RisøScan 1.0 - User manual and verification toolset

image

Contents

1. 80 Table B4 Dose values Dose measurement Response value Assigned Dose Ris Scan 20 1 40 3 60 5 80 7 100 10 120 13 140 16 160 20 180 25 200 30 220 41 240 80 19 0 241 0 64 Report number ex Ris R 1234 EN APPENDIX C How to perform Test 3 Ne Oe EOD en 10 11 12 13 14 15 16 17 18 19 20 2 22 23 24 Open the image Test3 bmp In the Image properties window set the DPI 100 Set the weighting factors for the RGB values to 0 1 0 Press Accept From the main window select the Select calibration button Select the calibration which was saved during Test 2 Press Select button From the main window select the Reference for image button Press Ctrl A to select the whole image Press Ctrl Z or right click to go into acquisition mode While clicking select an area inside the second rectangle at the bottom of the image This is the first reference value See Figure C1 Press the add button Select an area inside the next rectangle Press the second line in the upper table of calibration reference values B this will transfer the name of the reference to the table of image reference values middle one Press add Repeat the instructions from line 10 for the next two reference values Don t use the last rectangle Make sure that the lower table wi
2. Repeat the instructions from line 4 until all the weighting factors in table 1 has been evaluated Compare the result noted in table Al with the data in table 2 in the main text of this document Agreement in all values indicates that RiseScan functions as intended Press Done to return to the main window age properties TEST1 I it a t Figure A1 Weighting factors are typed in the upper right corner when an image is opened www risoe dk Figure A2 Press Ctrl A to select the whole image Then if the program is in Selection mode instead of Acquisition mode see upper left corner of the screen press Ctrl Z to change The response is written in the lower right corner of the screen Table A1 Weighting factors Measured response value Wa Wa We Square 1 Square2 Square 3 0 0 1 0 1 0 0 1 1 1 0 0 1 0 1 1 1 0 1 1 1 60 Report number ex Ris R 1234 EN APPENDIX B How to perform Test 2 NOR GJEV ca E Or 11 12 Open the image Test2 bmp In the Image properties window set the Scan Resolution DPI 100 Set the weighting factors for the RGB values to 0 1 0 Press Accept From the main window select the Perform calibration button Press Ctrl A to select the whole image Press Ctrl Z or right click to go into acquisition mode While holding down the left button select an area inside the first square upper left
3. A as this will be the first of the reference values Press Add Repeat the instructions from line 15 for the next three rectangles exclude the last one Give them the names B C and D Select the Create a calibration button to return to the calibration Press Save calibration reference to save the calibration Give the calibration the name Test calibration or something similar as long as it is not confused with a real calibration When back in the calibration window in the File menu select Load calibration This is necessary to be able to read dose values from the image Press the Residual button to view the residuals Press Ctrl P to make a print Make sure the Print information about calibration button is checked Note the residuals in Table B3 Compare the residuals in Table B3 with those listed in Table 7 Agreement in all values indicates that Ris Scan functions as intended Press Done to return to the main window Press Surface profile Make sure the program is in acquisition mode and the whole image is selected if not press Ctrl A and Ctrl Z www risoe dk 25 Show the image intensity as dose Press Ctrl M or Display gt Map dose 26 Click on each of the big squares and read the dose Note the doses in Table B4 Make a screen dump Ctrl S if necessary 27 Click on the first and last of the rectangles and note the dose value in table B4 Make a screen
4. Changed response value rounded to nearest integer The calibration created in Test 2 is assigned to the image used in Test 3 The reference values are pointed out and the reference correction enabled Each of the squares is measured using the tool in the Surface profile section of the program The squares are measured without and with the use of the reference correction and the results are shown in Table 11 and 12 50 Report number ex Ris R 1234 EN Table 11 Dose values calculated using Excel Original Dose without Dose with response value correction correction 20 1 05 1 00 40 3 12 2 90 60 5 37 5 02 80 7 85 7 33 100 10 46 9 89 120 13 09 12 71 140 16 05 16 05 160 19 27 19 92 180 22 83 24 59 200 27 47 30 67 220 34 56 40 24 240 47 21 74 99 Out of range check 19 1 00 1 00 241 49 48 80 00 Table 12 Dose values measured using Ris Scan Original Dose without Dose with response value correction correction 20 1 05 0 00 40 3 12 2 91 60 5 37 5 02 80 7 85 7 31 100 10 46 9 87 120 13 09 12 70 140 16 05 16 07 160 19 27 19 96 180 22 83 24 60 200 27 47 30 69 220 34 56 40 44 240 47 21 0 00 Out of range check 19 0 00 0 00 241 49 48 0 00 The agreement within 0 596 between dose calculated by Excel and RiseScan verifies that RiseScan functions The small differences
5. RISO Ris R 1487 EN Ris Scan 1 0 User Manual and Toolset for Retrospective Validation Jakob Helt Hansen Ris National Laboratory Roskilde Denmark December 2004 Author Jakob Helt Hansen Title RisoScan 1 0 User Manual and Toolset for Retrospective Validation Department Radiation Research Department Abstract max 2000 char The RiseScan software is used for dose measurements with radiochromic films that color visibly This report consists of two documents for use with the RiseScan software The User Manual tells how to use the program and the Toolset for Retrospective Validation describes how to perform a retrospective validation of the software Ris R 1487 EN December 2004 ISSN 0106 2840 ISBN 87 550 3387 3 Contract no Group s own reg no Sponsorship Cover Pages 72 Tables 30 References 3 Rise National Laboratory Information Service Department P O Box 49 DK 4000 Roskilde Denmark Telephone 45 46774004 bibl risoe dk Fax 45 46774013 www risoe dk Contents RiseScan 1 0 User manual 5 1 Introduction 7 1 1 Disclaimer and general precautions 8 1 2 Retrospective Validation 8 2 Requirements 9 3 Scanner 9 3 1 Reference 9 4 Installation 10 4 Uninstall 10 5 Open an image for analysis 11 5 1 Colour channel RGB selection 12 6 Selecting area of interest 13 6 1 General settings 14 7 Create a calibration 16 7 1 Select order of polynomial f
6. 9 right Figure 9 Left Order of polynomial fit Right Residual plot If you don t want to associate a reference to the calibration click on the Save calibration reference at this point 7 3 Include reference in calibration data It is optional to associate a reference to a calibration but it is strongly recommended since it gives the opportunity to check that scanner settings are correct and minor variations due to instabilities of the scanner can be corrected In selection mode select the area covering the reference Right click to get into acquisition mode Set the Create a Calibration Point out calibration Reference values button in the Point out calibration Reference values position See Figure 8 or 10 www risoe dk Write the name of the reference in the Specify the name of the reference field in this example Test Ref 1 Figure 10 Select an area of each of the colours of the reference Enter a name for each colour field e g R1 R2 and R3 Click on Add to enter the response of each reference colour in the table The standard deviations in response value and in percent are also displayed Ris Scan Calib B3 4 B3 4 j f y RI Y 108 86 Rz flies 282 16 p een Delete or Edt tined fv doo Test Ref 1 Figure 10 Measurement of reference values for the calibration function 7 4 Save calibration To save the
7. APPENDIX A How to perform Test 1 59 APPENDIX B How to perform Test 2 61 APPENDIX C How to perform Test 3 65 APPENDIX D How to perform Test 4 67 APPENDIX E How to perform Test 5 69 APPENDIX F How to perform Test 6 71 4 Report number ex Ris R 1234 EN RisoScan 1 0 User manual www risoe dk Report number ex Ris R 1234 EN 1 Introduction Ris Scan is a program for analysis of scanned images of radiochromic thin film dosimeters that colours visibly such as Rise B3 Far West and GAF The program is developed and maintained at Rise High Dose Reference Laboratory The software is distributed and supported worldwide by the GEX Corporation Ris Scan includes tools for Selection of weights for red green and blue colour channel of the scanned image to obtain an optimal signal to noise ratio Calibration pixel response value versus dose Inclusion of a reference to verify that scanner settings have been chosen correctly Minor variations due to scanner instability can be corrected for Depth dose profile analysis and calculation of electron radiation energy based on range measurements in aluminium Analysis of dose profiles including search for minimum and maximum values 3D surface plots of dose profiles Report generation For customer service and technical questions contact Customer Service GEX Corporation 7330 S Alton Way Suite 12 i Centennial Colorado 80112 USA Website www gexcorp co
