Home

EndoStim User Manual

image

Contents

1. Maintenance The Programmer Wand does not contain any user serviceable parts If any Programmer Wand parts become damaged or loose or 1t does not function properly return the Programmer Wand to EndoStim B V After each use it is recommended that you disconnect the Programmer Wand from the USB port of the laptop computer To clean the Programmer use a soft cloth dampened with distilled water methanol or isopropyl alcohol to wipe the exterior case of the Programming Wand Do not use solvents or cleaning cloths infused with chemical cleaning agents Warnings e The Programmer is not protected against the ingress of water IXPO Avoid immersing the Programmer in any fluids e Do not use the Programmer in the presence of flammable anesthetics e Do not sterilize the Programmer e Do not connect any other equipment to the Programmer e Do not modify the Programmer in any way Handling and Storage Do not use the Programmer if the package is damaged or if the Programmer has been dropped from a height of 1 meter or more Return damaged packages to EndoStim B V The Programmer environmental conditions are found in Table 1 Table 1 Environmental Conditions for the Programmer Condition Shipping Operating and Storage Temperature 20 70 C 4 158 F 5 37 C 41 104 F Humidity 15 93 non condensing 15 93 non condensing Atmospheric 54 0 101 3 kPa 7 8 14 7 54 0 101 3 kPa 7 8 14 7 Pressure psi psi Altitude 0 500
2. Range and Tolerance of Displayed Values oooo o o 51 Service and WaALranty ccccccscecscscscscscscscscscececececccceceees 52 V EndoStim Lower Esophageal Sphincter Stimulation System List of Acronyms and Abbreviations Radio Interference Magnetic Resonance Imaging Radio Frequency T2DM Type 2 Diabetes Mellitus TENS Transcutaneous Electrical Nerve Stimulation vi EndoStim Lower Esophageal Sphincter Stimulation System Device Description Contents of all Packages Model 1002 EndoStim LES Stimulator Package e EndoStim Implantable Pulse Generator e Torque wrench Allen type e Set peel off labels Model 1003 EndoStim Implantable Bipolar Lead Package e 1 Implantable bipolar lead e 2 Silicone stopper discs e 1 Set of peel off labels Model 1007 EndoStim LES Programmer System USB e Model 1504 EndoStim LES Programmer USB Wand e Model 1505 EndoStim LES Programmer Software Accessories e Magnet EndoStim Lower Esophageal Sphincter Stimulation System 1 Indications and Contraindications Indications The EndoStim LES stimulation system is intended for the treatment of patients with chronic Gastroesophageal Reflux Disease GERD with symptom duration of 6 months or longer Contraindication The EndoStim Lower Esophageal Sphincter Stimulation System is contraindicated for individuals with the following conditions or needs e Significant cardiac arrhythmia or ectopy or significant car
3. damage will occur Please note that any damage to the IPG may not be immediately detected In addition the electromagnetic fields generated by some therapeutic machines as part of the energy steering process can affect the operation of the IPG The effects of radiation therapy can range from temporary disturbance to permanent damage Therefore if such therapy is used protect the IPG with local radiation shielding and monitor its performance during and after treatment If tissue near the implant site must be irradiated it may be advisable to relocate the IPG 8 EndoStim Lower Esophageal Sphincter Stimulation System Magnetic Resonance Imaging MRI The EndoStim System is MR conditional and high specific conditions must be followed to ensure patient safety This includes using a 1 5 Tesla 64 MHz MR system and a transmit receive RF local coil e g head coil ONLY Other conditions must be followed to ensure patient safety Failure to carefully follow all labeling conditions may result in serious patient injury and or damage to the EndoStim System Lithotripsy Direct exposure of an IPG to lithotripsy shock waves can cause damage to the IPG If the implant site is outside of the shock wave path no clear contraindication to the use of lithotripsy can be established AS a precaution programming the IPG to the OFF mode reduces the possibility of adverse effects Check the IPG for proper operation immediately following the p
4. transmit receive RF local coil as shown in grey Under no circumstances is the transmit receive RF head coil allowed to overlap with the off limits area as shown in Figure 17 39 EndoStim Lower Esophageal Sphincter Stimulation System Figure 19 MR Imaging Region Knee and Ankle with Local RF Coil If these instructions are not followed higher than normal heating may occur at the lead electrodes If the patient has a broken lead higher than normal heating can also occur at the break if in the area of the transmission of the RF energy Excessive heating can cause tissue damage or serious patient injury Consult the MR system manufacturer for questions regarding the type of RF coil in use with the 1 5 Tesla 64 MHz MR system MRI Parameters The RF frequency must be 64 MHz ONLY Other frequencies have not been tested These could lead to tissue damage or serious patient injury Furthermore use of other transmit frequencies could cause device damage and or excessive heating The RF power should be limited to a specific absorption rate of 1 W kg or less in the head and whole body for each pulse sequence The spatial gradient should be limited to 250 Gauss cm or less 40 EndoStim Lower Esophageal Sphincter Stimulation System The isocenter location landmark should be the region of interest ONLY Preparing the Patient prior to MRI Procedure The patient should be informed to bring the Patient ID card to the MRI procedure appointm
5. 5 Tesla indicated that this is not a concern for the EndoStim System Warnings and Precautions EndoStim Programmer System The EndoStim Programmer contains ferromagnetic materials These materials could make the Programmer susceptible to the powerful static magnetic field of the MR system Therefore the EndoStim Programmer System is considered MR Unsafe meaning that it cannot be present and or used in the same room as the MR system The pre and post exposure implanted pulse generator programming requirements detailed below must be performed outside of the MR system room 36 EndoStim Lower Esophageal Sphincter Stimulation System Other Implanted Systems Prior to MRI examination the patient should be evaluated for other implanted medical devices especially electronically activated devices An x ray may be used to determine if the patient has other implanted medical devices The decision to perform an MRI examination must be made by the MRI trained radiologist and the patient s physician in consideration of the risks and benefits to the patient MR System Type A 1 5 Tesla 64 MHz MR system is the only type of equipment permitted for use in a patient with the EndoStim System There is no restriction on the MRI manufacturer Other MR systems e g 0 6 Tesla 3 Tesla etc are not permitted for use and may cause serious patient injury and or damage to the EndoStim System MR System RF Coil Use of Transmit Receive RF L
6. 80Hz 5Hz Active Time 00 00 01 to 1sec 2sec 23 59 59 Inactive Time 00 00 01 to 1sec 2sec 28 59 59 Duty Cycle 1 to 99 Dose Mode Parameters Parameter Range Increment Tolerance 00 00 01 to Block Time 0 5 to 4 0 hr Dose Schedule 00 00 to 23 30 Sensing Parameters Parameter Range Increment Tolerance Supine Time 1 5 30 or 60 N A 1 min min Supine Level 50 to 80 70 80 90 or 95 N A 1 Supine Time 70 80 90 or 95 N A 1 Yo 51 EndoStim Lower Esophageal Sphincter Stimulation System Service and Warranty EndoStim provides emergency device consultation on a 24 hour basis If you require emergency assistance please contact the following number 1 866 510 1003 EndoStim warrants that all IPGs and accessories including associated firmware and software will be free from defects in workmanship and materials for a period of 12 months after the original implantation of the IPG the Warranty Period If 1t appears that an IPG contains a defect in workmanship or materials or fails to conform to applicable specifications EndoStim will replace the defective or non conforming components free of charge The Warranty Period for a replaced component shall be the longer of the time remaining on the original Warranty Period or nine months from delivery of the replaced item If a system component completely fails to function within the first 72 hours of operation EndoStim will replac
