Home

Operating Instructions POINTRAC 31

image

Contents

1. This menu item appears only if you have selected Single point adjustment as adjustment mode Setup Adjustment mode In this menu item you can adjust at which percentage value of the max pulse rate the sensor should switch over Since in most cases the radiation is almost completely absorbed when the sensor is covered the pulse rate when the sensor is cov ered is very low The change between the two statuses is sufficiently clear Hence a percentage value of 90 for the Delta value is recom mended You select lower values for sensitive detection of material cones or buildup which cause only partial absorption of the radiation POINTRAC 31 Foundation Fieldbus 25 6 Set up with the display and adjustment module VELA Adjustment covered two point adjustment Adjustment uncovered two point adjustment Delta I x 1 00 100 00 EE This menu item appears only if you have selected under adjustment mode setup adjustment mode the Two point adjustment In this menu item you can set the min pulse rate ct s at which the sensor should switch over Fill the vessel until the POINTRAC 31 is covered You thus get the min pulse rate ct s for the covered adjustment Enter the requested pulse rate manually or let the rate be determined by POINTRAC 31 Automatic determination of the pulse rate should be given preference Adjustnent covered Adjustnent covered 0 ct s 0 Get count Act co
2. mom 8 OOOOO0O00 il Keri cL l vL SL 9L ZL 8L Lg Ta J Fig 8 Electronics and connection compartment with non Ex instruments and instruments with non intrinsically safe current output 1 Voltage supply 2 Relay output 3 Signal output FF bus 4 Signal input 4 20 mA active sensor 5 Switching input for NPN transistor 6 Switching input floating 7 Transistor output 8 Interface for sensor sensor communication MGC 9 Simulation switch 1 simulation on 10 Setting the bus address for sensor sensor communication MGC Fig 9 Adjustment and connection compartment with non Ex instruments and instruments with non intrinsically safe current output 1 Terminals for the external display and adjustment unit 2 Contact pins for the display and adjustment module or interface adapter Instruments with intrinsically safe current output You can find detailed information on the explosion protected versions Ex ia Ex d in the Ex specific safety instructions These safety MGC Multi Gauge Communication POINTRAC 31 Foundation Fieldbus 19 5 Connecting to power supply VEGA Electronics and connec tion compartment In struments with intrinsi cally safe current output Adjustment and connec tion compartment In struments with intrinsi cally safe current output instructions are part of the scope of del
3. mation of the procedure under www vega com downloads and Software You can find information about the installation in the download file Caution Instruments with approvals can be bound to certain software ver sions Therefore make sure that the approval remains effective with a software update You can find detailed information on www vega com downloads and Approvals 9 6 How to proceed in case of repair The following procedure refers only to the sensor Should a repair of the source container be necessary you can find the respective instructions in the operating instructions manual of the source con tainer You can find a repair form as well as detailed information on how to proceed under www vega com downloads and Forms and certifi cates By doing this you help us carry out the repair quickly and without hav ing to call back for needed information If a repair is necessary please proceed as follows e Print and fill out one form per instrument e Clean the instrument and pack it damage proof e Attach the completed form and if need be also a safety data sheet outside on the packaging e Please contact the agency serving you to get the address for the return shipment You can find the agency on our home page www vega com 40 POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 10 Dismounting 10 Dismounting 10 1 Dismounting steps Warning Before dismounti
4. AWG 24 0 2 1 5 mm AWG 24 14 16 Display and adjustment module Display element Measured value indication Number of digits Size of digits Adjustment elements Protection rating unassembled mounted into the housing without cover Materials Housing Inspection window Display with backlight 5 WxH 7x13mm 4 kevs IP 20 IP 40 ABS Polvester foil Integrated clock Date format Time format Time zone Ex factorv Rate deviation max Day Month Year 12 h 24 h CET 10 5 min vear Measurement electronics temerature Resolution Accuracv 1 C 1 8 F 1 C 1 8 F Voltage supply Operating voltage Sensor power supply FF bus 20 72 V DC or 20 253 V AC 50 60 Hz 9 32V DC POINTRAC 31 Foundation Fieldbus 45 11 Supplement Electrical protective measures Protection depending on housing ver IP 66 IP 67 sion Overvoltage category Protection class Approvals Instruments with approvals can have different technical data depending on the version For that reason the associated approval documents of these instruments must be carefully noted They are part of the delivery or can be downloaded under www vega com and VEGA Tools as well as under Downloads and Approvals 11 2 Supplementary information Foundation Fieldbus The following table gives you an overview of the instrument versions and the corresponding device description
5. Additional adjustments Info Start with the setup of POINTRAC 31 In the main menu point Setup the individual submenu points should be selected one after the other and provided with the correct parameters to ensure optimum adjustment of the measurement The procedure is described in the following Stick with the normal sequence of the menu items as closely as pos sible Setup In this menu item you can adjust the POINTRAC 31 to the integrated isotope in the source container For this purpose check which isotope is integrated in the source container You can find this information on the type label of the source container Isotope Cs 137 Through this selection the sensitivity of the sensor is adapted perfectly to the isotope The normal reduction of the emitter activity is hence considered through the radioactive decay The POINTRAC 31 requires this information of the automatic decay compensation This ensures an interference free measurement over the complete life time of the gamma emitter an annual recalibration is not necessary POINTRAC 31 Foundation Fieldbus 23 6 Set up with the display and adjustment module VELA Setup Application Setup Background radiation Setup Unit Setup Adjustment mode Enter the requested parameters via the appropriate keys save your settings with OK and jump to the next menu item with the ESC and the gt key Enter here the respective a
6. Low frequency potential equali sation currents are thus suppressed but the protective effect against high frequency interference signals remains The voltage supply and signal output are connected via the spring loaded terminals in the housing The connection to the display and adjustment module or to the inter face adapter is carried out via contact pins in the housing Proceed as follows The procedure applies to instruments without explosion protection 1 Unscrew the big housing cover 2 Loosen compression nut of the cable entry 3 Remove approx 10 cm 4 in of the cable mantle strip approx 1 cm 0 4 in of insulation from the ends of the individual wires 4 Insert the cable into the sensor through the cable entry Fig 7 Connection steps 4 and 5 1 Locking of the terminal blocks 5 Inserta small slotted screwdriver firmly into the rectangular lock openings of the respective connection terminal POINTRAC 31 Foundation Fieldbus 17 5 Connecting to power supply VEGA 6 Insert the wire ends into the round openings of the terminals ac cording to the wiring plan Information Solid cores as well as flexible cores with cable end sleeves are inserted directly into the terminal openings In case of flexible cores without end sleeves press the rectangular lock opening with a small screwdriver the terminal opening is freed When the screwdriver is released the terminal opening closes again 7
7. top view POINTRAC 31 lends itself well for level detection of bulk solids with low density Mount the instrument horizontally at the height of the requested switching point Mount the source container VEGASOURCE displaced by 90 in order to get the widest possible radiation angle When the sensor is covered by the medium the radiation damping is considerably stronger hence the switching point is all the more reliable If the max ambient temperature is exceeded you must take suitable measures to protect the instrument against overheating You can protect the instrument by providing a suitable insulation against the heat or mounting the instrument further away from the heat source Make sure these measures are taken into account already in the plan ning stage If you want to carry out such measures later on contact our specialists to ensure that the accuracy of the application is not impaired If these measures are not sufficient to maintain the max ambient temperature you could consider using the water cooling system we offer for POINTRAC 31 The water cooling must also be included in the calculations for the measuring point Contact our specialists regarding the dimensioning ofthe water cooling POINTRAC 31 Foundation Fieldbus 15 5 Connecting to power supply VEGA Safety instructions Voltage supply Connection cable Cable entry Cable gland 2 NPT Cable screening and grounding 5 Connecting