8. Relationship to FDA guidelines 1 1 General FDA guidelines Software used in production of medical devices or in the quality system of the manu facture shall be validated according to FDA guidelines The purpose of the validation is confirmation by examination and provision of objective evidence that software specifications conform to user needs and intended uses and that the particular requirements implemented through software can be consistently fulfilled Software validation usually consists of several separated tasks many of which are performed by the software developer To prevent errors in the final program code the developer shall use a set of documents to verify the successful completion of separate tasks in the development process These documents shall at least include a quality manual describing the intended development process a system requirements specification document describing requirements to the functionality of the program a software design specification describing how the requirements are implemented and a test scheme documenting that the output from modules of the program and the program itself is as expected The amount of documentation needed to validate a specific piece of software depends on the complexity of the program and the risks associated with its use 1 2 Retrospective validation As it is the case with RiseScan it is not always possible to present all development documentation needed for a proper val
9. The response value of the square is shown in box above the Add button See Figure B1 Type in the first dose value listed in Table B1 i e 1 Press Add Repeat the instructions from line 6 for every of the remaining 11 big squares Select the 2 order fit just left of the calibration function graph See Figure B1 Read the coefficients of the polynomial fit either on the screen just above the calibration function graph R 2 value is displayed in the upper right corner of the graph or by printing the values Press Ctrl P Note the values in Table B2 and make a screen dump Ctrl S if necessary Compare with Figure 3 and Table 5 in the main text of this document Agreement in all values indicates that RiseScan functions as intended Repeat the instructions from line 9 choosing the third and fourth order fit Instructions in line 13 17 tell how to store the reference values These are needed in Test 3 The middle four rectangles at the bottom line of the image are used for this The first and the last are used for out of range tests 13 14 15 16 17 18 19 20 2 22 23 24 Make sure that the 4 order fit is selected Select Point out calibration Reference values just below the image See Figure B2 While holding down the left button select an area inside the second rectangle The response value of that square is shown in the line below the table to the right Type the letter
10. 16 0 Set the weighting factors for the RGB values to 0 1 0 Press Accept From the main window select the Depth profile button Press Ctrl A to select the whole image Press Ctrl Z or right click to go into acquisition mode Make sure that the equations for average energy Rso Eso is set to be 6 2 x and the equation for most probable energy Rex Eex is set to be 5 09 x 0 2 The equations are displayed in the box below the graph If you are using another length unit than centimetres select cm as length unit Display gt Length unit The calculation of average and most probable energy is independent of the length scale but the cm setting is necessary here to evaluate the position of the segment selected for the linear regression In the Display menu select Settings for slope algorithm make sure that the first setting is 1 Press Accept See Figure Fl Press Offset and Slope button See Figure F2 The x axis has been shifted 0 2 cm if the starting point of the wedge thin spike in the response profile is placed at 0 0 cm depth Note the shift in Table F1 Read the interval of the linear fit as the positions of the vertical lines in the graph Note the start and end point of the linear fit L R in Table F1 Read the y axis offset and the maximum response value Y max in the upper right corner of the graph Note the values in Table F1 Read the values for Rso Es and Rex Eex in th
11. 3304E 01 9 8940E 01 3 9 4789E 04 1 7198E 01 1 0557E 01 1 0806E 01 9 9960E 01 4 5 5924E 06 1 7212E 03 2 0279E 01 1 0949E 01 9 7642E 00 9 9964E 01 Results from Ris Scan Table 5 Coefficients of the polynomial fit calculated using Ris Scan Coefficient Order 4 3 2 1 OR 2 2 6 4472E 02 7 8276E 00 2 3304E 01 9 8940E 01 3 9 4789E 04 1 7198E 01 1 0557E 01 1 0806E 01 9 9960E 01 4 5 5924E 06 1 7212E 03 2 0279E 01 1 0949E 01 9 7642E 0009 9964E 01 The exact agreement between coefficients calculated by Excel and Ris Scan verifies that Ris Scan functions 46 Report number ex Ris R 1234 EN Selecting the 4 order fit the following residuals are calculated in Excel and RiseScan respectively Table 6 and 7 Table 6 Residuals calculated using Excel Dose Image response Calculated response Calculated residual 96 1 20 20 5 2 56 3 40 40 8 2 08 5 60 59 6 0 58 7 80 77 0 3 69 10 100 100 6 0 64 13 120 121 4 1 21 16 140 139 7 0 21 20 160 160 5 0 31 25 180 181 4 0 80 30 200 197 7 1 17 41 220 220 6 0 27 80 240 240 0 0 00 Table 7 Residuals calculated using Ris Scan Dose Image response Measured residual 1 20 2 56 3 40 2 08 5 60 0 58 7 80 3 69 10 100 0 64 13 120 1 20 16 140 0 21 20 160 0 31 25 180 0 80 30 200 1 17 41 220 0 27 80 240 0 00 The 4 order polynomial fit
12. Image The scanned image as it is shown in Map dose mode i e intensity equal dose Print only the selected part of the image Only the selected part of the image is shown Scale graphs by a factor of The size of the graphics printed can be adjusted here Print the image as it appears on the screen Prints the image including guidelines etc Is useful to document which part of the image is used for the measurement Figure 26 Send report to default windows printer www risoe dk 13 2 Report generation HTML file The report can also be written in HTML file format by setting the switch to Write report to HTML file see Figure 27 Below the switch two boxes appear specifying the directory to write the HTML and graphics files and the name of the HTML file Two check boxes give the option to show the HTML report in the default web browser after the report is made and to clean up the directory to which the HTML file is written this is useful if an old report is to be replaced Scan 125125150dpi sq2b 3D Surface profile RisoeScan report html i 3 Figure 27 Write the report as a HTML file 13 3 Graph values to ASCII file Clicking on Ctrl W or File gt Graph to file gives the opportunity to write data from the graphs of the current task to an ASCII file The user can decide the number format if comments should be included and if there is more than one graph select what data to write Figure 28 30 Re
13. a calibration select Perform calibration on the main menu Left click on the image and select an area which covers the irradiated dosimeters Right click or click on Ctrl M to go into acquisition mode Make sure that the Create a Calibration Point out calibration Reference values switch is in the Create a Calibration position Figure 8 n RisoeScan Calib B3 4 B3 4 Di R 2 NaN Figure 8 Measuring dosimeters for creation of a calibration function Left click on the image and select an area of the first dosimeter Write the dose to which the dosimeter has been irradiated in the field above Add button Click on the return key or the Add button Continue the process with the rest of the dosimeters 16 Report number ex Ris R 1234 EN An entry in the table can be edited by selecting the entry in the table and clicking Edit or delete line 7 1 Select order of polynomial fit The calibration data sets response dose are fitted to a polynomial Use the button at the bottom of the screen to select the order of the polynomial Figure 9 left The polynomial relating dose and response is shown in the lower right figure 7 2 Residual plot The residual plot can be obtained by clicking the switch at the top left of the graph The residuals are calculated using the equation Deatculated Ddelivered Deeliverea 100 The uncertainty of the measured dose should be entered just below the table Figure
14. account for minor fluctuations between scans 4 4 Relationship between response value and dose A calibration is saved as dose and response values and the order of the polynomial used for fitting the response value as a function of dose The coefficients of the polynomial are re calculated each time the image using the calibration is opened The polynomial fit is written as I D 3 aix 3 i 0 where I D response value as function of dose n order of the polynomial a coefficients of the polynomial fit To find a dose value for a given response value the polynomial fit is shifted down by the response value and the first real positive root is found D I min Re root a x I 4 i 0 If a reference correction is used Eq 4 will be rewritten as i R oor D I min Re rool ax T calibration 5 i 0 R meer I where Reatibration linear extrapolation between the response values of the reference chart measured for the calibration R dosimeter linear extrapolation between the response values of the reference chart measured for the unknown dosimeter www risoe dk The first real positive root is found using the LabView subroutine Eval y f x optimal step vi by National Instruments It is verified that dose is be correctly calculated using a given calibration function by using Test 2 It is verified that the correction based on a comparison of reference values is calculated correctly by using Te
