Home
HOBO Waterproof Shuttle User's Guide
Contents
1. HOBOware 2 2 or greater to launch each logger you i l l wish to read out and relaunch with the shuttle later Read Reading out and relaunching loggers in the field Using the shuttle as a base station for instructions if you After you have ensured that the shuttle s batteries are good do not have another base station for the loggers The there is sufficient memory available and the shuttle s clock is shuttle cannot relaunch loggers that were last launched with synchronized follow these steps to read out and relaunch a an earlier version of HOBOware You only have to do this logger in the field fi h1 i once for each logger 1 Make sure the shuttle s large cap and center cap are closed 2 Plug the large end of a USB interface cable into a USB port securely Tighten the center cap until it is just flush with the on the computer Avoid using a USB hub if possible large cap or until the O ring is no longer visible 3 Unscrew the center cap on the shuttle If the cap is too tight 2 Make sure the communication end of the shuttle is clean to loosen by hand insert a screwdriver through the lanyard Attach the correct coupler for the logger and ensure that it hole and rotate counterclockwise until the cap is loosened is seated properly 4 Plug the small end of the USB interface cable into the USB 3 Insert the logger into the coupler following the instructions port in the shuttle If the shuttle has never been connected that came wi
2. function is available even when the batteries are depleted To use the shuttle as a base station 1 Connect the shuttle to the host computer running HOBOware 2 Attach a compatible logger and coupler Momentarily press the coupler s lever 4 The amber LED blinks momentarily then the green LED should glow steadily to indicate that the logger is ready to communicate with HOBOware If the red LED blinks instead the logger was not found Make sure the logger and coupler are aligned and seated properly and that there is no dirt or strong sunlight interfering with communications 5 When finished remove the logger from the coupler The green LED stops glowing when you disconnect the logger or the USB cable Important The Waterproof Shuttle cannot be used as a base station with Pendant logger models UA 001 and UA 003 including rain gauges RG3 and RG3 M with serial numbers less than 988278 These loggers require a BASE U I for communication with the host computer Indicator lights Green OK LED The green OK LED blinks when HOBOware recognizes it as a base station when it finishes reading out and relaunching a logger and when you press the coupler lever to check the shuttle s status see Checking shuttle status in the field for details Momentarily press the coupler lever to stop the blinking The green LED glows steadily when the shuttle is being used as a base station Amber Transfer LED
3. loggers store data in nonvolatile EEPROM memory Onset will if possible recover your data Tune Up Service Onset will examine and retest any HOBO data logger 2006 Onset Computer Corporation All rights reserved Part MAN U DTW 1 Doc 10264 D Onset HOBO and HOBOware are registered trademarks of Onset Computer Corporation Other products and brand names may be trademarks or registered trademarks of their respective owners
4. HOBO Waterproof Shuttle Part U DTW 1 Inside this package Thank you for purchasing a HOBO Waterproof HOBO Waterproof Shuttle Shuttle The HOBO Waterproof Shuttle performs several major functions e USB host cable e Reads out all logger information serial number e Set of couplers deployment number data etc from loggers in the For UA Pendant Part COUPLER2 A nes a aa Sron and stores each For U20 Water Level Part COUPLER2 B ogger s data In a ba For U22 Water Temp Pro v2 Part COUPLER2 C e Nonvolatile memory preserves data even if batteries For UTBI TidbiT v2 Part COUPLER2 D are depleted For U23 HOBO Pro v2 Part COUPLER2 E Relaunches the logger resetting the logger s time to the shuttle s time and synchronizing the logging interval on relaunch e Can be used as an optic to USB base station Although the HOBO Waterproof Shuttle is easy to use Onset strongly recommends that you spend a few minutes reading this manual and trying out the procedures described here before taking the shuttle into the field Doc 10264 D MAN U DTW 1 Onset Computer Corporation Specifications Compatibility All HOBO U Series loggers with optic USB Not compatible with the HOBO U Shuttle U DT 1 Requirements HOBOware 2 2 compatible logger and matching coupler 63 logger readouts of up to 64K each Operating temperature 0 to 50 C 32 to 122 F 20 to 50 C 4 to 122 F Poly
5. The amber Transfer LED blinks when the shuttle is reading out a logger and relaunching it Do not remove the logger when the Transfer light is lit Red Fail LED The red Fail LED blinks whenever the shuttle encounters an error condition Refer to Troubleshooting for details All LEDs All LEDs blink in unison when the shuttle has just been powered up either by installing fresh batteries or if batteries are not installed by connecting to the computer s USB port HOBO Waterproof Shuttle Troubleshooting This section describes problems you may encounter while using the shuttle Shuttle is not recognized by host computer If HOBOware does not recognize the shuttle when you connect it to the computer simply disconnect and reconnect the shuttle Red Fail LED blinks The red Fail LED blinks for 15 minutes or until you press the coupler lever whenever the shuttle encounters an error There are several conditions that might cause an error e Shuttle is full If the red LED blinks when you try to read out a logger check whether all of the banks are full as described in Checking shuttle status in the field Or use HOBOware to check the shuttle s memory Shuttle batteries are low If you cannot read out any loggers at all check the logger s status as described in Checking shuttle status in the field or use HOBOware to check the shuttle s batteries The batteries may simply n
