Home
        Flow Center User Manual
         Contents
1.                 M   095177    Current Mode  Flow       Change to  Flow x  DP    Gas flow  abs   Gas flow  norm     The user is free of choice to change the mode 1f this is wanted for any reasons  If this is done   the device must be restarted Power off power on   The simplest way is to just disconnect  the communication cable and re attach it and restart the communication with Flow Center   Errors in reading might occur if this is not done  Do not disconnect the unit from Flow Center  and connect it again  The unit is still powered up after disconnection to Flow Center     61A7E15    FLOW MONITORS          10  Mode descriptions    10 1 DP Mode    DP Mode configuration is used when you only need to measure the actual differential  pressure  Media is less important since you cannot display flow rate  Flow Meters in DP  mode is typically not used with our classic pipe sections that includes orifices  We normally  provide Flow Meters in DP mode with our connection blocks P8BR and P8SS  more  information about our connection blocks can be found in the M3 series manual     10 2 Flow Mode    Flow mode configuration is used when measuring flows of incompressible liquids  normally  water and oil     10 3 Gas Flow Mode  absolute norm     Gas flow mode configuration 1s used when measuring flows of gases  normally air  When  choosing gas flow mode there are two choices  absolute or normalized     Calculating the orifice plate used for the device when dealing with gases are most of the ti
2.      Sa          P Load transmitter settings from file  Action  Load previously saved device settings     ef Save transmitter settings to file  Action  Allows you to save device settings     61A7E15    FLOW MONITORS          5  Connect a device    Connect the Flow Meter to your PC with the communication cable for M3 series and open    Flow Center  w Add new device  The following input box will appear     9 Flow Center  Add transmitter  Add a new M transmitter    FLOW MONITORS                    Serial port     Add a new M transmitter with ID    250     V  Autodetect       Click on UPDATE to find all connected occupied USB ports  Choose the port where the  device is connected  Click on OK and the device will be uploaded for communication    The AUTODETECT can be unchecked to search and communicate with a specific device  with a certain Modbus ID  The MODBUS interaction is not yet fully developed and should be  used with pre caution           Once the device is connected it shows up with a green   online icon followed by the units  unique S N  When clicking on the   connected device you enter the read view which  consists of 5 different sub pages where you can oversee Overview  Displaying all  measurements in one page   Flow rate  Differential pressure  Temperature and Static  pressures  Note that the scaling is dependent on the settings saved on the device and  measurements will be performed immediately as the unit 1s connected              Overview   Flow rate   Differential 
3.    5        Open the  Setup flow _ center X X   choose installer language and follow the setup       wizard to install  Available languages are Swedish and English  Run the setup as  administrator  If you don   t have the necessary privileges  please contact your IT       support before installing the software        Flow Center does not work without the required plug in  If you are to communicate  with the M3 flow meter  open the  Setup Flow Center M plugin   X X   Choose  installer language and follow the setup wizard to install  Run the setup as  administrator  If you don t have the necessary privileges  please contact your IT          support before installing the software        Install the driver K 114B for the communication cable  The driver for the RS485   USB communication cable is available for download online at Eletta server  http   www eletta se en downloads softwaredrivers         Run the setup as administrator  If you don   t have the necessary privileges  please       contact your IT support before installing the software        Refer to the manual in the package for troubleshooting and system requirements for  the USB RS485 converter     Connect the communication cable to any USB port on your computer and the M3  unit     Run Flow Center X X    61A7E15    FLOW MONITORS          4  Overview    General Functions       File  M  Help          Flow Center              FLOW MONITORS                 File     Ze FC  Setup Ctrl Alt  Enter    Action  Change language  P