15. dump Ctrl S if necessary 28 Compare Table B4 with Table 9 Agreement in all values indicates that RiseScan functions as intended 29 Press Done to return to the main window IE RiseScan TEST2 Test2 po Piz 0 joo 00 Figure Bl After a section of a square has been selected the dose value is written in the box right above the press Add button It is checked that the response value shown next to it has the right value according to Table B1 Then the Add button is pressed After the order of the polynomial has been chosen the coefficients are shown just above the graph 62 Report number ex Riso R 1234 EN Ris Scan TEST2 Test2 Figure B2 The button used to toggle between the input for the calibration and the reference values is located to the left under the image The four middle rectangle are used for reference values and the letters A to D are assigned to each rectangle Table B1 Response value Dose 20 1 40 3 60 5 80 7 100 10 120 13 140 16 160 20 180 25 200 30 220 41 240 80 Table B2 Polynomial coefficients Coefficient www risoe dk Table B3 Residuals 4 order polynomial used The data is obtained using a print out Ctrl P Dose Image response Measured residual 96 01 08 7 10 13 16 20 25 30 41
16. is assigned as a calibration function to the image and the expected dose value for each square is calculated using an Excel Macro supplied by National Physical Lab UK www risoe dk Table 8 shows the calculated doses using Excel and table 9 shows the measured dose values using Ris Scan Table 8 Dose values calculated using Excel Calculated dose Response value Assigned Dose Excel Difference 20 1 VALUE VALUE 40 3 2 92 2 83 60 5 5 04 0 77 80 7 7 36 5 09 100 10 9 91 0 86 120 13 12 78 1 70 140 16 16 05 0 32 160 20 19 90 0 52 180 25 24 61 1 55 200 30 30 85 2 83 220 41 40 60 0 98 240 80 VALUE VALUE 19 0 VALUE VALUE 241 0 VALUE VALUE VALUE means that the calculated dose is not within the dose range 1 80 Table 9 Dose values measured using Ris Scan Dose measurement Response value Assigned Dose Rise Scan Difference 96 20 1 0 00 100 00 40 3 2 92 2 67 60 5 5 04 0 80 80 7 7 36 5 14 100 10 9 91 0 90 120 13 12 78 1 69 140 16 16 05 0 31 160 20 19 90 0 50 180 25 24 61 1 56 200 30 30 85 2 83 220 41 40 60 0 98 240 80 0 00 100 00 19 0 0 00 amp DIV O 241 0 0 00 amp DIV O The agreement within 0 296 between dose calculated by Excel and RiseScan verifies that RiseScan functions The small differences seen in difference are due
17. slope algorithm Figure 21 Normally there should be no reason to change the default values Settings for slope algorithm Figure 21 Settings for the slope algorithm www risoe dk 11 Surface dose profile Clicking on the Surface Profile button on the main menu brings up the dose profile analysis page On the image to the right is a cross hair that can be moved around while left clicking on it On the right is the dose or response value profile along the lines of the cross hair The dose value at the centre of the cross hair is shown at the lower left Figure 22 Ris Scan Depth Profile Scan 125125150dpi sq2b b3 med baggrund ref p X Dose 11 35 kGy at 1 76 4 10 cm Figure 22 Dose profile along the lines of the cross hair is shown in the graphs and the dose value at the centre is shown in the lower left corner By widening the cross hair the graphs will display an average along the width of the cross hair giving less noise in the dose profiles Figure 23 but at the expense of measurement resolution Minimum and maximum values can be found using the Find minimum or Find maximum button The search is performed averaging over an area equal to the centre square of the cross hair Choosing the width of the cross hair is a trade off between resolution and noise reduction The average dose of the centre area of the cross hair shown below the image in Fig 23 may be slightly different from the valu
18. the program The areas not addressed by the verification tests are transfer of data from the program to printer or ASCII files the data displayed in the graphs in the dose profile section of the program and finally the 3 D dose profile section The tests are only intended to verify that Ris Scan functions as intended the tests will not validate the scanner nor the quality of scanned images of dosimeters The tests have been carried out at Rise and the results are given in this document as baseline data The functions and calculations have been tested independently using Excel and the two sets of data are compared This comparison constitutes proof that the tested functions of RisoScan work as intended The Excel spreadsheet used to calculate the independent test is distributed together with this document so the user can see how the results are obtained Parts of the spreadsheet may also serve as templates in validation documents made by the user The user of RiseScan should perform the tests and compare the results with the data provided in this document in order to verify that RiseScan consistently operates as intended A few terms of glossary are defined in the first part of this document followed by a short description of the workflow in the program This gives an overview of the parts of the program on which the tests are focussed Some of the mathematics used in the program is also described Finally follows the tests and a step
19. the size of the cross hair centre square are evaluated The average pixel response is calculated and the location are stored if the value is below above the previous minimum maximum value The purpose of the test is to verify that the minimum maximum search tool works as expected and that it takes search the resolution chosen for the search into account The test image Figure 8 consists of one square with response value 100 In this square is placed three sets of two smaller squares each The first set has the size of 1x1 pixel with the response values 200 and 0 Second set has the size of 5x5 pixel with the response values 190 and 10 Third set has the size of 51x51 pixel with the response values 180 and 20 Figure 8 Image for Test 5 A search for minimum and maximum values is carried out with a resolution of 1x1 pixel 5x5 pixel and 51x51 pixel The search result should place the cross hair at the 1x1 5x5 and 51x51 squares respectively The last search can take a while Table 17 Expected and measured response values using minimum and maximum search tool in Ris Scan Resolution pixels Value sought Expected value Measured value Placement OK 1x1 Minimum 0 0 Y 1x1 Maximum 200 200 Y 5x5 Minimum 10 10 Y 5x5 Maximum 190 190 Y 51x51 Minimum 20 20 Y 51x51 Maximum 180 180 Y www risoe dk The exact agreement between expected and measured value and placement verifies that RiseScan functi
20. 0 200 4 0 200 0 4 0 Table 15 Test 4B Average value and standard deviation calculated using Excel Test 4B Excel Expected Expected std dev Mean response Std dev response mean response response 50 5 0 49 7 5 1 50 1 0 50 0 1 0 100 5 0 99 7 4 9 100 2 0 100 0 2 1 200 5 0 199 8 5 0 200 4 0 200 1 4 0 Table 16 Test 4B Average value and standard deviation calculated using Ris Scan Test 4B Ris Scan Expected Expected std dev Mean of centre Std dev response mean response response response 50 5 0 50 0 4 9 50 1 0 50 0 1 1 100 5 0 100 0 5 0 100 2 0 100 0 2 0 200 5 0 200 0 5 0 200 4 0 200 0 3 9 54 Report number ex Ris R 1234 EN The exact agreement in Test 4A and the good agreement in Test 4B between mean value and standard deviation calculated by Excel and RiseScan verifies that RiseScan functions Small differences in Test 4B are expected due to the random distribution of pixel values 5 5 TEST 5 Minimum and Maximum values Surface profile Refer to Appendix E on how to perform this test The surface profile tool in the RiseScan software has the ability to find minimum and maximum values Varying the width of the centre square of the cross hair chooses the resolution of the search A low resolution is useful to minimize the effect of noise in the image When a search is initiated all possible selections with
21. 234 EN The test image Figure 3 consists of 12 squares Only the green RGB value is used red and blue bit equal zero The squares have the following response values 20 40 60 80 100 120 140 160 180 200 220 240 Six additional squares have the response values 19 20 90 190 240 241 The four middle values are used as reference values while the first and the last are used for checks of out of calibration range Figure 3 Image for Test 2 The following dose values are assigned to each of the 12 response values Table 3 Table 3 Dose values assigned to the response values of test image 2 Response value Dose 20 1 40 3 60 5 80 7 100 10 120 13 140 16 160 20 180 25 200 30 220 41 240 80 www risoe dk Using the data above the polynomial fits of 2 3 and 4 order were calculated by Excel Figure 3 Response value 250 4 200 150 4 100 50 0 20 40 Dose 60 80 Data points 4th order fit 3rd order fit 2nd order fit 100 Figure 4 Calibration function for 2nd 3rd and 4th order fit calculated by Excel The coefficients of the polynomial fit 2 to 4 order is shown in Table 4 and 5 Table 4 Coefficients of the polynomial fit calculated using Excel Coefficient Order 4 3 2 1 OR 2 2 6 4472E 02 7 8276E 00 2
22. The noise in the pixel value distribution is sampled by using the following algorithm which as a result has a number x distributed according to Equation 7 52 Report number ex Ris R 1234 EN Let random 1 and random 2 be two random numbers in the interval 0 1 Calculate r o4 2 In random 1 8 gy 27 random 2 x r cos Q The pixel response value is then eal The standard deviation c is either set to 5 top row in image or 0 02 50 1 0 02 100 2 or 0 02 200 4 lower row Figure 7 Image for Test 4B In both tests a set of 10000 data points rounded to nearest number are generated using the algorithms above These data are analysed in Excel Since the response values are randomly distributed small differences between Excel and Ris Scan results are expected www risoe dk Table 13 Test 4A Average value and standard deviation calculated using Excel Test 4A Excel Expected Expected std dev Mean response Std dev response mean response response 50 5 0 50 0 5 0 50 1 0 50 0 1 0 100 5 0 100 0 5 0 100 2 0 100 0 2 0 200 5 0 200 0 5 0 200 4 0 200 0 4 0 Table 14 Test 4A Average value and standard deviation calculated using Ris Scan Test 4A Ris Scan Expected Expected std dev Mean of centre Std dev response mean response response response 50 5 0 50 0 5 0 50 1 0 50 0 1 0 100 5 0 100 0 5 0 100 2 0 100 0 2 0 200 5 0 200 0 5