7. Condition Condition Condition Condition 1 2 3 4 Fana 215 usec 215 usec 215 usec 215 usec e en 20 Hz 20 Hz 20 Hz 20Hz a 20y 2 0 V 1 6 V 7 5V se 28 5 0 2 5 24 uo ee 4000 2000 7500 pu ae 13 8 years 11 2years 12 6 years 1 5 years This is the recommended value Recommended to never exceed 0 01 Lead Impedance This assumes a 12 month shelf life prior to implant Recommended to never program continuous stimulation 24 hours outside the clinic environment Lead Characteristics and Specifications Table 4 Lead Physical Description and Materials Lead Specification Connector IS 1 BI Length 45 cm Mass 49 Electrodes Platinum iridium At 35 cm lead has two 10cm Bifurcation unipolar segments 44 EndoStim Lower Esophageal Sphincter Stimulation System Programmer Specifications Table 5 Programmer Specifications Description Value Height 140 mm Width 62 7 mm Thickness 30 5 mm Mass 251g Storage temperature 5 C to 37 C Safety Shutdown with a Magnet Therapy signal delivery of the IPG can be shut down for 24 hours if it is exposed to a magnetic field A standard pacemaker safety magnet can be used Electromagnetic Interference The center frequency of the EndoStim Programmer transmitter to the EndoStim IPG is 20 kHz with a bandwidth of 18 kHz to 22 kHz The center frequency of the EndoStim Programmer receiver from the EndoStim IPG is 10 kHz to 28 kHz 45 EndoStim Lower Esophageal Sphincter Stimulation System
8. Dialog 1 Verify that the System Current Time matches the PC time 2 Select Get Time The message Get Time OK appears Select Set With PC Time 4 If Daylight Savings Time is required select DST Enable and the Start date and Stop date 5 Select Close Note It is important to ensure that the time of day is always the same between the IPG and the Programmer If the time or clock is not synchronized the patient may receive therapy that is off schedule For example if there is a difference of 1 hour between the IPG and Programmer the patient may receive therapy during a meal 28 EndoStim Lower Esophageal Sphincter Stimulation System Measure Lead Impedance To measure lead impedance select the pull down menu Tools Impedance Measure There will be a slight delay while the impedance 1s being measured and the impedance will be displayed in the Lead Impedance icon aL If the impedance is too high or too low the text Out of Range appears in the icon Significant shifts in lead impedance that are not out of range may require adjustments in stimulation amplitude to accommodate for the changes Access IPG Statistics 1 To access the IPG statistics select the pull down menu Tools Read Statistics The Statistics window opens Figure 12 cms SUPINE UPRIGHT SUPINE CYCLIC SCHEDULE UPDATED SCHEDULE NOT MAGNET IMPEDANCE J DETECTED DETECTED DELIVERED DELIVERED PULSE VOLTAGE DELIVERED DEACTIVATION OUT OF RANGE 0 Figur
9. dec titusss 11 Cellular PNOMES darian 12 Airport Screening SyStems cscscscscsccceccccccscscscscececscees 12 Moenia coo 12 111 EndoStim Lower Esophageal Sphincter Stimulation System General Precaution cccceccccccscccccccccscccccccccccccccceccscccecs 12 System OvervieW isaac ii ins 13 Implantable Pulse Generator ss 13 Implantable Bipolar Stimulating Lead oooooo o o 16 Programme Sin aus 17 Connect the Programmer cccecscccscsccccccscscecsccscscececcecs 21 Maintenance nissan oies 22 Handling and StoOrage ccccccscscscecsscscsccccccscsceccccecscscees 22 Service Life and Disposal ccecsccscscecsccscsceccccscscsceececs 23 Instructions for Use Program an IPG cscscsssecscncees 24 Launch the Software usina ala 24 Initiate a Programming Session cecscscscecececcccscscscsceces 27 Prosram IPG Times ii 28 Measure Lead Impedance cscscscscscececcccccccccecececececs 29 Access IPG STATISTICS ri ia 29 Program Stimulation Amplitude and Polarity 29 Program Stimulation Session Times while Patient is Awake Ni 31 Program Stimulation Session Times while Patient is Asleep an la 32 Calibrate Posture Sensor ccsccscsccscsccsceccsceccsceccecsceeens 32 Advanced Parameters isis iia 33 Advanced Timing Parameters ccccscscscscscscscsceceseeesees 33 Advanced Stimulation Parameters csccscscsscscsscscsecees 33 Advance
10. is cremated Contact your local waste management officials for other information about the environmentally safe collection and disposal of the IPG Reuse The IPG and implanted lead are intended for single use only Environmental Hazards The following discussion reflects a conservative approach to the issue of patient safety in the presence of potential environmental hazards Design features in the IPG minimize the potential for such hazards but they cannot be ruled out completely Notes e Do not use any other electrical equipment adjacent to the EndoStim System If the components cannot be kept separate then monitor devices to assure normal operation e Portable and mobile RF radio frequency equipment can interfere with the normal operation of the EndoStim System The portable and mobile RF equipment should be considered in any situation where the EndoStim System devices are not acting as expected Other equipment may interfere with these devices even if that equipment complies with CISPR emission limits As with any medical device system all the components of the EndoStim System can be affected by magnetic electrical and electromagnetic signals of sufficient strength On rare occasions interfering signals could inhibit electrical stimulation delivery or alternatively trigger inappropriate delivery of electrical stimulation signals In addition certain sources can couple sufficient energy into the IPG to damage the circ
11. screen click Next 13 At the Completion screen click Finish e eae Powering the Programmer The Programmer is powered from the USB port of the laptop SV 0 5A Connecting the laptop to Mains 1s not allowed while the Programmer is connected Operate the Programmer using only the laptop computer battery Use the AC connection only when recharging the laptop and ensure that the USB cable is disconnected from the laptop 20 EndoStim Lower Esophageal Sphincter Stimulation System If the laptop 1s connected to Mains the Programmer software disables power to all computer USB ports and provides a popup warning Figure 7 The popup warning message terminates once the laptop 1s disconnected from the Mains and you press Retry The Programmer shuts down 1f you press Close A Mains Connected The Laptop is connected to the Mains Please disconnect it and press Retry Retry Figure 7 Programmer Connection to Mains Warning Notes e When recharging the laptop connect the laptop charger to the Mains per labeling on the laptop charger e Itis recommended that you routinely charge the battery of the laptop between uses Connect the Programmer Connect the USB end of the Programmer to the USB port of the laptop computer Note The Programmer should only be connected to the USB port of an IEC60950 certified laptop that runs on battery power only 21 EndoStim Lower Esophageal Sphincter Stimulation System
12. see Figure 14 below 26 EndoStim Lower Esophageal Sphincter Stimulation System Initiate a Programming Session To initiate a programming session place the Programming Wand over the IPG Red and green LED lights indicate 2 types of communication e Communication At least one of the Signal Strength LEDs blinks about once per second e Stronger communication Green LEDs at the top blink e Weaker communication Red LEDs at the bottom blink Press the Interrogate icon at the far left of the Information section The remaining 3 icons in the Information section of the Programmer screen should be filled in with appropriate values as shown in Figure 10 Device Model 56 Serial 237 Get DST ok Dst 7 21 20125 55 57PM Set IPG Time OK Time 7 21 2012 5 56 06 PM Interrogation OK Supine 7 21 2012 5 56 23 PM Interrogation OK Supine 7 21 2012 5 56 27 PM Impedance OK Impedance 7 21 2012 5 56 43 PM E OFF Every 2 hours v Every 3 hours Every 4 hours Other Figure 10 Information on the Main Screen 21 EndoStim Lower Esophageal Sphincter Stimulation System Program IPG Time Select the pull down menu Tools Time The Time dialog window opens Figure 11 Current Device Time GetTime 1 1 2001 12 00 42 AM Er Current PC Time 7 21 2012 3 26 37 PM Set with PC Current DST DST Enable GetDsT Start April 01 Set DST End October 01 dose Figure 11 EndoStim Programmer Time
