Home
        User Manual TDS 500B, TDS 600B & TDS 700A Digitizing
         Contents
1.                                                                                                                      Original Position Positioned Vertically Positioned Horizontally                                                                                                                                  Original Scale Scaled Vertically Scaled Horizontally    Figure 3 5  Scaling and Positioning    To Check the Check the Vertical Readout at the bottom left part of the display to read the  Vertical Scale    volts division setting for each displayed channel  the selected channel is in  inverse video    See Figure 3   6      To Change Vertical Scale The TDS Oscilloscope permits you to change vertical scale and position quickly  and Position from the front panel using dedicated control knobs  To change the vertical scale  and position     1  Turn the vertical SCALE knob  Note only the scale of the selected wave   form changes     As you turn the vertical SCALE knob clockwise  the value decreases  resulting in higher resolution because you see a smaller part of the wave   form  As you turn it counterclockwise  the scale increases allowing you to  see more of the waveform but with lower resolution     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 11    Acquiring and Displaying Waveforms       2  Turn the vertical POSITION knob  Again  note that only the selected  waveform changes position     3  To make positioning faster  press the SHIFT button  When the light above  the SHI
2.                                                Figure 3 27  InstaVu Display    To Set the InstaVu Style To change the InstaVu display style  do the following steps   1  Press DISPLAY     Mode  main   gt  Insta Vu  pop up   gt  Style  main      2  Select between Vectors and Dots in the side menu   Dots display is the  factory default setting      3  Select between Infinite Persistence and Variable Persistence in the side  menu   Variable Persistence is the factory default setting      4  Use the general purpose knob or keypad to adjust the persistence time  decay  rate  if you have selected Variable Persistence     To Set the InstaVu To change the InstaVu readout options  do the following steps     Readout Options    Press DISPLAY     Mode  main   gt  Insta Vu  pop up   gt  Readout  Options  main      2  Toggle Display T Trigger Point  Trigger Bar Style and Display Date   Time in the side menu to the settings desired     Incompatible Modes Several modes features are unavailable when InstaVu mode is selected     m FastFrame  Limit Testing  XY display  and Zoom modes    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 47    Acquiring and Displaying Waveforms       m Envelope  Average  Hi Res  and Single Acquisition Sequence acquisition  modes and Autosave mode    m Intensified time base  m Record lengths longer than 500 samples  m Interpolation  equivalent time sampling is used instead     m Vectors when in equivalent time mode  waveforms are displayed as Dots  instead    To det
3.                  f The Tektronix TDS Oscilloscope is a superb tool for acquiring  displaying  and  measuring waveforms  Its performance addresses the needs of both benchtop lab  and portable applications with the following features     m A maximum sample rate of up to 5 GS s per channel  depending on the  model  see Table 1 1 Key Features and Differences of by Models     m A analog bandwidth of 1 GHz or 500 MHz  depending on the model  see  Table 1 1     m Records lengths up to 50 K standard and 500 K with Option 1M  depending  on the model  see Table 1 1     m Four channel or 2   2 channel operation  depending on model   Two plus  Two channel operation allows two of four channels to be displayed  simultaneously   All channels have 8 bit resolution   See Table 1 1      m Trigger modes include edge  logic  and pulse  Video trigger modes  available  with option 05 only  include NTSC  SECAM  PAL  HDTV  and FlexFor   mat        m Dual Window Zoom  which shows a waveform magnified and unmagnified  on the same display     on all TDS 600B and TDS 700 Oscilloscopes    m Sample  envelope  average  high res  peak detect and InstaVu    acquisition  mode  which updates the display at rates rivaling the fastest analog oscillo   scopes  see Table 1   1 for models and modes available           KY    TDS 500B  TDS 600B   amp  TDS 700A User Manual 1 1             Product Description       Differences by Model    Table 1 1 lists some key TDS features and relates them to the different  TDS mode
4.         2 13  Figure 2 7  TRIGGER Controls           0  c cece cece eee eeeees 2 14  Figure 2 8  AUTOSET Button Location               0eceeeeeee 2 14  Figure 2 9  The Display After Pressing Autoset                6   2 15  Figure 2 10  Display Signals Requiring Probe Compensation        2 15  Figure 2 11  The Channel Buttons and Lights                     2 16  Figure 2 12  The Vertical Main Menu and Coupling Side Menu     2 17  Figure 2 13  The Menus After Changing Channels                2 18  Figure 2 14  Measure Main Menu and Select Measurement   Side  Menu crer save ea ee dinin seh ed sane sede adie EEEE 2 20  Figure 2 15  Four Simultaneous Measurement Readouts           2 21  Figure 2 16  General Purpose Knob Indicators                   2 22  Figure 2 17  Snapshot of Channel 1              0    cee eee eeee 2 23  Figure 2 18  Save Recall Setup Menu             cccccececceees 2 25  Figure 3 1  How Probe Compensation Affects Signals              3 4  Figure 3 2  P6139A Probe Adjustment                 eee e eee eee 3 5  Figure 3 3  The Channel Readout               ccceccscsccsees 3 9  Figure 3   4  Waveform Selection Priority               0e ce eeeee 3 10  Figure 3 5  Scaling and Positioning              cceccecseeeees 3 11  Figure 3 6  Vertical Readouts and Channel Menu                 3 12  Figure 3 7  Record View and Time Base Readouts                 3 14  Figure 3 8  Horizontal Controls             ccc ccece cece cceees 3 15    Figure 3 9  Acquisition  Input Anal
5.        cccceccscscceces 3 85  Figure 3 49  How the Delayed Triggers Work                 0   3 87  Figure 3 50  Delayed Trigger Menu             cececcecsccees 3 88    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Table of Contents       Figure 3 51   Figure 3 52   Figure 3 53   Figure 3 54   Figure 3 55   Figure 3 56   Figure 3 57   Figure 3 58   Figure 3 59   Figure 3 60   Figure 3 61   Figure 3 62   Figure 3 63   Figure 3 64   Figure 3 65   Figure 3 66   Figure 3 67   Figure 3 68   Figure 3 69   Figure 3 70   Figure 3 71   Figure 3 72     Figure 3 73   Hardcopy Device    Figure 3 74   aPC       Figure 3 75   Figure 3 76   Figure 3 77   Figure 3 78   Figure 3 79   Figure 3 80   Figure 3 81   Figure 3 82   Figure 3 83   Figure 3 84   Figure 3 85   Figure 3 86   Figure 3 87     Graticule  Cursor and Automated Measurements       Measurement Readouts              cc eee eee e ones  Measure Menu            cece rece c cece cree S  Measure Menu     Gating            cc cee ceececees  Measure Menu     Reference Levels                   Measure Delay Menu     Delay To                    Snapshot Menu and Readout                eeeeeee  Cursor Types    Cursor Modes    H Bars Cursor Menu and Readouts    Paired Cursor Menu and Readouts    Performing a Signal Path Compensation               Probe Cal Menu and Gain Compensation Display       Re use Probe Calibration Data Menu                  Save Recall Setup Menu             cece cceccccecs  Save Waveform Menu             cece ec 
6.       Set to TTL                pA   Set to ECL  Glitch                      i N Runt i T  am 1  width   Ms0 0ns Runt Chi  4     Slew Rate   ii                Oloo           Timeout      ONEN pi  Fier t _  hype Class source Polarity when hresholds   amp    lt  Pulse  gt   lt Runt gt  Occurs Holdoff    Kees   oe   oe  e   ee  ee                                                         Figure 3 45  Main Trigger Menu     Runt Class    Set the Thresholds  To set the two threshold levels used in detecting a runt pulse     1  Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Runt  pop up   gt  Thresholds  main      2  Use the general purpose knob or keypad to set the values for the high and  low thresholds     Hint  To use the Trigger Bar feature to set the threshold levels on the pulse  train  press DISPLAY     Readout Options  main   gt  Trigger Bar Style   side  until Long appears in that menu item     Note the position of the trigger indicator in Figure 3   45  Triggering occurs at the  point the pulse returns over the first  lower  threshold going negative without  crossing the second threshold level  upper   The polarity selected in the Polarity  side menu determines the order that the threshold must be crossed for a runt  trigger to occur     Positive requires that the Jower threshold must be first crossed going positive   then recrossed going negative without the upper threshold being crossed at all     Negative requires that the upper threshold
7.       Waveform Record  gt  10 K       Zero Phase  Reference    Figure 3 90  Waveform Record vs  FFT Time Domain Record    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 157    Using Features for Advanced Applications       3 158    FFTs Transform Time Records to Frequency Records  The FFT time domain  record just described is input for the FFT  Figure 3   91 shows the transformation  of that time domain data record into an FFT frequency domain record  The  resulting frequency domain record is one half the length of the FFT input  because the FFT computes both positive and negative frequencies  Since the  negative values mirror the positive values  only the positive values are displayed     FFT Time Domain Record    FFT Frequency Domain Record    Figure 3 91  FFT Time Domain Record vs  FFT Frequency Domain Record    FFT Frequency Range and Resolution  When you turn on an FFT waveform  the  oscilloscope displays either the magnitude or phase angle of the FFT frequency  domain record  The resolution between the discrete frequencies displayed in this  waveform is determined by the following equation     _ Sample Rate    AF   FFT Length    Where  AF is the frequency resolution   Sample Rate is the sample rate of the source waveform     FFT Length is the length of the FFT Time Domain waveform  record     The sample rate also determines the range these frequencies span  they span from  0 to   the sample rate of the waveform record   The value of 4 the sample rate is  often referre
8.      To make sure that the Main Trigger menu settings are compatible with Delayed  Triggerable and to select that mode  do the following steps     1  Press TRIGGER MENU     2  If Type is set to Logic  press Type  main  to change it to either Edge or  Pulse as fits your application  Logic type is incompatible with Delayed  Triggerable     3  If Source is set to Auxiliary  press Source  main   Select any source other  than Auxiliary from the side menu according to your application     4  If Type is set to Pulse  press Class  main  and change it to either Glitch or  Width as fits your application  Runt and Slew Rate pulse classes are  incompatible with Delayed Triggerable     5  Press HORIZONTAL MENU  gt  Time Base  main   gt  Delayed Only   side   gt  Delayed Triggerable  side      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       F Pretrigger Record Posttrigger Record  Delayed Runs After Main    Main Triger Point   lt  lt  Delayed Trigger Waveform Record                    gt     Main   Trigger   Source a  a ES     From Horiz Menu  Start Posttrigger Acquisition    Delayed Triggerable By Events       Main Trigger Point k alk Trigger Waveform Record                    Main j  Trigger    Source      Delayed N N   N N N N   N     N  Trigger      Source  N PSS ae  ers Start P osttrigger Acquisition  Trigger  Waiting for nth Event on nth Delayed Trigger Event    Where n 5     Delayed Triggerable By Time  k pa Trigger Waveform Record                 gt  
9.      ccc cece secs cece cesses cecens 3 49  Figure 3 29  Horizontal Menu     FastFrame Setup                3 49  Figure 3 30  Triggered Versus Untriggered Displays               3 51  Figure 3 31  Trigger Holdoff Time Ensures Valid Triggering        3 54  Figure 3 32  Slope and Level Controls Help Define the Trigger      3 55  Figure 3 33  TRIGGER Controls and Status Lights                3 56  Figure 3 34  Example Trigger Readouts     Edge Trigger Selected    3 58  Figure 3 35  Record View  Trigger Position  and Trigger Level Bar   READ OUUS oori 656s inetr se Dia oe ER SS eee Mee wine ee ee ee 3 58  Figure 3 36  Edge Trigger Readouts             ccccccecsceeees 3 59  Figure 3 37  Main Trigger Menu     Edge Type                   3 60  Figure 3 38  Violation Zones for Setup Hold Triggering            3 65  Figure 3 39  Logic Trigger Readouts     State Class Selected        3 66  Figure 3 40  Logic Trigger Menu              0  cece cece recone 3 67  Figure 3 41  Logic Trigger Menu     Time Qualified TRUE         3 69  Figure 3 42  Triggering on a Setup Hold Time Violation            3 72  Figure 3 43  Pulse Trigger Readouts            ccccccecscceces 3 73  Figure 3 44  Main Trigger Menu     Glitch Class                  3 75  Figure 3 45  Main Trigger Menu     Runt Class                   3 78  Figure 3 46  Main Trigger Menu     Slew Rate Class               3 82  Figure 3   47  Delayed Runs After Main                0  cece eeee 3 85  Figure 3 48  Delayed Triggerable       
10.     O    OFF  Real Time Only   Acquire menu  3 25  Off Bus  Utility menu  3 136  Offset  DC  See DC Offset  Vertical  3   14  vertical  3 159  3 170  3 175  Offset  Vertical menu  3   14  OK Confirm Clear Spool  Hardcopy menu  3 131  OK Create Math Wfm  More menu  3 149  OK Create Measurement  Measure Delay menu  3   99  OK Erase Ref  amp  Panel Memory  Utility menu  3 117  OK Store Template  Acquire menu  3 145  ON  Enable ET   Acquire menu  3 25          Index 10    ON STBY button  1 5  2 3   Optical probes  E 5   Options  A 1   Options  Color menu  3 38   OR  Glossary   7   OR  Main Trigger menu  3   68  3   70  Oscilloscope  Glossary   7   Overall  Display menu  3 31   Overshoot  Glossary   8   Overwrite Lock  File Utilities menu  3 125    P    P6205 Active Probe  1 3  Packaging  C 1  Paired cursor  3   101  PAL  Display menu  3 33  Palette  Color menu  3   35  Palette  Hardcopy menu  3 128  Passive voltage probes  E   1  Pattern trigger  3 62  How to setup  3 66  PCX  3 126  PCX Color  Hardcopy menu  3 127  PCX  Hardcopy menu  3 127  Peak detect acquisition mode  3 23  Glossary   7  Peak Detect  Acquire menu  3 25  Peak to peak  3 93  Glossary   8  Period  3   93  Glossary   8  Persistence  3   30  Persistence Palette  Color menu  3   35  Phase  3 93  Glossary   8  Phase suppression  3   163  Pixel  Glossary   8  Polarity and Width  Main Trigger menu  3 76  Polarity  Main Trigger menu  3 77  3 80  Pop up menu  2 8  Glossary 8  Port  Hardcopy menu  3 128  Port  Utility me
11.     The topics just listed contain steps that you perform to accomplish the task that  the topic defines  You should read Conventions on page xii of Preface before  reading about these tasks     Each topic just listed comprises more basic operation tasks and topics  A list of  these tasks follows     Coupling Waveforms to the Oscilloscope                  0 000000 3 3  Setting up Automatically  Autoset and Reset                   004  3 6  Selecting  Channels  acri smegenis oad sales Mea aed baw oe ee 3 8  Scaling and Positioning Waveforms                 000 c eee eens 3 10  Choosing an Acquisition Mode               0 0  e cece eee eens 3 17  Customizing the Display            0 0 00 cee cece 3 29  Customizing the Display Color            0 0    cece eee ene 3 34    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 1    Overview       Triggering on Waveforms    Measuring Waveforms    Saving Waveforms and  Setups    Determining Status and  Accessing Help    Using Features for  Advanced Applications    Zooming on Waveforms           0 0    eee eee eee eee eee ees 3 39  Using InstaVu   Acquisition Mode  TDS 600B and 700A Models Only  3 45  Using FastFrame    TDS 500B and 700A Models Only               3 48  Triggering Concepts isogo 0    cece eect ene 3 51  Triggering from the Front Panel               0 0 0    eee eee eee 3 55  Triggering on a Waveform Edge             0    0c eee eee eee 3 59  Triggering Based on Logic         0 0    cece cee eee 3 62  Triggering On Pulses   
12.    175  Clock Source  Main Trigger menu  3 71  Collision Contrast  Color menu  3   38  Color  3 35  How to set  3 34  Color Deskjet  3 126  Color Matches Contents  Color menu  3   37  3   38          TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index       Color menu  3 35   Bold  3 35   Change Colors  3 36   Collision Contrast  3   38   Color  3 36  3 37   Color Matches Contents  3   37  3   38   Hardcopy  3 35   Hue  3 36   Lightness  3 36   Map Math  3 37   Map Reference  3   37   Math  3 37   Monochrome  3 35   Normal  3 35   Options  3   38   Palette  3   35   Persistence Palette  3   35   Ref  3 37   Reset All Mappings To Factory  3   38   Reset All Palettes To Factory  3 38   Reset Current Palette To Factory  3 38   Reset to Factory Color  3   36   Restore Colors  3   38   Saturation  3   36   Spectral  3 35   Temperature  3 35   View Palette  3 35  Color  Color menu  3   36  3   37  Color  Display menu  3 35  Compare Ch1 to  Acquire menu  3 146  Compare Ch2 to  Acquire menu  3 146  Compare Ch3 to  Acquire menu  3 146  Compare Ch4 to  Acquire menu  3 146  Compare Math  to  Acquire menu  3 146  Compare Math2 to  Acquire menu  3 146  Compare Math3 to  Acquire menu  3 146  Compensation  of passive probes  3   4  Configure  Utility menu  3 127  3 136  Confirm Delete  File Utilities menu  3 125  Connector   AUX TRIGGER INPUT  2 5   Centronics  2   5   DELAYED TRIGGER OUTPUT  2 5   GPIB  2 5  3 135   MAIN TRIGGER OUTPUT  2 5   Power  2 5   RS 232  2 5   SIGNAL OUTPUT  2
13.    Main Trigger Point    Main    Trigger                l     Source         Delayed      Trigger    Source    ee re a Start P osttrigger Acquisition  ying  First Trigger After Delay   Delayed Triggerable By Events Time  Main Trigger Point k elis Trigger Waveform Record              gt    Main      Trigger    Source      l  Trigger    Source      a  a  Waiting for nth Event H  Time Delay     Start Posttrigger Acquisition   Where n 4   From Delay Trig Menu     Figure 3 49  How the Delayed Triggers Work    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 87    Triggering on Waveforms       3 88          NOTE  The Delayed Triggerable menu item is not selectable unless incompatible  Main Trigger menu settings are eliminated   See the steps at the beginning of  this procedure   If such is the case  the Delayed Triggerable menu item is dimmer  than other items in the menu        By pressing Intensified  side   you can display an intensified zone that shows  where the delayed timebase record may occur  a valid delay trigger event  must be received  relative to the main trigger on the main time base  For  Delayed Triggerable After mode  the start of the intensified zone corresponds  to the possible start point of the delayed time base record  The end of the  zone continues to the end of main time base  since a delayed time base  record may be triggered at any point after the delay time elapses     To learn how to define the intensity level of the normal and intensified  waveform  see
14.    TDS 500B  TDS 600B   amp  TDS 700A User Manual F 1    Appendix F  Cleaning       F 2    Table F 1  External Inspection Check List    Item    Cabinet  front panel   and cover    Inspect For Repair Action    Cracks  scratches  deformations    Send in for service    damaged hardware or gaskets       Front panel knobs    Missing  damaged  or loose Send in for service    knobs       Carrying handle  bail   cabinet feet        Correct operation Send in for service       Cleaning Procedure     Exterior     To clean the oscilloscope exterior  do the following steps     1  Remove loose dust on the outside of the oscilloscope with a lint free cloth     2  Remove remaining dirt with a lint free cloth dampened in a general purpose  detergent and water solution  Do not use abrasive cleaners     3  Clean the light filter protecting the monitor screen with a lint free cloth  dampened with either isopropyl alcohol or  preferably  a gentle  general  purpose detergent and water solution        CAUTION  To prevent getting moisture inside the oscilloscope during external  cleaning  use only enough liquid to dampen the cloth or applicator        Lubrication  There is no periodic lubrication required for this oscilloscope     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Glossary    ACM          2   2 channel operation  Two plus two channel operation limits the simultaneous display of channels  to two of the four channels provided  Channels not displayed can be used to  couple a triggering 
15.    the selected channel     Each menu item in the Vertical menu displays a side menu  Right now  the  Coupling item in the main menu is highlighted  which means that the side  menu shows the coupling choices  At the top of the side menu  the menu title  shows the channel affected by the menu choices  That channel always  matches the lighted channel button           Tek Run  eae Sample       ch2 Coupling      _ Side Menu Title  Impedance             Ch2 Reference Indicator             1                                           ch 200mv 100mv_ M 500s Chi f 264m       i       E 50   LEN Bandwidth moone Position   Offset   Deskew   Cal Probe  DC full 2 94 div ov as initialized    Oloo o        fdiv                                                  111 191        Figure 2 12  The Vertical Main Menu and Coupling Side Menu    TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 17    Tutorial       7  Press Q  side  to toggle the selection to 50 Q  That changes the input  coupling of channel 2 from 1 MQ to 50 Q  The channel readout for  channel 2  near the bottom of the graticule  now shows an Q indicator     Assign Controls to Pressing a channel  CH  button sets the vertical controls to that channel  It also  Another Channel adds the channel to the display if that waveform is not already displayed  To  explore assigning controls to different channels  do the following steps     1  Press CH 1     Observe that now the side menu title shows Ch1   See Figure 2   13   and that  the light ab
16.   3 76  AND  3 68  3 70  Auto  3 61  Ch1  Ch2      3 60  3 67  3 70  3 71  3 75  3 77   3 79  3   80  3 83  Class  3 79  3 83  Pattern  3   66  Runt  3 77  Setup Hold  3 70  Slew Rate  3 80  State  3 69  Class Glitch  3   75  Clock Source  3 71  Coupling  3 60  Data Source  3   70  DC  3 60  Define Inputs  3   67  3 69  3 71  Define Logic  3   68  3   70  Delta Time  3 81  Edge  3 59  3 60  Either  3 76  3 77  3 80  Falling edge  3 70  3 71  Glitch  3 76  Goes FALSE  3 68  3 70  Goes TRUE  3 68  3 70  HF Rej  3   60  Level  3   62  3 76  3   80  3   84  LF Rej  3   60  Mode  amp  Holdoff  3   61  NAND  3 68  3 70  Negative  3 76  3 77  3 80  Noise Rej  3   60  NOR  3 68  3 70  Normal  3   61          Index 8    OR  3 68  3 70   Polarity  3 77  3 80   Polarity and Width  3 76  Positive  3   76  3 77  3 80  Pulse  3   59  3 77  3 79  3 83  Reject Glitch  3   76   Rising edge  3 70  3 71  Runt  3 77   Set Thresholds  3   67  3   70             Set to 50   3 56  3 62  3 76  3 84  Set to ECL  3 62  3 76  3 80  3 84  Set to TTL  3   62  3 76  3 80  3 84        gt     Slope  3 62  Source  3   60  3 75  3 77  3 79  3 80  3 83  State  3 69  3 70  Thresholds  3   78  3   80  Time  3   84  Timeout  3 83  3 84  Trigger When  3   68  3 70  3 81  True for less than  3   68  True for more than  3   68  Type  3 59  3 60  3 79  3 83  Logic  3 66  3 69  3 70  Pulse  3   80  Pulse  3 77  Type Pulse  3 75  Width  3 76  3 79  MAIN TRIGGER OUTPUT  BNC  2 5  Map Math  Color menu  3   37  Map Refere
17.   Operation do the following tasks to make sure the instruments support GPIB protocols and  observe GPIB Interface requirements     Check for GPIB Protocols  Make sure the instruments to be connected support the  GPIB protocols  These protocols cover         Remote instrument control    Bidirectional data transfer  m Device compatibility   m Status and event reporting    To simplify the development of GPIB systems  include instruments that use  Tektronix defined codes and formats for messages that travel over the GPIB   Each device that follows these codes and formats  such as this oscilloscope   supports standard commands  Use of instruments that support these commands  can greatly simplify development of GPIB systems     Know the GPIB Interface Requirements  To prepare to connect the oscilloscope to  GPIB networks  read and follow these rules     m Connect no more than 15 devices  including the controller  to a single bus     m Connect one device load every two meters  about six feet  of cable length to  maintain bus electrical characteristics   Generally  each instrument represents  one device load on the bus      m Do not exceed 20 meters  about 65 feet  of the total cumulative cable length     m Turn on at least two thirds of the device loads present when you use your  network     3 134 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       m Include only one cable path between devices on your network   See  Figure 3 75   Do not create loop c
18.   Options and Accessories       Accessory Software The optional accessories listed in Table A   4 are Tektronix software products    recommended for use with your oscilloscope     Table A 4  Accessory Software       Software Part Number  Wavewriter  AWG and waveform creation  3FT400  WaveStar     Waveform capture and documentation WSTR31  LabWindows  3FG910  Telecommunication Package and i Pattern Software TTiP          Warranty Information Check for the full warranty statements for this product and the products listed  above on the first page after the title page of each product manual     Service Assurance Tektronix offers the following services that you can purchase any time during the    warranty period of this product     m REPXXXX provides one year of post warranty repair support  It is available    in One year increments up to three years     m CALXXXX provides one year of calibration support  It is available in one    year increments up to five years     TDS 500B  TDS 600B   amp  TDS 700A User Manual    A 7    Appendix A  Options and Accessories       A 8 TDS 500B  TDS 600B   amp  TDS 700A User Manual                          E   Appendix B  Algorithms    The TDS Oscilloscope can take 25 automatic measurements  By knowing how it  makes these calculations  you may better understand how to use your oscillo   scope and how to interpret your results     Measurement Variables    The oscilloscope uses a variety of variables in its calculations  These include     High  Low H
19.   Select Measrmnt  main   gt  Delay  side   gt   Edges  main   A side menu of delay edges and directions will appear   Choose from one of the combinations displayed on the side menu using the  following information     m The selection you choose defines which edges you want the delayed  measurement to be made between     m The upper waveform on each icon represents the from waveform and the  lower one represents the fo waveform     m The direction arrows on the choices let you specify a forward search on  both waveforms or a forward search on the from waveform and a  backwards search on the to waveform  The latter choice is useful for  isolating a specific pair of edges out of a stream     4  To take the measurement you just specified  press Delay To  main   gt  OK  Create Measurement  side      To exit the Measure Delay menu rather than creating a delay measurement   press CLEAR MENU  which returns you to the Measure menu     Take a Snapshot of Sometimes you may want to see all of the automated measurements on screen at  Measurements the same time  To do so  use Snapshot  Snapshot executes all of the single  waveform measurements available on the selected waveform once and displays  the results   The measurements are not continuously updated   All of the  measurements listed in Table 3   8 on page 3 92 except for Delay and Phase are  displayed   Delay and Phase are dual waveform measurements and are not  available with Snapshot      The readout area for a snapshot of measure
20.   Trigger Holdoff    Vertical Scale       Default Holdoff  S et equal to 5 horizontal divisions  Adjustable Holdoff  250 ns    Selection in Mode and Holdoff menu determines whether the  default holdoff value or the adjustable hold value is used     As determined by the signal level       TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       To Reset the  Oscilloscope    Selecting Channels    To Identify the  Selected Channel    3 8    Table 3 1  Autoset Defaults  Cont            Control Changed by Autoset to  Vertical Bandwidth Full   Vertical Offset 0 volts   Zoom Off       Do the following steps to reset the oscilloscope to its factory default settings     1  Press the Save Recall SETUP button to display the Setup menu  see Fig   ure 3 3   Press the button directly below the Recall Factory Setup menu item     2  Press the button to the right of the OK Confirm Factory Init side menu item     3  Press the SET LEVEL TO 50  button  front panel  to be sure the  oscilloscope triggers on the input signal     The TDS Oscilloscope applies all actions based on a specific waveform  such as  taking measurements or applying any changes it receives to the vertical control  settings  to the selected waveform  You can select a channel waveform  a math  waveform  or a reference waveform  This subsection describes how to select a  waveform and how you can turn the display of a waveform off     To determine which channel is currently selected  check the ch
21.   W t is the continuous function obtained by linear interpolation of W t   A and B are numbers between 0 0 and RecordLength 1 0    If A and B are integers  then     B Belgii i    woa   sx 570r Mes  A i A    where s is the sample interval   Similarly     is approximated by         2 where      WO  at       W t  dt    A    W t  is the sampled waveform    W t is the continuous function obtained by linear interpolation of W t   A and B are numbers between 0 0 and RecordLength 1 0    If A and B are integers  then     i A    2 A 2 B 1 Rey    f i       WO  dt  s x py O UA ee    where s is the sample interval     Measurements on Envelope Waveforms    Time measurements on envelope waveforms must be treated differently from  time measurements on other waveforms  because envelope waveforms contain so  many apparent crossings  Unless otherwise noted  envelope waveforms use either  the minima or the maxima  but not both   determined in the following manner     1  Step through the waveform from Start to End until the sample min and max  pair DO NOT straddle MidRef     2  If the pair  gt  MidRef  use the minima  else use maxima     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix B  Algorithms       If all pairs straddle MidRef  use maxima  See Figure B 4     The Burst Width measurement always uses both maxima and minima to  determine crossings        MidR ef            Both min and max  samples are above  MidRef  so use minima          Both min and max  samples are below  MidRef 
22.   main   gt  Width  pop up   gt  Polarity  main   gt  Positive or  Negative  side      Set to Trig When  To set the range of widths  in units of time  the trigger source  will search for and to specify whether to trigger on pulses that are outside this  range or within this range  do the following steps     1  Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Width  pop up   gt  Trig When  main      2  Press Within Limits  side  if you want the oscilloscope to trigger on pulses  that fall within the specified range  If you want it to trigger on pulses that are  outside the range  then press Out of Limits  side      3  To set the range of pulse widths in units of time  press Upper Limit  side   and Lower Limit  side   Enter the values with the general purpose knob or  keypad  The Upper Limit is the maximum valid pulse width the trigger  source will look for  The Lower Limit is the minimum valid pulse width   The oscilloscope will always force the Lower Limit to be less than or equal  to the Upper Limit     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 79    Triggering on Waveforms       3 80    To Trigger Based  on Slew Rate    Set the Level   Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Width  pop up   gt  Level  main   gt  Level  Set to TTL  Set to  ECL  or Set to 50   side      Set the Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer t
23.   you can also display the voltage in  IRE units     V Bars  The value after A shows the time  or frequency  difference between the  cursors  The value after   shows the time  frequency  of the selected cursor  relative to the trigger point  With the video trigger option  you can also display  the line number     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       TDS 500B and TDS 700A Models Only  In FastFrame mode  the   shows the  time position of the selected cursor relative to the trigger point of the frame that  the selected cursor is in  The A shows the time difference between the two  cursors only if both cursors are in the same frame     Paired  The value after one A shows the voltage difference between the two Xs   the other A shows the time  or frequency  difference between the two long  vertical bars  The value after   shows the voltage at the X of the selected cursor  relative to ground   See Figure 3 61      TDS 500B and TDS 700A Models Only  In FastFrame mode  the A shows the  time difference between the two cursors only if both cursors are in the same  frame        a i    Tek Run  100kS 5 Average          Cursor    Cursor Readout  H Bars  ee     o    ranetion                Non selected Cursor   Dashed Line                    Selected Cursor   Solid Line                                20 0m        lt  M 500us Chif 25 6mV  a ii rine rashipmeacte Sear  tunction Mone  Units Units   gt  seconds Base                                          
24.  2 03 s0 scoce beth et bh hath tek eed Gok da wanders 3 72  Delayed  Trig serine    iie snl e ately eae rane eee aaa 3 84  Taking Automated Measurements              0 0 00  e eee eee eee 3 91  Taking Cursor Measurements             0 00    eee eee eee eee 3 101  Taking Graticule Measurements               0 0 02  e eee ee ee eee 3 105  Optimizing Measurement Accuracy  SPC and Probe Cal              3 106  Saving and Recalling Setups       0    0 0    eee eee eee 3 115  Saving and Recalling Waveforms             0 0    cece eee eee eee 3 118  Managing the File System        0 0 0    eee eee eee eee 3 121  Printing a Hardcopy eccess 64 bees desi etn eee hed bee ite 3 125  Communicating with Remote Instruments                      04  3 134  Displaying Status isse 2 6  eee eee eet ee tee EE 3 139  Displaying Help rere tacos Cae eater eae awe Gee Lew teeudad 3 3 141  Limit festing eei etc  s ete oa shige ii hee dow desea a Eh es ete 3 143  Waveform Math ee  ctdcie ae taeeteeree seat bet Sma eedae dea 3 148  Fast Fourier Transforms           0 0    ce eee eee eee eee 3 150  Waveform Differentiation      0 0 2    eee eee eee 3 167  Waveform Integration       0 0    ee eee eee nee 3 171    TDS 500B  TDS 600B   amp  TDS 700A User Manual    SSS                   Acquiring and Displaying Waveforms    To use the TDS Oscilloscope to measure or monitor waveforms  you need to  know how to acquire  select  and display those waveforms properly  To help you  do so  this section describes how
25.  22   Knob  Glossary   6  General purpose  2 7  2 21  Glossary   4  Horizontal POSITION  2 13  3 15  Horizontal SCALE  2 13  3 15  Trigger MAIN LEVEL  2 14  3 55  Vertical POSITION  2 13  3 11  Vertical SCALE  2 13  3 11    L    Labelling menu  Enter Char  3 123  3 124  Landscape  Hardcopy menu  3 128  Language options  Option L  A   4  Laserjet  3 126  Laserjet  Hardcopy menu  3 127  Layout  Hardcopy menu  3 128  Level  Delayed Trigger menu  3   89  Level  Main Trigger menu  3   62  3   76  3 80  3   84  Level  Trigger  3 55  LF Rej  Main Trigger menu  3   60  Lightness  Color menu  3 36  Limit Test Condition Met  Acquire menu  3 147  Limit Test Setup  Acquire menu  3 146  3 147  Limit Test Sources  Acquire menu  3 146  Limit Test  Acquire menu  3 147  Limit Testing  Incompatible with InstaVu  3   47  Limit testing  3 143  Linear interpolation  3   20  3 32  Glossary   5  Linear interpolation  Display menu  3 32  Logic trigger  3 52  3 64   Definitions  3   64   Pattern  3 63  Glossary   6   Readout  3 66   State  3 64  Glossary   6       Index 7    Index       Logic triggering  3   62   Logic  Main Trigger menu  3 59  Low  3 93  Glossary   6   Low impedance Zo probes  E 2  Low Ref  Measure menu  3   98  Lubrication  F   2    Main menu  Glossary   6  Main menu buttons  2 3  Glossary   6  Main Scale  Horizontal menu  3   16  Main Trigger Menu  Falling edge  3 62  Rising edge  3 62  Main Trigger menu  3 59  3 60  3   66  3 69  3   70   3 75  3 77  3 80  AC  3 60  Accept Glitch
26.  3 74  Table 3 8  Measurement Definitions              c cece ce eeeees 3 92  Table 3 9  Probe Cal Status 2         ccc ccc cece cece ce eee ee eees 3 114  Table A 1  Options si ss 6sie od eretan tnika nee n Se ire dls God lee 88 A 1  Table A 2  Standard Accessories           ccccceccccscscceces A 4  Table A 3  Optional Accessories           0c cece cece ccc cceees A 5  Table A   4  Accessory Software         cccc ccc cece ccc ccscceees A 7  Table D 1  Factory Initialization Defaults                  0008  D 1  Table F   1  External Inspection Check List                00008  F 2    TDS 500B  TDS 600B   amp  TDS 700A User Manual vii    Table of Contents       viii TDS 500B  TDS 600B   amp  TDS 700A User Manual    ae a  General Safety Summary    Review the following safety precautions to avoid injury and prevent damage to  this product or any products connected to it  To avoid potential hazards  use the  product only as specified     Only qualified personnel should perform service procedures     Injury Precautions Use Proper Power Cord  To avoid fire hazard  use only the power cord specified  for this product     Avoid Electric Overload  To avoid electric shock or fire hazard  do not apply a  voltage to a terminal that is outside the range specified for that terminal     Avoid Electric Shock  To avoid injury or loss of life  do not connect or disconnect  probes or test leads while they are connected to a voltage source     Ground the Product  This product is grounded through t
27.  5   VGA  2 5  Contrast  Display menu  3 31  Conventions  xii  Copy  File Utilities menu  3 124    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Coupling  2 17  Ground  Glossary   5  Selecting  3 13  Trigger  3 54  Coupling Waveforms  3 3  Coupling  Delayed Trigger menu  3 89  Coupling  Main Trigger menu  3   60  Coupling  Vertical menu  3 13  Create Directory  File Utilities menu  3 124  Create Limit Test Template  Acquire menu  3 144  Create Measrmnt  Measure Delay menu  3   99  Cross Hair  Display menu  3 33  Current probes  E   4  Cursor  Horizontal bar  3 101  Measurements  3 101  modes  3   102  Paired  3 101  readout  3 102  Setting adjustment response  speed   3 105  Vertical bar  3 101  CURSOR button  3 103  Cursor menu  3 103  3 155  3 173  1 seconds  Hz   3 105  Amplitude Units  3 105  Base  3 105  Function  3 104  H Bars  3 104  Independent  3 104  IRE  NTSC   3 105  seconds  3   105  Time Units  3 105  Tracking  3 104  Video Line Number  3   105  Cursor readout  H Bars  3 155  3 169  3 173  Paired  3 169  Paired cursors  3 156  3 174  V Bars  3 155  3 169  3 174  Cursors  3 101  Glossary   3  How to use  3 103  with derivative waveforms  3 169  with FFT waveforms  3   155  with integral waveforms  3 173  Cycle area  3   92  Glossary   3  Cycle mean  3 92  Glossary   3  Cycle RMS  3 92  Glossary   3    D    Data Source  Main Trigger menu  3   70    Index 3    Index       Date Time   On hardcopies  3 128   To set  3 129  DC coupling  Glossary   3  DC offset  3 
28.  684B  TDS 782A  and TDS 784A oscilloscopes come without probes  but  for general  purpose measurements and to take advantage of the 1 GHz  bandwidth of these oscilloscopes  the P6245 Active Probe is recommended  This  manual lists the P6245 optional accessory probe in Appendix A  Options and  Accessories        Passive Voltage Probes    General Purpose  High  Input Resistance  Probes    Passive voltage probes measure voltage  They employ passive circuit compo   nents such as resistors  capacitors  and inductors  There are three common  classes of passive voltage probes     m General purpose  high input resistance   m Low impedance  Zo     m High voltage    High input resistance probes are considered    typical    oscilloscope probes  The  high input resistance of passive probes  typically 10 MQ  provides negligible  DC loading and makes them a good choice for accurate DC amplitude measure   ments     However  their 8 pF to 12 pF  over 60 pF for 1X  capacitive loading can distort  timing and phase measurements  Use high input resistance passive probes for  measurements involving     m    Device characterization  above 15 V  thermal drift applications     m Maximum amplitude sensitivity using 1X high impedance    TDS 500B  TDS 600B   amp  TDS 700A User Manual E 1    Appendix E  Probe Selection       Low Impedance  Zo   Probes    High Voltage Probes    E 2    m Large voltage range  between 15 and 500 V     m Qualitative or go no go measurements    Low impedance probes measure f
29.  Adjust Intensity on page 3 31     Now you need to bring up the Delayed Trigger menu so that you can define  the delayed trigger event     6  Press SHIFT DELAYED TRIG  gt  Delay by  main   gt  Triggerable After  Time  Events  or Events Time  side   See Figure 3   50         Tek Run  50 0MS s Sample Delay by Time  16ns          Delayed Edge  Delay          i  ef  Triggerable  after Time          Events          ae  PA  Events Time          Time             i     M 200us Chif 304mv  v  Events  D1 00us Chi 304mv 2             Delay by Source Coupling Slope Level  Time chi DC f 304mV    DIGIC OIC IO  e                                                    Figure 3 50  Delayed Trigger Menu    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       7  Enter the delay time or events using the general purpose knob or the keypad   If you selected Events Time  use Time  side  and Events  side  to switch  between setting the time and the number of events     Hint  You can go directly to the Delayed Trigger menu   See step 6   By  selecting one of Triggerable After Time  Events  or Events Time  the  oscilloscope automatically switches to Delayed Triggerable in the Horizontal  menu  You will still need to display the Horizontal menu if you want to leave  Delayed Triggerable     The Source menu lets you select which input will be the delayed trigger  source     8  Press Source  main   gt  Ch1  Ch2  Ch3  Ch4  or DC Aux  side         NOTE  Selecting DC Aux as source in B
30.  AeA 3 141  Using Features for Advanced Applications               e ee008  3 143  Lint Vesun gs  ovate ise Ge oe ee SSR a a eek as Shh Bees 3 143  Wavetorm   Math cn  3 3 4 2 eters hae E 60 a eR Be See RF es 6 3 148  Fast Fourier Transforms            0    cece cee eee eee eens 3 150  Waveform Differentiation           0 0    eect tenes 3 167  Waveform Integration         00    ee ec eee tenes 3 171   Appendices  Appendix A  Options and Accessories            ccc ce ese ce cece A 1  Appendix B  Algorithms         sssssssssssssesesseseseseseee B 1  Appendix C  Packaging for Shipment        esesssssesesesesee  C 1  Appendix D  Factory Initialization Settings                  000  D 1  Appendix E  Probe Selection             cc cece cc cc ccs ccsccecs E 1  Appendix F  Inspection and Cleaning         ssssssseseseseses  F 1   Glossary   Index    ii TDS 500B  TDS 600B   amp  TDS 700A User Manual    Table of Contents       List of Figures    Figure 1 1  Rear Panel Controls Used in Start Up                 1 5  Figure 1 2  ON STBY Button           0  cece cece c ce ere eeeens 1 6    Figure 2 1  Connecting a Probe for the Examples  P6245 shown     2 9    Figure 2 2  SETUP Button Location                  cece eee eee 2 10  Figure 2 3  The Setup Menu           ccccecccccccccscscceces 2 10  Figure 2   4  Trigger Controls            cece cece ccc ce sce cceees 2 11  Figure 2 5  The Display After Factory Initialization               2 12  Figure 2 6  The VERTICAL and HORIZONTAL Controls 
31.  Bits    t                         Parity  None          Hard Flagging  ga ON                   mi peec 1 1  oomvs M40  ous chi  7     78mv  3 1  GPIB 4 2020  3    conrigure  Hardware Software  Or  Hardcopy Rats Setup    Oooo o                                                  Dooa       Figure 3 70  Utility Menu     System I O    Set Hardcopy Parameters  To specify the hardcopy format  layout  and type of  port using the hardcopy menu  do the following steps     1  Press SHIFT  gt  HARDCOPY MENU to bring up the Hardcopy menu     2  Press Format  main   gt  Thinkjet  Deskjet  DeskjetC  Laserjet  Epson   DPU 411  DPU 412  PCX  PCX Color  TIFF  BMP Mono  BMP Color   RLE Color  EPS Mono Img  EPS Color Img  EPS Mono Pit  EPS Color  Pit  Interleaf  or HPGL  side    Press more   side  to page through all of  these format choices      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 127    Saving Waveforms and Setups       3 128       NOTE  Some formats  such as DeskJetC  require up to several minutes to process  and print the screen When using these formats  be careful not to inadvertently  abort the print by pressing the Hardcopy button for a second print before the  oscilloscope has finished processing and transmitting the first one        3  Press SHIFT  gt  HARDCOPY MENU  gt  Layout  main   gt  Landscape or  Portrait  side    See Figure 3 71      Landscape Format Portrait Format    Figure 3 71  Hardcopy Formats    4  Press SHIFT  gt  HARDCOPY MENU  gt  Palette  main   gt  Hardco
32.  G                Figure 3 44  Main Trigger Menu     Glitch Class    Select the Source  To specify which channel becomes the pulse trigger source     Press TRIGGER MENU     gt  Type  main   gt  Pulse  pop up   gt   Source  main   gt  Ch1  Ch2  Ch3  or Ch4  side   The source selected becomes  the trigger source for all four trigger classes     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 75    Triggering on Waveforms       Select the Polarity  amp  Width  To specify polarity  positive  negative  or either  and  width of the glitch  do the following steps     1  Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt  Polarity  amp   Width  main   gt  Positive  Negative  or Either  side      4 Positive looks at positive going pulses          a Negative looks at negative going pulses                    aa Either looks at both positive and negative pulses   ie 2  Press Width  side   and set the glitch width using the general purpose knob  or keypad     Set to Accept or Reject Glitch  To specify whether to trigger on glitches or ignore  glitches  press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt    Class  main   gt  Glitch  pop up   gt  Glitch  main   gt  Accept Glitch or Reject  Glitch  side      If you choose Accept Glitch  the oscilloscope will trigger only on pulses  narrower than the width you specified  If you select Reject Glitch  it will trigger  only on pulses wider than the specified width     Set the Level  To set the trigger level with the Level main 
33.  GIB    Iae e       Figure 3 60  H Bars Cursor Menu and Readouts    Paired cursors can only show voltage differences when they remain on screen  If  the paired cursors are moved off screen horizontally  the word Edge will replace  the voltage values in the cursor readout     Selectthe This procedure and those that follow detail the process for taking a cursor    Cursor Function measurement  To select the type of cursors you want  do the following steps     1  To display the cursor menu  press CURSOR   See Figure 3   60      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 103    Measuring Waveforms       3 104    Cursor Readout  Paired     2  Press Function  main   gt  H Bars  V Bars  Paired  or Off  side      Position of Vertical Bar Cursors  Useful for  Locating Cursors Outside the Display        Tek Run  250kS 5 Avprage       Cursor  Function             Non selected Cursor           Dashed Vertical Bar     Selected Cursor              Solid Vertical Bar     Set Mode and Adjust  the Cursors                            p   3  10 0m   O M 200ms  Chi F 25 6mV  B   i    E Sii richie  Funetion units  seconds           OIOI o                                                     aa  G          Figure 3 61  Paired Cursor Menu and Readouts    To select the cursor mode and adjust the cursors in either mode  do the following  steps     1  Press CURSOR     Mode  main   gt  Independent or Track  side      Independent makes each cursor positionable without regard to the position of  the o
34.  MHz Passive probe    P6217 Active  high speed digital voltage probe  FET  DC to 4 GHz  DC  offset  50 Q input  Use with 1103 TekProbe Power Supply for offset control    P6204 Active  high speed digital voltage probe  FET  DC to 1 GHz  DC  offset  50 Q input  Use with 1103 TekProbe Power Supply for offset control    P6563AS Passive  SMD probe  20X  500 MHz  P5100 High Voltage Passive probe  2 5 kV  DC to 250 MHz  P5205 High Voltage differential probe  1 kV  DC to 100 MHz    AM 503S     DC AC Current probe system  AC DC  Uses A6302 Current  Probe    AM 503S Option 03  DC AC Current probe system  AC DC  Uses A6303  Current Probe    TCP 202 Current Probe  DC to 100 Mhz  P6021 AC Current probe  120 Hz to 60 MHz  P6022 AC Current probe  935 kHz to 120 MHz    CT 1 Current probe     designed for permanent or semi permanent in circuit  installation  25 kHz to 1 GHz  50 Q input    CT 2 Current probe     designed for permanent or semi permanent in circuit  installation  1 2 kHz to 200 MHz  50 Q input    CT 4 Current Transformer     for use with the AM 503S  A6302  and P6021   Peak pulse 1 kA  0 5 Hz to 20 MHz with AM 503S  A6302     P6701A Opto Electronic Converter  500 to 950 nm  DC to 850 MHz   1 V mW    P6703A Opto Electronic Converter  1100 to 1700 nm  DC to 1 GHz   1 V mW    P6711 Opto Electronic Converter  500 to 950 nm  DC to 250 MHz  5 V mW    P6713 Opto Electronic Converter  1100 to 1700 nm  DC to 300 MHz   5 V mW    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix A
35.  Nd   Number of points per decimation interval   4   100    Resolution Enhancement  bits    0 5 X LOG  Nd      3 extra bits       Bits of Resolution   Resolution Enhancement  3 bits    8 bits   11 bits    Hi Res mode only works with real time  noninterpolated sampling  If you set the  time base so fast that it requires real time interpolation or equivalent time  sampling  the mode automatically becomes Sample even though the menu  selection will not change     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 23    Acquiring and Displaying Waveforms       Table 3 3  Additional Resolution Bits                Time Base Speed Bits of Resolution  100 ns and faster 8 bits   200 ns to 500 ns 9 bits   1 us to 2 us 10 bits   5 us to 10 us 11 bits   20 us to 50 us 12 bits   100 us to 200 us 13 bits   500 us 14 bits   1 ms and slower 15 bits   Checking the To determine the acquisition sampling rate  the acquisition state  running or    Acquisition Readout stopped   and the acquisition mode  check the Acquisition readout at the top of  the display   See Figure 3 14   The state    Run     shows the sample rate and  acquisition mode  The state    Stop     shows the number of acquisitions acquired  since the last stop or major change        Acquisition Readout   un  100kS s    I Sample             2oomv       tet Stop After Limit Test   Limit Test       fete    Limit Test  EGUI aS button Setup Sources    plate       Acquisition  Mode             Sample       Peak Detect    an    Laal  Envel
36.  Oscilloscope  pack it in the original shipping carton and  packing material  If the original packing material is not available  package the  instrument as follows     1     Obtain a corrugated cardboard shipping carton with inside dimensions at  least 15 cm  6 in  taller  wider  and deeper than the oscilloscope  The  shipping carton must be constructed of cardboard with 170 kg  375 pound   test strength     If you are shipping the oscilloscope to a Tektronix field office for repair   attach a tag to the oscilloscope showing the instrument owner and address   the name of the person to contact about the instrument  the instrument type   and the serial number     Wrap the oscilloscope with polyethylene sheeting or equivalent material to  protect the finish     Cushion the oscilloscope in the shipping carton by tightly packing dunnage  or urethane foam on all sides between the carton and the oscilloscope  Allow  7 5 cm  3 in  on all sides  top  and bottom     Seal the shipping carton with shipping tape or an industrial stapler        NOTE  Do not ship the oscilloscope with a disk inside the disk drive  When the  disk is inside the drive  the disk release button sticks out  This makes the button  more prone to damage than otherwise        TDS 500B  TDS 600B   amp  TDS 700A User Manual C 1    Appendix C  Packaging for Shipment       C 2 TDS 500B  TDS 600B   amp  TDS 700A User Manual    a  Se a ee ee    Appendix D  Factory Initialization Settings    Recalling the factory setup esta
37.  P6139A  TDS 724A  Two P6139A 10X  500 MHz Passive probes P6139A  TDS 6448  Four P6243 probes P6243  TDS 620B  Two P6243 probes P6243  TDS 540B  Four P6139A 10X  500 MHz Passive probes P6139A  TDS 520B  Two P6139A 10X  500 MHz Passive probes P6139A  TDS 680B  TDS 684B  TDS 784A  No probes standard  Front Cover 200 3696 01  Accessory Pouch  TDS 644B  TDS 684B  TDS 700A  016 1268 00  U S  Power Cord 161 0230 01          You can also order the optional accessories listed in Table A 3     Table A 3  Optional Accessories             Accessory Part Number  Service Manual 070 9721 XX  Plotter  GPIB and Centronics Standard  HC100  Oscilloscope Cart K420   Rack Mount Kit  for field conversion  016 1236 00  Accessory Pouch  TDS 500B  TDS 620B  TDS 680B  016 1268 00  Soft Sided Carrying Case 016 0909 01  Transit Case 016 1135 00  GPIB Cable  1 meter  012 0991 01  GPIB Cable  2 meter  012 0991 00  Centronics Cable 012 1214 00  RS 232 Cable 012 1298 00    The following optional accessory probes are recommended for use with your    oscilloscope        m P6245 Active  high speed digital voltage probe  FET  DC to 1 5 GHz  50 Q    input    TDS 500B  TDS 600B   amp  TDS 700A User Manual    A 5    Appendix A  Options and Accessories       A 6    P6243 Active  high speed digital voltage probe  FET  DC to 1 0 GHz  50 Q  input    P6101B 1X  15 MHz  Passive probe    P6156 10X  3 5 GHz  Passive  low capacitance   low impedance Zo  probe   Provides 100X  when ordered with Option 25    P6139A 10X  500
38.  Preparation    To ensure maximum accuracy for your most critical measurements  you should  know about signal path compensation and the proper use of the probe you choose  to use with your oscilloscope     Signal Path Compensation Be sure you compensate your oscilloscope for the surrounding temperature  This  action  called Signal Path Compensation  SPC   ensures maximum possible  accuracy for your most critical measurements  See Signal Path Compensation on  page 3   106 for a description of and operating information on this feature     Recommended Probes The TDS 680B  TDS 684B  TDS 782A  and TDS 784A oscilloscopes ship  without probes  To take advantage of the higher bandwidth of the TDS 680B   TDS 684B  TDS 782A  and TDS 784A Oscilloscopes  order the P6245 Active  Probe     The remaining TDS 500B  TDS 600B  and TDS 700A oscilloscopes ship with  general purpose probes     either the P6139A or the P6243  depending on the  oscilloscope model  The standard accessory probes and quantities shipped for  these oscilloscopes are listed on in Standard Accessories on page A 4     For a list of optional accessory probes for all TDS 500B  TDS 600B  and  TDS 700A oscilloscope models  see Accessory Probes on page A S     Probe Usage Be sure you use the appropriate probe for the measurement  For instance  do not  use the P6245 Active Probe to measure signals greater than  8 volts or with  more than  10 volts of offset  since errors in signal measurement will result   See  the User manual f
39.  Selecting this file in step 3 stores a setup  in a uniquely named  sequentially numbered file  For instance  the oscilloscope  saves the first setup you save in the file TEKOOO01 SET  the second   in TEK00002 SET  and so on           Tek Run  100kKS 5 Sample       A  Save Current  Setup             uJ  To File          To Setup1    factory               To Setup 2    factory          To Setup3    factory          20 0m    gt  M 500yus Chis 27 2mvV moren          3   3     o    n E H HH EIE PEE  DIAT a  saved Current Factory utilities  Setup Setu Setup    Loloa             ololoolo                                                    Figure 3 65  Save Recall Setup Menu  To Recall a Setup To recall a setup  do the following steps     1  To recall a setup stored internally  press SAVE RECALL SETUP  gt  Recall  Saved Setup  main   gt  Recall Setup 1  Recall Setup 2      side      3 116 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       2  To recall a setup stored on disk  press From File  side   Then use the general  purpose knob to select the exact file from the resulting scrollbar list  Only  files with  set extensions will be displayed  Finally  press Recall From  Selected File  side  to complete the operation     Recalling a setup will not change the menu that is currently displayed  If you recall  a setup that is labeled factory in the side menu  you will recall the factory setup    The conventional method for recalling the factory setup is de
40.  Set to TTL or Set to ECL trigger levels below standard TTL and  ECL levels  That happens because the trigger level range is fixed at  12  divisions from the center  At 100 mV  the next smaller setting after 200 mV  the  trigger range is  1 2 V which is smaller than the typical TTL   1 4 V  or ECL    1 3 V  level        3 90 TDS 500B  TDS 600B   amp  TDS 700A User Manual    E  Measuring Waveforms    To make the best use of the TDS Oscilloscope when taking measurements  you  need to know how to use the three types  or classes  of measurements it can take   This section describes how to take the following classes of measurements  Fig   ure 3   51 shows all three measurement classes      m Automated for automatically taking and displaying waveform measurements    m Cursor for measuring the difference  either in time or voltage  between two  locations in a waveform record    m Graticule for making quick estimates by counting graticule divisions on  screens    This section also tells you how to use Probe Cal  Channel Probe Deskew  and  Signal Path Compensation to optimize the accuracy of your measurements     Cursor Automated  Graticule Readouts Measurements       Chl  Frequency  100 MHz             Ch 1 Period  10 ns       Cursors                                           Figure 3 51  Graticule  Cursor and Automated Measurements    Taking Automated Measurements    The TDS Oscilloscope provides the feature Measure for automatically taking and  displaying waveform measurements  Thi
41.  TRIGGER LEVEL knob  the general purpose knob  or the  keypad to set each threshold     Define Logic  To choose the type of logic function you want applied to the input  channels     Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt  Class  main   gt   State  pop up   gt  Define Logic  main   gt  AND  OR  NAND  or NOR  side      Set Trigger When  To choose to trigger when the logic condition is met  Goes  TRUE  or when the logic condition is not met  Goes FALSE      Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt  Class  main   gt   State  pop up   gt  Trigger When  main   gt  Goes TRUE or Goes FALSE  side      To Set Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer to To Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see the descriptions  Trigger Modes and Trigger Holdoff on page 3 53     When you select the logic class Setup Hold  the oscilloscope uses one channel as  a data channel  the factory default setting is Chl   another channel as a clock  channel  default is Ch2   and triggers if the data transitions within the setup or  hold time of the clock   Pages 3   63 and 3   64 describe how setup hold triggers  work   To use setup and hold triggering  do the following procedures     Select Setup Hold Triggering  Press TRIGGER MENU  gt  Type  main   gt  Logic   pop up   gt  Class  main   gt  Setup Hold  pop up      Define the Data Source  To sel
42.  Voltage over time measurement  The area over the entire waveform or gated region in  volt seconds  Area measured above ground is positive  area below ground is negative     Voltage over time measurement  The area over the first cycle in the waveform  or the first cycle  in the gated region  in volt seconds  Area measured above ground is positive  area below  ground is negative     Timing measurement  The duration of a burst  Measured over the entire waveform or gated  region    Voltage measurement  The arithmetic mean over the first cycle in the waveform or the first cycle  in the gated region     Voltage measurement  The true Root Mean Square voltage over the first cycle in the waveform  or the first cycle in the gated region     Timing measurement  The time between the MidRef crossings of two different traces or the  gated region of the traces     Timing measurement  Time taken for the falling edge of the first pulse in the waveform or gated  region to fall from a High Ref value  default   90   to a Low Ref value  default  10   of its  final value              Frequency             frf High    Timing measurement for the first cycle in the waveform or gated region  The reciprocal of the  period  Measured in Hertz  Hz  where 1 Hz  1 cycle per second     The value used as 100  whenever High Ref  Mid Ref  and Low Ref values are needed  as in fall  time and rise time measurements   Calculated using either the min max or the histogram method   The min max method uses the maximum
43.  XY Format Pairs       XY Pair X Axis Source Y Axis Source  ChlandCh2 Chl Ch2   Ch3 and Ch 4  Axl and Ax2    Ch 3  Ax1  Ch 4  Ax2   Ref 1 and Ref 2 Refl Ref 2  Ref 3 and Ref 4 Ref3 Ref 4          Since selecting YT or XY affects only the display  the horizontal and vertical  scale and position knobs and menus control the same parameters regardless of  the mode selected  Specifically  in XY mode  the horizontal scale will continue  to control the time base and the horizontal position will continue to control  which portion of the waveforms are displayed     XY format is a dot only display  although it can have persistence  The Vector  style selection has no effect when you select XY format     You cannot display Math waveforms in XY format  They will disappear from the  display when you select XY        NOTE  Use of XY at higher room temperatures or with higher intensity display  formats  such as the white fields in the Hardcopy palette  can temporarily  degrade display quality        Customizing the Display Color  The TDS Oscilloscope can display information in different colors  This section    describes how to use the Color menu to choose the colors in which the various  display objects appear     3 34 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Change the Display Color To bring up the Color menu   1  Press DISPLAY to show the Display menu     2  Press Settings in the main menu until you select Color from the pop up  menu   See 
