1.                                                            7   3 SAFETY 10  Ere Lern 10   4 OPERATION 14  Installing the Mattress Replacement 5                                     4           4                      22              14  Installing the Side                                                                            EUNT CER HUE a a          l6  Installing the Power Unit     16  Connecting the lose     ataca          e n E aa              IN pee a                 16  Connecting the  Power  Cord ce ee      asa aedi ER DR M e URN cel ne uae P De diee b ua ada 18  Using the Front Panel    eo repeteret RR ge RUD ERI QE                   ov sette Ree Mr vaipundeldsdecnahsas EEE EEE Oa 19                                                                                                     ds Ca tee E eR ae          26                                                                                                                                                  pi da                       27  Transferring Patient From To                                                    8           8                     28  Transferring  Patient From Toia Wheelchair    ttt hte det aia li bb Hd ER de p actae          a ol 28                                                                                                       RUE                                          29   5 MAINTENANCE AND TROUBLESHOOTING 3l  Cleaning the System    esc pO UU Ee aa a da            ee a idee dedere t RM de
2.        Lock Mode   Pressing this button locks out all control unit functions  including the On button  to prevent any tampering with the settings   Press the button for approximately 3 5 seconds for the Lock LED to activate        The Lock LED illuminates when in this mode       Alarm Silence mode   Pressing this button silences the alarm that sounds in the event of power failure or when the hose is disconnected  from the control unit        The Alarm LED illuminates when in this mode     Part No 1148139 25 Invacare    MA90Z Series Mattresses    4 OPERATION  Power Fail LED    In the event of power outage  an alarm sounds and the Power Fail LED flashes AMBER  The control unit has internal memory and retains the previous    settings during the power outage        During a power outage  the MA90Z  MA90ZB42 and MA90ZB48 mattress retains the air as long as the mattress is    connected to the control unit     Low Pressure LED    In the event that the mattress hose disconnects  an alarm sounds and Low Pressure LED flashes AMBER  Once the low pressure problem is fixed  the    control unit resumes operation in the previously set mode     Displaying the Total Run Time  MA95Z Power Unit     Every time the control unit is plugged in  the LCD screen displays the Run Time in hours and minutes  For example  2 584 hours and 25 minutes    displays as 002584 Hours 000025 Minutes    To display the Run Time during use       Ensure the control unit is plugged in and the Standby LED is On   2  Pre
3.      fw     FIGURE 2 Installing the Power Unit    4 4 Connecting the Hose    CAUTION    Ensure that the hose connecting the control unit to the mattress is routed such that it cannot be stepped on  kinked   squeezed or otherwise damaged     Invacare    MA90Z Series Mattresses 16 Part No 1148139       Locate the hose at the foot end of the mattress   2  Locate the control unit connector on the right side of the control unit   3  Perform one of the following         907 Power Unit   Perform the following steps  Detail             i  Squeeze and hold the tabs on the hose connectors     ii  Insert the hose connectors into the control unit connector     4 OPERATION    iii  Push the hose connectors onto the control connector until an audible click is heard               MA95Z Power Unit   Perform the following steps  Detail             i  Lift and hold the connector cover up   ii    Push the hose connector onto the control connector     iii  Ensure the hose connector is flush with the control unit connector     DETAIL            MA90Z           The audible click indicates that the hose connector s  are properly engaged with the control connectors     DETAIL            MA95Z CONNECTOR       CONNECTOR MA95Z  MA95ZB42         MA9OZ  MA95ZB48 Hose  MA90ZB42 Connector    and  MA90ZB48  Hose Connector  Connector Cover  ontrol Unit  Control Unit Connectors  FIGURE 3 Connecting the Hose  Part No 1148139 17 Invacare    MA90Z Series Mattresses    4                     4 5 Connecting the Powe
4.    User Manual Gay    Invacare     44902 Series Mattresses  MA90Z Lateral Rotation with Alternating Pressure and On Demand Low Air Loss System  MA90ZB42 Lateral Rotation with Alternating Pressure and On Demand Low Air Loss Bariatric 42   MA90ZB48 Lateral Rotation with Alternating Pressure and On Demand Low Air Loss Bariatric 48     Invacare    MA95Z Series Mattresses  MA95Z Lateral Rotation with Alternating Pressure and True Low Air Loss System  MA95ZB42 Lateral Rotation with Alternating Pressure and True Low Air Loss Bariatric 42     MA95ZB48 Lateral Rotation with Alternating Pressure and True Low Air Loss Bariatric 48       This manual MUST be given to the user of the product        BEFORE using this product  read this manual and save for future reference     Yes  you can        2011 Invacare Corporation  All rights reserved  Republication  duplication or modification in whole or      part is prohibited without prior written  permission from Invacare  Trademarks are identified by     and 9   All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries  unless otherwise noted     Invacare    MA90Z Series Mattresses 2 Part No 1148139                  5         GENERAL 4  MALUI ee      T               oa 4  Limited Acura D aa ada cai d t   a o               a oa a aa           5   2 OVERVIEW 6  Mabel iL OCation                                                                 6  Typical Product Parameters peu                                                
