1. A Type B equipment A Attention consult this manual AKRON 9932 11 Transportation and Storage Handle with care Do not drop Avoid shock or violent impact The equipment should be stored in a clean dry and well ventilated area The following limits apply during transport and for a storage period of up to 15 weeks Ambient temperature 25 C to 70 C Relative humidity 10 to 75 Air pressure 50 kPa to 106 kPa The following limits apply to operating conditions or longer periods of storage Ambient temperature 10 C to 40 C Relative humidity 30 to 75 Air pressure 70 kPa to 106 kPa 12 Environmental Protection Dispose of unit in compliance with local regulations Huntleigh HEALTHCARE Huntleigh Akron 1 Farthing Road Ipswich Suffolk IP1 SAP England Tel 44 0 1473 461042 Fax 44 0 1473 462924 792 297 05
2. Huntleigh e HEALTHCARE AKRON AKRON GYNAECOLOGY 3 Model 9932 USER MANUAL AKRON 9932 Contents 1 10 11 12 Introduction Warnings Cautions General Warnings Installation Operation Elevation Controls Wheel Locking Seat Eztension Leg Supports Debris Tray Instrument Shelf Arm Loops Paper Roll Holder Head Support Stool amp Chair Accessories Operating Instructions Triple Footswitch Arm Support Arm Rests Cleaning and Decontamination Troubleshooting Maintenance Warranty and Service Revision Status Technical Specification Transport and Storage Environmental Protection CE The C mark on this product demonstrates conformity with the EC Directive 93 42 EEC NNN O 1 O LD A O Ma 10 11 11 12 13 13 14 15 15 AKRON 9932 1 Introduction Always keep this Operating Manual available for reference Warnings amp Cautions WARNINGS Warnings identify possible hazards in procedures or conditions which if not correctly followed could result in death injury or other serious adverse reactions CAUTIONS Cautions identify conditions or procedures which if not correctly followed could result in equipment failure or damage General Warnings WARNINGS Use only accessories that have been designed or approved for use with this couch Do not use electric couches in the presence of flammable gasses such as an anaesthetic age
3. a suitable cleaning area with castors disengaged or wheels locked l Wear protective clothing plastic disposable apron and gloves 2 Disconnect any mains supply and lock couch wheels 3 Control handsets can be wiped over using a 70 Isopropranol alcohol wipe 4 Prepare a solution of warm water and neutral detergent 5 Start at the top of the couch and work your way downwards 6 Wipe over head end working towards mid torso 7 Wipe over remaining horizontal patient surface working towards foot end 8 Wipe all horizontal parts under the patient surface 9 Wipe all vertical parts under the patient surface 10 Wipe over shroud if fitted EP Wipe over all parts of wheels 12 Ensure all parts of the couch are dry before re use 13 Safely discard cleaning materials 14 Wash hands thoroughly If DECONTAMINATING then repeat steps 1 to 14 using a dilute Hypochlorite warm water mixture 10 000ppm instead of warm water and soap Put couch back into service AKRON 9932 6 Troubleshooting Electric Actuator Actuator will not work Power disconnected Ensure mains supply is connected Fuse blown Replace fuse in socket Hand footswitch disconnected Check that the hand foot switch 1s fully inserted into the actuator If the problem persists consult your local Akron Supplier 7 Maintenance Check table visually before use Carry out routine servicing every six months or more frequently if in heavy use Rout
4. other Huntleigh Healthcare Akron Product Division Product please contact Huntleigh Healthcare Akron Product Division l Farthing Road Ipswich Suffolk IP 5AP England Telephone 44 0 1473 461042 Fax 44 0 1473 462924 Huntleigh Healthcare Akron Product Division A Member of the Huntleigh Technology PLC Group of Companies O Huntleigh Technology PLC 2003 If for any reason your couch 1s being returned please inform Huntleigh Akron prior to sending and l 3 4 Clean the product as described in the cleaning section Pack it 1n suitable packing Attach the decontamination certificate or other written statement declaring that the product has been cleaned to the outside of the package Mark the package Service Department Manufactured in the UK by Huntleigh Healthcare Akron Product Division As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice 9 Revision Status 9932 PAGES ISSUE WRITTEN DATE ALL PAGES UM 01 22 11 99 AKRON 9932 10 Technical Specification Akron Gynae 3 9932 Head Length cm 851 5 Mid Length cm Foot Length cm Width cm Overall Length cm Height Range cm 631 5 991 5 Max Load Weight 180 Kg Elevation Control Electrical Data 230v Voltage 220 240v 50 Hz Current Power 24v 6A Electrical Data 115v 115v 60 Hz 24v 6A IEC 601 1 Classification Class 1
