User Manual - HeartSine Samaritan Defibrillators

image

Contents

1. 22 swap it quickly Order a new Pad Pak as soon as possible Memory full warning If the device plays the message warning memory full then the memory can record no further ECG data or events However the device can still analyse and deliver a shock if required If you hear this message contact HeartSine Technologies technical support Audible warnings If the device emits 3 beeps rapidly when turned off it has sensed that the ambient temperature is outside of the specified operating range This beeping could also occur during the weekly self test During use if the status indicator changes from green to red and the device starts to beep there is insufficient battery capacity to deliver a shock The device will continue to analyse the patient s heart rhythm and advise when CPR is needed Device service required If the device plays the message device service required then it has detected a fault Contact your authorised distributor or HeartSine directly for further instruction Warning If you hear this message during use seek an alternative defibrillator immediately Sources of support If you have completed the troubleshooting steps above and you find the device is still not working correctly contact your authorised distributor or HeartSine Technologies Technical Support at support HeartSine com Warning No modification of A this equipment is allowed Warranty exclusion Hear
2. 1 Category 7 Pad Pak and Pediatric Pad Pak Shelf Life Weight Battery type 24 Check expiration date 0 2 kg 0 44 Ibs Meets ISO 60601 2 4 and AAMI DF80 2003 Patient analysis system Method Sensitivity Specificity User Interface Visual prompts Audible prompts Languages Controls Evaluates patient s ECG signal quality electrode contact integrity and patient impedance to determine if defibrillation is required Meets ISO 60601 2 4 and AAMI DF80 2003 Attach Pads Stand Clear Perform CPR Shock now Self Test Pass Ready State Extensive voice prompts guide the user through the operation sequence Contact your HeartSine authorised distributor Two buttons On Off and Shock Defibrillator performance Times to shock delivery fresh battery or after 6 shocks Charging time Following CPR Impedance range Battery Battery type Battery capacity Standby life Typically 150 J in lt 8 sec 200 J in lt 12 sec Typically 8 seconds 25 O to 230 Q Disposable single use combined battery and defibrillation electrode cartridge lithium manganese dioxide LiMnO2 18V gt 60 shocks at 200 J or 6 hours of continuous monitoring See the expiry date on the product 25 Technical Data Electrodes Type Placement Active area Cable length Shelf life Therapeutic shock Waveform Energy Event recording Type Memory Review 26 Single use pre attached combine
3. attack and SCA are not the same though sometimes a heart attack can lead to an SCA If you are experiencing symptoms of a heart attack chest pain pressure shortness of breath tight feeling in the chest or elsewhere in the body seek emergency medical attention immediately Ventricular fibrillation The normal electrical rhythm by which the heart muscle contracts to create blood flow around the body is known as normal sinus rhythm NSR Ventricular fibrillation VF caused by chaotic electrical signals in the heart is often the cause of SCA In victims of SCA it is possible to re establish normal sinus rhythm by means of an electric shock across the heart This treatment is called defibrillation Introduction CPR Quality When providing cardiopulmonary resuscitation CPR treatment to a victim of sudden cardiac arrest it is vital the chest compressions are of a good quality If the quality of the CPR provided is good the chances of successfully resuscitating a patient are greatly increased Research has demonstrated that non professional responders regularly provide ineffective CPR due to inexperience As a response to this problem HeartSine has developed the SAM 500P with CPR Advisor The SAM 500P with CPR Advisor can provide feedback to rescuers on the effectiveness of the cardiopulmonary resuscitation CPR they are providing to the victim The SAM 500P uses ICG measurements to analyse the effectivene
4. 5 860 8100 Toll Free 866 478 7463 Fax 215 860 8192 EMEA ASP HeartSine Technologies 203 Airport Road West Belfast Northern Ireland BT3 9ED Tel 44 0 28 9093 9400 Fax 44 0 28 9093 9401 CE 0120 H017 019 226 0 English UK