5.   Press and hold the Turn Angle and        Inflate buttons      The Fowler LED is lit and L displays in the LCD Display   ii  Raise the bed to 45          The control unit will beep and will return to Standby     Part No 1148139 23 Invacare    MA90Z Series Mattresses    4                       iii  Perform one of the following       Cancel   Press the Turn Angle button to exit Fowler Transmitter Setup without linking to a transmitter     Save   Wait 60 seconds without pressing a key        MA95Z Power Unit   Perform the following steps    i  Press and hold the Hold and        Inflate buttons    ii  AUTO FOWLER LEARN MODE will display          PLACE BED IN FLAT POSITION will display    iv  CONTINUE   gt  SELECT will display    v  Press the Select button    vi  PLACE BED IN FOWLER UPRIGHT POSITION will display    vii  TRANSMITTER DETECTED will display    viii  The control unit will beep    ix  STAND BY will display     Mode Button  MA90Z Power Unit        Press the Mode button t to select the Max Inflate Mode or the Low Air Loss mode              Inflate Mode   In this mode  the mattress inflates rapidly to maximum firmness  pressurized to 35   6 mmHg   A series of beeps sound  every three minutes as a reminder that the Max Inflate mode is active  After 30 minutes  the Max Inflate mode deactivates and the control  unit defaults to the previous setting        It takes 5 10 minutes for the MA90Z mattress to inflate fully  inflation time depends on size of mattress    Max Inflate 
6.  cycle duration can be set from   to 99 minutes  In this setting  the mattress will remain static at the current pres   sure settings for the time selected  Then  every two minutes  for the time selected  the pressure in the mattress will  decrease by 20  for five seconds  This pulse cycle repeats itself for the selected time   i e   if 10 minutes is selected  the  mattress will initially remain static for 10 minutes  Then  every two minutes  the pressure      the mattress will decrease  by 20  for five seconds  the mattress will then inflate back to the set pressure setting and remain static for the time  selected and the cycle is repeated            Part No 1148139 21 Invacare    MA90Z Series Mattresses    4                       FUNCTION DESCRIPTION       FOWLER  UPRIGHT   Fowler Mode  The pressure will increase to prevent the patient from bottoming out     3  Use the Select button to move the pointer to the times or attributes available for the selected function     4  Use the Firmer or Softer buttons to change the times or attributes     Hold Button  MA95Z Power Unit        Press the Hold button 0 to pause the mattress in the current position  The mattress will remain in this position until the Hold button is  pressed again     Firm Soft Buttons    l  Select comfort pressure settings by pressing Firm A or Soft M  buttons     Soft Button   Pressing this button reduces the pressure      the mattress     Firm Button   Pressing this button increases the pressure in the mattr
7.  filter cover to the back of the control unit using the two screws     Part No 1148139 33 Invacare    MA90Z Series Mattresses    5 MAINTENANCE AND TROUBLESHOOTING    Filter Cover    Screws          FIGURE   Changing the Filter    Invacare    MA90Z Series Mattresses 34 Part      1148139    5 3 Troubleshooting    PROBLEM    Mattress not inflating  Not alternating properly    5 MAINTENANCE AND TROUBLESHOOTING    Mattress hose disconnected    SOLUTION    Connect hose connectors and lock them  in place       Air hose kinked or split    Unkink hose or replace split hose       Major leak in the air cushions or overlay  pad    Replace leaking air cushions or overlay  pad       Kinked or split manifold    Unkink manifold or replace split manifold       Has power and fuse is good  control unit  does not come on    Send unit back for repair       Not alternating  solenoid or valve mal   function    Send control unit back to factory for  repair       No air  pump or blower malfunction     Send unit for repair       Part      1148139    Control unit off    Check power source and turn on unit       Power cord disconnected    Connect power cord to the power  source       No power in the power source    Check power source has power power  and turn it On       Power outage    Wait till the power source has power          Blown fuse       35       Replace blown fuse with an equivalent  fuse    Invacare    MA90Z Series Mattresses    Part      1148139    Rev D  1 11    Invacare Corporation    US