44.  a histogram of the record with one bin for each digitizing level  256  total      2  It splits the histogram into two sections at the halfway point between Min  and Max  also called Mid      3  The level with the most points in the upper histogram is the High value  and  the level with the most points in the lower histogram is the Low value    Choose the levels where the histograms peak for High and Low      If Mid gives the largest peak value within the upper or lower histogram  then  return the Mid value for both High and Low  this is probably a very low  amplitude waveform      If more than one histogram level  bin  has the maximum value  choose the  bin farthest from Mid     This algorithm does not work well for two level waveforms with greater than  about 100  overshoot     The user sets the various reference levels  through the Reference Level selection  of the Measure menu  They include     HighRef     the waveform high reference level  Used in fall time and rise time  calculations  Typically set to 90   You can set it from 0  to 100  or to a  voltage level     MidRef     the waveform middle reference level  Typically set to 50   You can  set it from 0  to 100  or to a voltage level     LowRef     the waveform low reference level  Used in fall and rise time  calculations  Typically set to 10   You can set it from 0  to 100  or to a  voltage level     Mid2Ref     the middle reference level for a second waveform  or the second  middle reference of the same waveform   Us
45.  a pulse edge  traverses  slews  between an upper and lower threshold  The edge of the  pulse may be positive  negative  or either  The oscilloscope can trigger on  slew rates faster or slower than a user specified rate     Slope  The direction at a point on a waveform  You can calculate the direction by  computing the sign of the ratio of change in the vertical quantity  Y  to the  change in the horizontal quantity  The two values are rising and falling     Tek Secure  This feature erases all waveform and setup memory locations  setup memories  are replaced with the factory setup   Then it checks each location to verify  erasure  This feature finds use where this oscilloscope is used to gather  security sensitive data  such as is done for research or development projects     Time base  The set of parameters that let you define the time and horizontal axis  attributes of a waveform record  The time base determines when and how  long to acquire record points     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Glossary          Sie Timeout trigger  A trigger mode in which triggering occurs if the oscilloscope does NOT find  a pulse  of the specified polarity and level  within the specified time period     Trigger  An event that marks time zero in the waveform record  It results in acquisi   tion and display of the waveform     yo Trigger level  The vertical level the trigger signal must cross to generate a trigger  on edge  trigger mode        i Vertical bar cursors  The two 
46.  a unique probe identification  number  and so on  Some optical probes and most active probes  such as the P6205  have TEKPROBE interfaces     3 114    TDS 500B  TDS 600B   amp  TDS 700A User Manual    ae a a  Saving Waveforms and Setups    The TDS Oscilloscope can save and recall the waveforms you measure and the  setups you use to measure them  It can also output or save a copy of its display  screen  This section describes how to use the following features to save  recall   and document your measurements     m Save Recall Setups  for saving the setups you create to internal memory or to  a floppy disk  so you can recall and reuse those setups    m   Save Recall Waveform for saving waveforms to internal memory or to a  floppy disk and for recalling those waveforms to the display    m Hardcopy for printing a copy of the oscilloscope display screen or for saving  it to floppy disk  hardcopies can be incorporated into documents using desk  top publishing software     m File Utilities for managing  copying  organizing into directories  and so on   the setups  waveforms  and display screens that you save to floppy disk    This section ends with details on how to connect your oscilloscope into a system  environment  so that it can communicate with remote instruments        NOTE  TDS 500B  TDS 620B  and TDS 680B oscilloscopes do not come  equipped with a disk drive unless you order Option 1F  See Options on  page A 1        Saving and Recalling Setups    The TDS Oscilloscope can s
47.  amp  TDS 700A User Manual    3 111    Measuring Waveforms       by service personnel  You can also check the error log by doing steps 13  through 14     18  After the clock icon is removed  verify the word Initialized changed to Pass  under Cal Probe in the main menu   See Figure 3   63      19  If desired  repeat this procedure beginning at step   to compensate for other  probe channel combinations  But before you do so  be sure you take note of  the following requirements     m Remember to first low frequency compensate any passive probe you  connect  see the prerequisites listed on page 3   109 at the beginning of  this procedure      m Remember to connect all but simple passive probes to the oscilloscope  for a twenty minute warm up before running Probe Cal     Changing Probes After a Probe Cal  If a Probe Cal has never been performed on  an input channel or if its stored Probe Cal data is erased using the Re use Probe  Calibration Data menu  discussed later   the oscilloscope displays Initialized  status in its vertical menu  It also displays initialized whenever you remove a  probe from an input     If you execute a successful Probe Cal on an input channel  the oscilloscope  stores the compensation data it derived in nonvolatile memory  Therefore  this  data is available when you turn the oscilloscope off and back on and when you  change probes     When you install a probe or power on the oscilloscope with probes installed  the  oscilloscope tests the probe at each inp
48.  care           Logic trigger logic   pattern and state     TDS 500B  TDS 600B   amp  TDS 700A User Manual       AND    D 3    Appendix D  Factory Initialization Settings       D 4    Table D 1  Factory Initialization Defaults  Cont      Control    Logic trigger pattern time qualification  Lower limit  Upper limit    Logic trigger S etup Hold times  Setup  Hold    Logic trigger sources and levels   Setup Hold     Logic trigger threshold  all channels    pattern  state  and S etup Hold     Changed by Factory Init to    5 ns  5 ns    3 ns  2 ns    Data Source  Channel 1  1 4 V  Clock Source   Channel 2  1 4 V     Source levels are clipped to 1 2 V atthe  default volts division setting established  by Factory Init     Clock Edge  Rising    1 4 V   clipped to 1 2 V at the default volts divi   sion setting when no 10X probe attached        Logic trigger triggers when       pattern and state     Goes TRUE       Math2 extended processing  Math3 definition    Main trigger mode Auto   Main trigger type Edge   Math1 definition Ch1 Ch2   Math1 extended processing No extended processing   Math2 definition Ch1  Ch2  FFT of Ch 1 for instruments    with Option 2F Advanced DSP Math   No extended processing  Inv of Ch 1       Math3 extended processing  Measure delay edges  Measure delay to    No extended processing  Both rising and forward searching  Channel 1  Ch1                       Measure gating Off   Measure high ref 90  and 0 V  units   Measure high low setup Histogram  Measure low re
49.  ciO    lt cpie gt  gece    Baao                                                    Figure 3 78  Utility Menu    To Find More Information See Printing a Hardcopy  on page 3 125     See the TDS Programmer Manual  Tektronix part number 070 9385 XX     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 137    Saving Waveforms and Setups       3 138 TDS 500B  TDS 600B   amp  TDS 700A User Manual    a ame iS Sn a  Determining Status and Accessing Help    Displaying Status    Displaying the Banner    The TDS Oscilloscope can display the status of its internal systems  It also  provides an on line help system  This section describes how to use the following  two features     m Status which displays a snapshot of system  display  trigger  waveform  and  I O settings      Help which displays a screen of brief information about each oscilloscope  control when that control is operated    Press SHIFT STATUS  gt  Status  main   gt  System  Display  Trigger   Waveforms  or I O  side      System displays information about the Horizontal  Zoom  Acquisition  Measure   and Hardcopy systems   See Figure 3   79   This display also tells you the  firmware version     Display provides parameter information about the display and color systems   Trigger displays parameter information about the triggers     Waveforms displays information about waveforms  including live  math  and  reference waveforms     T O displays information about the I O port s      To display the banner  lists firmware version  o
50.  feature you desire  Just be  sure the integrated waveform is the selected waveform   Press MORE  then  select the integrated waveform in the More main menu  Then use the Vertical and  Horizontal SCALE knobs to adjust the math waveform size      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 175    Using Features for Advanced Applications       If you want to see the zoom factor  2X  5X  etc    you need to turn Zoom on   press ZOOM  gt  On  side   The vertical and horizontal zoom factors appear on  screen     Whether Zoom is on or off  you can press Reset Zoom Factors  side  to return  the zoomed integral waveform to no magnification     3 176 TDS 500B  TDS 600B   amp  TDS 700A User Manual    a a    Appendix A  Options and Accessories    Options    Table A 1  Options    This appendix describes the various options as well as the standard and optional  accessories that are available for the TDS Oscilloscope     Tektronix will ship the options shown in Table A 1                                                                                      Option   Label Description  Al Universal European 220 V  50 Hz power cord  power cord  A2 UK power cord 240 V  50 Hz power cord  A3 Australian power cord 240 V  50 Hz power cord  A4 North American power 240 V  60 Hz power cord  cord  A5 Switzerland power cord 220 V  50 Hz power cord  05 Video trigger Oscilloscope comes with tools for investigating events that    occur when a video signal generates a horizontal or vertical  B b sync pul
51.  following notation to represent the sequence of  selections you made in steps 1  2 and 3  Press save recall SETUP  gt  Recall  Factory Setup  main   gt  OK Confirm Factory Init  side         D Note that a clock icon appears on screen  The oscilloscope displays this icon  when performing operations that take longer than several seconds     4  Press SET LEVEL TO 50   see Figure 2   4  to be sure the oscilloscope  triggers on the input signal     A TRIGGER    MAIN LEVEL    h    TRIG   D    READY  DELAYED TRIG    arm TRIGGER  MENU    SET LEVEL  TO 50     SET LEVEL TO 50  Button              gt  __     Figure 2 4  Trigger Controls    Examine the Display Read the following information to become familiar with the oscilloscope display  Elements before doing the examples     Figure 2   5 shows the display that results from the oscilloscope reset  There are  several important points to observe     m The trigger level bar shows that the waveform is triggered at a level near  50  of its amplitude  from step 4      m The trigger position indicator shows that the trigger position of the  waveform is located at the horizontal center of the graticule     TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 11    Tutorial       Trigger Level Bar    Trigger P osition Indicator  Channel Reference Indicator  Trigger Readout    Time Base Readout  Channel Readout    The channel reference indicator shows the vertical position of channel 1  with no input signal  This indicator points to the ground leve
52.  illustrating purposes  but you can only display one style at a time      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 31    Acquiring and Displaying Waveforms       The trigger bar is only displayed if the trigger source is an active  displayed  waveform  Also  two trigger bars are displayed when delay triggerable  acquisitions are displayed     one for the main and one for the delayed time  base  The trigger bar is a visual indicator of the trigger level     Sometimes  especially when using the hardcopy feature  you may want to  display the current date and time on screen   To find more information  displaying and setting date and time  see Date Time Stamp the Hardcopy on  page 3 128      4  Press Display Date Time  side  to turn it on or off  Push Clear Menu to see  the current date and time        Tek Run  50 0kS 5 Sample       Readout  Options                  Display    T       Trigger Point       Trigger P oint Indicator             Trigger Bar  Style             tong  Trigger Bar    Long 5 tyle tN tetantnfn been EEE oan bedded AAA EEEE AA can thn ea ede F D isp l ay     0r  g 4 3 i k ji i g     ON          Trigger Bar   Short Style             50 0mv      M1 ooms Chit   6amv             LOC C O    Style i Readout Filter Graticule   Format  Settings   Intensity 7 i     lt Display gt  Vectors leraticm  Sin x  x Full VF    IPDIC OC  G                                                       Figure 3 18  Trigger Point and Level Indicators    Select Interpolation Filter T
53.  in the  milliohm to low Q range  they are especially useful where low loading of the  circuit is important  Current probes can also make differential measurements by  measuring the results of two opposing currents in two conductors in the jaws of  the probe           Figure E 2  A6303 Current Probe Used in the AM 503S Opt  03       NOTE  Attempting to measure more than 40 amperes of total  in phase current   DC   peak AC  using three or more current probes installed on the input  channels can result in measurement or display errors        TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix E  Probe Selection       Optical Probes    Optical probes let you blend the functions of an optical power meter with the  high speed analog waveform analysis capability of an oscilloscope  You have the  capability of acquiring  displaying  and analyzing optical and electrical signals  simultaneously     Applications include measuring the transient optical properties of lasers  LEDs   electro optic modulators  and flashlamps  You can also use these probes in the  development  manufacturing  and maintenance of fiber optic control networks   local area networks  LANs   fiber based systems based on the FDDI and SONET  standard  optical disk devices  and high speed fiber optic communications  systems        NOTE  When you connect any probe with a TEKPROBE  level 2  interface to the  oscilloscope  the input impedance of the oscilloscope automatically becomes   50 Q  If you then connect a 
54.  items Ref2  Ref3  and Ref4 appear  shaded while Ref1 does not  References that are empty appear shaded in the  More main menu     To Recall a Waveform To recall a waveform from disk to an internal reference memory  press SAVE   From Disk RECALL WAVEFORM  gt  Recall Wfm To Ref  main   gt  Recall From File   side  Then use the general purpose knob to select the exact file from the  resulting scrollbar list  Only files with   WFM extensions are displayed  Finally   press To Ref1  To Ref2  To Ref3  or To Ref4  side  to complete the operation        We       Tek Run  100kKS 5 Sample       Math              Change Math  waveform  definition    No Extended  mee  Processsing                            Average  2                   300m             LOIDID OI oO                                                   DIG IP IOIC IC         Figure 3 67  More Menu    3 120 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       To Enable Autosave To use autosave  press SAVE RECALL WAVEFORM  gt  Autosave  main   gt   Autosave Single Seq ON  side      Also turn on Single Acquisition Sequence in the Acquire menu   See Stop After  on page 3 26      To disable this feature  simply press Autosave  main      Autosave Single Seq  OFF  side      If you enable both autosave and single sequence  the oscilloscope will save all  live channels to reference waveforms at the completion of each single sequence  event  All previous reference waveform data will be erased     To rearm 
55.  menu  You adjust those parameters to adjust slew rate  you can t  adjust slew rate directly        Set to Trig When  The oscilloscope compares the pulse edge of the trigger source  against the slew rate setting read out in the Trigger When menu  To select  whether to trigger on edges with slew rates faster than or slower than that  indicated in readout  do the following step     Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt  Class  main   gt   Slew Rate  pop up   gt  Trigger When  main   gt  Trigger if Faster Than or  Trigger if Slower Than  side    See Figure 3 46      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 81    Triggering on Waveforms       Components of Pulse Edge   dv and dt    Cursors Measure Slew Rate    Trigger Bar at Upper Threshold    Trigger Point at Second Crossing    3 82    Trigger Bar at Lower Threshold       NOTE  If you select Trigger if Faster Than and the oscilloscope does not trigger   it may be because the pulse edge is too fast rather than too slow  To check the  edge speed  switch to edge triggering  Then trigger on the pulse edge and  determine the time the edge takes to travel between the levels set in the slew rate  Thresholds menu  The oscilloscope cannot slew rate trigger on pulse edges that  traverse between threshold levels in 600 ps or less     Also  to reliably slew rate trigger  a pulse must have a width of 7 5 ns or more  A  pulse of less width may trigger on the wrong slope or not trigger at all  Switch to  edge trig
56.  must be first crossed going negative   then recrossed going positive without the lower threshold being crossed at all     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       Either requires only that either one of the thresholds must be first crossed going  in either direction  then recrossed going in the opposite direction without the  other threshold being crossed at all     For all three polarity settings  triggering occurs at the point the runt pulse  recrosses its first threshold     Set the Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer to Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see Trigger Modes  and Trigger Holdoff on page 3 53     Trigger Based on When you select the pulse class Width  the oscilloscope will trigger on a pulse  Pulse Width narrower  or wider  than some specified range of time  defined by the upper  limit and lower limit   To set up for width triggering  do the following proce   dures     Select Width Triggering  Press TRIGGER MENU     gt  Type  main   gt  Pulse   pop up   gt  Class  main   gt  Width  pop up      Select the Source  Press TRIGGER MENU  gt  Type  main   gt  Pulse   pop up   gt  Source  main   gt  Ch1  Ch2  Ch3  or Ch4  side   The source  selected becomes the trigger source for all four trigger classes     Select the Polarity  Press TRIGGER MENU  gt  Type  main   gt  Pulse   pop up   gt  Class
57.  of a Derivative Waveform    The settings you make for offset  scale  and position affect the math waveform  you obtain  Note the following tips for obtaining a good display     m You should scale and position the source waveform so it is contained on  screen   Off screen waveforms may be clipped  resulting in errors in the  derivative waveform      m You can use vertical position and vertical offset to position your source  waveform  The vertical position and vertical offset will not affect your  derivative waveform unless you position the source waveform off screen so  it is clipped     m When using the vertical scale knob to scale the source waveform  note that it  also scales your derivative waveform     Because of the method the oscilloscope uses to scale the source waveform before  differentiating that waveform  the derivative math waveform may be too large  vertically to fit on screen     even if the source waveform is only a few divisions  on screen  You can use Zoom to reduce the size of the waveform on screen  see  Using Zoom that follows   but if your waveform is clipped before zooming  it  will still be clipped after it is zoomed     If your math waveform is a narrow differentiated pulse  it may not appear to be  clipped when viewed on screen  You can detect if your derivative math wave   form is clipped by expanding it horizontally using Zoom so you can see the    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       Usi
58.  oscilloscope can acquire with no more than  3 dB  x  707  attenuation of the original  reference  signal     Burst width  A timing measurement of the duration of a burst     Channel  One type of input used for signal acquisition  The oscilloscope has four  channels     Channel probe deskew  A relative time delay for each channel  This lets you align signals to  compensate for the fact that signals may come in from cables of differing  length     Channel Reference Indicator  The indicator on the left side of the display that points to the position around  which the waveform contracts or expands when vertical scale is changed  This  position is ground when offset is set to 0 V  otherwise  it is ground plus offset     Coupling  The association of two or more circuits or systems in such a way that power  or information can be transferred from one to the other  You can couple the  input signal to the trigger and vertical systems several different ways     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Glossary                Cursors  Paired markers that you can use to make measurements between two waveform  locations  The oscilloscope displays the values  expressed in volts or time  of  the position of the active cursor and the distance between the two cursors     Cycle area  A measurement of waveform area taken over one cycle  Expressed in  volt seconds  Area above ground is positive  area below ground is negative     Cycle mean  An amplitude  voltage  measurement of the arit
59.  pop up   gt  Trigger When  main   gt  Goes TRUE or Goes FALSE   side      The side menu items TRUE for less than and TRUE for greater than are used to  time qualify a pattern trigger  See the procedure Define a Time Qualified Pattern  Trigger that follows for instructions     To Set Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer to To Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see the descriptions  Trigger Modes and Trigger Holdoff on page 3 53     You can also time qualify a pattern logic trigger  That is  you specify a time that  the boolean logic function  AND  NAND  OR  or NOR  must be TRUE  logic   high   To specify the time limit as well as the type of time qualification  greater  or less than the time limit specified  for a pattern trigger  do the following steps     1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  Pattern  pop up   gt  Trigger When  main   gt  TRUE for  less than or TRUE for more than  side      2  Use the knob and keypad to set the time in the side menu     When you select TRUE for less than and specify a time  the input conditions you  specify must drive the logic function high  TRUE  for less than the time you  specify  Conversely  the TRUE for more than menu item requires the boolean  function to be TRUE for longer than the time you specify     Note the position of the trigger indicator in Figure 3 
60.  prerequisites below and then  do the steps that follow     m If you are installing an active probe  such as the P6243 or P6245  there are  no prerequisites to performing this procedure  Start at step 1     m If you are compensating for a passive probe with this procedure you must  first compensate the low frequency response of the probe  First  do steps 1  and 2 below  and then compensate the probe by following the instructions  that came with your probe   Or see To Compensate Passive Probes on  page 3   4   Then continue with step 3 of this procedure     1  Install the probe on the input channel on which it is to be used     2  Power on the digitizing oscilloscope and allow a 20 minute warm up before  doing this procedure     3  Press SHIFT UTILITY  gt  System  main   gt  Cal  pop up      4  Look at the status label under Signal Path in the main menu  If the status  does not read Pass  perform a signal path compensation  Signal Path  Compensation  page 3 106   and then continue with this procedure     5  Press the front panel button corresponding to the input channel on which you  installed the probe     6  Press VERTICAL MENU  gt  Cal Probe  main      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 109    Measuring Waveforms          STOP  Your oscilloscope will detect the type of probe you have installed and  display screen messages and menu choices for compensation of probe gain   offset  or both   See Figure 3 63   The following steps will have you run probe  gain  off
61.  probes use active circuit elements in the probe design  to process signals from the circuit under test  All active probes require a source  of power for their operation  Power is obtained either from an external power  supply or from the oscilloscope itself        NOTE  When you connect an active probe to the oscilloscope  such as the  P6245   the input impedance of the oscilloscope automatically becomes 50 Q  If  you then connect a passive probe  you need to set the input impedance back to   1 MQ  The procedure To Change Vertical Scale and Position on page 3 11  explains how to change the input impedance  Also  please read Input Impedance  Considerations on page 3 5 for more information        High Speed Active Probes Active probes offer low input capacitance  1 to 2 pF typical  while maintaining  the higher input resistance of passive probes  10 KQ to 10 MQ   Like Zo probes   active probes are useful for making accurate timing and phase measurements   However  they do not degrade the amplitude accuracy  Active probes typically  have a dynamic range of  8 to  15 V     Differential Probes Differential probes determine the voltage drop between two points in a circuit  under test  Differential probes let you simultaneously measure two points and to  display the difference between the two voltages     Active differential probes are stand alone products designed to be used with  50 Q inputs  The same characteristics that apply to active probes apply to active  differential probe
62.  so use maxima              MidR ef    Figure B 4  Choosing Minima or Maxima to Use for Envelope Measurements    TDS 500B  TDS 600B   amp  TDS 700A User Manual B 13    Appendix B  Algorithms       Missing or Out of Range Samples    If some samples in the waveform are missing or off scale  the measurements will  linearly interpolate between known samples to make an    appropriate    guess as to  the sample value  Missing samples at the ends of the measurement record will be  assumed to have the value of the nearest known sample     When samples are out of range  the measurement will give a warning to that  effect  for example     CLIPPING        if the measurement could change by  extending the measurement range slightly  The algorithms assume the samples  recover from an overdrive condition instantaneously     For example  if MidRef is set directly  then MidRef would not change even if  samples were out of range  However  if MidRef was chosen using the   choice  from the Set Levels in   Units selection of the Measure menu  then MidRef  could give a    CLIPPING    warning        NOTE  When measurements are displayed using Snapshot  out of range warnings  are NOT available  However  if you question the validity of any measurement in  the snapshot display  you can select and display the measurement individually  and then check for a warning message        TDS 500B  TDS 600B   amp  TDS 700A User Manual    _             a   Appendix C  Packaging for Shipment    If you ship the TDS
63.  the delta time setting determine the  slew rate setting  To set these parameters     1  Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Slew Rate  pop up   gt  Thresholds  main      2  Press Set to TTL or Set to ECL  side  to set the upper and lower thresholds  to levels appropriate for those to logic families     OR       TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       3  Press the upper threshold button and  in turn  lower threshold button  side     Use the general purpose knob or keypad to set the values for the high and  low thresholds     Hint  To use the Trigger Bar feature to set the threshold levels on the pulse  edge  press DISPLAY     Readout Options  main   gt  Trigger Bar Style   side  until Long appears in that menu item     The threshold settings determine the voltage component of slew rate  Volts Se   cond   To finish specifying the slew rate  set the time component by doing the  following steps     4  Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Slew Rate  pop up   gt  Trigger When  main   gt  Delta  Time  side      5  Use the general purpose knob or keypad to set the delta time value for slew  rate        NOTE  The menu item Slew Rate in the side menu is not a button label  rather it  is a readout of the slew rate setting  This readout varies as you vary the Delta  Time setting this side menu and as you vary either of the threshold settings from  the Thresholds
64.  the main menu Restore Colors item   See Figure 3 22      2  Select the object s  you want to restore by pressing Reset Current Palette  To Factory  Reset All Palettes To Factory or Reset All Mappings To  Factory in the side menu     3 38 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms          Tek Run  100kS 5 Sample       Restore  Colors                Reset Current  Palette to  Factory          Reset All  Palettes to  Factory             Reset All  Mappings to  Factory                      s00mv C M 500s Chii  270mv    pete   Palette   Change ap ap   p Restore  Settings Math Reference   Options H   lt  Color  Normal Colors Colors Colors Colors          LOIDID ICO Ie                                              MIMIC  ClO  O       Figure 3 22  Display Menu     Restore Colors    Zooming on Waveforms    Using with Waveforms    The TDS Oscilloscope can expand or compress  zoom in or out  on a waveform  without changing the acquisition parameters  sample rate  record length  and so  on   This subsection describes how to use Zoom and how it interacts with the  selected waveform  It also describes how interpolation can affect Zoom     Use Zoom  press the ZOOM button  when you want to temporarily expand a  waveform to inspect small feature s  on that waveform  For example  to  temporarily expand the front corner of a pulse to inspect its aberrations  use  Zoom to expand it horizontally and vertically  After you are finished  you can  return to 
65.  to do the following tasks     m How to couple waveforms to the oscilloscope channels  m How to select channels to turn on and off their display  m How to size and position the selected channel on screen    m How to use the menus to set vertical  coupling  offset  and bandwidth  and  horizontal  time base  record length  and so on  parameters    This section also describes how to choose the appropriate acquisition mode for  acquiring your waveform  how to customize the display  including selecting the  color of the display elements   and how to use the Zoom  FastFrame  and InstaVu  features     Coupling Waveforms to the Oscilloscope    Tektronix produces a variety of probes and cables suitable for connecting various  types of signals to the input channels of this product  This subsection covers two  topics important to coupling  Probe Compensation and Input Impedance  Considerations     If your model oscilloscope ships with a probe  use it for general purpose  coupling of waveforms to the oscilloscope  For a list of other probes available  for use  see Accessory Probes on page A 5     The TDS 680B  TDS 684B  TDS 782A  and TDS 784A oscilloscopes ship  without probes  Tektronix recommends you order and use the P6245 Active  Probes to take advantage of the higher bandwidth of these models     The remaining TDS 500B  TDS 600B  and TDS 700A oscilloscopes ship with  general purpose probes     either the P6139A or the P6243  depending on the   oscilloscope model  The standard access
66.  trace     Delay measurements are actually a group of measurements  To get a specific  delay measurement  you must specify the target and reference crossing polarities  and the reference search direction     Delay   the time from one MidRef crossing on the source waveform to the  Mid2Ref crossing on the second waveform     Delay is not available in the Snapshot display     Timing measurement  The time taken for the falling edge of a pulse to drop from  a HighRef value  default   90   to a LowRef value  default   10       Figure B   2 shows a falling edge with the two crossings necessary to calculate a  Fall measurement     1  Searching from Start to End  find the first sample in the measurement zone  greater than HighRef     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix B  Algorithms       2  From this sample  continue the search to find the first  negative  crossing of  HighRef  The time of this crossing is THF   Use linear interpolation if  necessary      3  From THF  continue the search  looking for a crossing of LowRef  Update  THF if subsequent HighRef crossings are found  When a LowRef crossing is  found  it becomes TLF   Use linear interpolation if necessary      4  FallTime   TLF   THF                   Fall Time  as  THF TLF  High  HighR ef  LowRef  Low    Figure B 2  Fall Time    Frequency Timing measurement  The reciprocal of the period  Measured in Hertz  Hz   where 1 Hz   1 cycle per second        If Period   0 or is otherwise bad  return an error   
67.  uae  ype Class Inputs when  amp     Logic   lt State gt  Goes TRUE Holdoff                                              DIGIC OIC  G          Figure 3 39  Logic Trigger Readouts     State Class Selected       NOTE  When the trigger type Logic is selected  the trigger levels must be set for  each channel individually in the Set Thresholds menu  pattern and state classes   or the Levels  setup hold class  menu  Therefore  the Trigger Level readout will  disappear on the display and the Trigger Level knob can be used to set the  selected level while the Main Trigger menu is set to Logic        When you select the logic class Pattern  the oscilloscope will trigger on a  specified logic combination of the four input channels   Pages 3   62 through  3   64 describe how pattern triggers work   To use pattern triggering  do the  procedures that follow     Select Pattern Triggering  Press TRIGGER MENU     gt  Type  main   gt  Logic   pop up   gt  Class  main   gt  Pattern  pop up      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms          Tek Run  100kS 5 Sample chi  GL X       4 Logic Pattern  Inputs                                           Edge E ane ee a ene See T  Logic a O M S00ms Pattern Hxxx O       Pulse       Ooo JO           video        ot   H R ara E i   Ear    z gar PE   ode  Type Class tho ute Logic when hice held  amp   stogic gt    lt Pattern gt  AND Goes TRUE Holdoff                                                                Figure 3 
68.  use the Area measurement     If StartCycle   EndCycle then return the  interpolated  value at StartCycle     EndCycle     een   Waveform t dt    StartCycle    For details of the integration algorithm  see page B 12     Timing measurement  The duration of a burst   1  Find MCross1 on the waveform  This is MCrossStart     2  Find the last MCross  begin the search at EndCycle and search toward  StartCycle   This is MCrossStop  This could be a different value from  MCross1     3  Compute BurstWidth   MCrossStop     MCrossStart    TDS 500B  TDS 600B   amp  TDS 700A User Manual B 5    Appendix B  Algorithms             Cycle Mean   Cycle RMS   Delay   Fall Time  c     Amplitude  voltage  measurement  The mean over one waveform cycle  For  non cyclical data  you might prefer to use the Mean measurement     If StartCycle   EndCycle then return the  interpolated  value at StartCycle     EndCycle    Waveform t dt  Si    tartCycle    CycleMean         EndCycle     StartCycle  x Sampleinterval    For details of the integration algorithm  see page B 12     The true Root Mean Square voltage over one cycle   If StartCycle   EndCycle then CycleRMS   Waveform Start      Otherwise        EndCycle     Waveform t   dt  Si    tartCycle     EndCycle     StartCycle  x SampleInterval    CycleRMS      For details of the integration algorithm  see page B 12     Timing measurement  The amount of time between the MidRef and Mid2Ref  crossings of two different traces  or two different places on the same
69.  use the color  of the first constituent waveform     To return to the factory defaults  select Reset to Factory Color     Set Reference Waveform To define reference waveform colors     Color    Press Map Reference in the main menu   See Figure 3   21    Select one of the four reference waveforms by pressing Ref in the side menu     To assign the selected reference waveform to a specific color  press   repeatedly  Color and choose the value     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 37    Acquiring and Displaying Waveforms       4  To make the selected reference waveform the same color as the waveform it  is based on  select Color Matches Contents     To return to the factory defaults  select Reset to Factory Color        ee       Tek Run  250kS 5 Sample       Ref  Mappings             Ref  Refi                 Color Matches     ontents    chi Ch2 Ch3 Ch4 Math Color   Ref     Reset to  Factory  Color                               Palette   Change Math ap Restore    GAGGI Options    Settings  Colors Colors     lt  Color  Colors       Normal Colors                                           DIC ICIC lO  O  O       Figure 3 21  Display Menu     Map Reference Colors    Select Options To define what color to show where a waveform crosses another waveform   1  Press the Options main menu item     2  Toggle Collision Contrast to ON in the side menu to mark collision zones  with a special color     Restore Colors To restore colors to their factory default settings   1  Press
70.  value found  The histogram method uses the most  common value found above the mid point  Measured over the entire waveform or gated region        3 92    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       Table 3 8  Measurement Definitions  Cont                                                                                             Name Definition   ne Low The value used as 0  whenever High Ref  Mid Ref  and Low Ref values are needed  as in fall   eS time and rise time measurements   May be calculated using either the min max or the histogram  method  With the min max method itis the minimum value found  With the histogram method  it  refers to the most common value found below the midpoint  Measured over the entire waveform  or gated region    T Maximum Voltage measurement  The maximum amplitude  Typically the most positive peak voltage   Measured over the entire waveform or gated region     A Mean Voltage measurement  The arithmetic mean over the entire waveform or gated region    rf Minimum Voltage measurement  The minimum amplitude  Typically the most negative peak voltage    u Measured over the entire waveform or gated region    Negative Duty Cycle Timing measurement of the first cycle in the waveform or gated region  The ratio of the negative  pulse width to the signal period expressed as a percentage     Negative Width  S AAE    NegativeDutyCycle Period 100   Negative Overshoot Voltage measurement  Measured over the entire waveform or gate
71.  want  to delay an acquisition so it captures and displays events that follow other  events  See Jo Find More Information below     To Find More Information To perform tutorials that teach selecting  scaling  and positioning of waveforms   see Example 1  Displaying a Waveform on page 2 13 and Example 2  Displaying  Multiple Waveforms on page 2 15     To learn how to use delay with waveforms  see Delayed Triggering on  page 3 84  To learn how to magnify waveforms  see Zooming on Waveforms  on  page 3 39     Choosing an Acquisition Mode    The TDS Oscilloscopes are digital products that can acquire and process your  input signal in a variety of modes  To help you choose the best mode to use for  your signal measurement task  this section first describes     m How the oscilloscope samples and digitizes an input signal  m How the different acquisition modes  such as interpolation  affect this process  m How to select among these modes    Following these descriptions are procedures for selecting the sampling and  acquisition modes  beginning with Checking the Acquisition Readout on page 3   24     Sampling and Digitizing Acquisition is the process of sampling the analog input signal  digitizing it to  convert it into digital data  and assembling it into a waveform record   See  Figure 3 9   The oscilloscope creates a digital representation of the input signal  by sampling the voltage level of the signal at regular time intervals  The sampled  and digitized points are stored in mem
72.  waveform parameters automatically  and read out the results on screen  Do the following tasks to discover how to set  up the oscilloscope to measure waveforms automatically     Display Measurements To use the automated measurement system  you must have a stable display of  Automatically your signal  Also  the waveform must have all the segments necessary for the  measurement you want  For example  a rise time measurement requires at least  one rising edge  and a frequency measurement needs at least one complete cycle   To take automated measurements  do the following steps     1                    If you are not continuing from the previous example  follow the instructions  on page 2 9 under the heading Setting Up for the Examples     Press SETUP     Recall Factory Setup  main   gt  OK Confirm Factory  Init  side      Press AUTOSET   Press MEASURE to display the Measure main menu   See Figure 2 14      If it is not already selected  press Select Measrmnt  main   The readout for  that menu item indicates which channel the measurement will be taken from   All automated measurements are made on the selected channel     The Select Measurement side menu lists some of the measurements that can  be taken on waveforms  There are many different measurements available   up to four can be taken and displayed at any one time  Pressing the button  next to the  more    menu item brings up the other measurement selections     Press Frequency  side   If the Frequency menu item is not visibl
73.  with  Option 5  Video Line Number  side      Select Amplitude Units If you are measuring NTSC signals  you can choose to display vertical readings  in IRE units  If you are trying to measure such a signal  you should have  Option 05 Video Trigger installed as it would be difficult to trigger on composite  video waveforms without Option 05  To switch between IRE and base cursor  units  do the following steps     1  Press CURSOR     Amplitude Units  main   gt  IRE  NTSC      2  To return to normal  press CURSOR     Amplitude Units  main   gt  Base     To Find More Information To find instructions for using cursors with math waveforms  see Waveform Math  on page 3 148     To find instructions on using cursor with FFT waveforms  differentiated  waveforms  and integrated waveforms  see Fast Fourier Transforms on   page 3 150  Waveform Differentiation on page 3 167  and Waveform Integration  on page 3 171     To find information on cursor units with video waveforms  see the TDS Family  Option 05 Video Trigger Interface  if your oscilloscope is equipped with the  video trigger option     Taking Graticule Measurements    The TDS Oscilloscope provides a graticule for measuring the difference  either  in time or amplitude  between two points on a waveform record  Graticule  measurements provide you with quick  visual estimates  For example  you might  look at a waveform amplitude and say    it is a little more than 100 mV     This  section briefly describes how to take graticule measu
74.  would let you  store 10 records of 500 samples each into one record with a 5000 sample length     3 48 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       eas    ae eas   i aa                 l              Real Time                cep eee ee    eevee eases   L es    FS SS eS Sh q                    FastFrame i ot EE a ssns           be  Bye hes ee Ee      Figure 3 28  Fast Frame    If you are using the FastFrame mode  you can jump to the desired frame  To use    FastFrame  do the following steps     1  Press HORIZONTAL MENU  gt  FastFrame Setup  mai    n   gt  FastFrame     side  to toggle on or off the use of FastFrame  see Figure 3   29      S        FastFrame  Setup    z        Tek Run  50 0MS s Sampie  REI Frame Length  100                FastFrame  a ON          Frame            erin h       Frame  count  5                                  Fam        Position  Frame  3    FastFrame  Setup    C    oomvay M1 00qMs CHIA 404mv          Time Base    Trigger  positon  50      div                 CC OO  O                                         lt        Figure 3 29  Horizontal Menu     FastFrame Setup    TDS 500B  TDS 600B   amp  TDS 700A User Manual        Oooo o     3   49    Acquiring and Displaying Waveforms       3 50    2     Press Frame Length or Frame Count  side  and use the general purpose  knob to enter FastFrame parameters     Frame Length refers to the number of samples in each acquisition     Frame count refers to the numb
75. 00B or TDS 700A Oscilloscope can interleave its channels to attain  higher digitizing rates without equivalent time sampling or interpolating  The  oscilloscope applies the digitizing resources of unused channels  that is  channels  that are turned off  to sample those that are in use  turned on   Table 3 2 lists  how interleaving more than one digitizer to sample a channel extends the  maximum digitizing rate     Once you set horizontal scale to exceed the maximum digitizing rate for the  number of channels in use  see Table 3   2   the oscilloscope will not be able to  get enough samples to create a waveform record  At that point  the oscilloscope  will either interpolate to calculate additional samples or it will switch from real  to equivalent time sampling to obtain additional samples   See Interpolation on  page 3 19 and Equivalent Time Sampling on page 3 19      3 20 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       The Acquisition Modes    Table 3 2  How Interleaving Affects Sample Rate       No  of Maximum Digitizing Rate   Channels On TDS 520B  amp  P   TDS 724A TDS 540B TDS 754A TDS 782A TDS 784A  One 1GS sec 2 GS sec 2 GS sec 4 GS sec  Two 500 MS sec 2 GS sec 1 GS sec 2 GS sec  Three or Four   Not Available 1 GS sec NotAvailable   1GS sec             1 When real time sampling   GS   Gigasamples  MS   Megasamples      2 This TDS model  when purchased with Option 1G  is limited to a maximum sample  rate of 1GS second  See Opt
76. 0A User Manual 2 5    Operating Interface Maps       Display Map    The Indicates position of When present  the general   acquisition vertical bar cursors in the Trigger The value entered with purpose knob makes coarse   status  waveform record  position  T   the general purpose adjustments  when absent   page 3 24 page 3 104 page 3 58 knob or keypad  fine adjustments                                la    Tek Run  5 00GS 5 Samply   Trigger Level  252mV             X Shows what part of the waveform The waveform      record is displayed  page 3 14 record icon a    j                              Tek Run  5 00GS 5 Sample _Trigger Level  252mV  amp   Trig g er level on E __ oa APS ok A se ee ea E a       Coarse Kne  waveform  may be an pM ee aa  es O l  A  188s Tevet Cursor measurements   arrow at right side of Lag E I steep Biv eae page 3 101    screen instead of a bar         The side menu  Channel level and with choices of    waveform source  ee ea ihe Si ne set to ECL specific actions        Vertical scale   page 3 11    Trigger parameters   page 3 57                EE Ee i    Type Say  amp   eEige gt  ch DC F EM idoft       The main menu with Horizontal scale and time  choices of major actions base type  page 3 14    2 6 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Operating Interface Maps       To Operate a Menu    1 Press front panel menu button   Press SHIFT first if button label is blue                            Tektronix TDS 684B 22 2h32 8uoscox sa                  
77. 0B  TDS 600B   amp  TDS 700A User Manual 3 153    Using Features for Advanced Applications       10     11     12        Hamming     Very good window for resolving frequencies that are very  close to the same value with somewhat improved amplitude accuracy over  the rectangular window     Hanning     Very good window for measuring amplitude accuracy but  degraded for resolving frequencies     Blackman Harris     Best window for measuring the amplitude of  frequencies but worst at resolving frequencies     The topic Selecting a Window  on page 3   164  provides in depth information  on choosing the right window for your application     If you did not select Phase  deg  or Phase  rad  in step 7  skip to step 12   Phase suppression is only used to reduce noise in phase FFTs     If you need to reduce the effect of noise in your phase FFT  press Suppress  phase at amplitudes  lt   side      Use the general purpose knob to adjust the phase suppression level  FFT  magnitudes below this level will have their phase set to zero     The topic Adjust Phase Suppression  on page 3 163  provides additional  information on phase suppression     Press OK Create Math Wfm  side  to display the FFT of the waveform you  input in step 1   See Figure 3 88            Tek Run  50 0MS s Sample                                   Oloo        thi oova S ops    cht      s00my 29 Mar 1994  CME    20 0dE 2 50MHz 12 08 57           119  1  1  1 1c                                            Figure 3 88  FFT Ma
78. 1 MQ for any input signal not from a 50 Q system        TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 5    Acquiring and Displaying Waveforms       To Find More Information    To find a procedure for changing the coupling and input impedance settings  see  To Change Vertical Parameters on page 3 13     To find a list of available probes  see Accessory Probes on page A 5     To find a guide for selecting probes for a variety of applications  see Appen   dix E  Probe Selection on page E 1     Setting up Automatically  Autoset and Reset    To Autoset the  Oscilloscope    3 6    The TDS Oscilloscope can automatically obtain and display a stable waveform  of usable size  It can also be reset to its factory default settings  This subsection  describes how to execute Autoset and reset  and lists the default settings in effect  after an Autoset     Autoset automatically sets up the front panel controls based on the characteristics  of the input signal  It is much faster and easier than a manual control by control  setup  Autoset adjusts controls in these categories  Acquisition  Display   Horizontal  Trigger  and Vertical     Do the following steps to automatically set up the oscilloscope     1  Press the channel selection button  such as CH 1  corresponding to your  input channel to make it active     2  Press AUTOSET     If you use Autoset when one or more channels are displayed  the oscilloscope  selects the lowest numbered channel for horizontal scaling and triggering   Vertic
79. 159   for DC correction of FFTs  3 159   with math waveforms  3   159  3 175  DC  Main Trigger menu  3   60  Default Model s   xii  Define Inputs  Main Trigger menu  3 67  3 69  3 71  Define Logic  Main Trigger menu  3   68  3   70  Delay by Events  Delayed Trigger menu  3   88  Delay by Time  Delayed Trigger menu  3   88  Delay by  Delayed Trigger menu  3   88  Delay measurement  3   98  Glossary   3  Delay time  Glossary   3  Delay To  Measure Delay menu  3   98  Delayed Only  Horizontal menu  3 86  Delayed Runs After Main  3 55  Delayed Runs After Main  Horizontal menu  3 17    3 86   Delayed Scale  Horizontal menu  3   16  Delayed time base  Incompatible with InstaVu  3 48  DELAYED TRIG button  3 55  3 86  Delayed trigger  3 55  3 85   3 90   How to set up  3 86  Delayed Trigger menu  3   86   3   90   Chi  Ch2     3 89   Coupling  3 89   Delay by  3 88   Delay by Events  3   88   Delay by Time  3 88   Falling edge  3 89   Level  3   89   Rising edge  3 89   Set to 50   3 89   Set to ECL  3 89   Set to TTL  3 89   Slope  3 89   Source  3   89  DELAYED TRIGGER OUTPUT  BNC  2 5  Delayed Triggerable  3 55  Delayed Triggerable  Horizontal menu  3   17  3 86  Delete Refs  Save Recall Waveform menu  3   120  Delete  File Utilities menu  3 122  Delta Time  Main Trigger menu  3 81  Derivative math waveform  3 168   applications  3 168   derivation of  3 168   procedure for displaying  3 168   procedure for measuring  3   169   record length of  3 168    Index   4    Desk Top Publ
80. 2 10  3 8  3 115  3 122  Save Recall WAVEFORM  3 118  3 122  SELECT  3 104  Glossary   10  SET LEVEL TO 50   3 56  SINGLE TRIG  3 28  3 57  STATUS  3 139  TRIGGER MENU  3 59  3   60  3   66  3 69  3   70   3 75  3 77  3 80  UTILITY  3 107  3 127  3 136  VERTICAL MENU  2 17  WAVEFORM OFF  2 18  3 9  3 34          Index   2    ZOOM  3 39  3 40  Buttons  CH1  CH2     3 9  Channel selection  2   16  3   9  Main menu  2 3  Side menu  2 3  BW symbol  3 13    C    Cables  3 135  Cal Probe  Vertical menu  3   108  Cart  Oscilloscope  A 2  Centronics  2   5  Port  3 128  3 134  CH1  CH2     buttons  3 9  Ch1  Ch2      Delayed Trigger menu  3 89  Chl  Ch2      Main Trigger menu  3   60  3   67  3 70   3 71  3 75  3 77  3 79  3   80  3   83  Change Colors  Color menu  3 36  Channel  Glossary   2  Readout  2   6  3   8  3   40  Reference Indicator  2   6  3   8  Selection buttons  2   16  3   9  Trigger input  3   52  Channel readout  2   6  Channel reference indicator  Glossary   2  Channel    probe deskew  3 108  Glossary   2  Channels  Selecting  3 8  Circuit loading  Glossary   2  Class Glitch  Main Trigger menu  3 75  Class  Main Trigger menu  3 79  3 83  Pattern  3   66  Runt  3 77  Setup Hold  3 70  Slew Rate  3   80  State  3 69  Classes  Pulse triggers  3 74  CLEAR MENU button  2 3  2 8  2 12  2 20  2 21   3 99  Clear Spool  Hardcopy menu  3 131  Clipping  derivative math waveforms  3   170  FFT math waveforms  3 159  how to avoid  3 159  3 170  3 175  integral math waveforms  3
81. 3 156  DC correction  3 159  derivation of  3 151  displaying phase  3 153  frequency range  3 158  frequency resolution  3 158  interpolation mode  3 160  magnifying  3 160  phase display  setup considerations  3 162  phase suppression  3 154  3 163  procedure for displaying  3 152  procedure for measuring  3 155  record length  3 158  reducing noise  3 160  undersampling  3 161  zero phase reference  3 162  FFT time domain record  defined  3   157  File System  3 121  File Utilities menu  3 122  Confirm Delete  3   125  Copy  3 124  Create Directory  3 124  Delete  3 122  File Utilities  3 122  Format  3   125  Icons  3 116  3 118  3 122  3 125  Overwrite Lock  3 125  Print  3 124  Rename  3 123  File Utilities  File Utilities menu  3 122  File Utilities  Save Recall Setup menu  3 117  File Utilities  Save Recall Waveform menu  3 121  Filter  Display menu  3   32  Fine Scale  Vertical menu  3   13  Firmware version  3   139  Fit to screen  Horizontal menu  3   16  Fixtured active probes  E 3  Floppy disk drive  Option 1F  A 1  FORCE TRIG button  3 56  Format  File Utilities menu  3 125  Format  Hardcopy menu  3 127  Frame Count  Horizontal menu  3   50  Frame Length  Horizontal menu  3 50  Frame  Display menu  3 33  Frame  Horizontal menu  3   50  Frequency  2 19  3   92  Glossary   4  Front Cover removal  1   5  Front panel  2   4  Full  Display menu  3 33       Index 5    Index       Full  Vertical menu  3   13  Function  Cursor menu  3   104  Fuse  14  2 5    G    Gated 
82. 40  Logic Trigger Menu    To Define Pattern Inputs  To set the logic state for each of the input channels   Ch1  Ch2           1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  Pattern  pop up   gt  Define Inputs  main   gt  Ch1  Ch2   Ch3  or Ch    side      2  Repeatedly press each input selected in step   to choose either High  H    Low  L   or Don   t Care  X  for each channel     To Set Thresholds  To set the logic threshold for each channel     1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  Pattern  pop up   gt  Set Thresholds  main   gt  Ch1  Ch2   Ch3  or Ch    side      2  Use the MAIN TRIGGER LEVEL knob  the general purpose knob  or the  keypad to set each threshold     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 67    Triggering on Waveforms       To Define a Time Qualified    3 68    Pattern Trigger    To Define the Logic  To choose the logic function you want applied to the input  channels  see page 3   63 for definitions of the logic functions for both pattern  and state triggers      Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt  Class  main   gt   Pattern  pop up   gt  Define Logic  main   gt  AND  OR  NAND  or NOR  side      Set Trigger When  To choose to trigger when the logic condition is met  Goes  TRUE  or when the logic condition is not met  Goes FALSE   do the following  step     Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt  Class  main   gt   Pattern 
83. 41  Triggering occurs at the  point that the oscilloscope determines that the logic function you specify is  TRUE within the time you specify  The oscilloscope determines the trigger point  in the following manner     m It waits for the logic condition to become TRUE     m It starts timing and waits for the logic function to become FALSE     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       m It compares the times and  if the time TRUE is longer  for TRUE for more  than  or shorter  for TRUE for less than   then it triggers a waveform display  at the point the logic condition became FALSE  This time can be  and usually  is  different from the time set for TRUE for more than or TRUE for less  than     In Figure 3 41  the delay between the vertical bar cursors is the time the logic  function is TRUE  Since this time is more  216 us  than that set in the TRUE for  more than menu item  150 us   the oscilloscope issues the trigger at that point   not at the point at which it has been TRUE for 150 us        ei  S    TEk Stopped  4671 Acquisitions  True  gt   150 0000us      A  216us Logic Pattern  Sts Trig When    Goes  TRUE                   Time Logic Function is TRUE                            Logic Function  AND  Becomes TRUE ea       hi Goes  Dor FALSE          TRUE for  less than    3 7 i r r ens  Logic Function Becomes FALSE and ce  em E l   EA  Triggers Acquisition a ETO     chi    s00mvVa che    Ss0omva  M 100ps eae    PA nae ORENA a oak an A 
84. 43  Duty cycle  2 20  Glossary   7  Glossary   8    E    Edge trigger  3 52  3 59  Glossary   3   How to set up  3 60   Readout  3 59  Edge  Main Trigger menu  3 59  3   60  Edges  Measure Delay menu  3   99  Either  Main Trigger menu  3   76  3 77  3 80  empty  Saved waveform status  3 118  Encapsulated Postscript  3 126  Enter Char  Labelling menu  3 123  3 124  Envelope  Incompatible with InstaVu  3   48  Envelope acquisition mode  3   21  3 50  Glossary   3  Envelope  Acquire menu  3 25  EPS Color Img  Hardcopy menu  3 127  EPS Color Plt  Hardcopy menu  3 127  EPS Mono Img  Hardcopy menu  3 127  EPS Mono Plt  Hardcopy menu  3 127  Epson  3 126  Epson  Hardcopy menu  3 127  Equivalent time sampling  3 19  3 50  Equivalent time sampling  random  Glossary   4    F    Factory initialization settings  D 1 D   6  Factory Setup  How to execute  3 8  factory  Saved setup status  3 116  Fall time  3   92  Glossary   4  Falling edge  Delayed Trigger menu  3   89  Falling edge  Main Trigger menu  3   62  3 70  3 71  Fast Fourier Transforms  description  3 150  Fast Fourier Transforms  FFTs   applications  3 150  FastFrame  Incompatible with InstaVu  3 47  FastFrame interactions  3   50  3 103  FastFrame Setup  Horizontal menu  3   49  FastFrame  Horizontal menu  3 49  FFT frequency domain record  3 156   defined  3 157    TDS 500B  TDS 600B   amp  TDS 700A User Manual    length of  3 158  FFT math waveform  3 151  acquisition mode  3 160  aliasing  3 161  automated measurements of  
85. 60   comprise  one complete cycle of one of the waveforms  Waveforms measured should be  of the same frequency or one waveform should be a harmonic of the other     Pixel  A visible point on the display  The oscilloscope display is 640 pixels wide  by 480 pixels high     Pop up Menu  A sub menu of a main menu  Pop up menus temporarily occupy part of the  waveform display area and are used to present additional choices associated  with the main menu selection  You can cycle through the options in a pop up  menu by repeatedly pressing the main menu button underneath the pop up     Positive duty cycle  A timing measurement of the ratio of the positive pulse width to the signal  period  expressed as a percentage     Positive overshoot  Amplitude  voltage  measurement     Max     High    Amplitude TOO    PositiveOvershoot      Positive width  A timing measurement of the distance  time  between two amplitude  points     rising edge MidRef  default 50   and falling edge MidRef  default  50       on a positive pulse     Posttrigger  The specified portion of the waveform record that contains data acquired  after the trigger event     Pretrigger  The specified portion of the waveform record that contains data acquired  before the trigger event     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Glossary       Tae    Probe  An oscilloscope input device     Quantizing  The process of converting an analog input that has been sampled  such as a  voltage  to a digital value     Probe comp
86. 600B   amp  TDS 700A User Manual 2 1    Overview       2 2 TDS 500B  TDS 600B   amp  TDS 700A User Manual    E en  Operating Interface Maps    This section contains illustrations  or maps  of the display  the front and rear  panels  and the menu system of the TDS Oscilloscope  These maps will help you  understand and operate the oscilloscope  This section also contains a visual guide  to using the menu system     Front Panel Map     Left Side                 Tektronix TDS 684B Samire sie tsiuoscore bese                                                                                     File System   page 3 121 Side Menu Buttons   page 2 7  H       CLEAR MENU  o  es  es  fe   es   C      Removes Menus      CLEAR from the Display             ON STBY Switch  Main Menu Buttons   page 1 5 page 2 7    TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 3    Operating Interface Maps       Front Panel Map     Right Side    Measurement System  page 3 91    Cursor Measurements  page 3 101            Saving and Recalling  Waveforms  page 3 118  File System  page 3 121    Hardcopy  page 3 125    File System  page 3 121 Acquisition Modes  page 3 21               LO    SELECT       _ L    InstaVu    C lt             APPLICATION  MEASURE    ACQUIRE MENU    MENU      CRUN STOP    UTILITY  SHIFT  DISPLAY      SAVE RECALL   WAVEFORM          Autoset  page 3 6 TE  STATUS    HELP    SETUP    Help  page 3 141       Status  page 3 139    Saving and Recalling  Setups  page 3 115        a    Selecting 
