Home
        mDSLM User Manual 01
         Contents
1.  all the time  but the reading is only visible in the window  for the electronics box and in the log file     4    Transmission light      Use 760 nm position     Donotuse a filter     Intensity can only be controlled by changing the exposure time  laser  power does not work for transmission    6 position filter wheel    Semrock filters  not necessarily available in their catalogue but the Stelzer   group has a contact person whose details we should get      One position has to stay empty  the others contain bandpass or longpass  filters      Filter wheel can be turned manually when it is off but never try that when it  is on      Ludl Electronics filter wheel controller with green on off button     With time filters will become loose and start to rattle and need to be  tightened up again      The filter wheel must not touch the microscope to avoid vibration  If the  filter wheel does touch the microscope casing you will get vibration visible  by the generation of sinus waves on screen and the filter wheel position  will have to be adjusted     The air table needs to be well aligned and floating     5    Camera    Andor Clara  cooled CCD camera  With higher magnification objectives  camera vibration becomes noticeable  To avoid this use the software option  that allows turning camera cooling off during acquisition  Auto on off   There  are two holes at the back of the microscope casing  which contain screws   The one closest to the camera adjusts the camera angle  the other o
2.  to get beads in focus and adjust y height as well as y center  can    adjust Y height and center using beads        Use the z stage control to move the light sheet to the edge of the agarose  block that is closest to the detection objective      Set the correct z spacing     Pixel size in x and y x 4  objective magnification   6 45 um x 4  10   2 58 um z spacing    6 45 um x 4  20   1 29    6 45 um x 4  40   0 645    6 45 um x 4  63   0 402     Set Start plane  then set 300 planes by moving the z stage control until you  get ca  300 planes  By clicking on Set End Plane the number of planes is  updated      Set timelapse  long total duration  1h  interval  30 s      Untick Save next to Start Acquisition  tick z stack and time lapse and click  Start Acquisition     Use the detection Pifoc to focus beads  strongest signal  no rings        11     Use the illumination Pifoc to move the focussed beads into the middle of  the sheet   Change the laser line and align again        How to centre the stage   This needs to be done when a new chamber is fitted to the microscope  On  delivery the stage is centred     12           Remove the chamber and unscrew the basin  put the basin aside  Unscrew the head of the stage  Fit the head of the stage to the chamber in the least wobbly position   Make sure that the specimen holder keeps its correct position   Put the chamber back in with the head of the stage attached  The pin attached to the head of the stage will now not be centred to fit  t
3. Pifoc control    If not already correct  move the light sheet centre along its axis to the  centre of the camera image using the scanner offset control  top panel   and adjust the light sheet height to span the whole image or a smaller  ROI    Remove the alignment mirror using the reverse procedure of fitting it   Alignment is finished    Specimen can be inserted                             Camera view of the sample       left    top bottom    right             9    Alignment procedure for several laser lines using the mirror     Use the mirror only if you are not successful using beads for alignment  The   488 nm laser line is now too strong to be used with the mirror   The mirror is   used to correct chromatic aberration between different laser lines  Do the   alignment below using the mirror before the bead alignment if you want to  image several channels      Click on Preview  Insert the mirror with y 13 000 and z  1 500  turn to 45    using tweezers  adjust y to right height with tweezers if not inserted into the  specimen holder all the way  Then move to z 0      Adjust x with the x stage controls until the laser beam is visible  laser line of  your choice  integration time  50 ms  no filter   adjust dynamic range  bits   in between  With the 40x objective only one of the two reflections will be  visible  Move the x stage until you see the higher beam and concentrate  on that  If the beam is not perfectly horizontal it does not matter and is  caused by the mirror not be
