        Operating instructions School balance
         Contents
1. IN    KERN  amp  Sohn GmbH    WAAGEN   GEWICHTE   BALANCES   WEIGHTS Ziegelei 1 Tel   49  0 7433  9933 0  D 72336 Balingen Fax   49  0 7433 9933 149  E Mail  info kern sohn com Internet  www kern sohn com    Operating instructions  School balance    KERN 442    Version 1 5  07 2003  GB       442 BA e 0315    KERN 442   Version 1 5 07 2003  Operating instructions  School balance    Tabel of contents     1 Technical data  2 Declaration of conformty  3 Fundamental information  general   3 1 Intended use  3 2 Inappropriate use  3 3 Guarantee  3 4 Monitoring the test substances  d Fundamental safety information  4 1 Observe the information in the operating instructions  4 2 Staff training  5 Transport and storage  5 1 Acceptance check  5 2 Packaging  6 Unpacking  installation and commissioning  6 1 Place of installation  place of use    6 2 Unpacking  6 2 1 Installation       6 2 2 List of items supplied       6 3 Mains supply   6 4 Battery Operation   6 5 Initial start up   6 6 Adjustment   6 7 Adjusting   7 Operation   7 1 Overview of display   7 2 Operating elements  7 2 1 Weighing       ON  7 2 2 Taring  Late  key   7 2 3 Plus   Minus Weighings  7 2 4 Units             8 Maintenance  upkeep  disposal  8 1 Cleaning  8 2 Maintenance  upkeep  8 3 Disposal   9 Troubleshooting    2 15 08 200713442 BA e 0315    OO oo Oo oO Jr OO Om OO Om OO OO Om P Ga    1 Technical data                                     KERN 442 43N 442 51N  Readout 0 19 19  Weighing range 400g 4 000 g  Taring ra
2. d to the monitoring of balance test substances and the test  weights required for this  Test weights and balances can be adjusted quickly and at  a reasonable price in KERN   s accredited DKD calibration laboratory  return to na   tional normal      4 Fundamental safety information  4 1 Observe the information in the operating instructions    Please read the operating instructions carefully before erecting and commissioning   even if you already have experience with KERN balances     4 2 Staff training    The device may only be operated and looked after by trained members of staff     5 Transport and storage  5 1 Acceptance check    Please check the packaging immediately upon delivery and the device during un   packing for any visible signs of external damage     5 2 Packaging    Please retain all parts of the original packaging in case it should be necessary to re   turn items at any time    Only the original packaging should be used for return consignments    Before despatch  disconnect all attached cables and loose movable parts    Apply any intended transport security devices  Secure all parts  e g  glass windshield   weighing plate  power unit etc   to prevent slipping and damage     6 15 08 200713442 BA e 0315    6 Unpacking  installation and commissioning  6 1 Place of installation  place of use    The balance is constructed in such a way that reliable weighing results can be  achieved under normal application conditions    By selecting the correct location for your ba
3. estamos en la presente que el producto al que se refiere esta declaraci  n est      a de  acuerdo con las normas siguientes    Italiano Dichiariamo con cid che il prodotto al quale la presente dichiarazione si riferisce    confor   me alle norme di seguito citate     Electronic Scale  KERN 442 N             Mark applied EU Directive Standards  89 336EEC EMC EN 61000 6 3   2001  C     EN 61000 3 2   2000  EN 61000 3 3   1995 A1   2001  EN 61000 6 1   2001             Date  01  July 2003 Signature           Gottl  KERN  amp  Sohn GmbH  Management       Gottl  KERN  amp  Sohn GmbH  Ziegelei 1  D 72336 Balingen  Tel   49  0 7433 9933 0 Fax  49  0 7433 9933 149    4 15 08 200713442 BA e 0315    3 Fundamental information  general     3 1 Intended use    The balance you have acquired serves to determine the weighing value of the mate   rial to be weighed  It is intended to be used as a    non automatic    balance  i e  the  material to be weighed is manually and carefully placed in the centre of the weighing  plate  The weighing value can be read off after a stable weighing value has been ob   tained     3 2 Inappropriate use    Do not use the balance for dynamic weighing  In the event that small quantities are  removed or added to the material to be weighed  incorrect weighing results can be  displayed due to the    stability compensation    in the balance   Example  Slowly drain   ing fluids from a container on the balance     Do not leave a permanent load on the weighing plate  T
