User manual and technical description Altura Care

image

Contents

1. S NOTE If there is a substantial temperature difference between the bed due to transport storage and the installation site allow 24 hours for the difference to balance itself before connecting the bed to the mains 7 2 Battery Operation The battery supplied with the bed is delivered uncharged The battery serves as a backup during power failures or while transporting the patient NOTE Batteries will remain in fully functional condition only for a certain period of time which is dictated by the laws of physics and chemistry used in this type of Dry Lead Acid battery technology and their frequency and method of use 9 The user is obliged to monitor battery functionality and to change the batteries in accordance with the user amp service manuals Batteries must be checked according to the user manual at least once per month NOTE Manufacturer provides 6 months warranty for full function of batteries The manufacturer takes no responsibility for any damage to the bed or battery caused by m Not following manufacturers user manual instructions m Fitting batteries which are not approved by Linet m Batteries fitted by an unqualified service organisation To charge the battery Connect the bed to the mains NOTE Some bed adjustments cannot be carried out without a battery for example height adjustment under a load of above 200 kg Product Altura 1
2. Fig 9 CPRemergency backrest release Product Altura 25 48 Version 01 Winer 8 6 Siderails Full length or single collapsible siderails are components of the bed The nursing personnel are responsible for the siderails being folded up while the patient is in bed Fig 10 Mechanism Single Collapsible Siderail To fold siderails up down Grip the siderail by the upper tube Press both locking buttons to unlock Push siderail towards foot end Fold siderail up or down Push siderail towards head end 4 gt 4 Siderail latches automatically with a click Fig 11 Mechanism Full Length Siderail To fold siderails up down Grip the siderail by the upper tube in marked place 1 and lift slightly Press down locking pin 2 to unlock Fold siderail up or down When folded up siderail locks automatically with a click Product Altura 26 48 Version 01 Winer 8 7 Castor Control and Bed Transport Castor control for bed with control levers Fig 12 Positions of Castor Control Lever The control levers are located on both sides of the foot end Castor control lever positions 1 Forward Movement The front left castor is locked The bed moves straight ahead If the bed is equipped with a fifth castor this castor determines the direction of movement 2 Unrestricted Movement All of the castors are unlocked 3 Braked All of the castors are braked Produ
3. 14 21 20 19 18 Fig5 Supervisor Panel Central STOP Button Activating GO Button Thighrest Adjustment Button Thighrest Lock Button Thighrest Lock LED Backrest Adjustment Button Backrest Lock Button Backrest Lock LED Height Adjustment Buttons 10 Height Lock Button 11 Height Lock LED 12 Auto Contour Adjustment Buttons 13 Foot Control Lock Button 14 Foot Control Lock LED 15 Tilt Adjustment Buttons optional 16 Extra Low Position Button bed height below 43 cm 17 Cardiac Chair Position Button 18 CPR Resuscitation Position Button 19 Trendelenburg Position Button optional 20 Battery Charge Status LED 21 Mains Power LED Se Xo Ee Nx Product Altura Version 01 15 17 16 20 48 WLiner 8 1 1 Central STOP Button The central STOP button 1 immediately interrupts all bed movements in case of unauthorized bed positioning or an electronic failure When the central STOP button 1 is pressed the electronic stops all bed movements immediately 8 1 2 Activating GO Button The GO button 2 activates the keypad or the touchscreens of all control elements The GO button 2 is included on a number of different control elements The function of the GO button 2 is identical on all control elements After pressing the GO button 2 the keypad or touchscreen will remain active for 3 minutes During this time the following is possible m Adjusting individual mattress platform elements by pressing the
4. Backrest and calf rest handles E CPR release handle m Head and foot boards Siderails in highest position m Freely accessible mattress surface Accessory rails 9 2 2 Cleaning before Changing Patients Clean the following bed parts m All of the control elements for adjusting the bed m All handles Backrest and calf rest handles E CPR release handle Head and foot boards Siderails in highest position Freely accessible mattress surface Accessory rails All plastic mattress platform covers Plastic undercarriage covers Telescopic columns Mattress on all sides Freely accessible metal parts of mattress platform Cable ducts Lifting pole bushing Infusion stand bushing Wall deflection bumpers Castors Brakes Product Altura Version 01 32 48 Winer 9 2 3 Complete Cleaning and Disinfection Clean the following bed parts All of the control elements for adjusting the bed All handles Backrest and calf rest handles E CPR release handle Head and foot boards Siderails in highest position Freely accessible mattress surface Accessory rails All plastic mattress platform covers Plastic undercarriage covers Telescopic columns Mattress on all sides Freely accessible metal parts of mattress platform Cable ducts Lifting pole bushing Infusion stand bushing Wall deflection bumpers Castors Brakes Interior parts accessible after removing mattress platform covers Product Altura V
5. 10 4 1 Scope of Delivery 1 hh 10 a2 Bed VINAM rarere rena eee coe ls O eee ewes sees ewe eee eeeeeses 10 OD CUD foe ce Z ee ae Se et ees Be ea ee ers ee ne eee 11 XM Ico LPREMEREITCCP 11 PM KR S O T L E E E E EE TTE EEE TET 11 o2 NOU 64464654065 tone O O E E ew Oe O S O yee ee E 12 b Ded BIidS ced oue eH sen Goes Soe CHE RU OGG ee eho ae ae eR ae ee tee Rees eae ES eae REPE 14 G2 Potential EgqualisatOn caseiceeedheseresceceatedeasdesavets O sebheetae se 15 Y Oo 1 61 PE E d hs ee eee ye ee eye PETS RRUTTAESES EEG R geese sas 16 FEE ESO rU 25 204 orea cae ee eee ea Le eo ease O O ohn es 16 7 2 Battery Operation no 26 nahe d dk Rasa 6 eko donde Eun AGERE RO EG Ra Ed eens dd deka 16 7 3 Removing the Bed from Service s 18 o A PPU M E O O ee eee es Cees Sade se eRe Sees 19 8 1 Supervisor Panel 0 0 0 0 000 ee ee eee eet 20 Mariage PRPP 23 8 3 Satellite Panel 2 lllllleeeleeeeee RR Rh ss 24 8 4 Foot Control Bed Height eee hn 25 8 5 CPR Backrest Release 2 cc eee hh 25 C6 os PES oe ase take TITIO 26 8 7 Castor Control and Bed Transport ee s eee ee ee eee 27 Gio ACCESSOS iia eek dr 98 75 475 a O i Dre Je cale e ed be Oa oe aie ree on ee oS 28 9 Oleaning DISInIecllOEis s 2373 2 2 03 9 064286 os seeded eb teste heed pee bed tes ere epee eee ESSE 31 9 1 Preparing for Cleaning ee Rr ss 31 SE EEG SO S O O E E E E A EE EE E