8.  invacare com or your dealer    p n 1150708 Rev A       This label is on the foot  end of the mattress  cover  not shown                     Invacare    MA90Z Series Mattresses 6 Part No 1148139    22 Typical Product Parameters    2 OVERVIEW    MA90Z power unit is also used      MA90ZB42        MA90ZB48 systems   MA95Z power unit is also used on MA95ZB42 and MA95ZB48 systems     Electrical Parameters    INPUT VOLTAGE AC        INPUT FREQUENCY        CURRENT             2         MAXIMUM POWER CONSUMPTION     30   10 W    150   10 W       CIRCUIT PROTECTION     Dual fused   250 V     A fast blow fuses    Dual fused   250 V   5 A fast blow fuses          MODE OF OPERATION     Performance Parameters    WEIGHT CAPACITY  STANDARD MATTRESS   BARIATRIC MATTRESS  42 inch   48 inch     Part No 1148139             Continuous    350 Ibs    650 Ibs  1000 Ibs       Invacare    MA90Z Series Mattresses    2 OVERVIEW    PRESSURE ZONE        MAXIMUM FLOW     50   15 LPM 1275   100 LPM       MAXIMUM PRESSURE     35   5 mmHg       MAXIMUM FLOW TIMER     30 minutes 15 minutes       SUPPORT SURFACE INFLATION TIME     5   10 minutes 25   60 seconds          PATIENT COMFORT CONTROL PRESSURES     SOFT PRESSURE   FIRM PRESSURE     8   4 mmHg  32   4 mmHg       ROTATION TIME     5 minutes   4 hrs    10  20  30  60 minutes       ROTATION ANGLE     0   30     5      0   40     5                                              Invacare    MA90Z Series Mattresses       Control unit and mattress have  Latex F
9.  s    Perform one of the following       fblood is present  decontaminate the product   i  Remove all visible soil with disposable paper towels     ii    Scrub the area with freshly prepared effective phenolic detergent disinfectant solution     Part No 1148139 3l Invacare    MA90Z Series Mattresses    5 MAINTENANCE AND TROUBLESHOOTING      If blood is not present  remove any soil from the cover with paper towels      If soiled  the cover should be removed  cleaned and decontaminated     8  Perform the following steps to clean the mattress cover         Using a clean sponge or paper towel  wipe down the cover surface with a dilute detergent solution of quaternary cleaner disinfectant or  other germicidal detergent solution     B  Remove the cover and launder it using the following method     A CAUTION  Laundry workers should always wash their hands before working with clean bedding   DO NOT overload the machine   DO NOT use chlorine bleach because it may damage the fabric coating     High air temperatures will damage the fabrics and void the Invacare warranty     i  Place the cover in a washing machine   ii  Wash with warm water  below 120 F    iii  Add detergent and disinfectant according to the manufacturer s instructions   iv  Remove excess water   v  Set the dryer to the lowest setting  below 120 F    vi  Dry the cover until it is completely dry   9  Perform the following steps to clean the control unit and hose fittings       Wipe all controls  chassis and hose fittings 
10. A   One Invacare Way  Elyria  Ohio USA  44036 2125  800 333 6900    www invacare com    Canada   570 Matheson Blvd E  Unit 8   Mississauga Ontario  L4Z 4G4 Canada  800 668 5324       Yes  you can     
11. D FITNESS FOR A PARTICULAR PURPOSE  AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER  SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF  THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN  THE APPLICATION OF ANY IMPLIED WARRANTY  WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN  INVACARE SHALL  NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER    THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS     Part No 1148139 5 Invacare    MA90Z Series Mattresses    2 OVERVIEW  2 Overview    2 1 Label Location       DANGER EXPLOSION HAZARD  DO NOT use in the presence of flammable anesthetics     CAUTION  Equipment should be connected to a properly grounded receptacle  3 prong   Risk of  Electrical shock  DO NOT remove back  Disconnect air hose before administering CPR        This label is on the back of  the unit and also contains       the serial number                 ATTENTION  Clean Filter Every 5 months and Whenever Dirty   Remove Filter  Wash  Dry  amp  Replace               MA95Z  MA95ZB42 and  MA95ZB48 only        WARNING    Patient entrapment with bed side rails may cause injury or death  Mattress MUST fit bed frame and  side rails snugly to prevent patient entrapment  Follow the manufacturer   s instructions  Monitor  patient frequently  Read and understand the Owner   s Operator   s Manual prior to using this  equipment  Invacare product manuals are available at www
12. R           Pe                      95    PANEL  0    O O  STANDBY                        Alternating Rotation and Pre    Turn Time  Minutes  Center             Left            Power Button Right  Left    e  FUNCTION Down              e  ROTATION    Lock Alarm  Button    Rotation Button  Down Soft    Button LCD Up Firm Button  FIGURE 4 Using the Front Panel    Max Inflate Button    Part No 1148139 19 Invacare    MA90Z Series Mattresses    4 OPERATION  Power Button    l  To turn the control unit on or off  press and release the Power button O       Once the unit is plugged in  an AMBER LED on the control unit is lit indicating that the system is in STAND BY mode  Once  the Power button is pressed and released  a GREEN LED illuminates indicating that the control unit is on     If the power comes on after a power outage  the control unit will go through its system initialization routine for a few  seconds and then resume the desired function     Turn Button  MA90Z Power Unit        Press the Turn button    to select one of three rotation modes or to set the unit to No Turn  Static  mode        The rotation modes are right  left or both and all three modes include alternating pressure therapy     When No Turn is selected  all of the air cushions are maintained at a constant pressure     Turn Time Button  MA90Z Power Unit        Press the Turn Time button  amp g to select one of four rotation times        The rotation times are 10  20  30 and 60 minutes     Turn Angle Button  MA90