8. 131119 VEGA Contents 10 1 Dismounting Steps eren tren arte o ere rete oe ente e Fe ioris 41 10 2 Disposal Rn 41 11 Supplement a Ea TECH MCA Eo rice EREETEE 11 2 Supplementary information Foundation Fieldbus is BRECSEIBINIDE USENET ap IIT Safety instructions for Ex areas Please note the Ex specific safety information for installation and op eration in Ex areas These safetv instructions are part of the operating instructions manual and come with the Ex approved instruments Editing status 2013 11 11 POINTRAC 31 Foundation Fieldbus 3 1 About this document VEGA gt me 1 About this document 1 4 Function This operating instructions manual provides all the information you need for mounting connection and setup as well as important instruc tions for maintenance and fault rectification Please read this informa tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device 1 2 Target group This operating instructions manual is directed to trained specialist personnel The contents of this manual should be made available to these personnel and put into practice by them 1 3 Symbolism used Information tip note This symbol indicates helpful additional information Caution If this warning is ignored faults or malfunctions can result Warning If this warning is ignored injury to persons and or serious damage to the inst
9. Check the hold of the wires in the terminals by lightly pulling on them To loosen a line insert a small slotted screwdriver firmly into the rectangular locking opening according to the illustration 8 Connect the screen to the internal ground terminal connect the outer ground terminal to potential equalisation 9 Tighten the compression nut of the cable entry gland The seal ring must completely encircle the cable 10 Screw the housing cover back on The electrical connection is hence finished Information The terminal blocks are pluggable and can be detached from the electronics For this purpose loosen the two lateral locking levers of the terminal block with a small screwdriver When loosening the lock ing the terminal block is automatically squeezed out It must snap in place when re inserted 18 POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 VEGA 5 Connecting to power supply Electronics and con nection compartment Non Ex instruments and instruments with non intrinsically safe current output Adjustment and con nection compartment Non Ex instruments and instruments with non intrinsically safe current output b 5 2 Connection Non Ex instruments and instruments with non intrinsicallv safe current output t zi L L O N On lIGP Jo Oos el A ov oo O 0 Oe Os eL 52 m IZ 8
10. LLE L NJ 622 LY 41779 EN 131119 VEGA Check output signal Reaction after fault recti fication 24 hour service hotline 9 Diagnostics and service The following table describes possible faults that may not generate an error message Error Cause Rectification The instrument Voltage supply Check cables for breaks repair if nec signals covered missing essary without covering with the medium Operating voltage Check adapt if necessary too low The instrument l signals covered Electrical connec Check connection according to with covering with tion faulty chapter Connection steps and if the medium necessary correct according to chap ter Wiring plan Electronics de Change the switching behaviour of the fective sensors under Diagnosis Simulation If the instrument does not switch over send it in for repair Buildup on the in Remove buildup ner wall of the Check the Delta I value vessel Improve the switching threshold car ry out a double point adjustment Depending on the reason for the fault and the measures taken the steps described in chapter Setup must be carried out again or must be checked for plausibility and completeness Should these measures not be successful please call in urgent cases the VEGA service hotline under the phone no 49 1805 858550 The hotline is also available outside normal working hours seven days a week around
11. The exceptions are PIN language SIL and HART mode asic settings R fo fact Select Factory settinas iacu ga ory reset Peak val measurement settings Peak values temperature The following reset functions are available Basic settings Resetting of the parameter adjustments to default values at the time of shipment Order specific settings are deleted Default settings Resetting of the parameter adjustment like under Basic settings In addition special parameters are reset to default values Order specific settings are deleted Peak values of measured value Resetting of the parameter adjust ments in the menu item Setup to the default values of the respective instrument Order specific settings remain but are not taken over into the current parameters Peak values of temperature Resetting of the measured min and max temperatures to the actual measured value The following table shows the default values of the instrument The values apply for the application Level First of all you have to select the application Depending on the instrument version not all menu items may be available or they may be differently assigned POINTRAC 31 Foundation Fieldbus 29 6 Set up with the display and adjustment module VEGA Additional adjustments HART mode Additional adjustments Copy instrument settings Info Menu Menu item Default val
12. UNIT FF WEIGHT UNIT 24 Selected unit code for FF WEIGHT FF BELT SPEED 25 Belt speed FF BELT SPEED UNIT FF BELT SPEED 26 Selected unit code for FF BELT SPEED UNIT FF ELECTRONIC 27 Electronics temperature FF ELECTRONIC TEMPERATURE TEMPERATURE UNIT FF ELECTRONIC 28 Selected unit code for FF ELECTRONIC TEM TEMPERATURE UNIT PERATURE FF_COUNT_RATE 29 Count rate FF_COUNT_RATE_ UNIT FF_COUNT_RATE_ 30 Selected unit code for FF_COUNT_RATE UNIT DEVICE_TAG 31 Tagname kl DEVICE NAME 32 Device tvpe kl DEVICE STATE 33 Error code PEAK MEAS VAL MIN 34 Pulse rate min PEAK MEAS VAL 35 Pulse rate max MAX PEAK TEMP VAL MIN 36 Electronics temperature min PEAK TEMP VAL 37 Electronics temperature max MAX APPLICATION TYPE 38 Selected application SEL TEMP COMP UNIT 39 Selected unit code for process temperature DELTA I 40 Calculated percent delta I GAUGE TEMPERA 41 Electronics temperature TURE DECAY COMPENSA 42 Factor for the decay compensation TION FACTOR PMT VOLTAGE CALI 43 Photomultiplier voltage on delivery BRATION CORRELATION CO 44 Correlation coefficient for linearizer table EFF ERROR TEXT 45 Error text PMT VOLTAGE AC 46 Current photomultiplier voltage TUAL STANDARDIZATION 47 Factor for the real value correction FACTOR SERIALNUMBER 48 Serial number 48 POINTRAC 31 Foundation Fieldbus 6L LLE L NJ 622 LY 41779 E
13. accordingly Overfill protection the relay will deenergise safe condition when the max level is reached Dry run protection the relay will deenergise safe condition when the min level is reached Make sure that you have selected the correct characteristics See menu item Setup Current output mode Mode of operation mAOverfill protection Overfill prot hd Dru runnina protection With this menu item you safeguard the sensor parameters against unauthorized or unintentional modifications This menu item is described in the operating instructions manual Display and adjustment module Display With this parameter you can change the display language This parameter is described in the operating instructions manual Display and adjustment module With this parameter you can change the indication of the display You can choose if the display should show the actual pulse rate or the electronics temperature Displaued value v Count rate Electronics temp Diagnostics In this menu item you can enquire the status of your sensor In normal operation the sensor displays the message OK In case of fault you will find the corresponding fault code here This parameter is described in the operating instructions manual Display and adjustment module The peak value function holds the max and min values during opera tion POINTRAC 31 Foundation Fieldbus 27 6 Set up with the display and
14. cover You can simulate different values Simulation Simulation Start simulation Relay Pulse rate of the sensor Simulation running Count rate Count rate 124 ct s Switching function of the relay Simulation running Relay Closed 7 Simulation running Relay Open vE 28 POINTRAC 31 Foundation Fieldbus 6L LLE L NJ 622 Ly 41779 EN 131119 6 Set up with the display and adjustment module e i Diagnosis Calculated damping Additional settings PIN Additional adjustments Date Time Additional adjustments Reset Information 10 minutes after the key was pressed for the last time the simulation is interrupted automatically You can also interrupt the simulation with the switch on the electronics module The sensor calculates a suitable integration time automatically Calculated integration time 2s Additional adjustments In this menu item the PIN is permanently activated deactivated Thus you protect the sensor data against unauthorized access and unin tended changes The default setting of the PIN is 0000 This parameter is described in the operating instructions manual Display and adjustment module In this menu item you can set the actual date and time This parameter is described in the operating instructions manual Display and adjustment module When a reset is carried out all settings with only a few exceptions are reset
15. immediately at receipt Ascertained transit damage or con cealed defects must be appropriately dealt with Up to the time of installation the packages must be left closed and stored according to the orientation and storage markings on the outside Unless otherwise indicated the packages must be stored only under the following conditions e Notin the open Dry and dust free Not exposed to corrosive media Protected against solar radiation Avoiding mechanical shock and vibration e Storage and transport temperature see chapter Supplement Technical data Ambient conditions e Relative humidity 20 85 3 4 Accessories and replacement parts The display and adjustment module PLICSCOM is used for measured value indication adjustment and diagnosis It can be inserted into the sensor or the external display and adjustment unit and removed at any time You can find further information in the operating instructions Display and adjustment module PLICSCOM Document ID 27835 The interface adapter VEGACONNECT enables the connection of communication capable instruments to the USB interface of a PC For parameter adjustment of these instruments the adjustment software PACTware with VEGA DTM is required You can find further information in the operating instructions Interface adapter VEGACONNECT Document ID 32628 The VEGADIS 81 is an external display and adjustment unit for VEGA plics sensors For sensors with double cha
