Anton Telegan Sprint V1 combustion flue gas analyser user manual

image

Contents

1. ZZ V1 Multi function flue gas analyser Sprint V1 User Manual M09990 57 August 2006 TELEGAN gas monitoring Issue 1 Safety information e Read and understand all instructions in the operation section of this manual before use e Do not substitute components as this may impair safety and invalidate warranty e Observe all warnings and instructions marked on the unit and within this manual e If this product is not working properly read the troubleshooting guide or call Telegan e Ensure qualified service personnel change sensors and provide maintenance and calibration Additional information Sprint is designed to support the working practices defined in British Standard BS7967 and the Design Standards EN50379 BS7927 for flue gas analysers It is highly recommended that users are fully conversant with BS7967 when using a flue gas analyser for servicing or installing a boiler system Performing CO measurements When performing any CO measurements ensure the unit is zeroed in clean air in accordance with British Standard BS7967 A suitable location for sampling clean air will be outside of the building where the boiler system is installed 1 Repair of this equipment and gas sensor replacement shall be carried out by the manufacturer or certified service centre in accordance with the applicable code of practice 2 If the equipment is likely to come into contact with aggressive substances then it is
2. The result code is FAILED 2 CO Level Does Not Start Close to Zero It is important that the instrument is zeroed in clean air at switch on Failure to do so will invalidate the test result BS 7967 2 2005 also requires that the room is well ventilated prior to starting the test However it is possible that there may be a residual background CO level eg due to traffic fumes The instrument will display a pop up box prior to starting the test if the CO level is more than 2 5 ppm TELEGAN 23 gas monitoring Technical helpline Tel 44 0 870 428 0075 Manufactured by Telegan Gas Monitoring A division of Crowcon Detection Instruments Ltd 2 Blacklands Way Abingdon Business Park Abingdon Oxfordshire OX14 1DY United Kingdom Tel 44 0 1235 557700 Fax 44 0 1235 557749 Email sales telegangas co uk Web site www telegangas co uk Exclusive UK Distributor Anton Industrial Services Ltd Unit 6 Park House 15 23 Greenhill Crescent Watford Business Park Watford Hertfordshire WD18 8PH United Kingdom Tel 44 0 870 428 0075 Fax 44 0 870 428 0076 E mail sales anton group com web www anton group com ee HALMA ae el G RO U P Registration COM PA NY No 195583
3. date and time is correct If the cal due date is set to more than a year s time then return for service ASAP TELEGAN gas monitoring Appendix Carbon Monoxide Room Safety Tests Sprint V1 Appendix I Carbon Monoxide Room Safety Tests Sprint V instruments are designed to assist heating engineers to work to the BS 7967 2005 specification for carbon monoxide room safety testing You should refer to BS7967 which defines the requirements details the methods as well as the pass and fail criteria for various types of appliances CO Room Safety Test in the Sprint V instruments is designed to measure the build up of carbon monoxide levels in a room where a gas appliance is in use and record those values each minute for the duration of the test In addition at the end of the test it assists the engineer in an advisory capacity only to deter mine whether the test has passed or failed or if the results are invalid In certain circumstances where the results are borderline or open to interpretation the instrument will ask the operator to decide if the test has passed or failed and will record the operators decision Please note ultimately it is the responsibility of the operator to ensure that the test is correctly performed to the BS 7967 2005 specification If the data does not support the result or the operator suspects it is not reliable due to local conditions such as carbon monoxide level changes due to cigarette smoke