6. Castor Control and Bed Transport Move the bed on suitable floor surfaces only o Suitable surfaces B file m Hard linoleum m Poured flooring Unsuitable surfaces Too soft unsealed or defective flooring Soft wooden flooring Soft and porous stone floors Carpeted floors with underlay Soft linoleum For longer distances ensure that the castor steering function main control is activated Ensure that the brakes are released while moving the bed 5 2 Setup Set up the bed as follows Unpack the bed Check the delivery see Scope of Delivery and Bed Variants Install equipment and accessories see Assembly In case of delivery with dismantled bed ends mount the head and foot ends see Assembly Set up the bed on a suitable floor surface only see Transport Ensure that the mains cable does not collide or get stretched when adjusting the bed Check that the plug is inserted correctly Do not leave any extension cords or power strips loose on the floor Ensure that all of the required mechanical and electrical prevention mechanisms are available on site There is no mains switch on the bed i e the mains cable is the only means to isolate the bed from the mains Ensure that the mains cable is always accessible Have the separable plug of the mains cable changed and maintained exclusively by qualified and trained service technicians authorized by the manufacturer gt gt
7. EMC information given in this Manual gt i gt Altura is intended for the application in electromagnetic environment as specified below The customer or user of the bed is responsible for the fact that these requirements are met Product Altura 42 48 Version 01 Winer 15 4 1 Manufacturer s Manual and Declaration Electromagnetic Radiation Radiation Test Conformity Electromagnetic Environment Group 1 Satisfactory Product Altura 43 48 Version 01 WLiner 15 4 2 Manufacturer s Manual and Declaration Electromagnetic Resistance Resistance Test Test Level as per Level of Com Electromagnetic Environment IEC 60601 pliance Electrostatic 6 kV for contact 6 kV for contact Ensure that the following requirements discharge ESD 8 kV for air 8 kV for air are met IEC 61000 4 2 Floors wood concrete or ceramic tiles Relative humidity gt 30 Electrical fast tran 2 kV in feeder line kV in feeder line 4 Ensure that mains quality is suitable for sient response 1kVin input out kV in input out commercial or hospital environment group of impulses put line ut line IEC 61000 4 4 Shock pulse 1 kV between 1 kV in differential 4 Ensure that mains quality is suitable for IEC 61000 4 5 lines mode commercial or hospital environment 2 kV between line X 2 kV in co phasal lines and earth mode Short time voltage 5 Uy 5 UT Ensure that mains quality is suitable for drop short dur
8. degassing can occur In rare cases this can cause deformation of the battery packhousing and mains control box If this occurs the bed must be immediately taken out of service and taken to an adequately ventilatedroom without sparks electricity or fire Immediately inform service department of the manufacturer Warning Danger of reducing battery durability due to wrong use Use bed on battery only in crisis situations e g power blackout patient complications during transport etc After reconnecting bed to the mains charge battery to full capacity see chart Battery charge status Warning Battery must be replaced with new battery approved by the manufacturer Replacement must be done by qualified service organisation every 2 years at the latest Bed must only be fitted with batteries approved by the manufacturer To get more information on how to change battery please refer to the Service manual contact service department of the manufacturer Product Altura 17 48 Version 01 rris eT 7 3 Removing the Bed from Service Remove the bed from service as follows Disconnect the bed from the mains Disconnect the ground wire Deactivate the battery Remove accessories S te To prevent damage during storage Pack or cover the bed and accessories Ensure that storage conditions are the same as the operating conditions 7 3 1 Deactivating the Battery To avoi
9. gt ad gt Product Altura 11 48 Version 01 Winer 6 Assembly 9 8 765 4 Fig 1 Overview of Altura Domea Head end Support for IV pole lifting pole or extension system Single collapsible siderail Plastic cover for column protection Handset 5th castor Siderail release mechanism Standard castor Tente PJP 125 mm Column 10 Central castor control lever 11 Supervisor panel 12 Foot end 2 059 NOTE For safe easy handling Linet recommends that two technicians assemble the bed Product Altura 12 48 Version 01 Winer Fig 2 Overview of Altura Thema Head end Handset 5th castor Full length siderail Linen shelf with supervisor panel Foot end PDM NOTE For safe easy handling Linet recommends that two technicians assemble the bed Product Altura 13 48 Version 01 Winer 6 1 Bed Ends X A Fig 3 Locking unlocking the locking pin Dismount the bed ends as follows Unlock locking pin Pull out bed end Install the bed ends as follows Unlock locking pin Put bed end back Unlock locking pin Product Altura 14 48 Version 01 Winer 6 2 Potential Equalisation The bed is equipped with a standard protective connector This connector shall be used for potential equalisation between the bed and any intravascular or intracardiac device connected to the patient to protect the patient from
10. static electric shocks Fig 4 Potential Equalisation 1 Potential equalisation connector male 2 Potential equalisation connector female Use equalisation connector if m the patient is connected to any intravascular or intracardiac device Before connecting the patient to an intravascular intracardiac device Connect the ground wire of the device to the potential equalisation connector 2 on the bed on which the patient in question is lying Use a standard hospital connector 1 Make sure that the connectors match Make sure that there is no possibility for inadvertent disconnection SC Ss SS KS Before moving the bed 9 Disconnect the patient from the intravascular or intracardiac device Disconnect the potential equalisation connector Product Altura 15 48 Version 01 rris eT 7 Operation 7 1 Initial Operation Prepare the bed for service as follows Connect the bed to the mains Charge the battery Raise and tilt the mattress platform to the highest position Lower and tilt the mattress platform to the lowest position Check that the castors as well as main brake work correctly Check that the bed extension works correctly Check that the head and foot boards can be removed Check all of the functions on the control elements Multiboard etc Check that the siderails function properly Dispose of all packaging see Disposal e S kS