13. to 30 minutes Supine Level This 1s the angle in degrees that the patient must be to detect a supine position Patient standing vertical is considered 0 and lying down horizontal is 90 It is recommended to set this to 70 Minute This is the percentage of 1 minute segments within the Supine Time that the patient remains still and in the Supine Level It is recommended to set this to 90 Supine Time This is the percentage of time within Iminute that the patient remains still and in the Supine Level It is recommended to set this to 90 EndoStim Lower Esophageal Sphincter Stimulation System Terminate Stimulation To stop stimulation program the amplitude to OFF on the amplitude icon Wi and then select the Program button You can also use the Urgent Programming icon to stop stimulation at any time Reset the IPG To reset the IPG select the pull down menu 7oo s Reset IPG Shut Down the Programmer To turn off the Programmer e Click on the X at the top right of the screen OR select the pull down menu File Exit e Disconnect the USB cable from the computer e Turn the computer off Implant the EndoStim System Laparoscopic surgery is generally used for the implantation of the EndoStim System An open bariatric surgical approach may also be appropriate for some patients Minimal use of narcotics is recommended Follow standard sterile technique and adhere to standard operating room p
14. 0 m 0 16 404 ft 0 5000 m 0 16 404 ft 22 EndoStim Lower Esophageal Sphincter Stimulation System Environmental conditions for the laptop may be different than the Programmer Check laptop user manual for laptop environmental ranges and assume worst case condition Service Life and Disposal The life of service 1s expected to be 3 years The Programmer should be returned to EndoStim B V when disposal is required Environment of Use and Operator Profile The operators of the Programmer include physicians in charge of either implanting or monitoring an IPG and trained medical personnel who assist physicians Operators will be familiar with operation of electronic medical equipment particularly IPGs and programmers and or the operators will have been trained on the operation of the EndoStim Programmer The Programmer is used in an operating room where the IPG is being implanted When in the operating room the Programmer should be brought into the sterile field through a sterile sleeve while the laptop remains outside the sterile field The surgeon should place the Programmer over the IPG while an appropriate person outside the sterile field operates the Programmer The Programmer is also used in a clinical room where patients with an implanted IPG are monitored When in the clinic the clinician should place the Programmer over the IPG and have either an assistant or the patient hold the Programmer in place The patient can sit or l
15. DESCRIPTION Interrogate Click on this icon to interrogate the IPG Serial Number The serial number of the IPG is displayed here Battery Voltage The battery voltage of the IPG is displayed here Lead Impedance The lead impedance is displayed here 25 EndoStim Lower Esophageal Sphincter Stimulation System e Programming The lower section below the solid horizontal line contains icons for programming the device parameters These are labeled below ICON DESCRIPTION Stimulation Amplitude and Polarity OFF FE Click on the top button of this icon F to JL program the stimulation amplitude Click on the lower icon Jl to reverse polarity Program Click on this icon to program the IPG Urgent Programming The battery voltage of the IPG is displayed here Stimulation Session Timing Awake Click on the time icons below this icon to schedule stimulation sessions while the patient is awake Stimulation Session Timing Asleep Click on the options below this icon to schedule stimulation therapy while the patient is sleeping NOTE The Programmer automatically disables certain parameters and icons based upon values of other parameters or states of other icons For an icon example the Program icon will not be active unless a parameter was changed For a parameter example the Stimulation Session Timing Awake will not allow for programming two stimulation sessions immediately adjacent to each other
16. Electromagnetic Emissions Guidance and manufacturer s declaration electromagnetic emissions The EndoStim IPG and Programmer is intended for use in the electromagnetic environment specified in Table 6 The customer or user of the EndoStim IPG and Programmer should assure that it is used in such an environment Table 6 Electromagnetic Emissions The EndoStim IPG and Programmer are intended for use in the electromagnetic environment specified below The customer or user of EndoStim IPG and Programmer should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance The EndoStim IPG and Programmer use RF energy only for its internal function Therefore their RF emissions are very low and are not likely to cause any interference in nearby electronic equipment CISPR 11 Programmer are suitable for Harmonic use in all establishments other Emissions than domestic establishments IEC 61000 3 2 Not and those directly connected Voltage Applicable to the public low voltage Fluctuations Flicker PP supply network that supply Emissions buildings used for domestic IEC 61000 3 2 purposes RF emissions CISPR 11 46 EndoStim Lower Esophageal Sphincter Stimulation System ESD and Power Fluctuations Guidance and manufacturer s declaration electromagnetic immunity The EndoStim Programmer is intended for use in the electromagnetic environment specified in Table 7 The custome
17. EndoSt m Lower Esophageal sphincter Stimulation system Clinician Manual Revision B C q 0344 1588 EndoStim is a registered trademark of EndoStim B V Read all accompanying documentation before using the device al EndoStim B V Prinsessegracht 20 2514 AP The Hague The Netherlands The information contained in this document may be changed without notice No part of this manual may be reproduced or transmitted in any way or by any means either electronic or mechanical for any purpose without the express written consent of EndoStim B V EndoStim B V 2013 All Rights Reserved Explanation of Symbols on Labels DESCRIPTION Manufacturer m YYYY MM Date of Manufacture Conformit Europ enne 0344 Notified Body Number for 0344 AIMDD Conformit Europ enne 0344 Notified Body Number for AIMDD 0344 1588 1588 Notified Body Number for R amp TTE Consult instructions for use Do Not Use if Package is Damaged or Opened Transport Temperature Limits STERILE EO Sterilized with Ethylene Oxide gt lt Use By YYYY MM DD EndoStim Lower Esophageal Sphincter Stimulation System 1 SYMBOL DESCRIPTION Part Number LOT XX Lot Number XXXX Serial Number 2 Torque Wrench O Intentional Radiator Signal Strength Indicator USB Connector MR Conditional MR Unsafe 11 EndoStim Lower Esophageal Sphincter Stimulation System TABLE OF CONTENTS Explanation of Symbol
18. ce related issues to the implanting physician and EndoStim 42 EndoStim Lower Esophageal Sphincter Stimulation System System Specifications This section contains specifications and characteristics for all the devices included in the LES system IPG Bipolar lead and Programmer IPG Specifications Table 2 Implantable Pulse Generator Specifications Description Value Height 65 mm Width 48 mm Thickness 12 mm Mass 50 g Biocompatible materials in Titanium contact with human tissue Epoxy resin Silicone rubber set plugs Power source Lithium carbon monoflouride battery Storage temperature 20 C to 25 C IPG Battery Characteristics and Specifications The battery voltage at beginning of life is approximately 3 3V and the usable capacity is 2 5 Ah When battery voltage falls below 2 5V the device sends information to the Programmer upon interrogation that the battery 1s near the end of life In addition stimulation amplitude will be reduced if programmed at greater than 3 times the battery voltage When battery voltage falls below 2 3V the device will turn off all stimulation outputs When the battery voltage falls below 2 1V communication with the Programmer will no longer be possible The longevity of the battery can be estimated based on the different stimulating parameters applied as described in Table 3 43 EndoStim Lower Esophageal Sphincter Stimulation System Table 3 Estimated IPG Battery Longevity Parameter
