Enhance DX USER MANUAL
Contents
1. gt 95 dip in UT gt 95 dip in UT for 5 sec for 5 sec Power Power frequency magnetic fields frequency 3 A m 3 A m should be at levels characteristic of a 50 60 Hz typical commercial or hospital environment magnetic field IEC 61000 4 8 NOTE UT is the a c mains voltage prior to application of the test level 12 Guidance and manufacturer s declaration electromagnetic immunity The air pump is intended for use in the electromagnetic environment specified below The customer or the user of the air pump is responsible for making sure it is used in such an environment Immunity IEC60601 test Compliance Electromagnetic environment guidance level level test Portable and mobile RF communications equipment should be used no closer to any part of the air pump including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Conducted 3 Vrms Recommended separation distance RF 150 kHz to 80 IEC MHz 3 Vrms d 1 2 YP 61000 4 6 3 V m d 1 2 ve 80 MHz to 800 MHz Radiated 80 MHz to 2 5 3 V m RF GHz d 2 3 YP 800 MHz to 2 5 GHz IEC 61000 4 3 where Pis the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a sho
2. frame Connect the mattress air hose to pump Check air hoses under the mattress to make sure they are not kinked Plug in the control unit and turn it on Oo OI D The pump will now inflate the mattress 10 Press the control button on pump to select alternation or static therapy 11 Set the required pressure by pressing the to increase pressure or to reduce pressure 12 Pressure is constantly monitored by the control unit When pressure is less than the selected pressure the Low Pressure LED will illuminate and an alarm will sound 13 For emergency mattress deflation turn the CPR valve to the OPEN position The mattress should deflate within 20 seconds To close the CPR simply turn the valve to the CLOSE position Open Close 6 Cleaning The Mattress The mattress should be cleaned weekly using a damp soft cloth and mild detergent The mattress cover can be washed and thermally disinfected in a washing machine by following the procedure below Do not use a phenol based cleaning solution Industrial Pre Wash Cold 10 minutes Main Wash 60 140 H 6 minutes Main Wash 70 158 H 10 minutes Spin Cycle 2 minutes 3 Cold Rinses Spin Cycle 5 minutes Domestic Pre Wash Cold Main Wash 70 158 H 10 minutes Spin Cycle 2 minutes Cold Rinses Spin Cycle 5 minutes Tumble Drying or Tunnel Drying is not recommended Mattress cells can be wiped down with a solution of
3. sodium hypochlorite 1000ppm or any other non phenolic germicidal solution The Master Control Unit N CAUTION SWITCH OFF THE ELECTRICAL SUPPLY TO THE PUMP AND DISCONNECT THE POWER CORD FROM THE MAIN SUPPLY BEFORE CLEANING AND INSPECTION The pump unit should also be cleaned weekly using a damp soft cloth and mild detergent The pump casing is manufactured from ABS plastic If the case is soiled the pump can be wiped down with a solution of sodium hypochlorite dilution 1000ppm or any EPA approved hospital grade disinfectant Do not use a phenol based cleaning solution The air filter should also be cleaned and checked at a minimum of every six months The air filter can be removed by removing filter cover and pinching center of the filter and pulling outward from the back of the cover Replace Air Filter 1 Remove air filter and replace with a new one 2 Usea soft bristle to remove dust and difficult dried on soil J 0 bx Filter 1 EG amp oil ET Waste Disposal This Product has been supplied by an environmentally conscious manufacturer and complies with the WEEE This product may contain substances that can be harmful to the environment if disposed of in places that are not approved by your state local or federal laws Please be environmentally responsible and recycle this product through your recycling facility at its end of life E 7 Storage and Care Master Contr