5. D with CPR Advisor SAM 500P c Over the frequency range 150 kHz to 80 MHz field strength should be less than V1 V m 36 Appendix A Voice prompts Adult Patient Child Patient If a shock is not required Call for Medical Assistance No shock advised Remove Clothing from Patient s Chest Begin CPR to Expose Bare Skin It ls Safe to Touch the Patient Pull Green Tab to Remove Pads Place overlapping hands on Peel Pads from Liner centre of chest Apply Pads to Patient s Bare Chest as Press down in time with the Shown in Picture metronome Press Pads Firmly to Patient s Remain calm Bare Skin Push Harder or Push Faster or Assessing Heart Rhythm Do Not Push Slower or Good Compressions Touch the Patient 37 Appendix A If a shock is required Stand Clear of Patient Shock Advised Stand Clear of Patient Press the Orange Shock Button Now Shock Delivered Begin CPR It ls Safe to Touch the Patient Place overlapping hands on centre of chest Press down in time with the metronome Remain calm Push Harder or Push Faster or Push Slower or Good Compressions 39 Authorised Distributor www heartsine com info heartsine com US Americas HeartSine Technologies Inc 121 Friends Lane Suite 400 Newtown PA 18940 Tel 21
6. HeartSine samaritan PAD with CPR Advisor SAM 500P User Manual Contents Contents Warnings and Cautions Introduction The HeartSine Samaritan PAD with CPR Advisor SAM 500P Sudden Cardiac Arrest SCA Ventricular fibrillation CPR Quality Impedance cardiogram ICG Recommended training CPR metronome SAM 500P Overview Preparation Unpacking Checks before putting into service Guidelines for storage Register your SAM 500P Using the SAM 500P When to use Using the SAM 500P After use Service and Maintenance Tracking Requirements Data Management Troubleshooting Status indicator flashing red Low battery warning Memory full warning Audible warnings Device service required Sources of support Warranty exclusion Technical Data Appendix A Voice prompts Symbols used in this manual Warning Risk of death or serious injury A Caution Risk of injury Notice Risk of damage to data or material Further information Symbols used on this device On Off IP56 Ingress protection classified as IP56 according to EN 60529 Ti Consult operating instructions Follow instructions for use Q Single use item Do not re use R S ma IIS MC L h R oii c UL u 3XN6 Defibrillation protected Type BF connection Do not expose to high heat or open flame Do not incinerate Does not contain natural rubber latex Non sterile Recyclable Non rechargeable battery Do not short circuit b
7. Sine Technologies directly about the After Use Data Management Service 1 Connect the supplied USB cable to the SAM 500P see illustration below 2 Connect the USB cable to a PC 3 Launch the HeartSine Saver EVO utility Notice The SAM 500P should only be connected to an IEC60950 PC Caution You cannot defibrillate while the SAM 500P is connected to a PC For further information on this optional accessory contact your authorised distributor or HeartSine Technologies directly 21 Troubleshooting Status indicator flashing red If the status indicator is flashing red or if the device is emitting a beep check the expiry date on your Pad Pak see Preparation on page 13 If the expiry date has not been passed turn on the SAM 500P by pressing on the front panel and listen for the voice prompt call for medical assistance Then turn off by pressing on the front panel If this action does not correct the problem contact your authorised distributor or HeartSine Technologies immediately Low battery warning The first time the device plays the message warning low battery it will still continue to function properly However it may have fewer than 10 shocks left If you hear this message during use continue to use the SAM 500P until the emergency services arrive If available prepare the spare Pad Pak for use and be prepared to This message does not indicate a fault
8. age 19 Correct disposal of the device Dispose of the device in accordance with the European WEEE Directive or contact your HeartSine distributor Please follow the After use on page 17 Compliance with local regulations Check with the relevant local government health department for information about any requirements associated with ownership and use of a defibrillator in the region where it is to be used Introduction The HeartSine Samaritan PAD with CPR Advisor SAM 500P The HeartSine Samaritan PAD with CPR Advisor SAM 500P is a semi automatic external defibrillator designed to quickly deliver a defibrillation shock to victims of Sudden Cardiac Arrest SCA The SAM 500P is designed to operate in accordance with the joint European Resuscitation Council ERC and American Heart Association AHA 2010 guidelines on Cardiopulmonary Resuscitation CPR and Emergency Cardiovascular Care ECC Sudden Cardiac Arrest SCA Sudden cardiac arrest is a condition in which the heart suddenly stops pumping effectively due to a malfunction of the heart s electrical system Often victims of SCA have no prior warning signs or symptoms SCA can also occur in people with previously diagnosed heart conditions Survival from SCA depends on immediate cardiopulmonary resuscitation CPR The use of an external defibrillator within the first few minutes of collapse can greatly improve patient s chances of survival Heart
