Home
        Operating Instructions - OPTISWITCH 3000 C with relay
         Contents
1.      OPTISWITCH 3000 C   with relay output 21    Maintenance and fault rectification    22    Information    Make sure that the housing is not rotated during the electronics  exchange  Otherwise the plug may be in a different position  later     9 Insert the oscillator carefully  Make sure that the plug is in  the correct position     10 Screw in and tighten the two screws with a Phillips  screwdriver     11 Insert the wire ends into the open terminals according to  the wiring plan    12 Close the opening levers of the terminals  you will hear the  terminal spring closing    13 Check the hold of the wires in the terminals by lightly  pulling on them    14 Check the tightness of the cable entry  The seal ring must  completely encircle the cable     15 Screw the housing cover back on    The electronics exchange is finished     7 4 Instrument repair  If a repair is necessary  please proceed as follows     You can download a return form from our Internet homepage  http   www krohne mar com fileadmin media lounge PDF   Download Specimen_e pdf     By doing this you help us carry out the repair quickly and  without having to call for additional information     e Print and fill out one form per instrument   e Clean the instrument and pack it damage proof   e Attach the completed form and possibly also a safety data  sheet to the instrument    OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112    Dismounting    8 Dismounting  8 1 Dismounting steps    Warnin
2.    1     6 bar   100     600 kPa    14 5     87 psi    50      100  C   58      212  F     gt 0 08 g cm    gt 0 003 Ibs iN        e 1xcable entry M20x1 5  cable  5    9 mm    1x blind stopper M20x1 5  attached 1x  cable entry M20x1 5   or    e 1x cable entry   NPT  1x blind stopper  Yo NPT  1x cable entry 1   2 NPT    or   e 1x plug M12x1  1x blind stopper M20x1 5  for wire cross section up to 1 5 mm   0 0023 in      min  detection or dry run protection  max  detection or overfill protection    20     253 V AC  50 60 Hz  20     72 V DC  with  U  gt 60 V DC the ambiet temperature can be  max   50  C  122  F    1    8 VA  AC   approx  1 3 W  DC     25    Supplement    Electrical protective measures    Protection  Overvoltage category  Protection class    26    IP 66 IP 67  III       OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112     KROHNE    Supplement    9 2 Dimensions    OPTISWITCH 3000 C          1 13 64               220 5mm  8 11 46               Fig  11  OPTISWITCH 3000 C   Threaded version G1   2A    D All dimensions in mm  inch      OPTISWITCH 3000 C   with relay output 27       Supplement    9 3 Certificates    CE declaration of conformity             CE    Konformit  tserkl  rung    Declaration of conformity  D  claration de conformit      Krohne SAS   Le Ors HP 9S  F 26103 Romans Cedex    erkl  rt in alleiniger Verantwortung  da   das Produkt   declare under  our sole responsibility that our product   d  clare sous sa seule  responsab
3.  Adjustment    OPTISWITCH 3000 C can detect products with a density   gt 0 08 g cm    gt 0 003 Ibs iN         On the electronics module you will find the following indicating  and adjustment elements     e signal lamp for indication of the switching condition  green   red   e mode switch for selection of the output current    3 4 Storage and transport    Your instrument was protected by packaging during transport   Its capacity to handle normal loads during transport is assured  by a test acc  to DIN 55439     The packaging of standard instruments consists of environ   ment friendly  recyclable cardboard  In addition  the sensor is  provided with a protective cover of cardboard  For special  versions PE foam or PE foil is also used  Dispose of the  packaging material via specialised recycling companies     OPTISWITCH 3000 C   with relay output 7    Product description    Storage and transport tem   perature    e Storage and transport temperature see  Supplement    Technical data   Ambient conditions     e Relative humidity 20     85      OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112    Switching point    Moisture    Transport    Pressure Vacuum    Handling    Mounting    4 Mounting  4 1 General instructions    In general  OPTISWITCH 3000 C can be mounted in any  position  The instrument must be mounted in such a way that  the vibrating element is at the height of the requested switching  point     Use the recommended cable  see chapter  Connec