4. Enhance DX Alternating Anti Decubitus Overlay System USER MANUAL YA Prius Caremed Supply Inc 7F No 2 Lane 235 Bao Chiao Rd Xin Tien City Taipei 231 Taiwan Tel 886 2 2917 9808 Fax 886 2 2918 6505 Warning Connect the Master Control unit to a 220 240 volt to a power source Keep the pump and mattress away from open flame Keep the mattress away from sharp objects Do not place a heating device close to the mattress system A caution The Alternating System should always be used in accordance with your Institutions pressure care guidelines Re positioning of the patient is always recommended when using an Alternating Pressure Air Mattress APAM The Control unit can only be repaired by an authorized technician Do not drop the control unit Operation Temp 15 C 40 C R H 30 75 Yo Content 1 THE PURPOSE OF THIS MANUAL nesesvavnnvannesnavagnnenanennnrsaenennnvagonesennnnnsesenenent P 4 2 INTENDED USE sagnene eda euitet P 4 5 PRODUCT DESCRIPTION NG P 4 Master Control Unit Features Mattress Features A FECHNICAL DA P 6 Master Control Unit Enhance DX Mattress Replacement 5 OPERATION INSTRUCTION vesvssvrnvsnesvernesnennsnvennsnvsnvssesnesnrsnennesvennsnvsnnsnvsnesnenn P7 O CLEANING EN P 8 The Mattress The Master Control Unit Replace Air Filter Waste Disposal 7 STORAGE AND CARE NE P 10 8 MAINTENANCE AND TROUBLESHOOTING nmvsrsesrvrvvvnenenevevsvsesensvev
5. OSE position Replace damaged air cell if necessary Refer to qualified service technician if problem persist Pump is noisy Make sure pump is resting against a solid surface 1 Reposition the pump 2 Refer to qualified service technician if problem persist 10 9 EMC Related Notifications Guidance and manufacturer s declaration electromagnetic emissions The air pump is intended for use in the electromagnetic environment specified below The customer or the user of the air pump is responsible for making sure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions Group 1 The air pump uses RF energy only for its internal CISPR 11 function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The air pump is suitable for use in all CISPR 11 establishments including domestic and those Harmonic emissions Class A directly connected to the public low voltage IEC 61000 3 2 power supply network that supplies buildings Voltage fluctuations Complies Used LOR domes HC PURD OSE flicker emissions IEC 61000 3 3 Recommended distances between portable and mobile RF communications equipment and the air pump The air pump is intended for use in an electromagnetic environment in which radiated RF dis
6. arized design for easy air cell replacement vapor permeable oversized pliable quilted nylon top cover provides low shear friction and moisture protection CPR quick release for rapid deflation Integrated power cable management for organization and safety 4 Technical Data Master Control Unit Model No M16 6 Part No FC PHE0004 Size mm 258 L x110 W x210 H Weight Kg 2 5 kg Cycle Time min 12 min Min Max Pressure 16 32 mmHg 6 mmHg Max Flow rate Saer Rated Voltage AC 220 240 volt Max Current 0 1A Fuse Rating T1AH 250V Rated Frequency 50 Hz Protection Type Class II Type BF o Not AP or APG Type Mode of Operation Continuous Mattress Overlay Mattress type 5 one piece overlay Foam Model No FM PHE0003 Size mm 190cm L x 90cm W x 18cm H Weight Kg 5 5 Kg Cells Material PU coated Nylon Cover Type Full Cover with zipper Cover Material PU coated Nylon Base Material PVC mesh Cells Number 18 Weight capacity 160 kg 5 Operation instruction 1 Unpack the system and place the pump at the foot end of the bed 2 Remove the existing mattress from the bed frame W Place the Enhance DX mattress on the frame and position mattress so the air tube is at the foot of the bed Secure the Enhance DX mattress straps to the bed frame Hang the control unit on the foot board of the bed
7. ed in such an environment Immunity test IEC 60601 Compliance level Electromagnetic environment test level guidance Electrostatic 216 kV contact 216 kV contact Floors should be wood concrete or discharge ESD p8 kV air 28 kV air ceramic tile If floors are covered with IEC 61000 4 2 synthetic material the relative humidity should be at least 30 Electrical fast 22 kV for power 22 KV for power The Main power quality should be that transient burst supply lines supply lines of a typical commercial or hospital environment IEC 61000 4 4 211 kV for B1 kV for input output input output lines lines Surge 21 KV line s to 21 KV line s to The Main power quality should be that IEC 61000 4 5 line s line s of a typical commercial or hospital environment 22 kV line s to 22 kV line s to earth earth Interruptions lt 5 Yo UT lt 5 UT The Main power quality should be that and gt 95 dip in UT gt 95 dip in UT of a typical commercial or hospital voltage for 0 5 cycle for 0 5 cycle environment If the user of the air variations pump requires continued operation on power supply 40 UT 40 UT during power interruptions it is input lines 60 dip in UT 60 dip in UT recommended that the air pump is IEC 61000 4 11 for 5 cycles for 5 cycles powered from an_ uninterruptible power supply or a battery 70 UT 70 Yo UT 30 dip in UT 30 dip in UT for 25 cycles for 25 cycles lt 5 UT lt 5 UT
