MPT-250B SPECIFICATIONS AND OPERATING INSTRUCTIONS

image

Contents

1. 4 1 1 Check the operating voltage of the unit The correct operating voltage is on the identification label on the back of the unit Be sure that this voltage is the same as the voltage available at your location before you plug in the MPT 250B 4 1 2 Turn the power switch in the center of the back plate OFF 4 1 3 Plug the power cord into its receptacle on the back of the MPT 250B and into the incoming power receptacle 4 1 4 Startup Sequence NOTE Please read this section in its entirety prior to performing the startup sequence ANIELS NUFACTURING MA DA FE CORPORATION DATASHEET A FIRST TIME STARTUP ONLY Upon initial receipt from the factory a Prior to turning on the unit turn the zero adjustment knob to the maximum clockwise position b Turn the power switch ON Rear Panel The display will turn on and the motor control stop switch will illuminate c The unit will warm up for approximately 60 seconds d The unit will then display the serial number e The unit will need to be adjusted to zero NOTE It is imperative to move the zero adjustment knob accurately and slowly for proper operation of the Alphatron unit If the adjustment knob is turned too quickly the value may extend beyond the zero mark into the negative value range and cause an error message to be displayed As a general rule it is suggested to begin turning the adjustment knob slowly once the 002 0 display value is r
2. 855 6884 e www dmctools com Email dmc dmctools com Copyright 2007 All Rights Reserved Page 3 of 6 Rev J 05 10 MPT 250B DS MPT 250B SPECIFICATIONS AND OPERATING INSTRUCTIONS ON AUTO STOP enabled OFF AUTO STOP disabled the STOP switch must be used at the completion of a test NOTE The AUTOSTOP function will not operate if the test range is at or below approximately 1 51 b This is necessary to allow for the preload that an operator will normally apply when inserting a connector for the test If the preload is greater than the lower limit the unit will not start In this case depress the reset toggle switch next to the display to be able to begin test 3 5 2 POUND KILOGRAM NEWTON Dip switch 2 amp 3 of dip switch bank SW2 control the type of units that are measured by the tester If pounds are desired switch 2 is down and switch 3 is up If kilograms are desired switch 2 is up and switch 3 is down If Newton units are desired both switch 2 and 3 are up 3 6 REAR PANEL FUSE If the main fuse requires replacement install a 3 Amp 125 250V slow blow fuse The fuse compartment is located on the back of the unit and labeled as FUSE ON OFF Depress the rocker switch in order to turn the unit ON or OFF AC POWER Insert the proper end of the supplied power cord into the male receptacle 4 OPERATION In the instructions that follow operation of both standard and optional features will be covered 4 1 POWER
3. FRONT PANEL 3 4 1 SETPOINT The SETPOINT option enables an operator to set the force at which the MPT 250B is to stop automatically if the test proceeds to that force and the connection under test does not fail When the SETPOINT MEASURE switch is held in the MEASURE position it enables the operator to select the set point by turning the SET adjusting knob next to the switch As the knob is turned the set point force will be indicated on the display meter Clockwise adjustment increases the set point value and turning the knob counterclockwise decreases the value The force reading from the load cell is disabled and will not be enabled until the switch is in the SETPOINT raised position Tests are conducted with the switch in the raised position When the load reaches the set point level the motor will stop and the unit can be reset and reloaded for the next test NOTE The unit must be turned off and then back on for the set point measure changes to take effect This option can be used in conjunction with the AUTO STOP section 3 5 1 feature When both features are turned on the test will stop automatically if either condition is met AUTO STOP or SETPOINT If the peak reading of the force necessary to part a connection is desired the set point should be at 249 9lb before testing 3 5 SIDE PANEL DIP SWITCH SETTINGS The dip switches are accessed through the opening on the left side of the MPT 250B The access opening