13. SORIES WARNINGS    Invacare products are specifically designed and manufactured for use      conjunction with Invacare accessories  Accessories  designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare  products     NOTICE    THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE     Check all parts for shipping damage and test before using  In case of damage  DO NOT use  Contact Invacare Carrier for  further instruction     Contraindications    A WARNING  ALWAYS consult the patient   s physician before using the MA90Z and MA95Z systems     Invacare    MA90Z Series Mattresses 10 Part No 1148139    3 SAFETY    Installation    A WARNING    The MA90Z        MA95Z series  including bariatric  is recommended to be installed on medical bed frames with bed side or  assist rails  It is preferred that the rails be in the raised position whenever a patient is on the bed  Health care professionals  assigned to each case should make the final determination whether side or assist rails are warranted after assessing patient  risks of entrapment and falls in accordance with State patient restraint legislation or facility interpretation of such legislation     Controls on the footboard may be obstructed by the power unit on a few bed frames  It may be necessary to relocate the  power unit  Refer to Installing the Power Unit on page 16        Check that air hoses and power cord are clear of moving bed components bef
14. Z Power Unit     I  Press the Turn Angle button      p    to select one of four rotation angles        The rotation angles are 1 4     gt   3 4 and full rotation  45        Invacare    MA90Z Series Mattresses 20 Part No 1148139    4                     Rotation Button  MA95Z Power Unit     l  Press the Rotation button    to select Static mode      Alternating Rotation mode     When the LED is on  Alternating Rotation mode has been selected     In Static mode  the LED is off and all air cushions are maintained at a constant pressure     Select Button  MA95Z Power Unit     A     Press the Select button 3  move the pointer on the LCD to Function     2  Press the Firm or Soft button until the desired function displays  Available functions are     FUNCTION DESCRIPTION    LEFT Rotation to the Left side  The time indicates the rotation duration        CENTER Rotation to Center  The time indicates the rotation duration        RIGHT Rotation to the Right side  The time indicates the rotation duration        STATIC Static mode  All air cushions maintain constant pressure        PULSE        10 minute pulse time cycle  All of the air cushions in the mattress will remain static at the current pressure settings  for 10 minutes  then the pressure in the mattress will decrease by 20  for five seconds  After five seconds  the mat   tress will once again inflate to the set pressure settings for 10 minutes  then deflate by 20  for five seconds  and the  cycle is repeated        MULTI PULSE The
15. eel locks of the wheelchair before transferring the patient between  the bed and the wheelchair    Engage the wheel locks of the bed  Refer to the owner   s manual provided with the bed    Engage the wheel locks of the wheelchair  if applicable  Refer to the owner   s manual provided with the wheelchair    Press the Max Inflate button to achieve maximum mattress pressure     Raise or lower the bed to match the wheelchair height  Refer to the owner   s manual provided with the bed                  When the mattress has reached maximum firmness  perform one of the following      Bedto Wheelchair Transfer   Slide the patient onto the wheelchair       Wheelchair to Bed Transfer   Slide the patient onto the bed     Invacare    MA90Z Series Mattresses 28 Part No 1148139    4                     4 11 Preparing for CPR Procedure       For this procedure  refer to FIGURE 5 on page 30        Perform one of the following      MA90Z Power Unit   Press and hold the tabs      the hose connector  Detail    B           MA95Z Power Unit   Lift the connector cover  Detail  A       2  Pull the hose connector  Details  B  and  C   from the control unit connectors      MA95Z Only   If the hose connector is difficult to release  pull the hose connector down and then away     3      907 Only   Disconnect the RED CPR connector located      the hose  Detail    D         Part      1148139 29 Invacare    MA90Z Series Mattresses    4 OPERATION       DETAIL    A    DETAIL    B     Connector      907   Co