6. any Onset supplied product or equipment be used in or with any nuclear installation or activity or in or with any aeronautical or related application or activity Purchaser will indemnify Onset and hold Onset harmless from any liability or damage whatsoever arising out of the use of the product and or equipment in such manner Returns Please direct all warranty claims and repair requests to place of purchase Before returning a failed unit directly to Onset you must obtain a Return Merchandise Authorization RMA number from Onset You must provide proof that you purchased the Onset product s directly from Onset purchase order number or Onset invoice number Onset will issue an RMA number that is valid for 30 days You must ship the product s properly packaged against further damage to Onset at your expense with the RMA number marked clearly on the outside of the package Onset is not responsible for any package that is returned without a valid RMA number or for the loss of the package by any shipping company Loggers must be clean before they are sent back to Onset or they may be returned to you Repair Policy Products that are returned after the warranty period or are damaged by the customer as specified in the warranty provisions can be returned to Onset with a valid RMA number for evaluation ASAP Repair Policy For an additional charge Onset will expedite the repair of a returned product Data back Service HOBO data
7. are to check the shuttle s battery level available memory and clock You may need to change the batteries or offload the datafiles to the host computer and delete them from the shuttle to free up memory before you can continue reading out loggers Offloading data to the host computer You can offload the data stored in the shuttle even when the batteries are depleted Take the following steps 1 Connect the shuttle to a host computer running HOBOware 2 Follow the instructions in the HOBOware User s Guide to offload the new datafiles or access the Manage Shuttle dialog The Manage Shuttle dialog shows you how many banks are occupied and whether they have already been offloaded and saved to the host computer 3 Offload and save data from the banks of your choice Refer to the HOBOware User s Guide for details on saving datafiles offloaded from the shuttle 4 Review the list of banks and delete any that are no longer needed Make sure the battery level 1s good and change the batteries now if they are weak If you change the batteries in the field the shuttle s clock will stop and the shuttle will not read out loggers Update the shuttle s clock if necessary 5 When finished disconnect the shuttle from the computer and close the center cap securely HOBO Waterproof Shuttle Using the shuttle as a base station You can use the shuttle as a base station for any U Series logger with an optic USB interface This
8. carbonate case EPDM o rings and retaining loop Timebase Error ppm Wateproot fre 20m 6 ed Plot A minute per month at 25 C 77 F see Plot A Logger to shuttle transfer speed Reads out one full 64K logger in about 30 seconds Shuttle to host transfer speed Full shuttle offload 4 MB to host computer in 10 to 20 minutes depending on computer One year or at least 50 complete memory fills typical use 150 g 4 oz 15 2 x 4 8 cm 6 0 x19 inches The CE Marking identifies this product as complying with the relevant directives in the European Union EU aterproof ime accuracy to atteries 2 AA alkaline batteries required for remote operation attery life imensions 2006 Onset Computer Corporation Part MAN U DTW 1 Doc 10264 D HOBO Waterproof Shuttle HOBO Waterproof Shuttle features Coupler Pendant coupler shown Center cap Logger Pendant shown USB port Large cap Coupler lever Preparing to go on location 7 If you have used the shuttle before make sure there are enough banks available to accommodate the loggers you Before using the shuttle for the first time you must launch it plan to read out with HOBOware 2 2 or greater You must also launch any compatible loggers that were last launched with an earlier 8 Disconnect the USB cable from the shuttle and replace the version of HOBOware or have never been launched at all center cap securely 1 Use
