1.    The ProCair 3000 Alternating Mattress  Overlay System consists of the following   e Control Unit   e Mattress Overlay   e Power Cord   e User Guide   e Carry Bag     The Control Unit provides the air supply  to the Mattress Overlay  It is controlled   by a simple analogue interface with a dial  to control comfort settings and a visible  alarm to warn of system low pressure or  power failure     On the left side of the Control Unit are two  quick release air connectors for connecting  the overlay hoses        The mains supply to the Control Unit can  be easily disconnected and is designed to  detach if pulled too firmly   protecting the  internal wiring of the unit     The Mattress Overlay comprises 21 high  density cells whuch alternate over a 10 12  minute cycle  This provides regular periods  of pressure reduction to aid blood and  lymphatic flow to vulnerable tissue        Installation    Unpacking  amp  Inspection    NOTE  It is recommended that all packing  materials and instructions be kept in   the carry bag provided  in the event the  product has to be shipped to an approved  Medistore Service Centre     Carefully remove the Control Unit  Mattress  Overlay and accessories from the shipping  cartons  Inspect all items for any damage  that may have occurred during shipping   Any damage or missing components should  be reported to a Medistore Service Centre  as soon as possible     The box should contain the  following items     e Completely assembled  ProCair 3000 Al
2.   unit to prevent scratches      4  Carefully pry off the cover of the air  filter  remove and discard the filter  material and fit new  there may be a  small locking screw     use a small Philips  head to remove this first      5  Refit the air filter cover to the Control  Unit  The Control Unit is now ready for  re connection     NOTE  Good filter maintenance is critical to  maintain your system in optimal operating  condition  Failure to keep the filters   clean will result in system downtime and  increase repair costs  It is recommended  that the air filter be replaced annually    Air filters are available from a Medistore  Service Centre  Preventative inspection and  calibration is not required     Fuse Replacement  1  Ensure the Control Unit is deactivated     2  Remove the power cord from the  electrical socket on the side of the  Control Unit     3  Using a small flat head screwdriver  insert into the grove and turn  anti clockwise    turn      4  Remove the    blown    fuse from the fuse  holder clip and discard     5  Insert a new fuse into the plug  Push  against the force of the spring and turn  clockwise with the screwdriver a 1   4 turn        Technical Specifications    Cycle Control  Purpose designed distributor valve supplying operating air to the  inflatable cells    Cycle Time 10 12 minutes    Supply Voltage 230V  50Hz 120V  60Hz  0 2A for Control Unit    0 01   6A for power down alarm   availability of supply monitored by in line power down alarm  which pr
3.  for UL 2601 1 standard for Medical  Electrical Equipment     Li  medistore    Medistore Australia Pty Ltd  ABN 51 106 477 356  PO Box 347 Lane Cove NSW 1595 Australia  Freecall  1300 882 194 Freefax  1300 882 197    www medistore com au       May15    
4. User Guide  ProCair 3000          Important Notice    Before operating this medical equipment   it is important to read this User Guide and  understand the operating instructions and  safety precautions  Failure to do so could  result in patient injury and or damage to  the product        Page        Safety Precautions 4 5  Product Overview 6  Installation 7  Unpacking  amp  Inspection 7  System Installation 8  Control Unit Installation 8  Operation 9  Control Unit Panel 9  Removal and Transport Function 10  Troubleshooting 11  Maintenance and Cleaning 12 13  Technical Specifications 14    Warranty Information 15          Safety Precautions    The purpose of safety precautions are to  attract your attention to possible dangers   The safety symbols and the explanations  with them  require your careful attention  and understanding     The safety warnings by themselves do not  eliminate any danger  The instructions or  warnings they give are not substitutes for  proper accident prevention measures     The following symbols may appear in this  User Guide  on the Control Unit  or on its  accessories  Some of the symbols represent  standards and compliances associated with  its use     A SAFETY ALERT SYMBOL  Indicates  caution or warning     WARNING  Failure to obey a safety  warning can result in serious injury   to yourself or to others  Always follow  the safety precautions to reduce the  risk of fire  electric shock and   personal injury      A  DOUBLE INSULATION   Double insul
5. ation is a safety concept of  electrical equipment which eliminates   the need for earth grounding  Whenever  there is electric current in the Control Unit  there are two complete sets of insulation  to protect the user  All exposed metal parts  are isolated from the internal metal motor  components with protecting insulation     A WARNING  The double insulated  system is intended to protect the user from  shock resulting from a break in the Control  Unit s internal wiring  Observe all normal  safety precautions related to avoiding  electrical shock     A WARNING  Servicing a Control Unit  with double insulation requires extreme  care and knowledge of the system and  should only be performed by a qualified  service technician  For service we suggest  you return the Control Unit to your nearest  Medistore Authorised Service Centre for  repair  When servicing  use only identical  replacement parts        OPERATING INSTRUCTIONS    A WARNING  Do not attempt to operate  this product until you have thoroughly  read and completely understand the safety  rules  etc contained in this User Guide   Failure to comply can result in accidents  involving fire  electric shock or serious  personal injury  Save this User Guide   and review frequently for continual safe  operation and for instructing others who  may use this product     A Maximum recommended patient  weight for this system is 120kg     A Only the Control Unit and   Mattress combination as recommended   by Medistore should be us