16. the display and adjustment module The LC display indicates the individual menu items The functions of the individual keys are shown in the above illustration Ap prox 60 minutes after the last pressing of a key an automatic reset to measured value indication is triggered Any values not confirmed with OK will not be saved 6 3 Parameter adjustment The instrument is adapted to the application conditions via the pa rameter adjustment The parameter adjustment is carried out with an adjustment menu 22 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 VEGA Main menu Procedure Setup Isotope 6 Set up with the display and adjustment module The main menu is divided into five sections with the following func tions Displau Diagnostics Additional adjustments Info Setup Settings e g for measurement loop name isotope applica tion background radiation adjustment signal output Display Settings for example language measured value display Diagnosis Information for example of device status peak value simulation Additional adjustments Instrument unit reset date time copying function Info Instrument name hardware and software version date of manu facture instrument features Check if the correct language is already set for the display If not you can change the language in the menu item Display Language Display Setup Diagnostics Displayed value
17. 3 5 Connecting to power supply 5 1 Preparing the connection uincere oerte i E oes ananassae 16 5 2 JGONNE tj Mz sir rere t ERE TER CEDE ERE EGER d Ear te dE To Vg 19 6 Setup with the display and adjustment module 6 1 Insert display and adjustment module eene enne nennen nnn 21 6 2 Adjustment SySIOIm itti peser ite corse tere nec iet Lo aero cia da ev rg tes 22 6 3 Parameter adjustment tre rr n eta er da 22 6 4 Saving the parameter adjustment data nn ent 31 7 Setup with PACTware A Gonne ti the PO n sete ren nda A ree 7 2 Parameter adjustment with PACTware sss 7 8 Saving the parameter adjustment data sssssssssssssssse eee 33 8 Setup with other systems 8 1 DD adjustment programs eadeni dirdaia Taarna bieo niar enne nre nete net tenen nennen nin 34 8 2 Field Communicator 375 475 sseeesennnznnnnznnznnnzznzznnzznntrnzztitzizztitnizztnrnzzzinznzztnznzzztazzni 34 9 Diagnostics and service 9 1 Maintenarice ce rmt n ca en d n c n ire eR rear Od 9 2 Salis messag6S ii ma cte ev eiie td e cute ce ac rc e eee e n ee evacuees 9 3 Rectity faults n prre cr mta irent dee oz ar 9 4 Exchanging the electronics module Pe 9 5 Software UPCAC ses sues ie a eee eene c deed ne tee fee cre ri aec tr vin parir vd 9 6 How to proceed in case of repair nnne terna ne eraai 10 Dismounting POINTRAC 31 Foundation Fieldbus 6LELLEL N3 6ZZ17 41779 EN
18. Abrasion and mechanical influences You can find the specifications of the process conditions in chapter Technical data as well as on the nameplate Protective caps In the case of instrument housings with self sealing NPT threads it is not possible to have the cable entries screwed in at the factory The openings for the cable glands are therefore covered with red protec tive caps as transport protection Prior to setup you have to replace these protective caps with ap proved cable glands or close the openings with suitable blind plugs The suitable cable glands and blind plugs come with the instrument 12 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 VEGA 4 Mounting Installation position 4 2 Mounting instructions Note Along with the planning our specialists will analyse the conditions of the measurement loop to dimension the radiation source isotope accordingly You get a Source Sizing document specifying the required source activity and containing all relevant mounting information for your measuring point You must follow the instructions of this Source Sizing document in addition to the following mounting instructions The following mounting information is applicable as long as there is nothing else specified in the Source Sizing document Fig 3 Mounting position level detection 1 Vertical mounting 2 Mounting horizontally at right angles to container
19. CTware DTM or EDD Out of specification The measured value is unstable because the instrument specification is exceeded e g electronics temperature This status message is inactive by default It can be activated by the user via PACTware DTM or EDD POINTRAC 31 Foundation Fieldbus 35 9 Diagnostics and service VEGA Failure Maintenance Due to external influences the instrument function is limited The measurement is affected but the measured value is still valid Plan in maintenance for the instrument because a failure is expected in the near future e g due to buildup This status message is inactive by default It can be activated by the user via PACTware DTM or EDD The following table shows the error codes and text messages in the status message Failure and provides information on causes as well as corrective measures Example for a failure message Es Invalid linearizer table F025 span too small close together Code Cause Rectification Text mes sage F008 Additional sensors not Check wiring between the Error multi Switched on sensors sensorcom EMC influences Connect the sensors cor munication No other sensor available rectly and make them ready for operation F013 Error on the current input Check current input Sensor sig No valid measured value Check connected instru nals a fault Connected instruments ments Slaves without fu
20. LURE 15 Input out of bounds 13 OUT_OF_SPEC 14 Error while signal processing 68 FAILURE 13 Undefined 13 ki ki 12 Undefined 12 11 Undefined 11 10 Undefined 10 9 Undefined 9 8 Undefined 8 E 7 AITB simulated 29 FUNCTION CHECK 6 Undefined 6 a 5 Undefined 5 kl 4 Undefined 4 E 3 Undefined 3 z 2 Undefined 2 1 Error while trend recording 126 127 GOOD 0 Reserved Reserved Not displayed 33 45 71 86 11 3 Dimensions The following dimensional drawings represent only an extract of all possible versions Detailed dimensional drawings can be downloaded at www vega com downloads under Drawings 50 POINTRAC 31 Foundation Fieldbus 6L LLE L NJ 622 LY 41779 EN 131119 VELA 11 Supplement Aluminium and stainless steel housing HA Em els L Bn I M20x1 5 119mm 1 4 69 ya NPT 169 mm 6 65 1165 mm 175 mm 4 59 39 in un e E E e m y in S Q 8 S CLU E E uy RE wy H t 90 mm 100 mm 3 54 3 94 143 5 mm 5 65 Fig 19 Aluminium housing or stainless steel housing Precision casting POINTRAC 31 Foundation Fieldbus 51 11 Supplement POINTRAC 31 242 5 mm 9 55 9 76 2 mm 3 Fig 20 POINTRAC 31 L Measurin
21. MHTENMEKTVAJBHVIO co6crBeHHOCTb LanbHEMILUVIO wHcpopwauyro cvorpure Ha catre www vega com VEGAR JI MELREA AR P7 URP o 3t 548 Ri 2 PIE lt www vega com gt 11 5 Trademark All the brands as well as trade and company names used are property of their lawful proprietor originator 54 POINTRAC 31 e Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 VEGA INDEX INDEX A Cs 137 23 Adjustment System 22 L Adjustment data 28 Language 27 Adjustment mode 24 Adjustment point 25 M Application 24 Main menu 23 Mode 30 B Background radiation 24 N Block adjustment 27 NAMUR NE 107 Failure 36 C Function check 38 Cable entry 16 Maintenance 38 Check signal 39 Out of specification 38 Connection procedure 17 Connection technology 17 P Control areas 10 Peak value indicator 27 PIN 29 D Damping 29 R Date 29 Radiation protection 10 Date of manufacture 30 Radiation safety officer 10 Default values 29 Relay 27 Delta 25 Repair 40 Device name 30 Replacement parts Device settings Copy 30 Electronics module 9 Displayed value 27 Reset 29 E S EDD Enhanced Device Description 34 Sensor orientation 14 Emitter 23 Sensor status 27 Service F Hotline 39 Fault rectification 38 Setup FF parameter 47 Extended adjustment 23 Functional principle 8 Simulation 28 Function blocks Source container 9 Analog Input Al 47 Status messages NAMUR NE 107 35 H T Handling permit 10 Time 29 HART 30 Type plat
22. N Transducer block is in mode Manual 0x44 Uncertain Last Usable Value Three or more internal communications have failed while Process Data Update Last usable value is displayed 0x48 Uncertain Substitute On user error codes 29 Ox4C Uncertain Initial Value After startup of device or channel assignment in AIFB was changed The channel unit may be un known until next Process Data Update On user error codes 13 0x80 Good NC Non specific No errors concerning to channel handling On user error codes 0 33 45 71 126 127 POINTRAC 31 Foundation Fieldbus 49 11 Supplement VEGA Mapping of User Error Codes to FF Field Diagnostics Priority Description User Error Codes NE 107 Status 31 Hardware failure 40 41 FAILURE 30 Memory failure 34 35 36 37 FAILURE 29 Software failure 80 FAILURE 28 Paramererization corrupt 72 FAILURE 27 Undefined 27 kl ki 26 Undefined 26 ki ti 25 Paramererization error 16 17 25 52 53 57 66 117 FAILURE 120 24 Conflict in MGC 121 122 141 FAILURE 23 Communication error in MGC 8 FAILURE 22 MGC slave reports error 38 FAILURE 21 Undefined 21 20 Undefined 20 ki 19 Undefined 19 gt 18 Detector temperature critical 125 OUT_OF_SPEC 17 Error while auto standardization 73 FAILURE 16 Excessive radiation 123 124 FAI