5. ealthcare 310 312 Dallow Road Luton Beds Telephone 01582 745700 LUI 1TD Fax 01582 745869 For information on this product or any other Huntleigh Healthcare Akron Product Division s product please contact Huntleigh Healthcare Akron Product Division Farthing Road Ipswich Telephone 01473 461042 IP1 SAP Fax 01473 462924 Huntleigh Healthcare Akron Product Division A Member of the Huntleigh Technology PLC Group of Companies O Huntleigh Technology PLC 2003 If for any reason your couch is being returned please inform Huntleigh Healthcare Akron Product Division prior to sending and l Clean the product as described in the cleaning section 2 Pack it in suitable packing 3 Attach the decontamination certificate or other written statement declaring that the product has been cleaned to the outside of the package 4 Mark the package Service Department Manufactured m the UK by Huntleigh Healthcare Akron Product Division As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice AKRON 9932 8 Warranty and Service Outside UK Huntleigh Healthcare Akron Product Division s standard terms and conditions apply to this couch A copy 1s available upon request These contain full details of warranty terms and do not limit the statutory rights of the consumer For service maintenance and any questions regarding this or any
6. eg support to desired angle and secure by tightening T lever Debris Tray CAUTION Ensure debris tray s removed and frame positioned under seat prior to patient mounting or dismounting couch A To remove tray lift out of frame and slide frame under seat B To insert tray slide frame out and position debris tray into frame AKRON 9932 3 Operation Instrument Shelf CAUTION Do NOT use as a step as weight bearing may cause damage To use instrument shelf A Slide to position required B Ensure instrument shelf is pushed in when patient is dismounting couch Arm Rests CAUTION Ensure arm loop is secure prior to use A Slide din rail clamp onto din rail in desired position B Locate arm loop into din rail clamp and secure into position using locking knob C Adjust position by loosening lever on din rail clamp sliding to desired position and secure by tightening lever To replace paper roll A Take hold of paper roll B Place over plastic tube C Fit one end of tube over tab on holder D Gently spring holder to fit other tab into tube 3 Operation Head Support Operators Stool amp Chair AKRON 9932 Adjust Head Support to suit patient by A Unfastening Velcro strap B Position where required C Fasten Velcro straps to secure Either chair or stool available as standard Stipulated on order Alternatively either available as optional accesso
7. ine servicing or repair must only be performed by suitably qualified person WARNING Before commencing any maintenance activity disconnect the couch from the electrical supply Plug acts as disconnect device Examine the table for any signs of damage Make sure that all nuts bolts and other fasteners are tight Do not over tighten as this may cause binding Upholstery fasteners should be checked tightened or replaced 1f necessary Check section hinge pins to ensure that no pin protrudes Check gas strut controls throughout the full range of movement for locking and assistance Check hydraulic tables for smooth operation and signs of leakage WARNING All electrical checks should be performed by a suitably qualified and licensed electrician Neither BLUE or BROWN wires are to be connected to the earth terminal of a three pin plug Examine the flexible cables for cuts abrasions kinking or other deterioration and replace if necessary Ensure that the power supply connector plug is securely attached Check mains plug connections are tight and fuse 1s correctly rated 5A 11 AKRON 9932 8 Warranty and Service UK only Huntleigh Healthcare Akron Product Division s standard terms and conditions apply to this couch A copy 1s available upon request These contain full details of warranty terms and do not limit the statutory rights of the consumer For service or maintenance please contact Service Department Huntleigh H