9. arnings and Cautions Do not use if the pouch containing the electrodes is not sealed The Pad Pak is a single use item and you must replace it after each use or if pouch that seals defibrillation pads has been broken or compromised in any way If you suspect that the Pad Pak is damaged you must replace it immediately t Notice Susceptibility to electromagnetic interference To safeguard against interference you must operate the SAM 500P at least 2 m 6 feet away from all radio frequency devices and other susceptible equipment Alternatively switch off the equipment affected by or causing electromagnetic interference Temperature range for operation The SAM 500P with its battery pads and electrodes is designed to operate in the temperature range of 0 C to 50 C Use of the device outside this range may cause malfunction Environmental protection The IP56 rating does not cover the immersion of any part of the SAM 500P in water or any type of fluid Contact with fluids may seriously damage the device or cause fire or a shock hazard Prolonging standby life Do not turn the device on unnecessarily as this may reduce the standby life of the device Standby storage outside the range of 0 C to 50 C may decrease the shelf life of the device Do not test on simulators and manikins Our devices cannot be tested using industry standard simulators and manikins Standard simulators have a constant r r spa
10. around this icon are flashing Green tab Electrode drawer 12 Preparation Unpacking Checks before putting into service Check that the contents include the User Check the expiry date year month on the Manual soft case Pad Pak guarantee rear of the Pad Pak see the illustration card and Quick Start Guide below If the expiry date has passed you A Pad Pak is a single use removable milspieplace me Fad Pak battery and electrode pack in one unit It is available in two versions a pink coloured Pad Pak for use with children and a grey coloured Pad Pak for use with adults see the illustration below Adult Pad Pak Paediatric Pad Pak 13 Preparation Unpack the Pad Pak Retain the packaging in case you need to return the Pad Pak to HeartSine Insert the Pad Pak into the SAM 500P see the illustration below zaa A The SAM 500P will run a self test routine The action arrows will flash during this process On successful completion of the self test routine the green status indicator see SAM 500P Overview on page 12 will blink If so your SAM 500P is ready for use 1 14 Notice Do NOT pull the green tab on the Pad Pak If you have opened the electrode drawer you must replace your Pad Pak Turn on the SAM 500P by pressing on the front panel To check that the device is operating correctly listen for appropriate voice prompts and make sure that no warning messages are
11. attery Do not crush battery Temperature limitation as indicated Use by yyyy mm Dispose of in accordance with country requirements Automated External Defibrillator With respect to electrical shock fire and mechanical hazards only in accordance with e ANSI AAMI ES60601 1 2005 s CSA C22 2 NO 60601 1 2008 IEC60601 2 4 2010 UL60601 1 2006 CSA C22 2 No 601 1 M90 IEC60601 2 4 2002 l Warnings and Cautions A Warning Patients suitable for treatment The SAM 500P has been designed to work on unconscious nonresponsive patients If the patient is responsive or conscious do not use the SAM 500P to provide treatment The SAM 500P uses an interchangeable battery and electrode pack called Pad Pak The SAM 500P in combination with an adult Pad Pak is suitable for use on patients of over 25 kilograms 55 pounds in weight or equivalent to a child of approximately eight years old or over For use on children remove the adult Pad Pak and install a paediatric Pad Pak If a paediatric Pad Pak or an alternative suitable defibrillator is not available you may use an adult system If you treat a paediatric patient with an adult Pad Pak ignore any voice prompts regarding the quality of the cardiopulmonary resuscitation CPR The 4 CPR Advisor is currently only intended to provide feedback on adult patients Do not delay treatment trying to find out the patient s exact age and weight Risk of electric shock Th