4.  is 4 100 NI min and not 10 100 NI min                                                  Flow DP Analog Out  Lo  0 00   Lo 0 12   Lo  0 00    Min  0 00 Max  100 00 Min  0 00 Max  0 50 Min  4 00 Max  20 00  Min measurable Flow P2  Co  4 14   yd  10 28   yf  10 27    Min  0 00 Max  100 00 Min  0 00 Max  10 00 Min  0 00 Max  10 00  Turndown TP Coeffs Temperature  HE z 3  E Abs Preassure  6 013 bar  doai iii 38 57   Min   10 00 Max  80 00    11  Additional Functions    11 1 Save   Load transmitter settings    By clicking on  amp  Save transmitter settings to file you are able to save the configuration from  the settings page to a separate file on your computer  This can be convenient to use when  there is a need to configure lots of units with the same setting  where the user instead of  writing all settings manually instead reads a setting file and save the new settings to the unit     To load settings from file  click on     Load transmitter settings from file  If the unit is not set  to the same mode as to what the setting file was written for  the following will be prompted  on the screen              M 104588  Load settings x       You need to change mode before loading this file   Do you want to change mode   DP to Gas flow  norm     C Cance                By clicking no  the action is interrupted     By clicking yes  the unit is only changed to the new flow mode  In order to save new settings  the user must re load the settings file again  Then the unit can be saved with th
5.  or processing is not permitted     Any other use of data and information 1s strictly prohibited without prior written permission  from Eletta Flow AB  Sweden     61A7E15       FLOW MONITORS       1  General Information    Flow Center is a software developed by Eletta Flow to communicate with the Eletta M3   series Flow Meter  The M3 series Flow Meters is used to measure and control flows of  liquids and gases in pipes     Flow Center is currently only available for use with Microsoft Windows     The communication between the software and the Flow Meter is bi directional  This means  that the software reads existing data and parameters from the Flow Meter  It also allows you  to re configure the flow meter using the communication cable for M series     The communication cable for M series is a USB cable with a RS 485 converter and is  available to purchase from Eletta Flow     2  Technical specification for communication cable  e Supply via USB for one connected final consumer  U Out   11 8 VDC   I Out  max 40 mA   e Measuring channel  voltage input 0   12 VDC  e Visual status and configuration display  LED     e Galvanic isolation  e Total length  1 7 m  e Driver package included with package       61A7E15    FLOW MONITORS          3  Installation    Flow Center software  M plug in and driver K 114B for the communication cable are    available for download from http   www eletta se en downloads softwaredrivers   Check for  the latest versions online     1     2     3     4  
6. 0 000365  0 000447 4 000000 0   000000  2014 05 06 12 16 38 O   000000 0 000045 24 329246  0 000358  0 000403 4 800000 8 880000    FLOW MOMTORS          www eletta com    61A7E15    
7. 00                 20mA  100 0     4          Low pass filter  16s v                  Information   SIN  104588  Class  5 20  Version  13 17  Qmax  100 l min  DP max  0 500 bar  Prange  0  10 bar  Trange   10  80  C  Address  001   Mode  Flow   Port  COM2    Discard         Customized settings Factory defaults                   61A7E15    FLOW MONITORS          7 1 Measurement range    Custom tag       Measurement range                         Qrange 0  100 0 l min    DP range  0  500 0 mbar 7  Drop out  4 000 mbar x             Here you can change the max flow and corresponding DP range  Note that these values are  related to the orifice in the pipe section and should not be changed without ordering a new  orifice made according to new settings  The Max pressure is the maximum allowed static  pressure  This is defined by the pressure transmitter installed in the Flow Meter  Eletta  provides three different transmitters with static pressure limitations of 10  25 and 50 bar   There is also the option of custom tagging the unit to a specific name     7 1 1 Q range    States the maximum flow for your specific orifice  The value that is set as Q range will  always correspond to 100  of the set DP range     7 1 2 DP range    Determines the differential pressure span from where the transmitter calculates the flow  The  orifice plate installed must be calculated so that 100  flow corresponds to 100  DP that is  set in the DP range field  All orifice plates for the M3 series Flow Met
8. 21      Min  0 00 Max  100 00 Min  0 00 Max  500 00 Min  4 00 Max  20 00                         Min measurable Flow P1 P2    EE     O A     ma    LC  30 79   El 10 32     10 28        Min  0 00 Max  100 00 Min  0 00 Max  10000 00 Min  0 00 Max  10000 00                               8 2 2 Drop out set too high    If the cut out function is changed to for instance 15 mbar for a M310 with Flow mode setting   For instance to filter out line pressure disturbances at a non zero flow   the min value of the  flow will be around 17  of max flow  DP readings below 15 mbar will be considered as false  and the unit will show 0  flow  The turndown is 1 10 which should mean that the unit will  display flow from 10  up to 100   But the start min flow value is now limited by the drop  out function value               E Flow Center 10  M  108588   Eletta Flow AB   _ LL es      Si bam  zt   Gee  ELETTA  Flow Center                  M Overview Flow rate   Differential pressure Temperature   Static pressures    104588  Flow DP Analog Out    7  17 45   I 3 04   17 45      Min  0 00 Max  100 00 Min  0 00 Max  500 00 Min  4 00 Max  20 00                         Min measurable Flow P1 P2      Co  10 00   yf  10 43   fT  10 28      Min  0 00 Max  100 00 Min  0 00 Max  10000 00 Min  0 00 Max  10000 00                      Turndown TP Coeffs Temperature    EEEE Ea                         9  Changing mode of the device       Push the    button  An input box will appear     Change mode    FLOW MONITORS