11. take no responsibility for any damage to the products or any harm to patients staff or other individuals resulting from E Not following the instructions in the manual including warning notices m Using the product for a purpose other than the intended purpose stated in the relevant documentation provided by Linet i see Correct Use Product Altura 8 48 Version 01 Winer 3 Standards and Regulations The bed complies with the following standards and directives IEC 60601 1 2005 IEC 60601 1 2 2007 IEC 60601 1 6 2010 IEC 60601 2 52 2009 ISO 10993 5 2009 ISO 10993 10 2010 93 42 EEC The manufacturer adheres to a certified quality management system in compliance with the following standards m ISO 9001 2008 m ISO 14001 2004 m EN ISO 13485 2003 Product Altura 9 48 Version 01 rris eT 4 Scope of Delivery and Bed Variants 4 1 Scope of Delivery Delivery Upon receipt check that the shipment is complete as specified on the delivery note Notify the carrier and supplier about any deficiencies or damages immediately as well as in writing or enter them on the delivery note 4 2 Bed Variants Damage to the bed due to incorrect use Use 125 mm castors only to transport bed without patient Caution Bed features m Mattress platform type B m steel segments removable plastic parts m Siderails Type B E without siderails Altura Domea Altura Thema E single collapsible siderails Altu
12. 6 48 Version 01 WLiner The LED indicates the battery s charge status Yellow LED Battery status Not lit Battery capacity is sufficient charging completed Short flashing shortly lit Battery is charging continue charging until the LED is extinguished In longer not lit circa 1 8 sec emergency cases the battery can be used as a backup power source for a short period If LED is still flashing after 12 hours of charging or stops flashing but you can not position with bed battery is defective or broken Conect manufacturer Long flashing longer lit Low battery voltage battery can not be used as a backup power supply even shortly not lit circa 0 2 sec for a short period battery is completely discharged or defective if this type of signalisation persists it is necessary to replace the battery service action Lit continuously for several Battery absence or failure condition battery is connected incorrectly line hours circa 10 hours when between the power supply and battery is broken or battery fuses are faulty bed is connected to the mains contact service department of the manufacturer in case of such signalisation Replacing the battery Damage to the bed due to incorrect battery replacement Have the battery replaced exclusively by qualified personnel Exclusively use batteries approved by the manufacturer Caution Risk of damage or destruction of battery If battery is faulty
13. E EE E E 32 10 Troubleshooting X ace we 08 de m ee ede o e Coe ee Oe ee ee oe BA 34 TT WalMGNGNCe ZR S Sues Eme xr ESSu FPES teeta pette dena e dete abet s dudas d Red apa 35 11 1 Monthly maintenance lille hh nns 35 11 2 Maintenance every 3months cc RR hh 35 11 3 Maintenance every 12 months 0 0 06 aaa ee ee eens 35 113000 CMCCKs KECTNIPCTPRICTTPTx rrrTPPERE 37 BS o ANP aceon bee eo eee bee O ease eee 38 12 1 Environment Protection ied ac veto oboe Wh hbk EER ONE ER Eus d o bd Se Ge Grece m we De 38 12 2 1JISDOSd ae 352 9 2992 MPSSTUSPEAOCEPSSENICEUPeSFRU PS eet PUE KU TUME ERR eee ene 38 ESCUDO P E oh oe Ee hee ed en a a a hd ee we a 39 14 EC Declaration of Conformity 0 0 0 0 eee eet nee nee eens 40 15 Technical Specifications a na naa aa aa eee rrr 41 15 1 Mechanical Specifications n a aana aeaa ee ee eee ee eee teens 41 15 2 Electrical Specifications aa ee hrs 41 to 0 gt em ek ge oe O a eee ees Bee ee eee ee eee eee 42 15 4 PB 43 Electronic System of Altura ee rn 42 16 oo DET 46 16 1 DEINEN LOO CMPCrw 46 16 2 Service and Maintenance Log s ee ee ee ene eens 47 Product Altura 3 48 Version 01 Winer 1 Symbols 1 1 Warning Notices 1 1 1 Types of Warning Notices Warning notices are differentiated by the type of danger using the following key words Caution indicates potential damage to property Warning warns against bod
14. Number Serial Number Delivery Date Delivered By hereby confirm that the personnel has been familiarised with the correct operation of the bed Date Customer s Signature and Seal Supplier s Signature and Seal Product Altura 46 48 Version 01 LINET 16 2 Service and Maintenance Log Description of Service Date Performed By Product Altura 47 48 Version 01 Winer Contact Linet spol s r o Zelev ice 5 274 01 Slany Tel 420 312 576 111 Fax 420 312 522 668 E mail info linet cz http www linet com Product Altura 48 48 Version 01
15. a gt 95 short dura gt 95 96 short dura commercial or hospital environments moe Hs eae tion drop of U7 tion drop of Ur For permanent operation during a power der input line o o generator since the back up battery s IEC 61000 4 11 all ae 60 96 short dura 60 96 short dura capacity is limited tion drop of Ur tion drop of UT within 5 cycles within 5 cycles 70 UT 70 Ud 30 short dura 30 96 short dura tion drop of Ur tion drop of Uy within 25 cycles within 25 cycles lt 5 Uy lt 5 Ud gt 95 short dura gt 95 96 short dura tion drop of UT tion drop of UT within 5 s within 5 s Magnetic field of 3 A m 3 A m Ensure that the magnetic fields of the network frequency network frequency conform to the nor 50 60 Hz mal levels of commercial or hospital IEC 61000 4 8 environments NOTE U7 refers to the AC mains voltage before the test level is applied Product Altura 44 48 Version 01 WLiner Resistance Test Test Level as per Level of Com Electromagnetic Environment IEC 60601 pliance Conducted high 9 Vins DS Do not use portable and mobile HF frequency pheno 150 kHz to 80 MHz communication equipment near the bed mena a ues IEC 61000 4 6 Observe the distances indicated below Radiated high fre 3 V m Recommended distances quency phenomena 80 MHz to 2 5 GHz IEC 61000 4 3 d 1 2 JP d 1 2 JP 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz P is the r
