Home

Paradigm Real Time Troubleshooting Guide

image

Contents

1. Paradigm REAL time insulin pump amp continuous glucose monitoring system Troubleshooting Guide A practical guide for health professionals working with the Paradigm REAL Time Insulin pump and Continuous Glucose Monitoring System Jill Milliken RN CDE J Milliken RN CDE 2006 1 SN MINIMED Table of Contents 1 Overview of the Paradigm REAL Time components 2 Troubleshooting a Sensor b Skin and Taping Daily Self Management Education Data Management J Milliken RN CDE 2006 P MINIMED Overview of Paradigm REAL Time System Sen Serter Transmitter Sensor Communication ICONS 4 Sensor is on but not communicating y Antenna Icon sensor is ON and transmitting The sensor and the infusion set should be no less than 2 inches apart J Milliken RN CDE 2006 3 Troubleshooting Basics Troubleshooting the Paradigm REAL time system can be time consuming for the patient and their provider if a basic level of understanding has not been achieved during the training on the system This guide will help you understand and incorporate solutions to the most common issues related to this new technology Main technical issues include Sensor to meter glucose comparison Tools for Troubleshooting e Sensor Calibration All of the sensor troubleshooting needed can be done using e Error alarms information from the following three sources e Communication interruption e Adhes
2. to helping your patient manage these issues can make a difference J Milliken RN CDE 2006 7 gt Medtronic MINIMED Calibrating Calibrations are essential for the Paradigm system to function The quality of the data entered will help determine the quality of the information being made available to the patient Best calibration practices include using a reliable glucose meter and entering a glucose value before breakfast dinner and bedtime as a calibration Common pitfalls in terms of calibration are as follows Entering the same calibration value twice in a row within a few minutes Using more than one meter to enter calibrations Entering too many calibrations over a 24 hour period Rounding off calibration values Inserting the sensor in too late in the day to enter the recommended three calibrations in the first 6 hours Entering a calibration during a period of rapid glucose change as indicated by two trend arrows up or down Calibrations can be entered manually through the bolus or sensor menus If using the Bolus Wizard the pump will ask whether the entered blood glucose value should be entered as a calibration Suggest to the patient that they choose ho if it is a post meal value or during a time of rapid glucose fluctuation If using the BD Link meter the patient should should be encouraged to turn off the RF communication on the meter so that it does not automatically calibrate the s
3. all the tech support number on the back of the pump after exhausting all options The current software version for Solutions pumps and meters is 7 0c The CareLink website is http carelink minimed com J Milliken RN CDE 2006 20 SS Medtronic MINIMED
4. bad sensor within the initialization phase even as early as 5 minutes after connection The patient then is prompted to change the sensor and may use two or more and keep getting the same alarms Canada was the first country in the world to launch this product and we have the oldest transmitters in the field We learn from your experiences and that of your patients If your patient has any questions about warranty transmitter or sensor replacement please have them call Canadian Customer Service at 1 800 284 4416 J Milliken RN CDE 2006 16 SN Medtronic MINIMED Using Bolus Wizard amp real time glucose The Bolus Wizard tool becomes more important with the advent of real time glucose Patients must understand how to self manage using the new information available with the Paradigm REAL time Why 1 The temptation to correct with every threshold alarm 2 Lack of overall understanding of insulin action The basics of the Bolus Wizard include the use of Insulin to carbohydrate ratio Insulin sensitivity factor Target glucose pre meal Active insulin curve 2 8hrs default 6hrs The Bolus Wizard calculation is done when the glucose is entered into the pump The calculation is as follows Correction insulin if needed PLUS any bolus for food needed MINUS active insulin from correction insulin only not food bolus insulin Scenario Tom takes his glucose after receiving a high threshold alarm at 13 7mmo