23. N 131119 VEGA 11 Supplement FF descriptor Rel In Description Unit dex NAMUR_STATE 49 NAMUR state kl NULL COUNT RATE 50 Zero count rate COUNT RATE PMT 51 Pulse rate photomultiplier raw values ADJ DENSITY ABS 52 Process absorption coefficient COEFF DEV SW VER ASCII 53 Software version POINT LEVEL AD 54 Point level adjustment mode JUST MODE RELAY VALUE SEL 55 Relay basic value DIGITAL IN 56 Frequency of digital input kl DIGITAL IN BOOL 57 State of digital input ANALOG IN 58 Current on analog input FF CHANNEL AVAIL 59 Available channels ABLE FF CHANNEL USED 60 Used channels Mapping of Process Value Status Hex Qualitv Sub Status Condition 0x00 Bad Non specific Unexpected error 0x01 Bad Configuration Error It was tried to set a wrong unit with FF interface On user error codes 16 17 25 52 57 66 72 117 120 0x0C Bad Device Failure Five or more internal communications have failed while Process Data Update Last usable value is displayed On user error codes 8 34 35 36 37 38 73 80 86 121 122 141 0x10 Bad Sensor Failure On user error codes 40 41 53 68 123 124 125 Ox1C Bad Out of Service Transducer block is in mode Out of Service Channel is not assigned to an AIFB Channel is not available in running application 0x20 Bad Transducer in MA
24. Operating Instructions Radiation based sensor for level detection POINTRAC 31 Foundation Fieldbus BH Document ID 41779 Contents VEGA Contents 1 About this document IDEM I eir e 4 1 27 TA QOtGHOLIP iii in a pt a dia Sna d 4 T1 3 Symbolism Used iie de l le eb Hei Eee ads 4 2 For your safety 24 Authorised persotnel 2 2 erre eere cc use netu nies ka ee NAESER E iate PEDE RR n Seca eru e avis 22 Approprieto lise uci ned lesse cecus teet us nas ce eaen aE N Er E sa rasa ecu EU 2 3 Warning about incorrect use 2 4 General Safety instructions eere tenen etate So eoe RR ese oleae Ease eo ke pesa rais 2 5 AGE COD OLmily sec toe retreat a ARE 2 6 NAMUR recommendations 2 7 Erwironmerital IDSIEUctlonls coe erre bep pereo te kir a Br iYAEnE nEaN 3 Product description 93 sConfig ratiol iinaseecere heures Leo tarea ER ERR ae ERE D ae RE A p FE eR i PME RE ERE ARES 7 3 2 Principle of operallori eei acute rne turno levnsesecdssecedaatazstaccesdecasedzaasadsanaaeeneeezs 8 3 3 Packaging transport and storage ssssssssssssesee ener nnne 8 3 4 Accessories and replacement parts 3 5 Corresponding source container L seen nem nnnn nn mnn nn tr nn nene 9 4 Mounting 41 Gereraliristrucliohs Je edad eie e bete ah Al aie Pert Hd ease adda 12 42 MOUNTING instr ctlons 3e ipte Fico tetas Pa tipa e bebe tpa bie 1
25. You can find information on protective barriers and the mounting ofthe corresponding source container in the operating instructions manual of the source container e g VEGASOURCE For level detection the sensor is generally mounted horizontally at the height of the requested limit level Make sure that there are no struts or reinforcements at this position in the vessel Direct the exit beam of the source container exactly towards the measuring range of POINTRAC 31 Fasten the sensor in such a way that it cannot fall out of the holder If necessary provide the sensor with a support from below Mount the source container and POINTRAC 31 as close as possible to the vessel If there are gaps secure the area with a safety fence POINTRAC 31 Foundation Fieldbus 138 4 Mounting VELA and protective grating so that no one can reach into the dangerous area Sensor orientation Level detection max detection The POINTRAC 31 is suitable for level detection in liquids or bulk solids It is mounted at the height of the requested switching point Fig 4 POINTRAC 31 as max level detection uncovered Level detection min detection Fig 5 POINTRAC 31 as min level detection covered 14 POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 VEGA 4 Mounting Bulk solids with low density Protection against heat Fig 6 POINTRAC 31 as level detection
26. adjustment module VEGA Diagnosis Adjustment data Diagnosis Simulation This parameter is described in the operating instructions manual Display and adjustment module Here you can retrieve the adjustment value of the sensor This is the percentage value of the max pulse rate at which the sensor switches over If you have carried out a single point adjustment this is the entered value With a two point adjustment this is the calculated value The value is an indication for the reliability and reproducibility of the switching point The greater the difference in the pulse rate between covered and uncovered status the greater the differential value Delta and the more reliable the measurement The automatically calculated damp ing is also oriented around the Delta l value The higher the value the lower the damping A Delta value below 10 is an indication for a critical measurement Adjustment data DeltaI 90 00 96 In this menu item you can simulate measured values via the current output This allows the signal path to be tested e g through down stream indicating instruments or the input card of the control system Note To carry out a simulation with the display and adjustment module you have to switch on the simulation switch on the electronics module switch position 1 You can find the rotary switch on the electronics module in the elec tronics and connection department large
27. app VEGA Tools from the Apple App Store or the Google Play Store e Scanthe Data Matrix code on the type label of the instrument or e Enterthe serial number manually in the app This operating instructions manual applies to the following instrument versions e Hardware from 1 0 4 e Software from 1 5 0 POINTRAC 31 Foundation Fieldbus 7 3 Product description VELA e Modification status electronics as of 01 Electronics versions The instrument is available in different electronics versions Each ver sion can be identified via the product code on the type label e Standard electronics type PT30E XX Scope of delivery The scope of delivery encompasses e Radiation based sensor e Mounting accessory e Documentation this operating instructions manual Operating instructions manual Display and adjustment mod ule optional Ex specific Safety instructions with Ex versions if necessary further certificates 3 2 Principle of operation Application area The instrument is suitable for applications in liquids and bulk solids in vessels under difficult process conditions There are application pos sibilities in nearly all areas of industry The limit level is detected contactlessly through the vessel wall Nei ther a process fitting nor a vessel opening is required The instrument is thus ideal for retrofitting Functional principle In radiation based measurement a Caesium 137 or Cobalt 60 iso tope emits f
28. bility Deviation with bulk solids Deviation under EMC influence Ambient conditions 18 30 C 64 86 F 45 75 96 860 1060 mbar 86 106 kPa 12 5 15 4 psig 0 5 6 The values depend to a great extent on the application Binding specifications are thus not possible lt 1 Ambient storage and transport tempera 40 60 C 40 140 F ture Process conditions For the process conditions please also note the specifications on the type label The lower value always applies Process pressure Process temperature measured on the detector tube Vibration resistance Unpressurized 40 60 C 40 140 F With temperatures of more than 60 C we recommend the use of water cooling mechanical vibrations up to 1 g in the frequency range 5 200 Hz 3 Tested according to the guidelines of German Lloyd GL directive 2 44 POINTRAC 31 Foundation Fieldbus 6LELLEL NA 6ZZ1L7 41779 EN 131119 VEGA 11 Supplement Electromechanical data version IP 66 IP 67 Cable entry M20x 1 5 Ve NPT 2 x cable gland M20 x 1 5 cable 6 12 mm 4x blind plug M20 x 1 5 Included 1 x cable gland M20 x 1 5 5x closing cap red 2 NPT Included 3 x cable gland NPT cable 6 12 mm 4 x blind plug NPT Spring loaded terminals for wire cross section Massive wire cord Stranded wire with end sleeve 0 2 2 5 mm
29. cessary radiation exposure must be avoided An unavoidable radiation exposure must be kept as low as possible Take note of the following three important measures Fig 2 Measures for protection against radioactive radiation 1 Shielding 2 Time 3 Distance Shielding Provide good shielding between the radioactive source and yourself as well as all other persons Special source containers e g VEGASOURCE as well as all materials with high density e g lead iron concrete etc provide effective shielding Time Stay as short a time as possible in radiation exposed areas Distance Your distance to the source should be as large as possible The local dose rate of the radiation decreases in proportion to the square of the distance to the radiation source The plant operator must appoint a radiation safety officer with the necessary expert knowledge He is responsible that the radiation protection ordinance is maintained and that all radiation protection measures are implemented Control areas are areas in which the local dose rate exceeds a certain value Only persons who undergo official dose monitoring are allowed into these control areas You can find the respectively valid limit values 10 POINTRAC 31 Foundation Fieldbus 6GLLLEL NI 6 2Lp 41779 EN 131119 VEGA 3 Product description for control areas in the guideline of the respective authority in Ger many for example the radiation protection ordi
30. e 7 Heat 15 U l Unit 24 Installation position 13 Instrument features 30 V Isotope Voltage supply 45 Co 60 23 POINTRAC 31 2 Foundation Fieldbus 55 INDEX Ww Water cooling 15 56 POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 Notes POINTRAC 31 Foundation Fieldbus 57 Notes 58 POINTRAC 31 Foundation Fieldbus 6LELEL N3762Z v 41779 EN 131119 Notes POINTRAC 31 Foundation Fieldbus 59 Printing date VEGA All statements concerning scope of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior notice C 41779 EN 131119 VEGA Grieshaber KG Schiltach Germany 2013 VEGA Grieshaber KG Phone 49 7836 50 0 Am Hohenstein 113 Fax 49 7836 50 201 77761 Schiltach E mail info de vega com Germany www vega com
