Service Manual Perfusor Space

image

Contents

1. every pressure stage Motor power limitation Pressure stage 1 11 28N N Pressure stage 3 30 49N N Pressure stage 6 58 80N N O Error message Alarm drive blocked at every pressure stage M688 70 01 05 F04 3891 7319 GB B BRAUN ak SHARING EXPERTISE Perfusor Space 3 0 EN Technical Safety Check TSC Index g Master to be added to the documentation Mechanical Aids and Measuring Accessories Used Equipment Used O Syringe gauge serial No O Power Supply SP Calibrated until O Battery module O Service connector SP O Staff call lead 0 I Space PCA Kit PCA button g O Syringe Anti Removal Cap PSP g M688 70 01 05 F04 3891 7319 GB B BRAUN Sheet 3 of 3 SHARING EXPERTISE EI Perfusor Space 3 0 5 3 Functional Inspection Perfusor Space Procedural Instructions on the TSC Mechanical Inspection 1 Fit the unit to be tested on top of another Space device and check the proper functioning of the lock Fit the unit to be tested under another Space device and check the proper functioning of the lock Hold the device with the operating unit downwards The op erating unit must not open Check the battery compartment cover lock for proper opera tion Functional Check 1 Determine age of battery Open battery compartment cover and remove battery Read the batch number printed on the type plate of the battery and de
2. Perfusor Space Service Manual Version 3 0 English This document is not a complete Service Manual It only encompasses the pages that describe how to determine the age and how to check the condition of batteries as well as the accordingly adapted template for Technical Safety Check B BRAUN SHARING EXPERTISE 3 2 Battery Module Fig 3 8 Legend of fig 3 8 ItemDesignation 1 Battery compartment cover 2 Battery compartment cover lock EI Perfusor Space 3 0 Disassembly Assembly Designation Ord No Battery compartment cover PSP cpl 3452 0872 Battery module SP NiMH without fixing Pin 2 222222 eee ee eee eee 3452 0856 with fixing Pin s4s rare 3452 0856A Battery module SP Li lon with WLAN and fixing pin 0 8713182A Note Users have been informed as follows If a Perfusor Space pump used on battery only and showing full battery charge stops during syringe change and alarms users should remove the pump from use and return it to the Medical Engineering Department for replacement of the battery In this case the battery has to be replaced regardless of its age or any further test Disassembly Note Move the drive head to the extended end position before starting disassembly see General pg 3 1 1 Press the lock Fig 3 8 Item 2 on the battery compart ment cover Fig 3 8 Item 1 and remove the battery compartment
3. cover Note The battery compartment cover of more recent devices is shorter and can only be unlocked using a pointed object Technical Safety Check TSC Index g Master to be added to the documentation A Observe the Service Manual and the instructions for use All measured values are to be documented Checklist for Technical Safety Check Every 24 Months Unit Perfusor Space Manufacturer B Braun Melsungen AG Accessories used should be included in testing Make exclusive use of calibrated measuring equip ment Visual Inspection Electrical Safety Functional Inspection According to IEC EN 60601 1 or VDE 0750 and VDE 0751 U Perfusor Space The patient and housing leakage current of Locking with second unit Cleanliness completeness damage and the Perfusor Space is caused exclusively by Operating unit magnets defects affecting safety damage and the operating voltage supply Power Supply O Battery compartment cover er SP or SpaceStation readability of the labels Particularly The Technical Safety Checks of the power Determine age of battery O Syringe holder with blade supply SP drawing No M001321005 F04 J If older than 4 years O Syringe fastening or of the SpaceStation drawing No recommend replacement of battery Membrane in drive head M690000046 F04 serve to check whether or perform battery check O Axial clearance of drive both limit values are met refer to Procedural Inst