23. Z and right click toggles between Selection mode and Acquisition mode Note the difference in colour of the image frame Blue Selection mode Select area of interest Orange Acquisition mode Measurements can be performed on the selected part of the image The mode of operation is also indicated just above the image to the right Figure 5 www risoe dk ot B 125 125 100 dpi Figure 5 Whether the program is in Selection or Acquisition mode is shown above the image to the left and by the colour of the image frame 6 1 General settings In the Display menu the general number format can be changed e g the number of digits shown Figure 6 125 125 100 dpi Figure 6 Changing the way numbers are displayed in the program 14 Report number ex Riso R 1234 EN When a task is chosen e g surface profile the length scale unit can be changed in the Display gt Length unit menu The unit can either be cm inch or pixels Figure 7 Figure 7 Changing length scale IMPORTANT To return to the main menu press the blue Done button Figure 7 or File gt Done www risoe dk 7 Create a calibration The image used to create a calibration should contain dosimeters irradiated to known doses and a reference The recommended reference tablet is made from laminated B3 dosimeter films irradiated to three different doses in the expected dose range e g 12 25 and 50 kGy To create
24. ap dose If an area of the image has been irradiated to a dose outside the limits of the calibration that area will be shown as white An example is shown in Figure 13 150 dpi FarWest FWT 60 810 electron RIAS polystyrene Figure 13 If a response value of a pixel is outside the range of the calibration it will show the dose value 0 i e it appears white 20 Report number ex Ris R 1234 EN 9 Apply reference to an image After a calibration function has been assigned to an image the reference scanned together with the image can be compared with the reference used for the calibration function To measure the reference values for the image click on the Reference for Image button on the main menu Select the part of the image showing the reference and select Acquisition mode To the right of the screen Figure 14 is shown the reference values obtained during the calibration at the top the reference values obtained for the image at the middle and the ratio between the two at the bottom For each reference colour select an area Click on the corresponding reference item in the calibration reference table transferring the name of that reference item to the image reference table Click on the Add button Ris Scan Scan 125125150dpi sq2b b3 med baggrund ref FENE ES Figure 14 Measuring image reference values The graph and the lower table show the ratio between th
25. are due to different curve fitting algorithms The large deviations in the out of range check and in the result for response value 240 is due to the fact that just a small difference in calculated pixel value can be decisive whether or not the calculated dose in inside or outside the calibration range www risoe dk 5 4 TEST 4 Standard deviation Refer to Appendix D on how to perform this test This test is designed to verify that the average and standard deviation of a measurement is calculated correctly The test is divided into two parts A and B The test image for both tests is composed of three sets of two squares The average response values for the three sets are 50 100 and 200 The standard deviation for each set of two squares is either 5 absolute response value or 2 of the average pixel response value In Test 4A the spread about the average pixel response is simply obtained by alternating the pixel response by the standard deviation so that neighbouring pixels have a different value Figure 6 I e the pixel response value is described by r F o 6 where rij response value of pixel i j ij pixel index r average pixel response value o standard deviation Figure 6 Image for Test 4A In the second test Test 4B Figure 7 the pixel values are generated by a algorithm which approximate a Gaussian probability distribution with density 1 x P x ENS p3 7 where o variance
26. ble technologies within the fields of energy industrial technology and biotechnology The efforts made shall benefit Danish society and lead to the development of new multi billion industries www risoe dk
27. by step description on how to execute them can be found in the appendixes 38 Report number ex Ris R 1234 EN 3 Definitions RGB value A scanned image is composed of pixels and each pixel has a value for the red green and blue content RGB value The RGB value is thus a triplet of values one for each colour each being an integer value between 0 and 255 Pixel signal To enhance the signal to noise ratio the RGB value is assigned a weighting factor for each of the three colours The weighting factors have a value between 0 and 1 The pixel signal S in the equation below is defined as the sum of the weighted RGB value normalised by the sum of the weighting factors The pixel signal has a value between 0 and 255 The pixel signal is rounded to nearest integer R w G w B WG Wg 1 Wg t Wat Wp where S pixel signal R value for red colour G value for green colour B value for blue colour wr weighting factor red colour wc weighting factor green colour wg weighting factor blue colour Definition of response value White has the RGB value 255 255 255 and black 0 0 0 This means that the pixel signal will be a high value for light colours and a low value for dark colours contrary to the dose that will be higher the darker the dosimeter is To comply with intuition and to make the calibration curves more similar to those obtained at spectrophotometers or densitometers the response value of a pixel is
28. calibration data together with the reference data click on the Save calibration reference button In the pop up window it 1s possible to check comments and scanner settings Figure 11 Click on Save Write a name for the calibration and Click on Save IMPORTANT If a calibration is changed at any time after its initial save then the changed calibration should be saved under a new name This avoids conflicts at later time of use 18 Report number ex Ris R 1234 EN foc NI a B3 4 calibdata a Testl calibdata a B35 Electrons calibdata a Test2 calibdata Jos Eqs go i3 835 Gamma caibdats ai test3 calibdata Calibration 125 125 a calibdata al Calibration 125 125 b calibdata Test med ref calibdata This calibration is only valid for the ScanJet 4P f at Risoe ZT pin File name C Calibration not to usel Custom Pattern calibdata 00 100 200 300 40 0 500 600 700 800 300980 Dose kGy Figure 11 Saving a calibration 7 5 Possible calibration function conflict When a calibration is assigned to an image the calibration function is saved in the info file belonging to the image Then the image is opened at a later point the program checks if the calibration function assigned to the image matches the one saved on the hard drive If the calibration function exists and is equal to the assigned calibration function nothing happens If the calibration function e
29. defined as wp 255 R wq 255 G w 255 B Wg t Wg t Wg I 255 5 Q where I response value S pixel signal www risoe dk 4 Workflow The data flow in RiseScan has a low degree of complexity as sketched in Figure 1 Tests are designed to verify each step in the data flow As input is used computer generated images with known content The image format is Windows Bitmap Output values from RiseScan are compared against calculations made with Windows Excel Dose profile Calculate dose Colour channel Open Image Depth dose profile Measure reference Figure 1 Data flow in Ris Scan 4 1 Open an image When a scanned image is opened a binary file will be assigned to the image The file has the same name as the image but with the extension info and contains information about the weighting factors for the RGB value the assigned calibration the reference values etc The info file also contains data about the size and creation date of the image file This means that if an image file is altered it will be detected when the image is re opened and a warning message will appear When an image is opened the user provides information about the weighting factors scanner settings and if the image is from an energy measurement with a wedge the angle of the wedge This information is stored in the info file It is verified that response values in an image are calcu
30. e shown on the graphs This is because the graph values are averaged over a one pixel wide line along the width of the cross hair while the centre average is obtained over the area of the cross hair centre It is possible to move the cross hair by dragging the vertical lines on the graphs 26 Report number ex Ris R 1234 EN Ris Scan Depth Profile Scan 125125150dpi sq2b b3 med baggrund ref File Display Help Figure 23 Widening the cross hair gives less noise but also less measurement resolution Compare the graphs with Figure 22 It is possible to change the scale on the graphs by disabling Autoscale X or Autoscale Y By doing so boxes for minimum and maximum values for the axes will appear Figure 24 The axes of the two graphs will have the same scale if Interlock X or Interlock Y are activated By default Interlock Y is enabled to make the dose values directly comparable between the two graphs The Interlock option may overwrite manual settings of the axes scale Ris Scan Depth Profile Scan 125125150dpi sq2b b3 med baggrund ref E Qt Figure 24 Disabling Autoscale makes it possible to select the minimum and maximum values on the axes Note that if Interlock is enabled it may overwrite manual settings www risoe dk 12 3D surface dose profile The 3D Surface profile task shows the selected part of the image as a tree dimensional plot with contour colours Figure 25 Left clicking on t
