OPERATING INSTRUCTIONS FOR
Contents
1. button to decrease the over temperature cut out value 3 Once the value of your choice is displayed release all buttons and after a short delay the new value will be stored and retained in permanent memory SETTING AN UNDER TEMPERATURE CUT OUT VALUE If for example you would like to operate your incubator at 20 0 C an under temperature cut out value needs to be programmed into the cut out We recommend that the cut out value is set to 2 degrees lower than your chamber temperature In the example above the under temperature cut out set point needs to be set 18 C To do this follow steps 1 3 below 1 Press the C button twice Flashing AL2 will be displayed plus a value the under temperature cut out value 2 Press the 4 button to increase the under temperature cut out value or the Ly button to decrease the under temperature cut out value 3 Once the value of your choice is displayed release all buttons and after a short delay the new value will be stored and retained in permanent memory OVER amp UNDER TEMP NEON S BUZZER amp RESET BUTTONS If the temperature in the incubators chamber goes above it s upper set point the buzzer will sound and the over temperature neon will illuminate The heaters will automatically disconnect to prevent further over heating The buzzer 8 neon will remain on until a member of staff has acknowledged the alarm To reset the buzzer amp alarm press the reset button along side the il
2. 1 0 2 0 3 0 4 0 5 0 OPERATING INSTRUCTIONS FOR LEEC HUMIDITY CABINETS MODELS SFC2C AND SFC3C WITH JUMO dTRON 316 CONTROLLER 8 TLK38 OVER UNDER TEMP CUT OUT CONTENTS GENERAL DESCRIPTION INSTALLATION TEMPERATURE SETTING PROCEDURE HUMIDITY CONTROL MAINTENANCE Revised December 2007 LEEC Limited Private Road No 7 Colwick Industrial Estate Nottingham NG4 2AJ United Kingdom Tel 44 0 115 961 6222 Fax 44 0 115 961 6680 E mail sales leec co uk Web www leec co uk 1 0 1 1 1 2 1 3 1 4 1 5 1 6 GENERAL DESCRIPTION These cooled laboratory humidity cabinets provide the user with extremely accurate and safe temperature control over a wide temperature range HEATING Low wattage heaters are attached to the outer surface of the inner chamber in such a manner as to ensure even heating throughout the chamber COOLING Direct expansion coils are coupled to a hermetic condensing unit via a solenoid valve When the refrigeration switch located on the left of the control panel is switched on the compressor runs continuously TEMPERATURE CONTROL e000 The temperature is accurately controlled by a JUMO dTRON 316 microprocessor temperature controller The controller uses the signal from a PT100 sensor located in the chamber to control the current supplied to the heaters The temperature can be easily adjusted see section 3 for full details TLK38 OVER amp UNDER TEMPERATURE SAFETY C
3. UT OUT A TLK88 digital cut out protects the contents of the chamber When the chamber temperature goes out of limits the appropriate cut out will operate an audible alarm and visible red neon Two reset buttons are provided below the cut out for you to manually reset the out of limit system CIRCULATING FAN Air is drawn through the chamber by a circulating fan located in the top and returned to the base via a full width duct at the rear of the chamber A door micro switch stops the fan when the door is opened HUMIDITY A wide range of fixed relative humidities can be achieved by adding a suitable saturated salt solution directly into the base of the chamber See section 4 for more details 2 0 2 1 2 2 2 3 INSTALLATION The incubator has a factory fitted UK style 13 Amp fused moulded plug and requires a 240V 50Hz supply The wiring inside the plug and mains flex is colour coded as follows e BROWN LIVE BLUE NEUTRAL e GREEN YELLOW EARTH The green POWER light will illuminate when the supply is connected and the POWER switch is in the ON position The incubator is internally protected by a 10 Amp fuse The fuse holder is located on the rear of the cabinet and is clearly labelled as FUSE REFRIGERATION PRECAUTIONS e The incubator should be left to settle for 2 hours after final positioning before it is switched for the first time to allow the refrigerant to settle e Do not place the incubator in front of a wi