4. test Sprint will display T1 or T2 on the screen to represent the first and second reading 1 Place probe in position to make measurement T1 2 Press the soft key icon T1 to take a snapshot reading of T1 Do not remove probe until this snapshot is taken TELEGAN r gas monitoring Performing tests Sprint V1 3 Move the probe into position to take second reading T2 The screen will display the icon T2 to show temperature reading T2 is being taken The screen will display the snapshot temperature the live probe temperature and the differential temperature Use the soft keys to print or log values To end test press the ESC key 3 Room CO safety test Ensure the unit has been zeroed in clean air during switch on before performing this test Note For further details on performing Room CO Safety Tests see Appendix I 1 Select CO Room safety from the Test menu Sprint will display the Appliance menu Select the appliance from the list by using the up and down arrow Press the centre button to confirm appliance selection 2 Connect probe if required to the Sprint unit and place at the recommended height Refer to British Standard BS7967 if necessary 3 The pump will switch on in readiness for test NB The sound of the pump operating does not indicate the test has begun Press the Start key to start test During the test the screen will display the CO reading peak CO reading and duration of test The test will run for
5. the required duration by appliance according to BS7967 Sprint will emit an alarm if 30ppm or 90ppm threshold is exceeded Sprint is programmed with pass fail criteria for this test Refer to British Standard BS7967 for further details on performing room CO safety tests Once test can be complete press the Stop soft key to finish Use the soft keys to print or log readings To stop test at any time press the ESC key Use the soft keys Quit To stop or Cont to carry on When printing Room CO safety test results Sprint will also provide data showing for how long the recommended level of CO was exceeded This is expressed as hh mm gt allowed It also provides what the recommended maximum level is according to the standard 8 TELEGAN gas monitoring Sprint V1 Operation I Operation Before using the Sprint flue gas analyser on any heating appliance installation ensure you are familiar with the working practices defined in the BS7967 which details how to perform certain tests and safety issues to consider Auto zero setting Allow the Sprint unit to auto zero during switch on in clean air before perform ing any of the following tests flue gas analysis and room CO safety test Ensure all tubes and probes are connected before switch on and zero outside in clean air Smart purge Pump control Sprint runs the internal pump during purge at switch on and switch off and during some tests The speed and therefore the soun
6. Sprint unit ensure the water trap is empty and is fitted in the right direction Do not insert probe in flue till after auto zero Always start in clean air Press and hold the ON OFF ENTER key for two seconds Sprint will emit a few ris ing beeps performs a screen test and starts the pump to purge any residual gas from the unit Ensure the gas exhaust outlet is not blocked Auto zero The welcoming splash screens are displayed for approximately 30 seconds and will automatically go to the Auto zero screen The splash screens show software version serial number calibration due date and header details When Auto zero is complete the display will change to the Test menu Warm up should be extended if sensor needs to recover from previous gas exposure When Sprint has completed warm up the auto zero Tip auto zero screen will be displayed Ensure you are in clean air Sprint outside of before proceeding o ee a re Ensure the sensor reading is stable press any of the top td ant three buttons to begin auto zero Sprint will display gases Hg vicinity if the gas readings are stable Otherwise it will display affecting the auto or _ After a successful auto zero Sprint will display the zero process Test menu screen and switch off the pump Note Sprint will switch off the pump when not performing tests or purging Using the menu Press the ESC key to display the Main menu Use the soft keys to scroll select and edit menu ite
7. TELEGAN gas monitoring Sprint V1 Specification V Specification Instrument Operating temperature range 10 C to 50 C 14 F to 122 F Battery Lithium lon Life up to 9 hours dependant on test used 15 left warning Recharge time Minimum 6 hours from flat Charger input voltage 230 V 50 Hz AC Standard Fuels Natural gas LPG heavy oil light oil coal wood coke Biomass and Bagasse Display Back lit segment and matrix combined LCD Menu Intuitive structure tab selection on screen Dimensions 75 x 168 x 65 mm Function buttons key pad 5 button keypad Weight 440 g 0 971bs Pump Flow fail indication SmartPurge Enclosure Integrated robust protective case Protective rubber boot with integral magnets Standards BS7927 EN50379 BS7967 Gases Range Display Resolution Accuracy Detection Response Recovery Diagnostics limit time time t90 Oxygen 0 1 0 2 0 3 viv 50 sec 30 sec v 0 25 Carbon mon 1ppm lt 20ppm 1ppm 90 sec 60 sec v oxide 3ppm 0 10 000ppm gt 20ppm 3 Carbon dioxide 0 1 40 2 vv 0 2 VV 50 sec 30 sec Calculated 0 25 CO CO2 0 0001 Ratio 0 to 0 9999 TELEGAN 15 gas monitoring Specification Other Measurements Sprint V1 Measurement Range Temperature 50 to 1100 C 58 to 2012 F selectable C or F Efficiency 0 100 net or gross selectable XSAir 0 100 Commu
