press the Digital key on the front panel if the Digital Channel Menu is not already displayed 2 Press the Turn off or Turn on softkey for the D7 DO group Each time you press the softkey the softkey s mode toggles between Turn on and Turn off To change the logic threshold for digital channels 1 Press the Digital key so that the Digital Channel Menu is displayed 2 Press the Thresholds softkey 88 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 3 Press the D7 DO softkey then select a logic family preset or select User to define your own threshold Logic family Threshold Voltage TTL 1 44 V CMOS 2 55 V ECL 1 3 V User Variable from 8 V to 8 V The threshold you set applies to all channels within the selected D7 DO group Each of the two channel groups can be set to a different threshold if desired Values greater than t
241. urement setup and ensure a repeatable method for obtaining test data The 01650 61607 16 channel logic probe cable and the 01650 63203 termination adapter are designed to make it easy to connect to industry standard 20 pin board connectors The cable is a 2 m logic analyzer probe cable and the termination adapter provides the proper RC networks in a very convenient package These parts as well as the 1251 8106 20 pin low profile straight board connector can be ordered from Agilent Technologies To replace digital probe leads 98 If you need to remove a probe lead from the cable insert a paper clip or other small pointed object into the side of the cable assembly and push to release the latch while pulling out the probe lead Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 Table 3 Digital Probe Replacement Parts Part Number Description N6459 60001 Digital probe kit contains N6459 61601 8 channel cable 01650 82103 2 inch probe grounds qty 3 and 5090 4832 grabbers qty 10 N6459 61601 8 channel cable with 8 probe leads and 1 pod ground lead qty 1 5959 9333 Replacement probe leads qty 5 also contains 01650 94309 probe labels 5959 9334 Replacement 2 inch probe grounds qty 5 5959 9335 Replacement pod ground leads qty 5 5090 4833 Grabbers qty 20 01650 94309 Package of probe labels For other replacement parts
242. utio Center Function Operator Source 1 pan fit FFT 500kHe 250kHz Source 1 selects the source for the FFT See To perform a transform function on an arithmetic operation on page 64 for information about using g t as the source Span sets the overall width of the FFT spectrum that you see on the display left to right Divide span by 10 to calculate the number of Hertz per division It is possible to set Span above the maximum available frequency in which case the displayed spectrum will not take up the whole screen Press the Span softkey then turn the Entry knob to set the desired frequency span of the display Center sets the FFT spectrum frequency represented at the center vertical grid line of the display It is possible to set the Center to values below half the span or above the maximum available frequency in which case the displayed spectrum will not take up the whole screen Press the Center softkey then turn the Entry knob to set the desired center frequency of the display Scale lets you set your own vertical scale factors for FFT expressed in dB div decibels division See To adjust the math waveform scale and offset on page 65 Offset lets you set your own offset for the FFT The offset value is in dB and is represented by the center horizontal grid line of the display See To adjust the math waveform scale and offset on page 65 More FFT displays the More FFT Settings Menu 2 Pres
243. vertical horizontal and trigger setup information The display area contains the waveform acquisitions channel identifiers and analog trigger and ground level indicators Each analog channel s information appears in a different color Signal detail is displayed using 256 levels of intensity For more information about viewing signal detail see To adjust waveform intensity on page 101 For more information about display modes see Chapter 7 Display Settings starting on page 101 The information area normally contains acquisition analog channel automatic measurement and cursor results Menu line This line normally contains menu name or other information associated with the selected menu Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 39 1 Getting Started Softkey labels These labels describe softkey functions Typically softkeys let you set up additional parameters for the selected mode or menu Pressing the Back Up key at the top of the menu hierarchy turns off softkey labels and displays additional status information describing channel offset and other configuration parameters Access the Built In Quick Help To view Quick 1 Press and hold the key or softkey for which you would like to view Help help Agilent Acquisition uick Hel Sue p gt 5 BEME message _ sak Detect helb alahas e T is Fast Debug Channels All Normal Press and hold front pan
244. vice 35 USB CD device 243 USB eject device 35 USB storage device numbering 243 usb2 243 user cal 248 user calibration 248 user interface language 41 User defined threshold 89 utilities 237 Utility key 30 V V RMS FFT vertical units 70 values choosing 30 variable persistence 103 ventilation requirements 23 vernier channel 58 Vertical controls 34 vertical expansion 57 vertical position 57 vertical position knobs 34 vertical scale knobs 34 vertical sensitivity 34 57 Vertical Units FFT 70 VGA video output 37 video trigger 123 viewing tilt the instrument 22 VISA connect string 257 voltage measurements 173 WwW warranted specifications 273 warranty 252 Wave Gen key 30 35 waveform generator 205 waveform generator defaults restoring 215 waveform generator expected output load 209 waveform generator logic presets 209 waveform generator sync pulse 208 waveform generator sync pulse TRIG OUT signal 247 waveform generator waveform type 205 Waveform keys 33 waveform type waveform generator 205 waveform cursor tracking 161 waveform intensity 101 waveform printing 231 waveform reference point 243 waveform saving exporting 217 WAVEGEN license 280 web interface 257 web interface accessing 258 white noise adding to waveform generator output 210 Width measurement 183 Width measurement 183 Window FFT 70 X XY mode 47
