Home

0650 v1.4 Duolink II Fluorescence User Manual

image

Contents

1. whether to show the heavily expressed areas or to show all PLA signals even individual ones Duolink Il Fluorescence 21 8 3 DUOLINK IMAGETOOL IMAGE ANALYSIS To analyze the results use Duolink ImageTool to obtain objective quantification of PLA signals By using the software it is possible to obtain either the number of signals and cells per image allowing average measurements or to allocate each individual signal to a specific cell using single cell analysis The nuclei are automatically detected and cytoplasm size estimated enabling single cell statistical analysis of expression levels in tissue or cell populations A result can typically look as shown in Figure 12 Voca op Tos tropa dere LJ i 1 i la D Sey a Fig 12 Analysis of an image from a fluorescence microscope Nuclei are blue the red spots are the PLA signals representing the protein target of interest PLA signals marked with white circles and nuclei outlined in yellow are quantified at analysis The green outlines represent the user defined cytoplasm size When studying highly expressed proteins the density of PLA signals may be so high that it is impossible to discern the individual signals see Figure 11 Duolink ImageTool can not quantify regions of merged signals only separable signals can be counted For images taken with fluorescence microscopy a quality control function displays a warning if signals are merged In this case the
2. in close proximity This also enables you to verify the Duolink procedure 5 5 2 NEGATIVE CONTROL Ideally you should include a negative control with a cell line or tissue that does not express one or both of your targets Such a control will give you information on the specificity of your primary antibodies If this control gives you background signals in excess of signal obtained from omitting the primary antibody antibodies it means that your primary antibodies under the present conditions are binding to other targets than expected If facing problems with signals in your negative cell line tissue you might need to titrate your primary antibodies or look for more specific antibodies 5 5 3 NEGATIVE CONTROL TECHNICAL By omitting primary antibodies you will get a hint of how the PLA probe background looks like in your system Alternatively if you are using an approach with two primary antibodies you can also choose to omit only one of your primary antibodies If facing problems with background signals with the primary antibodies omitted consult the troubleshooting guide see section 9 Duolink Il Fluorescence 13 6 Reagent preparation 6 1 DUOLINK II REAGENTS Some Duolink Il reagents are supplied as concentrated stocks Dilute required volumes of the stocks Note Do not store diluted reagents Blocking Solution store at 4 C b Vortex before use Ready to use blocking solution one drop equals approximately
3. jars If amounts or fixation the problem remains sterile filter all washing solutions of non precipitates Wash your cells at least twice to ensure that the culture medium is amplification cause highly removed before adding the fixative dependent fluorescent background particles Fluorescent Ink type pen has Never label slides with ink type pens Use diamond or graphite pen mist over been used entire slide Inappropriate Use Duolink Il Wash Buffers A and B during the Detection protocol wash buffers Inappropriate Use Duolink Il Mounting Medium with DAPI for mounting of the mounting slides medium Contamination in Discard wash solutions and use new solutions and washing jars wash solutions Red Inappropriate Use Duolink Il Wash buffer B in the final washing step after the fluorescent wash buffers Amplification nuclei Mis Inappropriate Use Duolink II Mounting Medium with DAPI for mounting of the mounting slides medium Large Uneven spread Ensure that entire area of investigation is covered during each step variation of reagents of signals during different over slide steps individual reaction Partial drying of Check before each new step that no region has dried during parts of the area incubation If so use a better humidity chamber during incubation 24 Duolink Il Fluorescence PROBLEM PROBABLE CAUSE SUGGESTED SOLUTION The reaction Noor insufficient Use a grease pen or alike to enci
4. the slides Try to obtain equal residual volume for each slide as this will affect reproducibility Do not allow the samples to dry before addition of the primary antibodies as this will cause background c Immediately add the primary antibody solution d Incubate in a humidity chamber Use the optimal incubation temperature and time for your primary antibodies 3 PLA PROBES a Dilute the two PLA probes 1 5 in Antibody Diluent E g for a 40 ul reaction take 8 ul of PLA probe MINUS stock 8 ul of PLA probe PLUS stock and 24 ul of Antibody Diluent b Tap off the primary antibody solution from the slides c Wash the slides in 1x Wash Buffer A or in a wash buffer suitable for your primary antibody preferably 2 x 5 min Washing should be performed in a staining jar with a minimum volume of 70 ml on a shaker cradle with gentle orbital shaking Bring the wash buffers to room temperature before use d Add the PLA probe solution e Incubate the slides in a pre heated humidity chamber for 1 h at 37 C Proceed with the detection protocol in section 73 on page 18 Duolink II Fluorescence 17 7 3 DETECTION PROTOCOL Proceed from step 3e of the PLA probe protocols Use open droplet reactions without a cover slip and perform all incubations in a humidity chamber The volume examples are based on 40 ul reaction volume suitable for 1 cm reaction area Adjust the volumes corresponding to your specific d