31. ent you have to define the difference between the min and max adjustment points Delta I yourself during the following setup 24 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 6 Set up with the display and adjustment module Setup Adjustment uncovered single point adjustment Setup Delta l single point adjustment Adjustment type Adjustment type z Mi point 1 point 2 point This menu item appears only if you have selected Single point adjustment as adjustment mode Setup Adjustment mode In this menu item you determine the point at which the POINTRAC 31 should switch in uncovered status Empty the vessel until the sensor is uncovered For this enter the requested pulse rate manually or let the rate be determined by POINTRAC 31 Automatic determination of the pulse rate should be given preference The pulse rate is entered in ct s This is the number of counts per second i e the measured gamma radiation reaching the sensor Prerequisites e Radiation is switched on Source container is set to ON e There is no medium between source container and sensor Adjustrent uncovered Adjustnent uncovered 90000 ct s Get count Act count rate 44 et s You can enter the value for Adjustment uncovered ct s manually Adjustrent uncovered You can have the value for Adjustment uncovered determined by POINTRAC 31 Get count 610 ct s Escape
32. ent module is inserted into the replacement instrument and the data are likewise written into the sensor via the menu item Copy sensor data POINTRAC 31 Foundation Fieldbus 31 7 Setup with PACTware Via the interface adapter directly on the sensor Prerequisites 7 Setup with PACTware 7 1 Connect the PC Fig 14 Connection of the PC directly to the sensor via the interface adapter 1 USB cable to the PC 2 Interface adapter VEGACONNECT 4 3 gt Sensor Information The interface adapter VEGACONNECT 3 is not suitable for connec tion to the sensor 7 2 Parameter adjustment with PACTware For parameter adjustment of the sensor via a Windows PC the con figuration software PACTware and a suitable instrument driver DTM according to FDT standard are required The up to date PACTware version as well as all available DTMs are compiled in a DTM Collec tion The DTMs can also be integrated into other frame applications according to FDT standard Note To ensure that all instrument functions are supported you should always use the latest DTM Collection Furthermore not all described functions are included in older firmware versions You can download the latest instrument software from our homepage A description of the update procedure is also available in the Internet Further setup steps are described in the operating instructions manu al DTM Collection PACTware attached to each DTM Collection and wh
33. entation as well as a save function for measured value and echo curves In addition there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves The standard version is available as a download under www vega com downloads and Software The full version is avail able on CD from the agency serving you 7 3 Saving the parameter adjustment data We recommend documenting or saving the parameter adjustment data via PACTware That way the data are available for multiple use or service purposes POINTRAC 31 Foundation Fieldbus 33 8 Set up with other systems VEGA 8 Set up with other systems 8 1 DD adjustment programs Device descriptions as Enhanced Device Description EDD are available for DD adjustment programs such as for example AMS and PDM The files can be downloaded at www vega com downloads under Software 8 2 Field Communicator 375 475 Device descriptions for the instrument are available as EDD for pa rameter adjustment with the Field Communicator 375 or 475 For the integration of the EDD in the Field Communicator 375 or 475 the software Easy Upgrade Utility is required which is available from the manufacturer This software is updated via the Internet and new EDDs are automatically taken over into the device catalogue of this software after they are released by the manufacturer They can then be transferred to a Field Comm
34. g range Order length 152 mm or 304 mm 6 in 12 in 52 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp VELA 11 Supplement POINTRAC 31 Mounting example 70 mm E E ES ES 6 8 08 E sitt sit E e N N 5mm 0 20 __ 36mm 1 42 195 mm 7 68 152 5 mm 6 00 127 mm 5 00 gc 51mm EB 2 01 ele ac E h L A ze gcd SE rr E dl ge EL ED Ein gz 14 mm e 14 3 mm 0 55 ta 0 56 Fig 21 POINTRAC 31 with supplied mounting accessories 41779 EN 131119 POINTRAC 31 Foundation Fieldbus 11 Supplement VELA 11 4 Industrial property rights VEGA product lines are global protected by industrial property rights Further information see www vega com Only in U S A Further information see patent label at the sensor housing VEGA Produktfamilien sind weltweit geschutzt durch gewerbliche Schutzrechte Nahere Informationen unter www vega com Les lignes de produits VEGA sont globalement prot g es par des droits de propri t intellec tuelle Pour plus d informations on pourra se r f rer au site www vega com VEGA lineas de productos est n protegidas por los derechos en el campo de la propiedad indus trial Para mayor informaci n revise la pagina web www vega com Jluuna npogykynn cbupmb BETA 3augujarorcs no BceMy Mupy npaBaMM HA
35. iable Output Signal Physical layer Damping 63 of the input variable Channel Numbers Channel 1 Channel 8 Channel 9 Transmission rate Current value Resolution digital Relay output Output Switching voltage Min 152 mm 6 in or 304 mm 12 in 4 20 mA passive 2500 10 mA 100 mA digital output signal Foundation Fieldbus protocol according to IEC 61158 2 0 999 s adjustable Process value switching status Electronics temperature Pulse rate 31 25 Kbit s 10 mA 0 5 mA 0 1 mm 0 004 in Relay output SPDT floating spat 10 mV POINTRAC 31 Foundation Fieldbus 43 11 Supplement Max Switching current Min Max Breaking capacity Min Max Contact material relay contacts Transistor output Output Load current Voltage loss Switching voltage Blocking current 253V AC 253V DC 10 pA 3A AC 1 ADC 50 mW 750 VA AC 40 W DC If inductive loads or stronger currents are switched through the gold plating on the relay contact surface will be permanently damaged The contact is then no longer suitable for switching low level signal circuits AgNi or AgSnO and Au plated Floating transistor output permanently shortcircuit proof 400 mA lt 1V lt 55V DC 10 pA Accuracy according to DIN EN 60770 1 Process reference conditions according to DIN EN 61298 1 Temperature Relative humidity Air pressure Repeata
36. ich can also be downloaded from the Internet Detailed descrip tions are available in the online help of PACTware and the DTMs 32 POINTRAC 31 Foundation Fieldbus 6LLEEL N37622 Ly 41779 EN 131119 7 Setup with PACTware Standard Full version 1 PACTWare o Per File Edit View Project Device Extras Window Help agua 8s 0990 9 3 m 8 SENSOR Online parameterization dbx Device name POINTRAC 31 Deacon Radio based sensor for level detection VEGA Measurement loop name SENSOR Application Point level Bojs sayad B Adjustment of the switching thresholds Adjustment point Software version 1 0 1 PREQ Serial number 19245346 Adjustment mode A Single point E Device status OK Adjustment point 200 ctis Determine pulse rate OK Cancel Poply MR Connected QY BO Device anddataset A EA Administrator te kK NONAME Administrator n 4 Fig 15 Example of a DTM view All device DTMs are available as a free of charge standard version and as a full version that must be purchased In the standard version all functions for complete setup are already included An assistant for simple project configuration simplifies the adjustment considerably Saving printing the project as well as import export functions are also part of the standard version In the full version there is also an extended print function for complete project docum
37. ivery and come with the Ex approved instruments ZI EI PI ST OT ZT 8I 61 OCWIC EEA L T ec b 8 ls Or On O Of Ou Oo O O O O O TL e Fig 10 Electronics and connection compartment Ex d with instruments with intrinsically safe current output 1 GONDORON Voltage supply Relay output Signal input 4 20 mA active sensor Switching input for NPN transistor Switching input floating Transistor output Interface for sensor sensor communication MGC Simulation switch 1 simulation on Setting the bus address for sensor sensor communication MGC Fig 11 Adjustment and connection compartment Ex ia with instruments with oo C 2C 5678 intrinsically safe current output A amp ohN Terminals for intrinsically safe signal output FF bus Contact pins for the display and adjustment module or interface adapter Terminals for the external display and adjustment unit Ground terminal MGC Multi Gauge Communication 20 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 6 Set up with the display and adjustment module Mount Dismount display and adjustment module jie 6 Setup with the display and adjustment module 6 1 Insert display and adjustment module The display and adjustment module can be inserted into the sensor and removed again at any time I
38. mber housing the interface adapter DIS ADAPT is also required for VEGADIS 81 You can find further information in the operating instructions VE GADIS 81 Document ID 43814 The electronics module PTSOE XX is a replacement part for radiation based sensors POINTRAC 31 The electronics module can only be exchanged by VEGA service technician 3 5 Corresponding source container A radioactive isotope in a suitable source holder is the prerequisite for a radiation based measurement setup POINTRAC 31 Foundation Fieldbus 9 3 Product description VEGA Handling permit General instructions for radiation protection Radiation safety officer Control area The handling of radioactive substances is regulated by law The radia tion protection rules of the country in which the system is operated apply first and foremost In Germany for example the current radiation protection ordinance StrlSchV based on the Atomic Energy Law AtG applies The following points are important for measurement with radiation based methods A handling permit is required for operation of a system using gamma rays This permit is issued by the respective government office or the responsible authority in Germany for example offices for environ mental protection trade supervisory boards etc You can find further instructions in the operating instructions manual of the source container When handling radioactive sources unne
