1. Only use fuses that have the same specified rating see Technical Specifications Using fuses with higher ratings could result in damage and or injury The electronic controller is a precision electronic product Use care when handling or transporting Dropping or other sudden impacts may result in damage to the unit Risk of electric shock DO NOT OPEN Do not attempt to repair or service the Control Unit Repairs and service should be conducted by Novis Healthcare nominated Service Centres If the Control Unit is not functioning properly or has been damaged unplug the unit and take it out of service immediately Contact Lifehealthcare on 1300 133 120 for repairs and service information Do not place any objects items such as blankets on or over the Control Unit Excessive weight on the Control Unit could result in damage The power cord to the Control Unit should be positioned to avoid a tripping hazard and or damage to the cord Novis Healthcare recommends placing the cord under the bed frame and attaching it to an electrical outlet by the head of the bed Maintenance amp Cleaning Control unit WARNING Ensure the Control Unit is disconnected from the mains electricity supply before cleaning WARNING DO NOT spray disinfectant directly on to the Control Unit or immerse the controller in any type of liquid This could result in a severe electrical hazard as this equipment has no protection against ingre
2. 1 Apply an intermediate level disinfectant or a solution of Sodium Hypo chlorite or similar up to 10 000 ppm available chlorine to the Top Cover upper surface either by spraying or by hand application 2 Ensure the surface is completely covered with the disinfectant and remains in contact with the surface according to manufacturer s instructions 3 Remove disinfectant and rinse thoroughly 4 Allow to air dry before use Outside Shell Wipe down with disinfectant or a solution of Sodium Hypo chlorite or similar up to 10 000 ppm available chlorine ensuring that all surfaces come in contact with the disinfectant Rinse off well with a clean damp cloth and air dry CPR The exterior of the CPR assembly and tag can be periodically wiped using a cloth a dampened with disinfectant Air Cell Assembly CAUTION DO NOT machine wash or dry the air cells The air cell assembly does not routinely need to be cleaned or disinfected IMPORTANT DO NOT disassemble the overlay unless cleaning is required If cleaning or disinfecting is required DO NOT disconnect the hoses from the air cells Product Overview Ultimate75 Alternating Overlay System The Ultimate75 is an alternating mattress overlay system providing pressure application and release to patients with or vulnerable to pressure ulcers It is designed for use on top of an existing mattress and can be used on both standard and profiling bed frames The U
3. must be assigned a return authorisation number prior to return Returns should include information describing the problem and or requested repair and be sent to a Lifehealthcare Service Centre by prepaid transportation Lifehealthcare Hospital amp Rehab will return the repaired replaced product at no charge Warranty repairs do not extend the length of the warranty period Neither Lifehealthcare Hospital amp Rehab its officers directors employees or agents shall be liable for consequential or other damages including but no limited to personal injury loss or any other expense directly or indirectly arising from the use of its products The sole remedy for breach of the limited warranty granted herein shall be repair or replacement of the Lifehealthcare Hospital amp Rehab products Technical Specifications Ultimate75 Cycle Control Cycle Time 10 minutes Supply Voltage 230V 50Hz 120V 60Hz 0 2A for Control Unit 0 01 6A for power down alarm availability of supply monitored by in line power down alarm which provides failure indication Fuse Rating Mains Plug 3A Control Unit 1A Power Rating 12VA Overlay Order Code MC 3 PA Overlay Dimensions amp Weight Control Unit Dimensions amp Weight Length 2000mm Height 115mm Width 880mm Width 135mm Height 90mm Depth 270mm Weight 8 4kgs Weight 2 1kgs Cells 21 cells Cell material 0 15mm TPU film laminated on 210 Denier Nylon fabric Base material Cover Material M