8. d emitted by the pump will vary depending on the test being conducted To save on the battery life Sprint will turn off the pump when it is not required When the pump is running ensure the exit gas exhaust is not blocked and do not breath in the exhaust gases 1 1 Menu and operator button overview Menus Sprint provides a large custom LCD with blue backlight Navigation and func tions are provided by three soft key buttons which change according to what you are doing Use the soft keys to navigate menus select start and end tests change options and select actions such as print and log Use the escape key to exit menus exit tests and step back through screens After your Sprint unit has been switched on and performed an auto zero the display will show the Test menu screen ready for use To display the Main menu press the ESC key To enter a menu item use the up and down arrow soft keys below the screen icons to scroll the menu list and press the centre soft key to select Some menus have their own submenu Press the ESC key to exit a menu press TELEGAN 9 gas monitoring Operation Sprint V1 ESC twice to return to the Main menu from a submenu The soft keys control the function displayed above them on the screen These will change depending on the test or menu or if the ESC key is pressed The menu structure is shown below How to display unit serial number identity and software ver sion Press the ESC key
9. elect Tests from the Main menu Refer to section III Operation for more details 1 Flue gas analysis Before performing the flue gas analysis test check the water trap is clean and is upright arrow should point upwards To begin test select Flue Gas from the Test menu Sprint will ask you to select a fuel by using the up and down arrows on screen Press the centre button to make your selection Sprint will switch on the pump and begin making measure ments Use the soft keys to print or log the readings Press the centre button to display the three screens available Screen 1 CO CO2 CO CO2 ratio Screen 2 O2 excess air and efficiency Screen 3 O2 CO and temperature To end test press the ESC key Warning During testing ensure the combined filter and water trap is not blocked or full Failure to do so may result in an error message Note for personal safety a carbon monoxide alarm will Tip f activate at 90ppm CO This will deactivate when CO levels is displayed flue drop below 45ppm This is to protect the user from poten probe is not in flue tial hazardous exhaust gas levels or not connected 2 Differential temperature test Sprint can perform a differential temperature test with one thermocouple probe To begin test select Diff Temperature from the Test menu Check the units displayed are the correct temperature scale If no probes are connected Sprint will display Single probe
10. from the Main menu Main menu Press ESC to go to Main menu Test Menu Flue Gas Diff CO Room Safety Fuel selection Temperature Appliance I Temp units Eff sel Auto off Back light Key click Header 1 Header 2 Efficiency Settings Time date Lf Logs l Select a log Find log Clear logs 10 TELEGAN gas monitoring Sprint V1 Setting Up ll Setting Up 2 1 Configuration Options Fuel Options Sprint displays the current fuel selection before entering the flue gas analysis test screen To change the fuel option use the soft keys below the up and down arrow symbols to select the fuel and the centre button to confirm the selection Sprint then enters the flue gas analysis screen Fuel options available are natural gas LPG heavy oil light oil coal wood coke biomass bagasse To make other configuration changes press ESC to enter the Main Menu and select Settings Press the up and down arrow keys to move between selections and the centre key to confirm selections Temperature Units Use the soft keys to select the submenu Temp units Temperature units are degrees Centigrade C or degrees Fahrenheit F Efficiency Selection Use the soft keys to select the submenu Eff sel Efficiency settings are Net or Gross Auto off Selection Use the soft keys to select the submenu Auto off Auto off options are Disable 5 10 15 20 25 30 45 60 75 and 90 mi