12. by one of the following Soapy water Isopropyl alcohol 70 solution Caution Do not immerse any A part of the SAM 500P in water or any type of fluid Contact with fluids may seriously damage the device or cause a fire or a shock hazard Notice Do not clean the SAM 500P with abrasive materials cleaners or solvents 17 Using the SAM 500P Check the SAM 500P for damage If the SAM 500P is damaged replace it immediately Install a new Pad Pak Before installing check the Pad Pak expiry date see Preparation on page 13 After installation check the status indicator is blinking green 18 Service and Maintenance HeartSine recommends users perform regular maintenance checks A suggested maintenance check would be Weekly 1 Check the status indicator If the green status indicator is not flashing approximately every 5 seconds a problem has been detected See Troubleshooting on page 22 The SAM 500P performs a self test routine at midnight GMT on Sunday During this self test period the status light blinks red but returns to green on successful completion of the self test routine The self test takes no more than 10 seconds to complete If the status indicator continues to flash red the SAM 500P has a fault see Troubleshooting on page 22 Monthly 2 If the device shows any signs of physical damage contact your authorised distributor or HeartSine Technologies directly 3 Check the e
13. cing and do not produce the variability displayed in the normal human heart Our algorithm uses heart rate variability as one of its criteria for measuring ventricular fibrillation VF Consequently we do not recommend the use of normal simulators to test our device Further Information Use of this manual It is important that you read this manual carefully before using the HeartSine Samaritan PAD with CPR Advisor SAM 500P This manual is presented in support of any training you may have received If you have any questions contact your authorised distributor or HeartSine Technologies directly for advice or explanation The information in this manual is subject to change without notice and does not represent a commitment on behalf of HeartSine Technologies No part of this manual may be reproduced or transmitted in any form or by any means electrical or mechanical including photocopying and recording for any purpose without the express written permission of HeartSine Technologies l Warnings and Cautions Operator training HeartSine recommends that the SAM 500P is used by people trained in cardiopulmonary resuscitation defibrillation CPR D Use of accessories The SAM 500P is a self contained device Do not use any unauthorised accessories with it The SAM 500P may malfunction if non approved accessories are used Regular maintenance Check the device periodically See Service and Maintenance on p
14. d ECG sensor ICG sensor defibrillation pad Adult anterior lateral Paediatric electrodes anterior posterior or anterior lateral 100 cm 3 5 ft 1 m See the expiry date on the product SCOPE Self Compensating Output Pulse Envelope biphasic escalating waveform Optimised biphasic waveform compensates energy slope and envelope for patient impedance Pre configured factory settings for escalating energy are Version AHA ERC 2010 Adult 1 Shock 150 J 2 Shock 150 J 3 Shock 200 J Paediatric 1 Shock 50 J 2 Shock 50 J 3 Shock 50 J Internal memory 90 minutes of ECG full disclosure and event incident recording Custom USB cable directly connected to a PC and Saver EVO Windows based data review software Electromagnetic compatibility EMC EN 60601 1 2 Radiated Emissions EN55011 1999 A2 Electrostatic Discharge EN61000 4 2 8 kV RF Immunity EN61000 4 3 80 MHz 2 5 GHz 10 V m Magnetic Field Immunity _EN61000 4 8 3 A m Aircraft RTCA DO 160D Section 21 Category M RTCA DO 227 TSO C142a HeartSine Quality Management System ISO 9001 GB02 54194 ISO 13485 GB02 54195 EEC 92 43 GB02 54193 Technical Data SCOPE Biphasic Waveform The SAM 500P delivers a Self Compensating Output Pulse Envelope SCOPE biphasic waveform This waveform automatically optimises the waveform pulse envelope amplitude slope and duration for a wide range of patient impedances from 20 ohms to 230 ohms The deliver