16. ess        The patient comfort pressure ranges from Soft  6 4 mmHg  to Firm  32 6 mmHg   The Comfort Control LED displays the  patient comfort pressure levels and provides a guide to the caregiver to set approximate comfort pressure level depending  on the patient weight  If the patient   s weight to height ratio is above average  increase the pressure setting by approximately  20   MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from 6 to 32     Invacare    MA90Z Series Mattresses 22 Part No 1148139    4 OPERATION    Automatic Wireless Fowler    Manual Operation of the Fowler             Press the Select button 3 to move the pointer on the LCD to Function     2  Use the Firm and Soft buttons until Fowler displays        When this mode is activated  the control unit increases the pressure in the mattress to prevent the patient from bottoming  out     About Automatic Wireless Mode    When the bed articulates to 45    the transmitter in the head of the mattress signals the control unit to add 25  more air into the mattress to prevent  the patient from bottoming out    Programming the Automatic Wireless Fowler Transmitter   Ensure the transmitter is inside the mattress pocket located on the right side of the mattress at the head end    Ensure the mattress is in a flat position    Ensure the control unit is plugged in  and the LCD displays STAND BY     Perform one of the following                         907 Power Unit   Perform the following steps     i
17. h bed rail  product entry to access this bed safety guide     Refer to the owner   s manuals for beds and rails for additional product and safety information     After any adjustments  repair or service and before use  make sure all attaching hardware is tightened securely  Assist rails  with dimensions different from the original equipment supplied or specified by the bed manufacturer may not be  interchangeable and may result in entrapment or other injury    Invacare    MA90Z Series Mattresses 12 Part No 1148139    Fire Hazard          Electrical    A    3 SAFETY    DANGER  Smoking    DO NOT SMOKE while using this device  This system uses room air for circulation through the mattress  A cigarette can  burn a hole in the bed surface and cause damage to the mattress  Also  patient clothing  bed sheets  etc  may be  combustible and cause a fire  Failure to observe this warning can result in severe fire  property damage and cause physical  injury or death     Smoking by visitors in the room will contaminate the system  Therefore  visitor smoking is NOT permitted   Anesthesia Equipment   There is an explosion risk if used with flammable anesthetics    Oxygen    There is a possible fire hazard when used with oxygen administering equipment other than nasal mask or half bed tent type   The oxygen tent should NOT extend below mattress support level     DANGER    Electrical shock hazard  DO NOT remove cover  Refer to qualified service personnel   Before performing any maintenance to 
18. mode can be manually disengaged by pressing the Max Inflate button  This will deactivate the Max Inflate LED     It is recommended that Max Inflate setting be used during patient ingress egress  patient wound care  patient turning or  patient cleaning       Low Air Loss Mode   In this mode  the mattress goes into on demand low air loss relief mode     Invacare    MA90Z Series Mattresses 24 Part No 1148139    4                     Max Inflate Button  MA95Z Power Unit      o  l  Press the Max Inflate button       to select the max inflate mode and inflate the mattress rapidly to maximum firmness  pressurized to 35   6  mmHg   A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active  After 15 minutes  the Max Inflate mode  deactivates and the control unit defaults to the previous setting     It takes 25 60 seconds for the MA95Z mattress to inflate fully  inflation time depends on size of mattress      It is recommended that Max Inflate setting be used during patient ingress egress  patient wound care  patient turning or  patient cleaning     2  Perform one of the following   e   MA90Z Power Unit   Press the Max Inflate button again to manually disengage Max Inflate mode           957 Power Unit   Select a different setting to exit Max Inflate mode  or press the Max Inflate button again        This will deactivate the Max Inflate LED     Lock Alarm Silence Button       Press the Lock Alarm Silence button  A to select one of the following modes
19. model number  and the date of purchase   indicate nature of the defect and  if the product is serialized  indicate the serial number    Invacare Corporation will issue a return authorization  The defective unit or parts must be returned for warranty inspection using the serial number   when applicable  as identification within thirty days of return authorization date  DO NOT return products to our factory without our prior consent   C O D  shipments will be refused  please prepay shipping charges    LIMITATIONS AND EXCLUSIONS  THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO  ADHERE TO THE ENCLOSED INSTRUCTIONS  IN ADDITION  THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED  PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED  PRODUCTS SUBJECTED TO NEGLIGENCE  ACCIDENT   IMPROPER OPERATION  MAINTENANCE OR STORAGE  OR PRODUCTS MODIFIED WITHOUT INVACARE S EXPRESS WRITTEN CONSENT  INCLUDING  BUT NOT LIMITED TO  MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS  PRODUCTS  DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE  PRODUCTS  DAMAGED BY CIRCUMSTANCES BEYOND INVACARE S CONTROL  PRODUCTS REPAIRED BY ANYONE OTHER THAN AN INVACARE  DEALER  SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE    THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER EXPRESS WARRANTIES WHATSOEVER  WHETHER  EXPRESS OR IMPLIED  INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AN
