1. use the KSO salt as the high end reference Insert the probe into a suitable hole of the NaCl or K2S04 salt chamber The shorter the time the hole stays open before inserting the probe the shorter the stabilization time required Wait until the humidity reading stabilizes this takes about 10 30 minutes Note that in high humidities the risk for errors increases Therefore the stabilization time should be longer approximately 20 40 minutes Read the salt chamber temperature from the thermometer and then read the closest humidity value from the calibration table Greenspan s calibration table NaCl or K2SO solution Adjust the wet end WET gain to correspond to the value given in the calibration table 10 User s Guide 3 3 3 4 Greenspan s calibration table 1 MgC K2504 fo uf 33 7 40 3 75 5 0 3 98 8 1 1 33 6 0 3 75 7 0 3 98 5 0 9 M210185en B Lt 33 5 0 2 75 7 0 2 98 2 0 8 33 3 0 2 75 6 0 2 97 9 0 6 11 3 0 3 33 1 0 2 75 5 0 1 97 6 0 5 11 3 0 3 32 8 0 2 75 3 40 1 97 3 0 5 x 40 11 3 0 2 11 3 0 2 11 2 0 2 32 4 0 1 32 1 0 1 31 6 0 1 75 1 0 1 97 0 0 4 74 9 0 1 96 7 0 4 74 7 0 1 96 4 0 4 11 2 0 2 31 1 0 1 745 0 2 96 1 04 11 1 40 2 30 5 40 1 744 0 2 95 8 0 5 If the LiCl solution is used or stored in temperatures below 18 C its equilibrium humidity changes permanently The numbers on the right in each column i
2. C between the air in the chamber and the sensor causes an error of 3 RH at 50 RH and an error of 6 RH at 97 RH The more the transportation or process temperatures differ from the temperature at the calibration site the longer the stabilization time required In laboratory use the calibrator should be stored in that part of the room where the temperature is most stable The calibrator must be kept out of direct sunlight and away from localized heat sources such as spot lights heaters and soldering irons If the probe transmitter is checked against several humidity references the checking must first be made at the dry end More detailed instructions are given in individual manuals of transmitters and probes Handle the probe as little as possible Do not hold the salt chamber or other parts of the calibrator in hand during calibration as they warm up and cause errors in the readings NOTE even the smallest water drop on the probe near the sensors distorts the readings Make sure that chamber covers and plugs are carefully closed s HMP233 1o thermometer probe mo 4 ima o0 l GY i AS 4 VA WY O rings Wis sensor A e QT T NY U saturated sal _ SI zz solution maxX ny L KAS 1 crit WN undissolved i AANANANANASANSASNANAN salt Figure 3 1 Salt chamber User s Guide M210185en B 3 1 1 Th
3. downwards User s Guide CAUTION M210185en B Take off the grid or filter protecting the sensor Take care not to damage the sensor Some 12 mm probes need a adapter fitting to fit into the 13 5 mm hole Adapter fitting prevents sensor from soaking in the salt liquid In case you need an adapter fitting take off the grid or filter and replace it with the adapter fitting Two adapter fittings are attached on the base plate of the HMK1S5 see picture on page 1 Remember to check from the user s guide of your device if the adapter fitting is needed If sensor accidently soaks in salt liquid remove it quickly and rinse with clean water Let it dry before taking into use again Insert the probe into a suitable hole of the LiCl salt chamber Press it downwards until it passes the O rings The shorter the time the hole stays open before inserting the probe the shorter the stabilization time required Wait until the humidity reading stabilizes this takes about 10 30 minutes Read the salt chamber temperature from the thermometer and then read the closest humidity value from the calibration table Greenspan s calibration table LiCl solution Adjust the dry end DRY offset to correspond to the value given in the calibration table Insert the thermometer into the 13 5 mm hole of the NaCl salt chamber NOTE when calibrating probes that are being used for long term measurements over 1 hour in high humidities 90 100 RH
4. in Greenspan s calibration table see Chapter 3 3 at the calibration temperature e g the accuracy of LiCl salt at 20 C is 1 0 0 3 RH 1 3 RH Response time probe and calibrator at the same temperature with Vaisala sensors typically 10 min deviation of the final value lt 1 RH Accuracy of the thermometer with mercury 0 3 C with red capillary liquid 1 C Dimensions 20x9x 23cm Materials metal parts anodised aluminium Weight 1 kg without salt solutions 16 User s Guide M210185en B Warranty Vaisala hereby represents and warrants all Products manufactured by Vaisala and sold hereunder to be free from defects in workmanship or material during a period of twelve 12 months from the date of delivery save for products for which a special warranty is given If any Product proves however to be defective in workmanship or material within the period herein provided Vaisala undertakes to the exclusion of any other remedy to repair or at its own option replace the defective Product or part thereof free of charge and otherwise on the same conditions as for the original Product or part without extension to original warranty time Defective parts replaced in accordance with this clause shall be placed at the disposal of Vaisala Vaisala also warrants the quality of all repair and service works performed by its employees to products sold by it In case the repair or service works should appear inadequate or faulty and shoul