4. to reach operating pressure the pump Control Unit only emits a slight sound therefore it is necessary to hold the working Control Unit up to their cheek to feel the slight stream of air Also ensure that the quick release connectors are fully seated If the system is still not inflating and there is no air from the pump outlets then it may need a compressor replacement Q How do test for compressor failure and pressure switch status A Screw the female quick release connector out of the side of the pump the connector with the button Once unscrewed switch on the Control Unit and hold one finger over the male outlet and one finger over the hole left from unscrewing the female connector If a spare male connector is available you can connect this into the female outlet to seal it so that it does not have to be unscrewed out of the pump casing With both openings blocked the low pressure light should flick to normal within 10 seconds if this does not happen then the pump needs to be opened and inspected If unsure return it to Lifehealthcare to be properly pressure tested Q Control Unit will not start A Check that the Control Unit is connected to electrical outlet and that the unit is switched on Check the mains plug fuse 3 Amp Check both 1 Amp fuses in the base of the Control Unit Q The overlay is not alternating A The Control Units are sometimes mistreated on these systems If handled r
5. This product carries the CE mark in accordance with EC Directive on Medical Devices 93142IE EC Control Number 95GG Classified by Underwriter s Laboratory Inc for UL 2601 1 standard for Medical Electrical Equipment URE axe e mia eel Imate User Manual Joidacore Healthcare Equipment Solutions Important Notice Warranty Information Before operating this medical equipment it is important to read this manual and understand the operating instructions and safety precautions Failure to do so could result in patient injury and or damage to the product If you have any questions please contact Lifehealthcare on 1300 133 120 Lifehealthcare Hospital amp Rehab warrants each of its products to perform in accordance with established specifications for the following time periods starting from the date the product is shipped from our supplier Control Units 2 years Soft Goods 2 years During the warranty period Lifehealthcare Hospital amp Rehab will repair or replace at no charge any products that are not performing in accordance with established specifications unless the problem failure is due to 1 customer damage negligence and or misuse or 2 unauthorised repairs Items not covered under warranty include but are not limited to stains punctures cuts damages to electrical cords rips or tears dents and or lost missing parts All products returned for warranty repairs
6. elow CE Ultimate75 Status Lamps Power Switch Analogue Comfort Control Dial Overlay Function Establishing pressure supine patient Once the patient is in place adjust the dial for the best comfort setting that combines the required pressure support and comfort After approximately 20 minutes re assess whether the patient is comfortable and the unit is functioning correctly IMPORTANT When lowering the pressure setting ensure the patient is not bottoming out not sufficiently supported by the air cells and therefore coming in contact with mattress You should be able to slide your hand between the cells underneath the patient s sacral area bottom along a cell that is deflated Ensure patient is in a supine position for this assessment Establishing Pressure inclined patient When moving the patient to a more upright position pressure may need to be increased approximately 20 to provide added support and to avoid bottoming out NOTE It is important to return the original pressure setting when the patient returns to the supine position IMPORTANT Wait a minimum of 10 minutes between any adjustment and patient assessment as it may take a full cycle for the system to adjust to any new setting Operation System Removal Transport Function Overlay Removal 1 Switch off the Control Unit If applicable reset the alarm indication on the power down alarm Switch of
7. emove the blown fuse from the fuse holder clip and discard 5 Insert a new fuse into the plug Push against the force of the spring and turn clockwise with the screwdriver a turn Maintenance amp Cleaning Infection Control and routine cleaning must be carried out in accordance with your local Infection Control Policy It is suggested that all disinfection be done with a high grade disinfectant in accordance with manufacturer s instructions WARNING Do not use high temperature autoclave or use Phenolic based products for cleaning IMPORTANT It is recommended the system is cleaned between patients and approximately every 2 weeks if in constant use Overlay Top Cover Washing and Disinfecting If there are visible signs of body fluids and or substances present top covers should be washed Top covers can be machine washed up to 70 C using disinfectant according to manufacturers instructions To determine the amount of disinfectant to use determine the amount of water in the washer and then follow the manufacturers instructions for dilution Soak the Top Cover in the disinfectant during the wash cycle Rinse well in clean water and dry thoroughly before use CAUTION DO NOT dry the Top Cover using the HEAT cycle or a dryer Air dry or select a low or NON heat dry cycle Washing Alternative If there are no visible signs of body fluids and or substances on the top cover it can be sanitized