245. ware is distributed in the hope that it will be useful but WITHOUT ANY WARRANTY without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE See the GNU General Public License for more details This license is located on the Agilent InfiniiVision Oscilloscopes Documentation CD ROM RealVNC source code can be obtained from RealVNC or by contacting Agilent Agilent will charge for the cost of physically performing the source distribution Reference Waveform files use HDF5 HDF5 was developed by The HDF Group and by the National Center for Supercomputing Applications at the University of Illinois at Urbana Champaign Network printing uses the CUPS Common Unix Printing System library The CUPS and CUPS Imaging libraries are developed by Apple Inc and licensed under the GNU Library General Public License LGPL Version 2 This license is located on the Agilent InfiniiVision Oscilloscopes Documentation CD ROM CUPS network printing uses the mDNSResponder library The mDNSResponder library is developed by Apple Inc and licensed under the Apache License Version 2 0 This license is located on the Agilent InfiniiVision Oscilloscopes Documentation CD ROM Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 289 21 Reference 290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index Symbols Width measurement 183 Width measurement 183 A About
246. waveform When the oscilloscope triggers the single acquisition is displayed and the oscilloscope is stopped the Run Stop key is illuminated in red Press Single again to acquire another waveform If the oscilloscope doesn t trigger you can press the Force Trigger key to trigger on anything and make a single acquisition To display the results of multiple acquisitions use persistence See To set or clear persistence on page 103 Singlevs The maximum data record length is greater for a single acquisition than Running and when the oscilloscope is running or when the oscilloscope is stopped Record Length after running e Single Single acquisitions always use the maximum memory available at least twice as much memory as acquisitions captured when running and the oscilloscope stores at least twice as many samples At slower time div settings because there is more memory available for a single acquisition the acquisition has a higher effective sample rate e Running When running versus taking a single acquisition the memory is divided in half This lets the acquisition system acquire one record while processing the previous acquisition dramatically improving the number of waveforms per second processed by the oscilloscope When running a high waveform update rate provides the best representation of your input signal To acquire data with the longest possible record length press the Single key For mor
247. waveforms the scale knob acts like an analog channel vertical scale knob Multiplexed position knob This position knob is used with Math Ref or Digital waveforms whichever has the illuminated arrow to the left For math and reference waveforms the position knob acts like an analog channel vertical position knob 32 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Getting Started 1 12 Measure controls The measure controls consist of Cursors knob Push this knob select cursors from a popup menu Then after the popup menu closes either by timeout or by pushing the knob again rotate the knob to adjust the selected cursor position Cursors key Press this key to open a menu that lets you select the cursors mode and source Meas key Press this key to access a set of predefined measurements See Chapter 13 Measurements starting on page 169 13 Waveform keys The Acquire key lets you select Normal Peak Detect Averaging or High Resolution acquisition modes see Selecting the Acquisition Mode on page 149 and use segmented memory see Acquiring to Segmented Memory on page 155 The Display key lets you access the menu where you can enable persistence see To set or clear persistence on page 103 clear the display and adjust the display grid graticule intensity see To adjust the grid intensity on page 105 14 File keys Pre
248. xplorer 241 file format ASCII 218 file format BIN 218 file format BMP 218 file format CSV 218 file format PNG 218 File keys 33 file name new 226 file save recall load 241 fine adjustment horizontal scale 52 firmware updates 280 firmware upgrade files 269 firmware version information 257 firmware versions 269 flash drive 35 Flat top FFT window 70 FM frequency modulation waveform generator output 213 folding frequency 145 forcing a trigger 115 freeze display 253 freeze display Quick Freeze Display 253 French front panel overlays 36 French user interface and Quick Help 41 frequency deviation FM modulation 213 Frequency measurement 182 frequency modulation FM waveform generator output 213 frequency requirements power source 23 frequency Nyquist 145 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide frequency shift keying modulation FSK waveform generator output 214 front panel controls and connectors 28 front panel self test 251 front panel browser based remote 262 front panel language overlay 35 front panel real scope remote 260 front panel simple remote 261 FSK frequency shift keying modulation waveform generator output 214 G g t 64 gateway IP 239 Gaussian frequency response 146 German front panel overlays 36 German user interface and Quick Help 41 glitch capture 151 glitch trigger 121 golden wave