6. ed  otherwise  the correct function of this product cannot  be guaranteed        Safety Precautions  In General    A WARNING  This device is not suitable  for use in the presence of a flammable  anaesthetic mixture with air  or in the  presence of a flammable anaesthetic  mixture with oxygen or nitrous oxide     A WARNING  Bed frames used with the  systems can vary greatly depending on the  specific health care setting  ie hospitals   nursing homes  home care  etc   It is the  responsibility of the caregiver to take the  necessary precautions to ensure the safety  of the patient  This includes  but is not  limited to  the appropriate use of side rails  to prevent falls and or patient entrapment     IMPORTANT  Minimise articles between  the system surface and patient  and secure  bed sheets loosely so as not to affect the  alternating cell movement     A WARNING  Significant risks of  reciprocal interference may be posed by  the presence of the system during specific  investigations or treatments  Potential  electromagnetic or other interference  between the system and other devices  may occur  If interference is suspected   move equipment from sensitive devices or  contact Medistore     Control Unit    A Only plug into a grounded power  receptacle and use the power cord  supplied with the system       Exposure of the electronic  Control Unit to any liquid while it is  plugged in could result in a severe  electrical hazard     A    A    Medistore recommends placing   the cord 
7. est if the patient is not inclined but  supine  Turn the comfort control dial  up to maximum pressure and wait 10  minutes for the setting to take effect       Ensure that the overlay is not currently    undergoing an alternation  If not  then  turn off Control Unit and disconnect  both hoses  Immediately connect the air  hoses together       Once the air hoses are connected    together  the air pressure will average  out between the fully inflated cells  and the deflated cells  providing an  immersive surface for pressure relief  during patient transportation     Bottoming Out Test    1   2     Unzip one side of the top cover     Slide your hand underneath the patient   feeling for a deflated cell under the  patient s lower spine     your hand should  easily slide between patient and base       For optimal pressure  patient should be    immersed 30 50  into fully inflated cells       If patient is not appropriately immersed     decrease pressure setting and repeat  test every 10 12 minutes  until optimal  pressure is reached        Troubleshooting  Problem    Control unit does  not operate     Cause    The control unit may not be  attached to a power source    or a fuse may need replacing     Solution    Check the control unit is  connected to an electrical  power outlet     Check the control unit is  switched on     Check the mains plug fuse   3 AMP  and both control unit  fuses  1 AMP         Control unit  operational but red  Alarm LED illuminates     Mattress Overlay fai
8. hen lowering the pressure  setting  ensure the patient is not     bottoming out     not sufficiently supported  by the air cells and therefore coming in  contact with mattress   You should be   able to slide your hand between the   cells underneath the patient s sacral area   bottom  along a cell that is deflated   Ensure patient is in a supine position for  this assessment     Establishing Pressure   inclined patient     When moving the patient to a more  upright position  pressure may need to  be increased  approximately 20   to  provide added support and to avoid     bottoming out        NOTE  It is important to return the original  pressure setting when the patient returns  to the supine position     IMPORTANT  Wait a minimum of 10  minutes between any adjustment and  patient assessment as it may take a full  cycle for the system to adjust to any  new setting        Operation  System Removal    Overlay Removal    1  Switch off the Control Unit   If  applicable reset the alarm indication  on the power down alarm  Switch off  the mains supply and disconnect  the power down alarm from the  electrical outlet      2  Disconnect the air hoses from the  Control Unit to release air from  the system     4  Remove the Control Unit   power down  alarm if applicable  and power cord  from the bed and place on the overlay     5  Roll up the Mattress Overlay and  remove from the bed     Transport Function    Overlay Transport Function    1     When preparing for transport mode  it  is b