20. om the bed     2  If necessary  lower the side rails to facilitate installation of the mattress     3   Unroll the powered mattress and place it on the bed frame        Ensure that the hose is towards the foot end of the bed     4  Use the buckle straps to secure the powered mattress to the bed deck in the following locations     Head End   Head End Bed Deck     Center   Center of the Bed Frame       Foot End   Foot End Bed Deck    Part      1148139           Powered  Mattress    Head End    Mattress cover not  shown for clarity     Buckle Straps    Foot End  Hose    FIGURE   Installing the Mattress Replacement System    15 Invacare    MA90Z Series Mattresses    4                     4 2 Installing the Side Rails  A WARNING    Patient entrapment with bed side rails may cause injury or death  Mattress MUST fit bed frame and side rails snugly to  prevent patient entrapment  Follow the manufacturer   s instructions  Monitor patient frequently  Read and understand the  Owner s Operator s Manual prior to using this equipment  Invacare product manuals are available at www invacare com or  your dealer        Refer to the instructions provided with the side rails for the installation procedure     4 3 Installing the Power Unit    Bed Hook     Pull out the bed hook on the back of the control unit     2  Place the control unit on the footboard            Y 3       If the bed does not have a footboard  place the  control unit on a flat surface  leaving room for the  hose to hang down   
21. ore placing a patient on the bed  Operate all  bed frame motorized functions through their full range of motion to be certain that there is no pulling  interference or  pinching    DO NOT strap the mattress to the bed frame at the head and foot ends  Secure ALL mattress straps  Secure the straps to  the bed deck at the head and foot ends and to the frame at the center of the bed  Otherwise damage to the mattress will  occur when the head and foot ends are raised  Refer to Installing the Mattress Replacement System on page 14        Part No 1148139    Invacare    MA90Z Series Mattresses    3 SAFETY  Entrapment May Occur  A WARNING    Patient entrapment with bed side rails may cause injury or death  Mattress MUST fit bed frame and side rails snugly to  prevent patient entrapment  Follow the manufacturer   s instructions  Monitor patient frequently  Read and understand the  Owner   s Operator   s Manual prior to using this equipment  Invacare product manuals are available at www invacare com or  your dealer     Proper patient assessment and monitoring  and proper maintenance and use of equipment is required to reduce the risk of  entrapment  Variations in bed rail dimensions  and mattress thickness  size or density could increase the risk of entrapment   Visit the FDA website at http   www fda gov to learn about the risks of entrapment  Review    A Guide to Bed Safety      published by the Hospital Bed Safety Workgroup  located at www invacare com  Use the link located under eac
22. patient   s body   Ensure that there is 3 to 4 finger width clearance between the bottom of the patient and the bed frame     Adjust the comfort setting  if needed     mon    gt     Wait five minutes for the mattress pressure to stabilize           Repeat STEPS A E until patient has not bottomed out and patient comfort is achieved   If the patient feels that the bed is too soft hard  press the Soft or Firm button to adjust the comfort settings     Use a regular pillow to help support and stabilize the patient s head     Part No 1148139 27 Invacare    MA90Z Series Mattresses    4                     4 9 Transferring Patient From To    Gurney  A WARNING    ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney before transferring the patient between the  bed and the gurney    Engage the wheel locks of the bed  Refer to the owner s manual provided with the bed    Engage the wheel locks of the gurney  Refer to the owner s manual provided with the gurney    Press the Max Inflate button to achieve maximum mattress pressure     Raise or lower the bed to match the gurney height  Refer to the owner s manual provided with the bed          che ONS          When the mattress has reached maximum firmness  perform one of the following     Bed to Gurney Transfer   Slide the patient onto the gurney     Gurney to Bed Transfer   Slide the patient onto the bed     4 10 Transferring Patient From To a Wheelchair  A WARNING    ALWAYS engage the wheel locks of the bed and the wh
23. r Cord  A WARNING    DO NOT alter plug to fit a non conforming outlet  Instead  have an electrician install a properly grounded outlet  Failure to  use the correct plug and outlet can result in a potential safety hazard     CAUTION    Ensure that the power cord of the control unit is not pinched  or has any objects placed on it  and also ensure it is not  located where it can be stepped on or tripped over        Examine the hospital grade power cord supplied with the control unit   2  Perform one of the following     Ifthe plug is damaged   Call your supplier for a replacement hospital grade cord     Ifthe plug is not damaged   Plug the end of the supplied hospital grade power cord into the power outlet on the side of the control unit     3  Plug the other end of the plug into a properly grounded outlet on the wall        Once the unit is plugged in  an AMBER LED on the control unit is lit indicating that the system is in STAND BY mode     Invacare    MA90Z Series Mattresses 18 Part      1148139    4 OPERATION  4 6 Using the Front Panel  DETAIL    A       MA90Z POWER    UNIT FRONT  PANEL    Turn Angle Button Power Fail LED Low Pressure LED              microA    90    POWER    Low Pressure  Power Fail                Both 20  Minutes       Power  Button    e  Full Turn  TURN ANGLE    Turn Button Lock Alarm    ode  Max Inflate  Button  Firm Button Low Air Loss  Button        Turn Time Button Soft Button    DETAIL            Select Button Hold Button Power Fail LED    MA95Z POWE
