1. BlueLuxcer Digital Curing Light Blue or User Manual Table of Contents SET UP lle ia ia 1 OPERATION nari 1 HIGH POWER MODE 1 2 BUILT IN LIGHT METER 2 CLEANING amp STERILIZATION 3 DEVICE SPECIFICATIONS 3 TROUBLESHOOTING 4 WARRANTY li lea 4 SERVICE ul night 4 1 Unpack the curing light Place on firm flat surface or wall mount if preferred 2 Insert the fiber optic probe into the handpiece until firmly seated Attach the protective shield onto the probe 3 Place the handpiece in the holder 4 Plug the unit into an AC outlet within a range from 100V to 240V The light operates with a universal power supply 5 The CU 1000 curing light is ready for use OPERATION 1 Turn on the power switch located on the left side of base unit 2 The LED s lights will run through a power up test When the test is completed a beep sound will confirm unit is ready The previous curing mode setting is retained in memory 3 Apply composite in mm s as indicated by composite manufacturer Select curing mode to begin curing process HIGH POWER MODE Press the HIGH POWER button for full strength curing There 3 pre set curing times
2. 5 sec 10 sec and 20 sec The curing cycles are activated by pressing the trigger switch the handpiece which simultaneously activates the magnetic fan for cooling 5 sec The curing cycle ends with a beeping signal The fan will continue running for automatic cooling varies from mode to mode 10 beeping signal is sounded at 5 second intervals before the auto shutdown 20 sec beeping signal is sounded at 10 second intervals before the auto shutdown SOFT UP MODE The mode is designed to prevent composite shrinkage during polymerization It is also very beneficial to use when curing dual cure resin cement all ceramic and poly ceramic systems The SOFT UP button activates a curing cycle with increasing intensity from low to high power light output Pressing the trigger switch on the handpiece will start or stop the curing light at anytime 10 sec 20 sec 30 sec This curing cycle begins with 5 seconds of increasing intensity until reaching the maximum and then stays at maximum curing power for 5 seconds Beeping signal sounds at 5 second intervals This curing cycle begins with 7 seconds of increasing intensity until reaching the maximum and then stays at maximum curing power for 13 seconds Beeping signal sounds at 10 second intervals This curing cycle begins with 10 seconds of increasing intensity until reaching t
3. eplace the halogen bulb CLEANING amp STERILIZATION 1 Light Guide After each patient the light guide should be sterilized Autoclave at 135 C for 5 minutes 2 Use a cloth wipe and gentle cleansing agent to clean control unit and handpiece surfaces N CAUTION Never use a solvent based cleaner This will cause surface printing to fade DEVICE SPECIFICATIONS Voltage AC 110V 220V 50 60Hz Effect A Classification Class Protection against electric shock Type B applied part Protection against harmful ingress of water Ordinary equipment IPX0 Mode of operation Continuous operation or 30 seconds TROUBLESHOOTING PROBLEM CAUSE METHOD Inspect the power connection No power when switch is on Cord or plug connections and handpiece cable Inspect fuse and change if needed If it fails again send for repair No power when switch is on Fuse No light from Handpiece Halogen bulb not working Replace bulb Unexpected beeping Some buttons are pressed Check all buttons No light after a few uses if bulb okay Cooling fan not working Send for repair BUILT IN LIGHT METER This product is designed for use in a dental office and this warranty is not applicable to other uses The unit is warranted for twelve 12 months from date of purchase against manufacturing defects in materials and workmanship Henry Schein will repair or replace the unit at its sole discretion This war
4. he maximum and then stays at maximum curing power for 20 seconds Beeping signal sounds at 10 seconds 20 seconds BUILT IN LIGHT METER This feature is designed to measure the intensity of the lamp output and use these readings to auto adjust the curing time to compensate for any light deterioration This function is fully computerized and automated when necessary The light will automatically adjust the curing time based on the following guidelines GREEN YELLOW This color indicates that the light intensity is within accepted curing range at over the 600mW cm This color indicates that the light intensity has fallen into a weaker than normal range between 200 600mW cm At this point the curing time adjuster will extend the curing time relative to how much the halogen bulb has deteriorated using the following guidelines the light intensity is between 400 mW cm2 to 600mwicm the selected normal curing time will be extended by 3 seconds b If the light intensity is between 200 mW cm2 to 400mw cm the selected normal curing time will be extended by 6 seconds Any changes will be stored memory until unit is turned off Recallibrate the light intensity reading by turning the power off then on to reset the memory N CAUTION RED Indicates that the light intensity has fallen below 200mW cm At this point a continuous short beeping signal will sound as a reminder to r
5. ranty shall be limited to replacement or repair of the unit or its parts and shall not extend to any other claims including but not limited to loss of profit cost of removal or replacement incidental or consequential damages or other similar claims arising from the use of this product Damages to the product resulting from acts of God faulty installation misuse tampering accident abuse negligence or unauthorized repairs or alternations unrelated to problems with materials and workmanship are not covered by this warranty NAONITEX CO LTD 6F 70 GUANG FU ROAD SEC 1 SAN CHONG CITY TAIPEI TAIWAN TEL 886 2 29993869 FAX 886 2 29993647 E mail monitex fomed com tw http www monitex com tw