9. FLOW MONITORS    Flow Center User Manual    With plug in for M3 series       ye Flow Center 1 0   M   104588   Eletta Flow AB e   e a    File M Help    Ye  O  amp     ELETTA  Flow Center    FLOW MONITORS 4   104588                               4 ae M il Flow rate   Differential pressure   Temperature   Static pressures    104588                                                          Information   SN  104588  Class  5 20  Version  13 17  Qmax  100 l min  DP max  500 mbar   P range  0  10000 mbar  Trange   10  80  C  Address  001   Mode  Flow   Port  COM2                         61A7E15       FLOW MONITORS          About this Manual    This manual describes how to install and use Flow Center with the M plug in for the Eletta  M3 series Flow Meter     e The latest version of this manual is available for download on our website  www eletta com        e On our website you will also find other information such as leaflets  newsletters   application reports and software download     Proprietary Rights    This manual contains confidential technical data  including trade secrets and proprietary  information that are the property of Eletta Flow AB  Sweden     Any changes or alterations to downloaded or printed Eletta original documentation such as  manuals  drawings  leaflets  newsletters etc  are not permitted without a written permission  from Eletta Flow AB  Sweden     These data are only disclosed to you under permission of limited use within your company   Use for manufacturing
10. can    check in real time     61A7E15    FLOW MONITORS          Flow  Not active in DP mode    DP   Analog out   Min measurable flow  As determined by the dynamic turn down    P1 Pressure upstream   P2 Pressure downstream   Turndown  As determined by the ratio SQRT 100  FLOW DP DPmin device     TP Coeffs  Only active for Gas flow mode    Temperature  Measured on the sensor card of the device and not in the media     et ee ee a ee    Scaling and units can be altered real time and will be automatically displayed with new  settings on the screen  Note that these settings are only changed graphically and not saved to  the unit  If the device 1s disconnected and re connected  the units and scaling values will be  changed to match settings saved on the device     6 2 Flow rate    Overview Flow rate Differential pressure Temperature Static pressures    Total  971         Flow rate  l min     Show min max values       J  Show volume totalizer   Reset totalizer            Time frame   10s           Output signals    Current  6 23 mA       5 128 5 130 5 132 5 134 5 136  Time  s     The flow rate sub page is a real time graphically on going presentation of the flow in time   The time frame is changeable  There is also the option of showing volume totalizer or not   The output analogue signal is also shown on the page     6 3 Differential pressure    The differential pressure sub page is similar to flow rate sub page  The output analogue  signal is also shown on the page  The time frame 
11. e  more sensitive to changes and will detect a more true pressure flow variation  The signal will  vary more in size in between readings and the device will perhaps be considered restless in  output of indicated flow or differential pressure  A high filter value will highly damp the  output signal and will not be able to detect large variations in time and will be more or less a  mean signal of a Os filter value even though there will be a small variation in time     61A7E15    FLOW MONITORS                Overview Flow rate Differential pressure Temperature Static pressures    oO  o    Differential pressure  mbar        35 40 45 50 55 60  Time  s     An M310 with filter value of Os will be more sensitive to detect actual variations        Overview Flow rate Differential pressure Temperature Static pressures    0 555    0 545    Differential pressure  mbar     0 535       35 40 45 50 55 60  Time  s     The same M310 with filter value 26s will highly damp changes detections of variations     Default setting from factory is 1 6s damping value  This is a setting which should work for  most applications     61A7E15    FLOW MONITORS          8 2 Drop out function and turndown    The drop out function is a DP value under which the unit should consider measurements as a  non zero flow condition  The drop out function is to be considered as a min limit where the  PT compensation should stop working and where the meter should interpret the DP signal as  a non flow condition  The drop 