4. e the specimen holder to the front  right end  z  1 500 um and x 1 500 um  of the chamber using the  z and   x stage control  Insert the alignment mirror  using the provided forceps   so that its surface normal  its reflective surface is the one that intersects the  centre of the cylindrical rod it is glued to  roughly aligns with the detection  axis  mirror surface faces the detection objective   Adjust the insertion  height so that the mirror will  when back in the centre position  intersect  the illumination beam  centre line of the illumination objective   Move the  stage back to z 0 using the  z stage control              Turn on the Preview mode with acquisition parameters set to    Laser line of your choice   Power  0    Integration time  50ms   No filter   Now rotate the stage  carrying the specimen holder and the alignment  mirror  by 45   using the    stage control so that the illumination light is  intersected and directed towards the detection objective roughly parallel  with the detection axis  if necessary you can also use the x or z stage  controls to bring the mirror into that position   If not already visible   increase the power until you see a dim light sheet or light soot on the  camera image    10x objective  Use the x stage controls to bring the light sheet into the  centre of the camera image  Set the step size for the    control to 1   and  using the     and     stage controls find the angular acceptance  limits range of the detection lens and rotate t
5. he base   Using the software move the stage in x and z and later also in y to centre it  to the pin   Adjust the pin in such a way that it does not touch the sides of the hole   The stage connects with the table at y   10 000  Move to  10 000 and  tighten screw   To access the microscope controller open http   192 168 1 2  in Internet  Explorer  Remote file browser  button on the left  Open folder  Tara Settings  Stage Settings  Use Copy and Paste buttons to create a copy of the stage settings  Double click on the file to save it on the desktop and give it the same  name as before  Open file in Excel if installed  Home position is the important one   old value   movement   new value for the x and y axes  7800 1100 8900  Save the file and use download upload button to move the file to the  folder  Could have used the Edit button for all this instead  Restart realtime computer by pressing reset button on realtime controller  after exiting the acquisition software  Console button shows what the realtime computer is doing after restarting  Internet Explorer  Restart acquisition software  Unscrew head of stage again and re attach it to the table to check if it is  centred  Move the stage up to  10 000 and tighten the screw             Specimen  Multi Stage Position    Specimen 1  Define new specimen   Sync channels  sync basic setup  this will copy stack size and position as  well as spacing  or just fill in spacing         In specimen 1 and then in specimen 2 move to correct posi
6. he mirror to the centre of this  range    Higher magnification objectives  If the above is not possible  i e  light sheet  moves out of the field of view  use detection Pifoc and    stage controls to  align the rotation stage  When moving the Pifoc the light sheet moves out  of focus but also changes position if the mirror angle is not at 45    Use the     stage control to rotate the stage by 1    either direction is fine  and  check by moving the Pifoc over the full range whether the shift has  decreased or increased  If it has increased change the direction of the     step  otherwise continue in the same direction with another    step   Successive steps  minimising the position shift of the light sheet  bring the  mirror angle closer and closer to the 45   position    The mirror now intersects the illumination and the detection axes by 451     Use the detection Pifoc control to bring the mirror surface  not the light  sheet  into focus  which  when achieved  is apparent by well resolved   always present  dust particles on the surface of the mirror    If this is not possible  i e  the Pifoc range is not large enough  use the  scanner offset control together with the z stage control  use small step  sizes  e g  1Oum  to bring the surface of the mirror within the range of the  Pifoc while keeping the light sheet position centred on the camera range   Minimize the thickness of the light sheet visible on the mirror  i e  in the  centre of the camera image using the illumination 