31. e table below the graph Note the values in Table E1 Make a screen dump if necessary Compare Table F1 with Table 19 Agreement in all values indicates that Ris Scan functions as intended Settings for slope algorithm Figure F1 Off set on the y axis should be set to one www risoe dk Ris Scan TES F RS 160 62 9 95 Average energy Figure F2 The test is initiated by pressing the Offset and Slope button Y axis offset and maximum response value is found in the upper right corner of the graph Vertical purple sliders on the graph mark the interval covered by the linear regression The rest of the data is displayed in the table Table F1 Depth dose analysis Ris Scan X axis shifted cm Data points for L R starts at cm depth Data points for L R ends at cm depth Ymax Yoffset R50 cm Rex cm Average energy E50 Most probable energy Eex L R linear regression 72 Report number ex Riso R 1234 EN Mission To promote an innovative and environmentally sustainable technological development within the areas of energy industrial technology and bioproduction through research innovation and advisory services Vision Rise s research shall extend the boundaries for the understanding of nature s processes and interactions right down to the molecular nanoscale The results obtained shall set new trends for the development of sustaina
32. e two reference values If the scanner settings are correct the settings used for the calibration the deviation should be less than one or two percent If there are small deviations a correction can be made to the calibration function To do this click on the Enable reference button this can also be done on the main menu IMPORTANT The reference ensures traceability to the calibration scan Use the reference in every scan used for dose measurements www risoe dk 10 Depth dose profile If a wedge angle has been selected during the open image task see Figure 2 the Depth Profile button can be selected The energy of a high energy electron beam can be measured using the depth dose profile in an aluminium wedge Using properties of the descending slope of the depth dose profile and equations from ICRU or ASTM ISO the average and most probable irradiation energy can be determined RiseScan can find the relevant properties of the descending slope in a manual or automated way 1 e the starting point of the dosimeter film in the wedge the maximum dose value the linear fit of the descending slope with the highest coefficient of determination R59 and Rex Select an area of the image covering as much of the depth dose profile as possible without going outside the edge of the dosimeter Make sure that Intensity displays Dose IMPORTANT Do not calculate energy if intensity is displayed as
33. ference correction Original Dose without Dose with response value correction correction 20 40 60 80 100 120 140 160 180 200 220 240 Out of range check 19 241 66 Report number ex Ris R 1234 EN APPENDIX D How to perform Test 4 1 Open the image Test4A bmp In the Image properties window set the DPI 100 Set the weighting factors for the RGB values to 0 1 0 Press Accept From the main window select the Surface profile button Press Ctrl A to select the whole image Press Ctrl Z or right click to go into acquisition mode Change the Width of selection use the slider so centre square of the cross hair fills out most of one of the sample squares set the width to e g 75 pixels Move the cross hair to the first sample square Read the response value and the standard deviation in the line below the image See Figure D1 Note the Mean of centre and Std var values in Table D1 Make a screen dump Ctrl S if necessary 9 Move the cross hair to the sample square below read values move to sample square upper middle read values and so forth Six reading all in all See Figure DI 10 In the File menu select Open image and open the image Test4B bmp 11 In the Image properties window set the DPI 100 Set the weighting factors for the RGB values to 0 1 0 Press Accept 12 Pres
34. he plot while moving the mouse will rotate the plot Below the plot is a legend showing the units used in the plot and several options for the layout The Maximum total number of pixels in x and y direction box shows the maximum number of data points on each axis in the plot The 3D plot takes some time to compute and making a plot using all pixels in the original image will not be feasible The number of data points is therefore reduced lowering the resolution but speeding up the time to plot the 3D graph Ris Scan Scan 125125150dpi sq2b b3 med baggrund iit on X and Y axis cm Unit on Z axis kGy Figure 25 3D surface dose profile 28 Report number ex Ris R 1234 EN 13 Creating documentation Ris Scan offers several ways to document measurements 13 1 Report generation print to default windows printer To create a report click Ctrl P or File gt Print To print the report to the default windows print set the switch to Print report to printer see Figure 26 The contents of the report is chosen by enabling one or more of the check boxes Print information about the Image Information about creation date of the image operator etc Print information about the Calibration The data and graphs used to establish the calibration function Print information about the Reference The data and graphs concerning the reference values Print the Original Image The scanned image as it is Print the Calibrated
35. idation In these cases one must rely on black box testing of the software together with documentation of correct operation of the program Black box testing is also called functional or specification based testing since only specific functionality of the program is tested Whether the level of functional testing is enough to ensure that software works as intended depends on a risk analysis made by the end user Example A researcher using RiseScan for measuring the homogeneity of a dose profile may need very little documentation while medical device manufactures looking for dose distributions in products might need substantial documentation The user of the program shall carry out this type of tests and this report describes a set of functional tests intended to be used as part of a retrospective validation of RiseScan 1 3 Functional tests The purpose of functional testing is to produce documentation that the software works as intended A typical functional test will use a defined input and compare output with an expected result obtained from an independent source Objective pass fail criteria must be defined General Principles of Software Validation Final Guidance for Industry and FDA Staff U S Department Of Health and Human Services Food and Drug Administration Center for Devices and Radiological Health 2002 www risoe dk 2 Introduction This document describes a series of tests that verify key functionality of
36. it 17 7 2 Residual plot 17 7 3 Include reference in calibration data 17 7 4 Save calibration 18 7 5 Possible calibration function conflict 19 8 Apply calibration to an image 19 8 1 Change image from response to dose 20 9 Apply reference to an image 21 10 Depth dose profile 22 10 1 Manual energy measurement 22 10 2 Automated energy measurement 24 10 3 Settings for energy measurement 25 11 Surface dose profile 26 12 3D surface dose profile 28 13 Creating documentation 29 13 1 Report generation print to default windows printer 29 13 2 Report generation HTML file 30 13 3 Graph values to ASCII file 30 13 4 Screen dump 31 14 Bibliography 32 www risoe dk Ris Scan 1 0 Toolset for Retrospective Validation 35 1 Background Relationship to FDA guidelines 37 1 1 General FDA guidelines 37 1 2 Retrospective validation 37 1 3 Functional tests 37 2 Introduction 38 3 Definitions 39 4 Workflow 40 4 Open an image 40 4 2 Calibration 40 4 3 Dose measurement 41 4 4 Relationship between response value and dose 41 4 5 Electron beam energy measurement 42 5 Tests used for retrospective validation of Ris Scan 1 0 43 5 1 TEST 1 Applying weighting factors to RGB value and conversion to Response value 43 5 2 TEST 2 Calibration procedure 44 5 3 TEST 3 Reference correction 49 5 4 TEST 4 Standard deviation 52 5 5 TEST 5 Minimum and Maximum values Surface profile 55 5 6 TEST 6 Depth dose profile 56
37. lated correctly from the pixel signal by using Test 1 4 2 Calibration An image of dosimeters irradiated to known doses 1s used to create a calibration function An area of each dosimeter is selected and the average response value is assigned to the known dose A calibration function 1 to 4 order polynomial is chosen by the user and the residual is calculated It is verified that the calibration function is calculated correctly by using Test 2 40 Report number ex Risg R 1234 EN The image of a reference colour chart is recorded by the scanner with every image The response values for the different colours of the reference chart are stored together with the calibration function when a calibration is saved 4 3 Dose measurement When an image with dosimeters irradiated to unknown doses is opened the user will again supply information about the scanner etc When a calibration is chosen these data will be compared to the information given at the time when the calibration was made If there are discrepancies a warning massage will be displayed The response values of the reference colour chart is measured and compared with those from the calibration image Any disagreement between the reference values for the calibration and the image of the unknown dosimeters is displayed in percent The difference between the measurements of the reference values can be used to correct the calibration function as described in the next paragraph to