16. ated maximum output power of the transmitter in Watts W defined by the transmit ter s manufacturer d is the recommended separating distance in metres m Ensure that the field intensities of per manent HF transmitters determined by the summary of electromagnetic charac teristics for the given place do not exceed the satisfactory level b in each frequency range Interferences are possible in the vicinity of the instrument marked with the following symbol 9 a t is not possible to accurately indicate field intensities from permanent transmitters e g radio base stations of the radio phones and ground mobile and amateur radio stations AM and FM radio and television broadcasting To assess the electromagnetic environment for permanent HF transmitters take into account the on site electro magnetic characteristics If the measured field intensity is higher than the pertinent satisfactory HF level stated above observe whether the bed is functioning normally If any abnormal properties are observed move or relocate the bed The field intensity in the entire frequency range from 150 kHz to 80 MHz should be lower than 3 V m NOTE With 80 MHz and 800 MHz the higher frequency range is applicable NOTE The absorption and reflection of buildings objects and people will influence electromagnetic propagation Product Altura 45 48 Version 01 Winer 16 Log 16 1 Delivery Log Order Number Customer Model
17. corresponding function buttons Disabling individual functions with the lock buttons Each time a function button is pressed the keypad will remain active for another 3 minutes 8 1 3 Function Buttons The function buttons 3 6 9 12 and 15 adjust the position of individual elements as well as the height and tilt of the mattress platform NOTE If you press two function buttons at the same time the controller will recognise this as an error The controller immediately interrupts all bed movements Set the position as follows Activate the keypad by pressing the GO button 2 Press and hold function button until desired position is reached 8 1 4 Lock Buttons The lock buttons 4 7 10 and 13 on the supervisor panel disable individual functions To disable functions Activate the keypad by pressing the GO button 2 Press the respective lock button The respective LED is lit to indicate the lock NOTE The individual functions are locked in the central control panel the satellite control the handset and the siderail control If the bed is equipped with a foot control the foot control must be locked separately To enable disabled functions Activate the keypad by pressing the GO button 2 Press the respective lock button The function is enabled Product Altura 21 48 Version 01 WLiner 8 1 5 Position Buttons Risk of injury due to moving parts Ensure that no body parts are trapped b
18. covered Justified complaints will be fixed free of charge during the warranty period Proof of purchase with the date of purchase is required for all warranty service Our standard terms and conditions apply Product Altura 39 48 Version 01 Winer 14 EC Declaration of Conformity EC CONFORMITY DECLARATION Date of issue 5 12 2011 Conformity declaration issued by Commercial name Linet spol s r o Registered address elev ice 5 274 01 Slan Czech Republic Reg No 00507814 Telephone 420 312 576 111 Fax 420 312 522 668 As the producer of the product name Altura Thema Altura Domea brand Variants of the product 1SMB 15MD Variants are specified in the technical documentation of the product Description and function designation Electrically operated hospital and nursing bed intended for use in the standard care acute care only Altura Thema long term care and or nursing care including all applicable accessories Classification of the product as the medical Class I nonsterile without measuring function device A Declaration declare that the said product is safe under the conditions of common use in compliance with the instructions and that measures have been taken to ensure the conformity of all the products brought to market with basic requirements of directives related therefore stated in paragraph B B Fulfilled technical requirements This product s characteristics comply with the t
19. ct Altura 27 48 Version 01 rris EY Castor control for bed without control levers 1 2 Fig 13 Positions of Castor Control Lever Every individual castor has its own brake To brake a castor push brake down with foot To release brake push brake up with foot Transporting the bed Adjust bed height to at least 20 cm below maximum height Push bed by handles on head or foot end 8 8 Accessories The use of incompatible accessories can injure the patient e Use original accessories from the manufacturer only Warning NOTE The manufacturer is not responsible for the use of unapproved accessories NOTE All accessories meet the requirements of IEC 60601 2 52 2009 8 8 1 Mattress Altura is designed for the following standard mattress sizes m 200 cm x 86 cm x 14 cm m 200 cm x 86 cm x 20 cm Place the mattress exactly in the middle of the mattress platform Product Altura 28 48 Version 01 Winer 8 8 2 Lifting Pole Fig 14 Bushing for lifting pole and infusion stand Lifting pole variants E Lifting pole for beds with fixed bed end Lifting pole for beds without fixed bed end To ensure safe use of the lifting pole Never exceed the maximum load of 75 kg Never use the lifting pole for rehabilitation exercises To prevent the bed from tipping over ensure that the lifting pole does not project out from the bed Lifting pole positions m Over the backrest
20. d damaging the bed and the environment during storage Deactivate the battery on the supervisor panel To deactivate the battery on the supervisor Disconnect the bed from the mains Disconnect the ground wire Activate the keypad by pressing the GO button on the supervisor Press the Thigh Rest Up Thigh Rest Down Trendelenburg Position buttons at the same time and hold them for three seconds e The battery is deactivated Product Altura 18 48 Version 01 Winer 8 Manipulation The bed is operated by different control elements Control elements depending on the model Supervisor Handset Handset with adapter for easy connection Plug and Play Handset with illuminated buttons Foot control for height adjustment Patient control elements integrated in both middle sections of the siderails Disabling individual functions on the supervisor panel will affect all control elements If the bed does not react to individual position settings 9 Check whether the function is disabled on the supervisor panel Product Altura 19 48 Version 01 Winer 8 1 Supervisor Panel The supervisor panel is an optional control element The supervisor panel can be hung on the foot board or held in the hand while operating the bed JIK When not in use store the supervisor panel in the built in linen holder below the foot end if available 1 2 3 4 56 7 89 10 11 12 13 p