5. 0 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 Time min Figure 3 5 An example of RH and T stabilization when the probe is removed from a process 12 User s Guide M210185en B Humidity RH Temperature C 60 0 0 10 20 30 40 50 60 70 80 90 100 110 120 Time min Figure 3 6 An example of the RH and T stabilization when the transportation and calibration site temperatures are different User s Guide M210185en B Humidity RH In the following example the transportation temperature of the calibrator without the LiCl is 5 C the calibrator is then transferred into room temperature After this an HMP233 probe that has been stored at room temperature is inserted into a suitable hole of the NaCl salt chamber The figure shows that after 40 minutes the reading differs 1 4 RH from the final reading Temperature C 60 0 0 10 20 30 40 50 60 70 80 90 100 110 120 Time min Figure 3 7 An example of the RH and T stabilization when the transportation and calibration site temperatures are different 14 User s Guide M210185en B 4 MAINTENANCE Depending on the frequency of use and the general operating conditions the salt solutions maintain their characteristics for 6 12 months After that they must be replaced A visual check should be performed at intervals o
6. Humidity Calibrator HMK15 USER S GUIDE M210185en B September 2002 its VAISALA M210185en A User s Guide M210185en A User s Guide PUBLISHED BY Vaisala OyjPhone int 358 9 894 91 P O Box 26 Fax 358 9 8949 2227 FIN 00421 Helsinki Finland Visit our Internet pages at http www vaisala com Vaisala 2002 No part of this manual may be reproduced in any form or by any means electronic or mechanical including photocopying nor may its contents be communicated to a third party without prior written permission of the copyright holder The contents are subject to change without prior notice M210185en A User s Guide Contents WARRANTY M210185en A User s Guide This page intentionally left blank PRODUCT DESCRIPTION General The HMK15 humidity calibrator has been developed for calibration and checking of humidity probes and transmitters The functioning of the calibrator is based on the fact that certain salt solutions generate a certain relative humidity in the air above them The four holes in the chamber cover are designed for Vaisala probes and transmitters with 12 mm 13 5 mm 2 holes and 18 5 mm diameters The salt solutions suitable for the HMK15 calibrator are e g lithium chloride LiCl 11 RH magnesium chloride MgCl 33 RH sodium chloride NaCl 75 RH and potassium sulphate K SO 97 RH In calibration the sensor head is inserted into a salt chamber containing a saturate
7. ages are taken from a batch and salts are prepared according to the instructions given in this manual These salt solutions are then checked at Vaisala Oy s Measurement Standards Laboratory this calibration laboratory is accredited by FINAS a member of the European Cooperation for Accreditation of Laboratories The calibration certificate certifies that the equilibrium humidities generated by these salt solutions correspond to Greenspan s calibration table within the specified accuracy see Chapter l6 PREPARING THE SALT SOLUTIONS General instructions e Prepare the salts using the equipment provided with the calibrator or make sure that all accessories used are suitable for preparing salt solutions and are absolutely clean If necessary wash them carefully and rinse several times before preparing the salt solutions The last rinse must be made with distilled or ion exchanged water e The salts are quick and easy to prepare with Vaisala s ready dosed salt packages If you do not use ready dosed salt packages measure the salts using the measurement cup provided Make sure that the cup is clean before measuring the salts In the following instructions the required amounts are given in grammes and millilitres The salt solutions must be prepared from salts of pro analysis grade these are available in agencies selling industrial chemicals Note that the water must be distilled or ion exchanged conductivity lt 0 25 uS cm Ion exchanged wat