4.  the vessel where  no dead zones or mounds can form      gt  Check if the vibrating element is covered by buildup on  the socket       Wrong mode selected     gt  Set the correct mode on the mode switch  max  overfill  protection  min  dry run protection   Wiring should be  carried out acc  to the quiescent current principle     OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112         Maintenance and fault rectification    Signal lamp flashes red    Electronics has detected a failure     gt  Exchange instrument or return it for repair    7 3 Exchanging the electronics    In general  all oscillators of series VB60 can be interchanged  If  you want to use an oscillator with a different signal output  you  can download the corresponding operating instructions man   ual from our homepage under Downloads     Proceed as follows     1    2  3  4  5       Switch off power supply   Unscrew the housing cover   Lift the opening levers of the terminals with a screwdriver  Pull the connection cables out of the terminals    Loosen the two screws with a Phillips screwdriver  size 1              Fig     N O NS    10  Loosen the screws  Electronics module  Screws  2 pcs      Remove the old oscillator    Compare the new oscillator with the old one  The type label  of the oscillator must correspond to that of the old  oscillator     Compare the settings of the two oscillators  Set the  adjustment elements of the new oscillator to the settings of  the old oscillator
5. Instrumentrepaif       SES Es Se see 23  8 Dismounting  8 1 Dismounting steps    24  8 2 Disposal  eg ge dh cach oem ace Ae eae 24    OPTISWITCH 3000 C   with relay output    GL LO90 N3A 6S00       30059 EN 060112    Contents   9 Supplement  91 Technical data  25  92 Dimensiong  e  28  9 3 Certificates e  29    Supplementary operating instructions manuals    Information    OPTISWITCH 3000 C is available in different versions   Depending on the selected version  supplementary operating  instructions manuals will also be included in the delivery  The  supplementary operating instructions manuals are listed in  section  Product description      Operating instructions manuals for accessories and  replacement parts   Tip    To ensure reliable use and operation of your OPTISWITCH  3000 C we offer accessories and replacement parts  The  corresponding operating instructions manuals are     e Operating instructions manual  Oscillator     OPTISWITCH 3000 C   with relay output 3    About this document    EI Pme    d    1 About this document    1 1 Function    This operating instructions manual has all the information you  need for duick setup and safe operation  Please read this  manual before you start setup     1 2 Target group    This operating instructions manual is directed to trained  personnel  The contents of this manual should be made  available to these personnel and put into practice by them     1 3 Symbolism used    Information  tip  note  This symbol indicates helpful addit
6. Operating Instructions    OPTISWITCH 3000 C  with relay output       Variable area flowmeters   Vortex flowmeters   Flow controllers   Electromagnetic flowmeters  Ultrasonic flowmeters   Mass flowmeters   Level measuring instruments  Communications engineering  Engineering systems  amp  solutions    Switches  counters  displays and recorders    Heat metering  Pressure and temperature       Contents    Contents  1 About this document  11  FUNCION AE OE rte OE e 5  12 Target group             0 2 eee 5  13 Symbolism use  5  2 For your safety  2 1 Authorised personnel  6  2 2 Appropriate use naaa aaaea 6  23 Warning about misuse                    6  2 4 General safety instructions                 6  2 5 CE conformity       n a aana aeaa 6  3 Product description  3 1 Configuration       aaa a aaa aeae 7  3 2 Principle of operation                     7  3 3 Adjustment            0 0 00  cee ee 8  3 4 Storage and transport                 05  8  4 Mounting  4 1 General instructions                 006 10  4 2 Mounting information                  46  11  5 Connecting to power supply  5 1 Preparing the connection                  14  5 2 Connection procedure                 05  14  5 3 Wiring plans  signal chamber housing         16  6 Setup  6 1 General  18  6 2 Adjustment elements                  5  18  6 3 Function chat    19  7 Maintenance and fault rectification  7 1 Maintenance    21  7 2 Fault rectification    000 21  7 3 Exchanging the electronics                 22  7 4 