19. ction with the IPG The lead has an IS 1 BI connector at the proximal end for attachment to the IPG Figure 4 The lead is sterilized with ethylene oxide The lead delivers stimulation pulses to the tissue through stitch electrodes at the distal end During implantation the stitch electrodes of the leads are sutured into the LES and secured into place The electrodes of the bipolar lead are implanted in the sero muscular layer of the LES Figure 5 Figure 4 EndoStim Lead and Figure 5 Implanted EndoStim IPG Lead and IPG Implants for Surgery Cardiac Pacemakers Part 3 Low profile connectors IS 1 for implantable pacemakers ISO 5841 3 2000 E 16 EndoStim Lower Esophageal Sphincter Stimulation System AAA Physical Characteristics The implantable lead is constructed of biocompatible materials an inner and outer silicone rubber sheathing cobalt nickel conductors and platinum iridium stitch electrodes A curved stainless steel suture needle 1s affixed to the end of each stitch electrode The lead is 45 cm in length from the IS 1 BI connector tip to the end of the platinum iridium electrodes The lead bifurcates after 35 cm into individual unipolar leads with the length of each unipolar lead being 10 cm Handling and Storage Do not implant the lead if the package is damaged or if the lead has been dropped from a height of 30 cm or more Return damaged packages to EndoStim B V Store the lead at a temperature rang
20. d Sensing ParaMmeters cccccscscscscscscscscecereceeees 34 Terminate Stimulation ccccsccscscsccsccccsceccsceccsceccececs 35 Reset the IPG iia ad 35 Shut Down the ProgrammMeP cecsccscscecsccscsceccccscscsceccecs 35 Implant the EndoStim System 35 MRI Information for the EndoStim System 35 Potential Interactions with the EndoStim System 36 Warnings and PrecautiOns ccccscscscscscscscscscscececececceees 36 Preparing the Patient prior to MRI Procedure 41 Beginning of the MRI Procedure sense 41 During the MRI Procedure cccscscsscscscessccscscececcscscscees 42 1V EndoStim Lower Esophageal Sphincter Stimulation System After the MRI Procedure ccccccccscccccccccccccccccccsccscees 42 System Specifications ceceecsceccccccsceccccccscececcecscecees 43 IPGSDECITICATONS net onda 43 IPG Battery Characteristics and Specifications 43 Lead Characteristics and Specifications cscscscsscscseees 44 Programmer Specifications ccsccccccscscecsccscsceccccscscscees 45 Safety Shutdown with a Magnet 45 Electromagnetic Interference scsccscscecsccscscsceccecscscees 45 Electromagnetic Emissions ccscscsscscscecsccscscececcecscscees 46 ESD and Power Fluctuations ss 47 Radiated Electromagnetic Fields 48 Recommended Separation Distances cscscccsececececececs 50
21. diovascular disease e Pregnant or nursing The following treatments are contraindicated for patients implanted with the EndoStim device e Medical diathermy e Transcutaneous Electrical Nerve Stimulation TENS exposure in the abdominal region Precautions The EndoStim Lower Esophageal Sphincter Stimulation System has not been evaluated in the following populations Patients with the following conditions should be considered prior to implant e Large greater than 3 cm hiatal hernia that has not been repaired e Severe Grade D esophagitis e Long segment Barrett s esophagus or Barrett s esophagus with dysplasia e Significant uncontrolled autoimmune disorder such as Scleroderma Dematomyositis CREST syndrome Sjogren s Syndrome or Sharp s Syndrome that effects esophageal motility 2 EndoStim Lower Esophageal Sphincter Stimulation System Severe obesity that in the opinion of the surgeon increases the risk of a laparoscopic implant procedure Type 1 diabetes mellitus Uncontrolled type 2 diabetes mellitus T2DM defined as HbA Ic gt 9 5 in the previous 6 months or has T2DM for more than 10 years Suspected or confirmed esophageal or gastric cancer Esophageal or gastric varices Dysphagia due to severe esophageal peptic structure excluding Schatzk1 s ring History of any advanced malignancy History of previous surgery that directly involves the lower esophageal sphincter or the EndoStim electrode implant site such a
22. e 12 EndoStim Programmer Statistics Window 2 The statistics window displays the timestamp and total number of stimulation therapy sessions delivered and those not delivered 3 To reset the statistics select the pull down menu Tools Clear Statistics Program Stimulation Amplitude and Polarity OFF l Select the upper icon inside Stimulation or Amplitude and Polarity icon Ba This will provide a popup menu of stimulation amplitude options Figure 13 29 EndoStim Lower Esophageal Sphincter Stimulation System Figure 13 Stimulation Amplitude Options Select amplitude from the menu and click the Program icon a 2 Select the lower icon tL inside stilato Amplitude and Polarity icon CA This icon inverts and changes the polarity of stimulation between the re electrodes Click the Program icon Mo 30 EndoStim Lower Esophageal Sphincter Stimulation System Program Stimulation Session Times while Patient is Awake 1 Click on the desired stimulation time boxes for the patient The boxes fill followed by an X in boxes below indicating that therapy 1s not possible at those times Figure 14 abbreviated 2 Click on the Program icon Figure 14 Stimulation Session Times while Patient is Awake Example is for stimulation sessions at 08 30 12 30 and 20 30 31 EndoStim Lower Esophageal Sphincter Stimulation System Program Stimulation Session Times while Patient is Asleep The IPG includes a s
23. e of 20 C to 25 C and ina dry location Do not allow the packaging to become damp or wet because sterilization of the contents may be compromised Re Sterilization and Re Use The lead is a single use device Do not re sterilize the lead or silicone stoppers Do not re use an explanted lead the device must not be used for implantation in another patient Programmer General Description Use the EndoStim LES System Programmer to interrogate and program the IPG The Programmer software runs on an IEC60950 certified laptop personal computer PC which runs on battery power Communication between the Programmer and the IPG is accomplished with the Programmer placed directly over the implant site The Programmer communicates via magnetic induction telemetry with the IPG implanted in the patient 17 EndoStim Lower Esophageal Sphincter Stimulation System Note Avoid skin contact between the patient s skin and the Programmer Wand when possible to avoid any potential cross contamination from previous use of the Programmer The Programmer can e Read interrogate IPG parameters as currently programmed e Modify IPG parameters e Retrieve statistics accumulated by the IPG as it operates e Log the activity of the IPG e Store standard programs for future use The Programmer has 2 components Figure 6 e Programmer Wand e Programmer Software Figure 6 EndoStim LES System Programmer 18 EndoStim Lower Esophageal Sphincter Stimulati
24. e the failed item with a new one EndoStim shall not be liable under this warranty if testing and examination discloses that the alleged defect or non conformity in the system component does not exist or was caused by the end user s misuse neglect improper implantation or testing unauthorized attempts to repair or by accident fire lightning or other hazard 52 EndoStim Lower Esophageal Sphincter Stimulation System 53 EndoStim Lower Esophageal Sphincter Stimulation System EndoSt m EndoStim B V Prinsessegracht 20 2514 AP The Hague The Netherlands 1 866 510 1003 Internet www endostim com Part Number CM 04 ENG Rev B