9. ls  to inflate     The alarm illuminates when  Mattress Overlay pressure  fails to reach or drops below  the accepted minimum level   this can be an indication of  an air leak in the system     Check both air hoses are  properly attached to the  control unit     Check all hoses along the  inside of the overlay     each  should be firmly connected     Check each air cell is securely  attached to the connecting  air pipe     Check all cells  pipes and hoses  for any air leakage     Check that the air filter cover  is correctly secured and the air  filter is clean     A Do not try to open the  control unit  Opening the unit  could cause personal injury or  equipment damage        Patient is sinking or     bottoming out     while lying flat on  the mattress overlay     The pressure may be set too  low for the patient s weight     Increase pressure setting      usually an incremental increase  is sufficient  however wait  10 12 minutes  one full cycle   before checking the increase  was sufficient     Check effective system  performance by conducting  a    bottoming out    test as  described on page 10     If the problem cannot be resolved  please contact a Medistore Service Centre for    further support        Maintenance  amp  Cleaning  Infection Control and routine cleaning  must be carried out in accordance with  your local Infection Control Policy    It is suggested that all disinfection   be done with a high grade   disinfectant in accordance with  manufacturer s instruction
10. ovides failure indication     Fuse Rating Mains Plug   3A  Control Unit   1A   Overlay Dimensions  amp  Weight Control Unit Dimensions  amp  Weight  Length 2000mm Height 115mm  Width 880mm Width 135mm  Height 90mm Depth 270mm  Weight 8 4kgs Weight 2 1kgs  Power Rating 12VA   Cells 21 cells   Cell material 0 15mm TPU film laminated on 210 Denier Nylon fabric  Base material Nylon fabric 420 Denier with a 0 1mm TPU coating  Cover Material PU coated polyester   Connection Quick release connectors   Standards CE  UL   Mode of Operation Continuous    Environment    Air humidity 30  to 75   Ambient temperature 0  C to 70  C  Altitude  lt  2000m    All product specifications are subject to change without notice        Warranty Information    Our goods come with guarantees that  cannot be excluded under the Australian  Consumer Law  You are entitled to a  replacement or refund for a major failure  and for compensation for any other  reasonably foreseeable loss or damage   You are also entitled to have the goods  repaired or replaced if the goods fail to be  of acceptable quality and the failure does  not amount to a major failure     Medistore warrants each of its products  to perform in accordance with established  specifications for the following time  periods  starting from the date the  product is shipped from Medistore     Control Units 2 years    Soft Goods 2 years    During the warranty period  Medistore   through its distributors  will repair or  replace at no charge  any prod
11. own with disinfectant  or a solution  of Sodium Hypochlorite or similar up to  10 000 ppm available chlorine   ensuring  that all surfaces come in contact with the  disinfectant  Rinse off well with a clean  damp cloth and air dry     Air Cell Assembly    A CAUTION  DO NOT machine wash or  dry the air cells  The air cell assembly does  not routinely need to be cleaned   or disinfected     IMPORTANT  DO NOT disassemble the  overlay  unless cleaning is required    If cleaning or disinfecting is required   DO NOT disconnect the hoses from the  air cells        Maintenance  amp  Cleaning  Control unit    A WARNING  Ensure the Control Unit  is disconnected from the mains electricity  supply before cleaning     A WARNING  DO NOT spray disinfectant  directly on to the Control Unit  or immerse  in any type of liquid  This could result in a  severe electrical hazard as this equipment  has no protection against ingress of water     A WARNING  This equipment is not  suitable for use in the presence of a  flammable anesthetic mixture with air or  with oxygen or nitrous oxide     External cleaning    Wipe down Control Unit with warm water  containing detergent  or with a solution  of Sodium Hypochlorite or similar   dry  thoroughly before use     Air Filter Replacement    1  Switch off the power supply to the  Control Unit     2  Disconnect the power cord and  air hoses     3  Place the Control Unit on a flat  surface with its back panel uppermost   NOTE  be sure to place soft cloth under
12. s     A WARNING  Do not use high  temperature autoclave  or use  Phenolic based products for cleaning     IMPORTANT  It is recommended the  system is cleaned between patients  and approximately every two weeks if  in constant use     Overlay Top Cover  Washing and Disinfecting    If there are visible signs of body fluids  and or substances present  top covers  should be washed  Top covers can be  machine washed  up to 70  C  using  disinfectant according to manufacturer s  instructions  To determine the amount of  disinfectant to use  determine the amount  of water in the washer and then follow the  manufacturers    instructions for dilution   Soak the Top Cover in the disinfectant  during the wash cycle  Rinse well in clean  water and dry thoroughly before use     A CAUTION  Do NOT dry the Top Cover  using the HEAT cycle or a dryer  Air dry or  select a low or NON heat dry cycle     Washing Alternative    If there are no visible signs of body fluids  and or substances on the top cover  it can  be sanitised     1  Apply an intermediate level disinfectant   or a solution of Sodium Hypochlorite  or similar up to 10 000 ppm available  chlorine  to the Top Cover upper  surface either by spraying or by  hand application     2  Ensure the surface is completely covered  with the disinfectant and remains in  contact with the surface according to  manufacturer s instructions     3  Remove disinfectant and  rinse thoroughly     4  Allow to air dry before use     Outside Shell    Wipe d