7. are parallel to the product movement              Fig  4  Orientation of the tuning fork in case of flow  1 Marking with screwed version  2 Direction of flow    OPTISWITCH 3000 C   with relay output 11    Mounting    Adhesive products    12    In case of horizontal mounting in adhesive products  the  surfaces of the tuning fork should be vertical in order to reduce  buildup on the tuning fork  On the screwed version you will find  a marking on the hexagon  With this  you can check the  position of the tuning fork when screwing it in  When the  hexagon touches the seal  the thread can be still turned by  approx  half a turn  This is sufficient to reach the recom   mended installation position                                                                                         Fig  5  Vertical installation   marking  1 Marking on top with screwed version    OPTISWITCH 3000 C   with relay output    GL LO9O NA 6SOOE       30059 EN 060112    Note safety instructions    Select power supply    Select connection cable    Connecting to power supply    5 Connecting to power supply  5 1 Preparing the connection    Always observe the following safety instructions   e Connect only in the complete absence of line voltage    Connect the power supply acc  to the following diagrams   Oscillator VB 6OR is designed in protection class 1  To  maintain this protection class  it is absolutely necessary that  the ground conductor be connected to the internal ground  terminal  Take note of the ge
8. ed    Wen   3 4 5    7   8   Relay energized   fg   3 4 5    7   8     Relay deener   gized       OPTISWITCH 3000 C   with relay output    GL LO9O NA 6SOOE       30059 EN 060112       Setup    Failure of the    supply voltage In 1   min  max  mode     ei o ei    Relay deener   gized    Failure any        3 4 5 O   6   7   8   Relay deener  flashes red  gized    OPTISWITCH 3000 C   with relay output 19    Maintenance and fault rectification    Checking the switching signal    20    7 Maintenance and fault rectification    7 1 Maintenance    When used as directed in normal operation  OPTISWITCH  3000 C is completely maintenance free     7 2 Fault rectification    d OPTISWITCH 3000 C signals  covered  when the vibrating  element is not submerged  overfill protection     d OPTISWITCH 3000 C signals  uncovered  when the  vibrating element is submerged  dry run protection       Supply voltage too low   gt  Check the power supply    Electronics defective     gt  Push the mode switch  min  max    If the instrument  then changes the mode  the instrument may be  mechanically damaged  Should the switching function  in the correct mode still be faulty  return the instrument  for repair      gt  Push the mode switch  If the instrument then does not  change the mode  the oscillator may be defective   Exchange the oscillator      gt  Check if there is buildup on the vibrating element  and if  so  remove it       Unfavourable installation location        Mount the instrument at a location in
9. er must take note of the safety instructions in this operating  instructions manual  the country specific installation standards   e g  the VDE regulations in Germany  as well as all prevailing  safety regulations and accident prevention rules     2 5 CE conformity    OPTISWITCH 3000 C is in CE conformity with EMC  89 336   EWG   fulfils the NAMUR recommendation NE 21 and is in CE  conformity with NSR  73 23 EWG      Conformity has been judged acc  to the following standards     e EMC       Emission EN 61326  1997  class B       Susceptibility EN 61326  1997 A1  1998    e LVD  EN 61010 1  2001    OPTISWITCH 3000 C   with relay output 5    Product description    Scope of delivery    Components    Area of application    3 Product description  3 1 Configuration    The scope of delivery encompasses     e OPTISWITCH 3000 C level sensor  e Documentation      this operating instructions manual      Supplementary instructions manual    Plug connector for  level sensors    optional    OPTISWITCH 3000 C consists of the following components     e Housing cover  e Housing with electronics  e process fitting with tuning fork             Fig  1  OPTISWITCH 3000 C   with plastic housing  1 Housing cover   2 Housing with electronics   3 Process fitting    3 2 Principle of operation    OPTISWITCH 3000 C is a level sensor with tuning fork for level  detection     It is designed for industrial use in all areas of process  technology and is preferably used for bulk solids     Typical application