38. m Tel 1 303 400 9640 Fax 1 303 400 9831 e mail sales gexcorp com Technical questions may also be directed to Rise National Laboratory Att Jakob Helt Hansen Postbox 49 DK 4000 Roskilde Denmark Website http www risoe dk nuk risoescan risoescan htm Tel 45 4677 4908 Fax 45 4677 4959 e mail jakob helt hansen risoe dk www risoe dk 1 1 Disclaimer and general precautions The Ris Scan software program is intended for use in accordance with this manual The software was not designed to provide database or information storage in any manner The user is responsible for verification of the accuracy of the data contained in the reports generated using the Ris Scan software Users are also responsible for determining the suitability and performance of the software for their specific applications Rise and GEX do not warrant the data input or output associated with this software All scan measurements and dose results and any subsequent usage of the data derived from usage of this software are the sole responsibility of the user Further by acceptance implementation and use the user assumes any and all risks associated with such usage of this user manual and the associated software product Ris Scan users assume complete responsibility for the application of the software product and agree to use the software program in accordance with the information and instructions contained in this user manual The examples given in
39. mber ex Risg R 1234 EN 5 Open an image for analysis When Ris Scan starts the main menu Figure 1 is displayed To the left is the RiseScan logo or the image that is analyzed To the right is a list of tasks To open an image click on Open Image and navigate to the image file RisoeScan 2 01 Beta Licence NN Figure 1 Open image In the Image properties window Figure 2 fill in the required information about Scanning device Write the name of the scanner used e g HP Scanjet 4P Exposure adjustment units Click the switch to select the scanner exposure units Brightness Contrast Gamma or Shadow Highlight and Midtone Write the values used during the scanning of the image Scan resolution DPI Write the resolution used e g 100 dpi This information is required to keep track of the length scale in the image Wedge angle deg If the image is a depth dose profile fill in the angle of the wedge used Dose unit The dose unit to be used for the image the unit used during calibration kGy or Gy Note The dose unit used has to be specified at this stage It is recommended to write comments and to identify the operator www risoe dk Image properties Beamspot B 125 125 100 dpi Shadow Highlight Midtone Jo ss geez Remember to specify angle if any Figure 2 Image properties 5 1 Colour channel RGB selection Each pixel in the scanned image has a RGB value that
40. nergy Rex 1 73 5 09 x 0 2 8 99 Most probable Figure 18 Defining the dose off set by moving the horizontal line 10 2 Automated energy measurement A Click on the Off set and slope button Figure 19 B Adjust for dose off set if any using the vertical green line C The average and most probable energy are displayed below the graph to the right If the identification of the wedge surface fails point it out manually and then click on Slope only to find the best fit to the slope RSO 1 34 6 2 x 8 31 Rex 1 72 5 09 x 0 2 8 94 Figure 19 The slope with the highest coefficient of determination is found using the Off set and Slope button 24 Report number ex Riso R 1234 EN 10 3 Settings for energy measurement If the program chooses a wrong direction for the depth dose display the direction can be changed using the Scan direction button The equations used to calculate the energy can be changed using the Edit Equations button Figure 20 Edit Equations 5 09 x 0 2 Figure 20 Changing the equations for the energy calculations The automated energy measurement uses an algorithm to find the segment of the slope to use The algorithm makes a linear fit to every possible segment of the descending curve and then selects the segment which gives the highest coefficient of determination There are some constrains in the algorithm that can be changed from the menu Display gt Settings for
41. ng different weighting factors for the RGB value and the response values of each of the squares are measured using the tool in the Surface profile section of the program www risoe dk Figure 2 Image for Test 1 Table 1 and 2 shows the expected and measured response values Table 1 Expected response values calculated using Equation 2 in Excel Weighting factors Expected response value Eq 2 Wp Wa Wp Square 1 Square 2 Square 3 0 0 1 150 100 50 0 1 0 100 50 150 0 1 1 125 75 100 1 0 0 50 150 100 1 0 1 100 125 75 1 1 0 75 100 125 1 1 1 100 100 100 Table 2 Measured response values using Ris Scan Weighting factors Response value RiseScan results Wa Wa We Square 1 Square 2 Square 3 0 0 1 150 100 50 0 1 0 100 50 150 0 1 1 125 75 100 1 0 0 50 150 100 1 0 1 100 125 75 1 1 0 75 100 125 1 1 1 100 100 100 The exact agreement between response values calculated in Excel and by RiseScan verifies that Ris Scan functions 5 2 TEST 2 Calibration procedure Refer to Appendix B on how to perform this test A key feature in the Ris Scan software is the calibration A polynomial fits the response values as a function of dose This test is used to verify that the polynomial fit and the residuals are calculated correctly and that dose measurements are carried out according to the polynomial fit 44 Report number ex Riso R 1
42. nnection with RiseScan is assured though measurement of the reference tablet supplied with RiseScan The reference tablet is measured at the time of the measurement of the dosimeters for calibration and at any subsequent measurement of other dosimeters Any differences between the two measurements are used for correction of the measured result The reference tablet supplied with Ris Scan is made from B3 dosimeter irradiated to 12 25 and 50 kGy and laminated for protection This reference can be used for www risoe dk measurements with B3 dosimeter films Far West and GAF dosimeters Other colour charts can be used as long as they are long time colour stable Tests at Rise have proven the B3 reference to be stable for more than one year and it is recommended to re measure or replace the reference once a year Protect the B3 reference from exposure to UV light 4 Installation To install RiseScan and the Runtime engine for LabView 6 1 from National Instruments run the Setup exe file in the directory corresponding to the Microsoft Office package installed on the computer If there is no Office package installed either Setup file can be used 4 1 Uninstall To uninstall RiseScan remove the program by the add remove programs in the Control Panel If RiseScan and Runtime engine has been installed separately remove RiseScan by deleting the program and the Runtime engine using add remove programs in the Control Panel 10 Report nu
43. ons 5 6 TEST 6 Depth dose profile Refer to Appendix F on how to perform this test The energy of a high energy electron beam can be measured using the depth dose profile in an aluminium wedge Using properties of the descending slope of the depth dose profile and equations from ICRU or ASTM ISO the average and most probable irradiation energy can be determined Ris Scan has the option of finding the relevant properties of the descending slope in a automated way and the purpose of the test is to verify that the starting point of the wedge is found that the maximum dose value is found that the linear fit with the lowest coefficient of determination is found that Rso and Rex are measured and that Eso and E are calculated correctly The test image is an artificial depth dose curve as depicted in the Figure 9 and 10 200 4 150 100 4 i Rresponse value 50 0 T T T T T 1 0 00 0 50 1 00 1 50 200 250 3 00 Depth cm Figure 9 Response values of the test dose profile Figure 10 Image for Test 6 The curve consists of 11 segments of different lengths and slopes The test image is designed so it scales with a scan resolution of 200 dpi and an aluminium wedge angle of 16 0 degrees 56 Report number ex Ris R 1234 EN Table 18 shows the composition of the test image Table 18 Composition of the depth dose profile image Length pi
44. or L R ends at cm depth 2 0 Ymax 153 Yoffset 11 R50 cm 1 60 Rex cm 2 05 Average energy E50 9 95 Most probable energy Eex 10 64 The agreement within 0 5 between values calculated by Excel and RiseScan verifies that RiseScan functions The small differences are due to round off errors 58 Report number ex Ris R 1234 EN APPENDIX A How to perform Test 1 l 2 Naw 10 11 12 Im Open the image Test1 bmp Notice that the image is shown in grey scale colour only In the Image properties window set the Scan Resolution DPI 100 Set the weighting factors for the RGB values upper right corner of the window see Figure A1 Start with the first value 0 0 1 shown in the first column of table Al Press Accept From the main window select the Surface profile button Press Ctrl A to select the whole image Press Ctrl Z or right click to go into acquisition mode Click on the first square and read the response value in the lower left corner See Figure A2 Note the response value in Table A1 If needed save a screen dump using Ctrl S Click on the second square and third square Note the response values in Table A and make a screen dump if necessary In the File menu select Open image Re open the image Test1 bmp In the Image properties window set the weighting factors for the RGB values to the next value in the first column of table Al Press Accept