25. ensor that detects when the patient is sleeping A repeating stimulation session can be programmed while the patient is sleeping This therapy will automatically terminate when the IPG detects that the patient is awake and upright To enable stimulation sessions when the patient is asleep click on the desired frequency of stimulation sessions Figure 15 If no therapy is required for when the patient is y asleep then click OFF Click on the Program icon OFF Every 2 hours Every 3 hours Every 4 hours Other Figure 15 Stimulation Session Times while Patient is Asleep Example is for stimulation sessions once every 3 hours starting when patient goes to sleep Calibrate Posture Sensor With the patient standing upright and not moving select the pull down menu Tools Set Stand Vector This calibrates the sensor to the upright position 32 EndoStim Lower Esophageal Sphincter Stimulation System Advanced Parameters If more therapy flexibility is required more options can be found by selecting the pull down menu Tools Advanced Figure 16 Timing Parameters Stimulation Parameters Sensing Parameters Dose Time f Ohs 3 30 min Pulse Width 215 us Supine Time 30 min Block Time Frequency Supine Level 70 60 min 20 0 Hz 70 Minute 90 imp Tracker y Recommended Supine Time 90 On Time 0 hs 30 min Os Max Volt 30V Off Time 2 hs 30 min Os Min Volt 0 5V z 2 Figure 16 Advanced Stimulation Parameters Adva
26. ent and to inform the MR system operator that he she has an active implantable medical device prior to entering the MR system room The patient should be informed that the therapy will be turned off until after the MRI procedure is completed The IPG should be programmed to OFF reference section Program Stimulation Amplitude and Polarity Important Note If the IPG is in End of Life reference section IPG Battery Characteristics and Specifications then the IPG mode cannot be programmed or confirmed via telemetry and the MRI procedure cannot be performed The patient should be informed of all risks associated with undergoing an MRI procedure with a transmit receive RF local coil only reference the Warning and Precautions section in this chapter The patient should be informed to immediately inform the MR system operator of any discomfort stimulation shocking or unusual sensation that may occur during the MRI procedure The patient should be conscious throughout the MRI examination 1 e no sedation or anesthesia is allowed Beginning of the MRI Procedure Verify the patient s weight is entered correctly into the MR system s console to ensure that the regional SAR is correctly calculated Verify that the MR system s transmit receive RF local coil is being used Verify that the MR system transmit receive RF body coil is not being used 41 EndoStim Lower Esophageal Sphincter Stimulation System Verify the MRI pa
27. er frequency range applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the EndoStim Programmer is used exceeds the applicable RF compliance level above the EndoStim Programmer should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as relocating the EndoStim Programmer Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m 49 EndoStim Lower Esophageal Sphincter Stimulation System Recommended Separation Distances Recommended separation distances between portable and mobile RF communications equipment and the EndoStim Programmer The EndoStim Programmer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled Table 9 The customer or user of the EndoStim Programmer can help prevent electromagnetic interference by maintaining a minimum distance between portable and m
28. ery inhibition Instruct patients to walk at a normal pace and avoid lingering when passing through the entrances and exits of these establishments Industrial Machinery High voltage power lines electric arc welders electric smelting furnaces and power generating equipment can interfere with the operation of the IPG For this reason the intensities and modulation characteristics of the electromagnetic fields encountered by patients as a result of their occupation and lifestyle should be considered When appropriate give specific warnings Radio Transmitters Communications equipment such as radio and TV transmitters including amateur ham transmitters microwave transmitters and CB transmitters with high power linear amplifiers and radar transmitters can interfere with the operation of the IPG Discuss this with patients whose occupations may expose them to these electromagnetic fields When appropriate give specific warnings 11 EndoStim Lower Esophageal Sphincter Stimulation System Cellular Phones Cellular and other portable telephones can interfere with the operation of the IPG Potential effects may result from either the radio frequency emitted by these telephones or the magnet within the phone s speaker These effects may include inhibition or inappropriate triggering of electrical stimulation delivery when the phone is in close proximity within 25 cm to the IPG and the lead Advise patients to hold the phone t
29. es as far from the IPG and lead as possible The TENS electrodes should also be placed as close as possible to each other to reduce current spread The operation of the IPG should be monitored closely during TENS use As a precaution programming the IPG to the OFF mode reduces the possibility of adverse effects Contraindication Patients and physicians should be advised that exposure to TENS in the abdominal region is contraindicated for patients implanted with the EndoStim System Home Appliances Home and commercial microwave ovens in good condition and used as intended will not affect the IPG Even a defective oven that exposes the IPG to direct microwave energy may not damage the unit itself Ovens using electromagnetic induction can cause the device to go into magnet mode disable stimulation therapy output 10 EndoStim Lower Esophageal Sphincter Stimulation System Inform patients about the possibility of interference from some electric razors electric power tools and electrical ignition systems including those used on gasoline powered devices In general patients who have an IPG may operate gasoline powered devices if protective hoods shrouds and other shielding remain in place Antitheft Systems Certain types of antitheft devices such as those used at entrances exits of retail stores libraries and other establishments can interfere with the IPG Most commonly interference may result in electrical stimulation deliv
30. ie down The laptop should be outside the patient environment at least 1 5 meters away and operated by the clinician To be compliant with electromagnetic compatibility requirements interference characteristics the Programmer should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary verify that the Programmer functions properly It is recommended to have an additional Programmer present at each implant and clinic visit 23 EndoStim Lower Esophageal Sphincter Stimulation System Instructions for Use Program an IPG Launch the Software Double click the EndoStim icon on the desktop to launch the Programmer software The password screen appears Figure 8 o E Enter Password Figure 8 EndoStim Programmer Password Screen Enter the password and press Accept or press Enter The Programmer Main Screen appears Figure 9 a Endostim NeuroSt Fil ols wo a a Es jm OFF 8 20 OFF 9 21 Every 2 hours 10 12 Every 3 hours Ll 41 D Every 4 hours na Other 17 5 18 6 A y EndoSt m Figure 9 EndoStim Programmer Main Screen 24 EndoStim Lower Esophageal Sphincter Stimulation System The screen is divided into two sections Information and Programming e Information The top section above the solid horizontal line contains icons for retrieving and displaying device information These are labeled below ICON