9. e clock at relaunch For most accurate results make sure the relaunch host computer s clock is correct before launching the E Ea shuttle If you need to adjust the computer s clock quit taading sang j emy mag iskis HOBOware set the computer s clock then reopen HOBOware and launch the shuttle 10 00 10 15 10 30 10 45 11 00 Important If the logger was launched with multiple logging intervals there will be no synchronizing delay The logger will start immediately with the last defined logging interval 6 When the relaunch has completed the green LED blinks for 15 minutes or until you momentarily press the coupler lever to stop it If the red LED blinks instead there was an error and the logger may have stopped Refer to Troubleshooting in this manual for details 7 Remove the logger from the coupler Checking shuttle status in the field The shuttle s memory has 63 banks One logger readout can be stored in each bank To check the shuttle s memory and batteries in the field remove the logger and press the coupler s lever for at least three seconds When you release the lever the green LED blinks once for each unoccupied bank in the shuttle s memory Press the lever momentarily to stop the blinking If the shuttle s batteries are running low all of the shuttle banks are full or the clock has not been set the red LED blinks Press the lever momentarily to stop the blinking Use HOBOw
10. e expose to extreme heat or mix with other battery types as the batteries may explode or leak Contents of an open or leaking battery can cause chemical burn injuries Replace all used batteries at the same time Recycle or dispose of batteries according to applicable federal state and local regulations HOBO Waterproof Shuttle Service and Support As part of Onset s ongoing efforts to provide 100 customer satisfaction our Continuing Engineering Group constantly monitors and evaluates all of our products and software In the unlikely event any significant defect is found Onset will notify you If you find a defect please e mail us at loggerhelp onsetcomp com HOBO products are easy to use and reliable In the unlikely event that you have a problem with this instrument contact the company where you bought the logger Onset or an Onset Authorized Dealer Before calling you can evaluate and often solve the problem if you write down the events that led to the problem are you doing anything differently and if you visit the Technical Support section of the Onset web site at www onsetcomp com support html When contacting Onset ask for technical support and be prepared to provide the product number and serial number for the logger and software version in question Also completely describe the problem or question The more information you provide the faster and more accurately we will be able to respond Onset Computer Corpora
11. eed to be replaced Compatibility The shuttle cannot read out or relaunch loggers that were last launched from HOBOware prior to version 2 2 You will need to read out these loggers on the host computer and relaunch them in HOBOware 2 2 or greater before you can use them with the shuttle Shuttle clock is not set The shuttle has experienced a power failure that caused the clock to reset You must use HOBOware to offload the files that are already on the shuttle then relaunch the shuttle before you can read out another logger Can t communicate with logger Remove the logger and coupler Inspect them and the shuttle to ensure that all are free of dirt that could block the optic communication sensor Carefully reassemble the shuttle coupler and logger and make sure they are all seated properly Shield the shuttle from strong sunlight if applicable which can interfere with optic communications Other logger problems If you can read out some loggers but not others or if you cannot read out any loggers even with fresh batteries in the shuttle check the loggers in HOBOware Make sure their batteries are at acceptable levels and that there is no corrupted header message Amber Transfer LED stays on without blinking The amber light is magnetically activated when you press the coupler lever If it glows steadily at any other time the magnet in the lever may be too close to the magnetic switch in the shuttle or another
12. ll the rubber loop free of the large cap The large cap cannot be removed while the rubber loop is in place 4 Turn the large cap counter clockwise slightly then pull it off continued 5 Turn the shuttle over and tap it gently The circuit board should slide into your hand 6 Remove the old batteries and install two new ones in the correct orientation Both batteries should be turned the same way with their positive ends facing the USB port on the board When the second battery makes contact all of the shuttle s LEDs will blink in unison 10 HOBO Waterproof Shuttle Put the board back into the case taking care not to bend the communication LEDs Align the circuit board with the runners in the case The USB port should face the open end of the shuttle and the LEDs should show through the window on the label Close the shuttle s case Line up the tabs on the large cap with the slots on the case press gently and turn slightly clockwise until the large cap is closed securely Replace the rubber loop and center cap Tighten the center cap until it is just flush with the large cap or until the O ring is no longer visible Using HOBOware offload any datafiles that are on the shuttle and launch the shuttle before going into the field again The shuttle will not read out and relaunch loggers until the clock has been synchronized A WARNING Do not install batteries backwards recharge put in fir