5. to set up assay J e with high sensitivity In this case only one antibody needs to bind s the target under the conditions used fixation retrieval buffer etc Using one primary antibody is recommended for detection of single target proteins only when a well performing specific primary antibody is available If you use an unspecific antibody you will see high background due to the sensitivity of the assay Using only one primary antibody is more efficient than using two primary antibodies when detecting low abundant targets Fig 6 A Single recognition B Two primary antibodies double recognition Usage of two primary antibodies will give you an assay with superior specificity When using two primary antibodies they must be directed against different non competing epitopes on the same target molecule The two primary antibodies must have been raised in different species Also both primary antibodies must bind to the target under the same conditions fixation retrieval buffer etc Fig 7 B Double recognition Duolink Il Fluorescence 9 5 1 2 PRIMARY ANTIBODIES FOR DETECTION OF PROTEIN MODIFICATIONS Detection of protein modifications such as phosphorylations f J often suffers from low specificity Our recommended strategy is ts to use two primary antibodies one against the target protein and one against a modification site on the same protein Preferably a specific modification site is t
6. 1 The same agent used for blocking of the sample should be present also in the antibody incubation buffer preferably together with some detergent to reduce background staining as well as in the incubation buffer for the PLA probes Note Ensure that you do not use bulk IgG from the same species as your primary antibodies as blocking reagent as this will cause false signals from the PLA probes PLA probe protocol Duolink II Solutions Use this protocol if you have no previous experience or no recommendation from the antibody vendor regarding blocking and antibody diluents see section 72 Note The Duolink Il Antibody Diluent contains salt blocking agent and detergents All to prevent unspecific binding of the antibodies It is important that you first verify the function of your primary antibodies one at the time using Duolink single recognition or possibly IF or IHC Use proper controls to verify that your individual antibodies do bind the correct target under the conditions used Duolink Il Fluorescence 5 5 CONTROLS To be able to properly evaluate your results it is advisable to include both biological and technical controls Examples of suitable controls are given below 5 5 1 POSITIVE CONTROL Ideally you should include a cell or tissue type that you know contains your target protein and that your primary antibodies should bind to If you are using two primary antibodies the primary antibodies should be known to bind
7. 2 min under gentle agitation Tap off all wash solution after the last washing d Remove the Polymerase from the freezer using a freezing block 20 C Add Polymerase to the Amplification solution from step a ata 1 80 dilution and vortex E g for a 40 ul reaction add 0 5 ul of Polymerase to 39 5 ul of Amplification solution e Add the Amplification Polymerase solution to each sample f Incubate the slides in a pre heated humidity chamber for 100 min at 37 C 18 Duolink Il Fluorescence 6 FINAL WASH STEP a b c d Note Light sensitive reagents Wash and dry the slides protected from light Ensure to have 1x and 0 01x Wash Buffer B available Tap off the Amplification Polymerase solution from the slides Wash the slides in 1x Wash Buffer B for 2 x 10 min Wash the slides in 0 01x Wash Buffer B for 1 min Let the slides dry at room temperature in the dark 7 PREPARATION FOR IMAGING Mount your slides with a cover slip using a minimal volume of Duolink II Mounting Medium with DAPI ensuring no air bubbles get caught under the cover slip Nail polish can be used to seal the edges Wait for approximately 15 min before analyzing in a fluorescence or confocal microscope After imaging store the slides at 20 C in the dark Duolink Il Mounting Media with DAPI is aqueous and does not solidify Duolink Il Fluorescence 19 8 Results 8 1 TYPICAL RESULTS The result from a Duolink I
8. 40 ul Antibody Diluent store at 4 C Vortex before use For dilution of primary antibodies and PLA probes PLA probe 5x store at 4 C b Vortex before use Dilute 1 5 in Antibody Diluent or custom diluent immediately before use and vortex the solution before addition to the sample Ligation 5x store at 20 C b Thaw at room temperature and vortex before use gt Dilute 1 5 in high purity water immediately before use Note The buffer contains DTT that may precipitate during freezing Vortex to dissolve homogenize Ligase 1 U l store at 20 C The Ligase should be kept at 20 C at all times Use a freezing block 20 C when removing the enzyme from the freezer b Add the Ligase to the reaction mix at a 1 40 dilution immediately before addition to the sample Ensure that the Ligation Ligase reaction solution is thoroughly vortexed before addition to the sample Amplification 5x store at 20 C gt Thaw at room temperature and vortex before use b Dilute 1 5 in high purity water immediately before use Polymerase 10 U l store at 20 C The Polymerase should be kept at 20 C at all times Use a freezing block when removing the enzyme from the freezer Add the Polymerase to the reaction mix at a 1 80 dilution immediately before addition to the sample gt Ensure that the Amplification Polymerase reaction solution is thoroughly vortexed before addition to the sample 14 Duoli