4. is covered with a plate that can be swung out of the way Turn the large captivated screw counter clockwise until it comes loose from the side cover The label below the cover depicts the switch banks and lists the switch functions The detailed description of the switch settings is as follows am A OH ANUFACTURING DFE CORPORATION NOTE The unit must be turned off and then back on for dip switch changes to take effect Bank SW1 Toward rear of tester Switch 1 Not Used Switch 2 Not Used Switch 3 Not Used Switch 4 Not Used Switch 5 Factory Setting DO NOT adjust Switch 6 Down Normal Up Set point measure enabled Switch 7 Auto Stop Down Off Up On Bank SW2 Toward front of tester Switch 1 Not Used Switch 2 Ib units Down Section 3 5 2 Kg units Up N units Up Switch 3 Ib units Up Section 3 5 2 Kg units Down N units Up Switch 4 Factory Setting DO NOT adjust Switch 5 Factory Setting DO NOT adjust Switch 6 Factory Setting DO NOT adjust Switch 7 Factory Setting DO NOT adjust 3 5 1 AUTO STOP This feature stops the pull test when the piece under test fails When the force applied to the piece drops below 70 of the displayed reading the test is completed and the unit stops This feature is turned on or off using switch 7 of dip switch bank SW1 Daniels Manufacturing Corporation 526 Thorpe Road Orlando FL 32824 U S A Phone 407 855 6161 e Fax 407
5. 0B SPECIFICATIONS AND OR ea OPERATING INSTRUCTIONS __ DATASHEET Daniels Manufacturing Corporation warrants each new unit sold by it to be free from defects in material and workmanship under normal use and service Its obligation under this warranty is limited to the free correction or at its option the refund of the purchase price of any such unit which proves defective within ninety 90 days after delivery to the first user provided that the unit is returned with all transportation charges prepaid and which shall appear to its satisfaction upon inspection by it to have been defective in material or workmanship This warranty shall not cover any damage to such products which in the opinion of Daniels Manufacturing Corporation was caused by normal wear misuse improper operation or accident This warranty is in lieu of all other warranties express or implied No warranty express or implied is made or authorized to be made or assumed with respect to products of Daniels Manufacturing Corporation other than that herein set forth as defined by PL93 637 Daniels Manufacturing Corporation 526 Thorpe Road Orlando FL 32824 U S A Phone 407 855 6161 e Fax 407 855 6884 e www dmctools com Email dmc dmctools com Copyright 2007 All Rights Reserved Page 6 of 6 Rev J 05 10 MPT 250B DS Glenair UK Limited Connector Tooling Division Alphatron by DMC Check List CUSTOMER O NO ENQ NO DMC O NO ze ea See ee Option Tick MPT 250B F
6. 3 UPPER GRIP do not exceed 12 months between factory calibration 4 3 1 If the upper grip is not positioned in the home intervals position as shown in the diagram on page 2 press the RESET button 4 3 2 Depress the operating lever of the self tightening cam type upper grip to open far enough to accept the test sample wire When you release the lever the cam wheel will grip the wire 4 2 3 Ring Terminal Lower Grip Optional Rotate the grip to place the correct stud in the uppermost The internal calibration test sequence is a reliable verification that the tester is accurate but it does not assure traceability to NIST 4 4 SAMPLE OPERATION The following is an example procedure for testing a wire specimen after the unit has successfully completed the STARTUP SEQUENCE section 4 1 4 4 4 1 Place the terminal end of the wire sample in the Lower Grip section 4 2 LIMITATION OF LIABILITY LIMITED WARRANTY DANIELS MANUFACTURING CORPORATION IS NOT LIABLE FOR CONSEQUENTIAL OR SPECIAL DAMAGES OF ANY NATURE OR KIND RESULTING FROM THE USE OF ANY OF ITS PRODUCTS OWNERS AND USERS OF DMC PRODUCTS ASSUME FULL RESPONSIBILTY FOR INSTRUCTING THEIR EMPLOYEES IN THE PROPER AND SAFE USE OF SUCH PRODUCTS Daniels Manufacturing Corporation 526 Thorpe Road Orlando FL 32824 U S A Phone 407 855 6161 e Fax 407 855 6884 e www dmctools com e Email dmc dmctools com Copyright 2007 All Rights Reserved Page 5 of 6 Rev J 05 10 MPT 250B DS MPT 25