15. e SAM 500P delivers therapeutic electrical shocks that can cause serious harm to either operators or bystanders Take care to ensure that nobody touches the patient when a shock is to be delivered Avoid opening or repairing The SAM 500P has no serviceable parts Do NOT open or repair the device under any circumstances as there could be danger of electric shock If damage is suspected replace the SAM 500P immediately Avoid explosive or flammable gases It has been determined that the SAM 500P is safe to use with oxygen mask delivery systems However to avoid the risk of an explosion it is strongly advised that you do NOT use the SAM 500P in the vicinity of explosive gases including flammable anaesthetics or concentrated oxygen A Caution Correct placement of the electrode pads Proper placement of the SAM 500P s electrode pads is critical You must strictly observe the instructions shown in the Quick Start Guide and on the device Wrong placement or the presence of air hair surgical dressings or medicine patches between the pads and the skin could cause skin burns Slightly red skin after shock therapy is normal Do not touch the patient during analysis Touching the patient during the analysis phase of treatment can cause interference with the diagnostic process Avoid contact with the patient while analysis is being carried out The device will instruct you when it is safe to touch the patient l W
16. ed waveform to the patient is an optimised impedance compensated biphasic truncated exponential waveform that incorporates an escalating energy protocol of 150 joules 150 joules and 200 joules The duration of each phase is automatically adjusted to compensate for varying patient impedances The first phase T1 duration is always equivalent to the second phase T3 duration The interphase pause is always a constant 0 4 ms for all patient impedances The specific SCOPE waveform characteristics for a 150 joules pulse are listed opposite 28 i a ha T T_T 23 4 5 7 8 9 T 6 ration msec Du Waveform Voltages volts Waveform Resistance Duration ms ohms Tilt 25 1630 63 1 50 1640 52 7 75 1650 51 4 6 5 6 5 100 1660 48 7 8 8 125 1660 50 4 10 5 10 5 150 1660 48 7 12 12 175 1660 48 7 14 14 200 1660 47 6 15 5 15 5 225 1670 46 7 17 17 Adult Pad Pak waveform specification Resistance ohms Energy joules Waveform Voltages volts Waveform Duration Cut 25 47 5 514 55 6 7 8 5 4 50 51 3 671 50 4 8 8 6 75 52 1 751 47 1 10 6 6 100 51 8 813 44 3 10 8 6 8 125 52 4 858 41 4 11 5 7 3 Paediatric Pad Pak waveform specification Arrhythmia analysis algorithm The SAM 500P uses the HeartSine samaritan ECG arrhythmia analysis algorithm This algorithm will evaluate the patient
17. ents rescuers should compress the lower half of the sternum but not compress over the xiphoid For patients at the upper end of the paediatric range adult style compressions should be performed The force required for the paediatric patients is less than that required in adult CPR CPR advisor is currently configured only to advise compressions at a force and rate suitable for adult patients over eight years old weighing more than 25 kgs 55 Ibs Electrode placement may also differ in paediatric patients Depending on the patient size the electrodes may be placed anterior posterior front and back or anterior lateral standard adult placement Differing electrode positions may result in different ICG readings Current technology does not support CPR advisor in determining which electrode placements are being used and therefore electrodes must be placed anterior lateral for CPR advisor to function correctly For these reasons CPR advisor is disabled when a paediatric Pad Pak is used in the SAM 500P Notice The ECG readings used to determine if the patient requires a defibrillation shock are not affected by the electrode positions selected in paediatric patients ignore the prompts provided The CPR Advisor is currently only intended Warning If a paediatric patient is treated with an adult Pak Pad you must to provide feedback on adult patients 32 idance and manufacturer s declaration electromagnetic emissions The HeartSi