9. Fluorescence USER MANUAL The protocols in this manual are compatible with all Duolink Il PLA probes Duolink Il Detection Reagents Green art no 92014 Orange art no 92007 Red art no 92008 and Far Red art no 92013 Table of content 1 2 9 APPENDIX A DUOLINK II WASH BUFFERS FORMULAS INTRODUCTION sos cotes eene ete teet PRINCIPLE OF THE ASSAY occitano APPLICATIONS adiac eterna 3 1 Detect and quantify protein interactions 3 2 Detect and quantify protein modifications 3 3 Detect and quantify protein expression REAGENTS AND EQUIPMENT 41 Duolink Il reagents 4 22 Reagents to be supplied by the user 43 Equipment needed 44 Duolink ImageTool software for image analysis ASSAY CONSIDERATIONS 5 1 Choice of primary antibodies 5 2 Primary antibody optimization 5 3 Sample type 5 4 Pre treatment 5 5 Controls REAGENT PREPARATION cio ccoo init 6 1 Duolink Il reagents 6 2 Duolink II Wash Buffers 6 3 Samples 6 4 Reaction volume ASSAY PROTOCOL ccoo 7 1 PLA probe protocol CUSTOM solutions 7 2 PLA probe protocol Duolink II Solutions 73 Detection protocol tenentes RESULTS 8 1 Typical results 82 Image acquisition 83 Duolink ImageTool image analysis s TROUBLESHOO FI NG asco tt tette reset Duolink Il Wash Buffer A Duolink Il Wash Buffer B REFERENCES
10. argeted but it is also possible to use a generic antibody against a modification site Using two primary antibodies will give you superior specificity for your modification assay compared to using one single primary antibody The two primary antibodies must be raised in different species Also both primary antibodies must bind to the target under the same conditions fixation retrieval buffer etc Fig 8 Protein modifications 5 1 3 PRIMARY ANTIBODIES FOR DETECTION OF PROTEIN INTERACTIONS Duolink provides an excellent way to detect protein interactions This c is done using two primary antibodies each directed against one of F the targets of interest The two primary antibodies must be raised in different species and must bind to the target under the same conditions fixation retrieval buffer etc Ne Note Detection of homodimers is challenging The Duolink Il Probemaker reagents can be used to directly conjugate primary antibodies with oligonucleotide PLUS or MINUS avoiding the prerequisites of the secondary PLA probe approach Contact Olink Bioscience for further information support olink com Fig 9 Protein interactions 5 2 PRIMARY ANTIBODY OPTIMIZATION You will save time and increase your success rate if you optimize your primary antibodies to be sure that they work properly in your material The conditions for your primary antibodies should be optimized with respect to Sample fixation Ant
11. artered in Uppsala Sweden For Technical Support please contact us at support olink com Please be welcome to visit our web page at www olink com This product is for research use only Not for use in human diagnostic or therapeutic procedures This product includes a license for non commercial use of the Duolink product Commercial users will require additional licenses Please contact Olink AB for details There are no warranties expressed or implied which extend beyond this description Olink AB is not liable for property damage personal injury or economic loss caused by this product The following trademarks are owned by Olink AB Olink Olink Bioscience Duolink and PLA This product is covered by several patents and patent applications including US 6 511 809 US 6 558 928 US 6 8785 15 US 7 074 564 US 5 665 539 and related US and foreign patents Olink Bioscience Dag Hammarskj lds v 54A SE 751 83 Uppsala Sweden BIOSCIENCE www olink com 0650 v 1 4 2010 11 25
12. ble for your primary antibodies Washing should be performed in a staining jar with a minimum volume of 70 ml on a shaker cradle with gentle orbital shaking Bring the wash buffers to room temperature before use d Add the diluted PLA probe solution e Incubate the slides in a pre heated humidity chamber for 1 h at 37 C gt gt Proceed with the detection protocol in section 73 on page 18 16 Duolink Il Fluorescence 7 2 PLA PROBE PROTOCOL DUOLINK II SOLUTIONS Use this protocol if you use the Duolink II Blocking Solution for blocking of your sample and Duolink II Antibody Diluent for dilution of your primary antibodies Before you start your samples should be deposited on glass slides and pre treated with respect to fixation retrieval and or permeabilization Use open droplet reactions without a cover slip and perform all incubations in a humidity chamber The volume examples are based on 40 ul reaction volume suitable for 1 cm Use volumes corresponding to your delimited reaction area see the Reaction Volume Guide Art no 80520 or section 6 4 1 BLOCKING a Addone drop of Blocking Solution per 1 cm Ensure to cover the entire reaction area with Blocking Solution b Incubate the slides in a pre heated humidity chamber for 30 min at 37 C 2 PRIMARY ANTIBODIES a Dilute your primary antibody or antibodies to suitable concentration in the Antibody Diluent b Tap off the Blocking Solution from
13. carita uenerat petet aen Duolink Il Fluorescence 3 4 1 Introduction Duolink I reagents from Olink Bioscience enable detection visualization and quantification of individual proteins protein modifications and protein interactions in tissue and cell samples prepared for microscopy The target is detected using one or two primary antibodies depending on the application In the case that two primary antibodies are used they must have been raised in different species The Duolink reagents are based on in situ PLA which is a proximity ligation assay technology A pair of oligonucleotide labeled secondary antibodies PLA probes generates a signal only when the two PLA probes have bound in close proximity either to the same primary antibody or two primary antibodies that have bound to the sample in close proximity The signal from each detected pair of PLA probes is visualized as an individual fluorescent spot These PLA signals can be quantified counted and assigned to a specific subcellular location based on microscopy images Duolink Il Fluorescence 2 Principle of the assay Typical starting materials are adherent cells cytospin preparations or tissue sections on a glass slide fixed pre treated and blocked with a blocking reagent according to the requirements of the primary antibodies used A E D The samples are incubated with primary antibodies that bind to the protein s to be detected Sec
