1. IST 300 A02 MAN REV 1 03 10 2008 Page 1 of 12 meet en regeltechniek User s Manual IST 300 2 channel portable function generator Output This manual is applicable for all calibrators with model number IST300 A02 001 Publication Number IST 300 A02 MAN Edition 1 September 2008 Copyright Istec International B V Netherlands IST 300 A02 MAN REV 1 03 10 2008 Page 2 of 12 Index LENT 019 0 24 ON anaE ne eee E ee eee ee er ee eee ee eee 3 LE AUTO Sgt ccs eset waa se os cece Gen cyan os eles canes A T chee oie eects 3 Le QURON NON een ee A ne AEE E A ee eee eee eee 4 LS PECCO ON eon ee E EA ener 5 LI MOGE ea E E E E EE EN E E 5 e ACAI NO scattered E E ETN 6 el FTOR PONE e E E EE AA ENE R 6 e OP PONET arr E E EE EA EEE EEOAE 7 LS SETIN UN eo E E EE E TE S 8 Co CONNEC 0 ea E EA E N E EE 8 Lo DCO E ranae A A E A EER 8 Z393 AMDITUGE oriri E E TEE E AEAEE OR 9 eF NCQ CIN CY oeiee A e E E E A ENESA 9 Z309 Combined OUT DUR eeissrsseiepi rirnan r erenn A i E N EE 10 3 Declaration of CONFOPMITY ccccccsccscessessescesccscesecsesscescsscesceecescscsscesseeseeseeeseeseeseens 12 it IST 300 A02 MAN REV 1 03 10 2008 Page 3 of 12 meet en regeltechniek 1 Introduction 1 1 Purpose In a number of situations determining the loop accuracy in the field is required A simple battery powered DC voltage simulator allows only static signals to be simulated e g axial displacement and or bias voltage For dynamic acc
2. calibrator Also next to the output a signal reverse switch is available for fast DC reverse selection J ISTEC INTERNATIONAL B V OUTPUT 2 ISTEC INTERNATIONAL B OUTPUT 1 nea X Ar AL 300 A02 001 neg g Mag pos SERIAL No 108090515 001 eg Frequenc 0 19999Hz K Freq iud y z s Amplitude 0 4V RMS ae Offset 0 16V 2a Mascot NiMH charger ape 2215 0 4A ccu pack 21 6V 2300mA Send wee eo ee a Figure 2 IST 300 A02 Top view meet en regeltechniek 1 3 Specification Digital Amplitude 14 bit Accuracy Digital Offset 14 bit Accuracy Temperature drift Maximum output current Frequency Range Tolerance Operating temperature range Storage 95 Humidity Battery operation 1 4 Models IST 300 AXX AXX Number of channels O1 Not Applicable 02 2 channel portable unit 03 Not Applicable 04 4 channel lab unit IST 300 A02 MAN REV 1 03 10 2008 Page 5 of 12 O 4 000 mV Pk Pk RMS selectable 100 mV 3 200 mV 4000 mV better than 1 O 16 000 mV selectable Full range 1 30 uV C Typical 20 mAmp O 19 999 Hz 19 999 Hz lt 0 001 1 Hz Dp C 599 C 10 80 C with batteries removed Up to 8 hour continuously NI MH fast charge type Other combinations are not applicable IST 300 A02 MAN REV 1 03 10 2008 Page 6 of 12 meet en regeltechniek 2 Operation 2 1 Front panel Output Figure 3 IST 300 A02 Two channel cal
3. eleration velocity and or radial vibration signals a function generator is needed The standard available function generators specific portable versions do not allow a sufficient DC span in combination with AC superimposed signals The IST 300 vibration calibrator solves these issues The compact two channel battery powered vibration calibrator has been build based on the input from field service engineers and for its purpose is truly one of the most easy to use and complete field service instruments in the field Following models are available e 2 channel portable IST 300 Type AO2 Battery operated e 4 channel lab instrument IST 300 Type A04 Not part of this manual Figure 1 IST 300 AO02Two channel calibrator with charge adapter IST 300 A02 MAN REV 1 03 10 2008 Page 4 of 12 meet en regeltechniek 1 2 Quick Operation Way of operation is for both models the same Per channel a toggle switch is used to select the appropriate display 1 Frequency 2 Amplitude 3 Offset Pk Pk or RMS output is selected by means of a second toggle switch The channel is set by means in an incremental dial Depending of the speed of dialling the output value is adjusted either in increments of 1 or 50 A Led indicates when the output has reached its maximum or 16 000 mVolt All value s are continuous visible in their assigned display The outputs are available by means of a 50 Ohm BNC output connector at the top of the vibration