8. d salt solution The probe transmitter reading is then adjusted to the correct value i e to the humidity that the specific salt solution generates at that particular temperature Calibration is usually performed at least at two different humidities to ensure the sensor accuracy over the entire humidity range 0 100 RH The HMK15 calibrator is suitable for both laboratory and field usage During transportation the chambers can be tightly closed with specially designed transit covers The optional transit bag order code HM27032 helps to transport the calibrator in the vertical position Furthermore the bag can be used for housing the calibrator also during calibration The following photo shows the HMK15 calibrator with some accessories marked with lon exchanged water Measurement cup Mixing spoon Calibration certificate for 1 thermometer Salt chamber Thermometer Adapter fitting Plugs Ready dosed salt package Base plate with calibration certificate Figure 1 1 The HMK15 calibrator with some accessories User s Guide M210185en B 1 2 2 1 Accessories include additional salt chambers ion exchanged water transit bag and ready dosed salt packages LiCl 11 RH MgCl 33 RH NaCl 75 RH K2S04 97 RH Calibration certificate of the salt packages Vaisala s ready dosed salt packages are delivered with calibration certificates given on the basis of a batch check In other words a sample of pack
9. d this cause malfunction or nonfunction of the product to which the service was performed Vaisala shall at its free option either repair or have repaired or replace the product in question The working hours used by employees of Vaisala for such repair or replacement shall be free of charge to the client This service warranty shall be valid for a period of six 6 months from the date the service measures were completed This warranty is however subject to following conditions a A substantiated written claim as to any alleged defects shall have been received by Vaisala within thirty 30 days after the defect or fault became known or occurred and b The allegedly defective Product or part shall should Vaisala so require be sent to the works of Vaisala or to such other place as Vaisala may indicate in writing freight and insurance prepaid and properly packed and labelled unless Vaisala agrees to inspect and repair the Product or replace it on site This warranty does not however apply when the defect has been caused through a normal wear and tear or accident b misuse or other unsuitable or unauthorized use of the Product or negligence or error in storing maintaining or in handling the Product or any equipment thereof c wrong installation or assembly or failure to service the Product or otherwise follow Vaisala s service instructions including any repairs or installation or assembly or service made by unauthorized personnel not appr
10. er can also be ordered as an accessory for the HMK15 calibrator order code 19767HM e Handle the salts and equipment carefully and keep them absolutely clean so that the salts are not contaminated Make sure that the salts do not mix e At the bottom of the salt chamber there must not be more than 1 cm of undissolved salt and liquid Otherwise the sensor to be calibrated can be immersed into the solution If necessary verify the amount e g with the measurement spoon e Various standards ASTM E104 85 DIN 50008 JIS Z8806 include recommendations and instructions on how to prepare and store salt solutions User s Guide M210185en B 2 2 Preparing the solutions Lithium chloride generates a reference humidity of approximately 11 RH It is normally used as the dry end offset reference NOTE Never add water to dry LiCl salt the salt may heat up so rapidly that it splashes out of the chamber NOTE LiCl is harmful when swallowed the solution is also corrosive NOTE If the LiCl solution is used or stored in temperatures below 18 C its equilibrium humidity changes permanently Magnesium chloride generates a reference humidity of approximately 33 RH It is often used as a check point if calibration is performed at more than two points Sodium chloride generates a reference humidity of approximately 75 RH It is used as the wet end reference gain for probes measuring in applications with normal humidities Po
11. ermometer The calibrator includes a thermometer which can contain either mercury or red capillary liquid If the thermometer contains mercury it is delivered with a calibration certificate The accuracy of the mercury thermometer is sufficient for checking the transmitter s temperature channel if desired The mercury thermometer has been calibrated at five points and the result is traceable to the Measurement Standards Laboratory of the Centre for Metrology and Accreditation in Finland For more accurate results use the corrections given in the calibration certificate or their interpolated values Consider also the parallax error which may occur while reading the results The mercury thermometer should be recalibrated within three years in order to maintain the traceability The manufacturer of the thermometer with red capillary liquid provides it with a factory calibration certificate As the accuracy and stability of this kind of a thermometer are inferior to those of a mercury thermometer it is not recommended to be used for checking the temperature channel of the transmitter The sleeve at the end of the thermometer s protective tube can be used in two different positions The thermometer is delivered with the sleeve in such a position that it protects the part containing the capillary liquid When turned the other way round it acts as an adapter when the thermometer is inserted into the 13 5 mm hole see Figure 3 2 Figure 3 2 The