31. lation can cause the IPG to revert to its DOWN mode with possible loss of statistics data If sufficient energy 1s coupled into the system the unit may be damaged Application of RF ablation in close proximity to the electrodes of an implanted IPG can also cause a direct coupling of radio frequency energy through the leads and electrodes to the LES tissue possibly producing burns 6 EndoStim Lower Esophageal Sphincter Stimulation System If RF ablation is necessary position the ground plate as far from the IPG and lead as possible Avoid direct contact with the ablation catheter and the IPG and lead Program the IPG to the OFF mode to reduce the possibility of adverse effects Check the IPG for proper operation immediately following the procedure If the unit is found in the DOWN mode follow the Reset procedure Medical Diathermy Medical diathermy short wave thermal induction is generally contraindicated for patients implanted with active medical de vices It is unknown how the IPG will operate if subjected to the intense fields of energy Although damage to either the circuitry of the IPG or LES tissue is improbable it cannot be ruled out If medical diathermy is required apply it away from the immediate vicinity of the IPG and lead Programming the IPG to the OFF mode reduces the possibility of adverse effects Check the IPG for proper operation immediately following the procedure If the unit is in the DOWN mode follow the Rese
32. nced Timing Parameters e Dose Time This sets the duration of the stimulation session It is recommended to set this to 30 minutes e Block Time This sets the duration of time after a stimulation session where another stimulation session cannot occur for any reason It is recommended to set this to 60 minutes e Active Time This sets the duration of the stimulation session It is recommended to set this to 30 minutes e Inactive Time This sets the duration of the time between consecutive stimulation sessions Advanced Stimulation Parameters e Pulse Width This sets the width of the stimulation pulse It is recommended to set this to 215 usec e Frequency This sets the rate or frequency at which the pulses are delivered It is recommended to set this to 20 Hz e g 20 pulses per second Impedance Tracker When enabled this automatically adjusts the stimulation voltage to accommodate changes in lead impedance This ensures the programmed current is delivered independent of lead impedance Parameters Max Volt Min Volt and Max A Volt place boundaries on the amount that changing lead impedance can change stimulation voltage Click on the Recommended icons to use suggested parameters 33 EndoStim Lower Esophageal Sphincter Stimulation System Advanced Sensing Parameters 34 Supine Time This 1s the amount of time the patient must maintain the Supine Level and be still It 1s recommended to set this
33. o the ear opposite the side of the implanted IPG Patients should not carry the phone in a breast pocket or on a belt over or within 25 cm of the implanted IPG because some phones emit signals when they are turned on though not in use Airport Screening Systems Globally passenger screening systems encountered in airports may interfere with the IPG Most commonly interference may result in electrical stimulation inhibition Instruct patients to inform security personnel about the implanted medical device to show their identification card and to walk at a normal pace when passing through the portal of these systems No damage to the implanted system should occur Magnets Application of a magnet directly over the IPG for at least 2 consecutive seconds followed by 2 consecutive seconds without a magnet will terminate therapy for 24 to 48 hours General Precaution The EndoStim System may fail to properly operate for any number of reasons including but not limited to random component failure including the battery lead failure including electrical shorts opens and insulation faults and software errors The frequency of these events cannot be predicted 12 EndoStim Lower Esophageal Sphincter Stimulation System System Overview This section provides descriptions of all the components for the Model 1008 EndoStim Lower Esophageal Sphincter Stimulation System Throughout the instructions in this manual you may see the terms
34. obile RF communications equipment transmitters and the EndoStim Programmer as recommended according to the maximum output power of the communications equipment Table 9 Recommended Separation Distances Between RF Equipment and EndoStim Programmer Rated Separation distance according to maximum frequency of transmitter output 150 kHz to 80 MHz to 800 MHz to power of 80 MHz transmitter 7 117 P d 0 35VP d 0 70VP Notes For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated by using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer e At 80 MHz and 800 MHz the higher frequency range applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 50 EndoStim Lower Esophageal Sphincter Stimulation System Range and Tolerance of Displayed Values IPG and Lead Parameters Parameter Range Increment Tolerance 0001 to 9999 2 10 to 3 10 V 0 01V 0 05V Max of Impedance 200 to 2000 Q 10 1000 and 20 Stimulation Parameters Parameter Range Tolerance Increment Pulse Width Max of 5 and 15usec Amplitude 0 5mA and 20 Hz Pulse Rate 2 to 80 Hz 2 10Hz 1Hz Max of 1 10 40Hz and 2 Hz 100usec 40
35. ocal Coil ONLY A transmit receive RF local coil is the only type of RF coil permitted for use in a patient with the EndoStim System The transmit receive RF body coil must not be used No portion of the RF transmit receive local coil can cover any part of the implantable system The diagram below illustrates the region that is considered unsafe for MRI exposure 37 EndoStim Lower Esophageal Sphincter Stimulation System ey pa LEAD ms a J NAH EXCLUSION ZONE A I PULSE L GENERATOR A q tw F W Si N AA Figure 17 Unsafe MR Imaging Region highlighted in grey This exclusion zone is dependent on the typical placement of the EndoStim IPG and Lead Implanted system components may be located as close as Ocm to the lower caudad edge of the transmit receive RF local coil but no part may be inside of it Brain head imaging should be performed using a transmit receive RF head coil only The diagram below illustrates the required placement of the transmit receive RF head coil as shown in grey Under no circumstances is the transmit receive RF head coil allowed to overlap with the off limits area as shown in Figure 18 38 EndoStim Lower Esophageal Sphincter Stimulation System Figure 18 Brain MR Imaging Region with Head RF Coil Extremity imaging should be performed using an appropriate transmit receive RF local coil only The diagrams below illustrate several examples of the required placement of the
36. of 30 cm or more Return damaged packages to EndoStim B V Store the IPG at a temperature range of 20 C to 25 C and in a dry location Do not allow the sterile packaging to become damp or wet because sterilization of the contents may be compromised Exposure to temperatures below 0 C may cause a change in the parameter values of the IPG to the DOWN mode no output Once the IPG is removed from the extreme environment it will either return to the programmed settings or it will remain in the DOWN mode In the latter case follow the Reset procedure If unsuccessful the unit should be returned to EndoStim B V Re sterilization and Re Use The IPG and torque wrench are single use devices Do not re sterilize the IPG or torque wrench Do not re use an explanted IPG the device must not be used for implantation in another patient X Ray Identification A radiopaque marker placed within the IPG allows the model number and year of manufacture to be identified by normal X ray techniques The manufacturer s identification code for EndoStim is E the code for the EndoStim IPG model is A and the code for the year of manufacture is XX where XX is replaced by the last two digits of the calendar year in which the device was manufactured 15 EndoStim Lower Esophageal Sphincter Stimulation System Implantable Bipolar Stimulating Lead General Description The EndoStim lead called the Implantable Bipolar Stimulating Lead is used in conjun