5. ch an inch e Use the upper outer buttocks even if there is not much fat there it seems to be less sensitive to insert a sensor into than the abdomen e The three ways to insert a sensor are using the serter stretching the skin between the feet of the sen serter manually by pinching and using the serter while pinching The latter can be tricky to do and takes practice Don t worry if you happen to hit a surface capillary and it bleeds a bit It may not mean that the sensor is no good under the skin surface Practice makes perfect The first few sensor insertions are often awkward for the patient They can press the site too hard as they try and remove the sen serter or introducer needle Encourage them to get the introducer needle out as quickly as possible just after insertion to minimize the amount of mobility of the needle tip in the tissues This can lead to lingering discomfort For insertions in the back of the body recommend to the patient that they use a mirror to guide the insertion of the sensor Placement in the outer buttock area will save pump sites Inserting the sensor into scar tissue such as commonly seen on the abdomen of a long term insulin user can yiela unreliable sensor values Evaluate best sensor sites with patient J Milliken RN CDE 2006 14 gt Medtronic MINIMED Sensor Insertion Cont d eChoose a site with enough body fat for sensor insertion As a rule wing avoid frequent inje
6. ction or pump sites or areas of scarred lumpy or K hard tissue amp 2 p SL TM Jof ta Needle wit a guard on F Stretch skin between the feet of the Sen Serter Feet gt O Secure the sensor once inserted with your fingers and slide the Sen Serter laterally off the sensor eRemove the needle by pulling the wing firmly Dispose of the needle in a biohazardous sharps disposal unit J Milliken RN CDE 2006 15 MINIMED Transmitter end of life behaviour The current transmitter has an internal battery which has a variable life depending on the degree of use The following information will help you troubleshoot the transmitter when it is nearing the end of it s use Note as the transmitter battery begins to lose strength it does not cause unreliable sensor glucose values A low transmitter battery notice can be found in the sensor status screen only when the sensor is in use It may take up to two weeks before the transmitter completely fails If the sensor is not in use the sensor status screen cannot be accessed to provide this information The test plug may be used to verify battery life If it does not beep when connected to the test plug the transmitter is likely finished On some occasions however even with a beep when the transmitter is connected there is not enough battery life to fully initialize a sensor The behaviour in this scenario is that the pump will alarm
7. d fluctuation the sensor can be off in comparison to blood glucose this is called sensor lag and is different than the acceptable variance between sensor and blood glucose Many patients appreciate this technical explanation of what is happening and why there is a difference in values J Milliken RN CDE 2006 6 E gt Medtronic TE MINIMED Making sense of BG vs Sensor Descrepancy With early sensor use comes gradual increasing understanding of the system Support is essential during the early days as the patient recognizes differences between meter and sensor relate to several things Rapid fluctuations cause greater discrepancy in values than stable ones What an acceptable variable is 2 mmol at the hyperglycemic range is more acceptable than a 2 mmol variance at the low end of the range However in each scenario the system is functioning normally This is a problem with diabetes not the system There are some issues that can cause greater issues of meter vs BG discrepancy which relate to the meter As the patient becomes more familiar and comfortable with the system they will experience greater satisfaction This is due to the realization that the trends indicated by the arrows are more important than the point to point comparison the alert thresholds can help them identify behaviors which lead to alarms change the behaviors control the excursions and get closer meter to sensor values Your approach
8. d patience can help af J Milliken RN CDE 2006 5 SN Medtronic MINIMED te Sensor vs Meter BG Troubleshooting Discrepancy between sensor and meter BG Possible causes e New sensor has to settle in to tissue requires 3 calibrations within first 6 hours e Sensor placed late in the day evening and not calibrated till morning e Meter variance e Rapidly rising blood glucose Sensor Lag between fluid compartments e Unrealistic expectations Reinforce to patient that there will always be a difference between interstitial glucose and blood glucose Sometimes it helps to explain what is actually happening when a value is provided The sensor is a relative scale of blood glucose The sensor reads in electrical values the blood in mmol L When a calibration is entered the electrical signal also known as an Input Signal ISIG is referenced or stamped to that blood glucose entry Then as the ISIG rises or falls the sensor provides a value based on this relative scale ie 100na sensor ISIG a blood glucose is entered as a calibration at 5 2mmol L The algorithm calculates and accepts this pairing When a few minutes later the ISIG has risen to 119 na the sensor will read something like 5 7mmol L or conversely the ISIG drops to 75 na the value read off the pump will be more in the range of 4 6mmol L as an example Because it takes time for the body s fluids to equalize during a rapi