8. f the mains supply and disconnect the power down alarm from the electrical outlet 2 Disconnect the inflation hoses from the control unit to release air from the system 4 Remove the Control Unit power down alarm if applicable and power lead from the bed and place on the overlay 5 Roll up the mattress and remove from the bed Overlay Transport Function 1 When preparing for transport mode it is best if the patient is not inclined but supine Turn the pressure control knob up to maximum pressure and wait 10 minutes for the setting to take effect Ensure that the overlay is not currently undergoing an alternation If not then turn off control unit and disconnect both hoses Immediately connect the air hoses together Once the air hoses are connected together the air pressure will average out between the fully inflated cells and the deflated cells providing an immersive surface for pressure relief during transport Bottoming Out Test 1 Ensure system is on alternation mode but is not undergoing an alternation Unzip top cover just past sacral region Slip hand under top cover and along a deflated cell under the patients sacrum If hand can pass under patient then patient is adequately suspended
9. icro Air Loss cells Connection Emergency Electrical Safety Standards Mode of Operation Environment Air humidity Ambient temperature All product specifications are subject to change without notice Contents Purpose designed distributor valve supplying operating air to the inflatable cells Nylon fabric 420 Denier with a 0 1mm TPU coating Nylon with PVC membrane PU leather base is polyester N A Quick release connectors N A EN 60601 1 1996 UL 2601 1 safe CE UL Continuous 30 to 75 0 C to 70 C Safety Precautions In General Control Unit CAUTION Do not use this equipment in the presence of flammable anesthetics Explosions could result IMPORTANT Bed frames used with the systems can vary greatly depending on the specific health care setting ie hospitals nursing homes home care etc Therefore it is the responsibility of the caregiver to take the necessary precautions to ensure the safety of the patient This includes but is not limited to the appropriate use of side rails to prevent falls and or patient entrapment IMPORTANT Minimise articles between the systems surface and patient and secure bed sheets loosely so as not to affect the alternating cell movement Only plug into a grounded power receptacle and use the power cord supplied with the system Exposure of the electronic controller to any liquid while it is plugged in could result in a severe electrical hazard
10. l unit positioning c Secure end straps completely under the mattress Ensure end straps are not twisted or pulled towards the corners of the mattress iY XN e Confirm there are no sharp objects in the immediate area which may risk damage to the overlay mattress Important Make sure that the attachment of the overlay does not interfere with the movement or operation of the bed a Position Control Unit by its hanging hooks over foot head board of the bed or on the floor under the bed b Attach two inflation hoses to Control Unit with the press on the quick release connectors Ensure air hoses do not kink between overlay mattress bed frame and Control Unit TT Insert power cord into Control Unit then into a grounded 240 Vac 50 Hz electrical outlet pen C d Switch on the power and turn the pressure dial up to maximum to commence inflation Allow up to 40 minutes for full inflation lt e Once ready the low pressure light will go out lay the patient on the bed and select the appropriate comfort level f Perform a bottom out test to ensure that patient is properly suspended by slipping hand under top cover along a deflated cell in sacral area g Secure sheets loosely enough to ensure they do not interfere with cell alternation Operation Basic Features Control Panel The Ultimate75 Control Unit front panel is shown in the diagram b
11. ltimate75 alternating overlay system consists of the following e Control Unit e Mattress Overlay e Power Cord e Instruction Manual e Carry Bag The Control Unit provides the air supply to the mattress It is controlled by an analogue dial to control comfort levels A lighted on off switch activates deactivates the control unit On the left side of the Control Unit are two quick release air connectors for connecting the overlay hoses The mains supply to the Control Unit can be easily disconnected and is designed to detach if tugged too firmly protecting the internal wiring of the unit The overlay comprises 21 high density cells which alternate over a 10 12 minute cycle This provides regular periods of pressure reduction to aid blood and lymphatic flow to vulnerable tissue A comforter cover can be used to provide additional comfort and warmth to the patient m E Installation Unpacking amp Inspection NOTE It is recommended that all packing materials and instructions be kept in the bag provided in the event that the product has to be shipped to an approved Lifehealthcare Service Centre Carefully remove the control unit overlay and accessories from the shipping cartons Inspect all items for any damage that e Ultimate75 overlay e Carry Bag e Analogue Control Unit e Medical Grade Power Cord e Quick Setup Guide e User Manual Analogue Control Unit Quick setup guide may have