39. n Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 6 Set up with the display and adjustment module e Date of manufacture shows calibration date and the date of the last change e Device ID shows the device ID and the sensor TAG PD TAG e Instrument features shows further instrument features These parameters are described in the operating instructions manual Display and adjustment module 6 4 Saving the parameter adjustment data We recommended noting the adjusted data e g in this operating instructions manual and archiving them afterwards They are thus available for multiple use or service purposes If the instrument is equipped with a display and adjustment module the data in the sensor can be saved in the display and adjustment module The procedure is described in the operating instructions manual Display and adjustment module in the menu item Copy sensor data The data remain there permanently even if the sensor power supply fails The following data or settings for adjustment of the display and ad justment module are saved e All data of the menu Setup and Display e Inthe menu Additional adjustments the items Sensor specific units temperature unit and linearization e The values of the user programmable linearization curve The function can also be used to transfer settings from one instru ment to another instrument of the same type If it is necessary to exchange a sensor the display and adjustm
40. nance We are at your disposal for further information concerning radiation protection and regulations in other countries POINTRAC 31 Foundation Fieldbus 11 4 Mounting VELA 4 Mounting 4 1 General instructions Switch off source The source container is part of the measuring system In case the Source container is already equipped with an active isotope the Source container must be locked before mounting Danger IN Before mounting make sure that the source is securelv closed Use a padlock to secure the source container in the closed condition and prevent it from being inadvertentiv opened Protection against mois Protect your instrument against moisture penetration through the fol ture lowing measures e Use the recommended cable see chapter Connecting to power supply e Tighten the cable gland e Loop the connection cable downward in front of the cable gland This applies particularly to e Outdoor mounting e Installations in areas where high humidity is expected e g through cleaning processes e Installations on cooled or heated vessels Suitability for the process Make sure that all parts of the instrument exposed to the process are conditions suitable for the existing process conditions These are mainly Active measuring component e Process fitting e Process seal Process conditions are particularly e Process pressure e Process temperature e Chemical properties of the medium e
41. nction F016 Values of the min and max Correct adjustment data Adjustment adjustment exchanged data ex changed F017 The values of the min and Correct adjustment data Adjustment max adjustment are too error FO25 Wrong value in the lineari Correct linearization table Invalid lineari Zation table zation table F030 Process values are not Repeat adjustment Process value within the adjusted measur out of limits Ing range F034 Electronics defective Exchanging the electronics EPROM hard ware error FO35 Error in the internal instru Carry out a reset EPROM data Ment communication Exchanging the electronics 36 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 9 Diagnostics and service Code Cause Rectification Text mes sage F036 Error during software Repeat software update Faulty pro update Exchanging the electronics gram memory F037 Error in RAM Exchanging the electronics RAM hard ware error F038 Connection cable to the Define instrument as Slave Slave signals Slave instrument interrupted Check the connection cable failure Instrument not defined as to the Slave instrument Slave instrument F040 Sensor defective Exchanging the electronics Hardware error F041 Error in the measured value Exchanging the electronics Photomulti
42. ng be aware of dangerous process conditions such as e g pressure in the vessel or pipeline high temperatures cor rosive or toxic products etc Take note of chapters Mounting and Connecting to power supply and carry out the listed steps in reverse order 10 2 Disposal The instrument consists of materials which can be recycled by spe cialised recycling companies We use recyclable materials and have designed the parts to be easily separable Correct disposal avoids negative effects on humans and the environ ment and ensures recycling of useful raw materials Materials see chapter Technical data If you have no way to dispose of the old instrument properly please contact us concerning return and disposal WEEE directive 2002 96 EG This instrument is not subject to the WEEE directive 2002 96 EG and the respective national laws Pass the instrument directly on to a spe cialised recycling company and do not use the municipal collecting points These may be used only for privately used products according to the WEEE directive POINTRAC 31 Foundation Fieldbus 41 11 Supplement 11 Supplement 11 1 Technical data General data 316L corresponds to 1 4404 or 1 4435 Materials non wetted parts Detector tube Scintillation material Aluminium die casting housing Stainless steel housing Seal between housing and housing cover Inspection window in housing cover optional Ground
43. ocussed gamma rays that are attenuated when penetrat ing the vessel wall and the medium The PVT detector on the opposite side of the tank receives the radiation When the intensity of the radiation drops below a defined value e g due to damping then the POINTRAC 31 switches The measuring principle has proven itself well under extreme conditions because it measures contactlessly from outside through the vessel wall The measuring system ensures maximum safety reliability and plant availability independent of the medium and its properties 3 3 Packaging transport and storage Packaging Your instrument was protected by packaging during transport Its capacity to handle normal loads during transport is assured by a test based on ISO 4180 The packaging of standard instruments consists of environment friendly recyclable cardboard For special versions PE foam or PE foil is also used Dispose of the packaging material via specialised recycling companies Transport Transport must be carried out in due consideration of the notes on the transport packaging Nonobservance of these instructions can cause damage to the device 8 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 VEGA 3 Product description Transport inspection Storage Storage and transport temperature PLICSCOM VEGACONNECT VEGADIS 81 Electronics module The delivery must be checked for completeness and possible transit damage
44. pli recording er error F045 Error on the current output Check wiring of the current Error on the output current output Exchanging the electronics F052 Invalid parameter adjust Carry out a reset Faulty config ment uration F114 Discharge accumulator Readjust real time clock Error real time clock Fi22 Instrument addresse was Change instrument Double ad assigned several times addresses dress on the multisensor communica tion bus F123 External instruments cause Determine reason for X ray X ray alarm radioactive radiation alarm In case of a brief X ray alarm shut down the instru ment switching outputs for this time F124 Radiation dose too high Determine reason for Alarm due to increased radiation increased ra diation F125 Ambient temperature on Cool heat the instrument Ambient tem the housing outside the or protect it with isolation perature too high specification material against cold or radiation heat POINTRAC 31 Foundation Fieldbus 37 9 Diagnostics and service VEGA Function check Out of specification Maintenance Reaction when malfunc tions occur Procedure for fault recti fication The following table shows the error codes and text messages in the status message Function check and provides information on causes as well as corrective measures Code Cause Rec
45. pplication This menu item enables adaptation of the sensor to the requested ap plication You can choose between the following applications Level Point level or Summation Slave Application Point level ki mx Standardization The natural radiation on earth influences the accuracy of the meas urement With this menu item the natural background radiation can be faded out For this purpose the POINTRAC 31 measures the natural back ground radiation and sets the pulse rate to zero In the future the pulse rate from this background radiation will be automatically deducted from the total pulse rate This means only the component of the pulse rate originating from the source will be displayed The source container must be closed for this setting Background radiation Backaround radiation Oct s Get count In this menu item you can select the temperature unit Temperature unit Temperature unit vI in this menu item you can select if you want to carry out a single or double point adjustment on the sensor With the double point adjustment the Delta value is selected auto matically We recommend selecting the double point adjustment To use this you must be able to change the level of the vessel so as to carry out the adjustment of the sensor with full status covered and with empty status uncovered Hence you will get a very reliable switching point With single point adjustm