14. chamber Use volumes corresponding to your delimited reaction area see the Reaction Volume Guide Art no 80520 or section 6 4 1 BLOCKING Use your previously tested blocking solution a Add blocking solution to each sample b Incubate the slides 2 PRIMARY ANTIBODIES Use your previously tested buffer to dilute your primary antibodies a Dilute your primary antibody to a suitable concentration in antibody incubation buffer If using two primary antibodies dilute the two antibodies in the same solution b Tap off the blocking solution from the slides Try to obtain equal residual volume for each slide as this will affect reproducibility Do not allow the samples to dry before addition of the primary antibodies as this will cause background c Add the primary antibody solution d Incubate in a humidity chamber Use temperature and time optimal for your primary antibodies 3 PLA PROBES Use the same buffer as for your primary antibodies or use Duolink Il Antibody Diluent Note The buffer should contain the blocking agent that was used for blocking the samples a Mixand dilute the two PLA probes 1 5 in your chosen buffer Allow the mixture to sit for 20 min at room temperature E g for a 40 ul reaction take 8 ul of PLA probe MINUS stock 8 ul of PLA probe PLUS stock and 24 ul of the antibody buffer b Tap off the primary antibody solution from the slides c Wash the slides in a wash buffer suita
15. data is best analyzed by fluorescence intensity measurement or area fraction of sample with signal using software for traditional fluorescence analysis The primary antibodies used in the assay can be titered down to reduce the number and incidence of merged signals 22 Duolink Il Fluorescence 9 Troubleshooting Some general guidelines are given below PROBLEM PROBABLE CAUSE No or few No or insufficient signals in binding of SUGGESTED SOLUTION Optimize your primary antibodies individually with the Duolink II single recognition approach IHC or IF in positive control samples Optimize parameters such as fixation retrieval protocol blocking and buffer conditions etc for the subsequent Duolink Il double recognition assay Evaluate and optimize fixative retrieval incubation temperature time concentration and buffer composition positive primary samples antibodies Insufficient reaction volume Ensure that your reaction area corresponds to the reaction volume The droplet must cover the reaction area Encircle your reaction area using a hydrophobic barrier and incubate the slides in a humidity chamber to prevent evaporation Do NOT use a cover slip to disperse the droplet Mounting media The intensity of the PLA signals will decrease quickly in some mounting media Duolink II Mounting Med optimized for preserving the PLA signals ium with DAPI is Wrong filter used for acquisition The fluoro
16. elimited reaction area see the Reaction Volume Guide Art no 80520 or section 6 4 4 LIGATION a Dilute the Ligation stock 1 5 in high purity water and mix Wait to add the Ligase until immediately before addition to the samples Take the addition of Ligase into account when calculating the amount of water added E g for a 40 ul reaction take 8 ul of the 5x Ligation stock and 31 ul of high purity water b Tap off the PLA probe solution from the slides c Wash the slides in a wash buffer suitable for your primary antibody or 1x Wash Buffer A for 2x5 min under gentle agitation d Remove the Ligase from the freezer using a freezing block 20 C Add Ligase to the Ligation solution from step a at a 1 40 dilution and vortex E g for a 40 ul reaction add 1 of Ligase to 39 ul of Ligation solution e Add the Ligation Ligase solution to each sample f Incubate the slides in a pre heated humidity chamber for 30 min at 37 C 5 AMPLIFICATION Note Light sensitive reagents a Dilute the Amplification stock 1 5 in high purity water and mix Wait to add the Polymerase until immediately before addition to the sample Take the addition of Polymerase into account when calculating the amount of water added E g for a 40 ul reaction take 8 ul of the 5x Amplification stock and 31 5 ul of high purity water b Tap off the Ligation Ligase solution from the slides c Wash the slides in 1x Wash Buffer A for 2 x
17. ester K Hydbring P Bahram F Larsson LG and Landegren U Direct observation of individual endogenous protein complexes in situ by proximity ligation Nat Methods 3 995 1000 2006 Gullberg M Gustafsdottir SM Schallmeiner E Jarvius J Bjarneg rd M Betsholtz C Landegren U and Fredriksson S Cytokine detection by antibody based proximity ligation Proc Natl Acad Sci USA 101 8420 24 2004 Fredriksson S Gullberg M Jarvius J Olsson C Pietras K Gustafsdottir SM stman A and Landegren U Protein detection using proximity dependent DNA ligation assays Nat Biotechnol 20 473 77 2002 Duolink Il Fluorescence 27 Olink Bioscience is based on a strong intellectual property portfolio focused towards the development of cutting edge technologies for biomolecular analysis in research and diagnostics The company s proprietary Duolink product line enables researchers in academia and the pharmaceutical industry to visualize and accurately quantify individual proteins their interactions and modifications in unmodified cells and tissues Such information is crucial to help elucidate disease mechanisms leading to the development of novel diagnostics and therapeutics The Duolink line is marketed and sold through a growing distributor network A second revenue stream comes from the out licensing of other proprietary technologies to major biotech supply companies Olink Bioscience is a privately held company founded in 2004 and headqu