37. on System Electrical Characteristics The Programmer is internally powered Type B equipment that is suitable for continuous use The Programmer is considered an applied part and the Programmer laptop 1s considered out of the patient environment at least 1 5 meters from the patient 19 EndoStim Lower Esophageal Sphincter Stimulation System Loading the Software You must install the Programmer software on an IEC60950 certified laptop that has Microsoft Windows 7 The minimum functional requirements of the computer are the same as those required for Microsoft Windows 7 The computer must support a screen resolution of 1024 x 768 Follow these steps to install the software 1 Close all open applications 2 Insert the CD into the CD or DVD drive 3 Wait for the computer to recognize the CD and open a folder 4 Double click the icon EndoStim Neurostimulator Setup Win Click Yes 1f prompted to allow software access The computer restarts immediately At the Setup Welcome screen click Next At the Choose User screen choose the appropriate user s At the Choose Components screen ensure that all boxes are checked 10 At the Choose Install Location screen choose the desired location for the software to install 11 At the Insert Application Password screen type the password you want to use in both locations then click Accept 12 At the Installation Complete
38. r or user of the EndoStim Programmer should assure that it is used in such an environment Table 7 ESS and Power Fluctuations Electromagnetic Immunity Immunity IEC 60601 Compliance Electromagnetic test Test level level environment guidance Electrostatic 6 kV 6 kV contact Floors should be wood discharge ESD contact concrete or ceramic tile 8 kV air If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast Not Not Applicable Mains power should only transient burst Applicable be used to recharge the laptop computer battery IEC 61000 4 4 pe the unit is powered IEC 61000 4 2 8 kV air Surge Not Not Applicable Mains power should only Applicable be used to recharge the laptop computer battery while the unit is powered off Voltage dips Not Not Applicable Mains power should only short Applicable be used to recharge the interruptions and laptop computer battery voltage while the unit is powered variations on off power supply input lines IEC 61000 4 5 IEC 61000 4 11 Power frequency Power frequency magnetic fields should be at levels characteristic of a typical location in a residential environment 50 60 Hz IEC 61000 4 8 47 EndoStim Lower Esophageal Sphincter Stimulation System Radiated Electromagnetic Fields Guidance and manufacturer s declaration electromagnetic immunity The EndoStim Programmer is intended for use in the
39. radiated electromagnetic fields specified in Table 8 The customer or user of the EndoStim Programmer should assure that it is used in such an environment Portable and mobile RF communications equipment should be used no closer to any part of the EndoStim Programmer including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Table 8 Radiated Electromagnetic Fields Electromagnetic Immunity Immunity test IEC Compliance Recommended 60601Test level Separation level Distance Conducted RF IEC 3 Vrms 3Vrms 61000 4 6 150 kHz d 1 17VP to 80 MHz outside ISM bands Conducted RF IEC 3 Vrms 3Vrms EE 20VP 61000 4 6 150 kHz to 80 MHz inside ISM bands Radiated RF IEC 3 hie 3 ill Lt 61000 4 3 80 MHZ 1 20 to 800 MHz Radiated RF IEC 3 eal 3 a 61000 4 3 800 MHz oo 2 304P to 2 5 GHZ 48 EndoStim Lower Esophageal Sphincter Stimulation System Notes e Pis the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m e Field strengths from fixed radio frequency RF transmitter as determined by an electromagnetic site survey should be less than the compliance level in each frequency range e Interference may occur in the vicinity of equipment marked with the ionizing radiation symbol gt At 80 MHz and 800 MHz the high
40. rameters for compliance with the requirements in this section If the parameters are not compliant they should be adjusted to meet the requirements If this is not possible the MRI procedure must not be performed Verify that the Programmer is not in the same room as the MR system During the MRI Procedure The EndoStim System should not activate turn on or conduct any energies into a stimulation There should be no visual artifacts from the EndoStim System as it is not to be placed within the transmit receive RF local coil Monitor the patient visually and audibly during and between the MR imaging sequences to ensure the patient is not having any difficulty During the MRI procedure the patient may feel vibration of the IPG Discontinue the MRI procedure immediately if the patient reports any issues 1s unable to respond to any questions or the IPG movement or vibration is too uncomfortable Do not under any circumstances attempt to program the IPG while the patient is in the same room as the MR system After the MRI Procedure Verify that the patient has not experienced any adverse effects during the MRI procedure Once outside of the MR system room reset the IPG using the Programmer Reprogram the parameter values to those prior to the MRI procedure and enable stimulation Any problem with these programming steps indicates that the IPG may have been damaged due to the MRI examination Immediately report any devi
41. re or the patient s incision could become infected System component malfunction or other clinical circumstances e g sepsis may require noninvasive corrective actions or possibly even a surgical revision repositioning replacement or removal of the malfunctioning component s It is recommended to shut the system down if the patient experiences severe sensation or muscle or diaphragm stimulation Notes e If necessary use the Programmer to adjust the stimulation amplitude to eliminate any pain discomfort or sensation that may be related to stimulation Reducing stimulation amplitude is the only recommended adjustment After verifying the IPG is functioning as previously programmed it is recommend reducing stimulation amplitude incrementally in voltage steps of 0 3mA until the adverse effect resolves If such attempts are not successful it may be necessary to use the Programmer or a magnet to shut down the system e A dramatic increase of the lead impedance may indicate a partial or complete lead dislodgement Patient symptoms should be monitored and if deteriorating x ray or fluoroscopy is recommended to assess the lead and electrode location 4 EndoStim Lower Esophageal Sphincter Stimulation System Warnings and Precautions Cremation and Incineration The IPG contains a sealed chemical power cell lithium battery For this reason never incinerate an IPG Be sure that the IPG is explanted before a deceased patient
42. rocedure If the unit is found in the DOWN mode follow the Reset procedure Therapeutic and Diagnostic Ultrasound Direct exposure of an IPG to diagnostic ultrasound can cause damage to the IPG The IPG may inadvertently concentrate the ultrasonic field and cause harm to the patient Therapeutic ultrasound may be used if the implant site 1s distant and clearly outside of the ultrasonic field Programming the IPG to OFF mode reduces the possibility of adverse effects Check the IPG to assure operation immediately following the procedure If the unit is found in the DOWN mode follow the Reset procedure 9 EndoStim Lower Esophageal Sphincter Stimulation System Effects on Other Implanted Devices The EndoStim system may affect the operation of other implanted devices such as cardiac pacemakers implantable cardioverter defibrillators ICDs other neurostimulation systems and implantable drug infusion pumps Physical proximity may cause sensing problems and inappropriate device responses Clinicians involved with both devices should evaluate any potential interference problems before surgery Transcutaneous Electrical Nerve Stimulator Transcutaneous Electrical Nerve Stimulator TENS therapy is generally contraindicated for patients who have active implantable medical devices The high voltage pulses delivered by TENS units to the body can interfere with the operation of the IPG If a TENS device must be used place the TENS electrod