4. ibrator Front View The front panel consists of two parallel simulator units Each unit consists of e LCD display 1 Frequency e LCD Display 2 AC Amplitude e LCD Display 3 DC offset e Switch Display 1 2 3 Operation mode selector e Switch Output Output mode RMS or PEAK e Dial Incremental Dial e Peak output LED lights up when the max output has been reached IST 300 A02 MAN REV 1 03 10 2008 Page 7 of 12 meet en regeltechniek 2 2 Top Panel e oe Ol ISTEC INTERNATIONAL B V OUTPUT 1 ee gee eters neg OS art nr n neg See PO SERIAL No 108090515 001 g Frequency 0 19999Hz be Amplitude 0 4V RMS BS Offset 16V 2 Mascot NiMH charger ype 2215 0 ccu pack 21 6V 2300mA Figure 4 IST 300 A02 Two channel calibrator Top View The top panel consists of e Two BNC outputs output one and output 2 e Two polarity change switches e On off switch e Power input socket e Product label containing product number serial number charger make and model CE label IST 300 A02 MAN REV 1 03 10 2008 Page 8 of 12 meet en regeltechniek 2 3 Setting up To operate the unit either the batteries need to be charged or the unit needs to be connected to the power adapter Note Make and model are registered on the label on top of the IST 300 unit Operation is very simple and for each simulator identical Below the operation is described for one simulator For other simulators these steps need to be
5. red an ace calibrator Peak Output Function In the example above the DC value the RMS AC value exceed the 16 000 mVolTt 13 014 mV DC 2 624 mV RMS 1 414 16 724 m Volt Note The factor 1 414 is the square root of two Which is the difference between the RMS and PEAK value for pure sinus shaped signals IST 300 A02 MAN REV 1 03 10 2008 Page 12 of 12 meet en regeltechniek 3 Declaration of conformity Declaration of Conformity Istec International B V Keer en Duin amp 2163 HA Lise Declare under our sole responsibility that the product range ISTCAL 300 AXX XXX Field Calibrators Note AXX XXX indicates all models which are covered by any number variation of the X s To which this declaration relates is in conformity with the following standard s EN 53022 EN 6 1 4 2 EN W014 ENG 14 4 ENY S014 Following provisions of the Electromagnetic Compatibility 89 336 EEC Directive Lisse Netherlands 30 09 2708 O O sesssessessresserseessermeesms D Verschuren Managing Director Declaration of Conformity ISTCAL 300 AKX AXA M September 2008 Gemaakt op 30 9 1008 2 22 Pagina van Istec International B V Meer en Duin amp 2163 HA Lisse Netherlands
6. reproduced 2 3 1 Connection Connect the BNC output to the device to be tested Note power output signals needs either to be disconnected or switched off For ICP loops the power mode on the device to be tested needs to be switched to off or to external 2 3 2 DC offset To select the DC display switch the Display selector to Display 3 Note In the top left corner of the LCD display a indicator is shown indicating which display is selected Figure 5 IST 300 A02 Two channel calibrator Display The Dial is used to change the value IST 300 A02 MAN REV 1 03 10 2008 Page 9 of 12 meet en regeltechniek When the output switch on the top is set to negative the output is negative voltage E g required for Eddy Current Probe loops using 24 volt power When the output switch on the top is set to positive the output is positive voltage E g required for ICP sensors loops using 4 mAmp Power supply and 12 Volt Bias Voltage Using the dial any value between O and 16 000 m Volt may be entered Depending on the polarity switch the output is either O 16 000 mVolt or O 16 000 mVolTt 2 3 3 Amplitude Once the DC value is set the display switch may be changed to Amplitude Figure 6 IST 300 A02 Two channel calibrator Display The indicator now moves to the amplitude display Using the dial any value between O and 4 000 m Volt may be entered The entered value will be super imposed on the previously entered value When the DC outp
7. ut and the amplitude output together exceed 16 000 mVolt the Peak led will turn red Figure 7 IST 300 A02 Two channel calibrator Peak Output 2 3 4 Frequency Once the Amplitude is set the display switch may be changed to Frequency The indicator now moves to the frequency display Using the dial any value between O and 19 999 Hz may be entered IST 300 A02 MAN REV 1 03 10 2008 Page 10 of 12 meet en regeltechniek 2 3 5 Combined output The 3 different settings are combined into one output DC value AC value are defining the voltage output Either in RMS or PEAK depending on the RMS PEAK switch PIP Output Figure 8 IST 300 A02 Two channel calibrator RMS Pk Pk Switch The output polarity may any time be changed from positive to negative polarity ore vice versa OUTPUT 1 neg pos Figure 10 IST 300 A02 Two channel calibrator Polarity Switch IST 300 A02 MAN REV 1 03 10 2008 Page 11 of 12 meet en regeltechniek Each value may be changed at any time by selecting the applicable display and changing the value by using the incremental dial meee Pte es A E l er SH er i ray iat Figure 11 IST 300 A02 Two channel calibrator Incremental Dial Limitations are either the range values of each property or in case of the AC and DC selection the sum of both peak values When these reach a maximum of 16 000 mVolt peak the out put will be limited to this value and the Peak indicator LED will turn