7. MPT 250B SPECIFICATIONS AND OPERATING INSTRUCTIONS a SPECIFICATIONS PERFORMANCE Capacity 250lb 113kg 1112N Safe Overload 150 of capacity Readout Accuracy 0 5 of full scale Resolution 0 1lb Speed Control Accuracy 1 8 min 1 5 min rate of pull 1 4 min 6 10 min rate of pull Standard Display Pound Kilogram or Newton Units XXX X Operating Temperature 50 F to 100 F 10 C to 38 C Auto Stop Selectable to be enabled or disabled PHYSICAL Weight 26lb less options Height 8in Width 10in Depth 15in Cabinet Painted Carbon Steel ELECTRICAL Standard Power 115VAC 50 60Hz OPTIONAL FEATURES AND ACCESSORIES Universal self tightening cam type lower grip P N 15 3202 Ring terminal lower grip P N 15 0066 Adjustable set point for motor stop at preset force Optional Power 230VAC 50 60Hz ANIELS NUFACTURING MA Du FE CORPORATION DATASHEET 1 SAFETY The MPT 250B Wire Crimp Pull Tester is a force measurement device and as such should be operated with due caution Operators should wear safety glasses for eye protection because the crimp under test may break suddenly and fragments may fly off Do not overload The gear motor in the pull tester can exceed the load limits of the load cell component in this device For your protection the MPT 250B has been factory set to stop if the peak force exceeds 250lb The unit is set to automatically stop at read
8. Self Tightening Cam Type Lower Grip optional Depress the operating lever of the self tightening cam type lower grip to open it far enough to accept the sample under test When you release the lever the cam wheels will grip the sample NOTE DO NOT POSITION THE GRIP PRESSURE ONTO ANIELS NUFACTURING MA Du FEA CORPORATION DATASHEET 4 4 2 Pull the sample to remove any slack in the wire and place the opposite end of the wire sample in the Upper Grip section 4 3 4 4 3 Once the sample is taut between the grips press the motor control START switch section 3 3 1 The pull wheel will turn in a clockwise motion thereby tightening the grip on the sample and applying tension to the wire 4 4 4 The unit will cease operation once the sample has failed section 3 5 1 or the force limit has been reached Record the value indicated on the display panel 4 4 5 Press the motor control RESET switch section 3 3 1 to return the pull wheel to the start position to allow for the next test 4 4 6 Press the display control RESET switch section 3 3 3 to return the display to zero 4 4 7 Repeat steps 4 4 1 to 4 4 6 THE CRIMP BARREL OF THE TEST SAMPLE 5 CALIBRATION The MPT 250B wire crimp pull tester is factory calibrated with equipment traceable to the NIST Itis recommended practice to return the unit to DMC for position and place the terminal over the stud calibration In order to maintain traceability to NIST 4
9. eached so as to not pass zero i Turn the zero adjustment knob counterclockwise slowly until zero is displayed ii Once zero is reached the unit will automatically perform an internal calibration sequence iii If the calibration is OK the unit will be ready for operation iv If the Alphatron does not zero calibrate properly an error message will display prompting the end user to contact DMC Prior to contacting DMC the following steps should be taken 1 Turn the power to the unit OFF Rear Panel 2 Turn the zero adjustment knob clockwise one complete turn 3 Follow steps 4 1 4 A b through 4 1 4 A e 4 If the calibration is OK the unit will be ready for use 5 If the Alphatron does not zero calibrate properly an error message will display Daniels Manufacturing Corporation 526 Thorpe Road Orlando FL 32824 U S A Phone 407 855 6161 e Fax 407 855 6884 e www dmctools com Email dmc dmctools com Copyright 2007 All Rights Reserved Page 4 of 6 Rev J 05 10 MPT 250B DS MPT 250B SPECIFICATIONS AND OPERATING INSTRUCTIONS prompting the end user to contact DMC B FOR DAILY USE After the Alphatron has been in operation a Follow steps 4 1 4 A b through 4 1 4 A e 4 2 LOWER GRIP 4 2 1 Standard Grip Rotate the lower grip to place the correct slot for the wire terminal under test in the uppermost position Select a slot that is the same width as the wire diameter or one size larger 4 2 2