87. 73  Glossary   10  600 ps limitation  3 82   3 90  7 5 ns limitation  3   82   3   90  How to set up  3   80    3 90  Slope  Glossary   10  Slope  Delayed Trigger menu  3 89  Slope  Main Trigger menu  3   62  Slope  Trigger  3 55  Snapshot  Readout  3 99  Snapshot of Measurements  2   23  3   99  Snapshot  Measure menu  3   99  Software version  3 139  Source  Delayed Trigger menu  3 89  Source  Main Trigger menu  3   60  3 75  3 77  3 79   3 80  3 83  Spectral  Color menu  3 35  Spooler  Hardcopy  3 131  Start up  1 3  State trigger  3 63  3 69 3 90  How to set up  3 69    3 90  State  Main Trigger menu  3   69  3   70  Status  Determining setup  3 139       TDS 500B  TDS 600B   amp  TDS 700A User Manual    STATUS button  3 139  Status menu  3    139   3 142  Display  3 139  Firmware version  3 139  T O  3 139  System  3 139  Trigger  3 139  Waveforms  3 139  Stop After Limit Test Condition Met  Acquire menu   3 146  Stop After  Acquire menu  3   26  3 147  Style  Display menu  3   29  Switch  principal power  1 5  2 5  System  Status menu  3 139  System  Utility menu  3 127    T    Talk Listen Address  Utility menu  3 136  Tek Secure  3 117  Glossary   10  Tek Secure Erase Memory  Utility menu  3 117  Temperature compensation  3   106  Temperature  Color menu  3   35  Template Source  Acquire menu  3 144  Text Grat  Display menu  3 31  Thinkjet  3 126  Thinkjet  Hardcopy menu  3 127  Thresholds  Main Trigger menu  3 78  3   80  TIFF  3 126  TIFF  Hardcopy menu  3 127  Time 
88. 8  3 70  3 81  Trigger  delayed  How to set up  3 86  Trigger  edge  How to set up  3   60  Trigger  glitch  How to set up  3 75  Trigger  runt  How to set up  3   77   3 90  Trigger  slew rate  How to set up  3   80   3   90  Trigger  Status menu  3 139  Trigger  timeout  How to set up  3   83 3 90  Trigger  width  How to set up  3   79 3   90  Triggering on Waveforms  3 51  Triggering  Auto mode  Incompatible with InstaVu   3 47  True for less than  Main Trigger menu  3   68  True for more than  Main Trigger menu  3   68  Type Logic  Main Trigger menu  Logic  3 66  3 69  3 70  Pulse  3   80  Type Pulse  Main Trigger menu  3 75  Type  Main Trigger menu  3 59  3   60  3 79  3   83  Pulse  3   77    U    Undershoot  Glossary   7  user  Saved setup status  3 116                Index 14    UTILITY button  3 107  3 127  3 136  Utility Menu   OK Erase Ref  amp  Panel Memory  3 117   Tek Secure Erase Memory  3 117  Utility menu  3 127   Configure  3 127  3 136   GPIB  3 136   Hardcopy  3 136   Hardcopy  Talk Only   3 127   VO  3 127   Off Bus  3 136   Port  3 136   System  3 127   Talk Listen Address  3 136    V    V Limit  Acquire menu  3 145  Variable Persistence  Display menu  3   30  Vectors  3   30  Vectors display  Incompatible with InstaVu  3   48  Vectors  Display menu  3   30  Vertical   Bar cursors  3 101  Glossary   11   Offset  3 14   Position  3 11  3 12   Readout  3 11   Scale  3 11   SCALE knob  2 13  3 11   System  2 13  Vertical deskew  3 108  Glossary   2  Vertica
89. APPLICATION MENU   ACQUIRE MEY    MEASURE HARDCOPY  RUN STOP     CO CO CGC     CURSOR                      Tek Run  5 00GS s_ Sample Trigger Level   4mv  E F               am          Set to TTL          Set to ECL              Set to 50   SINGLE TRIG  FORCE TRIG    200mv M10 0ns Chi       4mv  Co  oa eo  n ae rn  a   Source  coupling   siop Ere Far Co Ten od   lt Euge gt   gt    T       JN s CUR  7 7       See C                                                                                                                                                                                             2 Press one of these buttons to select from main menu        3 Press one of these buttons to select from side menu  if displayed      4 if side menu item has an adjustable value  shown in reverse video   adjust it with the general purpose knob or keypad     TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 7       Operating Interface Maps    To Operate a Pop Up Menu          COLOR FOUR CHANNEL  DIGITAL REAL TIME OSCILLOSCOPE      Tektronix TDS 684          Tek Run  100kKS s Sample  E F  Edge Source    Lanre    280m      more     1of 2                                     M 500ums Chi         Logic  Pulse  Level oge    video  Source Coupling Slope  Type DE T 2gemv   Hofioti                                                                               Press here to  remove menus from    Press to display pop up menus   screen     Press it again  to make selection        Alternatively  pr
90. Auxiliary Trigger input connector on the  oscilloscope rear panel     The digitizing oscilloscope provides three standard triggers for the main trigger  system  edge  pulse  and logic  Option 05 provides a video trigger  The standard  triggers are described individually starting on page 3 59  A brief definition of  each type follows     Edge is the    basic    trigger  You can use it with both analog and digital test  circuits  An edge trigger event occurs when the trigger source  the signal the  trigger circuit is monitoring  passes through a specified voltage level in the  specified direction  the trigger slope      Pulse is a special purpose trigger primarily used on digital circuits  The five  classes of pulse triggers are glitch  runt  width  slew rate and timeout  Pulse  triggering is available on the main trigger only     Logic is a special purpose trigger primarily used on digital logic circuits  Two of  the classes  pattern and state  trigger based on the Boolean operator you select  for the trigger sources  Triggering occurs when the Boolean conditions are  satisfied  A third class  setup hold  triggers when data in one trigger source  changes state within the setup and hold times that you specify relative to a clock    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       in another trigger source  Logic triggers are available on the main trigger system  only     Video  available as option 05  is a special trigger used on video circuits  I
91. CH 1     Instruction steps are numbered  The number is omitted if there is only one step     When steps require that you make a sequence of selections using front panel  controls and menu buttons  an arrow         marks each transition between a  front panel button and a menu  or between menus  Also  whether a name is a  main menu or side menu item is clearly indicated  Press VERTICAL    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Preface       MENU  gt  Coupling  main   gt  DC  side   gt  Bandwidth  main   gt   250 MHz  side      Using the convention just described results in instructions that are graphically  intuitive and simplifies procedures  For example  the instruction just given  replaces these five steps     1   2  3   4    5     Press the front panel button VERTICAL MENU   Press the main menu button Coupling    Press the side menu button DC    Press the main menu button Bandwidth     Press the side menu button 250 MHz     Sometimes you may have to make a selection from a pop up menu  Press  TRIGGER MENU  gt  Type  main   gt  Edge  pop up   In this example  you  repeatedly press the main menu button Type until Edge is highlighted in the  pop up menu     TDS 500B  TDS 600B   amp  TDS 700A User Manual xiii    Preface       xiv TDS 500B  TDS 600B   amp  TDS 700A User Manual    E  Product Description                                              lololololo                                                                                      9 alol                    
92. Ch2  Math1  Math2  Math3  Ref1  Ref2  Ref3  or Ref4  side    See  Figure 3   83         NOTE  The template will be smoother if you acquire the template waveform using  Average acquisition mode  If you are unsure how to select Average  see Selecting  an Acquisition Mode on page 3 25        3  Once you have selected a source  select a destination for the template  press  Template Destination  side   gt  Ref1  Ref2  Ref3  or Ref4     3 144 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       Tek Run  100kKS 5 Sample       Limit Test  Template             Template  Source  chi          Template   4 Destination  Refi           V Limit  4omaiv           H Limit  4omdiv          200mv M 500s Chis 252mv ye  i         i ore             Ololooo       Template    Mode  Stop After Ea Limit Test PRES  Sample  R S button Setup Sources Limi  Test    PIP IC ICO  S  O   O                                                       Figure 3 83  Acquire Menu     Create Limit Test Template    4  Press  V Limit  side   Enter the vertical  voltage  tolerance value using the  general purpose knob or keypad     5  Press  H Limit  side   Enter the horizontal  time  tolerance value using the  general purpose knob or keypad     Tolerance values are expressed in fractions of a major division  They  represent the amount by which incoming waveform data can deviate without  having exceeded the limits set in the limit test  The range is from 0  the  incoming waveform mu
93. Channels   page 3 8    SINGLE TRIG  FORCE TRIG    Waveform Math    page 3 148        PROBE COMPENSATION            SIGNAL GND    PROBE  AT iy                War A 88 CH3                              Vertical Controls   page 3 11                            Ground    Zoom  page 3 39 Triggering  page 3 51   Delay Triggering  page 3 84  Edge Triggering  page 3 59  Logic Triggering  page 3 62  Pulse Triggering  page 3 72       Horizontal Controls   page 3 14            Cursor  Measurements   page 3 101    InstaVu   page 3 45   TDS 500B   amp  TDS 700A  models only     Color  page 3 34  Display Settings   page 3 29  Remote  Communication   page 3 134    Probe Calibration   page 3 108    2 4 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Operating Interface Maps       Rear Panel Map    Principal Power  Switch  GPIB Connector   page 1 5 Centronics Connector RS 232 Connector page 3 134 VGA Output                                                                               POWER N    a                                                                                               Fuse  Serial Number Power Connector  Rear Panel Connectors Security Bracket    page 1 4 page 1 4 SIGNAL OUTPUT     Provides Analog Signal Output  from CH3   or AX1    10 mV div     AUX TRIGGER INPUT     Provides Auxiliary Trigger Signal Input     MAIN TRIGGER OUTPUT     Provides Main Trigger  TTL  Output     DELAYED TRIGGER OUTPUT     Provides Delayed Trigger  TTL  Output     TDS 500B  TDS 600B   amp  TDS 70
94. DISCLAIM ANY  IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE   TEKTRONIX    RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND  EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY  TEKTRONIX  AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT  SPECIAL  INCIDENTAL  OR  CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS  ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES     EC Declaration of Conformity    Tektronix Holland N V   Marktweg 73A   8444 AB Heerenveen  The Netherlands    declare under sole responsibility that the  TDS 520B  TDS 540B  TDS 724A  TDS 754A  TDS 782A and  TDS 784A Digitizing Oscilloscopes  meet the intent of Directive 89 336 EEC for Electromagnetic Compatibility and Low  Voltage Directive 73 23 ECC for Product Safety  Compliance was demonstrated to the  following specifications as listed in the Official Journal of the European Communities   EMC Directive 89 336 EEC   EN 55011 Class A Radiated and Conducted Emissions  EN 50081 1 Emissions   EN 60555 2 AC Power Line Harmonic Emissions    EN 50082 1 Immunity     IEC 801 2 Electrostatic Discharge Immunity   IEC 801 3 RF Electromagnetic Field Immunity  IEC 801 4 Electrical Fast Transient Burst Immunity  IEC 801 5 Power Line Surge Immunity    Low Voltage Directive 73 23 EEC        EN 61010 1 Safety requirements for electrical equipment for  measurement  control  and laboratory use       EC Declaration of Conformity    Tektr
95. DS 620B  and TDS 724A  models         TDS 500B  TDS 600B   amp  TDS 700A User Manual    A 3    Appendix A  Options and Accessories       Table A 1  Options  Cont      Option   Label    2F    FFT  INTG  DIFF       Advanced DSP math    31 and 3P Color printer    Description    Add advanced DSP math features such as FFT  integration  and  differentiation     Option applies only to TDS 520B  TDS 540B  TDS 620B and  TDS 680B models      Tektronix Phaser 140  360 dpi  inkjet  color printer   Order option 31 for 220V use or option 3P for 110 V use        4D    L1  L3  L5  L9       Standard Accessories    A 4    Four probes    Manuals in French  Manuals in German  Manuals in J apanese    Manuals in Korean    Delete the four standard probes shipped with the model    Option applies only to TDS 540B  TDS 644B  and TDS 754A  models      Provides Language versions of User and Reference Manuals   according to option number chosen     The oscilloscope comes standard with the accessories listed in Table A 2     Table A 2  Standard Accessories          Accessory Part Number  User Manual 070 9719 XX  Programmer Manual 070 9556 XX  Reference 070 9382 XX  Technical Reference  Performance Verification and Specifications 070 9720 XX       TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix A  Options and Accessories       Table A 2  Standard Accessories  Cont         Optional Accessories    Accessory Probes    Accessory Part Number  Probes   TDS 754A  Four P6139A 10X  500 MHz Passive probes
96. Existing Cal Data     To use the probe uncalibrated and erase the  existing cal data  press Erase Probe Cal Data     To use the probe uncalibrated while retaining OK  the existing cal data for later use  press Erase Probe     Clear Menu  sake Cal Data  chit T 00V T 00V   M 500us Chi                           LOO IDID ID                                                              1119  1  1 1       Figure 3 64  Re use Probe Calibration Data Menu    If the Re use Probe Calibration data  menu is displayed  you can choose one of  the following options     m Press OK Use Existing Data  side  to use the Probe Cal data last stored to  compensate the probe     m Press OK Erase Probe Cal Data  side  to erase the Probe Cal data last  stored and use the probe uncompensated     m Press CLEAR MENU on the front panel to retain the Probe Cal data last  stored and use the probe uncompensated        NOTE  If the Re use Probe Calibration data menu is displayed  do not select OK  Use Existing Data if the probe currently installed is not of the same impedance  stored for the Probe Cal  For example  if the last Probe Cal stored for a channel  was done with a passive 50 Q probe installed  do not install a passive 1 MQ  probe and select OK Use Existing Data if the menu appears  If you do so  most  of any signal you attempt to measure will not be coupled to the input channel  because of the probe to oscilloscope impedance mismatch        TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 113    Measu
97. FT button is on and the display says Coarse Knobs in the upper  right corner  the POSITION knob positions waveforms more quickly     The POSITION knob simply adds screen divisions to the reference point of  the selected waveform  Adding divisions moves the waveform up and  subtracting them moves the waveform down  You also can adjust the  waveform position using the offset option in the Vertical menu  discussed  later in this section         7       Tek Run  TOOKS sample        Ch  Coupling  Impedance             DC          AC                               Oloo     Vertical Readout 00m   OM Soos Chit zem  Q  T a a a AR E 50    Em en candwidth  M2 Scale  position   offset    Deskew   cal Probe  Full zoomy a div au   as Pass    DID I C IC IO  C O                                                     Figure 3 6  Vertical Readouts and Channel Menu    By changing the vertical scale  you can focus on a particular portion of a  waveform  By adjusting the vertical position  you can move the waveform up or  down on the display  Adjustment of vertical position is particularly useful when  you are comparing two or more waveforms     3 12 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       To Change  Vertical Parameters    DC  ACM  GND    To select the coupling  bandwidth  and offset for the selected waveform  use the  Vertical menu  Figure 3   6   This menu also lets you numerically change the  position or scale instead of using the vertical knobs  
98. Factors  side  to return  the zoomed derivative waveform to no magnification     The Advanced DSP Math capabilities of the TDS Oscilloscope include  waveform integration  This capability allows you to display an integral math  waveform that is an integrated version of the acquired waveform  This section  describes how to setup the oscilloscope to display and measure integral math  waveforms     Integral waveforms find use in the following applications   m Measuring of power and energy  such as in switching power supplies    m Characterizing mechanical transducers  as when integrating the output of an  accelerometer to obtain velocity    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 171    Using Features for Advanced Applications       To Create a Integral Math  Waveform    3 172    The integral math waveform  derived from the sampled waveform  is computed  based on the following equation     n  y n    scale  gt  L Dy  i l  Where  x i  is the source waveform    y n  is a point in the integral math waveform  scale is the output scale factor  T is the time between samples    Since the resultant math waveform is an integral waveform  its vertical scale is in  volt seconds  its horizontal scale is in seconds   The source signal is integrated  over its entire record length  therefore  the math waveform record length equals  that of the source waveform     To obtain an integral math waveform  do the following steps     1  Connect the waveform to the desired channel input and sele
99. Figure 3 19         Tek Run  250kKS 5 Sample       Palettes    4    Bold                           Hardcopy  Preview          Monochrome             260mvV  Persistence  Palettes    ee  Change ap ap 3 Restore  Settings    Math Reference   Options   lt  Colors Normal Colors Colors Colors Colors                                                                Gpe       Figure 3 19  Display Menu     Setting    The Color menu allows you to alter color settings for various display compo   nents such as waveforms and text  The Display menu allows you to adjust the  style  intensity level  graticule  and format features  To find more information on  display  see Change the Display Settings on page 3 29     Choose a Palette To choose a palette of 13 colors from a menu of preset palettes   1  Choose the starting palette by selecting Palette from the main menu     2  Select one of the available palettes in the side menu  Choose from Normal   Bold  Hardcopy Preview or Monochrome     3  Ifyou are using a persistence display and want to vary the color of each  point depending on its persistence  choose Persistence Palettes  Then  choose Temperature  Spectral  or Gray Scale from the resulting side menu   Choose View Palette to preview your selection on the display  Press    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 35    Acquiring and Displaying Waveforms       Change the Palette Colors    3 36    Persistence Palette to quit preview mode  Press Clear Menu to return to the  Palette men
100. Frequency   1 Period  High 100   highest  voltage reference value   See High  Low on page B 1    Tf Using the min max measurement technique     High   Max    TDS 500B  TDS 600B   amp  TDS 700A User Manual B 7    Appendix B  Algorithms       Low   Maximum   Mean   Minimum   Negative Duty Cycle                   Negative Overshoot       0   lowest  voltage reference value calculated   See High  Low on page B 1    Using the min max measurement technique     Low   Min    Amplitude  voltage  measurement  The maximum voltage  Typically the most  positive peak voltage     Examine all Waveform    samples from Start to End inclusive  and set Max  equal to the greatest magnitude Waveform    value found     The arithmetic mean for one waveform  Remember that one waveform is not  necessarily equal to one cycle  For cyclical data you may prefer to use the cycle  mean rather than the arithmetic mean     If Start   End then return the  interpolated  value at Start     Otherwise     End    Waveform t dt  Mean  Start        End     Start  x SampleInterval  For details of the integration algorithm  see page B 12     Amplitude  voltage  measurement  The minimum amplitude  Typically the most  negative peak voltage     Examine all Waveform    samples from Start to End inclusive  and set Min  equal to the smallest magnitude Waveform    value found     Timing measurement  The ratio of the negative pulse width to the signal period  expressed as a percentage     NegativeWidth is defined in Negative W
101. Index       Reference Indicator  Channel  3   8  Reference levels  2   21  Defining for Measure  3   97  Reference Levels  Measure menu  3   97  Reference memory  Glossary   9  Reject Glitch  Main Trigger menu  3   76  Remote communication  3   134   3    138  Remote operation  Communicating with Remote Instruments  3 134  Connecting to the GPIB  3 136  GPIB interface requirements  3 134  GPIB Protocols  3 134  Interconnect cabling  3 135  Preparation for  3 134  Procedures for  3 135  Selecting and configuring the port  3 136  Remove Measrmnt  Measure menu  3   95  3   99  Rename  File Utilities menu  3 123  Repetitive Signal  Acquire menu  3   25  Reset  How to execute  3   8  Reset All Mappings To Factory  Color menu  3 38  Reset All Palettes To Factory  Color menu  3   38  Reset Current Palette To Factory  Color menu  3   38  Reset to Factory Color  Color menu  3 36  Reset Zoom Factors  Zoom menu  3 42  Restore Colors  Color menu  3   38  Ring Bell if Condition Met  Acquire menu  3 146  Rise time  2   20  3 93  Glossary   9  Rising edge  Delayed Trigger menu  3   89  Rising edge  Main Trigger menu  3   62  3 70  3 71  RLE Color  Hardcopy menu  3 127  RMS  3 93  Glossary   9  RS 232  2 5  RS 232  Port  3 128  3 134  RS232C Centronics Hardcopy Interface  Option 13   A 2  RUN STOP  3 50  RUN STOP  Acquire menu  3 27  Runt trigger  3 73  3 74  Glossary   9  How to set up  3   77   3 90  Runt  Main Trigger menu  3 77    S    Sample acquisition mode  3   21  Glossary   9  Sampl
102. Math capabilities of the TDS Oscilloscope include  waveform differentiation  This capability allows you to display a derivative math  waveform that indicates the instantaneous rate of change of the waveform  acquired  This section describes how to setup the oscilloscope to display and  measure derivative math waveforms     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 167    Using Features for Advanced Applications       To Create a Derivative of a    3 168    Waveform    Derivative waveforms are used in the measurement of slew rate of amplifiers and  in educational applications  You can store and display a derivative math  waveform in a reference memory  then use it as a source for another derivative  waveform  The result is the second derivative of the waveform that was first  differentiated     The math waveform  derived from the sampled waveform  is computed based on  the following equation     Y     Xa   z x4    Where  X is the source waveform  Y is the derivative math waveform  T is the time between samples    Since the resultant math waveform is a derivative waveform  its vertical scale is  in volts second  its horizontal scale is in seconds   The source signal is differen   tiated over its entire record length  therefore  the math waveform record length  equals that of the source waveform     To obtain a derivative math waveform   Connect the waveform to the desired channel input and select that channel     2  Adjust the vertical and horizontal scales and trigger t
103. Measurements  3 95  Glossary 4  Gating  Measure menu  3 95  General purpose  high input resistance  probes  E 1  General purpose knob  2 7  2   21  Glossary   4  Glitch trigger  3 73  3 74  Glossary   4  How to set up  3 75  Glitch  Main Trigger menu  3   76  Goes FALSE  Main Trigger menu  3   68  3   70  Goes TRUE  Main Trigger menu  3 68  3   70  GPIB  2 5  3   134   3 138  Glossary   4  Connecting to  3 136  Interconnect cabling  3 135  Interface requirements  3 134  Procedures for using  3 135  Protocols  3 134  Selecting and configuring the port  3 136  Typical configuration  3 135  GPIB  Hardcopy menu  3 128  GPIB  Utility menu  3 136  Graticule  3 33  Glossary4  Graticule measurements  3 105  Graticule  Display menu  3 33  Grid  Display menu  3   33  Ground coupling  Glossary   5  GROUP 1  GROUP 2    buttons  3 40    H    H Bars  Cursor menu  3   104  H Limit  Acquire menu  3 145  Hamming window  3 154  Hanning window  3 154  Hardcopy  3 125  Glossary   5   How to print  controller   3 132   How to print  no controller   3 130   How to save to disk  3 131   How to set up for  3 127   Printing a hardcopy  3 125   Spool  3 130  3 131  Hardcopy  Talk Only   Utility menu  3 127  HARDCOPY button  3 122  3 127  3 136  Hardcopy if Condition Met  Acquire menu  3 146  Hardcopy menu   BMP Color  3 127   BMP Mono  3 127   Clear Spool  3 128  3 131    Index 6    Deskjet  3 127   DeskjetC  3 127   DPU411 II  3 127   DPU412  3 127   EPS Color Img  3 127   EPS Color Plt  3 127   EPS 
104. Mono Img  3 127   EPS Mono Plt  3 127   Epson  3 127   Format  3   127   GPIB  3 128   HPGL  3 127   Interleaf  3 127   Landscape  3 128   Laserjet  3 127   Layout  3 128   OK Confirm Clear Spool  3 131   Palette  3 128   PCX  3 127   PCX Color  3 127   Port  3 128   Portrait  3 128   RLE Color  3 127   Thinkjet  3 127   TIFF  3 127  Hardcopy  Color menu  3 35  Hardcopy  Utility menu  3 136  Help  Accessing  3 139  HELP button  3 141  Help system  3 139  HF Rej  Main Trigger menu  3   60  Hi Res  Incompatible with InstaVu  3 48  Hi Res acquisition mode  3   23  Glossary   5  Hi Res  Acquire menu  3 25  High  3 92  Glossary   5  High Ref  Measure menu  3 98  High speed active probes  E 3  High voltage probes  E 2  High Low Setup  Measure menu  3   96  HiRes acquisition mode  3   50  Histogram  Measure menu  3   96  Holdoff  trigger  3 53  Glossary   5  Horiz Pos  Horizontal menu  3   17  Horiz Scale  Horizontal menu  3   16  Horizontal   Bar cursors  3 101  Glossary   5   Control  3   14 3 50   Menu  3 55   Position  3   14   Readouts  3   14   Scale  3 14   SCALE knob  2 13   System  2 13    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index       Horizontal menu  3   86   Delayed Only  3 86   Delayed Runs After Main  3 17  3   86   Delayed Scale  3   16   Delayed Triggerable  3   17  3   86   FastFrame  3   49   FastFrame Setup  3   49   Fit to screen  3 16   Frame Count  3   50   Frame Length    3 50   Frame   3   50   Horiz Pos  3   17   Horiz Scale  3   16   Intensif
105. OTH the main and delayed triggering  menus forces main and delayed trigger levels to adjust in tandem  As long as  their source remains DC Aux  adjusting the trigger level for either system adjusts  it for both systems        9  Press Coupling  main   gt  Main Trigger  DC  or Noise Rej  side  to define  how the input signal will be coupled to the delayed trigger     Main Trigger sets delayed trigger coupling to match the main trigger  coupling setting  For descriptions of the DC and Noise Rej coupling types   see To Specify Coupling on page 3 60     10  Press Slope  main  to select the slope that the delayed trigger will occur on   Choose between the rising edge and falling edge slopes     When using Delayed Triggerable mode to acquire waveforms  two trigger  bars are displayed  One trigger bar indicates the level set by the main trigger  system  the other indicates the level set by the delayed trigger system     11  Press Level  main   gt  Level  Set to TTL  Set to ECL  or Set to 50   side      Level lets you enter the delayed trigger level using the general purpose knob  or the keypad     Set to TTL fixes the trigger level at  1 4 V   Set to ECL fixes the trigger level at    1 3 V     Set to 50  fixes the delayed trigger level to 50  of the peak to peak value  of the delayed trigger source signal     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 89    Triggering on Waveforms          NOTE  When you set the Vertical SCALE smaller than 200 mV  the oscilloscope  reduces the
106. Once these basics are covered  this section describes how to trigger using the  various trigger types provided by the Main trigger system  edge  logic  and pulse     m To use the    general purpose    trigger type  edge  see Triggering on a  Waveform Edge on page 3 59     m To logic trigger based on an input pattern  state  or setup hold violation  see  Triggering Based on Logic on page 3 62     m To pulse trigger based on various pulse types  glitch  runt  or their parame   ters  width  slew rate  see Triggering on Pulses on page 3 72     This section concludes with details about and instructions for using the Delayed  time base and Delayed trigger system to delay the acquisition of a waveform  relative to a trigger event   See Delayed Triggering on page 3 84      Triggers determine when the oscilloscope stops acquiring and displays a  waveform  They help create meaningful waveforms from unstable jumbles or  blank screens   See Figure 3 30   The oscilloscope has four types of triggers   edge  logic  pulse  and  with option 5  video                                                                                                                                   Triggered Waveform Untriggered Waveforms    Figure 3 30  Triggered Versus Untriggered Displays    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 51    Triggering on Waveforms       3 52          The Trigger Event    Trigger Sources    ACM    Trigger Types       The trigger event establishes the time zero point in the 
107. Optimizing Measurement Accuracy  SPC and Probe Cal    3 106    Signal Path  Compensation    The TDS Oscilloscope provides three features that optimize measurement  accuracy  Signal Path Compensation  SPC  lets you compensate the internal  signal path used to acquire the waveforms and measure based on the ambient  temperature  Channel Probe Deskew lets you compensate for the fact that signals  may come in from cables of different length  Probe Cal lets you compensate the  entire signal path  from probe tip to digitized signal  to improve the gain and  offset accuracy of the probe  This section tells you how to use both features     The TDS Oscilloscope lets you compensate the internal signal path used to  acquire the waveforms you measure  SPC optimizes the oscilloscope capability  to make accurate measurements based on the ambient temperature     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       Run an SPC anytime you wish to ensure that the measurements you make are  made with the most accuracy possible  You should also run an SPC if the  temperature has changed more than 5   C since the last SPC was performed        NOTE  When making measurements at volts division settings less than or equal to  5 mV  you should run SPC at least once per week  Failure to do so may result in  the oscilloscope not meeting warranted performance levels at those volts div  settings   Warranted characteristics are listed in the Performance Verification  and Specification
108. S    Selected Graticule              gt  I           A oom  f                      o  of       Zoomed Waveform Edges          Zoom Boxes                   On             ut    Dual Zoom  ON         E                   praview  Q e AA     oe ea          Unzoomed Waveform       M 100ns  Chi    DZ Offset  472 685             LOJO O C O                                                    MIGIC  G   G  e       Figure 3 25  Dual Zoom     Shown Dual Window  Preview  Mode       NOTE  To make setting up Dual Zoom easier  turn on Preview in the side menu   In this dual window mode  the zoomed display appears in the top graticule   while the lower graticule shows the two zoomed portions enclosed in two boxes  on the unzoomed waveform  Adjusting offset moves the right box relative to the  left box  which remains stationary  The associated zoomed waveform in the  upper graticule moves to track the offset changes  You can also adjust the  waveform relative to the zoom boxes by selecting the lower graticule and  adjusting the vertical and horizontal control knobs  See Using Dual Window  Mode on page 3 42        3 44 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Using InstaVu    Acquisition Mode  TDS 500B and 700A Models Only     The TDS 500B and 700A Oscilloscopes can use InstaVu acquisition mode to  reduce the dead time between waveform updates that normally occur when  digitizing storage oscilloscopes  DSOs  acquire waveforms  InstaVu mode can  
109. S 600B   amp  TDS 700A User Manual 3 63    Triggering on Waveforms       3 64    Setup and Hold Class    For state triggering  the oscilloscope waits until the end of trigger holdoff and  then waits until the edge of channel 4 transitions in the specified direction  At  that point  the oscilloscope samples the inputs from the other channels and  triggers if the conditions defined in Table 3   6 are met     Table 3 6  Pattern and State Logic    AND Clocked AND If allthe preconditions selected for the  3 logic inputs3 are TRUE  then the  oscilloscope triggers     NAND af 1   Clocked NAND If not all of the preconditions selected  ae for the logic inputs  are TRUE  then the  oscilloscope triggers                          OR a gt    Clocked OR If any of the preconditions selected for  e the logic inputs  are TRUE  then the  oscilloscope triggers     7                NOR 3 1 Clocked NOR If none of the preconditions selected for  y the logic inputs3 are TRUE  then the  oscilloscope triggers     7                         1 Note that for state class triggers  the definition must be met at the time the clock  input changes state     2 The definitions given here are correct for the Goes TRUE setting in the Trigger When  menu  If that menu is set to Goes False  swap the definition for AND with that for  NAND and for OR with NOR for both pattern and state classes     3 The logic inputs are channels 1  2  3  and 4 when using pattern logic triggers  For  State Logic Triggers  channel 4 beco
110. TUP  gt  Recall Factory Setup  main   gt  OK Confirm Factory  Init  side      3  Press AUTOSET   4  Press CH2     The display shows a second waveform  which represents the signal on  channel 2  Since there is nothing connected to the CH 2 input connector  this  waveform is a flat line  There are several other important things to observe     m The channel readout on the display now shows the settings for both Ch1  and Ch2     m There are two channel indicators at the left edge of the graticule  Right  now  they overlap     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Tutorial       m The light above the CH 2 button is now on  and the CH 1 light is off   Because the knobs control only one channel at a time  the vertical  controls are now set to adjust channel 2     m The trigger readout still indicates that the trigger is detecting trigger  events on Ch1  The trigger source is not changed simply by adding a  channel   You can change the trigger source by using the TRIGGER  MENU button to display the trigger menu      5  Turn the vertical POSITION knob clockwise to move the channel 2  waveform up on the graticule  You will notice that the channel reference  indicator for channel 2 moves with the waveform     6  Press VERTICAL MENU  gt  Coupling  main      The VERTICAL MENU button displays a menu that gives you control over  many vertical channel parameters   See Figure 2   12   Although there can be  more than one channel displayed  the vertical menu and buttons only adjust
111. To make such changes  do  the following procedures     Coupling  To choose the type of coupling for attaching the input signal to the  vertical attenuator for the selected channel and to set its input impedance     Press VERTICAL MENU  gt  Coupling  main   gt  DC  AC  GND  or Q  side    m DC coupling shows both the AC and DC components of an input signal    m AC coupling shows only the alternating components of an input signal    m Ground  GND  coupling disconnects the input signal from the acquisition     m Input impedance lets you select either 1 M Q or 50 Q impedance        NOTE  If you select 50 Q impedance with AC coupling  the digitizing oscillo   scope will not accurately display frequencies under 200 kHz     Also  when you connect an active probe to the oscilloscope  such as the P6245    the input impedance of the oscilloscope automatically becomes 50 Q  If you then  connect a passive probe  like the P6139A   you need to set the input impedance  back to 1 MQ     The maximum volts div setting is reduced from 10 V to 1 V when you select 50 Q  impedance  See the discussion Input Impedance Considerations on page 3 5        Bandwidth  Bandwidth refers to the range of frequencies that an oscilloscope can  acquire and display accurately  that is  with less than 3 dB attenuation   If you  limit the upper limit for the higher frequency components by selecting 250 MHz  or 20 MHz  a Bw symbol will appear in the lower part of the display  To change  the bandwidth of the selected 
112. User Manual    Tektronix       TDS 500B  TDS 600B  amp  TDS 700A  Digitizing Oscilloscopes    070 9719 00    This document applies for firmware version 1 0  and above     ce    Copyright    Tektronix  Inc  All rights reserved  Licensed software products are owned by Tektronix or its suppliers and  are protected by United States copyright laws and international treaty provisions     Use  duplication  or disclosure by the Government is subject to restrictions as set forth in subparagraph  c  1  ii  of the  Rights in Technical Data and Computer Software clause at DFARS 252 227 7013  or subparagraphs  c  1  and  2  of the  Commercial Computer Software     Restricted Rights clause at FAR 52 227 19  as applicable     Tektronix products are covered by U S  and foreign patents  issued and pending  Information in this publication supercedes  that in all previously published material  Specifications and price change privileges reserved     Printed in the U S A   Tektronix  Inc   P O  Box 1000  Wilsonville  OR 97070 1000    TEKTRONIX  TEK  FastFrame    FlexFormat    and InstaVu    are registered trademarks of Tektronix  Inc     WARRANTY    Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three  3  years  from the date of shipment  If any such product proves defective during this warranty period  Tektronix  at its option  either  will repair the defective product without charge for parts and labor  or will provide a replacement i
113. Using Features for Advanced Applications       Normal Waveform of an       Tek Stopped  820 Acquisitions          i i    Mdth3 Zoom   0 5X Ve       Impulse Response    FFT Waveform of the  Magnitude Response    FFT Waveform of the  Phase Response                                                 Oloo     Ch  250mV  WENE  90 0    2 50kHz       M s00ps Chis 252V                                           oaoa       Figure 3 86  System Response to an Impulse    To Create an FFT To obtain an FFT of your waveform  do the following steps     1   2     3 152    Connect the waveform to the desired channel input and select that channel     Adjust the vertical and horizontal scales and trigger the display  or press  AUTOSET      The topic Offset  Position  and Scale  on page 3 159  provides in depth  information about optimizing your setup for FFT displays     Press MORE to access the menu for turning on math waveforms     Select a math waveform  Your choices are Math1  Math2  and  Math3  main      If the selected math waveform is not FFT  press Change Math Definition   side   gt  FFT  main   See Figure 3   87     Press Set FFT Source to  side  repeatedly until the channel source selected  in step 1 appears in the menu label     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       Tek Run  50 0MS s Sample       Define FFT  Waveform             Set FFT  Source to  chi          Set FFT     Vert Scale to  dBV RMS             Set FFT  Window to    S
114. WARNING  To avoid electrical shock  be sure that the power cord is disconnected  before checking the fuse        3     Check the fuse to be sure it is the proper type and rating  see Figure 1 1    You can use either of two fuses  Each fuse requires its own cap  see  Table 1 2   The oscilloscope is shipped with the UL approved fuse installed     Check that you have the proper electrical connections  The oscilloscope  requires 90 to 250 VACprms  continuous range  45 Hz to 440 Hz  and may  require up to 300 W     Connect the proper power cord from the rear panel power connector  see  Figure 1 1  to the power system     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Start Up                   Power Connector                                                                      Principal Power Switch                   ogo                                                 Fuse       Figure 1 1  Rear Panel Controls Used in Start Up    Table 1 2  Fuse and Fuse Cap Part Numbers    Fuse    0 25 inch x 1 25 inch  UL 198 6  3AG   6 A  FAST  250 V    5mm x 20 mm  IEC 127   5 A  T   250 V       Fuse Part Number  159 0013 00    159 0210 00          Fuse Cap  Part Number    200 2264 00    200 2265 00    Front Cover Removal To remove the front cover  grasp the left and right edges and snap the cover off  of the front subpanel   To reinstall the cover  align it to the front subpanel and    snap it back on      Power On To power on the oscilloscope  do the following steps     1  Check that the 
115. Waveforms             0c e cece cece cece cere eeees 3 51  Triggenng Concepts  ied ei Eile iin ee teeth On ee a E 3 51  Triggering from the Front Panel            00    eee eee cee ee eee 3 55  Triggering on a Waveform Edge               0 0  e cece eee eens 3 59    TDS 500B  TDS 600B   amp  TDS 700A User Manual i    Table of Contents       Triggering Based on Logic    1 0 0      ee cee eee 3 62  Triggering on Pulses isso  etsy id tpo o bag ES ag Re Sa 3 72  Delayed  Ting Sern gy  sinnti kee si bag sae Fees Pew eats Yes ees 3 84  Measuring Waveforms          cccceccccsccsccccccsscsceeees 3 91  Taking Automated Measurements                0 0 0 0  c eee eee eee 3 91  Taking Cursor Measurements           0 0 0    e eee eee eee eee 3 101  Taking Graticule Measurements               0 0    c eee eee eee eee ee 3 105  Optimizing Measurement Accuracy  SPC and Probe Cal                    3 106  Saving Waveforms and Setups            cece ccc ceccccsscscees 3 115  Saving and Recalling Setups          0 0 0    cee eee ee eee 3 115  Saving and Recalling Waveforms             0 0 0  c eee eee eee eee 3 118  Managing the File System           00 0    eee eee eee 3 121  Printing a Hardcopy       0 0 0    ce eee eee teens 3 125  Communicating with Remote Instruments                       000000005 3 134  Determining Status and Accessing Help               ceceeeeees 3 139  Displaying Status  s eon ere buisa nee pewiay Cana ee bees 3 139  Displaying Help  stab Aves ee Meee ela ee oo as a a
116. able from 250 ns  minimum holdoff available  to 12 seconds   maximum holdoff available   To see how to set holdoff  see To Set Mode  amp   Holdoff on page 3   61            You can also set a default holdoff  The default hold is the    general purpose     holdoff for most triggering signals and varies with the horizontal scale  It is  equal to 5 divisions times the current time division settings     Trigger Coupling Trigger coupling determines what part of the signal is passed to the trigger    circuit  All trigger types except edge triggering use only DC coupling  edge  triggering can use all available coupling types  AC  DC  Low Frequency  Rejection  High Frequency Rejection  and Noise Rejection  See To Specify  Coupling on page 3   60 for a description of each coupling mode     Trigger Position The adjustable feature trigger position defines where on the waveform record the    trigger occurs  It lets you properly align and measure data within records  The  part of the record that occurs before the trigger is the pretrigger portion  The part  that occurs after the trigger is the posttrigger portion     T To help you visualize the trigger position setting  the top part of the display has    an icon indicating where the trigger occurs in the waveform record  You select in  the Horizontal menu what percentage of the waveform record will contain        pretrigger information     Displaying pretrigger information can be valuable when troubleshooting  For  example  if you are try
117. acquisitions   The oscilloscope saves the highest and  lowest values in two adjacent intervals similar to the Peak Detect mode  But  Envelope mode  unlike Peak Detect  gathers peaks over many trigger events     After each trigger event  the oscilloscope acquires data and then compares the  min max values from the current acquisition with those stored from previous  acquisitions  The final display shows the most extreme values for all the  acquisitions for each point in the waveform record     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 21    Acquiring and Displaying Waveforms          Single Waveform Acquisition                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     Samples Acquired in Four Acquisition Displayed Waveform Drawn  Acquisition Intervals Mode Record Points on CRT  Interval 1 2 3 4 Interval 1 2 3 4  e P       e rig Q  e e              0  Coo meee  pen it Sample Q  Uses first sample in  gt   interval  Use for fastest acquisition rate  This is the default mode  S
118. ak Peak Amplitude Measurement of a   Derivative Waveform             cece ew cece cece cere eee eens 3 170  Figure 3 97  Integral Math Waveform              eceeeeeceees 3 173  Figure 3 98  H Bars Cursors Measure an Integral Math Waveform 3 174  Figure B   1  MCross Calculations              cccceccecscceces B 4  Figure B   2  Fall Time             ccc cece ccc ccc ec cence ceeces B 7  Figure B 3  Rise Time          ossssesesesesesesoecococosoeoo B 11  Figure B 4  Choosing Minima or Maxima to Use for Envelope   Measurements  saoe  scs 0 5 oeenn o R OREN as B 13  Figure E 1  Typical High Voltage Probes                 0eeeeee E 2    Figure E 2  A6303 Current Probe Used in the AM 503S Opt  03     E 4    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Table of Contents       List of Tables    Table 1 1  Key Features and Differences of Models                1 2  Table 1 2  Fuse and Fuse Cap Part Numbers                005  1 5  Table 3 1  Autoset Defaults             ccc ccc cece cece eee eeeene 3 7  Table 3 2  How Interleaving Affects Sample Rate                 3 21  Table 3 3  Additional Resolution Bits             c cece eee e eens 3 24  Table 3 4  TDS 500B and TDS 700A Sampling Mode Selection    When Fit to Screen is Off           ccc ccc c cece ccc ce ec ceces 3 26  Table 3 5  XY Format Pairs            ccc cece cece cece eeeees 3 34  Table 3 6  Pattern and State Logic            ccc e cece wesc eeeee 3 64  Table 3 7  Pulse Trigger Definitions               cece eceeeeees
119. aliasing is 4  the sample frequency  Since source waveforms often have a fundamental  frequency that does not alias but have harmonic frequencies that do  you should  have methods for recognizing and dealing with aliases     m Be aware that a source waveform with fast edge transition times creates  many high frequency harmonics  These harmonics typically decrease in  amplitude as their frequency increases     m Sample the source signal at rates that are at least 2X that of the highest  frequency component having significant amplitude     m Filter the input to bandwidth limit it to frequencies below that of the Nyquist  frequency     m Recognize and ignore the aliased frequencies     If you think you have aliased frequencies in your FFT  select the source channel  and adjust the horizontal scale to increase the sample rate  Since you increase the  Nyquist frequency as you increase the sample rate  the alias signals should  appear at their proper frequency     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 161    Using Features for Advanced Applications       Considerations for Phase    3 162    Displays    Nyquist Frequency  Point    Amplitude    allied atest    Frequency        Aliased Frequencies Actual Frequencies    Se    Figure 3 92  How Aliased Frequencies Appear in an FFT    When you set up an FFT math waveform to display the phase angle of the  frequencies contained in a waveform  you should take into account the reference  point the phase is measured against  You ma
120. ally  all channels in use are individually scaled  If you use Autoset when no  channels are displayed  the oscilloscope will turn on channel one  CH 1  and  scale it        NOTE  Autoset may change vertical position in order to position the waveform  appropriately  It always sets vertical offset to O V        TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       List of Autoset Defaults    Table 3   1 lists the autoset defaults     Table 3 1  Autoset Defaults    Control  Selected channel  Acquire Mode    Changed by Autoset to  Numerically lowest of the displayed channels  Sample       Acquire Repetitive Signal   TDS 500B and 700A Models  Only     On          Acquire Stop After RUN STOP button only  Deskew  Channel P robe Unchanged   Display Style Vectors   Display Intensity     Overall If less than 50   set to 75   Display Format YT   FastFrame  TDS 500B and Off   700A Models Only        Horizontal Position  Horizontal Scale          Centered within the graticule window  As determined by the signal frequency                Horizontal Time Base Main Only   Horizontal Record Length Unchanged   Horizontal Lock Unchanged   InstaVu Acquisitions Unchanged    TDS 500B and 700A Models   Only    Limit Test Off   Trigger Position Unchanged   Trigger Type Edge   Trigger Source Numerically lowest of the displayed channels  the selected  channel    Trigger Level Midpoint of data for the trigger source   Trigger Slope Positive   Trigger Coupling DC     
121. als where you are only concerned with monitoring amplitude  level  Although the unsynced waveform may    roll    across the display  it will not  freeze as it would in normal trigger mode  Monitoring of a power supply output  is an example of such an application     Trigger Holdoff When the oscilloscope recognizes a trigger event  it disables the trigger system  until acquisition is complete  In addition  the trigger system remains disabled  during the holdoff period that follows each acquisition  You can set holdoff time  to help ensure a stable display     For example  the trigger signal can be a complex waveform with many possible  trigger points on it  Though the waveform is repetitive  a simple trigger might  get you a series of patterns on the screen instead of the same pattern each time     A digital pulse train is a good example of a complex waveform   See Fig    ure 3 31   Each pulse looks like any other  so many possible trigger points exist   Not all of these will result in the same display  The holdoff period allows the  oscilloscope to trigger on the correct edge  resulting in a stable display     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 53    Triggering on Waveforms       3 54    Acquisition Acquisition  Interval Interval       Trigger Level voa O m E one 6111 7  Trigger P oints    yY YY yY  Holdoff Holdoff Holdoff             Triggers are not recognized during holdoff time     Figure 3 31  Trigger Holdoff Time Ensures Valid Triggering    Holdoff is sett
122. ample          L A  2 72ms        1 77ms      Chi Brstwd    2 039ms                   Gate with  V Bar Cursors                            mY       M SO0Ms chi 7 252mv    VEIT INE a a TOE a POTTED  Remove Gating Reference   Measrmnt setup   for cht a o E ae             Ooo JO                                      CoCo ool       Figure 3 54  Measure Menu     Gating    The oscilloscope provides two settings  histogram and min max  for specifying  how measure determines the High and Low levels of waveforms  To use the  high low setup do the following step     Press MEASURE     gt  High Low Setup  main   gt  Histogram or Min Max   side   If you select Min Max  you may also want to check and or revise values  using the Reference Levels main menu     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       Histogram sets the values statistically  It selects the most common value either  above or below the midpoint  depending on whether it is defining the high or  low reference level   Since this statistical approach ignores short term aberrations   overshoot  ringing  etc    histogram is the best setting for examining pulses     Min max uses the highest and lowest values of the waveform record  This setting is  best for examining waveforms that have no large  flat portions at a common value   such as sine waves and triangle waves     almost any waveform except for pulses     Define Reference Levels Once you define the reference levels  the oscilloscope will use the
123. annel readout  It  shows the selected channel in inverse video in the lower left corner of the  display  The channel reference indicator for the selected channel also appears in  reverse video along the left side of the display   See Figure 3 3      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Channel Reference  Indicator    Channel Readout    To Select and  Remove Waveforms                 Tek Run  100kS 5 Sample       Recall  Factory                                         amom  O M 500ps Chive 27 2mV OK    i   l     i       Confirm    _  Factory Init    scat pei H iat we ne ey ie  Saved Current Factory Utilities  Setup Setup Setu    CIC   C lO  C              LOIDID COle                                                 Figure 3 3  The Channel Readout    To select a channel  use the channel selection buttons on the right of the display   These buttons labeled CH 1  CH 2  CH 3  CH 4  and MORE select a channel  and display it if its off   The MORE button allows you to select internally stored  Math and Ref waveforms for display and manipulation   The selected channel is  indicated by the lighting the LED above the button of the selected channel     Do the following steps to first display and then remove waveforms from the  display     1  Press CH 1  CH 2  CH 3  or CH 4 to turn on as many of these channels as  desired  The one you select last  or first if you only select one  becomes the  selected channel  Selecting a channel turns 
124. araia     aa  Type Class Logic when  amp   stogic gt    lt Pattern gt  Inputs AND True  gt   Thresholds Holdoff    Oil  OlM  e                                              LOIDID OI O      Time Logic Function Must be TRUE                                                       Figure 3 41  Logic Trigger Menu     Time Qualified TRUE    To State Trigger When you select the logic class State  the oscilloscope uses channel 4 as a clock  and triggers on a logic circuit made from the rest of the channels  pages 3   62  through 3   64 describe how state triggers work   To use state triggering  do the  following procedures     Select State Triggering  Press TRIGGER MENU  gt  Type  main   gt  Logic   pop up   gt  Class  main   gt  State  pop up      Define Inputs  To set the logic state for each of the input channels  Ch1  Ch2           1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  State  pop up   gt  Define Inputs  main      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 69    Triggering on Waveforms       3 70    To Trigger on Setup   Hold Time Violations    2  Choose either High  H   Low  L   or Don   t Care  X   side  for the first three  channels  The choices for Ch4 are rising edge and falling edge     Set Thresholds  To set the logic threshold for each channel     1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  State  pop up   gt  Set Thresholds  main   gt  Ch1  Ch2   Ch3  or Ch    side      2  Use the MAIN
125. ardcopy in Process   Press HARDCOPY to abort     remains displayed  You can abort only the  ast hardcopy sent by pressing the  button while the message is still displayed  When the printer empties enough of  the spool to finish adding the last hardcopy  it does so and then removes the  message     Clear the Spool  To remove all hardcopies from the spool  press SHIFT  gt   HARDCOPY MENU  gt  Clear Spool  main   gt  OK Confirm Clear Spool   side      The oscilloscope takes advantage of any unused RAM when spooling hardcopies  to printers  The size of the spool is  therefore  variable  The number of hardco   pies that can be spooled depends on three variables     m The amount of unused RAM  m The hardcopy format chosen  m The complexity of the display    Although not guaranteed  usually about 2 5 hardcopies can be spooled before the  oscilloscope must wait to send the rest of the third copy     To Save toa Floppy Disk To send hardcopies to a floppy disk  do the following steps     1  Set up the oscilloscope communication and hardware parameters as outlined  in To Set Up for Making Hardcopies on page 3 127     2  Insert a formatted 720 Kbyte or 1 44 Mbyte floppy disk into the slot at the  left of the oscilloscope display        NOTE  To format disks  delete hardcopy files you save to disk  and otherwise  manage the disk storage  see Managing the File System on page 3 121        3  Press SHIFT  gt  HARDCOPY MENU  gt  Port  main   gt  File  side  to  specify that any hardcopy made 
126. ay     The cursor units will be in dB or volts for FFTs measuring magnitude and in  degrees or radians for those FFTs measuring phase  The cursor unit depends  on the selection made for Set FFT Vert Scale to  side   See step 7 on page  3 153 for more information     Press V Bars  side   Use the general purpose knob to align one of the two  vertical cursors to a point of interest along the horizontal axis of the  waveform     Press SELECT to select the alternate cursor   Align the selected cursor to another point of interest on the math waveform     Read the frequency difference between the cursors from the A  readout  Read  the frequency of the selected cursor relative to the zero frequency point from  the    readout     The cursor units will always be in Hz  regardless of the setting in the Time  Units side menu  The first point of the FFT record is the zero frequency  point for the    readout     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 155    Using Features for Advanced Applications       To Take Automated    Measurements of an FFT    3 156    The FFT Frequency  Domain Record    Tek Run  50 0MS s Sample    ae                                           LOIDID IC  O      DE         M i 00us Chi    1 34 V          Math 10 0 dB 2 50MHz           111 1  1 EaI                                           Figure 3 89  Cursor Measurement of an FFT Waveform    10  Press Function  main   gt  Paired  side      11  Use the technique just outlined to place the vertical bar of each 
127. base  Glossary   10  Time Base  Horizontal menu  3   86  Time Units  Cursor menu  3   105  Time  Main Trigger menu  3   84  Time to voltage converter  E 5  Timeout Trigger  3   74  Timeout trigger  3 73  3 83  Glossary   11  How to set up  3 83 3 90  Timeout  Main Trigger menu  3   83  3 84  Tracking  Cursor menu  3 104  Trigger  3 5 1 3   90  Glossary   11  AC Line Voltage  3 52  Auxiliary  3   52  Coupling  3 54  Delay  3 55  Delayed  3   85   3   90  Edge  3 52  3 59  Glossary   3  Glitch  3 73  3 74  Glossary   4  Holdoff  3   53  Level  3 55  Glossary   11  Logic  3 52  3 62  3   64  Mode  3 53    Index 13    Index       Pattern  3   62  3   66  Position  3   16  3   50  3 54  Pulse  3 52  3 72  Readout  3   57  3   58  Runt  3 73  3 74  Glossary   9  Setup Hold  3   63  3   64  3   70 3    90  Slew Rate  3   74  Glossary   10  Slew rate  3   73  Slope  3 55  Source  3   52  State  3 63  3 69 3 90  Status Lights  3 57  Timeout  3 73  3 74  3 83  Glossary   11  Types  3 59 3 90  Video  3   53  Width  3 73  3 79  Trigger Bar  2   6  3 50  Trigger Bar Style  Display menu  3 31  Trigger if Faster Than  Main Trigger menu  3 81  Trigger if Slower Than  Main Trigger menu  3 81  Trigger Level Bar  Readout  3 31  3 50  Trigger MAIN LEVEL knob  2 14  3 55  3 56  TRIGGER MENU button  3 59  3   60  3 66  3   69   3 70  3 75  3 77  3 80  Trigger Point  Readout  3 31  3 50  Trigger Position  Horizontal menu  3   16  Trigger Status Lights  3   57  Trigger When  Main Trigger menu  3   6
128. be    2 ns  lt  e  gt    Clock Level  Clock Signal   Setup Hold    lt        Violation        gt    T Zone  e   Ty  o  gt   Clock Level  Clock Signal  Setup Hold   lt     Violation     gt    Zone T   IH   s E  aes     Clock Level   Clock Signal  Positive Ts  Negative Ty  Negative Ts  Positive Ty          Figure 3 38  Violation Zones for Setup Hold Triggering    Negative settings for setup or hold times skew the setup hold violation zone to  locate it before or after the clocking edge   See the bottom and center waveforms    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 65    Triggering on Waveforms       To Check Logic  Trigger Status    To Trigger on a Pattern    3 66    of Figure 3 38   The oscilloscope can then detect and trigger on violations of a  time range that occurs before or one that occurs after the clock        NOTE  Keep hold time to at least 2 ns less than the clock period or the oscillo   scope cannot trigger        To quickly check if logic triggers are selected and if so  what class  check the  Trigger readout  When logic triggers are selected  the trigger readout displays the  selected logic trigger class  Pattern  State  or StlHld  Setup Hold    See   Figure 3 39      Ch 1  2  3 Inputs   High  Don   t Care  High    Ch 4 Input  Rising Edge  Trigger Class   State                                  Logic  OR      i        i   U  Edge           i     EEI omy O M S00us State HKS S 0  ulse Car Sa Pee a A a ee  video EEE EEE EAEE  FN ELEA L EAE    n Define ae Set
129. be output to a file on the floppy drive  The  file list and its scrollbar will appear     4  Turn the general purpose knob to place the scroll bar over the file in which to  store the hardcopy     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 131    Saving Waveforms and Setups       To Print Using a Controller    3 132       NOTE  Upon power on  the oscilloscope creates the    wild card    file    hardcopy format you select  Selecting this file and pressing Hardcopy stores a  hardcopy in a uniquely named  sequentially numbered file  For instance  the  oscilloscope saves the first hardcopy you save to the file TEKO0001 FMT  the  second to TEK00002 FMT  and so on        5  Press HARDCOPY to print your hardcopy to the selected file     Saving files to the floppy disk provides a convenient way to store hardcopies   You can print hardcopies stored on disk at a site remote from where the  hardcopies were captured using another TDS oscilloscope  Or you might load  stored hardcopies from disk into your desktop publishing software that runs on a  PC compatible computer     To make your hardcopies  use the procedures that follow     Connect to a Hardcopy Device  To connect a controller with two ports between the  oscilloscope and the hardcopy device  connect from the oscilloscope GPIB  connector  rear panel  to the controller GPIB port and from the controller  RS 232 or Centronics port to the hardcopy device   See Figure 3   74   Use the  GPIB port to remotely request and receive a 