10. g    Before dismounting  be aware of dangerous process con   ditions such as e g  pressure in the vessel  high temperatures   corrosive or toxic products etc     Take note of chapters  Mounting  and  Connecting to power  supply  and carry out the listed steps in reverse order     8 2 Disposal    OPTISWITCH 3000 C consists of materials which can  recycled by specialised recycling companies  We have  purposely designed the electronic modules to be easily  separable     WEEE directive 2002 96 EG   This instrument is not subject to the WEEE directive 2002 96   EG and the respective national laws  in Germany  e g   ElektroG   Pass the instrument directly on to a specialised  recycling company and do not use the municipal collecting  points  These may only be used for privately used products  acc  to the WEEE directive     Correct disposal avoids negative effects to persons and  environment and ensures recycling of useful raw materials     Materials  see  Technical data     If you cannot dispose of the instrument properly  please  contact us about disposal methods or return     OPTISWITCH 3000 C   with relay output 23    Supplement    9 Supplement    9 1 Technical data    General data    Material 316L corresponds to 1 4404 or 1 4435    Materials  wetted parts      Process fitting   Thread        Process seal   Thread        Tuning fork    Materials  non wetted parts        Housing        Seal ring between housing and    housing cover        Ground terminal    Weight  Max  lateral l
11. ilit   que le produit    OPTISWITCH 3000 C  mit Relaisausgang   with relay output   avec sortie relais  VB60R   mit Transistorausgang   with transistor output   avec sortie transistor  VB60T   mit Zweileiterausgang   with two wire output   avec sortie bifilaire  VB60Z     auf das sich diese Erkl  rung bezieht  mit den folgenden Normen     bereinstimmt   to which this declaration relates is in conformity   with the following standards   auquel se ref  re cette declaration  est conforme aux normes    Emission   Emission   Emission   EN 61326   1997  Klasse B   Immission   Susceptibility   Immission     EN 61326  1997   A1   1998  EN 61010     1  2001    gem     den Bestimmungen der Richtlinien   following the provision  of Directives   conform  ment aux dispositions des Directives    73 23 EWG  89 336 EWG    Ges g  03 02 2005 ra Aa    LN p p  Florian Steneele    Fig  12  CE declaration of conformity    28    OPTISWITCH 3000 C   with relay output    GL LO9O NA 6SOOE       30059 EN 060112    OPTISWITCH 3000 C   with relay output    Supplement    29    Supplement    30    OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112    OPTISWITCH 3000 C   with relay output    Supplement    31    Subject to change without notice    30059 EN 060112    
12. ional information     Caution  warning  danger   This symbol informs you of a dangerous situation that could  occur  Ignoring this cautionary note can impair the person and   or the instrument     Ex applications  This symbol indicates special instructions for Ex applications     List  The dot set in front indicates a list with no implied sequence     Action  This arrow indicates a single action     Sequence  Numbers set in front indicate successive steps in a procedure     OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112    For your safety    2 For your safety    2 1 Authorised personnel    All operations described in this operating instructions manual  must be carried out only by trained  specialist personnel  authorised by the operator  For safety and warranty reasons   any internal work on the instruments must be carried out only  by personnel authorised by the manufacturer     2 2 Appropriate use  OPTISWITCH 3000 C is a sensor for level detection     Detailed information on the application range of OPTISWITCH  3000 C is available in chapter  Product description      2 3 Warning about misuse    Inappropriate or incorrect use of the instrument can give rise to  application specific hazards  e g  vessel overfill or damage to  system components through incorrect mounting or adjustment     2 4 General safety instructions    OPTISWITCH 3000 C is a high tech instrument requiring the  strict observance of standard regulations and guidelines  The  us
13. is applies particularly to use  with adhesive products     Install the instrument in such a way that the tuning fork does  not protrude directly into the filling stream  Should such an  installation location be necessary  mount a suitable baffle  above or in front of the tuning fork     If such an installation location should be necessary  mount a  suitable protective sheet above or in front of the vibrating  element   see illustration a       In abrasive solids  mounting acc  to illustration b  has proven  A  spout forms in the concave protective sheet preventing wear of  the protective sheet     To achieve a very precise switching point  you can install  OPTISWITCH 3000 C horizontally  However  if the switching  point can have a tolerance of a few centimeters  we  recommend mounting OPTISWITCH 3000 C approx  20    inclined to the vessel bottom to avoid buildup     OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112    Inflowing material    Flow    Mounting                Fig  3  Horizontal installation  a Protective sheet  b Concave protective sheet for abrasive solids    If OPTISWITCH 3000 C is mounted in the filling stream   unwanted switching signals can be generated  Mount OPTI   SWITCH 3000 C at a location in the vessel where no disturbing  influence from e g  filling openings  agitators  etc  can occur     To minimise flow resistance caused by the tuning fork   OPTISWITCH 3000 C should be mounted in such a way that  the surfaces of the fork 