11. ide sss sssseiisseeiiseserieesrrrrrerrrr rren 18 Appendix Carbon Monoxide Room Safety Tests 20 Unpacking Sprint V1 Sprint V1 Multi function Flue Gas Analyser Thank you for purchasing the Sprint V1 Multi function Flue Gas Analyser Sprint has redefined flue gas analysis and will give you years of unparalleled service and reliability Please read the instructions carefully before use Keep the manual for future reference Unpacking Important ensure unit is fully charged Remove the Sprint unit from the packaging The Sprint accessories will be located in the carry case Check the contents are complete you should have e Carry case e Sprint unit e Mains battery charger power supply e Rubber boot e Probe including water trap and tubes e User manual e Certificate of calibration Warning Do not attempt to use any other charger power supply with this unit except the one s supplied Failure to comply could invalidate the warranty and may result in permanent damage to the unit 2 TELEGAN gas monitoring Sprint V1 Overview Overview Display Battery icon Three bars show Soft keys Use these unmarked buttons complete charge to make selections from the menu start and stop tests print and store results See section Operation for more details ON OFF ENTER key Press and hold button for two seconds to switch on and switch off Spri
12. ings Note that resetting the time and date changes your cal due date by the same amount 12 TELEGAN gas monitoring Sprint V1 Battery Charging lll Battery Charging The Sprint has a lithium ion rechargeable battery and will operate for up to 9 hours when fully charged depending on the type of test used Low battery When the battery is low Sprint will display a low battery icon If the battery gets too low then Sprint will give further warning before switching off Use of the charger as a power adaptor The Sprint charger can be used to power the unit and will continue to charge whilst operating the unit Automatic battery saver The Sprint unit will automatically power down if left unused for 15 minutes adjustable Sprint will warn user when power down is imminent Charging the batteries 1 Plug the charger into a mains socket 2 Connect the charger to the Sprint using the power jack on the side of the unit Switch on the power at the mains socket The unit would normally be left switched off for charging The display will show the battery charging icon Warning Do not attempt to use any other charger with this unit except the one s supplied Failure to comply could invalidate the warranty and may result in permanent damage to the unit TELEGAN 3 gas monitoring Maintenance and Calibration Sprint V1 IV Maintenance and Calibration General To keep the display panel and operato
13. l or unstable FAILED 2 CO unacceptably high for too long FAILED 3 CO dangerously high FAILED 4 Unacceptable or incomplete Press ESC key to continue Pass Cases Normal Acceptable Peak of CO Normally for a test to pass the CO levels must peak without exceeding the maximum allowed CO level and then fall by at least 0 5 ppm below the peak value before the end of the test Note that it is not necessary for the CO level to reach or be close to zero at the end of the test so long as it remains below the maximum allowed CO level The result code is PASSED Very Low Levels of CO If the CO levels remain below 2 5 ppm ie close to clean air or background noise levels for the duration of the test then the test is considered to have passed The result code is PASSED Failure Cases Excessive Levels of CO If the CO level exceeds the CO alarm level then the test is considered to have failed and should be immediately aborted The CO alarm may be triggered at any stage before during and after completion of the test and continues to annunciate until the CO level returns to a safe level The result code is FAILED 3 TELEGAN a gas monitoring Appendix Carbon Monoxide Room Safety Tests Sprint V1 NB The alarm should prompt the operator to take appropriate action according to BS7967 and safety procedures The sensors in the instrument should be purged with clean air and allowed to rec