21. d stored in an indoor environment m where there is a risk of explosion E containing inflammable anaesthetic The bed is designed for use in rooms for medical purposes Electrical installations must therefore meet local norms laying down the necessary conditions for electrical installations 9 Disconnect the bed from the mains in exceptional cases i e a storm 2 3 Risk Assessment 9 Have a qualified clinician or medical provider conduct a risk assessment to ensure maximum safety of the patient Product Altura T A8 Version 01 rris eT 2 4 Specifications of use Altura is a care bed designed for supporting the patient in several positions such as horizontal Trendelenburg Cardiac Chair etc 2 4 1 Correct Use The care bed is suitable for m Patients m older that 15 years m with a weight of up to 165 kg m Personnel qualified medical staff E any person familiar with the manual m patient condition permitting m Use m standard care E uninterrupted operation m Location E nursing homes m care facilities 2 4 2 Incorrect Use The bed is not suitable for m Patients E younger than 15 years m Use E intensive care NOTE For information concerning uses other than those outlined in the Correct Use section above please contact Linet Linet s efforts in research design and manufacture make sure Linet P products are of the highest quality and fit for their intended purpose However Linet can
22. earch design and manufacturing of beds to ensure that their products are fit for purpose and producd to the highest level of quality However Linet can take no responsibility for any damage caused to the product or harm to patients staff or other individuals as a result of m Not following the instructions for use including warning and caution statements provided in their user manuals m Use of the product for other than its intended purpose as stated in any documentation provided by Linet NOTE On reguest the manufacturer will provide circuit diagrams component part lists descriptions calibration instructions etc for service personnel for the repair of ME eguipment designated by the manufacturer as repairably by service personnel Product Altura 37 48 Version 01 rris EY 12 Disposal 12 1 Environment Protection Linet is aware of the important role that the protection of our environment plays for future generations The materials of this product are environmentally compatible It does not contain hazardous substances on the basis of cadmium mercury asbestos PCB or CFC The noise emission and the vibrations meet the directives for premises None of the wooden parts are made of tropical woods for example mahogany jacaranda ebony teak etc or of woods from the Amazonian region or similar rainforests The packaging materials are produced according to the respective directives Dispose of the packaging material acco
23. echnical parameters related to it and stated in MDD 93 42 EEC as amended which stipulates the technical parameters for healthcare products in directive 2004 108 ES as amended which stipulates the technical parameters of products concerning their electromagnetic compatibility and directive 2006 42EC as amended which stipulates the technical parameters for machinery device C Means of assessing conformity Conformity was assessed by the procedure stated in MDD 93 42 EEC Annex VII D Used standards The said product fulfils the requirements of these harmonized technical standards which were used for assessing of conformity EN 60601 1 2005 EN 60601 1 2 2007 EN 60601 1 6 2010 EN 60601 2 52 2010 EN 14971 2007 Ing Zbyn k Frol k managing director Product Altura 40 48 Version 01 WLiner 15 Technical Specifications 15 1 Mechanical Specifications Dimensions With full length siderail Altura Thema With single collapsible siderail Altura Domea Mattress platform dimensions mattress standard Max mattress height For full length siderail Altura Thema For single collapsible siderail Altura Domea Siderail height over mattress platform without mattress With full length siderail Altura Thema With single collapsible siderail Altura Domea Mattress platform extension With full length siderail Altura Thema With single collapsible siderail Altura Domea Mattress platform height adjustment Maximum backrest ang
24. ersion 01 33 48 Oyun NET 10 Troubleshooting Error Fault The bed cannot be adjusted with the position buttons Faulty mattress platform height tilt adjustment Backrest cannot be lowered from the upright position The siderails cannot be adjusted Faulty brakes Inserting head or foot board is not possible Danger Product Altura Version 01 Cause GO button was not pressed Function disabled on supervisor panel Drive motors have no power Defective drive motors Defective battery Plug inserted incorrectly Drive motors have no power Defective drive motors Defective battery Plug inserted incorrectly Insert the mains plug correctly Faulty power source Notify the service department Faulty control element Notify the service department There is an object under the backrest or in the drive mechanism Locking handle is defective Defective mechanism Risk of mortal injury due to electric shock Solution Press the GO button Enable disabled function Check the mains connection Notify the service department Insert the mains plug correctly Faulty power source Notify the service department Faulty control element Notify the service department There is an object on the Remove the object undercarriage cover Function disabled on supervisor Enable disabled function panel Check the mains connection Notify the service department Remove the object Not