14. l lamp  4     Note    As a rule  always set the mode with the mode switch  1  before  starting setup OPTISWITCH 3000 C  The switching output will  change if you set the mode switch  1  afterwards  This could  possibly trigger other connected instruments or devices     6 2 Adjustment elements                   Fig  9  Oscillator   Relay output   1 DIL switch for mode adjustment  2 Ground terminal   3 Terminals   4 Control lamp    With the mode adjustment  min  max   you can change the  switching condition of the relay  You can set the required mode    OPTISWITCH 3000 C   with relay output 17    Setup    Signal lamp  4     18    acc  to the  Function chart   max    max  detection or overfill  protection  min    min  detection or dry run protection      We recommend connecting acc  to the quiescent current  principle  replay contact deenergizes when reaching the  switching point   because the relay takes on the same  safe   condition if a failure is detected     Control lamp for indication of the switching condition        green   relay energized  e red   relay deenergized  e red  flashing    failure    6 3 Function chart    The following chart provides an overview of the switching  conditions depending on the adjusted mode and level     max  mode  Overfill protection    max  mode  Overfill protection    min  mode  Dry run protection    min  mode  Dry run protection                Vly    20 gt     N    o ei    l  Relay energized Green    rc   3 4 5   7   8     Relay deener   giz
15. neral installation regulations  As a  rule  connect OPTISWITCH 3000 C to vessel ground  PA   or  in case of plastic vessels  to the next ground potential  On the  side of the instrument housing there is a ground terminal  between the cable entries  This connection serves to drain off  electrostatic charges     The data for power supply are stated in the Technical data in  the Supplement     OPTISWITCH 3000 C is connected with standard cable with  round cross section  An outer cable diameter of 5    9 mm   0 2     0 35 in  ensures the seal effect of the cable entry     If cable with a different diameter or wire cross section is used   exchange the seal or use an appropriate cable connection     5 2 Connection procedure   Proceed as follows    1 Unscrew the housing cover   2 Loosen compression nut of the cable entry    3 Remove approx  10 cm  4 in  of the cable mantle  strip  approx  1 cm  0 4 in  insulation from the ends of the  individual wires    4 Insert the cable into the sensor through the cable entry    5 Lift the opening levers of the terminals with a screwdriver   see following illustration     6 Insert the wire ends into the open terminals according to  the wiring plan    7 Press down the opening levers of the terminals  you will  hear the terminal spring closing    OPTISWITCH 3000 C   with relay output 13    Connecting to power supply    14    8 Check the hold of the wires in the terminals by lightly  pulling on them    9 Tighten the compression nut of the cable entr
16. oad    Output variable    Output  Turn on voltage        min       max     Switching current      min        max    Breaking capacity      min        max     24    plastic PBT  Polyester   Klingersil C 4400  316L    plastic PBT  Polyester   silicone    316L    1500 g  53 oz   600 N  135 Ibf  longitudinal to the fork side    relay output  DPDT   2 floating spdts    10 mV  253 V AC  253 V DC    10 HA  3AAG 1ADG    50 mW    750 VA AC  54 W DC   If inductive loads or stronger currents are  switched through  the gold plating on the relay  contact surface will be permanently damaged   The contact is then no longer suitable for  switching low level signal circuits     OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112    Contact material  relay contacts   Modes  adjustable   Integration time        when immersed      when laid bare    Ambient conditions    Ambient temperature on the housing  Storage and transport temperature    Process conditions    Parameter   Process pressure  Process temperature  Density    Electromechanical data    Cable entry plug  dependent on the version       Single chamber housing    Spring loaded terminals    Adjustment elements  Mode switch       min        max     Voltage supply  Power supply    Power consumption    OPTISWITCH 3000 C   with relay output    Supplement    AgNi or AgSnO and Au plated  min max    approx  0 5 s  approx  1s     40      80  C   40       40      80  C   40         176  F    176  F     level of solids 