18. h Buffers contains 3 pouches of Wash Buffer A and 1 pouch of Wash Buffer B to be dissolved in high purity water 4 2 REAGENTS TO BE SUPPLIED BY THE USER Reagents required for fixation and antigen retrieval of the sample according to your own protocol specific for each antigen and antibody used Primary antibody or antibodies matching a set of PLA probes 4 3 EQUIPMENT NEEDED b Fluorescence microscope equipped as follows Excitation emission filters compatible with fluorophore and nuclear stain excitation emission Camera and software for image acquisition optional gt Staining jars Pen or mask for delimitation of reaction area grease pen or silicon mask e g ImmEdge Pen from Vector Laboratories Shaker Humidity chamber moist chamber Freeze block for enzymes Incubator 37 C Pipettes covering the range from 1 ul to 1000 pl Cover slips compatible with fluorescence microscopy v v v v v v v High purity water sterile filtered MilliQ or similar 4 4 DUOLINK IMAGETOOL SOFTWARE FOR IMAGE ANALYSIS The result from a Duolink Il experiment is typically a number of distinct fluorescent spots PLA signals of sub micrometer size in various locations of the studied cells Use Duolink ImageTool to obtain objective quantification of PLA signals The nuclei are automatically detected and cytoplasm size estimated enabling single cell statistical analysis of expression levels in tissue or cell p
19. ification site on the same protein 3 3 DETECT AND QUANTIFY PROTEIN EXPRESSION Detection and quantification of protein expression can be done with two different approaches A To detect and quantify one single protein with high sensitivity use only one primary antibody B To detect and quantify one single protein with high specificity use two different primary antibodies directed against two different epitopes on the same protein Duolink Il Fluorescence Fig 2 Protein interactions Fig 3 Protein modifications Fig 4 A Single recognition Fig 5 B Double recognition 4 Reagents and equipment 4 1 DUOLINK II REAGENTS The Duolink Il reagents are generic reagents using secondary antibodies for detecting presence of analyte specific primary antibodies provided by the user To run a Duolink Il assay you need the following Duolink Il components b PLA probe MINUS Wash Buffers A and B PLA probe PLUS Mounting Medium with DAPI P Detection Reagents 4 1 1 DUOLINK II PLA PROBES The choice of PLA probes depends on the species of your primary antibodies and your application see section 3 For a complete list of PLA probes visit www olink com Each Duolink Il PLA probe product contains the following gt Blocking Solution For blocking of the sample if you have not already optimized your primary antibody with another blocking solution Supplied in ready to
20. igen retrieval Blocking solution Antibody incubation buffer v v v v v Primary antibody concentration Sample pre treatment for Duolink is identical to procedures used for immunohistochemical IHC and or immunofluorescence IF staining If you already have a working assay for IHC or IF use the same pre treatment protocol The primary antibody concentration may need to be titered and optimized for the Duolink conditions When using two primary antibodies optimization is best done for one primary antibody at the time using the Duolink single recognition approach These conditions can then be used as a start for optimizing your assays using both primary antibodies in the same Duolink Il assay 10 Duolink Il Fluorescence When using two primary antibodies it may in some cases be difficult to find conditions compatible with both of the selected primary antibodies Under these circumstances it is advisable to search for other antibodies recognizing alternative epitopes If you have used the previous version of the Duolink kits and are converting into Duolink II you may need to increase the primary antibody concentration to obtain a comparable number of PLA signals This increase varies depending on the application and the affinity of the primary antibodies but could typically be in the range of 1 5x to 2x The signal to background ratio for Duolink will still exceed the one for the previous kit For further advice on immunostaining optimiza
21. l experiment is typically a number of discrete fluorescent spots PLA signals in various locations of the studied cells see Figure 10 In some cases when studying highly expressed proteins the density of signals may be so high that the signals coalesce see Figure 11 A B Fig 10 Detection of EGFR in cytospin preparations of A431 cells using Duolink with one primary antibody where the primary antibody has been titrated to a very low concentration to give individual signals The pictures show a maximum intensity projection of the raw image based on 20 z planes PLA signals are shown in red and the nuclei in blue The nucleus image was acquired in one z plane A Positive reaction B Negative control without primary antibodies B Fig 11 Detection of Her2 in FFPE preparations of SKBR 3 high expression cells using Duolink with two primary antibodies In this case the antibodies have been titrated to give signals in low expression cell lines The pictures show a maximum intensity projection of the raw image based on 20 z planes PLA signals are shown in red and the nuclei in blue The nucleus image was acquired in one z plane A Positive reaction B Negative control without primary antibodies 20 Duolink Il Fluorescence 8 2 IMAGE ACQUISITION The PLA signal is recognized as a fluorescent spot see Figure 10 in a fluorescence microscope using the appropriate filters for the detection fluorophore used An individual signal is of s
