Home

SERVICE MANUAL

image

Contents

1. 700 SERIES METER CALIBRATION The procedure to change the calibration of the meter 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 Check for repeatability A Run a fast high gallons per minute test in your prover and record how much the meter is over or under registering Do NOT change the meter s calibration B Run a slow low gallons per minute test and again record how much the meter is over or under registration If the results from the fast and the slow test are the same or close to the same the meter can be successfully calibrated If the test results differ beyond the rated tolerance of the meter then repairs are necessary before the meter can be calibrated A valve or nozzle must be at the end of the delivery hose at the prover Turn on the pump and purge the system of air Leave the pump on Shut off the valve at the end of the delivery hose and let the system pressurize Wet the prover and empty it letting it drip for 30 seconds Reset the register on the meter to 0 Fill the prover to the line marked 0 Calculate the percentage increase or decrease of registration required For example if a 50 gallon test prover indicates exactly 50 gallons have been run through the meter against a meter registration of 49 7 gallons the percentage increase of registration would be 0 6 3 10 s in 50 gallons is 0 6 If meter registration is less than the test prover measured volume an in
2. adjuster barrel is equal to 1 gallon in 100 gallons or 1 of delivered volume The adjuster body will show divisions of 1 while the adjuster barrel has divisions of 0 02 0 02 Each Division ta es _ Tighten Screw 0 1 Each Division Le Alir Adjusting Locking Collar P Adjuster Barrel For volumes other than 100 gallons the following formula may be used to calculate the Adjuster percent to increase or decrease prover volume Volume on Prover minus Volume on the Meter Counter Volume on the Meter Counter TOTAL CONTROL SYSTEMS 2515 Charleston Place Fort Wayne Indiana 46808 800 348 4753 1 260 484 0382 Fax 1 260 484 9230 service tcsmeters com
3. crease adjustment is required If meter registration is greater than the test prover measured volume a decrease adjustment is required Remove the seal and seal wire Remove the front cover plate Loosen the screw on the clamping ring The adjuster barrel will now be free to rotate Each small hash mark corresponds to 0 02 adjustment Each of the larger numbers on the rotating barrel corresponds to 0 1 The large numbers engraved into the adjuster body correspond to 1 The entire range of adjustment is 5 with the meter being set at the factory very close to the middle of this range To increase the registered volume rotate the adjuster barrel to the left by the percentage increase desired It is recommended that the barrel be rotated 0 02 more than the desired amount and then backed down to the desired figure This accounts for any slack present in the adjuster mechanism In the example above the barrel would be rotated 16 tick marks on the barrel 15 tick marks x 0 02 0 3 3 10 s plus the one tick mark for slack With the slack removed the barrel would then be rotated one tick mark back to the right to achieve the final adjustment To decrease the registered volume the adjuster barrel is rotated to the right Tighten the screw and clamping ring Run another test and adjust if necessary When calibration is achieved ensure that the screw and clamping ring are tight and then replace the cover plate One complete turn of the

Download Pdf Manuals

image

Related Search

Related Contents

Bedienungsanleitung herunterladen  Hoist Fitness H210 User's Manual  Retail Loan Origination User Manual  Stellungnahme des Senats zum Jahresbericht 2000 (Drs. 14/1328)  expansion tanks tanques de expansión réservoirs d  Leadmgmmvata。n))) 東芝白熱灯器具取扱眼明書  Cornelius P/N 620043512 User's Manual  

Copyright © All rights reserved.
Failed to retrieve file