OPERATING INSTRUCTIONS FOR THE HAND HELD

image

Contents

1. Key 1 Pressure port 3 8 BSP Parallel female connection to take master instruments e g digital or analogue gauge PC6 Calibrator Fine adjustment valve Pressure release valve i Pressure Vacuum selector Adjustable stroke for varying maximum pressure output over pressure protection BSP or NPT adaptor set Flexible hose to item under test Nylon seals for BSP adaptors see seal kit provided DO NOT use PTFE tape for sealing with parallel threads 9 Knurled quick fit connectors 10 Pump Handles aARwN OND Specification Output pressure 0 to 40 bar 0 to 600 psi Adjustable Output Vacuum 0 to 960 mbar 0 to 29 inHg Materials Bright nickel plated brass clear anadised aluminium Nylon Adjustment Fine volumetric pressure vacuum adjuster Dimensions 220mm L x 105mm W X 63mm D Weight 650 grams The system is a portable dual source of vacuum and pressure Each system incorporates a vacuum pressure selector a volume control for fine adjustment and adjustable stroke to provide over pressure protection TEST INSTRUMENTS AVAILABLE TPG1 TPG6 range of Analogue gauges 1 0 accuracy Digital Indicator DTG2K 0 2 accuracy PCB PC6 PRO PC6 IS Pressure Calibrators up to 0 025 accuracy DPM Digital Pressure Module up to 0 025 accuracy HOSE ADAPTORS The hoses 7 and adaptors 6 are fitted by simply screwing them into the connectors 9 by turning the knuried
2. fine adjustment valve 2 beyond the red line indicator 1 8 Reductions in pressure can also be achieved by careful use of the pressure release valve 3 1 9 Vacuum is achieved using the above procedure and having the changeover valve 4 pushed completely towards the vacuum position Note release pressure before changing mode 2 Use With High Resolution Pressure Calibrators When used with instruments such as the PC6 Calibrator the connections and use are as for gauges above however the higher resolution available will amplify the visibility of the thermodynamic effects as mentioned in paragraph 1 7 These will settle to useable values within one minute of pressurisation Note On very high resolutions such as mbar or 0 1 inches of water smal movements of the pipe may result in noticeable pressure changes 3 Fault Investigation In the event that the system appears to lose pressure then the procedure above should be repeated ensuring new seals are used adaptors are tightened sufficiently and the pressure release valve 3 is tightened firmly Note The connections to the hand held test system are sealed with o ring or bonded seals and should not leak The pipe to body connection can be checked but tightened no more than 2 Nm DO NOT attempt to tighten the other fittings to the test system as this could lead to damage of sealed joints When testing for leaks it may be noticed that air is drawn in or expelled from around the changeover
3. knut on the connector fully anti clockwise RELEASE VALVE 3 This can be used to reduce or release the pressure in the system The rate of pressure reduction is dependent upon the degree of rotation when opening the valve Minimal force is required to seal the system VOLUME CONTROL 2 2 The pressure generated can be finely adjusted by turning the fine adjustment valve 2 either clockwise or anticlockwise to increase or decrease pressure accordingly 3 OVER PRESSURE PROTECTION 5 To adjust the maximum output pressure of the system turn the nuts 5 to increase or decrease the stroke length IMPORTANT Under no circumstances should the fine adjustment valve 2 be wound back beyond the red line indicator on the body Should this occur then the pressure must be released from the system before attempting to re engage the fine adjustment valve PRESSURE VACUUM SELECTION 4 Press the selector 4 as indicated on the label to engage the desired mode Ensure that the release valve 3 is open before changing mode Red Line NOTE Indicator The system should only be used for pressurising small volumes due to its small displacement lf the system has not been used for a period of time it could be difficult to operate on the first stroke The cylinder has been lightly greased on assembly but if additional lubrication should ever be required then apply a minimal amount to the inside of the cylinder Access is via the three retaining screw
4. s located under the black collar For seal replacement refer to service kit TPK1 40 instructions PRESSURE INSTRUMENTS 31 0302 40 Iss 6 DIM 5312 SS a A WARNING DO NOT CONNECT TP1 40 TO EXTERNAL PRESSURE SOURCE Guidelines for use 1t Calibration Comparison against Analogue gauge 1 1 Fit a S Test Gauge to the top of the test system 1 for other makes fit by using adaptor 18111 not supplied and correct seal Note The retaining nut should be screwed fully down but needs to be no more than finger tight as seal is achieved by o ring 1 2 Connect item under test using appropriate adaptor and sealing 6 at the end of the flexible hose 7 or directly to ee adaptors tightened to a maximum torque of 15 Nm 1 3 Screw fine adjustment valve 2 fully clockwise 1 4 Screw fine adjustment valve 2 4 6 full turns anticlockwise 1 5 Screw pressure release valve 3 fully clockwise tightening to ensure good seal 1 6 Operate handles 10 until the pressure is close to that finally required Ensure handles are fully squeezed together on each stroke to achieve maximum pressure output 1 7 Wind the fine adjustment valve 2 clockwise to increase pressure or anticlockwise to decrease pressure until required pressure is reached Note The pressure may settle for up to 30 seconds after increasing pressure due to thermodynamic effects settling of seals and expansion of the flexible hose Caution NEVER screw the
5. valve This is normal and should cause no concern Ordering Codes TP1 40 Hand held pressure test system in rigid carrying case with 1 metre flexible hose and adaptor set PC6 PC6 PRO A range of Pressure Calibrators available between 1 and 1000 bar 15 000 psi PC6 AV PC6 IS DTG 2K range 5 ranges of Digital Indicator up to 1000 bar DPM A range of Digital Pressure Modules available between 1 and 1000 bar TPG1 2 3 4 5 6 G6 ranges of Analogue Gauge between 1 and 40 bar TPK1 40 Service kit containing a set of seals O rings retaining screws and allen key LTPA BSP OQF Set of 7 BSP Quick_fit adaptors LTPA NPT GF Set of 6 NPT Quick fit adaptors 18114 3 8 BSP male to 4 BSP female swivel adaptor HTP1 Hydraulic hand held test system for pressures up to 1000 bar HTPG1 2 3 4 4 ranges of Analogue Gauges 0 5 accuracy between 0 to 1000 bar URE HTPK1 Service Kit containing a complete set of seals for HTP1 PRESS L INSTRUMENTS