12. f 2 3 months There must be a minimum of approximately 10 of undissolved salt at the bottom of the chamber max 90 and the salt must be clean otherwise it must be reprepared Please note that the LiCl solution may crystallize on the surface However there may still be solution under the surface Therefore in such cases stir the solution and check it again the next day It is advisable to keep a maintenance log which shows the dates of salt replacements and other maintenance procedures For correct calibration it is essential that salt chambers are tightly closed Check the O rings at each salt replacement If they are damaged replace them with new ones The O rings used are of the following types O ring location dimensions _ type measurement hole 12 mm 12x 2 5 mm NBR70 measurement hole 13 5 mm 13 5x 2 5mm NBR70 measurement hole 18 5 mm 18 3 x 2 4mm NBR70 41 2x3 0mm NBR70 50 0 x 2 0mm___ NBR70 5 SPARE PARTS AND ACCESSORIES 15 User s Guide M210185en B TECHNICAL DATA Operating temperature range 0 50 C NOTE If the LiCl solution is used or stored in temperatures below 18 C its equilibrium humidity changes permanently Accuracy lithium chloride LiCl 1 0 RH Greenspan s uncertainty magnesium chloride MgCl 1 0 RH Greenspan s uncertainty natrium chloride NaCl 1 4 RH Greenspan s uncertainty potassium sulphate K2SO4 1 5 RH Greenspan s uncertainty The uncertainty given
13. he chamber with the chamber cover Figure 2 4 Figure 2 4 Close the chamber with the chamber cover User s Guide M210185en B 6 Fasten the salt chamber on the holder in the base plate and close the measurement holes with rubber plugs The chambers can also be used as individual checkers without the base plate NOTE the plugs have three steps each of which is suitable for a certain hole diameter the first step for the 12 mm the second step for the 13 5 mm and the third step for the 18 5 mm hole Keep the holes always closed when not calibrating 7 Write the preparation date on a sticker and mark the chamber with it If you use a ready dosed salt package use the sticker with the batch code Mark all parts of the salt chamber chamber chamber cover and transit cover with stickers This way various covers and different salts do not get mixed Figure 2 5 Mark the chamber and covers with stickers 8 Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity 9 See Chapter Fon how to maintain the salt solutions NOTE If you do not use the instrument for a longer period of time close the chambers with transit covers User s Guide 3 3 1 M210185en B INSTRUCTIONS FOR THE CALIBRATION OF HUMIDITY INSTRUMENTS General instructions Usually the errors during humidity calibration are due to temperature differences A temperature difference of 1 C at 20
14. ndicate the uncertainty of the reference humidity of the salt in that temperature Greenspan L Journal of Research of the National Bureau of Standards A Physis and Chemistry Vol 81A No 1 January February 1977 pp 89 95 On site calibrations and transportation The HMK15 calibrator is easily trasferred from one place to another Two O rings seal the transit cover on the chamber The optional transit bag helps to transport the calibrator so that chambers stay in the vertical position Furthermore the bag can be used for housing the calibrator during calibration When transferring the calibrator from one place to another follow the instructions below e Turn the protective sleeve on the thermometer and place the thermometer in the holder e Replace chamber covers with transit covers Press chamber covers on vacant chamber holders for transportation e During transportation keep the calibrator so that the chambers stay as upright as possible This way very little salt solution gets on transit covers and cleaning is easier The closer the transportation temperature is to the temperature of the the shorter the time required for temperature stabilization If the transportation temperature is below 18 C the calibration site 11 User s Guide M210185en B LiCl salt chamber should be transported separately to keep the solution warm e When the calibrator is at the calibration site remove transit covers and fa