24. ree components       Part No 1148139    Mechanical Parameters    2 OVERVIEW       CONTROL UNIT DIMENSIONS  L X W X H      I5  x 65  x II          CONTROL UNIT WEIGHT     15 Ibs       CONTROL UNIT POWER CORD     10   14 Feet Long  Hospital Grade       CONTROL UNIT CONNECTION     Triple Flow   Two  gt     and    Quad Flow   Four 4    One  4    Single Magnetic  Single Quick Coupling Quick Coupling       CONTROL UNIT PACKAGING       piece box       AIR FILTER     Environmental Parameters       Charcoal Air Filter with  None Fire Retardant       MA90Z MA95Z    OPERATING CONDITIONS  AMBIENT TEMPERATURE   RELATIVE HUMIDITY   ATMOSPHERIC PRESSURE     50     95   F  30    75  Non Condensing  70   106 kPa       STORAGE AND SHIPPING CONDITIONS  AMBIENT TEMPERATURE    RELATIVE HUMIDITY    ATMOSPHERIC PRESSURE     Part      1148139        40     158   F  10    100   50   106 kPa       Invacare    MA90Z Series Mattresses    3 SAFETY  3 Safety    3 1 General Guidelines  A WARNING    DO NOT use this product or any available optional equipment without first completely reading and understanding these  instructions and any additional instructional material such as owner s manuals  service manuals or instruction sheets  supplied with this product or optional equipment  If you are unable to understand the warnings  cautions or instructions   contact a healthcare professional  dealer or technical personnel before attempting to use this equipment   otherwise  injury  or damage may occur     A ACCES
25. se d 3l  Changing the EID RONDE E DD DD Sl 33  Een i    M                                                           M  35    Part No 1148139 3 Invacare    MA90Z Series Mattresses      GENERAL      General         Warnings    Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage   See the information below for definitions of the signal words     A DANGER  Danger indicates an imminently hazardous situation which  if not avoided  will result in death or serious injury     A WARNING    Warning indicates a potentially hazardous situation which  if not avoided  could result in death or serious injury     A CAUTION    Caution indicates a potentially hazardous situation which  if not avoided  may result in property damage or minor injury or    both     IMPORTANT  Indicates a hazardous situation that could result in damage to property if it is not avoided      Gives useful tips  recommendations and information for efficient  trouble free use     Invacare    MA90Z Series Mattresses 4 Part      1148139      GENERAL    1 2 Limited Warranty    PLEASE NOTE  THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS  MANUFACTURED AFTER JULY 4  1975    This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare or a dealer  This warranty is  not extended to any other person or entity and is not transferable or assignable to any 
26. ss and hold the Soft and Firm Buttons at the same time     3         LCD screen displays the Run Time as described above   47 Powering Up the System     For this procedure  refer to FIGURE 4 on page 19        Turn on the power to the system by pressing the Power button on the control unit        Once the button is released  a green led illuminates when the unit is on     Invacare    MA90Z Series Mattresses 26    Part      1148139    4 8       x    t    9   10     4                     Placing the Patient on the Mattress    Press the Mode button or Max Inflate button to turn on the control unit to maximum flow      In this mode  the Max Inflate Led lights up     After the mattress is fully inflated  place the patient on the bed     Ensure that the patient   s feet are towards the foot end of the mattress  the end with the hose      Center the patient on the bed from side to side and head to foot      Special positioning may be required with contracted patients to provide comfortable positions     After placing the patient  make certain no objects will fall under the patient  such as feeding tubes  IV s etc   Wait five minutes for the mattress pressure to stabilize    Set the comfort pressure to the desired comfort level    Verify that the patient has not bottomed out by performing the following steps    Ensure that the patient is lying flat on his her back in the middle of the mattress    Place four fingers between the air cushions directly underneath the sacral region of the 