43. rocedures Warning The implanting physician is required to undergo formal training by an EndoStim expert representative prior to implanting the system in humans MRI Information for the EndoStim System It is important to read the section in its entirety before conducting any magnetic resonance imaging MRI examination on a patient implanted with the EndoStim System Contact an EndoStim representative with any questions regarding the following MRI information 35 EndoStim Lower Esophageal Sphincter Stimulation System If all the instructions below are followed MRI examinations of the regions using a transmit receive RF local coil e g head coil only may be safely performed at 1 5 Tesla 64 MHz Failure to carefully follow the specific information presented in this document may result in serious patient injury and or damage to the EndoStim System Potential Interactions with the EndoStim System MRI Related Heating The MRI uses radiofrequency RF energy which may cause excessive heating of the pulse generator and lead of the EndoStim System during the MRI examination Notes High specific absorption rate SAR RF power levels can increase the risk of heating Excessive heating of the lead may occur even if the IPG 1s explanted Magnetic Field Interactions MRI uses a powerful static magnetic field which may exert force and torque on the metallic objects that could result in movement or displacement Tests performed at 1
44. s fundoplication or an esophageal myotomy Other implanted electro medical devices Younger than 21 years old Potential Adverse Effects Events Implantation of the System Potential adverse effects events associated with the implantation of the implantable pulse generator IPG and lead include but are not limited to the following death pulmonary embolism partial or complete ileus peritonitis esophageal perforation by the electrodes infection inflammation injury to organs within the abdominal cavity intravenous site complications pneumonia bleeding incisional hernia allergic or abnormal reaction to anesthetic agents pain and fever EndoStim Lower Esophageal Sphincter Stimulation System Use of the System Additional adverse effects that could be associated with the EndoStim System include but are not limited to the following lead electrode dislodgement lead erosion or perforation into the esophagus or stomach IPG migration in the subcutaneous space IPG erosion through the skin diaphragmatic stimulation stimulation of abdominal muscle irritation and or inflammatory response to the IPG and or the lead allergic reaction to materials hematoma infection dysphagia odynophagia cardiac arrhythmia nausea and discomfort LES stimulation ceases when the battery in the IPG 1s completely discharged There is a potential that any system component could malfunction e g software bug become damaged e g lead fractu
45. s on Labels i TABLE OF CONTENTS siii riana iii List of Acronyms and Abbreviations ccscscsssssscscees vi Device Descrip naaa ais 1 Contents of all Packages ccsccscscscsccscsceccccscscececcecscscees 1 Indications and CONntraindications csceccccccscscscscsceces 2 INICIAMOS ic 2 Contral nadlca HO Mica linia 2 PFEeCAUTION Sii ia 2 Potential Adverse Effects Events ccccscsscscceccscscceees 3 Implantation of the System cscsscscscsccscscsceccccscececccsces 3 Use of the System a 4 Warnings and Precautions cccscscsccscsccccccscececcccscscees 5 Cremation and Incineration sessessessessessssescsessessessessssses 5 ROUSE sirce E 5 Environmental Hazards csccscscsccscscsccccscscecccsccscsccccccsces 5 Electrocute ecsisccscsccccssocassasteacesensscavessennsecessosacsucsensaenasess 6 READ OM moros 6 Medical Diathermy ccccsccscscessscscscsccccscsceccccccscsseccceces 7 DOTIDIINACI ON srta 8 Therapeutic Radiation ccccccccscsssccscscsccccscsceccccccsceceececs 8 Magnetic Resonance Imaging MRI sccscscsscsssscsceccsceees 9 510 DS RE E A 9 Therapeutic and Diagnostic Ultrasound cccecsecscsceeeees 9 Effects on Other Implanted Devices 10 Transcutaneous Electrical Nerve Stimulator 10 Home Appliances scarico cid 10 Antitheft Systems cirios 11 Industrial Machiner y aurons 11 Radio Transmitter ini aus items
46. stimulator and IPG implantable pulse generator These terms refer to the same type of device The system consists of an IPG bipolar stimulating lead and a Programmer Figure 1 EndoStim LES System Implantable Pulse Generator General Description The EndoStim implantable pulse generator IPG Figure 2 is an internally powered lithium battery device that delivers electrical stimulation pulses to the lower esophageal sphincter LES The IPG is hermetically sealed in a titanium case Figure 2 It delivers electrical pulses to the LES via a bipolar IS 1 BI header connector port Figure 3 that connects the IPG and lead The IPG is sterilized with ethylene oxide 13 EndoStim Lower Esophageal Sphincter Stimulation System y rod e A i o CE AAA EndoSiNm LES Stimulator Figure 2 EndoStim LES Figure 3 IPG Header Stimulator Connector The programmable IPG communicates with the external Programmer via telemetry Medical and technical personnel use the Programmer to program parameters that control the IPG function The IPG may be programmed specifically for individual patients by changing the settings parameters within the implanted device In addition the Programmer can access performance data that the IPG collects during its normal operation 14 EndoStim Lower Esophageal Sphincter Stimulation System Handling and Storage Do not implant the IPG if the package is damaged or if the IPG has been dropped from a height
47. t procedure Contraindication Patients and physicians should be advised that exposure to medical diathermy is contraindicated in patients implanted with the EndoStim System 7 EndoStim Lower Esophageal Sphincter Stimulation System Defibrillation Any implanted active medical device can be damaged by cardiac defibrillation procedures In addition the defibrillation current can cause damage to LES tissue adjacent to the electrodes and or to tissue surrounding the IPG The defibrillation current may also cause the IPG to revert to 1ts DOWN mode with possible loss of statistics data If sufficient energy is coupled into the system the unit may be damaged If defibrillation is necessary position the paddles as far away from the implanted system as possible avoid placing the IPG in the defibrillation current path between the paddles Following defibrillation closely monitor the performance of the IPG If an operational abnormality is detected consider repositioning or replacing the lead and or reprogramming or replacing the IPG If the IPG changes to the DOWN mode follow the Reset procedure Therapeutic Radiation Therapeutic equipment that produces ionizing radiation such as linear accelerators and cobalt machines used in cancer treatment can damage the type of circuitry used in most active implantable medical devices Since the effect is cumulative both dose rate and total radiation dosage determine whether and to what extent
48. uitry of the IPG and or LES tissue adjacent to the electrodes The physician may wish to discuss these risks with the patient 5 EndoStim Lower Esophageal Sphincter Stimulation System The susceptibility of a particular unit will also depend on the location of the IPG pocket the nature of the interference and the programmed operating parameters Because of the diversity of potential causes of electromagnetic interference EndoStim cannot characterize and describe within this manual the effects of all potential sources of interference Warning Advise patients to be cautious when in the vicinity of equipment that generates electrical or magnetic fields and to seek medical advice before entering an area posted with a warning for pacemaker patients or other medical implantable devices Electrocautery The surgical use of electrocautery can cause the IPG to become inactive and possibly lose statistical data Electrocautery may damage the IPG and lead Application of electrocautery close to an IPG can also cause damage to the LES tissue possibly producing burns If electrocautery is required apply it in short bursts position the ground plate on the patient to minimize current flow through the IPG and lead To reduce the possibility of adverse effects program the IPG to the OFF mode Check the IPG for proper operation immediately following the procedure If the unit is in the DOWN mode follow the Reset procedure RF Ablation RF Ab

Download Pdf Manuals

image

Related Search

Related Contents

Hamilton Beach 04493P User's Manual  2. - Psn-web.net  User Manual  Samsung RB1955VQ Bottom Freezer Refrigerator  Printer Management Utility Scanner Mode for Fiery X3e 31C-M  Catalogo para PDF (2007)  消費生活用製品の重大製品事故に係る公表について  21,1km  Wi-Ex YX027-PCS-CEL network antenna  

Copyright © All rights reserved.
Failed to retrieve file