14. ms Press the up or down arrows to make change or the ESC key to cancel Press the centre button to accept the change or the ESC key to cancel TELEGAN gas monitoring a Quickstart guide Sprint V1 Changing fuel and units Fuel selection is made on entry into the Flue Gas Analysis test Fuel Options select one of the following options natural gas LPG heavy oil light oil coal wood coke Biomass and Bagasse Temperature selection is from the settings menu From the Main menu select Settings From Settings menu select Temp units Temperature scale Centigrade or Fahrenheit Switch off Turn off unit in clean air and ensure any gas is purged from sensors Press and hold the ON OFF ENTER button for approximately two seconds The power off screen will be displayed and the pump will run to purge the sensors The pump will run for up to 30 seconds to purge unit if gas is present Sprint will normally switch off in 10 seconds Press the ESC key to cancel the switch off sequence Charging battery Plug the supplied charger into the charger socket The batteries will recharge in six hours from flat You may leave the unit on charge for longer periods e g overnight without damaging the unit A fully charged unit will give up to nine hours of operation A shorter charge time can be applied such as 1 2 hour to allow a more limited length of operation 6 TELEGAN gas monitoring Sprint V1 Performing tests Performing tests S
15. n utes Press Set or disable to select your preferred auto off timeout period Backlight Control Use the soft keys to select the submenu Back light Back light settings are Controlled Default Off Dim and Bright Keyclick control Use the soft keys to select the submenu Key click Keyclick control settings are Enable and Disable This controls whether Sprint emits a beep when you press a keypad button TELEGAN i gas monitoring Setting Up Sprint V1 Header 1 and Header 2 Use the soft keys to select the submenu Header 7 or Header 2 These are your details shown on the log printout This sets the top line 1 or bottom line 2 of the header and is displayed at the top and bottom of the display during start up The A button sequences back through alphabet and symbols and the Z button sequences forward through alphabet and symbols The button saves current letter and moves cursor to set next letter Once cursor is moved through all previously saved spaces the button changes to Save Press Save to complete changes or or buttons to con tinue up to maximum number of symbols 24 Time and Date Use the soft keys to select the submenu Time Date Use the and keys to set the Hour Min Day Month and Year sequen tially Once you have set the Hour press the button marked Hour and Sprint sequences to Min Likewise press Min to save minutes and move to Day Press ESC to save your date time sett
16. nications IR Port 16 TELEGAN gas monitoring Sprint V1 Accessories and Spare Parts X Accessories and Spare Parts Telegan General Spares Part number Description CAS29003 PRB29000 ENC29001 CHG29001 FIL29001 FIL99008 TELEGAN gas monitoring SPRINT V CARRY CASE SPRINT V FLUE PROBE INC WATER TRAP WITH FILTER SPRINT V PROTECTIVE BOOT WITH MAGNETS SPRINT V UK CHARGER SPRINT V WATER TRAP INC FILTER FILTER ELEMENT 17 Troubleshooting Guide Sprint V1 Vill Troubleshooting Guide Sprint will provide on screen messages which advise clear actions Contact Anton if unsure on how to proceed Symptom Cause Recommended User Action Instrument will not Battery flat Connect charger and retry turn on when on off Battery may be flat The instru button is pressed and ment is designed to prevent held for 2 seconds deep discharge occurring and will turn it self off when battery level gets too low Instrument will not Battery flattened Ensure charger is correct type turn on and charger symbol is not displayed when charger is con nected beyond standard charging point If so plug in and leave con nected Check to see if charg ing symbol appears every 4 6 hours If it does not and unit does not switch on return both unit and charger for ser vice Pump sometimes stays on at a faster rate par ticularly when exiting the Flue Gas Analysis or CO Room Test Smart
17. nt unit Use this key to finish edits and save changes ESCAPE key Use this button to quit tests and step back through menus Thermocouple socket K type connector Exhaust exit Warning do not inhale exit gas exhaust or block exit IR window Printer output senses ambient light to control backlight Tip to quickly activate backlight cover this window TELEGAN 3 gas monitoring Probe connections Sprint V1 Probe connections Flue gas analysis Connect the flue probe to the flue sample inlet and the thermocouple to the k connector marked FLOW Efficiency Differential temperature test Connect one thermocouple probe to the k type connector Sprint will display a soft key option to switch between T1 and T2 snapshot measuring points Room CO safety test Connect CO room safety test probe optional where suitable to the flue sam ple inlet Sprint is supplied with a rubber boot for protection of your Sprint unit A mag net on the reverse of the rubber boot can be used to place the Sprint unit in location on the boiler system for easy hands free operation Note The pump will operate at different speeds or switch off depending on the test being performed This will vary the pitch of the sound from the pump and does not indicate the pump is performing incorrectly 7 TELEGAN gas monitoring Sprint V1 Quickstart guide Quickstart Guide Important please connect flue probe first before switching on your