17. s are overfill and dry run protection  Thanks  to its simple and robust measuring system  OPTISWITCH  3000 C is virtually unaffected by the chemical and physical  properties of the solid     It functions even when exposed to strong external vibration or  changing products     OPTISWITCH 3000 C   with relay output    col LO90 N3 6S00       30059 EN 060112    Physical principle    Power supply    Packaging    Product description    Fault monitoring  The electronics of OPTISWITCH 3000 C continuously mon   itors the following criteria     e Correct vibrating frequency  e Line break to the piezo drive    If one of these faults is detected or in case the power supply  fails  the electronics takes on a defined switching condition  i e   the relay deenergizes  safe condition      The tuning fork is piezoelectrically energised and vibrates at its  mechanical resonance frequency of approx  150 Hz  When the  tuning fork is submerged in the product  the vibration  amplitude changes  This change is detected by the integrated  oscillator and converted into a switching command     OPTISWITCH 3000 C is a compact instrument  i e  it can be  operated without external evaluation system  The integrated   electronics evaluates the level signal and outputs a switching  signal  With this switching signal  a connected device can be  operated directly  e g  a warning system  a PLC  a pump etc       The exact range of the voltage supply is specified in the   Technical data  in the  Supplement      3 3
18. ting to  power supply   and tighten the cable gland     You can give your OPTISWITCH 3000 C additional protection  against moisture penetration by leading the connection cable  downward in front of the cable entry  Rain and condensation  water can thus drain off  This applies mainly to mounting  outdoors  in areas where moisture is expected  e g  by  cleaning processes  or on cooled or heated vessels                    Fig  2  Measures against moisture penetration    Do not hold OPTISWITCH 3000 C on the vibrating element   The sensor can be damaged by the weight of the instrument     Remove the protective cover just before mounting     The process fitting must be sealed if there is gauge or low  pressure in the vessel  Before use  check if the seal material is  resistant against the measured product and the process  temperature     The vibrating level switch is a measuring instrument and must  be treated accordingly  Bending the vibrating element will  destroy the instrument     OPTISWITCH 3000 C   with relay output 9    Mounting    Socket    Filling opening    Horizontal installation    10    Warning    The housing must not be used to screw the instrument in   Applying tightening force on the housing can damage its  internal mechanical parts     To screw in  use the hexagon above the thread     4 2 Mounting information    The vibrating element should protrude into the vessel to avoid  buildup  For that reason  avoid using mounting bosses for  flanges and screwed fittings  Th
19. y  the seal  ring must completely encircle the cable    10 Screw the housing cover back on    The electrical connection is finished              Fig  6  Connection steps 5 and 6    OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112     KROHNE   Comestngtopowersuppy    5 3 Wiring plans  signal chamber housing                      Electronics and connection  compartment  Fig  7  Electronics and connection compartment  1 DIL switch for mode adjustment  2 Ground terminal  3 Terminals  4 Control lamp  Wiring plan We recommend connecting OPTISWITCH 3000 C in such a    way that the switching circuit is open when there is a level  signal  line break or failure  Safe condition      The relays are always shown in non operative condition     OPTISWITCH 3000 C   with relay output 15    Connecting to power supply    16       min max N                                                             CH   cl   ol   ol                                                                                                        Fig  8  Wiring plan  1 Relay output   2 Relay output   3 Voltage supply    OPTISWITCH 3000 C   with relay output    GL LO90 N3 6S00       30059 EN 060112    Function Configuration    Mode adjustment  1     Setup    6 Setup    6 1 General    The numbers in brackets refer to the following illustrations     On the electronics module you will find the following indicating  and adjustment elements     e DIL switch for mode adjustment   min  max   1   e Signa
    