13. ternating  Mattress Overlay System    e Carry Bag  e Analogue Control Unit  e Medical Grade Power Cord    e User Guide     The Mattress Overlay is treated as an  applied part           Installation  System Installation    The following describes the procedures to  follow upon setting up the system for the  first time     a  Remove all covers and sheets from the  existing mattress                  b  Position overlay  with its top cover  on  top of existing mattress and position  hose ends at the foot or head end of bed  for preferred Control Unit positioning              c  Secure end straps completely under  the mattress  Ensure end straps are not  twisted or pulled towards the corners of  the mattress     ae           d  Confirm there are no sharp objects in  the immediate area which may risk  damage to the overlay mattress     IMPORTANT  Make sure that the attachment  of the overlay does not interfere with the  movement or operation of the bed     These systems are designed to operate in  a controlled environment  which is   free from extreme temperatures  high  humidity and or excessive amounts of  airborne particles  such as dust and smoke     Control Unit Installation    a        b     c     Position Control Unit by its hanging  hooks over foot board of the bed or on  the floor under the bed        Attach two inflation air hoses to Control  Unit with the press on quick release  connectors  Ensure air hoses do not kink  between bed frame and Control Unit        Position 
14. the power cord through the  cord retention loops along the side of  the mattress base and insert power cord  into Control Unit  then into a grounded  240 V AC 50 Hz electrical outlet     d  Switch on the power and turn the     lt     e    f     pressure dial up to maximum to  commence inflation     Allow up to 40  minutes for  full inflation        Once ready  the  second LED as well  as the alternating LED will illuminate  to indicate that the system is ready for  use  system automatically defaults to  alternation mode after startup      Once the Mattress Overlay is fully  inflated  bedding can be replaced   Secure sheets loosely enough to  ensure they do not interfere with  cell alternation     g  Transfer the patient to the Mattress    Overlay and perform a    bottoming  out    test to ensure that the patient is  properly suspended  Refer to page 10  for more details        Operation  Basic Features    Control Panel    The ProCair 3000 Analogue Control Unit  front panel is shown in the diagram below           Status Lamps Power Switch    Analogue Comfort Control Dial    Overlay Function  Establishing pressure  supine patient     Once the patient is in place  adjust the dial   for the best comfort setting that combines   the required pressure  support and comfort   refer to patient weight guide labelling on   the side of the Control Unit      After approximately 20 minutes  re assess  whether the patient is comfortable and the  unit is functioning correctly     IMPORTANT  W
15. ucts that  are not performing in accordance with  established specifications  unless the  problem failure is due to     e customer damage  negligence  and or misuse    e unauthorised repairs     Items not covered under warranty include   but are not limited to  stains  punctures   cuts  damages to electrical cords  rips or  tears  dents and or lost missing parts     All products returned for warranty repairs  must be assigned a Return Authorisation  Number prior to return  Returns should  include information describing the problem  and or requested repair and be sent to   a Medistore Service Centre by prepaid  transportation  Medistore will return the  repaired replaced product at no charge  Warranty repairs do not extend beyond the  length of the warranty period     Neither Medistore  nor its distributors   officers  directors  employees and agents  shall be liable for consequential or other  damages  including but not limited to  personal injury  loss  or any other expense   directly or indirectly arising from the use of  its products  The sole remedy for breach of  the limited warranty granted herein shall  be repair of the Mattress or Control Unit or  related parts     Medistore has a policy of continuous  product improvement and reserves  the right to amend specifications  presented in this brochure     This product carries the CE mark  in accordance with EC Directive on  Medical Devices  93142IE EC       Control Number 95CG  Classified  by Underwriter s Laboratory Inc  
16. under the bed frame and  attaching it to an electrical outlet by  the head of the bed     The power cord to the Control Unit  should be positioned to avoid a  tripping hazard and or damage to  the cord     Do not place any objects or items   such as blankets  on or over the  Control Unit     The electronic Control Unit is a  precision electronic product  Use  care when handling or transporting   Dropping or other sudden impacts  may result in damage to the unit   Before attaching the Control Unit  to the floor or the foot board of  the bed  ensure they are sufficiently  robust and free of damage     WARNING  Do not open the   Control Unit     risk of electrical shock   Do not attempt to repair or service  the Control Unit  If parts need to be  repaired  circuit diagrams and parts  lists can be provided by Medistore to  suitably qualified service personnel     WARNING  Only use fuses that   have the same specified rating  see  Technical Specifications   Using fuses  with higher ratings could result in  damage and or injury     NOTE  No special skills  training or  knowledge are required to operate the  Control Unit        Product Overview    ProCair 3000 Alternating Mattress Overlay System    The ProCair 3000 is an Alternating Mattress  Overlay System providing pressure  application and relief for patients with    or vulnerable to  pressure ulcers  It is  designed to be used on top of an existing  single bed mattress and can be used on  both standard and profiling bed frames  
    