4. e saturated 2 D A solution A A i i 3 10 10 Lithium chloride saturated solution oA 11 11 Silica gel dried 20 30 40 50 60 Temperature C Salts Graph The graph above shows the temperatures and relative humidities for several of the more common saturated salts The distilled water saturated salt solution should be topped up regularly and should fully cover the floor at all times Do not allow the water saturated salt solution to evaporate or the humidity control will be affected Allow the chamber temperature and humidity to stabilise before placing products inside the chamber MAINTENANCE No routine maintenance is necessary but regular cleaning of the inner chamber is essential Keep ventilation grilles clear and unobstructed The chamber should be regularly de frosted as a substantial build up of ice in the air circulation duct will restrict the airflow and affect the incubators performance It is recommended that the service ports on the right of the incubator remain sealed with the threaded red bungs supplied to reduce the tendency for humidity in the air to condense out in the chamber
5. ller shows when heat is being supplied The yellow 5 LED on the temperature controller shows when cooling is being supplied The incubator operating temperature or set point SP can be altered by simply pressing either the A button to increase the temperature or the W button to decrease the temperature When the required temperature has been entered release all buttons and the green display will blink once to confirm that a new target temperature has been stored TLK38 OVER amp UNDER TEMPERATURE SAFETY CUT OUT A TLK38 digital safety cut out unit provides essential over amp under temperature protection for the products inside the chamber The cut out has buttons labelled as follows gt Enters the programming mode No function INCREASES the displayed value DECREASES the displayed value Ld e E El 3 4 3 5 3 6 3 7 SETTING AN OVER TEMPERATURE CUT OUT VALUE If for example you would like to operate your incubator at 20 0 C an over temperature cut out value needs to be programmed into the cut out We recommend that the cut out value is set to 2 degrees higher than your chamber temperature In the example above the over temperature cut out set point needs to be set 22 C To do this follow steps 1 3 below 1 Press the P button once only Flashing AL1 will be displayed plus a value the over temperature cut out value 2 Press the a button to increase the over temperature cut out value or the Ly
6. luminated neon If the temperature in the incubators chamber drops below it s lower set point the buzzer will sound and the under temperature neon will illuminate The refrigeration system will automatically disconnect to prevent further over cooling The buzzer amp neon will remain on until a member of staff has acknowledged the alarm To reset the buzzer amp alarm press the reset button along side the illuminated neon Allow the chamber temperature to stabilise before placing products inside the chamber 4 0 5 0 5 1 5 2 HUMIDITY CONTROL A wide range of fixed and relative humidities can be achieved by adding suitable saturated salt solutions to the base reservoir The relative humidity of air in contact with saturated salt solutions reaches known equilibrium values at certain temperatures Temperature and relative humidities normally attainable in 20 C ambient a temperature Material to be placed in tray to 1 Distilled water 2 obtain indicated values of relative humidity 2 Potassium nitrate saturated shown in graph Curve Material solution e 3 Potassium chloride saturated ET 4 1 solution CGT Cis O Sodium chloride saturated solution z 53 os E 3 5 Sodium nitrite saturated solution 4 6 Sodium dichromate saturated or TES solution 7 Potassium carbonate saturated 3 Se et solution i aaa 8 Magnesium chloride saturated E 8 solution 3 e ee pvp E 9 Potassium acetat
7. ndow in direct sunlight e A qualified electrician or other competent person must carry out any electrical work required to install your incubator e Make sure that your incubator is not standing on its electrical cable e There are parts in your incubator that heat up Always ensure there is adequate ventilation as failure to do so can result in component failure Keep all vent grilles clear e Before any cleaning or maintenance work is carried out the mains supply must be switched off and the plug removed from the socket 3 0 3 1 3 2 3 3 TEMPERATURE SETTING PROCEDURE Switch the incubator on using the POWER switch located at the left hand side of the control panel press the RESET button to cancel the buzzer if it sounds after initial switch on The COOLING system will need to be ON unless you wish to use your incubator as a HEAT ONLY humidity cabinet This is important for the correct operation of the humidity cabinet The JUMO dTRON 316 microprocessor temperature controller on your incubator has two LED displays The upper display which is red represents the actual chamber temperature The lower display which is green represents the target temperature The controller has 4 buttons marked as follows 0000 PGM Program button enters the Programming Mode A Increases the chamber temperature v Decreases the chamber temperature EXIT Exits the Programming Mode The yellow 3 LED on the temperature contro