18. er frequency Power frequency magnetic fields 50 60 Hz 3 A m Complies should be at levels characteristic magnetic field P of a typical location in a typical IEC 61000 4 8 commercial or hospital environment Surge 1 kV line s to line s Not applicable IEC 61000 4 5 2 kV line s to earth Not applicable 34 uidance and manufacturer s declaration electromagnetic immunity The HeartSine Samaritan PAD with CPR Advisor SAM 500P is intended for use in the electromagnetic environment specified below The customer or user of the HeartSine Samaritan PAD with CPR Advisor SAM 500P should assure that it is used in such an environment IEC 60601 Compliance test level level Immunity test Electromagnetic environment guidance d 1 2 VP 800 Mhz to 800 Mhz d 2 3 VP 800 Mhz to 2 5 Ghz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation 40 Vim distance in metres m Radiated RF Field strengths from fixed RF transmitters as determined IEC 61000 4 3 E phe 10 Wit by an electromagnetic site survey should be less than to2 5 z the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol o Notes At 80 MHz and 800 MHz the higher frequency range applies These guidelines may not apply in all situations Electromagnetic propa
19. gation is affected by absorption and reflection from structures objects and people See footnotes on next page 35 Technical Data Guidance and manufacturer s declaration electromagnetic immunity a The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 5 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause interference if it is inadvertently brought into patient areas For this reason an additional factor of 10 3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges b Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location where the HeartSine Samaritan PAD with CPR Advisor SAM 500P is used exceeds the applicable RF compliance level see above the HeartSine Samaritan PAD with CPR Advisor SAM 500P should be observed to verify that it is operating normally If abnormal operation is observed additional measures may be necessary such as re orienting or relocating the HeartSine Samaritan PA
20. good quality of life following a cardiac arrest and having the misfortune of suffering neurological impairment due to inadequate cerebral oxygenation Most modern defibrillators and mechanical resuscitation systems have a metronome facility to ensure an operator administers CPR at the correct rate A feedback system to ensure the operator delivers compressions at the correct depth enabling adequate re filling time would optimise coronary perfusion pressures Impedance cardiography ICG measures the changes of shape and movement of blood in the thorax which could could be a useful indicator of perfusion levels during external cardiac massage The impedance cardiogram can be accurately measured using two standard defibrillator electrodes Combining both the FORCE and SPEED CPR management tools will enhance CPR efficacy for both lay users and minimally trained bystanders CPR Speed Good 82377 Sensitivity gt 90 Sensitivity gt 95 38 Sensitivity gt 83 40 Specificity gt 90 Specificity gt 93 11 Specificity gt 82 19 CPR 108728 Sensitivity gt 90 Sensitivity gt 99 96 Sensitivity gt 99 54 Force Adequate Specificity gt 90 Specificity gt 98 47 Specificity gt 96 29 31 Technical Data Paediatric restriction Use of the CPR Advisor function must be restricted to adult patients only Chest compression techniques differ for the different ages and sizes of paediatric patients up to eight years old For younger paediatric pati
21. ne Samaritan PAD with CPR Advisor SAM 500P is intended for use in the electromagnetic environment specified below The customer or user of the HeartSine Samaritan PAD with CPR Advisor SAM 500P should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance The HeartSine Samaritan PAD with CPR Advisor SAM RF emissions CISPR 11 Group 1 500P uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable 33 Technical Data Guidance and manufacturer s declaration electromagnetic immunity The HeartSine Samaritan PAD with CPR Advisor SAM 500P is intended for use in the electromagnetic environment specified below The customer or user of the HeartSine Samaritan PAD with CPR Advisor SAM 500P should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance Electromagnetic environment level guidance Electrostatic 6 kV contact Complies Floors should be wood concrete or ceramic tile If floors are covered discharge ESD ith heti ial th lati IEC 61000 4 2 wit synt etic material the relative t8 kV air Complies humidity should be at least 30 Pow