15. oved by Vaisala or replacements with parts not manufactured or supplied by Vaisala d wm modifications or changes of the Product as well as any adding to it without Vaisala s prior authorization e other factors depending on the Customer or a third party Notwithstanding the aforesaid Vaisala s liability under this clause shall not apply to any defects arising out of materials designs or instructions provided by the Customer This warranty is expressly in lieu of and excludes all other conditions warranties and liabilities express or implied whether under law statute or otherwise including without limitation any implied warranties of merchantability or of fitness for a particular purpose and all other obligations and liabilities of Vaisala or its representatives with respect to any defect or deficiency applicable to or resulting directly or indirectly from the Products supplied hereunder which obligations and liabilities are hereby expressly cancelled and waived Vaisala s liability shall under no circumstances exceed the invoice price of any Product for which a warranty claim is made nor shall Vaisala in any circumstances be liable for lost profits or other consequential loss whether direct or indirect or for special damages 17
16. sleeve protects the part with capillary liquid on the left or acts as an adapter on the right During calibration the thermometer is inserted into the 13 5 mm hole of a salt chamber Press it downwards until it passes the O rings The thermometer is correctly in place when you can feel a resistance while pressing it downwards User s Guide M210185en B Figure 3 3 During calibration the thermometer is inserted into the 13 5 mm hole When the thermometer is not in use or the calibrator is transferred from one place to another place the thermometer in holders Figure 3 4 Figure 3 4 Place the thermometer in holders when not in use 3 2 Calibration Perform the calibration according to the following instructions e Leave the HMK15 calibrator and the probe at the calibration site for at least 30 minutes before starting the calibration in order to let the probe temperature stabilize to the room temperature e With lithium chloride it is not necessary to use the thermometer as the humidity reading changes only very slightly in the temperature range of 25 30 C However if you use it make sure that the sleeve is in the correct position and insert it into the 13 5 mm hole of the LiCl salt chamber NOTE handle the thermometer as little as possible and do not handle it by the measuring end Press the thermometer downwards until it passes the O rings The thermometer is correctly in place when you feel a resistance while pressing it
17. sten chamber covers on salt chambers Close the holes with plugs NOTE the plugs have three steps each of which is suitable for a certain hole diameter the first step for the 12 mm the second step for the 13 5 mm and the third step for the 18 5 mm hole e Clean the transit covers with a damp cloth and press them on vacant chamber holders e Take the thermometer off the holder turn the sleeve the other way round and perform the calibration according to Chapter In on site calibrations it is important to allow enough time for the calibrator and probe temperatures to stabilize Depending on the temperature differences between the transportation and the calibration site or between the probe removed from the process and the calibration site a two point calibration takes about 0 5 2 hours If there are several frequently calibrated instruments at the same site knowing the stabilization times is useful The following figure illustrates an example of temperature and humidity stabilization In this example an HMP233 probe has been transferred from a temperature of 75 C oven to a suitable hole of the NaCl salt chamber the calibrator is at room temperature The figure shows that after 40 minutes the humidity reading differs 0 2 RH from the final reading 78 90 77 80 76 70 5 75 60 2 D 74 50 E 5 5 E 73 40 I 72 30 71 20 70 f f 1
18. tassium sulphate generates a reference humidity of approximately 97 RH Potassium sulphate is used as the wet end reference gain for probes measuring in applications with very high humidities 90 100 RH e g outdoor and concrete measurements Prepare the salt solutions according to the following instructions Measure the salts carefully if you do not use the ready dosed packages User s Guide M210185en B 1 Take the calibrator out of the box Open the transit cover of the chamber Remove the measurement cover from the chamber holder and press the transit cover on the holder see Figure 2 1 Press the transit cover on a chamber holder 2 Pour ion exchanged water into the chamber the required amounts are given in the table Figure 2 2 Pour a correct dose of ion exchanged water into the chamber dry the measurement cup 2 Sprinkle the contents of a salt package or measure the salt according to the table in small quantities into the chamber stirring constantly When measuring with the measurement cup make sure that the cup is clean and dry Rinse and dry the cup after every use User s Guide M210185en B 15 gor 18 ml 30 g or 30 ml 20 g or 15 ml 30 g or 20 ml Figure 2 3 Sprinkle a correct dose of salt into the chamber stirring constantly 4 When all salt has been sprinkled into the chamber the saturated salt solution should have the ratio of 60 90 undissolved salt to 10 40 liquid 5 Close t