130. bed occurs if you set the Horizontal SCALE knob to a  time base setting that is faster than 10 ns  TDS 600B    The setting varies with  the number of channels for TDS 500B and TDS 700A models  see Table 3   4 on  page 3 26   The oscilloscope then interpolates to create the intervening points in  the waveform record  There are two options for interpolation  linear or sin x  x    TDS 500B and TDS 700A models can also equivalent time sample to acquire  more samples  see Equivalent Time Sampling on page 3 19      Linear interpolation computes record points between actual acquired samples by  using a straight line fit  It assumes all the interpolated points fall in their  appropriate point in time on that straight line  Linear interpolation is useful for  many waveforms such as pulse trains     Sin x  x interpolation computes record points using a curve fit between the actual  values acquired  It assumes all the interpolated points fall along that curve  That  is particularly useful when acquiring more rounded waveforms such as sine  waves  Actually  it is appropriate for general use  although it may introduce some  overshoot or undershoot in signals with fast rise times        NOTE  When using either type of interpolation  you may want to set the display  style so that the real samples are displayed intensified relative to the interpolated  samples  The instructions under Select the Display Style on page 3 29 explain  how to turn on intensified samples        Interleaving A TDS 5
131. blishes the initialization settings shown in  Table D 1 to set the TDS Oscilloscope to a known default state     Table D 1  Factory Initialization Defaults    Control    Changed by Factory Init to    Acquire mode  Acquire repetitive signal    Sample  TDS 500B  amp  TDS 700A  ON  Enable ET        Acquire stop after    RUN STOP button only       Acquire  of averages       Acquire  of envelopes  Channel selection    16  10  Channel 1 on  all others off       Cursor H Bar 1 position  Cursor H Bar 2 position    Cursor V Bar 1 position       Cursor V Bar 2 position    10  of graticule height    3 2 divs from the center     90  of the graticule height    3 2 divs from the center     10  of the record length  90  of the record length                         Cursor amplitude units Base   Cursor function Off   Cursor mode Independent  Cursor time units Seconds  Date and time No change  Delay trigger average   16   Delay trigger envelope   10   Delay events  triggerable after main 2   Delay time  16 ns  delayed runs after main   Delay time  16 ns  delayed triggerable after main   Delayed  delay by     Delay by Time  Delayed edge trigger coupling Main Trigger  Delayed edge trigger level OV   Delayed edge trigger slope Rising       TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix D  Factory Initialization Settings       Table D 1  Factory Initialization Defaults  Cont      Control    Delayed edge trigger source    Changed by Factory Init to  Channel 1       Delayed  time base mo
132. cal offset  all channels  OV   Vertical position  all channels  0 divs              TDS 500B  TDS 600B   amp  TDS 700A User Manual    D 5    Appendix D  Factory Initialization Settings       D 6    Table D 1  Factory Initialization Defaults  Cont         Control Changed by Factory Init to  Vertical volts division  all channels  100 mV division  Zoom dual Off       Zoom dual offset    Minimum available       Zoom  dual window  selected graticule Upper  Zoom graticule Upper  Zoom horizontal  all channels  2 0X  Zoom horizontal lock All    Zoom horizontal position   all channels     50   0 5  the middle of the display              Zoom mode Off  Zoom vertical  all channels  2 0X  Zoom vertical position  all channels  0 divisions    TDS 500B  TDS 600B   amp  TDS 700A User Manual    ar ar  Appendix E  Probe Selection    The TDS Oscilloscope can use a variety of Tektronix probes for taking different  kinds of measurements  To help you decide what type of probe you need  this  section introduces the five major types of probes  passive  active  current   optical  and time to voltage probes  See Appendix A  Options and Accessories  for a list of the optional probes available  see your Tektronix Products Catalog  for more information about a given probe        NOTE  With some TDS models  Tektronix ships the recommended general pur   pose probes as standard accessories   The model probe and number shipped  depends on the model     see Probes in Table A 2 on page A 4   The TDS 680B   TDS
133. capture and display transient deviations  such as glitches or runt pulses  often  missed during longer dead times that accompany normal DSO display  This  section describes how to use InstaVu mode and how it differs from normal  acquisition mode     Waveform Capture Rate Figure 3 26 illustrates how InstaVu acquisition mode differs from the normal  acquisition mode used by digital storage oscilloscopes  Note that normal mode  follows a    capture waveform    digitize waveform   update waveform memory    dis   play waveform    cycle  Normal mode misses short term deviations occurring  during the long dead times  Typical waveform capture rates are 50 waveforms  per second     InstaVu mode increases the waveform capture rate to up to 400 000 waveforms  per second  maximum waveform rate depends on oscilloscope model   updating  the waveform array many times between displays  This very fast frame rate  greatly increases the probability that runts  glitches  and other short term changes  will accumulate in waveform memory  The oscilloscope then displays the  waveform at the normal display rate using variable or infinite persistence  You  can control how long the waveform persists on screen by selecting variable  persistence and setting a decay constant     To Use InstaVu Mode To turn on InstaVu mode  do either of the two following steps   1  Press the front panel button Insta Vu   See Figure 3 27      2  Press DISPLAY  gt  Mode  main   Push Mode again to toggle to InstaVu  mode   Se
134. ccessive samples in a time base  For real time  digitizers  the sample interval is the reciprocal of the sample rate  For  equivalent time digitizers  the time interval between successive samples  represents equivalent time  not real time     TDS 500B  TDS 600B   amp  TDS 700A User Manual Glossary 9    Glossary       Glossary 10    Sampling  The process of capturing an analog input  such as a voltage  at a discrete  point in time and holding it constant so that it can be quantized  Two general  methods of sampling are  real time sampling and equivalent time sampling     Setup Hold trigger  A mode in which the oscilloscope triggers when a data source changes state  within the setup or hold time relative to a clock source  Positive setup times  precede the clock edge  positive hold times follow the clock edge  The clock  edge may be the rising or falling edge     Select button  A button that changes which of the two cursors is active     Selected waveform  The waveform on which all measurements are performed  and which is  affected by vertical position and scale adjustments  The light over one of the  channel selector buttons indicates the current selected waveform     Side menu  Menu that appears to the right of the display  These selections expand on  main menu selections     Side menu buttons  Bezel buttons to the right of the side menu display  They allow you to select  items in the side menu     Slew Rate trigger  A mode in which the oscilloscope triggers based on how fast
135. ch channel becomes the pulse trigger source     Press TRIGGER MENU     gt  Type  main   gt  Pulse  pop up   gt   Source  main   gt  Ch1  Ch2  Ch3  or Ch4  side   The source selected becomes  the trigger source for all four trigger classes     Select the Polarity  To specify the direction of the runt pulse     Press TRIGGER MENU     gt  Type  main   gt  Pulse  pop up   gt  Class  main   gt   Runt  pop up   gt  Polarity  main   gt  Positive  Negative  or Either  side      Positive looks for positive going runt pulses   Negative looks for negative going runt pulses     Either looks for both positive and negative runt pulses     Set to Trig When  To determine how wide a runt pulse the oscilloscope will  trigger on     1  Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Runt  pop up   gt  Trig When  main      2  Press Occurs to trigger on all runt pulses regardless of width     3  Press Runt is Wider Than  side  to trigger only on runt pulses that exceed  the width you set  Enter the width using the general purpose knob or keypad     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 77    Triggering on Waveforms       Runt    Selected trigger bar at  upper threshold     Unselected trigger bar at lower  threshold     Pulse Crosses First Threshold    Only  Recrosses First Threshold    3 78    Level  and Triggers Acquisition            Tek Run  aur Sample Runt High Threshold  3 00 V          Pulse Runt  Thresholds                            1 00 V       
136. channel     Press VERTICAL MENU  gt  Bandwidth  main   gt  Full  250 MHz  or  20 MHz  side      Fine Scale  To make fine adjustments to the vertical scale  press VERTICAL  MENU  gt  Fine Scale  main  and use the general purpose knob or the keypad     Position  To adjust the vertical position to a specific number of divisions  press  VERTICAL MENU  gt  Position  main  and use the general purpose knob or the  keypad to set the offset value  Press Set to 0 divs  side  if you want to reset the  reference point of the selected waveform to the center of the display     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 13    Acquiring and Displaying Waveforms       To Check the  Horizontal Status    Offset  Use offset to subtract DC bias before examining a waveform  For example   you might want to display a small ripple  for example  100 mV of ripple  on a  power supply output  for example  a  15 V output   Adjust offset to keep the ripple  on screen while setting the vertical scale sensitive enough to best display the ripple     To adjust offset  press VERTICAL MENU  gt  Offset  main   Then use the  general purpose knob or keypad to set the vertical offset  Press Set to 0 V  side   if you want to reset the offset to zero     Check the Record View to determine the size and location of the waveform  record and the location of the trigger relative to the display   See Figure 3 7      Check the Time Base readout at the lower right of the display to see the  time division settings and th
137. cilloscope to a Controller    Select GPIB Port  To select the GPIB port  press SHIFT  gt  UTILITY  gt   System  main   gt  I O  pop up   gt  Port  main   gt  GPIB  pop up      Configure the GPIB Port  You must set two important GPIB parameters  mode  and address  To set those parameters     Press SHIFT     UTILITY  gt  System  main   gt  I O  pop up   gt  Port  main   gt   GPIB  pop up   gt  Configure  main   gt  Talk Listen Address  Hardcopy  Talk  Only   or Off Bus  side    See Figure 3 78      Talk Listen Address configures the port for controller based system operation   Use the general purpose knob or the keypad to define the address     Hardcopy  Talk Only  configures the port for the hardcopy output without  controller supervision  Once so configured  the oscilloscope will send the  hardcopy data to any listeners on the bus when the HARDCOPY button is  pressed     Pressing HARDCOPY with the port configured any other way causes an error   and the oscilloscope responds with a message saying the selected hardcopy port  is currently unavailable     Off Bus disconnects the oscilloscope from the bus     3 136 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups          Tek Run  100kS 5 Sample GPIB Address  1       GPIB  Configuration    E  Talk Listen  mal Address             GPIB Configuration Menu               Hardcopy   Talk Only           Off Bus                         Oloo        Zoomy   l CO M 500s Chif  248mV    Configure  System Port ithe 
138. condition for each logic input  channels 1  2  and 3    m The direction of the state change for the clock input  channel 4       J H 4 m The Boolean logic function     select from clocked AND  NAND  OR  and  NOR           S   S                          S  5    m The condition for triggering     whether the trigger occurs when the Boolean  function becomes TRUE  logic high  or FALSE  logic low     A setup hold trigger occurs when a logic input changes state inside of the setup and  hold times relative to the clock  When you use setup hold triggering  you define     m The channel containing the logic input  the data source  and the channel  containing the clock  the clock source     m The direction of the clock edge to use    m The clocking level and data level that the oscilloscope uses to determine if a  clock or data transition has occurred    m The setup and hold times that together define a time range relative to the clock    Pattern and State Classes Pattern and state triggers apply boolean logic functions to the logic inputs   Table 3   6 defines these four logic functions     For pattern triggering  the oscilloscope waits until the end of trigger holdoff and  then samples the inputs from all the channels  The oscilloscope then triggers if  the conditions defined in Table 3   6 are met   Goes TRUE or Goes FALSE must  be set in the Trigger When menu  The other settings in that menu are described  in To Define a Time Qualified Pattern Trigger on page 3 68      TDS 500B  TD
139. cope   s  ambient temperature has changed by more  than 5  C or once a week if vertical settings  of 5mV div or less are used                    y    Input signals OK  with AC components  Disconnect or otherwise  remove these signals prior to running SPC  Compensate  SPC will take 15 minutes to run      Signal Paths  oltage  Frequency ulse  system Path Reference   Response   Trigger   lt Cal gt  Pass Pass Pass Pass             Oooo oj                                                           E    1 1        Figure 3 62  Performing a Signal Path Compensation    The TDS Oscilloscopes allow you to adjust a relative time delay for each  channel  This feature lets you align the signals to compensate for the fact that  signals may come in from cables of differing lengths     The oscilloscope applies deskew values after it completes each acquisition   therefore  the deskew values do not affect logic triggering  Also  deskew has no  affect on XY display format     To set a channel probe deskew  do the following steps   m Press VERTICAL MENU  gt  Deskew  main      m Then use the general purpose knob or the keypad to set the deskew time  You  can also eliminate any deskew setting by pressing Set to 0 S  side      The TDS Oscilloscope lets you compensate the probe  based on the channel it is  connected to  to improve the gain and offset accuracy of the probe  By executing  Probe Cal on a channel with its probe installed  you can optimize the oscillo   scope capability to make accurate m
140. ct Single Acquisition Sequence without displaying the  Acquire and Stop After menus  press SHIFT FORCE TRIG  Now the RUN STOP  button operates as just described   You still must display the Acquire menu and  then the Stop After menu to leave Single Acquisition Sequence operation           Press Limit Test Condition Met  side  to acquire waveforms until  waveform data exceeds the limits specified in the limit test  Then acquisition  stops  At that point  you can also specify other actions for the oscilloscope to  take  using the selections available in the Limit Test Setup main menu        NOTE  For the oscilloscope to stop an acquisition when limit test conditions have  been met  limit testing must be turned ON using the Limit Test Setup main menu        Setting up limit testing requires several more steps  See Limit Testing on  page 3 143     Under certain conditions  a waveform may be aliased on screen  Read the  following description about aliasing and the suggestions for preventing it     About Aliasing  When a waveform aliases  it appears on screen with a frequency  lower than the actual waveform being input or it appears unstable even though  the light next to TRIG   D is lighted  Aliasing occurs because the oscilloscope  cannot sample the signal fast enough to construct an accurate waveform record    See Figure 3   16      Actual High Frequency Waveform  gt     3 28    Apparent Low frequency  Waveform Due to Aliasing    Sampled P oints    Figure 3 16  Aliasing    Metho
141. ct that channel     2  Adjust the vertical and horizontal scales and trigger the display  or press  AUTOSET      3  Press MORE  gt  Math1  Math2  or Math3  main   gt  Change Math  waveform definition  side   gt  Single Wfm Math  main      4  Press Set Single Source to  side   Repeatedly press the same button until the  channel source selected in step 1 appears in the menu label     5  Press Set Function to  side   Repeatedly press the same button until intg  appears in the menu label     6  Press OK Create Math Waveform  side  to turn on the integral math  waveform     You should now have your integral math waveform on screen  See Fig   ure 3   97  Use the Vertical SCALE and POSITION knobs to size and  position your waveform as you require     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications          Tek Run  5 00MS s Sample       Math              Integral Math Waveform J  Change Math   waveform     definition    re ee ees ee eee A  Source Waveform                    Average  2                                                 Oloo     ch SMN   3   M 10 0ys  chi 320mV  DHUA _5 0opvs  __ To ops E E    Math1 Math2 Matha g eae T    difftch     FTRT  Ref1 Ref2 Ref3                                                 oloo       Figure 3 97  Integral Math Waveform    To Take Cursor Once you have displayed your integrated math waveform  use cursors to measure  Measurements its voltage over time     1  Be sure MORE is selected  lighted  in the cha
142. d     The FFT windowing acts like a bandpass filter between the FFT time domain  record and the FFT frequency domain record  The shape of the window controls  the ability of the FFT to resolve  separate  the frequencies and to accurately  measure the amplitude of those frequencies     Selecting a Window  You can select your window to provide better frequency  resolution at the expense of better amplitude measurement accuracy in your FFT   better amplitude accuracy over frequency resolution  or to provide a compromise  between both  You can choose from these four windows  Rectangular  Hamming   Hanning  and Blackman Harris     In step 8  page 3 153  in To Create an FFT  the four windows are listed in order  according to their ability to resolve frequencies versus their ability to accurately  measure the amplitude of those frequencies  The list indicates that the ability of a  given window to resolve a frequency is inversely proportional to its ability to  accurately measure the amplitude of that frequency  In general  then  choose a  window that can just resolve between the frequencies you want to measure  That  way  you will have the best amplitude accuracy and leakage elimination while  still separating the frequencies     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       FFT Window we    FFT Time Domain Record After  Windowing    FFT Frequency Domain Record           Figure 3 93  Windowing the FFT Time Domain Record    You can 
143. d Applications          NOTE  The button labeled Stop After Limit Test Condition Met corresponds to the  Limit Test Condition Met menu item in the Stop After main menu  You can turn  this button on in the Limit Test Setup menu  but you cannot turn it off  To turn it  off  press Stop After and specify one of the other choices in the Stop After side  menu        3  Ensure that Limit Test  side  reads ON  If it reads OFF  press Limit Test   side  once to toggle it to ON     When you set Limit Test to ON  the oscilloscope compares incoming  waveforms against the waveform template stored in reference memory  according to the settings in the Limit Test Sources side menu     Single Waveform You can compare a single waveform against a single template  When making a  Comparisons single waveform versus a single template comparison  consider the following  operating characteristics     m The waveform will be repositioned horizontally to move the first sample in  the waveform record that is outside of template limits to center screen     m The position of the waveform template will track that of the waveform     Multiple Waveform You can also compare more than one waveform against a single template  or  Comparisons more than one waveform with each one compared against its own template or a  common template  When setting up for such comparisons  consider the following  operating characteristics     m You should set Horizontal Lock to None in the Zoom side menu  push  ZOOM and press  repea
144. d Setups       3 124    To Copy    To Print    To Create a Directory       Tek Run  100kS 5 Sample       Enter Label                                        Name  f OK  ECDEFGHI  KLMNOPQRSTUVWAYZ0123456789    amp  Accept  Enter Char     EE Delete Clear    SSIS S                                                          Figure 3 69  File System     Labeling Menu    To copy a file or directory  turn the general purpose knob until it scrolls the  cursor over the name of the file to copy  Then  press the side menu Copy button   The file menu will reappear with the names of directories to copy to  Select a  directory and press the side menu button labelled Copy  lt name gt  to Selected  Directory     To copy all files  select the     entry     The oscilloscope copies all directories recursively  That means it copies both the  directories and all their contents     To print a file  turn the general purpose knob until it scrolls the cursor over the  name of the file to print  Then  press the side menu Print button     The Print to side menu should appear  Select the port to print to from GPIB   RS 232  or Centronics  Then the oscilloscope will send the file in its raw form  out the port  The device  printer  receiving the file must be capable or printing  the particular file format     To create a new directory  press the side menu Create Directory button     The labeling menu should appear  Turn the general purpose knob or use the  main menu arrow keys to select each letter  Press En
145. d be  with respect to the other frequencies  Also  the selection of window type  introduces errors in the DC value of an FFT     The displayed output of the FFT on TDS oscilloscopes is corrected for these  errors to show the true value for the DC component of the input signal  The  Position and Offset must be set to zero for the source waveform in the Vertical  menu  When measuring the amplitude at DC  remember that 1 VDC equals   1 Vrms and the display is in dB     Record Length Most often  you will want to use a short record length because more of the FFT  waveform can be seen on screen and long record lengths can slow oscilloscope  response  However  long record lengths lower the noise relative to the signal and  increase the frequency resolution for the FFT  More important  they might be  needed to capture the waveform feature you want to include in the FFT     To speed up oscilloscope response when using long record lengths  you can save  your source waveform in a reference memory and perform an FFT on the saved  waveform  That way the DSP will compute the FFT based on saved  static data  and will only update if you save a new waveform     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 159    Using Features for Advanced Applications       Acquisition Mode    Zoom and Interpolation    3 160    Selecting the right acquisition mode can produce less noisy FFTs     Set up in Sample  Use sample mode until you have set up and turned on your  FFT  Sample mode can acquire repet
146. d only be used on repetitive signals     Record Points ier aia or  Ist Acquisition Cycle JL JL LL  2nd Acquisition Cycle       ee    3rd Acquisition Cycle   I  nth Acquisition Cycle       oe      Figure 3 12  Equivalent Time Sampling       The type of equivalent time sampling the oscilloscope uses is called random  equivalent time sampling  Although it takes the samples sequentially in time  it  takes them randomly with respect to the trigger  Random sampling occurs  because the oscilloscope sample clock runs asynchronously with respect to the  input signal and the signal trigger  The oscilloscope takes samples independently  of the trigger position and displays them based on the time difference between  the sample and the trigger     Interpolation Your oscilloscope can interpolate between the samples it acquires  Like for  equivalent time sampling  it does so only when it cannot obtain all the real  samples it needs to fill up its waveform record  For instance  setting the  horizontal SCALE to progressively faster acquisition rates leaves progressively  shorter time periods for the waveform record  Therefore  the oscilloscope must  sample faster to acquire the samples  record points  needed to fill up the record     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 19    Acquiring and Displaying Waveforms       Eventually the time period established by scale setting does not allow enough  time to get all the real samples needed to fill the record     The situation just descri
147. d region     _ Low     Min 7  NegativeOvershoot      Amplitude x 100   Negative Width Timing measurement of the first pulse in the waveform or gated region  The distance  time   between MidR ef  default 50   amplitude points of a negative pulse   P Peak to Peak Voltage measurement  The absolute difference between the maximum and minimum amplitude  Bo in the entire waveform or gated region   GAT Phase Timing measurement  The amount one waveform leads or lags another in time  Expressed in  degrees  where 360   comprise one waveform cycle   Period Timing measurement  Time it takes for the first complete signal cycle to happen in the waveform  or gated region  The reciprocal of frequency  Measured in seconds   Positive Duty Cycle Timing measurement of the first cycle in the waveform or gated region  The ratio of the positive  pulse width to the signal period expressed as a percentage   Pa _ PositiveWidth     PositiveDutyCycle   Period x 100   Positive Overshoot Voltage measurement over the entire waveform or gated region   PositiveOvershoot   V     HOA X 100   Amplitude  Positive Width Timing measurement of the first pulse in the waveform or gated region  The distance  time   between MidR ef  default 50   amplitude points of a positive pulse    Le Rise time Timing measurement  Time taken for the leading edge of the first pulse in the waveform or  gated region to rise from a Low Ref value  default   10   to a High Ref value  default   90   of  its final value       rp RMS Voltage 
148. d to as the Nyquist frequency or point   For example  a sample rate  of 20 Megasamples per second would yield an FFT with a range of 0 to 10 MHz   The sample rates available for acquiring data records vary over a range  TDS  oscilloscopes display the sample rate in the acquisition readout at the top of the  oscilloscope screen     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       Offset  Position  and Scale The following topics contain information to help you display your FFT properly     Adjust for a Non Clipped Display  To properly display your FFT waveform  scale  the source waveform so it is not clipped     m You should scale and position the source waveform so it is contained on  screen   Off screen waveforms may be clipped  resulting in errors in the FFT  waveform      Alternately  to get maximum vertical resolution  you can display source  waveforms with amplitudes up to two divisions greater than that of the  screen  If you do  turn on Pk Pk in the measurement menu and monitor the  source waveform for clipping     m Use vertical position and vertical offset to position your source waveform   As long as the source waveform is not clipped  its vertical position and  vertical offset will not affect your FFT waveform except at DC   DC  correction is discussed below      Adjust Offset and Position to Zero for DC Correction  Normally  the output of a  standard FFT computation yields a DC value that is twice as large as it shoul
149. d to linear  following the instructions on  page 3   41     To read about the two interpolation methods  see Interpolation on page 3 19  To  differentiate between the real and interpolated samples  set the display style to  Intensified Samples   See Select the Display Style on page 3 29      To quickly determine the zoom factor of a zoomed waveform  select it and check  the Zoom readout  It shows the selected waveform by number  along with the  horizontal and vertical expansion factors     The Zoom readout appears at the top of the display when zoom is on   See  Figure 3   23 on page 3 41   The TDS 600B and 700A models do not display the  Zoom readout when in dual window  preview  mode     To use Zoom  select a waveform  turn Zoom on  and magnify that waveform  using the vertical and horizontal scale knobs     1  Press any of waveform selection buttons CH 1 through CH 4 on the right  side of the display  Or press MORE and select a math or reference waveform  from the More menu     2  Press ZOOM     Press ZOOM     Mode  main   gt  ON  side   The ZOOM front panel button  should light up  Toggle Dual Zoom to OFF in the side menu     3  Adjust the vertical zoom factor for the selected waveform using the vertical  SCALE knob  Adjust the vertical position of the zoomed waveform using  the vertical POSITION knob     4  Adjust the horizontal zoom factor using the horizontal SCALE knob  Adjust  the horizontal position of the zoomed waveform using the horizontal  POSITION knob     Depe
150. d your derivative math waveform  you can use automated  measurements to make various parameter measurements  Do the following steps  to display automated measurements of the waveform     1  Be sure MORE is selected in the channel selection buttons and that the  differentiated math waveform is selected in the More main menu     2  Press MEASURE  gt  Select Measrmnt  main      3  Select up to four measurements in the side menu   See Figure 3   96      You can also use cursors to measure derivative waveforms  Use the same  procedure as is found under To Take Cursor Measurements on page 3 173  When  using that procedure  note that the amplitude measurements on a derivative  waveform will be in volts per second rather than in volt seconds as is indicated  for the integral waveform measured in the procedure     TDS 500B  TDS 600B   amp  TDS 700A User Manual    3 169    Using Features for Advanced Applications       Offset  Position  and Scale    3 170    Tek Run  5 00MS s Average       Select  Measurement              Ni pk 3  478KV s dla          1 Amplitude          Positive    Overshoot          he oep panen e ear a ee a ee hee Shay a ee es ol  ONegathe  L o mana  n o Overshoot    Chi 200mVvaQ M 10 045 Chi f 52mY TS       i o          DEUME _100kv s   _ 10 0p 7 O i i   Remove   Gating   LOW  Reference  asrmnt OFF Histagram Levels          LOIDID O lO      Snapshot                                                     I  19  1 1 a        Figure 3 96  Peak Peak Amplitude Measurement
151. de    Delayed Runs After Main                                                 Deskew  Channel P robe 0 seconds   Display clock No Change   Display color   collision contrast TDS 624B  684B   amp  TDS 700A  Off   Display color   map math colors TDS 624B  684B   amp  TDS 700A   Color    Math      Display color   map reference colors TDS 624B  684B   amp  TDS 700A   Color    Ref   Display color   palette TDS 624B  684B   amp  TDS 700A  Normal   Display color   palette colors TDS 624B  684B   amp  TDS 700A   The colors of each palette are reset to  factory hue  saturation  and lightness   HLS  values   Display color   persistence palette TDS 624B  684B   amp  TDS 700A   Temperature   Display format YT   Display graticule type Full   Display instavu persistence TDS 500B  amp  TDS 700A  Varpersist   Display instavu style TDS 500B  amp  TDS 700A  Vectors   Display instavu varpersist TDS 500B  amp  TDS 700A  500 e 3   Display intensity   contrast TDS 620B  amp  680B  175    Display intensity   overall TDS 620B  amp  680B  85    Display intensity   text TDS 624B  684B   amp  TDS 700A  100   TDS 500B  TDS 620B  amp  680B  60    Display intensity   waveform TDS 624B  684B   amp  TDS 700A  100   TDS 500B  TDS 620B  amp  680B  75    Display interpolation filter Sin x  x   Display mode Normal   TDS 500B  amp  TDS 700A  InstaVu off    Display style Vectors   Display trigger bar style Short   Display trigger    T    On   Display variable persistence 500 ms   Edge trigger coupling DC   Edge t
152. deletes directories recursively  That means it deletes both the  directories and all their contents     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups          Tek Run  1 00GS 5 Sample                Use  amp  to choose Ee i E  Press SELECT to change G F  File  Directory  fdo   Free  690KB Utilities                               0     WEMS 94 03 17 08 15 58  TEK  TEK00003  WFM 94 03 08 12 40 46  SETHLD1 SET    94 03 17 06 53 56  TEK00001 SET z 94 03 17 06 56 14  TEK00002 5ET A 94 03 17 06 56 32  TEK00000  WFM   94 03 17 06 57 18  TEK00001  WFM z 94 03 17 06 57 34    Delete          Rename                       more     1of2    eca ave eca File  Saved Current Factory Utilities  Setup Setup Setup    DIGIC IO  O O O    CJO C C O                                                               Figure 3 68  File Utilities    To Rename To rename a file or directory  turn the general purpose knob until it scrolls the  cursor over the name of the file or directory to delete  For example  to rename  the target file whose default name is TEK      set the cursor over its name   Then  press the side menu Rename button   See Figure 3 69      The labeling menu should appear  Turn the general purpose knob or use the  main menu arrow keys to select each letter  Press Enter Char from the main  menu to enter each letter  When you have entered the name  press the side menu  OK Accept item     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 123    Saving Waveforms an
153. displays waveform record points as dots     Intensified Samples style also displays waveform record points as dots  However   the points actually sampled are displayed in the color labeled    Zone    in the  Display Colors menus     In addition to choosing Intensified Samples in the side menu  the oscilloscope  must be interpolating  equivalent time must be off for TDS 500B and TDS 700A  models  or Zoom must be on with its horizontal expansion greater that 1X  See  Interpolation on page 3 19  see Zooming on Waveforms on page 3 39     Variable Persistence style accumulates the record points on screen and displays  them only for a specific time interval  In that mode  the display behaves like that  of an analog oscilloscope  You enter the time for that option with the keypad or  the general purpose knob  Record points are also displayed with colors that vary  depending on the persistence of the point  See Choose a Palette on page 3 35     Infinite Persistence style accumulates the record points until you change some  control  such as scale factor  causing the display to be erased        NOTE  TDS 500B and TDS 700A models only  See Using InstaVu   Acquisition  Mode  on page 3 45  to see how Style setup differs for InstaVu mode        TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms          Tek Run  100kS s Hi Res       Display Style    7  m  vectors    Dots                                  Intensified  Samples          Infinite     Persistenc
154. ds to Check and Eliminate  To quickly check for aliasing  slowly increase  the horizontal scale  time per division setting   If the shape of the displayed  waveform changes drastically or becomes stable at a faster time base setting   your waveform was probably aliased     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       To avoid aliasing  be sure to sample the input signal at a rate more than twice as  fast as the highest frequency component  For example  a signal with frequency  components of 500 MHz would need to be sampled at a rate faster than   1 Gigasamples second to represent it accurately and to avoid aliasing  The  following tips may help you eliminate aliasing on a signal     m Try adjusting the horizontal scale   m Try pressing the AUTOSET button     m Try switching the acquisition mode  in the acquisition menu  to Envelope or  Peak Detect  Envelope searches for samples with the highest and lowest  values over multiple acquisitions  Peak Detect mode does the same but in a  single acquisition  Either can detect faster signal components over time     m Try pressing the InstaVu acquisition button  TDS 500B and TDS 700A  models only   InstaVu mode results in waveform displays similar to those  obtained using an analog oscilloscope  due to its fast waveform capture rate     Customizing the Display    The TDS Oscilloscope can display waveform records and other display elements  in different ways  This section describes how to ad
155. e          20 0mv C M 500ps Chif 26 4mv parables    Style A Readout Filter Graticule  Settings f wan  Intensity 5 i   lt Display gt  Vectors Options   Sin x  x Full    DIPIC IO  C    O               Ollo                                                       Figure 3 17  Display Menu     Style    Adjust Intensity Intensity lets you set text graticule and waveform intensity  brightness  levels   To set the intensity     Press DISPLAY     Intensity  main   gt  Text Grat or Waveform  side   Enter  the intensity percentage values with the keypad or the general purpose knob     All intensity adjustments operate over a range from 20   close to fully off  to  100   fully bright      Set Display Readout Readout options control whether the trigger indicator  trigger level bar  and  Options current date and time appear on the display  The options also control what style  trigger level bar  long or short  is displayed     1  TDS 600B  Press DISPLAY     Readout Options  main      TDS 500B and TDS 700A Models  Press DISPLAY     Mode  main   gt   Normal  pop up   gt  Format RO  main      2  Toggle Display    T      Trigger Point  side  to select whether or not to  display    T    indicating the trigger point  You can select ON or OFF   The  trigger point indicates the position of the trigger in the waveform record      3  Press Trigger Bar Style  side  to select either the short or the long trigger  bar or to turn the trigger bar off   See Figure 3   18  Note that both styles are  shown for
156. e     EndCycle   MCross3            MCross1   StartCycle     MCross3  MCross2  EndCycle              Figure B 1  MCross Calculations    Waveform  lt 0 0     RecordLength 1 0 gt       holds the acquired data     TPOS     is the location of the sample just before the trigger point  the time  reference zero sample   In other terms  it contains the domain reference location   This location is where time   0     TSOFF     is the offset between TPOS and the actual trigger point  In other  words  it is the trigger sample offset  Values range between 0 0 and 1 0 samples   This value is determined by the instrument when it receives a trigger  The actual    zero reference  trigger  location in the measurement record is at   TPOS TSOFF      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix B  Algorithms       Measurement Algorithms    Amplitude    Area       Cycle Area       Burst Width          UU          The automated measurements are defined and calculated as follows     Amplitude   High     Low    The arithmetic area for one waveform  Remember that one waveform is not  necessarily equal to one cycle  For cyclical data you may prefer to use the cycle  area rather than the arithmetic area     if Start   End then return the  interpolated  value at Start     Otherwise     Area  p     Waveform t dt    Start    For details of the integration algorithm  see page B 12     Amplitude  voltage  measurement  The area over one waveform cycle  For data  not cyclical  you might prefer to
157. e  press     more   side  repeatedly until the Frequency item appears  Then press  Frequency  side      TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 19    Tutorial       Observe that the frequency measurement appears within the right side of the  graticule area  The measurement readout includes the notation Ch1  meaning  that the measurement is taken on the channel   waveform   To take a  measurement on another channel  select that channel  and then select the  measurement         Tek Run  100kKS 5 Sample       Select  Measurement                                         Frequency                Positi  widt                      Negative  width          lot     see Tira tevr ee ee  i Remove Gating t Reference    Measrmnt Setup Snapshot  arms Vieasrmnat OFF Histogram Levels             OII oj        200mv      F M 3004s chi 7  248mv    more                      Figure 2 14  Measure Main Menu and Select Measurement Side Menu                7  Press Positive Width  side      more   side   gt  Rise Time  side   gt   of Positive Duty Cycle  side                     All four measurements are displayed  Right now  they cover a part of the  graticule area  including the displayed waveforms     8  To move the measurement readouts outside the graticule area  press CLEAR  MENU   See Figure 2 15      Remove Measurement Use the Measure menu to remove waveforms you no longer want  To remove a  Readouts measurement individually  you can also remove them  as a group   do the  following st
158. e Figure 3   27      To turn InstaVu off  press InstaVu  Alternatively  press DISPLAY  gt   Mode  main   and then press Mode again to toggle to Normal mode     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 45    Acquiring and Displaying Waveforms       Normal DSO Mode    Ist Acquired  Waveform  Record          Next Acquired  Waveform  Record                      Next Acquired  Waveform  Record                                ai yN     Dead Time Dead Time Dead Time   Waveform Waveform Waveform   Memory Memory Memory   Display Updated Display Updated Display   InstaVu Mode   Ist Set of Acquired Next Set of Acquired Next Set of Acquired   Waveform Records Waveform Records Waveform Records  A AC A  1     0  N                   Waveform  Memory       BitMap       Variable Persistence  Display    Figure 3 26  Normal DSO Acquisition and Display Mode Versus InstaVu Mode    3   46                   Waveform  Memory  Bit Map       Updated Variable  Persistence Display                      Waveform  Memory  BitMap       Updated Variable  Persistence Display    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms          Tek Run  spose Sample BEAY Pergist  nce  10 00 5          Instavu  Style    JIL  Vectors    Dots                                                                   Infinite   4 Persistence    Siua gg  lt   M Toons Chis zgony persistance  l           10 00       re a mensi  O   Sou  I 11  1  1  1  1a  1                    Oooo            
159. e does not detect a pulse edge  when it expected to     Figure 3   43 shows the pulse trigger readout  Table 3 7  on page 3 74  describes  the choices for pulse triggers     Trigger Class  Runt      Set to ECL                    Glitch   Runt i  fay il Width  ool s  ons    Runt chi                  Slew Rate  Timeout             Tri ger ode  when hresholds  amp   Occurs Holdoff    MPIC ICG OIC              Polarity  Positive         Source  chi        Class   lt Runt gt              Tape   lt Pulse gt                                                     Figure 3 43  Pulse Trigger Readouts    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 73    Triggering on Waveforms       3 74    Table 3 7     Name       h                   Pulse Trigger Definitions    Glitch positive  Glitch negative  Glitch either    Runt positive    Definition    Triggering occurs if the oscilloscope detects positive spike  widths less than the specified glitch time     Triggering occurs if the oscilloscope detects negative spike  widths less than the specified glitch time     Triggering occurs if the oscilloscope detects positive or  egative widths less than the specified glitch time          Triggering occurs if the oscilloscope detects a positive pulse  that crosses one threshold going positive but fails to cross a  econd threshold before recrossing the first going negative     wn       Runt negative    Triggering occurs if the oscilloscope detects a negative  going pulse that crosses one threshold go
160. e greatly increases the probability  that runts  glitches  and other short term changes will accumulate in  waveform memory  The oscilloscope then displays the waveform at the  normal display rate using variable or infinite persistence     Interleaving  TDS 500B and TDS 700A Models Only  A method by which these  oscilloscopes attain higher digitizing speeds  The oscilloscope applies the  digitizing resources of unused channels  that is  channels that are turned off   to sample those that are in use  turned on   Table 3   2 on page 3 21 lists  acquisition rates vs  number of channels that are on     Knob  A rotary control     Logic state trigger  The oscilloscope checks for defined combinatorial logic conditions on  channels 1  2  and 3 on a transition of channel 4 that meets the set slope and  threshold conditions  If the conditions of channels 1  2  and 3 are met then  the oscilloscope triggers     Logic pattern trigger  The oscilloscope triggers depending on the combinatorial logic condition of  channels 1  2  3  and 4  Allowable conditions are AND  OR  NAND  and NOR     Low  The value used as 0  in automated measurements  whenever high ref  mid  ref  and low ref values are needed as in fall time and rise time measure   ments   May be calculated using either the min max or the histogram  method  With the min max method  most useful for general waveforms   it is  the minimum value found  With the histogram method  most useful for  pulses   it refers to the most common value 
161. e interval  Glossary   9  Sample Rate  Maximum  3 21  Option 1G  A 1  Sample  Acquire menu  3 25  Sampling  3 18  Glossary   10  Sampling and acquisition mode  3 25    Index 12    Sampling and digitizing  3   17  Saturation  Color menu  3   36  Save  Setups  3 115  Save Current Setup  Save Recall Setup menu  3 115  Save Format  Save Recall Waveform menu  3   119  Save Waveform  Save Recall Waveform menu  3   118  Save Recall SETUP button  2   10  3   8  3 115  3 122  Save Recall Setup menu  3 115  factory status  3 116  File Utilities  3 117  Recall Factory Setup  3 117  Recall Saved Setup  3 116  Save Current Setup  3 115  user status  3 116  Save Recall WAVEFORM button  3 118  3 122  Save Recall Waveform menu  3 118  active status  3 118  Autosave  3 121  Delete Refs  3 120  empty status  3 118  File Utilities  3 121  Ref1  Ref2  Ref3  Ref4  File  3 120  Save Format  3 119  Save Waveform  3 118  Saving  Waveforms  3 118  Saving and recalling setups  2   24  3 115  Saving and recalling waveforms  3 118  Saving Waveforms and Setups  3 115  Scale  vertical  3 159  3 170  3 175  seconds  Cursor menu  3   105  Security bracket  2 5  SELECT button  3 104  Glossary   10  Select Measrmnt  Measure menu  3   94  3   98  Selected waveform  Glossary   10  Self test  1 6  Serial number  2   5  Service Assurance  A   7  Set 1st Source to  More menu  3   149  Set 2nd Source to  More menu  3 149  Set Function to  More menu  3 149  SET LEVEL TO 50  button  3 56  Set Levels in   units  Measu
162. e menu  Figure 3 84   The More menu  allows you to display  define  and manipulate three math waveforms  the  following steps explain how to create a math waveform based on a single source  waveform     1  Press MORE  gt  Math   Math2  or Math3  main   gt  Change Math  waveform definition  side   gt  Single Wfm Math  main      2  To define the source waveform  press Set Single Source to  side  repeatedly  to cycle it to the desired channel or reference waveform        Tek Run  100kKS 5 Sample       Math                   Change Math  waveform    definition    No Extended  S  Processing                            Average  2                   chi 100mV       M 500Ms Chi            LOJOJO C C      Math2 Math3  Chitche   FFT ChI    inych                                                      DID ClO            Figure 3 84  More Menu    3 148 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       3  Press Set Function to  side  repeatedly to cycle it to inv  invert   intg  or  diff  Waveform integration  intg  is described on page 3 171  and waveform  differentiation  diff  is described on page 3 167     4  To create the math waveform  press OK Create Math Wfm  side      To Use Dual Wim Math To create a math waveform that requires two waveform sources  do the following  steps     1  Press MORE  gt  Math1  Math2  or Math3  main   gt  Change Math  waveform definition  side   gt  Dual Wfm Math  main      2  To define the first source waveform  
163. e second waveform  specified in the Delay or Phase Measurements  The default is 50      Take a Delay The delay measurement lets you measure from an edge on the selected waveform  Measurement to an edge on another waveform  To take a delay measurement  do the following  steps     1  Press MEASURE  gt  Select Measrmnt  main   gt  Delay  side   gt  Delay  To  main   gt  Measure Delay to     2  Press Measure Delay to  side  repeatedly to choose the delay to waveform   The choices are Ch1  Ch2  Ch3  Ch4  Math1  Math2  Math3 Ref1  Ref2   Ref3  and Ref4     The steps just performed select the waveform you want to measure fo  note  that the waveform you are measuring delay from is the selected waveform    See Figure 3 56         Tek Run  100kS 5 sure Delay to  Chl            Delay from  Selected Wim             Measure    mat to                EPN T ad EAN E a    ENA En nT SEET          verify or change the waveform to which Delay  will be measured  the Search Direction  and the  Edges at which to measure  when you are  satisfied with the selections  press Create  Measurement     E Press CLEAR MENU to return to the Measure  menu without creating the Delay measurement  agmv OK    f Create  Measurement    Delay To Edges  chi  gt F  gt F    a a    O                               Oooo o                                                        Figure 3 56  Measure Delay Menu     Delay To    3 98 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       3  Press MEASURE  gt
164. e time base  main or delayed  being referred to   See  Figure 3 7  Also see Figure 3   5 on page 3 11   Since all live waveforms use the  same time base  the oscilloscope only displays one time base and time division  setting for all the active channels     ie a    TEk Stopped               Record View Readout    32253 Acquisitions Delay TMN 16 5018  amp                   Time Base             Main Only             Intensified          ae  Delayed Only    Delayed Runs  16 50n5  After Main                   Time Base Readout    To Change Horizontal  Scale and Position                  we og ae    _ Delayed  chi f   200m     Triggerable   see Delayed    Trig Menu           Trigger  osition  50                                               Olola       Figure 3 7  Record View and Time Base Readouts    The TDS Oscilloscope provides control of horizontal position and scale using  either the horizontal front panel knobs     By changing the horizontal scale  you can move the waveform right or left to see  different portions of the waveform  That is particularly useful when you are  using larger record sizes and cannot view the entire waveform on one screen     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       To change the horizontal scale and position     1  Turn the horizontal POSITION and horizontal SCALE knobs   See  Figure 3 8      2  If you want the POSITION knob to move faster  press the SHIFT button   When the light above the shift but
165. eadout  When  Trigger Status edge triggers are selected  the trigger readout displays the trigger source  as well  as the trigger slope and level   See Figure 3 36      Main Trigger  Slope  Rising Edge    Main Trigger  Source  Ch 1            Main Time Base Time Div    Main Trigger  Level    Main Time Base       m 2o mva CF M 500us Chif                  coupling  DE        TYPE    Source   lt Edge gt  Chi                                          IDC OC  O   CS           Figure 3 36  Edge Trigger Readouts    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 59    Triggering on Waveforms       To Select Edge Triggering    To Select a Source    To Specify Coupling    DC    ACM    3 60    Use the edge trigger menu to select edge triggering and to perform the proce   dures for source  coupling  slope  trigger level  mode  and holdoff that follow     To bring up the Edge Trigger menu  press TRIGGER MENU  gt   Type  main   gt  Edge  pop up    See Figure 3 37         Tek Run  100kS s Hi Res       Edge Source                                           Logic lomv  O M 500qms Chi 26 8mv   more   Pulse     A         i 1of2  yden       P    See coupling   Slope Level ode  Type A  amp     Eiger   cal BC f   26 3mV   ot ioe    CICIC OIC  O C                  Oloo                                                          Figure 3 37  Main Trigger Menu     Edge Type    To select which source you want for the trigger     Press TRIGGER MENU  gt  Type  main   gt  Edge  pop up   gt   Source  main   
166. easurements and does not update those measure   ments unless it is performed again        Tek Run  100kKS 5 Sample    Snapshot   aa i fa us Again  g snapshot on Si  Period    1 000ms Freq  1 00008kH2   Width   500ps  Width   500ys  Brstwd  4 500ms  Rise   6MS Fall   BUS   Duty   50 0   Duty   50 0    Over   11 3   Over 38 1   High   504m   Low  SmvV  Max   560mvV Min   32mvV  Ampl  496mV Pk Pk  592m    i Mean  255 6m   cycMean  262 4m    RMS  359 6m   CycRMS   362 4m   ny  R Area   1 27548mV5 cycArea   262 20uUVSs    ICICI IOC  S  e                                                                                                                 Figure 2 17  Snapshot of Channel 1    2  Press Again  side  to do another snapshot and update the snapshot measure   ments     TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 23    Tutorial       3  Press Remove Measrmnt  main  to remove the snapshot display   You can  also press CLEAR MENU  but a new snapshot will be executed the next  time you display the Measure menu      Example 4  Saving Setups    The TDS Oscilloscope can save its controls settings and recall them later to  quickly re establish a setup  It provides ten storage locations to store up to ten  setups  It also provides a file system  so that you can also save setups to a floppy  disk  Do the following procedures to learn how to save  and then recall  a setup        NOTE  Besides being able to save several complete setups  the oscilloscope  remembers all the parameter se
167. easurements using that channel and probe     Run a Probe Cal anytime you wish to ensure that the measurements you make  are made with the most accuracy possible  You should also run a Probe Cal if  you have changed to a different probe since the last Probe Cal was performed     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       Some Probes Cannot Be Compensated  Some types of probes can be gain  compensated  some can be offset compensated  and some can be compensated  for both  Some probes cannot be compensated at all  Note the following  restrictions     m The oscilloscope cannot compensate probes that have an attenuation factor of  greater than 20X  If you attempt to compensate such a probe you will get an  error message     m The oscilloscope cannot compensate probes that have gain and or offset  errors that are too great   gt  2  gain and or  gt 50 mV offset   If these errors  are within specified limits for your probe  you may want to use another  probe  If they are not within specification  have your probe checked by  service personnel        NOTE  Probe Cal is not recommended with the P6139A passive probe  This  probe typically has little gain and offset error  and therefore  the improvement in  performance after a Probe Cal is not worth the time needed to do the Probe Cal   Probe Cal makes significant performance improvements when performed with  active probes or older passive probes        To run a probe cal  follow the instructions regarding
168. ececeeeee  More Menu  File Utilities  File System     Labeling Menu                  05   Utility Menu     System I O    Hardcopy Formats    Date and Time Display    Connecting the Oscilloscope Directly to the    Connecting the Oscilloscope and Hardcopy Device Via    Typical GPIB Network Configuration  Stacking GPIB Connectors  Connecting the Oscilloscope to a Controller  LOE 1108  20 5   2 11 Gene eee ener amea ane ire  Status Menu     System    Banner Display    Initial Help Screen             ccc cee cece eens  Comparing a Waveform to a Limit Template           Acquire Menu     Create Limit Test Template           More Menu  Dual Waveform Math Main and Side Menus  System Response to an Impulse                 000     Define FFT Waveform Menu             cecceccece    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Table of Contents       vi    Figure 3 88  FFT Math Waveform in Math                   0   3 154  Figure 3 89  Cursor Measurement of an FFT Waveform           3 156  Figure 3 90  Waveform Record vs  FFT Time Domain Record       3 157  Figure 3 91  FFT Time Domain Record vs  FFT Frequency Domain   R  cord pierin ehee wheres dncce E TE gave  e erse ip EERE EN Bree eee 3 158  Figure 3 92  How Aliased Frequencies Appear in an FFT           3 162  Figure 3 93  Windowing the FFT Time Domain Record            3 165  Figure 3 94  FFT Windows and Bandpass Characteristics          3 167  Figure 3 95  Derivative Math Waveform               ceeeeeees 3 169  Figure 3 96  Pe
169. ect the channel that is to contain the data signal     1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  Setup Hold  pop up   gt  Data Source  main      2  Press any one of Ch1  Ch2  Ch3  or Ch4  side   Do not select the same  channel for both the data and clock sources     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       Define the Clock Source and Edge  To select the channel that is to contain the  clock signal and the edge to use to clock     1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  Setup Hold  pop up   gt  Clock Source  main   gt  Ch1   Ch2  Ch3  or Ch4  side      2  Press any one of Ch1  Ch2  Ch3  or Ch4  side   Do not select the same  channel that you selected for the clock source     3  Press Clock Edge  side  to toggle between the rising and falling edges     Set the Data and Clock Levels  To set the transition levels that the clock and data  must cross to be recognized by the oscilloscope     1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  Setup Hold  pop up   gt  Levels  main   gt  Clock Level or  Data Level  side      2  Turn the general purpose knob or use the keypad to set values for the clock  level and for the data level you select     If you prefer  you can set both clock levels to a value appropriate to either of two  logic families  To do so     3  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  mai
170. ed in delay time calculations   Typically set to 50   You can set it from 0  to 100  or to a voltage level     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix B  Algorithms       Other Variables The oscilloscope also measures several values itself that it uses to help calculate  measurements     RecordLength     is the number of data points in the time base  You set it with  the Horizontal menu Record Length item     Start     is the location of the start of the measurement zone  X value   It is 0 0  samples unless you are making a gated measurement  When you use gated  measurements  it is the location of the left vertical cursor     End     is the location of the end of the measurement zone  X value   It is   RecordLength     1 0  samples unless you are making a gated measurement   When you use gated measurements  it is the location of the right vertical cursor     Hysteresis     The hysteresis band is 10  of the waveform amplitude  It is used  in MCross1  MCross2  and MCross3 calculations     For example  once a crossing has been measured in a negative direction  the  waveform data must fall below 10  of the amplitude from the MidRef point  before the measurement system is armed and ready for a positive crossing   Similarly  after a positive MidRef crossing  waveform data must go above 10   of the amplitude before a negative crossing can be measured  Hysteresis is useful  when you are measuring noisy signals  because it allows the oscilloscope to  ignore minor fl
171. eform record  All three records are described below     Waveform Record     the complete waveform record acquired from an input  channel and displayed from the same channel or a reference memory  The length  of this time domain record is user specified from the Horizontal menu  The  waveform record is not a DSP Math waveform     FFT Time Domain Record     that part of the waveform record input to the FFT   This time domain record waveform becomes the FFT math waveform after it   s  transformed  Its record length depends on the length of the waveform record defined  above     FFT Frequency Domain Record     the FFT math waveform after digital signal  processing converts data from the FFT time domain record into a frequency  domain record     Figure 3 90 compares the waveform record to the FFT time domain record  Note  the following relationships     m For waveform records  lt 10 K points in length  the FFT uses all of the  waveform record as input     m For waveform records  gt 10 K points  the first 10 K points of the waveform  record becomes the FFT time domain record     m Each FFT time domain record starts at the beginning of the acquired  waveform record     m The zero phase reference point for a phase FFT math waveform is in the middle  of the FFT time domain record regardless of the waveform record length     FFT Time Domain Record                 gt   Waveform Record    Waveform Record  lt  10 K       Zero Phase  Reference     lt           FFT Time Domain Record   10k 
172. elect the type of Math function to define from Hanning    the bottom row of bezel buttons  Define the  new function using the side buttons  ate  When you are satisfied with the selections  press         amplitudes  lt   Create Math wfm  3dB    Press CLEAR MENY to return to the More menu  without creating a Math waveform  OK    Create  Math wim                      LOIOIDIC O                                                         G 1 1 1        Figure 3 87  Define FFT Waveform Menu    7  Press Set FFT Vert Scale to  side  repeatedly to choose from the following  vertical scale types     dBV RMS     Magnitude is displayed using log scale  expressed in dB  relative to 1 Vpms where 0 dB  1 Vrms     Linear RMS     Magnitude is displayed using voltage as the scale     Phase  deg      Phase is displayed using degrees as the scale  where degrees  wrap from    180   to  180       Phase  rad      Phase is displayed using radians as the scale  where radians  wrap from  T to  7     The topic Considerations for Phase Displays  on page 3 162  provides in  depth information on setup for phase displays     8  Press Set FFT Window to  side  repeatedly to choose from the following  window types     Rectangular     Best type of window for resolving frequencies that are very  close to the same value but worst for accurately measuring the amplitude of  those frequencies  Best type for measuring the frequency spectrum of  nonrepetitive signals and measuring frequency components near DC     TDS 50