22. ning provider If potential users of the SAM 500P are not trained in these techniques contact your authorised distributor or HeartSine Technologies directly Either can arrange for training to be provided Alternatively contact your local government health department for information on certified training organisations in your region CPR metronome The SAM 500P will play an audible click and flash the Safe To Touch indicator at a rate compliant with current AHA ERC guidelines This feature is referred to as the CPR metronome Responders should use this as a guide to the rate for performing compressions 11 SAM 500P Overview Attach pads icon Status indicator When the indicator is flashing green the SAM 500P is ready for use Data port The action arrows around this icon will flash to instruct you to attach the electrode pads to Shock button the patient as indicated Do not touch the patient Press this button to deliver a ee therapeutic shock when the action arrows fe around this icon are to flashing The SAM 500P may be analysing the patient s heart rhythm or ARs V Q h about to charge preparing Paediatric Q ho to deliver a shock ready icon Ss t eg N Action arrows CPR advisor A icon Se Z Do not touch icon On Off button Press this button to turn the device on or off Safe to touch icon Speaker It is safe to touch the patient when the action arrows a
23. patients of over 25 kg 55 Ibs in weight or equivalent to a child of approximately eight years old or over For use on younger children remove the adult Pad Pak and install a paediatric Pad Pak If a paediatric Pak Pad or an alternative suitable defibrillator is not available you may use an adult Pad Pak 16 Using the SAM 500P Refer to the separate Quick Start Guide During use the SAM 500P will give extensive voice prompts to guide the user For full list of voice prompts see Appendix A on page 37 Notice PAD aborts a ready to shock condition once a non shockable rhythm is detected After use Turn off the SAM 500P by pressing on the front panel Remove the electrode pads from the patient and stick them together face to face The electrodes may be contaminated with human bodily tissue fluid or blood Dispose of the electrodes separately as an infectious waste material The Pad Pak is a single use item and must be replaced after each use Remove the Pad Pak by pressing the two tabs on either side of the Pad Pak The Pad Pak will slide forward see the illustration below Do not dispose of the SAM 500P or Pad Pak in the normal waste Dispose of it at an appropriate recycling facility according to local requirements Alternatively return it to your distributor for disposal or replacement Check the SAM 500P for dirt or contamination If necessary clean it using a soft cloth dampened
24. played Notice Only turn the SAM 500P on ONCE If you turn it on and off repeatedly you will exhaust the batteries prematurely and you will need to replace the Pad Pak Turn off the SAM 500P by pressing on the front panel Check the status indicator see SAM 500P Overview on page 12 is flashing green If you have heard no warning messages and the status indicator is flashing green the device is ready for use Guidelines for storage Place the SAM 500P in its supplied soft carry case Store the SAM 500P in an unobstructed secure location in a clean dry environment Be sure to store according to specifications see Technical Data on page 24 Notice HeartSine recommends that you keep a spare Pad Pak with your SAM 500P You can store it in the rear section of the soft carry case Register your SAM 500P Complete the guarantee card and return it to your authorised distributor or HeartSine Technologies directly see Tracking Requirements on page 20 15 Using the SAM 500P When to use The HeartSine Samaritan PAD with CPR Advisor SAM 500P is indicated for use on victims of sudden cardiac arrest who are exhibiting the following signs Unconscious Not breathing No life signs The SAM 500P has been designed to work on unconscious nonresponsive patients If the patient is responsive or conscious do not use the SAM 500P to provide treatment The SAM 500P is suitable for use on
25. s ECG to ascertain if a therapeutic shock is appropriate If a shock is required the SAM 500P will charge and advise the user to press the shock button If no shock is advised the device will pause to allow the user to deliver CPR The SAM 500P ECG arrhythmia analysis algorithm performance has been extensively evaluated by using several databases of real life ECG traces Included in this are the American Heart Association s AHA database and the Massachusetts Institute of Technology MIT NST database The SAM 500P ECG arrhythmia analysis algorithm s sensitivity an specificity meet the AAMI DF80a 2003 requirements and AHA recommendations 29 Technical Data The SAM 500P ECG arrhythmia analysis algorithm performance is summarised in the table below 90 One Rhythm Class ee Performance Performance Sided Lower y p Specifications Results Confidence ize aes Limit Shockable rhythm eee wr 2453 Sensitivity gt 90 93 48 90 58 ventricular tachycardia VT Non shockable rhythm Geshoceen layin 1902 Specificity gt 95 100 100 Asystole Non shockable rhythm y 46711 Specificity gt 95 99 11 95 04 All other rhythms No error to measure 30 The following summary shows the results produced by the CPR component of the diagnostic algorithm when run against the clinical database The importance of administering effective chest compressions can mean the difference between a patient having a