10. ings above 249 9lb 2 SETUP The Alphatron MPT 250B is shipped from the factory pre calibrated and tested To assure consistent correct results users should familiarize themselves with the setup and operation of the unit before placing it in service To operate set the MPT 250B on a flat level surface in an upright position The unit can be tilted with the adjustable front feet to ease operator viewing of the display Do not handle pick up or move the unit by exerting leverage against any front or rear panel controls fixtures or connections Always lift by the base plate preferably at the mid point along either side DO NOT use the lower grip as a handle for lifting or moving the MPT 250B This can result in permanent damage to the load cell sensing unit 3 ASSEMBLY 3 1 UPPER GRIP The MPT 250B is shipped with the upper grip installed 3 2 LOWER GRIP The MPT 250B is shipped with the standard lower grip installed If an optional lower grip was ordered it will be necessary for the purchaser to install it 3 2 1 Self tightening cam type lower grip optional To install the self tightening cam type lower grip proceed as follows A Remove the screw holding the standard lower grip in place B Place the screw in the mounting hole of the self tightening cam type lower grip To get the screw head past the operating gears into its recess use the operating lever to rotate the cams into the fully open position C Screw the assembly fir
11. mly into place in the load cell sensing unit Daniels Manufacturing Corporation 526 Thorpe Road Orlando FL 32824 U S A Phone 407 855 6161 Fax 407 855 6884 e www dmctools com Email dmc dmctools com Copyright 2007 All Rights Reserved Page 1 of 6 Rev J 05 10 MPT 250B DS MPT 250B SPECIFICATIONS AND ANIELS NUFACTURING MA Du FEA CORPORATION OPERATING INSTRUCTIONS DATASHEET SHOWN IN THE HOME POSITION FOR TEST ALPHATRON lt AS i MPT 250B fi WIRE CRIMP PULL TESTER M 250LB MAX CAPACITY Hi J ZERO RESET IX ae J f N JEN ee Ne A J Yv MOTOR CONTROL 4 i ra CRIMP GRIP A SL START RESET STOP 3 s SA ate oF PULL o 1 NCH MINUTE 7 i 1 4 J orao oo 2 5 1a Y A J SET POINT l i Le eo MEASURE NN D 5 5 SS Ye S7 lt A YW Pa S s a OPTIONAL 3 2 2 Ring Terminal Lower Grip Optional If the ring terminal lower grip has been ordered installation is as follows A Remove the screw holding the standard lower grip in place B Substitute the optional grip for the standard grip and tighten the screw until the lower grip ro
12. tates with resistance but is not loose or does not need to be forced to turn 3 3 FRONT PANEL CONTROLS 3 3 1 MOTOR CONTROLS NOTE PRESS AND HOLD SWITCHES FOR APPROXIMATELY 1 SECOND TO ALLOW FOR PROPER ACTIVATION START Motor start switch to begin test RESET Resets display meter to zero and resets pull wheel to START position for the next test RESET must be performed after each test STOP Stops the motor and the meter retains the current peak value 3 3 2 RATE OF PULL CONTROLS LO HI SWITCH LO Green selects rate of pull of 1 5 in min HI Yellow selects rate of pull of 6 10 in min NUMBERED RATE SWITCHES Select the rate of pull 1 5 in min LO Green or 6 10 in min HI Yellow as chosen by setting of LO HI switch 3 3 3 DISPLAY CONTROLS ZERO Adjusts the display to true zero When turned counterclockwise the meter reading will decrease When turned clockwise the meter reading will increase The display will be checked for zero Daniels Manufacturing Corporation 526 Thorpe Road Orlando FL 32824 U S A Phone 407 855 6161 e Fax 407 855 6884 e www dmctools com Email dmc dmctools com Copyright 2007 All Rights Reserved Page 2 of 6 Rev J 05 10 MPT 250B DS MPT 250B SPECIFICATIONS AND OPERATING INSTRUCTIONS adjustments at startup and after each test when the RESET switch is pressed RESET Resets the display clears peak values to zero 3 4 OPTIONAL CONTROLS ON THE
13. ully CE Approved Adj Load Set Point for Motor Stop 30 VAC 50 60Hz Power elf Tightening Cam Type Lower Grip Heavy Duty Swivel Lockdown Upper Grip Both Parts Needed hort Wire Grip Miniature Terminal Grip Ring Terminal Lower Grip N aS k d d Heavy Duty Swivel Lockdown Upper Grip Both Parts Needed i y Short Wire Grip y Miniature Terminal Grip Ring Terminal Lower Grip Country of Use Please Tick and Return with any Quotation or Order you request or place for any Alphatron Machine Please Return to Glenair UK Limited 40 Lower Oakham Way Oakham Business Park Mansfield Nottinghamshire NG18 5BY Tel 01623 638100 Fax 01623 661044

Download Pdf Manuals

image

Related Search

Related Contents

Debit Card Interface User Manual  JELD-WEN H31302 Use and Care Manual  GARANTIE 8 ANS DEFIBTECH Avec échange standard  Hewlett Packard Enterprise AF447A uninterruptible power supply (UPS)  Weather Information System Version 2.0 User Manual    Portable LED Light System User Manual  TOASTER  AC-8000 - キューマット株式会社  STK / SEET R  