4. ge the Units with the key   Press an hold the key   few seconds later change the unit     442 BA e 0315 11    8 Maintenance  upkeep  disposal    8 1 Cleaning   Please disconnect the device from the operating voltage before cleaning    Only use a cloth dampened with mild suds and not aggressive cleaning agents  sol   vents or similar   Please ensure that fluids are not able to get into the device and rub  off using a clean  soft cloth    Loose sample residue powder can be removed carefully using a brush or hand vac   uum cleaner    Remove any spilt material to be weighed immediately     8 2 Maintenance  upkeep    The device may only be opened by trained service engineers authorised by KERN   Disconnect from the mains supply before opening     8 3 Disposal    The operating company shall dispose of the packaging and the device in compliance  with the valid national or regional law of the operating location     12 15 08 200713442 BA e 0315    9 Troubleshooting    The balance should be switched off for a short time following an interruption in the  programme sequence and disconnected from the mains supply  It is then necessary  to repeat the weighing process from the beginning     Help   Interruption  Weight display is not illuminated  e    The weight display changes continu  e  ally    The weighing result is obviously in  e  correct    Possible cause    The balance is not switched on     The mains supply connection has been  interrupted  mains cable not plugged  in faulty      Po
5. his can damage the measur   ing equipment    Be sure to avoid impact shock and overloading the balance in excess of the pre   scribed maximum load rating  max    minus any possible tare weight that is already  present  This could cause damage to the balance    Never operate the balance in hazardous locations  The series design is not explo   sion proof    Structural alterations may not be made to the balance  This can lead to incorrect  weighing results  faults concerning safety regulations as well as to destruction of the  balance    The balance may only be used in compliance with the described guidelines  Varying  areas of application planned use must be approved by KERN in writing     3 3 Guarantee    The guarantee is not valid following     non observation of our guidelines in the operating instructions    use outside the described applications     alteration to or opening of the device     mechanical damage and damage caused by media  liquids     natural wear and tear     inappropriate erection or electric installation     overloading of the measuring equipment    442 BA e 0315 5    3 4 Monitoring the test substances    The metrology features of the balance and any possible available adjusting weight  must be checked at regular intervals within the scope of quality assurance  For this  purpose  the answerable user must define a suitable interval as well as the nature  and scope of this check  Information is available on KERN   s home page  www kern   sohn com  with regar
6. ht at the bottom of the display    If the object should be heavier than the weighing range allowance  the symbol    E      overload  will appear on the display     7 2 2 Taring 6 key   Switch the balance on by pressing the key  then wait for the    0    indication     Place the tare cup on the weighing pan and press the key  Display again shows     0     Now the weight of the tare cup is memorised internally     By pressing the key after a weighing procedure     0    will appear on the display  again    The taring procedure can be repeated continuously  for instance when mixing several  components    The limit is reached when the full weighing range is used    After having removed the tare cup the total weight will appear as a negative value     10 15 08 200713442 BA e 0315    7 2 3 Plus   Minus Weighings  For instance to control piece weights  filling process control etc    Switch the balance on by pressing the key  then wait until    O    is indicated     Place rated weight on the weighing pan and tare on    O    by pressing the key   Remove rated weight    Place the objects on the weighing pan successively  the balance will show any devia   tion from the rated weight    According to the same procedure packages with the same weight  related to a rated  weight  can be produced     Return to the weighing mode by pressing the key     7 2 4 Units    Different foreign weighing units are integrated into the various balance models  de   tails capture   Technical data          Chan
7. ighing pan   Battery or mains adapter  Operating instructions    6 3 Mains supply    Electric power supply is by means of the external mains supply circuit  The printed  voltage level must comply with the local voltage    Only use original KERN mains supply circuits  The use of other makes is subject to  approval by Kern     6 4 Battery Operation    Remove the battery cover from the bottom of the balance    Connect 9 V block batteries    Re insert the battery cover    Battery conservation through automatic power off 1   2 minutes after the last weigh   ing operation     When the battery power is used up the display will show    LO     Press the key and  change the batteries at once    When the balance is not in use for a longer period of time remove batteries and keep  them separately  Leakage of battery liquid might damage the balance     6 5 Initial start up    A warm up time of 1 minutes stabilises the measured values after switching on   The accuracy of the balance depends on the local acceleration of the fall   Please be sure to observe the information in the chapter on adjusting     8 15 08 200713442 BA e 0315    6 6 Adjustment    As the acceleration value due to gravity is not the same at every location on earth   each balance must be coordinated     in compliance with the underlying physical  weighing principle   to the existing acceleration due to gravity at its place of location    only if the balance has not already been adjusted to the location in the factory   Th