22. nk Il Fluorescence 6 2 DUOLINK II WASH BUFFERS The washing steps should be performed in a staining jar with a minimum volume of 70 ml on a shaker cradle with gentle orbital shaking Washing after the primary antibodies and PLA probes should be performed in Duolink Il Wash Buffer A or in the wash buffer optimal for your primary antibodies In all subsequent wash steps ensure to use Duolink Il Wash Buffers Wash Buffer A store at room temperature To prepare a 1x buffer dissolve the content of one pouch in high purity water to a final volume of 1000 ml Store pouches at room temperature 1x solutions may be kept at room temperature for short time storage one week or less For long time storage store at 4 C Bring the solutions to room temperature before use gt Alternatively make your own Duolink ll Wash Buffer A according to appendix A Wash Buffer B store at room temperature To prepare a 1x buffer dissolve the content of one pouch in high purity water to a final volume of 1000 ml Store pouches at room temperature 1x solutions may be kept at room temperature for short time storage one week or less For long time storage store at 4 C Bring the solutions to room temperature before use Forthe final wash step prepare 0 01x Wash Buffer B by diluting 1x buffer 1 100 in high purity water gt Alternatively make your own Duolink II Wash Buffer B according to appendix A 6 3 SAMPLES Before you start the D
23. of wash solution remains on the slide before addition of amplification reagents Keep amplification time and temperature Ensure Polymerase is active i e has been kept at 20 C and that correct dilution of the reagents have been used Prepare fresh dilutions just before use do not allow mix with enzyme to stand for more than 5 minutes before use Duolink Il Fluorescence 23 PROBLEM PROBABLE CAUSE SUGGESTED SOLUTION High Incomplete If you use paraffin embedded samples incomplete removal of background retrieval protocol paraffin can cause background Use fresh solutions if necessary and general ensure that correct times during deparaffinization have been used cause Custom blocking If you have used your own blocking solution allow primary antibody solution and PLA probes to be in contact with your blocking reagent before addition to the sample Insufficient Ensure sufficient washing under gentle agitation a minimum volume washing of slides of 70 ml is recommended Use fresh washing solutions and jars Drying of sample Ensure good humidity during all incubation steps and never let slides dry out after washes and before addition of reagents Unspecific Titrate primary antibodies IHC IF or PLA with respect to binding of concentration temperature time and buffer and ensure appropriate primary fixation protocol was used antibodies Large Dust salt Discard wash solutions and use new solutions and washing
24. ondary antibodies conjugated with oligonucleotides PLA probe MINUS and PLA probe PLUS are added to the reaction and incubated The Ligation solution consisting of two oligonucleotides illustrated as red bands and Ligase is added and the oligonucleotides will hybridize to the two PLA probes and join to a closed circle if they are in close proximity The Amplification solution consisting of nucleotides not shown and fluorescently labeled oligonucleotides is added together with Polymerase The oligonucleotide arm of one of the PLA probes acts as a primer for a rolling circle amplification RCA reaction using the ligated circle as a template generating a concatemeric repeated sequence product The fluorescently labeled oligonucleotides will hybridize to the RCA product The signal is easily visible as a distinct fluorescent spot and analyzed by fluorescence microscopy Fig 1 Assay principle Duolink Il Fluorescence 5 3 Applications 3 1 DETECT AND QUANTIFY PROTEIN INTERACTIONS Detection and quantification of interacting proteins can be done using two different primary antibodies against each of the two proteins of interest 3 2 DETECT AND QUANTIFY PROTEIN MODIFICATIONS Detection and quantification of a protein and its specific post translational modification can be done using two different primary antibodies one directed against the target protein and one against a mod
25. ons 0 01 M Tris 0 15 M NaCl and 0 05 Tween 20 Filter the solution through a 0 22 um filter and store at 4 C Bring the solutions to room temperature before use DUOLINK Il WASH BUFFER B Alternative 1 use Duolink II Wash Buffer B Art no 82048 To prepare a 1x buffer dissolve the content of one pouch in high purity water to a final volume of 1000 ml Store pouches at room temperature Expiry date is marked on each individual lot 1x solutions may be kept at room temperature for short time storage one week or less For long time storage store at 4 C Bring the solutions to room temperature before use Alternative 2 make your own Duolink II Wash Buffer B Dissolve 5 84 g NaCl 4 24 g Tris base and 26 0 g Tris HCI in 500 ml high purity water Adjust pH to 75 using HCI Add high purity water to 1000 ml final concentrations 0 2 M Tris and 0 1 M NaCl Filter the solution through a 0 22 um filter and store at 4 C Bring the solutions to room temperature before use 26 Duolink Il Fluorescence References Jarvius M Paulsson J Weibrecht Leuchowius KJ Andersson AC W hlby C Gullberg M Botling J Sj blom T Markova B Ostman A Landegren U S derberg O In situ detection of phosphorylated platelet derived growth factor receptor p using a generalized proximity ligation method Molecular and Cellular Proteomics 6 1500 1509 2007 S derberg O Gullberg M Jarvius M Ridderstr le K Leuchowius KJ Jarvius J W