25. etween moving parts of bed and mattress platform Ensure that no persons or body parts are close to the bed or the accessories e g infusion stand lifting pole when the mattress platform is moving Warning Damage to property due to moving parts Ensure that no objects e g cables are trapped between moving parts of bed and mattress platform Ensure that no objects are close to the bed or the accessories e g infusion stand lif ting pole when the mattress platform is moving Caution The therapeutic and safety related positions are preprogrammed When a position is set several parts of bed and mattress platform move simultaneusly You can set the programmed positions with the following buttons m Cardiac Chair Position pre programmed 17 m CPR Resuscitation Position pre programmed 18 Set the position as follows Activate the keypad by pressing the GO button 2 Press and hold the function button until the desired position is reached Cardiac Chair Position The cardiac chair position is for patients with cardiac arrhythmia and breathing difficulties Settings after pressing position button 17 E The foot of the mattress platform is tilted to the lowest position The backrest and thighrest are moved to the upright position You can disable this function If individual functions for example thighrest adjustment are disabled they are not performed CPR Resuscitation Position The CPR position is f
26. ify the service department Clean the locking mechanism Notify the service department Clean the brake system Notify the service department Wrong position of head or foot board Check locking mechanism Position head or foot board correctly Notify the service department Ifa fault occurs have the electric motor power box or other electrical parts repaired by qualified personnel exclusively Do not open the protective covers of the electric motor or the power box 34 48 Winer 11 Maintenance NOTE 11 2 9 X1 X1 9 9 11 3 Linet 9 recommends that you attach the maintenance plaque to the bed Ensure that the following maintenance work is performed every 12 months by the manufacturer or by a qualified service organisation trained and certified by the manufacturer Monthly maintenance Check all movable parts for wear Maintenance every 3 months Check function of brake lever Clean the piston shafts and lubricate with silicone oil Clean the bolts and hinges of the mattress platform backrest thigh rest calf rest and lubricate with silicone oil Check mechanism of mattress platform extension Check bolts on castors and tighten them if necessary Maintenance every 12 months 11 3 1 Spare Parts The product label is located on the inside of the longitudinal rail of the mattress platform frame The product label contains information for claims and orderi
27. ily injury Danger warns against potential mortal injury 1 1 2 Structure of Warning Notices 1 2 Other Symbols 1 2 1 Instructions Structure of instructions Perform this step Results if necessary 1 2 2 Lists Structure of bulleted lists m List level 1 m List level 2 Structure of numbered lists a List level 1 b List level 1 1 Listlevel 2 2 Listlevel 2 Product Altura 4 48 Version 01 Winer 1 3 Symbols and Labels on the Product Read instructions for use Attention consult accompanying documents Thermal protection for transformer Only suitable for indoor use Protection against accidents due to electrical current Type B instruments Safety isolating transformer general DDM CE mark CN TX Jack for attachment of conductor for potential V Y equalisation rve aues Product label EET Foye Greece Safe working load Warning against crushing or trapping Maximum weight of patient ordi Use mattress recommended by manufacturer j CP Location of CPR backrest release lever Product Altura 5 48 Version 01 WLiner 2 Safety and Dangers 2 1 Safety Instructions Follow the instructions carefully Only use the bed if it is in perfect working order If necessary check the bed functions daily or at each shift change Use the bed with the correct mains supply only Ensure that the bed is operated by qualified personnel only Ensure that the pa
28. le Auto regression length compensation of backrest Maximum calf rest angle Trendelenburg Anti Trendelenburg position Weight depending on equipment Safe working load including mattress and accessories With battery Without battery Max weight Patient Accessories Mattress Max Lifting Pole Load Environmental conditions Temperature Humidity Atmospheric Pressure 15 2 Electrical Specifications Input Voltage Maximum Power Input DIN EN 60529 Safety Protection Safety Class Applied parts Electrical Motor Operating Time Product Altura Version 01 218 9 cm x 102 5 cm 218 5 cm x 98 3 cm siderail length 166 5 200 cm x 86 cm 20 cm standard 14 cm with extender 24 cm 42 cm 36 cm 15 cm 15 cm 38 5 cm 0 5 cm 6 5 cm 0 5 cm 70 11 cm 30 16 16 140 kg 230 kg 180 kg 165 kg 45 kg 20 kg 75 kg 10 C 40 C 30 75 96 700 1060 hPa 230 VAC 10 50 60 Hz max 1 6 A 370 VA IP 54 Class Type B 10 max 2 min out of 18 min 41 48 NOTE Upon request Linet can deliver hospital beds with electrical specifications that comply with regional standards custom voltage different mains plugs 9 A No special precautions regarding EMC are required for the equipment gt Altura requires special preliminary measures pertaining to EMC that necessitate installation and commissioning in conformity with the
29. n the cable protect the cable from mechanical wear and tear Improper handling of mains cable can cause an electric shock hazard other serious injuries or damage the bed Ensure that the stipulated duty cycle on time is not exceeded Ensure that the moving parts of the bed are not blocked To prevent failures use the manufacturer s original accessories and mattresses only Ensure that the stipulated safe working load is not exceeded If the patient s condition could lead to an entrapment leave the mattress support platform in the flat position whilst unattended Adjust bed height to approx 20 cm below maximum height when transporting the bed in order to facilitate overcoming possible obstacles Do not exceed maximum load of 80 kg for mattress platform extension Ensure that the bed and its components are exclusively modified with the manufacturer s approval hesg hesg hesa hesa hesg hesa hesg gt hesg hesg Product Altura 6 48 Version 01 Winer 2 2 Conditions of use The bed may be used and stored in an indoor environment where the temperature range is between 10 C and 40 C m the relative humidity is between 30 and 75 E the atmospheric pressure is between 700 hPa and 1060 hPa The bed may not be used an