12. e new settings     61A7E15    FLOW MONITORS          11 2 Change MODBUS ID    Click on  amp  Change MODBUS ID to set a specific identification N   for the transmitter  You  can set an ID between 001 250  This is a usable function when you have many Flow Meters  connected to your system at the same time using a MODBUS Protocol communication  For  more information on how to set up a MODBUS communication  contact Eletta Flow  The  MODBUS interaction is not yet fully developed and should be used with pre caution        11 3 Creating log files    Push     to access the ability to log to file  From here you are able to create log files that will  be stored as a  txt file on your computer  At the bottom right you can choose where you want  to store the file on your computer  The time interval for each log point can be set between 1   second to 60 minutes  To start logging  check the checkbox located next to the time interval     e i i      M   Plugin settings  Data logger   Select the trasnmitters you want to log from                   The green   online icon changes during the logging session     i  ro   M 095177    The log file information consists of date time  flow rate  DP  temperature  P1 and P2 static  pressures  mA output signal and accumulated volume     Date and Time flow rate DP Temperature P1 static P2 static mA signal Acc volume  2014 05 06 12 16 36 9   000000 9 000080 24 327991  0 000335  0 000415 4 000000 0   000000  2014 05 06 12 16 37 0   000000 0 000081 24 327991  
13. er are engraved with  the max DP range that corresponds to max flow rate     Normally the value in the DP range field is not to be changed  Do not change this field  without checking the orifice plate versus DP range     7 1 3 Drop out    This field shows when the Flow Meter stops to measure  When the differential pressure goes  below the set value  the output signal will go down to 4 mA and the indicated flow will go  down to zero  This is to prevent the Flow Meter showing erratic flow at zero condition  The  default value is set just below 10  of the max DP range  If you see an unstable flow when  you have a non flow condition  you can raise this value to avoid seeing it  For example   vibrations can cause these erratic small values  therefore you might want to cut them off     61A7E15    FLOW MONITORS          7 2 Pressure settings    7 2 1 Pressure setting DP FLOW mode    Pressure settings       SS    a   e z       ta   e z     Default values for P1 and P2                           For incompressible flow and DP mode there is the option of setting the transmitters to gauge  pressure  Zero bar  g  will remove the small pressure differences in between the two  transmitters and then set the DP to zero bar  g  as a start value to gauge pressure  measurement  The calibration will take some time before stabilizing the values  Click Save  when you are satisfied with the calibration  This is done at the factory at the on site  atmospheric condition and will not be representative 
14. ettings for the orifice plate  If the DP range is changed  the turndown will also be re   calculated  An M310 user setting of 100  flow at 250 mbar will result in a turndown of 1 7   where the 5 mbar now will represent a flow of 14  of full scale value     The min measurable flow will be also automatically be re calculated with changed turndown   Note that the turndown is rounded up to nearest integer     Recommended settings are factory default values   8 2 1 No flow condition but device indicate flow    If the M310 is used for gas flow showing flow when there is no flow in the system  one  should tweak the cut out value  The reason why the device is showing flow is that the device  is actually measuring a differential pressure reading  thus interpreting this as a corresponding  flow  If the normal running state is say 8 bar g   gasmode  abs  and the device is kept at  atmospheric condition  the device will try to compensate for the change in pressure and re   scale the 100  DP so this state will match the same volumetric flow  now at atmospheric  pressure  The 4 5 mbar will now correspond to a much higher flow value than before    around 30   If the conditions here vary between 0 5 mbar  which could be the case   the  device will show flow values 0 and 30  flow   when there is no flow in the system     61A7E15    FLOW MONITORS             Overview   Gas flow rate  abs    Differential pressure Temperature   Static pressures    Flow DP Analog Out    ae  28 21   0 83   yy  28 