13. strong magnet may be present Try bending the lever away from the coupler to reduce the magnet s effect LEDs do not function If the LEDs are not functioning at all the batteries may be completely exhausted To test this attach the shuttle to the host computer and check the battery level The shuttle should be 4 able to communicate with the host computer blink its LEDs normally and perform as a base station even when the batteries are missing or depleted Replacing the shuttle s batteries The shuttle s batteries should last about one year or at least 50 complete memory fills in typical conditions When the shuttle s batteries run low 2 2 V or less any logger data that is already in the shuttle will remain safe but the shuttle will not read out another logger until its batteries are replaced To avoid battery problems always check the shuttle s batteries in HOBOware before going into the field and replace them if needed If you cannot replace the bad batteries right away you should remove them as soon as possible to ensure that they do not leak and damage the shuttle To change the shuttle s batteries 1 Work over a clean surface to provide a safe platform for the disassembly 2 Unscrew the center cap on the shuttle If the cap is too tight to loosen by hand insert a screwdriver through the lanyard hole and rotate counterclockwise until the cap is loosened 3 Use the center cap to help you carefully pu
14. t exceed the Purchaser s price of the product s The determination of suitability of products to the specific needs of the Purchaser is solely the Purchaser s responsibility THERE ARE NO WARRANTIES BEYOND THE EXPRESSED WARRANTY OFFERED WITH THIS PRODUCT EXCEPT AS SPECIFICALLY PROVIDED IN THIS DOCUMENT THERE ARE NO OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTIES OF MERCHANTIBILITY OR FITNESS FOR A PARTICULAR PURPOSE NO INFORMATION OR ADVICE GIVEN BY ONSET ITS AGENTS OR EMPLOYEES SHALL CREATE A WARRANTY OR IN ANY WAY INCREASE THE SCOPE OF THE EXPRESSED WARRANTY OFFERED WITH THIS PRODUCT Indemnification Products supplied by Onset are not designed intended or authorized for use as components intended for surgical implant or ingestion into the body or other applications involving life support or for any application in which the failure of the Onset supplied product could create or contribute to a situation where personal injury or death may occur Products supplied by Onset are not designed intended or authorized for use in or with any nuclear installation or activity Products supplied by Onset are not designed intended or authorized for use in any aeronautical or related application Should any Onset supplied product or equipment be used in any application involving surgical implant or ingestion life support or where failure of the product could lead to personal injury or death or should
15. th the coupler to the computer before it may take a few seconds for the new hardware to be detected 4 Momentarily press the coupler lever Readout should begin immediately The amber LED blinks continuously while 5 Follow the instructions in the HOBOware User s Guide to readout and relaunch are in progress Do not remove the access the Manage Shuttle dialog Make sure the battery logger when the amber LED 1s blinking level i h he batteri if th k l Everts ana a nae areca 5 After reading out the logger the shuttle synchronizes the Important If you change the batteries in the field the logger s clock to the shuttle s internal clock and relaunches shuttle s clock will stop and the shuttle will not read out the logger using the description channels to log logging loggers again until you relaunch it in HOBOware interval and other settings that are already in the logger If the logger was launched with multiple logging intervals the final defined logging interval will be used The logger is launched with a slight delay that causes its readings to be synchronized with those of the previous deployment as shown in the following diagram 6 If you are using the shuttle for the first time launch the shuttle as described in the HOBOware User s Guide Launching synchronizes the shuttle s clock to the host computer and initializes the shuttle s header Important The shuttle s clock is used to set the logger s ea
16. tion 470 MacArthur Blvd Bourne MA 02532 Mailing PO Box 3450 Pocasset MA 02559 3450 Phone 1 800 LOGGERS 1 800 564 4377 or 508 759 9500 Fax 508 759 9100 E mail loggerhelp onsetcomp com Internet www onsetcomp com Warranty Onset Computer Corporation Onset warrants to the original end user purchaser for a period of one year from the date of original purchase that the HOBO product s purchased will be free from defect in material and workmanship During the warranty period Onset will at its option either repair or replace products that prove to be defective in material or workmanship This warranty shall terminate and be of no further effect at the time the product is 1 damaged by extraneous cause such as fire water lightning etc or not maintained in accordance with the accompanying documentation 2 modified 3 improperly installed 4 repaired by someone other than Onset or 5 used in a manner or purpose for which the product was not intended THERE ARE NO WARRANTIES BEYOND THE EXPRESSED WARRANTY ABOVE IN NO EVENT SHALL ONSET BE LIABLE FOR LOSS OF PROFITS OR INDIRECT CONSEQUENTIAL INCIDENTAL SPECIAL OR OTHER SIMILAR DAMAGES ARISING OUT OF ANY BREACH OF THIS CONTRACT OR OBLIGATIONS UNDER THIS CONTRACT INCLUDING BREACH OF WARRANTY NEGLIGENCE STRICT LIABILITY OR ANY OTHER LEGAL THEORY Limitation of Liability The Purchaser s sole remedy and the limit of Onset s liability for any loss whatsoever shall no
Download Pdf Manuals
Related Search
Related Contents
Analox aspida Configuration Software User Manual Sony PFM-50C1 50 in. Plasma HDTV Hoja de datos técnicos Herramientas de montaje Copyright © All rights reserved.