173. eneral Purpose Interface Bus   An interconnection bus and protocol that allows you to connect multiple  instruments in a network under the control of a controller  Also known as  IEEE 488 bus  It transfers data with eight parallel data lines  five control  lines  and three handshake lines     Graticule  A grid on the display screen that creates the horizontal and vertical axes  You  can use it to visually measure waveform parameters     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Glossary       GND Ground  GND  coupling  Coupling option that disconnects the input signal from the vertical system     Hardcopy  An electronic copy of the display in a format useable by a printer or plotter                 Hi Res acquisition mode  TDS 500B and TDS 700A Models Only  An acquisition mode in which the  oscilloscope averages all samples taken during an acquisition interval to  create a record point  That average results in a higher resolution  lower band   width waveform  That mode only works with real time  non interpolated  sampling     TW High  The value used as 100  in automated measurements  whenever high ref   mid ref  and low ref values are needed as in fall time and rise time measure   ments   May be calculated using either the min max or the histogram  method  With the min max method  most useful for general waveforms   it is  the maximum value found  With the histogram method  most useful for  pulses   it refers to the most common value found above the mid point  See  Ap
174. ensation  Adjustment that improves low frequency response of a probe     Pulse trigger  A trigger mode in which triggering occurs if the oscilloscope finds a pulse   of the specified polarity  with a width between  or optionally outside  the  user specified lower and upper time limits     Real time sampling  A sampling mode where the oscilloscope samples fast enough to completely  fill a waveform record from a single trigger event  Use real time sampling to  capture single shot or transient events     Record length  The specified number of samples in a waveform     Reference memory  Memory in a oscilloscope used to store waveforms or settings  You can use  that waveform data later for processing  The oscilloscope saves the data even  when the oscilloscope is turned off or unplugged     Rise time  The time it takes for a leading edge of a pulse to rise from a LowRef value   typically 10   to a HighRef value  typically 90   of its amplitude     RMS  Amplitude  voltage  measurement of the true Root Mean Square voltage     Runt trigger  A mode in which the oscilloscope triggers on a runt  A runt is a pulse that  crosses one threshold but fails to cross a second threshold before recrossing  the first  The crossings detected can be positive  negative  or either     Sample acquisition mode  The oscilloscope creates a record point by saving the first sample during  each acquisition interval  That is the default mode of the acquisition     Sample interval  The time interval between su
175. ep     1  Press MEASURE     Remove Measrmnt  main   gt  Measurement 1   Measurement 2  and Measurement 4  side  to remove those measurements   Leave the rise time measurement displayed     2 20 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Tutorial          Tek Run  100kKS 5 Sample       Ch1 Freq  999 44 HZ       chi  width    501s          chi Rise  7usS    Low     resolution          chi  Duty    50 0                              i om a 0ps Chi F AEM                                              wa          Press to Remove Menus From Screen  Figure 2 15  Four Simultaneous Measurement Readouts    Change the Measurement By default  the measurement system will use the 10  and 90  levels of the  Reference Levels waveform for taking the rise time measurement  You can change these values to  other percentages or change them to absolute voltage levels     To examine the current values  press Reference Levels  main   gt  High Ref       side      The General Purpose Knob  The general purpose knob  the large knob  is now set  to adjust the high reference level  Figure 2 16      There are several important things to observe on the screen       m The knob icon appears at the top of the screen  The knob icon indicates that  the general purpose knob has just been set to adjust a parameter     m The upper right corner of the screen shows the readout High Ref  90      m The High Ref side menu item is highlighted  and a box appears around the  90  readout in the High Ref menu item  The b
176. er of acquisitions to store in the acquisition  memory of the channel  The oscilloscope will set the record length to a value  greater than or equal to the product of the frame count and the frame length   If the product exceeds the maximum available record length  the oscilloscope  will reduce the frame length or frame count in size such that the product will  fit the record length     Press Horiz Pos  main   then Frame  side   and use the general purpose  knob to enter the number of a specific frame to view  After you press Enter   that frame should appear on the display     If you shift the waveform right or left with the front panel HORIZONTAL  POSITION knob  the window next to the side menu Frame button will indicate  the frame number of the waveform at the center of the screen     FastFrame Operating Characteristics  Consider the following operating character   istics when using FastFrame     Envelope  Average  and Hi Res form the envelope or average following the  last frame of the concatenated record  For example  if Average or Hi Res  acquisition modes are selected and the frame count is 10  segments 1 through  10 will show Sample or Hi Res frames  and frame 11 will show the average  of frames 1 through 10  If there is not room for one additional frame  the  envelope or average of the frames replaces the display of the last acquired  frame  Average and envelope counts have no affect in FastFrame     You can press RUN STOP to terminate a FastFrame sequence  If any fra
177. erform the following steps     1     Fast Fourier Transforms    Press MORE     Math   Math2  or Math3  main  to select the math  waveform to be averaged     Press Average  side  and enter a value with the general purpose knob or the  keypad  Any math operations you select for the waveform are performed on  an average of multiple acquisitions     To turn off math averaging for the selected math waveform  press No  Extended Processing  side   Any math operations you select for the  waveform are performed on only one acquisition     The Advanced DSP Math capabilities of the TDS Oscilloscope include taking  the Fast Fourier Transform  FFT  of a waveform  This section describes FFTs  and how to set up the oscilloscope to display and measure FFTs     The FFT allows you to transform a waveform from a display of its amplitude  against time to one that plots the amplitudes of the various discrete frequencies  the waveform contains  Further  you can also display the phase shifts of those  frequencies  Use FFT math waveforms in the following applications     3 150    Testing impulse response of filters and systems   Measuring harmonic content and distortion in systems  Characterizing the frequency content of DC power supplies  Analyzing vibration   Analyzing harmonics in 50 and 60 cycle lines    Identifying noise sources in digital logic circuits    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       The FFT computes and displays the frequenc
178. erification and Specifica   tions manual for typical minimum and maximum values     3  To change to the factory default holdoff setting for the current horizontal  scale setting  press Default Holdoff  side      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 61    Triggering on Waveforms       To Set Slope      a  An    To Set Level       NOTE  If you select Default Holdoff  the default holdoff time will vary with the  horizontal scale setting to maintain a good value for general purpose triggering  at that scale  However  if you select Holdoff  as opposed to Default Holdoff   the  time set in the Holdoff menu item is used at all horizontal scale settings        To select the slope that the edge trigger will occur on     1  Press the TRIGGER MENU     gt  Type  main   gt  Edge  pop up   gt   Slope  main      2  Select the rising or falling edge from the side menu     Press the TRIGGER MENU  gt  Type  main   gt  Edge  pop up   gt   Level  main   gt  Level  Set to TTL  Set to ECL  or Set to 50   side      Level lets you enter the trigger level using the general purpose knob or the  keypad     Set to TTL fixes the trigger level at  1 4 V   Set to ECL fixes the trigger level at  1 3 V        NOTE  When you set the volts div smaller than 200 mV  the oscilloscope reduces  the Set to TTL or Set to ECL trigger levels below standard TTL and ECL levels   This reduction occurs because the trigger level range is fixed at  12 divisions  from the center  At 100 mV  the next smaller settin
179. ermine under what conditions the oscilloscope normally  interpolates or uses equivalent time  see Selecting Repetitive Sampling on  page 3 25      If you select any of these modes before or while InstaVu is on  they will remain  selected in their respective menus  but the oscilloscope will ignore them  It will  put the modes into effect when you turn InstaVu off  If the oscilloscope setup is  not as you expect when you turn InstaVu off  this may be because the oscillo   scope quit ignoring these InstaVu incompatible modes     InstaVu mode disallows selection of Math waveforms  If you attempt to select a  math waveform from the MORE menu  the oscilloscope will display an error  message  Either switch InstaVu off and create the math waveform  or select a  channel waveform and continue using InstaVu mode     InstaVu displays using a persistence display style  see display menu   If you  select Intensified time base  the intensified zone is controlled by Horizontal  Scale and Delay time settings as when InstaVu is off  but the zone is masked by  the persistence display mode and cannot be seen  Turn off InstaVu to display the  intensified zone     Using FastFrame   TDS 500B and 700A Models Only     You can define and enable FastFrame  also called segmented memory   This  feature lets you capture multiple acquisitions in the acquisition memory of a  single channel  Figure 3   28 shows how FastFrame combines the desired  captured records into one larger record  For example  FastFrame
180. es To run file utilities  see the Managing the File System on page 3 121     To Find More Information See Example 4  Saving Setups  on page 2 24     See Appendix D  Factory Initialization Settings on page D 1     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 117    Saving Waveforms and Setups       Saving and Recalling Waveforms    To Save a Waveform    3 118    TDS Oscilloscope provides four internal reference memories in any of which you  can store a waveform  Waveforms thus stored are retained even when you turn  the oscilloscope off or unplug it  The oscilloscope also can save waveforms to  floppy disk  This section describes how to save  delete  and display reference  waveforms     The oscilloscope can display up to 11 waveforms at one time  That includes  waveforms from the four input channels  four reference waveforms  and three  math waveforms  You can save any combination of different size waveform  records     You will find saving waveforms useful when working with many waveforms and  channels  If you have more waveforms than you can display  you can save one of  the waveforms and then stop acquiring it  By doing so  you free an input channel  to display another waveform without losing the first one     To save a waveform  do the following steps     1  Select the channel that has the waveform you want to save        STOP  Before doing step 2 that follows  note that if you choose a reference  memory location labeled active  see Figure 3   66   you will overwrite t
181. ess SHIFT first to make  selection in the opposite direction     A pop up selection changes the other  main menu titles     TDS 500B  TDS 600B   amp  TDS 700A User Manual    2 8     E a   Tutorial    This section quickly makes you acquainted with some of the fundamental  operations required to use the TDS Oscilloscope to take measurements  Start this  tutorial by doing Setting Up for the Examples on this page     Setting Up for the Examples    Perform the following tasks to connect input signals to the TDS Oscilloscope  to  reset it  and to become acquainted with its display screen  Once completed  these  tasks ready the oscilloscope for use in the examples that follow     Connect the Input Signal Remove all probes and signal inputs from the input BNC connectors along the  lower right of the front panel  Then  using an appropriate probe  such as the  P6245   connect from the CH 1 connector of the oscilloscope to the PROBE  COMPENSATION connectors   See Figure 2 1                                                                     a DODO G oO f N        DOD OO Soy   am  oooO  0 CoD O    i  W olol  oi aes K    9000         r       S                                                                                        Figure 2 1  Connecting a Probe for the Examples  P6245 shown     TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 9    Tutorial          NOTE  See Appendix A  Options and Accessories for optional probes you can  order and use with this product        Reset the Osc
182. et                   Zoomy   C M 500s Chif  248mV             ololooo    mode IEC E Si Limit Test   Limit Test     amp reate    Limit Test  Sample PERE  Setup Sources   7  plate     I 101  1  1                                                       Figure 3 15  Acquire Menu     Stop After    m Press RUN STOP button only  side  to use the RUN STOP button to start  or stop acquiring  Pressing the RUN STOP button once will stop the  acquisitions  The upper left hand corner in the display will say    Stop    and  show the number of acquisitions  If you press the button again  the oscillo   scope will resume taking acquisitions       Press Single Acquisition Sequence  side   That selection lets you run a  single sequence of acquisitions by pressing the RUN STOP button  In  Sample  Peak Detect  or Hi Res mode  the oscilloscope will acquire a  waveform record with the first valid trigger event and stop   Hi Res is  available only on TDS 500B and TDS 700A models      In Envelope or Average mode  the oscilloscope will make the specified  number of acquisitions to complete the averaging or enveloping task     TDS 500B and TDS 700A models only  If the oscilloscope is in equivalent   time mode and you press Single Acquisition Sequence  side   it will  continue to recognize trigger events and acquire samples until the waveform  record is filled     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 27    Acquiring and Displaying Waveforms       Preventing Aliasing       NOTE  To quickly sele
183. et  and third  MCross3  in the  source  target  waveform are found     2  The period of the target waveform is calculated  see Period above      3  The first MidRefCrossing  MCross1Ref  in the reference waveform crossing  in the same direction  polarity  as that found MCross1 Target for the target  waveform is found     4  The phase is determined by the following     TDS 500B  TDS 600B   amp  TDS 700A User Manual B 9    Appendix B  Algorithms       Positive Duty Cycle                   Positive Overshoot       Positive Width                Rise Time    aa    B 10    _ MCrosslRef     MCross1Target x    Phase Period    360       If the target waveform leads the reference waveform  phase is positive  if it lags   negative     Phase is not available in the Snapshot display     Timing measurement  The ratio of the positive pulse width to the signal period   expressed as a percentage     PositiveWidth is defined in Positive Width  following     If Period   0 or undefined then return an error     PositiveWidth    1090     PositiveDutyCycle   Period    Amplitude  voltage  measurement   Max     High    Amplitude mee    PositiveOvershoot      Note that this value should never be negative     Timing measurement  The distance  time  between MidRef  default   50    amplitude points of a positive pulse     If MCross1Polarity         then   PositiveWidth    MCross2     MCross1   else    PositiveWidth    MCross3     MCross2     Timing measurement  Time taken for the leading edge of a pulse t
184. f 10  and 0 V  units   Measure mid ref 50  and 0 V  units   Measure mid2 ref 50  and 0 V  units   Pulse glitch filter state On  Accept glitch   Pulse glitch trigger polarity Positive    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix D  Factory Initialization Settings       Table D 1  Factory Initialization Defaults  Cont                  Control Changed by Factory Init to   Pulse glitch width 2 0 ns   Pulse runt high threshold 12V   Pulse runt low threshold 0 8 V   Pulse runt trigger polarity Positive   Pulse runt triggers when     Occurs   Pulse runt width 2 0 ns   Pulse slew rate delta time 2 0 ns   Pulse slew rate polarity Positive   Pulse slew rate thresholds Trig if faster than  Upper 1 80 V  Lower 800 mV          Pulse slew rate triggers when        Trig if faster than       Pulse timeout polarity  Pulse timeout time  Pulse trigger class  Pulse trigger level    TDS 600B  Either  TDS 600B  2 0 ns  Glitch  0 0 V       Pulse trigger source   Glitch  runt  width  and slew rate    Timeout   TDS 600B     Channel 1  Ch 1                    Pulse width lower limit 2 0 ns  Pulse width trigger polarity Positive  Pulse width trigger when Within limits  Pulse width upper limit 2 0 ns  Repetitive signal TDS 500B  amp  TDS 700A  On  RS 232 parameters No change  Saved setups No change  Saved waveforms No change  Stop after R S button  Vertical bandwidth  all channels  Full   Vertical coupling  all channels  DC   Vertical impedance  termination  1MQ    all channels    Verti
185. f equivalent time  sampling called random equivalent time sampling  which uses an internal  clock that runs asynchronously with respect to the input signal and the signal  trigger  The oscilloscope takes samples continuously  independent of the  trigger position  and displays them based on the time difference between the  sample and the trigger  Although the samples are taken sequentially in time   they are random with respect to the trigger     Fall time  A measurement of the time it takes for the trailing edge of a pulse to fall  from a HighRef value  typically 90   to a LowRef value  typically 10   of  its amplitude     Frequency  A timing measurement that is the reciprocal of the period  Measured in Hertz   Hz  where 1 Hz   1 cycle per second     Gated Measurements  A feature that lets you limit automated measurements to a specified portion  of the waveform  You define the area of interest using the vertical cursors     General purpose knob  The large front panel knob with an indentation  You can use it to change the  value of the assigned parameter     Glitch positive trigger  Triggering occurs if the oscilloscope detects positive spike widths less than  the specified glitch time     Glitch negative trigger  Triggering occurs if the oscilloscope detects negative spike widths less than  the specified glitch time     Glitch either trigger  Triggering occurs if the oscilloscope detects either positive or negative spike  widths less than the specified glitch time     GPIB  G
186. found below the mid point  See  Appendix B  Algorithms for details     Main menu  A group of related controls for a major oscilloscope function that the  oscilloscope displays across the bottom of the screen     Main menu buttons  Bezel buttons under the main menu display  They allow you to select items  in the main menu     Maximum  Amplitude  voltage  measurement of the maximum amplitude  Typically the  most positive peak voltage     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Glossary                                           7    7       Mean  Amplitude  voltage  measurement of the arithmetic mean over the entire  waveform     Minimum  Amplitude  voltage  measurement of the minimum amplitude  Typically the  most negative peak voltage     NAND  A logic  Boolean  function in which the output of the AND function is  complemented  true becomes false  and false becomes true   On the  oscilloscope  that is a trigger logic pattern and state function     Negative duty cycle  A timing measurement representing the ratio of the negative pulse width to  the signal period  expressed as a percentage     Negative overshoot measurement  Amplitude  voltage  measurement     Low     Min x 100     NegativeOvershoot   Atplinide    Negative width  A timing measurement of the distance  time  between two amplitude  points     falling edge MidRef  default 50   and rising edge MidRef  default  50       on a negative pulse     Normal trigger mode  A mode on which the oscilloscope does not acq
187. g after 200 mV  the trigger  range is  1 2 V  which is smaller than the typical TTL   1 4 V  or ECL     1 3 V   level        Set to 50  fixes the trigger level to approximately 50  of the peak to peak  value of the trigger source signal     Triggering Based on Logic    3 62    The TDS Oscilloscope can trigger on a logic or binary pattern and on the state of  a logic pattern at the time it is clocked  It can also trigger on data that violates  setup and hold times relative to a clock  This subsection describes how to use  these three classes of logic triggering  pattern  state  and setup hold     A pattern trigger occurs when the logic inputs to the logic function you select    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       cause the function to become TRUE  or at your option FALSE   When you use a  pattern trigger  you define     m The precondition for each logic input     logic high  low  or do not care  the  logic inputs are channels 1  2  3  and 4     m The Boolean logic function     select from AND  NAND  OR  and NOR          m The condition for triggering     whether the trigger occurs when the Boolean  function becomes TRUE  logic high  or FALSE  logic low   and whether the  TRUE condition is time qualified             A state trigger occurs when the logic inputs to the logic function cause the  function to be TRUE  or at your option FALSE  at the time the clock input  changes state  When you use a state trigger  you define     m The pre
188. g sampling or let it choose between real time or equivalent time sampling     Press SHIFT ACQUIRE MENU  gt  Repetitive Signal  main   gt   ON or OFF  side      m ON  Enable ET  uses both the real time and the equivalent time sampling  as appropriate  see Table 3   4        OFF  Real Time Only  limits the oscilloscope to real time sampling  If the  oscilloscope cannot accurately get enough samples for a complete waveform   it will interpolate        NOTE  The oscilloscope will use the interpolation method selected in the display  menu to fill in the missing record points     either linear or sin x  x interpolation   See Interpolation on page 3 19 for a discussion of interpolation        The sampling speeds and the number of channels you choose affect the mode the  oscilloscope uses to sample waveforms     m The oscilloscope always real time samples at slower time base settings   faster time settings force the oscilloscope to switch from real time sampling    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 25    Acquiring and Displaying Waveforms       to equivalent time sampling or interpolation  depending on whether ET is on    or off     m The oscilloscope extends the limit at which it must switch from real time  sampling by using the digitizers of channels that are turned off to sample the    channel or channels that are turned on     Check Table 3   4 below to determine the time base setting s  at which the switch  from real time sampling  RT  to equivalent time sampling 
189. gering and check the pulse width if you can t slew rate trigger as  expected                    Run  S00MS s Average Delta Time  250 0ns                                 iL  Slew Rate  Trigger When               Trig if     Faster Than                     Trig if  Slower Than                       Delta Time              Slew Rate              eee et EB OVA                            1 00 YQ loons  Slew chi    oan Seen eeeweewews E cer a _  hype Class mou    Polarity when  amp    lt Pulse gt   lt Slew gt  Faster Holdoff             LOOID CI aQ                                                       OM ClO ClO          Figure 3 46  Main Trigger Menu     Slew Rate Class    To understand what happens when you slew rate trigger  study Figure 3 46 as  you consider the following points     m The main menu shows the oscilloscope is set to trigger based on the slew  rate of a pulse input to the trigger source  Ch 1  It is set to monitor the  positive polarity pulse edges of the trigger source and to trigger on any edge  with a slew rate faster than the slew rate setting     m The Trigger When side menu displays the readout Slew Rate that indicates  the slew rate setting  The slew rate setting is not the slew rate of the pulse   instead  it is the slew rate against which the oscilloscope compares the slew    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       rate of pulse  see above   You set the slew rate setting indirectly by setting  the ratio of delta vol
190. gs are  listed in step 4 on page 3 40        Set Interpolation To change the interpolation method used  press DISPLAY  gt  Settings  main   gt   Display  pop up   gt  Filter  main   gt  Sin x  x Interpolation or Linear  Interpolation  side      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 41    Acquiring and Displaying Waveforms       Reset Zoom    Using Dual Window Mode    3 42    To reset all zoom factors to their defaults  do the following step     Press ZOOM  gt  Reset  main   gt  Reset Live Factors or Reset All Factors   side   Reset Live Factors resets only for live waveforms  as opposed to reference  waveforms  Reset All Factors resets for all waveforms     The oscilloscope can display and control a waveform that is both zoomed and  unzoomed  magnified and unmagnified   To do so  it creates two 1 2 height  graticules  or windows  and displays the magnified waveform in the upper  and  the unmagnified waveform in the lower graticule  To use Dual Window Zoom   also called zoom preview mode   do the following steps     1     Press Zoom     Mode  main      Preview  side   Note that the oscilloscope  displays the box enclosed area on the waveform as magnified in the top  graticule   See Figure 3   24      To scale or position the unmagnified waveform  press Selected Grati   cule  main   gt  Lower  side   Use the vertical and horizontal knobs to scale  and position the unmagnified waveform in the box     Note that as you scale or move the unmagnified waveform relative to 
191. gt  Ch1  Ch2  Ch3  Ch4  AC Line  or DC Aux  side      To select the coupling you want  press TRIGGER MENU     Type  main   gt   Edge  pop up   gt  Coupling  main   gt  DC  AC  HF Rej  LF Rej  or Noise Rej   side      DC passes all of the input signal  In other words  it passes both AC and DC  components to the trigger circuit     AC passes only the alternating components of an input signal  It removes the DC  component from the trigger signal     HF Rej removes the high frequency portion of the triggering signal  That allows  only the low frequency components to pass on to the triggering system to start an  acquisition  High frequency rejection attenuates signals above 30 kHz     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       Nt LF Rej removes the low frequency portion of the triggering signal  Low  frequency rejection attenuates signals below 80 kHz        Noise Rej provides lower sensitivity  Noise Rej requires additional signal  amplitude for stable triggering  reducing the chance of falsely triggering on  noise        NOTE  When you select Line as the trigger source  the oscilloscope uses AC  coupling to couple a sample of the AC line voltage to the trigger circuits   Although trigger coupling can be set to other than AC when in Line  the  oscilloscope ignores the setting until another source  one of Chl through ChA4  is  selected     In similar fashion  when you select DC Aux  Rear Panel  as the trigger source   the oscilloscope uses DC c
192. gt  Time  main     2  Turn the general purpose knob or use the keypad to set values for the timeout  time     Set the Level   Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Timeout  pop up   gt  Level  main   gt  Level  Set to TTL  Set to  ECL  or Set to 50   side      m If you select Level  you set the trigger level by entering a value with the  general purpose knob or the keypad     m If you select Set to TTL  the oscilloscope sets the trigger level to the TTL  switching threshold     m If you select Set to ECL  the oscilloscope sets the trigger level to the ECL  switching threshold     m If you select Set to 50   the oscilloscope searches for the point halfway  between the peaks of the trigger source signal and sets the trigger level to  that point     Set the Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer to Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see Trigger Modes  and Trigger Holdoff on page 3 53     The TDS Oscilloscope provides a main time base and a delayed time base  The  delayed time base  like the main time base  requires a trigger signal and an input    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       source dedicated to that signal  You can only use delay with respect to the main  edge trigger and certain classes of main pulse triggers  This section describes  how to delay the acqui
193. h a perfectly flat top on the  display  Figure 3   2 shows where the adjustment is located     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms             Figure 3 2  P6139A Probe Adjustment    Input Impedance To ensure proper coupling of your input signals to the oscilloscope  consider the  Considerations following points when you use 50 Q coupling with any channel     m The oscilloscope does not accurately display frequencies under 200 kHz if  AC coupling is selected     m   The oscilloscope reduces the maximum volts division setting for the channel  to 1 V from 10 V  to 10 V from 100 V with a X10 probe attached   since  input amplitudes appropriate for the higher settings would overload the  50 Q input     m The oscilloscope switches to 50 Q and disables AC coupling  and switches  coupling to DC if AC is selected  if you connect an active probe  such as the  P6245 probe  Such probes also reduce the maximum volts div to 10 V as just  described  This behavior results in 50 Q  nonAC coupling  which is  appropriate for such probes        NOTE  If you remove an active probe  the oscilloscope does not switch coupling  back to 1 MQ  nor AC if it was previously selected   Nor does the oscilloscope   when you restore 1 MQ coupling  return to a volts division setting that was  reduced due to the 50 Q selection  In general  you must set channel scale  input  coupling  and impedance appropriate for your input coupling scheme  Be sure to  switch to 
194. h2 to  3 146  Compare Math3 to  3 146  Create Limit Test Template  3 144  Envelope  3 25  H Limit  3 145  Hardcopy if Condition Met  3 146  Hi Res  3 25  Limit Test  3 147  Limit Test Condition Met  3 147  Limit Test Setup  3 146  3 147  Limit Test Sources  3 146  Limit Testing  3 144  OFF  Real Time Only   3   25  OK Store Template  3 145  ON  Enable ET   3 25  Peak Detect  3 25  Repetitive Signal  3 25  Ring Bell if Condition Met  3 146  RUN STOP  3 27  Sample  3 25  Single Acquisition Sequence  3   27  Stop After  3 26  3 147    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Stop After Limit Test Condition Met  3 146  Template Source  3 144  V Limit  3 145  ACQUIRE MENU button  3 25  3 144  Acquiring and Displaying Waveforms  3 3  Acquisition  3 17  Glossary   1  Interval  Glossary   1  Mode  envelope  3 21  Modes  Average  3   23  3 50  Envelope  3 50  Hi Res  3 23  HiRes  3   50  Peak detect  3   23  Sample  3 21  Readout  3   24  Acquisition mode  Choosing an  3   17  Acquisition modes  How to select  3   25  Incompatible with InstaVu  3 48  Active cursor  Glossary   1  Active voltage probes  E 3  active  Saved waveform status  3 118  Advanced applications  Features for  3 143  Advanced DSP Math  Option 2F  A 4  Algorithms  B 1  Aliasing  3 28  3 161  Glossary   1  Amplitude  3 92  Glossary   1  Amplitude Units  Cursor menu  3 105  AND  Glossary   1  AND  Main Trigger menu  3 68  3 70  Applications  derivative math waveforms  3 168  FFT math waveforms  3 150  integra
195. hardcopy from the oscilloscope  Use  the RS 232 or the Centronics port on the controller to print output     Digitizing Oscilloscope                                         Centronics or    GPIB Cable RS 232 Cable                   Figure 3 74  Connecting the Oscilloscope and Hardcopy Device Via a PC    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       Print  If your controller is PC compatible and it uses the Tektronix GURU   or  S3FG210  National Instruments GPIB PCII IIA  GPIB package  do the  following steps to print a hardcopy     1  Use the MS DOS cd command to move to the directory that holds the  software that came with your GPIB board  For example  if you installed the  software in the GPIB PC directory  type  cd GPIB PC     2  Run the IBIC program that came with your GPIB board  Type  IBIC     3  Type  IBFIND DEV1 where    DEV1    is the name for the oscilloscope you  defined using the IBCONF EXE program that came with the GPIB board        NOTE  If you defined another name  use it instead of    DEV1     Also  remember  that the device address of the oscilloscope as set with the IBCONF EXE program  Should match the address set in the oscilloscope Utility menu  typically  use   7 2         4  Type  IBWRT    HARDCOPY START          NOTE  Be sure the oscilloscope Utility menu is set to Talk Listen and not  Hardcopy  Talk Only  or you will get an error message at this step  Setting the  oscilloscope Utility menu is described under Set C
196. he  waveform that was previously stored there  You can store waveforms in reference  locations labeled empty without disturbing previously stored waveforms        2  To store a waveform internally  press SAVE RECALL WAVEFORM  gt   Save Wfm  main   gt  To Ref1  To Ref2  To Ref3  or To Ref4  side      3  To store a waveform to disk  press To File  side   Then use the general  purpose knob to select the exact file from the resulting scrollbar list  Finally   press Save To Selected File  side  to complete the operation        NOTE  Upon power on  the oscilloscope creates the    wild card    file  marked in     shown left   for storing waveforms  Selecting this file in step 3 stores a  waveform in a uniquely named  sequentially numbered file  For instance  the  oscilloscope saves the first waveform you save in the file TEKO0001 WFM  the  second in TEK00002 WFM  and so on        TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups          Tek Run  100kKS 5 Sample       Save  Waveform    To File    To Ref1                   empty          To Ref2  empty          To Ref3  empty       20 0mv i   MM 5004s Chi f 26  8mV Tahora                Oooo o     Recall Wim EAI Delete   sutosave File  to Ref Refs Single Seq utilities    Slaa                                                    Figure 3 66  Save Waveform Menu    To Change Format To select the format that the oscilloscope uses to save waveforms to a disk  press  save recall WAVEFORM  gt  Save Format  main   g
197. he display  or press  AUTOSET      3  Press MORE  gt  Math1  Math2  or Math3  main   gt  Change Math  Definition  side   gt  Single Wfm Math  main    See Figure 3 95      4  Press Set Single Source to  side   Repeatedly press the same button  or use  the general purpose knob  until the channel source selected in step 1 appears  in the menu label     5  Press Set Function to  side   Repeatedly press the same button  or use the  general purpose knob  until diff appears in the menu label     6  Press OK Create Math Wfm  side  to display the derivative of the  waveform you input in step 1     You should now have your derivative math waveform on screen  Use the  Vertical SCALE and POSITION knobs to size and position your waveform  as you require     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       Derivative Math Waveform    Source Waveform    To Take Automated  Measurements    To Take Cursor  Measurements       Tek Run  5 00MS s Sample                   chi 20 0mva o       AOR  200kV s    Mathi Math2 Math3  FFFCCh1    inv chi           10  ous      Ref1          M 10 0   chi T         ov    Ref2 Ref3         Change Math           Math              waveform       definition    No Extended     Processing                Average  2                                Ref4                 Oloo o                           ons   ee  oe                                                Figure 3 95  Derivative Math Waveform    Once you have displaye
198. he display filter types are sin x  x interpolation and linear interpolation  To  switch between interpolation filters     Press DISPLAY  gt  Filter  main   gt  Sin x  x Interpolation or Linear Inter   polation  side      3 32 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms          NOTE  When the horizontal scale is set to the faster rates  10 ns div for the  TDS 600B  see Table 3 4 on page 3 26 for rates specific to the TDS 500B and  TDS 700A models  or when using the ZOOM feature to expand waveforms  horizontally  interpolation occurs   The filter type  linear or sin x x   depends  on which is set in the Display menu   Otherwise  interpolation is not used  See  Interpolation on page 3 19 for a discussion of interpolation        Select the Graticule Type To change the graticule     Press DISPLAY     Graticule  main   gt  Full  Grid  Cross Hair  Frame  NTSC  or PAL  side               Full provides a grid  cross hairs and a frame              Grid displays a frame and a grid                 Cross Hair provides cross hairs  and a frame                 Frame displays just a frame        NTSC provides a grid useful for measuring NTSC class waveforms     PAL provides a grid useful for measuring PAL class waveforms        NOTE  Selecting either NTSC or PAL graticules automatically changes the  vertical scale  position settings  coupling  and sets to zero any vertical offset of  any channel displayed  These settings are not restored after switch
199. he grounding conductor  of the power cord  To avoid electric shock  the grounding conductor must be  connected to earth ground  Before making connections to the input or output  terminals of the product  ensure that the product is properly grounded     Do Not Operate Without Covers  To avoid electric shock or fire hazard  do not  operate this product with covers or panels removed     Use Proper Fuse  To avoid fire hazard  use only the fuse type and rating specified  for this product     Do Not Operate in Wet Damp Conditions  To avoid electric shock  do not operate  this product in wet or damp conditions     Do Not Operate in an Explosive Atmosphere  To avoid injury or fire hazard  do not  operate this product in an explosive atmosphere     Keep Probe Surface Clean and Dry  To avoid electric shock and erroneous  readings  keep probe surface clean and dry     Product Damage Use Proper Power Source  Do not operate this product from a power source that  Precautions applies more than the voltage specified     Provide Proper Ventilation  To prevent product overheating  provide proper  ventilation     Do Not Operate With Suspected Failures  If you suspect there is damage to this  product  have it inspected by qualified service personnel     TDS 500B  TDS 600B   amp  TDS 700A User Manual ix    General Safety Summary       Symbols and Terms    A  A    Certifications and  Compliances    Terms in this Manual  These terms may appear in this manual        WARNING  Warning statements ident
200. high input impedance passive probe  you need to set  the input impedance back to 1 MQ  The procedure To Change Vertical Parame   ters  on page 3 13  explains how to change the input impedance        Time to Voltage Converter    The instantaneous time interval to voltage converter  TVC  continuously  converts consecutive timing measurements to a time interval versus time  waveform     Timing variations typically appear as left to right motion  or jitter  on an  oscilloscope  Time base or trigger holdoff adjustments may improve display  stability  but they do not show timing dynamics  The TVC untangles the often  confusing waveforms and delivers a coherent real time view     The TVC adds three measurement functions to the voltage versus time capability  of your oscilloscope  time delay versus time  pulse width versus time  and period  versus time     TDS 500B  TDS 600B   amp  TDS 700A User Manual E 5    Appendix E  Probe Selection       E 6 TDS 500B  TDS 600B   amp  TDS 700A User Manual    a a  Appendix F  Inspection and Cleaning    General Care    Inspection and Cleaning  Procedures    A    Inspect for dirt and damage on and clean the exterior of the TDS Oscilloscope   When done regularly  this preventive maintenance may prevent oscilloscope  malfunction and enhance its reliability     How often to do this preventive maintenance depends on the severity of the  environment in which the oscilloscope is used  A proper time to perform  preventive maintenance is just before oscillo
201. hmetic mean over one cycle     Cycle RMS  The true Root Mean Square voltage over one cycle     DC coupling  A mode that passes both AC and DC signal components to the circuit   Available for both the trigger system and the vertical system     Delay measurement  A measurement of the time between the middle reference crossings of two  different waveforms     Delay time  The time between the trigger event and the acquisition of data     Digitizing  The process of converting a continuous analog signal such as a waveform to a  set of discrete numbers representing the amplitude of the signal at specific  points in time  Digitizing is composed of two steps  sampling and quantizing     Display system  The part of the oscilloscope that shows waveforms  measurements  menu  items  status  and other parameters     Edge Trigger  Triggering occurs when the oscilloscope detects the source passing through a  specified voltage level in a specified direction  the trigger slope      Envelope acquisition mode  A mode in which the oscilloscope acquires and displays a waveform that  shows the variation extremes of several acquisitions     TDS 500B  TDS 600B   amp  TDS 700A User Manual Glossary 3    Glossary                                                                   Glossary 4    Equivalent time sampling  ET   TDS 500B and TDS 700A Models Only  A sampling mode in which the  oscilloscope acquires signals over many repetitions of the event  The  TDS 500B and TDS 700A Oscilloscopes use a type o
202. hoot  Glossary   8   Peak to peak  3 93  Glossary   8  Period  3   93  Glossary   8   Phase  3 93  Glossary   8   Positive duty cycle  3 93   Positive overshoot  3   93   Positive width  3   93   Propagation delay  3 92   Readout  3   94   Reference levels  2   21   Rise time  2 20  3 93  Glossary   9  RMS  3 93  Glossary   9  Undershoot  Glossary   7   Width  2 20  Glossary   7  Glossary   8    Measurement Accuracy  Ensuring maximum  3 106     3 108    Measurements    Algorithms  B   1  Automated  2 19  automated  3   91  Classes of  3 91  Cursor  3 101  Gated  3 95    TDS 500B  TDS 600B   amp  TDS 700A User Manual    graticule  3 105   List of automated  3   92   Snapshot of  3 99  Measuring Waveforms  3 91  Memory  Waveform  3 120  Menu   Acquire  3   25  3 144   Color  3 35   Cursor  3 103   Delayed Trigger  3    86   3   90   Display  3 29  3 35   File Utilities  3 122   Horizontal  3   55  3   86   Main  2   6   Main Trigger  3 59  3   60  3   66  3   69  3   70  3 75    3 77  3 80  Measure  3   94  3   99  More  3   120  3 148  3 152  3 168  See also More menu   Operation  2   7   Pop up  2 8  Glossary   8   Save Recall  3 115   Save Recall Waveform  3 118   Setup  2 10  3 8   Status  3 139 3 142   Utility  3 127  Mid Ref  Measure menu  3   98  Mid  Ref  Measure menu  3   98  Min Max  Measure menu  3   96  Minimum  3   93  Glossary   7  Mode  amp  Holdoff  Main Trigger menu  3   61  Model number location  2   3  Models  Manual references to  xii  Models  key features a
203. hreshold  The pulse must travel  between the two levels ata rate faster or slower than  user  specifies  the user specified slew rate for triggering to  occur        iz    Timeout stays high    Triggering occurs if the signal stays higher than the trigger  level for longer than the timeout value           TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       Table 3 7  Pulse Trigger Definitions  Cont      Name Definition    4  Timeout stays low Triggering occurs if the signal stays lower than the trigger  os level for longer than the timeout value           Timeout either Triggering occurs if the signal stays higher or stays lower  ma than the trigger level for the timeout value                 To Trigger ona Glitch When you select the pulse class Glitch  the oscilloscope will trigger on a pulse  narrower  or wider  in width than some specified time  To set up for glitch  triggering  do the following procedures     Select Glitch Triggering  Press TRIGGER MENU  gt  Type  main   gt  Pulse   pop up   gt  Class  main   gt  Glitch  pop up         5       Tek Run  TS Sample  SEDE E AAEE EE ETET TE       A  Pulse Glitch  Source    po    ch2                               Runt                   Slew Rate  Timeout    a   Eme oei   cien   Level ode  Type Class A  amp   amp    lt Pubse gt     lt Glitch gt   e width        cept   252mV   Holdoff          Kas  es   es  es  ey    2q_Width i M 500qs Glitch Chi                                                 MID ClO 
204. ic pattern and  state function     Area  Measurement of the waveform area taken over the entire waveform or the    TDS 500B  TDS 600B   amp  TDS 700A User Manual Glossary 1    Glossary       Glossary 2                                  UU          gated region  Expressed in volt seconds  Area above ground is positive  area  below ground is negative     Attenuation  The degree the amplitude of a signal is reduced when it passes through an  attenuating device such as a probe or attenuator  That is  the ratio of the input  measure to the output measure  For example  a 10X probe will attenuate  or  reduce  the input voltage of a signal by a factor of 10     Automatic trigger mode  A trigger mode that causes the oscilloscope to automatically acquire if  triggerable events are not detected within a specified time period     Autoset  A function of the oscilloscope that automatically produces a stable waveform of  usable size  Autoset sets up front panel controls based on the characteristics of  the active waveform  A successful autoset will set the volts div  time div  and  trigger level to produce a coherent and stable waveform display     Average acquisition mode  In this mode  the oscilloscope acquires and displays a waveform that is the  averaged result of several acquisitions  Averaging reduces the apparent noise   The oscilloscope acquires data as in the sample mode and then averages it  according to a specified number of averages     Bandwidth  The highest frequency signal the
205. ice  press  HARDCOPY to print your date time stamped hardcopy        Tek Run  100kKS 5 Sample                                           gil oom soos Chi A OW 25 Mar 1994    08 39 47  cano ey     Ke  e   oe  es   ee  es        Date and Time Display       i                                              Figure 3 72  Date and Time Display    Set the Date and Time  You might need to set the date and time of the oscillo   scope  To set those parameters  do the following steps     1  Press SHIFT  gt  UTILITY     Config  pop up   gt  Set Date  amp   Time  main   gt  Year  Day Month  Hour  or Minute  side      2  Use the general purpose knob or the keypad to set the parameter you have  chosen to the value desired   The format when using the keypad is  day month  For example  use 23 6 for the 23 4 of June      3  Repeat steps 1 and 2 to set other parameters as desired     4  Press OK Enter Date Time  side  to put the new settings into effect  This  sets the seconds to zero     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 129    Saving Waveforms and Setups          NOTE  When setting the clock  you can set to a time slightly later than the  current time and wait for it to catch up  When current time catches up to the time  you have set  pressing Ok Enter Date Time  side  synchronizes the set time to  the current time        5  Press CLEAR MENU to see the date time displayed with the new settings     To Print Directly toa To make your hardcopies  use the procedures that follow   Hardc
206. idth  below     If Period   0 or undefined then return an error   NegativeWidth    Period BUN    NegativeDutyCycle    Amplitude  voltage  measurement     NegativeOvershoot   Amplitude    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix B  Algorithms       Note that this value should never be negative  unless High or Low are set  out of range      Negative Width Timing measurement  The distance  time  between MidRef  default   50    amplitude points of a negative pulse                 If MCross1Polarity        gt   then   NegativeWidth    MCross2     MCross1   else    NegativeWidth    MCross3     MCross2     Peak to Peak Amplitude measurement  The absolute difference between the maximum and  minimum amplitude        PeaktoPeak   Max     Min    Period Timing measurement  Time taken for one complete signal cycle  The reciprocal  of frequency  Measured in seconds                    Period   MCross3     MCross1    Phase Timing measurement  The amount of phase shift  expressed in degrees of the  target waveform cycle  between the MidRef crossings of two different wave   flr forms  Waveforms measured should be of the same frequency or one waveform  should be a harmonic of the other     Phase is a dual waveform measurement  that is  it is measured from a target  waveform to a reference waveform  To get a specific phase measurement  you  must specify the target and reference sources     Phase is determined in the following manner     1  The first MidRefCrossing  MCrossI Targ
207. ied  3 86  3   88   Main Scale  3   16   Record Length  3 16   Set to 10   3 17   Set to 50   3 17   Set to 90   3 17   Time Base  3 86   Trigger Position  3 16  HORIZONTAL MENU button  3 55  3 86  Horizontal POSITION knob  3   15  3   40  Horizontal Readouts  3   14  Horizontal SCALE knob  3 15  3   40  HPGL  3 126  HPGL  Hardcopy menu  3 127  Hue  Color menu  3   36    I O  Status menu  3   139  TO  Utility menu  3 127  Icons  File Utilities menu  3 116  3 118  3 122  3 125  Independent  Cursor menu  3   104  Infinite Persistence  Display menu  3   30  Installation  1   4  InstaVu  3   45  Glossary   6  InstaVu mode  How to use  3   45  Modes incompatible with  3   47  Vs  Normal DSO mode  illustrated   3   46  Waveform capture rate  3   45  Integral math waveform  3 172  applications  3 171  automated measurements of  3 175  derivation of  3 172  magnifying  3 171  3 175  procedure for displaying  3 172  procedure for measuring  3 173  record length of  3 172  Integration  Waveform  3 171       TDS 500B  TDS 600B   amp  TDS 700A User Manual    Intensified Samples  Display menu  3   30  Intensified  Horizontal menu  3   86  3   88  Intensity  Glossary   5  Intensity  Display menu  3 31  Interleaf  3 126  Interleaf  Hardcopy menu  3 127  Interleaving  3 20  Glossary   6  Interpolation  3 19  3   20  3 32  Glossary   5  And zoom  3   40  FFT distortion  3 161  Incompatible with InstaVu  3 48  linear versus sin x  x  3 161  IRE  NTSC   Cursor menu  3 105    K    Keypad  2 7  2
208. ify conditions or practices that could result  in injury or loss of life           CAUTION  Caution statements identify conditions or practices that could result in  damage to this product or other property        Terms on the Product  These terms may appear on the product     DANGER indicates an injury hazard immediately accessible as you read the  marking     WARNING indicates an injury hazard not immediately accessible as you read the  marking     CAUTION indicates a hazard to property including the product     Symbols on the Product  The following symbols may appear on the product     A    A m    DANGER Protective Ground ATTENTION Double  High Voltage  Earth  Terminal Refer to Manual Insulated    Refer to the specifications section in the Performance Verfication manual for a  listing of certifications and compliances that apply to this product     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Preface    Related Manuals    This is the User Manual for the TDS 500B  TDS 600B   amp  TDS 700A Digitizing  Oscilloscopes     The chapter Getting Started briefly describes the TDS Oscilloscope  prepares  you to install it  and tells you how to put it into service     The chapter Operating Basics covers basic principles of the operation of the  oscilloscope  The operating interface illustrations and the tutorial examples  rapidly help you understand how your oscilloscope operates     The chapter Reference teaches you how to perform specific tasks  See page 3   1  for a complete l
209. igger point indicator and level bar are displayed from the Display  menu  See Set Display Readout Options on page 3   31 for more information     The trigger point indicator shows position  It can be positioned horizontally off  screen  especially with long record length settings  The trigger level bar shows  only the trigger level  It remains on screen  regardless of the horizontal position   as long as the channel providing the trigger source is displayed     Trigger Status Screen  To see a more comprehensive status listing of the settings  for the main and delayed trigger systems  press SHIFT STATUS  gt  STA   TUS  main   gt  Trigger  side      Trigger Menu Each trigger type  edge  logic  and pulse  has its own main trigger menu  which  is described as each type is discussed in this section  To select the trigger type   press TRIGGER MENU  gt  Type  main   gt  Edge  Logic  or Pulse  pop up      Triggering on a Waveform Edge    The TDS Oscilloscope can trigger on an edge of a waveform  An edge trigger  event occurs when the trigger source passes through a specified voltage level in a  specified direction  the trigger slope   You will likely use edge triggering for  most of your measurements  This subsection describes how use edge trigger    ing     how to select edge type  source  coupling  slope  and level  It also details  how to select trigger mode  auto or normal  for all trigger types     To Check Edge To quickly check if edge triggers are selected  check the Trigger r
210. igh is the value used as the 100  level in measurements such as fall time and  rise time  For example  if you request the 10  to 90  rise time  then the  oscilloscope will calculate 10  and 90  as percentages with High representing  100      Low is the value used as the 0  level in measurements such as fall time and rise  time     The exact meaning of High and Low depends on which of two calculation  methods you choose from the High Low Setup item of the Measure menu   These are Min max and Histogram     Min Max Method     defines the 0  and the 100  waveform levels as the lowest  amplitude  most negative  and the highest amplitude  most positive  samples   The min max method is useful for measuring frequency  width  and period for  many types of signals  Min max is sensitive to waveform ringing and spikes   however  and does not always measure accurately rise time  fall time  overshoot   and undershoot     The min max method calculates the High and Low values as follows   High   Max  and  Low   Min  Histogram Method     attempts to find the highest density of points above and  below the waveform midpoint  It attempts to ignore ringing and spikes when    determining the 0  and 100  levels  This method works well when measuring  square waves and pulse waveforms     TDS 500B  TDS 600B   amp  TDS 700A User Manual B 1    Appendix B  Algorithms       HighRef  MidRef  LowRef   Mid2Ref    The oscilloscope calculates the histogram based High and Low values as  follows     1  It makes
211. illoscope Do the following steps to reset the oscilloscope to a known factory default state  before doing the examples   You can reset the oscilloscope anytime you begin a  new task and need to    start fresh    with known default settings      1  Press the save recall SETUP button to display the Setup menu   See  Figure 2 2            SAVE RECALL  AUTOSET     WAVEFORM    STATUS          APPLICATION C MENU     ACQUIRE MENU  MEASURE  HARDCOPY   RUN STOP         UTILITY  SETUP CURSOR DISPLAY Sar  ae Oe E a    SETUP Button f    Qs VERTICAL      A HORIZONTAL   yi                Figure 2 2  SETUP Button Location    The oscilloscope displays main menus along the bottom of the screen   Figure 2   3 shows the Setup main menu        OK Confirm Factory Init  Menu Item and Button                                                 Recall Factory Setup  Menu Item and Button       Figure 2 3  The Setup Menu    2  Press the button directly below the Recall Factory Setup menu item     2 10 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Tutorial       The display shows side menus along the right side of the screen  The buttons  to select these side menu items are to the right of the side menu     Because an accidental instrument reset could destroy a setup that took a long  time to create  the oscilloscope asks you to verify the Recall Factory Setup  selection   See Figure 2 3      3  Press the button to the right of the OK Confirm Factory Init side menu item        NOTE  This manual uses the
212. ing from the Front Panel    The trigger buttons and knob let you quickly adjust the trigger level or force a  trigger   See Figure 3   33   The trigger readout and status screen lets you quickly  determine the state of the trigger system  You use the following trigger controls  and readouts for all trigger types except where noted     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 55    Triggering on Waveforms       3 56    Trigger Status Lights    To set MAIN LEVEL    To Set to 50     To Force a Trigger       4 TRIGGER    MAIN LEVEL    fh      DELAYED TRIG    TRIGGER  MENU    J    SET LEVEL  TO 50     J    SINGLE TRIG  FORCE TRIG    J    Figure 3 33  TRIGGER Controls and Status Lights    To manually change the trigger level when edge triggering  or certain threshold  levels when logic or pulse triggering   turn the MAIN LEVEL knob  It adjusts  the trigger level  or threshold level  instantaneously no matter what menu  if any   is displayed     To quickly obtain an edge trigger or a glitch or width pulse trigger  press SET  LEVEL TO 50   The oscilloscope sets the trigger level to the halfway point  between the peaks of the trigger signal  Set Level to 50  has no effect when  trigger type is logic or video     You can also set the level to 50  in the Trigger menu under the main menu item  Level if edge trigger or glitch or width pulse trigger is selected     Note that the MAIN LEVEL knob and menu items apply only to the main trigger  level  To modify the delayed trigger leve
213. ing negative but  fails to cross a second threshold before recrossing the first  going positive           Runt either    Width positive       Triggering occurs if the oscilloscope detects a positive or  negative going pulse that crosses one threshold but fails to  cross a second threshold before recrossing the first     Triggering occurs if the oscilloscope finds a positive pulse  with a width between  or optionally outside  the user speci   fied lower and upper time limits              A    Width negative    Slew positive    Slew negative    Slew either    Triggering occurs if the oscilloscope finds a negative pulse  with a width between  or optionally outside  the user speci   fied lower and upper time limits     Triggering occurs if the oscilloscope detects a positive  pulse edge that after first crossing the lower threshold then  crosses the upper threshold  The pulse must travel  between the two levels ata rate faster or slower than  user  specifies  the user specified slew rate for triggering to  occur                       a       Triggering occurs if the oscilloscope detects a negative  pulse edge that after first crossing the upper threshold then  crosses the lower threshold  The pulse must travel between  the two levels ata rate faster or slower than  user  specifies  the user specified slew rate for triggering to  occur     Triggering occurs if the oscilloscope detects a positive or  negative pulse edge that first crosses one threshold and  then crosses the other t
214. ing to find the cause of an unwanted glitch in your test  circuit  it might trigger on the glitch and make the pretrigger period large enough    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       to capture data before the glitch  By analyzing what happened before the glitch   you may uncover clues about its source     Slope and Level The slope control determines whether the oscilloscope finds the trigger point on  the rising or the falling edge of a signal   See Figure 3 32      a You set trigger slope by first selecting Slope in the Main Trigger menu and then  selecting between the rising or falling slope icons in the side menu that appears     2 The level control determines where on that edge the trigger point occurs   See  Figure 3   32   The oscilloscope lets you set the main trigger level with the trigger  MAIN LEVEL knob     Positive Going Edge Negative Going Edge    Trigger level  can be adjusted  vertically     X J  N    Trigger slope can be positive or negative        Figure 3 32  Slope and Level Controls Help Define the Trigger    Delayed Trigger System The oscilloscope also has a delayed trigger system that provides an edge trigger   no pulse or logic triggers   When using the delayed time base  you can also  delay the acquisition of a waveform for a user specified time or a user specified  number of delayed trigger events  or both  after a main trigger event  See  Delayed Triggering on page 3   84 to learn how to use delay     Trigger
215. ing to other  graticule types  Therefore  you might want to recall the factory setup or other  stored setup after selecting a different graticule        Select the Format The oscilloscope displays waveforms in either of two formats  YT and XY  To  set the display axis format     TDS 600B  Press DISPLAY     Format  main   gt  XY or YT  side      TDS 500B and TDS 700A models  Press DISPLAY     Mode  main   gt  Normal   pop up   gt  Format RO  main   gt  XY or YT  side         YT is the conventional oscilloscope display format  It shows a signal voltage  the  vertical axis  as it varies over time  the horizontal axis               TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 33    Acquiring and Displaying Waveforms       eo XY format compares the voltage levels of two waveform records point by point   That is  the oscilloscope displays a graph of the voltage of one waveform record  against the voltage of another waveform record  This mode is particularly useful  for studying phase relationships     When you choose the XY format  any channel or reference displayed is assigned  to the axis indicated in Table 3 5 and displayed as part of an XY pair  If only one  source in an XY pair is displayed  the oscilloscope automatically turns on the  other source to complete the XY pair when you select XY  Moreover  once XY is  on  selecting either source in a pair turns the pair on  pressing WAVEFORM OFF  for either source in the pair removes both sources from the display     Table 3 5 