7. ing perfectly vertical  Do not infer that the  camera angle is wrong      Lower laser intensity to O      Use y height to reduce lines to dots  y height changes the height of the  light sheet  the higher the light sheet the shorter the time per point  which  decreases the laser intensity and increases the length of the lines on  screen       Use y center to change left right positioning of the dots to the centre of  the screen  Y center moves the whole light sheet up or down      Z position moves the lines up or down  Do not use      Now use y height to increase the length of the lines on screen so that the  ends are outside the field of view and the bright end is not visible  anymore      Concentrate on the upper line  When moving the detection Pifoc a long  way the light sheet may move in x  If this happens the mirror is not at  precisely 45    Change   in 1   steps and move the detection Pifoc until the  light sheet stops moving in x      Use the detection Pifoc to focus on dirt on the upper light sheet      Activate the cross hairs and move the upper light sheet to the centre  using the x stage control                                  Use the illumination Pifoc to find the position where the light sheet is  slimmest      Now adjust the second channel by choosing the relevant laser line  no  filter and laser power 0      The upper light sheet of the first channel has to be in the centre of the  cross hairs  see above       lf the laser power in the second channel is too st
8. ion  tick Time Lapse  tick Save  Start Acquisition                Alternatively  use Next Specimen for all the different angles   Rotation works only if the needle for compensation  the capillary and the  agarose are absolutely vertical and if there is no wobble  One should mount  samples in micropipettes for the rotation        14   Structured Illumination  This generates 3 images per plane  phases are moved by 120       Structured illumination  Scanning Properties  Laser Power Modulation    Current SI Frequency  modulates laser on off  depends on the sample  and on how deep you are inside the sample  30 60 are good numbers    15   Preparation of Tetraspeck beads for mDSLM    objective lens final conc   beads mL  bead stock 150 uL    10x NA 0 3 3 15E 07 5 19 ul 1 um beads  20x NA 0 5 3 45E 08 56 85 uL 1 um beads  40x NA 0 75 6 29E 09 40 95 uL 200 nm beads  63x NA 1 0 2 77E 10 180 uL 200 nm beads    Set the heatblock to 70   C  Prepare 2  low melting agarose in 2x the buffer  you are going to image your samples in  Vortex your bead stock thoroughly  and make up your beads in H2O dest  to a total volume of 75 uL   For the 63x  lens the beads need to be dried down in a speed vac before use   Warm the  diluted beads up in the heat block  Add 75 uL of the low melting agarose and  vortex thoroughly  Put back into the heat block and draw the liquid up into  the capillary using the Teflon plug  Hold capillary against ice until the agarose  is set        TetraSpeckTM Fluorescent Micr
9. mDSLM Manual    Start up procedure   1  Switch on computer    2  Change objectives before switching on microscope  1   2      3  If you are a using a different chamber from the previous user the stage  must be realigned by a member of staff before you continue  11      4  Screw chamber in  fill with appropriate medium  switch microscope on  and start acquisition software  6   7    Check for leaks    5  If you are using more than one laser line you need to align the different  laser lines  3   4   9    If you are using one laser line move on to the next  item    6  Align the appropriate beads  10      7  Calibrate your rotation compensation tool if you want record multiple  angles of your specimen  13      8  Insert your sample           Shut down procedure  1  Set    to P 0  set x stage control to x 0  set z stage control to z 0  set y  stage control to y 0  Remove sample  Exit acquisition software   2  Remove medium and unscrew chamber  Clean chamber and tools to  prevent rust   Switch microscope off before removing any objective   Unscrew detection objective and clean it   5  Shut the computer down at the end of the day        Rw    1    llumination objectives and detection objectives    2 5x  NA 0 06  illumination objective generates more homogeneous light for   10x  NA 0 3  and 20x  NA 0 5  detection objectives   If you use the 5x   illumination objective for the 20x detection objective the light sheet will be   slimmer but only over a short distance in the middle and onl