9. ensor Refer to the BD link meter user guide for instructions on this meter function More is not better in the case of calibrations Too many can throw off the algorithm that is being used to calculate sensor J Milliken RN CDE 2006 g R Medtronic MINIMED Bad Sensor Alarm When did alarm happen Within the first 12 hours At sensor insertion Were the sensor and transmitter connected PRIOR to hitting new sensor After 12 hours of use Did you get 2 cal error s What are the ISIG s previous to this alarm ISIG s lt 5 sensor may not be settled in enough Did you calibrate 1 2 hours after a meal today Have you calibrated the system too much today lf a Cal error occurs wait at least 2 hours to calibrate again eck sensor transmitter connections ake sure the two are connected it New Sensor again Review optimal Sensor has reached Insert a new sensor J Milliken RN CDE 2006 9 gt Medtronic Skin adhesion and dressings Skin care taping and adhesion become necessary areas to address during training on the Paradigm REAL time system Actual skin allergy is rare Most skin issue reports are related to irritation and lack of adhesion What have we learned The transmitter adhesive pad can cause the most irritation especially if it is left in one place more than 2 days After contact with water most adhesives will loosen It is very commo
10. holds should be considered dynamic The longer on the sensor the tighter the thresholds can be set J Milliken RN CDE 2006 18 R Medtronic MINIMED Follow up and support Several key areas should be addressed at the first subsequent visit after the first sensor use Some areas for discussion include e General activities of daily life and sensor wear sports sleeping clothing questions from other people ability to hear the alarm during work or overnight e Any skin issues Local discomfort Tape not sticking Rotating transmitter adhesion area hair itchy e Calibrations are they well correlated What is the difference they noticed If more than 2mmo was that during a rapid rise in BG Did they calibrate pre meal and bedtime General observations about calibration e Alarms If the alert thresholds were being used were they effective at notifying the patient for events Were they aware of any event before the alarm went off or did the alarm warn them Are the alarms keeping them up Do they need to be adjusted Is the snooze interval effective Errors Were there any error alarms Did they verify the ISIG Troubleshoot Re start sensor Call the tech support line e Glucose awareness Did you view any trend arrows Did you intervene What did you do Verify by finger stick Reinforce diabetes basics treating lows and highs Did you identify any trends which you were not aware of Yo
11. it can appear as a red weepy rash Use antibiotic ointment and keep the area dry Suggest alternative ways to secure the transmitter J Milliken RN CDE 2006 10 gt Medtronic MINIMED Allergy vs sensitivity True allergy to the sensor or other components of the system should be reported to the physician Sensor allergy is very rare but is independent of any dressings worn or tried and is directly coming from the sensor insertion site Localized swelling and oozing from the sensor site is an indication of a sensor allergy Asking the patient the following questions will help you decide what kind of issue it is Is this the first time you have used this product sensor IV 3000 transmitter tape Is the redness in the shape of the dressing itself What are your symptoms Itchy Swelling redness Is there any oozing from the sensor site itself How long has it been in u ew N How long had your transmitter been taped into one place Based on the answers to these questions you will be able to help determine whether the issue is localized or generalized symptomatic or non symptomatic possible allergy or sensitivity to a dressing or product true allergy to sensor Ask the patient to report any sensor allergy to the 1 O 800 technical support line J Milliken RN CDE 2006 11 R Medtronic MINIMED Preparation for Sensor Insertion Checklist Gather the below supplies and equipment in