216. ion 1G on page A 1     All oscilloscopes in this manual support the following four acquisition modes   Sample  Envelope  Average  and Peak Detect  TDS 500 B and TDS 700A  Oscilloscopes also support Hi Res  Keep in mind which modes apply to your  model oscilloscope as you read the following descriptions     Sample  the mode most commonly used   Peak Detect  and Hi Res modes  operate in real time on a single trigger event  provided that the oscilloscope can  acquire enough samples for each trigger event  Envelope and Average modes  operate on multiple acquisitions  the oscilloscope averages or envelopes several  waveforms on a point by point basis   For TDS 500B and TDS 700A models  only  Hi Res  Envelope  and Average modes are not available when in InstaVu  mode  see Incompatible Modes on page 3   47      Figure 3   13 illustrates the different modes and lists the benefits of each  It will  help you select the appropriate mode for your application  Refer to it as your  read the following descriptions of each mode     Sample Mode  In Sample mode  the oscilloscope creates a record point by saving  the first sample  of perhaps many  during each acquisition interval   An  acquisition interval is the time covered by the waveform record divided by the  record length   Sample mode is the default mode     Envelope Mode  In Envelope mode  the oscilloscope acquires and displays a  waveform record that shows the extremes in variation over several acquisitions   you specify the number of 
217. ishing  3 126  Hardcopy  3 115  Deskew  3 108  Glossary   2  Deskew  Vertical menu  3 108  Deskjet  3 126  Deskjet  Hardcopy menu  3 127  DeskjetC  Hardcopy menu  3 127  Differential active probes  E 3  Differentiation  of a derivative  3 168  waveform  3 167  Digitizing  Glossary   3  Disk  How to save a hardcopy to  3 131  Disk drive  3 121  Display  2   6  Hardcopy of  3 125  Options  3   29 3 50  Record View  3   58  System  Glossary   3  Display    T      Trigger Point  Display menu  3 31  DISPLAY button  3   29  3 35  Display menu  3 29  3 35  Color  3 35  Contrast  3   31  Cross Hair  3 33  Display  3 29  Display    T      Trigger Point  3   31  Dots  3 30  Dots style  3 146  Filter  3   32  Frame  3 33  Full  3 33  Graticule  3   33  Grid  3 33  Infinite Persistence  3   30  Intensified Samples  3   30  Intensity  3 31  Linear interpolation  3   32  NTSC  3 33  Overall  3   31  PAL  3 33  Readout  3 31  3 33  Settings  3 29  3 35  Sin x  x interpolation  3   32  Style  3 29  Text Grat  3 31  Trigger Bar  3 31  Variable Persistence  3   30  Vectors  3 30  Waveform  3 31  XY  3 33  YT  3 33    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index       Display  Display menu  3   29   Display  Status menu  3 139   Dots  3 30   Dots style  Display menu  3 146  Dots  Display menu  3 30  DPU411 II  Hardcopy menu  3 127  DPU412  Hardcopy menu  3 127  Dual Wfm Math  More menu  3   149  Dual Window Zoom  3 42   Dual Zoom  Zoom menu  3   43   Dual Zoom Offset  Zoom menu  3   
218. ist of operating tasks covered in that chapter     The Appendices provide an options listing  an accessories listing  and other  useful information     The following documents are related to the use or service of the oscilloscope     m The TDS Family Digitizing Oscilloscopes Programmer Manual  Tektronix  part number 070 9556 XX  describes using a computer to control the  oscilloscope through the GPIB interface     m The TDS 500B  TDS 600B   amp  TDS 700A Reference  Tektronix part number  070 9382 XX  gives you a quick overview of how to operate the oscillo   scope     m The TDS 500B  TDS 600B   amp  TDS 700A Technical Reference  Performance  Verification and Specifications   Tektronix part number 070 9720 XX  tells  how to verify the performance of the oscilloscope and lists its specifications     m The TDS Family Option 05 Video Trigger Instruction Manual  Tektronix  part number 070 8748 XX  describes use of the video trigger option  for  TDS oscilloscopes equipped with that option only      m The TDS 500B  TDS 600B   amp  TDS 700A Service Manual  Tektronix part  number 070 9721 XX  provides information for maintaining and servicing  the oscilloscope to the module level     TDS 500B  TDS 600B   amp  TDS 700A User Manual xi    Preface       Default Model    Model References    Conventions    xii    This manual documents the TDS 500B  TDS 600B   amp  TDS 700A Digitizing  Oscilloscopes  Take special note of the following conventions     Some TDS models have two auxiliary channe
219. it on if it is not already on     You do not use the channel selection buttons to select the trigger source   Instead you select the trigger source in the Main Trigger menu or Delayed  Trigger menu     2  Press WAVEFORM OFF to turn OFF the display of the selected channel  waveform  It will also remove from the display any automated measurements  being made on that waveform     3  To select a math waveform you have created or a reference waveform you  have stored  press MORE and select the waveform from the More menu   Press WAVEFORM OFF while the MORE button is lit to remove the  display of the waveform selected in the More menu     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 9    Acquiring and Displaying Waveforms       Waveform Priority    To Find More Information    When you turn off a waveform  the oscilloscope automatically selects the next  highest priority waveform  Figure 3   4 shows the order of priority     1  CH1 1  MATH1  2  CH2 2  MATH2  3  CH3 or AX1 3  MATH3  4  CH4 or AX2 4  REF1  5  REF2  6  REF3  7  REF4    Figure 3 4  Waveform Selection Priority    Note Figure 3   4 shows two orders of priority due to the following rules  If you  are turning off more than one waveform and you start by turning off a channel  waveform  all channels will be turned off before going to the MORE waveforms   If you start by turning off the MORE waveforms  all the MORE waveforms will  be turned off before going to the channel waveforms     If you turn off a channel that is a t
220. ition  8 di       labeled Probe Compensation SIGNAL and GAD              Probe       Cal             Clear  Probe Cal    Compensate    Cal Probe  initialized                         OK       LOIDID CIO                                 GaS                      Cc        Figure 3 63  Probe Cal Menu and Gain Compensation Display    11  If the Probe Offset Compensation message is displayed  continue with step    15  otherwise  continue with step 12     12  If the Compensation Error message is displayed  continue with step 13     otherwise continue with step 18     13  Press SHIFT UTILITY  gt  System  main   gt  Diag Err  pop up   gt  Error  Log  main   If there are too many error messages to be seen on screen  rotate  the general purpose knob clockwise to scroll to the last message     14  Note the compensation error amount  Skip to step 19     15  Disconnect the probe from any signal you may have connected it to  Leave    the probe installed on its channel     16  Press OK Compensate Offset  side      D 17  Wait for offset compensation to complete  one to three minutes      When offset compensation completes  the following occurs     m The clock icon will disappear     m If offset compensation did not complete successfully  you may get the  message    Compensation Error     This error implies that the probe offset  scale  10  error  and or offset  50 mV  is too great to be compensated   You can substitute another probe and continue  Have your probe checked    TDS 500B  TDS 600B  
221. itive and nonrepetitive waveforms and does  not affect the frequency response of the source waveform     Hi Res and Average Reduce Noise  If the pulse is repetitive  Average mode may be  used to reduce noise in the signal at a cost of slower display response  Average  operates on repetitive waveforms only  and averaging does affect the frequency  response of the source waveform     For TDS 500B and TDS 700A models only  after the FFT is set up and  displayed  you can to turn on Hi Res mode to reduce the effect of noise in the  signal  Hi Res operates on both repetitive and nonrepetitive waveforms   however  it does affect the frequency response of the source waveform     Peak Detect and Envelope Add Distortion  Peak Detect and Envelope mode can  add significant distortion to the FFT results and are not recommended for use  with FFTs     Once you have your waveform displayed optimally  you may magnify  or  reduce  it vertically and horizontally to inspect any feature you desire  Just be  sure the FFT waveform is the selected waveform   Press MORE  then select the  FFT waveform in the More main menu  Then use the Vertical and Horizontal  SCALE knobs to adjust the math waveform size      If you wish to see the zoom factor  2X  5X  etc   you need to turn Zoom on   press ZOOM  gt  On  side   The vertical and horizontal zoom factors appear on  screen     Whether Zoom is on or off  you can press Reset Zoom Factors  side  to return  the zoomed FFT waveform to no magnification     Zoo
222. just the oscilloscope display  style  intensity level  graticule  and format        NOTE  TDS 500B and 700A models only  This section assumes you are using  Normal acquisitions mode and gives display settings for this mode  If you select  InstaVu acquisitions  procedures for making Style  Format  and Readout display  settings differ and some selections are not permitted  See Using InstaVu     Acquisition Mode  on page 3 45  for setup differences and Incompatible Modes on  page 3 47        Change Display Settings To bring up the Display menu   Press DISPLAY  gt  Settings  main   gt  Display  pop up      The Display menu allows you to adjust the style  intensity level  graticule  and  format features described below  The Color menu allows you to alter color  settings for various display components such as waveforms and text  To find  more information on color  see Customizing the Display Color on page 3 34     Select the Display Style TDS 600B only  Press DISPLAY  gt  Style  main   gt  Vectors  Dots  Intensified  Samples  Infinite Persistence  or Variable Persistence  side    See Fig   ure 3 17      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 29    Acquiring and Displaying Waveforms       3 30    TDS 500B and TDS 700A models only  Press DISPLAY     Mode  main   gt   Normal  pop up   gt  Style  main   gt  Vectors  Dots  Intensified Samples   Infinite Persistence  or Variable Persistence  side     Vectors style displays vectors  lines  between the record points   Dots style 
223. k  For later field conversions  order        kit  016 1136 00   20 TCP202 CurrentProbe   Add current probe calibrator  Calibrator  Option is only available for TDS 500B and TDS 700A models     l 21 TCP202 CurrentProbe   Add current probe calibrator probe and HV differential probe    A 2    Calibrator and  HV Differential P robe     Option is only available for TDS 500B and TDS 700A models      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix A  Options and Accessories       Table A 1  Options  Cont      Option   Label  22 Two passive probes    Description    Add two 500 MHz P6139A passive probes   Option applies only to TDS 520B and TDS 724A models         23 Two active probes    Add two 1 GHz P6243 active probes   Option applies only to TDS 500B  TDS 724A  and TDS 754A  models         24 Four passive probes    26 Four active probes    27 Two active probes    28 ADA 400A Differential  Preamplifier    Add four 500 MHz P6139A passive probes   Option applies only to TDS 600B  TDS 782A  and TDS 784A  models      Add four 1 5 GHz  probe only  P6245 active probes   Option applies only to TDS 684B and TDS 784A models      Add two 1 5 GHz  probe only  P6245 active probes   Option applies only to TDS 680B and TDS 782A models      Add one differential preamp  switchable gain   Option applies only to TDS 500B and TDS 700A models         2D Two active probes    a AEA Arab    Delete two standard probes normally shipped with the  oscilloscope    Option applies only to TDS 520B  T
224. l  use the Level item in the Delayed  Trigger menu     To force the oscilloscope to immediately start acquiring a waveform record even  without a trigger event  press the FORCE TRIG front panel button     Forcing a trigger is useful when in normal trigger mode and the input signal is  not supplying a valid trigger  By pressing FORCE TRIG  you can quickly  confirm that there is a signal present for the oscilloscope to acquire  Once that is  established  you can determine how to trigger on it  press SET LEVEL TO  50   check trigger source setting  and so on      The oscilloscope recognizes and acts upon FORCE TRIG even when you press it  before the end of pretrigger holdoff  However  the button has no effect if the  acquisition system is stopped     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       To Single Trigger To trigger on the next valid trigger event and then stop  press SHIFT FORCE  TRIG  Now press the RUN STOP button each time you want to initiate the  single sequence of acquisitions     To leave Single Trig mode  press SHIFT ACQUIRE MENU  gt  Stop Af   ter  main    RUN STOP Button Only  side      See the description under Stop After on page 3   26 for further discussion of  single sequence acquisitions      Single sequence triggering is not available in InstaVu mode  see Incompatible  Modes on page 3 47      To Check Trigger Status To ascertain the state and setup of the triggering circuit  use the trigger status  lights  readout  and sc
225. l for the channel  when its vertical offset is set to 0 V in the vertical menu  when vertical offset  is not set to O V  it points to the vertical offset level     The trigger readout shows that the oscilloscope is triggering on channel 1   Ch1  on a rising edge  and that the trigger level is about 200 300 mV     The time base readout shows that the main time base is set to a horizontal  scale of 500 us div     The channel readout indicates that channel 1  Ch1  is displayed with DC  coupling   In AC coupling    appears after the volts div readout   The  oscilloscope always displays channel 1 at reset                                                                    Tek Run  100kKS 5 Sample  Recall  Factory  ne       cr Nr S OK  1 00 Y        M 5004S chi T 220mV  a a aaa i   Factory Init  eca ave eca File m  Saved Current Factory Utilities  Setup Setup Setu                                                                      ceeiseels       Figure 2 5  The Display After Factory Initialization    Right now  the channel  time base  and trigger readouts appear in the graticule  area because a menu is displayed  You can press the CLEAR MENU button at  any time to remove any menus and to move the readouts below the graticule     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Tutorial       Example 1  Displaying a Waveform    The TDS Oscilloscope provides front panel knobs for you to adjust a waveform   or it can automatically set up its controls to display a waveform  Do the fo
226. l math waveforms  3 171  Area  3   92  Glossary   1  Attenuation  Glossary   2  Auto  Main Trigger menu  3 61  Automated Measurements  Snapshot of  2   23  Automated measurements  2   19  3   91  of derivative math waveforms  3 169   procedure   3 169  of FFT math waveforms  3 156  of integral math waveforms  3 175  Automatic trigger mode  3 53  Glossary   2  Autosave  Save Recall Waveform menu  3 121  Autoset  2 14  3 6  Glossary   2  Default settings  3   7  How to execute  3   6    Index   1    Index       AUTOSET button  2 14   AUX TRIGGER INPUT  BNC  2 5   Auxiliary trigger  3   52   Average  Incompatible with InstaVu  3   48  Average acquisition mode  3   23  3 50  Glossary   2  Average mode  Acquire menu  3 144   Average  Acquire menu  3   25   Average  More menu  3 150    Bandwidth  Glossary   2  Selecting  3 13  Bandwidth  Vertical menu  3   13  Base  Cursor menu  3   105  Blackman Harris window  3   154  BMP  3 126  BMP Color  Hardcopy menu  3 127  BMP Mono  Hardcopy menu  3 127  BNC  AUX TRIGGER INPUT  2 5  DELAYED TRIGGER OUTPUT  2 5  MAIN TRIGGER OUTPUT  2 5  SIGNAL OUTPUT  2 5  Bold  Color menu  3 35  Burst width  3   92  Button  ACQUIRE MENU  3 25  3 144  AUTOSET  2 14  CLEAR MENU  2 3  2 8  2 12  2 20  2 21  3 99  CURSOR  3 103  DELAYED TRIG  3 55  3   86  DISPLAY  3 29  3 35  FORCE TRIG  3 56  HARDCOPY  3 122  3 127  3 136  HELP  3 141  HORIZONTAL MENU  3 55  3 86  InstaVu  3   45  MEASURE  3 94  MORE  3 9  3 120  3 148  ON STBY  1 5  2 3  Save Recall SETUP  
227. l menu   20 MHz  3 13   250 MHz  3 13   Bandwidth  3 13   Cal Probe  3   108   Coupling  3 13   Deskew  3 108   Fine Scale  3 13   Full  3 13   Offset  3   14   Position  3 13   Set to Zero  3 13  VERTICAL MENU button  2   17  Vertical position  for DC correction of FFTs  3 159  Vertical POSITION knob  3 11  3   40  Vertical Readout  3 11  Vertical SCALE knob  3 11  3 40  VGA Output  2 5  Video Line Number  Cursor menu  3 105  Video Trigger  Option 5  A 1  Video trigger  3   53    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index       View Palette  Color menu  3 35    W    Waveform  Glossary   11  Acquiring and Displaying of  3   3  Autoset on  3   6  Coupling to the oscilloscope  3   3  Interval  Glossary   11  Math  3 148  Priority for turning off  3 10  Save Formats  3 119  Saving  3 115  Triggering on  3 51  Waveform clipping  See Clipping  Waveform differentiation  3 167  Waveform FFTs  3   150  Waveform integration  3 171  Waveform memory  3 120  WAVEFORM OFF button  2 18  3 9  3 34  Waveform record  FFT  3 156  FFT frequency domain  3 157  length of  3 157  FFT source  3 156  acquisition mode  3 160  defined  3 157  long versus short  3 159  FFT time domain  3   157  Waveform  Display menu  3 31  Waveforms  And zoom  3   39  Math  3 150  Measuring  3 91  Scaling and positioning  3   10  Waveforms  Status menu  3   139  Width  2   20  Glossary   7  Glossary   8  Width trigger  3 73  3 79  How to set up  3   79 3 90  Width  Main Trigger menu  3   76  3   79  Windo
228. lative to the trigger event  To define the trigger point position     Press HORIZONTAL MENU  gt  Trigger Position  main   gt  Set to 10   Set    to 50   or Set to 90   side   or use the general purpose knob or the keypad to  change the value     Record Length  The number of points that make up the waveform record is  defined by the record length  To set the waveform record length     1  Press HORIZONTAL MENU  gt  Record Length  main   Select the record  length desired from the side menu  Press  more  to see additional choices     m All TDS 500B  TDS 600B  and TDS 700A Oscilloscopes provide  standard record lengths of 500  1000  2500  5000  and 15000 points     m For instruments shipped with Option 1M  the TDS 500B and TDS 700A  Oscilloscopes provide additional extended record lengths up to 500 000  points  depending on the model  For specific record lengths available   see the 1M option in the Table A 1 on page A 2  Option 1M is available  only at the time of original purchase and is not available for the  TDS 600B models        NOTE  TDS 500B and TDS 700A Models  Hi Res acquisition mode requires twice  the acquisition memory of other acquisition modes  Therefore  the maximum  record length available is 15 000 points without option 1M and 50 000 points  with option 1M  Turning Hi Res on switches the setting for record length  accordingly  thereby keeping the oscilloscope from running out of memory        2  To fit an acquired waveform to the visible screen  regardless of reco
229. layed signal may look rounded or pointed   See Figure 2   10   If so   then you may need to compensate your probe  See To Compensate Passive  Probes on page 3 4        AMETE ee    Figure 2 10  Display Signals Requiring Probe Compensation    Example 2  Displaying Multiple Waveforms    The TDS Oscilloscope can display up to four channels  three math waveforms   and four reference waveforms at one time  Do the following tasks to learn how to  display and control more than one waveform at a time     TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 15    Tutorial       Add a Waveform    The VERTICAL section of the front panel contains the channel selection  buttons  These buttons are CH 1  CH 2  CH 3  CH 4  and MORE   See Fig   ure 2 11    CH 3 and CH 4 will be replaced by AUX1 and AUX2 on some  models  see Default Model on page xii and Differences by Model on page 1 2      ae Es VERTICAL       CH1       O     CH 2    VERTICAL ZOOM    ap E E      A SCALE  vZ              u    J    CH3    i    CH 4    i    MORE  WAVEFORM OFF    2    MATH   REF       J    Figure 2 11  The Channel Buttons and Lights    Each of the channel  CH  buttons has a light behind its label  Right now  the  CH 1 light is on  That light indicates that the vertical controls are set to adjust  channel 1  Do the following steps to add a waveform to the display     1  Ifyou are not continuing from the previous example  follow the instructions  on page 2 9 under the heading Setting Up for the Examples     2  Press SE
230. lications       m For records with a 100 K length  TDS 500 B and 700A models with Option  1M only   set the trigger position to 5   Use the horizontal position knob to  move the trigger T on screen to the center horizontal graticule line     m Do not use the 15 K  all models  length  nor  if your oscilloscope model is so  equipped  and of the record lengths 30 K  75 K  or 130 K to impulse test  using FFTs  These record lengths do not allow easy alignment of the zero  reference point for phase measurements     m Trigger on the input impulse     Adjust Phase Suppression  Your source waveform record may have a noise  component with phase angles that randomly vary from    pi to pi  This noise  could make the phase display unusable  In such a case  use phase suppression to  control the noise     You specify the phase suppression level in dB with respect to 1 Vrms  If the  magnitude of the frequency is greater than this threshold  then its phase angle  will be displayed  However  if it is less than this threshold  then the phase angle  will be set to zero and be displayed as zero degrees or radians   The waveform  reference indicator at the left side of the graticule indicates the level where phase  is zero for phase FFTs      It is easier to determine the level of phase suppression you need if you first  create a frequency FFT math waveform of the source and then create a phase  FFT waveform of the same source  Do the following steps to use a cursor  measurement to determine the s
231. lling Waveforms  on page 3 118     See Printing a Hardcopy  on page 3 125     Printing a Hardcopy    The TDS Oscilloscope can provide you with hardcopies of its display  To obtain  a hardcopy  you need to know how to configure the communication and  hardcopy parameters of the oscilloscope  how to connect it to one of the many  hardcopy devices it supports  and how to print the hardcopy  This subsection  describes how to do these tasks and how to save a hardcopy to a floppy disk     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 125    Saving Waveforms and Setups       Supported Formats The oscilloscope prints hardcopies of its display in many formats  which allows  you to choose from a wide variety of hardcopy devices  It also makes it easier for  you to place oscilloscope screen copies into a desktop publishing system  The  oscilloscope supports the following formats     m HP Thinkjet inkjet printer   m HP Deskjet inkjet printer   m HP Color Deskjet inkjet printer   m HP Laserjet laser printer   m Epson   m DPU 411 II portable thermal printer   m DPU 412 portable thermal printer   m PCX    PC Paintbrush        PCX Color  PC Paintbrush        TIFF    Tag Image File Format      BMP   Mono  Microsoft Windows file format      BMP   Color  Microsoft Windows file format    m RLE Color  Microsoft Windows color image file format     compressed   m EPS Mono Image  Encapsulated Postscript  mono image   m EPS Color Image  Encapsulated Postscript  color image   m EPS Mono Plot  Encaps
232. llowing  tasks to learn how to adjust a waveform and how to autoset the TDS Oscilloscope     Adjust the Waveform The display shows the probe compensation signal  It is a 1 kHz square wave of  Display approximately 0 5 V amplitude     Figure 2   6 shows the main VERTICAL and HORIZONTAL sections of the front  panel  Each has SCALE and POSITION knobs  Do the following steps to adjust  the size and placement of the waveform using the front panel knobs     1  Turn the vertical SCALE knob clockwise  Observe the change in the  displayed waveform and the channel readout at the bottom of the display     EE VERTICAL A HORIZONTAL   gt  O        VERTICAL ZOOM HORIZONTAL  Z SCALE  lt   SCALE   gt   zZ    WAVEFORM OFF    E  Figure 2 6  The VERTICAL and HORIZONTAL Controls    2  Turn the vertical POSITION knob first one direction  and then the other   Observe the change in the displayed waveform  Then return the waveform to  the center of the graticule     3  Turn the horizontal SCALE knob one click clockwise  Observe the time  base readout at the bottom of the display  The time base should be set to  200 us div now  and you should see two complete waveform cycles on the  display     TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 13    Tutorial       Autoset the Oscilloscope    MAIN LEVEL Knob    AUTOSET Button    When you first connect a signal to a channel and display it  the signal displayed  may not be scaled and triggered correctly  Use the autoset function and you  should quickly ge
233. llowing requirements for obtaining a  good display     m You should scale and position the source waveform so it is contained on  screen   Off screen waveforms may be clipped  which will result in errors in  the integral waveform      m You can use vertical position and vertical offset to position your source  waveform  The vertical position and vertical offset will not affect your  integral waveform unless you position the source waveform off screen so it  is clipped     m When using the vertical scale knob to scale the source waveform  note that it  also scales your integral waveform     The source waveforms that you connect to the oscilloscope often have a DC  offset component  The oscilloscope integrates this offset along with the time  varying portions of your waveform  Even a few divisions of offset in the source  waveform may be enough to ensure that the integral waveform saturates  clips    especially with long record lengths     You may be able to avoid saturating your integral waveform if you choose a  shorter record length   Press HORIZONTAL MENU  gt  Record   Length  main    Reducing the sample rate  use the HORIZONTAL SCALE  knob  with the source channel selected might also prevent clipping  You can also  select AC coupling in the vertical menu of the source waveform or otherwise DC  filter it before applying it to the oscilloscope input     Once you have your waveform optimally displayed  you may magnify  or  reduce  it vertically and horizontally to inspect any
234. lolololfofolk  e      _       e 5  e     ar Ay 0  e des    Peak Detect L _      4     gt     0    y ti v A   v AN y M Ted i Ho  Uses highest and lowest samples in 0  two intervals  Use to reveal aliasing and for glitch detection  P rovides the benefits of enveloping with the speed of a single acquisition  l  leolelelelele  _   ieee  TEE     E l ai ie  esi    t Hi Res      TO  Calculates average of all samples in HoF  interval  TDS 500B and TDS 700A 0  Models Only   Use to reduce apparent noise  Provides the benefits of averaging with the speed of a single acquisition  S el e e e ck       Multiple Waveform Acquisitions  Three Acquisitions from One Source Acquisition Waveform Drawn  Mode on CR  Acquisition 1 2 3  wa  W 5  ae Finds highest and 0  Uses Peak Detect Mode for Each Acquisition lowest record points over TO  eae many acquisitions  Use to reveal variations in the signal across time  Slolololololk                                                                         ae Calculates average value for  Uses Sample Mode for Each Acquisition each record point over    many acquisitions                                                    Seco                      Use to reduce apparent noise in a repetitive signal                                                        Figure 3 13  How the Acquisition Modes Work    3 22 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms                   Average Mode  Average mode lets you acquire and display a 
235. loscopes use only real time sampling  the TDS 500B and  TDS 700A Oscilloscopes use both real  and equivalent time sampling     In real time sampling  the oscilloscope digitizes all the points it acquires after  one trigger event  see Figure 3   11   Always use real time sampling to capture  single shot or transient events     en ee a e  Sampling Rate UUUUUUUUUL    Figure 3 11  Real Time Sampling    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Equivalent Time Sampling A TDS 500B or TDS 700A model oscilloscope  the TDS 600B models use only  real time sampling  uses equivalent time sampling to extend its sample rate over  its real time maximum sampling rate  but only under two conditions     m You must have turned equivalent time on in the Acquisition menu     m You must have set the oscilloscope to a sampling rate that is too fast to allow  it to get enough samples with which to create a waveform record using  real time sampling    If both conditions are the case  the oscilloscope takes a few samples with each   trigger event and eventually obtains enough samples to construct a waveform   record  In short  the oscilloscope makes multiple acquisitions of a repetitive  waveform to obtain the sample density required for a waveform record   See   Figure 3 12   By doing so  the oscilloscope lets you accurately acquire signals    with frequencies much higher than its maximum real time bandwidth would  allow  Equivalent time sampling shoul
236. ls called AUX 1 and AUX 2   instead of CH 3 and CH 4  References to these channels default to CH 3 and  CH 4  if your oscilloscope is one of models TDS 520B  TDS 620B    TDS 680B  TDS 724A  or TDS 782A  read AUX 1 and AUX 2 respectively  for all references to CH 3 and CH 4 in this manual     The TDS 684B display screen appears as the default screen wherever a  display screen is illustrated in this manual     This manual documents the TDS 500B  TDS 600B   amp  TDS 700A Digitizing  Oscilloscopes  Take note of the following conventions used when referencing  these oscilloscopes     The name    TDS 500B    is used when providing information common to the  TDS 520B and TDS 540B model oscilloscopes     The name    TDS 600B    is used when providing information common to the  TDS 620B  TDS 644B  TDS 680B  and TDS 684B model oscilloscopes     The name    TDS 700A    is used when providing information common to the  TDS 724A  TDS 754A  TDS 782A  and TDS 784A model oscilloscopes     In this manual  you will find various procedures which contain steps of  instructions for you to perform  To keep those instructions clear and consistent   this manual uses the following conventions     In procedures  names of front panel controls and menu labels appear in  boldface print     Names also appear in the same case  initial capitals or all uppercase  in the  manual as is used on the oscilloscope front panel and menus  Front panel names  are all upper case letters  for example  VERTICAL MENU and 
237. ls that this manual covers     Table 1 1  Key Features and Differences of Models    Record Lengths  max  To 50K Channel3  InstaVu Acquisitions   Yes o w E Yes   Hi Res Acquisitions Coos e S w Yes  Advanced DSP Math4 Std    Storage  Floppy Disk Std      0 includes RS 232 and Opt 13 Std  Opt  13 Std   Centronics      Two plus Two channel operation allows two of four channels to be displayed simultaneously  Channels not displayed can  be used to couple a triggering signal to the oscilloscope     2 This TDS model can be purchased with Option 1G  which limits the maximum digitizing rate to 1GS second  This option  allows sales in export controlled countries  See Option 1G on page A 1     3 This TDS model can be purchased with Option 1M  which adds longer record length settings  See Option 1M on  page A 2     Advanced digital signal processing provides FFTs  integrals  and differentials of waveforms  See Option 2F on page A 4   5 Std  denotes a standard product feature as opposed to a feature included as part of an option   6 GPIB I O included with all models     Product Specification    The product specification is found in the technical reference TDS 500B   TDS 600B   amp  TDS 700A Technical Reference  Performance Verification and  Specifications  that is shipped as a standard accessory with the TDS Oscillo   scope     1 2 TDS 500B  TDS 600B   amp  TDS 700A User Manual     S  Start Up    Before you use the TDS Oscilloscope  ensure that it is properly installed and  powered on    
238. m  Factory Init  side      TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 25    Tutorial       2 26 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Overview    Acquiring and Displaying  Waveforms    This chapter describes in detail how to perform the operating tasks you must do  to measure  test  process  or save and document your waveforms  It leads with  three sections on the fundamental tasks of acquiring  stably displaying  and  taking measurements on waveforms     m Acquiring and Displaying Waveforms  m Triggering on Waveforms  m Measuring Waveforms    Once you have acquired and measured waveforms  you may want to save and  restore them or the control setups used to acquire and measure them  Or you may  want to save the display screen  complete with waveform and setup information   to include them with the documents you produce with your desk top publishing  system  You may even want to digitally process them  add  multiply  or divide  them  integrate  differentiate or take an FFT of them   The following two topics  cover these tasks     m Saving Waveforms and Setups  m Using Features for Advanced Applications    When performing any operation task  you might want to display a comprehen   sive listing of its current control settings on screen  Or you may find it handy to  display operating information about front panel controls and menus instead of  looking them up in this manual  The following topic tells you how to do both     m Determining Status and Accessing Help
239. m always uses either sin x  x or linear interpolation when expanding  displayed waveforms  To select the interpolation method  press DISPLAY  gt   Setting  main   gt  Display  pop up   gt  Filter  main   gt  Sin x  x or Linear   side      If the source waveform record length is 500 points  the FFT will use 2X Zoom to  increase the 250 point FFT frequency domain record to 500 points  Therefore   FFT math waveforms of 500 point waveforms are always zoomed 2X or more  with interpolation  Waveforms with other record lengths can be zoomed or not  and can have minimum Zooms of 1X or less     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       Sin x  x interpolation may distort the magnitude and phase displays of the FFT  depending on which window was used  You can easily check the effects of the  interpolation by switching between sin x  x and linear interpolation and  observing the difference in measurement results on the display  If significant  differences occur  use linear interpolation     Undersampling  Aliasing  Aliasing occurs when the oscilloscope acquires a source waveform with  frequency components outside of the frequency range for the current sample rate   In the FFT waveform  the actual higher frequency components are under   sampled  and therefore  they appear as lower frequency aliases that    fold back     around the Nyquist point   See Figure 3   92      The greatest frequency that can be input into any sampler without 
240. m for all  measurements requiring those levels  To set the reference levels  do the  following steps     1  Press MEASURE  gt  Reference Levels  main   gt  Set Levels  side  to  choose whether the References are set in   relative to High  100   and Low   0   or set explicitly in the units of the selected waveform  typically volts   See  Figure 3 55  Use the general purpose knob or keypad to enter the values       is the default selection  It is useful for general purpose applications     Units helps you set precise values  For example  if your are measuring  specifications on an RS 232 C circuit  set the levels precisely to RS 232 C  specification voltage values by defining the high and low references in units        Tek Run  100kKS 5 Sample       Reference  Levels               Set Levels    in   units                                           200mv CoM Soone chi 248mV    Remove    eating Migr tow Reference  Measrmnt Setup Snapshot  forchi  Measrmnt ORF Histogram evels    LOJO C C O                                                      Figure 3 55  Measure Menu     Reference Levels    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 97    Measuring Waveforms       2  Press High Ref  Mid Ref  Low Ref  or Mid2 Ref  side    High Ref     Sets the high reference level  The default is 90    Mid Ref     Sets the middle reference level  The default is 50      Low Ref     Sets the low reference level  The default is 10      TE    Mid2 Ref     Sets the middle reference level used on th
241. m of four measurements at a time  To add a  fifth  you must remove one or more of the existing measurements     m To vary the source for measurements  simply select the other channel and  then choose the measurements you want     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       m Becareful when taking automatic measurements on noisy signals  You might  measure the frequency of the noise and not the desired waveform  Your  oscilloscope helps identify such situations by displaying a low signal  amplitude or low resolution warning message        A       Tek Run  100kKS 5 Sample       Select  Measurement                                        Frequency                Positi  widt                      Negative  width             200mv    mM 500us Chif 268mvV    more  p      4             o    see Tighter Gay nt Saad Sy  5 Remove Gating t Reference    Measrmnt Setup Snapshot  irae Vieasrmnt OFF Histogram Levels          LOO  OC  C O                                        CQOoloaolco le             Figure 3 53  Measure Menu    Remove Measurements The Remove Measrmnt selection provides explicit choices for removing  measurements from the display according to their readout position  To remove  measurements  do the following steps     1  Press MEASURE     Remove Measrmnt  main      2  Select the measurement to remove from the side menu  If you want to  remove all the measurements at one time  press All Measurements  side      Gate Measurements The gating feat
242. me difference between the cursors from the A  readout  Read the  time difference between the selected cursor and the trigger point for the  source waveform from the    readout     11  Press Function  main   gt  Paired  side      12  Use the technique just outlined to place the long vertical bar of each paired  cursor to the points along the horizontal axis you are interested in     13  Read the following values from the cursor readouts     m Read the integrated voltage over time between the Xs of both paired  cursors in volt seconds from the A  readout     m Read the integrated voltage over time between the X of the selected  cursor and the reference indicator of the math waveform from the     readout     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       To Take Automated  Measurements    Offset  Position  and Scale    DC Offset    Using Zoom    m Read the time difference between the long vertical bars of the paired  cursors from the A  readout     You can also use automated measurements to measure integral math waveforms   Use the same procedure as is found under To Take Automated Measurements on  page 3 169  When using that procedure  note that your measurements on an  integral waveform will be in volt seconds rather than in volts per second as is  indicated for the differential waveform measured in the procedure     When creating integrated math waveforms from live channel waveforms   consider the following topics  Note the fo
243. measurement  The true Root Mean Square voltage over the entire waveform or gated    region        TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 93    Measuring Waveforms       Measurement Readouts    Measurement Readout Area    With no menus displayed  the measurement readouts appear far right of the  display graticule   See Figure 3 52   You can display and continuously update as  many as four measurements at any one time  With any menu displayed  the  readouts move to the right side of the graticule area        va       Tek Run  100kKS 5 Sample          chi Period  1 000ms          chi  width  E 500s       Display Measurements    3 94            chi  Duty  50 0            chi  Over  6 5                       oloooo     i om a 0ps Chi F 208m                                           OIC   COlC  e       Figure 3 52  Measurement Readouts    Measurement 1 is the top readout  Measurement 2 is below it  and so forth  Once  a measurement readout is displayed in the screen area  it stays in its position  even when you remove any measurement readouts above it     To use the automatic measurements you first need to obtain a stable display of  the waveform to be measured   Pressing AUTOSET may help   Once you have a  stable display  press MEASURE to bring up the Measure menu   See   Figure 3 53      1  Press MEASURE  gt  Select Measrmnt  main      2  Select a measurement from the side menu  Note the following rules for  taking automatic measurements     m You can only take a maximu
244. ments is a pop up display that covers  about 80  of the graticule area when displayed   See Figure 3   57   You can  display a snapshot on any channel or ref memory  but only one snapshot can be  displayed at a time     To use Snapshot  obtain a stable display of the waveform to be measured   pressing AUTOSET may help   Then do the following steps     1  Press MEASURE     SNAPSHOT  main    2  Press either SNAPSHOT  main  or AGAIN  side  to take another snapshot        NOTE  The Snapshot display tells you the channel that the snapshot is being  made on        3  Push Remove Measrmnt     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 99    Measuring Waveforms       3 100        aaa Sanne  b       a T   m      ae    snapshot on Si 4  Period    1 000ms Freq  1 00008kKH2 j  Snapshot Display      Width   500ps  Width   500s              Tek Run  100kKS 5 Sample                                      Over 14 8    Over 34 8   High  512m   Low  SmvV  Max  536m   Min   16mVv  Ampl  504mvV Pk Pk    552m            i Mean  255 6m   cycMean  252 0mv  RMS  359 2mv CYCRMS   356 4mV hv    Area  1 27528mVs CycArea   251  80yVs                   Brstwd   4 490ms ste  Rise   BMS Fall   BUS   Duty   50 0   Duty   50 0                 elec 7 igh Low    Remove Gating Reference   Measrmnt Setup Snapshot  forech   Measrmnt OFF Histogram Levels   Snapshot                              Slao       Figure 3 57  Snapshot Menu and Readout    Consider the following rules when taking a snapshot     Be sure t
245. menu  or the front  panel trigger LEVEL knob   press TRIGGER MENU  gt  Type  main   gt  Pulse   pop up      Level  main      Level  Set to TTL  Set to ECL  or Set to 50    side      m If you select Level  you set the trigger level by entering a value with the  general purpose knob or the keypad     m If you select Set to TTL  the oscilloscope sets the trigger level to the TTL  switching threshold     m If you select Set to ECL  the oscilloscope sets the trigger level to the ECL  switching threshold     m If you select Set to 50   the oscilloscope searches for the point halfway  between the peaks of the trigger source signal and sets the trigger level to  that point     To Set Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer to To Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see the descriptions  Trigger Modes and Trigger Holdoff on page 3 53     3 76 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       To Trigger ona Runt Pulse When you select the pulse class Runt  the oscilloscope will trigger on a short  pulse that crosses one threshold but fails to cross a second threshold before  recrossing the first  To set up for runt triggering  do the following procedures     Select Runt Triggering  Press TRIGGER MENU  gt  Type  main   gt  Pulse   pop up      Class  main   gt  Runt  pop up    See Figure 3   45      Select the Source  To specify whi
246. mes  were acquired  they are displayed  If no frames were acquired  the previous  FastFrame waveform is displayed     Because FastFrame waveforms contain many triggers  trigger position  indicators are removed from both the waveform and the record view when  the selected channel  reference  or math waveform is a FastFrame waveform     In Equivalent Time  the oscilloscope ignores FastFrame mode     Because FastFrame introduces additional processing time into the operation  cycle of acquire  process  and display  its best to use Single Sequence  Acquisition  see Acquire menu  Stop After menu   With Single Sequence  selected  you will see the current acquisition sequence  otherwise  the display  lags the current sequence by one sequence  You can also see the current  sequence by pressing the RUN STOP button to stop the acquisition     TDS 500B  TDS 600B   amp  TDS 700A User Manual    a ery  Triggering on Waveforms    Triggering Concepts    To use the TDS Oscilloscope to measure or monitor waveforms  you need to  know how to trigger a stable display of those waveforms  Toward that end  this  section first covers the following topics     m Trigger Concepts which details some basic principles of triggering and  describes triggering elements  type  source  coupling  holdoff  mode   and so on    m Triggering from the Front Panel which describes how to use the front panel  triggering controls each of which is common to most  if not all  the trigger  types the oscilloscope provides    
247. mes the clock input  leaving the remaining  channels as logic inputs     Setup hold triggering uses the setup and hold times to define a    setup hold  violation zone    relative to the clock  Data that changes state within this zone  triggers the oscilloscope  Figure 3 38 shows how the setup and hold times you  choose positions this zone relative to the clock     Setup hold triggering uses the setup hold violation zone to detect when data is  unstable too near the time it is clocked  Each time trigger holdoff ends  the  oscilloscope monitors the data and clock sources  When a clock edge occurs  the  oscilloscope checks the data stream it is processing  from the data source  for  transitions occurring within the setup hold violation zone  If any occur  the  oscilloscope triggers with the trigger point located at the clock edge     Positive settings for both setup and hold times  the most common application   locate the setup hold violation zone so it spans the clocking edge   See the top  waveform in Figure 3   38   The oscilloscope detects and triggers on data that    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       does not become stable long enough before the clock  setup time violation  or  that does not stay stable long enough after the clock  hold time violation                                          Setup Hold   Ts  Setup Time  lt        Violation        gt     Ty  Hold Time Zone   Setup Hold Violation Zone  Ts  T4 ts    4   Ts  Ty must
248. mit adjusts the  other time to maintain the limit        5         Tek Run  5 00GS 5 Average             Hold Time  30 0ns  amp   3   30   i    Setup Hold  Times                                     E aoe  Setup Time     i fr  72 0ns     C1 C2 Dly E  Ton fo fA 10 54ns 7  gold ie    4                   1 00 ve M 10 0ns  StHid          ch2                LOJO  C O o     oh i  Source    Class   sour   Kag pane   lt StHid gt  Cha e Hotioft   11911  91              Type   lt Logic gt                                               Figure 3 42  Triggering on a Setup Hold Time Violation    In most cases  you will enter positive values for both setup and hold time   Positive values set the oscilloscope to trigger if the data source is still settling  inside the setup time before the clock or if it switches inside the hold time after  the clock  You can skew this    setup hold violation zone    that the setup and hold  times form by entering negative values  See Figure 3   38 on page 3 65     To Set Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer to To Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see the descriptions  Trigger Modes and Trigger Holdoff on page 3 53     The TDS Oscilloscope can trigger on glitch or runt pulses  or it can trigger based  on the width  slew rate  or timeout period of a pulse  These capabilities make the  oscilloscope suitable for such tasks as unat
249. n   gt  Setup Hold  pop up   gt  Levels  main   gt  Set Both to TTL  or Set Both to ECL  side      The oscilloscope uses the clock level you set to determine when a clock edge   rising or falling  depending on which you select  occurs  The oscilloscope uses  the point the clock crosses the clock level as the reference point from which it  measures setup and hold time settings     Set the Setup and Hold Times  To set the setup time and the hold time relative to  the clock     1  Press TRIGGER MENU  gt  Type  main   gt  Logic  pop up   gt   Class  main   gt  Setup Hold  pop up   gt  Set Hold Times  main   gt  Setup  Time or Hold Time  side   See Figure 3   42     2  Turn the general purpose knob or use the keypad to set values for the setup  and for the hold times        NOTE  Keep hold time to at least 2 ns less than the clock period or the oscillo   scope cannot trigger        TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 71    Triggering on Waveforms       Cursors measure the setup hold  violation zone which equals setup  time   hold time  30 ns      Data  Ch1  transition occurs  within   10 ns after the clock  violating hold time limit     The oscilloscope recognizes the  violation and triggers at the clock edge     Triggering on Pulses    3 72    Positive setup time always leads the clock edge  positive hold time always  follows the clocking edge  Setup time always leads the hold time by at least 2 ns   Ts   Ty 22 ns   Attempting to set either time to reduce the 2 ns li
250. n exchange for the  defective product     In order to obtain service under this warranty  Customer must notify Tektronix of the defect before the expiration of the  warranty period and make suitable arrangements for the performance of service  Customer shall be responsible for  packaging and shipping the defective product to the service center designated by Tektronix  with shipping charges prepaid   Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the  Tektronix service center is located  Customer shall be responsible for paying all shipping charges  duties  taxes  and any  other charges for products returned to any other locations     This warranty shall not apply to any defect  failure or damage caused by improper use or improper or inadequate  maintenance and care  Tektronix shall not be obligated to furnish service under this warranty a  to repair damage resulting  from attempts by personnel other than Tektronix representatives to install  repair or service the product  b  to repair  damage resulting from improper use or connection to incompatible equipment  or c  to service a product that has been  modified or integrated with other products when the effect of such modification or integration increases the time or  difficulty of servicing the product     THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY  OTHER WARRANTIES  EXPRESSED OR IMPLIED  TEKTRONIX AND ITS VENDORS 
251. n information on any front panel button  knob or  menu item   See Figure 3   81         Help Mode is on     Push HELP to exit     Press any button or turn any Knob for information about that  control  Changing a control while in help mode will not affect  the digitizing oscilloscope settings  Press HELP again to exit the  help mode     Press the SHIFT button before pressing front   panel buttons  that display help screens for shifted menus  The shifted menus  are the Acquire menu  Application menu  Hardcopy menu   Delayed Trigger menu  Status menu  and Utility menu              With Help off  press SH  FT  lighted when on  before turning       the general purpose  horizontal position  or vertical position  knobs when you want increased knob response  Less knob  rotation will produce greater changes in settings           with Help off  press SHIFT after pressing any menu button that  toggles through multiple settings  such as those for pop up  menus   Pressing SHiF7 reverses the direction in which the pop  up menu or button label toggles through the menu or    parameter settings     Sai  ooma M 500yus Chi   OV 24 Jul 1995  14 01 31              OIOI                                                    aa          Figure 3 81  Initial Help Screen    When you press that button  the instrument changes mode to support on line  help  Press HELP again to return to regular operating mode  Whenever the  oscilloscope is in help mode  pressing any button  except HELP or SHIFT    turning a
252. nce  Color menu  3   37  Math waveform  derivative  See Derivative math waveform  FFT  See FFT math waveform  integral  See Integral math waveform  Math waveforms  3   150  Disallowed in InstaVu  3   48  Math  Color menu  3   37  Math1 2 3  More menu  3 150  Maximum  3 93  Glossary   6  Mean  3   93  Glossary   7  MEASURE button  3   94  Measure Delay menu  Create Measrmnt  3   99  Delay To  3 98  Edges  3 99  Measure Delay To  3   98  OK Create Measurement  3   99  Measure Delay To  Measure Delay menu  3   98  Measure menu  3   94  3   99  Gating  3 95  High Ref  3 98  High Low Setup  3   96          TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index       Histogram  3   96   Low Ref  3   98   Mid Ref  3   98   Mid2 Ref  3   98   Min Max  3   96   Reference Levels  3   97  Remove Measrmnt  3   95  3   99  Select Measrmnt  3   94  3   98  Set Levels in   units  3   97  Snapshot  3   99    Measurement    Amplitude  3 92  Glossary   1  Area  3   92  Glossary   1   Burst width  3 92  Glossary   2  Cycle area  3   92  Glossary   3  Cycle mean  3 92  Glossary   3  Cycle RMS  3 92  Glossary   3  Delay  3   98  Glossary   3   Duty cycle  2 20  Glossary   7  Glossary   8  Fall time  3   92   Frequency  2 19  3   92  Glossary   4  Gated  Glossary   4   High  3   92  Glossary   5   Low  3   93  Glossary   6  Maximum  3 93  Glossary   6  Mean  3   93  Glossary   7  Minimum  3   93  Glossary   7  Negative duty cycle  3   93  Negative overshoot  3   93  Negative width  3   93   Overs
253. nd differences  1 2  Monochrome  Color menu  3   35  MORE button  3   9  3 120  3 145  3 148  More menu  3 120  3 148  3 168   Average  3 150   Blackman Harris  3 154   Change Math waveform definition  3 152  3 168    3 172   dBV RMS  3 153   diff  3 168   Dual Wfm Math  3 149   FFT  3 152   Hamming  3 154   Hanning  3 154   intg  3 172   Linear RMS  3 153   Math1  Math2  Math3  3 152  3 168  3 172   Math1 2 3  3 150   No Process  3   150   OK Create Math Waveform  3 149  3 172             Index 9    Index       Phase  deg   3 153   Phase  rad   3   153   Rectangular  3 153   Reference waveform status  3 120  Set 1st Source to  3 149   Set 2nd Source to  3 149   Set FFT Source to   3 152   Set FFT Vert Scale to   3 153   Set FFT Window to   3 153   Set Function to  3 149   Set Function to   3   168  3 172   Set operator to  3 149   Set Single Source to  3 148  3 149  Set Single Source to   3 168  3 172  Single Wfm Math  3 148  3 168  3 172    N    NAND  Glossary   7  NAND  Main Trigger menu  3   68  3 70  Negative duty cycle  3 93  Negative overshoot  3   93  Negative width  3   93  Negative  Main Trigger menu  3   76  3 77  3 80  No Process  More menu  3   150  Noise   reducing in FFTs  3 160   reducing in phase FFTs  3 154  3 163  Noise Rej  Main Trigger menu  3   60  NOR  Glossary   7  NOR  Main Trigger menu  3 68  3   70  Normal trigger mode  3   53  Glossary   7  Normal  Color menu  3 35  Normal  Main Trigger menu  3 61  NTSC  Display menu  3 33  Nyquist frequency  3 161
254. nding on the selection for Horizontal Lock in the side menu  Zoom  affects the displayed waveforms as follows     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       None     only the waveform currently selected can be magnified and  positioned horizontally  Figure 3 23      Live     all    live     as opposed to reference  waveforms can be magnified and  positioned horizontally at the same time  If a reference or math waveform is  selected and Horizontal Lock set to Live  only the selected reference or math  waveform is magnified and positioned     All     all waveforms displayed  live  math  and reference  can be magnified  and positioned horizontally at the same time        Tek Run  100kS s Sample  a    Chl Zoom     1 0X Vert 2 0X Horz Ki T Horizontal  i Lock                      Only the selected A  waveform  the top one    ei  changes size  cL def Ppa ok    Le oak toate aa Meck Gane sa la Ie   Jive          None                         2oomv SM 250s Chi   264mv    oe na iiri iaaa  Mode lock Graticule Reset     Upper     11      1                                                        Figure 3 23  Zoom Mode with Horizontal Lock Set to None    5  Press ZOOM  gt  Lock  main   gt  AIl  Live  or None  side         NOTE  Although Zoom must be turned on to control which waveforms Zoom  affects  the setting for Horizontal Lock affects which waveforms the horizontal  control positions whether Zoom is on or off  The rules for the three settin
255. ng Zoom    Waveform Integration    clipped portion  Also  the automated measurement Pk Pk will display a clipping  error message if turned on  see To Take Automated Measurements on  page 3 169      If your derivative waveform is clipped  try either of the following methods to  eliminate clipping     m Reduce the size of the source waveform on screen   Select the source channel  and use the vertical SCALE knob      m Expand the waveform horizontally on screen   Select the source channel and  increase the horizontal scale using the horizontal SCALE knob   For  instance  if you display the source waveform illustrated in Figure 3   95 on  page 3 169 so its rising and falling edges are displayed over more horizontal  divisions  the amplitude of the corresponding derivative pulse will decrease     Whichever method you use  be sure Zoom is off and the zoom factors are reset   see Using Zoom below      Once you have your waveform optimally displayed  you can also magnify  or  contract  it vertically and horizontally to inspect any feature  Just be sure the  differentiated waveform is the selected waveform   Press MORE  then select the  differentiated waveform in the More main menu  Then use the Vertical and  Horizontal SCALE knob to adjust the math waveform size      If you wish to see the zoom factor  2X  5X  etc    you need to turn zoom on   press ZOOM  gt  ON  side   The vertical and horizontal zoom factors appear on  screen     Whether zoom is on or off  you can press Reset Zoom 
256. ng into  Service cse teeri ai e eee E E toed E ea 1   4  Operating Basics  OVERVIEW eose eroate o esa E E Ea E EE E 2 1  Operating Interface Maps       ssesssesssossosssossesossoess 2 3  T  torial sees ectnars sence nesi we wee eae en ee Rees E E E SOR E 2 9  Setting Up for the Examples              0 0    cece cee eee ene 2 9  Example 1  Displaying a Waveform           2 00    cece eee eee ee eee 2 13  Example 2  Displaying Multiple Waveforms               0 2 00  ee eee eee 2 15  Example 3  Taking Automated Measurements                 0 00 eee eee 2 19  Example 4  Saving Setups               0  ccc cece ect e eee n en eee 2 24  Reference  OVERVIOW ieee eg 5 SR ee PEGS eee SOs hee RESORTS 3 1  Acquiring and Displaying Waveforms                eeeeeeeees 3 3  Coupling Waveforms to the Oscilloscope           2 0 0 0 cee eee eee eee ee 3 3  Setting up Automatically  Autoset and Reset                0 0000 e eens 3 6  Selecting Channels          0    eee een eee een eens 3 8  Scaling and Positioning Waveforms             00    cee eee ee eee eee 3 10  Choosing an Acquisition Mode               0 0 c eee c eee eens 3 17  Customizing the Display           2 0    cece eee cece teens 3 29  Customizing the Display Color               0 00  eee eee eee nee 3 34  Zooming on Waveforms sess seeria cece eee eee eens 3 39  Using InstaVu   Acquisition Mode  TDS 500B and 700A Models Only        3 45  Using FastFrame    TDS 500B and 700A Models Only                     3   48  Triggering on 
257. nnel selection buttons and that  the integrated math waveform is selected in the More main menu     2  Press CURSOR     Mode  main   gt  Independent  side   gt  Func   tion  main   gt  H Bars  side      3  Use the general purpose knob to align the selected cursor  solid  to the top   or to any amplitude level you choose      4  Press SELECT to select the other cursor     5  Use the general purpose knob to align the selected cursor  to the bottom  or  to any amplitude level you choose      6  Read the integrated voltage over time between the cursors in volt seconds  from the A  readout  Read the integrated voltage over time between the  selected cursor and the reference indicator of the math waveform from the     readout   See Figure 3   98      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 173    Using Features for Advanced Applications       3 174    Integral Math Waveform    Source Waveform    Se N    Tek Run  5 00MS s Sample          Cursor  Function                                        M 10 045 Chis 520m  v                                                          ceS       Figure 3 98  H Bars Cursors Measure an Integral Math Waveform    7  Press Function  main   gt  V Bars  side   Use the general purpose knob to  align one of the two vertical cursors to a point of interest along the horizon   tal axis of the waveform     8  Press SELECT to select the alternate cursor   9  Align the alternate cursor to another point of interest on the math waveform     10  Read the ti
258. nu  3 136  Portrait  Hardcopy menu  3 128  Position  Vertical  3   12  vertical  3 159  3 170  3 175  Position  Vertical menu  3   13  Positive duty cycle  3 93  Positive overshoot  3   93  Positive width  3 93  Positive  Main Trigger menu  3   76  3 77  3 80  Postscript  3 126  Posttrigger  Glossary   8  Power connector  14  2   5  Power cords  A 1    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index       Power off  1   6  Power on  1 5  Pretrigger  Glossary   8  Preview  Zoom menu  3   42  Principal power switch  1 5  2 5  Print  File Utilities menu  3 124  Printer  Color   Option 3I and 3P  A 4  Printing a Hardcopy  Reference Article  3 125  Probe Cal  3 108  Probe usage  1 3  Probe   channel deskew  3 108  Glossary   2  Probes  Accessories  A   5  Active voltage  E 3  Compensation  2   15  Glossary   9  Compensation of  3 4  Compensation Signal  2 13  Connection  2   9  Current  E4  Definition  Glossary   9  Differential active  E 3  Fixtured active  E 3  General purpose  high input resistance   E   1  High speed  E 3  High voltage  E 2  Low impedance Zo  E 2  Optical  E 5  Option 20 to add  A 2  Option 21 to add  A 2  Option 22 to add  A 3  Option 23 to add  A 3  Option 24 to add  A 3  Option 26 to add  A 3  Option 27 to add  A 3  Option 28 to add  A 3  Option 2D to delete  A 3  Option 4D to delete  A 4  P6205 Active  1 3  Passive  3   4  Passive voltage  E   1  Selection  E   1  Time to voltage converter  E 5  Procedures  Inspection and Cleaning  F       F   2  P