8. nt Maximum Load 180 Kg 396 lbs distributed evenly Do not concentrate weight on either side of the couch Plug acts as a disconnect device Do not allow children to play with the couch at any time Adequately supervise children in the proximity of the couch AKRON 9932 2 Installation Electric Models WARNING Do not insert the 2 pin mains connector directly into a 3 pin 13A mains outlet Always use the adapter provided Plug acts as a disconnect device Use adapter 1 3 pin 13A plug 1s required Order reference 5804097 Ensure mains supply cable 1s not stretched when connected Ensure during couch use no obstacles are in its path Ensure hand foot switch tubes or leads are not trapped by moving parts of the couch CAUTION When not in use switch couch off Battery Backup CAUTION Battery Backup is for the unlikely event of mains power failure not normal use The couch should always be used with mains power connected The light on the control box indicates mains power is on Whilst on batteries are trickle charged In case of power failure To charge the accumulators the control unit must be activated by shortly pressing one of the switch keys AKRON 9932 3 Operation Elevation Controls WARNING Safely position patient before elevating couch Position all electrical cables away from the moving parts Position Handswitch where accidental operation is not possible Pres
9. ries Operators Stool Model 8500 Operators Chair Model 8600 AKRON 9932 4 Accessories Operating Instructions Triple Footswitch Optional Extra A B A B C D Arm Rests Optional Extra To use triple footswitch Remove existing handswitch from control box if fitted Plug in D sub lead into handset port on control box To adjust arm support Slide din rail clamp onto din rail in desired position secure by tightening lever Locate support into din rail clamp and secure in position using locking knob To adjust angle of arm support pad loosen lever on underside of pad adjust to desired position and secure by tightening lever If further angle adjustment required loosen T lever by turning anti clockwise rotate support to desired angle and secure by tightening T lever Arm Rest down Arm Rest up A To adjust angle rotate the knob on the underside of the armrest B To raise or lower the armrest simply move to position up or down AKRON 9932 5 Cleaning and Decontamination WARNING Do not use Alkalis or other organic solvents for routine cleaning Replace torn or damaged vinyl immediately Always cover vinyl with fresh paper or sheet between patients If contamination is known or suspected always decontaminate couch before use Frequency of routine couch cleaning will depend on exact application CLEANING DECONTAMINATION Locate couch in
10. s to Raise Couch Press to Lower Couch Press to Raise Backrest O Press to Lower Backrest V Handswitch Press 1 to Raise Backrest Ps Press to Lower Backrest es Backrest Angles 75 0 3 Seat Angles 15 0 3 AKRON 9932 3 Operation Wheel Locking WARNING Lock couch wheels castors prior to patient transfer or treatment Disconnect and coil electrical cord away from moving parts prior to repositioning couch Wheels Locked Directional Lock Wheels Unlocked Seat Extension CAUTION Prior to patient positioning ensure seat extension if used is secure To remove seat extension A Loosen locking knob on underside B Slide seat extension horizontally away from couch To refit seat extension A Locate seat extension into tubes B Secure by turning locking knob on underside clockwise to tighten AKRON 9932 3 Operation Leg Supports CAUTION Ensure all locking knobs and levers have been sufficiently tightened before using leg supports To adjust leg support A Slide din rail clamp onto din rail in desired position secure by tightening lever B Locate leg support into din rail clamp and secure in position using locking knob C To adjust angle of leg support pad loosen lever on underside of pad adjust to desired position and secure by tightening lever D If further angle adjustment required loosen T lever by turning anti clockwise rotate l