30. ndset using the handset for another bed Product Altura 23 48 Version 01 Winer 8 3 Satellite Panel The satellite panel is optional The satellite panel is attached to the backrest with a flexible arm Fig 7 Satellite Panel 1 GO Button 2 Thighrest Adjustment Button 3 Backrest Adjustment Button 4 Height Adjustment Buttons Set the position as follows Activate the keypad by pressing the GO button Press and hold the function button until the desired position is reached NOTE The nursing staff must decide whether the patient can adjust the bed If the patient s condition requires it you can prevent the patient from adjusting the bed by Moving the satellite panel out of the patient s reach Or Disabling functions NOTE The satellite panel can be fixed on the right or left side of the bed Product Altura Version 01 24 48 Winer 8 4 Foot Control Bed Height The foot control is optional and allows setting the height of the bed using one s feet 1 4 3 2 Fig 8 Foot Control Bed Height 1 Protection Frame against Unwanted Activation 2 Foot Switch Raise Mattress Platform 3 Foot Switch Examination Position 4 Foot Switch Lower Mattress Platform Set the position as follows Press and hold foot switch until desired position is reached 8 5 CPR Backrest Release The bed permits quick mechanical lowering of the backrest for emergency resuscitation CPR procedures
31. ng replacement parts Information about spare parts is available from m Customer service m Sales E Our technical support department Product Altura 35 48 Version 01 Winer 11 3 2 Completeness 9 Perform a visual check with delivery note if necessary Have any missing parts replaced 11 3 3 Wear Check all bolts and tighten if necessary Check all locking mechanisms Check the bed for wear scratches or rub marks Eliminate the cause if necessary Have any defective parts replaced hesg 11 3 4 Functioning 9 4 Check that all bed adjustments reach the maximum position If necessary clean lubricate or replace any worn spots and parts 11 3 5 Electric Control Plug connections Replace O rings on connectors Check plug connections for dirt and defects Clean or replace if necessary Check that the plug connectors are properly seated te te te Motors Check motor movement adjust bed positions Check for incorrect and interrupted movements Have defective motors replaced if necessary Check cables for signs of wear and entanglement Install a new cable or have it replaced if necessary kS kS Battery Check that the battery is working properly disconnect the bed from the mains every 2 months Have the battery replaced if necessary Product Altura 36 48 Version 01 WLiner 11 3 6 Cast
32. or resuscitating the patient in an emergency Settings after pressing position button 18 m The mattress platform is moved to the horizontal position NOTE All mattress platform elements must be retracted to their lowest position by function buttons NOTE For rapid mechanical positioning see CPR Backrest Release Product Altura 22 48 Version 01 Winer 8 2 Handset A handset is included with the bed as an optional feature The position of the handset depends on the patient s condition The handset is available with and without button illumination The button illumination of the illuminated handset is active when the bed is connected to the mains The functions of both handsets are identical 1 23 Fig 6 Handset Buttons Thigh Rest Position LED Thigh Rest Backrest Lock Button Backrest Position GO Button Button Autocontour Button Flashlight LED Height Lock Buttons Height Adjustment 005 To switch on the flashlight Press flashlight button 6 Set the position as follows Activate the keypad by pressing the GO button Press and hold function button until desired position is reached NOTE The nursing staff must decide whether the patient can adjust the bed If the patient s condition requires it preventing the patient from adjusting the bed is possible by 4 Disabling functions NOTE An adapter for the handset is available The adapter enables quick mounting and dismounting e g replacing a defective ha
33. ors Clean the castors completely Grease the castors if necessary Use Caro EP 2 by DEA or an equivalent grease Check that the castors work properly Forward Movement Unrestricted Movement Braked Have the brakes adjusted if necessary Have any defective castors replaced NEB 11 3 7 Accessories Check that all accessories for example lifting pole siderails infusion stand etc are working properly Replace if necessary 11 4 Safety Checks Incorrect safety checks can cause injuries Ensure that safety checks are performed by customer service or authorised personnel certified by the manufacturer only Ensure that the safety checks are recorded in the service and maintenance log Warning A defective bed can cause injuries gt Have a defective bed repaired immediately Ifthe defect cannot be repaired do not continue to use the bed Warning In accordance with 6 of the Medical Devices Operator Ordinance the operator is required to perform a technical safety check on the hospital bed every 12 months The procedure for performing the safety check is stipulated in VDE 0751 or IEC 601 1 The manufacturer will give a certificate to service organisations in which the manufacturer declares that the service organisation is qualified to perform maintanance on Linet products Linet makes every resonable effort within res
34. ra Domea E full length aluminium siderails without extensions Altura Thema E full length aluminium siderails with extensions Altura Thema m Bed ends Type B E design F05 fixed or removable Altura Domea E design B05 wood fixed or removable Altura Domea design H05 removable Altura Thema m Castors Tente PJP 125 mm with individual brakes Altura Domea Tente PJP 125 mm with central brake Altura Domea Tente Motion 125 mm with central brake Altura Domea Altura Thema Tente Motion 125 mm with central brake and 5th castor Altura Domea Altura Thema Tente Motion 125 mm with plastic cover and central brake Altura Domea Altura Thema Tente Motion 125 mm with plastic cover central brake and 5th castor Altura Domea Altura Thema m Control elements Type B E supervisor panel m handset m foot controls m satellite control panel m Other special features m linen shelf m mattress platform extension m 1 pair of urine bag holders BE 1 pair of universal accessory rails m CPR emergency backrest release m Colour powder coated metal and plastic parts RAL 9006 and RAL 9002 Product Altura 10 48 Version 01 WLiner 5 Setup 5 1 Transport For a safe transport observe the following Ensure that no cables are run over when moving a bed Ensure that the mains cable is attached with a hook at the head end of the bed 4 Ensure that the castors are unlocked before moving the bed during the loading unloading process see