46. rmity The device fulfills the legal requirements of the applicable EC guide lines By affixing the CE marking VEGA confirms successful testing of the product Only with class A instruments The device is a class A instrument designed for use in an industrial environment When used in a different environment e g in a living area the electromagnetic compatibility must be ensured by the user If necessary suitable screening measures against conducted and emitted disturbances must be taken You can find the conformity certificate in the download section under www vega com 2 6 NAMUR recommendations NAMUR is the automation technology user association in the process industry in Germany The published NAMUR recommendations are accepted as the standard in field instrumentation The device fulfills the requirements of the following NAMUR recom mendations e NE 21 Electromagnetic compatibility of equipment e NE 43 Signal level for malfunction information from measuring transducers e NES53 Compatibility of field devices and display adjustment components e NE 107 Self monitoring and diagnosis of field devices For further information see www namur de 2 7 Environmental instructions Protection of the environment is one of our most important duties That is why we have introduced an environment management system with the goal of continuously improving company environmental pro tection The environment management system is cer
47. rument can result Danger If this warning is ignored serious injury to persons and or destruction of the instrument can result Ex applications This symbol indicates special instructions for Ex applications List The dot set in front indicates a list with no implied sequence Action This arrow indicates a single action Sequence of actions Numbers set in front indicate successive steps in a procedure Battery disposal This symbol indicates special information about the disposal of bat teries and accumulators POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 2 For your safety 2 For your safety 2 1 Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator During work on and with the device the required personal protective equipment must always be worn 2 2 Appropriate use The POINTRAC 31 is a sensor for point level detection You can find detailed information on the application range in chapter Product description Operational reliability is ensured only if the instrument is properly used according to the specifications in the operating instructions manual as well as possible supplementary instructions 2 3 Warning about incorrect use Inappropriate or incorrect use of the instrument can give rise to application specific hazards e g vessel overfill or damage to s
48. s the electrical characteristics of the bus system as well as the applied function blocks Revisions Data DD Revision Rev 01 CFF File 010101 cff Device Revision 0101 ffo 0101 sym Cff Revision Xx xx 01 Device software revision 1 5 0 ITK Interoperability Test Kit Number 6 0 1 Electricial Characteristics Physicial Layer Type Low power signaling bus pow ered FISCO I S Input Impedance 3000 Ohms between 7 8 KHz 39 KHz Unbalanced Capacitance 250 pF to ground from either in put terminal Output Amplitude 0 8 V P P Electrical Connection 4 Wire Polarity Insensitive Yes Max Current Load 10mA Device minimum operating voltage 9v Transmitter Function Blocks Resource Block RB 1 Transducer Block TB 1 Standard Block AI 3 Execution Time 30 ms Diagnostics Standard Yes Advanced Yes Performance No Function Blocks Instantiable No A suitable cable is the prerequisite for maintaining the protection rating 46 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 VEGA 11 Supplement General Information LAS Link Active Scheduler No Master Capable No Number of VCRs Virtual Communication 24 Relationships Function block Analog Input Al The function block Analog Input AI takes the original measured value selected by a Channel Number and makes it available to additional f
49. sufficient protection against moisture and must be replaced with blind plugs The suitable cable glands and blind plugs come with the instrument Make sure that the cable screening and ground is executed accord ing to the Fielbus specification If electromagnetic interference is expected which is above the test values of EN 61326 1 for industrial areas we recommend to connect the cable screen on both ends to ground potential In systems with potential equalisation connect the cable screen directly to ground potential at the power supply unit in the connection box and at the sensor The screen in the sensor must be connected directly to the internal ground terminal The ground terminal outside on the housing must be connected to the potential equalisation low impedance In systems without potential equalisation with cable screening on both sides connect the cable screen directly to ground potential at the power supply unit and at the sensor In the connection box or 16 POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 5 Connecting to power supply Connection technology Connection procedure T distributor the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen The cable Screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a ceramic capacitor e g 1 nF 1500 V
50. t is not necessary to interrupt the power supply Proceed as follows 1 Unscrew the small housing cover 2 Place the display and adjustment module in the desired position on the electronics you can choose any one of four different posi tions each displaced by 90 3 Press the display and adjustment module onto the electronics and turn it to the right until it snaps in 4 Screw housing cover with inspection window tightly back on Removal is carried out in reverse order The display and adjustment module is powered by the sensor an ad ditional connection is not necessary 1 Fig 12 Insert display and adjustment module Note If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication a higher cover with an inspection glass is required POINTRAC 31 Foundation Fieldbus 21 6 Set up with the display and adjustment module VELA Key functions Adjustment system 6 2 Adjustment system Fig 13 Display and adjustment elements 1 LC display 2 Adjustment keys e OK key Move to the menu overview Confirm selected menu Edit parameter Save value e key Presentation change measured value Select list entry Select editing position e key Change value of the parameter e ESC key Interrupt input Jump to next higher menu The device is adjusted via the four keys of
51. terminal Mounting accessorv Process fittings Fastening lugs Weight Aluminium housing with electronics Stainless steel housing with electron ics Measuring length 152 mm 6 in Measuring length 304 mm 12 in 316L PVT Polyvinyltoluene Aluminium die casting AISi1OMg powder coated basis Polyester 316L NBR stainless steel housing investment casting silicone Aluminium housing Polycarbonate 316L 316L 9 mm 0 35 in hole centre distance 119 mm 4 69 in 3 4 kg 7 5 Ibs measuring length 8 36 kg 18 43 Ibs measuring length 0 98 kg 2 16 Ibs 1 95 kg 4 3 Ibs Max torque mounting screws fastening 50 Nm 36 88 Ibf ft lugs on the sensor housing Max torque for NPT cable glands and Conduit tubes Aluminium Stainless steel housing Input variable Measured variable 50 Nm 36 88 Ibf ft The measured variable is the intensity of the gamma radiation When the intensity of the radiation is below the stipulated value due to a damping by the medium the POINTRAC 31 switches 42 POINTRAC 31 Foundation Fieldbus 6LLLEL NI 6 2Lp 41779 EN 131119 11 Supplement Fig 17 Data of the input variable L Measuring range range in which the switching point must lie Measuring range Analogue input Input type Internal load Switching input Type of input Open Collector Type of input Relay contact Output var
52. the clock Since we offer this service worldwide the support is provided in English The service itself is free of charge the only costs involved are the normal call charges 9 4 Exchanging the electronics module If the electronics module is defective it can be replaced by the user In Ex applications only instruments and electronics modules with ap propriate Ex approval may be used If there is no electronics module available on site the electronics module can be ordered through the agency serving you The electron ics modules are adapted to the respective sensor and differ in signal output or voltage supply The new electronics module must be loaded with the default settings ofthe sensor These are the options e Inthe factory e Oronsite by the user In both cases the serial number of the sensor is needed The serial numbers are stated on the type label of the instrument on the inside of the housing as well as on the delivery note POINTRAC 31 Foundation Fieldbus 39 9 Diagnostics and service VEGA When loading on site first of all the order data must be downloaded from the Internet see operating instructions manual Electronics module 9 5 Software update The following components are required to update the sensor soft ware e Sensor Voltage supply Interface adapter VEGACONNECT PC with PACTware Current sensor software as file You can find the actual sensor software as well as detailed infor
53. tification Text mes sage C029 Simulation active Finish simulation Simulation Wait for the automatic end after 60 mins The following table shows the error codes and text messages in the status message Out of specification and provides information on causes as well as corrective measures Slave outside the specifica tion tion Code Cause Rectification Text mes sage S017 Accuracy outside the Correct adjustment data Accuracy specification outside the specification S025 Bad linearization table Carry out linearization Bad lineariza tion table S038 Slave outside the specifica Check Slave S125 Ambient tem perature too high too low Ambient temperature too high too low Protect instrument with isolating material against extreme temperatures The instrument has no status messages to the section Maintenance 9 3 Rectify faults The operator of the system is responsible for taking suitable meas ures to rectify faults The first measures are e Evaluation of fault messages via the indicating and adjustment module e Checking the output signal e Treatment of measurement errors Further comprehensive diagnosis possibilities are offered by a PC with the software PACTware and suitable DTM In many case the failure reasons can thus be determined and faults rectified 38 POINTRAC 31 Foundation Fieldbus 6L