259. ny knob  or pressing any menu item displays help text on the screen that  discusses that control     The menu selections that were displayed when HELP was first pressed remain on  the screen  On line help is available for each menu selection displayed at the  time the HELP button was first pressed  If you are in help mode and want to see  help on selections from menus not displayed  you first exit help mode  display  the menu you want information on  and press HELP again to re enter help mode     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 141    Determining Status and Accessing Help       3 142 TDS 500B  TDS 600B   amp  TDS 700A User Manual    ee ey  Using Features for Advanced Applications    The TDS Oscilloscope provides powerful features for testing and digitally proces   sing the waveforms you acquire  This section describes how to use the following  features     Limit Testing     for testing the waveforms you acquire against a template  you create  on this page     Waveform Math     for inverting  adding  subtracting  and multiplying of  waveforms  see page 3 148     Fast Fourier Transforms     for displaying the frequency content of  waveforms  see page 3 150     Waveform Differentiation     for displaying the derivative of a waveform   see page 3 167     Waveform Integration     for displaying the integral of a waveform  see  page 3 171        NOTE  If InstaVu mode is on  the features listed above are not available  press  InstaVu on the front panel to toggle Ins
260. o Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see Trigger Modes  and Trigger Holdoff on page 3 53     When you select the pulse class Slew Rate  the oscilloscope will trigger on a  pulse edge that traverses between an upper and lower threshold faster or slower  than a slew rate you specify  To set up for slew rate triggering  do the following  procedures     Select Slew Rate Triggering  Press TRIGGER MENU  gt  Type  main   gt  Pulse   pop up   gt  Class  main   gt  Slew Rate  pop up    See Figure 3   46 on  page 3 82      Select the Source  Press TRIGGER MENU     gt  Type  main   gt  Pulse   pop up      Source  main   gt  Ch1  Ch2  Ch3  or Ch4  side   The source  selected becomes the trigger source for all four trigger classes     Select Polarity  To specify the direction of the pulse edge  press TRIGGER  MENU  gt  Type  main   gt  Pulse  pop up   gt  Class  main   gt  Slew Rate   pop up   gt  Polarity  main   gt  Positive  Negative  or Either  side      Positive monitors the slew rate of the positive going edges of pulses  The edge  must first cross the lower threshold and then cross the upper threshold     Negative monitors the slew rate of the negative going edges of pulses  The edge  must first cross the upper threshold and then cross the lower threshold     Either monitors positive  and negative going edges of pulses  The edge may  first cross either threshold and then cross the other     Set the Slew Rate  The threshold levels and
261. o display the waveform properly before taking a snapshot  Snapshot  does not warn you if a waveform is improperly scaled  clipped  low signal  amplitude  low resolution  etc       To vary the source for taking a snapshot  simply select another channel   math  or ref memory waveform and then execute snapshot again     Note that a snapshot is taken on a single waveform acquisition  or acquisi   tion sequence   The measurements in the snapshot display are not continu   ously updated     Be careful when taking automatic measurements on noisy signals  You might  measure the frequency of the noise and not the desired waveform     Note that pushing any button in the main menu  except for Snapshot  or any  front panel button that displays a new menu removes the snapshot from  display     Use High Low Setup  page 3   96   Reference Levels  page 3   97   and   Gated Measurements  page 3   95  with Snapshot exactly as you would  when you display individual measurements from the Select Measrmnt  menu     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms       To Find More Information    To perform a tutorial that shows you how to take automatic measurements  see  Example 3  Taking Automated Measurements on page 2 19     To learn how the oscilloscope calculates each automatic measurement  see  Appendix B  Algorithms on page B 1     Taking Cursor Measurements    Cursor Types          Horizontal Bar Cursors    Figure 3 58  Cursor Types    The TDS Oscilloscope provides cu
262. o rise from a  LowRef value  default   10   to a HighRef value  default   90       Figure B   3 shows a rising edge with the two crossings necessary to calculate a  Rise Time measurement     1  Searching from Start to End  find the first sample in the measurement zone  less than LowRef     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix B  Algorithms       RMS     2  From this sample  continue the search to find the first  positive  crossing of  LowkRef  The time of this crossing is the low rise time or TLR   Use linear  interpolation if necessary      3  From TLR  continue the search  looking for a crossing of HighRef  Update  TLR if subsequent LowRef crossings are found  If a HighRef crossing is  found  it becomes the high rise time or THR   Use linear interpolation if  necessary      4  RiseTime   THR   TLR                      Rise Time   TLR THR  High  HighR ef  LowR ef  Low    Figure B 3  Rise Time    Amplitude  voltage  measurement  The true Root Mean Square voltage   If Start   End then RMS   the  interpolated  value at Waveform Start      Otherwise        End     Waveform t   dt  RM  S   Start        End     Start  x SampleInterval    For details of the integration algorithm  see below     TDS 500B  TDS 600B   amp  TDS 700A User Manual B 11    Appendix B  Algorithms       Integration Algorithm    The integration algorithm used by the oscilloscope is as follows     2 is approximated by f    where     W t dt   W t dt  A    A    W t  is the sampled waveform  
263. often determine the best window empirically by first using the window  with the most frequency resolution  rectangular   then proceeding toward that  window with the least  Blackman Harris  until the frequencies merge  Use the  window just before the window that lets the frequencies merge for best compro   mise between resolution and amplitude accuracy        NOTE  If the Hanning window merges the frequencies  try the Hamming window  before settling on the rectangular window  Depending on the distance of the  frequencies you are trying to measure from the fundamental  the Hamming  window sometimes resolves frequencies better than the Hanning        TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 165    Using Features for Advanced Applications       3 166    Window Characteristics  When evaluating a window for use  you may want to  examine how it modifies the FFT time domain data  Figure 3   94 shows each  window  its bandpass characteristic  bandwidth  and highest side lobe  Consider  the following characteristics     m The narrower the central lobe for a given window  the better it can resolve a  frequency     m The lower the lobes on the side of each central lobe are  the better the  amplitude accuracy of the frequency measured in the FFT using that  window     m Narrow lobes increase frequency resolution because they are more selective   Lower side lobe amplitudes increase accuracy because they reduce leakage     Leakage results when the time domain waveform delivered to 
264. og Signal  Sample  and Digitize   3 18  Figure 3 10  Several Points May be Acquired for Each Point Used   3 18    Figure 3 11  Real Time Sampling               ccce cess cceees 3 18  Figure 3 12  Equivalent Time Sampling               eceeeeeee 3 19  Figure 3 13  How the Acquisition Modes Work                   3 22    TDS 500B  TDS 600B   amp  TDS 700A User Manual iii    Table of Contents          Figure 3 14  Acquisition Menu and Readout                  66  3 24  Figure 3 15  Acquire Menu     Stop After                 eee eee 3 27  Figure 3 16  Aliasing            05 2 cccnses one cee ee creates cee eee ee 3 28  Figure 3 17  Display Menu     Style              ccc eee ee ee eeee 3 31  Figure 3 18  Trigger Point and Level Indicators                   3 32  Figure 3 19  Display Menu     Setting                cece eee eee 3 35  Figure 3 20  Display Menu     Palette Colors                 66  3 37  Figure 3 21  Display Menu     Map Reference Colors              3 38  Figure 3 22  Display Menu     Restore Colors                0065 3 39  Figure 3 23  Zoom Mode with Horizontal Lock Set to None         3 41  Figure 3 24  Dual Window  Preview  Mode               seeee8  3 43  Figure 3 25  Dual Zoom     Shown Dual Window  Preview  Mode   3 44  Figure 3 26  Normal DSO Acquisition and Display Mode Versus   Trista Vu Mod   2 eis vee o ic vince sencese He ee hee ends oeio oe ee 3   46  Figure 3 27  InstaVu Display       essssssssseseseseseseseseeo 3 47  Figure 3 28  Fast Frame       
265. ommunication Parameters on  page 3 127        5  Type  IBRDF  lt Filename gt  where  lt Filename gt  is a valid DOS file name with  which you want to label your hardcopy file  It should be  lt 8 characters long  and up to a 3 character extension  For example  you could type    ibrdf  screenl        6  Exit the IBIC program by typing  EXIT    7  Copy the data from your file to your hardcopy device  Type   COPY  lt Filename gt   lt Output port gt   lt  B gt  where      lt Filename gt  is the name you defined in step 5 and     lt Output port gt  is the PC output port your hardcopy device is connected to   such as LPT1 or LPT2      For example  to copy  print  a file called screen  to a printer attached to the   pt  parallel port  type    copy screen  Iptl   B        Your hardcopy device should now print a picture of the oscilloscope screen     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 133    Saving Waveforms and Setups          NOTE  If you transmit hardcopy files across a computer network  use a binary   8 bit  data path        Communicating with Remote Instruments    The TDS Oscilloscope can connect into a system environment  so that you can  control it remotely or exchange measurement or waveform data between it and a  computer  This subsection explains how to prepare and setup the oscilloscope for  control and operation over the IEEE Std 488 2 1987  GPIB  interface     To Prepare for Remote To transfer data between the oscilloscope and other instruments over the GPIB 
266. onfigurations     GPIB Device GPIB Device  GPIB Device  GPIB Device    Figure 3 75  Typical GPIB Network Configuration    GPIB Device                       GPIB Device        GPIB Device    Obtain the Proper Interconnect Cabling  To connect the oscilloscope to a GPIB  network  obtain at least one GPIB cable  Connecting two GPIB devices requires  an IEEE Std 488 1 1987 GPIB cable  available from Tektronix  part number  012 0991 00      The standard GPIB cable connects to a 24 pin GPIB connector located on the  rear panel of the oscilloscope  The connector has a D type shell and conforms to  IEEE Std 488 1 1987  You can stack GPIB connectors on top of each other   See  Figure 3 76                                 Figure 3 76  Stacking GPIB Connectors    To Set Up for Remote To set up remote communications  be sure your setup will meet GPIB protocol  Operation and interface requirements just described  Then do the following procedures     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 135    Saving Waveforms and Setups       Connect the Oscilloscope to the GPIB  To connect the oscilloscope  plug an IEEE  Std 488 2 1987 GPIB cable into the GPIB connector on the oscilloscope rear  panel and into the GPIB port on your controller   See Figure 3 77                                                                                                         Controller  Digitizing Oscilloscope  Rear P anel   ra Eo    Io  amp    OCIO        GPIB Connector       Figure 3 77  Connecting the Os
267. onix Holland N V   Marktweg 73A   8444 AB Heerenveen  The Netherlands    declare under sole responsibility that the    TDS 620B  TDS 644B  TDS 680B  and TDS 684B Digitizing Oscilloscopes    meet the intent of Directive 89 336 EEC for Electromagnetic Compatibility and Low  Voltage Directive 73 23 ECC for Product Safety  Compliance was demonstrated to the  following specifications as listed in the Official Journal of the European Communities     EMC Directive 89 336 EEC    EN 55011 Class B Radiated and Conducted Emissions  EN 50081 1 Emissions    EN 60555 2 AC Power Line Harmonic Emissions    EN 50082 1 Immunity     IEC 801 2 Electrostatic Discharge Immunity   IEC 801 3 RF Electromagnetic Field Immunity  IEC 801 4 Electrical Fast Transient Burst Immunity  IEC 801 5 Power Line Surge Immunity          ca af    Table of Contents    General Safety Summary              ccc ccc c cece cece ecscees ix  Preface fsb ooren ORR ee BA RKEY NE Rees eee Ree EEEO xi  Related  Manuals s     0022 400  iiew ese eee ee eed eee ses ddan xi  Default Model  ri ranee at pea ae eile petit baie a aiin e xii  Model References    peri a ek ee bra ee aai oa iaaii xii  Conventions Sarrea en A a ae eae ed ee a xii  Getting Started  Product Description          sssssssssesssesossosesseeesossoo 1 1  Differences by Model EE ES ETE T ES E 1 2  Product Specification seais arnas A elke S A ae Aa OA 1 2  Start UP sco bake EEEE E E EE E E OR E RE S 1 3  Preparation  oeo t e pick See E ene A E E E A wet E 1 3  Putti
268. ope  t0                            Average  16          M 500s Chis  256mV                                              oloo a       Figure 3 14  Acquisition Menu and Readout    3 24    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Selecting an The oscilloscope provides several modes  see The Acquisition Modes on  Acquisition Mode page 3 21  for acquiring and converting analog data into digital form  To choose  how the oscilloscope will create points in the waveform record     1  Press SHIFT ACQUIRE MENU  gt  Mode  main    See Figure 3   14    2  TDS 600B  Press Sample  Envelope  Average  or Peak Detect  side  or        TDS 500B and TDS 700A Models  Press Sample  Peak Detect  Hi Res   Envelope  or Average  side    InstaVu mode must be off to use Hi Res   Envelope  or Average modes      3  If you selected Envelope or Average  enter the number of waveform records  to be enveloped or averaged using the keypad or the general purpose knob        NOTE  TDS 500B and 700A models only  Selecting Hi Res mode in the Acquire  menu automatically reduces long record length settings to prevent overflow of  acquisition memory  Because Hi Res mode uses twice the acquisition memory  used by other acquisition modes  allowing selection of the longer horizontal  record lengths with Hi Res mode would cause the oscilloscope to run out of  memory        Selecting Repetitive TDS 500B and TDS 700A models only  To limit the oscilloscope to real time  Samplin
269. ope will trigger on a  Pulse Timeout pulse change that does NOT occur within the specified limits  That is  the trigger  will occur when  depending on the polarity you select  the signal stays higher or  stays lower than the trigger level for the timeout value  To set up for timeout  triggering  do the following procedures     Select Timeout Triggering  Press TRIGGER MENU     gt  Type  main   gt  Pulse   pop up      Class  main   gt  Timeout  pop up      Select the Source  Press TRIGGER MENU     gt  Type  main   gt  Pulse   pop up      Source  main   gt  Ch1  Ch2  Ch3  or Ch4  side   The source  selected becomes the trigger source for all four trigger classes     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 83    Triggering on Waveforms          Delayed Triggering    3 84    Select the Polarity  Press TRIGGER MENU     Type  main   gt  Pulse   pop up   gt  Class  main   gt  Timeout  pop up   gt  Polarity  main   gt  Stays  High  Stays Low  or Either  side      Stays High causes a trigger to occur if the signal stays higher than the  trigger level for longer than the timeout value     Stays Low causes a trigger to occur if the signal stays lower than the trigger  level for longer than the timeout value     Either causes a trigger to occur if the signal stays lower or stays higher than  the trigger level for longer than the timeout value     Time  To set the timeout time     1  Press TRIGGER MENU  gt  Type  main   gt  Pulse  pop up   gt   Class  main   gt  Timeout  pop up   
270. opy Device    Connect to a Hardcopy Device  To connect the oscilloscope directly to a hardcopy  device  determine which interface and cable the device uses  and connect  accordingly   See Figure 3 73      Some devices  such as the Tektronix HC100 Plotter  use the GPIB interface   Many printers  such as the Tektronix HC200  use Centronics interfaces  Many  hardcopy devices  including the HC100 and HC200 with option 03  provide  RS 232 support   Check the documentation for your hardcopy device      Digitizing Oscilloscope Hardcopy Device                      GPIB  RS 232   or Centronics Cable    Figure 3 73  Connecting the Oscilloscope Directly to the Hardcopy Device    Print  To print a single hardcopy or send additional hardcopies to the oscilloscope  spool  queue  while waiting for earlier hardcopies to finish printing  press  HARDCOPY     While the hardcopy is being sent to the printer  the oscilloscope will display the  message    Hardcopy in process     Press HARDCOPY to abort        Abort  To stop and discard the hardcopy being sent  press HARDCOPY again  while the hardcopy in process message is still on screen     Add to the Spool  To add additional hardcopies to the printer spool  press  HARDCOPY again after the hardcopy in process message is removed from the  screen     3 130 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       You can add hardcopies to the spool until it is full  When adding a hardcopy fills  the spool  the message    H
271. or interpolation  ETI     occurs for your model     Table 3 4  TDS 500B and TDS 700A Sampling Mode Selection  When Fit to Screen is Off              Time Base    gt 50 ns RT RT RT RT RT RT  50 ns RT RT  25 ns RT ETl  12 5 ns ETI ETI   lt 25ns ETI ETI ETI                1 Note that the TDS 520B  TDS 724A  and TDS 782A can have no more that two channels on at a time     2     gt     means    slower than          lt     means    faster than        3    RT    abbreviates    Real Time Sampling    throughout this table     4    ETI    abbreviates    Equivalent Time Sampling or Interpolation    throughout this table     5 Sample mode is ETI for TDS 540B models purchased with Option 1G  Option 1G limits these models to a maximum    sample rate of 1GS second  See Option 1G on page A 1     6 Time base setting not available for this model     3 26    Stop After    To choose the event that stops the acquiring waveforms  do the following step     Press SHIFT ACQUIRE MENU  gt  Stop After  main   gt  RUN STOP button  only  Single Acquisition Sequence  or Limit Test Condition Met  side    See  Figure 3 15    TDS 500B and TDS 700A models only  single acquisition and    limit testing are ignored in InstaVu mode  see Incompatible Modes on  page 3 47      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Tek Run  100kKS 5 Sample       Stop  After    RUN STOP  button only    4 Single   4 Acquisition  Sequence                         Limit Test  Condition M
272. or the probe for more information   Instead  use a passive  probe  such as P6139A passive probe  that allows higher voltage measurements   The P6139A probe is for measurements up to  500 volts        CAUTION  Using the P6243 or P6245 Active Probe to measure signals greater  than  40 volts may damage the probe        TDS 500B  TDS 600B   amp  TDS 700A User Manual 1 3    Start Up       Input Coupling    Putting into Service    Installation    1 4    Be sure to choose the proper input coupling and impedance for the probe or other  cabling you use to couple signals to your oscilloscope  You should read Input  Impedance Considerations on page 3 5 for information needed to ensure proper  coupling of your input signals     To learn how to install  access the front panel  power on  do a self test  and  power off the oscilloscope  do the following procedures     To properly install and power on the oscilloscope  do the following steps     1     Be sure you have the appropriate operating environment  Specifications for  temperature  relative humidity  altitude  vibrations  and emissions are  included in the TDS 500B  TDS 600B   amp  TDS 700A Technical Reference   Performance Verification and Specifications  manual  Tektronix part  number 070 9720 xx      Leave space for cooling  Do this by verifying that the air intake and exhaust  holes on the sides of the cabinet  where the fan operates  are free of any  airflow obstructions  Leave at least 5 1 cm  2 inches  free on each side        
273. ords are placed in time  relative to the main trigger     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 85    Triggering on Waveforms       3 86    To Run After Delay    To Trigger After Delay       NOTE  Due to hardware limitations  the delayed time base cannot be made  triggerable when the main trigger type is Logic  any class  or when the main  trigger type is Pulse with Runt or Slew Rate classes selected  For these settings   the oscilloscope will force the delayed time base to be in Runs After mode        You use the Horizontal menu to select and define either delayed runs after main  or delayed triggerable  Delayed triggerable  however  requires further selections  in the Delayed Trigger menu  Do the following steps to set the delayed time base  to run immediately after delay     1  Press HORIZONTAL MENU  gt  Time Base  main   gt  Delayed Only   side   gt  Delayed Runs After Main  side      2  Use the general purpose knob or the keypad to set the delay time     If you press Intensified  side   you display an intensified zone on the main  timebase record that shows where the delayed timebase record occurs  relative to the main trigger  For Delayed Runs After Main mode  the start of  the intensified zone corresponds to the start of the delayed timebase record   The end of the zone corresponds to the end of the delayed record        NOTE  The intensified zone is not visible when in InstaVu mode  TDS 500B and  TDS 700A models only   see Incompatible Modes on page 3 47   
274. ory along with corresponding timing  information  You can use this digital representation of the signal for display   measurements  or further processing     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 17    Acquiring and Displaying Waveforms       3 18    Real time Sampling    on oN  5 0 V  5 0 V      X        P A a e a      d    OV ov OV OV  wv wl vad E  4 in  5 0 V  5 0 V  Input Signal Sampled Points Digital Values    Figure 3 9  Acquisition  Input Analog Signal  Sample  and Digitize    The oscilloscope uses the samples it takes  see Figure 3   11  to create a wave   form record containing a user specified number of data or record points  Each  record point represents a certain voltage level that occurs a determined amount of  time from the trigger event     The oscilloscope may take more samples than the number of points in your  waveform record  In fact  the oscilloscope may take several samples for each record  point  see Figure 3   10   The digitizer can use any extra samples to perform  additional processing  such as averaging or looking for minimum and maximum  values  The methods of sampling and acquisition modes you choose determine how  the oscilloscope assembles the sample points it acquires into the waveform record      lt        Interval for one waveform record point     Samples for a record point     Figure 3 10  Several Points May be Acquired for Each Point Used    The two general methods of sampling are real time and equivalent time  The  TDS 600B Oscil
275. ory probes and quantities shipped for  these oscilloscopes are listed in Standard Accessories on page A 4     Tektronix also offers a variety of BNC cabling and connectors to couple a variety  of signal sources to its input channels  See your Tektronix Sales representative  for the specific items offered for signal coupling     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 3    Acquiring and Displaying Waveforms       To Compensate  Passive Probes    3 4    When using a passive probe with any product  compensate it to ensure maximum  distortion free input to the oscilloscope and to avoid high frequency amplitude  errors  see Figure 3   1   To compensate your probe  do the following steps     1     Connect the probe to the probe compensation signal on the front panel   Connect the probe ground lead to the ground terminal on the front panel     Press AUTOSET   Press VERTICAL MENU  gt  Bandwidth  main   gt  20 MHz  side              Probe Compensated Correctly    Probe Overcompensated            Probe Undercompensated    Figure 3 1  How Probe Compensation Affects Signals    If you need to change the input impedance  press Coupling  main   Then  toggle the side menu selection Q to get the correct impedance     TDS 500B and 700A models only  Press SHIFT ACQUIRE MENU  gt   Mode  main   gt  Hi Res  side      TDS 600B models only  Press SHIFT ACQUIRE MENU  gt   Mode  main   gt  Average  side   Use the keypad to set Averages to 5     Adjust the probe until you see a square wave wit
276. ou order Option IF  See Options on page A 1        TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 121    Saving Waveforms and Setups       To Access the File Utilities    3 122    To Delete    EA E    The File Utilities menu lets you delete  rename  copy  print files  create a new  directory  operate the confirm delete and overwrite lock  and format disks     To bring up the File Utilities menu     1  Press the SAVE RECALL SETUP button to bring up the Save Recall Setup  menu  or press the SAVE RECALL WAVEFORM button to bring up the  Save Recall Waveform menu  or press the SHIFT HARDCOPY button to  bring up the Hardcopy menu     2  Press File Utilities in the main menu to bring up the File Utilities side menu    See Figure 3   68         NOTE  The amount of free space on the disk is shown in the upper right corner of  the display  The oscilloscope shows the amount in Kbytes  or in Mbytes if the  free space is 1 Mbyte or more   To convert the amount to bytes  you simply  multiply the Kbytes amount times 1024  Thus  the 690 Kbytes shown in Figure  3 68   690 Kbytes x 1024 bytes Kbyte   706 560 bytes        To delete a file or directory  turn the general purpose knob until it scrolls the  cursor over the line marked with both the name of the file or directory to delete  and the file icon or directory icon as shown to the left of this page  Then  press  the side menu Delete button     To delete all files in the file list  set the cursor to the     selection     The oscilloscope 
277. oupling to couple an auxiliary signal to the trigger  circuits  Although trigger coupling can be set to other than DC when in DC Aux   the oscilloscope ignores the setting until one of Ch1 through Ch4 is selected        To Set Mode  amp  Holdoff   You can change the holdoff time and select the trigger mode using this menu  item  See Trigger Modes and Trigger Holdoff beginning on page 3 53 for a  description of these features  To set mode and holdoff  do the following steps     1  Press the TRIGGER MENU  gt  Mode  amp  Holdoff  main   gt  Auto or  Normal  side   The modes operate as follows     m In Auto mode the oscilloscope acquires a waveform after a specific time has  elapsed even if a trigger does not occur  The amount of time the oscilloscope  waits depends on the time base setting     m In Normal mode the oscilloscope acquires a waveform only if there is a valid  trigger     2  To change the holdoff time  press Holdoff  side   Enter the value in time  using the general purpose knob or the keypad     If you want to enter a large number using the general purpose knob  press the  SHIFT button before turning the knob  When the light above the SHIFT button  is on and the display says Coarse Knobs in the upper right corner  the general  purpose knob speeds up significantly     You can set holdoff from 250 ns  minimum holdoff available  to 12 seconds   maximum available   See Holdoff  Variable  Main Trigger in the TDS 500B   TDS 600B   amp  TDS 700A Oscilloscopes Performance V
278. ove CH 1 is lighted  The highlighted menu item in the side menu  has changed from the 50 Q channel 2 setting to the 1 MQ impedance setting  of channel 1     2  Press CH 2  gt  Q  side  to toggle the selection to 1 MQ  That returns the  coupling impedance of channel 2 to its initial state        Tek Run  er Sample                Bandwidth D  Fu      coupling Position Offset Deskew   Cal Probe  it Cony a div au   as Pass    e Scale T       200mY   ch2 100mVQ M 500ys chi 7  264mv        ich  Coupli A    RT podace    y         _ Side Menu Title                                        HEA 50          a        C             Cc           Co                         an  as        Figure 2 13  The Menus After Changing Channels    Remove a Waveform Pressing the WAVEFORM OFF button removes the waveform for the currently  selected channel  If the waveform you want to remove is not already selected   select that channel using the channel  CH  button     1  Press WAVEFORM OFF  under the vertical SCALE knob      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Tutorial       2     Since the CH 2 light was on when you pressed the WAVEFORM OFF  button  the channel 2 waveform was removed     The channel  CH  lights now indicate channel 1  Channel 1 has become the  selected channel  When you remove the last waveform  all the CH lights are  turned off     Press WAVEFORM OFF again to remove the channel 1 waveform     Example 3  Taking Automated Measurements    The TDS Oscilloscope can measure many
279. ox indicates that the general  purpose knob is currently set to adjust that parameter     Turn the general purpose knob left and right  and then use it to adjust the high  level to 80   That sets the high measurement reference to 80      TDS 500B  TDS 600B   amp  TDS 700A User Manual 2 21    Tutorial       2 22    Hint  To make large changes quickly with the general purpose knob  press the  SHIFT button before turning the knob  When the light above the SHIFT button  is on and the display says Coarse Knobs in the upper right corner  the general  purpose knob speeds up significantly     General Purpose Knob  Setting and Readout    i      eeu one a Tere   General  e 2      S Reference Knob Icon  Leve    Set Levels  in  x units 0                             Highlighted  Menu Item  with Boxed  Readout  Value                 Low 7  resolution                               ic hit zoomy i   a a M Sois    chi      248mV  Pa  Remove Gating 1g Reference  Measrmnt getup Snapshot  forech   Measrmnt OFF Histogram Levels    IDC OC aoa       ClO                                                      Figure 2 16  General Purpose Knob Indicators    The Numeric Keypad  Any time the general purpose knob is set to adjust a  numeric parameter  you can enter the value as a number using the keypad instead  of using the knob  Always end the entry of a number by pressing ENTER  47      The numeric keypad also provides multipliers for engineering exponents  such as  m for milli  M for mega  and u for mic
280. paired cursor  to the points along the horizontal axis you are interested in     12  Read the amplitude between the X of the two paired cursors from the  top most A  readout  Read the amplitude of the short horizontal bar of the  selected  solid  cursor relative to either 1 Vrms  0 dB   ground  0 volts   or  zero phase level  0 degrees or 0 radians  from the    readout  Read the  frequency between the long horizontal bars of both paired cursors from the  bottom A  readout     You can use automated measurements to measure FFT math waveforms  Use the  procedure To Take Automated Measurements found in Waveform Differentiation  on page 3 169     There are several characteristics of FFTs that affect how they are displayed and  should be interpreted  Read this topic to learn about the FFT frequency domain  record     how the source waveform relates to the record length  frequency  resolution  and frequency range of that record   The FFT frequency domain  waveform is the FFT math waveform that you display   Continue reading the  topics that follow to learn how to optimize the oscilloscope setup for good  display of your FFT waveforms     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       FFTs May Not Use All of the Waveform Record  The FFT math waveform is a  display of the magnitude or phase data from the FFT frequency domain record  This  frequency domain record is derived from the FFT time domain record  which is  derived from the wav
281. pendix B  Algorithms for details        Holdoff  trigger  A specified amount of time after a trigger signal that elapses before the  trigger circuit will accept another trigger signal  Trigger holdoff helps ensure  a stable display     a Horizontal bar cursors  The two horizontal bars that you position to measure the voltage parameters  of a waveform  The oscilloscope displays the value of the active  moveable   cursor with respect to ground and the voltage value between the bars     Interpolation  The way the oscilloscope calculates values for record points when the  oscilloscope cannot acquire all the points for a complete record with a single  trigger event  That condition occurs when the oscilloscope is limited to real  time sampling and the time base is set to a value that exceeds the effective  sample rate of the oscilloscope  The oscilloscope has two interpolation  options  linear or sin x  x interpolation     Linear interpolation calculates record points in a straight line fit between the  actual values acquired  Sin x  x computes record points in a curve fit  between the actual values acquired  It assumes all the interpolated points fall  in their appropriate point in time on that curve     Intensity  Display brightness     TDS 500B  TDS 600B   amp  TDS 700A User Manual Glossary 5    Glossary                      Glossary 6    InstaVu acquisition mode  A mode that increases the waveform capture rate to up to 400 000 wave   forms per second  This very fast capture rat
282. press Set 1st Source to  side  repeated   ly to cycle it to the desired channel or reference waveform     3  To define the second source waveform  press Set 2nd Source to  side   repeatedly to cycle it to the desired channel or reference waveform     4  To enter the math operator  press Set operator to  side  repeatedly to cycle it  through the choices  Supported operators are           and          NOTE  If you select    for multiply  in step 4  the cursor feature will measure  amplitude in the units volts squared  VV  rather than in volts  V           Tek Run  100kKS 5 Sample         Change Math  Definition             Set ist  Source to  chi        4         i i Set  ae T Ee e EET x D od    operator to       Set 2nd  Source to  Ch2                         Select the type of Math function to define from  the bottom row of bezel buttons  Define the  new function using the side buttons     When you are satisfied with the selections  press     create Math Wim       Press CLEAR MENU to return to the More menu  chi  without creating a Math waveform     DOIL  S                   OOOO                                                             Figure 3 85  Dual Waveform Math Main and Side Menus    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 149    Using Features for Advanced Applications       5     Press OK Create Math Wfm  side  to perform the function     To Average a Math You can also select whether or not you wish to average a certain math waveform   Waveform to do so  p
283. ptions  copyright  and patents      Press SHIFT STATUS     Banner  main    See Figure 3 80      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 139    Determining Status and Accessing Help          ie      Tek Run  auc Sample      A Status  Firmware Version           apres Ev v4 oe snapshot  Horiz Scale de Eri                       vhode iq Fo Rec length Pos  Main 300us Main 50  300   10 divs 30 0  system  Dlyd S ys Runs After  t6ns       zoom Status Vert Horiz Lock  OFF 2 0 20 All OFF                Display       Mode Stop  Sample R S button          OFE OFF OFF    Limit Source Chi Ch2 Ch3 Ch4 Mathi Math2  Refi None None None None None None  Measure Gate Hi low  OFF Histogram waveforms  90  50   70  50      Trigger                            Format Layout Port Palette  Interleaf Portrait File Hardcopy 1 0    Status Banner      IDC  O O O             TIAICIC la                                                       Figure 3 79  Status Menu     System       fe a    Tek Run  aman Sample             Tektronix  7    TDS 6846 Digitizing Oscilloscope  F   v4 0e           Options  13 1F 2F        c  Copyright  Tektronix  Inc   1991 1995   All rights reserved                      LOJOJO  CO          oa  oe    oe   ee                                                           Figure 3 80  Banner Display    3 140 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Determining Status and Accessing Help       Displaying Help  To use the on line help system     Press HELP to provide on scree
284. py or  Current  side  to specify a hardcopy palette  Current uses the current  palette settings to create the hardcopy  while Hardcopy sets the hardcopy  palette to an optimal setting for hardcopy devices     5  Press SHIFT  gt  HARDCOPY MENU  gt  Port  main  to specify the output  channel to send your hardcopy through  The choices are GPIB  RS 232   Centronics  and File     The menu item File chooses the floppy disk drive as the destination for  hardcopies  See To Save to a Floppy Disk on page 3 131        NOTE  TDS 500B  TDS 620B  and TDS 680B oscilloscopes do not come  equipped with a disk drive unless you order Option 1F  See Options on  page A 1        Date Time Stamp the Hardcopy  You can display the current date and time on  screen so that they appear on the hardcopies you print  To date and time stamp  your hardcopy  do the following steps     1  Press DISPLAY  gt  Settings  main   gt  Display  pop up   gt  Readout  Options  main   gt  Display Date and Time  side  to toggle the setting to  On     2  If you want to set the date and time  skip steps 3 and 4 and continue with  step 1 of Set the Date and Time below  Then redo this procedure     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       3  Press Clear Menu to remove the menu from the display so the date and time  can be displayed   See Figure 3 72    The date and time are removed from  the display when menus are displayed      4  Once the oscilloscope is connected to a hardcopy dev
285. r  Indicating the Trigger Position  on the Waveform Record    Trigger Bar Indicating the Trigger  Level on the Waveform Record    3 58    Main Trigger Main Trigger    Main Time Base Time Div cource  Ch1 Slope   Rising Edge         Main Trigger    Main Time Base  Level    U    5                                              Dooa       Figure 3 34  Example Trigger Readouts     Edge Trigger Selected    Record View  To determine where the trigger point is located in the waveform  record and with respect to the display  check the record view at the top of the  display   See Figure 3 35                        Tek Run  1 00GS 5 MO             Select  Measurement             Period                            Frequency    Pang                       Positive  width                      a       Negative  width                    lot   esc Tira ter ESIE ER    Remove Gating t Reference    Measrmnt Setup Snapshot  arms vieasrmnt OFF Histogram Levels     111919  1 a           200mv       i  M 50 0ns  chi T 196mVY    more       LOO IDID ID                                                       Figure 3 35  Record View  Trigger Position  and Trigger Level Bar Readouts    Trigger Position and Level Indicators  To see the trigger point and level on the  waveform display  check the graphic indicators Trigger Position and Trigger Bar   Figure 3   35 shows the trigger point indicator and trigger level bar     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       Both the tr
286. ral purpose knob or  keypad to select the lightness you desire  A value of 0 results in black  A  value of 50 provides the nominal color  A value of 100 results in white     6  Choose Saturation from the side menu and use the general purpose knob or  keypad to select the saturation you desire  A value of 100 provides a pure  color  A value of 0 provides gray     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms                Tek Run  250kS 5 Sample       Change  Colors             color     Ch                Chi   Ch2 Ch3 Ch4 Math Ref Clsn Saturation  o    om       A pony  Reset to    Zone Grat Text Scroll ScrTxt Color    Eog Palette  erties Wap Wap     Restore  Settings 5 Math Reference   Options   lt  Colors Normal Colors Colors Colors Colors       Lightness  65                                                                               MIDICIO  S  O   O       Figure 3 20  Display Menu     Palette Colors    Set Math Waveform Color To define math waveform colors     1     Choose to define math waveform colors by selecting the main menu Map  Math item     Select one of the three math waveforms by pressing Math in the side menu     If you want to assign the selected math waveform to a specific color  press  Color and cycle through the choices     If you want the selected math waveform to be the same color as the  waveform it is based on  select Color Matches Contents  If the math  waveform is based on dual waveforms  the math waveform will
287. rd  length  press HORIZONTAL MENU  gt  Record Length  main   Then  toggle Fit to Screen to ON from the side menu  This feature fits the  waveform automatically much like you could do manually     by turning zoom  mode on and changing the time division until the waveform fits the screen  To  turn off this feature  toggle Fit to Screen to OFF     Horizontal Scale  To change the horizontal scale  time per division  numerically  in the menu instead of using the Horizontal SCALE knob     Press HORIZONTAL MENU  gt  Horiz Scale  main   gt  Main Scale or    Delayed Scale  side   and use the keypad or the general purpose knob to change  the scale values     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Horizontal Position  To set the horizontal position to specific values in the menu  instead of using the Horizontal POSITION knob     Press HORIZONTAL MENU  gt  Horiz Pos  main   gt  Set to 10   Set to  50   or Set to 90   side  to choose how much of the waveform will be  displayed to the left of the display center     You can also control whether changing the horizontal position setting affects all  displayed waveforms  just the live waveforms  or only the selected waveform   See Zoom a Waveform  on page 3   40 for the steps to set the horizontal lock feature     To Select the You also can select Delayed Runs After Main or Delayed Triggerable  Use the  Delayed Time Base main time base for most applications  Use the delayed time base when you
288. re menu  3   97  Set operator to  More menu  3 149  Set Single Source to  More menu  3 148  3 149  Set Thresholds  Main Trigger menu  3   67  3   70  Set to 10   Horizontal menu  3 17  Set to 50   Delayed Trigger menu  3 89  Set to 50   Horizontal menu  3 17  Set to 50   Main Trigger menu  3   56  3 62  3 76   3 84  Set to 90   Horizontal menu  3 17  Set to ECL  Delayed Trigger menu  3 89       TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index       Set to ECL  Main Trigger menu  3   62  3 76  3 80   3 84  Set to TTL  Delayed Trigger menu  3   89  Set to TTL  Main Trigger menu  3   62  3 76  3 80   3 84  Set to Zero  Vertical menu  3 13  Set Hold Trigger  3 64  Setting Up for the Examples  2   9  Settings  Display menu  3 29  3 35  Setup menu  2 10  3 8  Setup Hold trigger  Maximum hold time  NOTE   3   66  Negative setup or hold times  3   65  Positive setup or hold times  3 64  Trigger point location  3 64  Setup Hold trigger  3   63  How to setup  3   70    3   90  Setups  Save and recall  3 115  Shipping  C 1  Side menu  Glossary   10  Side menu buttons  2   3  Glossary   10  SIGNAL OUTPUT  BNC  2 5  Signal Path Compensation  1 3  3 106  Sin x  x interpolation  3   20  3 32  Glossary   5  Sin x  x interpolation  Display menu  3   32  Single Acquisition Sequence  Acquire menu  3   27  SINGLE TRIG button  3   28  3 57  Single Wfm Math  More menu  3 148  Single Shot sampling  3   18  Slew rate setting  How derived  3   82  Slew Rate Trigger  3 74  Slew rate trigger  3 
289. rear panel principal power switch is on  see Figure 1   1   The  principal power switch controls all AC power to the instrument     2  Ifthe oscilloscope is not powered on  the screen is blank   push the  front panel ON STBY button to toggle it on  see Figure 1 2      TDS 500B  TDS 600B   amp  TDS 700A User Manual    Start Up             Tektronix TDS 684B iaren mme oss uoscore  WA             Tek Run  100kS s5 Sample       Tektronix  7          TDS 6846 Digitizing Oscilloscope  F   v4 0e   Options  13 1F 2F              c  Copyright  Tektronix  Inc   1991 1995   All rights reserved          Power On self check PASSED   Push CLEAR MENU to proceed                             M 500s Chi F OV 17Jul 1995  09 29 30                                              ON STBY Button       Figure 1 2  ON STBY Button    The ON STBY button controls power to most of the instrument circuits  Power  continues to go to certain parts even when this switch is set to STBY     Once the oscilloscope is installed  it is typical to leave the principal power  switch on and use the ON STBY button instead of the power switch     Self Test The oscilloscope automatically performs power up tests each time it is turned on   It will come up with a display screen that states whether or not it passed the self  test  To determine the self test results  check the screen   If the self test passed   the status display screen will be removed after a few seconds      If the self test fails  call your local Tektronix Se
290. reen     Trigger Status Lights  To quickly determine trigger status  check the three status  lights TRIG   D  READY  and ARM in the Trigger control area   See Fig   ure 3 33      m When TRIG   D is lighted  it means the oscilloscope has recognized a valid  trigger and is filling the posttrigger portion of the waveform     m When READY is lighted  it means the oscilloscope can accept a valid trigger  event and the oscilloscope is waiting for that event to occur     m When ARM is lighted  it means the trigger circuitry is filling the pretrigger  portion of the waveform record     m   When both TRIG   D and READY are lighted  it means the oscilloscope has  recognized a valid main trigger and is waiting for a delayed trigger  When  the oscilloscope recognizes a delayed trigger  it will fill in the posttrigger  portion of the delayed waveform     m When ARM  TRIG   D  and READY are all off  the digitizer is stopped     m When ARM  TRIG   D  and READY are all lighted  TDS 500B and  TDS 700A models only   FastFrame or InstaVu modes are in effect  No  trigger status monitoring is taking place     Trigger Readout  To quickly determine the settings of some key trigger parame   ters  check the Trigger readout at the bottom of the display   See Figure 3 34    The readouts differ for edge  logic  and pulse triggers     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 57    Triggering on Waveforms       Trigger Position Relative to the  Display and Waveform Record    Trigger Point Indicato
291. rements     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 105    Measuring Waveforms       Measure Waveform  Amplitude    Measure Waveform Time    To measure the amplitude of a waveform  do the following steps     1  Press the channel selection button of the channel you wish to measure  Note  the vertical scale factor for the channel in the channel readout on screen     2  Count the graticule divisions between two features to be measured and  multiply by the vertical scale factor     For example  if you count five major vertical graticule divisions between the  minimum and maximum values of a waveform at a scale factor of 100 mV di   vision  then you can easily calculate your peak to peak voltage as     5 divisions x 100 mV division   500 mV        NOTE  When you select the NTSC graticule  the volts per division of all selected  channels is set to 143 mV div  152 mV div for PAL  where the divisions are those  of the conventional graticule  not the divisions of the video graticules  For  NTSC  the actual grid lines represent 10 IRE  and for PAL the lines are 100 mV  apart        To measure the time of a waveform  repeat the process just described  but count  the horizontal divisions and multiply by the horizontal scale factor  For example   if you count five major horizontal graticule divisions for one waveform cycle at a  horizontal scale factor of 50 uS division  then you can easily calculate the  waveform period as     5 divisions x 50 wS division   250 us  or 4 kHz     
292. requency more accurately than general purpose  probes  but they make less accurate amplitude measurements  They offer a higher  bandwidth to cost ratio     These probes must be terminated in a 50 Q scope input  Input capacitance is  much lower than high Z passive probes  typically 1 pF  but input resistance is  also lower  500 to 5000 Q typically   Although that DC loading degrades  amplitude accuracy  the lower input capacitance reduces high frequency loading  to the circuit under test  That makes Zo probes ideal for timing and phase  measurements when amplitude accuracy is not a major concern     Zo probes are useful for measurements up to 40 V     High voltage probes have attenuation factors in the 100X to 1000X range  The  considerations that apply to other passive probes apply to high voltage probes  with a few exceptions  Since the voltage range on high voltage probes varies  from 1 kV to 20 kV  DC   peak AC   the probe head design is mechanically  much larger than for a passive probe  High voltage probes have the added  advantage of lower input capacitance  typically 2 3 pF         Figure E 1  Typical High Voltage Probes    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix E  Probe Selection       Active Voltage Probes    Active voltage probes  sometimes called    FET    probes  use active circuit  elements such as transistors  There are three classes of active probes     m High speed active  m Differential active  m Fixtured active    Active voltage measuring
293. rigger level 0 0V    D 2       TDS 500B  TDS 600B   amp  TDS 700A User Manual    Appendix D  Factory Initialization Settings       Table D 1  Factory Initialization Defaults  Cont               Control Changed by Factory Init to  Edge trigger slope Rising  Edge trigger source Channel 1  GPIB parameters No change  Hardcopy Format Unchanged   Layout Unchanged   Palette Unchanged   Port Unchanged  Horizontal   delay time division 50 us  Horizontal   delay trigger position 50   Horizontal   delay trigger record length 500 points  10 divs     Horizontal   FastFrame  Horizontal   FastF rame  frame count  Horizontal   FastFrame  frame length    TDS 500B  amp  TDS 700A  Off  TDS 500B  amp  TDS 700A  2  TDS 500B  amp  TDS 700A  500                            Horizontal   fit to screen Off  Horizontal   main time division 500 us  Horizontal   main trigger position 50   Horizontal   main trigger record length 500 points  10 divs   Horizontal   time base Main only  Limit template  V Limit 40 mdiv   H Limit 40 mdiv  Limit template destination Refl  Limit template source Chl  Limit test sources Chl compared to Ref   all others compared to none   Limit testing Off  Limit testing   hardcopy if condition met Off  Limit testing   ring bell if condition met Off  Logic pattern trigger Ch4  Ax2  input X  don   t care   Logic state trigger Ch4  Ax2  input Rising edge  Logic trigger class Pattern    Logic trigger input   pattern and state     Channel 1  H  high    Channels 2  amp  3  Ax1   X  don t
294. rigger source  it continues to be the trigger  source even though the waveform is not displayed     To read about selecting reference waveforms  see Saving and Recalling  Waveforms on page 3 118     To read about selecting  and creating  math waveforms  see Waveform Math on  page 3 148     Scaling and Positioning Waveforms    3 10    To Check Position    The TDS Oscilloscope allows you to scale  change the vertical or horizontal  size  and position  move up  down  left  or right  waveforms on screen for best  display   Figure 3   5 shows the results of both vertical and horizontal scaling and  positioning   This section first tells you how to quickly check and set vertical  and horizontal scales  positions  and other parameters  such as vertical bandwidth  and horizontal record length     To quickly see the position of the waveform in the display  check the Channel  Reference and the Record View icons   See Figure 3 5      The Channel Reference icon  at the left side of the display  points to ground on  the waveform record when offset is set to 0 V  The oscilloscope contracts or  expands the selected waveform around this point when you change the vertical  scale     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       The Record View  at the top of the display  indicates where the trigger occurs and  what part of the waveform record is displayed     Record View                     __f HH eF           Channel Reference Icon                
295. ring Waveforms       Table 3 9  Probe Cal Status         Probe Cal   d    No Doesn t Matter    User Action        Type Probe Connected     Initialized    Simple Interface     Table 3 9 shows the action the oscilloscope takes based on the probe connected  and user operation performed             TEKPROBE Interface  Initialized          Power off          Initialized   probe data is retained            Initialized   probe data is retained                                          Yes Poweron Can not detect differ  Display Re use Probe   Cal d Probe  Pass  ent probe  Calibration Data menu  Different probe  Initialized Different probe  Initialized  Yes Disconnect Probe Initialized Initialized  Yes Connect Probe Can not detect differ  Display Re use Probe   Cal d Probe  Pass  ent probe  Calibration Data menu  Different probe  Initialized Different probe  Initialized                1 Refers to a channel input that was successfully compensated at the time Probe Cal was last executed for the input channel     2 Ifo probe is connected  the probe status in the vertical main menu is always initialized     3 A probe with a simple interface is a probe that can convey very limited information to the oscilloscope  Most passive  probes  such as the P6139A  have simple interfaces     4 A probe with a TEKPROBE interface is a probe that can convey additional information  For instance  it might automatical   ly set the oscilloscope input channel impedance to match the probe  send the oscilloscope
296. ro  To enter these multiplier values  press  the SHIFT button  and then press the multiplier     1  Press Low Ref  side      2  On the numeric keypad  press the 2  the 0  and the ENTER     buttons   which sets the low measurement reference to 20   Observe that the rise time  value has changed     3  Press Remove Measrmnt  main   gt  All Measurements  side   That returns  the display to its original state     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Tutorial       Displaying a Snapshot of You have seen how to display up to four individual automated measurements on  Automated Measurements screen  You can also pop up a display of almost all of the automated measure   ments available in the Select Measrmnts side menus  This snapshot of measure   ments is taken on the waveform currently selected using the channel selection  buttons     As when displaying individual measurements  you must have a stable display of  your signal  and that signal must have all the segments necessary for the  measurement you want     1  Press Snapshot  main  to pop up a snapshot of all available single waveform  measurements   See Figure 2 17      The snapshot display includes the label Ch 1  meaning that the measure   ments displayed are taken on the channel 1 waveform  You take a snapshot  of a waveform in another channel by first selecting that channel using the  channel selection buttons     The snapshot measurements do not continuously update  Snapshot executes a  one time capture of all m
297. roduct description  1   1  Propagation delay  3   92  Pulse trigger  3 52  3 72  definition of classes  3   72  Pulse triggers  definitions of  3   74  Pulse  Main Trigger menu  3 59  3 77  3 79  3 83    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Q    Quantizing  Glossary   9    R    Rack mounting  A 2  Readout  Acquisition  3   24  Channel  2   6  3   8  3   40  Cursors  2   6  Edge trigger  3 59  General purpose knob  2   6  Logic trigger  3   66  Measurement  3   94  Record view  2   6  Snapshot  3   99  Time base  2   6  Trigger  2 6  3 57  3 58  Trigger Level Bar  3 31  Trigger Point  3 31  Readout  Cursor  Paired  3 169  Readout  cursor  H Bars  3 155  3 169  3 173  Paired cursors  3 156  3 174  V Bars  3 155  3 169  3 174  Readout  Display menu  3 31  3 33  Real time sampling  3 18  Real time sampling  Glossary   9  Rear panel  2 5  Recall  Setups  3 115  Recall Factory Setup  Save Recall Setup menu  3 117  Recall Saved Setup  Save Recall Setup menu  3 116  Recalling  Waveforms  3 118  Record Length  Limit in Hi Res mode  3 16  Record length  3   16  Glossary   9  derivative math waveforms  3 168  integral math waveforms  3 172  Option 1M  A 2  Record Length  Horizontal menu  3   16  Record lengths  Incompatible with InstaVu  3 48  Record View  2   6  3   10  3 14  3 58  Rectangular window  3   153  Ref  Color menu  3   37  Ref1  Ref2  Ref3  Ref4  File  Save Recall Waveform  menu  3   120  Ref1  Ref2  Ref3  Ref4  Reference waveform status   3 120    Index 11    
298. rsors that measure the difference  either in  time or voltage  between two locations in a waveform record  This section  describes cursors     how to select their type and mode  how to display them  and  how to use them to take measurements     Cursor measurements are fast and easy to take  Cursors are made up of two  markers that you position with the general purpose knob  You move one cursor  independently or both cursors in tandem  depending on the cursor mode  As you  position the cursors  readouts on the display report and update measurement  information     There are three cursor types  horizontal bar  vertical bar  and paired     see Figure 3 58                                YV          Vertical Bar Cursors Paired Cursors       Horizontal bar cursors measure vertical parameters  typically volts    Vertical bar cursors measure horizontal parameters  typically time or frequency      Paired cursors measure both vertical parameters  typically volts  and horizontal  parameters  typically time  simultaneously     Look at Figure 3 58  Note that each of the two paired cursors has a long vertical  bar paired with an X  The Xs measures vertical parameters  typically volts   the  long vertical bars measure horizontal parameters  typically time or frequency     See Cursor Readouts on page 3   102 for more information      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 101    Measuring Waveforms       3 102    Cursor Modes    Independent Mode       NOTE  When cursors measure cer
299. rvice Center  Depending on the    type of failure  you may still be able to use the oscilloscope before it is serviced     Power Off To power off the oscilloscope  toggle the ON STBY switch     1 6 TDS 500B  TDS 600B   amp  TDS 700A User Manual    D    Overview    This chapter describes the basic concepts of operating the TDS Oscilloscope   Understanding the basic concepts of your oscilloscope will help you use it much  more effectively     The first section  Operating Interface Maps  quickly shows you how the  oscilloscope controls are organized and where you can read about them  It also  illustrates the general procedures for operating the menu system  This section  includes the titles     m   Front Panel Map   m Rear Panel Map   m Display Map   m To Operate a Menu   m To Operate a Pop Up Menu    The second section  Tutorial  contains example procedures that lead you through  the fundamental tasks needed to display a waveform measurement  It also  includes an example procedure that teaches you how to store a setup of the  oscilloscope controls for later use  This section includes the following tuto    rial examples     m Setting Up for the Examples   m Example 1  Displaying a Waveform   m Example 2  Displaying Multiple Waveforms  m Example 3  Taking Automated Measurements  m Example 4  Saving Setups    To explore these topics in more depth and to read about topics not covered in this  section  see Reference  A list of the topics covered begins on Page 3 1     TDS 500B  TDS 
300. s     Fixtured Active Probes In some small geometry or dense circuitry applications  such as surface mounted  devices  SMD   a hand held probe is too big to be practical  You can instead use  fixtured  or probe card mounted  active probes  or buffered amplifiers  to  precisely connect your instrument to your device under test  These probes have  the same electrical characteristics as high speed  active probes but use a smaller  mechanical design     TDS 500B  TDS 600B   amp  TDS 700A User Manual E 3    Appendix E  Probe Selection       Current Probes    E 4    Current probes enable you to directly observe and measure current waveforms   which can be very different from voltage signals  Tektronix current probes are  unique in that they can measure from DC to 1 GHz     Two types of current probes are available  one that measures AC current only  and AC DC probes that utilize the Hall effect to accurately measure the AC and  DC components of a signal  AC only current probes use a transformer to convert  AC current flux into a voltage signal to the oscilloscope and have a frequency  response from a few hundred hertz up to 1 GHz  AC DC current probes include  Hall effect semiconductor devices and provide frequency response from DC to  50 MHz     Use a current probe by clipping its jaws around the wire carrying the current that  you want to measure   Unlike an ammeter which you must connect in series with  the circuit   Because current probes are noninvasive  with loading typically