35. rding to the symbols and by delivering it to an authorised person The product consists of recyclable steel plastic and electronic components a A CA Y N A em C E 12 2 1 Within Europe 12 2 Disposal To dispose of the appliance When you dispose of your appliance do not put it into the household waste Send the appliance to the recycling of electrical appliances mm The materials of the appliance are reusable By reusing material recycling or other forms of use of old appliances you give an important contribution to the protection of our enviroment Ask the responsible environmental protection authorities for the appropriate disposal point 12 2 2 Outside Europe Dispose of the bed or its components in accordance with local laws and regulations m After using the bed E Following maintenance and installation work Hire an approved waste disposal company for disposal Product Altura 38 48 Version 01 Winer 13 Warranty Linet 9 will only be held responsible for the safety and reliability of products that are regularly serviced and used in accordance with the safety guidelines Should a serious defect arise that cannot be repaired during maintenance Stop using the bed This product is covered by a 24 month warranty from the date of purchase The warranty covers all material and manufacturing related failures and errors Failures and errors caused by incorrect use and external effects are not
36. rn ET User manual and technical description Altura Care bed CE 9200 0364 Version 01 Print date 02 2012 Wruiner Altura Care bed Author LINET s r o Related links www linet com Version 01 Print date 02 2012 Copyright LINET s r o 2012 Translation O LINET s r o 2012 All rights reserved All trademarks and brands are the property of the appropriate owners The manufacturer reserves the right to changes in the contents of this manual that relate to the product s technical regulations It is for this reason that the contents of this manual may indicate differences from the current manufacture of the product WLiner MEE 000 rrr 4 11 Waming NOCES RPCT BBem 4 1 2 Other Symbols 262444252 eu s 5 dc a EP Ed A EE S Rd 8 od B rs mr Ea Ro ERE dono 4 1 3 Symbols and Labels on the Product 0 0 0 cc eee 5 2 Safety and Dancers M TITIO TEILT TOT TITOLI DTE 6 2 1 Salely INSIMICIONS e ocea bubs a we Bote 609 b hee Oh eho on Se 00 E ee HES Beto Cb dot he gon decko be Ob RO 6 2 2 COHdHIOFS Ol Uses 42 392291 oe area P PPRECPPerIRePEETPRESEPPEEFPEbS SE SU PES RUE PT T 2 9 ISK ASSOSSImOUD ua 33 9 5 8006 5h b 2h5 54 08 5 ee dun dice her a E A pince 8 BORE a de eh die T ZA Specifications Of USE sss 4 z sedan 9 We retirer nd Boj ahaideei arae rekia 8 3 Standards and Regulations 0 0 ee eee ee eens 9 4 Scope of Delivery and Bed Variants ee e
37. tient health permitting has been informed about the operation of the bed and all applicable safety instructions Move the bed on even hard surfaced floors only Replace any damaged parts immediately with original spare parts only Ensure that maintenance and installation are performed exclusively by qualified personnel who have been trained by the manufacturer Do not apply excess weight or loads to the bed During peak loads or unavoidable excess loads CPR place mattress platform in the lowest position Ensure that only one adult patient uses the bed at any time Take care to avoid injuries or squeezing when operating moving parts When using lifting poles or infusion stands ensure that nothing will be damaged when you move or adjust the bed Brake the castors when the bed is occupied Keep the mattress platform in the lowest position at any time when the healthcare personnel are not treating the patient in order to prevent the patient from falling or injuries Ensure that siderails are operated by healthcare personnel only Never use the bed in areas where there is a hazard of explosion Enable or disable functions on patient controls using the supervisor panel as appropriate for the patient s physical and mental state Verify that the function is actually disabled Never handle the mains plug with wet hands Remove the mains cable by pulling on the plug only Position the mains cable so that there are no loops or bends i
38. ts that are suitable for cleaning medical equipment finish Never use any corrosive or caustic detergents Never use detergents that deposit calcium carbonate Never use detergents with solvents that might affect the structure and consistency of the plastics benzene toluene acetone etc Clean the bed with a well wrung damp cloth Clean electrical components carefully and allow them to dry sufficiently S Linet recommends the following cleaning agents Cleaning agents Manufacturer Sch lke amp Mayr BODE Chemie LYSOFORM Ecolab Schulke 9 1 Preparing for Cleaning Prepare for cleaning as follows Put the mattress platform in the highest position Adjust the back and thigh rests so that the reverse sides are accessible Disable the function buttons on the control elements using the supervisor panel Disable the foot controls using the supervisor panel Disconnect the bed from the mains Move the bed to the location where it will be cleaned Lock the brakes on the bed SA t hesa e hesg Product Altura Version 01 Do not use abrasive powders steel wool or other material and cleaning agents that might damage the Winer 9 2 Cleaning 9 2 1 Daily Cleaning Clean the following bed parts m All of the control elements for adjusting the bed m All handles
39. working position m Parallel to the head board To install the lifting pole Fig 15 Safety pin locked in place Insert the lifting pole in the corresponding bushings at the head of the bed corners Ensure that the safety pin locks into place A plastic grab handle with an adjustable strap is attached to the lifting pole NOTE The date of manufacture is marked on the grab handle Linet 9 recommends that you replace the plas tic grab handle every four years Product Altura 29 48 Version 01 Winer 8 8 3 Infusion Stands It is possible to insert infusion stands into the sleeve fittings or the multifunctional accessory adapters at the head end Use exclusively infusion stands with four hooks for hanging IV bags or for hanging a basket for intravenous solutions Ensure that the weight bearing capacity of the infusion stand and the four hooks is not exceeded The maximum weight bearing capacity per hook is 2 kg 8 8 4 Accessory rails Fig 16 Accessory Rail Load capacity m Maximum load of 5 kg without leverage E Maximum load of hook pair 10 kg Accessories for hanging on the accessory rail E Urine bag holder E Urine bottle basket m Cannula holders m Stainless steel rails Product Altura 30 48 Version 01 Wruiner 9 Cleaning Disinfection For safe and gentle cleaning Disconnect the bed from the mains Do not use any strong acids or bases optimum pH range 6 8 Only use detergen