15. in the process  The software has no ability to do this summation     61A7E15    FLOW MONITORS    At normal running conditions  the actual and normalized flow rate will be exactly the same  It  is only when P T conditions vary from normal running conditions where they deviate from    each other     It is important to understand that these modes compensate inversely of each other     Flow l min    T_std  Density on the media    pstd STP Lo             a 2  At lower pressure  the volume of the  media is expanding  In order to maintain  the same volumetric flow as run under  normal running condition  the device re   scale the 100  dP to a lower value to try  to compensate for the change in  pressure or temperature  This will result  in a lower turndown since the device now  require a lower dP value  thus lower flow     GASFLOW  ABS      e   e   e       o  act lt p c  P_ P  Oo OOOO OOO       At lower pressure the density of the media is  less dense than normal running conditions    In order to maintain the equal amount of     massflow per unit time   the device re scale the    Normal running condition  100  dP to a higher value to try to compensate     Calc  coefficients  00000000    Te 00000000    00000000  p_c    GASFLOW  NORM     for the change in pressure or temperature   This will result in a higher turndown   since the device now require demand   a higher flow rate to pass the same  amount of molecules per unit time    thus a higher dP max        Assume a unit M310 set for n
16. is changeable and so also the units  Note  that these settings are only changed graphically and not saved to the unit  If the device is  disconnected and re connected  the units and scaling values will be changed to match settings  saved on the device     61A7E15    FLOW MONITORS          6 4 Temperature    The temperature sub page is similar to flow rate sub page  The output analogue signal is also  shown on the page  The time frame 1s changeable and so also the units  Note that these  settings are only changed graphically and not saved to the unit  If the device is disconnected  and re connected  the units and scaling values will be changed to match settings saved on the  device     6 5 Static pressures    Overview Flow rate Differential pressure Temperature Static pressures        Show pressure sensor P1    ic E    1035    Min  0 Max  10000           Show pressure sensor P2    o  8    Min  0 Max  10000       Static pressures  mbar     en  O  N  w    Show min max values       Time frame  30s z       1020    Output signals    Current  4 00 mA       Time  s     The static pressure sub page is similar to the other sub pages  Here you can show both  pressure measurements or choose to uncheck one of them  The output analogue signal is also  shown on the page  The time frame is changeable and so also the units  Note that these  settings are only changed graphically and not saved to the unit  If the device is disconnected  and re connected  the units and scaling values will be cha
17. me  related to a STP calculation  STP is Standard conditions for Temperature and Pressure  If a  flow value is related to this kind of condition  the flow is always denoted as NI min or other  unit  This relation lock the calculation for a specific volumetric flow of a certain kind of gas  with known density at STP state  The STP is different world wide     The orifice plate used in the device is then calculated with user settings  1 e  calculation  settings for temperature and pressure in which the system is working under normal  conditions  The device is then locked at this state and will from this Pressure Temperature  compensate due to variations in the process  The M3 unit has 2 different modes as to what the  user prefers to compensate for or measure  The denomination used in Flow Center is absolute  or normalized flow  Other denominations can be actual volumetric  and standard flow     Standard or normalized flow is the equivalent flow rate the gas would be moving if the  temperature and pressure were at standard conditions  It 1s usually the most useful measure of  gas flow because it defines the mass flow  number of molecules and heat carrying capacity of  the gas     Volumetric flow rate or actual flow rate 1s the actual volume flow of the gas exiting the flow  meter     The M3 device do not have the ability to function directly as a mass flow meter  The  summation must be done elsewhere and if necessary also be temperature adjusted if this  parameter is not stable 
18. nged to match settings saved on the  device     61A7E15    FLOW MONITORS          7  Change settings on a device       Push the Z Setup button to unlock the settings on the device  The settings loaded will  automatically be the settings on the device     Do not change DP  Q  flow  or Calc  coefficients  They are parameters connected to the bore  size of the orifice plate  Changes will cause errors in output           Q range is set to 100 l min when delivered from Eletta factory regardless of which flow range  the orifice plate is specified for  This means that when you first connect your Flow Meter to  Flow Center  the flow rate reading will most likely appear incorrect unless you have a 100  l min specified orifice plate installed  This does not mean that the mA output signal is  incorrect  The mA output signal is determined by that the installed orifice plate and the set  DP range corresponds  To get a correct reading in Flow Center you need to set Q range to the  maximum range of your installed orifice plate     File M Help       a   be    Flow Center           FLOW MON M   104588             M 104588   Flow Settings        104588                                                                   Customta       Pressure settings  Measurement range   Zero bar  g  for P1 and P2  Max pressure    4  bar X   Pl  HHA    bar x  Qrange 0  100 0 Imin x  P2  HHA    bar X  DP range  0  0 500  bar z    Default values for P1 and P2  Dropout  0 004  bar      Output signal       4mA  0 0