27. subsequent purchaser or owner  Coverage under this warranty  will end upon any such subsequent sale or other transfer of title to any other person    This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state    Invacare warrants the mattress and cover when purchased new and unused to be free from defects in materials and workmanship for a period of one  year from the date of purchase from Invacare or a dealer  with a copy of the seller s invoice required for coverage under this warranty  Invacare  warrants the control unit when purchased new and unused to be free from defects in materials and workmanship for a period of two years from the date  of purchase from Invacare or a dealer  with a copy of the seller s invoice required for coverage under this warranty If within such warranty period any  such product shall be proven to be defective  such product shall be repaired or replaced  at Invacare s option  This warranty does not include any labor  or shipping charges incurred in replacement part installation or repair of any such product  Invacare s sole obligation and your exclusive remedy under  this warranty shall be limited to such repair and or replacement    For warranty service  please contact the dealer from whom you purchased your Invacare product  In the event you do not receive satisfactory warranty  service  please write directly to Invacare at the address on the back cover  Provide dealer s name  address  
28. the power unit  disconnect the power cord from the wall outlet     DO NOT insert items into any openings of the control unit  Doing so may cause fire or electric shock by shorting the  internal components    The control unit must be kept away from all heat sources and radiators during operation     Connect the equipment to properly grounded three prong wall outlet using 10 14 ft hospital grade power cord provided  with the product     Grounding reliability depends upon a properly grounded receptacle  3 prong      Part No 1148139 13 Invacare    MA90Z Series Mattresses    4 OPERATION  4 Operation    A CAUTION    The control unit and mattress      the MA90Z series are designed to be used as a system  DO NOT replace mattresses or  control units with other models or other brands  Otherwise  damage to the system may occur  Contact your supplier to  get the correct replacement if needed     41 Installing the Mattress Replacement System    CAUTION    DO NOT strap the mattress to the bed frame at the head and foot ends  Secure ALL mattress straps  Secure the straps to  the bed deck at the head and foot ends and to the frame at the center of the bed  Otherwise damage to the mattress will  occur when the head and foot ends are raised           Head End Bed Deck       Foot End Bed Deck                Invacare    MA90Z Series Mattresses 14               1148139    4                          The powered mattress comes with ten nylon buckle straps        Remove the original foam mattress fr
29. ver MA90ZB42  Tab and  MA90ZB48  Hose   Connector  Control Unit Control Unit  Connectors          DETAIL    D             MA95Z  Mattress  MA95ZB42 and  MA95ZB48 Hose    Connector           DETAIL             RED CPR Connector   MA90Z  MA90ZB42 and  MA90ZB48 Only     FIGURE 5 Preparing for CPR Procedure    Invacare    MA90Z Series Mattresses 30 Part No 1148139    5 MAINTENANCE AND TROUBLESHOOTING    5 Maintenance and Troubleshooting    5 1 Cleaning the System  A WARNING    Before cleaning or disassembling the      90 series  check the underside of the mattress folds for sharp objects such as  scissors  needles  etc  These objects should be removed and discarded before proceeding with further cleaning or  disassembly  otherwise injury or damage to the product may occur    Because of potential risk of infectious exposure  cleaning with the patient on the bed is not recommended         equipment should be inspected  Any item that is visibly soiled with the patient s blood or other body fluids should be  properly cleaned or removed     Laundry workers should treat all soiled bedding as if it were contaminated with pathogenic microorganisms     Remove the bedding    If necessary  inflate the mattress    Ensure that the control unit is off    Unplug the power cord from the wall outlet    Ensure that the underside of the mattress is clear of all sharp objects     Examine the surface of the control unit and mattress assembly components for visible blood or body fluids     SO iin br E
30. with a quaternary disinfectant solution   B  Using a nylon brush  gently clean all crevices as they can harbor microorganisms     C  Air dry all treated surfaces     Invacare    MA90Z Series Mattresses 32 Part No 1148139    5 MAINTENANCE AND TROUBLESHOOTING    10  Perform the following steps to clean the mattress components  air cushions  mattress base  etc       A  Using a clean sponge or paper towel  wipe down the mattress components with a dilute detergent solution of quaternary cleaner disinfectant  or other germicidal detergent solution     B  Wipe the mattress components with a clean dry cloth     Storing the System   l  Ensure that the control unit is off and disconnect the power cord from the wall outlet    Clean the system  Refer to Cleaning the System on page 3l    Disconnect the air hose connector from the control unit and allow air to vent from the mattress   Gently roll up the mattress with minimal handling and agitation     Ensure the cover surface is inside the roll     DY  Ur duco m    Store the unit  keeping the mattress with the control unit     5 2 Changing the Filter     For this procedure  refer to FIGURE    on page 34     Clean the filter every 5 months or whenever it is dirty     l  Remove the two screws securing the filter cover to the back of the control unit   Remove the filter    Wash the filter with water and a mild detergent    Dry the filter completely     Install che filter into the back of the control unit     Qr                      Secure the
    