54. tified according to DIN EN ISO 14001 Please help us fulfill this obligation by observing the environmental instructions in this manual e Chapter Packaging transport and storage e Chapter Disposal POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 VEGA 3 Product description Type plate Serial number Scope of this operating instructions manual 3 Product description 3 1 Configuration The nameplate contains the most important data for identification and use of the instrument s n 19997099 77761 Schiitach Germany WWW vega Com Fig 1 Lavout of the tvpe label example 1 Instrument type 2 Product code 3 Electronics 4 Protection rating 5 Process and ambient temperature process pressure 6 Instrument length 7 Hardware and software version 8 Order number 9 Serial number of the instrument 10 ID numbers instrument documentation The type label contains the serial number of the instrument With it you can find the following data on our homepage Product code of the instrument HTML Delivery date HTML Order specific instrument features HTML Operating instructions at the time of shipment PDF Order specific sensor data for an electronics exchange XML Test certificate pressure transmitters PDF Go to www vega com VEGA Tools and Serial number search As an alternative you can find the data via your Smartphone e Download the smartphone
55. to power supply 5 1 Preparing the connection Always keep in mind the following safety instructions e Connect only in the complete absence of line voltage e lf overvoltage surges are expected overvoltage arresters should be installed Power supply and digital bus cable via separate two wire connection cables Power is supplied via the H1 voltage supply Connection is carried out with screened cable according to Fieldbus specification Use cable with round cross section A cable outer diameter of 5 9mm 0 2 0 35 in ensures the seal effect of the cable gland If you are using cable with a different diameter or cross section exchange the seal or use a Suitable cable gland Make sure that the entire installation is carried out according to the Fieldbus specification In particular make sure that the bus is termi nated with suitable terminating resistors Generally provide all unused cable entries with suitable blind plugs The thin foam rubber washers in the cable glands only serve as a dust cover during transport In the case of instrument housings with self sealing NPT threads it is generally not possible to have the cable glands screwed in at the factory The openings for the cable glands are therefore covered with red protective caps as transport protection Before setup you have to replace these protective caps with approved cable glands or close the openings with suitable blind plugs Unused cable glands do not provide
56. ue Setup Measurement loop name X Sensor Isotope Cs 137 Application Limit level Adjustment mode Single point adjustment Adjustment uncovered 90000 ct s Adjustment covered 9000 ct s only with two point adjust ment Delta 90 Background radiation 0 ct s Temperature unit C Adjustment mode 1 point Adjustment uncovered 900000 ct s Delta 90 Mode relay Overfill protection Block adjustment Released Display Language Selected language Displayed value Pulse rate Display unit ct s With this function you can select the mode The sensor offers the HART modes standard and multidrop If the measured value is outputted via the 4 20 mA output you must not switch over to HART Multidrop The mode Standard with fixed address 0 factory setting means output of the measured value as 8 16 mA signal This parameter is described in the operating instructions manual Display and adjustment module With this function Load parameter adjustment data from the sensor into the display and adjustment module e Write parameter adjustment data from the display and adjustment module into the sensor This parameter is described in the operating instructions manual Display and adjustment module Info In this menu you will find the following menu items e Instrument name shows instrument name and serial number e Instrument version shows hardware and software version of the instrument 30 POINTRAC 31 Foundatio
57. unction blocks on its output Simulate SIMULATE L_TY FIELD_VAL Fig 18 Schematic presentation function block Analog Input Al Parameter list The following table gives you an overview of the parameters used process value after min max adjustment and Lin earization with the status of the transducer block The unit is defined in PRIMARY VALUE UNIT FF descriptor Rel In Description Unit dex PRIMARY VALUE 13 PRIMARY VALUE Linearized value This isthe FF PRIMARY VAL UE UNIT FLOW FF PRIMARY VALUE 14 Selected unit code for PRIMARY VALUE UNIT FF VAPOR DENSITY 115 Density with Temperature correction FF VAPOR DENSGSI TV UNIT FF VAPOR DENSITY 16 Selected unit code for FF VAPOR DENSITY UNIT FF PROCESS TEM 17 Process temperature FF PROCESS TEM PERATURE PERATURE UNIT FF PROCESS TEM 18 Selected unit code for FF PROCESS TEMPER PERATURE UNIT ATURE FF DENSITY 19 Density FF DENSITY UNIT FF DENSITY UNIT 20 Selected unit code for FF DENSITY FF VOLUMETRIC 21 Volumetric flow FF VOLUMETRIC FLOW UNIT POINTRAC 31 Foundation Fieldbus 47 11 Supplement FF descriptor Rel In Description Unit dex FF VOLUMETRIC 22 Selected unit code for FF VOLUMETRIC FLOW FLOW UNIT FF WEIGHT 23 Weight on belt FF WEIGHT
58. unicator 34 POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 9 Diagnostics and service Status messages 9 Diagnostics and service 9 1 Maintenance If the device is used correctly no maintenance is required in normal operation The corresponding source container must be checked in regular intervals You can find further information in the operating instructions manual of the source container 9 2 Status messages The instrument features self monitoring and diagnostics according to NE 107 and VDI VDE 2650 In addition to the status messages in the following tables there are more detailed error messages available under the menu item Diagnostics via the display and adjustment module PACTware DTM and EDD The status messages are divided into the following categories Failure Function check Out of specification Maintenance requirement and explained by pictographs sara Fig 16 Pictographs of the status messages Failure red Out of specification yellow Function check orange Maintenance blue ROMA Failure Due to a malfunction in the instrument a failure message is outputted This status message is always active It cannot be deactivated by the user Function check The instrument is in operation the measured value is temporarily invalid for example during simulation This status message is inactive by default It can be activated by the user via PA
59. unt rate 101 et sj You can enter the adjustment point ct s manually Adjustnent covered You can let the adjustment point be determined by POINTRAC 31 Get count 610 ct s Escape This menu item appears only if you have selected under adjustment mode setup adjustment mode the Two point adjustment In this menu item you can set the max pulse rate ct s at which the sensor should switch over Empty the vessel until the POINTRAC 31 is uncovered You thus get the max pulse rate ct s for the uncovered adjustment Enter the requested pulse rate manually or let the rate be determined by POINTRAC 31 Automatic determination of the pulse rate should be given preference Adjustrent uncovered Adjustrent uncovered 70000 ct s Get count Act count rate 93 ct s You can enter the adjustment point ct s manually Adjustnent uncovered 26 POINTRAC 31 Foundation Fieldbus 6LELEL N3 6ZZ1L7 41779 EN 131119 6 Set up with the display and adjustment module Setup Relay Lock setup adjustment Display Language Display Displayed value Diagnostics Device status Diagnosis Peak value You can let the adjustment point be determined by POINTRAC 31 Get count 610 ct s Escape In this menu item you can select which mode the sensor should oper ate in You can choose between overfill and dry run protection The relay outputs of the sensor react
60. ystem components through incorrect mounting or adjustment 2 4 General safety instructions This is a state of the art instrument complying with all prevailing regulations and guidelines The instrument must only be operated in a technically flawless and reliable condition The operator is responsible for the trouble free operation of the instrument During the entire duration of use the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula tions The safety instructions in this operating instructions manual the na tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user For safety and warranty reasons any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer Arbi trary conversions or modifications are explicitly forbidden The safety approval markings and safety tips on the device must also be observed This measuring system uses gamma rays Therefore take note of the instructions for radiation protection in chapter Product description All work on the source container may only be carried out under the supervision of a qualified radiation protection officer POINTRAC 31 Foundation Fieldbus 5 2 For your safety VEGA 2 5 CE confo

Download Pdf Manuals

image

Related Search

Related Contents

Handbuch - Speidel Gartengeräte Fachmarkt für Garten  Liebert Load Bus Synchronization  FBC-007DX取付け取扱説明書  Calculated Industries Scale Master Pro XE  Manual de instalación  FICHA TÉCNICA PRODUCTO LIMPIADOR PARA TECA FASE 1  Samsung 22" Business moniteur avec grande fonction ergonomique Manuel de l'utilisateur  Manuel PSE V2 - Ministère de l`écologie, du développement durable  Toshiba Satellite L655D-S5110  

Copyright © All rights reserved.
Failed to retrieve file