12. ive and skin issues Test Plug e Software and downloading Screen Icons 7 Input Signal ISIG measured in nA This guide is meant to augment the existing User Guides for the Paradigm REAL Time and Transmitter Look for this symbol for quick O tips or other information J Milliken RN CDE 2006 4 R Medtronic MINIMED Tips for Top Ten Technical Troubleshooting Frequent Bad Sensor Alarms in first 12 hours Sensor was put in the body cold Leave out of fridge overnight Normal to some point Re calibrate and wait If continues may be related to end of transmitter life eLost sensor during initialization eNo Transmitter ID entered or entered wrong Connections not clicked in place eNo antenna icon after sensor insertion Forgot to go to start new sensor and press ACT eNo ISIG only in sensor status l l Sensor not fully connected to transmitter on both sides two clicks no transmitter ID have not started sensor or dead sensor Weak Signal Lost sensor Transmitter and pump not in close enough proximity Re approximate and wait Calibration Error eA calibration was entered that is outside of the acceptable range of the current algorithm May be caused by Sensor end e Great discrepancy between entered meter value and sensed value Low sensor ISIG Often the solution to the problem is overlooked due to it s simplicity Trial and error an
13. l 2 5 hours after his meal He enters that value into the bolus wizard and enters 0 for food Based on his ratios his pump estimates he needs 3 u to correct but then subtracts 1 7 u from the total bolus and suggests a correction bolus amount of 1 3u At this point Tom has a choice he can ACT and deliver the correction or wait until his next meal to correct Some practitioners will suggest that the patient NOT correct a mid meal glucose and instead wait to see where he lands however others will recognize that glucose exposure during this hyperglycemic time can cause complications by virtue of higher Alc over the long term and as long as the Bolus Wizard is used for the next meal bolusing again will not cause overlapping insulin stacking J Milliken RN CDE 2006 17 gt Medtronic ee MINIMED Setting Alert Thresholds Thresholds for hypoglycemia or hyperglycemia alerts can be turned ON or Off raised and lowered at an increased or decreased alarm interval snooze There are some considerations when setting threshold alarms 1 What kind of control does the patient currently have e Most patients are unaware of their overall daily control and may find frequent alarming a nuicance This is a problem with diabetes not the sensor or pump system Intervention set the hyperglycemia target high or leave it off altogether for the first sensor A suggested initial threshold is 13 or 14 mmol L 2 How do you as the hea
14. lth provider wish the patient to intervene with the alerts e The alerts are only as effective as the decisions or behaviour changes they can prompt If your patient is having trouble learning the therapy or is going through a time when evaluation of glucose is not a priority it may be best to leave the alerts off The information can be downloaded and observed retrospectively with the care provider e In the case of pediatrics the parent may choose to leave these alarms off during the school day so that there is no confusion on what is to be done about them Scrolling back through the day s values can be sufficient in some children 5 Does the patient have hypoglycemia unawareness or a history of lack of hypo symptoms e In the case of severe hypoglycemia or lack of hypo symptoms the low threshold may be set a little higher than the recommended 4 5mmol L in order to increase the chances of capturing a low glucose The higher the low threshold alert is set however the more frequent the false alarms to the patient For example if the low threshold is set at 5 0mmol any sensor glucose up to 7 mmol is still an acceptable reading The frequent alarming within the normal range may cause the boy that cried wolf phenomenon ignoring the alarms e The trade off of more frequent alarming may still be acceptable to the patient with hypo unawareness Discussion of the alert threshold with the health provider is warranted Alert thres
15. n that the edges peel up from the IV 3000 after a bath shower or swimming It may or may not mean that the dressing has to be replaced Based on trial and error with hundreds of people using the transmitter we have developed the following tips or suggestions Use a piece of IV 3000 under the two sided transmitter adhesive tape Remove the transmitter adhesive tape every 2 days and move it to another area of skin without moving the sensor site Pump users are familiar with site rotation for their infusion sets this is rotation of the site for the transmitter You may also suggest leaving the transmitter un taped to the skin altogether tucked into a waistband or placed in a baby sock and pinned to inside of clothing After immersion in water re taping is probably needed The old dressing may be removed while holding a finger on the sensor site and gently stretching and pulling the dressing off Use an IV prep pad around the skin surrounding the sensor and replace the IV 3000 Check the ISIG and the calibration variant if there is question whether water has entered the sensor site Unusually high gt 300 or poorly correlated BG vs sensor values all may be due to water entering the sensor Though unusual it is worth suggesting to patients that they reference a finger stick to their sensor value after bathing or immersing the transmitter and sensor in water If the transmitter adhesive pad has been left on the skin for more than 2 days
