Contents
1. PENETRANT PROCESSING INSPECTION EQUIPMENT PENETRANT COMPARATOR Product Description The Penetrant Comparator is generally used to compare the quality of working penetrant against a retained sample with the same batch number Part number 070C001 Operating Instructions Note Before use mark each half of the comparator with an appropriate label Dip one half up to the saw cut in the tank of penetrant Dip the other half again up to the saw cut in the retained penetrant Process the comparator block in the same way as you do for your components Either inspect visually or in the case of fluorescent penetrant under UV A light for difference in the clarity brightness of defects Alternatively record the results photographically a a a If complying with ASME B amp PV Code Section V refer to T 650 Procedure Technique Magnaflux A Division of ITW Ltd Faraday Road South Dorcan Industrial Estate AS S Swindon Wiltshire SN3 5HE UK REGISTRARS Tel 44 0 1793 524566 Fax 44 0 1793 490459 Email sales magnaflux co uk www magnaflux com Issue 2 14 05 2012 A MAGNAFLUX Operating Instructions Penetrant Comparator page 1 A Division of ITW Ltd