19. of the pressure conditions of all  customer installations  Always check the pressure settings before starting the measurement  If  necessary  zero bar g   the unit           Original setting  fabrication settings  values can be retrieved by returning to default values        This is mainly used for troubleshooting and should not be used as a run setting for this mode        7 2 2 Pressure setting GAS mode    Pressure settings       Set P1 and P2 to atmospheric pressure  Li PB           z  pm ou Oo y o ea z    Default values for P1 and P2                            Gauge pressure setting for gas flow is not possible  Here you can either reset the transmitters  to default values or re calibrate and save the current atmospheric pressure condition as  measured with the unit  Setting the P1 and P2 to atmospheric pressure condition will do the  same as the zero bar  g  action for DP flow  It will measure the absolute pressures and  remove small differences in between the transmitters  The calibration will take some time  before stabilizing the values  Click Save when you are satisfied with the calibration  This is  done at the factory at the on site atmospheric condition and will not be representative of the       pressure conditions of all customer installations  Always check the pressure settings before  starting the measurement  If necessary  set P1 and P2 to atmospheric pressure        Original setting  fabrication settings  values can be retrieved by returning to default value
20. ormal running conditions at 5 bar  g   25 deg C and 100 NI min     This is the response from Flow Center at atmospheric condition for Gas flow  ABS   The gas  will be expanded at this state and will only measure between 5 and around 80 mbar   Turndown 1 4 in Q   The device has been forced to re scale itself in order to maintain the  volumetric flow rate at this pressure to match it to the calculation coefficients  This will affect  the units ability to measure flow and the flow range from standard 10 100 NI min is now 24   100 Nl min   Remember that the turndown is rounded to nearest integer      Overview Gas flow rate  abs    Differential pressure Temperature   Static pressures                                                       Flow DP Analog Out  Lo  0 00   Lo  0 15   Lo  0 00    Min  0 00 Max  100 00 Min  0 00 Max  0 50 Min  4 00 Max  20 00  Min measurable Flow P2  Co  24 18   yf  10 27    Min  0 00 Max  100 00 Min  0 00 Max  10 00 Min  0 00 Max  10 00  Turndown TP Coeffs Temperature     Li    l    q Abs Preassure  6 013 bar  Temperature  25  C    61A7E15       FLOW MONITORS          This is the response from Flow Center at atmospheric condition for Gas flow  NORM   The  gas has now become less dense and the device now demands a higher DP in order to maintain  the same amount of  mass flow  per unit time  The device will now measure between 5 and  2900 mbar  Turndown 1 24 in Q   This will actually increase the units ability to measure the  flow range since the range now
21. out value will be set to 4mbar for the M310  10 mbar for the  M325 and 20 mbar for the M350    This value is somewhat related to the accuracy of the absolute transmitter which is around  0 05  of the pressure rating  which for the M310 will be a measurement of around Smbar   The 5 mbar will be the DP min device value for calculating the turndown  One can change  the drop out value  but this could affect the displayed Q flow in relation to what the turndown  is at the moment  The drop out is convenient to use as a filter value when there are  disturbances in line pressure at a zero flow state  This will set a min dP limit where the flow  will start to display  For gas mode there is no actual min gauge value of the static pressure  where the unit will not try to compensate  And a small disturbance in line pressure could  result in a false detection of flow  especially if the unit is set to gasflow  abs  where a  turndown of 1 3 is a normal turndown at atmospheric pressure  The 5 mbar DP value will  then represent as much as 33  flow     The turndown is an accuracy range of the device and should act as a guide of the capacity   This value is a dynamic value which will be automatically re scaled to represent 100  flow   The M310  M325 and M350 will be from factory always configured for a turndown of 1 10  which for the DP measurement represent a measurement range of 1 100 in DP  5 500 mbar  for M310  12 5 1250 mbar for M325 and 25 2500 mbar for M350   These are the calculation  s