26. opulations Furthermore regions of interest can be defined a feature of particular relevance when studying tissue samples Duolink ImageTool supports images generated from both fluorescence and brightfield microscopy Raw imaging data can be imported directly from the four major microscope vendors Olympus Leica Nikon and Zeiss The results data can be exported into an excel sheet for further evaluation Duolink Il Fluorescence 5 Assay considerations 5 1 CHOICE OF PRIMARY ANTIBODIES The Duolink Il reagents are generic reagents using secondary antibodies for detecting presence of analyte specific primary antibodies Your choice of primary antibodies is crucial when setting up the PLA assay The primary antibodies should be of IgG class specific for the target to be detected and preferably affinity purified The primary antibodies could be either polyclonal or monoclonal To maximize your success rate choose antibodies that are IHC and or IF classified and follow the optimization guidelines given in section 5 2 Primary antibodies could also be qualified individually in a Duolink single recognition experiment section 5 1 1 before a double recognition assay 5 1 1 PRIMARY ANTIBODIES FOR DETECTION OF PROTEIN EXPRESSION When single protein targets are to be detected you can use either one or two primary antibodies against your target A One primary antibody single recognition Usage of one primary antibody gives you an easy
27. phore in wavelength o can be detec The fluorophore in wavelength o can be detec The fluorophore in wavelength o can be detec The fluorophore in wavelength o can be detec ed using 554 nm ed using 598 nm ed using 644 nm ed using he Amplification Green 501 nm and an emission wavelength of 523 nm and he same filters as he Amplification Orang and an emission wavelength of 576 nm and he same filters as he same filters as he same filters as Make sure that you use appropriate filters has an excitation or e g Cy2 or FITC e has an excitation or e g Cy3 he Amplification Red has an excitation and an emission wavelength of 634 nm and or e g Texas Red he Amplification Far Red has an excitation and an emission wavelength of 669 nm and or e g Cy5 Inefficient ligation Keep ligation incubation time and tempera excessive amount of wash solution remains on slide before addition ure Ensure that no of ligation reagents Ensure that the Ligase is active i e has been kept at 20 C and that correct dilution of the reagents have been used Prepare fresh dilutions just before use do not allow mix with enzyme to stand for more than 5 minutes before use Inefficient amplification If the signals are very weak they can appear to be few as only a fraction reaches above detection threshold Ensure that no excessive amount
28. rcle reaction area spreads over delimitation a too large around reaction surface area is unable to cover Inappropriate Adjust reaction volume according to your reaction area For guidance sample reaction volume on reaction volumes please refer to section Reagent Preparation or the separate Reaction Volume Guide No nuclear Mounting Ensure using Duolink Il Mounting Medium with DAPI staining is without observed mounting media containing DAPI Wrong filter used Nuclear staining is performed with DAPI excitation 360 nm and for acquisition emission 460 nm Correct filters must be used If problems remain please contact us at support olink com or 46 18 444 3970 Duolink Il Fluorescence 25 Appendix A Duolink Il Wash Buffers formulas DUOLINK II WASH BUFFER A Alternative 1 use Duolink II Wash Buffer A Art no 82047 To prepare a 1x buffer dissolve the content of one pouch in high purity water to a final volume of 1000 ml Store pouches at room temperature Expiry date is marked on each individual lot 1x solutions may be kept at room temperature for short time storage one week or less For long time storage store at 4 C Bring the solutions to room temperature before use Alternative 2 make your own Duolink II Wash Buffer A Dissolve 8 8 g NaCl 1 2 g Tris base and 0 5 ml Tween 20 in 800 ml high purity water Adjust pH to 74 using HCI Add high purity water to 1000 ml final concentrati
29. s compatible with all fixations typically used for IHC including Ethanol Zinc b Acetone gt Formalin gt Paraformaldehyde PFA Duolink ll Fluorescence 11 12 5 4 2 ANTIGEN EPITOPE RETRIEVAL Antigen epitope retrieval is mostly used for formalin fixed paraffin embedded FFPE material Formalin fixation is masking the epitopes which then can be un masked by antigen epitope retrieval Different methods for antigen epitope retrieval will influence the performance of your primary antibodies and must be optimized for your application Duolink is compatible with all antigen retrieval methods typically used for IHC including Heat Induced Epitope Retrieval HIER b Enzyme treatment with Proteinase K Pepsin Trypsin etc For HIER it might be necessary to try a few different solutions of various pH 5 4 3 BLOCKING SOLUTION AND ANTIBODY INCUBATION BUFFER It is essential to use a proper blocking solution and antibody incubation buffer Use the recommendations from the vendor of your primary antibodies if available If you have previously optimized your assay in for example IHC use the same conditions for Duolink There are two different protocols depending on whether you use your own blocking solution and antibody diluent or the Duolink Il Blocking Solution and Antibody Diluent PLA probe protocol CUSTOM solutions gt Use this protocol if you use your own blocking and antibody buffer solution recommended see section 7