4. play on the LC display Indicator lamps LEDs yellow green blue light up for a short moment LED changes from green to red Changed order for units from SW J on LEDs yellow red blue light up for a short moment LED changes from red to green Audible alarm A deep and a high sound Status display Battery capacity mains operation service connector SP Lighting Illumination of syringe compartment LC display and but tons Syringe fastening a Axial fastening holds syringe wing b Syringe holder blade fixes syringe piston c Both claws in drive head catch the syringe piston plate Check syringe recognition a Insert approved 2 ml 3 ml syringe The syringe size is recognized b Insert approved 50 ml 3 ml syringe The syringe size is recognized Carry out infusion and bolus with any syringe and press all buttons at least once Infusion and bolus are performed and all buttons trigger the function desired Perfusor Space 3 0 EN EI Perfusor Space 3 0 Procedural Instructions on the TSC Trigger bolus by pressing the PCA button For this purpose select a PCA drug from the database fit the PCA button to P2 and trigger bolus Confirm that the green LED on the PCA button lights up Bo lus is delivered Staff call a Plug service connector SP on connector P2 b Open syringe holder while the infusion is administered The red LED on the service connector SP lights up 9 PCA lock on the Perfuso
5. r Space a Check the PCA lock with an approved 50 60ml syringe 10 PCA lock of the Syringe Anti Removal Kit PSP a Place on the drive head and lock b Check the Syringe Anti Removal Kit PSP with an ap proved 50 60ml syringe c Unlock and remove from the drive head Battery Alarm 1 Switch device off 2 Pull off the power supply 3 Switch on unit Self test is carried out 4 Open battery compartment cover during operation An alarm is activated 5 Remove battery An alarm sounds with the piezo buzzer for at least 3 minutes Revision Documentation Version 3 0 Changed TSC of the unit Revised procedural instructions on the TSC Release date 25 10 2013 Perfusor Space 3 0 EN B Braun Melsungen AG Carl Braun Strasse 1 34209 Melsungen Germany Tel 49 0 5661 71 0 Fax 49 0 5661 71 4567 www bbraun de Mat No M688_001152 Doc No 8713 9020 Version 3 0 Effective 25 10 2013
6. ructions O Screw covers on the TSC O Connectors P2 and P3 O Protection cap at drive head Switch on unit with power supply O Self test Accessories Indicator lamps LEDs Cleanliness completeness damage and O Audible alarm faults affecting safety damage and O Visual alarm readability of the label O Status display O Check the unit and the accessories for O Lighting compatibility Battery alarm M688 70 01 05 F04 3891 7319 GB EI Perfusor Space 3 0 B BRAUN SHARING EXPERTISE Sheet 1 of 3 Technical Safety Check TSC Index g Master to be added to the documentation Visual Inspection Electrical Safety According to IEC EN 60601 1 or VDE 0750 and VDE 0751 Functional Inspection Operation O Syringe fastening O Syringe recognition O Infusion O Buttons on the operating unit O Staff call O Trigger bolus at the device O Trigger bolus by pressing the PCA button O PCA lock bottom side of pump with inserted 50 60 ml syringe O PCA lock Syringe Anti Removal CAP PSP with inserted 50 60 ml syringe Switch on unit without power supply O Self test O Magnetic function of the battery compartment cover Pressure cut off Syringe type LehreOPS50 Delivery rate 200 ml h Strain gauge pressure measurement Pressure stage 1 9 15N N Pressure stage 3 25 33N N Pressure stage 8 63 76N N O Error message Alarm pressure too high at
7. termine the age of the battery a For NiMH batteries see Fig 6 2 b For Li Ion batteries see Fig 6 3 Example Example a 3 Produced in week 42 2008 B BRAUN Produced in week 41 2011 Space DABEN EEEn Paion manant 3 metxcven A g g e Of fire or buen hazar o Do damages battery a NiMH accu 8713180 TE 0 oak na opos r A Oooo D P1 142H2 0 S1 52 AAI AML DSI ec Em is 3 D 2003 8713162A007240A41 1 1A 187 un A A IM s E 2004 8713180147079P114220152 ill Y f F 12005 Made in Germany avec 1620A 120m E G 2006 T z P1 142H2 0 S1 52 A 4111 A H1 1 S1 87 H 2007 I 12008 iite f The digits calender week Second two digits year of production j Fi igits cal k oz Letter year of production irst two digits calender wee according to left table Fig 6 2 NiMH batteries Fig 6 3 Li lon batteries EI Perfusor Space 3 0 Note If the battery is older than 4 years recommend the replacement of the battery or perform the battery check Procedural Instructions on the TSC Battery check Use the battery maintenance mode to discharge the battery to battery pre alarm the battery status indicator displays no bar Perform 5 syringe changes with emptied syringe Replace the battery if a battery alarm occurs during the syringe changes Switch on unit with the power supply connected and check the following details Self test Dis