45. port number ex Ris R 1234 EN lunit on X and Y axis cm Unit on Z axis kGy o g g reo Figure 26 Writing graph values to an ASCII file 13 4 Screen dump An effective way to document a measurement is to make a screen dump Clicking on Ctrl S or File gt Save will display a dialog box where it is possible to write a comment Figure 29 Select the type of image file to create and click on Save Figure 30 shows an example of a screen dump Calibration name b3 med baggrund Figure 29 A comment for the screen dump can be written in the dialog box www risoe dk on X and Y axis cm Unit on Z axis kGy RisoeScan 2 02c Beta 2003 1 2 Jakob Helt Hansen Image created 18 04 2002 14 11 1251251 Date 07 01 2004 15 48 Scanner HP Scanjet 4P Operator JAHH Resolution 150 dpi Calibration name b3 med baggrund Figure 30 Example of a screen dump 14 Bibliography Helt Hansen J and Miller A 2004 Ris Scan A new Dosimetry Software Rad Phys Chem 71 359 362 2004 Peter Sharpe and Arne Miller Guidelines for the calibration of dosimeters for use in radiation processing NPL Report CIRM 29 National Physical Laboratory UK 1999 32 Report number ex Riso R 1234 EN www risoe dk 34 Report number ex Ris R 1234 EN Ris Scan 1 0 Toolset for Retrospective Validation www risoe dk 36 Report number ex Ris R 1234 EN 1 Background
46. r every possible segment of the depth dose profile with the constraints that the segment is going through R5 and lies within the depth of maximum and minimum dose The segment with the highest coefficient of determination is chosen The method is not very sensitive to noise and does not depend on any mathematical fitting of the depth dose curve itself If the program is asked to find the offset on the dose axis it will find the minimum value of a segment of the later part of the depth dose curve where there 1s no abrupt changes To the minimum value is added a user defined constant times the square root of the standard deviation of the mean standard error on the linear fit of the slope The user will have to make sure that offsets in depth and dose are chosen correctly It is verified that the algorithms used to find the best slope of a depth dose profile work correctly and that Ea Ep Rso and Rex are calculated correctly by using Test 6 42 Report number ex Riso R 1234 EN 5 Tests used for retrospective validation of RisoScan 1 0 Each test is based on artificial data that is the images to be analysed are generated by software and are not images of real dosimeter films The RGB values for each pixel in each test image are known precisely and the analysis of a test image must therefore lead to the predicted values without any uncertainty Any deviations from the predicted values are interpreted as a malfunction of RiseScan Note To sa
47. response Since the calibration function is non linear the calculated energy will be misleading 10 1 Manual energy measurement 22 A Point out the location of the start of the dosimeter film at the wedge surface on the depth dose profile To do so left click on the vertical red line in the graph and move it If the line cannot be found click on the Guideline to Centre button Figure 15 To fine tune the position of the wedge surface click on the Zoom In button Figure 16 Click on the Adjust for Off set button Note that the x axis shifts so zero is at the wedge surface Use the two vertical purple lines to mark the beginning and end of the segment of the slope to be used in the calculation of the energy Figure 17 If the vertical lines are not visible click on the Guidelines to Centre button Use the horizontal line to mark any dose off set Figure 18 The average and most probable energy are displayed below the graph to the right Report number ex Ris R 1234 EN R50 0 00 6 2 Rex 0 00 5 09 x 0 2 Figure 15 Finding the surface of the wedge 03 4 125 125 200 dpi b3 4 ref R50 123 62 762 Average energy Rex 162 509402 8 45 Most probable Figure 17 Pointing out the segment of the depth dose curve to use for the energy measurement Moving the vertical lines changes the start and the end point www risoe dk R50 1 34 6 2 8 31 Average e
48. s Ctrl A to select the whole image 13 Press Ctrl Z or right click to go into acquisition mode 14 Repeat the instructions from line 7 to 9 except the measurements are now written in Table D2 15 Compare Table D1 with Table 13 and Table D2 with Table 15 Agreement in all values indicates that RiseScan functions as intended Small deviations between Table D2 and Table 15 are acceptable m BO uB m Ris Scan TEST4a File Display Help Uu i 1 i u LU 1 1 UE on dde 20 25 30 35 40 vertical pixels 1 24 cm sponse 45 at 0 81 0 81 m Mean of center 50 0 Std var 5 0 gt Figure D1 The squares are measured in this order Average value and standard deviation is displayed in the line in the lower left corner www risoe dk Table D1 Test 4A RiseScan Pixel response Expected std dev Average response Std dev response response 50 5 0 50 1 0 100 5 0 100 2 0 200 5 0 200 4 0 Table D2 Test 4B Ris Scan Pixel response Expected std dev Average response Std dev response response 50 5 0 50 1 0 100 5 0 100 2 0 200 5 0 200 4 0 68 Report number ex Ris R 1234 EN APPENDIX E How to perform Test 5 Ne 906 Qv tA d s 10 11 12 13 Open the image Test5 bmp In the Image properties window set the DPI 100 Set the weighting factors for the RGB values to 0 1 0 Press Accept From
49. specifies the amount of red green and blue The signal to nose ratio can be enhanced by selecting a ratio of red green and blue that gives the highest contrast For B3 and FWT dosimeters 100 green and 0 red and blue gives the best result Figure 3 The values written here will select the RGB value to be used for the analysis of the image The same values as used during calibration must be selected Image properties Beamspot B 125 125 100 dpi zs T f HP Scanjet 5400c aio Units Shadow Highlight Midtone ad 4 a n ERE F fr Remember to specify anale if any Figure 3 Weighting factors for red green and blue colour channel 12 Report number ex Riso R 1234 EN When all initial settings are selected click on Accept to proceed RiseScan returns to the main menu with the selected image displayed At this menu different tasks can be selected 6 Selecting area of interest To select a area of interest on the scanned image left click and draw a rectangle to select the area you want to analyse This can be done anytime e g on the front menu Figure 4 or after a task has been selected Beamspot B 125 125 100 dpi Figure 4 A region of interest is marked while holding the left mouse button down IMPORTANT Press CTRL Z or right click to activate the selection Acquisition mode You can de activate again Selection mode by pressing CTRL Z or right click That is Ctrl
50. st 3 A measurement in RiseScan will usually consist of a collection of pixel signals and the response or dose values are displayed as an average value with a standard deviation Itis verified that average and standard deviation are calculated correctly by using Test 4 Itis verified that minimum and maximum signal in an image can be found correctly by using Test 5 4 5 Electron beam energy measurement The energy of an electron beam can be measured using the depth dose profile through e g an aluminium wedge The average 4 and most probable Ep energy can be found based on a measurement of the depth where 50 of maximum dose is reached 859 and the extrapolated range Rex based on the descending slope of the depth dose profile Differentiating the first part of the curve and finding the first local minimum value locate the starting point on the dosimeter often marked by a line or scratch To this value is added a length of 1 pixel i e the offset in depth is located just after the offset mark To find the maximum dose value used for Rso each data point on the depth dose curve is averaged together with the two nearest data points on each side to minimize the influence of noise The maximum value is found as the first local maximum in the negative direction of the depth dose curve The linear equation best fitting the descending slope of the depth dose profile is found using a brute force method A linear fit is performed fo
51. th correction values now have 6 entries 4 reference values plus 0 and 255 Do not press the Enable reference button Press Done to return to the main window Press Surface profile Make sure the program is in acquisition mode and the whole image is selected if not press Ctrl A and Ctrl Z Show the image intensity as dose Press Ctrl M or File gt Map dose Click on each of the big squares and read the dose Note the dose in Table C1 and make a screen dump Ctrl S if necessary Click on the first and last of the rectangles Note the dose value in Table C1 and make a screen dump Ctrl S if necessary Press Done to return to the main window Press the Enable reference button Repeat the instructions from line 16 to 19 to read the doses with the reference correction enabled Press Done to return to the main window Compare Table C1 with Table 12 Agreement in all values indicates that RiseScan functions as intended www risoe dk E Ris Scan TEST3 Test2 KN EE a aoo B 9000 om 00 IS 190 00 0 00 0 00 240 00 0 00 0 0 a a SE None 21 00 0 00 0 0 95 00 0 00 0 0 181 00 0 00 0 0 o L loo joo joo 0 00 0 00 0 00 0 00 Figure C1 Each of the four reference values is measured The response value for each rectangle is assigned the letters A to D There are 6 entries in the lower table when all reference values are measured Table C1 Re