30. tion we recommend the Education Guide from Dako http pri dako com 08002 25may06 ihc guide book pdf 5 3 SAMPLE TYPE Duolink ll reagents can be used with a broad range of sample types depending on the preferences of your primary antibodies Duolink has so far shown to be successful with the following sample types b Adherent cell cultures gt Formalin fixed paraffin embedded Cytospin preparations Ste SODUIODS b Formalin fixed paraffin embedded Frozen tissue sections cell samples b Fresh tissue sections 5 4 PRE TREATMENT Before you start your samples should be deposited on glass slides and sufficiently pre treated with respect to fixation retrieval and permeabilization It is crucial for the performance of the assay to optimize the conditions for the primary antibodies Also the choice of wash buffer may be dependent on your primary antibodies Use the recommendations from the vendor of your primary antibodies if available 5 4 1 FIXATION Different fixations will influence the antigenicity and thus the performance of your primary antibodies If there is no fixation recommendation for your antibody you must optimize this for your application For cell materials like cells grown on slides or cytospin preparations you should choose the fixation most suitable for your primary antibody The best choice of fixation may vary depending on your cell material antibody and the type of protein you want to study Duolink i
31. ub micrometer size For images taken in one focal plane several signals can be either above or below the current focus A true PLA signal is easy to go through by changing the focus making it appear and disappear This does not apply to coalesced signals which may occur for highly expressed proteins see below To detect all PLA signals it is thus necessary to obtain images throughout the entire thickness of the sample However you may acquire images in one plane as long as all images to be compared are acquired in a similar position within the sample As with all fluorescent imaging it is important to keep settings constant during an experiment with regard to exposure time and filters used etc It is recommended to get a feeling for the setting by use of one acquisition time needed for your specific microscope setting by use of one positive slide and one negative slide For proper positive and negative controls please refer to section 5 5 Controls on page 13 It is important to remember that a negative slide without primary antibodies but with PLA probes is likely to contain a few signals one or two signals every tenth cell and therefore the image from the negative reaction should not necessarily be completely blank The same setting should be used for all the images If the number of PLA signals is large it could happen that they coalesce all over the image or in certain cells Care must then be taken to choose for how to set the acquisition time
32. uolink Il protocol the sample should have been deposited on a glass slide sufficiently pre treated to fit your primary antibodies with respect to fixation retrieval and permeabilization see section 5 2 Your reaction area must be delimited with e g a grease pen or silicon mask 6 4 REACTION VOLUME Table 1 Suitable reaction volume for different reaction areas Use open droplet reactions a E P AREA TOTAL REACTION VOLUME b Perform the incubations without a cover slip 0 2 cm 15 ul b Perform all incubations in a pre heated Um mm humidity chamber Sul la m 2cm 80 ul Use volumes corresponding to your delimited 3 cm 120 ul reaction area Never use less than 15 ul of total reaction volume see Table 1 or Reaction 4 cm 160 ul Volume Guide Art no 80520 6 em 240 ul Note t is important that all incubations are performed in a humid environment to prevent 8 cm 320 yl excessive evaporation If the sample goes dry this 10 cm 400 ul will give rise to severe artifacts Duolink Fluorescence 15 7 Assay protocol 7 1 PLA PROBE PROTOCOL CUSTOM SOLUTIONS Use this protocol if you use your own blocking solution and antibody buffer that you know work with your primary antibodies Before you start your samples should be deposited on glass slides and pre treated with respect to fixation retrieval and or permeabilization Use open droplet reactions without a cover slip and perform all incubations in a humidity
33. use concentration gt Antibody Diluent For dilution of PLA probes and in some cases the primary antibodies section 5 4 3 Supplied in ready to use concentration PLA probe 5x stock Secondary antibody conjugated with a PLA oligonucleotide 4 1 2 DUOLINK II DETECTION REAGENTS Duolink II Detection Reagents are available with four different fluorophores DETECTION REAGENTS EXCITATION nm EMISSION nm FILTERS FOR IMAGE ACQUISITION Detection Reagents Green 501 523 same filters as e g Cy2 or FITC Detection Reagents Orange 554 576 same filters as e g Cy3 Detection Reagents Red 598 634 same filters as e g Texas Red Detection Reagents Far Red 644 669 same filters as e g Cy5 Each Duolink II Detection Reagents box contains the following gt Ligation bx Contains oligonucleotides that hybridize to the PLA probes and all components needed for ligation except the Ligase Ligase 1 U l gt Amplification bx Contains all components needed for Rolling Circle Amplification except the Polymerase Included are also oligonucleotide probes labeled with a fluorophore that hybridize to the RCA product b Polymerase 10 U ul Duolink Il Fluorescence i 4 1 3 DUOLINK II MOUNTING MEDIUM WITH DAPI Optimal mounting medium for preservation of the PLA signals It also contains DAPI nuclear stain excitation 360 nm and emission 460 nm Aqueous mountant which does not solidify 4 1 4 DUOLINK II WASH BUFFERS Duolink II Was

Download Pdf Manuals

image

Related Search

Related Contents

Bedienungsanleitung (PDF Format)    APS-U User Manual - Higgs Hydrographic Tek  Freedom 6000 - User Manual  HB400 User Manual revision 1 - North Carolina Instruments, LLC  ApeosPort-IV C5570/C4470/C3370/C2270  IMPLEMENTACION DE UN PROTOTIPO PARA CONTROL DE  SIW 22T-A    HP Pavilion dv4-2045dx User's Manual  