18. or vehicle traffic or incorrect then either the test should be repeated or the operator should seek expert advice CO Room Test Pass and Fail Test Specifications Type C Type B Type A Type A Type A Room sealed Boiler Cooker Water heater Space heater appliance open flue flueless flueless flueless Max Allowed CO 10 ppm 10 ppm 30 ppm 10ppm 10 ppm Max Peak Duration 60 secs 60 secs 20 mins 30 secs 60 secs exceeding Max Allowed CO CO Alarm Level 30 ppm 30 ppm 90 ppm 30 ppm 30 ppm Min Test Duration 15 mins 15 mins 20 mins 5 mins 30 mins Max Test Duration 30 mins 30 mins 30 mins 10 mins 30 mins Result Codes The pass or fail result is displayed when the test completes and is printed on the report as well as recorded in the log If the test fails a code number is also displayed printed on the report and recorded in the log This fail code identifies the way in which the test failed and can help identify the cause Also when the 20 TELEGAN gas monitoring Sprint V1 Appendix Carbon Monoxide Room Safety Tests test completes a short text message associated to this code is displayed in a pop up prompt dialog screen to explain the reason for failure The result codes and associated prompt dialogue messages are as follows RESULT amp CODE POP UP PROMPT ON SCREEN PASSED None ANY FAILED Warning CO Room Safety test failed FAILED 1 CO levels did not fal
19. over Unacceptable Levels of CO A peak duration timer records whenever the CO level exceeds the maximum allowed CO level during the test If the total peak duration time exceeds the max peak duration allowed then the test failed due to unacceptable levels of CO The result code is FAILED 2 Acceptable Levels of CO with no Peak If there is no peak in CO level even though it remains below the maximum allowed CO level then the test is failed The result code is FAILED 1 Operator Pass Fail Cases The following results are considered to be operator determined whether the test passes or fails CO Level Exceeds Max Allowable Level for a Short Duration For some appliances eg cooker a peak exceeding the maximum allowed CO level may be acceptable provided the CO level falls back below this level within the max peak duration time It is up to the operator to determine if the result is acceptable or not If the operator chooses to fails the test the result code is FAILED 4 Otherwise the result code is PASSED Other Cases When a test is not performed correctly the results are inconclusive or there is insufficient or unreliable data Sprint V attempts to interpret the readings detected and fail the test It is possible for the results of a test to appear to be valid when it was performed incorrectly or the data collected was unreliable in some way Please refer to BS7967 2005 and ensure tests are carried out cor rectly S
20. print V tries to reject incorrectly taken test readings but should not be relied on to instruct engineers on correct working practice 22 TELEGAN gas monitoring Sprint V1 Appendix Carbon Monoxide Room Safety Tests Multiple Peaks of CO The overall peak CO reading recorded will be the latest peak CO reading that was higher than any previous The peak duration recorded will be the total time the CO readings exceeded the maximum allowed CO level The pass fail criteria are applied as before based on this information Sprint V does not expect to record multiple peaks of CO where the reading goes up and down more than once Unstable or Rising Levels of CO If the CO level in the last minute of the test is not lower than the overall peak CO reading then Sprint V assumes that the CO level has not fallen so the test will fail The test will also fail if there is a significant build up of CO levels at the end of the test and levels are still rising or the CO level in the last minute of the test is not lower than the overall peak CO reading The result code is FAILED 1 CO Level Exceeds Max Allowable Level and Peaks for Unknown Duration If the CO level exceeds the maximum allowed too close to the end of the test then the test is failed This can occur if there is a build up of CO levels towards the end of the test or the appliance fires up late on in the test or the test is stopped too early The test should be repeated if this occurs