10. ne  adjusts the tube length        How to check the camera angle on assembly      Special metal tool inserted into probe slot at 45   angle just like the mirror    Use transmission light  760 nm  150 ms exposure  no filter     Focus on the right  pointed  edge of the tool using the z control            Draw an ROI around the edge of the blade or use the cross hairs     lf the blade does not look horizontal on screen loosen screw  1 3 mm  in  hole at back of casing closest to the camera  which fixes the camera in  place      Slowly rotate the camera until the blade in the field of view is perfectly  horizontal and fasten screw      When you move the blade in  y in 100 um steps it must not move in x        The stage movers   The cables that connect the stage movers  x y z  must not hang down and  pull  Ensure that the heavy connectors are fixed to the table     Galvo mirror    1 screw for alignment  no alignment should be necessary  If there is a problem  contact Alexander Atzberger    alexander atzberger physikalischebiologie de           DIA illumination    Transmission light with power switch set to maximum and therefore fixed  intensity  can only be controlled by changing the exposure time and  changing the camera bits     The chamber      Finger gloves are available from pharmacies  Latex without powder  size  M       Coverslips  300 um thickness  18 mm diameter  glued in with nail varnish  from the outside     Grease for the chamber  Glisseal N  Swiss  grease for laborato
11. osphere Standards  Life Technologies   T7280  0 2 um diameter   T7282  1 0 um diameter    Samples are mounted in low melting agarose at 0 8     1 2   Cool the agarose  down  pipet it into the capillary and add the specimen    Arabidopsis seedlings    Grow seedlings vertically on 1  pytagel  Scoop seedling with pytagel into  capillary leaving the shoot exposed  Dip bottom end of the capillary in 1  low  melting agarose to lift up the root until it is exposed as well  Add carbon pin to  support root     Perfusion system  Ismatec ISM831C  Capillaries  Hilgenberg  ask Alex     
12. ries suitable  for vacuum  without silicone    6    How to change objectives    Only change objectives when the microscope is turned off      Remove water from chamber  set x stage control to x 0  set z stage  control to z 0  set y stage control to y 0      Unscrew the chamber      Change objectives      Check that the finger glove is in the right position and not ruptured  It is  better to change the finger glove when changing objectives  You need a  special tool to unscrew the rings that hold it in place  Fit a new glove over  the inner ring and cut off the finger leaving ca  1 5 cm  Screw the finger  glove in so that the fat end points to the inside of the chamber  Push the  cut off end in as well      Screw the chamber back in  fill with water and check for leaks  Check  again after 15 min      Turn microscope back on and activate Preview in the acquisition  software      Set the y stage control to y 13 000  set the z stage control to z  1 500      Fit the sample capillary      Set the z stage control to z 0     7   How to start the microscope and the acquisition software   There is a red start button on the console for the microscope that also controls  the realtime computer  The acquisition software computer is separate  When  the microscope crashes press Reset button on the realtime computer   mDSLM LightMicroscopy Group EMBL Heidelberg    Acquisition software Computer sign in  your Biochemistry login  password  your  Biochemistry password   embl  password  mds m   sof
13. rong even when using the  full dynamic range  14 bit  increase y height temporarily to the highest  setting      Enable both channels and use light sheet position offset to perfectly  superimpose both  least amount of wobble  match both intensities as well  as possible   It is always best to adjust the outer    to the position of the 561  nm laser line      If you want to adjust three channels set the light sheet position offset of  the middle wavelength  561 nm  to 0      Move the top line of the light sheet of the 561 nm laser into the centre of  the crosshairs using the x stage control      Switch channel and move the second laser to the same position using  light sheet position offset followed by the third              10    mDSLM alignment using beads    Perform a bead alignment every time you change objectives  If you are not   changing objectives perform the alignment once a day  Use Tetraspec beads   in 1  low melting agarose    um beads for the 10x and 20x objectives and   200 nm beads for the 40x and 63x objectives      The default stage settings before adding the beads are y 0  x 0  z 0     0      Click on preview to activate stage settings and laser  Move the y stage  control to y 13 000  then move the z stage control to z  1 500      Add the beads and move the z stage control to z 0      Use any laser line  laser power  gt 0  integration time  50 ms  appropriate  emission filter      For coarse adjustments go to Settings under Scan Contoller  Use detection  Pifoc