8. ersenenens P 10 9 EMC RELATED NOTIFICATION wwanveanwwanzanuna waaina misa manuaa aaa P 11 1 The Purpose of this Manual This operation manual is mainly focused on the set up cleaning and routine maintenance of the Enhance DX Alternating Anti Decubitus System We recommend keeping this manual available to answer questions related to the system 2 Intended use The Enhance DX Alternating Pressure Overlay System is primarily used for the treatment and prevention of decubitus ulcers up to stage II The System comprises of the latest technology in alternating mattress therapy which enables the mattress overlay to perform accurate pressure setting to individual patients needs The cells of the mattress are spatially oriented above a two inch convoluted foam base to provide extra protection and comfort to the patient Enhance DX Overlay System is one of the most advanced active pressure area Management systems 3 Product Description The Enhance DX Control Unit Features LED indicator for function status amp Electronic adjustable comfort setting amp 2 in 1 alternation with 10 minutes cycle time Function Status LED Indication Adjustable Comfort Setting Alternate Static ON OFF Control Switch Switch POWER POWER FAILURE LOW PRESSURE STATIC Mattress Features Therapeutic micro low air loss helps manage moisture and provides alternating therapy to prevent and treat pressure ulcers Modul
9. ol Unit amp Check the power cord and plug for abrasions and excessive wear Plug in the unit and verify air flow from the hose connection ports Place in plastic bag for storage Mattress amp Check the air manifold for kinks or breaks and replace if necessary amp Set CPR valve to OPEN and disconnect the air hose to the pump The mattress will now deflate and can be packed for storage It is recommended that the following proceedings are used whenever the system is being stored or transported to another location Temperature limitations 5 C 41 F 60 C 140 F Relative humidity 30 75 8 Maintenance amp Troubleshooting No daily maintenance is required This equipment should only be serviced by a qualified and authorized technician refer to the chart below For common trouble shooting tips please Symptom Inspection Procedures Possible Solution The pump is not functioning Check power source connection Check fuse Connect to proper power source 2 Replace fuse Refer to qualified service technician if problem persist Low pressure LED is constantly illuminated or the mattress is not inflating while pump is in operation Check hoses and hose connections 2 Check CPR valve 3 Check air cells for holes or tears other than where designed Make sure all connections are secure Make sure CPR valve is in the CL
10. turbances are controlled The customer or the user of the air pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the air pump as recommended below according to the maximum output power of the communications equipment Rated maximum Separation distance according to frequency of transmitter output power of m transmitter 150 kHz to 80 MHz 80MHzto 800 800 MHz to 2 5 GHz W d 1 2 YP MHz d 2 3 VP d 1 2 YP 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 11 Guidance and manufacturer s declaration electromagnetic immunity The air pump is intended for use in the electromagnetic environment specified below The customer or user of the air pump is responsible for making sure it is us
11. uld be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol HD NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered Ifthe measured field strength in the location in which the air pump is used exceeds the applicable RF compliance level above the air pump should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the air pump b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m 13