26. ss and rate of compressions given and then based on this analysis advises the rescuer to push harder faster or slower as appropriate The SAM 500P uses both audible and visual prompts to 10 give the responder feedback on the quality of the compressions administered Warning The CPR Advisor function is intended for use on adult patients only If a paediatric Pad Pak is used the CPR function is disabled In this case the rescuer is prompted to begin CPR but receives no CPR Advisor feedback Impedance cardiogram ICG The impedance cardiogram is a method of measuring changes in the patient s impedance due to motion blood flow and changes to the shape of the heart The SAM 500P uses these measurements to estimate the impedance changes in the chest and so as a way to determine the effectiveness of compressions given during CPR Recommended training SCA is a condition requiring immediate emergency medical intervention Due to the nature of the condition this intervention can be performed before seeking the advice of a physician To properly diagnose this condition HeartSine recommends that all potential users of the SAM 500P are fully trained in cardiopulmonary resuscitation CPR basic life support BLS and in particular the use of an automated external defibrillator HeartSine also recommends that this training be kept up to date by regular refresher courses as and when recommended by your trai
27. tSine or its authorised distributors are not obliged to replace or repair under warranty if one or more of the following conditions apply The device has been opened Unauthorised modifications have been made The device has not been used in accordance with the instructions provided in this manual The serial number has been removed defaced altered or by any other means made unreadable The device has been used or stored outside its indicated temperature range The Pad Pak packaging is not returned The device has been tested using unapproved methods or inappropriate equipment see Warnings and Cautions on page 4 23 Technical Data Physical parameters with Pad Pak installed Size Weight Environmental Operating temperature Standby temperature Transport temperature Relative humidity Enclosure Altitude Shock Vibration 20 x 18 4 x 4 8 cm 8 0 x 7 25 x 1 9 in 1 1 kg 2 4 Ibs 0 to 50 C 32 to 122 F 0 to 50 C 50 to 122 F 10 to 50 C 14 to 122 F for up to two days If the device has been stored below 0 C 32 F it should be returned to an ambient temperature of between 0 to 50 C 32 to 122 F for at least 24 hours before use 5 to 95 non condensing IEC 60529 EN 60529 IP56 0 to 15 000 feet 0 to 4575 metres MIL STD 810F Method 516 5 Procedure 40G s MIL STD 810F Method 514 5 Procedure 1 Category 4 MIL STD 810F Method 514 5 Procedure
28. xpiry date of the SAM 500P Pad Pak see Preparation on page 13 for the location of the date If the date has expired replace with a new Pad Pak or contact your local HeartSine distributor for a replacement 4 If you hear a warning message when you turn on your SAM 500P or if for any reason you have suspicions that your SAM 500P is not working correctly contact your authorised distributor or HeartSine Technologies directly support HeartSine com 19 Tracking Requirements The Medical Devices Regulations require us to track the location of all medical devices sold It is important that you complete the warranty card with your details and return it to your authorised distributor or HeartSine Technologies directly Alternativley please send an email to support heartsine com containing the following information Name Address Device serial number or use our on line registration tool at https secure heartsine com UserRegistration html Your participation will allow us to contact you with any important notifications about the SAM 500P such as any future software updates or field safety corrective actions If there is a change in the information you have provided to us such as a change of address or a change in ownership of your SAM 500P please contact us with the updated information 20 Data Management The HeartSine Saver EVO software is an optional accessory Contact your authorised distributor or Heart