14. tion  set start  plane set end plane  start at surface and end further into the sample   moving in the other direction is possible       Set time lapse  this is valid for both channels       Click Save and decide where to save     Set Position Here  button without function     Start acquisition     When finished open file in Explorer     Open Fiji     Highlight file in Explorer and drop into Fiji     Stacks   Manipulation  Concatenate     When you have your concatenated stack    Stacks  Hyperstack  Stack to Hyperstack     Forz projection over time    Image  Stacks  z project    13    Rotation   There are two tools for rotation compensation  a metal needle for the 10x and   20x objectives and a glass needle for the  20x and  40x objective  To make   the glass needle break a bit off a Femto tip and glue it on to the holder for   thin capillaries      Fit the needle into the specimen holder  760 nm  150 ms exposure  filter  empty  and centre it using the crosshairs      Focus on the needle and then fit an ROI to the tip of the needle      Advanced  in top menu    Manual Calibration  follow instructions and when finished press Finish     Remove the needle and insert your sample  focus on it and set up a stack  that starts way outside the specimen and ends way outside the specimen      Rotation  start angle 0    end angle 315    angular step 45    steps 8   alternatively  start angle 0    end angle 330    angular step 30        Possibly set camera gain     Tick Z stack  tick Rotat
15. tware  Lsfm Contr 4 9 1  Recorded Data  Set up network path  Copy only  copies data from computer to network without deleting   Save  in folder on computer   File name  saves on computer  File structure   Multiplane  1 Tiff per z stack and channel and time point  Hierarchical  1 folder for timepoint containing 1 folder per  specimen  Flat  1 folder and every file is in there  1 file for each plane and  each channel  Format  Tiff 16 regular  can contain up to 4 GB  Big Tiff  can hold bigger files  for sSCMOS cameras only    Scan controller  settings can be saved and loaded  do not touch Gamma  and Angle    The minimum exposure time is 50 ms because the scanner moves too  slowly for less     We will get software upgrades and will be notified by email beforehand           8    mDSLM alignment procedure using the mirror    The chamber must be filled with water to above detection lens level      Click enter after typing each number to set scan  and Pifoc parameters  to     Lightsheet height  1000  Illumination Pifoc  50   Detection Pifoc  50   x centre and y centre  0      Move stage up  using the  y stage control button  x z and     angle  must  be in their default position i e  at zero  until it couples fully to the specimen  holder  at ca  y 10 000 um  and then lift the specimen holder  by ca   1 500 um  again using the  y stage control to a point where it is positioned  well above its parking position and where it can now be freely moved in  the horizontal  x z  plane  Now mov
16. y slim objects can   be imaged with this combination     5x  NA 0 16  illumination objective generates more homogeneous light for   40x  NA 0 75  and 63x  not currently available  detection objectives      Only change objectives when the microscope is switched off and the  chamber has been taken off      Every time you change the detection objective change the finger glove  or at least check it has not acquired any holes                 How to fit the laser cable on assembly    The cable has two discernable ends  The microscope end is just pushed in  To  fit the end at the laser unit press button to push end in  but do not move the  adjustable screws  which have been adjusted and fixed in the factory     2    Objective Pifocs    The Pifocs vibrate when no objective is mounted but will be still when the  objective is on  Always switch the Pifocs off before changing objectives  In  order to do that switch the whole microscope off because the Pifoc power  switch is hidden     3    Laser lines    488 nm diode laser   561 nm DPSS solid state laser  AOM    638 nm diode laser   The lasers are not quite stable during the first hour of operation  Switch on the  electronics with the lasers and wait 20 min before starting an experiment   Even at 0   power the lasers emit a bit of light  They can be unclicked and so  turned off in the software at random  but it is best to leave the solid state laser  on  otherwise you have to wait 10 min before imaging  A photodiode  measures laser power
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
取扱説明書 液晶カラーモニター  English  John Deere TY25862 User's Manual  GR-ADK II - グレープシステム  DENTAL, AEROPULIDOR    Divar 2 - Digital Versatile Recorder  Untitled - セイコーウオッチ  Foundry Networks AR1204 User's Manual      Copyright © All rights reserved. 
   Failed to retrieve file