21. purge is oper ating CO sensor is recovering from expo sure to gas Remove probe from flue Continue to use the instrument as normal Pump flow fail alert Filter water trap or sample line blocked Empty and clean filter water trap Ensure sample line is free from blockage Auto zero failed Sensors exposed to gas or faulty Switch off and on ensuring you zero in clean air and sen sors are purged If unit contin ues to fail auto zero return for service Alarm activates in CO Room safety test CO is reading greater than 30ppm Dependant on safety proce dures 18 TELEGAN gas monitoring Sprint V1 Troubleshooting Guide Symptom Cause Recommended User Action During auto zero the CO reading does not stabilise CO sensor faulty or recovering from high CO exposure Return for service ASAP or ensure unit is purged and allow sensor to recover Sometimes negative CO readings are dis played High CO exposure zeroed in auto zero sequence Turn off and on and repeat auto zero in clean outside air allowing time for the CO sen sor to recover and stabilise Sometimes gt or I lt is displayed in place of a number Sensor is out of range Contact support for advice Return for service if problem persists or other failures are observed Cal due date has changed Time date has been amended Check current
22. r buttons free from dirt build up regularly wipe over your Sprint unit with a slightly damp cloth Protective boot To protect Sprint from dirt and knocks a rubber boot is supplied This boot is supplied with concealed magnetics in the rear 4 1 Unit The Sprint should be calibrated once a year Your Sprint unit will also display the calibration due date at switch on and will warn you when calibration due date is drawing near If the calibration due date has passed the Sprint unit will display a message Calibration overdue 4 2 Water trap The combined filter and watertrap is used in line between the probe and the unit Before performing any tests check that the filter is clean and there is no water inside the trap Filter Bung Arrow keep arrow upright a during flue gas analysis q oo ZZ ICV a f x gt L 4 o J I amp I qq V j LJ a j Sooo N gt A ab Sa Residual water Filter The filter element should be changed if the filter has Warning Filters become contaminated or dirty If the filter has become must be used at all soaked with water remove from trap and leave to dry times Failure to do before reusing Ensure O ring remains in place so may invalidate the warranty To change the filter Unscrew the filter housing remove the old filter and replace Water To empty the water trap unscrew the filter housing remove the bung and empty Replace bung after emptying 14
23. the respon sibility of the user to take suitable precautions that prevent it from being adversely affected 3 The equipment is only certified for use in ambient temperatures in the range 10 C to 50 C and should not be used outside this range Use only charger s supplied by Telegan 5 Use only the appropriate Telegan supplied cables for connection to the sockets on the instru ment Copyright Telegan Gas Monitoring 2006 All rights are reserved No part of the document may be photocopied reproduced or translated to another lan guage without the prior written consent of Telegan Gas Monitoring Publication number M09990 First edition August 2006 TELEGAN gas monitoring Contents Sprint V1 Multi function Flue Gas Analyser 2 Quickstart GUIde ec ceceeeeccce eee eeeeeeeeeeeaneneeeeeeaaneess 5 I OG Fatt OM mpira ETE OEE 9 1 1 Menu and operator button Overview cece ccccccccecseeeeeessteeeeteeeeees 9 Me SN tee retet tat ee ieone E eee mancouee 11 2 1 Configuration OptiOnS ccceceeeeeceeeeeeeeeeeseeeeeeeseeeesseeeseeeneees 11 lll Battery ChargiNg essi mo aneneen e GE EES 13 IV Maintenance and Calibration ccccececeeeeeeteeeeeeees 14 A Wine AE EE A elas 14 A 2 Water tapas eene na ERRER TEIE E EEEE EER 14 V Specification ss nnssssknisekiireeerarttttrrettrrnrrrrinterrin errr rrer 15 X Accessories and Spare Parts cccccecceeeeeeteeeeesteeeees 17 VIII Troubleshooting Gu

Download Pdf Manuals

image

Related Search

Related Contents

Philips RFX2001WM IR/RF home control Universal remote control  Untitled  6. Technical specifications of Wireless LAN  MANUEL DE L`UTILISATEUR  安全上のご注意  los proyectos hacen los objetos eternos, las modas los  SIMATIC IPC Remote Manager - Service, Support  VM 740 - Domyos  Harman/Kardon Nova  Heating System  