52. the main window select the Surface profile button Press Ctrl A to select the whole image Press Ctrl Z or right click to go into acquisition mode Make sure that the Width of selection is 1 pixel use the slider Press Find minimum button See Figure El Read the response value in the text below the image Note the response value in Table E1 make a screen dump if necessary Note in Table E1 if the cross hair has been placed over the correct part of the image Press Find maximum button Read the response value in the text below the image Note the response value and placement in Table E1 Repeat the instructions from line 6 changing the width of selection to 5 and later 51 pixels Compare Table E1 with Table 16 Agreement in all values indicates that Ris Scan functions as intended Figure El Minimum and maximum values are found using the buttons to the right The width of the cross hair is adjusted with the slider in the middle www risoe dk Table El Resolution pixels Value sought Expected value Measured value Placement OK 1x1 Minimum 0 1x1 Maximum 200 5x5 Minimum 10 5x5 Maximum 190 51x51 Minimum 20 51x51 Maximum 180 70 Report number ex Ris R 1234 EN APPENDIX F How to perform Test 6 arpes 11 12 13 14 15 Open the image Test6 bmp In the Image properties window set the DPI 200 Angle between surface and image
53. this manual are based on Rise recommended program uses other uses of the program are possible Contact Rise or GEX to discuss your specific application needs Users of the Ris Scan software program must provide and validate their own computer and scanner system and have full and complete responsibility for any and all verification and validation activities related to the use of the Ris Scan software A Toolkit for Retrospective Validation 1s included with the software product to facilitate user with Installation Qualification testing to demonstrate that the software is functioning properly Users must establish and perform their own validation of the dosimetry system itself which includes the PC scanner software and associated procedures to be developed by the user to control image quality and dose traceability 1 2 Retrospective Validation Ris Scan ver 1 0 has been black box validated through the use of a Toolkit for Retrospective Validation The validation has been done retrospectively and should be repeated for any new installation and at time intervals specified by the user Use of the Retrospective Validation Toolkit ensures that the software works as intended and meets user needs and intended uses according to General Principles of Software Validation Final Guidance for Industry and FDA Staff section 6 3 Validation of Off the Shelf software and automated equipment 8 Report number ex Ris R 1234 EN 2 Req
54. to different curve fitting algorithms 48 Report number ex Ris R 1234 EN 5 3 TEST 3 Reference correction Refer to Appendix C how to perform this test Different lamp temperature of the scanner and drift in electronics may lead to small variations in multiple recordings of the same dosimeter To correct for this a reference colour chart is scanned together with the dosimeters used for the calibration At later scans the same reference is measured and assuming that the reference colours are stable a correction for changes in response values can be made The purpose of this test is to evaluate if the correction is executed correctly The image Figure 5 used in Test 3 is similar to the one in Test 2 except that the pixel response values are changed according to Table 10 below In general response values between 20 and 90 are shifted 5 up between 190 and 240 are shifted 5 down between 91 and 189 a linear fit is used to calculate the shift Figure 5 Image for Test 3 www risoe dk Table 10 Change in response values Response Change in Changed response value response value value 19 4 75 20 20 5 21 40 5 42 60 5 63 80 5 84 100 4 104 120 2 122 140 0 140 160 2 157 180 4 173 200 5 190 220 5 209 240 5 228 241 4 77 230 Reference values and out of range check 19 4 75 20 20 5 21 90 5 95 190 5 181 240 5 228 241 4 77 230
55. uirements Windows 98x 2000 or XP 200 Mb disc space 256 Mb RAM 500 MHz processor The program is best viewed with a screen resolution of 600x800 pixels Other resolutions may cause graphics to be misaligned 3 Scanner Most scanners use an auto configuration of exposure parameters i e brightness contrast and gamma not all scanners use the same terms It is important to use a scanner that can switch this feature off Some scanner software allows the user to save the exposure parameters for later use Saving the exposure parameters minimizes the risk of scanning with different settings for calibration and dose measurements If the scanner has an option for pre heating the scanner lamp it should be used The surface on which the dosimeters are placed should be as homogeneous as possible High quality glossy paper used for ink jet printers is recommended IMPORTANT Use bmp or png as file format for the scanned images The image file size depends on the size of the scanned area and on the chosen resolution An image file size of 0 5 to 1 5 Mb is usually sufficient Large image files may cause Ris Scan to halt Not enough memory to complete this operation error massage or work extremely slow Image file size should not exceed 4 Mb 3 1 Reference Reproducibility of the instrument used to measure the dosimeters is essential in order to maintain the measurement traceability The stability of the scanner used in co
56. ve a measurement using RiseScan for documentation purposes one can make a screen dump by selecting Save in the File menu or pressing Ctrl S 5 1 TEST 1 Applying weighting factors to RGB value and conversion to Response value Refer to page 1 for definitions of RGB and response values Refer to Appendix A on how to perform this test The purpose of the test is to verify that the response value is calculated correctly for a given set of weighting factors of the RGB value Each pixel in a bitmap image consists of an unsigned long integer that is split up into four unsigned bytes The first byte is not used the last three bytes of value 0 255 describe the amount of red green and blue in that pixel these three bytes are called the RGB value To maximise the signal to noise ratio in a measurement the three bytes are reduced to one byte using a linear equation Eq 2 wp 255 R wq 255 G w 255 B Wg t Wg t Wg I 255 5 Q where I response value S pixel signal R value for red colour G value for green colour B value for blue colour wr weighting factor red colour wc weighting factor green colour wg weighting factor blue colour The test image Figure 2 consists three squares with following RGB values 205 155 105 105 205 155 and 155 105 205 The response value in the red green and blue colour is 50 100 150 150 50 100 and 100 150 50 The image is opened usi
57. xels Length cm Depth cm Response 0 0 00 0 00 99 54 0 69 0 19 99 56 0 71 0 20 119 57 0 72 0 20 99 171 2 17 0 60 99 229 2 91 0 80 139 286 3 63 1 00 153 345 4 38 1 21 149 769 9 77 2 69 10 857 10 88 3 00 10 The segment from pixel 486 to 628 marked by yellow in Table 18 is the longest segment in the depth dose profile one pixel longer than the segment from 345 to 486 It is therefore the segment from pixel 486 to 628 which should be covered by the linear regression found by the automated measurement performed by RiseScan The equation for average energy Rso Eav is set to be Eay 6 2 Rso The equation for most probable energy Rex Ep is set to be Ep 5 09 R 0 2 The Offset and Slope function is used to perform an automated measurement Results calculated by Excel and RiseScan is shown in Table 19 and Table 20 respectively Table 19 Results from depth dose analysis calculated using Excel Excel X axis shifted cm 0 2 Data points for L R starts at cm depth 1 5 Data points for L R ends at cm depth 2 0 Y max 153 Y offset 10 R50 cm 1 61 Rex cm 2 06 Average energy E50 9 99 Most probable energy Eex 10 69 L R linear regression www risoe dk Table 20 Results from depth dose analysis measured using Ris Scan Ris Scan X axis shifted cm 0 2 Data points for L R starts at cm depth 1 5 Data points f
58. xists but 1s different to the assigned function a warning will appear In this case the user can either save the calibration in the info file to the hard drive use the calibration on the hard drive i e copy it to the info file or use the calibration in the info file without saving it to the hard drive It is strongly recommended not to use the later option since there can be confusion about which calibration function is in use If the calibration function in the info file does not exist on the hard drive the user is given the option to save it An easy way to transfer a calibration between different computers is therefore to copy the info file together with the image 8 Apply calibration to an image After an image has been loaded a calibration can be selected by clicking on the Select calibration button on the main menu or File gt Load Calibration Select the calibration to be assigned to the image and click on Select Figure 12 www risoe dk Select Calibration File SS B 125125100 d r NE AAAA SSJAILOLI SS GE a N 5502 am ene Aena ter 0597 zenean We SHM 0 255 1 00 MP Scanjet 5400 6 00 1 00 Tm 5 125 1 00 HP Scanjet 4P 0 00 1 00 0 00 Yes 125 125 1 00 HP Scanjet 4P gera 1 e 00 Figure 12 Assigning a calibration to an image 8 1 Change image from response to dose To display dose instead of pixel response or vice versa click on Ctrl M or Display gt M