8. is  adjustment process must be carried out during the initial start up  after change in lo   cation and variation of surrounding temperature  It is also recommendable to adjust  the balance periodically during weighing operation in order to obtain exact measured  values     6 7 Adjusting    Using a precision weight  the accuracy of the balance can be checked at any time  and adjusted     Adjustment procedure     Check that the surrounding conditions are stable    A short warm up time of about 1 minutes are recommended for stabilisation    Press key approx  3 seconds   CAL 0    and then the exact weight of the adjusting  weight will appear the display     Place the adjusting weight on the centre of the weighing plate   Shortly after    CAL F    appears  and then the balance switched off     In case of an adjusting error or a wrong adjusting weight     E    appears in the display   and then turn off     442 BA e 0315 9    7 Operation    7 1 Overview of display       7 2 Operating elements    7 2 1 Weighing  Switch the balance on by pressing the key     The balance will show    8888  for approx  3 seconds and then change to    0     Now it is  ready for use     Important  If the display flashes or does not show    0    press the key     Only now     place object on the weighing pan  Make sure that the weighing object  does not stripe or touch the housing or base    Now the weight will be indicated  After a successful stability control a small triangle  will appear on the far rig
9. lance  you will be able to work quickly and  precisely     Therefore please observe the following at the place of installation     Place the balance on a firm  level surface     Avoid extreme heat as well as temperature fluctuation caused by installing  next to a radiator or in the direct sunlight     Protect the balance against direct draughts due to open windows and doors   Avoid jarring during weighing   Protect the balance against high humidity  vapours and dust     Do not expose the device to extreme dampness for longer periods of time  In   admissible bedewing  condensation of air moisture on the device  can occur if  a cold device is taken into a significantly warmer environment  In this case   please acclimatise the device for approx  2 hours at room temperature after it  has been disconnected from the mains     Avoid static charging of the material to be weighed  weighing container and  windshield     Major display deviations  incorrect weighing results  are possible if electromagnetic  fields occur as well as due to static charging  currents and instable power supply  It is  then necessary to change the location     442 BA e 0315 7    6 2 Unpacking    Carefully remove the balance from its packaging  remove the plastic wrapping and  position the balance in its intended working location     6 2 1 Installation    Install the balance in such a fashion that the weighing plate is absolutely horizonta    6 2 2 List of items supplied    Standard accessories     Balance   We
10. nge  subtractive  4009 4 000 g  Reproducibility 0 19 19  Linearity  0 29  29  Adjusting weight  not included  Adjust  200g  M2  20009  M2   ment tolerance class   Stabilisation time  typical  3 sec    Units g  tol  tlt g tol  tlt       Optimal temp  of operation      10  C       30  C       Humidity    max  80    non condensing     Casing  W x D x H  mm    145 x 185 x 55       Weighing pan mm       125          Weight kg  net        0 5       442 BA e 0315       2 Declaration of conformty    K E Fe N    KERN  amp  Sohn GmbH  D 72322 Balingen Frommern Tel  0049  0 7433  9933 0   A    WAAGEN   GEWICHTE   BALANCES   WEIGHTS    Weg Postfach 4052 Fax  0049  017433 9933 149  E Mail  info  kern sohn de Internet  www kern sohn de    Konformitatserklarung    Declaration of conformity for apparatus with CE mark  Konformitatserklarung fur Gerate mit CE Zeichen  D  claration de conformit   pour appareils portant la marque CE  Declaraci  n de conformidad para aparatos con disitintivo CE  Dichiarazione di cofnromita per apparecchi contrassegnati con la marcatura CE    English We hereby declare that the product to which this declaration refers conforms with the fol   lowing standards     Deutsch Wir erklaren hiermit  daB das Produkt  auf das sich diese Erklarung bezieht  mit den nach   stehenden Normen   bereinstimmt     Francais Nous d  clarons avec cela responsabilit   que le produit  auquel se rapporte la pr  sente  d  claration  est conforme aux normes cit  es ci apres     Espanol Manif
11. wer supply interrupted       The batteries are wrongly inserted  the  batteries are empty    No batteries are attached    Draught air movement    Table floor vibrations    The weighing plate is in contact with for   eign matter     Electromagnetic fields   static charging   choose different location switch off inter   fering device if possible     The balance display is not set to zero    Adjustment is no longer correct   Great fluctuations in temperature     Electromagnetic fields   static charging   choose different location switch off inter   fering device if possible     Switch the balance off if other error messages should appear and then switch on  again  Contact the manufacturer if the error message does not disappear     442 BA e 0315    13    
    