22. pressure   Temperature   Static pressures         61A7E15    FLOW MONITORS          On the left bottom side of the screen is the device information frame     ee    Information   SIN  104588  5 20   13 17   100 l min  0 500 bar  0  10 bar  Trange   10  80  C  Address  001   Mode   Port     Class   Version   Q max   DP max     P range     Flow  COM2    Here you can see    e S N  Serial number    e Class  Transmitter class    e Version  Software version written to the sensor card of the device  Keller software     e Qmax  Maximum flow saved to the unit    e DP max  Differential pressure over orifice to match Qmax    e Prange  Static pressure range min  max       T range  Temperature range min  max  e Address  ModBus ID  e Mode  What mode the device is set to measure and report values for  e Port  What COM port the device is connected to    6  Measurement Sub pages    6 1 Overview    Overview Flow rate   Differential pressure Temperature    Flow           0 00      Min  0 00 Max  100 00    Min measurable Flow          Ld  10 00      Min  0 00 Max  100 00    Turndown      Fl  l  tu      Static pressures    DP     0 06      Min  0 00 Max  500 00          P1    E      107     Min  0 00 Max  10000 00          TP Coeffs    Analog Out    in      L Joo     Min  4 00 Max  20 00          P2    fd 10 27      Min  0 00 Max  10000 00          Temperature    Ld  39 38      Min   10 00 Max  80 00          The overview is a collection of all vital parameters as measured by the device  Here you 
23. rogram must be restarted     Quit Ctrl Q   Action  Quit program   M       M  Setup Ctrl Return    Action  Access log function settings  Also accessible using    button   sje M  Add a new transmitter Ctrl N   Action  Add a new M to Flow Center Also accessible using 7 button    Help       FC  About    Action  Info about the software and version    T M  About plugin  Action  Info about plug in    61A7E15    FLOW MONITORS             ye Flow Center 1 0   M   104588   Eletta Flow AB          Ye    GOT    Flow Center    FLOW MONITORS                     a 7e M   Overview   Flowrate   Differential pressure   Temperature   Static pressures      104588       New functions will first appear when a device is connected with Flow Center     Ye Setup    Action  Configure settings for the device     a Add new device  Action  Connects a new device to Flow Center     Remove device communication  Action  Disconnects a device     a  i  Change MODBUS ID  Action  Allows you to set a specific identification No for each devise  The MODBUS  interaction is not yet fully developed and should be used with pre caution           w Change mode  Action  Change the mode you use the M3 device  4 different modes are supported with the  M3 unit     DP  Differential pressure measurement    Flow  UIncompressible flow    Gas flow  ABS  Volumetric flow mode    Gas flow  NORM   Mass flow  mode  Flow value must be density multiplied real   time wrt current temperature and absolute pressure  Not supported by software
24. s        This is mainly used for troubleshooting and should not be used as a setting for this mode        61A7E15    FLOW MONITORS          7 3 Calculation coefficients for GAS mode    Gas flow modes have an extra setting to define the standard  ordinary or stationary customer  operation settings  line pressure and temperature  This is information the device must know  in order to have a P T compensation that will work satisfactory     Do not change these values  They are connected to the orifice bore  Changes will disrupt  measurement                                   Pressure settings  o   Set P1 and P2 to atmospheric pressure       aA oi    E 5   C mi JE    Default values for P1 and P2    Calc  coefficients  Absolute Pressure  9013 0   mbar      Temperature  0 000     c              7 4 Output signal                Output signal  4mA  0 000    A  20mA  100 0     ed    Low pass filter   26s 7         The user can re scale the 4 20 mA output signal  The output can be set inversely also  letting  the 0  flow be 20 mA and 100  flow be 4 mA  You can change the low pass filter value if   there is a need to make the device more sensitive to display changes or to make the unit more  slow to detect changes in differential pressure     8  Impact of settings    8 1 Low pass filter setting    The low pass filter is a damping that will be applied to the calculation of the output signal  If  the filter value is set to 0 there will be no damping of the signal and will thus make the devic
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Bedienungsanleitung Air Joy Cool  Multimedia Interface INFINITY FX / NISSAN Murano  notice of function change (pdf:442 kb)  C-9891 - TXRobotic  ICカードリーダライタ(CLOUD2700) 状態確認方法  FX2N-20PSU DC POWER SUPPLY UNIT USER`S MANUAL  Bedienungsanleitung  Descargar el archivo PDF  Marshall electronic V-R102DP-2C User's Manual  MODE D`EMPLOI Composants de manipulation Unité    Copyright © All rights reserved. 
   Failed to retrieve file