249. y knob to select a value for the digit If a digit is made up of less than four bits then the value of the digit will be limited to the value that can be created by the selected bits 5 You can use the Setall Digits softkey to set all digits to a particular value When a hex bus digit contains one or more don t care X bits and one or more bit with a value or 0 or 1 the sign will be displayed for the digit For information regarding digital bus display when Pattern triggering see Bus values are displayed when using Pattern trigger on page 92 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Triggers 9 Pulse Width Trigger Pulse Width glitch triggering sets the oscilloscope to trigger on a positive or negative pulse of a specified width If you want to trigger on a specific timeout value use Pattern trigger in the Trigger Menu see Pattern Trigger on page 118 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Pulse Width Channels gt 5 00us 3 Press the Source softkey then rotate the Entry knob to select a channel source for the trigger The channel you select is shown in the upper right corner of the display next to the polarity symbol The source can be any analog or digital channel available on your oscilloscope 4 Adjust the trigger level For analog channels turn the Trigger Level knob Agilent InfiniiVision 2000
250. y the proper mechanical design of the probe tip assembly This provides the maximum input impedance at high frequencies Probe Grounding A probe ground is the low impedance path for current to return to the source from the probe Increased length in this path will at high frequencies create large common mode voltages at the probe input The voltage generated behaves as if this path were an inductor according to the equation pai voi Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Digital Channels 6 Increasing the ground inductance L increasing the current di or decreasing the transition time dt will all result in increasing the voltage V When this voltage exceeds the threshold voltage defined in the oscilloscope a false data measurement will occur Sharing one probe ground with many probes forces all the current that flows into each probe to return through the same common ground inductance of the probe whose ground return is used The result is increased current di in the above equation and depending on the transition time dt the common mode voltage may increase to a level that causes false data generation Probe 1 lt Zin C iy L GND Probe Vn Common Mode Ground 4 i Itlatlns igtin Probe 2 lt 4 Zin a Probe N lt Zin Figure 12 Common Mode Input Voltage Model In addition to the com
251. yed Setup files have the extension SCP These extensions appear when using the File Explorer see File Explorer on page 241 but they do not appear when using the Recall Menu To save BMP or PNG image files Image files can be saved to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select 8 bit Bitmap image bmp 24 bit Bitmap image bmp or 24 bit image png 2 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 226 3 Press the Settings softkey In the File Settings Menu you have these softkeys and options Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 219 17 Save Recall Setups Screens Data Setup Info setup information vertical horizontal trigger acquisition math and display settings is also saved in a separate file with a TXT extension Invert Grat the graticule in the image file has a white background instead of the black backgound that appears on screen Graticule Not Inverted Graticule Inverted Palette lets you choose between Color or Grayscale images 4 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed When saving screen images the oscilloscope uses the last menu visited before pressing the Save Recall key This lets you save any relevant information within
252. ys 36 Traditional Chinese user interface and Quick Help 41 transforms math 68 transient withstand capability 275 Transparent backgrounds 244 TRIG OUT connector 38 246 Trig d trigger indicator 137 Trig d trigger indicator 137 Trigger controls 30 trigger coupling 138 trigger indicator Auto 137 trigger indicator Trig d 137 trigger indicator Trig d 137 trigger level 114 trigger mode auto or normal 136 trigger mode Quick Trigger Mode 253 trigger output 246 trigger output mask test 193 247 trigger type edge 115 trigger type glitch 121 trigger type hex bus 120 trigger type pattern 118 trigger type pulse width 121 trigger type slope 115 trigger type video 123 trigger types 113 trigger definition 114 trigger external 140 trigger forcing a 115 trigger general information 114 trigger holdoff 140 trigger mode coupling 135 trigger source 115 triggers TRIG OUT signal 247 TTL threshold 89 turn channel on 34 U under sampled signals 145 units cursor 162 units external trigger probe 141 296 units math 65 units probe 60 updating software and firmware 280 upgrade files 269 upgrade options 279 upgrading the oscilloscope 280 upload new firmware 257 usb 243 USB device port 38 USB device port remote control 237 USB host port 38 231 USB host ports 35 USB printer 231 USB printers supported 231 USB storage de