12. occurred during shipping Any damages or missing components should be reported to a Lifehealthcare Service Centre as soon as possible The box should contain completely assembled overlay system components consisting of Mattress Overlay Carry Bag Medical Grade Power Cord P User Manual Troubleshooting Q The low pressure warning light keeps coming on A The warning light often activates when the Control Unit senses a loss of pressure somewhere in the system This can occur if weight suddenly shifts off a particular set of cells E g If the patient is getting into or being lifted out of bed If the light doesn t go out after 10 minutes then there could be a leak somewhere in the system Q The Control Unit is not working A Firstly check that the power cord is properly connected to both wall outlet and Control Unit If there is a power outage then press the fuse buttons to see if that re activates the system If this fails to re activate the Control Unit then the pump needs to be serviced NOTE If problem is not resolved please contact Lifehealthcare Service Centre Troubleshooting Below are solutions to some of the most frequently asked questions and answers to problems customers have experienced Solving a problem over the phone is naturally more cost effective than a site visit Q I have turned the pump on and the overlay is not inflating A The systems can take up to 40 minutes
13. oughly or turned with too much force the pressure control knob can be lifted and override the maximum pressure stopper ridge If this occurs the spring loaded pressure mechanism becomes fully seated and prevents the Control Units from alternating the overlay You need to gently pry away the control knob and twist the underlying axle slowly anti clockwise until you hear the hiss of air again Keep slowly twisting until the spring feels totally released and then tighten slightly clockwise until you feel it catch then replace the plastic knob with the indicator pointing to the lowest setting If unsure return it to Novis Healthcare to be properly pressure tested If this occurs on a system check to see that the static switch has not been activated Q Control Unit running amber Low Pressure Alert light illuminated overlay fails to inflate loses pressure A Check mattress air hoses are connected to the Control Unit correctly Check CPR sealing connectors are pushed fully onto the air hoses Check all mattress cells are pushed fully onto the air hoses Installation System Installation Control Unit Activation The following describes the procedures to follow upon setting up the system for the first time a Remove all covers and sheets from the existing mattress ie b Position overlay with its top cover on top of existing mattress and position hose ends at foot or head end of bed for preferred contro
14. ss of water WARNING This equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide External cleaning Wipe down Control Unit with warm water containing detergent or with a solution of Sodium Hypo chlorite or similar dry thoroughly before use Air Filter Replacement 1 Switch off the power supply to the Control Unit 2 Disconnect the electrical lead and air hoses 3 Place the Control Unit on a flat surface with its back panel uppermost NOTE be sure to place soft cloth under unit to prevent scratches 4 Carefully pry off the cover of the air filter remove and discard the filter material and fit new there may be a small locking screw use a small philips head to remove this first 5 Refit the air filter cover to the Control Unit The Control Unit is now ready for re connection NOTE Good filter maintenance is critical in keeping your system in optimal operating condition Failure to keep the filters clean will result in system downtime and increase repair costs It is recommended that the air filter be replaced annually Replacement air filters are available from Lifehealthcare Service Centre Fuse Replacement 1 Ensure the control unit is deactivated 2 Remove the power cord from the electrical socket on the side of the control unit 3 Using a small flat head screwdriver insert into the grove and turn anti clockwise 1 4 turn 4 R