301. s manual         To run an SPC  do the following steps     1  Power on the digitizing oscilloscope and allow a 20 minute warm up before  doing this procedure     2  Disconnect any input signals you may have connected from all four input  channels        STOP  When doing steps 3 and 4  do not turn off the oscilloscope until signal  path compensation completes  If you interrupt  or lose  power to the instrument  while signal path compensation is running  a message is logged in the oscillo   scope error log  If such a case occurs  rerun signal path compensation        3  Press SHIFT UTILITY     System  main   gt  Cal  pop up   gt  Signal  Path  main    OK Compensate Signal Paths  side      D 4  Wait for signal path compensation to complete  up to 15 minutes   While it  progresses  a    clock    icon  shown at left  is displayed on screen  When  compensation completes  the status message will be updated to Pass or Fail  in the main menu     5  Verify the word Pass appears under Signal Path in the main menu   See  Figure 3 62      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 107    Measuring Waveforms       Channel Probe Deskew    Probe Cal    3 108       Tek Run  5 00GS 5 Sample          4 Signal Path     ompensatio                ENEA AEE E AATE AEA i i i fe  Signal Path Compensation  SPC  corrects for DC  inaccuracies caused by temperature variations  and or long term drift  SPC can be run any  time after the oscilloscope is armed up and  should be run whenever the oscillos
302. s section describes how to set up the  oscilloscope to let it do the work of taking measurements for you     Because automatic measurements use the waveform record points  they are  generally more accurate and quicker than cursor and graticule measurements   The oscilloscope will continuously update and display these measurements     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 91    Measuring Waveforms       Measurement List    Automatic measurements are taken over the entire waveform record or  if you  specify gated measurements  see page 3   95   over the region specified by the  vertical cursors  Automated measurements are not taken just on the displayed  portions of waveforms     The oscilloscope can also display almost all of the measurements at once     see  Take a Snapshot of Measurements on page 3 99     The TDS Oscilloscope provides you with 25 automatic measurements in three  categories  amplitude  typically in volts  sometimes in     time  typically in  seconds or hertz   and area  in volt seconds   Table 3   8 lists brief definitions of  the automated measurements in the oscilloscope  for more details see Appen   dix B  Algorithms  page B 1      Table 3 8  Measurement Definitions    Name  tt Amplitude    Area       Cycle Area       fl Burst Width  a   Cycle Mean    yy Cycle RMS       Delay       whee Fall Time    Definition  Voltage measurement  The high value less the low value measured over the entire waveform or  gated region    Amplitude   High     Low   
303. scope adjustment     The cabinet helps keep dust out of the oscilloscope and must be in place when  operating the oscilloscope  The oscilloscope front cover protects the front panel  and display from dust and damage  Install it when storing or transporting the  oscilloscope     Inspect and clean the oscilloscope exterior as often as operating conditions  require     Send the oscilloscope in for service if it requires an interior cleaning  The  collection of dirt on components inside can cause them to overheat and  breakdown  Dirt acts as an insulating blanket  preventing efficient heat dissipa   tion  Dirt also provides an electrical conduction path that could cause an  oscilloscope failure  especially under high humidity conditions        CAUTION  Avoid the use of chemical cleaning agents which might damage the  plastics used in this oscilloscope  Use only deionized water when cleaning the  menu buttons or front panel buttons  Use a 75  isopropyl alcohol solution as a  cleaner and rinse with deionized water  Before using any other type of cleaner   consult your Tektronix Service Center or representative        Inspection  Inspect the outside of the oscilloscope for damage  wear  and missing  parts  using Table F   1 as a guide  Oscilloscopes that appear to have been  dropped or otherwise abused should be checked thoroughly to verify correct  operation and performance  Immediately repair defects that could cause personal  injury or lead to further damage to the oscilloscope  
304. scribed below      To Recall the To reset your oscilloscope to the factory defaults     Factory Setu  y P Press SAVE RECALL SETUP  gt  Recall Factory Setup  main   gt  OK  Confirm Factory Init  side      See Factory Initialization Settings  on page D 1  for a list of the factory defaults     To Delete All Setups and Sometimes you might use the oscilloscope to acquire waveforms that are  Waveforms     Tek confidential  Furthermore  before returning the oscilloscope to general usage   Secure   you might want to remove all such waveforms and any setups used to acquire  them   Be sure you want to remove all waveforms and setups  because once they  are removed  you cannot retrieve them   To use Tek Secure to remove all  reference setups and waveforms  does not affect mass storage disk      Press SHIFT UTILITY  gt  System  main   gt  Config  pop up   gt  Tek Secure  Erase Memory  main   gt  OK Erase Setup  amp  Ref Memory  side      Executing Tek Secure accomplishes the following tasks   m Replaces all waveforms in reference memories with zero sample values     m Replaces the current front panel setup and all setups stored in setup memory  with the factory setup     m Calculates the checksums of all waveform memory and setup memory  locations to verify successful completion of setup and waveform erasure     m Ifthe checksum calculation is unsuccessful  displays a warning message  if  the checksum calculation is successful  displays a confirmation message     To Run the File Utiliti
305. se  These tools allow investigation of a range of NTSC   PAL  SECAM  and high definition TV signals     IF Floppy disk drive Add 3 5 inch  1 44 MByte floppy disk drive   I  Option applies to TDS 500B  TDS 620B  amp TDS 680B only    1G 1G second sample max  Limitsample rate to 1 GS s maximum  This option allows sales    TDS 500B  TDS 600B   amp  TDS 700A User Manual    in export  controlled countries    Option is only available for the TDS 540B      Appendix A  Options and Accessories       Table A 1  Options  Cont                                                                                                                                                                                             Option   Label Description  13 RS 232 Centronics Hard  Add RS 232 C and Centronics interface ports   Ce oo copy Interface Ports  Option applies to TDS 500B  TDS 620B  amp  TDS 680B only    1K Scope cart K420 scope cart  This cart can help transport the oscilloscope  Jt   around many lab environments   0 i  1M 130 000 record length Extend record length from 50 000 samples standard as follows   a TDS 520B  TDS 724A   amp TDS 782A  To 250 000 samples on  one channel and 130 000 on two channels  TDS 540B  TDS 754A   amp  TDS 784A  To 500 000 samples on  one channel  250 000 on two channels  and 130 000 samples  on three or four channels   Option is only available for the models listed above    1R Rackmount Oscilloscope comes configured for installation in a 19 inch  wide instrument rac
306. set  or both depending on the probe the oscilloscope detects        7  Ifthe message on screen is Probe Offset Compensation rather than Probe  Gain Compensation  skip to step 15     8  Connect the probe tip to PROBE COMPENSATION SIGNAL  connect  the probe ground lead to PROBE COMPENSATION GND     9  Press OK Compensate Gain  side    D 10  Wait for gain compensation to complete  one to three minutes    When gain compensation completes  the following actions occur   m The clock icon will disappear     m If offset compensation is required for the probe installed  the Probe  Offset Compensation message will replace the Probe Gain Compensa   tion message     m  If gain compensation did not complete successfully  you may get a     Probe is not connected    message  examine the probe connections to the  digitizing oscilloscope  be sure the probe tip is properly installed in its  retractor  etc   and repeat step 9      m  If gain compensation did not complete successfully  you may get the  message    Compensation Error     This error implies that the probe gain   2  error  and or offset  50 mV  is too great to be compensated  You can  substitute another probe and continue  Have your probe checked by  service personnel     3 110 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms          Tek Run  er Sample          Probe Gain Compensation    1  Connect probe to front panel signal outputs    2  Press Compensate Gain to proceed   Press Clear Menu to exit           Pos
307. side menu button to  select one of the reference memories in which you have stored a template     Valid selections are any of the four reference waveforms Ref1 through Ref4  or None  Choosing None turns limit testing off for the specified channel or  math waveform        NOTE  Specify the same reference memory you chose as the template destination  if you want to use the template you just created        If you have created more than one template  you can compare one channel to  one template and the other channel to another template     Now specify the action to take if waveform data exceeds the limits set by the  limit test template and turn on limit testing     1  Press SHIFT ACQUIRE MENU  gt  Limit Test Setup  main  to bring up a  side menu of possible actions     2  Ensure that the side button corresponding to the desired action reads ON     m If you want to send a hardcopy command when waveform data exceeds  the limits set  toggle Hardcopy if Condition Met  side  to ON  You can  set the hardcopy system to send the hardcopy to the file system   Do not  forget to set up the hardcopy system  See Hardcopy on page 3 125 for  details      m If you want the bell to ring when waveform data exceeds the limits set   toggle Ring Bell if Condition Met  side  to ON     m If you want the oscilloscope to stop when waveform data exceeds the  limits set  toggle Stop After Limit Test Condition Met  side  to ON     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advance
308. signal to the oscilloscope     AC coupling  A type of signal transmission that blocks the DC component of a signal but  uses the dynamic  AC  component  Useful for observing an AC signal that is  normally riding on a DC signal     Accuracy  The closeness of the indicated value to the true value     Acquisition  The process of sampling signals from input channels  digitizing the samples  into data points  and assembling the data points into a waveform record  The  waveform record is stored in memory  The trigger marks time zero in that  process     Acquisition interval  The time duration of the waveform record divided by the record length  The  oscilloscope displays one data point for every acquisition interval     Active cursor  The cursor that moves when you turn the general purpose knob  It is  represented in the display by a solid line  The   readout on the display  shows the absolute value of the active cursor     Aliasing  A false representation of a signal due to insufficient sampling of high  frequencies or fast transitions  A condition that occurs when a oscilloscope  digitizes at an effective sampling rate that is too slow to reproduce the input  signal  The waveform displayed on the oscilloscope may have a lower  frequency than the actual input signal     Amplitude  The High waveform value less the Low waveform value     AND  A logic  Boolean  function in which the output is true when and only when  all the inputs are true  On the oscilloscope  that is a trigger log
309. sition of waveforms     There are two different ways to delay the acquisition of waveforms  delayed runs  after main and delayed triggerable  Only delayed triggerable uses the delayed  trigger system  Delayed runs after main looks for a main trigger  then waits a  user defined time  and then starts acquiring   See Figure 3 47      Wait for Wait User specified Acquire  Main Time Data  Trigger       Figure 3 47  Delayed Runs After Main    Delayed triggerable looks for a main trigger and then  depending on the type of  delayed trigger selected  makes one of the three types of delayed triggerable mode  acquisitions  After Time  After Events  or After Events Time  Study Figure 3   48 to  understand the sequence the oscilloscope goes through for each delayed mode                                     Wait for Wait for  Delayed Triggerable Main Wait User specified Delayed Acquire  After Time Triager Time Trigger A Data  gge Event  Wait the  Delayed Triggerable User specified  After Events Number of Delayed  Trigger Events  Wait the Wait  Delayed Triggerable User specified Uceesnceiiea  After Events Time Number of Delayed p    Time    Trigger Events       Figure 3 48  Delayed Triggerable    The oscilloscope is always acquiring samples to fill the pretrigger part of the  waveform record  When and if delay criteria are met  it takes enough posttrigger  samples to complete the delayed waveform record and then displays it  Refer to  Figure 3   49 for a more detailed look at how delayed rec
310. st be exactly like the template source  to 5 major  divisions of tolerance     6  When you have finished specifying the limit test template  press OK Store  Template  side   This action stores the specified waveform in the specified  destination  using the specified tolerances  Until you have done so  the  template waveform has been defined but not created     To avoid overwriting the template you have just created  store any new  template you create in a different destination from that just stored     To view the template you have created  press the MORE button  Then press  the button corresponding to the destination reference memory you have used   The waveform appears on the display     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 145    Using Features for Advanced Applications       To Select a Limit Test    Source    To Specify the Limit Test    3 146    Response       NOTE  To view the waveform data as well as the template envelope  it might be  useful to select the Dots display style   See Select the Display Style on  page 3 29         Now specify the channel that will acquire the waveforms to be compared against  the template you have created     1  Press SHIFT ACQUIRE MENU  gt  Limit Test Sources  main   gt   Compare Ch1 to  Compare Ch2 to  Compare Ch3 to  Compare Ch4 to   Compare Math1 to  Compare Math2 to or Compare Math3 to  side      2  Once you have selected one of the four channels or a math waveform as a  waveform source from the side menu  press the same 
311. t  Internal  MathCad  or  Spreadsheet  side      Internal creates files   WFM  in the internal format of the oscilloscope   MathCad creates files   DAT  in a format usable by MathCad       Spreadsheet creates files   CSV  in a format usable by spreadsheets  Excel     Lotus 1 2 3    and Quattro Pro        If you are writing a MathCad program  note that the TDS MathCad file is an  ASCII files  the first four values of which contain header information     m The first header value holds the TDS record length   m The second header value holds time  in seconds  between samples     m The third header value holds the trigger position   expressed as an index in the data position      m The fourth header value refers to the fractional trigger position     Also note that the delimiters are carriage returns     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 119    Saving Waveforms and Setups       To Delete Waveforms To delete a reference waveform s  that you no longer need     Press SAVE RECALL WAVEFORM  gt  Delete Refs  main   gt  Delete Ref1   Delete Ref2  Delete Ref3  Delete Ref4  or Delete All Refs  side      To Delete All Waveforms To remove all stored reference waveforms and setups  use the feature called Tek  and Setups Secure  See To Delete All Setups and Waveforms on page 3 117     To Display a To display a waveform in internal reference memory     Saved Waveform  Press MORE  gt  Ref1  Ref2  Ref3  or Ref4  main    See Figure 3 67      Note that in Figure 3   67  the main menu
312. t a meaningful display     You should have a stable display of the probe compensation waveform from the  last step  Do the following steps to first create an unstable display and then to  autoset the display     1  To create an unstable display  slowly turn the trigger MAIN LEVEL knob   see Figure 2   7  first one direction  and then the other  Observe what  happens when you move the trigger level above the highest part of the  waveform  Leave the trigger level in that untriggered state     4  TRIGGER  MAIN LEVEL  TRIG D    READY  DELAYED TRIG    arm TRIGGER  MENU    SET LEVEL  TO 50            Figure 2 7  TRIGGER Controls    2  Press AUTOSET  see Figure 2   8  and observe the stable waveform display           SAVE RECALL    ACQUIRE MENU             APPLICATION   MENU   C AUTOSET_    WAVEFORM MEASURE HARDCOPY   RUN STOP_   STATUS UTILITY SHIFT  C HELP   SETUP CURSOR DISPLAY    SG  as E    Qs VERTICAL   A HORIZONTAL      y      Figure 2 8  AUTOSET Button Location    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Tutorial       Figure 2   9 shows the display after pressing AUTOSET  If necessary  you can  adjust the waveform now by using the knobs discussed earlier in this example        Ze       Tek Run  at Sample                                  ai om sons cai F 260mv                                              PIG ICO IC          Figure 2 9  The Display After Pressing Autoset       NOTE  If you are using a passive probe  such as the P6139A probe  the corners  on your disp
313. t helps   apd you investigate events that occur when a video signal generates a horizontal or  vertical sync pulse  Supported classes of video triggers include NTSC  PAL   SECAM  and high definition TV signals     Trigger Modes The trigger mode determines how the oscilloscope behaves in the absence of a  trigger event  The oscilloscope provides two trigger modes  normal and automatic     Normal trigger mode enables the oscilloscope to acquire a waveform only when  it is triggered  If no trigger occurs  the oscilloscope will not acquire a waveform    You can push FORCE TRIGGER to force the oscilloscope to make a single  acquisition      Automatic trigger mode  auto mode  enables the oscilloscope to acquire a  waveform even if a trigger does not occur  Auto mode uses a timer that starts  after a trigger event occurs  If another trigger event is not detected before the  timer times out  the oscilloscope forces a trigger anyway  The length of time it  waits for a trigger event depends on the time base setting     Be aware that auto mode  when forcing triggers in the absence of valid triggering  events  does not sync the waveform on the display  In other words  successive  acquisitions will not be triggered at the same point on the waveform  therefore   the waveform will appear to roll across the screen  Of course  if valid triggers  occur the display will become stable on screen     Since auto mode will force a trigger in the absence of one  auto mode is useful in  observing sign
314. taVu mode off        Limit Testing    The TDS Oscilloscope provides limit testing  which can automatically compare each  incoming or math waveform against a template waveform  You set an envelope of  limits around a waveform and the oscilloscope finds waveforms that fall outside  those limits   See Figure 3 82   When it finds such a waveform  the oscilloscope  can generate a hardcopy  ring a bell  and stop and wait for your input                                                     Figure 3 82  Comparing a Waveform to a Limit Template    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 143    Using Features for Advanced Applications       To use limit testing  you must do four tasks    m Create the limit test template from a waveform    m Specify the channel to compare to the template    m Specify the action to take if incoming waveform data exceeds the set limits     m Turn limit testing on so that the parameters you have specified will take  effect     To do the tasks just listed  do the following procedures     To Create Limit Test To use an incoming or stored waveform to create the limit test template  first you  Template select a source and specify a template destination  Then you create the template  envelope by specifying the amount of variation from template you will tolerate   To do these tasks  perform the following steps     1  Press SHIFT ACQUIRE MENU to bring up the Acquire menu     2  Press Create Limit Test Template  main   gt  Template Source  side   gt   Ch1  
315. tage to delta time as     Upper Threshold Setting     Lower Threshold Setting    Sie AC CER Delta Time Setting       Substituting the threshold and delta time settings for the setup in Fig   ure 3   46     Slew Rate Setting   oof or   16 0 mV ns    m The trigger bar indicators  long horizontal bars  point to the upper and lower  thresholds  The pair cursors  which are aligned to threshold levels  read out a  delta voltage of approximately 4 V and a delta time of 200 ns between the  threshold levels  Therefore  the slew rate of the pulse edge triggered on is     dv _ 4 Volts   dt 200 ns   m The Trigger When side menu indicates the oscilloscope will trigger on pulses  with slew rates slower than the slew rate setting  Since the pulse edge slews  at 20 mV ns  which is faster than the slew rate setting of 16 mV ns  the  oscilloscope triggers     Slew Rate Measured           20 mV  ns    m The trigger point indicator shows where the oscilloscope triggers  For a slew  rate triggered waveform  the trigger point is always at the threshold crossed last   the upper threshold for positive polarity settings  the lower for negative  settings      Set the Mode and Holdoff  Mode and holdoff can be set for all standard trigger  types and classes  To set mode and holdoff  refer to Set Mode  amp  Holdoff on  page 3 61  To learn more about trigger mode and holdoff  see Trigger Modes  and Trigger Holdoff on page 3 53     Trigger Based on When you select the pulse class Timeout  the TDS oscillosc
316. tain math waveforms  the measurement may not  be of time  frequency  or voltage  Cursor measurement of those math waveforms  that are not of time  frequency  or voltage is described in Waveform Math  which  begins on page 3 148        There are two cursor modes  independent and tracking   See Figure 3   59      In independent mode  you move only one cursor at a time using the general  purpose knob  The active  or selected  cursor is a solid line  Press SELECT to  change which cursor is selected     In tracking mode  you normally move both cursors in tandem using the general  purpose knob  The two cursors remain a fixed distance  time or voltage  from  each other  Press SELECT to temporarily suspend cursor tracking  You can then  use the general purpose knob to adjust the distance of the solid cursor relative to  the dashed cursor  A second push toggles the cursors back to tracking                Tracking Mode                Only Selected Cursor Moves  lt    ___   gt     Cursor Readouts    Both Cursors Move  in Tandem    Figure 3 59  Cursor Modes    The cursor readout shows the absolute location of the selected cursor and the  difference between the selected and non selected cursor  The readouts differ  depending on the cursor type you select  H Bars  V Bars  or Paired     H Bars  The value after A shows the voltage difference between the cursors  The  value after   shows the voltage of the selected cursor relative to ground   See  Figure 3   60   With the video trigger option
317. tedly  Horizontal Lock to None   See Zoom a  Waveform  on page 3   40 for more information on horizontal lock     m With horizontal lock set as just described  the oscilloscope will reposition  each waveform horizontally to move the first sample in the waveform record  that is outside of template limits to center screen     m If you are comparing each waveform to its own template  the position of  each waveform template will track that of its waveform     m If you are comparing two or more waveforms to a common template  that  template will track the position of the failed waveform  If more than one  waveform fails during the same acquisition  the template will track the  position of the waveform in the highest numbered channel  For example   CH 2 is higher than CH 1     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 147    Using Features for Advanced Applications       Waveform Math    The TDS Oscilloscope provides a means for you to mathematically manipulate  your waveforms  For example  you might have a waveform clouded by  background noise  You can obtain a cleaner waveform by subtracting the  background noise from your original waveform     This section describes the invert  add  subtract  divide  and multiply waveform  math features  See Fast Fourier Transforms on page 3 150  Waveform Differ   entiation on page 3 167  and Waveform Integration on page 3 171 for informa   tion on Advanced DSP Math features     To Use Single Wfm Math To perform waveform math  use the Mor
318. tended monitoring for  and capturing  of  a power supply glitch or GO NO GO slew rate testing of operational    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Triggering on Waveforms       amplifiers  This subsection describes how to use each of the five classes of pulse  triggers  glitch  runt  width  and slew rate  and timeout triggering     A glitch trigger occurs when the trigger source detects a pulse narrower  or  wider  in width than some specified time  It can trigger on glitches of either  polarity  Or you can set the glitch trigger to reject glitches of either polarity     A runt trigger occurs when the trigger source detects a short pulse that crosses  one threshold but fails to cross a second threshold before recrossing the first  You  can set the oscilloscope to detect positive or negative runt pulses     A width trigger occurs when the trigger source detects a pulse that is inside or   optionally  outside some specified time range  defined by the upper limit and  lower limit   The oscilloscope can trigger on positive or negative width pulses     A slew rate trigger occurs when the trigger source detects a pulse edge that  traverses  slews  between two amplitude levels at a rate faster than or slower  than you specify  The oscilloscope can trigger on positive or negative slew rates   You can also think of slew rate triggering as triggering based on the slope   change in voltage change in time  of a pulse edge     A timeout trigger occurs when the trigger sourc
319. ter Char from the main    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       menu to enter each letter  When you have entered the name  press the side menu  OK Accept item   See Figure 3 69      To Set Confirm Delete To turn on or off the confirm delete message  toggle the side menu Confirm  Delete button     When the confirm delete option is OFF  the oscilloscope can immediately delete  files or directories  When the confirm option is ON  the oscilloscope warns you  before it deletes files and gives you a chance to reconsider    To Set Overwrite Lock To turn on or off the file overwrite lock  toggle the side menu Overwrite Lock  button     When overwrite lock is on  the oscilloscope will not permit you to write over an  existing file of the same name  An important reason to allow overwriting is to let  you write files using a target file name that contains wild card characters             This means the oscilloscope creates sequential files whose names are similar  except for the sequential numbers that go in the real name in the place of the  question marks     To Format To format a 720 Kbyte or 1 44 Mbyte disk  turn the general purpose knob until it  scrolls the cursor over the line marked with both the name of the drive to format  in  fd0   and the disk drive icon as shown to the left of this page  Then  press the       side menu Format button     To Find More Information See Saving and Recalling Setups  on page 3 115   See Saving and Reca
320. th Waveform in Math1    3 154    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications       To Take Cursor Once you have displayed an FFT math waveform  use cursors to measure its  Measurements of an FFT frequency amplitude or phase angle     1     Be sure MORE is selected in the channel selection buttons and that the FFT  math waveform is selected in the More main menu     Press CURSOR     Mode  main   gt  Independent  side   gt  Func   tion  main   gt  H Bars  side      Use the general purpose knob to align the selected cursor  solid line  to the  top  or to any amplitude on the waveform you choose      Press SELECT to select the other cursor  Use the general purpose knob to  align the selected cursor to the bottom  or to any amplitude on the waveform  you choose      Read the amplitude between the two cursors from the A  readout  Read the  amplitude of the selected cursor relative to either 1 Vrms  0 dB   ground   0 volts   or the zero phase level  0 degrees or 0 radians  from the     readout   The waveform reference indicator at the left side of the graticule  indicates the level where phase is zero for phase FFTs      Figure 3   89 shows the cursor measurement of a frequency magnitude on an  FFT  The    readout reads 0 dB because it is aligned with the 1 Vgms level   The A  readout reads 24 4 dB indicating the magnitude of the frequency it is  measuring is    24 4 dB relative to 1 Vrms  The source waveform is turned  off in the displ
321. the  box  the oscilloscope alters the magnified display accordingly to include  only the waveform portion within the box     To scale or position the magnified waveform  press Selected Grati   cule  main   gt  Upper  side   Use the vertical and horizontal knobs to scale  and position the magnified waveform     Note that as you scale or move the magnified waveform  the oscilloscope  scales or moves the box relative to the unmagnified waveform  so the box  encloses only the waveform portion magnified in the upper graticule     In Dual Window Zoom mode  the oscilloscope does not display the zoom  magnification factors  however  it does display the scale factors  volts divi   sion and time division  for the zoomed waveform     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Acquiring and Displaying Waveforms       Zoomed  Magnified  Waveforms    Nonzoomed Waveforms  with Box    Indicators at Corners Denoting  the Selected Graticule    Dual Zoom a Waveform    To Set Dual Zoom Offset               Tek Run  ee Sample                          Fay  So omv   ee E  T   Preview                                                                            idom E A enpinerin ear     OIOI    electe  Graticule  tower           119  1  1  1                                              Figure 3 24  Dual Window  Preview  Mode    To select Dual Zoom  press ZOOM     Mode  main   gt  Dual Zoom  side  to  toggle it to ON   See Figure 3 25      Dual zoom displays a second zoomed view of the selected 
322. the FFT  function contains a non integer number of waveform cycles  Since there are  fractions of cycles in such records  there are discontinuities at the ends of the  record  These discontinuities cause energy from each discrete frequency to     leak    over on to adjacent frequencies  The result is amplitude error when  measuring those frequencies     The rectangular window does not modify the waveform record points  it  generally gives the best frequency resolution because it results in the most  narrow lobe width in the FFT output record  If the time domain records you  measured always had an integer number of cycles  you would only need this  window     Hamming  Hanning  and Blackman Harris are all somewhat bell shaped widows  that taper the waveform record at the record ends  The Hanning and Blackman   Harris windows taper the data at the end of the record to zero  therefore  they are  generally better choices to eliminate leakage     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced Applications        al   hn    Rectangular Window    Hamming Window    SN    Hanning Window    Blackman Harris Window       Figure 3 94  FFT Windows and Bandpass Characteristics    Care should be taken when using bell shaped windows to be sure that the most  interesting parts of the signal in the time domain record are positioned in the  center region of the window so that the tapering does not cause severe errors     Waveform Differentiation    The Advanced DSP 
323. the oscilloscope for taking a new autosave single acquisition sequence   press RUN STOP     To avoid loss of reference waveforms  you can save them to floppy disk  use the  SAVE RECALL WAVEFORM menu   before rearming the oscilloscope     Consider the following operating characteristics when using autosave     m Autosave saves all    live    waveforms  that is  waveforms displayed in CH 1      CH 4  To be saved  the live waveforms must be displayed on screen     m Autosave saves each live waveform into the reference memory that  corresponds to the channel  CH 1 to Refl  CH 2 to Ref2  and so on      m Autosave  when executing  erases all four reference memories  To avoid loss  of important waveforms  you may want to save them to a floppy disk file  before enabling a single acquisition sequence     m Autosave is not available in InstaVu mode  TDS 500B and TDS 700A  models only   see Incompatible Modes on page 3 47     To Run the File Utilities To run file utilities  see the Managing the File System on page 3 121     Managing the File System    The TDS Oscilloscope provides file utilities and a floppy disk drive for saving  hardcopies  setups  and waveforms  This section describes how to manage   delete  rename  etc   these files using the file system  Read the sections listed  under To Find More Information on page 3 125 for information on saving  hardcopies  setups  and waveforms        NOTE  TDS 500B  TDS 620B  and TDS 680B oscilloscopes do not come with a  disk drive unless y
324. ther cursor     Track makes both cursors positionable in tandem  that is  both cursors move in  unison and maintain a fixed horizontal or vertical distance between each other     2  Adjust the cursors according to the mode you have selected     m To adjust either cursor in independent mode  use the general purpose knob to  move the selected  active  cursor  A solid line indicates the adjustable cursor  and a dashed line the fixed cursor  Press Select to toggle selection between  the two cursors     m To adjust both cursors in tracking mode  use the general purpose knob to  move both cursors     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms         To adjust the distance between cursors in tracking mode  press SELECT to  temporarily suspend cursor tracking  Then use the general purpose knob to  adjust the distance of the solid cursor relative to the dashed cursor  Press  SELECT again to resume tracking     Select Cursor Speed To change the cursors speed  press SHIFT before turning the general purpose  knob  The cursor moves faster when the SHIFT button is lighted and the display  reads Coarse Knobs in the upper right corner     Select Time Units You can choose to display vertical bar cursor results in units of time or fre   quency  If you have Option 5 Video  you can also display the results in terms of  video line number  To choose vertical bar cursor units  do the following step     Press CURSOR     Time Units  main   gt  seconds or 1 seconds  Hz  or 
325. ton is on and the display says Coarse Knobs  in the upper right corner  the POSITION knob positions waveforms more  quickly     A HORIZONTAL   lt Jposition gt     O    HORIZONTAL  MENU      SCALE  gt        Figure 3 8  Horizontal Controls    When you select a channel  the horizontal SCALE knob scales all channel  waveforms displayed at the same time  If you select a math or reference  waveform  the knob scales only the selected waveform     When you select a channel  the horizontal POSITION knob positions all channel   reference  and math waveforms displayed at the same time when Horizontal Lock is  set to Lock in the Zoom menu  See Zoom a Waveform on page 3   40     To Change Horizontal To select the waveform record length and the trigger position  use the Horizontal  Parameters menu  You can also use this menu to change the horizontal position or scale  instead of using the horizontal knobs  You can select the delayed time base  see  Delayed Triggering on page 3   84  or choose the frames on the waveform that  you want to see  see Using FastFrame on page 3 48      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 15    Acquiring and Displaying Waveforms       3 16    Trigger Position  The trigger point marks time zero in a waveform record  All  record points before the trigger event make up the pretrigger portion of the  waveform record  Every record point after the trigger event is part of the  posttrigger portion  All timing measurements in the waveform record are made  re
326. tore up to ten instrument setups in internal memory  that you may later recall  This section describes how you save and recall a setup   and how you can recall the factory default setup     Save a setup when you want to reuse it later  For example  after changing the  setting during the course of an experiment  you may want to quickly return to  your original setup  Save setups are retained even when you turn the oscilloscope  off or unplug it     To Save a Setup To save the current setup of the oscilloscope     1  Press SAVE RECALL SETUP  gt  Save Current Setup  main      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 115    Saving Waveforms and Setups          STOP Before doing step 2 that follows  note that if you choose a setup location  labeled user  you will overwrite the user setup previously stored there  You can store  setups in setup locations labeled factory without disturbing previously stored setups        2  To store to a setup internally  choose one of the ten internal storage locations  from the side menu To Setup 1  To Setup 2       see Figure 3   65   Now the  current setup is stored in that location     3  To store a setup to disk  press To File  side   Then use the general purpose  knob to select the exact file from the resulting scrollbar list  Finally  press  Save To Selected File  side  to complete the operation        NOTE  Upon power on  the oscilloscope creates the    wild card    file  marked in the    on the left of this page  for storing setups 
327. ttings when you power it off  That feature lets  you power on and continue where you left off without having to reconstruct the  setup in effect when you powered off the oscilloscope        Save a Setup First  you need to create an instrument setup you want to save  Perform the  following steps to create and save a setup that is complex enough that you might  prefer not to go through all these steps each time you want that display     1  If you are not continuing from the previous example  follow the instructions  on page 2 9 under the heading Setting Up for the Examples     2  Press SETUP  gt  Recall Factory Setup  main   gt  OK Confirm Factory  Init  side      3  Press   AUTOSET                    4  Press MEASURE  gt  Select Measrmnt  main   gt  Frequency  side    Press  the  more  side menu item if the Frequency selection does not appear in  the side menu      5  Press CH 2  gt  CLEAR MENU     6  Press SAVE RECALL SETUP     Save Current Setup  main  to display  the Setup main menu   See Figure 2 18         contain a stored setup or with the label factory if they do not  To avoid overwrit   ing  and losing forever  a saved setup  choose a setup location labeled factory    Setup locations labeled factory have the factory setup stored as a default and  can be used to store current setups without disturbing previously stored setups      A CAUTION  Setup locations in the side menu appear with the label user if they       2 24 TDS 500B  TDS 600B   amp  TDS 700A User Manual    T
328. u        NOTE  Use at higher room temperatures or with higher intensity display formats   such as the white fields in the Hardcopy Preview palette  can temporarily  degrade display quality     You can select the Hardcopy Preview palette when using certain color hardcopy  formats  The default colors in the this palette comprise a white background and  fully saturated primary colors which generally produce the best result        To change the color of the current palette  select a color and vary these attributes     Hue  which is the wavelength of light reflected from the surface  It varies  continuously along the color spectrum as produced by a rainbow     Lightness  which is the amount of light reflected from the surface  It varies from  black  to the nominal color  to white     Saturation  which is the intensity of color  Completely desaturated color is gray   Completely saturated color of any hue is that color at its most intense level     1  Select the main menu Change Colors item   See Figure 3   20    2  Select one of the 13 colors by pressing  repeatedly  Color in the side menu     3  If you want to use the factory default for this color  press the side menu  Reset to Factory Color     4  Choose Hue from the side menu and use the general purpose knob or keypad  to select the desired hue  Values range from 0 to 359  Sample values are   0   blue  60   magenta  120   red  180   yellow  240   green  and  300   cyan     5  Choose Lightness from the side menu and use the gene
329. uctuations in the signal     MCross Calculations MCross1  MCross2  and MCross3     refer to the first  second  and third MidRef  cross times  respectively   See Figure B   1      The polarity of the crossings does not matter for these variables  but the  crossings alternate in polarity  that is  MCross1 could be a positive or negative  crossing  but if MCross  is a positive crossing  MCross2 will be a negative  crossing     The oscilloscope calculates these values as follows     1  Find the first MidRefCrossing in the waveform record or the gated region   This is MCross1     2  Continuing from MCross   find the next MidRefCrossing in the waveform  record  or the gated region  of the opposite polarity of MCross1  This is  MCross2     3  Continuing from MCross2  find the next MidRefCrossing in the waveform  record  or the gated region  of the same polarity as MCross1  This is  MCross3     TDS 500B  TDS 600B   amp  TDS 700A User Manual B 3    Appendix B  Algorithms       MidR ef    Hysteresis x Amplitude   MidR ef  MidR ef    Hysteresis x Amplitude        MCross1Polarity     is the polarity of first crossing  no default   It can be rising or  falling     StartCycle     is the starting time for cycle measurements  It is a floating point    number with values between 0 0 and  RecordLength     1 0   inclusive     StartCycle   MCross1    EndCycle     is the ending time for cycle measurements  It is a floating point  number with values between 0 0 and  RecordLength     1 0   inclusiv
330. uire a waveform record  unless a valid trigger event occurs  It waits for a valid trigger event before  acquiring waveform data     NOR  A logic  Boolean  function in which the output of the OR function is  complemented  true becomes false  and false becomes true   On the  oscilloscope  that is a trigger logic pattern and state function     OR  A logic  Boolean  function in which the output is true if any of the inputs are  true  Otherwise the output is false  On the oscilloscope  that is a trigger logic  pattern and state function     Oscilloscope  An instrument for making a graph of two factors  These are typically voltage  versus time     Peak Detect acquisition mode  A mode in which the oscilloscope saves the minimum and maximum  samples over two adjacent acquisition intervals  For many glitch free signals     TDS 500B  TDS 600B   amp  TDS 700A User Manual Glossary 7    Glossary       Glossary 8                                                    that mode is indistinguishable from the sample mode   Peak detect mode works  with real time  non interpolation sampling only      Peak to Peak  Amplitude  voltage  measurement of the absolute difference between the  maximum and minimum amplitude     Period  A timing measurement of the time covered by one complete signal cycle  It  is the reciprocal of frequency and is measured in seconds     Phase  A timing measurement between two waveforms of the amount one leads or  lags the other in time  Phase is expressed in degrees  where 3
331. ulated Postscript  mono plot    m EPS Color Plot  Encapsulated Postscript  color plot    m Interleaf   m HPGL Color Plot    Depending on the output format selected  the oscilloscope creates either an  image or a plot  Images are direct bit map representations of the oscilloscope  display  Plots are vector  plotted  representations of the display     Some formats  particularly Interleaf  EPS  TIFF  PCX  BMP  and HPGL  are  compatible with various desktop publishing packages  Such compatibility means  you can paste files created from the oscilloscope directly into a document on any  of those desktop publishing systems     3 126 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Saving Waveforms and Setups       EPS Mono and Color formats are compatible with Tektronix Phaser Color  Printers  HPGL is compatible with the Tektronix HC100 Plotter  and Epson is  compatible with the Tektronix HC200 Printer     To Set Up for Making Before you make a hardcopy  you need to set up communications and hardcopy  Hardcopies parameters  Do the following procedures to set up for making hardcopies     Set Communications Parameters  To set up the communication parameters for a  printer attached directly to the oscilloscope GPIB  RS 232 or Centronics port     Press SHIFT     UTILITY  gt  System  main   gt  I O  pop up   gt  Confi   gure  main   gt  Hardcopy  Talk Only   side    See Figure 3 70         i 5    Tek Run  2 50MS s Sample RS232 Baud Rate  9600          R 32  HW Setup    Baud Rate    Stop
332. unzoomed waveform   The second zoomed view is offset in time from the first zoomed view  Also   zoom must be enabled  side menu set to On or Preview  to see the Dual Zoom  displays     To set the offset in time of the second zoomed waveform from the first  press  ZOOM     Mode  main   gt  Dual Zoom Offset  side   Then turn the general  purpose knob or use the keypad to set the offset     Dual Zoom offset is always positive  The oscilloscope sets the offset as close to  the requested value as possible  An offset request of 0 0 insures that the zoom  boxes are butted up against each other  regardless of the zoom factor     The horizontal zoom and scale factors determine the minimum offset time  available  Both zoom boxes always enclose equal amounts of time with the  second box always offset from the first by a time equal to one box  Doubling the  zoom factor halves the time enclosed by either box and  therefore  halves the  minimum offset time     The oscilloscope retains any value input that is less than the minimum time  available as a    request    if you enter that value using the keypad  Increasing the  zoom factor or decreasing the horizontal scale to a setting that allows the    TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 43    Acquiring and Displaying Waveforms       requested value sets offset time to that value  You cannot set offset to less than  the minimum offset time available when using the general purpose knob           Tek Run  ICOM Sample oe Time  412 6n
333. uppression level     1  Do steps 1 through 7 of To Create an FFT that begins on page 3 152  Select  dBV RMS  side  for the Set FFT Vert Scale to  side      2  Press CURSOR     Mode  main   gt  Independent  side   gt  Func   tion  main   gt  H Bars  side   Use the general purpose knob to align the  selected cursor to a level that places the tops of the magnitudes of frequen   cies of interest above the cursor but places other magnitudes completely  below the cursor     3  Read the level in dB from the    readout  Note the level for use in step 5     4  Press MORE  main   gt  Change Waveform Definition menu  side   Press  Set FFT Vert Scale to  side  repeatedly to choose either Phase  rad  or  Phase  deg      5  Press Suppress Phase at Amplitudes  side   Use the general purpose knob  to set phase suppression to the value obtained using the H Bar cursor  Do not  change the window selection or you will invalidate the results obtained using  the cursor     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 163    Using Features for Advanced Applications       3 164    FFT Windows    To learn how to optimize your display of FFT data  read about how the FFT  windows data before computing the FFT math waveform  Understanding FFT  windowing can help you get more useful displays     Windowing Process  The oscilloscope multiplies the FFT time domain record by  one of four FFT windows before it inputs the record to the FFT function  Figure  3 93 shows how the time domain record is processe
334. ure lets you limit measurements to a specified portion of the  waveform  When gating is Off  the oscilloscope makes measurements over the  entire waveform record     When gating is activated  vertical cursors are displayed  Use these cursors to  define the section of the waveform you want the oscilloscope to measure   This  section is called the gated region   Do the following steps to gate a measurement     1  Press MEASURE  gt  Gating  main   gt  Gate with V Bar Cursors  side     See Figure 3 54      TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 95    Measuring Waveforms       Define High Low Setup    3 96    2  Using the general purpose knob  move the selected  the active  cursor  Press  SELECT to change which cursor is active     Displaying the cursor menu and turning V Bar cursors off will not turn  gating off   Gating arrows remain on screen to indicate the area over which  the measurement is gated   You must turn gating off in the Gating side menu        NOTE  Cursors are displayed relative to the selected waveform  If you are making a  measurement using two waveforms  this behavior can be a source of confusion  If  you turn off horizontal locking and adjust the horizontal position of one waveform  independent of the other  the cursors appear at the requested position with respect  to the selected waveform  Gated measurements remain accurate  but the displayed  positions of the cursors change when you change the selected waveform           a A    Tek Run  100kKS 5 S
335. ut  Depending on the probe it finds on  each input  it takes one of the following actions     m If the probe has a TEKPROBE interface  such an interface can convey  additional information  such as a unique identification number   the  oscilloscope determines whether it is the same probe for which data was  stored  If it is  the oscilloscope sets status to pass  if not  it sets the status to  Initialized     m Ifa probe has a simple oscilloscope interface  the oscilloscope can usually  determine if it has a different probe attenuation factor than that stored for the  last Probe Cal  It can also determine if the last Probe Cal was for a probe  with a TEKPROBE interface  If either is the case  the probe installed is  different from that stored for the last Probe Cal  Therefore  the oscilloscope  sets the status to Initialized      m Ifa probe has a simple oscilloscope interface and the probe attenuation factor  is the same as was stored at the last Probe Cal  the oscilloscope cannot  determine whether it is the same probe  Therefore  it displays the Re use  Probe Calibration data  menu   See Figure 3   64      3 112 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Measuring Waveforms          Tek Running  100kS 5 Sample        Re use Probe  Cal              OK    Use Existing  Cal Data                   u Re use Probe Calibration data       Probe cal data exists for MiA  but it may not be  for the attached probe     To use the existing probe cal data for this  probe  press Use 
336. utorial          Tek Run  100kS 5 Sample          Save Current  Setup             uJ  To File          To Setup1    user          To Setup 2    factory          To Setup3    factory                      chi 200m   100m   M 5004s Chil  248m  V Nora  i 5   s   E          o    Si E i HHH TEREA IETT     Saved Current Factory Utilities  Setup Setu Setup    DIG ICO IC IC         Ooo oj                                                       Figure 2 18  Save Recall Setup Menu    7  Press one of the To Setup side menu buttons to store the current instrument  settings into that setup location  Remember which setup location you  selected for use later     There are more setup locations than can be listed at one time in the side  menu  The    more     side menu item gives you access to all the setup  locations     Once you have saved a particular setup  you can change the settings as you  wish  knowing that you can come back to that setup at any time                 8  Press MEASURE  gt  Positive Width  side  to add that measurement to the  display     Recall a Setup To recall the setup  press SAVE RECALL SETUP  gt  Recall Saved Set   up  main   gt  Recall Setup  side  for the setup location you used in the last  exercise  The positive width measurement is now removed from the display  because you selected it after you saved the setup     The step just performed completes the examples  You can restore the default  settings by pressing SETUP  gt  Recall Factory Setup  main   gt  OK Confir
337. vertical bars you position to measure the time parameter of a  waveform record  The oscilloscope displays the value of the active  move   able  cursor with respect to the trigger and the time value between the bars     Waveform  The shape or form  visible representation  of a signal     Waveform interval  The time interval between record points as displayed     C XY format    A display format that compares the voltage level of two waveform records  point by point  It is useful for studying phase relationships between  two waveforms     Ema YT format    The conventional oscilloscope display format  It shows the voltage of a  waveform record  on the vertical axis  as it varies over time  on the  horizontal axis                  TDS 500B  TDS 600B   amp  TDS 700A User Manual Glossary 11    Glossary       Glossary 12 TDS 500B  TDS 600B   amp  TDS 700A User Manual    Index    Numbers    1 seconds  Hz   Cursor menu  3   105   2   2 channel operation  xii  1 1  1 2  Glossary   1  20 MHz  Vertical menu  3   13   250 MHz  Vertical menu  3   13    A    AC coupling  Glossary 1  AC line voltage  trigger input  3   52  AC  Main Trigger menu  3   60  Accept Glitch  Main Trigger menu  3   76  Accessories  A   1  Optional  A S   A 8  Probes  A 5  Software  A   7  Standard  A   4  A 7  Accuracy  Glossary   1  Acquire menu  3   25  Average  3 25  Average mode  3 144  Compare Ch1 to  3 146  Compare Ch2 to  3 146  Compare Ch3 to  3 146  Compare Ch4 to  3 146  Compare Math1 to  3 146  Compare Mat
338. w  3 164  Blackman Harris  3 154  3 164  3 167  characteristics of  3 166  Hamming  3 154  3 164  3 167    TDS 500B  TDS 600B   amp  TDS 700A User Manual    Hanning  3 154  3 164  3 167  rectangular  3 153  3 164  3 167  rectangular vs  bell shaped  3 166  selecting  3 164  Windowing  process  3 164  Windows  descriptions of  3 153    X    XY   Format  3 33   Incompatible with InstaVu  3   47  XY format  Glossary   11  XY  Display menu  3 33    Y    YT  Format  3   33  YT format  Glossary   11  YT  Display menu  3 33    Z    Zero phase reference point  3 157  3 162  establishing for impulse testing  3   162   3 163  Zoom  3   39 3   50  And interpolation  3   40  And waveforms  3   39  derivative math waveforms  3 171  Dual Window mode  3 42  Dual Zoom  3   43  Dual Zoom Offset  3   43  Incompatible with InstaVu  3 47  on FFT math waveforms  3 160  on integral math waveforms  3 175  ZOOM button  3   40  Zoom feature  3   39  Zoom menu  Dual Zoom  3   43  Dual Zoom Offset  3   43  Preview  3   42  Reset Zoom Factors  3   42    Index 15    Index       Index 16 TDS 500B  TDS 600B   amp  TDS 700A User Manual    
339. waveform record  All  points in the record are located in time with respect to that point  The oscillo   scope continuously acquires and retains enough sample points to fill the  pretrigger portion of the waveform record  that part of the waveform that is  displayed before  or to the left of  the triggering event on screen   When a trigger  event occurs  the oscilloscope starts acquiring samples to build the posttrigger  portion of the waveform record  displayed after  or to the right of  the trigger  event   Once a trigger is recognized  the digitizing oscilloscope will not accept  another trigger until the acquisition is complete     You can derive your trigger from the following sources     Input channels provide the most commonly used trigger source  You can select  any one of the four input channels  The channel you select as a trigger source  will function whether it is displayed or not     AC Line Voltage is the trigger source most often used when you are looking at  signals related to the power line frequency  Examples include devices such as  lighting equipment and power supplies  Because the oscilloscope generates the  trigger  you do not have to input a signal to create the trigger     Auxiliary Trigger is the trigger source most often used in doing digital design  and repair  For example  you might want to trigger with an external clock or with  a signal from another part of the circuit  To use the auxiliary trigger  connect the  external triggering signal to the 
340. waveform record  that is the averaged result of several acquisitions  This mode reduces random  noise  The oscilloscope acquires data after each trigger event using Sample  mode  It then averages the record point from the current acquisition with those  stored from previous acquisitions     mal Peak Detect Mode  Peak Detect mode alternates between saving the highest  sample in one acquisition interval and lowest sample in the next acquisition  interval  Peak Detect mode only works with real time  noninterpolated sampling     If you set the time base so fast that it requires real time interpolation or  equivalent time sampling  the mode automatically changes from Peak Detect to  Sample  although the menu selection will not change              Hi Res Mode  TDS 500B and 700A models only  Hi Res mode averages all  samples taken during an acquisition interval to create a record point  The average  results in a higher resolution  lower bandwidth waveform        A key advantage of Hi Res is its potential for increasing resolution regardless of  the input signal  Table 3 3 and the equations shown below illustrate how you  can obtain up to 15 significant bits with Hi res mode  Note that the resolution  improvements are limited to speeds slower than 100 ns div  Also  resolutions  above 15 bits are not allowed by internal hardware and computation limitations     Si   Sampling Interval for TDS 754A   I ns  E _ Time  Div _  Sps Div _  pe CHa   Number Of Points Div 50 Points   Div FALOU   
341. y also need to use phase suppression  to reduce noise in your FFTs     Establish a Zero Phase Reference Point  The phase of each frequency is measured  with respect to the zero phase reference point  The zero reference point is the  point at the center of the FFT math waveform but corresponds to various points  on the source  time domain  record   See Figure 3   90 on page 3 157      To measure the phase relative to most source waveforms  you need only to center  the positive peak around the zero phase point   For instance  center the positive  half cycle for a sine or square wave around the zero phase point   Use the  following method     m First be sure the FFT math waveform is selected in the More menu  then set  horizontal position to 50  in the Horizontal menu  This positions the zero  phase reference point to the horizontal center of the screen     m In the Horizontal menu  vary the trigger position to center the positive peak  of the source waveform at the horizontal center of screen  Alternately  you  can adjust the trigger level  knob  to bring the positive peak to center screen  if the phase reference waveform has slow enough edges     When impulse testing and measuring phase  align the impulse input into the  system to the zero reference point of the FFT time domain waveform     m Set the trigger position to 50  and horizontal position to 50  for all record  lengths less than 15 K     TDS 500B  TDS 600B   amp  TDS 700A User Manual    Using Features for Advanced App
342. y content of a waveform you  acquire as an FFT math waveform  This frequency domain waveform is based on  the following equation     N    pie  _ j2ank  1 N  X k    x n e for k   0 toN     1  je   2  Where  x n  is a point in the time domain record data array    X k  is a point in the frequency domain record data array  n is the index to the time domain data array   k is the index to the frequency domain data array   N is the FFT length   jis the square root of    1    The resulting waveform is a display of the magnitude or phase angle of the  various frequencies the waveform contains with respect to those frequencies  For  example  Figure 3   86 shows the untransformed impulse response of a system in  channel 2 at the top of the screen  The FFT transformed magnitude and phase  appear in the two math waveforms below the impulse  The horizontal scale for  FFT math waveforms is always expressed in frequency per division with the  beginning  left most point  of the waveform representing zero frequency  DC      The FFT waveform is based on digital signal processing  DSP  of data  which  allows more versatility in measuring the frequency content of waveforms  For  example  DSP allows the oscilloscope to compute FFTs of source waveforms  that must be acquired based on a single trigger  making it useful for measuring  the frequency content of single events  DSP also allows the phase as well as the  magnitude to be displayed     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 151    
343. your original horizontal scale setting by pressing one menu button    Zoom is also handy if you have acquired a waveform at the fastest time per  division and want to further expand the waveform horizontally      To help you use zoom effectively  consider how it operates on waveforms  When  zooming vertically  the oscilloscope expands or contracts the selected waveform  only  Also  the oscilloscope only positions the selected waveform when in Zoom     When zooming horizontally  Zoom expands either the selected waveform  all live  waveforms  or all live and reference waveforms  depending on the setting for  Horizontal Lock in the Zoom menu     TDS 500B  TDS 600B   amp  TDS 700A User Manual 3 39    Acquiring and Displaying Waveforms       Interpolation and Zoom    Checking the Zoom Factor    3   40    Zoom a Waveform    When zooming horizontally or vertically  Zoom expands or contracts the  waveform by the zoom factor in 1X  2X  and 5X steps     To help you use Zoom effectively  consider how it is affected by interpolation   When you zoom on a waveform  you expand a portion of it  If the expansion  requires the oscilloscope to show more points for that portion than it has  acquired  it interpolates     The method the oscilloscope uses to interpolate  linear or sin x  x  can affect the  way Zoom displays your waveform  If you selected sin x  x  the default   it may  introduce some overshoot or undershoot to the waveform edges  If such is the  case  change the interpolation metho
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Samsung SM-N900U 用戶手冊(KK)  Manuel d`utilisation de la carte environnementale agricole 2014  Super LoiLoScope Ver.1.8 Você pode exibir o Super LoiLoScope    CUISINIÈRE À GAZ  EXBP-MP3 - Korg Professional Arranger  RD 16 Démontage et remontage  User Manual - Newegg.com  STH・500  一般競争入札公告(売払)    Copyright © All rights reserved. 
   Failed to retrieve file