16. preparation for sensor insertion Paradigm REAL Time Monitor Ensure your BG meter has been calibrated with the Transmitter lab and is functioning well sen Serter Use the same BG meter throughout each 72 hour l Transmitter adhesive pad sensor E 2 Alcohol pads Wash your hands 2 IV 3000 occlusive dressing eYou may open the sensor package and leave the sensor on the inner sterile surface 1 Sensor Open the outer wrapping of the occlusive dressing If Smith amp Nephew IV 3000 is not available Op site or Tegaderm can be used eYou will be placing one IV 3000 horizontally and one vertically over the sensor Allow the skin to dry well after it has been cleansed Do not use IV prep pads on the skin with alcohol before placing the sensor Use it after the sensor is in place to help secure the IV 3000 or tegaderm dressing Milliken RN CDE 2006 12 gt Medtronic MINIMED Procedure for inserting the sensor 1 Fit sensor into Sen serter 9 Hold the Sen serter on the skin at the angle 2 Load Sen serter by pushing firmly downwards on the indicated by the feet of the Sen serter about 45 sensor base until you hear a click 10 Unlock the Sen serter if locked by turning the button a quarter turn 3 You may lock the Sen serter by turning the white ar j button on the top end a quarter turn This ensures it 11 Insert the sensor by pushing slowly and firmly on the will not be
17. released accidentally white button 4 Remove first layer of clear tape surrounding the 12 Hold sensor in place and remove the Sen serter by needle sliding it laterally along the skin 5 Remove the clear needle guard from the needle 13 Remove the second layer of tape by pulling firmly on SSS the white tab at the base of the sensor 6 Choose insertion site lower abdomen outer quadrant of the buttock or upper arm 14 Connect the sensor to the transmitter and make sure both sides have clicked into place 2 Clean the intended area of insertion with an alcohol g we ae swab or isopropyl alcohol in circular motion 15 Tuck the transmitter into a waist band to prevent the weight of it pulling out the sensor before you 8 Allow skin to dry Stretch skin slightly between the have it taped down feet of the Sen Serter using your fingers 16 Tape the sensor site with IV 3000 or other occlusive dressing Dispose of needle in a sharps container J Milliken RN CDE 2006 13 gt Medtronic MINIMED Sensor Insertion The sensor should be left out of the fridge even as long as overnight and sensors are useable at room temperature up to 5 days This will help avoid the irritating bad sensor alarms in the first few hours Sensors are best placed in areas of fatty tissue however many children teens and adults who use the sensor are lean There are a couple suggestions for helping place sensors in those who cant quite pin
18. ur patient will need help learning how to use this new information Many people are unaware of their daily glucose variations J Milliken RN CDE 2006 19 SS Medtronic MINIMED Using Data Management There are exciting options for viewing sensor glucose and other parameters using either Solutions pumps and meters software or CareLink web based software Patients can access either system with the following components e Internet access high speed is best e Paradigm REAL Time pump e Com Link or BD Link Meter and cable For radio frequency RF interface There are some common issues which patients can encounter when downloading their data which can interfere in the completion of the data transfer These include interference caused by other radio frequency devices such as optical mice and the computer itself Stretch the cable away from the computer and place the pump right beside the RF interface dial up connection can cause a delay in the data transfer and the server hangs up on the download There is no communication between the pump and the software usually caused by not having the proper Serial numbers of the pump and or the link meter entered into the system There is activity going on with the pump at the time a bolus a temp basal low battery etc In rare cases there is a problem located within the pump software which causes a problem with communication to the software Ask your patient to c